Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

Size: px
Start display at page:

Download "Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon"

Transcription

1 (2009) 23, & 2009 Macmillan Publishers Limited All rights reserved /09 $ ORIGINAL ARTICLE Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon GS Stergiou, P Lourida, D Tzamouranis, NM Baibas Hypertension Center, Third University Department of Medicine, Sotiria Hospital, Athens, Greece Oscillometric devices are being widely used for ambulatory, home and office blood pressure (BP) measurement. However, even successfully validated oscillometric devices fail to provide accurate measurements in some patients. This study investigated the prevalence, the reproducibility and the characteristics of the phenomenon of unreliable oscillometric BP (UOBP) measurement. A total of 5070 BP measurements were obtained simultaneously (Y connector) using a professional oscillometric device (BpTRU) and a mercury sphygmomanometer in 755 patients (1706 visits). UOBP readings were defined as those with 410 mm Hg difference (systolic or diastolic) between the two methods. UOBP was found in 15% of systolic and 6.4% of diastolic BP measurements. In all, 18% of the participants had their first but not their second visit, or the reverse. However, 49% of these participants had at least one more UOBP visit after their second visit within the study database. Patients with persistent UOBP were more likely to be female and had lower arm circumference. The systolic BP discrepancy between the two methods was associated with pulse pressure (r ¼ 0.41) and inversely with diastolic BP (r ¼ 0.40) and arm circumference (r ¼ 0.30), whereas the diastolic discrepancy with diastolic BP (r ¼ 0.61) and inversely with pulse pressure (r ¼ 0.32). There was a consistent significant trend for larger systolic BP discrepancy and smaller diastolic from the lower to the higher pulse pressure quintile (Po0.0001). A decreasing arm circumference was a significant predictor of persistent UOBP. These data suggest that the UOBP measurement is particularly common, not very reproducible and mainly affected by pulse pressure and arm circumference. (2009) 23, ; doi: /jhh ; published online 26 March 2009 Keywords: blood pressure measurement; blood pressure monitoring; electronic devices; oscillometric devices Introduction Devices that measure blood pressure (BP) using the oscillometric technique are being widely used in clinical practice. This technique has almost completely dominated the area of ambulatory BP monitoring, 1,2 and home BP monitoring seems to follow the same direction. 3,4 More recently, professional oscillometric devices for office/clinic BP measurement appeared on the market and are expected to be increasingly used, particularly because of the recent ban of mercury sphygmomanometers in several countries. 5,6 Thus, any limitations of oscillometric BP measurement deserve close attention and thorough investigation. Correspondence: Professor GS Stergiou, Hypertension Center, Third University Department of Medicine, Sotiria Hospital, 152 Mesogion Avenue, Athens 11527, Greece. gstergi@med.uoa.gr Received 14 November 2008; revised 20 February 2009; accepted 22 February 2009; published online 26 March 2009 For the assessment of the accuracy of oscillometric devices, validation protocols have been developed by the American Association for the Advancement of Medical Instrumentation, 7 the British Hypertension Society 8 and the European Society of Hypertension. 9 In the last two decades several oscillometric BP monitors have successfully fulfilled the validation criteria of one or more of these protocols. 10 However, it is recognized that, even the best oscillometric BP monitors that comfortably pass the requirements of established validation protocols fail to provide accurate BP measurements in some patients. 11,12 This drawback of oscillometric BP measurement is well recognized and, therefore, the validation protocols allow for a rather high level of inaccuracy for devices that successfully pass the protocol requirements. 11,12 For example, the European Society of Hypertension International Protocol (ESH-IP) 9 accepts successfully validated devices to have up to 24 out of 99 BP measurements (24%) to differ by 410 mm Hg from the reference auscultatory BP measurement and up to 11 of 33 patients

2 (33%) to have a 45 mm Hg BP difference from auscultation in two of their three comparisons. Thus, it is accepted that successfully validated and therefore recommended oscillometric devices fail to provide accurate BP measurements in a significant proportion of patients. There is no doubt that this limitation of the oscillometric technique has important implications in the application of such monitors in clinical practice. However, factors responsible for the failure of the oscillometric technique to provide accurate BP measurements remain largely unknown. This prospective study was designed to investigate the prevalence and the reproducibility of the phenomenon of unreliable oscillometric BP (UOBP) measurement and the characteristics of these patients. Patients and methods Consecutive patients referred to an outpatients hypertension clinic for elevated BP, treated or untreated, were recruited. Participants had simultaneous same arm BP measurements (Y connector) using a validated professional oscillometric device (BpTRU Medical Devices, Ltd, Coquitlam, BC, Canada) 13,14 and a standard mercury sphygmomanometer (Baumanometer; WA Baum Co. Inc., New York, NY, USA). This is a standard procedure for routine office BP measurement in the outpatients hypertension clinic since Three cuffs of different size were used to fit the arm of each individual (inflatable bladder to cover % of the individual s arm circumference). BP measurements were performed in the clinic by three trained observers experienced in BP monitoring research. Before the study initiation and twice during the study period the observers were re-tested for agreement in BP measurement according to the British Hypertension Society protocol, 8 and the components of the mercury sphygmomanometers were carefully checked. Triplicate BP measurements were taken per participant at each visit after at least 5 min sitting rest and with at least 30 s between measurements. Information about the presence of cardiovascular disease (stroke, transient ischaemic attack, coronary heart disease, heart failure and peripheral artery disease), left ventricular hypertrophy (yearly echocardiography and electrocardiogram), renal failure (serum creatinine 41.2 mg/dl) and smoking (current, never and ex 46 months) were obtained from the patients records. Arrhythmia was detected by the investigators during the auscultatory BP measurements. The protocol was approved by the hospital scientific committee, which did not require an informed consent for the analysis of the outpatients BP data and the retrospective collection of demographic and other data from the patients records. Unreliable oscillometric BP (UOBP) readings were defined as individual readings with 410 mm Hg difference (systolic and/or diastolic) between the oscillometric and the auscultatory measurement. UOBP visits were defined as those with at least two UOBP readings of the triplicate readings performed at each visit. Patients with UOBP were defined as those with at least one UOBP visit. The reproducibility of the UOBP was assessed in patients with two or more visits by comparing the first versus the second visit. Patients with persistent UOBP were defined as those with at least two UOBP visits (any visits within the study database). A case control comparison of the characteristics of these patients (systolic and diastolic BP, pulse pressure, body mass index, arm circumference, cuff size, diabetes mellitus, renal failure, cardiovascular disease, left ventricular hypertrophy, arrhythmia, smoking, time from hypertension diagnosis, number of antihypertensive drugs and hypolipidemic drug treatment) was carried out against the first consecutive patients, sex and age (within 5 years) matched, but without none of their two or more study visits. Unpaired t-tests were used to compare continuous parameters of different subgroups of participants and w 2 -tests for categorical parameters. Paired t-tests were used for the case control comparison of patients with persistent UOBP against their controls. The kappa statistic was applied to assess the agreement in the detection of different visits of the same patients (first versus second visit of patients with at least two visits during the study). Pearson s correlations were used to investigate the relationship between the oscillometric auscultatory BP measurement discrepancy and systolic and diastolic BP, pulse pressure and arm circumference. Conditional logistic regression models were used to explore the association of variables with cases of persistent UOBP versus age- and sex-matched controls with non-discrepant BP measurements, in a case control analysis. The variables explored were auscultatory systolic and diastolic BP, arm circumference, pulse rate and auscultatory pulse pressure. Separate models were fitted for systolic/diastolic BP and pulse pressure to avoid collinearity between the variables. One-way analysis of variance was used to compare the oscillometric auscultatory BP discrepancy across the quintiles of auscultatory pulse pressure. Statistical analysis was carried out using the MINITAB INC Statistical Software (release 13.31) (Stage College, PA, USA) and conditional logistic regression using the STATA statistical package (StataCorp LP, College Station, TX USA). A probability level Po0.05 was regarded as statistically significant. Results Patients and BP measurements A total of 5154 BP readings were collected within 14 months. Erroneous readings and visits with fewer 795

