Rapid-VIDITEST. Strep A Blister. One step Strep A Blister for the detection of Group A Streptococcal antigen from throat swabs or culture.
|
|
- Eunice Mosley
- 6 years ago
- Views:
Transcription
1 Rapid-VIDITEST Strep A Blister One step Strep A Blister for the detection of Group A Streptococcal antigen from throat swabs or culture. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, Jesenice, Czech Republic, Tel.: , INTENDED USE: The Rapid-VIDITEST Strep A Blister is a one step coloured chromatographic immunoassay for the qualitative detection of Group A Streptococcal antigen. It can be used with throat swabs or confirmation of presumptive Group A Streptococcal colonies recovered from culture. INTRODUCTION: Group A Streptococcal (Strep) infections are caused by Group A Streptococcus, a bacterium responsible for a variety of health problems. These infections can range from mild skin infection or sore throat to severe, life-threatening conditions such as toxic shock syndrome (multi-organ failures) and necrotizing fasciitis (soft tissue disease), commonly known as flesh eating disease. Most people are familiar with strep throat, which along with minor skin infection, is the most common form of the disease. Health experts estimate that more than 10 million mild infections (throat and skin) like these occur every year. Conventional identification procedures for Group A Streptococcus from throat swabs involve the isolation and subsequent identification of viable pathogens by techniques that require 24 to 48 hours or longer. The Rapid-VIDITEST Strep A Blister detects either viable or nonviable organisms directly from a throat swab, providing results within 5 minutes. PRINCIPLE OF THE TEST: The Rapid-VIDITEST Strep A Blister is a qualitative immunochromatographic assay for the determination of Group A Streptococcal antigen. The membrane is pre-coated with mouse monoclonal antibodies, on the test band region, to recognize this antigen. During testing, the sample is allowed to react with the coloured conjugate (anti-group A Streptococcal antibodies-red microspheres) which was pre-dried on the test. The mixture then moves upward on the membrane by capillary action. As the sample flows through the test membrane, the coloured particles migrate. In the case of a positive result the specific antibodies present on the membrane will capture the coloured conjugate (result region). The mixture continues to move across the membrane to the immobilized antibody placed in the control band region, a blue coloured band always appears. The presence of this blue band serves as: 1) verification that sufficient volume is added, 2) that proper flow is obtained and 3) as an internal control for the reagents
2 STORAGE AND STABILITY: Store as packaged in the sealed pouch at 15-30ºC. The test is stable through the expiration date printed on the sealed pouch. The test must remain in the sealed pouch until use. Do not freeze. MATERIALS PROVIDED: - Blister tests - Reagent A (2M Sodium Nitrite) and Reagent B (0.15M Acetic Acid) - Testing tubes - Sterile swabs - Disposable pipettes - Instructions for use MATERIALS REQUIRED BUT NO PROVIDED: - Disposable gloves - Timer SPECIMEN COLLECTION AND PREPARATION: Collect specimens with a sterile swab from the tonsils and/or the back of the throat. Take care to avoid the teeth, gums, tongue or cheek surfaces. Process the swab as soon as possible after collecting the specimen. If your lab requires a culture result as well as the Rapid-VIDITEST Strep A Blister result, streak the culture plate with the swab before starting the Rapid-VIDITEST Strep A Blister procedure. The extraction reagents will cause the specimen to become nonviable. The swabs can be stored at room temperature or in the refrigerator (2-4ºC) for up to 24 hours prior to testing. The swabs and the test kit must be brought to room temperature before testing. CULTURE CONFIRMATION: The Rapid-VIDITEST Strep A Blister can also be used to confirm the identification of Group A Streptococcus on blood agar plates. The plates must be less than 72 hours old. Lightly touch 1-3 suspect colonies (showing characteristic beta hemolysis) using a sterile swab. Do not sweep the plate. Follow the instructions in the test procedure section to test the swab. TEST PROCEDURE: Allow the tests, swabs and controls to reach to room temperature (15-30ºC) prior to testing. Do not open the package until ready to perform the assay. Only bring to room temperature the number of tests required to assay before opening it. 1. Add 5 drops (1) Reagent A (pink) and 5 drops (2) Reagent B in a testing tube (the solution should turn light yellow). 2. Immediately put the swab into the tube. 3. Mix the solution by rotating the swab forcefully against the side of the tube at least 10 times. Best results are obtained when the specimen is vigorously extracted in the solution (3). 4. Let stand for 1 minute. 5. Extract as much liquid as possible from the swab, pressing the sides of the tube. 6. Discard the swab
