1 2 Cardiovascular Computed Tomography: Current and Future Scanning System Design Wm. Guy Weigold Introduction The heart can be visualized in gross form on any standard chest computed tomography (CT), but for detailed evaluation of cardiac anatomy exacting specifications of scanner hardware and software must be in place because: the heart is continually in motion, the structures of interest (i.e., coronary arteries) are small, and their curvilinear course requires multiplanar reformatting for analysis. For these specifications to be met, the scanner hardware must be robust, the scanner must be fitted with ECGmonitoring equipment, and cardiac-specific software and image reconstruction servers must be installed. Elements of Scanner Design The general design of a CT scanner equipped for cardiac imaging is the same as that of any modern scanner: an X-ray tube mounted on a rotating gantry opposite a detector array generates X-rays which are emitted across the bore, are attenuated as they pass through the patient s chest, and are scintillated by the detectors as the gantry rotates around the patient obtaining attenuation data from multiple view angles. These data are transmitted to an image reconstruction server which then computes the axial images. To perform cardiac CT, however, a number of additional capabilities are required, [1, 2] namely, high spatial resolution, high temporal resolution, and fast scan speed, while delivering an acceptable radiation exposure. Special Design Considerations for Cardiac CT Spatial Resolution Spatial resolution is the product of multiple variables including detector characteristics and collimation, sampling rate, and reconstruction methods. State of the art scanners use thin mm collimation. This reduces volume averaging thus increasing image detail of small objects, and allows reconstruction of isotropic datasets, thereby preserving z-axis resolution and permitting multiplanar reformatting necessary for analysis of cardiac anatomy. GE Healthcare has focused their system development on spatial resolution, developing a new detector scintillation material with a fast decay time (30 ns), minimal afterglow (one-quarter that of conventional Gd 2 O 2 S crystals), and increased efficiency [3]. This new detector is one element of a multipronged approach to increasing image resolution, which also includes new electronics that permit increased sampling,[4] and a new iterative reconstruction algorithm [5, 6] (Figure 2.1). These innovations increase spatial resolution by line pairs per centimeter (lp/cm), which is substantial given that current resolution of systems is approximately lp/cm. The enhanced resolution is particularly useful for imaging stented and/or calcified coronary arteries. Temporal Resolution In order to reconstruct an axial image, attenuation data from multiple view angles that encompass half a rotation around the patient are required. The amount of time needed to obtain these data is the nominal, heart-rate independent temporal resolution of the scanner. Hence, a gantry that requires 330 ms to make one complete rotation has a baseline temporal resolution of 165 ms. This is analogous to shutter speed: just as shutter speed must be sufficiently fast to capture motion-free images of a moving subject, temporal resolution must be sufficiently fast to capture the data needed for image reconstruction within the relatively brief period of cardiac diastasis. If temporal resolution is insufficient, motion artifact occurs (Figure 2.2). A temporal resolution of 165 ms is relatively slow for cardiac imaging, so systems overcome this by various means: faster gantry rotation, multicycle reconstruction, and a second X-ray tube. 21

2 22 Cardiac CT Imaging: Diagnosis of Cardiovascular Disease Figure 2.1. Increased sampling improves image quality. Axial images of line-pair phantom demonstrating improved spatial resolution using increased sampling density (on the right) compared to standard sampling frequency (on the left). Note the improved edge detail and resolution of the line pairs of the image derived from the high density sampling dataset. (Reproduced with permission from Chandra [4]). Currently, the fastest gantry on the market resides in the Philips ict, which has a rotation time of 270 ms using an innovative frictionless design in which the gantry is actually suspended in air [7]. As this design has only recently been released, it is likely that future iterations will allow even faster rotation. Multicycle reconstruction combines data from multiple contiguous heart beats to complete the half-scan of views (Figure 2.3). This technique is heart-rate dependent: it can only be used within certain ranges of heart rates, since the cardiac cycle and gantry rotation must be asynchronous. Figure 2.2. Better temporal resolution improves image quality of moving structures. 3D volume reconstructions of the right coronary artery in three CT scans representing three generations of CT scanner (4-, 16-, and 64-slice, from left to right, respectively), possessing temporal resolution of 400, 250, and 180 ms from left to right, respectively. Note that the vessel appears distorted and disjointed, due to motion artifact, when temporal resolution is insufficient (far left), while motion artifact is minimally present, and small anatomic details are clearly imaged when temporal resolution is improved (far right). (Reproduced with the kind permission from Springer Science + Business Media from Hurlock et al [2]). An approach adopted by Siemens in 2006 uses two X-ray tubes mounted on the gantry at 90 angles; therefore, only a quarter turn of the gantry is required to collect the 180 of attenuation data (Figure 2.4). Hence, a dual-source system with a gantry rotation time of 330 ms has a baseline central temporal resolution of approximately 85 ms. This obviates the need for multicycle reconstruction, making the system less susceptible to variation in heart rate, and providing high heart rate independent temporal resolution. Initial reports demonstrated very good image quality, [8] and, in a comparison of 64 MDCT and DSCT scans, the dual-source

