Contrast Enhancement in Cardiovascular MDCT: Effect of Body Weight, Height, Body Surface Area, Body Mass Index, and Obesity

Size: px
Start display at page:

Download "Contrast Enhancement in Cardiovascular MDCT: Effect of Body Weight, Height, Body Surface Area, Body Mass Index, and Obesity"

Transcription

1 Vascular Imaging Original Research Bae et al. Contrast Enhancement in Cardiovascular MDCT Vascular Imaging Original Research Kyongtae T. Bae 1 Brian A. Seeck 2 Charles F. Hildebolt 3 Cheng Tao 1 Fang Zhu 1 Masayuki Kanematsu 4 Pamela K. Woodard 3 Bae KT, Seeck BA, Hildebolt CF, et al. Keywords: aorta, cardiac CT, CT angiography, vascular imaging DOI: /AJR Received June 21, 7; accepted after revision September 26, 7. 1 Department of Radiology, University of Pittsburgh School of Medicine, Lothrop St., Ste. 4895, Pittsburgh, PA Address correspondence to K. T. Bae (baek@upmc.edu). 2 Division of Cardiology, Washington University School of Medicine, St. Louis, MO. 3 Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis, MO. 4 Department of Radiology, Gifu University School of Medicine, Gifu, Japan. AJR 8; 190: X/08/ American Roentgen Ray Society Contrast Enhancement in Cardiovascular MDCT: Effect of Body Weight, Height, Body Surface Area, Body Mass Index, and Obesity OBJECTIVE. The purpose of our study was to evaluate the effect of body weight, height, body surface area (BSA), body mass index (BMI), and obesity on aortic contrast enhancement in cardiac MDCT. MATERIALS AND METHODS. Seventy-three consecutive patients underwent cardiac CT angiography on a 64-MDCT scanner. Seventy-five ml of contrast medium ( mg I/ ml) was injected at 4.5 ml/s, followed by a 40-mL saline flush at 4.5 ml/s. The scanning delay of CT was determined with a bolus tracking technique. Aortic attenuation was measured over the aortic-root lumen. BMI and BSA were calculated from the patient s body weight and height. The patients were divided into low- (BMI < 30) and high- ( 30) BMI groups. Associations of aortic attenuation with body weight, height, BMI, and BSA were evaluated with regression analysis and the Student s t test. RESULTS. Strong inverse correlations were seen between aortic attenuation and body weight (r = 0.73), height (r = 0.47), BMI (r = 0.63), and BSA (r = 0.74) (p < for all). The regression formula of aortic attenuation versus body weight suggests that 1.0 ml/kg of contrast medium would yield a mean aortic attenuation of 355 H. The mean aortic attenuation was significantly higher in the low-bmi (352.6 ± 59.1 H) than in the high-bmi (286.2 ± 55.5 H) group. The regression formula for aortic attenuation on body weight was aortic attenuation = body weight (p < 0.001) for the low-bmi group and aortic attenuation = body weight (p < 0.001) for the high-bmi group, suggesting that the amount of contrast medium required with increased body weight is less in the high-bmi group. This group difference was less pronounced for the regression of aortic attenuation on BSA. CONCLUSION. To achieve a consistent contrast enhancement in cardiac CT angiography (CTA), contrast-medium dose should be adjusted with the body weight or the BSA (which accounts for both the body weight and height factors) to provide adjustment of iodine dose over a wide range of body sizes. M DCT angiography is increasingly used for the diagnostic evaluation of patients with clinically suspected obstructive coronary artery disease or abnormal cardiac or coronary anatomy. Advances in CT and ECGgating technology allow us to acquire highresolution, motion-free images of the heart and coronary arteries during a single short breath-hold [1, 2]. Markedly improved temporal resolution of CT has contributed to reducing motion artifacts and improving contrast enhancement and image quality [3 6]. Fast CT such as 64-MDCT allows us to obtain an entire scan of the coronary artery and heart in 10 seconds or less. As a result, scan timing becomes far more critical and challenging than with older and slower CT scanners [2, 7 9]. The other consideration with this challenge is that the short scanning time of MDCT may provide us an opportunity to improve contrast enhancement and to use the contrast medium more efficiently [9, 10]. With MDCT, the amount of contrast medium injected during some clinical applications may be reduced without decreasing contrast enhancement. Diverse protocols have been used for contrast administration and scan timing in cardiac and coronary CT angiography [8, 11 28]. Contrast-medium volumes ranging from 45 to 160 ml and injection rates ranging 2.5 to 5 ml/s have been used. This diversity in contrast-medium administration protocols reflects the rapid technical evolution of cardiac CT technology. As scanning duration shortens AJR:190, March 8 777

2 Bae et al. from 30 to 40 seconds with 4-MDCT to 6 12 seconds with 64-MDCT, there is a trend for using less contrast medium typically, ml with fast MDCT for an averagesize adult [13, 15, 18]. As CT technology evolves, contrast medium injection protocols must be adjusted and optimized. A patient s body weight and the amount of contrast medium are closely related to the degree of contrast enhancement [10, 29 36]. When consistent contrast enhancement is desired, the amount of administered iodine should be adjusted according to the patient s body weight. A large patient needs more iodine than a small patient to achieve the same magnitude of enhancement. One simple scheme for adjusting the amount of iodine mass with the body weight is to use a simple linear scale for example, to double the iodine mass when the patient s body weight doubles. However, this simple body-weight-based linearity may not provide an accurate estimate of the required contrast medium dose, particularly in obese patients who have a lot of body fat that is not metabolically active and thus contributes little to dispersing or diluting the contrast medium in the blood. Linear-weightbased dosing may overestimate the amount of contrast medium needed in these patients. Other body size parameters such as height, body mass index (BMI), and body surface area (BSA) should be considered for optimal dosing of contrast medium. To our knowledge, however, the association of various body size parameters with the degree of contrast enhancement for a given iodine dose has not been previously addressed. Therefore, the purpose of our study was to evaluate the effect of body weight, height, BMI, BSA, and obesity on aortic contrast enhancement in 64-MDCT cardiac angiography. Materials and Methods Patients This study was approved by our institutional internal review board, and informed consent was obtained from each patient. Patients who had prior coronary artery bypass surgery, valve replacement surgery, atrial fibrillation, or contraindications to contrast-enhanced CT were excluded. The study group consisted of 73 consecutive patients (41 men and 32 women; mean age, 59.4 ± 11 [SD] years; age range, years) who were scheduled to undergo catheter-based conventional coronary angiography and who agreed to participate in coronary CT angiography as a part of the research protocol. The patients body weight and height were recorded. CT All CT was performed on a 64-MDCT scanner (Sensation 64, Siemens Medical Solutions). The heart rate was measured in each patient before the CT examination. If the heart rate was higher than 65 beats per minute, IV β-blockers were administered to lower the heart rate. The scanning protocol consisted of 120 kvp, effective mas, tube current modulation, mm section collimation (z-flying focal spot), 330-millisecond gantry rotation time, and retrospective ECG gating. The images were reconstructed with a standard soft-tissue-kernel algorithm at 0.6-mm section thickness (without overlap) at every 10% of the R-R intervals. Contrast enhancement was achieved by administering mg I/mL contrast medium (Optiray [ioversol], Tyco Health/Mallinckrodt) into an antecubital vein through a 20-gauge angiocatheter. A fixed volume of 75 ml of con trast medium followed by a 40-mL saline flush was continuously injected at an injection rate of 4.5 ml/s for all patients. The scanning delay for diagnostic CT was determined using a bolus tracking method: A circular region of interest of 5 10 mm in diameter was placed over the ascend ing aorta, and diagnostic CT was started auto matically 5 seconds after the contrast enhancement exceeded a predefined threshold of 100 H. Aortic Attenuation Measurement The diagnostic cardiac CT images were retrieved from the institutional PACS and sent to a clinical workstation (Leonardo, Siemens Medical Solutions). Multiplanar reformatted images with 10-mm transverse section thickness were generated from the CT images reconstructed at 70% of the R-R interval of each cardiac cycle. Mean CT values (H) of the ascending aorta on these thick, transverse CT images were measured by a radiologist who had 12 years of experience in interpreting chest CT images. A 1.0-cm 2 circular region of interest was used to make measurements. Body Mass Index The BMI is defined as the weight in kilograms divided by the square of the height in meters. Although the BMI calculation does not take into account factors such as frame size and body tissue compositions, BMI categories are generally used as a means of estimating adiposity and assessing how much an individual s body weight departs from what is normal or desirable for a person of the same height. A BMI greater than or equal to 30 is generally considered to indicate obesity [37]; thus, in our study, for the evaluation of the effect of BMI, patients were divided into two groups for analysis: low BMI (BMI < 30) and high BMI (BMI 30). Body Surface Area BSA is a commonly used index in clinical practice to correct for patient size differences in various physiologic measurements and in calculating drug dosage. BSA is a better indicator of metabolic mass than body weight because it is less associated with excessive body fat. Various formulas have been proposed to estimate the BSA from a patient s weight and height, and these formulas result in slightly different values [38 43]. The most commonly used formula in day-to-day clinical practice is the Mosteller formula [42]: BSA (m 2 ) = (square root of product of weight [kg] height [cm]) / 60. This formula is simplified from a formula produced by Gehan and George [40] and has become a common standard because it is easy to memorize, and its use requires only a handheld calculator. More recently, however, a new formula was developed by Livingston and Lee [43] to relate BSA to body weight alone without using height data: BSA (m 2 ) = weight (kg) to the power of This scaling formula was reported to provide a more accurate estimate of the BSA in obese patients. Both the Mosteller and the scaling formulas were used in our study to calculate the BSA for each patient. Statistical Analysis The patient data for the following variables were fitted with normal distributions and tested for normality using the Shapiro-Wilk W tests: patient age, body weight, height, BMI, BSA, and aortic attenuation. Aortic attenuation values and ages for low- and high-bmi groups were also tested for normality and for equality of variances using the O Brien, Brown-Forsythe, Levene, and Bartlett tests. Differences in sex between the lowand high-bmi groups were tested using the Fisher s exact test. Means and 95% CIs were calculated for variables. To evaluate the effect of the patient s body size parameters on contrast enhancement, we performed linear regression analyses between aortic attenuation versus each of the following: body weight, height, BMI, and BSA. Ninety-five percent CIs were plotted for the regression lines. We used the Pearson product moment correlation coefficient (r) to assess the strengths of associations involving normal data distributions and the Spearman s rank (ρ) correlation coefficient to assess strengths of associations involving nonnormal data distributions. In addition, linear regression analyses were performed separ ately for the low- and high-bmi groups and 95% CIs were calculated for the resulting regression slopes. Differences in aortic attenuation between the lowand high-bmi groups were tested for statistical significance using the Student s t test. Alpha was set at Statistical analyses were performed with 778 AJR:190, March 8