3 a) Using the blister Test single pack as a Card test: 7. Cut the blister to obtain a test single pack, hold the non sealed side and open it peeling off the upper foil. Don t remove the test from the blister cavity and use it as soon as possible. 8. Place the test on a flat surface. Use a separate pipette and device for each sample or control. Dispense exactly 4 drops or 125 µl from the testing tube, on the white end of the test (4a). 9. Read the result at 10 minutes. b) Using the blister Test single pack as a Strip test: By immersion: 7. Cut the blister to obtain a test single pack, hold the non sealed side and open it peeling off the upper foil. 8. Place the test vertically with the white end submerged into the sample taking care of not surpassing the limit of immersion indicated with the arrows (4b). 9. Read the result at 10 minutes. INTERPRETATION OF RESULTS (please refer to the illustration below): - 3 -
4 NEGATIVE: Only one blue band (control line) appears in the white central zone of the test (control region). POSITIVE: In addition to the blue control band, a red band (test line) also appears in the white central zone of the test (result region). INVALID: A total absence of the control coloured band (blue) regardless the appearance or not of the result line (red). Insufficient specimen volume, incorrect procedural techniques or deterioration of the reagents are the most likely reasons for control line failure. Review the procedure and repeat the test with a new test. If the problem persists, discontinue using the test kit and contact your local distributor. NOTES ON THE INTERPREATION OF RESULTS: The intensity of the red coloured band in the result region will vary depending on the concentration of antigens present in the specimen. However, neither the quantitative value, nor the rate of increase in antigens can be determined by this qualitative test. QUALITY CONTROL: Internal procedural controls are included in the test. - A blue line appearing in the control region is the internal procedural control. It confirms sufficient specimen volume and correct procedural technique. - The color of the liquid changes from pink to light yellow as you add Reagent B to Reagent A. This is an internal extraction reagent control. The color change means that you mixed the extraction reagent control, that you mixed the extraction reagent properly and that the reagents are functioning properly. External quality control testing. LIMITATIONS: 1. The test must be carried out within 2 hours of opening the sealed pack. 2. This test provides a presumptive diagnosis for Group A Streptococcus infections. A confirmed infection diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated. 3. The Rapid-VIDITEST Strep A Blister should be used only with throat swabs or colonies taken directly from a plate. The use of swab specimens taken from other sites or the use of other samples such as saliva, sputum or urine has not been established. The quality of the test depends on the quality of the sample; proper throat swab specimens must be obtained. 4. This test does not differentiate between carriers and acute infection. Pharyngitis may be caused by organisms other than Group A Streptococcus. 5. A negative result may be obtained if the specimen is inadequate or antigen concentration is below the sensitivity of the test. Therefore, it is recommended that all negative Rapid-VIDITEST Strep A Blister test results undergo confirmatory testing using a culture method. EXPECTED VALUES: Although strep throat can occur at any age and at any time of the year, it mainly affects school-age children during the winter and spring
5 PERFORMANCE: Sensitivity Different inactivated Group A Streptococcus cultures dilutions were tested directly in the sample diluent or spiked in a negative specimens in accordance with the kit instructions. Detection limit: 1,9x 10 4 CFU/mL. The detection of Group A Streptococcus with Rapid-VIDITEST Strep A Blister test showed >99% of sensitivity compared with other commercial rapid tests. Specificity The detection of Group A Streptococcus with Rapid-VIDITEST Strep A Blister test showed >99% of specificity compared with other commercial rapid tests. PRECAUTIONS: - Rapid-VIDITEST Strep A Blister is intended for professional in vitro diagnostic use. - Do not use after expiration date. - Do not use test devices if the aluminium has been damaged during storage. - All the specimens should be considered potentially hazardous and handled in the same manner as an infectious agent. A new swab must be used for each sample to avoid contamination errors. - The tests should be discarded in a proper biohazard container after testing. REFERENCES: 1. Dos Santos, Ana Gabriele P., Berezin, Fitan N.; Comparative analysis of clinical and laboratory methods for diagnosing Streptococcal sore throat. J. Pediatr. (Rio J.) 2005; 81 (1): Borbeau, Paul P. and Heither, Barbara J.; Use of swab without Transport media for the Gen-Probe Group A Strep Direct Test. Journal of Clinical Microbiology, July 2004, p Petts, D.N.; Evaluation of a modified nitrous acid extraction latex agglutination kit for grouping beta-hemolytic Streptococci and Enterococci. Journal of Clinical Microbiology, Apr. 1995, p Sheeler, Robert P. et al., Accuracy of Rapid Strep testing in patients who have had recent Streptococcal Pharyngitis. JABFP, July-August 2002, Vol 15 No 4, p SYMBOLS FOR IVD COMPONENTS AND REAGENTS: In vitro diagnostic device Use by Batch code Manufacturer - 5 -
Rapid-VIDITEST. Strep A Card. One step Strep A Card for the detection of Group A Streptococcal antigen from throat swabs or culture.
