Imaging of the Cardiac Venous System: Comparison of MDCT and Conventional Angiography

Size: px
Start display at page:

Download "Imaging of the Cardiac Venous System: Comparison of MDCT and Conventional Angiography"

Transcription

1 Mühlenbruc h et al. Imaging the Cardiac Venous System Cardiac Imaging Clinical Observations Georg Mühlenbruch 1 Ralf Koos 2 Joachim E. Wildberger 1 Rolf W. Günther 1 Andreas H. Mahnken 1 Mühlenbruch G, Koos R, Wildberger JE, Günther RW, Mahnken AH DOI: /AJR Received August 3, 2004; accepted after revision November 10, Department of Diagnostic Radiology, RWTH Aachen University Hospital, Pauwelsstrasse 30, Aachen 52057, Germany. Address correspondence to G. Mühlenbruch (gmuehlen@rad.rwth-aachen.de). 2 Department of Cardiology, RWTH Aachen University Hospital, Aachen 52057, Germany. AJR 2005; 185: X/05/ American Roentgen Ray Society Imaging of the Cardiac Venous System: Comparison of MDCT and Conventional Angiography OBJECTIVE. Diagnostic and therapeutic strategies in electrophysiology and interventional cardiology include the coronary venous system. The purpose of this study was to compare MDCT angiography with conventional coronary sinus angiography in terms of detailed anatomic display of the coronary veins. CONCLUSION. MDCT angiography is a reliable alternative to conventional coronary sinus angiography for detailed anatomic display of the coronary veins. n addition to coronary arteries, the I coronary venous system is a field of growing interest in diagnostic and interventional cardiology. This interest is based on a growing number of diagnostic and therapeutic procedures that include the coronary venous system: cardiac electrophysiological studies with placement of a catheter in the coronary sinus (CS), radiofrequency ablation of posteroseptal or leftsided accessory pathways [1], or retroperfusion therapy in high-risk or complicated coronary angioplasty [2]. Further therapeutic approaches include the use of the coronary venous system as a conduit for bypassing coronary artery stenoses [3] and for local drug and gene delivery [4]. In addition, the coronary venous system is increasingly being used for left ventricular or biventricular pacing in patients with severe heart failure [5]. Because the coronary veins are variable in number, caliber, and course, detailed knowledge of the patient s individual anatomy is essential for optimum planning of the intended procedure. Thus, there is a need for imaging the cardiac venous system. The current gold standard for imaging of the coronary veins is retrograde coronary venography via the coronary sinus. It is frequently performed before placement of left ventricular pacemaker leads or in patients with accessory pathways who are undergoing radiofrequency ablation. Conventional angiography requires central venous access and may be technically challenging because of difficult intubation of the coronary sinus and vigorous backwash of contrast material into the right atrium [5]. Because only the venous system is visualized, conventional angiography is of limited value for some interventional cardiac procedures, for example, using the coronary venous system for conduits to bypass coronary artery stenoses. Before these procedures, an additional catheter-based coronary angiography is needed. Contrast-enhanced MDCT angiography with high spatial and temporal resolution allows for diagnostic evaluation of the coronary arteries [6]. MDCT angiography is less invasive than conventional angiography and has fewer complications; it may therefore be an alternative tool for visualization of the coronary veins. The usefulness of electron beam CT for visualization of the coronary veins has already been shown [7]. To the best of our knowledge, this is the first study comparing MDCT angiography to conventional angiography as the reference standard. Materials and Methods Patients Within 5 months from March 2002 until August 2002, 22 patients (6 men, 16 women; mean age, 62.9 ± 10.6 years) underwent conventional angiography. Seven of these patients also underwent MDCT angiography and were included in the study. The mean time interval between the two examinations was 4.3 ± 3.6 days. In five patients, placement of a left ventricular lead for biventricular pacing was planned; two patients underwent radiofrequency ablation. MDCT angiography was conducted to gain additional information on coronary arteries, cardiac chamber morphology, and cardiac function. Exclusion criteria were renal insufficiency with creatinine 1252 AJR:185, November 2005

2 Imaging the Cardiac Venous System levels higher than 1.3 mg/dl, hyperthyroidism, intolerance to iodinated contrast agents, atrial fibrillation, previous placement of cardiac pacemaker leads, and inability to suspend breathing for about 40 sec. The study was approved by the local institutional review board, and informed consent was obtained from each patient. A C Fig year-old man with left bundle branch block prior to resynchronization therapy. LA = left atrium, LV = left ventricle, RA = right atrium, RV = right ventricle, RCA = right coronary artery. A, Example (multiplanar reformation) of measurement of vessel diameter at orifice of coronary sinus (arrow). B D, Maximum intensity projections (MDCT angiography) of different cardiac veins: B, posterior cardiac vein (arrow); C, left marginal vein (arrow); D, small cardiac vein (arrowhead); arrow in D denotes RCA. Coronary Sinus Angiography Retrograde coronary catheter venography was performed via femoral venous access by an experienced cardiologist. Estimated radiation dose was 2 msv, and 50 to 60 ml of iodinated contrast agent was administered. Venograms were obtained in right anterior oblique (60 ) and left anterior oblique (30 ) views. The different image series were stored digitally (ACOM version 3.0, Siemens Medical Solutions) and analyzed at a separate workstation. Vein diameters were measured after catheter-based image calibration (QuantCor, CASS II; Siemens Medical Solutions) by an experienced electrophysiologist. The catheter was always centered and as- B D AJR:185, November

