Procurement Support to Ministry of Health, Ukraine. Invitation to Bid for:

Size: px
Start display at page:

Download "Procurement Support to Ministry of Health, Ukraine. Invitation to Bid for:"

Transcription

1 Procurement Support to Ministry of Health, Ukraine Invitation to Bid for: PROGRAMME XIX: MEDICAL PRODUCTS FOR HEALTHCARE INSTITUTIONS FOR TREATMENT OF PATIENTS WITH CARDIOVASCULAR AND CEREBROVASCULAR DISEASES Part 2 Electrophysiology & Cardio Stimulation ITB No: CA/107403/0001/001 Clarification No 2 This is Clarification No 2 dated 6 th January 2017 to the ITB dated 28th December 2106, ITB No: CA/107403/0001/001 Please find attached the questions received from potential Suppliers and Crown Agents clarifications. Note that the deadline for seeking further clarifications is 19 th January Please acknowledge receipt of this clarification. Aysel Abasova Supply Chain Specialist Crown Agents 6 th January 2017 Page 1

2 CA/107403/0001/001 Clarification No 2 No 11. Attached you can also find the list of questions specifically related to Appendix C (STATEMENT OF COMPLIANCE and TECHNICAL OFFER) Answer Please see answers attached on the relevant bidding forms. Page 2

3 APPENDIX C STATEMENT OF COMPLIANCE and TECHNICAL OFFER STATEMENT OF COMPLIANCE Item Part 2 Electrophysiology and Cardio Stimulation 1 Single chamber pacemakers (SSI) with bipolar and monopolar electrode should correspond to the following medical-technical - Pacemaking regime single chamber (SSI) - AP rhythm intruding control option and automated amplitude control during ventricular pacing - Supply with bipolar steroid electrode with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a kit/set (1 unit) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended 2 Single chamber rate-responsive pacemaker with automatic amplitude control during ventricular pacing (SSIR) should correspond to the following medicaltechnical - Pacemaking regime single chamber (SSIR) Discrepancy of lot name and requirements (SSI vs. SSIR) SSIR is correct Page 3

4 - AP rhythm intruding control option and automated amplitude control during ventricular pacing - Supply with bipolar steroid electrode with active fixing (draw-out spiral) and possibility of implantation through the introducer in a kit/set (1 unit) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended 3 Single chamber rate-responsive pacemaker (SSIR) should correspond to the following medical-technical - Pacemaking regime single chamber (SSIR) Discrepancy of lot name and requirements (SSI vs. SSIR) SSIR is correct episodes ATA/AF and pacemaking system status - Supply with bipolar steroid electrode with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a kit/set (1 unit) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended 4 Dual chamber pacemaker without autocapture function (DDD) should correspond to the following medical-technical - Pacemaking regime dual chamber (DDD) Page 4

5 episodes ATA/AF and pacemaking system status - Supply with two bipolar steroid electrodes with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a ket/set. (2 units) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended 5 Dual chamber pacemaker with the autocapture function (DDD) should correspond to the following medical-technical?? with the autocapture function - Pacemaking regime dual chamber (DDD) - AP rhythm intruding control option and automated amplitude control during ventricular pacing Discrepancy of lot name and requirements (without autocapture vs. amplitude control during VP meaning autocapture) Lot name corrected episodes ATA/AF and pacemaking system status - Function allowing noninvasive pacing for the patient with artificial pacemaker implanted. - Automated tracing of intrinsic conduction to decrease ventricular pacing - Supply with two bipolar steroid electrodes with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a kit/set (2 units) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended 6 Dual chamber pacemaker (DDDR) which should correspond to the following medical-technical No longevity requirements while superior is mentioned in the lot name Page 5

6 Wording amended - Pacemaking regime dual chamber (DDDR) Discrepancy of lot name and requirements (DDD vs. DDDR) Should read DDDR - AP rhythm intruding control option and automated amplitude control during ventricular pacing episodes ATA/AF and pacemaking system status - Automated tracing of intrinsic conduction to decrease ventricular pacing - Supply with two bipolar steroid electrodes with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a kit/set (2 units) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended 7 Dual chamber rate-adaptive pacemaker (DDDR) for neonates should correspond to the following medicaltechnical - Pacemaking regime dual chamber (DDDR) Discrepancy of lot name and requirements (DDD vs. DDDR) Should read DDDR - Maximal possible pacemaking rate (not less than 200 BPM) - AP rhythm intruding control option and automated amplitude control during ventricular pacing Page 6