3 796 than three valid readings were discarded. The analysis was on the basis of 5070 valid BP readings taken from 755 patients in 1706 clinic visits. In this dataset 40% of the patients had one clinic visit, 26% had two and 34% three or more visits. The characteristics of the study participants are presented in Table 1. Average auscultatory BP was 141.7±19.0/86.9±11.8 mm Hg (systolic/diastolic) and average oscillometric BP was 136.6±19.9/ 83.6±11.3 (mean difference systolic 5.0±6.2 mm Hg and diastolic 3.4±5.0 mm Hg). The proportion of UOBP readings as well as the proportion of UOBP visits and of participants with UOBP is presented in Figure 1 (separately for systolic and diastolic BP). UOBP reproducibility A total of 453 patients had at least two office visits during the study. Assessment of the UOBP reproducibility by comparing the first with the second visit of these 453 patients showed that 39 (9%) had UOBP in their first but not their second visit, 41 had the reverse (9%), 16 (4%) had both the first and the second visit and 357 (79%) in none (disagreement in UOBP detection between the two visits in 18% of patients; kappa statistic 0.18, s.e. 0.06, 95% confidence intervals (CI) 0.07, 0.30). Patients with UOPB in the first but not the second visit, or the reverse, (n ¼ 80) were compared with those with both visits (n ¼ 16), regarding age, gender, duration of hypertension, left ventricular hypertrophy, cholesterol, smoking, auscultatory BP, pulse rate, BP difference and antihypertensive treatment change between the two visits (initiation, addition or substitution). No differences between the two groups were found (data not shown), apart from a larger proportion of treatment changes between the two visits in the group with both visits (37 versus 14%, P ¼ 0.02). Thus, none of these factors seemed to explain the poor reproducibility of the UOBP phenomenon between the first and second visits. It is interesting that, of the 80 patients with disagreement regarding the UOBP diagnosis in the first versus the second visit, 39 (49%) had another visit after the second visit within the study database. Furthermore, as shown in Table 2, patients with at least one UOBP visit had a good chance to repeat this phenomenon if they had additional visits within the study database. For example, those with at least one UOBP visit and overall four or more visits within the database (on an average 4.8 visits) had 50% chance to have at least one more UOBP visit within the study database, 21% chance to have at least another two UOBP visits and 9% chance to have another three UOBP visits within the study database (Table 2). UOBP patients characteristics Fifty-nine patients had persistent UOBP defined as at least two study visits (any visits within the study database) with at least two UOBP readings per visit. A comparison of the characteristics of these patients against the rest of the study population showed that patients with persistent UOBP were more likely to be female and had lower arm circumference (Table 1). A case control comparison of these patients against the first consecutive 59 patients, sex and age matched, but without none of their two or more study visits, showed UOBP patients to have lower arm circumference (mean difference 1.2±3.1 cm, 95% CI 0.26, 2.1, P ¼ 0.01), lower body mass index (mean difference 3.0±4.4 kg/m 2, 95% CI 1.8, 4.1, Po0.0001) and a tendency for higher auscultatory systolic BP (mean difference 2.5±26.8 mm Hg, P ¼ NS) and higher pulse pressure (mean difference 2.7±22.4 mm Hg, P ¼ NS). No differences in other study parameters were found. BP Readings (n=5.070) Office Visits (n=1.706) Subjects (n=755) Systolic BP Diastolic BP Proportion (%) Figure 1 Prevalence of unreliable oscillometric blood pressure measurement in individual readings, visits or patients (for definitions see Patients and methods; BP, blood pressure). Table 1 Characteristics of subjects with persistent unreliable oscillometric blood pressure (UOBP) measurement compared with subjects without persistent UOBP and the total study population Persistent UOBP (n ¼ 59) Other non-uobp (n ¼ 696) Total database (n ¼ 755) Age (years) 58.2± ± ±12.9 Women 38 (64%) 319 (46%) * 367 (49%) Body mass index (kg/m 2 ) 26.0± ± ±4.3 Arm circumference (cm) 28.7± ±3.3 ** 29.7±3.3 Antihypertensive drugs 1.8± ± ±1.0 Diabetes mellitus 5 (9%) 64 (9%) 69 (9%) *P ¼ 0.01; **P ¼ compared with subjects with persistent UOBP.