3 Cardiovascular Computed Tomography: Current and Future Scanning System Design 23 z 180 t z t Figure 2.4. Dual source CT. Schematic diagram of dual-source CT scanner: two X-ray source and corresponding detector arrays are offset by approximately 90. In this way, the half-scan of data required for axial image reconstruction can be acquired over a quarter-rotation of the gantry, thereby halving the nominal heart rate independent temporal resolution of the scan. (Reproduced with permission from the RSNA from McCollough et al [22]). Figure 2.3. Multicycle reconstruction. Single cycle recon (a) the duration of the acquisition window (gray bar) is approximately equivalent to one-half the gantry rotation time, since this is the time required to obtain 180 of attenuation data. Multicycle recon (b) When multiple detector rows are present, the axial slice position in the z-position can be imaged at multiple different times, using multiple, shorter, acquisition windows distributed across multiple contiguous beats. These data can then be combined to provide 180 of attenuation data and used to reconstruct the axial image. Temporal resolution of the image is improved because the duration of the each acquisition window is shorter. (Reproduced with permission from Wolters Kluwer from Vembar et al [1]). system produced better image quality at higher heart rates [9]. Image quality is still improved, however, by reducing heart rate before scanning. Siemens recently announced a new scanner with an even faster (280 ms) gantry which is capable of a central temporal resolution of 75 ms [10]. Detector Coverage and Scan Speed Since the advent of multislice CT, manufacturers have increased z-axis coverage by the addition of more rows of detectors. By increasing coverage, scan times have been reduced to just a few seconds, and even to single-beat scanning. This has been achieved by Toshiba s Aquilion One scanner, which currently has the largest z-axis coverage [11]. This scanner has 320 rows of detectors, with 0.5 mm collimation, 16 cm of coverage, and can perform an axial 1-beat acquisition of the entire heart when the rate is well controlled. The Philips ict is a 128-row scanner with 8 cm of coverage, which allows axial acquisition of the entire heart in 2 beats [7]. GE offers a 64-row scanner with 4 cm of coverage. The current Siemens system is a 32-row scanner, with approximately 2 cm of coverage, but Siemens has recently announced a new 64-row scanner offering approximately 3.8 cm of coverage. Radiation Exposure Traditional cardiac CT uses a helical scan mode and very low pitch to perform retrospective ECG-gated reconstructions. This delivers an undesirably high radiation exposure, so various methods have been designed to reduce radiation exposure. One of the first was ECG-based tube current modulation, which fluctuates tube current in sync with the heart rate, maintaining high tube current during diastasis and providing best image quality in that phase, then lowering tube current during other phases when high-resolution detail is not required (Figure 2.5). This results in a dose savings of approximately 40%. A great advance in dose reduction was made with the advent of prospective, ECG-triggered, axial cardiac CT (Figure 2.6). By keeping the X-ray tube off during most of the scan, and only turning it on during diastasis, as triggered by the ECG, radiation exposure is dramatically reduced (Figure 2.7). GE first published results using this method [12] with good clinical results and radiation doses of 1 2 msv, [13] and this approach is now offered by all manufacturers. Using a 64-slice system, the entire heart can be covered in 3 4 acquisitions; however this leaves the system vulnerable to fluctuations in heart rate and rhythm during the scan. Larger coverage mitigates this vulnerability by reducing the number of beats

4 24 Cardiac CT Imaging: Diagnosis of Cardiovascular Disease Tube output 100% 20% Time Figure 2.5. ECG-based tube current modulation reduces radiation exposure. Once the desired acquisition phase is determined, based on heart rate, in this case 75% phase, scanning proceeds with maintenance of 100% tube output during that phase, while tube current is reduced outside of that phase. This can reduce effective radiation dose by 40% or more; however, note that images reconstructed from the low tube output phases (left image) will be excessively noisy, and are usually considered unusable for coronary interpretation. (Reproduced with permission from Wolters Klower from Vembar et al [1]). Number of Patients PGA RGH Table move Effective Dose (msv) 75% 75% Figure 2.6. Prospective ECG-triggered acquisition method. Cardiac rhythm is monitored while table remains stationary. When cardiac cycle reaches predetermined acquisition phase (in this case 75%, diastolic, phase), X-ray source is briefly turned on (<1 s, blue bars) and acquires attenuation data of a length equivalent to the craniocaudal coverage of the detector array, and is then turned off. The table is advanced almost the length of craniocaudal coverage (minus a small amount of overlap), and the process repeats again. Depending on the craniocaudal coverage of the scanner, the entire heart can be scanned in 1 4 acquisitions. (Reproduced with the kind permission from Springer Science + Business Media from Weigold et al [7]). required for data acquisition. However, larger coverage lends to higher radiation doses, as the detectors are less efficient when having to expose a wider detector array. Hence, Philips 128-row ict system can cover the entire heart in 2 beats, with Figure 2.7. Prospective cardiovascular computer tomographic angiographty (CCTA) reduces radiation dose. Histogram of effective radiation dose, grouped by acquisition method. Using prospectively gated axial scanning, average effective dose is dramatically reduced to 2 3 msv. Even the highest dose PGA scan still delivers a lower effective dose than the lowest-dose RGH scan. PGA prospectively gated axial; RGH retrospectively gated helical. (Reproduced with permission from RSNA from Earls et al [13]). a reported radiation exposure of 3 5 msv [7]. Toshiba s 320-row Aquilion One scanner can cover the entire heart in one beat, with a reported radiation exposure of approximately 6 msv. However, lower radiation doses and dependable coronary scanning using 1-beat acquisition requires heart rate reduction. At higher heart rates (>65 bpm), optimal image quality requires a 2- or 3-beat acquisition (and use of multicycle recon) which increases radiation exposure to approximately 13 (2-beat) to 19 (3-beat) msv [14, 15]. Acquiring with a wide exposure window ( padding ) can ensure the capture

5 Cardiovascular Computed Tomography: Current and Future Scanning System Design 25 a Low heart rate Table move Figure 2.8. Prospective CCTA phase selection and padding. Acquisition phase and padding depend on heart rate: For low heart rates (a), target late diastole (75% phase) for prospective acquisition, but if heart rate is high (b) (>75 bpm), target end-systole (40% phase) which is more likely to produce motion-free images. By restricting acquisition to the minimum exposure window (dark gray), radiation dose is minimized, but ability to reconstruct adjacent cardiac phases is obliterated. Widening the acquisition window ( padding, light gray shading) allows reconstruction of a small number of additional phases, which can help interpretation of motion artifact, but increases the radiation dose. (Reproduced with the kind permission from Springer Science + Business Media from Weigold et al [7]). b 75% 75% High heart rate Table move 40% 40% 83ms Figure 2.9. Method of high-pitch coronary CCTA. Single source helical CT requires a pitch 1.5; a faster pitch results in gaps in the data (left panel). A dual-source system can scan at a higher pitch (up to 3.2), using the second X-ray source-detector system to fill in what would otherwise be gaps in the data. Since the table speed is so high, the entire heart can be imaged in a fraction of a second, from data derived from a single heartbeat (right panel). If the image acquisition is triggered appropriately to synchronize acquisition with diastasis, and the heart rate is sufficiently low, motion-free images can be obtained with a very low radiation dose. (Reproduced with permission from Elsevier from Achenbach et al [18]). of motion-free data, but increases radiation exposure (Figure 2.8). A narrower exposure window can be used with preservation of image quality if heart rate is controlled [16] and reduces radiation exposure. Siemens has recently introduced a new innovative scanning method using a high pitch helical mode which takes advantage of the dual-source design [17, 18] (Figure 2.9). The high pitch ( ) would normally produce gaps in the attenuation data using a single-source system, but, in a dual-source system, these are compensated for by data gathered from the second detector. Scan time is less than 1 s, with a radiation exposure of approximately 2 msv. Initial studies performed on the first-generation dualsource system proved the feasibility of the method, though the authors noted that these first-generation systems are not suitable for this high pitch technique. High pitch scanning with their new scanner, with faster gantry rotation (280 ms) and larger coverage (64 rows), has been reported to produce very good coronary image quality with radiation exposure of approximately 1 msv [10]. Of note, in this initial investigational phase, low heart rates (<60 bpm) are absolutely required, as the entire data collection takes place