7 - Automated tracing of intrinsic conduction to decrease ventricular pacing episodes and pacemaking system status - The least weight and dimensions of device out of all introduced Not exact requirement Agreed - deleted 8 - Supply with two bipolar steroid electrodes with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a kit/set (2 units) with a diameter not exceeding 6 Fr Dual chamber rate-responsive pacemaker (DDDR) for children and adolescents should correspond to the following medical-technical This row was missing in the original Appendix C - Pacemaking regime dual chamber (DDDR) Discrepancy of lot name and requirements (DDD vs. DDDR) Should read DDDR - Maximal possible pacemaking rate (not less than 180 BPM) - AP rhythm intruding control option and automated amplitude control during ventricular pacing - Automated tracing of intrinsic conduction to decrease ventricular pacing episodes and pacemaking system status - The least weight and dimensions of device out of all introduced Page 7

8 - Supply with two bipolar steroid electrode with active fixing (draw-out spiral) and possibility of implantation through the introducer in a kit/set (2 units) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended 9 Dual chamber rate-responsive pacemaker with regime (DDDR) should correspond to the following medicaltechnical No leads/introducers in the kit?? If leads/introducers are supposed to be in the kit, please add this to requirements. Amended - Pacemaking regime dual chamber rate adaptive (DDDR) episodes and pacemaking system status - Automated tracing of intrinsic conduction to decrease ventricular pacing 10 - Supply with two bipolar steroid electrodes with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a kit/set (2 units) with a diameter not exceeding 7 Fr Dual chamber rate-adaptive pacemaker with preventive pacing algorithms (DDDRP) for bradycardia patients with atrial fibrillation should correspond to the following medical-technical This row was missing in the original Appendix C - Pacemaking regime dual chamber rate adaptive (DDDR) - AP rhythm intruding control option and automated amplitude control during ventricular pacing episodes ATA/AF and pacemaking system status - Automated tracing of intrinsic conduction to decrease ventricular pacing Page 8

9 - Atrial pacing self-regulation algorithm to prevent and treat atrial arrhythmia - Supply with two bipolar steroid electrodes with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a kit/set (2 units) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended 11 Cardiac resynchronization therapy pacemaker with the function of intrathoracic impedance measurement should correspond to the following medical-technical Discrepancy of lot name and requirements (intrathoracic impedance measurement vs. no in requirements) See below - Pacemaking regime - triple chamber biventricular - Cardiac resynchronization therapy option - Sensor for rate adaptation before physical activity - Interventricular break regulation option episodes ATA/AF and pacemaking system status - Supply with two bipolar steroid electrodes for right atrium and right ventricule with active fixing (draw-out spiral) and possibility of implantation through the introducer sheath in a kit/set (2 units) with a diameter not exceeding 7 Fr - Electrode for left ventricle (LV) pacemaking should be supplied with the system of delivery to the cardiac veins What are the parts of the system of delivery? Name each part in requirements exactly, please. Page 9

10 The electrode should have the capability (additional system) which enables the delivery of the electrode to the cardiac veins. - Electrode for left ventricle (LV) pacemaking should have the option of delivery on the conductor - Option of recording and memorizing of intracardial tracing, adjustment of duration and number of intracardial tracing recordings 12 Single chamber implantable cardioverter-defibrillator should correspond to the following medical-technical - Pacemaking regime - triple chamber biventricular (VVIR) episodes AT/AF/VT/VF and pacemaking system status - Maximal power of defibrillating impulse not less than 30.0 J - Directed discharge inversion option - Ventricular and supraventricular tachyarrhythmia differencing algorithm - Supply with single spiral defibrillating electrode not less than 9 Fr thick with active fixing and introducer sheath not more than 10 Fr thick with connector DF-1 or DF-4. Where we can have clients requirements? Deleted - Function of ventricular tachycardia elimination by defibrilator antitachycardic pacemaking 13 Dual chamber implantable cardioverter-defibrillator should correspond to the following medical-technical Discrepancy of lot name and requirements (Single chamber vs. dual Page 10

11 requirements chamber rate adaptive) Corrected should be dual chamber - Pacemaking regime dual chamber rate adaptive (DDDR) - Maximal power of defibrillating impulse not less than 30.0 J - Directed discharge inversion option DDDR is correct - Ventricular and supraventricular tachyarrhythmia differencing algorithm episodes AT/AF/VT/VF and pacemaking system status - Function to decrease the extensive stimulation of right ventricle - Function of ventricular tachycardia elimination by defibrillator antitachycardic stimulation - Supply with bipolar steroid electrode with active fixing (draw-out spiral) and possibility of implantation through the introducer in a kit/set (2 units) with a diameter not exceeding 7 Fr What does stand CPL. For. Complete wording amended - Supply with single spiral defibrillating electrode not more than 9 Fr thick with active fixing and introducer sheath not more than 10 Fr thick with connector DF-1 or DF-4 Where we can have clients requirements? Should be "more" and not "less" Client requirement deleted. Page 11

