1 Devices and technology 223 Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol Paolo Verdecchia, Fabio Angeli, Roberto Gattobigio, Enrica Angeli, Silvia Pede and Loretta Pittavini Objectives We tested the accuracy of the UA-705 blood pressure semi-automatic monitor. Methods Device evaluation was performed according to the modified British Hypertension Society protocol released in Eighty-five patients with characteristics outlined in the British Hypertension Society protocol were recruited among those attending our out-patient clinic. Sequential readings were taken for the main validation test. Outcome was classified according to the British Hypertension Society criteria, which are based on four zones of accuracy differing from the mercury standard by 5, 10 and 15 mmhg, or more. Results The mean blood pressure difference ( ± 1 SD) between device and observers was 0.4 mmhg (SD 7.7) for systolic blood pressure and 0.3 mmhg (SD 8.2) for diastolic blood pressure. Overall, 96% (observer 1) and 95% (observer 2) of readings between device and observers differed by 15 mmhg or less for both systolic and diastolic blood pressure. The device achieved a grade A for both systolic and diastolic blood pressure. Conclusions This study shows that the A&D UA-705 device satisfies the British Hypertension Society standard for accuracy by achieving a grade A for both systolic and diastolic blood pressure. Blood Press Monit 11: c 2006 Lippincott Williams & Wilkins. Blood Pressure Monitoring 2006, 11: Keywords: device, home blood pressure measurement, hypertension, validation Department of Cardiology, Hospital R. Silvestrini, Perugia, Italy Correspondence and requests for reprints to Paolo Verdecchia, MD, FACC, Dipartimento Malattie Cardiovascolari, Ospedale R. Silvestrini, Località S. Andrea delle Fratte, Perugia 06156, Italy Tel: , , ; fax: ; verdec@tin.it Sponsorship: Supported in part by grants from Associazione Umbra Cuore e Ipertensione, Perugia, Italy. Received 9 December 2005 Accepted 10 January 2006 Introduction The interest in semi-automatic devices for home blood pressure (BP) measurement is supported by the growing evidence that home BP measurements significantly refine cardiovascular risk stratification [1,2]. Among the semi-automatic devices, the models that use oscillometric techniques are increasingly popular. Usually, these devices require little training and may be particularly suitable for patients with infirmities such as arthritis and deafness. Only a few semi-automatic devices, however, successfully passed rigorous validation studies according to internationally established criteria [3]. The list of BP devices that have been tested according to the British Hypertension Society (BHS) protocol and met the BHS criteria is available on the web ( The BHS provided a standard protocol for validating BP monitors in order to establish their accuracy when compared with standard mercury sphygmomanometers c 2006 Lippincott Williams & Wilkins [4]. This protocol, first released in 1990, was subsequently updated in 1993 [4]. We adopted this revised version of the protocol to execute a validation study of the UA-705 compact desktop semi-automatic monitor produced by the A&D Company (Toshimi-ku, Tokyo, Japan) for self-measurement of BP. Such devices measure BP at arm level, using an oscillometric algorithm. Measurement range is mmhg for BP and beat/min for heart rate. The device works with one battery (1.5 V, type AA). The dimensions of the device are approximately 81 mm 54 mm 105 mm and its weight is approximately 120 g excluding the batteries. Operating conditions range from +10 to +401C and storage conditions from 10 to +601C. The UA-705 has a large display (BP measurement is displayed as numerals) and a single start button for switching the device on and off. The arm cuff is connected by an air connector plug to a rubber bulb used to pressurize. Deflation is automatic and different cuff sizes are available. The device can store up to 30