4 Table 2 Persistence of the unreliable oscillometric blood pressure measurement (UOBP) phenomenon in repeated visits within the study dataset in subjects with at least one UOBP visit 797 Number of visits in the study dataset Average number of visits Xone visit Xtwo visits Number of subjects (%) Xthree visits Xfour visits X (31) X (38) 13 (14) X (50) 12 (21) 5 (9) X (59) 7 (24) 4 (14) UOBP associations In patients with persistent UOBP the systolic BP discrepancy between the oscillometric and the auscultatory measurement was strongly correlated with auscultatory pulse pressure (r ¼ 0.41, P ¼ 0.001) and inversely correlated with auscultatory diastolic BP (r ¼ 0.40, P ¼ 0.002) and arm circumference (r ¼ 0.30, P ¼ 0.05), whereas a weak association was found with auscultatory systolic BP (r ¼ 0.17, P ¼ NS). On the other hand, the diastolic BP discrepancy was strongly correlated with auscultatory diastolic BP (r ¼ 0.61, Po0.0001) and inversely with auscultatory pulse pressure (r ¼ 0.32, P ¼ 0.01), whereas no relationship was found with arm circumference (r ¼ 0.11) or auscultatory systolic BP (r ¼ 0.04). Table 3 Conditional logistic regression models investigating independent predictors of persistent unreliable oscillometric blood pressure (BP) measurement cases versus age- and sexmatched controls Variable Odds ratio P (95% CI) Model 1 Arm circumference (0.51, 0.95) Auscultatory systolic BP (0.98, 1.04) Auscultatory diastolic BP (0.91, 1.04) Pulse rate (0.97, 1.09) Model 2 Arm circumference (0.55, 0.96) Pulse pressure (0.98, 1.03) Pulse rate (0.97, 1.08) Abbreviation: CI, confidence intervals. UOBP predictors In conditional logistic regression analysis of patients with persistent UOBP measurements (n ¼ 59) versus age- and sex-matched controls with non-discrepant BP measurements, arm circumference was the only variable that showed a significant inverse association with persistently discrepant UOBP cases in both models that were fitted (Table 3). Pulse pressure Q1 Q2 Q3 Q4 Q5 P< Systolic blood pressure discrepancy (mmhg) BP discrepancy and pulse pressure The levels of systolic and diastolic BP discrepancy in quintiles of auscultatory pulse pressure in the total 5070 readings are shown in Figure 2 (1014 BP readings per quintile; Q1: mean pulse pressure 33.4 mm Hg, range 12 40; Q2: mean 43.8, range 40 48; Q3: mean 52.0, range 48 56; Q4: mean 62.2, range 56 68; Q5: mean 82.6, range ). There was a consistent trend for larger systolic BP discrepancy and smaller diastolic across the pulse pressure quintiles from Q1 to Q5 (Po for both). Discussion This study investigated the phenomenon of UOBP measurement in patients attending an outpatients BP linic. The main findings are, first, that this Pulse pressure Q1 Q2 Q3 Q4 Q5 P< Diastolic blood pressure discrepancy (mmhg) Figure 2 Systolic and diastolic blood pressure discrepancy between the oscillometric and the auscultatory measurement across the pulse pressure quintiles (Q1 Q5; 1014 readings per quintile; mean difference with 95% confidence intervals). phenomenon is particularly common, second that it is poorly reproducible and third that seems to be affected by pulse pressure.

5 798 This study has neither the elegance nor the complexity of a formal validation study of an oscillometric BP monitor. In contrast with validation studies (a) BP measurements were taken at an outpatients clinic instead of a research setting, (b) simultaneous instead of sequential BP measurements were obtained to compare the oscillometric versus the reference auscultatory BP measurements, and (c) a single instead of two observers performed each set of measurements. This simplified protocol was necessary to shorten the validation procedure in order to be applicable in a large number of patients. On the other hand, in order to increase the accuracy of the study in detecting the UOBP phenomenon, carefully standardized observers obtained three pairs of measurements and the threshold for UOBP was set at 410 mm Hg difference between the oscillometric and the auscultatory BP measurement instead of the conventional 5 mm Hg threshold. 7 9 The latter aimed to eliminate the expected underestimation of auscultatory systolic and overestimation of diastolic BP because of the faster deflation rate of the electronic device ( mm/s) compared with that recommended for auscultatory BP measurement (2 3 mm/s). 15 Furthermore, at least two of the triplicate measurements were required to define an UOBP visit, and two such visits were required to define persistent UOBP that was used to investigate predictors of this phenomenon. Oscillometric BP measurements in this study were obtained using a professional device (BPM- 100 or BpTRU) that has fulfilled the requirements of American Association for the Advancement of Medical Instrumentation validation protocol and achieved an A grade of the British Hypertension Society validation protocol for both systolic and diastolic BP measurement. 13,14 Although not formally investigated, the average oscillometric auscultatory BP discrepancy found in this study seems to fulfill the American Association for the Advancement of Medical Instrumentation validation criterion of o5 mm Hg with s.d. o8 mm Hg. 7 Furthermore, the 85% of the total of BP readings lying within 10 mm Hg from the reference method (Figure 1) also complies with the ESH-IP phase 2.1 criteria. 9 Despite the use of a successfully validated professional oscillometric device, the prevalence of the UOBP phenomenon in this study is striking. Irrespective of how this phenomenon was approached (in terms of single BP readings, single visits or individual patients), in about 10 18% of the cases for systolic BP and 5 8% for diastolic the oscillometric measurement failed to provide an accurate assessment of BP (Figure 1). These data confirm the findings of multiple validation studies of oscillometric BP monitors using the ESH-IP (phase 2.2), showing about 20% of each study s participants having a 45 mm Hg difference between the oscillometric and the reference BP method in two or more of their triplicate BP readings. 11 Although a looser threshold (10 mm Hg) was applied in this study, as mentioned above this was deemed necessary to improve the accuracy of the detection of the UOBP phenomenon in the conditions of a clinic rather than a research setting. The most puzzling finding of this study is the poor reproducibility of the UOBP phenomenon. About 9% of the participants with their first study visit were not classified as such in their second visit, whereas another 9% did the reverse. Thus, in 18% of the participants there was disagreement between the first and the second visit in the UOBP yes/no classification, with kappa statistic suggesting only slight agreement. 16 It is interesting that, although the appearance of a UOBP visit was often sporadic, as suggested by the poor reproducibility of the UOBP phenomenon when only the first and second visits were assessed, it seems that once the phenomenon appears, it is very likely that it will be found again if repeated visits are performed (Table 2). This study failed to identify factors related with the poor reproducibility of the UOBP phenomenon. The imperfect reproducibility of the BP measurement in the clinic probably is one of the factors responsible. 17 It should be noted that no earlier information on the reproducibility of the UOBP is available because in both the validation studies, and in other studies that assessed the UOBP phenomenon, all BP measurements of each participant were obtained in a single occasion. Several different approaches have been used in this study to identify factors related with the UOBP phenomenon (results of UOBP patients characteristics, associations, predictors and BP discrepancy and pulse pressure). The most interesting finding of this analysis is that pulse pressure, which is a reliable marker of arterial stiffness, seemed to be related with the oscillometric auscultatory BP discrepancy. Although not significant in all the study analyses, the trend was always the same, with increasing pulse pressure being associated with larger oscillometric versus auscultatory BP discrepancy. As shown in Figure 1, with increasing pulse pressure the systolic BP discrepancy was increased, whereas the diastolic discrepancy reduced. Clinically important (45 mm Hg) discrepancy between the oscillometric and the auscultatory BP measurement appeared at pulse pressure quintiles Q4 Q5 (pulse pressure 456 mm Hg) for systolic BP and Q1 for diastolic (pulse pressure o40 mm Hg). Thus it seems that the algorithm of the tested oscillometric device is accurate within a given range of pulse pressure (arterial stiffness). When pulse pressure is higher than this range the systolic BP measurement accuracy is reduced and when it is lower the diastolic. Four earlier studies, each one with entirely different design, support the view that arterial stiffness is related with UOBP measurement First is a population-based cohort study of 1808