3 Mühlenbruch et al. sessed in the middle of the view in all patients. Thus, variations were minimized. MDCT Angiography All MDCT examinations were performed with a 4-MDCT scanner (Somatom Volume Zoom, Siemens) during a single end-expiratory breath-hold of sec (median, 42.3 sec) in a supine position. A standardized examination protocol with 4 1 mm collimation, 1.5-mm table feed per rotation (normalized pitch: 0.375), and a gantry rotation time of 500 msec was used. Tube voltage was 120 kv with a tube current of 400 mas eff. Contrast material was administered via an 18-gauge needle in the right cubital vein. The scanning delay was determined by injection of a 20-mL test bolus with a flow rate of 2.5 ml/sec and repeated scanning at the level of the aortic root. The time to peak enhancement plus 5 sec was chosen as delay time. For vessel enhancement, 120 ml of nonionic contrast material (Ultravist 370 [iopromide], Schering) was injected at a flow rate of 2.5 ml/sec. Both injections were followed by a 50-mL saline chaser injected at the same flow rate. Average heart rate during the MDCT examination was 73.2 ± 13.3 bpm (range, bpm); no medication was given to slow down the heart rate. Estimated radiation dose using WinDose 2.1 (IMP) [8] was 6.5 ± 1.2 msv for men and 10.3 msv for women. Axial image series were reconstructed at 0 100% of the R-R interval in steps of 5% with an effective slice thickness of 1.25 mm and a reconstruction increment of 0.8 mm. A field of view of mm 2, a matrix of , and a mediumsmooth convolution kernel (B30f) were applied. All image series were transferred to an external workstation (Leonardo, Siemens) and analyzed using the standard software package (Syngo CT, Siemens). From the image series presenting with the least motion artifacts, multiplanar reconstructions (MPR) and maximum intensity projections (MIP) along the course of the coronary veins and 3D volume-rendering technique images (3D-VRT) were obtained. Image analysis was performed by an experienced radiologist using all of the described images and blinded to the results of conventional angiography. Thus, the final MDCT angiography interpretation of the coronary veins represents an integrated opinion based on all image-display techniques mentioned above. In both techniques (conventional angiography and MDCT angiography), visibility of the coronary veins was graded visually using a 4- point scale for each coronary vein (grade 0, not visible; grade 1, visible with discontinuity; grade 2, visible with irregular borders; grade 3, visible as smoothly bordered vascular structure). The diameter of each of these coronary veins was measured at its orifice (Fig. 1). Statistical Analysis Statistical analysis was performed using SAS System for Windows (Microsoft), release 8.02 TS level 02M0 (SAS Institute). Venous visibility is given as median and range. Statistical analysis for visibility comparing conventional angiography and MDCT angiography was performed using Pearson s correlation coefficient. Values for vessel diameter are given as mean ± SD. An equivalence test for conventional angiography and MDCT angiography was applied and a difference of 0.5 mm was considered acceptable. Bland-Altman plots for method comparison were calculated. Results In all patients, conventional angiography and MDCT angiography were successfully completed without the occurrence of any complications. Comparing both imaging techniques, the following coronary veins were visualized (conventional angiography/mdct angiography) (Fig. 1): coronary sinus (7/7); great cardiac vein (7/7); middle cardiac vein (6/7); running in the posterior interventricular groove, left posterior vein (7/7); draining the inferior posterior wall of the left ventricle, left marginal vein (6/6); running along the free wall of the left ventricle, the anterior interventricular vein (7/7); and near the anterior descending artery and small cardiac veins (2/3) with direct drainage into the right atrium. For both techniques, the median coronary vein visibility was 2 (conventional angiography, mean 2.38 ± 0.70; MDCT angiography, 2.33 ± 0.67). Results are summarized in detail in Table 1 and Figure 2. With decreasing size and peripheral course of the vessel, the visibility of the cardiac veins decreases. Especially for visualization of the small cardiac veins draining directly into the right atrium, MDCT angiography was superior to conventional angiography. TABLE 1: Distribution of Visibility Grades of the Coronary Veins Coronary Sinus Angiography MDCT Angiography Visibility Grade (%) Visibility Grade (%) Target Visualized Mean Mean CS GCV MCV LPV LMV AIVV SCV Note Data are percentages of veins rated at each grade. CS = coronary sinus, GCV = great cardiac vein, MCV = middle cardiac vein, LPV = left posterior vein, LMV = left marginal vein, AIVV = anterior interventricular vein, SCV = small cardiac vein. Visibility Grade CS GCV MCV LPV LMV AIVV SCV Fig. 2 Average visibility grades of all cardiac veins. White bars indicate coronary sinus angiography, gray bars indicate MDCT angiography. CS = coronary sinus, GCV = great cardiac vein, MCV = middle cardiac vein, LPV = left posterior vein, LMV = left marginal vein, AIVV = anterior interventricular vein, SCV = small cardiac vein AJR:185, November 2005

4 Imaging the Cardiac Venous System In selective retrograde conventional angiography, vessel evaluation was sometimes hindered by insufficient contrast of the cardiac veins. Mainly the middle cardiac vein, with its orifice into the coronary sinus very close to the right atrium, was affected. Because of its anatomy, only a small amount of contrast material A C Fig year-old man with cardiac arrhythmias. RA = right atrium, RV = right ventricle, LV = left ventricle, CS = coronary sinus, GCV = great cardiac vein, MCV = middle cardiac vein. A D, Samples of different image displays in MDCT angiography compared with conventional angiography: A, axial image; B, multiplanar reformation; C, maximum intensity projection; and D, 3D volume rendering display. (Fig. 3 continues on next page) is delivered into the middle cardiac vein in retrograde perfusion angiography with selective intubation of the coronary sinus. In MDCT angiography, the visually determined optimum time of the R-R interval for image reconstruction for evaluation of the coronary venous system with fewest motion artifacts was 65% (range, 60 80%). Three-dimensional VRT images proved helpful for global orientation and vessel identification. However, detailed information on the venous anatomy and diameter measurements were provided by the axial slices, individually adapted MPRs, and MIPs (Fig. 3). B D AJR:185, November

