SIGNAL PROCESSING and KNOWLEDGE ACQUISITION in OBJECTIVE BLOOD PRESSURE MEASUREMENT with CONVENTIONAL CLINICAL SPHYGMOMANOMETRY

Size: px
Start display at page:

Download "SIGNAL PROCESSING and KNOWLEDGE ACQUISITION in OBJECTIVE BLOOD PRESSURE MEASUREMENT with CONVENTIONAL CLINICAL SPHYGMOMANOMETRY"

Transcription

1 SIGNAL PROCESSING and KNOWLEDGE ACQUISITION in OBJECTIVE BLOOD PRESSURE MEASUREMENT with CONVENTIONAL CLINICAL SPHYGMOMANOMETRY Ventzas, D. 1 - TEI Larisa Angelopoulos A. Biomedical Eng. Cardiologist, MD ABSTRACT: The paper presents and compares the principles of auscultatory and oscillometric techniques. It focus on the signal processing and knowledge acquisition needs and techniques and finally concludes on their accuracy and reliability. KEYWORDS: Blood pressure, signal, knowledge, Korotkoff sounds, errors, auscultatory, oscillometric. INTRODUCTION: Non-invasive (i.e. non-surgical) methods sacrifice a degree of accuracy for patient safety, comfort, and convenience. Possible uses for Self Blood Pressure Measurement (SBPM) devices are white coat hypertension, systolic hypertension (higher in older), identification of hypotension, pregnancy, hypertensive patients with diabetes mellitus, blood pressure control in diabetes with macro vascular complications, resistant hypertension, failure to reduce BP with hypotensive agents. Blood pressure measurements that are chronically elevated may indicate hypertension and is a major contributing factor in heart disease and stroke since the level of arterial pressure is the basis for major diagnostic and therapeutic decisions in medicine, the measurement must be correct and maximally reproducible. PHYSIOLOGY of BLOOD PRESSURE MEASUREMENT: The introduction of innovative technologies in clinical sphygmomanometry allow not only accurate and reproducible non-invasive measurement of blood pressure but also an assessment of blood pressure as a dynamic phenomenon of the cardiovascular system. The difference between the systolic and diastolic pressure is called the pulse pressure. However, if we want to think in terms of the average force exerted against the artery wall by blood, it is worthwhile to consider a single average pressure. The mean arterial pressure (MAP) is a calculated average arterial pressure. The mean arterial pressure is defined as the diastolic pressure plus one third of the pulse pressure. Individuals blood pressure values have wide distribution. Fig. 1. Time series and averaged autospectra of beat-to-beat changes in mean arterial pressure (MAP) Table 1. Geometrical Definitions of Blood Pressure Systolic Blood Pressure (SBP): maximum blood pressure / arterial wall distension. Mean Arterial Pressure (MAP) Diastolic Blood Pressure (DBP): minimum artery pressure, smaller inward distension A normal objective blood pressure measurement depends on the person s age, sex, heredity, fitness, weight, physical conditioning, past illness, time of day, altitude, activity, climate and environment. In general, blood pressure is lower in the morning and increases throughout the day; it is lower in warm and higher in cold weather. Physical activity can have a significant short term impact on blood 1 D. Ventzas, Informatics & Telecommunications Dpt, TEI Larisa, Larisa

2 pressure, while work, exercise, smoking, diet, eating, drinking beverages containing caffeine or alcohol - even talking, laughing, or crying and emotional stress will all affect a person s blood pressure. Some useful evidence from epidemiological studies and clinical trials support for nurses and health visitors suggest, Table 2. Classifying stages of hypertension on the basis of average blood pressure levels, clinicians should specify presence or absence of target organ disease and additional risk factors. Table 2. Blood Pressure for Adults Systolic BP Diastolic BP Follow up recommended for Adults Category <120 mmhg <80 optimal <130 mmhg <85 Recheck in 2 years normal Recheck in 1 year High normal Confirm within 2 months mild Evaluate or refer to source of care within 1 month moderate Evaluate or refer to source of care within 1 week severe < 210 < 120 Evaluate or refer to source of case immediately In general, blood pressure is dependent on four main factors: cardiac output (the efficiency of the pumping process), the peripheral resistance of the vascular system (vasodilators and vasoconstrictors, viscosity), vessel elasticity and the blood volume (electrolytes). Blood pressure may increase to compensate for the reduced flow. Fig. 2. Typical recording of automatic cuff deflation (female, age = 20). The first/last notch wave is the systolic/diastolic pressure. Pulsus alternant, or beat variation in the intensity of the Korotkoff sounds, occurs in people with fluctuating systolic pressure. Fig. 3. Relative Intensity of Korotkoff sounds and phases 2

