1 The nfluence of Electromagnetic Environment on the Performance of Artificial Cardiac Pacemakers Seymour Furman, M.D., Bryan Parker, Martin Krauthamer, M.D., and Doris J. W. Escher, M.D. T he danger of producing ventricular fibrillation by passage of 60-cycle alternating current (AC) through pacemaker leads attached to a patient s heart has been outlined by Burchell [l], Weinberg [71, Noordjik [6], and Furman [3]. Each of them described patients who had died because alternating current had leaked in minute quantity to the heart via direct wired connection and had produced ventricular fibrillation. Whalen [8] has described the techniques required to insure that no leakage of current can occur into the heart. A parallel problem has been observed somewhat more recently and described by Carleton et al. [2], Lichter et al. [4], and Mansfield [5]. This is the response of implanted cardiac pacemakers to the presence of electromagnetic fields or intracorporeal alternating currents without direct contact with the pacemaker or its lead system. Each of the earlier studies [ 1, 3, 6-83 included only asynchronous pacemakers. With the development of atrial synchronous, ventricular synchronous, and demand pacemakers, an effort has been made to determine the sensitivity of 10 types of pacemakers to the effects of low-frequency alternating current in the range of l cycle per second to 100 kilocycles per second and to the effects of impulses resulting from electrical corona. n each instance the variety of response, i.e., increase in rate, development of erratic rate, change from a triggered to a fixed rate, or cessation of pacemaker function, has been determined. Case 1. As a patient with an implanted Cordis Atricor r. (atrial synchronous pacemaker) lay on a fluoroscope table that was connecte d to an alternating current-operated electrocardiograph, he promptly developed a pacemaker-induced rate of 145 beats per minute. On investigation, the frame of the fluoroscope table was found to be connected to the neutral terminal of the AC line, rather than to ground; also, a potential of 0.5 volts AC existed between the table and From the Cardiothoracic Service. Division of Surgery, and the Cardiology Service, Division of Medicine, Montefiore Hospital and Medical Center, Bronx, N.Y. Supported in part by U.S. Public Health Service Grants HE and HE Presented at the Fourth Annual Meeting of The Society of Thoracic Surgeons, New Orleans, La., Jan , THE ANNALS OF THORACC SURGERY

2 NOTE: Electromagnetic Znfluence on Pacemakers the properly grounded electrocardiograph. This resulted in a 50-microampere, 60-cycle current between the patient's right arm, which touched the table, and his right leg, which was grounded via the electrocardiograph. This small current (below the level of the patient's perception) was sufficient to interfere with pacemaker operation. Case 2. This patient had an implanted Cordis Atricor; he developed an irregular pulse and a sense of cardiac fluttering when standing within three feet of the back of an opened television set (although not in front of the screen with the set properly assembled). He found that his erratic pulse accompanied corona discharges from the high voltage system of the television set. When he stepped back from the set, or to its side, no cardiac effect occurred. METHOD All pacemakers were bench tested to determine the effects of an alternating current generated by a Hewlett-Packard Model 202C low-frequency oscillator applied directly to the electrodes. The maximum voltage was restricted to 3 volts root mean square (RMS) to avoid damage to the new and unused pacemakers. n each instance the sensitivity of response of each pacer was determined and a curve plotted for millivolts against frequency. Three in vivo tests were carried out: (1) An alternating current was passed from the right arm to the right leg via standard pasted ECG electrodes. Cardiac response was observed from the electrocardiogram, or where the ECG was obliterated by the applied current, from the record of the pulse from a finger plethysmograph. n all instances the applied voltage was kept below 3 volts to remain below the threshold of sensation. (2) A %"-long spark generated by a 6-volt automobile coil was moved toward and away from the implanted pacemaker. (3) Three operating electric shavers were brought up to the pacemaker. These were Norelco Models SC7970 and SC7960 and a Ronson CFL. The electrocardiogram was continually monitored in all instances. RESULTS ASYNCHRONOUS PACEMAKERS Cordis Ventricor 111C. These pacemakers were unaffected in vivo by any interfering signal. During bench testing, direct application of 60-cycle alternating current to the pacemaker output terminals induced a rapid emission rate. General Electric Single and Dual Rate Pacemakers. n vivo, alternating current produced no effect; the spark at 9 inches and the razors at 2 inches produced an erratic increase in rate (Fig. 1). Direct application of alternating current to the pacemaker produced only a slight increase in rate. Electrodyne TR-14. No interference effects were noted. Airborne nstrument Laboratory's Radio Frequency Pacemaker. This partially implanted system was subjected to interference near the external pulse generating unit. Only the spark coil produced rate increases to 95 impulses per minute and then only when it was within 6 inches of the pacemaker; however, this range was increased to two feet when the spark coil was placed on a steel dolly. Medtronic 5870C. No interference effects were noted. NONCOMPETTVE (TRGGERED PACEMAKERS) American Optical Co. Demand Pacer-Bench test. Frequencies below 20 cycles per second inactivated the pacemaker. At higher frequencies-including 60-cycle alternating current-the pacemaker lost its sensitivity to R wave activity and reverted to its fixed-rate mode, even when cardiac activity was more rapid. VOL. 6, NO. 1, JULY, 1968 gi