6 26 Cardiac CT Imaging: Diagnosis of Cardiovascular Disease within a ms acquisition window of one heart beat; hence diastasis must be at least this long to provide motion free data acquisition, and therefore the requirement for a very low heart rate. The previously described new detector and reconstruction system offered by GE aims to lower radiation exposure by preserving image quality while scanning with less radiation (<1 msv), because images are of higher resolution and lower noise than would otherwise be achievable with a lowexposure scan. energy CT could take the field to a new level of diagnostic power if it can be used for refined tissue characterization, such as differentiating plaque characteristics. This is difficult to do in the coronary arteries, and initial studies have not yielded any breakthroughs, but at the least it has the potential to enhance coronary lumen visualization in heavily calcified vessels by differentiating calcium and iodine [20] (Figure 2.11). Dual-energy CT may also yield new applications for noncoronary imaging, such as imaging of myocardial perfusion [21]. Future Directions Future scanner designs are closely guarded industry secrets, but some concepts have been openly discussed. Flat-panel volume CT systems replace detector rows with a large area detector, and provide coverage of the entire heart in one axial acquisition and extremely high spatial resolution of up to 26 lp/cm, comparable to invasive angiography [19] (Figure 2.10). However, the contrast resolution is inferior to that of multidetector CT, a high radiation exposure is needed to achieve a sufficient contrast-to-noise ratio, and scintillation times are slow and temporal resolution insufficient for cardiac imaging. Dual energy scanning is being developed by all vendors by various means, including rapid fluctuation of tube energy, stacked detectors, or the dual-tube design in which each tube emits photons of different energy levels. Dual Conclusions Since the turn of the twenty-first century, there has been an explosion in technological development of cardiac CT systems, with concomitant gain in reliability and accuracy, especially of coronary imaging. The theme has been one of progressively improved spatial and temporal resolution, reduced scan time, and, most recently, a focus on driving down radiation exposure. In appropriately selected patients, using careful technique, it is easily achievable to perform a cardiac CT using less radiation exposure than that of a standard chest CT or an invasive coronary angiogram. The goal of future systems will be to make this more widely achievable in a larger group of patients without requiring patient selection or stringent patient prep. Given the history of rapid technological advancement, we can expect to see this goal achieved in the near future. X-Ray Tube 57 cm Source Detector Panel Isocenter 93 cm Detector Figure Flat panel CT. Schematic of flat panel scanner: In a flat panel system, the traditional multirow detector array is replaced by an area detector consisting of millions of detector elements, each with dimensions < mm. Large coverage allows whole-organ imaging with tremendous spatial resolution (up to 150 mm). Limiting factors of these prototypes for cardiac imaging have been insufficient temporal resolution, and requirement of high radiation dose. (Reproduced with permission from the RSNA from Gupta et al [19]).