5 Mühlenbruch et al. Diameter (mm) Pearson s correlation coefficient for vessel visibility was r = For evaluation of vessel diameter before measurement, an aberrance of 0.5 mm was considered acceptable for comparison of conventional angiography and MDCT angiography values. The average vessel diameter on conventional angiography was 3.98 ± 1.82 mm, and 4.18 ± 1.84 mm on MDCT angiography (Fig. 4). Equivalence of vessel diameter comparing conventional angiography and MDCT angiography could be shown with a 95% confidence interval inside the tolerated difference. A Bland-Altman plot of the paired vessel diameters is shown in Figure 5. Besides evaluation of the coronary venous system, MDCT angiography with the contrast bolus optimized for the coronary arteries also E Fig. 3 (continued) 52- year-old man with cardiac arrhythmias. RA = right atrium, RV = right ventricle, LV = left ventricle, CS = coronary sinus, GCV = great cardiac vein, MCV = middle cardiac vein. E, Conventional angiography, view: right anterior oblique 60. CS GCV MCV LPV LMV AIVV SCV Fig. 4 Average cardiac vein diameter comparing conventional angiography and MDCT angiography values. White bars indicate coronary sinus angiography, gray bars indicate MDCT angiography. CS = coronary sinus, GCV = great cardiac vein, MCV = middle cardiac vein, LPV = left posterior vein, LMV = left marginal vein, AIVV = anterior interventricular vein, SCV = small cardiac vein. allows for detailed analysis of the coronary arteries and the cardiac chambers. Overall, three high-grade stenoses were detected. In two patients, intact coronary bypass grafts were visualized; two patients showed impaired left ventricular function with wall thinning; and in two other patients, coronary artery disease could be excluded. Discussion The usefulness of electron beam CT for visualization of the coronary veins has already been shown [7]. In previous reports, excellent sensitivity and specificity of MDCT coronary angiography have been shown in detecting coronary artery stenoses or coronary bypass graft patency [6]. However, to the best of our knowledge there are no data re- garding the feasibility of dedicated examination of the coronary venous system using MDCT angiography in comparison with conventional angiography as the reference standard. Cardiac resynchronization therapy (CRT) acutely improves hemodynamics in patients with systolic heart failure and ventricular conduction delay [9]. It has been shown to reduce symptoms, increase exercise tolerance, and reduce hospitalization in such patients [10]. The implantation of a CRT pacemaker or defibrillator usually requires left ventricular stimulation, achieved either by direct epicardial lead placement via a limited lateral thoracotomy or, mainly, by a transvenous approach using the tributaries of the coronary sinus. Thus, the availability and location of cardiac veins for potential placement of a left ventricular lead should be assessed before implantation of a CRT device. Because the posterolateral left ventricular wall is usually the optimal location for pacing [11], attention is focused on the presence and availability of coronary veins on the posterolateral aspect of the left ventricle. Depending on the diameter and angle with the great cardiac vein, cannulating a specific left ventricular vein may be difficult. Thus, prior knowledge of venous anatomy may aid in anticipating the degree of difficulty and in selecting special equipment (e.g., steerable pacing guidewires) [7]. Furthermore, vessel size is an important criterion for selecting the target vein in procedures such as left ventricular pacemaker lead insertion. The results of this study prove that evaluation of the cardiac venous system is feasible using MDCT angiography, even with a scanning protocol optimized for the visualization of the coronary arteries. A good correlation of vessel visibility and diameter was shown. Remaining aberrances may have different reasons, including differences in the ways images display in conventional angiography and MDCT angiography. Vessel diameter in conventional angiography can only be assessed in a 2D projection. Although different imaging planes of the same vessel are addressed, the reviewer needs to be aware of the tendency of this imaging technique to underestimate the maximum vessel diameter [12]. In contrast, using MPRs of MDCT angiography with the image plane orthographic to the vessel course, the maximum vessel diameter can easily be measured. MDCT angiography may suffer from motion artifacts, although, in 1256 AJR:185, November 2005

6 Imaging the Cardiac Venous System Coronary Sinus Angiography MDCT Angiography (mm) SD 1.24 Mean SD Vessel Diameter Using Average of Combined Coronary Sinus Angiography and MDCT Angiography (mm) Fig. 5 The Bland-Altman analysis of the paired vessel diameters shows a mean deviation of 0.2 mm with lower and upper limits of agreement (dotted lines) of 1.63 and 1.24 mm (±1.96 SD), respectively. our study, image analysis was performed at the point of the R-R interval presenting with the fewest artifacts and substantial coronary vein filling. Systematic overestimation of the vessel diameter by use of MDCT angiography may be explained in that the vessel wall of the coronary vein is supposed to enhance with the use of contrast material and this was taken into account during the measurements, whereas conventional angiography only depicts the true vessel lumen. However, in conventional angiography, this vessel lumen might be dilated and its diameter overestimated by retrograde injection of contrast material. Compared with conventional angiography, MDCT angiography is less invasive and has a lower rate of complications. Besides increased acceptance of the procedure by the patient population, reduced hospital stays and cost-reduction may be expected. Cardiac MDCT angiography provides additional information on cardiac function and thoracic abnormalities and may replace other invasive procedures, in some cases even catheter-based coronary angiography. Technical improvements, including the introduction of new MDCT scanners with higher spatial and temporal resolution exceeding the temporal resolution of electron beam CT scanners, will lead to even better visualization of the coronary veins. In conclusion, MDCT angiography offers a reliable tool for evaluation of the coronary veins before interventional electrophysiological procedures, pacemaker lead placement, and other procedures and should be considered as an alternative diagnostic technique to invasive conventional angiography. References 1. Wen MS, Yeh SJ, Wang CC, King A, Lin FC, Wu D. Radiofrequency ablation therapy of the posteroseptal accessory pathway. Am Heart J 1996; 132: Kar S, Nordlander R. Coronary veins: an alternate route to ischemic myocardium. Heart Lung 1992; 21: Oesterle SN, Reifart N, Hayase M, et al. Catheterbased coronary bypass: a development update. Catheter Cardiovasc Interv 2003; 58: Ryden L, Tadokoro H, Sjoquist PO, et al. Pharmacokinetic analysis of coronary venous retroinfusion: a comparison with anterograde coronary artery drug administration using metoprolol as a tracer. J Am Coll Cardiol 1991; 18: Gerber TC, Kantor B, Keelan PC, Hayes DL, Schwartz RS, Holmes DR. The coronary venous system: an alternate portal to the myocardium for diagnostic and therapeutic procedures in invasive cardiology. Curr Interv Cardiol Rep 2000; 2: Giesler T, Baum U, Ropers D, et al. Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR 2002; 179: Gerber TC, Sheedy PF, Bell MR, et al. Evaluation of the coronary venous system using electron beam computed tomography. Int J Card Imaging 2001; 17: Kalender WA, Schmidt B, Zankl M, Schmidt M. A PC program for estimating organ dose and effective dose values in computed tomography. Eur Radiol 1999; 9: Auricchio A, Stellbrink C, Block M, et al. Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. Circulation 1999; 99: Cazeau S, Leclercq C, Lavergne T, et al. Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay: Multisite Stimulation in Cardiomyopathies (MUSTIC) Study Investigators. N Engl J Med 2001; 12: Koos R, Sinha AM, Markus K, et al. Comparison of left ventricular lead placement via the coronary venous approach versus lateral thoracotomy in patients receiving cardiac resynchronization therapy. Am J Cardiol 2004; 94: Chen SJ, Carroll JD. 3-D reconstruction of coronary arterial tree to optimize angiographic visualization. IEEE Trans Med Imaging 2000; 19: AJR:185, November