3 FURMAN, PARKER, KRAUTHAMER, AND ESCHER FG. 1. This pacemaker is designed to respond to external stimuli in order to modify its function. Electromagnetic interference increased its rate-in this instance to 150 impulses per second. With increase in rate, pacer output fell. The entire frequency response of the pacemaker was the narrowest of any of the tested units. Cordis Bentricor Z. This pacemaker, similar in mode of operation to the American Optical Co. unit, was readily stopped by all tested interference signals both in vivo and on the bench. n vivo, alternating voltages less than one volt stopped pacemaker output at a wide range of frequencies (Figs. 2, 3). The spark coil and the electric razor both caused cessation of pacer output at a range of up to 9 inches. This unit also ceased to function in close proximity to a fluorescent light tube. Cordis Ectocor (Ventricular Synchronous). Alternating currents applied directly to the pacemaker caused the unit to revert to fixed-rate operation at its automatic rate (70 impulses per minute). With a frequency and amplitude of 40 CORDS VENTRCOR FG. 2. Electromagnetic interference in this highly sensitive demand pacemaker produced inactivity and asystole during the period of interference. With removal of the interfering signal, the pacer returned to its normal sensitivity to cardiac action. 92 THE ANNALS OF THORACC SURGERY

4 NOTE: Electromagnetic Znfluence on Pacemakers ma 7r VENTRCOR N VVO ,^ to 60 cycles and 1 to 1.2 millivolts per second, the unit increased to its maximum synchronous rate (125 per minute) (Fig. 4). n vivo, both maximum synchronous rate and fixed-rate operation were induced with intracorporeal alternating currents. Erratic operation was initiated by a spark at distances less than one foot, and intermittent extra impulses were emitted with an electric razor on the skin over the pacemaker. Medtronic Demand n vivo, this unit was unaffected by any of the tests. On the bench, alternating voltages applied to the electrodes inhibited the unit s output with the frequency response curve shown in Figure 5. When this unit was placed in a saline-filled plastic vessel, 35 volts of 60-cycle alternating current were needed to affect its operation. Cordis Atricor and Atricor Jr. (Atrial Synchronous). This variety of pacemaker was the type implanted in the patients referred to earlier. n vivo tests were applied to patients with both transvenous and myocardial electrodes. Both systems readily produced tachycardia at the maximum synchronous rate (125 for Atricor, 145 for Atricor Jr.). At more intense applied alternating current, both reverted to fixed-rate, asynchronous operation because of amplifier saturation. The spark coil and shavers all caused these units to operate at maximum synchronous rate-in one patient at a distance of 4 feet. CORDS ECTOCOR FG. 4. The ventricular synchronous pacemaker responded to interfering signals by increasing its rate of impulse formation to its maximum synchronous rate, or reverting to fixed stimulation at its automatic rate (70 per minute). VOL. 6, NO., JULY,