2 224 Blood Pressure Monitoring 2006, Vol 11 No 4 measurements for monitoring the trend of BP measurements and the average value. An average of all the stored data is shown at each start up, giving a condensed reading history at a glance. Data are retained as long as the battery is in the device and measurements can be recalled easily. After each measurement, the device displays automatically, on a bar indicator, a segment based on the current data corresponding to the World Health Organization classification (optimal, normal, high normal and mild, moderate and severe hypertension) [5]. The device also has an irregular heart beat indicator that can help users find an irregular heartbeat. Methods We used the revised version of the BHS protocol to validate the device. Briefly, this protocol is based on sequential measurement comparisons. The protocol consisted of four phases and successful completion of each phase was required to continue to the next [4]. In phase 1, pressure calibration of the devices was performed. We received three devices from the manufacturer, who gave a written declaration that the devices were standard production models. Each device underwent dynamic calibration using three observers who were blinded from each other. The device was connected in parallel to two sphygmomanometers, and all three manometers were then connected to a cuff, which was placed around a cylinder. By using the calibration table in the protocol, the widest range of BP applicable to the device was selected. The first observer inflated the cuff and called out at five randomly pre-selected pressure points during deflation. The second observer noted the pressure on the mercury column and the third observer noted the pressure on the device. This procedure was repeated six times, for a total of 30 readings. Readings obtained from observers 2 and 3 were then compared; the difference had to be less than 3 mmhg in at least 28 of the readings. In phase 2 (in-field phase), the three devices were put into clinical use. Each device underwent at least 400 inflations in the out-patient clinic. In phase 3, calibration procedures were repeated again after in-field use. At least two of the three devices had to pass calibration successfully in order to continue to phase 4. In phase 4, one single device was randomly selected for validation. Phase 4 was carried out in 85 study participants recruited according to the criteria outlined by the BHS protocol. The range of BP and number of participants required in each category are shown in Table 1. Exclusion criteria were cardiac arrhythmias or Korotkoff Table 1 Minimum number of study participants required within each blood pressure group and actual number of participants recruited in each group Systolic blood pressure Diastolic blood pressure Minimum number of participants required Number of participants examined <90 < > 180 > sounds that did not disappear. Two observers were trained for BP measurement with a mercury sphygmomanometer using a CD ROM produced by the BHS. Interobserver accuracy was monitored by comparing differences in simultaneous mercury readings after every 20 patients. The participants were seated for at least 5 min in a warm, quiet room before the measurements were taken. BP measurements were taken, alternating between simultaneous mercury sphygmomanometer readings (by both observers) and the device. The effect of venous congestion and variability of BP were minimized by allowing more than 30 s and less than 1 min between measurements. The first mercury reading was used only to assign the patient to one of the predefined pressure ranges (Table 1) and was not used in analysis. The remaining seven readings (four by mercury sphygmomanometer and three by device) were subsequently used in the sequential analysis according to the BHS protocol [4]. Differences between the device and the observers were then established by calculating three sets of differences for each observer and for systolic and diastolic BP, respectively: (1) the difference between the device reading and the previous mercury reading; and (2) the difference between the device reading and the subsequent mercury reading. For sequential analysis, the lowest set of differences (one or two as above) for each participant was used to calculate the mean and standard deviation for systolic and diastolic BP separately. BHS grading was determined by the percentage of differences r 5, r 10 and r 15 mmhg. The BHS protocol indicates that the device needs to grant a grade A or B for both systolic and diastolic BP measurement in order to be recommended (Table 2). Results Eighty-five participants were recruited in our out-patient clinic according to the criteria set by the BHS protocol.

3 Validation of the A&D UA-705 device Verdecchia et al. 225 At the beginning of validation, interobserver accuracy was satisfactory. Indeed, 98% of readings ranged within 5 mmhg, and 100% within 10 mmhg, for both systolic and diastolic BP. Similar results were noted after every 20 patients analyzed during phase 4. In the first phase, the difference between readings obtained from observers 2 and 3 was less than 3 mmhg in 30 of the readings. The second phase was also successfully completed with a good performance of the device. After this phase, the calibration procedure was repeated (phase 3) and all the three devices passed calibration successfully. To assess static device validation, 85 participants were selected to fulfill criteria specified by the BHS protocol (Table 1). Systolic BP ranged from 85 to 198 mmhg and diastolic BP from 50 to 118 mmhg. The main clinical and demographic characteristics of the participants recruited are listed in Table 3. The mean difference and standard deviation were 0.4 mmhg (7.7) for systolic and 0.3 mmhg (8.2) for diastolic BP in the sequential analysis (Table 4). Overall, the device achieved an A grade for both systolic and diastolic BP. An additional tool of analysis, the difference between device and observer readings was plotted against the mean pressure of both for systolic and diastolic BP according to the Bland Altman [6] method (Figs 1 and 2). Horizontal lines are drawn at the mean difference, and at the mean difference ± 1.96 times the standard deviation of the differences. As depicted in the figures, there was no systematic trend in the differences between device and observers with progressively higher BP values. All three devices passed a post-validation calibration test, as described previously, by achieving differences within 2 mmhg in 29 out of 30 readings. Discussion The UA-705 device tested in the present study met the requirements set by the revised version of the BHS protocol [3], which look quite rigorous. Our results show that the device met the requirements for use in adults over a wide range of BP values. Table 2 British Hypertension Society grading criteria Absolute difference between standard and test device r 5 r 10 r 15 The device achieved a grade A for both systolic and diastolic BP and, as shown in Figs 1 and 2, was accurate over a wide range of BP values. Overall, the device was simple to use and readings were easy to read owing to the large screen. Cumulative percentage of readings (%) Grade A Grade B Grade C Grade D Worse than C Worse than C Worse than C To obtain a grade, the device must achieve percentages greater than or equal to those reported in the table in the three categories ( r 5, r 10 and r 15 mmhg). Table 3 Main characteristics of participants recruited Age (years) 50 ± 13 Male sex (%) 55 Systolic blood pressure 139 ± 29 Diastolic blood pressure 85 ± 19 Heart rate 73 ± 12 Recently, two other A&D devices (UA-774 and UA-704) have been validated [7,8]. These devices have the same BP determination algorithm as the UA-705 (A&D, written communication). Both devices yielded a BHS A/A grade [7,8]. Current indications for self-measurement of BP at home include screening for white-coat hypertension, resistant hypertension, evaluation of long-term follow-up and improving adherence to treatment [9]. Other potential indications include hypertension in the elderly, hypertension in diabetic individuals, and hypertension in Table 4 British Hypertension Society grading Difference between standard and test device Grade r 5 r 10 r 15 Mean ( ± SD) Mean ( ± SD) of differences Systolic blood pressure Observer 1 A (23.3) 0.5 (7.9) Observer 2 A (23.1) 0.4 (7.7) Diastolic blood pressure Observer 1 A (14.7) 0.3 (8.2) Observer 2 A (14.6) 0.4 (7.7) Observer comparison Systolic blood pressure A (0.7) Diastolic blood pressure A (0.9) Numbers under the difference between standard and test device denote the percentages of test device readings within r 5, r 10 and r 15 mmhg. The mean blood pressure values by observer and the mean of differences between the test device and the observers are also shown.