Rapid-VIDITEST Strep A Card One step Strep A Card for the detection of Group A Streptococcal antigen from throat swabs or culture. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, Vestec,
More informationRapid-VIDITEST. Influenza A+B
Rapid-VIDITEST Influenza A+B (One step Influenza A+B blister Test for the detection of Influenza type A and type B from nasal swabs, nasal wash or nasal aspirate specimens). Instruction manual Producer:
More informationRapid-VIDITEST. Influenza A
Rapid-VIDITEST Influenza A (One step Influenza A Card test for the detection of Influenza type A antigen from human nasopharyngeal specimens (swab, nasopharyngeal wash and aspirate). Instruction manual
More informationRapid-VIDITEST Shigella dysenteriae
Rapid-VIDITEST Shigella dysenteriae One step Shigella dysenteriae test for the qualitative detection of Shigella dysenteriae in faeces. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365,
More informationRapid-VIDITEST. Helicobacter pylori. One step Helicobacter pylori Blister test. Instruction manual
Rapid-VIDITEST Helicobacter pylori One step Helicobacter pylori Blister test. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, 252 50 Vestec, Czech Republic, Tel.: +420 261 090 565,
More informationRapid-VIDITEST FOB Blister
Rapid-VIDITEST FOB Blister One Step Fecal Occult Blood Blister test. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic, Tel.: +420 261 090 565,
More informationRapid-VIDITEST Swine Flu
Rapid-VIDITEST Swine Flu One Step Influenza type A Antigen Card test. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, 252 50 Vestec, Czech Republic, Tel.: +420 261 090 565, www.vidia.cz
More informationRapid-VIDITEST. Astrovirus Card
Rapid-VIDITEST Astrovirus Card One step Astrovirus Card Test. Instruction manual INTENDED USE: The Rapid-VIDITEST Astrovirus Card is a one step coloured chromatographic immunoassay for the qualitative
More informationRapid-VIDITEST. Rota-Adeno Blister. One Step Rotavirus and Adenovirus Antigen Blister test. Instruction manual
Rapid-VIDITEST Rota-Adeno Blister One Step Rotavirus and Adenovirus Antigen Blister test. Instruction manual INTENDED USE: The Rapid-VIDITEST Rota-Adeno Blister test is a rapid chromatographic immunoassay
More informationStrep-a-Test Twister Test
Strep-a-Test Twister Test Code: 24524 A rapid test for the qualitative detection of Strep A antigen in throat swab specimens. For professional in vitro diagnostic use only. INTENDED USE The Strep A Twist
More informationRapid-VIDITEST. Rota-Adeno. One Step Rotavirus and Adenovirus Antigen Blister test. Instruction manual
Rapid-VIDITEST Rota-Adeno One Step Rotavirus and Adenovirus Antigen Blister test. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic, Tel.: +420
More informationRapid-VIDITEST C. difficile Ag (GDH) Card/Blister
Li StarFish S.r.l. Via Cavour, 35-20063 Cernusco S/N (MI), Italy Tel. +39-02-92150794 - Fax. +39-02-92157285 info@listarfish.it -www.listarfish.it Rapid-VIDITEST C. difficile Ag (GDH) Card/Blister One
More informationRapid-VIDITEST Calprotectin
Rapid-VIDITEST Calprotectin One Step Calprotectin Card Test. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, 252 50 Vestec, Czech Republic, Tel.: +420 261 090 565, www.vidia.cz INTENDED
More informationRapid-VIDITEST Enterovirus
Rapid-VIDITEST Enterovirus A rapid one step Enterovirus Card test for the qualitative detection of Enterovirus antigens in human feces. Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365,
More informationHuman Influenza A (Swine Flu) Rapid test
Human Influenza A (Swine Flu) Rapid test Cat.No: DTSXY-Z9 Lot. No. (See product label) Size 20T Intended use The Influenza A (Swine Flu) test is a rapid chromatographic immunoassay for the qualitative
More informationSAMPLE PROCEDURE 907-9, 06/11
SAMPLE PROCEDURE This Sample Procedure is not intended as a substitute for your facility s Procedure Manual or reagent labeling, but rather as a model for your use in customizing for your laboratory s
More informationRapid-VIDITEST. FOB+Tf. One step Fecal Occult Blood Card Test. Instruction manual
Rapid-VIDITEST FOB+Tf One step Fecal Occult Blood Card Test Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic, Tel.: +420 261 090 565, www.vidia.cz