5 FURMAN, PARKER, KRAUTHAMER, AND ESCHER FXED RATE REGON - SWTCH OFF REGON (T! FREQUENCY (cycles/sec) FG. 5. Frequency plot of alternating voltages applied to electrodes required to interfere with the operation of three triggered pacemakers. Voltages in the area above the curves disable the pacemaker, either by causing pacer inactivity or its reversion to an unsynchronized fixed-rate mode. The mode of failure is indicated by interrupted or solid lines. Note that the American Optical demand unit ceased function below 20 cycles and reverted to asynchronous operation above this frequency. All three pacemakers returned to normal function with cessation of the interfering stimu 1 i. DSCUSSON Ten different pacer units were tested for their sensitivity to alternating current applied to the human and to signals generated by a spark coil or by the small-brush variety of electric motors. Bench tests were used to find the sensitivity of the pacemaker units to 60-cycle alternating current and to determine the energy frequency range that would produce a response. A wide variety of response to extraneous stimuli was found among pacemakers of different manufacturers. The common asynchronous pacers were all insensitive to interference, except for those two designed to allow external radio frequency control of operation. n analyzing the data, the pacemakers must be divided into triggered and untriggered units. The untriggered or asynchronous (fixedrate) units do not respond to body electrical activity and should, if possible, remain fixed in their output despite any provocation. The triggered units, i.e., demand, atrial, and ventricular synchronous pacers, modify their activity as a result of body electrical activity. Their ability to respond only to appropriate (physiologically generated) electrical stimuli while at the same time rejecting extraneous stimuli is an impor- 94 THE ANNALS OF THORACC SURGERY

6 NOTE: Electromagnetic Znfluence on Pacemakers tant consideration, as is the variety of response. An atrial or ventricular synchronous pacemaker that increases its rate to 120 per minute, the maximum synchronous rate, or that returns to its automatic rate with increasing current input is far less hazardous than a demand pacemaker that perceives 60-cycle alternating current as ventricular contractions and ceases to function. All pacemakers tested reverted to normal function when interference was discontinued. Of great interest was the mode of operation of the American Optical Co. demand pacemaker. t ceased function only at low frequencies, which are rare. At higher frequencies it was almost unaffected except in a narrow frequency range; there it lost its demand function and returned to a nondemand, fixed-rate function. This unit seemed to combine greatest resistance to interference with the most satisfactory variety of response when affected by such interference. SUMMARY Ten cardiac pacemakers, triggered and untriggered, were tested in vivo and on the bench for sensitivity to radio-frequency signals and intracorporeal 60-cycle alternating current. All untriggered units were insensitive in vivo except for those designed to be controlled by radiofrequency signals. Demand and synchronous units exhibited greater sensitivity and greater reversion to inactivity, fixed-rate operation, or maximum synchronous rate in the presence of interference. Two demand units were unaffected by the interfering signals in vivo. REFERENCES 1. Burchell, H. B. Hidden hazards of cardiac pacemakers. (Editorial) Circulation 24:161, Carleton, R. A., Sessions, R. W., and Graettinger, J. S. Environmental influence on implantable cardiac pacemakers. J.A.M.A. 190:938, Furman, S., Schwedel, J. B., Robinson, G., and Hurwitt, E. S. Use of an intracardiac pacemaker in the control of heart block. Surgery 49:98, Lichter,., Borrie, J., and Miller, W. M. Radio-frequency hazards with cardiac pacemakers. Brit. Med. J. 1:1513, Mansfield, P. B. On interference signals and pacemakers. Amer. J. Electronics First Quarter. Page 61, Noordjik, J. A., Oey, F. T.., and Tebra, W. Myocardial electrodes and the danger of ventricular fibrillation. Lancet 1 :975, Weinberg, J., Artley, J. A., Whalen, R. E., and Mcntosh, H. D. Electric shock hazards in cardiac catheterization. Circ. Res. 11: 1004, Whalen, R. E., Starmer, F., and Mcntosh, H. D. Electrical hazards associated with cardiac pacing. Ann. N.Y. Acad. Sci. 11:922, VOL. 6, NO. 1, JULY,