6 healthy patients aged 455 years. 18 BP was measured twice using an oscillometric device (Dinamap, GE Healthcare, Chalfont St Gilles, UK) in the supine position by a physician (always first), and again 15 mins later using a Hawksley random-zero sphygmomanometer (Hawksley and Sons, Lancing, Sussex, UK) in the sitting position by a nurse. As expected, because of the measurement protocol the oscillometric BP was by 11/4 mm Hg (systolic/ diastolic) higher. Yet, the BP discrepancy was related to arterial stiffness assessed by carotidfemoral pulse wave velocity, independently of age, gender and BP level. 18 Second is a study in 305 patients with BP measured three times using an oscillometric device (Dinamap) and a mercury device in random order, which showed that the BP discrepancy between the two devices tended to increase in very old participants and in those with high pulse pressure. 19 Third is a retrospective analysis of six simultaneous BP measurements (SpaceLabs Inc, Redmont, Washington, WA, USA oscillometric versus mercury device) obtained in 192 young adults (diabetics type-1 and healthy controls) that showed the BP discrepancy between the two devices to differ in the two groups, and to be influenced mainly by the presence of diabetes and less so by age and diabetes duration. 20 Last is a validation study of an oscillometric home BP monitor in patients with end-stage renal disease (Microlife AG, Widnau, Switzerland), who are known to have advanced arterial stiffness. 21 Although the device fulfilled the ESH-IP validation criteria, the degree of the arterial stiffness assessed by measuring the augmentation index affected the accuracy of oscillometric BP measurement. 21 In this study, a relationship of an increasing oscillometric auscultatory BP discrepancy with small arm circumference was consistently found in three different analyses (results of UOBP patients characteristics, associations and predictors). An earlier validation study of another oscillometric device (Omron Corp., Kyoto, Japan) in 197 patients aged 6 16 years showed better accuracy for diastolic BP measurement at arm circumference cm compared with cm. 22 More research is needed to show the effect of arm/cuff size on the accuracy of oscillometric BP measurement. Taken together the above findings suggest an important role of pulse pressure and arterial stiffness on the phenomenon of UOBP. This study confirms this concept supported through different approaches applied in earlier studies, and in addition provides a real-life, prospectively collected and large database that allowed the investigation of the true prevalence, the reproducibility and the characteristics of this clinically important and common phenomenon. The conclusion is that oscillometric BP measurement technology is still imperfect and caution is needed when such monitors are used in patients with arterial stiffness. What is known about this topic K Oscillometric devices are being widely used for ambulatory, home and office blood pressure (BP) measurement. K Even successfully validated oscillometric devices fail to provide accurate measurements in some patients. Factors related to unreliable oscillometric BP (UOBP) measurement remain largely unknown. What this study adds K The prevalence, the reproducibility and the characteristics of the UOBP phenomenon were investigated in a prospective study involving BP measurements. K The findings suggest that in clinical practice the UOBP phenomenon is particularly common, poorly reproducible and mainly affected by pulse pressure and arm circumference. References 1 O Brien E, Asmar R, Beilin L, Imai Y, Mallion JM, Mancia G et al. European Society of Hypertension Working Group on Blood Pressure Monitoring. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. J Hypertens 2003; 21: Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN et al. Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension 2005; 45: Pickering TG, Miller NH, Ogedegbe G, Krakoff LR, Artinian NT, Goff D. American Heart Association; American Society of Hypertension; Preventive Cardiovascular Nurses Association. Call to action on use and reimbursement for home blood pressure monitoring: a joint scientific statement from the American Heart Association, American Society Of Hypertension, and Preventive Cardiovascular Nurses Association. Hypertension 2008; 52: Parati G, Stergiou GS, Asmar R, Bilo G, de Leeuw P, Imai Y et al. ESH Working Group on Blood Pressure Monitoring. European Society of Hypertension guidelines for blood pressure monitoring at home: a summary report of the Second International Consensus Conference on Home Blood Pressure Monitoring. J Hypertens 2008; 26: Pickering TG. What will replace the mercury sphygmomanometer? Blood Press Monit 2003; 8: Stergiou GS. Office blood pressure measurement with electronic devices: has the time come? Am J Hypertens 2008; 21: White WB, Berson AS, Robbins C, Jamieson MJ, Prisant LM, Roccella E et al. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers. Hypertension 1993; 21: O Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, O Malley K et al. The British Hypertension Society protocol for the evaluation of automated and semi- 799