More informationRapid-VIDITEST. Calprotectin-Lactoferrin
Rapid-VIDITEST Calprotectin-Lactoferrin One step calprotectin and lactoferrin Card Test Instruction manual Producer: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic, Tel.:
More informationCDIA TM Rubella IgG/IgM Rapid Test Kit
CDIA TM Rubella IgG/IgM Rapid Test Kit Cat.No: DTSJZ024 Lot. No. (See product label) Intended Use The CDIA TM Rubella IgG/IgM Rapid Test Kit is a rapid chromatographic immunoassay for the qualitative detection
More informationAN IMMUNOASSAY TEST FOR THE QUALITATIVE DETECTION OF STREP A ANTIGEN IN THROAT SWAB SPECIMENS
CLARITY Strep A Dipsticks FOR LABORATORY AND PROFESSIONAL USE AN IMMUNOASSAY TEST FOR THE QUALITATIVE DETECTION OF STREP A ANTIGEN IN THROAT SWAB SPECIMENS CLIA COMPLEXITY: Waived CLARITY Strep A Dipsticks:
More informationRotavirus Test Kit. Instructions For Use. Format: Cassette Specimen: Fecal Extract Catalog Number: VEL-001-ROTA
Rotavirus Test Kit Instructions For Use Format: Cassette Specimen: Fecal Extract Catalog Number: VEL-001-ROTA * Please read the instructions carefully before use INTENDED USE Velotest Rotavirus Test is
More informationChlamydia trachomatis (CHLa)Test Kit
Chlamydia trachomatis (CHLa)Test Kit Instructions For Use Format: Cassette Specimen: Urethral/Genital Swab Catalog Number: VEL-001-CHLa * Please read the instructions carefully before use INTENDED USE
More informationChlamydia Rapid Screen Test (RAP-2858) RUO in the USA. Revised 28 Jul 2006
INDICATION For the rapid detection of Chlamydia Trachomatis antigens in swab specimens. For in vitro diagnostic use only, except in the United States where it is intended for Research Use Only. SUMMARY
More informationCLIA Complexity: WAIVED
CLIA Complexity: WAIVED INTENDED USE The is intended for the rapid, qualitative detection of Group A Streptococcal antigen from throat swabs or confirmation of presumptive Group A Streptococcal colonies
More informationHepatitis C Virus (HCV) Antibody Test
Hepatitis C Virus (HCV) Antibody Test Instructions For Use Format: Cassette For: Catalog Number: VEL-001-HCV Specimen: Serum/Plasma * Please read the instructions carefully before use INTENDED USE Hepatitis
More informationTreponema Pallidum (TP) Antibody Test
Treponema Pallidum (TP) Antibody Test Instructions For Use Format: Cassette For: Catalog Number: VEL-001-TP Specimen: Serum/Plasma * Please read the instructions carefully before use INTENDED USE Velotest
More informationMalaria Pf/pan antigen Rapid Test
Malaria Pf/pan antigen Rapid Test Cat. No.:IVDTS003 Pkg.Size:10T/50T Intended use The Malaria Pf/pan antigen Rapid Test is a self-performing, qualitative, sandwich immunoassay, utilizing whole blood for
More informationFecal H. pylori Antigen Rapid Test (Strip)
Fecal H. pylori Antigen Rapid Test (Strip) Cat. No.:DTS526 Pkg.Size: Intended use This H. pylori Antigen Rapid Test is intended for the direct qualitative detection of the presence of H. pylori antigen
More informationUSO PROFESSIONALE PROFESSIONAL USE
Gima S.p.A. - Via Marconi, 1-20060 Gessate (MI) Italy Italia: tel. 199 400 401 - fax 199 400 403 Export: tel. +39 02 953854209/221/225 fax +39 02 95380056 gima@gimaitaly.com - export@gimaitaly.com www.gimaitaly.com
More informationHelicobacter pylori Antigen Test
Helicobacter pylori Antigen Test Instructions For Use Format: Cassette For: Catalog Number: VEL-001-HP(s) Specimen: Fecal Specimen * Please read the instructions carefully before use INTENDED USE Helicobacter
More informationIVD Revised 22 July, 2008
INTENDED USE The DRG One Step HCG Pregnancy Test is an in vitro diagnostic test for the qualitative detection of human chorionic gonadotropin (hcg) in urine. The test is a two-site immunoassay employing
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx 0159-055-00 OraQuick HIV 1/2 Rapid Antibody Test with product codes 5x4-0010
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx 0159-055-00 OraQuick HIV 1/2 Rapid Antibody Test with product codes 5x4-0010
More informationDengue IgG/IgM/NS1 Combo Rapid Test Device