3 AUSCULTATORY METHOD PRINCIPLES - KOROTKOFF SOUNDS: With the palpatory method, the changes in blood flow associated with systolic and diastolic pressures are felt, rather than being heard. With the ausculatory method, a stethoscope is used to listen to the blood flow in the arteries. The auscultatory technique is based on the ability of the human ear to detect and distinguish sounds. This is a great advantage since it allows the clinician to determine the quality of each measurement. However, inherent in this is the possibility for measurement error due to differences in hearing acuity from clinician to clinician. Unqualified or inexperienced personnel may be more susceptible to outside noise, other interference, or inconsistent assessment of Korotkoff sounds. In an attempt to increase reproducibility, some automated devices have replaced the human ear with a microphone. In the auscultatory method the bladder is rapidly and steadily inflated to a pressure mmhg above the level, then partially unscrewed (opened) the valve and deflated the bladder at 2 mm/sec while listening for the appearance of Korotkoff sounds. The clinician uses the stethoscope to listen for the Korotkoff sounds as the cuff deflates. The sounds are divided into five phrases based on the loudness and quality of the sounds. Phase Sound Indicates 1 1 st appearance of clear repetitive loud, clear sound, faint tapping heard with increasing intensity 2 Start swishing sound, softer and longer, with the quality of an intermittent murmur, succession of murmurs, blows, or swishes may disappear if cuff is deflated slowly. 3 Intensity increases, crisper and louder, with thumping quality similar to phase 1, distinct tapping. 4 Sounds are abrupt muffling, less distinct and softer blowing quality, disappear completely, muffled sound abruptly replaces the This coincides approximately with the appearance of a palpable pulse. The pressure at which the sounds first appear corresponds to the systolic pressure. At pressures below diastolic, the cuff does not occlude the artery. Therefore, blood flow is no longer impeded. Forces on the arterial wall are dominated by haemodynamic factors. This region is useful for calculations of arterial compliance or blood flow. Diastolic Pressure (the beginning of Phase 5) in children < 13 years old and in adults that are exercising, pregnant or hyperthyroid (See Phase 5) Diastolic pressure in adults - occurs 5-10 mm Hg below Phase 4 in normal adults. In states of increased rate of blood flow it may be greater than 10 mm Hg below Phase 4, in these cases the Phase 4 sound should be used as Diastolic Pressure in adults. thumping sounds of Phase 3. 5 All sounds disappear. Phase IV, Phase V, or a combination of the two best represent diastolic pressure. Some patients may not have audible Phase IV, and in others Phase V is difficult to determine. Usually the pressures at which muffling and disappearance occur often are only within 10 mm of Hg. The onset of Phase V defines the diastolic pressure in adults. At rest, blood flow is laminar and silent. The sound of turbulent blood flow can be heard The vibrations in the artery walls are called Korotkoff vibrations Korotkoff sound may be audible all the way to 0 mmhg The BP cuff is completely deflated before and after measurement If BP is to be retaken in the same arm, wait 2 minutes before cuff reinflated to allow the release of blood trapped in the veins to enter back into circulation In children less than 13 years of age, pregnant women, and patients with high cardiac output or severe aortic regurgitation and peripheral vasodilatation, sounds are often heard at levels far below those at which muffling occurs, sometimes to levels approaching 0 mmhg. Korotkoff sounds may be difficult to hear and arterial pressure difficult to measure when arterial pressure rises at a slow rate (as in aortic stenosis), when the vessels are markedly constricted (as in shock), and when the stroke volume is reduced (as in severe heart failure). Very soft or inaudible Korotkoff sounds can often be accentuated by dilating the blood vessels of the upper extremities simply by opening and closing the fist repeatedly. Korotkoff sounds can usually be detected by auscultating with the stethoscope (or by palpation, or the Doppler ultrasound blood flow detecting) over the radial artery to obtain both systolic and diastolic pressures. A multi-processor scheme blood pressure measurement with built-in microcomputer that analyses the digitized wave, recognizes the systolic and diastolic blood pressure, samples data converted into digital data and stored in the memory, deflates step wise with certain speed. Initially the 3

4 CPU checks the signal level, and if it is sufficient for the measurement, the wave-forms of the vascular sounds are analyzed to measure the blood pressure; otherwise another attempt may be tried automatically. The cuff pressure, the amplifiers gain, etc are readjusted by the CPU, through RS-232C interface. Fig. 4. Indirect blood pressure measurements: oscillometric and auscultatory measurement. AUSCULTATORY METHOD ERRORS: Errors in the auscultatory blood pressure method is due to equipment, observer, subject, interface-operation, interpretation measurement and technique due to poor sound transmission, distorted sounds, sounds not well heard, extraneous noise, danger of missing auscultatory gap, inaccurate reading, with results such as blood pressure level too high / low / variable, under / over estimated of systolic / diastolic pressure. Common errors are: ERRORS, ACTION and CAUSE do not highlight the 1 st sec when the heart rate calculated artifacts or poor averaged information pressure take in both arms, but differ by 5-10 mmhg we prefer the highest pressure indicating arm 20 mmhg differences orthostatic hypotension standing drops systolic pressure and increases diastolic pressure excessive patient motion leaky cuff or hose oscillometric cycling ischemia, neuropathy. Inaccuracies may be caused by arm engorged with blood daily and seasonal basis lowest during sleep and rises in the morning and throughout the day. Do measure your pressure the same time each day. Record the date and time Do rest 5 to 10 minutes before measuring your blood pressure. allow 5 to 10 minutes rest BETWEEN subsequent measurements. blood pressure history Home measurements control of the deflation rate pressure is released too quickly vacuum is formed above the mercury and the meniscus is distorted. Low sound levels indicates low systolic, high diastolic pressure Differences at 5 mmhg rejects measurements Normal pressures might indicate hypotension, if previous pressures re high Erroneous placement of stehoscope Failure to make full skin contact with its sensing bell Possibility of shock Venous engorgement of the patients arm from repeated inflations of the cuff Patient s effort to support arm raise pressures Brachial artery to heart anatomy changes pressures sound-dependent algorithms 4

5 Fig. 5. Geometry problems a. Placing the cuff on the arm b. arm and cuff level with heart and patient orthostatic positions Sphygmomanometer arrangement Fig. 6. Occasionally the Korotkoff sounds become inaudible during Phase II or III, only to reappear as the pressure in the cuff is reduced further. This period of silence is called the auscultatory gap and is common in older and hypertensive patients. Raising the arm before inflating the cuff at max systolic+30 mmhg eliminates it. Unrecognized ausculatory gap may lead to underestimation of systolic pressure or overestimation of diastolic pressure PRINCIPLES of OSCILLOMETRIC TECHNIQUE: Most automated devices make use of oscillometry. Pressure changes in the artery are recorded with every heartbeat and displayed as an oscillogram. The oscillogram consists of two parts: 1. pressure curve as the cuff inflates and deflates 2. oscillations/vibrations of the arterial wall Unlike auscultatory techniques, which measure systolic and diastolic but estimate mean arterial pressure (MAP), oscillometric devices measure the mean but estimate systolic and diastolic. The point of maximum amplitude is considered mean arterial pressure. Therefore, an error in MAP may produce inaccurate values for systolic and diastolic. The Pulse Dynamic waveform is directly affected by variations in blood flow and blood pressure. The waveform reflects, to a certain extent, the state of the cardiovascular and peripheral system. Oscillometric devices for blood pressure (BP) measurement at the wrist are becoming more widely used in clinical practice. However, systematic comparisons with standard upper-arm auscultatory BP mercury sphygmomanometry measurement at the brachial artery are scarce. A motorized pump inflates the cuff to approximately 180 mmhg (120 mmhg for neonatal patients) initially. Then we deflate cuff pressure gradually. A pressure transducer detects air pressure and sends a signal to the measurement circuits. As pressure in the cuff is reduced, blood flows in the previously occluded artery and the pressure is read by the transducer. The point at which oscillation increases sharply is defined as the 5