7 800 automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990; 8: O Brien E, Pickering T, Asmar R, Myers M, Parati G, Staessen J, et al., On behalf of the Working Group on Blood Pressure Monitoring of the European Society of Hypertension Working Group on Blood Pressure Monitoring of the European Society of Hypertension. International Protocol for validation of blood pressure measuring devices in adults. Blood Press Monit 2002; 7: dabl s Educational Trust. Devices for blood pressure measurement. Assessed 2 November O Brien E, Atkins N. Validation and reliability of blood pressure monitors. In: W White (ed). Blood Press Monit Cardiovasc Med Ther. Humana Press Inc.: Totowa, NJ, USA, 2007, pp Gerin W, Schwartz AR, Schwartz JE, Pickering TG, Davidson KW, Bress J et al. Limitations of current validation protocols for home blood pressure monitors for individual patients. Blood Press Monit 2002; 7: Wright JM, Mattu GS, Perry Jr TL, Gelferc ME, Strange KD, Zorn A et al. Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor. Blood Press Monit 2001; 6: Mattu GS, Perry Jr TL, Wright JM. Comparison of the oscillometric blood pressure monitor (BPM-100 (Beta)) with the auscultatory mercury sphygmomanometer. Blood Press Monit 2001; 6: Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN et al. Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension 2005; 45: Viera AJ, Garrett JM. Understanding interobserver agreement: the kappa statistic. Fam Med 2005; 37: Stergiou GS, Alamara CV, Salgami EV, Vaindirlis IN, Dacou-Voutetakis C, Mountokalakis TD. Reproducibility of home and ambulatory blood pressure in children and adolescents. Blood Press Monit 2005; 10: Van Popele NM, Bos WJ, de Beer NA, van Der Kuip DA, Hofman A, Grobbee DE et al. Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer. Hypertension 2000; 36: Ni H, Wu C, Prineas R, Shea S, Liu K, Kronmal R et al. Comparison of Dinamap PRO-100 and mercury sphygmomanometer blood pressure measurements in a population-based study. Am J Hypertens 2006; 19: Van Ittersum FJ, Wijering RM, Lambert J, Donker AJ, Stehouwer CD. Determinants of the limits of agreement between the sphygmomanometer and the SpaceLabs device for blood pressure measurement in healthy volunteers and insulin-dependent diabetic patients. J Hypertens 1998; 16: Thompson AM, Eguchi K, Reznik ME, Shah SS, Pickering TG. Validation of an oscillometric home blood pressure monitor in an end-stage renal disease population and the effect of arterial stiffness on its accuracy. Blood Press Monit 2007; 12: Stergiou GS, Yiannes NG, Rarra VC. Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: the Arsakion School Study. Blood Press Monit 2006; 11:

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation (2009), 1 5 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room. Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room Abstract Ahmad Mirdamadi (1), Mostafa Etebari (2) Original Article BACKGROUND:

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? O R I G I N A L A R T I C L E Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? I. Aksoy, J. Deinum, J.W.M. Lenders, Th. Thien *# Department of

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Original article http://dx.doi.org/10.3345/kjp.2011.54.11.463 Korean J Pediatr 2011;54(11):463-469 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

Automatic oscillometric blood pressure devices are frequently

Automatic oscillometric blood pressure devices are frequently Arterial Stiffness as Underlying Mechanism of Disagreement Between an Oscillometric Blood Pressure Monitor and a Sphygmomanometer Nicole M. van Popele, Willem Jan W. Bos, Nicole A.M. de Beer, Deirdre A.M.

More information

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER Naser KA. Teaching Hospital Peradeniya, Peradeniya, Sri Lanka Zawahir S

More information

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009 42 Original article Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP, the Omron HBP T and the Pic Indolor Professional Roland

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 16 Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-71-E)

More information

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure?

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure? International Journal of Nursing December 2018, Vol. 5,. 2, pp. 94-99 ISSN 2373-7662 (Print) 2373-7670 (Online) Copyright The Author(s). All Rights Reserved. Published by American Research Institute for

More information

Home blood pressure measurement with oscillometric upper-arm devices

Home blood pressure measurement with oscillometric upper-arm devices REVIEW Home blood pressure measurement with oscillometric upper-arm devices R.L. Braam, Th. Thien *# Department of Internal Medicine, University Medical Centre St Radboud, PO Box 9101, 6500 HB Nijmegen,

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice.

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice. AJH 26; 19:81 89 BP Measurement Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice Tony C. Y. Pang and Mark A. Brown Background: To determine the extent of discrepancies between

More information

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015 Original article 1 Clinical practice of ambulatory versus home blood pressure monitoring in hypertensive patients Jorge A. Paolasso, Florencia Crespo, Viviana Arias, Eduardo A. Moreyra, Ariel Volmaro,

More information

In clinical practice, blood pressure (BP) treatment

In clinical practice, blood pressure (BP) treatment Original Paper Does Hidden Undercuffing Occur Among Obese Patients? Effect of Arm Sizes and Other Predictors of the Difference Between Wrist and Upper Arm Blood Pressures Hardik Doshi, MD; Alan B. Weder,

More information

How do we diagnose hypertension today? Presentation Subtitle

How do we diagnose hypertension today? Presentation Subtitle How do we diagnose hypertension today? Presentation Subtitle Renata Cífková Case 1 JM, a 64-year-old lady referred to our center because of undesirable effects of her antihypertensive medication Personal

More information

How well do office and exercise blood pressures predict sustained hypertension? A Dundee Step Test Study

How well do office and exercise blood pressures predict sustained hypertension? A Dundee Step Test Study (2000) 14, 429 433 2000 Macmillan Publishers Ltd All rights reserved 0950-9240/00 $15.00 www.nature.com/jhh ORIGINAL ARTICLE How well do office and exercise blood pressures predict sustained hypertension?