4 226 Blood Pressure Monitoring 2006, Vol 11 No 4 Fig secretarial assistance and Mr Mariano Cecchetti for nursing assistance. Difference in blood pressure device observer Mean of device and observer blood pressure Plot of the difference in systolic blood pressure between the device and the better observer for sequential comparison. Fig. 2 Difference in blood pressure device observer Mean of device and observer blood pressure Plot of the difference in diastolic blood pressure between the device and the better observer for sequential comparison. pregnancy [9]. It is generally accepted that target levels of 135 mmhg for systolic BP and 85 mmhg for diastolic BP should be achieved. These values are the average of several measurements taken in the morning and in the evening over at least three days of a week [10]. A limitation of the present study is that the UA-705 was not tested in newborns and infants or in pregnant women. Acknowledgements The devices used in this study were provided by A&D Co. Ltd, Tokyo, Japan and were chosen at random from the production line. We thank Miss Francesca Saveri for References 1 Ohkubo T, Imai Y, Tsuji I, Nagai K, Kato J, Kikuchi N. Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure: a population-based observation in Ohasama, Japan. J Hypertens 1998; 16: Bobrie G, Chatellier G, Genes N, Clerson P, Vaur L, Vaisse B, et al. Cardiovascular prognosis of masked hypertension detected by blood pressure self-measurement in elderly treated hypertensive patients. JAMA 2004; 291: O Brien E, Weber B, Parati G, Staessen J, Myers MG, on behalf of the European Society of Hypertension. Blood pressure measuring devices: recommendations of the European Society of Hypertension. BMJ 2001; 322: O Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, Altman DG, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11(Suppl 2):S43 S62. 5 Guidelines Committee European Society of Hypertension European Society of Cardiology guidelines for the management of arterial hypertension. J Hypertens. 2003; 21: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; i: Kobalava ZD, Kotovskaya YV, Rodionov E. Validation of semi-automatic device UA-704 for self-measurement of blood pressure. Blood Press Monit 2005; 10: Verdecchia P, Angeli F, Poeta F, Reboldi GP, Borgioni C, Pittavini L, Porcellati C. Validation of the A&D UA-774 (UA-767Plus) device for selfmeasurement of blood pressure. Blood Press Monit 2004; 9: O Brien E, Asmar R, Beilin L, Imai Y, Mallion JM, Mancia G, et al., representing the European Society of Hypertension Working Group on Blood Pressure Monitoring. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. J Hypertens 2003; 21: Participants of the First International Consensus Conference. Guidelines for the use of self-blood pressure monitoring: a summary report of the First International Consensus Conference. J Hypertens 2000; 18: Appendix In this appendix the basic information on the device is reported in accordance with the suggestions of the BHS protocol [3]. Device identification UA-705, A&D Company Ltd, Tokyo, Japan Compliance with Standard The device conforms to the European Directive 93/42 EEC for Medical Products (CE mark 0366) Validation studies No other validation studies Instructions for use, care, and maintenance These are reported in detail in the instruction manual Power supply One alkaline battery 1.5 V (type LR6, type AA) Number of measurements 400 Service facilities Japan: A&D Company Ltd, Higashi-Ikebukuro, Toshima-ku, Tokyo , Japan. Tel: ; fax: USA: A&D Engineering Inc., 1555 McCandless Drive, Milpitas, CA USA. Tel: ; fax: Europe: A&D Instruments Ltd, Abingdon Science Park, Abingdon, Oxford OX14 3YS, UK. Tel: + 44 (0) ; fax: + 44 (0) Italy: INTERMED Srl, P.zza C Donegani, Milan, Italy. Tel: ; fax: Australia: A&D Mercury Pty Ltd, 32 Dew Street, Thebarton, South Australia 5031, Australia. Tel: ; fax: Dimensions 81 mm 54 mm 105 mm excluding cuff