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Noninvasive Visualization of the Cardiac Venous System Using Multislice Computed Tomography

Noninvasive Visualization of the Cardiac Venous System Using Multislice Computed Tomography Journal of the American College of Cardiology Vol. 45, No. 5, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2004.11.035

More information

Imaging congestive heart failure: role of coronary computed tomography angiography (CCTA)

Imaging congestive heart failure: role of coronary computed tomography angiography (CCTA) Imaging congestive heart failure: role of coronary computed tomography angiography (CCTA) Gianluca Pontone, MD, PhD, FESC, FSCCT Director of MR Unit Deputy Director of Cardiovascul CT Unit Clinical Cardiology

More information

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population Journal of the American College of Cardiology Vol. 49, No. 9, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.10.066

More information

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING Non-invasive coronary angiography along with multidetector computed tomography or magnetic resonance imaging is attracting increasing interest

More information

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie 268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre Pushing the limits of cardiac CT Steven Dymarkowski Radiology / Medical Imaging Research Centre 5 X 2013 Introduction Rapid technological advances and new clinical applications in cardiovascular imaging

More information

An Unreported Type of Coronary Artery Anomaly in Congenitally Corrected Transposition of Great Arteries 선천성수정대혈관전위환자에서새롭게보고된관상동맥변이

An Unreported Type of Coronary Artery Anomaly in Congenitally Corrected Transposition of Great Arteries 선천성수정대혈관전위환자에서새롭게보고된관상동맥변이 Case Report pissn 1738-2637 / eissn 2288-2928 J Korean Soc Radiol 2016;75(1):62-67 http://dx.doi.org/10.3348/jksr.2016.75.1.62 An Unreported Type of Coronary Artery Anomaly in Congenitally Corrected Transposition

More information

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing Cademartiri et al. Heart Rhythm Irregularities on MDCT Angiography Cardiac Imaging Original Research A C M E D E N T U R I C A L I M A G I N G AJR 2006; 186:634 638 0361 803X/06/1863 634 American Roentgen

More information

2004;77:800 4 MSCT OF CORONARY ARTERY BYPASS GRAFTS. Results. CABG With Adequate Diagnostic Quality

2004;77:800 4 MSCT OF CORONARY ARTERY BYPASS GRAFTS. Results. CABG With Adequate Diagnostic Quality Isotropic Half-Millimeter Angiography of Coronary Artery Bypass Grafts With 16-Slice Computed Tomography Marc Dewey, MD, Alexander Lembcke, MD, Christian Enzweiler, MD, Bernd Hamm, MD, and Patrik Rogalla,

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats

Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats Amolak Singh 1*, Yash Sethi 1, Sonya Watkins 1, Angela Youtsey 1, Angie Thomas 1 1. Department of Radiology, University

More information

A rare case: Coronary sinus thrombosis

A rare case: Coronary sinus thrombosis A rare case: Coronary sinus thrombosis Poster No.: P-0085 Congress: ESTI 2014 Type: Educational Poster Authors: B. Özkul, N. Inan, Ö. Özkul, H. T. Sarisoy, G. Akansel, A. Akça, #. Çam; Kocaeli/TR Keywords:

More information

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease Horizon Scanning Technology Summary National Horizon Scanning Centre Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease April 2007 This technology summary is based

More information

What every radiologist should know about cardiac CT: A case-based pictorial review

What every radiologist should know about cardiac CT: A case-based pictorial review What every radiologist should know about cardiac CT: A case-based pictorial review Poster No.: C-0555 Congress: ECR 2010 Type: Educational Exhibit Topic: Cardiac Authors: C. M. Capuñay, P. Carrascosa,

More information

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography Journal of the American College of Cardiology Vol. 46, No. 3, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.05.056

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

Variation in Cardiac Vein System is Associated with Coronary Artery Calcium A Venous-Atherosclerosis Paradox?

Variation in Cardiac Vein System is Associated with Coronary Artery Calcium A Venous-Atherosclerosis Paradox? Original Article Acta Cardiol Sin 2015;31:536 542 doi: 10.6515/ACS20150204B Atherosclerosis Variation in Cardiac Vein System is Associated with Coronary Artery Calcium A Venous-Atherosclerosis Paradox?

More information

The Management of Heart Failure after Biventricular Pacing

The Management of Heart Failure after Biventricular Pacing The Management of Heart Failure after Biventricular Pacing Juan M. Aranda, Jr., MD University of Florida College of Medicine, Division of Cardiovascular Medicine, Gainesville, Florida Approximately 271,000

More information

Dr Felix Keng. Imaging of the heart is technically difficult because: Role of Cardiac MSCT. Current: Cardiac Motion Respiratory Motion

Dr Felix Keng. Imaging of the heart is technically difficult because: Role of Cardiac MSCT. Current: Cardiac Motion Respiratory Motion Siemens Philips Dr Felix Keng GE Toshiba Role of Cardiac MSCT Current: Structural / congenital heart imaging Extra-cardiac / Great vessel imaging Volumes and ejection fractions (cine + gating) Calcium

More information

Renal vascular evaluation with 64 Multislice Computerized Tomography Daniela Stoisa, Fabrizzio E. Galiano, Andrés Quaranta, Roberto L.