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 223 Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol Paolo Verdecchia, Fabio Angeli, Roberto Gattobigio,

More information

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License Zheng DC, Liu CY, Amoore J, Mieke S, Murray A. Need for Re-validation of Automated Blood Pressure Devices for use in Unstable Conditions. In: 215 Computing in Cardiology Conference (CinC). 215, Nice, France:

More information

There is convincing evidence in clinical studies

There is convincing evidence in clinical studies AJH 1998;11:1413 1417 Reliability of Reporting Self-Measured Blood Pressure Values by Hypertensive Patients Thomas Mengden, Rosa Maria Hernandez Medina, Belen Beltran, Elena Alvarez, Karin Kraft, and Hans

More information

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008 Analytical methods and statistical analyses 285 PA.NET International Quality Certification Protocol for blood pressure monitors Stefano Omboni a,b, Carlo Costantini a,b,c, Claudio Pini b,d, Roberto Bulegato

More information

Nomogram of the Relation of Brachial-Ankle Pulse Wave Velocity with Blood Pressure

Nomogram of the Relation of Brachial-Ankle Pulse Wave Velocity with Blood Pressure 801 Original Article Nomogram of the Relation of Brachial-Ankle Pulse Wave Velocity with Blood Pressure Akira YAMASHINA, Hirofumi TOMIYAMA, Tomio ARAI, Yutaka KOJI, Minoru YAMBE, Hiroaki MOTOBE, Zydem

More information

Within-Home Blood Pressure Variability on a Single Occasion Has Clinical Significance

Within-Home Blood Pressure Variability on a Single Occasion Has Clinical Significance Published online: May 12, 2016 2235 8676/16/0041 0038$39.50/0 Mini-Review Within-Home Blood Pressure Variability on a Single Occasion Has Seiichi Shibasaki a, b Satoshi Hoshide b Kazuomi Kario b a Department

More information

Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus Devices Microlife A2 Basic (BP3GQ1-3P) 9 BP3BT0-A 9 Microlife BP A100 Plus 9 Image 10 10 10 Validation

More information

The introduction of oscillometric automated blood pressure

The introduction of oscillometric automated blood pressure Atrial Fibrillation Impact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring Nikolaos Pagonas,* Sven Schmidt,* Jörg Eysel, Friederike Compton, Clemens Hoffmann, Felix Seibert,

More information

RESEARCH INTRODUCTION

RESEARCH INTRODUCTION Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial Martin G Myers, professor of medicine, 1 Marshall

More information

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials. AJH 1995; 8:311-315 BRIEF COMMUNICATIONS Lack of Placebo Effect on Ambulatory Blood Pressure Giuseppe Mancia, Stefano Omboni, Gianfranco Parati, Antonella Ravogli, Alessandra Villani, and Alberto Zanchetti

More information

Arm position and blood pressure: a risk factor for hypertension?

Arm position and blood pressure: a risk factor for hypertension? (3) 17, 389 395 & 3 Nature Publishing Group All rights reserved 09-9240/03 $25.00 www.nature.com/jhh ORIGINAL ARTICLE Arm position and blood pressure: a risk factor for hypertension? A Mourad, S Carney,

More information

The increasing awareness of hypertension as a serious

The increasing awareness of hypertension as a serious AJH 2002; 15:787 792 Evaluation of the Performance of a Wrist Blood Pressure Measuring Device With a Position Sensor Compared to Ambulatory 24-Hour Blood Pressure Measurements Sakir Uen, Burkard Weisser,

More information

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA)

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) THE NEW ARMENIAN MEDICAL JOURNAL Vol.5 (2011), Nо 2, p.29-34 DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) Zelveian P.H. 1, 2, Podosyan G.A. 2 1 Institute

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04) SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: MEASURING BLOOD PRESSURE - MANUAL Nursing DATE: REVIEWED: PAGES: 2/80 7/17 1 of 5 RESPONSIBILITY: RN, LPN, Patient Care Technician Multi-skilled

More information

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods Vol. 2, Issue 1, pages 31-36 ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY By Alejandro de la Sierra, MD Luis M. Ruilope, MD Hypertension Units, Hospital Clinico, Barcelona & Hospital 12 de Octubre,

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

THE ACCURATE ASSESSMENT

THE ACCURATE ASSESSMENT ORIGINAL INVESTIGATION Consistency of Blood Pressure Differences Between the Left and Right s Kazuo Eguchi, MD; Mona Yacoub, MD; Juhee Jhalani; William Gerin, PhD; Joseph E. Schwartz, PhD; Thomas G. Pickering,

More information

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice (2005) 19, 801 807 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Prognostic significance of blood pressure measured in the office, at home and

More information

Validation Study Result Form

Validation Study Result Form Validation Study Result Form Please complete Section 1 to Section 3 of this form and return it to dabl Educational Trust with copies of the validation plots. The requirements for each of these sections

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

greater than 90% of patients having blood pressure measured consistently according to the protocol.

greater than 90% of patients having blood pressure measured consistently according to the protocol. Clinical methods and pathophysiology 1 Reliability and validity of blood pressure measurement in the Secondary Prevention of Small Subcortical Strokes study Pablo E. Pérgola a, Carole L. White b, John

More information

Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus Devices Microlife A6 PC (BP3GU1-8Y) 9 BP3BT0-A 9 Microlife BP A100 Plus 9 Image 10 10 10 Validation

More information

DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI.

DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI. DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI. Franco Cipollini, Carlo Porta, Enrica Arcangeli, Carla Breschi, & Giuseppe Seghieri Azienda USL 3, Ambulatorio

More information

Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations

Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations The American Journal of Medicine (2006) 119, 69.e13-69.e18 CLINICAL RESEARCH STUDY Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations

More information

Association of Isolated Systolic, Isolated Diastolic, and Systolic-Diastolic Masked Hypertension With Carotid Artery Intima-Media Thickness

Association of Isolated Systolic, Isolated Diastolic, and Systolic-Diastolic Masked Hypertension With Carotid Artery Intima-Media Thickness ORIGINAL PAPER Association of Isolated Systolic, Isolated Diastolic, and Systolic-Diastolic Masked Hypertension With Carotid Artery Intima-Media Thickness Efstathios Manios, MD; 1 Fotios Michas, MD; 1

More information

CARDIOVASCULAR RISK FACTORS & TARGET ORGAN DAMAGE IN GREEK HYPERTENSIVES

CARDIOVASCULAR RISK FACTORS & TARGET ORGAN DAMAGE IN GREEK HYPERTENSIVES CARDIOVASCULAR RISK FACTORS & TARGET ORGAN DAMAGE IN GREEK HYPERTENSIVES C. Liakos, 1 G. Vyssoulis, 1 E. Karpanou, 2 S-M. Kyvelou, 1 V. Tzamou, 1 A. Michaelides, 1 A. Triantafyllou, 1 P. Spanos, 1 C. Stefanadis