Electronic Analysis for Pacemaker Failure

Electronic Analysis for Pacemaker Failure Electronic Analysis for Pacemaker Failure Seymour Furman, M.D., Doris J. W. Escher, M.D., Bryan Parker, and Norman Solomon, M.D. T he physician or surgeon who implants a cardiac pacemaker incurs a continuing

More information

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

The heart's natural pacemaker is called the sinoatrial (SA) node or sinus node. PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in

More information

A Mechanism for "False" Inhibition of Demand Pacemakers

A Mechanism for False Inhibition of Demand Pacemakers A Mechanism for "False" Inhibition of Demand Pacemakers By KENNETH C. LASSETER, M.D., JACK W. BUCHANAN, JR., M.S.E.E., AND KARL F. YOSHONIS, M.D. SUMMARY Certain variations in discharge rate of demand

More information

also carried out with a portable high duty electric drill (Black and Decker), a short wave diathermy apparatus (Magatherm Junior 4), a 5 watt neon

also carried out with a portable high duty electric drill (Black and Decker), a short wave diathermy apparatus (Magatherm Junior 4), a 5 watt neon British Heart journal, I970, 32, 626-632. Electrical interference in non-competitive pacemakers E. Sowton, K. Gray, and T. Preston From The Institute of Cardiology, London W.I Patients with 4I implanted

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul, Mayo Clin Proc, August 2002, Vol 77 Unintentional Deactivation of ICDs 855 Case Report Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care Settings MARY JANE RASMUSSEN,

More information

Factors Determining Vulnerability to Ventricular Fibrillation Induced by 60-CPS Alternating Current

Factors Determining Vulnerability to Ventricular Fibrillation Induced by 60-CPS Alternating Current Factors Determining Vulnerability to Ventricular Fibrillation Induced by 60-CPS Alternating Current By Tsuneoki Sugimoto, M.D., Stephen F. School, M.D., and Andrew G. Wallace, M.D. ABSTRACT Very weak,

More information

The Electrocardiogram

The Electrocardiogram The Electrocardiogram Chapters 11 and 13 AUTUMN WEDAN AND NATASHA MCDOUGAL The Normal Electrocardiogram P-wave Generated when the atria depolarizes QRS-Complex Ventricles depolarizing before a contraction

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value

More information

Radioelectrocardiography in Diagnosis of Rhythm

Radioelectrocardiography in Diagnosis of Rhythm Brit. Heart J., 1969, 31. 480. Radioelectrocardiography in Diagnosis of Rhythm Disturbances in Patients with Implanted Cardiac Pacemakers MARIUSZ JAN STOPCZYK, ZYTA KRASZEWSKA, AND RYSZARD JACEK ZOCHOWSKI

More information

Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D.

Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D. NOTES An Implantable Demand Pacemaker Lawrence I. Zaroff, M.D.,* Barouh V. Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D. E lectrical stimulation in the treatment of heart

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD External Pacemaker Set-up and Revised Date: 11/13/2014 Page 1 of 5 POLICY: The transcutaneous external pacemaker will be used as a noninvasive method of delivering electrical stimuli to the myocardium

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish

More information

Project Title Temporary Pacemaker Training Simulator

Project Title Temporary Pacemaker Training Simulator Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially

More information

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Biomedical Instrumentation

Biomedical Instrumentation University of Zagreb Faculty of Electrical Engineering and Computing Biomedical Instrumentation Safety of electrical medical devices prof.dr.sc. Ratko Magjarević Medical technology Achievements Significant

More information

RN-BC, MS, CCRN, FAHA

RN-BC, MS, CCRN, FAHA Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:

More information

Epicardial Pacemaker Implantation for Complete Heart Block

Epicardial Pacemaker Implantation for Complete Heart Block for Complete Heart Block Donald G. Mulder, M.D., and C. Gordon Frank, M.D. S udden death or progressive cardiac failure threatens life in the patient with complete heart block. The slow heart rate and

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES

STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES March, 2007 Ministry of Internal Affairs and Communications (MIC) Introduction We are moving toward the creation of a ubiquitous network society