3 Contrast Enhancement in Cardiovascular MDCT patients were in the high-bmi group (BMI 30). The mean aortic attenuation was ± 66.1 H (range, H). Cardiac CT images from three patients (small, medium, and large body size) are presented in Figure 1 to illustrate the different degrees of aortic attenuation. JMP Statistical Software (version 6, SAS Institute) and StatXact 7 Statistical Software for Exact Nonparametric Inference (Cytel). Results Data Distributions: Age, Weight, Height, BMI, BSA, and Aortic Attenuation Data were normally distributed for age, height, BSA, and aortic attenuation (Shapiro Wilk W test, p > 0.05). Body weight and BMI were not normally distributed (Shapiro Wilk W test, p < 0.05). For the low- and high-bmi groups, aortic attenuation and body weight were normally distributed (Shapiro Wilk W test, p > 0.05). For the lowand high-bmi groups, aortic attenuation and body weight had equal variances (O Brien, Brown-Forsythe, Levene, and Bartlett tests, p > 0.05). No significant difference was seen in the patient age and sex between the lowand high-bmi groups (p 0.34). A wide range of patient ages ( years), weights ( kg; mean, 88.5 ± 22.0 kg), heights ( cm; mean, ± 9.5 cm), BMI ( ; mean, 30.2 ± 6.1), and BSA ( m 2 ; mean, 2.04 ± 0.29m 2 ) were observed. Forty-three patients were in the low-bmi group (BMI < 30) and 30 A C Fig. 1 Transverse CT images are shown for patients having three body sizes in whom different degrees of aortic attenuation were seen (image display window width, 800 H; center, H). A, 59-year-old woman weighing 55.8 kg and having body mass index (BMI) of 21.1 and body surface area (BSA) of 1.59 m 2 (aortic attenuation, 469 H). B, 61-year-old man weighing 91.4 kg and having BMI of 30.6 and BSA of 2.09 m 2 (aortic attenuation, 379 H). C, 59-year-old man weighing kg and having BMI of 42.4 and BSA of 2.85 m 2 (aortic attenuation, 172 H) B Effect of Body Weight, Height, BMI, and BSA on Aortic Attenuation Strong inverse correlation was noted between the aortic attenuation and body weight (r = 0.73, ρ = 0.74, p < 0.001), indicating reduced aortic attenuation in heavier patients (Fig. 2). The regression formula (aortic attenuation [H] = weight [kg]) suggests that 1.0 ml/kg (e.g., 75 ml for a 75-kg patient) of mg I/mL contrast medium would yield the aortic attenuation of 355 H. A moderately strong inverse correlation was noted between aortic attenuation and height (r = 0.47, p < 0.001), which indicates reduced aortic attenuation in taller patients (Fig. 3). The regression formula (aortic attenuation [H] = height [cm]) (p < 0.001) suggests that, for each 10-cm increase in height, there is a decrease in enhancement of 33 H in the aortic attenuation. A relatively strong inverse correlation existed between aortic attenuation and the BMI (r = 0.63, ρ = 0.64, p < 0.001). This indicates reduced aortic attenuation in patients with higher BMI (Fig. 4). The regression formula was aortic attenuation (H) = BMI (p < 0.001). The strongest inverse correlation (r = 0.74, p < 0.001) existed between the aortic attenuation Weight (kg) Fig. 2 Plot of aortic attenuation (H) versus body weight (kg). Strong inverse correlation existed between aortic attenuation and body weight (r = 0.73, ρ = 0.74, p < 0.001). This indicates aortic attenuation is reduced in heavier patients. Regression formula (aortic attenuation [H] = weight [kg]) indicates that 1.0 ml/kg (e.g., 75 ml for a 75-kg patient) of mg I/mL contrast medium is required to achieve aortic attenuation of 355 H. + = low-body mass index (BMI) group (BMI < 30), o = high-bmi group (BMI 30). Ninety-five percent CIs (dotted fitting lines) are fit to regression line (solid fitting line). AJR:190, March 8 779