12 14 Cardiac resynchronization therapy implantable dual chamber cardioverter- defibrillator should correspond to the following medical-technical - Pacemaking regime triple chamber rate adaptive. - VV break programming option - Ventricular and supraventricular tachyarrhythmia differencing algorithm - Supply with bipolar steroid electrodes for atrium and right ventricle not more than 7 Fr thick. Electrode should have active fixing - Electrode for left ventricle (LV) pacemaking should be bipolar and supplied with the system of its delivery to the cardiac veins on the conductor If that the Right Ventricle lead (asked in the req below), please remove it. What are the parts of the system? Do we need to provide catheter? Or balloon as well? There is no further information. Each bidder should offer their product that gives the solution required by MOH Ukraine. - Maximal power of defibrillating impulse not less than 30.0 J - Directed discharge inversion option - Ventricular and supraventricular tachyarrhythmia differencing algorithm episodes AT/AF/VT/VF and pacemaking system status - Supply with single spiral defibrillating electrode not less than 9 Fr thick with active fixing and introducer sheath not more than 9 Fr thick with connector DF-1 or DF-4. Where we can have clients requirements? Because DF1/DF4 connectors are the different types of devices Each bidder should offer their product that gives the solution required by Page 12

13 MOH Ukraine - Function of ventricular tachycardia elimination by defibrillator antitachycardic pacemaking 15 Epicardial chronic steroid eluting electrode should correspond to the following medical-technical - Bipolar with ventricle or atrium pacemaking option - Active fixation with stitching option - Electrode length from 35 to 80 cm Will the length of 35, 50 and 60 be accepted The length should be within the range specified - Connector type IS-1 Were these items lost from the ITB appendix C?. Our source copy is correct. 16 Conventional electrodes for ablation catheters should correspond to the following medical-technical - 4 poles of electrode with the length of distal pole 4 mm - Electrode thickness not more than 7 Fr - Temperature indicator - Curve dimensions according to the Client s requirements - Electrode spacing : 2-5-2mm 17 Irrigated electrodes for ablation catheters should correspond to the following medical-technical Page 13

14 - 4 poles of electrode with the length of distal pole 4 mm - Electrode thickness not more than 7 Fr - Temperature indicator - Should have external cooling contour - Curve dimensions according to the Client s requirements - Electrode spacing : 2-5-2мм 18 Steerable double-curved ablation catheter with bidirectional tip should correspond to the following medical-technical 18.1 Distal electrode length 4 mm - 4 poles of electrode with the length of distal pole 4 mm - Electrode thickness not more than 7 Fr - Temperature indicator - Proximal and distal curve. Bi-directional distal tip - Electrode spacing : mm - Curve dimensions according to the Client s requirements 18.2 Distal electrode length 8 mm - 4 poles of electrode with the length of distal pole 8 mm - Electrode thickness not more than 7 Fr - Temperature indicator - Proximal and distal curve. Bi-directional distal tip - Electrode spacing : mm Page 14

15 - Curve dimensions according to the Client s requirements 19 Navigation irrigated diagnostic ablation catheter should correspond to the following medical-technical Purpose For electrophysiological mapping (pacing and recording) and for cardiac ablation, if used together with RF generator Compatibility Should be compatible with Carto navigation system and RF generator Stockert EP-Shuttle Possibilities Electrode cooling (irrigation) technology should be available during the ablation procedure Electrode spacing mm Insertion depth Not less than 115 cm Curve dimensions according to the Client s requirements 20 Navigation system reference electrodes should correspond to the following medical-technical Temporarily removed from tender whilst specification is checked. It may be re-introduced through an amendment to this ITB 21 Temporary cardiac pacing wire should correspond to the following medical-technical 21.1 Length 0.65 m - Temporary cardiac pacing wire with length not less than 65 cm and not more than 75sm - Micro wire bladed or twisted, insulated - 2 needles at the end of the wire Page 15