7 Cardiovascular Computed Tomography: Current and Future Scanning System Design 27 Figure Dual energy CT. Dual energy CT can be used to characterize tissue: vessel crosssectional images derived from high- (140 kev) and low-energy (90 kev) attenuation profiles (a, b) demonstrate the influence of photon energy on photoattenuation. The contrast-filled lumen (arrow) exerts greater photoattenuation on lowenergy photons, and hence appears brighter in (b), while calcification (asterisk) appears highdensity in both images. In the subtracted image (c), dense calcification has been removed. By adding in the low-energy (90 kev) attenuation data to this subtracted image, depiction of the lumen edge is enhanced (because of reduced contrast blooming), and visualization of a small side branch (arrowhead) is improved (d). (Reproduced with permission from Wolters Kluwer from Boll et al [23]). References 1. Vembar M, Walker MJ, Johnson PC. Cardiac imaging using multislice computed tomography scanners: technical considerations. Coron Artery Dis. 2006;17: Hurlock GS, Higashino H, Mochizuki T. History of cardiac computed tomography: single to 320-detector row multislice computed tomography [Review] [50 refs]. Int J Cardiovasc Imaging. 2009;25: Jiang H. Gemstone The Ultimate Scintillator for Computed Tomography [White Paper]. Waukesha, WI: GE Healthcare; Chandra N. CT Sampling Technology [White Paper]. Waukesha, WI: GE Healthcare; Thibault JB, Sauer KD, Bouman CA, Hsieh J. A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys. 2007;34: Hsieh J. Adaptive Statistical Iterative Reconstruction [White Paper]. Waukesha, WI: GE Healthcare; Weigold W, Olszewski M, Walker MJ. Low-dose prospectively gated 256-slice coronary computed tomographic angiography. Int J Cardiovasc Imaging. 2009;25(suppl 2): Achenbach S, Ropers D, Kuettner A, et al. Contrast-enhanced coronary artery visualization by dual-source computed tomography initial experience. Eur J Radiol. 2006;57: Achenbach S, Ropers U, Kuettner A, et al. Randomized comparison of 64-slice single- and dual-source computed tomography coronary angiography for the detection of coronary artery disease. JACC Cardiovas Imaging. 2008;1: [see comment]. 10. Lell M, Marwan M, Schepis T, et al. Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual-source CT: technique and initial experience. Eur Radiol. 2009; 19: Mather R. Dynamic Volume CT [White Paper]. Tustin, CA: Toshiba America Medical Systems; Hsieh J, Londt JH, Vass M, Li J, Tang X, Okerlund D. Step and shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys. 2006;33: Earls JP, Berman EL, Urban BA, et al. Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology. 2008;246: Hoe J, Toh KH. First experience with 320-row multidetector CT coronary angiography scanning with prospective electrocardiogram gating to reduce radiation dose. J Cardiovasc Comput Tomogr. 2009;3: Rybicki FJ, Otero HJ, Steigner ML, et al. Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging. 2008;24: [see comment]. 16. Steigner ML, Otero HJ, Cai T, et al. Narrowing the phase window width in prospectively ECG-gated single heart beat 320-detector row coronary CT angiography. Int J Cardiovasc Imaging. 2009;25: Hausleiter J, Bischoff B, Hein F, et al. Feasibility of dual-source cardiac CT angiography with high-pitch scan protocols. J Cardiovasc Comput Tomogr. 2009;3: [see comment]. 18. Achenbach S, Marwan M, Schepis T, et al. High-pitch spiral acquisition: a new scan mode for coronary CT angiography. J Cardiovasc Comput Tomogr. 2009;3: Gupta R, Cheung AC, Bartling SH, et al. Flat-panel volume CT: fundamental principles, technology, and applications. Radiographics. 2008; 28: Barreto M, Schoenhagen P, Nair A, et al. Potential of dual-energy computed tomography to characterize atherosclerotic plaque: ex vivo assessment of human coronary arteries in comparison to histology. J Cardiovasc Comput Tomogr. 2008;2: [see comment]. 21. Ruzsics B, Schwarz F, Schoepf UJ, et al. Comparison of dual-energy computed tomography of the heart with single photon emission computed tomography for assessment of coronary artery stenosis and of the myocardial blood supply. Am J Cardiol. 2009;104: McCollough CH, Primak AN, Saba O, et al. Dose performance of a 64-channel dual-source CT scanner. Radiology. 2007;243: Boll DT, Hoffmann MH, Huber N, Bossert AS, Aschoff AJ, Fleiter TR. Spectral coronary multidetector computed tomography angiography: dual benefit by facilitating plaque characterization and enhancing lumen depiction. J Comput Assist Tomogr. 2006;30(5):

8

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Cardiac CT - Coronary Calcium Basics Workshop II (Basic) Cardiac CT - Coronary Calcium Basics Workshop II (Basic) J. Jeffrey Carr, MD, MSCE Dept. of Radiology & Public Health Sciences Wake Forest University School of Medicine Winston-Salem, NC USA No significant

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography Journal of the American College of Cardiology Vol. 52, No. 18, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.07.048

More information

Cardiopulmonary Imaging Original Research

Cardiopulmonary Imaging Original Research Cardiopulmonary Imaging Original Research Bischoff et al. Versus Helical Mode for Coronary CTA Cardiopulmonary Imaging Original Research Bernhard Bischoff 1 Franziska Hein 1 Tanja Meyer 1 Markus Krebs

More information

CT Coronary Angiography: 256-Slice and 320-Detector Row Scanners

CT Coronary Angiography: 256-Slice and 320-Detector Row Scanners CT Coronary Angiography: 256-Slice and 320-Detector Row Scanners The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters. Citation Published

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Dual Energy CT of the Heart: Perfusion and Beyond

Dual Energy CT of the Heart: Perfusion and Beyond Dual Energy CT of the Heart: Perfusion and Beyond U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant

More information

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing Cademartiri et al. Heart Rhythm Irregularities on MDCT Angiography Cardiac Imaging Original Research A C M E D E N T U R I C A L I M A G I N G AJR 2006; 186:634 638 0361 803X/06/1863 634 American Roentgen

More information

Adapted Transfer Function Design for Coronary Artery Evaluation

Adapted Transfer Function Design for Coronary Artery Evaluation Adapted Transfer Function Design for Coronary Artery Evaluation Sylvia Glaßer 1, Steffen Oeltze 1, Anja Hennemuth 2, Skadi Wilhelmsen 3, Bernhard Preim 1 1 Department of Simulation and Graphics, University

More information

CT Perfusion. U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging

CT Perfusion. U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging CT Perfusion U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant for / research support from Bayer Bracco

More information

The radiation dose in retrospective

The radiation dose in retrospective The radiation dose in retrospective gated tdcoronary computed td tomography (CCT) Saeed AL Ahmari, Ghormallah AL Zahrani, Sumiah AL Helali, Samir AL Dulikan, Abdullah Bafagih, HibaKhashojji Prince Sultan

More information

Clinical Image Gallery Next Generation Volume 1

Clinical Image Gallery Next Generation Volume 1 Clinical Image Gallery Next Generation Volume 1 Dr. Russell Bull Royal Bournemouth Hospital, Bournemouth, United Kingdom After long experience with the first generation, a next generation Aquilion ONE

More information

Computed Tomography Imaging of the Coronary Arteries

Computed Tomography Imaging of the Coronary Arteries Chapter 6 Computed Tomography Imaging of the Coronary Arteries G.J. Pelgrim, M. Oudkerk and R. Vliegenthart Additional information is available at the end of the chapter http://dx.doi.org/10.5772/54044

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population Journal of the American College of Cardiology Vol. 49, No. 9, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.10.066

More information

Aquilion ONE: Pediatric Imaging. Richard Mather, PhD. Senior Manager, CT Clinical Science Toshiba America Medical Systems, Inc.

Aquilion ONE: Pediatric Imaging. Richard Mather, PhD. Senior Manager, CT Clinical Science Toshiba America Medical Systems, Inc. Aquilion ONE: Pediatric Imaging Richard Mather, PhD Senior Manager, CT Clinical Science Toshiba America Medical Systems, Inc. The use of CT in pediatric diagnostic procedures has increased significantly

More information

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located?