5 Validation of the A&D UA-705 device Verdecchia et al. 227 Weight 120 g excluding batteries List of components UA-705 device including its cuff, carrying case, one alkaline battery, and instruction manual Method of BP measurement Oscillometric Factors affecting accuracy Sources of inaccurate measurements may be arrhythmia or noise due to arm or wrist movements Operator training requirements The instrumentation does not require specific expertise because it is very easy to operate

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Original article http://dx.doi.org/10.3345/kjp.2011.54.11.463 Korean J Pediatr 2011;54(11):463-469 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

More information

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009 42 Original article Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP, the Omron HBP T and the Pic Indolor Professional Roland

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 16 Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-71-E)

More information

Home blood pressure measurement with oscillometric upper-arm devices

Home blood pressure measurement with oscillometric upper-arm devices REVIEW Home blood pressure measurement with oscillometric upper-arm devices R.L. Braam, Th. Thien *# Department of Internal Medicine, University Medical Centre St Radboud, PO Box 9101, 6500 HB Nijmegen,

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

The increasing awareness of hypertension as a serious

The increasing awareness of hypertension as a serious AJH 2002; 15:787 792 Evaluation of the Performance of a Wrist Blood Pressure Measuring Device With a Position Sensor Compared to Ambulatory 24-Hour Blood Pressure Measurements Sakir Uen, Burkard Weisser,

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval BPU22 with the AViTA BPM63S Devices Item 9 HARTMANN Veroval BPU22 AViTA BPM63S Pictures Display Image Validation Equivalence ESH 2010 ESH 2002 BHS AAMI Category Arm Type

More information

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? O R I G I N A L A R T I C L E Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? I. Aksoy, J. Deinum, J.W.M. Lenders, Th. Thien *# Department of

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval upper arm blood pressure monitor with the AViTA BPM63S Devices Pictures HARTMANN Veroval upper arm blood pressure monitor AViTA BPM63S

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval BPW22 with the AViTA BPM15S Devices Item 9 HARTMANN Veroval BPW22 AViTA BPM15S Pictures Display Image Validation Equivalence ESH 2010 ESH 2002 BHS AAMI Category Wrist

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval wrist blood pressure monitor with the AViTA BPM15S Devices HARTMANN Veroval wrist blood pressure monitor AViTA BPM15S Pictures Display

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the SAFETY PRONTEX INTEGRA (Nissei DS-B33-03) with the Nissei DSK-1011 Devices Item 9 SAFETY PRONTEX INTEGRA (Nissei DS-B33-03) Nissei DSK-1011 Pictures Display Image Validation ESH IP2010

More information

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure? MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents 01... Blood Pressure Monitor Intended Use 02... 1. What is blood pressure? 02... 2. Why is it useful measure blood pressure at home?... A. WHO blood pressure

More information

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA)

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) THE NEW ARMENIAN MEDICAL JOURNAL Vol.5 (2011), Nо 2, p.29-34 DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) Zelveian P.H. 1, 2, Podosyan G.A. 2 1 Institute

More information

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home.