4 Bae et al. Fig. 3 Plot of aortic attenuation (H) versus height (cm). Moderately strong inverse correlation existed between aortic attenuation and body weight (r = 0.47, p < 0.001). This indicates aortic attenuation is reduced in taller patients. Regression formula (aortic attenuation [H] = height [cm]) indicates that, for each 10-cm increase in height, there is decrease in enhancement of 33 H in aortic attenuation. + = low-body mass index (BMI) group (BMI < 30), o = high-bmi group (BMI 30). Ninety-five percent CIs (dotted fitting lines) are fit to regression line (solid fitting line). and the BSA (estimated with the Mosteller formula). This indicated reduced aortic attenuation with increased BSA (Fig. 5). The regression analysis was repeated with the BSA values, which were estimated using the scaling formula [43] from the body weight data only. The computed r value was nearly identical. The regression formula was aortic attenuation (H) = BSA (m 2 ) (p < 0.001). This formula implies that, with our study injection protocol, the aortic attenuation of 382 H would be achieved for a patient with an average adult BSA of 1.7 m 2. The BSA that corresponds to Height (cm) the aortic attenuation of 355 H (i.e., the aortic attenuation achieved with 1.0 ml/kg of the weight-based dose) is 1.86 m 2. Comparison Between the Low- and High-BMI Groups The mean aortic attenuation for the low- BMI group (352.6 ± 59.1 H) was significantly higher (p < 0.001) than that for the high-bmi group (286.2 ± 55.5 H) (Fig. 6). This trend was anticipated from the strong inverse correlation observed in the aortic attenuation versus BMI, as shown in Figure Body Mass Index Fig. 4 Plot of aortic attenuation (H) versus body mass index (BMI). Relatively strong inverse correlation existed between aortic attenuation and BMI (r = 0.63, ρ = 0.64, p < 0.001). This indicates aortic attenuation is reduced in patients with higher BMI. Regression formula was aortic attenuation (H) = BMI (p < 0.001). + = low-bmi group (BMI < 30), o = high-bmi group (BMI 30). Ninety-five percent CIs (dotted fitting lines) are fit to regression line (solid fitting line). For the low-bmi group, the regression formula for aortic attenuation on body weight was aortic attenuation (H) = weight (kg) (r = 0.60, p < 0.001); for the high-bmi group, the regression formula was aortic attenuation (H) = weight (kg) (r = 0.69, p < 0.001) (Fig. 7). Note that the regression slope (H/kg) of the high-bmi group (1.9 [ , 95% CI]) was less steep than that for the low-bmi group (3.1 [ ]), but the 95% CIs overlap. The lower slope in the high-bmi group suggests a trend for a slower decline of contrast enhancement with increased body weight and for less contrast medium to be required per kilogram of body weight to maintain contrast enhancement. The regression formula for aortic attenuation on BSA was aortic attenuation (H) = BSA (m 2 ) (r = 0.58, p < 0.001) for the low- BMI and aortic attenuation (H) = BSA (m 2 ) (r = 0.71, p < 0.001) for the high- BMI group (Fig. 8). Note that the regression slope (H/m 2 ) of the high-bmi group (151) was less steep than that for the low-bmi group (187), just as was the trend observed in the regression of the aortic attenuation on body weight. However, the relative ratio of the slopes for the high- to low-bmi groups for the BSA regression is smaller (151/187 [81%]) than that for the body weight regression (1.9/3.1 [61%]). This suggests that BSA may serve better than body weight for the adjustment of iodine dose over a wide range of body sizes. Discussion The magnitude of contrast enhancement for CTA depends on a number of patientrelated and injection-related factors, including body weight and cardiac output, contrast medium volume and concentration, injection rate, type of contrast medium, and saline flush [10]. The most important patient-related factor affecting the magnitude of vascular and parenchymal contrast enhancement is body weight [30, 34, 35, 44]. The significance of body weight on contrast enhancement was confirmed by our study, which showed a strong trend of declining aortic attenuation proportional to patient weight for a given contrast medium administration protocol. To achieve a consistent degree of contrast enhancement, the amount of administered contrast medium should be adjusted according to the patient s body weight. To devise a contrast medium dosing protocol, we first must determine the desirable level of contrast enhancement for aortic and coronary CT angiography. We consider attenuations of H 780 AJR:190, March 8

5 Contrast Enhancement in Cardiovascular MDCT Fig. 5 Plot of aortic attenuation (H) versus body surface area (BSA) (m 2 ). Strongest inverse correlation (r = 0.74, p < 0.001) existed between aortic attenuation and BSA (estimated with Mosteller formula [42]). This indicates aortic attenuation is reduced as BSA increases. Regression formula was aortic attenuation (H) = BSA (m 2 ) (p < 0.001). + = low body mass index (BMI) group (BMI < 30), o = high-bmi group (BMI 30), Ninety-five percent CIs (dotted fitting lines) are fit to regression line (solid fitting line). (i.e., H net contrast enhancement) to be diagnostically adequate for aortic and coronary CT angiography [17]. Although slightly lower attenuations ( H) were advocated for coronary CT angiography to differentiate the enhanced lumen from calcification (> H) [24], higher vascular attenuations (> H) were reported as necessary to improve visualization of small coronary arteries and stenoses [17]. On the basis of our results, we think that an aortic attenuation of 355 H (i.e., approximately H net contrast enhancement) would be achieved with 1.0 ml/kg Body Surface Area (m 2 ) of mg I/mL contrast medium injected at 4.5 ml/s (i.e., 0.35 g I/kg of contrast medium injected at 1.6 g I/s), followed by a saline flush. Using this weight-based protocol, we are be able to estimate the amount of required iodine loads for large patients (i.e., use of higher volumes or longer injections) or for small patients to generate equivalent degrees of contrast enhancement. Alternatively, we can modify iodine delivery rates (i.e., use different injections or different iodine concentrations in contrast mediums) to achieve different degrees of contrast enhancement. < Body Mass Index Fig. 6 Ninety-five percent mean diamond plots of aortic attenuation (H) for low- and high-body mass index (BMI) groups. Mean aortic attenuation of low-bmi group (352.6 ± 59.1 H) was significantly higher (p < 0.001) than that of high-bmi group (286.2 ± 55.5 H). + = low-bmi group (BMI < 30), o = high-bmi group (BMI 30). Horizontal line is grand mean. Heights of diamonds represent 95% CIs and widths of diamonds are proportional to sample sizes. Although numerous studies have been conducted to investigate the effect of body weight on contrast enhancement [9, 30, 31, 34, 35, 45 48], the effect of the patient s height on contrast enhancement has rarely been studied. Our study revealed a moderately strong inverse correlation between aortic attenuation and height. This indicates reduced aortic attenuation in taller patients. Although the correlation is weaker than that for the body weight, this inverse relationship is expected because height and weight are proportional (i.e., taller people are generally heavier than shorter people). This is supported by a recent study that showed that for weights less than 80 kg, height and weight are highly correlated and both increase proportionally: height (cm) = weight (kg) to the power of (r = 0.99) [43]. This study also reported that the height weight relationship is complex because beyond 80 kg, weight increases without any significant increase in height, as commonly seen in patients with obesity. Considering this nonlinear relationship and individual height variation in adult patients being less than body weight variation, it is not surprising that the correlation of aortic attenuation with height was weaker than that with body weight. Nevertheless, to precisely determine the required volume of contrast medium for a consistent enhancement, not only body weight but also height and obesity should be taken into consideration. Although BMI is commonly used because it is easy to calculate and is strongly correlated with total body fat content in adults, its accuracy in relation to actual levels of body fat is easily distorted by such factors as fitness level, muscle mass, bone structure, sex, and ethnicity. Muscular people have high BMIs because muscle is denser than fat. Despite its limitations, the BMI has been widely applied for the identification, evaluation, and treatment of obesity in adults [37]. Our study result show a relatively strong inverse correlation between aortic attenuation and BMI. The mean aortic attenuation was significantly lower in high- rather than low-bmi patients. These findings intuitively make sense because contrast enhancement is expected to be lower in obese patients than in nonobese patients. However, the findings are less obvious from the definition of BMI (i.e., weight divided by the square of height) because both weight and height showed an inverse correlation with aortic attenuation. Our explanation for this is that although the square of the height constitutes the denominator for the calculation of the BMI, the contribution of AJR:190, March 8 781