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located? Typical PET Image Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located? PET/CT Oncology Imaging Anatometabolic fusion images are useful in the management

More information

The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6

The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6 The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6 n&list_uids=17433949 64-Multislice Detector Computed Tomography Coronary Angiography as Potential Alternative

More information

B. CT protocols for the spine

B. CT protocols for the spine B. CT protocols for the spine Poster No.: A-003 Congress: ECR 2010 Type: Invited Speaker Topic: Neuro Authors: B. Tins; Oswestry/UK Keywords: CT, spine, diagnostic imaging protocol DOI: 10.1594/ecr2010/A-003

More information

Noncoronary Cardiac MDCT

Noncoronary Cardiac MDCT Noncoronary Cardiac MDCT David A. Bluemke, M.D., Ph.D. Professor, of Radiology and Medicine Johns Hopkins University School of Medicine Baltimore, Maryland Toshiba Disclosures Grant support Noncoronary

More information

Soft and Intermediate Plaques in Coronary Arteries: How Accurately Can We Measure CT Attenuation Using 64-MDCT?

Soft and Intermediate Plaques in Coronary Arteries: How Accurately Can We Measure CT Attenuation Using 64-MDCT? 64-MDCT Measurement of Coronary Artery Plaques Cardiac Imaging Original Research Jun Horiguchi 1 Chikako Fujioka 1 Masao Kiguchi 1 Yun Shen 2 Christian E. Althoff 3,4 Hideya Yamamoto 5 Katsuhide Ito 3

More information

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography Supalerk Pattanaprichakul, MD 1, Sutipong Jongjirasiri, MD 2, Sukit Yamwong, MD 1, Jiraporn Laothammatas,

More information

Ultralow Dose Chest CT with MBIR

Ultralow Dose Chest CT with MBIR Ultralow Dose Chest CT with MBIR Ella A. Kazerooni, M.D. Professor & Director Cardiothoracic Radiology Associate Chair for Clinical Affairs University of Michigan Disclosures Consultant: GE Healthcare

More information

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03 Introduction Cardiac Imaging Modalities Arnaud Bistoquet 12/19/03 Coronary heart disease: the vessels that supply oxygen-carrying blood to the heart, become narrowed and unable to carry a normal amount

More information

Disclosure Information

Disclosure Information Coronary CTA Pearls and Pitfalls Ricardo C. Cury, MD, FSCCT, FAHA, FACC Chairman of Radiology Radiology Associates of South Florida Director of Cardiac Imaging Miami Cardiac and Vascular Institute Past-President

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT

SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT Journal of Cardiovascular Computed Tomography (2011) 5, 198 224 Guidelines SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT Sandra S. Halliburton, PhD a, *, Suhny

More information

La Cardio TC nella Malattia Coronarica Risultati del Progetto di Adozione della Regione Emilia-Romagna e Prospettive Future Parma 15 Gennaio 2010

La Cardio TC nella Malattia Coronarica Risultati del Progetto di Adozione della Regione Emilia-Romagna e Prospettive Future Parma 15 Gennaio 2010 20 minuti La Cardio TC nella Malattia Coronarica Risultati del Progetto di Adozione della Regione Emilia-Romagna e Prospettive Future Parma 15 Gennaio 2010 Dr. Filippo Cademartiri,, MD, PhD EMC Rotterdam,

More information

Gender differences in CT calcium scoring: A phantom study

Gender differences in CT calcium scoring: A phantom study Gender differences in CT calcium scoring: A phantom study Nicholas Petrick, Qin Li, Benjamin Berman, Marios A Gavrielides, Rongping Zeng, Berkman Sahiner CDRH/OSEL/DIDSR U.S. Food and Drug Administration

More information

CT angiography techniques. Boot camp

CT angiography techniques. Boot camp CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary

More information

Dose and image quality comparison between prospectively gated axial and retrospectively gated helical coronary CT angiography

Dose and image quality comparison between prospectively gated axial and retrospectively gated helical coronary CT angiography The British Journal of Radiology, 84 (2011), 51 57 Dose and image quality comparison between prospectively gated axial and retrospectively gated helical coronary CT angiography 1 C HLAIHEL, MD, 1 L BOUSSEL,

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Disclosures. GETTING TO THE HEART OF THE MATTER WITH MULTIMODALITY CARDIAC IMAGING Organ Review Meeting 25 September. Overview

Disclosures. GETTING TO THE HEART OF THE MATTER WITH MULTIMODALITY CARDIAC IMAGING Organ Review Meeting 25 September. Overview GETTING TO THE HEART OF THE MATTER WITH MULTIMODALITY CARDIAC IMAGING Organ Review Meeting 25 September Disclosures None relevant to this presentation Mini Pakkal Assistant Professor of Radiology University

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

Integrated PET/CT systems State of the art and Clinical Applications

Integrated PET/CT systems State of the art and Clinical Applications Integrated PET/CT systems State of the art and Clinical Applications V. Bettinardi Nuclear Medicine Dep. Scientific Institute San Raffaele Hospital Milan Italy The announcement of a new Diagnostic Imaging

More information

Pediatric Imaging Original Research

Pediatric Imaging Original Research Pediatric Imaging Original Research Podberesky et al. Dose Estimates of Pediatric Cardiac CT Angiography Using 320-MDCT Scanner Pediatric Imaging Original Research Daniel J. Podberesky 1 Erin Angel 2 Terry

More information

Magnetic Resonance Angiography

Magnetic Resonance Angiography Magnetic Resonance Angiography 1 Magnetic Resonance Angiography exploits flow enhancement of GR sequences saturation of venous flow allows arterial visualization saturation of arterial flow allows venous

More information

Safety, Efficacy and Indications of β-blockers to Reduce Heart Rate prior to Coronary CT Angiography-An Overview

Safety, Efficacy and Indications of β-blockers to Reduce Heart Rate prior to Coronary CT Angiography-An Overview Safety, Efficacy and Indications of β-blockers to Reduce Heart Rate prior to Coronary CT Angiography-An Overview WA Jahan 1, A Azam 2,S Deena 1, W Begum 1, WA Jahan 1, M Rahman 1 1 Department of radiology,