More information

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work Electrical Activity of the Human Heart Oguz Poroy, PhD Assistant Professor Department of Biomedical Engineering The University of Iowa Outline Basic Facts about the Heart Heart Chambers and Heart s The

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

BEDSIDE ECG INTERPRETATION

BEDSIDE ECG INTERPRETATION BEDSIDE ECG INTERPRETATION Presented by: Ryan Dean, RN, MSN, CCRN, CCNS, CFRN Flight Nurse 2017 Based on presentations originally by Gennifer DePaoli, RN Objectives Hospital policies Electrical conduction

More information

Patient Safety During Cardiac Catheterization

Patient Safety During Cardiac Catheterization 119 CIRCULATION VOL 58, No 6, DECEMBER 1978 32. Neff MS, Kim KE, Persoff M, Onesti G, Swartz C: Hemodynamics of uremic anemia. Circulation 43: 876, 1971 33. Cangiano JL, Ramirez-Mux, Ramirez-Gonzalez R,

More information

BIPN100 F15 Human Physiology I (Kristan) Problem set #5 p. 1

BIPN100 F15 Human Physiology I (Kristan) Problem set #5 p. 1 BIPN100 F15 Human Physiology I (Kristan) Problem set #5 p. 1 1. Dantrolene has the same effect on smooth muscles as it has on skeletal muscle: it relaxes them by blocking the release of Ca ++ from the

More information

Biomedical Instrumentation

Biomedical Instrumentation Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Specified Tolerance (ms)

Specified Tolerance (ms) SigmaPace TM 1000: Medtronic Temporary External Pacemaker Model 5388 Checkout Procedure 1. PRELIMINARY SET UP a. Connect the pacemaker output to the analyzer s ATRIAL and VENTRICULAR 4 mm interface jacks

More information

Defibrillator. BiomedGuy

Defibrillator. BiomedGuy Defibrillator BiomedGuy Medtronic Physiocontrol LifePak 10 Introduction This life-support system is used by paramedic, hospital staff, and other trained authorized healthcare providers. Provides, ECG,

More information

Transtelephone Pacemaker Monitoring

Transtelephone Pacemaker Monitoring Transtelephone Pacemaker Monitoring Five Years Later Seymour Furman, M.D., and Doris J. W. Escher, M.D. ABSTRACT Six hundred nineteen patients have been followed by remote monitoring of pacemaker function

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

Pacemaker Concepts and Terminology*

Pacemaker Concepts and Terminology* Pacemaker Concepts and Terminology* BAROUH V BERKOVITS, E.E., Ing. Associate in Surgery, Harvard Medical School, Boston, Mass., Associate in Electrophysiology, Miami University, Miami, Fla., Senior Research

More information

ECG Interpretation Cat Williams, DVM DACVIM (Cardiology)

ECG Interpretation Cat Williams, DVM DACVIM (Cardiology) ECG Interpretation Cat Williams, DVM DACVIM (Cardiology) Providing the best quality care and service for the patient, the client, and the referring veterinarian. GOAL: Reduce Anxiety about ECGs Back to

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Arrhythmias. Pulmonary Artery

Arrhythmias. Pulmonary Artery Arrhythmias Introduction Cardiac arrhythmia is an irregularity of the heart beat that causes the heart to beat too slowly, too fast, or irregularly. There are different types of arrhythmias. Most arrhythmias

More information

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular

More information

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee -RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University The Normal Conduction System Sinus Node Normal Sinus Rhythm (NSR) Sinus Bradycardia Sinus Tachycardia

More information

Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs

Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs Faruk Cingoz, MD, George Yanulis, MS, Elizabeth Ching, RN, Kiyotaka Fukamachi,

More information

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net

More information

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

Interpreting Electrocardiograms (ECG) Physiology Name: Per: Interpreting Electrocardiograms (ECG) Physiology Name: Per: Introduction The heart has its own system in place to create nerve impulses and does not actually require the brain to make it beat. This electrical

More information

Basic Dysrhythmia Interpretation

Basic Dysrhythmia Interpretation Basic Dysrhythmia Interpretation Objectives 2 To understand the Basic ECG To understand the meaning of Dysrhythmia To describe the normal heart conduction system. To describe the normal impulse pathways.