Dengue IgG/IgM/NS1 Combo Rapid Test Device Cat.No: DTSXY-L2 Lot. No. (See product label) Intended Use The Dengue IgG/IgM/NS1 Combo Rapid Test Device is a lateral flow chromatographic immunoassay for the
More informationTB LAM Ag Lateral Flow Assay Standard Operating Procedure
TB LAM Ag Lateral Flow Assay Standard Operating Procedure 1.0. Purpose The purpose of this standard operating procedure (SOP) is to detail the steps for correctly performing, interpreting, and documenting
More informationMono Test 1 DROP 1 DROP. For fingertip blood: For serum, plasma or whole blood samples in tubes: Rev , 05/09
1 DROP 1 DROP For serum, plasma or whole blood samples in tubes: For fingertip blood: Rev. 38102, 05/09 FOR LABORATORY AND PROFESSIONAL IN VITRO DIAGNOSTIC USE ONLY. INTENDED USE The SureVue Signature
More informationA rapid test for the qualitative detection of Malaria pf and pv antigen in human blood sample
A rapid test for the qualitative detection of Malaria pf and pv antigen in human blood sample For in vitro use only Intended Use For the rapid qualitative determination of Malaria P. falciparum specific
More informationQUICK REFERENCE INSTRUCTIONS. THYROCHEK TSH Cassette
QUICK REFERENCE INSTRUCTIONS THYROCHEK TSH Cassette A certificate of CLIA waiver is required to perform the testing in a waived setting. If the laboratory does not have a Certificate of Waiver, the Application
More informationNEGATIVE POSITIVE. Rev , 06/11. 5 min. Mono Test. For fi ngertip. blood: 1 DROP 1 DROP For serum, whole blood. plasma or. samples.
Mono Test 1 2 1 DROP 1 DROP For serum, plasma or whole blood samples in tubes: For fi ngertip blood: 3 4 5 min POSITIVE NEGATIVE Rev. 3078-0, 06/11 Mono Test CLIA Complexity: Waived for Whole Blood Non-Waived
More informationProduct Training & Certification
Product Training & Certification Oratect Oral Fluid Drug Screen Device For Professional Point of Care Use Oratect Training Page 1 of 12 MKT-045 (C) Oratect Oral Fluid Drug Screen Device Training and Certification
More informationHAV IgG/IgM Rapid Test
HAV IgG/IgM Rapid Test Cat. No.:DTS649 Pkg.Size: Intended use CD HAV IgG/IgM Rapid Test is a solid phase immunochromatographic assay for the rapid, qualitative and differential detection of IgG and IgM
More informationMalaria (Pan-LDH) W/B
CORTEZ DIAGNOSTICS, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 USA Tel: (818) 591-3030 Fax: (818) 591-8383 E-mail: onestep@rapidtest.com Web site: www.rapidtest.com See external label
More informationInfluenza A & B Screening
Chapter 8 Influenza A & B Screening CHAPTER 8: INFLUENZA TYPE A AND B Page 1 of 6 Flu 3-5-02.doc 6/24/2005 Procedure: Nasal Swab for Influenza type A and B by Quidel influenza test PRINCIPLE: The QuickVue
More informationWHO Prequalification of Diagnostics Programme PUBLIC REPORT
WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: ABON HIV 1/2/O Tri-Line Human Immunodeficiency Virus Rapid Test Device Number: PQDx 0141-051-00 Abstract ABON HIV 1/2/O Tri-Line Human
More informationCrAg Lateral Flow Assay Standard Operating Procedure
CrAg Lateral Flow Assay Standard Operating Procedure 1.0. Purpose The purpose of this standard operating procedure (SOP) is to detail the steps for correctly performing, interpreting, and documenting valid
More informationT R A I N I N G G U I D E
TRAINING GUIDE InstaCube Oral Fluid Drug Test For Forensic Use Only Powered By: Contents 3 The information in this presentation is a general overview on performing and interpreting the InstaCube Oral Fluid
More informationOpiates Rapid Test. Cat. No.:DTS137 Pkg.Size:50T. Intended use. General Description. Principle Of The Test. Reagents And Materials Provided
Opiates Rapid Test Cat. No.:DTS137 Pkg.Size:50T Intended use The CD One Step Opiates Screening Test is a rapid, qualitative immunoassay for the detection of opiates and opiate metabolites in urine. The
More informationHAV IgM Rapid Test(Cassette)
HAV IgM Rapid Test(Cassette) Cat. No.:DTS586 Pkg.Size:25 T Intended use The HAV IgM Rapid Test is a lateral flow chromatographic immunoassay for the qualitative detection of IgM antibody to Hepatitis A
More informationHBeAg and HBeAg Ab ELISA Kit
HBeAg and HBeAg Ab ELISA Kit Catalog Number KA0290 96 assays Version: 17 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3