6 systolic pressure. As the cuff deflates further, oscillation amplitude increases to a maximum, then decreases. The point of peak oscillation amplitude is defined as the mean arterial pressure. The point at which the system detects a loss of oscillation amplitude is defined as the diastolic pressure. The oscillometric technique uses a pressure transducer and electronic data processing to non invasively determine systolic, mean, and diastolic pressures. Fig. 7. Oscillatory blood pressure measurement the oscillations begin at approximately the level of systolic pressure and reach their greatest amplitude at the level of mean arterial pressure. Diastolic pressure is a derived value. Fig. 8. Digital Representation of Pressure Waveform in Brachial Artery in the oscillometric method Section I Section II Section III Section IV Cuff Pressure > Systolic Systolic Pressure > Cuff Cuff Pressure Between Mean Cuff Pressure Less than Blood Pressure Pressure > Mean and Diastolic Pressures Diastolic Pressure artery remains fully the blood spurts through the occluded-no blood flow artery at a high velocity Pressure waves generated by heartbeat create pulsatile pressure signal; highly representative of aortic activity "super-systolic" portion, the artery remains fully occluded. Pressure waves generated by cardiovascular activity create pulsatile arterial distension proximal to the occlusion. Unlike later oscillations, the occlusion prevents blood flow from contributing to the measured pressures (i.e. the pressures are dominated by forces directly generated by the heart). creates a Bernoulli effect causing a phase shift in the pressure signal As the cuff deflates, increasing portions of the cardiac cycle generate pressures that exceed cuff pressure. Therefore, an increasing volume of blood flows through the artery. The increased flow causes an increase in the amplitude and phase shift of the pressure signal. When cuff pressure reaches mean arterial pressure, the forces produced by blood flow balance the cuff pressure. Mean arterial pressure is the average pressure in the artery, and is marked by the second triangle icon. As cuff pressure continues to decrease, it is no longer sufficient to occlude the artery. The alleviation of cuff pressure causes the magnitude of the pressure signal to decrease. This is reflected in the decreasing magnitude of the waveform. At pressures lower than diastolic, the cuff does not occlude the artery. Therefore, blood flow is no longer impeded. This final portion of the waveform is known as the sub-diastolic range of measurement In this region, the forces on the arterial wall are dominated by hemodynamic factors. Therefore, any arterial wall distention can be used for calculations of arterial compliance (change in volume/change in pressure), or blood flow based on changes in hemodynamic parameters. 6

7 source of additional cardiovascular information Pattern recognition is used to determine systolic blood pressure Systolic pressure is marked by the first triangle icon.. When the time-dependent signal reaches a characteristic DBP point, the pattern recognition algorithm identifies DBP (marked by the third triangle icon). ACCURACY OF PULSE DYNAMICS: In clinical trials, pulse dynamics correlated well with both auscultatory and invasive methods. Pulse dynamics produces measurements that are usually higher than auscultatory measurements for the same patient with expected differences of 5-15 mmhg systolic and 5-10 mmhg diastolic. Pulse dynamics compares well with invasive measurements, which are considered standard for accuracy. Clinical studies produced correlation coefficients of 0.94/0.95/ 0.91 for systolic/mean/diastolic. Movement artifact and certain arrhythmia are easily detectable by glancing at the measurement waveform. Signal noise (artifact) due to patient movement, inadequate cuff placement, or other factors will produce waveform irregularities that may be visually detected. This allows clinicians to discard suspicious measurements, and gives them more insight into the cardiovascular condition of the patient. Fig. 9. Pressure Waveform in the oscillometric method patient movement (a) pressure arrhythmia (b) The existence of positive or negative spike which appears in the first wave of Tone-Segment, especially steep up rising spike, for the diastolic blood pressure, which appears in up-slope of the wave of Tone-Segment. Several factors influence the accuracy of an automated device such as distorted signal. Systematic and random errors in measurement can occur at each of the interactionary points of the technique (arteries, cuff, sensor, processing). accurate transmission and interpretation of a signal (Korotkoff sound or pulse wave) readings should be accurate, representative, interpretable, reproducible, standardised Fig. 10. Pulse Dynamic Oscillometrics vs. Auscultatory Method (ANSI/AAMI SP ). Studies and pattern recognition technology validated the use of Pulse Dynamics in the non invasive low cost measurement of arterial compliance and left ventricular contractility. 7

8 The mercury manometer is preferred and recommended over the aneroid manometer (prone to mechanical alterations that affect its accuracy) and is less likely to become decalibrated. The mercury reservoir should be full so that the upper curve of the meniscus is at the zero level before the cuff is inflated. The mercury should rise and fall freely in the column. The column should be kept vertical. supine to standing results in slight decrease in systolic blood pressure, slight increase or no change in diastolic blood pressure and a mild increase in heart rate postural hypotension, observed in elderly patients under antihypertensive therapy secondary hypertension. systolic pressure in the legs is up to 20 mmhg higher than in the arms the diastolic pressures in the legs is virtually identical to this in the arms In elderly patients who have sclerotic, calcified vessels, it is likely that the systolic pressure is overestimated by the indirect method of measurement. "pseudohypertension" blood pressure of pregnant women is more complicated because of the wide pulse pressure and the need to record both Phase IV and Phase V as sounds is heard to 0 mmhg. In dialysis patients or undergone a mastectomy, blood pressure measured in opposite arm In patients with cardiac dysrhythmias and atrial fibrillation or frequent premature beats, the systolic pressure varies widely from beat to beat In a patient with slow heart rate, reduce the pressure in the cuff gradually (2 mmhg per heartbeat) to avoid underestimation of systolic pressure and over-estimation of diastolic pressure Disparity in pressure between the two arms may occur in patients with congenital heart disease, peripheral vascular disease, unilateral neurological and musculoskeletal abnormalities, and aortic dissection. In hypertensive patients the arms with the higher pressure should be used The Device The Patient Interface-Operation Interpretation Measurement Appropriate cuff size, Blood pressure levels are with the palpatory lowest and most stable Auto Zero - larger cuff needed for affected by emotional, method avoid under readings Automatically those with a large arm environmentaland physical inflation of the cuff establishes zero circumference stimuli, so effort should in patients with pressure be made to standardize auscultatory gap reference before conditionsof measurement, and over inflation in each or series of keeping extraneous those with very low measurements influences to a minimum. blood pressure Bladder length of The subject must be Palpate the brachial Ambulatory Blood Inflate cuff 30 approximately comfortably seated with artery and place the Pressure Measurement mm Hg above times the bladder width the midpoint of the upper cuff so that the refers to repeated Doppler systolic and cuff length 80% x arm at the level of the midline of the measurements made pressure circumference heart (approximately the bladder is over the away from the medical Recommended cuff size fourth intercostals space, arterial pulsation, environment with a Newborn 3 x 6 cm when the individual is then wrap and portable, automatic seated). secure the cuff (self-inflating) recorder Infant 5 x 10 cm snugly around the in patients engaging in Child/lean cm subjects'bare upper their usual activities adult arm. Cuff loose Adult cm Adult large cm alarm limits operating mode cuff / hose connections isometric exercise during the measurement will transiently raise the blood pressure level subject lying or seated, heart level blood pressure should be determined with the patient standing as well as while sitting. When the patient is standing, support the raised arm at the level of the heart. When the subject is lying down, the arm should be at the side of the body, slightly raised application results in overestimation of the pressure. Place the head of the stethoscope over the brachial artery pulsation, just above and medical to the antecubital fossa but below the lower edge of the cuff, and hold firmly in place In arrhythmia, blood pressure is level variable and we take multiple measurements and average eye level with the manometer If reading gives systolic pressure below 140 / diastolic below 85, no further readings needed systolic greater than mean pulse rate out of range Deflate the cuff quickly and completely 8