16 21.1 Length 0.90 m - One needle is straight, with length not less than 60 mm, with notch for easier breakage - Second needle is bladed ½ circle, with length 17 to 26 mm - Temporary cardiac pacing wire with length not less than 90 cm and not more than 100cm - Micro wire bladed or twisted, insulated - 2 needles at the end of the wire 21.3 Length 1.20 m - One needle is straight, with length not less than 60 mm, with notch for easier breakage - Second needle is bladed ½ circle, with length 17 to 26 mm - Temporary cardiac pacing wire with length not less than 120 cm and not more than 130 cm - Micro wire bladed or twisted, insulated - 2 needles at the end of the wire - One needle is straight, with length not less than 60 mm, with notch for easier breakage - Second needle is bladed ½ circle, with length 17 to 26 mm Page 16

17 TECHNICAL OFFER Crown Agents Reference: CA/107403/0001/001 Bidders Reference: Bid Closing Date: 26 January 2017 Currency: USD To:- Crown Agents Limited St Nicholas House, St Nicholas Road Sutton, Surrey SM1 1EL United Kingdom Details of item offered Item INN Form of presentation Shelf Life (expiry date of product dd/mm/yy) Manufacturer Brand Name Country of Manufacture & Certified Country of Origin Delivery lead time from contract signature* Can we put item description as the lot (kit)? Or it shall be item by item:it should be item by item based on the list provided in ITB 1 Single chamber AP (SSI) with bipolar lead Kit with lead and introducer that includes: Not less than 75% of total UBD (without dd/mm/yy) We have several manufacturi ng facilities for one type of device and it could come from all of them) The origin should comply with the product documents registered in Ukraine 4. Page 17

18 Or? 1 Single chamber AP (SSI) with bipolar lead pcs Not less than 75% of total UBD (without dd/mm/yy) NO Page 18

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

A Closer Look Product Education at a glance

A Closer Look Product Education at a glance A Closer Look Product Education at a glance an ICD or CRT-D When Atrial Information is Not Used SUMMARY Boston Scientific dual-chamber ICDs and multi-chamber CRT-Ds will respond to atrial sensing regardless

More information

Pediatric Pacemaker Implantation Endocardial or Epicardial

Pediatric Pacemaker Implantation Endocardial or Epicardial Pediatric Pacemaker Implantation Endocardial or Epicardial HAITHAM BADRAN, MD, FEHRA CONSULTANT OF INTERVENTIONAL CARDIOLOGY CONSULTANT OF CARDIAC PACING AND ELECTROPHYSIOLOGY LECTURER OF CARDIOLOGY AIN

More information

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers

ISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers INTERNATIONAL STANDARD ISO 14708-2 Second edition 2012-08-15 Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers Implants chirurgicaux Dispositifs médicaux implantables actifs

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

Cardiac Rhythm Management Coder 2017

Cardiac Rhythm Management Coder 2017 Cardiac Rhythm Management Coder 2017 An easy-to-use tool for coding and reimbursement compliance Prepared and Published By: MedLearn Publishing, A Division of Panacea Healthcare Solutions, Inc. 287 East

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]

More information

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections.

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections. 2015 Cardiology Survival Guide Chapter 15: Pacemakers With more than 30 codes and many variables to choose from, you must weigh your options carefully when reporting insertion, revision, or removal of

More information

Kadlec Regional Medical Center Cardiac Electrophysiology

Kadlec Regional Medical Center Cardiac Electrophysiology Definition of electrophysiology study and ablation Kadlec Regional Medical Center Cardiac Electrophysiology Electrophysiology Study and Ablation An electrophysiology, or EP, study is a test of the heart

More information

Complex Left Atrial (CLA) Best Practices for Navigant

Complex Left Atrial (CLA) Best Practices for Navigant Complex Left Atrial (CLA) Best Practices for Navigant Ischemic Ventricular Tachycardia (VT) Best Practices for Navigant Ischemic VT Ablation Procedures Set Up In the absence of a septal defect or other

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Different indications for pacemaker implantation are the following:

Different indications for pacemaker implantation are the following: Patient Resources: ICD/Pacemaker Overview ICD/Pacemaker Overview What is a pacemaker? A pacemaker is a device that uses low energy electrical pulses to prompt the heart to beat whenever a pause in the

More information

2009 CPT Codes for Cardiac Device Monitoring

2009 CPT Codes for Cardiac Device Monitoring 2009 CPT Codes for Cardiac Device Monitoring December 2008 Notices Current Procedural Terminology (CPT ) is copyright 2008 American Medical Association. All Rights reserved. No fee schedules, basic units,

More information

New generations pacemakers and ICDs: an update

New generations pacemakers and ICDs: an update Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology

More information

Timing cycles. Objectives Upon completion of this program the participant will be able to:

Timing cycles. Objectives Upon completion of this program the participant will be able to: Timing Cycles Timing cycles Objectives Upon completion of this program the participant will be able to: Identify the basic timing cycles of a single and dual chamber pacemaker. Describe the characteristics