More information

European Society of Hypertension Practice Guidelines for home blood pressure monitoring

European Society of Hypertension Practice Guidelines for home blood pressure monitoring European Society of Hypertension Practice Guidelines for home blood pressure monitoring G Parati 1, GS Stergiou 2, R Asmar 3, G Bilo 1, P de Leeuw 4, Y Imai 5, K Kario 6, E Lurbe 7, A Manolis 8, T Mengden

More information

Importance of Ambulatory Blood Pressure Monitoring in Adolescents

Importance of Ambulatory Blood Pressure Monitoring in Adolescents Importance of Ambulatory Blood Pressure Monitoring in Adolescents Josep Redon, MD, PhD, FAHA Internal Medicine Hospital Clinico Universitario de Valencia University of Valencia CIBERObn Instituto de Salud

More information

White coat and masked hypertension

White coat and masked hypertension White coat and masked hypertension Conflict of interest Support from Spacelabs, Microlife. Honoraria from Novartis, Elpen, Boeringer-Ingelheim, CANA, Lilly, MSD, Sanofi, Menarini, Ciezi, Astra-Zeneca.

More information

High-dose monotherapy vs low-dose combination therapy of calcium channel blockers and angiotensin receptor blockers in mild to moderate hypertension

High-dose monotherapy vs low-dose combination therapy of calcium channel blockers and angiotensin receptor blockers in mild to moderate hypertension (2005) 19, 491 496 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE High-dose monotherapy vs low-dose combination therapy of calcium channel blockers

More information

Primary hypertension in adults

Primary hypertension in adults Primary hypertension in adults NICE provided the content for this booklet which is independent of any company or product advertised Hypertension Welcome NICE published an updated guideline on the diagnosis

More information

5.2 Key priorities for implementation

5.2 Key priorities for implementation 5.2 Key priorities for implementation From the full set of recommendations, the GDG selected ten key priorities for implementation. The criteria used for selecting these recommendations are listed in detail

More information

No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death

No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death STROKE statistics No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death Every 6 seconds a person in the world has a STROKE regardless of age and gender. Worldwide 15 million people suffer

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

Blood Pressure Response Under Chronic Antihypertensive Drug Therapy

Blood Pressure Response Under Chronic Antihypertensive Drug Therapy Journal of the American College of Cardiology Vol. 53, No. 5, 29 29 by the American College of Cardiology Foundation ISSN 735-197/9/$36. Published by Elsevier Inc. doi:1.116/j.jacc.28.9.46 Hypertension

More information

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE General practice Indicators for the NICE menu

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE General practice Indicators for the NICE menu NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE General practice Indicators for the NICE menu Indicator area: Pulse rhythm assessment for AF Indicator: NM146 Date: June 2017 Introduction There is evidence

More information

Ambulatory Blood Pressure Monitoring Clinical Practice Recommendations

Ambulatory Blood Pressure Monitoring Clinical Practice Recommendations Acta Medica Marisiensis 2016;62(3):350-355 DOI: 10.1515/amma-2016-0038 UPDATE Ambulatory Blood Pressure Monitoring Clinical Practice Recommendations Mako Katalin *, Ureche Corina, Jeremias Zsuzsanna University

More information

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood SPECIAL ARTICLE Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Eoin O Brien a,thomas

More information

Please do not hesitate to contact our dedicated service unit for more details regarding this program at

Please do not hesitate to contact our dedicated service unit for more details regarding this program at The Blue Cross and Blue Shield Federal Employee Program (FEP) and the American Medical Association (AMA) have come together in a collaborative effort to provide physicians with resources designed to improve

More information

Home monitoring of blood glucose

Home monitoring of blood glucose Home Monitoring of Glucose and Blood Pressure JAMES R. TAYLOR, PharmD, CDE, and KENDALL M. CAMPBELL, MD, University of Florida, Gainesville, Florida Home monitoring of blood glucose and blood pressure

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

The accurate measurement of blood pressure

The accurate measurement of blood pressure Position Paper ASH Position Paper: Home and Ambulatory Blood Pressure Monitoring When and How to Use Self (Home) and Ambulatory Blood Pressure Monitoring Thomas G. Pickering, MD, D Phil; 1 William B. White,

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Indirect ambulatory blood pressure monitoring devices

Indirect ambulatory blood pressure monitoring devices AJH 1992;5:880-886 Comparison of Ambulatory and Clinic Blood Pressure and Heart Rate in Older Persons With Isolated Systolic Hypertension Gale H. Rutan, Robert H. McDonald, and Lewis H. Kuller We compared

More information

Maker Spacelabs Healthcare Manufacturer Spacelabs Healthcare. Brand Spacelabs Model OnTrak

Maker Spacelabs Healthcare Manufacturer Spacelabs Healthcare. Brand Spacelabs Model OnTrak Validation Study Result Form Please complete Section 1 to Section 3 of this form and return it to dabl Educational Trust with copies of the validation plots and a digital photograph of the device used

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the Omron M6 Comfort IT (HEM-7322U-E) with the Omron M6 Comfort (HEM-7321-E) Devices Omron M6 Comfort IT (HEM-7322U-E) Omron M6 Comfort (HEM-7321-E) Pictures Display Validation ESH 2010 Device

More information

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines TITLE: Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines DATE: 25 September 2012 CONTEXT AND POLICY ISSUES A mercury manual sphygmomanometer

More information

Hypertension is a common medical disorder, affecting. Is Resistant Hypertension Really Resistant? Mark A. Brown, Megan L. Buddle, and Allison Martin

Hypertension is a common medical disorder, affecting. Is Resistant Hypertension Really Resistant? Mark A. Brown, Megan L. Buddle, and Allison Martin AJH 2001; 14:1263 1269 Is Resistant Hypertension Really Resistant? Mark A. Brown, Megan L. Buddle, and Allison Martin Background: Managing resistant hypertension is difficult and mostly involves expensive

More information

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION*

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Progress in Clinical Medicine 1 ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Keishi ABE** Asian Med. J. 44(2): 83 90, 2001 Abstract: J-MUBA was a large-scale clinical