More information

5- The normal electrocardiogram (ECG)

5- The normal electrocardiogram (ECG) 5- The (ECG) Introduction Electrocardiography is a process of recording electrical activities of heart muscle at skin surface. The electrical current spreads into the tissues surrounding the heart, a small

More information

Changing Indications for Pacemakers in Children

Changing Indications for Pacemakers in Children Changing Indications for Pacemakers in Children William H. Fleming, M.D., Lynne B. Sarafian, M.N., John D. Kugler, M.D., Philip J. Hofschire, M.D., and Edward B. Clark, M.D. ABSTRACT Permanent pacemakers

More information

This presentation will deal with the basics of ECG description as well as the physiological basics of

This presentation will deal with the basics of ECG description as well as the physiological basics of Snímka 1 Electrocardiography basics This presentation will deal with the basics of ECG description as well as the physiological basics of Snímka 2 Lecture overview 1. Cardiac conduction system functional

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

Pacemaker and AV Node Ablation Patient Information

Pacemaker and AV Node Ablation Patient Information Melbourne Heart Rhythm Pacemaker and AV Node Ablation Patient Information The Heart The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two upper chambers

More information

Cardiac Cycle. Each heartbeat is called a cardiac cycle. First the two atria contract at the same time.

Cardiac Cycle. Each heartbeat is called a cardiac cycle. First the two atria contract at the same time. The Heartbeat Cardiac Cycle Each heartbeat is called a cardiac cycle. First the two atria contract at the same time. Next the two ventricles contract at the same time. Then all the chambers relax. http://www.youtube.com/watch?v=frd3k6lkhws

More information

Patient Resources: Arrhythmias and Congenital Heart Disease

Patient Resources: Arrhythmias and Congenital Heart Disease Patient Resources: Arrhythmias and Congenital Heart Disease Overview Arrhythmias (abnormal heart rhythms) can develop in patients with congenital heart disease (CHD) due to thickening/weakening of their

More information

ICD Implantation Patient Information

ICD Implantation Patient Information Melbourne Heart Rhythm ICD Implantation Patient Information The Heart The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two upper chambers (the right atrium

More information

UNDERSTANDING YOUR ECG: A REVIEW

UNDERSTANDING YOUR ECG: A REVIEW UNDERSTANDING YOUR ECG: A REVIEW Health professionals use the electrocardiograph (ECG) rhythm strip to systematically analyse the cardiac rhythm. Before the systematic process of ECG analysis is described

More information

CORONARY ARTERIES HEART

CORONARY ARTERIES HEART CARDIAC/ECG MODULE THE HEART CORONARY ARTERIES FIBRILLATING HEART CORONARY ARTERIES HEART PRACTICE RHYTHMS PRACTICE RHYTHMS ELECTRICAL CONDUCTION SA Node (60 100) Primary pacemaker AV Node (40 60) ***Creates

More information

ELECTROCARDIOGRAPHY (ECG)

ELECTROCARDIOGRAPHY (ECG) ELECTROCARDIOGRAPHY (ECG) The heart is a muscular organ, which pumps blood through the blood vessels of the circulatory system. Blood provides the body with oxygen and nutrients, as well as assists in

More information

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and

More information

Perioperative Management of Cardiac Implantable Devices

Perioperative Management of Cardiac Implantable Devices Financial Disclosures Breandan Sullivan MD Assistant Professor Co Director Cardiothoracic and Vascular Surgery ICU Department of Anesthesiology and Critical Care Medicine University of Colorado None Perioperative

More information

Interpretation of complex demand pacemaker arrhythmias

Interpretation of complex demand pacemaker arrhythmias British Heart Journal, I972, 34, 312-3I7. Interpretation of complex demand pacemaker arrhythmias S. Serge Barold, John J. Gaidula, Richard L. Banner, George I. Litman, and Sidney Goldstein From the Department