More information

Cardiac Imaging in abnormal rhythm Role of MDCT

Cardiac Imaging in abnormal rhythm Role of MDCT Cardiac Imaging in abnormal rhythm Role of MDCT Cardiac Imaging in abnormal rhythm Role of MDCT Scope of the problem CT in Atrial Fibrillation CT and pacing Ventricular arrhythmia Other applications 1

More information

Acknowledgments. Introduction. Managing the Care of Radiotherapy Patients Implanted with Cardiac Devices. Outline

Acknowledgments. Introduction. Managing the Care of Radiotherapy Patients Implanted with Cardiac Devices. Outline Managing the Care of Radiotherapy Patients Implanted with Cardiac Devices Acknowledgments Members of AAPM TG203 Joann Prisciandaro Moyed Miften, PhD Dimitris Mihailidis, PhD Professor and Chief Physicist

More information

Summary of the Final 2013 Policy, Payment Changes For Hospital Outpatient Departments

Summary of the Final 2013 Policy, Payment Changes For Hospital Outpatient Departments Summary of the Final 2013 Policy, Payment Changes For Hospital Outpatient Departments A. Introduction On November 1, 2012, the Centers for Medicare & Medicaid Services (CMS) issued a final rule that will

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Active Fixation Models: 7740, 7741 and 7742 Passive Fixation Models: 7731, 7732, 7735 and 7736

Active Fixation Models: 7740, 7741 and 7742 Passive Fixation Models: 7731, 7732, 7735 and 7736 The INGEVITY MRI pacing leads are 6F (2.0 mm) steroid-eluting, endocardial pace / sense leads designed for permanent implantation for either atrial or ventricular applications. INGEVITY MRI is the only

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Transvenous Pacemaker Procedures

Transvenous Pacemaker Procedures Cardiology: Pacemaker and Defibrillator Coding Presented By: Moderate Sedation 2017 99151 : under age 5, initial 15 minutes by MD performing intervention 99152: age 5 or older, initial 15 minutes by MD

More information

Patient Resources: Arrhythmias and Congenital Heart Disease

Patient Resources: Arrhythmias and Congenital Heart Disease Patient Resources: Arrhythmias and Congenital Heart Disease Overview Arrhythmias (abnormal heart rhythms) can develop in patients with congenital heart disease (CHD) due to thickening/weakening of their

More information

Cardiac Resynchronisation Therapy Patient Information

Cardiac Resynchronisation Therapy Patient Information Melbourne Heart Rhythm Cardiac Resynchronisation Therapy Patient Information Normal Heart Function The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two

More information

CONSULTA CRT-P Model C4TR01

CONSULTA CRT-P Model C4TR01 CONSULTA CRT-P Model C4TR01 Physical characteristics Volume a 15 cm 3 Mass H x W x D Surface area of titanium device can Surface area of each titanium nitride LECG electrode Radiopaque ID Materials in

More information

CRT Implantation Techniques 부천세종병원순환기내과박상원

CRT Implantation Techniques 부천세종병원순환기내과박상원 Cardiac Venous System and CRT Implantation Techniques 부천세종병원순환기내과박상원 Cardiac Resynchronization Therapy (CRT) Goal: Atrial synchronous biventricular pacing Transvenous approach for left ventricular lead

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

Biomedical Instrumentation

Biomedical Instrumentation Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous

More information

CATHETER ABLATION CODING & REIMBURSEMENT GUIDE. Updated September 2018

CATHETER ABLATION CODING & REIMBURSEMENT GUIDE. Updated September 2018 CATHETER ABLATION CODING & REIMBURSEMENT GUIDE Updated September 2018 TABLE OF CONTENTS Diagnosis Codes...3 ICD-10-CM Diagnosis Codes Coverage for Catheter Ablation Procedures....4 Medicare Other Payers

More information

Cardiac Rhythm Management Coder 2018

Cardiac Rhythm Management Coder 2018 Cardiac Rhythm Management Coder 2018 An easy-to-use tool for coding and reimbursement compliance Prepared and Published By: MedLearn Publishing, A Division of MedLearn Media, Inc. 445 Minnesota Street,

More information

Friedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing.

Friedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing. Friedman, Rott, Wokhlu, Asirvatham, Hayes 201 Figure.7 Shortening of the AV interval during pacing. 202 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization Figure.8 is obtained from

More information

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie

Girish M Nair, Seeger Shen, Pablo B Nery, Calum J Redpath, David H Birnie 268 Case Report Cardiac Resynchronization Therapy in a Patient with Persistent Left Superior Vena Cava Draining into the Coronary Sinus and Absent Innominate Vein: A Case Report and Review of Literature

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

A request for a log book extension must be put in writing and sent to BHRS, Unit 6B, Essex House, Cromwell Business Park, Chipping Norton,

A request for a log book extension must be put in writing and sent to BHRS, Unit 6B, Essex House, Cromwell Business Park, Chipping Norton, 7 7. A request for a log book extension must be put in writing and sent to BHRS, Unit 6B, Essex House, Cromwell Business Park, Chipping Norton, Oxfordshire OX7 5SR. E-mail: admin@bhrs.com. Tel: 01789 867

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net

More information

Pacing Codes and Modes Concepts

Pacing Codes and Modes Concepts Pacing Codes and Modes Concepts Pacing codes and modes concepts Objectives Upon completion of this program the participant will be able to: State what the first four positions of the NBG code represent.

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

Implantation of a Cardiac Resynchronisation Therapy-Defibrillator (CRT-D)

Implantation of a Cardiac Resynchronisation Therapy-Defibrillator (CRT-D) PLEASE PRINT WHOLE FORM DOUBLE SIDED ON YELLOW PAPER Patient Information to be retained by patient Implantation of a Cardiac Resynchronisation Therapy-Defibrillator (CRT-D) affix patient label What is

More information

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece Leadless pacing P.E. Vardas Professor of Cardiology University of Crete, Greece DISCLOSURES Small teaching fees from: Bayer/J&J, Boeringer, Menarini Small consulting fees from: Medtronic 1957: First External

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

Is This Thing Working?

Is This Thing Working? Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu Disclosures I currently receive research support from Medtronic, Inc. I have been compensated

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES

STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES March, 2007 Ministry of Internal Affairs and Communications (MIC) Introduction We are moving toward the creation of a ubiquitous network society

More information

How to Approach the Patient with CRT and Recurrent Heart Failure

How to Approach the Patient with CRT and Recurrent Heart Failure How to Approach the Patient with CRT and Recurrent Heart Failure Byron K. Lee MD Associate Professor of Medicine Electrophysiology and Arrhythmia Section UCSF Update in Electrocardiography and Arrhythmias

More information

Antibradycardia Pacemakers

Antibradycardia Pacemakers #90802 Antibradycardia Pacemakers COURSE #90802 15 CONTACT HOURS/CREDITS Release Date: 09/01/17 Expiration Date: 08/31/20 Antibradycardia Pacemakers HOW TO RECEIVE CREDIT Read the enclosed course. Complete

More information

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. The Miracle of Living February 21, 2018 Matthew Ostrom MD,FACC,FHRS Division of

More information

COGNIS. Specifications. COGNIS: Model N119 ZIP TELEMETRY CARDIAC RESYNCHRONIZATION THERAPY

COGNIS. Specifications. COGNIS: Model N119 ZIP TELEMETRY CARDIAC RESYNCHRONIZATION THERAPY COGNIS Specifications COGNIS: Model N119 The COGNIS cardiac resynchronization therapy defibrillator (CRT-D) provides ventricular tachyarrhythmia therapy and cardiac resynchronization therapy (CRT) for

More information

Implant Procedure Concepts

Implant Procedure Concepts 1 4 7 2 5 LV Lead 8 3 6 Implant Procedure Concepts 9 ICD and CRT LV Lead This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is

More information

8/2/2017. Acknowledgement. Disclaimer. How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations

8/2/2017. Acknowledgement. Disclaimer. How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations Moyed Miften, Ph.D. Department of Radiation Oncology University of Colorado School of Medicine Disclaimer

More information

Clinical Cardiac Electrophysiology

Clinical Cardiac Electrophysiology Clinical Cardiac Electrophysiology Certification Examination Blueprint Purpose of the exam The exam is designed to evaluate the knowledge, diagnostic reasoning, and clinical judgment skills expected of

More information

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network INTERNAL CARDIOVERSION Lancashire & South Cumbria Cardiac Network DC Cardioversion AF & AFL safe 1,2 efficacious 3,4 (60-94%) 5 SR - Increases exercise tolerance 6 Maintainence SR unlikely in patients

More information

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Francesco Zanon, MD, FESC, FHRS Arrhythmia and Electrophysiology Unit, Cardiology Dept. Santa Maria della Misericordia General

More information

Leadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital

Leadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Leadless pacing: Technical challenges and concerns Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Modest consultancy fees/honoraria from Bayer Boehringer Ingelheim