More information25 OH Vitamin D Rapid Test
INSTRUCTION FOR USE REF:GDB 7120-25T 25 OH Vitamin D Rapid Test A Rapid Sandwich Immunochromatographic Test for Quantitative Detection of total 25-OH Vitamin D in human finger-prick blood For In Vitro
More informationINSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY
INTENDED USE INSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY INSTANT-VIEW H. pylori Rapid Test is a rapid qualitative immunoassay
More informationProcine sphingomyelin ELISA Kit
Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH
More informationMALARIA P. FALCIPARUM / P. VIVAX
MALARIA P. FALCIPARUM / P. VIVAX 1. EXPLANATION OF THE TEST: Malaria is a serious, sometimes fatal, parasitic disease characterized by fever, chills, and anemia and is caused by a parasite that is transmitted
More informationSAMPLE PROCEDURE , 07/11
SAMPLE PROCEDURE This Sample Procedure is not intended as a substitute for your facility s Procedure Manual or reagent labeling, but rather as a model for your use in customizing for your laboratory s
More informationQuickTox Drug Screen Dipcard (with and without Adulteration Tests)
QuickTox Drug Screen Dipcard (with and without Adulteration Tests) Training and Certification Program Presented by CLIAwaived.com, San Diego, CA Distributed by CLIAwaived.com www.cliawaived.com 1-858-481-5031
More informationToxCup Drug Screen Cup (with and without Adulteration Tests) Training and Certification Program
ToxCup Drug Screen Cup (with and without Adulteration Tests) Training and Certification Program ToxCup Training Page 1 of 8 Rev. B ToxCup Drug Screen Cup ToxCup Drug Screen Cup with Adulteration Tests
More informationProduct Training & Certification
B R A N A N M E D I C A L C O R P O R A T I O N Product Training & Certification Oratect III Oral Fluid Drug Screen Device Catalog # HM11 & HM12 For Forensic Use Only Branan Medical Corporation 140 Technology
More informationENG MYCO WELL D- ONE REV. 1.UN 29/09/2016 REF. MS01283 REF. MS01321 (COMPLETE KIT)
ENG MYCO WELL D- ONE MYCO WELL D-ONE System for the presumptive identification and antimicrobial susceptibility test of urogenital mycoplasmas, Gardnerella vaginalis, Trichomonas vaginalis, Candida albicans
More informationFastect II Drug Screen Dipstick Test Training and Certification Program
Fastect II Drug Screen Dipstick Test Training and Certification Program Fastect II Training Page 1 of 7 Rev. B Fastect II Drug Screen Dipstick Test Training and Certification for Test Administrators The
More information25 OH Vitamin D Rapid test
INSTRUCTION FOR USE REF:GDB 7120-25T 25 OH Vitamin D Rapid test Rapid Whole Blood Vitamin D Test A Rapid Sandwich Immunochromatographic Test for the Quantitative Detection of total 25-OH Vitamin D in human
More informationQuickTox Drug Screen Dipcard (with and without Adulteration Tests)
QuickTox Drug Screen Dipcard (with and without Adulteration Tests) Training and Certification Program Presented by Branan Medical Corporation Branan Medical Corporation 10015 Muirlands Road Irvine, California
More informationStep-by-Step Instructions For OraQuick HCV Rapid Antibody Test
Step-by-Step Instructions For OraQuick HCV Rapid Antibody Test Complexity: WAIVED for fingerstick whole blood and venipuncture whole blood. A Certificate of CLIA Waiver is required to perform the test
More informationSummary of Significant Changes at this Revision. Items Required. Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
COPY Summary of Significant Changes at this Revision 1. Change author from M Stubberfield to N Hodges Purpose and Scope Items Required The One Step Pregnancy Test urine test strip is a rapid chromatographic
More informationAVIAN INFLUENZA COMMODITIES TRAINING GUIDE MODULE 3 - SAMPLING, RAPID TESTING AND PACKAGING PARTICIPANT HANDOUTS
AVIAN INFLUENZA COMMODITIES TRAINING GUIDE MODULE 3 - SAMPLING, RAPID TESTING AND PACKAGING PARTICIPANT HANDOUTS MARCH 2007 MODULE 3 PARTICIPANT HANDOUTS Participant Handouts Participant Handout # 1 Flu
More informationIV2-113E Use by. Invitron Glargine ELISA Kit REF LOT IVD. Definitions. English. For in-vitro diagnostic use. Instructions for use.