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home. Welch Allyn ABPM 6100S Comprehensive blood pressure monitoring that brings patient comfort home. Features and Benefits of the Welch Allyn ABPM 6100S Pictorial application instructions, artery markers and

More information

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION*

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Progress in Clinical Medicine 1 ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Keishi ABE** Asian Med. J. 44(2): 83 90, 2001 Abstract: J-MUBA was a large-scale clinical

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the Omron M6 Comfort IT (HEM-7322U-E) with the Omron M6 Comfort (HEM-7321-E) Devices Omron M6 Comfort IT (HEM-7322U-E) Omron M6 Comfort (HEM-7321-E) Pictures Display Validation ESH 2010 Device

More information

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008 Analytical methods and statistical analyses 285 PA.NET International Quality Certification Protocol for blood pressure monitors Stefano Omboni a,b, Carlo Costantini a,b,c, Claudio Pini b,d, Roberto Bulegato

More information

Validation Study Result Form

Validation Study Result Form Validation Study Result Form Please complete Section 1 to Section 3 of this form and return it to dabl Educational Trust with copies of the validation plots. The requirements for each of these sections

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

Digital Blood Pressure Monitor. Model UA-767 BT

Digital Blood Pressure Monitor. Model UA-767 BT Digital Blood Pressure Monitor Model UA-767 BT Instruction Manual WM+PD4001054B 13 Contents Dear Customers... 2 Preliminary Remarks... 2 Precautions... 2 Parts Identification... 3 Symbols... 4 Using The

More information

White-coat hypertension (WCH), 1 referred to as office 2

White-coat hypertension (WCH), 1 referred to as office 2 Short- and Long-Term Incidence of Stroke in White-Coat Hypertension Paolo Verdecchia, Gian Paolo Reboldi, Fabio Angeli, Giuseppe Schillaci, Joseph E. Schwartz, Thomas G. Pickering, Yutaka Imai, Takayoshi

More information

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials. AJH 1995; 8:311-315 BRIEF COMMUNICATIONS Lack of Placebo Effect on Ambulatory Blood Pressure Giuseppe Mancia, Stefano Omboni, Gianfranco Parati, Antonella Ravogli, Alessandra Villani, and Alberto Zanchetti

More information

Declaration of Equivalence Form

Declaration of Equivalence Form Declaration of Equivalence Form DECLARATION OF BLOOD PRESSURE MEASURING DEVICE EQUIVALENCE 2013 SECTION A - Please complete all items. I Bill Huang,, Name of a Company Director A SIGNED COPY WILL BE POSTED

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

greater than 90% of patients having blood pressure measured consistently according to the protocol.

greater than 90% of patients having blood pressure measured consistently according to the protocol. Clinical methods and pathophysiology 1 Reliability and validity of blood pressure measurement in the Secondary Prevention of Small Subcortical Strokes study Pablo E. Pérgola a, Carole L. White b, John

More information

MX3 Plus English. Introduction. Table of contents. 1 How to obtain meaningful blood pressure readings

MX3 Plus English. Introduction. Table of contents. 1 How to obtain meaningful blood pressure readings X3 Plus English X3 Plus Introduction The ORON X3 Plus measures your blood pressure and pulse simply and quickly from the arm, without the use of an inflation bulb or stethoscope. The measurement values

More information

Model UB-512. Wrist Digital Blood Pressure Monitor 使用手冊 翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Model UB-512. Wrist Digital Blood Pressure Monitor 使用手冊 翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni Wrist Digital Blood Pressure Monitor Model UB-512 Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción 1WMPD4000980C Manuale di Istruzioni Traduzione 使用手冊 翻譯

More information

Model UB-511. Wrist Digital Blood Pressure Monitor 使用手冊 翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Model UB-511. Wrist Digital Blood Pressure Monitor 使用手冊 翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni Wrist Digital Blood Pressure Monitor Model UB-511 Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción 1WMPD4000888C Manuale di Istruzioni Traduzione 使用手冊 翻譯

More information

There is convincing evidence in clinical studies

There is convincing evidence in clinical studies AJH 1998;11:1413 1417 Reliability of Reporting Self-Measured Blood Pressure Values by Hypertensive Patients Thomas Mengden, Rosa Maria Hernandez Medina, Belen Beltran, Elena Alvarez, Karin Kraft, and Hans

More information

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

!#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 !"#$ Blood Pressure Monitor Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 Introduction

More information

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room. Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room Abstract Ahmad Mirdamadi (1), Mostafa Etebari (2) Original Article BACKGROUND:

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

Use of ambulatory and home blood pressure (BP) measurements

Use of ambulatory and home blood pressure (BP) measurements Long-Term Risk of Mortality Associated With Selective and Combined Elevation in Office, Home, and Ambulatory Blood Pressure Giuseppe Mancia, Rita Facchetti, Michele Bombelli, Guido Grassi, Roberto Sega