More information

The recent introduction of multirow detector computed

The recent introduction of multirow detector computed ORIGINAL ARTICLE Gauging Effective Spatial Resolution in Multirow Helical Cardiac Computed Tomography With a Dynamic Phantom Friedrich D. Knollmann, MD,* Tarkan Cangöz, MA,* Erdogan Cesmeli, PhD, Thomas

More information

Imaging ischemic heart disease: Role of CCTA

Imaging ischemic heart disease: Role of CCTA Plenary Session II Ischemic Heart Disease Imaging ischemic heart disease: Role of CCTA Florian Wolf Medical University of Vienna Department of Biomedical Imaging and Image Guided Therapy Division of Cardi0vascular

More information

Protocols in Cardiac CT Dr. Bruce Precious Dalhousie University Friday, April 15, 2016

Protocols in Cardiac CT Dr. Bruce Precious Dalhousie University Friday, April 15, 2016 Protocols in Cardiac CT Dr. Bruce Precious Dalhousie University Friday, April 15, 2016 Disclosure Statement: No Conflict of Interest I do not have an affiliation, financial or otherwise, with a pharmaceutical

More information

Coronary angiography is the standard way of visualizing

Coronary angiography is the standard way of visualizing Clinical Investigation and Reports Coronary Artery Fly-Through Using Electron Beam Computed Tomography Peter M.A. van Ooijen, MSc; Matthijs Oudkerk, MD, PhD; Robert J.M. van Geuns, MD; Benno J. Rensing,

More information

MR Advance Techniques. Cardiac Imaging. Class IV

MR Advance Techniques. Cardiac Imaging. Class IV MR Advance Techniques Cardiac Imaging Class IV Heart The heart is a muscular organ responsible for pumping blood through the blood vessels by repeated, rhythmic contractions. Layers of the heart Endocardium

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine Non Contrast MRA Mayil Krishnam Director, Cardiovascular and Thoracic Imaging University of California, Irvine No disclosures Non contrast MRA-Why? Limitations of CTA Radiation exposure Iodinated contrast

More information

Coronary Calcium Screening Using Low-Dose Lung Cancer Screening: Effectiveness of MDCT with Retrospective Reconstruction

Coronary Calcium Screening Using Low-Dose Lung Cancer Screening: Effectiveness of MDCT with Retrospective Reconstruction Cardiac Imaging Original Research Kim et al. Coronary Calcium Screening Using Lung Cancer Screening Cardiac Imaging Original Research Sung Mok Kim 1 Myung Jin Chung 1 Kyung Soo Lee 1 Yeon Hyun Choe 1 Chin

More information

Dose Reduction Options in Cardiac CT

Dose Reduction Options in Cardiac CT Dose Reduction Options in Cardiac CT Doyle P, Ball P*, Donnelly P # Radiological Sciences & Imaging, Forster Green Hospital *Department of Radiology, Ulster Hospital # Department of Cardiology, Ulster

More information

Disclosures CORONARY CALCIUM SCORING REVISITED. Learning Objectives. Scoring Methods. Consultant for M2S, Inc. Coronary Calcium Scoring: Software

Disclosures CORONARY CALCIUM SCORING REVISITED. Learning Objectives. Scoring Methods. Consultant for M2S, Inc. Coronary Calcium Scoring: Software CORONARY CALCIUM SCORING REVISITED Disclosures Consultant for M2S, Inc. Julianna M. Czum, MD Director, Division of Cardiothoracic Imaging Department of Radiology Dartmouth Hitchcock Medical Center Assistant

More information

Patient preparation and coronary CTA techniques. Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging

Patient preparation and coronary CTA techniques. Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging Patient preparation and coronary CTA techniques Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging Overview 1. Patient preparation 2. Scanning techniques Patient preparation Preparation

More information

Cardiac spiral dual-source CT with high pitch: a feasibility study

Cardiac spiral dual-source CT with high pitch: a feasibility study Eur Radiol (2009) 19: 2357 2362 DOI 10.1007/s00330-009-1503-6 PHYSICS Dirk Ertel Michael M. Lell Frank Harig Thomas Flohr Bernhard Schmidt Willi A. Kalender Cardiac spiral dual-source CT with high pitch:

More information

Diagnostic and Prognostic Value of Coronary Ca Score

Diagnostic and Prognostic Value of Coronary Ca Score Diagnostic and Prognostic Value of Coronary Ca Score Dr. Ghormallah Alzahrani Cardiac imaging division, Adult Cardiology department Prince Sultan Cardiac Center ( PSCC) Madina, June 2 Coronary Calcium

More information

MUGA Scan. A Patient s Guide. Copyrighted Material. HeartWise Patient Education

MUGA Scan. A Patient s Guide. Copyrighted Material. HeartWise Patient Education MUGA Scan Copyrighted Material HeartWise Patient Education 800-747-1606 A Patient s Guide What Is a MUGA Scan? A multiple gated acquisition (or MUGA) scan is a test that uses a radioactive substance, called

More information

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview CARDIAC MRI Dr Yang Faridah A. Aziz Department of Biomedical Imaging University of Malaya Medical Centre Cardiovascular Disease Diseases of the circulatory system, also called cardiovascular disease (CVD),

More information

Medical Policy An Independent Licensee of the Blue Cross and Blue Shield Association

Medical Policy An Independent Licensee of the Blue Cross and Blue Shield Association CTA for Coronary Artery Evaluation Page 1 of 22 Medical Policy An Independent Licensee of the Blue Cross and Blue Shield Association Title: Contrast-Enhanced Computed Tomography Angiography (CTA) for Coronary

More information

A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering

A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering DOI 10.1007/s10554-015-0802-z ORIGINAL PAPER A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering J. D. van Dijk 1,4 E. D. Huizing 1,4 P.