9 cuff application 1. Cuff Wrapped too closely Blood pressure recording too high 2. Applied over clothing - Inaccurate reading 3. Palpatory pressure omitted - Danger of missing auscultatory gap 4. Inflation level too low - Underestimation of systolic pressure 5. Inflation rate too slow - Diastolic pressure too high 6. Deflation rate too fast - Systolic pressure too low 7. Deflation rate too slow - Diastolic pressure too high cuff orientation periodic recalibration. based on detection of Korotkoff sounds, and others use the oscillometric technique. Badly located brachial pulse; Doppler assists Automatic Calibration Sensor dependent geometry from the bed, at the level of the middle of the chest. Standing readings are needed for those in whom a significant postural drop could occur (e.g. the elderly, in hypertensives under treatment, or in the presence of other situations where postural hypotension is possible) With the subject standing for at least a minute, take at least one reading of the standing blood pressure with the arm extended and supported, plus a record of the pulse rate. Most people exhibit this circadian rhythm, which consists of a blood pressure decrease of approximately 15-25% during the night-time, with increases to daytime levels again in the morning. patient should as far as possible be relaxed and the arm well supported longer and wider cuff needed for adequate compression of the brachial artery in the obese patient with large upper arm - place the centre of the bladder over the brachial artery pulse Pulsus paradoxus is a normal fall in systolic pressure that occurs during inspiration, in patients with constrictive pericarditis and severe pulmonary disease or restrictive cardiomyopathy arrhythmia, heart valve disease Patients with arterial disease as SLE (Lupus Erythematosis), Raynaud syndrome have decreased elasticity of arterial wall. Restricted movement / vibration of the arterial wall causes a flat signal. quality of the stethoscope patient motion or physiology Inflate the cuff rapidly to 70 mmhg, and increase by 10 mmhg increments while palpating the radial pulse. Note the level of pressure at which the pulse disappears and subsequently reappears during deflation. Wait 30 seconds before a 2nd measurement if the 1st is unsatisfactory. error of overestimating the pressure when measuring with a cuff that is too small for an obese arm, can be considerable and can lead to unnecessary therapy An auscultatory gap, often detected in patients with high systolic pressure, is normal relationship between the blood pressure and pulse rate If levels are higher (hypertensive), repeated readings at 2 minute intervals required with a concurrent record of the pulse rate Since blood pressure to be more labile in elderly patients, it is particularly important to obtain several base line determinations before making diagnostic or therapeutic decisions. conventional mercury sphygmomanometer automated devices are more inaccurate at higher pressures. This could be due to various factors including a higher nonlinear resistance in the vessels. operator dependent Maximal Inflation Level (MIL) Determine the pressure to which to inflate the cuff for the measurement of the systolic blood pressure. Distorted sounds due to long stethoscope tubing considerable extraneous noise, surrounding recordings averaged Digit Inaccurate reading Record blood pressure to nearest 2 mmhg automatic and selfinflating adult-paediatricneonatal mode Adjusting Measurement Interval [min], EAN/PULSE RATE / MEAN CYCLE SUGGESTIONS FOR FURTHER WORK: The pressure signal is caused by the cuff / artery interaction. This signal can be analysed to measure more information than the mean arterial pressure or the estimated systolic and diastolic blood pressures. combined intermittent and continuous measurement techniques improve SNR more or no microphones for differential Korotkoff recognition ECG gating techniques may be used to minimize artifacts Constant pressure oscillometric measurement of pulses generated at a low, constant pressure, permit continuous beat-to-beat arterial blood pressures to be measured, whereas the traditional 9