More information

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice (2005) 19, 801 807 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Prognostic significance of blood pressure measured in the office, at home and

More information

Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus Devices Microlife A2 Basic (BP3GQ1-3P) 9 BP3BT0-A 9 Microlife BP A100 Plus 9 Image 10 10 10 Validation

More information

M1 Classic. Digital Blood Pressure Monitor. Instruction manual page 1 SYS. mmhg DIA. mmhg. PULSE/min

M1 Classic. Digital Blood Pressure Monitor. Instruction manual page 1 SYS. mmhg DIA. mmhg. PULSE/min 1 Classic Digital Blood Pressure onitor Instruction manual page 1 1 Classic English Introduction The ORON 1 Classic measures your blood pressure and pulse simply and quickly from the arm, without the use

More information

Model UA Digital Blood Pressure Monitor 使用手冊翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Model UA Digital Blood Pressure Monitor 使用手冊翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni Digital Blood Pressure Monitor Model UA-1020 1WMPD4002224C Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione 使用手冊翻譯 Contents

More information

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood SPECIAL ARTICLE Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Eoin O Brien a,thomas

More information

Model UA-767. Digital Blood Pressure Monitor 使用手冊翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Model UA-767. Digital Blood Pressure Monitor 使用手冊翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni Digital Blood Pressure Monitor Model UA-767 1WMPD4000561D Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione 使用手冊翻譯 Contents

More information

Validation Study Result Form

Validation Study Result Form Validation Study Result Form Please complete Section 1 to Section 5 of this form and return it to dabl Educational Trust with copies of the validation Study Ref. plots. In Section 3 to Section 5, complete

More information

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice.

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice. AJH 26; 19:81 89 BP Measurement Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice Tony C. Y. Pang and Mark A. Brown Background: To determine the extent of discrepancies between

More information

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon (2009) 23, 794 800 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Unreliable oscillometric blood pressure measurement: prevalence, repeatability

More information

BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN?

BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN? BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN? DECLARATION OF INTEREST Medical Director and Chairman, Advisory Board dabl Ltd., Dublin, Ireland. BLOOD PRESSURE MEASUREMENT

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations

Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations The American Journal of Medicine (2006) 119, 69.e13-69.e18 CLINICAL RESEARCH STUDY Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations

More information

Digital Blood Pressure Monitor. Model UA-852

Digital Blood Pressure Monitor. Model UA-852 Digital Blood Pressure Monitor Model UA-852 Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione Bedienungsanleitung Übersetzung

More information

Data Collection Worksheet

Data Collection Worksheet Data Collection Worksheet 1. Has a doctor or nurse ever said that you have: High blood pressure or hypertension? [ ] No [ ] Yes [ ] Not Sure 1.a. If Yes, then at what age were you first told this? Age

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus Devices Microlife A6 PC (BP3GU1-8Y) 9 BP3BT0-A 9 Microlife BP A100 Plus 9 Image 10 10 10 Validation

More information

Managing Blood Pressure: It Takes a Team

Managing Blood Pressure: It Takes a Team Managing Blood Pressure: It Takes a Team Module Two How to Take Blood Pressure October 2011 MODULE TWO: How to Take Blood Pressure Module Two: HOW TO TAKE BLOOD PRESSURE Objectives Demonstrate proper technique

More information

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods Vol. 2, Issue 1, pages 31-36 ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY By Alejandro de la Sierra, MD Luis M. Ruilope, MD Hypertension Units, Hospital Clinico, Barcelona & Hospital 12 de Octubre,

More information

Does the reduction in systolic blood pressure alone explain the regression of left ventricular hypertrophy?

Does the reduction in systolic blood pressure alone explain the regression of left ventricular hypertrophy? (24) 18, S23 S28 & 24 Nature Publishing Group All rights reserved 95-92/4 $3. www.nature.com/jhh ORIGINAL ARTICLE Does the reduction in systolic blood pressure alone explain the regression of left ventricular

More information

Special articles from the ESH working group on blood pressure monitoring 23

Special articles from the ESH working group on blood pressure monitoring 23 Special articles from the ESH working group on blood pressure monitoring 23 European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices

More information

Protocol. Automated Ambulatory Blood Pressure Monitoring for the Diagnosis of Hypertension in Patients with Elevated Office Blood Pressure

Protocol. Automated Ambulatory Blood Pressure Monitoring for the Diagnosis of Hypertension in Patients with Elevated Office Blood Pressure Automated Ambulatory Blood Monitoring for the Diagnosis of Hypertension in Patients with Elevated Office Blood (10102) Medical Benefit Effective Date: 07/01/14 Next Review Date: 03/15 Preauthorization