More information

...a view. Aquilion ONE... of the future. Introducing our. The European Scanning Centre Newsletter. Special Edition

...a view. Aquilion ONE... of the future. Introducing our. The European Scanning Centre Newsletter. Special Edition Introducing our Aquilion ONE... ESC acquires a 640 slice CT scanner 2 Imaging the Heart - Current Status of CT Angiography 4...a view of the future Toshiba and ESC sign Research Partnership Agreement 8

More information

Computed Tomography (CT) - Head

Computed Tomography (CT) - Head Scan for mobile link. Computed Tomography (CT) - Head Computed tomography (CT) of the head uses special x-ray equipment to help assess head injuries, severe headaches, dizziness, and other symptoms of

More information

Vascular and Interventional Radiology Original Research

Vascular and Interventional Radiology Original Research Vascular and Interventional Radiology Original Research Apfaltrer et al. High-Pitch Versus Standard-Pitch CTA of the Aorta Vascular and Interventional Radiology Original Research Paul Apfaltrer 1,2 E.

More information

Computed Tomography - From basics to clinical use Diagnosis and Treatment of Valvular Heart Disease. David Messika-Zeitoun

Computed Tomography - From basics to clinical use Diagnosis and Treatment of Valvular Heart Disease. David Messika-Zeitoun Computed Tomography - From basics to clinical use Diagnosis and Treatment of Valvular Heart Disease David Messika-Zeitoun Disclosures Lectures fees for Edwards, Abbott Consultant fees from Symetis You

More information

Update on Acute Aortic Syndrome

Update on Acute Aortic Syndrome SUNDAY Update on Acute Aortic Syndrome Diana Litmanovich, MD Learning objectives To be familiar with the definition, natural history, and imaging findings of acute aortic syndrome, including: I. Aortic

More information

SPECT-CT: Τι πρέπει να γνωρίζει ο Καρδιολόγος

SPECT-CT: Τι πρέπει να γνωρίζει ο Καρδιολόγος SPECT-CT: Τι πρέπει να γνωρίζει ο Καρδιολόγος Δρ Αναστασία Κίτσιου Διευθύντρια, Καρδιολογική Κλινική, Σισμανόγλειο ΓΝΑ Chair, Education Committee, Section on Nuclear Cardiology & Cardiac CT, EACVI, ESC

More information

Lei Zhao 1,2, Fabian Plank 2, Moritz Kummann 2, Philipp Burghard 2, Andrea Klauser 2, Wolfgang Dichtl 3, Gudrun Feuchtner 2.

Lei Zhao 1,2, Fabian Plank 2, Moritz Kummann 2, Philipp Burghard 2, Andrea Klauser 2, Wolfgang Dichtl 3, Gudrun Feuchtner 2. Original Article Improved non-calcified plaque delineation on coronary CT angiography by sonogram-affirmed iterative reconstruction with different filter strength and relationship with BMI Lei Zhao 1,2,

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Diagnostics consultation document

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Diagnostics consultation document NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Diagnostics consultation document Aquilion ONE, Brilliance ict, Discovery CT750 HD and Somatom Definition Flash for cardiac imaging in people with

More information

Pitfalls and Remedies in PET/CT imaging for RT planning

Pitfalls and Remedies in PET/CT imaging for RT planning Pitfalls and Remedies in PET/CT imaging for RT planning Tinsu Pan, Ph.D. M.D. Anderson Cancer Center The University of Texas Outlines Background Average CT (< 1 msv) to reduce mis-alignment of PET and

More information

Noninvasive Coronary Artery Imaging: Current Clinical Applications

Noninvasive Coronary Artery Imaging: Current Clinical Applications The Cardiac Society of Australia and New Zealand Noninvasive Coronary Artery Imaging: Current Clinical Applications These guidelines were revised by Dr Gary Liew, Profs Michael Feneley and Stephen Worthley

More information

Measurement of Ventricular Volumes and Function: A Comparison of Gated PET and Cardiovascular Magnetic Resonance

Measurement of Ventricular Volumes and Function: A Comparison of Gated PET and Cardiovascular Magnetic Resonance BRIEF COMMUNICATION Measurement of Ventricular Volumes and Function: A Comparison of Gated PET and Cardiovascular Magnetic Resonance Kim Rajappan, MBBS 1,2 ; Lefteris Livieratos, MSc 2 ; Paolo G. Camici,

More information

Computed Tomography (CT) - Sinuses

Computed Tomography (CT) - Sinuses Scan for mobile link. Computed Tomography (CT) - Sinuses Computed tomography (CT) of the sinuses uses special x-ray equipment to evaluate the paranasal sinus cavities hollow, air-filled spaces within the

More information

Computed Tomography (CT) - Chest

Computed Tomography (CT) - Chest Scan for mobile link. Computed Tomography (CT) - Chest What is CT Scanning of the Chest? Computed tomography, more commonly known as a CT or CAT scan, is a diagnostic medical test that, like traditional

More information

Echo in a Heartbeat. Article from the customer magazine Medical Solutions, April

Echo in a Heartbeat. Article from the customer magazine Medical Solutions, April Echo in a Heartbeat Article from the customer magazine Medical Solutions, April 2009 www.siemens.com/healthcare-magazine Echo in a Heartbeat Recently, Siemens introduced a new echocardiography system that

More information

Justification of the use of CT for individual health assessment of asymptomatic people: the Chinese experience

Justification of the use of CT for individual health assessment of asymptomatic people: the Chinese experience Justification of the use of CT for individual health assessment of asymptomatic people: the Chinese experience Liang Wang, MD, PhD Professor, attending radiologist wang6@tjh.tjmu.edu.cn Dept. of Radiology

More information

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Computed Tomography to Detect Coronary Artery Calcification File Name: computed_tomography_to_detect_coronary_artery_calcification Origination: 3/1994 Last CAP Review 10/2017 Next

More information

Coronary Artery Calcium Score

Coronary Artery Calcium Score Coronary Artery Calcium Score August 19, 2014 by Axel F. Sigurdsson MD 174 Comments essential for living organisms. Calcium is a chemical element that is Most of the calcium within the human body is found

More information

Evaluation and treatment of intracranial aneurysms using Dual Energy CT Angiography (DECTA) and rotational Digital Subtraction Angiography (DSA).