Definitions Instructions for use REF Catalogue number IV2-113E Use by English Invitron Glargine ELISA Kit For in-vitro diagnostic use Σ 96 LOT IVD Lot/Batch Code Storage temperature limitations In vitro
More informationELISA-VIDITEST anti-chlamydia trachomatis IgA
1023 ELISA-VIDITEST anti-chlamydia trachomatis IgA ODZ-182 Instruction manual PRODUCER: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic, tel.: +420 261 090 565, www.vidia.cz
More informationInsulin (Porcine/Canine) ELISA
Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01
More informationELISA-VIDITEST anti-mycoplasma pneumoniae IgM
ELISA-VIDITEST anti-mycoplasma pneumoniae IgM Cat. No. ODZ-011 Instruction manual PRODUCER: VIDIA spol. s r.o., Nad Safinou II 365, 252 42 Jesenice, Czech Republic, Tel.: +420 261 090 565, www.vidia.cz
More informationMouse C-Peptide ELISA Kit
Mouse C-Peptide ELISA Kit Cat.No: DEIA4507 Lot. No. (See product label) Size 96T Intended Use The Mouse C-Peptide ELISA kit is for the quantitative determination of c-peptide in mouse serum, plasma, and
More informationCotinine (Mouse/Rat) ELISA Kit
Cotinine (Mouse/Rat) ELISA Kit Catalog Number KA2264 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle
More informationIVD information *Droppers for the sensitized and control cells. Not for use other than dispensing the sensitized and control cells.
In Vitro Diagnostic Reagent Instruction Manual of Diagnostic Reagent for Determination of anti-hbs Thoroughly read this instruction manual before use of this kit Background of the development and features
More informationTo provide you with necessary knowledge and skills to accurately perform 3 HIV rapid tests and to determine HIV status.
Module 9 Performing HIV Rapid Tests Purpose To provide you with necessary knowledge and skills to accurately perform 3 HIV rapid tests and to determine HIV status. Pre-requisite Modules Module 3: Overview
More informationBChek Strep A Test (Dipstick) For detection of group A streptococcal antigen directly from throat swabs / /04 INTENDED USE The BD Chek St
Revisions Rev from Rev to ECO # NEW 01 2742-04 SO 0191-5 Notes: 1. BD Cat. Number 256018 2. Blank (Sheet) Size : Length: 7.5 Width: 18 Number of Pages: 8 Number of Sheets: 1 Page Size: Length 7.5 Width
More informationToxCup Drug Screen Cup (with and without Adulteration Tests) Training and Certification Program
ToxCup Drug Screen Cup (with and without Adulteration Tests) Training and Certification Program ToxCup Training Page 1 of 8 Rev. B ToxCup Drug Screen Cup ToxCup Drug Screen Cup with Adulteration Tests
More informationInstructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600
Instructions for use TSH rat ELISA AR E-8600 TSH rat ELISA 1. INTRODUCTION 1.1 Intended use The TSH rat ELISA is an enzyme immunoassay for the quantitative measurement of TSH in rat serum. For research
More informationOratectPlus Oral Fluid Drug and Alcohol Screen Device Training and Certification Program
OratectPlus Oral Fluid Drug and Alcohol Screen Device Training and Certification Program OratectPlus Training Page 1 of 10 Rev. A OratectPlus Oral Fluid Drug and Alcohol Screen Device Training and Certification
More informationInfectious Mononucleosis IM Cassette Test RAPU04A830
Infectious Mononucleosis IM Cassette Test RAPU04A830 DIAsource ImmunoAssays S.A. - Rue de l'industrie, 8 - B-1400 Nivelles - Belgium : 090714/1 en DIAsource IM (Mononucleosis) Test Cassette for whole blood,
More informationELISA-VIDITEST anti-mycoplasma pneumoniae IgG assay Lot:
ELISA-VIDITEST anti-mycoplasma pneumoniae IgG assay Lot: 010810 Instruction manual PRODUCER: VIDIA Ltd., Nad Safinou II 365, 252 42 Jesenice, Czech Republic, Tel.: +420 261 090 565, Fax: +420 261 090 566
More informationRubella Latex Agglutination Test
Rubella Latex Agglutination Test Cat. No.:DLAT1088 Pkg.Size:30T Intended use The Rubella Latex Agglutination Test is a rapid latex particle agglutination test for the qualitative and semi-quantitative
More informationELISA-VIDITEST anti-tbev IgM ODZ-194
ELISA-VIDITEST anti-tbev IgM ODZ-194 Instruction manual PRODUCER: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic, Tel.: +420 261 090 565, www.vidia.cz, E-mail: info@vidia.cz
More informationSee external label. 2 C-30 C for test card and Sample tube Cat.# OneStep Rotavirus Antigen RapiCard InstaTest.