Renal vascular evaluation with 64 Multislice Computerized Tomography Daniela Stoisa, Fabrizzio E. Galiano, Andrés Quaranta, Roberto L. Renal vascular evaluation with 64 Multislice Computerized Tomography Daniela Stoisa, Fabrizzio E. Galiano, Andrés Quaranta, Roberto L. Villavicencio Footnote Diagnóstico Médico Oroño. Bv. Oroño 1515. 2000.

More information

General Cardiovascular Magnetic Resonance Imaging

General Cardiovascular Magnetic Resonance Imaging 2 General Cardiovascular Magnetic Resonance Imaging 19 Peter G. Danias, Cardiovascular MRI: 150 Multiple-Choice Questions and Answers Humana Press 2008 20 Cardiovascular MRI: 150 Multiple-Choice Questions

More information

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Original Article Open Access The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Ziqiao Lei 1, Jin Gu 2, Qing Fu 3, Heshui Shi 4, Haibo Xu 5, Ping Han 6, Jianming

More information

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Cardiac CT David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Courtesy Tom Kracus Courtesy Kettering Tom Medical Kracus Cente Kettering Medical Center 2003-2006, David Stultz, MD Courtesy Tom Kracus

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience

Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience Ximing Wang, M.D., Zhaoping Cheng, M.D., Dawei Wu, M.D., Lebin

More information

Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease

Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease Hendrik Treede, MD, Christoph Becker, MD, Hermann Reichenspurner,

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

Cardiac Imaging in abnormal rhythm Role of MDCT

Cardiac Imaging in abnormal rhythm Role of MDCT Cardiac Imaging in abnormal rhythm Role of MDCT Cardiac Imaging in abnormal rhythm Role of MDCT Scope of the problem CT in Atrial Fibrillation CT and pacing Ventricular arrhythmia Other applications 1

More information

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital Coronary Artery Imaging Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital Larger array : cover scan area Detector size : spatial resolution Rotation speed : scan time Retrospective

More information

Noncoronary Cardiac MDCT

Noncoronary Cardiac MDCT Noncoronary Cardiac MDCT David A. Bluemke, M.D., Ph.D. Professor, of Radiology and Medicine Johns Hopkins University School of Medicine Baltimore, Maryland Toshiba Disclosures Grant support Noncoronary

More information

Case 47 Clinical Presentation

Case 47 Clinical Presentation 93 Case 47 C Clinical Presentation 45-year-old man presents with chest pain and new onset of a murmur. Echocardiography shows severe aortic insufficiency. 94 RadCases Cardiac Imaging Imaging Findings C

More information

Biventricular Pacing - Hemodynamic Benefit for Patients with Congestive Heart Failure

Biventricular Pacing - Hemodynamic Benefit for Patients with Congestive Heart Failure 428 December 2000 Biventricular Pacing - Hemodynamic Benefit for Patients with Congestive Heart Failure K. MALINOWSKI Helios Clinics, Aue, Germany Summary Congestive heart failure afflicts a large and

More information

b. To facilitate the management decision of a patient with an equivocal stress test.

b. To facilitate the management decision of a patient with an equivocal stress test. National Imaging Associates, Inc. Clinical guidelines EBCT HEART CT & HEART CT CONGENITAL CCTA CPT4 Codes: 75571 EBCT 75572, 75573 Heart CT & Heart CT Congenital 75574 - CCTA LCD ID Number: L33559 J K

More information

CT angiography techniques. Boot camp

CT angiography techniques. Boot camp CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary

More information

Cardiac MRI in ACHD What We. ACHD Patients

Cardiac MRI in ACHD What We. ACHD Patients Cardiac MRI in ACHD What We Have Learned to Apply to ACHD Patients Faris Al Mousily, MBChB, FAAC, FACC Consultant, Pediatric Cardiology, KFSH&RC/Jeddah Adjunct Faculty, Division of Pediatric Cardiology

More information

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor Cardiothoracic Radiology Disclosure I have no disclosure pertinent to this presentation.

More information

THE ROLE OF HIGH END MULTI DETECTOR CT IN CORONARY IMAGING ESSAY

THE ROLE OF HIGH END MULTI DETECTOR CT IN CORONARY IMAGING ESSAY THE ROLE OF HIGH END MULTI DETECTOR CT IN CORONARY IMAGING ESSAY Submitted for partial fulfillment of Master degree in Radiodiagnosis By Ahmed Yehia Ahmed (M.B.B.Ch., Cairo University) Supervisors Prof.

More information

Biventricular Pacing: A Promising Therapeutic Alternative for Patients with Severe Congestive Heart Failure

Biventricular Pacing: A Promising Therapeutic Alternative for Patients with Severe Congestive Heart Failure September 2001 353 Biventricular Pacing: A Promising Therapeutic Alternative for Patients with Severe Congestive Heart Failure T. SZILI-TOROK, D. THEUNS, P. KLOOTWIJK, M.F. SCHOLTEN, G.P. KIMMAN, L.J.