More information

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers INTERNATIONAL STANDARD ISO 14708-2 Second edition 2012-08-15 Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers Implants chirurgicaux Dispositifs médicaux implantables actifs

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

THE HEART THE CIRCULATORY SYSTEM

THE HEART THE CIRCULATORY SYSTEM THE HEART THE CIRCULATORY SYSTEM There are three primary closed cycles: 1) Cardiac circulation pathway of blood within the heart 2) Pulmonary circulation blood from the heart to lungs and back 3) Systemic

More information

The Results of Demand Pacing

The Results of Demand Pacing The Results of Demand Pacing in Cardiac Arrhythmias Peter Allen, M.D., and C. Eve Rotem, M.D. I n the majority of patients with established 3" heart block the fixedrate cardiac pacemaker has successfully

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

Dry body skin has a ( higher / lower ) resistance to the flow of the electric current than the internal organs of the body.

Dry body skin has a ( higher / lower ) resistance to the flow of the electric current than the internal organs of the body. RESISTANCE OF INTERNAL ORGANS (SUCH AS HEART, LUNGS, LIVER, ETC.) Dry body skin has a ( higher / lower ) resistance to the flow of the electric current than the internal organs of the body. 19. Electricity

More information

BASIC CONCEPT OF ECG

BASIC CONCEPT OF ECG BASIC CONCEPT OF ECG Electrocardiogram The electrocardiogram (ECG) is a recording of cardiac electrical activity. The electrical activity is readily detected by electrodes attached to the skin. After the

More information

The protection you need - without touching your heart

The protection you need - without touching your heart The S ICD System The protection you need - without touching your heart listen to your heart Patient information You have options An implantable defibrillator, commonly known as an ICD, is a device designed

More information

CRC 431 ECG Basics. Bill Pruitt, MBA, RRT, CPFT, AE-C

CRC 431 ECG Basics. Bill Pruitt, MBA, RRT, CPFT, AE-C CRC 431 ECG Basics Bill Pruitt, MBA, RRT, CPFT, AE-C Resources White s 5 th ed. Ch 6 Electrocardiography Einthoven s Triangle Chest leads and limb leads Egan s 10 th ed. Ch 17 Interpreting the Electrocardiogram

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

Reversion of ventricular tachycardia by pacemaker stimulation

Reversion of ventricular tachycardia by pacemaker stimulation British Heart Journal, 1971, 33, 922-927. Reversion of ventricular tachycardia by pacemaker stimulation M. A. Bennett and B. L. Pentecost From the General Hospital, Birmingham 4 Reversion of ventricular

More information

Pacemaker and ICD Interrogation

Pacemaker and ICD Interrogation Pacemaker and ICD Interrogation To receive the maximum benefit from your pacemaker, you will need to have regular follow-up appointments to ensure that it is working properly. This follow up can be arranged

More information

Diploma in Electrocardiography

Diploma in Electrocardiography The Society for Cardiological Science and Technology Diploma in Electrocardiography The Society makes this award to candidates who can demonstrate the ability to accurately record a resting 12-lead electrocardiogram

More information

Clinical Data Summary: Avoid FFS Study

Clinical Data Summary: Avoid FFS Study Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead Clinical Data Summary:

More information

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. The Miracle of Living February 21, 2018 Matthew Ostrom MD,FACC,FHRS Division of

More information

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES PURPOSE Guidelines for caring for and monitoring patients with a temporary external epicardial pacemaker. Epicardial pacemaker monitoring is performed on a continuous basis by a nurse to evaluate the effectiveness

More information

WHAT S THAT RHYTHM I AM HEARING? GUIDE TO AUSCULTATION OF ARRHYTHMIAS IN HORSES

WHAT S THAT RHYTHM I AM HEARING? GUIDE TO AUSCULTATION OF ARRHYTHMIAS IN HORSES WHAT S THAT RHYTHM I AM HEARING? GUIDE TO AUSCULTATION OF ARRHYTHMIAS IN HORSES Michelle Henry Barton DVM, PhD, DACVIM University of Georgia, Athens, GA INTRODUCTION The purpose of this talk is to review