More information

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter Caution: Federal (USA) law restricts this device to sale by or on the order of a physician. DEVICE DESCRIPTION

More information

Pacemakers, Defibrillators, and Electrophysiology in 2012

Pacemakers, Defibrillators, and Electrophysiology in 2012 Pacemakers, Defibrillators, and Electrophysiology in 2012 Presented by: David B. Dunn, MD, FACS, CIRCC, CPC-H, CCC National Coding Standards Sources of information Centers for Medicare & Medicaid (CMS)

More information

(51) Int Cl.: A61N 1/37 ( )

(51) Int Cl.: A61N 1/37 ( ) (19) TEPZZ 654888B_T (11) EP 2 654 888 B1 (12) EUROPEAN PATENT SPECIFICATION (45) Date of publication and mention of the grant of the patent: 12.10.2016 Bulletin 2016/41 (21) Application number: 11793970.2

More information

Charging Considerations. Agenda 1/30/2013

Charging Considerations. Agenda 1/30/2013 February 2013 Jean C. Russell, MS, RHIT jrussell@epochhealth.com Richard Cooley, BA, CCS rcooley@epochhealth.com 518-430-1144 2 Agenda Cardiology special considerations 2013 Cardiology Codes Deleted Codes

More information

February Jean C. Russell, MS, RHIT Richard Cooley, BA, CCS

February Jean C. Russell, MS, RHIT Richard Cooley, BA, CCS February 2013 Jean C. Russell, MS, RHIT jrussell@epochhealth.com Richard Cooley, BA, CCS rcooley@epochhealth.com 518-430-1144 2 Agenda Cardiology special considerations 2013 Cardiology Codes Deleted Codes

More information

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012 IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT201203 JANUARY 24, 2012 The IHCP to reimburse implantable cardioverter defibrillators separately from outpatient implantation Effective March 1, 2012, the

More information

UNDERSTANDING ELECTROPHYSIOLOGY STUDIES

UNDERSTANDING ELECTROPHYSIOLOGY STUDIES UNDERSTANDING ELECTROPHYSIOLOGY STUDIES Testing and Treating Your Heart s Electrical System A Problem with Your Heart Rhythm The speed and pattern of a heartbeat is called the heart rhythm. The rhythm

More information

Biventricular Pacemakers Sensing

Biventricular Pacemakers Sensing Biventricular Pacemakers Sensing Sergio Dubner Director Arrhythmias and Electrophisiology Service, Clinica y Maternidad Suizo-Argentina, Buenos Aires, Argentina Left ventricular pacemaker lead technology

More information

BIVENTRICULAR PACEMAKER (CARDIAC RESYNCHRONIZATION THERAPY) FOR THE TREATMENT OF HEART FAILURE

BIVENTRICULAR PACEMAKER (CARDIAC RESYNCHRONIZATION THERAPY) FOR THE TREATMENT OF HEART FAILURE FOR THE TREATMENT OF HEART FAILURE Non-Discrimination Statement and Multi-Language Interpreter Services information are located at the end of this document. Coverage for services, procedures, medical devices

More information

Implantation of an Implantable Cardioverter Defibrillator (ICD)

Implantation of an Implantable Cardioverter Defibrillator (ICD) PLEASE PRINT WHOLE FORM DOUBLE SIDED ON YELLOW PAPER Patient Information to be retained by patient affix patient label What is an Implantable? Your doctor has recommended that you have an Implantable.

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE GUIDANCE EXECUTIVE (GE)

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE GUIDANCE EXECUTIVE (GE) NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE GUIDANCE EXECUTIVE (GE) Review of TA88; Dual-chamber pacemakers for symptomatic bradycardia due to sick sinus syndrome and/or atrioventricular block

More information

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only. SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000 1 APPROVED MRI SCAN CONDITIONS 1.5 TESLA FULL BODY 2 SETUP SJM MRI ACTIVATOR HANDHELD DEVICE The SJM MRI Activator handheld device must be enabled

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

CLINICAL CARDIAC ELECTROPHYSIOLOGY Maintenance of Certification (MOC) Examination Blueprint

CLINICAL CARDIAC ELECTROPHYSIOLOGY Maintenance of Certification (MOC) Examination Blueprint CLINICAL CARDIAC ELECTROPHYSIOLOGY Maintenance of Certification (MOC) Examination Blueprint ABIM invites diplomates to help develop the Clinical Cardiac Electrophysiology MOC exam blueprint Based on feedback