10 oscillometric method is limited to certain oscillometric pulses at different pressures allowing only for intermittent measurement of blood pressure. Volume-oscillometric method is based on arterial volume oscillations instead of pressure oscillations. The infrasound, ultrasound, plethysmography, electrical conductivity (impedance) and arterial tonometry and combined techniques attempt to improve the auscultatory method by detecting low frequency Korotkoff vibrations below 50 Hz, including sub-audible vibrations; a waveform is produced, similar to the pressure-generated oscillometric waveform. CONCLUSION: Despite the relative lack of evidence for the widespread use of monitoring home blood pressure in hypertension it is clear that the market is growing for such devices. The greater availability and advertising has resulted in an increasing interest by the general public. As a result cheaper, lighter and easier to use devices have been produced. SBPM should ensure that their operation is optimised and reduce patients cardiovascular risk. The paper compared the principles of auscultatory and oscillometric techniques and presented their signal processing needs and knowledge acquisition for accuracy and reliability. REFERENCES 1. The British Hypertension Society. Blood pressure measurement CD ROM. London: BMJ Books, Eckert S, Gleichmann U, Zagorski O, Klapp A. Validation of the Omron R3 blood pressure self-measuring device through simultaneous comparative invasive measurements according to protocol of the German Institute for Validation. Blood Pressure Monit; 2: , Geddes LA, Voelz M, Combs C, Reiner D, Babbs CF. Characterization of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng 1982; 10: B. P. Imholz, G. J. Langenwouters, G. A. van Montrans, G. Parati, J. van Goudoever, K. H. Wesseling, W. Wieling, and G. Mancia, Feasibility of ambulatory, continous 24-hour finger arterial pressure recording, Hypertension, 21(1), 65 73, Jones DW, Frohlich ED, Grim CM, Grim CE, Taubert KA, for the Professional Education Committee, Council for High Blood Pressure Research. Mercury sphygmomanometers should not be abandoned: An advisory statement from the Council for High Blood Pressure Research, American Heart Association. Hypertens 2001; 37(2): O'Brien E, Petrie J, Littler WA, de Swiet M, Padfield PD, Dillon MJ. Blood pressure measurement: recommendations of the British Hypertension Society. 3rd ed. London: BMJ Publishing Group, O'Brien E. State of the market in 2001 for blood pressure measuring devices. Blood Pressure Monitoring 2001; 6: O'Brien E, Waeber B, Parati G, Staessen J, Myers M (2001) Blood pressure measuring devices: recommendations of the European Society of Hypertension. BMJ; 322: P. Rogers, V. Burke, P. Stroud, I.B. Puddey Comparison of oscillometric blood pressure measurements at the wrist with an upper-arm auscultatory mercury sphygmomanometer, Clinical & Experimental Pharmacology & Physiology, Vol 26 Is. 5-6 p. 477, May/June Sykes M K, Vickers M D, Hull C J 1991 Principles of measurement and monitoring in anaesthesia and intensive care. Blackwell Scientific Publications, Oxford 11. Staessen J, Thijs L and the participants of the First International Consensus Conference on SBPM. (2000) Developments of diagnostic thresholds for automated self-measurement of blood pressure in adults. Blood Press Monitor 5: Stergiou G, Skeva I, Zourbaki A, Mountokalakis D (1998) Self-monitoring of blood pressure at home: how many measurements needed?, Journal of Hypertension 16:6;

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04) SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: MEASURING BLOOD PRESSURE - MANUAL Nursing DATE: REVIEWED: PAGES: 2/80 7/17 1 of 5 RESPONSIBILITY: RN, LPN, Patient Care Technician Multi-skilled

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Section 03: Pre Exercise Evaluations and Risk Factor Assessment

Section 03: Pre Exercise Evaluations and Risk Factor Assessment Section 03: Pre Exercise Evaluations and Risk Factor Assessment ACSM Guidelines: Chapter 3 Pre Exercise Evaluations ACSM Manual: Chapter 3 Risk Factor Assessments HPHE 4450 Dr. Cheatham Purpose The extent

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

Ambulatory Services Orientation & Skill Review Skill Performance Checklist: Vital Signs. Name: Date:

Ambulatory Services Orientation & Skill Review Skill Performance Checklist: Vital Signs. Name: Date: Evaluation Levels Behavior Level 3 Met Performed all skills competently. No behaviors omitted. Level 2 Satisfactory Omitted no more than three (3) non-critical behaviors. Level 1 Unsatisfactory Omitted

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information

Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations

Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations ighapmlre33apg269_274 5/12/04 3:10 PM Page 269 impos03 302:bjighapmL:ighapmLrevshts:layouts: NAME Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations LAB TIME/DATE REVIEW SHEET exercise

More information

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

2. Measure a subject's blood pressure and heart rate both at rest and during exercise. Lab Activity 11 The Cardiovascular System Student Learning Objectives After completing this lab, you should be able to: 1. Define, explain and correctly use the key terms. 2. Measure a subject's blood

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement Direct Care Staff Trained in Accurate BP Measurement 1. 2. 3. 5. 22 to 26 cm Small adult (12X22 cm) 27 to 34 cm Adult (16X30 cm) 35 to 44 cm Large adult (16X36 cm) 45 to 52 cm Adult thigh (16X42 cm) widths

More information

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine Techniques of Vital Signs John Gazewood, MD, MSPH Department of Family Medicine Objectives Define normal in several ways. Describe correct technique of measuring vital signs. Describe correct technique

More information

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System!

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System! LAB: Blood Measurable Indicator of the Health of the Circulatory System! Lab Objectives. At the completion of the lab, you should be able to: measure pulse or heart rate (HR) and respiratory rate (RR);

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust.

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust. AN INTRODUCTION TO DOPPLER Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust. THE DOPPLER EFFECT The Doppler Principle was described by Physicist and mathematician

More information

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS LAB 8: CARDIOVASCULAR PHYSIOLOGY PART 1. HEART SOUNDS AND PULSE DETERMINATIONS Introduction Two distinct sounds can be heard during each cardiac cycle. These sounds are commonly described as lub and dup

More information

Physiology of the Circulatory System modified from

Physiology of the Circulatory System modified from Physiology of the Circulatory System modified from http://www.ekcsk12.org/science/aplabreview/aplab10.htm Introduction The circulatory system functions to deliver oxygen and nutrients to tissues for growth

More information

PHONOCARDIOGRAPHY (PCG)

PHONOCARDIOGRAPHY (PCG) PHONOCARDIOGRAPHY (PCG) The technique of listening to sounds produced by the organs and vessels of the body is called auscultation. The areas at which the heart sounds are heard better are called auscultation

More information

Managing Blood Pressure: It Takes a Team

Managing Blood Pressure: It Takes a Team Managing Blood Pressure: It Takes a Team Module Two How to Take Blood Pressure October 2011 MODULE TWO: How to Take Blood Pressure Module Two: HOW TO TAKE BLOOD PRESSURE Objectives Demonstrate proper technique

More information

Lab #10 Physiology of the Circulatory System

Lab #10 Physiology of the Circulatory System Lab #10 Physiology of the Circulatory System Introduction The circulatory system functions to deliver oxygen an nutrients to tissues for growth and metabolism, and to remove metabolic wastes. The heart

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure #6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure Objectives: Observe slide of artery and vein cross-section Auscultate heart sounds using a stethoscope Measure

More information

SECTION 16: BLOOD PRESSURE MEASUREMENT

SECTION 16: BLOOD PRESSURE MEASUREMENT CHRONIC KIDNEY DISEASE IN CHILDREN COHORT STUDY SECTION 16: BLOOD PRESSURE MEASUREMENT 16.1 OVERVIEW OF BLOOD PRESSURE MEASUREMENT This section describes the procedures that are to be performed to obtain

More information

Lecture #3 - Blood Pressure Recording Procedure. Equipment

Lecture #3 - Blood Pressure Recording Procedure. Equipment Lecture #3 - Blood Pressure Recording Procedure CKiD Blood Pressure Training Equipment Mabis Medic-Kit aneroid sphygmomanometer Mabis Medic-Kit assortment of cuff sizes Stethoscope with bell (pediatric

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

BIOE221. Session 2. Vital Signs. Biosciences Dept.