6 Bae et al Fig. 7 Plot of aortic attenuation (H) versus body weight (kg) with bivariate fits for low- and high-body mass index (BMI) groups. Regression formula of aortic attenuation versus body weight was aortic attenuation (H) = weight (kg) (p < 0.001) for low-bmi group (+, solid fitting line) and aortic attenuation (H) = weight (kg) (p < 0.001) for high-bmi group (o, dotted fitting line). Regression slope (H/kg) of high-bmi group (1.9 [ , 95% CI]) was less steep than that for low-bmi group (3.1 [ ]), but 95% CIs overlap. So that regression lines can be better seen, 95% CIs are not included in figure. weight is greater than that of height. This is because height is proportional to the power of weight for weights < 80 kg [43]. The BMI still has a net positive correlation with body weight and thus has an inverse correlation with the aortic attenuation. The correlation of aortic attenuation with BSA was higher than those with body weight, height, and BMI. Although the relationship between contrast enhancement and BSA has not been addressed in previous studies, this strong relationship is not surprising because Weight (kg) for many physiologic and clinical purposes, BSA has been used as a better indicator of metabolic mass than body weight or other indicators. In our study, we found no difference between the BSA values estimated with the Mosteller [42] and the scaling [43] formulas for the correlation with aortic attenuation. When we consider contrast medium as an IV drug that distributes and disperses pharmacokinetically throughout the body, it is conceivable that contrast medium could be dosed according to the patient s BSA rather Body Surface Area (m 2 ) Fig. 8 Plot of aortic attenuation (H) versus body surface area (BSA) (m 2 ) with bivariate fit of low- and high- body mass index (BMI) groups. Regression formulas of aortic attenuation versus BSA were aortic attenuation (H) = BSA (m 2 ) (p < 0.001) for low-bmi group (+, solid fitting line) and aortic attenuation (H) = BSA (m 2 ) (p < 0.001) for high-bmi group (o, dotted fitting line). Note that these two regression lines are more closely approximated than two regression lines in Figure 7. This suggests that discrepancy between high- and low-bmi groups in decline rate of aortic attenuation was less pronounced with BSA than with body weight. than other body size parameters. With a direct 1:1 weight-based linear proportionality, contrast medium may be overdosed for obese patients and underdosed in pediatric patients. Support for a BSA-based contrast medium dosing scheme is as follows. In our study, the aortic attenuation of 355 H was obtained with 75 ml of contrast medium for a patient with a BSA of 1.86 m 2. This 75 ml of contrast medium normalized by the BSA corresponds to 40.3 ml/m 2. In our study population, BSA ranged from 1.45 to 2.85 m 2. When these are multiplied by 40.3 ml/m 2, the volumes of contrast medium corresponding to the aortic attenuation of 355 H equal 58.7 ml for the minimum BSA and ml for the maximum BSA. This estimation based on BSA can be measured against that based on body weight as follows: It is assumed that through linear proportionality, a weight-based dosing scheme can be used to estimate the amount of contrast volume required to achieve a certain degree of enhancement (i.e., 1.0 ml/kg of mg I/mL contrast medium results in an aortic attenuation of 355 H). In our study population, the body weight ranged from 50 to 154 kg; thus, the estimated volumes of contrast medium would be 50 and 154 ml. When these are compared with the volumes estimated with BSA, the lightest patient would be underdosed by 8.7 ml, and the heaviest patient overdosed by 39 ml. Instead of linear proportionality, because the BSA is directly related to weight (kg) to the power of according to the scaling formula [43], weight scaling can be used as a proportionality to estimate the contrast medium dose. If 75 kg is used as the reference weight and 1.0 ml/(kg) to the power of as the proportionality, the amount of contrast medium required for a 50-kg patient is 57.5 ml (i.e., 75 ml [50 / 75] to the power of ) and that required for a 154-kg patient is ml (i.e., 75 ml [154 / 75] to the power of ). Not surprisingly, these values are similar to the 58.7 and ml that were estimated from the BSA. For obese patients, the relationship between iodine dose and the degree of contrast enhancement is not well established. An obese patient may have a high proportion of body fat and thus a relatively small blood volume and a small, well-perfused extracellular compartment. As a result, in obese patients, if the amount of iodine dose is estimated and increased linearly in proportion to body weight, the resulting contrast enhancement may be 782 AJR:190, March 8

7 Contrast Enhancement in Cardiovascular MDCT higher than that of nonobese patients, who receive the iodine dose with the same linear body weight proportionality. We propose a contrast medium dosing scheme for which a reference dose is used for a standard body size to achieve a clinically desirable degree of contrast enhancement, and subsequent contrast-medium doses are adjusted using the two-thirds power of the weight, as suggested in the BSA scaling formula [43]. This scaling rule of adjusting contrast medium dose with the two-thirds power of the weight is applicable not only across a wide range of human body sizes (from infants to a markedly obese adults) but also across species ranging in size from rats to cows [49]. However, further clinical studies are necessary to validate this. Because the blood volume per body weight is relatively smaller in obese patients, the rate of declining aortic attenuation with body weight in obese patients (BMI 30) is expected to be slower than that in nonobese patients (BMI < 30). This was confirmed in our results that the regression slope of the high-bmi patients was less than that for the low-bmi patients. In other words, for patients with high BMIs compared with those with low BMIs, a smaller decrease in contrast enhancement occurred with increased body weight (i.e., less contrast medium was required per body weight to maintain contrast enhancement). Furthermore, we observed that, with BSA instead of weight as the regression variable, the regression slope discrepancy between the high- and low-bmi patients was less pronounced. This implies that BSA is a better parameter than body weight for estimating contrast medium dose for both the high- and low-bmi patients. Our study had some limitations. First, it was a retrospective study and was not specifically designed to test the clinical impact of different contrast enhancements on the accuracy of diagnosing coronary artery disease or stenosis. No follow-up or further clinical evaluation was performed to investigate the extent to which the image (diagnostic quality) was associated with the degree of contrast enhancement. Second, we used BMI to divide the patients into obese and nonobese groups. Although the BMI is easy to use and readily calculated from the weight and height data, it is limited in characterizing obesity because it does not distinguish between muscle and fat. Percentage of body fat is a better measure to quantify adipose tissue; however, it requires techniques such as skin-fold measurements and specialized body fat scales, which were not available in our study. Third, although aortic attenuation is affected by cardiovascular function and cardiac output [10, 50], this was not specifically investigated in our study. We recognize that because of variations in cardiac output among patients, it is critical to individualize the scanning delay for CT angiography. We used a bolus tracking method to determine the scanning delay. We think that the effect of cardiac output variation on aortic attenuation is minimized by means of an individualized scanning delay and use of a fixed injection duration. Also, the attenuation value we measured does not necessarily represent the peak attenuation value but rather attenuation values obtained with diagnostic scans using a specific injection protocol. In our clinical diagnostic scan of 5- to 7-seconds duration, we could neither explore the temporal course of the enhancement nor determine peak enhancement. Fourth, our CT attenuation measurements were limited to the ascending aorta. Although the attenuation or contrast enhancement of the coronary arteries is clinically valuable, we think that the ascending-aortic attenuation measurement is highly reliable and representative of the coronary artery attenuation measurements. The measurement of CT attenuation in small arteries is technically difficult and highly variable and is affected by the severity of vascular disease. In conclusion, on the basis of our specific contrast injection protocol and study results, we found that the degree of aortic contrast enhancement in MDCT angiography declines with increases in body weight, height, body mass index, and body surface area. This finding may be interpreted and used in clinical practice as follows: To achieve a consistent aortic and coronary artery enhancement, the amount of contrast medium should be adjusted according to the patient s body weight or body surface area. Body surface area accounts for both the body weight and height factors and perhaps is more appropriate for obese patients because a 1:1 direct increase in contrast volume with body weight may overestimate the contrast medium dose needed in obese patients. We propose that one approach to adjust the amount of administered contrast medium with respect to the patient s body surface area is to estimate the required amount of contrast medium proportional to the two-thirds power of the body weight. This proposed method would require further prospective investigative work for its validation. References 1. Ohnesorge BM, Hofmann LK, Flohr TG, Schoepf UJ. CT for imaging coronary artery disease: defining the paradigm for its application. Int J Cardiovasc Imaging 5; 21: Schoepf UJ, Becker CR, Ohnesorge BM, Yucel EK. CT of coronary artery disease. Radiology 4; 232: Achenbach S, Ropers D, Kuettner A, et al. Contrast-enhanced coronary artery visualization by dual-source computed tomography: initial experience. Eur J Radiol 6; 57: Hoffmann MH, Lessick J. Multidetector-row computed tomography for noninvasive coronary imaging. Expert Rev Cardiovasc Ther 6; 4: Cury RC, Nieman K, Shapiro MD, Nasir K, Cury RC, Brady TJ. Comprehensive cardiac CT study: evaluation of coronary arteries, left ventricular function, and myocardial perfusion is it possible? J Nucl Cardiol 7; 14: Herzog C, Zangos S, Zwerner P, Costello P, Vogl TJ, Schoepf UJ. CT of coronary artery disease. J Thorac Imaging 7; 22: Bae KT. Test-bolus versus bolus-tracking techniques for CT angiographic timing. Radiology 5; 236: ; author reply Cademartiri F, Nieman K, van der Lugt A, et al. Intravenous contrast material administration at 16-detector row helical CT coronary angiography: test bolus versus bolus-tracking technique. Radiology 4; 233: Bae KT, Tao C, Gurel S, et al. Effect of patient weight and scanning duration on contrast enhancement during pulmonary multidetector CT angiography. Radiology 7; 242: Bae KT, Heiken JP. Scan and contrast administration principles for MDCT. Eur Radiol 5; 15[suppl 5]:E46 E Funabashi N, Suzuki K, Terao M, et al. New acquisition method to exclusively enhance the left side of the heart by a small amount of contrast material achieved by multislice computed tomography with 64 data acquisition system. Int J Cardiol 7; 114: Rist C, Nikolaou K, Kirchin MA, et al. Contrast bolus optimization for cardiac 16-slice computed tomography: comparison of contrast medium formulations containing and milligrams of iodine per milliliter. Invest Radiol 6; 41: Hamoir XL, Flohr T, Hamoir V, et al. Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time. Eur Radiol 5; 15: Haberl R, Tittus J, Bohme E, et al. Multislice spiral computed tomographic angiography of coronary arteries in patients with suspected coronary artery disease: an effective filter before catheter AJR:190, March 8 783