More information

Vascular CT Protocols

Vascular CT Protocols Vascular CT Protocols V 1D: Chest and abdominal CT angiogram (aortic dissection protocol) V 1T: Chest CT angiogram (aortic trauma protocol) V 2: Abdominal and pelvis CT angiogram (aortic aneurysm protocol)

More information

Multidetector computed tomography in the evaluation of atrial septal defects

Multidetector computed tomography in the evaluation of atrial septal defects Multidetector computed tomography in the evaluation of atrial septal defects Poster No.: C-0502 Congress: ECR 2010 Type: Educational Exhibit Topic: Cardiac Authors: S. Espejo, R. Ysamat, B. Cajal, M. Pan,

More information

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT Eur Radiol (2006) 16: 1964 1972 DOI 10.1007/s00330-006-0262-x CARDIAC Sebastian Leschka Lars Husmann Lotus M. Desbiolles Oliver Gaemperli Tiziano Schepis Pascal Koepfli Thomas Boehm Borut Marincek Philipp

More information

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT Ralf Koos 1 Andreas Horst Mahnken 2 Anil Martin Sinha 1 Joachim Ernst Wildberger 2 Rainer Hoffmann 1 Harald Peter Kühl 1 Received March 3, 2004; accepted after revision May 18, 2004. 1 Department of Cardiology,

More information

2010 Canadian Cardiovascular Society/ Canadian Heart Rhythm Society Training and Maintenance of Competency in Adult Clinical Cardiac

2010 Canadian Cardiovascular Society/ Canadian Heart Rhythm Society Training and Maintenance of Competency in Adult Clinical Cardiac 2010 Canadian Cardiovascular Society/ Canadian Heart Rhythm Society Training and Maintenance of Competency in Adult Clinical Cardiac Electrophysiology Martin S. Green, Chair, CHRS Education Committee Peter

More information

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation Poster No.: C-2014 Congress: ECR 2014 Type: Educational Exhibit Authors: M. A. Ottesen; Oslo/NO Keywords: Cardiac, Arteries / Aorta, CT, CT-Angiography,

More information

National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION. Original Date: October 2015 Page 1 of 5

National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION. Original Date: October 2015 Page 1 of 5 National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION CPT Codes: 93451, 93452, 93453, 93454, 93455, 93456, 93457, 93458, 93459, 93460, 93461 LCD ID Number:

More information

Index. radiologic.theclinics.com. Note: Page numbers of article titles are in boldface type.

Index. radiologic.theclinics.com. Note: Page numbers of article titles are in boldface type. Index Note: Page numbers of article titles are in boldface type. A ALCAPA. See Anomalous left coronary artery from the pulmonary artery. Angiosarcoma computed tomographic assessment of, 809 811 Anomalous

More information

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial Poster No.: C-0595 Congress: ECR 2013 Type: Authors:

More information

X-ray coronary angiography is considered the diagnostic

X-ray coronary angiography is considered the diagnostic Noninvasive Coronary Angiography by Retrospectively ECG-Gated Multislice Spiral CT Stephan Achenbach, MD; Stefan Ulzheimer, MS; Ulrich Baum, MD; Marc Kachelrieß, PhD; Dieter Ropers, MD; Tom Giesler, MD;

More information

Blood supply of the Heart & Conduction System. Dr. Nabil Khouri

Blood supply of the Heart & Conduction System. Dr. Nabil Khouri Blood supply of the Heart & Conduction System Dr. Nabil Khouri Arterial supply of Heart Right coronary artery Left coronary artery 3 Introduction: Coronary arteries - VASAVASORUM arising from aortic sinuses

More information

IMAGES. in PAEDIATRIC CARDIOLOGY

IMAGES. in PAEDIATRIC CARDIOLOGY IMAGES in PAEDIATRIC CARDIOLOGY Images Paediatr Cardiol. 2005 Jan-Mar; 7(1): 12 17. PMCID: PMC3232568 Stent implantation for coarctation facilitated by the anterograde trans-septal approach N Sreeram and

More information

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Int J Cardiovasc Imaging (2008) 24:547 556 DOI 10.1007/s10554-007-9282-0 ORIGINAL PAPER Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Jaap M. Groen Æ Marcel J.

More information

Case Report Preoperative Assessment of Anomalous Right Coronary Artery Arising from the Main Pulmonary Artery

Case Report Preoperative Assessment of Anomalous Right Coronary Artery Arising from the Main Pulmonary Artery Case Reports in Medicine Volume 2011, Article ID 642126, 4 pages doi:10.1155/2011/642126 Case Report Preoperative Assessment of Anomalous Right Coronary Artery Arising from the Main Pulmonary Artery Marshall

More information

Chapter. Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography

Chapter. Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography Chapter 3 Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography Joanne D. Schuijf, Jeroen J. Bax, Liesbeth P. Salm, J. Wouter Jukema, Hildo J. Lamb,

More information

CRT-D or CRT-P: HOW TO CHOOSE THE RIGHT PATIENT?

CRT-D or CRT-P: HOW TO CHOOSE THE RIGHT PATIENT? CRT-D or CRT-P: HOW TO CHOOSE THE RIGHT PATIENT? Alessandro Lipari, MD Chair and Department of Cardiology University of Study and Spedali Civili Brescia -Italy The birth of CRT in Europe, 20 years ago

More information

Cardiac Imaging Tests

Cardiac Imaging Tests Cardiac Imaging Tests http://www.medpagetoday.com/upload/2010/11/15/23347.jpg Standard imaging tests include echocardiography, chest x-ray, CT, MRI, and various radionuclide techniques. Standard CT and

More information

Ch.15 Cardiovascular System Pgs {15-12} {15-13}

Ch.15 Cardiovascular System Pgs {15-12} {15-13} Ch.15 Cardiovascular System Pgs {15-12} {15-13} E. Skeleton of the Heart 1. The skeleton of the heart is composed of rings of dense connective tissue and other masses of connective tissue in the interventricular

More information

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications SYMPOSIA Indications, Patient Selection, and Clinical Implications Christian Thilo, MD,* Mark Auler, MD,* Peter Zwerner, MD,w Philip Costello, MD,* and U. Joseph Schoepf, MD* Abstract: Recent technical

More information

Disclosure Information

Disclosure Information Coronary CTA Pearls and Pitfalls Ricardo C. Cury, MD, FSCCT, FAHA, FACC Chairman of Radiology Radiology Associates of South Florida Director of Cardiac Imaging Miami Cardiac and Vascular Institute Past-President

More information

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3 Coronary Artery Bypass Graft Imaging with CT Angiography and Iterative Reconstruction: Quantitave Evaluation of Radiation Dose Reduction and Image Quality Simon Nepveu 1, Irina Boldeanu 1, Yves Provost