BIOE221. Session 2. Vital Signs. Biosciences Dept. BIOE221 Session 2 Vital Signs Biosciences Dept. www.endeavour.edu.au Session Objectives cont o Understand the physiology related to the vital signs of temperature, pulse, respiration and blood pressure.

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

Data Collection Worksheet

Data Collection Worksheet Data Collection Worksheet 1. Has a doctor or nurse ever said that you have: High blood pressure or hypertension? [ ] No [ ] Yes [ ] Not Sure 1.a. If Yes, then at what age were you first told this? Age

More information

LAB 9: Metabolic Rates

LAB 9: Metabolic Rates LAB 9: Metabolic Rates Introduction: The cardiovascular (circulatory) system functions to deliver oxygen and nutrients to tissues for growth and metabolism, and to remove metabolic wastes. The heart pumps

More information

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients Vital signs (pulse, blood pressure, temperature, respiratory rate, pain) are physiological parameters that a healthcare professional requires when dealing with patients. Accurate measurement of vital signs

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM AP Biology Laboratory Date: Name and Period: AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM OVERVIEW In this lab you will: 1. in Exercise 10A you will learn how to measure blood pressure. 2. in

More information

HUMAN ANATOMY AND PHYSIOLOGY

HUMAN ANATOMY AND PHYSIOLOGY HUMAN ANATOMY AND PHYSIOLOGY NAME Detection of heart sounds. Clean the ear pieces of the stethoscope before using. The ear pieces should be pointing slightly forward when inserted into the ears because

More information

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 64 CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 4.1 GENERAL This chapter presents the methodologies that are usually adopted for the measurement of blood pressure, heart rate and pulse

More information

How To Measure Vital Signs

How To Measure Vital Signs How To Measure Vital Signs How to Use This Tutorial This tutorial is intended for healthcare providers or students to teach basic vital signs skills Use the navigation buttons below to move through the

More information

1 Non-invasive measurement of arterial pressure

1 Non-invasive measurement of arterial pressure Non-invasive measurement of arterial pressure I. Background A. Circulatory systems Human circulation takes place in a closed system that consists of two subsystems, pulmonary circulation and systemic circulation,

More information

5.2. MEASUREMENT OF BLOOD PRESSURE

5.2. MEASUREMENT OF BLOOD PRESSURE 5.2. MEASUREMENT OF BLOOD PRESSURE Purpose of experiment To examine a person s blood pressure. Tasks of experiment To record the pressure changes in the vascular system: using the traditional method measure

More information

Physiology lecture 15 Hemodynamic

Physiology lecture 15 Hemodynamic Physiology lecture 15 Hemodynamic Dispensability (D) : proportional change in volume per unit change in pressure D = V/ P*V It is proportional (divided by the original volume). Compliance (C) : total change

More information

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation J. JILEK*, M. STORK** *Carditech, Culver City, California, USA **Department of Applied Electronics

More information

Blood Pressure Laboratory

Blood Pressure Laboratory Introduction The blood that circulates throughout the body maintains a flow and pressure. The nervous system can change the flow and pressure based on the particular needs at a given time. For example,

More information

IB TOPIC 6.2 THE BLOOD SYSTEM

IB TOPIC 6.2 THE BLOOD SYSTEM IB TOPIC 6.2 THE BLOOD SYSTEM THE BLOOD SYSTEM TERMS TO KNOW circulation ventricle artery vein 6.2.U1 - Arteries convey blood at high pressure from the ventricles to the tissues of the body Circulation

More information

IB TOPIC 6.2 THE BLOOD SYSTEM

IB TOPIC 6.2 THE BLOOD SYSTEM IB TOPIC 6.2 THE BLOOD SYSTEM TERMS TO KNOW circulation ventricle artery vein THE BLOOD SYSTEM 6.2.U1 - Arteries convey blood at high pressure from the ventricles to the tissues of the body Circulation

More information

Recommendations for Blood Pressure Measurement in Humans and Experimental Ani...

Recommendations for Blood Pressure Measurement in Humans and Experimental Ani... Recommendations for Blood Pressure Measurement in Humans and Experimental Ani... Página 1 de 41 Published online before print December 20, 2004, doi:10.1161/01.hyp.0000150859.47929.8e (Hypertension. 2005;45:142.)

More information

PART I: HEART ANATOMY

PART I: HEART ANATOMY Lab 7: Heart Sounds and Blood Pressure PART I: HEART ANATOMY a) You should be able to identify the following structures on an adult human heart diagram. the 4 chambers the bicuspid (mitral) and tricuspid

More information

Chapter 15: Measuring Height, Weight, and Vital Signs. Copyright 2012 Wolters Kluwer Health Lippincott Williams & Wilkins

Chapter 15: Measuring Height, Weight, and Vital Signs. Copyright 2012 Wolters Kluwer Health Lippincott Williams & Wilkins Chapter 15: Measuring Height, Weight, and Vital Signs Height and Weight Weight Baseline measurement at patient s first visit Measured in kg or lbs Common types of scales Balance beam Dial Digital Height

More information

BLOOD PRESSURE ASSESSMENT

BLOOD PRESSURE ASSESSMENT BLOOD PRESSURE ASSESSMENT Course Principles of Health Science Unit X Vital Signs Course Health Science Unit VI CPR / AED & First Aid Essential Question How do health care skills help to promote health

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS CARDIAC MECHANICAL SUPPORT PROGRAM GUIDELINES CARDIAC MECHANICAL SUPPORT: LVAD BASICS FREQUENT SCENARIOS AND TROUBLESHOOTING Review Date: July 2011

More information

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013

More information

HEART SOUNDS & BLOOD PRESSURE

HEART SOUNDS & BLOOD PRESSURE LABORATORY HANDBOOK HEART SOUNDS & BLOOD PRESSURE For the course Systems Biology Karolinska Institutet Department of Physiology and Pharmacology Physiology Education Postal address: 171 77 Stockholm Visiting