8 Bae et al. angiography? Am Heart J 5; 149: Leber AW, Knez A, von Ziegler F, et al. Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 5; 46: Datta J, White CS, Gilkeson RC, et al. Anomalous coronary arteries in adults: depiction at multi-detector row CT angiography. Radiology 5; 235: Cademartiri F, Mollet NR, van der Lugt A, et al. Intravenous contrast material administration at helical 16-detector row CT coronary angiography: effect of iodine concentration on vascular attenuation. Radiology 5; 236: Achenbach S, Ropers D, Pohle FK, et al. Detection of coronary artery stenoses using multi-detector CT with collimation and 375 ms rotation. Eur Heart J 5; 26: Nikolaou K, Knez A, Sagmeister S, et al. Assessment of myocardial infarctions using multidetector-row computed tomography. J Comput Assist Tomogr 4; 28: van Ooijen PM, Dorgelo J, Zijlstra F, Oudkerk M. Detection, visualization and evaluation of anomalous coronary anatomy on 16-slice multidetectorrow CT. Eur Radiol 4; 14: Kuettner A, Trabold T, Schroeder S, et al. Noninvasive detection of coronary lesions using 16- detector multislice spiral computed tomography technology: initial clinical results. J Am Coll Cardiol 4; 44: Kopp AF, Kuttner A, Trabold T, Heuschmid M, Schroder S, Claussen CD. Multislice CT in cardiac and coronary angiography. Br J Radiol 4; 77[spec no 1]:S87 S Hong C, Chrysant GS, Woodard PK, Bae KT. Coronary artery stent patency assessed with instent contrast enhancement measured at multidetector row CT angiography: initial experience. Radiology 4; 233: Becker CR, Hong C, Knez A, et al. Optimal contrast application for cardiac 4-detector-row computed tomography. Invest Radiol 3; 38: Pannu HK, Flohr TG, Corl FM, Fishman EK. Current concepts in multi-detector row CT evaluation of the coronary arteries: principles, techniques, and anatomy. RadioGraphics 3; 23[spec no]:s111 S Vogl TJ, Abolmaali ND, Diebold T, et al. Techniques for the detection of coronary atherosclerosis: multi-detector row CT coronary angiography. Radiology 2; 223: Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PM, de Feyter PJ. Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 2; 106: Kopp AF, Schroeder S, Kuettner A, et al. Non-invasive coronary angiography with high resolution multidetector-row computed tomography: results in 102 patients. Eur Heart J 2; 23: Dean PB, Violante MR, Mahoney JA. Hepatic CT contrast enhancement: effect of dose, duration of infusion, and time elapsed following infusion. Invest Radiol 1980; 15: Kormano M, Partanen K, Soimakallio S, Kivimaki T. Dynamic contrast enhancement of the upper abdomen: effect of contrast medium and body weight. Invest Radiol 1983; 18: Berland LL, Lee JY. Comparison of contrast media injection rates and volumes for hepatic dynamic incremented computed tomography. Invest Radiol 1988; 23: Chambers TP, Baron RL, Lush RM. Hepatic CT enhancement. Part I. Alterations in the volume of contrast material within the same patients. Radiology 1994; 193: Small WC, Nelson RC, Bernardino ME, Brummer LT. Contrast-enhanced spiral CT of the liver: effect of different amounts and injection rates of contrast material on early contrast enhancement. AJR 1994; 163: Heiken JP, Brink JA, McClennan BL, Sagel SS, Crowe TM, Gaines MV. Dynamic incremental CT: effect of volume and concentration of contrast material and patient weight on hepatic enhancement. Radiology 1995; 195: Platt JF, Reige KA, Ellis JH. Aortic enhancement during abdominal CT angiography: correlation with test injections, flow rates, and patient demographics. AJR 1999; 172: Bae KT. Peak contrast enhancement in CT and MR angiography: when does it occur and why? Pharmacokinetic study in a porcine model. Radiology 3; 227: NHLBI Obesity Education Initiative Expert Panel. Clinical guidelines on the identification, evaluation, and treatment of overweight and obesity in adults: the evidence report. Bethesda, MD: National Institutes of Health, National Heart, Lung, and Blood Institute, DuBois D, DuBois EF. A formula to estimate the approximate surface area if height and weight are known. Arch Intern Med 1916; 17: Boyd E. The growth of the surface area of the human body. Minneapolis, MN: University of Minnesota, Gehan EA, George SL. Estimation of human body surface area from height and weight. Cancer Chemother Rep 1970; 54: Haycock GB, Schwartz GJ, Wisotsky DH. Geometric method for measuring body surface area: a height weight formula validated in infants, children, and adults. J Pediatr 1978; 93: Mosteller RD. Simplified calculation of bodysurface area. N Engl J Med 1987; 317: Livingston EH, Lee S. Body surface area prediction in normal-weight and obese patients. Am J Physiol Endocrinol Metab 1; 281: E586 E Bae KT. Technical aspects of contrast delivery in advanced CT. Appl Radiol 3; 32[suppl]: Bae KT, Heiken JP, Brink JA. Aortic and hepatic contrast medium enhancement at CT. Part I. Prediction with a computer model. Radiology 1998; 207: Yamashita Y, Komohara Y, Takahashi M, et al. Abdominal helical CT: evaluation of optimal doses of intravenous contrast material prospective randomized study. Radiology 0; 216: Takeshita K. Prediction of maximum hepatic enhancement on computed tomography from dose of contrast material and patient weight: proposal of a new formula and evaluation of its accuracy. Radiat Med 1; 19: Awai K, Hori S. Effect of contrast injection protocol with dose tailored to patient weight and fixed injection duration on aortic and hepatic enhancement at multidetector-row helical CT. Eur Radiol 3; 13: Dawson TH. Engineering design of the cardiovascular systems of mammals. Englewood Cliffs, NJ: Prentice Hall, Bae KT, Heiken JP, Brink JA. Aortic and hepatic contrast medium enhancement at CT. Part II. Effect of reduced cardiac output in a porcine model. Radiology 1998; 207: AJR:190, March 8

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Coronary CT Angiography

Coronary CT Angiography Coronary CT Angiography Byoung Wook Choi, M.D. Department of Diagnostic Radiology Yonsei University College of Medicine, Severance Hospital E mail : bchoi@yumc.yonsei.ac.kr Abstract With the advent of

More information

Angio-CT: heart and coronary arteries

Angio-CT: heart and coronary arteries European Journal of Radiology 45 (2003) S32/S36 www.elsevier.com/locate/ejrad Angio-CT: heart and coronary arteries Andreas F. Kopp * Tübingen University Hospital, Tübingen, Germany Received 22 November

More information

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing Cademartiri et al. Heart Rhythm Irregularities on MDCT Angiography Cardiac Imaging Original Research A C M E D E N T U R I C A L I M A G I N G AJR 2006; 186:634 638 0361 803X/06/1863 634 American Roentgen

More information

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Poster No.: B-0742 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Paper

More information

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography Supalerk Pattanaprichakul, MD 1, Sutipong Jongjirasiri, MD 2, Sukit Yamwong, MD 1, Jiraporn Laothammatas,

More information

Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients

Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients (2006) 30, 569 573 & 2006 Nature Publishing Group All rights reserved 0307-0565/06 $30.00 www.nature.com/ijo ORIGINAL ARTICLE Image quality and diagnostic accuracy of 16-slice multidetector computed tomography

More information

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16.