More information

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology,

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology, Chapter 4 Impact of Coronary Calcium Score on Diagnostic Accuracy of Multislice Computed Tomography Coronary Angiography for Detection of Coronary Artery Disease Gabija Pundziute, 1,3 Joanne D. Schuijf,

More information

Multisclice CT in combination with functional imaging for CAD. Temporal Resolution. Spatial Resolution. Temporal resolution = ½ of the rotation time

Multisclice CT in combination with functional imaging for CAD. Temporal Resolution. Spatial Resolution. Temporal resolution = ½ of the rotation time Multisclice CT in combination with functional imaging for CAD Prof. Juhani Knuuti, MD, FESC Turku University Hospital and University of Turku Turku, Finland MSCT and functional imaging for CAD Practical

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]

More information

Computed Tomography of the Coronary Arteries

Computed Tomography of the Coronary Arteries Cardiology Update DAVOS 2011 Computed Tomography of the Coronary Arteries Anders Persson M.D., Ph.D Director, Assoc. Professor Center for Medical Image Science and Visualization Linköping University SWEDEN

More information

CRT Implantation Techniques 부천세종병원순환기내과박상원

CRT Implantation Techniques 부천세종병원순환기내과박상원 Cardiac Venous System and CRT Implantation Techniques 부천세종병원순환기내과박상원 Cardiac Resynchronization Therapy (CRT) Goal: Atrial synchronous biventricular pacing Transvenous approach for left ventricular lead

More information

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03 Introduction Cardiac Imaging Modalities Arnaud Bistoquet 12/19/03 Coronary heart disease: the vessels that supply oxygen-carrying blood to the heart, become narrowed and unable to carry a normal amount

More information

Pacing in patients with congenital heart disease: part 1

Pacing in patients with congenital heart disease: part 1 Pacing in patients with congenital heart disease: part 1 September 2013 Br J Cardiol 2013;20:117 20 doi: 10.5837/bjc/2013.028 Authors: Khaled Albouaini, Archana Rao, David Ramsdale View details Only a

More information

Utility of CT angiography for pre-operative evaluation of robotic-assisted minimally invasive mitral valve surgery.

Utility of CT angiography for pre-operative evaluation of robotic-assisted minimally invasive mitral valve surgery. Utility of CT angiography for pre-operative evaluation of robotic-assisted minimally invasive mitral valve surgery. Poster No.: C-2214 Congress: ECR 2014 Type: Educational Exhibit Authors: M. Muthuvelu,

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 5: , 2013

EXPERIMENTAL AND THERAPEUTIC MEDICINE 5: , 2013 636 Adaptive prospective ECG-triggered sequence coronary angiography in dual-source CT without heart rate control: Image quality and diagnostic performance CHANG-JIE PAN, NONG QIAN, TAO WANG, XIAO-QIANG

More information

Review of Cardiac Imaging Modalities in the Renal Patient. George Youssef

Review of Cardiac Imaging Modalities in the Renal Patient. George Youssef Review of Cardiac Imaging Modalities in the Renal Patient George Youssef ECHO Left ventricular hypertrophy (LVH) assessment Diastolic dysfunction Stress ECHO Cardiac CT angiography Echocardiography - positives

More information

Angio-CT: heart and coronary arteries

Angio-CT: heart and coronary arteries European Journal of Radiology 45 (2003) S32/S36 www.elsevier.com/locate/ejrad Angio-CT: heart and coronary arteries Andreas F. Kopp * Tübingen University Hospital, Tübingen, Germany Received 22 November

More information

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011 Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011 Outline Background Hypothesis Study Population Methodology

More information

Coronary artery disease remains

Coronary artery disease remains CT scanning of the coronary arteries: How to do it and how to interpret it Leo P. Lawler, MD, FRCR Coronary artery disease remains one of the leading killers in the western world. Given that many of those

More information

Indications and Technique

Indications and Technique Robotic LV Epicardial Lead Placement: Indications and Technique Sandhya K. Balaram, M.D., Ph.D. Division of Cardiothoracic Surgery St. Luke s-roosevelt Hospital Center Assistant Professor of Clinical Surgery

More information

CORONARY ANATOMY WITH MULTIDETECTOR COMPUTED TOMOGRAPHY ANGIOGRAPHY*

CORONARY ANATOMY WITH MULTIDETECTOR COMPUTED TOMOGRAPHY ANGIOGRAPHY* Updating CORONRY NTOMY WITH MULTIDETECTOR COMPUTED TOMOGRPHY NGIOGRPHY* Joalbo Matos ndrade 1 * Study developed at Instituto do Coração, Fundação Zerbini, rasília, DF, razil. 1. MD, Radiologist at Departament

More information

Interpreting CT Angiography: Three-Dimensional Reconstruction Techniques

Interpreting CT Angiography: Three-Dimensional Reconstruction Techniques 4 Interpreting CT Angiography: Three-Dimensional Reconstruction Techniques Matthew J. Budoff This chapter will take you through the performance and evaluation of the non-invasive coronary angiogram. Using

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015 2384 Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients GUIRU HE, XIAOPEI LIU, YAN LIU, WEI

More information

Advanced imaging of the left atrium - strain, CT, 3D, MRI -

Advanced imaging of the left atrium - strain, CT, 3D, MRI - Advanced imaging of the left atrium - strain, CT, 3D, MRI - Monica Rosca, MD Carol Davila University of Medicine and Pharmacy, Bucharest, Romania Declaration of interest: I have nothing to declare Case

More information

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease Perspectives of new imaging techniques for patients with known or suspected coronary artery disease Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands Correspondence: Jeroen

More information

ADVANCED CARDIOVASCULAR IMAGING. Medical Knowledge. Goals and Objectives PF EF MF LF Aspirational

ADVANCED CARDIOVASCULAR IMAGING. Medical Knowledge. Goals and Objectives PF EF MF LF Aspirational Medical Knowledge Goals and Objectives PF EF MF LF Aspirational Know the basic principles of magnetic resonance imaging (MRI) including the role of the magnetic fields and gradient coil systems, generation