More information

EUROPEAN SOCIETY OF HYPERTENSION GUIDELINES FOR THE USE OF HOME BLOOD PRESSURE MONITORING. A SUMMARY REPORT OF THE

EUROPEAN SOCIETY OF HYPERTENSION GUIDELINES FOR THE USE OF HOME BLOOD PRESSURE MONITORING. A SUMMARY REPORT OF THE 1 EUROPEAN SOCIETY OF HYPERTENSION GUIDELINES FOR THE USE OF HOME BLOOD PRESSURE MONITORING. A SUMMARY REPORT OF THE SECOND INTERNATIONAL CONSENSUS CONFERENCE ON HOME BLOOD PRESSURE MONITORING Short title:

More information

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA Case 1 What should be your BP goal for an elderly (> 75 yrs of

More information

Comparison of arbitrary definitions of circadian time periods with those determined by wrist actigraphy in analysis of ABPM data

Comparison of arbitrary definitions of circadian time periods with those determined by wrist actigraphy in analysis of ABPM data Journal of Human Hypertension (1999) 13, 449 453 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE Comparison of arbitrary definitions of

More information

How Low Do We Go? Update on Hypertension

How Low Do We Go? Update on Hypertension How Low Do We Go? Update on Beth L. Abramson, MD, FRCPC, FACC As presented at the University of Toronto s Saturday at the University Session (September 2003) Arecent World Health Organization report states

More information

Special articles from the ESH working group on blood pressure monitoring 23

Special articles from the ESH working group on blood pressure monitoring 23 Special articles from the ESH working group on blood pressure monitoring 23 European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices

More information

Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements

Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements Original Article J Clin Med Res. 2018;10(4):309-313 Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements Kannayiram Alagiakrishnan a, d, Ruojin

More information

Relationship of home blood pressure with target-organ

Relationship of home blood pressure with target-organ (11) 3, & 11 The Japanese Society of Hypertension All rights reserved 91-93/11 $3. www.nature.com/hr ORIGINAL ARTICLE Relationship of home blood pressure with target-organ damage in children and adolescents

More information

STATE OF THE ART BP ASSESSMENT

STATE OF THE ART BP ASSESSMENT STATE OF THE ART BP ASSESSMENT PROFESSOR MOLECULAR PHARMACOLOGY CONWAY INSTITUE UNIVERSITY COLLEGE DUBLIN CO-CHAIRMAN BLOOD PRESCSURE MANAGEMENT IN LOW RESOURCE SETTINGS CENTRE FOR INTERNATIONAL HUMANITARIAN

More information

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Anthony J. Viera, MD, MPH, FAHA Department of Family Medicine Hypertension Research Program UNC School of Medicine Objectives Review limitations

More information

Low fractional diastolic pressure in the ascending aorta increased the risk of coronary heart disease

Low fractional diastolic pressure in the ascending aorta increased the risk of coronary heart disease (2002) 16, 837 841 & 2002 Nature Publishing Group All rights reserved 0950-9240/02 $25.00 www.nature.com/jhh ORIGINAL ARTICLE Low fractional diastolic pressure in the ascending aorta increased the risk

More information

a Hypertension Center, Third University Department of Medicine, Sotiria Hospital,

a Hypertension Center, Third University Department of Medicine, Sotiria Hospital, Special articles from the ESH working group on blood pressure monitoring 39 European Society of Hypertension International Protocol for the validation of blood pressure monitors: a critical review of its

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the Microlife BP A150 AFIB with the Microlife BP A100 Plus and Microlife BP A100 Device Equivalence Evaluation Form Devices Microlife BP A150 AFIB 9 Microlife BP A100 Plus 9 Microlife BP

More information

Home blood pressure (BP) measurement has been reported

Home blood pressure (BP) measurement has been reported Prognostic Value of the Variability in Home-Measured Blood Pressure and Heart Rate The Finn-Home Study Jouni K. Johansson, Teemu J. Niiranen, Pauli J. Puukka, Antti M. Jula Abstract The objective of the

More information

Céline Guggiari a, Christophe Büla a, Katia Iglesias b and Bernard Waeber b,c

Céline Guggiari a, Christophe Büla a, Katia Iglesias b and Bernard Waeber b,c Published in Blood Pressure Monitoring 19, issue 11, 32-37, 2014 which should be used for any reference to this work 1 Measurement with an automated oscillometric wrist device with position sensor leads

More information

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM Faculty Disclosure I have no financial interest to disclose No off-label use of medications will be discussed FIFTH ANNUAL SYMPOSIUM Recognize changes between

More information

Prevalence, awareness, treatment and control of hypertension in employees of factories of Northern Greece: the Naoussa study

Prevalence, awareness, treatment and control of hypertension in employees of factories of Northern Greece: the Naoussa study (2004) 18, 623 629 & 2004 Nature Publishing Group All rights reserved 0950-9240/04 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Prevalence, awareness, treatment and control of hypertension in employees of

More information

ORIGINAL INVESTIGATION. C-Reactive Protein Concentration and Incident Hypertension in Young Adults

ORIGINAL INVESTIGATION. C-Reactive Protein Concentration and Incident Hypertension in Young Adults ORIGINAL INVESTIGATION C-Reactive Protein Concentration and Incident Hypertension in Young Adults The CARDIA Study Susan G. Lakoski, MD, MS; David M. Herrington, MD, MHS; David M. Siscovick, MD, MPH; Stephen

More information

Arterial Age and Shift Work

Arterial Age and Shift Work 340 Arterial Age and Shift Work Ioana Mozos 1*, Liliana Filimon 2 1 Department of Functional Sciences, Victor Babes University of Medicine and Pharmacy, Timisoara, Romania 2 Department of Occupational

More information

Use and Interpretation of Home Blood Pressure Monitoring

Use and Interpretation of Home Blood Pressure Monitoring Use and Interpretation of Home Blood Pressure Monitoring BLOOD PRESSURE in year 2015 PARAMETERS TODAY Gianfranco Parati University of Milano-Bicocca & Ospedale S.Luca, IRCCS, Istituto Auxologico Italiano

More information

Prognostic significance of blood pressure measured on rising

Prognostic significance of blood pressure measured on rising (2001) 15, 413 417 2001 Nature Publishing Group All rights reserved 0950-9240/01 $15.00 www.nature.com/jhh ORIGINAL ARTICLE Prognostic significance of blood pressure measured on rising P Gosse, C Cipriano,

More information