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16. NIH Public Access Author Manuscript Published in final edited form as: Acad Radiol. 2008 January ; 15(1): 40 48. doi:10.1016/j.acra.2007.07.023. Analysis of Heart Rate and Heart Rate Variation During Cardiac

More information

ECG-Gated 16-MDCT of the Coronary Arteries: Assessment of Image Quality and Accuracy in Detecting Stenoses

ECG-Gated 16-MDCT of the Coronary Arteries: Assessment of Image Quality and Accuracy in Detecting Stenoses ECG-Gated 16-MDCT of the Coronary Arteries ECG-Gated 16-MDCT of the Coronary Arteries: Assessment of Image Quality and Accuracy in Detecting Stenoses Martin Heuschmid 1 Axel Kuettner 1 Stephen Schroeder

More information

Studies with electron beam computed tomography (EBCT) Imaging

Studies with electron beam computed tomography (EBCT) Imaging Imaging Predictive Value of 16-Slice Multidetector Spiral Computed Tomography to Detect Significant Obstructive Coronary Artery Disease in Patients at High Risk for Coronary Artery Disease Patient- Versus

More information

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT Eur Radiol (2006) 16: 1964 1972 DOI 10.1007/s00330-006-0262-x CARDIAC Sebastian Leschka Lars Husmann Lotus M. Desbiolles Oliver Gaemperli Tiziano Schepis Pascal Koepfli Thomas Boehm Borut Marincek Philipp

More information

Non-invasive Coronary Angiography: the Role, Limitations and Future of 64-Slice Spiral Computed Tomography Coronary Angiography

Non-invasive Coronary Angiography: the Role, Limitations and Future of 64-Slice Spiral Computed Tomography Coronary Angiography HOSPITAL CHRONICLES 2009, 4(3): 105 109 Review Non-invasive Coronary Angiography: the Role, Limitations and Future of 64-Slice Spiral Computed Tomography Coronary Angiography Arkadios C. Roussakis, MD

More information

Cardiac CT Techniques in Neonates (and infants)

Cardiac CT Techniques in Neonates (and infants) Cardiac CT Techniques in Neonates (and infants) Siddharth P. Jadhav, MD Director, Body CT and MRI Edward B. Singleton Department of Pediatric Radiology Texas Children s Hospital Disclosures None Objectives

More information

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography African Journal of Biotechnology Vol. 10(19), pp. 3854-3858, 9 May, 2011 Available online at http://www.academicjournals.org/ajb DOI: 10.5897/AJB10.983 ISSN 1684 5315 2011 Academic Journals Full Length

More information

Contrast material enhanced computed tomography (CT) is the most commonly used imaging modality for the detection and characterization of liver metasta

Contrast material enhanced computed tomography (CT) is the most commonly used imaging modality for the detection and characterization of liver metasta Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Hiroshi Kondo, MD

More information

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology,

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology, Chapter 4 Impact of Coronary Calcium Score on Diagnostic Accuracy of Multislice Computed Tomography Coronary Angiography for Detection of Coronary Artery Disease Gabija Pundziute, 1,3 Joanne D. Schuijf,

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6

The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6 The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6 n&list_uids=17433949 64-Multislice Detector Computed Tomography Coronary Angiography as Potential Alternative

More information

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography Journal of the American College of Cardiology Vol. 46, No. 3, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.05.056

More information

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications SYMPOSIA Indications, Patient Selection, and Clinical Implications Christian Thilo, MD,* Mark Auler, MD,* Peter Zwerner, MD,w Philip Costello, MD,* and U. Joseph Schoepf, MD* Abstract: Recent technical

More information

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography Journal of the American College of Cardiology Vol. 52, No. 18, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.07.048

More information

General Cardiovascular Magnetic Resonance Imaging

General Cardiovascular Magnetic Resonance Imaging 2 General Cardiovascular Magnetic Resonance Imaging 19 Peter G. Danias, Cardiovascular MRI: 150 Multiple-Choice Questions and Answers Humana Press 2008 20 Cardiovascular MRI: 150 Multiple-Choice Questions

More information

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T. The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume

More information

Accuracy of Multislice Computed Tomography in the Preoperative Assessment of Coronary Disease in Patients With Aortic Valve Stenosis

Accuracy of Multislice Computed Tomography in the Preoperative Assessment of Coronary Disease in Patients With Aortic Valve Stenosis Journal of the American College of Cardiology Vol. 47, No. 10, 2006 2006 by the American College of Cardiology Foundation ISSN 0735-1097/06/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.11.085

More information

CT angiography techniques. Boot camp

CT angiography techniques. Boot camp CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary

More information

Patient-Tailored Scan Delay for Multiphase Liver CT: Improved Scan Quality and Lesion Conspicuity With a Novel Timing Bolus Method

Patient-Tailored Scan Delay for Multiphase Liver CT: Improved Scan Quality and Lesion Conspicuity With a Novel Timing Bolus Method Gastrointestinal Imaging Original Research Schneider et al. Multiphase Liver CT Gastrointestinal Imaging Original Research J. Gabriel Schneider 1 Zhen J. Wang Wilbur Wang Judy Yee Yanjun Fu Benjamin M.

More information

Soft and Intermediate Plaques in Coronary Arteries: How Accurately Can We Measure CT Attenuation Using 64-MDCT?

Soft and Intermediate Plaques in Coronary Arteries: How Accurately Can We Measure CT Attenuation Using 64-MDCT? 64-MDCT Measurement of Coronary Artery Plaques Cardiac Imaging Original Research Jun Horiguchi 1 Chikako Fujioka 1 Masao Kiguchi 1 Yun Shen 2 Christian E. Althoff 3,4 Hideya Yamamoto 5 Katsuhide Ito 3

More information

Impact of 64-Slice Multidetector Computed Tomography on Other Diagnostic Studies for Coronary Artery Disease

Impact of 64-Slice Multidetector Computed Tomography on Other Diagnostic Studies for Coronary Artery Disease CLINICAL RESEARCH STUDY Impact of 64-Slice Multidetector Computed Tomography on Other Diagnostic Studies for Coronary Artery Disease Alex J. Auseon, DO, Sunil S. Advani, MD, Charles A. Bush, MD, Subha

More information

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760 Can Weight-Adapted IV Contrast Media Protocols Reduce Iodine Dose and Still Produce a Diagnostic Contrast Enhancement Level in Abdomino-Pelvic CT Scans? Poster No.: C-1760 Congress: ECR 2015 Type: Authors:

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

Cardiac Imaging Tests

Cardiac Imaging Tests Cardiac Imaging Tests http://www.medpagetoday.com/upload/2010/11/15/23347.jpg Standard imaging tests include echocardiography, chest x-ray, CT, MRI, and various radionuclide techniques. Standard CT and

More information

M Marwan, D Ropers, T Pflederer, W G Daniel, S Achenbach

M Marwan, D Ropers, T Pflederer, W G Daniel, S Achenbach Department of Cardiology, University of Erlangen, Erlangen, Germany Correspondence to: Dr M Marwan, Innere Medizin II, Ulmenweg 18, 91054 Erlangen, Germany; mohamed.marwan@ uk-erlangen.de Accepted 17 November

More information

Radiation dose of cardiac CT what is the evidence?

Radiation dose of cardiac CT what is the evidence? Eur Radiol (2009) 19: 1311 1315 DOI 10.1007/s00330-009-1312-y CARDIAC Hatem Alkadhi Radiation dose of cardiac CT what is the evidence? Received: 10 December 2008 Revised: 2 January 2009 Accepted: 12 January

More information

Clinical Medicine Insights: Cardiology

Clinical Medicine Insights: Cardiology Open Access: Full open access to this and thousands of other papers at http://www.la-press.com. Clinical Medicine Insights: Cardiology Supplementary Issue: Cardiovascular Imaging: Current Developments

More information

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Original Article Open Access The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Ziqiao Lei 1, Jin Gu 2, Qing Fu 3, Heshui Shi 4, Haibo Xu 5, Ping Han 6, Jianming

More information

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING Non-invasive coronary angiography along with multidetector computed tomography or magnetic resonance imaging is attracting increasing interest

More information

Coronary Calcium Screening Using Low-Dose Lung Cancer Screening: Effectiveness of MDCT with Retrospective Reconstruction

Coronary Calcium Screening Using Low-Dose Lung Cancer Screening: Effectiveness of MDCT with Retrospective Reconstruction Cardiac Imaging Original Research Kim et al. Coronary Calcium Screening Using Lung Cancer Screening Cardiac Imaging Original Research Sung Mok Kim 1 Myung Jin Chung 1 Kyung Soo Lee 1 Yeon Hyun Choe 1 Chin

More information

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease Horizon Scanning Technology Summary National Horizon Scanning Centre Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease April 2007 This technology summary is based

More information

A Noninvasive Assessment of CAD

A Noninvasive Assessment of CAD : A Noninvasive Assessment of CAD In this article, Dr. Heilbron and Dr. Forster look at the noninvasive assessment of coronary artery disease (CAD), by means of coronary computed tomography angiography

More information

Spiral Multislice Computed Tomography Coronary Angiography: A Current Status Report

Spiral Multislice Computed Tomography Coronary Angiography: A Current Status Report Clin. Cardiol. 30, 437 442 (2007) Spiral Multislice Computed Tomography Coronary Angiography: A Current Status Report P. J. De Feyter, M.D., PH.D., W. B. Meijboom, M.D., A. Weustink, M.D., C. Van Mieghem,