More information

Use and Interpretation of Home Blood Pressure Monitoring

Use and Interpretation of Home Blood Pressure Monitoring Use and Interpretation of Home Blood Pressure Monitoring BLOOD PRESSURE in year 2015 PARAMETERS TODAY Gianfranco Parati University of Milano-Bicocca & Ospedale S.Luca, IRCCS, Istituto Auxologico Italiano

More information

Welcome to the S410. Blood Pressure System. Instruction Manual

Welcome to the S410. Blood Pressure System. Instruction Manual Welcome to the S410 Blood Pressure System Instruction Manual The simulator is to be used as part of an approved educational program for healthcare. It is not a substitute for traditional learning methods

More information

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator MSc Thesis in Clinical Science and Technology Sara Rose Newell Department of Health Science

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

UPPER ARM AUTOMATIC BLOOD PRESSURE MONITOR. Model: Blood Pressure Monitor INSTRUCTION MANUAL

UPPER ARM AUTOMATIC BLOOD PRESSURE MONITOR. Model: Blood Pressure Monitor INSTRUCTION MANUAL UPPER ARM AUTOMATIC BLOOD PRESSURE MONITOR Model: 2005 Blood Pressure Monitor INSTRUCTION MANUAL TABLE OF CONTENTS Introduction...2 Notes on safety...2 A Few Words About Blood Pressure...3 Precautions

More information

Automated versus manual blood pressure measurement: A randomized crossover trial

Automated versus manual blood pressure measurement: A randomized crossover trial Automated versus manual blood pressure measurement: A randomized crossover trial Author Heinemann, Megan, Sellick, Ken, Rickard, Claire, Reynolds, Pam, McGrail, Matthew Published 2008 Journal Title International

More information

Slide notes: References:

Slide notes: References: 1 2 3 Cut-off values for the definition of hypertension are systolic blood pressure (SBP) 135 and/or diastolic blood pressure (DBP) 85 mmhg for home blood pressure monitoring (HBPM) and daytime ambulatory

More information

European Society of Hypertension Practice Guidelines for home blood pressure monitoring

European Society of Hypertension Practice Guidelines for home blood pressure monitoring European Society of Hypertension Practice Guidelines for home blood pressure monitoring G Parati 1, GS Stergiou 2, R Asmar 3, G Bilo 1, P de Leeuw 4, Y Imai 5, K Kario 6, E Lurbe 7, A Manolis 8, T Mengden

More information

Please do not hesitate to contact our dedicated service unit for more details regarding this program at

Please do not hesitate to contact our dedicated service unit for more details regarding this program at The Blue Cross and Blue Shield Federal Employee Program (FEP) and the American Medical Association (AMA) have come together in a collaborative effort to provide physicians with resources designed to improve

More information

ORIGINAL INVESTIGATION. Is Isolated Home Hypertension as Opposed to Isolated Office Hypertension a Sign of Greater Cardiovascular Risk?

ORIGINAL INVESTIGATION. Is Isolated Home Hypertension as Opposed to Isolated Office Hypertension a Sign of Greater Cardiovascular Risk? ORIGINAL INVESTIGATION Is Isolated Home as Opposed to Isolated Office a Sign of Greater Cardiovascular Risk? Guillaume Bobrie, MD; Nathalie Genès, MD; Laurent Vaur, MD; Pierre Clerson, MD; Bernard Vaisse,

More information

The accurate measurement of blood pressure

The accurate measurement of blood pressure Position Paper ASH Position Paper: Home and Ambulatory Blood Pressure Monitoring When and How to Use Self (Home) and Ambulatory Blood Pressure Monitoring Thomas G. Pickering, MD, D Phil; 1 William B. White,

More information

Accurate blood pressure (BP) measurement is essential

Accurate blood pressure (BP) measurement is essential AJH 2006; 19:468 474 BP Measurement A Comparison of Home Measurement and Ambulatory Monitoring of Blood Pressure in the Adjustment of Antihypertensive Treatment Teemu J. Niiranen, Ilkka M. Kantola, Risto

More information

Microlife WatchBP Home A

Microlife WatchBP Home A Microlife The Microlife is designed to provide reliable and unbiased blood pressure measurement results, and strictly follows European Society of Hypertension (ESH) and American Heart Association (AHA)