Evaluation and treatment of intracranial aneurysms using Dual Energy CT Angiography (DECTA) and rotational Digital Subtraction Angiography (DSA). Evaluation and treatment of intracranial aneurysms using Dual Energy CT Angiography (DECTA) and rotational Digital Subtraction Angiography (DSA). L. Testaverde, G. Pelle, A. Saltarelli, P. Rabuffi, M.

More information

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association MEDICAL POLICY SUBJECT: CARDIAC/CORONARY COMPUTED TOMOGRAPHIC ANGIOGRAPHY PAGE: 1 OF: 6 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial

More information

Medical Imaging. Alex Elliott Western Infirmary Glasgow

Medical Imaging. Alex Elliott Western Infirmary Glasgow Medical Imaging Alex Elliott Western Infirmary Glasgow History of medical imaging X-rays - Roentgen, 1895 Nuclear medicine - Cassen, 1951 Ultrasound Donald, 1962 SPECT - Kuhl, Edwards, 1963 PET Ter-Pogossian,

More information

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Alice B. Smith, MD

More information

Computed tomography of the chest: I. Basic principles

Computed tomography of the chest: I. Basic principles BJA Education, 15 (6): 299 304 (2015) doi: 10.1093/bjaceaccp/mku063 Advance Access Publication Date: 2 February 2015 Matrix reference 1A03, 2A12 Computed tomography of the chest: I. Basic principles P

More information

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT Ralf Koos 1 Andreas Horst Mahnken 2 Anil Martin Sinha 1 Joachim Ernst Wildberger 2 Rainer Hoffmann 1 Harald Peter Kühl 1 Received March 3, 2004; accepted after revision May 18, 2004. 1 Department of Cardiology,

More information

CT or PET/CT for coronary artery disease

CT or PET/CT for coronary artery disease CT or PET/CT for coronary artery disease Rotterdam 2012 Juhani Knuuti, MD, PhD, FESC Turku PET Centre University of Turku Turku, Finland Juhani.knuuti@utu.fi Turku PET Centre University of Turku Åbo Akademi

More information

General Imaging. Imaging modalities. Incremental CT. Multislice CT Multislice CT [ MDCT ]

General Imaging. Imaging modalities. Incremental CT. Multislice CT Multislice CT [ MDCT ] General Imaging Imaging modalities Conventional X-rays Ultrasonography [ US ] Computed tomography [ CT ] Radionuclide imaging Magnetic resonance imaging [ MRI ] Angiography conventional, CT,MRI Interventional

More information

Patient dose assessment of CT perfusion scanning at the RSCH

Patient dose assessment of CT perfusion scanning at the RSCH Patient dose assessment of CT perfusion scanning at the RSCH Lesley Leavesley, Emma Whitehead, Matthew Pryor, Debbie Peet Regional Radiation Protection Service Royal Surrey County Hospital, Guildford Overview

More information

Metal Artefact Reduction in CT

Metal Artefact Reduction in CT Metal Artefact Reduction in CT DANIEL MARRINER Metal Artefact Reduction in CT Metal Artefact Clinical Indications for MAR SEMAR and How It Works Technical Considerations Case Studies utilising SEMAR Metal

More information

Assessment of Non-Calcified Coronary Plaques Using 64-Slice Computed Tomography: Comparison With Intravascular Ultrasound

Assessment of Non-Calcified Coronary Plaques Using 64-Slice Computed Tomography: Comparison With Intravascular Ultrasound ORIGINAL ARTICLE DOI 10.4070 / kcj.2009.39.3.95 Print ISSN 1738-5520 / On-line ISSN 1738-5555 Copyright c 2009 The Korean Society of Cardiology Assessment of Non-Calcified Coronary Plaques Using 64-Slice

More information

Preview of Presentation

Preview of Presentation Preview of Presentation Discuss Healthcare Environment Clinical Implementation of Technical Innovations SNMMI/ASNC Joint Statement on Rb-82 Cardiac PET Imaging Protocol Principles of Rb-82 Cardiac Imaging

More information

Diagnostic Accuracy of Angiographic View Image for the Detection of Coronary Artery Stenoses by 64-Detector Row CT

Diagnostic Accuracy of Angiographic View Image for the Detection of Coronary Artery Stenoses by 64-Detector Row CT Diagnostic Accuracy of Angiographic View Image for the Detection of Coronary Artery Stenoses by 64-Detector Row CT A Pilot Study Comparison With Conventional Post-Processing Methods and Axial Images Alone

More information

Objectives 8/17/2011. Challenges in Cardiac Imaging. Challenges in Cardiac Imaging. Basic Cardiac MRI Sequences

Objectives 8/17/2011. Challenges in Cardiac Imaging. Challenges in Cardiac Imaging. Basic Cardiac MRI Sequences 8/17/2011 Traditional Protocol Model for Tomographic Imaging Cardiac MRI Sequences and Protocols Frandics Chan, M.D., Ph.D. Stanford University Medical Center Interpretation Lucile Packard Children s Hospital

More information

Tremendous progress in computed tomography (CT) technology

Tremendous progress in computed tomography (CT) technology ORIGINAL ARTICLE Demonstration of Dose and Scatter Reductions for Interior Computed Tomography Deepak Bharkhada, MS,* Hengyong Yu, PhD,Þ Robert Dixon, PhD, Yuchuan Wei, PhD,* J. Jeffrey Carr, MD, # J.

More information

Head-to-Head Comparison of Coronary Plaque Evaluation Between Multislice Computed Tomography and Intravascular Ultrasound Radiofrequency Data Analysis

Head-to-Head Comparison of Coronary Plaque Evaluation Between Multislice Computed Tomography and Intravascular Ultrasound Radiofrequency Data Analysis JACC: CARDIOVASCULAR INTERVENTIONS VOL. 1, NO. 2, 2008 2008 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-8798/08/$34.00 PUBLISHED BY ELSEVIER INC. DOI: 10.1016/j.jcin.2008.01.007 Head-to-Head

More information

Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease

Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease Hendrik Treede, MD, Christoph Becker, MD, Hermann Reichenspurner,

More information