More information

Normal and Variant Coronary Arterial and Venous Anatomy on High-Resolution CT Angiography

Normal and Variant Coronary Arterial and Venous Anatomy on High-Resolution CT Angiography CT ngiography of Coronary rterial and Venous natomy Cardiac Imaging Pictorial Essay Sunil Kini 1, 2 Kostaki G. is 2 Leroy Weaver 2, 3 Kini S, is KG, Weaver L Keywords: anatomy, anomalies, arteriography,

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Coronary Artery Anomalies from Birth to Adulthood; the Role of CT Coronary Angiography in Sudden Cardiac Death Screening

Coronary Artery Anomalies from Birth to Adulthood; the Role of CT Coronary Angiography in Sudden Cardiac Death Screening Coronary Artery Anomalies from Birth to Adulthood; the Role of CT Coronary Angiography in Sudden Cardiac Death Screening E O Dwyer 1, C O Brien 1, B Loo 1, A Snow Hogan 1, O Buckley1 2, B 1. Department

More information

Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure. Coronary Artery Calcium Volume?

Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure. Coronary Artery Calcium Volume? Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure Coronary Artery Calcium Volume? 1 Objective: The purpose of this study is to investigate how accurately we can measure CAC volume

More information

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Mathieu Sabbagh, R3 Michigan State University Radiology Garden City Hospital

More information

PERFORMANCE MEASURE TECHNICAL SPECIFICATIONS. HRS-3: Implantable Cardioverter-Defibrillator (ICD) Complications Rate

PERFORMANCE MEASURE TECHNICAL SPECIFICATIONS. HRS-3: Implantable Cardioverter-Defibrillator (ICD) Complications Rate PERFORMANCE MEASURE TECHNICAL SPECIFICATIONS HRS-3: Implantable Cardioverter-Defibrillator (ICD) Complications Rate Measure Title Description Measure Type Data Source Level of Analysis Numerator HRS-3:

More information

Cigna - Prior Authorization Procedure List Cardiology

Cigna - Prior Authorization Procedure List Cardiology Cigna - Prior Authorization Procedure List Cardiology Category CPT Code CPT Code Description 33206 Insertion of new or replacement of permanent pacemaker with transvenous electrode(s); atrial 33207 Insertion

More information

Coronary angiography is the standard way of visualizing

Coronary angiography is the standard way of visualizing Clinical Investigation and Reports Coronary Artery Fly-Through Using Electron Beam Computed Tomography Peter M.A. van Ooijen, MSc; Matthijs Oudkerk, MD, PhD; Robert J.M. van Geuns, MD; Benno J. Rensing,

More information

Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides

Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides Cardiac Devices CRT,ICD: Who is and is not a Candidate? Who Decides Colette Seifer MB(Hons) FRCP(UK) Associate Professor, University of Manitoba, Cardiologist, Cardiac Sciences Program, St Boniface Hospital

More information

Case Report. Heon Lee, MD, PhD 1, Seok-Yeon Kim, MD 2, U. Joseph Schoepf, MD 3 CASE REPORT INTRODUCTION

Case Report. Heon Lee, MD, PhD 1, Seok-Yeon Kim, MD 2, U. Joseph Schoepf, MD 3 CASE REPORT INTRODUCTION Case Report http://dx.doi.org/10.3348/kjr.2012.13.2.244 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2012;13(2):244-248 Isolated Non-Compaction of the Left Ventricle in a Patient with New-Onset Heart

More information

Studies with electron beam computed tomography (EBCT) Imaging

Studies with electron beam computed tomography (EBCT) Imaging Imaging Predictive Value of 16-Slice Multidetector Spiral Computed Tomography to Detect Significant Obstructive Coronary Artery Disease in Patients at High Risk for Coronary Artery Disease Patient- Versus

More information

Coronary artery bypass grafting has been a historically. Multislice CT Evaluation of Coronary Artery Bypass Graft Patients SYMPOSIA

Coronary artery bypass grafting has been a historically. Multislice CT Evaluation of Coronary Artery Bypass Graft Patients SYMPOSIA SYMPOSIA Multislice CT Evaluation of Coronary Artery Bypass Graft Patients Robert Chapman Gilkeson, MD* and Alan H. Markowitz, MDw Abstract: Continuous improvement in multislice computed tomography technology

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Chapter 5 Section 1.1. Diagnostic Radiology (Diagnostic Imaging)

Chapter 5 Section 1.1. Diagnostic Radiology (Diagnostic Imaging) Radiology Chapter 5 Section 1.1 Issue Date: March 7, 1986 Authority: 32 CFR 199.4(a), (b)(2)(x), (c)(2)(viii), (e)(14) and 32 CFR 199.6(d)(2) 1.0 CPT 1 PROCEDURE CODES 70010-72292, 73000-76499, 77071-77084,

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients

Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients (2006) 30, 569 573 & 2006 Nature Publishing Group All rights reserved 0307-0565/06 $30.00 www.nature.com/ijo ORIGINAL ARTICLE Image quality and diagnostic accuracy of 16-slice multidetector computed tomography

More information

Jae Hoon Lim, M.D., Song Choi, M.D. 2, Yang Jun Kang, M.D. 2, Hyun Ju Seon, M.D., Yun Hyeon Kim, M.D.

Jae Hoon Lim, M.D., Song Choi, M.D. 2, Yang Jun Kang, M.D. 2, Hyun Ju Seon, M.D., Yun Hyeon Kim, M.D. J Korean Soc Radiol 2010;62:113-117 The Noninvasive Diagnosis and Postoperative Evaluation of nomalous Right Coronary rtery from the Pulmonary rtery (RCP) using Coronary MDCT: Case Report 1 Jae Hoon Lim,

More information

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine Non Contrast MRA Mayil Krishnam Director, Cardiovascular and Thoracic Imaging University of California, Irvine No disclosures Non contrast MRA-Why? Limitations of CTA Radiation exposure Iodinated contrast

More information