More information

A Snapshot on Nuclear Cardiac Imaging

A Snapshot on Nuclear Cardiac Imaging Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

Low-dose CT coronary angiography in the step-andshoot mode: diagnostic performance

Low-dose CT coronary angiography in the step-andshoot mode: diagnostic performance 1 Institute of Diagnostic Radiology, University Hospital Zurich, Zurich, Switzerland; 2 Clinic for Cardiovascular Surgery, University Hospital Zurich, Zurich, Switzerland; 3 Cardiovascular Centre, University

More information

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Cardiac CT David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Courtesy Tom Kracus Courtesy Kettering Tom Medical Kracus Cente Kettering Medical Center 2003-2006, David Stultz, MD Courtesy Tom Kracus

More information

Sang Ho Lee, Byoung Wook Choi, Hee-Joung Kim*, Member, IEEE, Haijo Jung, Hye-Kyung Son, Won-Suk Kang, Sun Kook Yoo, Kyu Ok Choe, Hyung Sik Yoo

Sang Ho Lee, Byoung Wook Choi, Hee-Joung Kim*, Member, IEEE, Haijo Jung, Hye-Kyung Son, Won-Suk Kang, Sun Kook Yoo, Kyu Ok Choe, Hyung Sik Yoo Analysis of Heart Rate and its Variation Affecting Image Quality and Optimized Reconstruction Window in Retrospective ECG-gated Coronary Angiography Using Multi-detector Row CT Sang Ho Lee, Byoung Wook

More information

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial Poster No.: C-0595 Congress: ECR 2013 Type: Authors:

More information

Diagnostic and Prognostic Value of Coronary Ca Score

Diagnostic and Prognostic Value of Coronary Ca Score Diagnostic and Prognostic Value of Coronary Ca Score Dr. Ghormallah Alzahrani Cardiac imaging division, Adult Cardiology department Prince Sultan Cardiac Center ( PSCC) Madina, June 2 Coronary Calcium

More information

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 1, FEBRUARY

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 1, FEBRUARY IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 1, FEBRUARY 2004 225 Analysis of the Heart Rate and Its Variation Affecting Image Quality and Optimized Reconstruction Window in Retrospective ECG-Gated

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination Int J Clin Exp Med 2015;8(11):21898-21905 www.ijcem.com /ISSN:1940-5901/IJCEM0015005 Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out

More information

An Introduction to Dual Energy Computed Tomography

An Introduction to Dual Energy Computed Tomography An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

Disclosure Information

Disclosure Information Coronary CTA Pearls and Pitfalls Ricardo C. Cury, MD, FSCCT, FAHA, FACC Chairman of Radiology Radiology Associates of South Florida Director of Cardiac Imaging Miami Cardiac and Vascular Institute Past-President

More information

Adapted Transfer Function Design for Coronary Artery Evaluation

Adapted Transfer Function Design for Coronary Artery Evaluation Adapted Transfer Function Design for Coronary Artery Evaluation Sylvia Glaßer 1, Steffen Oeltze 1, Anja Hennemuth 2, Skadi Wilhelmsen 3, Bernhard Preim 1 1 Department of Simulation and Graphics, University

More information

Assessment of Non-Calcified Coronary Plaques Using 64-Slice Computed Tomography: Comparison With Intravascular Ultrasound

Assessment of Non-Calcified Coronary Plaques Using 64-Slice Computed Tomography: Comparison With Intravascular Ultrasound ORIGINAL ARTICLE DOI 10.4070 / kcj.2009.39.3.95 Print ISSN 1738-5520 / On-line ISSN 1738-5555 Copyright c 2009 The Korean Society of Cardiology Assessment of Non-Calcified Coronary Plaques Using 64-Slice

More information

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital Coronary Artery Imaging Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital Larger array : cover scan area Detector size : spatial resolution Rotation speed : scan time Retrospective

More information

Computed Tomography of the Coronary Arteries

Computed Tomography of the Coronary Arteries Cardiology Update DAVOS 2011 Computed Tomography of the Coronary Arteries Anders Persson M.D., Ph.D Director, Assoc. Professor Center for Medical Image Science and Visualization Linköping University SWEDEN

More information

QUANTITATIVE ASSESSMENT OF IMAGE QUALITY IN 64-SLICE-COMPUTED TOMOGRAPHY OF CORONARY ARTERIES IN SUBJECTS UNDERGOING SCREENING

QUANTITATIVE ASSESSMENT OF IMAGE QUALITY IN 64-SLICE-COMPUTED TOMOGRAPHY OF CORONARY ARTERIES IN SUBJECTS UNDERGOING SCREENING QUANTITATIVE ASSESSMENT OF IMAGE QUALITY IN 64-SLICE-COMPUTED TOMOGRAPHY OF CORONARY ARTERIES IN SUBJECTS UNDERGOING SCREENING FOR CORONARY ARTERY DISEASE Li-Hwa Yang, 1 Ding-Kwo Wu, 1,2 Chiao-Yun Chen,

More information

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Poster No.: C-0863 Congress: ECR 2015 Type: Scientific Exhibit Authors: A. Nakamoto,

More information

Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control

Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control Eur Radiol (2006) 16: 2739 2747 DOI 10.1007/s00330-006-0474-0 CARDIAC Hans Scheffel Hatem Alkadhi André Plass Robert Vachenauer Lotus Desbiolles Oliver Gaemperli Tiziano Schepis Thomas Frauenfelder Thomas

More information

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Int J Cardiovasc Imaging (2008) 24:547 556 DOI 10.1007/s10554-007-9282-0 ORIGINAL PAPER Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Jaap M. Groen Æ Marcel J.

More information

b. To facilitate the management decision of a patient with an equivocal stress test.

b. To facilitate the management decision of a patient with an equivocal stress test. National Imaging Associates, Inc. Clinical guidelines EBCT HEART CT & HEART CT CONGENITAL CCTA CPT4 Codes: 75571 EBCT 75572, 75573 Heart CT & Heart CT Congenital 75574 - CCTA LCD ID Number: L33559 J K

More information

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population Journal of the American College of Cardiology Vol. 49, No. 9, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.10.066

More information

Chapter. Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography

Chapter. Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography Chapter 3 Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography Joanne D. Schuijf, Jeroen J. Bax, Liesbeth P. Salm, J. Wouter Jukema, Hildo J. Lamb,

More information

Multisclice CT in combination with functional imaging for CAD. Temporal Resolution. Spatial Resolution. Temporal resolution = ½ of the rotation time

Multisclice CT in combination with functional imaging for CAD. Temporal Resolution. Spatial Resolution. Temporal resolution = ½ of the rotation time Multisclice CT in combination with functional imaging for CAD Prof. Juhani Knuuti, MD, FESC Turku University Hospital and University of Turku Turku, Finland MSCT and functional imaging for CAD Practical

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

Evaluation of Ascending Aortic Atherosclerosis with 16-Multidetector Computed Tomography Is Useful before Total Endoscopic Coronary Bypass Surgery

Evaluation of Ascending Aortic Atherosclerosis with 16-Multidetector Computed Tomography Is Useful before Total Endoscopic Coronary Bypass Surgery 9 (5), 2006 [Epub June 2006] doi:10.1532/hsf98.20051103 Online address: http://cardenjennings.metapress.com/link.asp?id=112496 Evaluation of Ascending Aortic Atherosclerosis with 16-Multidetector Computed

More information

Purpose. Methods and Materials

Purpose. Methods and Materials The dose of contrast medium should be determined on the basis of estimated lean body weight (LBW) instead of the body weight--excessive amounts are currently given to females. Poster No.: C-2641 Congress:

More information

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre Pushing the limits of cardiac CT Steven Dymarkowski Radiology / Medical Imaging Research Centre 5 X 2013 Introduction Rapid technological advances and new clinical applications in cardiovascular imaging

More information

Characteristics of Subclinical Coronary Artery Disease in Diabetic Patients without Known Coronary Artery Disease

Characteristics of Subclinical Coronary Artery Disease in Diabetic Patients without Known Coronary Artery Disease IBIMA Publishing Journal of Research in Diabetes http://www.ibimapublishing.com/journals/diab/diab.html Vol. 2014 (2014), Article ID 322292, 12 pages DOI: 10.5171/2014.322292 Research Article Characteristics

More information

Diagnostic Accuracy of Angiographic View Image for the Detection of Coronary Artery Stenoses by 64-Detector Row CT

Diagnostic Accuracy of Angiographic View Image for the Detection of Coronary Artery Stenoses by 64-Detector Row CT Diagnostic Accuracy of Angiographic View Image for the Detection of Coronary Artery Stenoses by 64-Detector Row CT A Pilot Study Comparison With Conventional Post-Processing Methods and Axial Images Alone

More information

Computed tomography in coronary imaging: current status

Computed tomography in coronary imaging: current status 7 Computed tomography in coronary imaging: current status ARJUN NAIR AND ANAND DEVARAJ Recent technological advances have led to improvements in the use of computerised tomography for coronary imaging.