More information

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

Protocol 4: Measuring Blood Pressure

Protocol 4: Measuring Blood Pressure Zool 430L Protocols Page 8 of 12 Protocol 4: Measuring Blood Pressure 4.1 Via Auscultation (listening through stethoscope) The estimate of blood pressure is obtained by cutting off blood flow with a known

More information

Blood Pressure and Exercise Lab

Blood Pressure and Exercise Lab Blood Pressure and Exercise Lab Rob MacLeod March 9, 2006 1 Purpose and Background Purpose: To learn about external means to measure blood pressure, observe features of venous circulation, and observe

More information

Lesson 16 BLOOD PRESSURE

Lesson 16 BLOOD PRESSURE Physiology Lessons for use with the Lesson 16 BLOOD PRESSURE Indirect measurement Ventricular Systole & Diastole Korotkoff sounds Mean Arterial pressure Manual Revision 3.7.3 061808 Richard Pflanzer, Ph.D.

More information

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE Pamela B. Morris, MD, FACC, FAHA, FASPC, FNLA Professor Medicine and Cardiovascular Diseases Director, Seinsheimer Cardiovascular Health

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (  The story so far. Medical Electronics Dr Neil Townsend Michaelmas Term 2001 (wwwrobotsoxacuk/~neil/teaching/lectures/med_elec) Blood Pressure has been measured (in some way) for centuries However, clinical interpretation

More information

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms Chapter 29 Vital Signs Key Terms Blood pressure Body temperature Diastolic pressure Hypertension Hypotension Pulse Pulse rate Respiration Sphygmomanometer Stethoscope Systolic pressure Vital signs Measuring

More information

a central pulse located at the apex of the heart Apical pulse Apical-radial pulse a complete absence of respirations Apnea

a central pulse located at the apex of the heart Apical pulse Apical-radial pulse a complete absence of respirations Apnea Afebrile absence of a fever Apical pulse a central pulse located at the apex of the heart Apical-radial pulse measurement of the apical beat and the radial pulse at the same time Apnea a complete absence

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

to use the proper technique often lead to the wrong diagnosis, which may indirect measurement ofblood pressure. 240k

to use the proper technique often lead to the wrong diagnosis, which may indirect measurement ofblood pressure. 240k BRITISH MEDICAL JOURNAL VOLUME 293 6 SEPTlMBER 1986 611 J C PETRIE E T O'BRIEN British ypertension Society W A LITTLER RECOMMENDATIONS ON BLOOD PRESSURE MEASUREMENT M DE SWIET Incorrect technique-- False

More information

High Blood Pressure Blood Pressure Monitoring at Home -- F...

High Blood Pressure Blood Pressure Monitoring at Home -- F... Return to Web version High Blood Pressure Blood Pressure Monitoring at Home Why should I measure my blood pressure at home and keep a record of it? Measuring your blood pressure at home and keeping a record

More information

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1982 Characterization of the Oscillometric Method for Measuring Indirect

More information

Blood pressure. Arterial and venous circulation 1

Blood pressure. Arterial and venous circulation 1 Blood pressure. Arterial and venous circulation 1 Introduction Hales first measured blood pressure in 1733 by inserting tubes directly into the arteries of animals. Non invasive techniques for the measurement

More information

Physiology - 8 Hemodynamics - 1 M.jafar 24/3/2016 Turquoise Team

Physiology - 8 Hemodynamics - 1 M.jafar 24/3/2016 Turquoise Team 21 Physiology - 8 Hemodynamics - 1 M.jafar 24/3/2016 Turquoise Team Hemodynamics Today we will take about hemodynamics which is the study of the movement of blood and of the forces concerned. Now how the

More information

Chapter 7. Blood Pressure and Sound

Chapter 7. Blood Pressure and Sound Chapter 7. Blood Pressure and Sound Dr. Kurtuluş Erinç Akdoğan kurtuluserinc@cankaya.edu.tr Blood-pressure values in the various chambers of the heart and in the peripheral vascular system help the physician

More information

Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno. Biophysics of cardiovascular system

Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno. Biophysics of cardiovascular system Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno Biophysics of cardiovascular system 1 Lecture outline Mechanical properties of blood vessels Reynolds

More information

Hemodynamic Monitoring

Hemodynamic Monitoring Perform Procedure And Interpret Results Hemodynamic Monitoring Tracheal Tube Cuff Pressure Dean R. Hess PhD RRT FAARC Hemodynamic Monitoring Cardiac Rate and Rhythm Arterial Blood Pressure Central Venous

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

BP-600 Noninvasive Blood Pressure Sensor

BP-600 Noninvasive Blood Pressure Sensor Technical Note BP-600 Overview A person's cardiac output, peripheral vascular resistance, blood pressure, and other cardiovascular parameters change in response to the activities and events taking place

More information

Chapter 12 - Vital_Signs_and_Monitoring_Devices

Chapter 12 - Vital_Signs_and_Monitoring_Devices Introduction to Emergency Medical Care 1 OBJECTIVES 12.1 Define key terms introduced in this chapter. Slides 13 15, 17, 21 22, 26, 28, 30, 32 33, 35, 44, 47 48, 50, 55, 60 12.2 Identify the vital signs

More information

Lab Period: Name: Physiology Chapter 14 Blood Flow and Blood Pressure, Plus Fun Review Study Guide

Lab Period: Name: Physiology Chapter 14 Blood Flow and Blood Pressure, Plus Fun Review Study Guide Lab Period: Name: Physiology Chapter 14 Blood Flow and Blood Pressure, Plus Fun Review Study Guide Main Idea: The function of the circulatory system is to maintain adequate blood flow to all tissues. Clinical

More information

An Automated System for Blood Pressure Determination During Exercise

An Automated System for Blood Pressure Determination During Exercise An Automated System for Blood Pressure Determination During Exercise SEPHEN P. GLASSER, M.D., AND MAYNARD R. RAMSEY III, M.D., PH.D. with the technical assistance ofblanche Price SUMMARY Automation of

More information

WISEWOMAN UPDATE. December The Importance of Accurate Blood Pressure Measurements

WISEWOMAN UPDATE. December The Importance of Accurate Blood Pressure Measurements WISEWOMAN UPDATE December 2005 The Importance of Accurate Blood Pressure Measurements As you know, accurate blood pressure measurement is important! Even a few mmhg can make the difference between a person