More information

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants EKG Pop Quiz Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants 1 Disclosures No financial relationships to disclose. EKG #1 75 y/o woman with a dual-chamber pacemaker

More information

VALLIAMMAI ENGINEERING COLLEGE

VALLIAMMAI ENGINEERING COLLEGE VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur 603 203 DEPARTMENT OF ELECTRONICS AND COMMUNICATION ENGINEERING QUESTION BANK VI SEMESTER EC6001 MEDICAL ELECTRONICS Regulation 2013 Academic Year

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) The story so far

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) The story so far Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) The story so far The heart pumps blood around the body. It has four chambers which contact

More information

EC6001 MEDICAL ELECTRONICS

EC6001 MEDICAL ELECTRONICS EC6001 MEDICAL ELECTRONICS UNIT I ELECTRO-PHYSIOLOGY AND BIO-POTENTIAL RECORDING The origin of Bio-potentials; bio-potential electrodes, biological amplifiers, ECG, EEG, EMG, PCG, lead systems and recording

More information

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report Cardiol Ther (2017) 6:139 143 DOI 10.1007/s40119-017-0086-6 CASE REPORT Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report Abhishek Shenoy. Abhishek

More information

This leaflet is intended for patients with permanent pacemakers and their parents.

This leaflet is intended for patients with permanent pacemakers and their parents. Permanent PACEMAKER This leaflet is intended for patients with permanent pacemakers and their parents. It serves as an information tool and a guide that offers some advice in order to live safely. General

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

Cardiac Resynchronisation Therapy Patient Information

Cardiac Resynchronisation Therapy Patient Information Melbourne Heart Rhythm Cardiac Resynchronisation Therapy Patient Information Normal Heart Function The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two

More information

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2 Temporary Transvenous Pacing Guideline Classification: Clinical Guideline Lead Author: Dr Peter Woolfson Additional author(s): Dr Alan Fitchet Sister Joanne Hughes, Matron Julie Winstanley Authors Division:

More information

About atrial fibrillation (AFib) Atrial Fibrillation (AFib) What is AFib? What s the danger? Who gets AFib?

About atrial fibrillation (AFib) Atrial Fibrillation (AFib) What is AFib? What s the danger? Who gets AFib? Understanding AFib Atrial Fibrillation (AFib) About AFib 3 How Your Heart Works 4 Types of AFib 5 Symptoms 5 Risk Factors 5 How is AFib Diagnosed? 6 Treatment 6 What to Ask Your Doctor 7 A normal heartbeat

More information

J. Physiol. (I957) I37, I4I-I53

J. Physiol. (I957) I37, I4I-I53 141 J. Physiol. (I957) I37, I4I-I53 EFFECTS OF NORADRENALINE AND ADRENALINE ON THE ATRIAL RHYTHM IN THE HEART-LUNG PREPARATION BY J. H. BURN, A. J. GUNNING AND J. M. WALKER From the Department of Pharmacology,

More information

SPLITTING OF HEART SOUNDS FROM VENTRICULAR

SPLITTING OF HEART SOUNDS FROM VENTRICULAR Brit. Heart J., 1965, 27, 691. SPLITTING OF HEART SOUNDS FROM VENTRICULAR ASYNCHRONY IN BUNDLE-BRANCH BLOCK, VENTRICULAR ECTOPIC BEATS, AND ARTIFICIAL PACING* BY EDGAR HABER AND AUBREY LEATHAM From the

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Notification and Prior Authorization Program: Electrophysiology Implant Classification Table The following chart contains the codes that require notification or prior authorization as part of UnitedHealthcare

More information

Panorama. Arrhythmia Analysis Frequently Asked Questions

Panorama. Arrhythmia Analysis Frequently Asked Questions Panorama Arrhythmia Analysis Frequently Asked Questions What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Chapter 9: Ventricular Dysrhythmias 2012 The Companies, Inc. All rights reserved. Learning Outcomes 9.1 Describe the various ventricular dysrhythmias 9.2

More information