More information

Clinical material and methods. Clinical Departments of 1 Radiology II and 2 Cardiology, Innsbruck Medical University, Innsbruck, Austria

Clinical material and methods. Clinical Departments of 1 Radiology II and 2 Cardiology, Innsbruck Medical University, Innsbruck, Austria Aortic Valve Calcification as Quantified with Multislice Computed Tomography Predicts Short-Term Clinical Outcome in Patients with Asymptomatic Aortic Stenosis Gudrun M. Feuchtner 1, Silvana Müller 2,

More information

Purpose. Methods and Materials

Purpose. Methods and Materials Comparison of iterative and filtered back-projection image reconstruction techniques: evaluation of heavily calcified vessels with coronary CT angiography Poster No.: C-1644 Congress: ECR 2011 Type: Scientific

More information

Recent developments in coronary computed tomography imaging

Recent developments in coronary computed tomography imaging REVIEW Recent developments in coronary computed tomography imaging Cardiac computed tomography (CT) has recently emerged as a noninvasive alternative to catheter angiography for the assessment of coronary

More information

Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure. Coronary Artery Calcium Volume?

Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure. Coronary Artery Calcium Volume? Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure Coronary Artery Calcium Volume? 1 Objective: The purpose of this study is to investigate how accurately we can measure CAC volume

More information

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Radiation Exposure in Pregnancy John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Illustrative Clinical Scenario 32 year old female 34 weeks pregnant with recent onset shortness of breath and central chest pain

More information

Noncoronary Cardiac MDCT

Noncoronary Cardiac MDCT Noncoronary Cardiac MDCT David A. Bluemke, M.D., Ph.D. Professor, of Radiology and Medicine Johns Hopkins University School of Medicine Baltimore, Maryland Toshiba Disclosures Grant support Noncoronary

More information

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease Perspectives of new imaging techniques for patients with known or suspected coronary artery disease Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands Correspondence: Jeroen

More information

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Mathieu Sabbagh, R3 Michigan State University Radiology Garden City Hospital

More information

Computed tomography. Department of Radiology, University Medical School, Szeged

Computed tomography. Department of Radiology, University Medical School, Szeged Computed tomography Department of Radiology, University Medical School, Szeged voxel +1-4 +2 +5 +3 +1 0-2 pixel -2 0 +1-4 -6 +5 +2 +1 Department of Radiology, University Medical School, Szeged

More information

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA SYMPOSIA Image Acquisition and Interpretation Josef Matthias Kerl, MD,*w Lars K. Hofmann, MD,z Christian Thilo, MD,*y Thomas J. Vogl, MD,w Philip Costello, MD,* and U. Joseph Schoepf, MD*y Abstract: Computed

More information

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011 Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011 Outline Background Hypothesis Study Population Methodology

More information

Image quality and, hence, the diagnostic value of cardiac. Imaging

Image quality and, hence, the diagnostic value of cardiac. Imaging Imaging Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily Practice Impact of Different Scanning Protocols on Effective Dose Estimates Jörg Hausleiter, MD; Tanja Meyer, MD; Martin

More information

ECG-Gated 64-MDCT Angiography in the Differential Diagnosis of Acute Chest Pain

ECG-Gated 64-MDCT Angiography in the Differential Diagnosis of Acute Chest Pain Johnson et al. ECG-Gated CTA in Chest Pain Cardiopulmonary Imaging Original Research 076.fm 11/30/06 Thorsten R. C. Johnson 1 Konstantin Nikolaou 1 Bernd J. Wintersperger 1 Andreas Knez 2 Peter Boekstegers

More information

The Impact of Warmed Intravenous Contrast Material on the Bolus Geometry of Coronary CT Angiography Applications

The Impact of Warmed Intravenous Contrast Material on the Bolus Geometry of Coronary CT Angiography Applications The Impact of Warmed Intravenous Contrast Material on the Bolus Geometry of Coronary CT Angiography Applications Tuncay Hazirolan, MD 1 Baris Turkbey, MD 1 Erhan Akpinar, MD 1 Murat Canyigit, MD 1 Musturay

More information

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT Ralf Koos 1 Andreas Horst Mahnken 2 Anil Martin Sinha 1 Joachim Ernst Wildberger 2 Rainer Hoffmann 1 Harald Peter Kühl 1 Received March 3, 2004; accepted after revision May 18, 2004. 1 Department of Cardiology,

More information

Cardiopulmonary Imaging Original Research

Cardiopulmonary Imaging Original Research Cardiopulmonary Imaging Original Research Bischoff et al. Versus Helical Mode for Coronary CTA Cardiopulmonary Imaging Original Research Bernhard Bischoff 1 Franziska Hein 1 Tanja Meyer 1 Markus Krebs

More information

A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT

A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT Poster No.: C-0623 Congress: ECR 2017 Type: Authors: Keywords: DOI:

More information

Diagnostic Accuracy of Computed Tomography Angiography in Patients After Bypass Grafting

Diagnostic Accuracy of Computed Tomography Angiography in Patients After Bypass Grafting JACC: CARDIOVASCULAR IMAGING VOL. 2, NO. 7, 2009 2009 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-878X/09/$36.00 PUBLISHED BY ELSEVIER INC. DOI:10.1016/j.jcmg.2009.02.010 Diagnostic Accuracy

More information

Sixty-four-slice multidetector computed tomography: the future of ED cardiac care

Sixty-four-slice multidetector computed tomography: the future of ED cardiac care Thomas Jefferson University Jefferson Digital Commons Department of Radiology Faculty Papers Department of Radiology October 2006 Sixty-four-slice multidetector computed tomography: the future of ED cardiac

More information

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography 633 CARDIOVASCULAR MEDICINE Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography A W Leber, A Knez, C Becker, A Becker, C White, C Thilo, M Reiser,

More information

Cardiac computed tomography: indications, applications, limitations, and training requirements

Cardiac computed tomography: indications, applications, limitations, and training requirements European Heart Journal (2008) 29, 531 556 doi:10.1093/eurheartj/ehm544 SPECIAL ARTICLE Cardiac computed tomography: indications, applications, limitations, and training requirements Report of a Writing

More information

Computed Tomography Imaging of the Coronary Arteries

Computed Tomography Imaging of the Coronary Arteries Chapter 6 Computed Tomography Imaging of the Coronary Arteries G.J. Pelgrim, M. Oudkerk and R. Vliegenthart Additional information is available at the end of the chapter http://dx.doi.org/10.5772/54044

More information

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Poster No.: C-1207 Congress: ECR 2013 Type: Scientific Exhibit Authors: Y. Shimomiya, M. Kondo, M.

More information

Department of Cardiology, Grosshadern Clinic, University of Munich, Marchioninistrasse 15, Munich, Germany. Department of Cardiology,

Department of Cardiology, Grosshadern Clinic, University of Munich, Marchioninistrasse 15, Munich, Germany. Department of Cardiology, Eur Radiol (2002) 12:1532 1540 DOI 10.1007/s00330-002-1394-2 CARDIAC B. Ohnesorge T. Flohr R. Fischbach A. F. Kopp A. Knez S. Schröder U. J. Schöpf A. Crispin E. Klotz M. F. Reiser C. R. Becker Reproducibility

More information

Impact of Body Mass Index and Metabolic Syndrome on the Characteristics of Coronary Plaques Using Computed Tomography Angiography

Impact of Body Mass Index and Metabolic Syndrome on the Characteristics of Coronary Plaques Using Computed Tomography Angiography Impact of Body Mass Index and Metabolic Syndrome on the Characteristics of Coronary Plaques Using Computed Tomography Angiography Cardiovascular Division, Faculty of Medicine, University of Tsukuba Akira

More information

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation Poster No.: C-2014 Congress: ECR 2014 Type: Educational Exhibit Authors: M. A. Ottesen; Oslo/NO Keywords: Cardiac, Arteries / Aorta, CT, CT-Angiography,

More information