More information

Assessment of techniques for measurement of blood pressure in infants and children

Assessment of techniques for measurement of blood pressure in infants and children Archives of Disease in Childhood, 1973, 48, 932. Assessment of techniques for measurement of blood pressure in infants and children A. M. ELSEED, E. A. SHINEBOURNE, and M. C. JOSEPH From the Paediatric

More information

2. The heart sounds are produced by a summed series of mechanical events, as follows:

2. The heart sounds are produced by a summed series of mechanical events, as follows: Heart Sounds. Phonocardiography 1 Objectives 1. Phonocardiography - Definition 2. What produces the heart sounds 3. Where to listen for the heart sounds 4. How to record a phonocardiogram 5. Normal heart

More information

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License Pan F, Zheng DC, He PY, Murray A. Does the Position or Contact Pressure of the Stethoscope Make Any Difference to Clinical Blood Pressure Measurements: An Observational Study. Medicine 214, 93(29). Copyright:

More information

Clinical Research Coordinator Skills Program

Clinical Research Coordinator Skills Program Clinical Research Coordinator Skills Program Vital Signs This Skill Requires Provider direction Standard Precautions Using purell before and after contact with the patient or the patient s environment

More information

Chapter 08. Health Screening and Risk Classification

Chapter 08. Health Screening and Risk Classification Chapter 08 Health Screening and Risk Classification Preliminary Health Screening and Risk Classification Protocol: 1) Conduct a Preliminary Health Evaluation 2) Determine Health /Disease Risks 3) Determine

More information

9/27/2016. Adherence and Monitoring of Hypertensive Patients. Conflicts of Interest. Brought to you though a partnership:

9/27/2016. Adherence and Monitoring of Hypertensive Patients. Conflicts of Interest. Brought to you though a partnership: Adherence and Monitoring of Hypertensive Patients. Conflicts of Interest Speaker has no conflicts of interest to disclose I have had no financial relationship over the past 12 months with any commercial

More information

Chapter 24 Vital Signs. Copyright 2011 Wolters Kluwer Health Lippincott Williams & Wilkins

Chapter 24 Vital Signs. Copyright 2011 Wolters Kluwer Health Lippincott Williams & Wilkins Chapter 24 Vital Signs Vital Signs Temperature Pulse Respiration Blood pressure When to Assess Vital Signs Upon admission to any healthcare agency Based on agency institutional policy and procedures Anytime

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

The Cardiac Cycle Clive M. Baumgarten, Ph.D. The Cardiac Cycle Clive M. Baumgarten, Ph.D. OBJECTIVES: 1. Describe periods comprising cardiac cycle and events within each period 2. Describe the temporal relationships between pressure, blood flow,

More information

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model Gye-rok Jeon, Jae-hee Jung, In-cheol Kim, Ah-young Jeon, Sang-hwa Yoon, Jung-man Son, Jae-hyung Kim, Soo-young Ye,

More information

Cardiovascular Physiology Laboratory Clive M. Baumgarten, Ph.D.

Cardiovascular Physiology Laboratory Clive M. Baumgarten, Ph.D. Cardiovascular Physiology Laboratory Clive M. Baumgarten, Ph.D. To pump blood, the heart has a rhythmical sequence of both electrical and mechanical events, the cardiac cycle. The electrical activity,

More information

Blood Pressure Protocol Training

Blood Pressure Protocol Training Blood Pressure Protocol Training May 31, 2018 Pat Spier, RN-BC, PCMH-CCE Barb Rice, RN-BC BCBS of North Dakota Fargo, ND 2013 Template and icons provided by The Advisory Board Company. Hypertension Education

More information

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT Blood Pressure Measurement Can Isik Electrical Engineering and Computer Science Department Syracuse University Syracuse, NY 13244 cisik@syr.edu Abstract Blood pressure is one of the fundamental vital signs,

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

!#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 !"#$ Blood Pressure Monitor Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 Introduction

More information

LAB 4. Human Cardiovascular Lab

LAB 4. Human Cardiovascular Lab 35 Assignments: LAB 4 Human Cardiovascular Lab Due before lab: Do IP exercises (Factors that Affect Blood Pressure and Arterial Baroreceptor Reflex (Pgs. 46-47). Quiz: Define the terms on pg 36 and IP

More information

Vital Signs. Vital Signs. Vital Signs

Vital Signs. Vital Signs. Vital Signs Vital Signs Vital Signs Why do vital signs? Determine relative status of vital organs Establish baseline Monitor response to Rx, meds Observe trends Determine need for further evaluation, Rx, intervention

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

SMALL GROUP SESSION 18A January 17th or January 19th. Groups 1-12: VS and Chest Exam and Harvey Stethophone Session

SMALL GROUP SESSION 18A January 17th or January 19th. Groups 1-12: VS and Chest Exam and Harvey Stethophone Session SMALL GROUP SESSION 18A January 17th or January 19th Groups 1-12: VS and Chest Exam and Harvey Stethophone Session Readings: Complete the cardiac examination web module. Mosby s Physical Examination, 4

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

Webinar Logistics. Now What Do I Need To Do? Blood Pressure in Kids January NC Department of Health and Human Services Division of Public Health

Webinar Logistics. Now What Do I Need To Do? Blood Pressure in Kids January NC Department of Health and Human Services Division of Public Health NC Department of Health and Human Services Division of Public Health Now What Do I Need To Do? Blood Pressure in Kids January 2018 Gerri Mattson, MD, MSPH, FAAP Pediatric Medical Consultant Children and

More information

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018 WAVES CATCH A WAVE.. W I S C O N S I N P A R A M E D I C S E M I N A R A P R I L 2 0 1 8 K E R I W Y D N E R K R A U S E R N, C C R N, E M T - P Have you considered that if you don't make waves, nobody

More information

GUIDELINES FOR THE MEASUREMENT OF ANKLE BRACHIAL PRESSURE INDEX USING DOPPLER ULTRASOUND

GUIDELINES FOR THE MEASUREMENT OF ANKLE BRACHIAL PRESSURE INDEX USING DOPPLER ULTRASOUND GUIDELINES FOR THE MEASUREMENT OF ANKLE BRACHIAL PRESSURE INDEX USING DOPPLER ULTRASOUND AIM To provide evidence based principles for the measurement of Ankle Brachial Pressure Index (ABPI) using a BACKGROUND/EVIDENCE

More information