CORTEZ DIAGNOSTICS, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 USA Tel: (818) 591-3030 Fax: (818) 591-8383 E-mail: onestep@rapidtest.com Web site: www.rapidtest.com See external label
More informationANTIBODY SCREENING by Uni-Gold Recombigen HIV
ANTI-HIV SPECIMEN 1 REQUIREMENTS ANTIBODY SCREENING by Uni-Gold Recombigen HIV PRINCIPLE: The Uni-Gold Recombigen HIV was designed as a rapid immunoassay and is intended to detect antibodies to HIV- 1
More informationELISA-VIDITEST anti-chlamydia trachomatis IgM
1023 ELISA-VIDITEST anti-chlamydia trachomatis IgM ODZ-181 Instruction manual PRODUCER: VIDIA spol. s r.o., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic, tel.: +420 261 090 565, www.vidia.cz
More informationELISA-VIDITEST anti-chlamydia trachomatis IgA
1023 ELISA-VIDITEST anti-chlamydia trachomatis IgA Instruction manual PRODUCER: VIDIA Ltd., Nad Safinou II 365, Vestec, 252 42 Jesenice, Czech Republic, tel.: +420261 090565, www.vidia.cz 1. TITLE: ELISA-VIDITEST
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: SD BIOLINE HCV WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: SD BIOLINE HCV WHO reference number: PQDx 0257-012-00 SD BIOLINE HCV with product code 02FK10, manufactured by Standard Diagnostics,
More informationHuman HBcAb IgM ELISA kit
Human HBcAb IgM ELISA kit Catalog number: NR-R10163 (96 wells) The kit is designed to qualitatively detect HBcAb IgM in human serum or plasma. FOR RESEARCH USE ONLY. NOT FOR DIAGNOSTIC OR THERAPEUTIC PURPOSES
More informationTESTS, REF TESTS)
ENGLISH Read the Package Insert completely before using the product. Follow the instructions carefully when performing testing. Failure to do so may result in inaccurate test results. For In Vitro Diagnostic
More informationHuman HIV (1+2) antigen&antibody ELISA Kit
Human HIV (1+2) antigen&antibody ELISA Kit Catalog Number. CSB-E18042h For the qualitative determination of human HIV (1+2) antibody and P24 antigen concentrations in serum, plasma. This package insert
More informationMouse C3 (Complement Factor 3) ELISA Kit
Mouse C3 (Complement Factor 3) ELISA Kit Cat. No.:DEIA8289 Pkg.Size:96T Intended use The Mouse C3 (Complement Factor 3) ELISA Kit is a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring
More informationMurex Chlamydia Verification Kit
1F83-01 E C06DK05GB Murex Chlamydia Verification Kit DK05 For the confirmatory testing of clinical samples reactive in Murex Chlamydia (1F82-01) Note Changes Highlighted 2005 Abbott / Printed in the UK
More informationCLIA Complexity: WAIVED INTENDED USE SUMMARY AND EXPLANATION
CLIA Complexity: WAIVED INTENDED USE The QuickVue ifob (immunochemical Fecal Occult Blood) test is an immunochemical device intended for the qualitative detection of fecal occult blood by laboratories
More informationSerum Amyloid A ELISA
Serum Amyloid A ELISA For the quantitative determination of serum amyloid A (SAA) in serum plasma For Research use Only. Not for Use in Diagnostic Procedures Please see Appendix A for Reference Serum information
More informationH.Pylori IgG
DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com
More informationMouse HBsAg(Hepatitis B Virus Surface Antigen) ELISA Kit
Mouse HBsAg(Hepatitis B Virus Surface Antigen) ELISA Kit Catalogue No.: EM0002 Size: 96T Reactivity: Mouse Application: This immunoassay kit allows for the qualitative determination of HBsAg in Mouse serum
More informationH.Pylori IgG Cat # 1503Z
DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external
More informationIBD EZ VUE I. PRINCIPLE
IBD EZ VUE INTENDED USE The IBD EZ VUE test is an immunochromatographic test for the qualitative detection of elevated levels of lactoferrin, a marker for fecal leukocytes and an indicator of intestinal
More informationCLIA Complexity: Waived
CLIA Complexity: Waived INTENDED USE The QuickVue ifob (immunochemical Fecal Occult Blood) test is an immunochemical device intended for the qualitative detection of fecal occult blood by laboratories
More informationInstruction Manual Test kit for 20 tests individually pouched Including Positive Control REF 41115
QuickStripe + Positive control A rapid test device for the qualitative detection of Chlamydia trachomatis antigen in female cervical swab, male urethral swab and male urine specimens. Instruction Manual
More informationApolipoprotein A-1 ELISA
Apolipoprotein A-1 ELISA For the quantitative determination of apolipoprotein A1 in serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical
More information