More information

Tachycardia Devices Indications and Basic Trouble Shooting

Tachycardia Devices Indications and Basic Trouble Shooting Tachycardia Devices Indications and Basic Trouble Shooting Peter A. Brady, MD., FRCP Cardiology Review Course London, March 6 th, 2014 2011 MFMER 3134946-1 Tachycardia Devices ICD Indications Primary and

More information

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2 Temporary Transvenous Pacing Guideline Classification: Clinical Guideline Lead Author: Dr Peter Woolfson Additional author(s): Dr Alan Fitchet Sister Joanne Hughes, Matron Julie Winstanley Authors Division:

More information

Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations

Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations Journal of Geriatric Cardiology (2018) 15: 193 198 2018 JGC All rights reserved; www.jgc301.com Case Report Open Access Repetitive narrow QRS tachycardia in a 61-year-old female patient with recent palpitations

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

Automatic assessment of atrial pacing threshold in current medical practice

Automatic assessment of atrial pacing threshold in current medical practice Europace (212) 14, 1615 1619 doi:1.193/europace/eus76 CLINICAL RESEARCH Leads and lead extraction Automatic assessment of atrial pacing threshold in current medical practice Jean Luc Rey 1, Serge Quenum

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

Arrhythmia Care in the DGH What Still Needs to be Done? Dr. Sundeep Puri Consultant Cardiologist

Arrhythmia Care in the DGH What Still Needs to be Done? Dr. Sundeep Puri Consultant Cardiologist Arrhythmia Care in the DGH What Still Needs to be Done? Dr. Sundeep Puri Consultant Cardiologist LOTS!!! This presentation confines itself to the situation in the North West. The views expressed are my

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Electrophysiology Implant Code Classification Table The

More information

Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD

Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD Harish Doppalapudi, MD Assistant Professor of Medicine Director, Clinical Cardiac Electrophysiology Training Program University of Alabama

More information

PROCEDURAL ASSESSMENT CHECKLIST PACEMAKER IMPLANT PROCEDURE. Fellow: Supervising Physician: Date of Observation:

PROCEDURAL ASSESSMENT CHECKLIST PACEMAKER IMPLANT PROCEDURE. Fellow: Supervising Physician: Date of Observation: PACEMAKER IMPLANT PROCEDURE 9. Placement of diagnostic catheters limitation of fluoroscopy 10. Placement of diagnostic catheters electrogram interpretation 11. Measurement of baseline intervals 12. Assessment

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

Case Report What Next After Failed Septal Ventricular Tachycardia Ablation?

Case Report What Next After Failed Septal Ventricular Tachycardia Ablation? www.ipej.org 180 Case Report What Next After Failed Septal Ventricular Tachycardia Ablation? Laurent Roten, MD 1, Nicolas Derval, MD 1, Patrizio Pascale, MD 1, Pierre Jais, MD 1, Pierre Coste, MD 2, Frederic

More information

BHRS Prep course Pub style Quiz NOT A

BHRS Prep course Pub style Quiz NOT A BHRS Prep course Pub style Quiz NOT A Round 1 Quiz questions Common ECG s in CRM Colin Cunnington Over the past 3 months, a 24-yearold man who works as a truck driver has had frequent episodes of pre-syncope

More information

FRONTIER II. User s Manual. Model 5586 Model Biventricular Cardiac Resynchronization Device with AF Suppression Algorithm

FRONTIER II. User s Manual. Model 5586 Model Biventricular Cardiac Resynchronization Device with AF Suppression Algorithm FRONTIER II Model 5586 Model 5596 Biventricular Cardiac Resynchronization Device with AF Suppression Algorithm User s Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed

More information

Icd 10 for pacemaker pocket infection

Icd 10 for pacemaker pocket infection Search Search pages & people Search Search Search pages & people Search Icd 10 for pacemaker pocket infection Oct 29, 2014. 9.1.2 In generator pocket infection, ICED-LI and ICED-IE? 9.1.3 What is the.

More information

VT Ablation in Structural Heart Disease Patient Information

VT Ablation in Structural Heart Disease Patient Information Melbourne Heart Rhythm VT Ablation in Structural Heart Disease Patient Information Ventricular Tachycardia in Structural Heart Disease (VT-SHD) Ventricular tachycardia (VT) is an abnormal rapid heart rhythm

More information

Intermittent episodes of paced tachycardia: what is the cause?

Intermittent episodes of paced tachycardia: what is the cause? Intermittent episodes of paced tachycardia: what is the cause? Lindsey Parkinson and Lucy Broadhurst, Rotherham NHS Foundation Trust Contact: Lindsey.Parkinson@rothgen.nhs.uk Introduction A hospital in-patient

More information

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation

More information