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

The introduction of oscillometric automated blood pressure

The introduction of oscillometric automated blood pressure Atrial Fibrillation Impact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring Nikolaos Pagonas,* Sven Schmidt,* Jörg Eysel, Friederike Compton, Clemens Hoffmann, Felix Seibert,

More information

Home monitoring of blood glucose

Home monitoring of blood glucose Home Monitoring of Glucose and Blood Pressure JAMES R. TAYLOR, PharmD, CDE, and KENDALL M. CAMPBELL, MD, University of Florida, Gainesville, Florida Home monitoring of blood glucose and blood pressure

More information

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015 Original article 1 Clinical practice of ambulatory versus home blood pressure monitoring in hypertensive patients Jorge A. Paolasso, Florencia Crespo, Viviana Arias, Eduardo A. Moreyra, Ariel Volmaro,

More information

Journal of Hypertension 2004, 22: a Hypertension and Cardiovascular Rehabilitation Unit, Catholic University of

Journal of Hypertension 2004, 22: a Hypertension and Cardiovascular Rehabilitation Unit, Catholic University of Original article 81 Relationship between ambulatory blood pressure and followup clinic blood pressure in elderly patients with systolic hypertension Robert H. Fagard a, Jan A. Staessen a, Lutgarde Thijs

More information

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure?

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure? International Journal of Nursing December 2018, Vol. 5,. 2, pp. 94-99 ISSN 2373-7662 (Print) 2373-7670 (Online) Copyright The Author(s). All Rights Reserved. Published by American Research Institute for

More information

High Blood Pressure Blood Pressure Monitoring at Home -- F...

High Blood Pressure Blood Pressure Monitoring at Home -- F... Return to Web version High Blood Pressure Blood Pressure Monitoring at Home Why should I measure my blood pressure at home and keep a record of it? Measuring your blood pressure at home and keeping a record

More information

From measurement to profiles, phenomena and indices: a workshop of the European Society of Hypertension Eoin O Brien

From measurement to profiles, phenomena and indices: a workshop of the European Society of Hypertension Eoin O Brien Introductory note 291 From measurement to profiles, phenomena and indices: a workshop of the European Society of Hypertension Eoin O Brien Blood Pressure Monitoring 2005, 10:291 295 Blood Pressure Unit

More information

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License Zheng DC, Liu CY, Amoore J, Mieke S, Murray A. Need for Re-validation of Automated Blood Pressure Devices for use in Unstable Conditions. In: 215 Computing in Cardiology Conference (CinC). 215, Nice, France:

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the A comparison of blood pressure measurement over a sleeved arm versus a bare arm Grace Ma MD, Norman Sabin MD, Martin Dawes MBBS MD @ See related article page 591 DOI:10.1503/cmaj.070975 Abstract Background:

More information

5.2. MEASUREMENT OF BLOOD PRESSURE

5.2. MEASUREMENT OF BLOOD PRESSURE 5.2. MEASUREMENT OF BLOOD PRESSURE Purpose of experiment To examine a person s blood pressure. Tasks of experiment To record the pressure changes in the vascular system: using the traditional method measure

More information

Pulse pressure (PP) is a well-established marker of cardiovascular

Pulse pressure (PP) is a well-established marker of cardiovascular Different Prognostic Impact of 24-Hour Mean Blood Pressure and Pulse Pressure on Stroke and Coronary Artery Disease in Essential Hypertension Paolo Verdecchia, MD; Giuseppe Schillaci, MD; Gianpaolo Reboldi,

More information

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Anthony J. Viera, MD, MPH, FAHA Department of Family Medicine Hypertension Research Program UNC School of Medicine Objectives Review limitations

More information

STATE OF THE ART BP ASSESSMENT

STATE OF THE ART BP ASSESSMENT STATE OF THE ART BP ASSESSMENT PROFESSOR MOLECULAR PHARMACOLOGY CONWAY INSTITUE UNIVERSITY COLLEGE DUBLIN CO-CHAIRMAN BLOOD PRESCSURE MANAGEMENT IN LOW RESOURCE SETTINGS CENTRE FOR INTERNATIONAL HUMANITARIAN

More information

Review Paper. ID: 8418

Review Paper.   ID: 8418 Review Paper Call to Action on Use and Reimbursement for Home Blood Pressure Monitoring: Executive Summary A Joint Scientific Statement From the American Heart Association, American Society of Hypertension,

More information

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT Blood Pressure Measurement Can Isik Electrical Engineering and Computer Science Department Syracuse University Syracuse, NY 13244 cisik@syr.edu Abstract Blood pressure is one of the fundamental vital signs,

More information