Noninvasive Coronary Angiography with 16 Detector Row CT: Effect of Heart Rate 1

Size: px
Start display at page:

Download "Noninvasive Coronary Angiography with 16 Detector Row CT: Effect of Heart Rate 1"

Transcription

1 Cardiac Imaging Radiology Martin H. K. Hoffmann, MD Heshui Shi, MD Robert Manzke, MSc Florian T. Schmid, MD Leon De Vries, PhD Michael Grass, PhD Hans-Jürgen Brambs, MD Andrik J. Aschoff, MD Noninvasive Coronary Angiography with 16 Detector Row CT: Effect of Heart Rate 1 Published online before print /radiol Radiology 2005; 234:86 97 Abbreviations: ECG electrocardiography MPR multiplanar reformation 3D three-dimensional 1 From the Department of Diagnostic Radiology, University Hospital of Ulm, Steinhoevelstrasse 9, D Ulm, Germany (M.H.K.H., H.S., F.T.S., H.J.B., A.J.A.); Philips Medical Systems, Cleveland, Ohio (L.D.V.); and Philips Research Laboratories, Hamburg, Germany (R.M., M.G.). Received September 1, 2003; revision requested November 11; final revision received April 7, 2004; accepted April 28. Address correspondence to M.H.K.H. ( martin.hoffmann@medizin.uni -ulm.de). Author contributions: Guarantors of integrity of entire study, H.J.B., L.D.V., A.J.A., M.H.K.H.; study concepts, M.H.K.H., M.G., L.D.V.; study design, M.H.K.H., R.M.; literature research, M.H.K.H., H.S., F.T.S.; clinical studies, M.H.K.H., F.T.S.; data acquisition and analysis/interpretation, H.S., F.T.S., M.H.K.H.; statistical analysis, M.H.K.H., M.G.; manuscript preparation, M.H.K.H., R.M.; manuscript definition of intellectual content, A.J.A.; manuscript editing, F.T.S., R.M.; manuscript revision/review, A.J.A., M.G., R.M.; manuscript final version approval, M.H.K.H., M.G., A.J.A., H.J.B., L.D.V. RSNA, 2004 PURPOSE: To evaluate the effect of heart rate on the quality of coronary angiograms obtained with 16 detector row computed tomography (CT) by using temporally enhanced three-dimensional (3D) approaches. MATERIALS AND METHODS: The local ethics committee approved the study, and informed consent was obtained from all patients. Fifty patients underwent coronary CT angiography (heart rate range, beats per minute). Raw data from helical CT and electrocardiography (ECG) were saved in a combined data set. Retrospectively ECG-gated images were reconstructed at preselected phases (50% and 80%) of the cardiac cycle. A 3D voxel-based approach with cardiac phase weighting was used for reconstruction. Testing for correlation between heart rate, cardiac phase reconstruction window, and image quality was performed with Kruskal-Wallis analysis. Image quality (freedom from cardiac motion related artifacts) was referenced against findings at conventional angiography in a secondary evaluation step. Regression analysis was performed to calculate heart rate thresholds for future -blocker application. RESULTS: A significant negative correlation was observed between heart rate and image quality (r 0.80, P.001). Motion artifact free images were available for 44 (88%) patients and were achieved consistently at a heart rate of 80 or fewer beats per minute (n 39). Best image quality was achieved at 75 or fewer beats per minute. Segmental analysis revealed that 97% of arterial segments (diameter 1.5 mm according to conventional angiography) were assessable at 80 or fewer beats per minute. Premature ventricular contractions and rate-contained arrhythmia did not impede diagnostic assessment of the coronary arteries in 10 (83%) of the 12 patients affected. CONCLUSION: Motion-free coronary angiograms can be obtained consistently with 16 detector row CT scanners and adaptive multicyclic reconstruction algorithms in patients with heart rates of less than 80 beats per minute. RSNA, 2004 Contrast material enhanced multi detector row computed tomography (CT) is a promising noninvasive technique for the detection, visualization, and characterization of stenotic disease in the coronary arteries (1 4). However, multi detector row CT has relatively poor temporal resolution (approximately 210 msec) compared with that of other noninvasive methods of coronary imaging, such as electron-beam CT (approximately 50 msec) (5) and magnetic resonance imaging (approximately 78 msec) (6). Therefore, multi detector row CT angiography remains sensitive to cardiac motion induced artifacts (7). Currently recommended protocols for coronary CT angiography, based on the results of studies performed with four detector row scanners, include reduced heart rate as a prerequisite (7). Sixteen detector row CT scanners have two characteristics that, in theory, should enable a slight increase in temporal resolution: slightly increased gantry rotation speed (0.42 second, compared with 0.50 second for four detector row scanners), which, with conventional half-scan (180 ) reconstruction, provides a temporal resolution of

2 Figure 1. Curves show mean temporal resolution in milliseconds according to constant heart rate at specified number of beats per minute. Resolution values reflect weighted contributions of projection data for each voxel and were calculated for all voxels in an arbitrary volume by using a numeric approach based on the mode of action of the reconstruction algorithm. Temporal resolution varies according to number of cardiac cycles used for reconstruction (order of segmentation): With reconstruction at end systole (50% of the cardiac cycle; dashed line) for minimal coronary artery tree motion, temporal resolution declines linearly with increasing heart rate. With middiastolic reconstruction (80% of the cardiac cycle; solid line), resolution follows the general rule of diastolic phase duration and declines with a curvilinear function against increasing heart rate. Curves for end-systolic and middiastolic phases show that middiastolic reconstruction is especially useful in patients with slow heart rates ( 65 beats per minute) but renders motion-compromised results at faster heart rates. With multicyclic reconstruction (notched line), curve for temporal resolution plotted against increasing heart rate has a wavelike appearance but shows less variation in temporal resolution, over a wider range of heart rates, than with middiastolic reconstruction. Plotting of cardiac cycle windows was modified according to Herzog et al (23). msec (compared with 250 msec for four detector row scanners) (8). In addition, the combination of three-dimensional (3D) reconstruction algorithms with adaptive multicyclic parameter settings theoretically allows better temporal resolution (9) at optimal heart rates (Fig 1). Thus, the purpose of this study was to evaluate the effect of heart rate on the quality of coronary angiograms obtained with 16 detector row CT by using temporally enhanced 3D reconstruction approaches. MATERIALS AND METHODS Patients Multi detector row CT for noninvasive coronary angiography was undertaken in 50 consecutive patients (mean age, 59.3 years; age range, years), including 32 men (mean age, 54 years; age range, years) and 18 women (mean age, 62 years; age range, years). Patients either were suspected of having coronary artery disease (n 15) or had manifested recurrent symptoms after percutaneous coronary interventions or surgery (n 35). The major inclusion criterion was indication for conventional coronary angiography. From November 2002 to January 2003, 53 patients were referred with this indication. Conventional angiograms were available for 45 patients. Eight patients referred for initial work-up for suspected coronary artery disease, in whom the result of multi detector row CT was negative (normal scan with adequate distal coverage), refused subsequent invasive coronary angiography. Exclusion criteria were allergy to iodine-containing contrast medium, renal failure (serum creatinine level of more than 120 mol/l), pregnancy, severe respiratory impairment, and pronounced cardiac failure. Three patients were excluded from the study because of elevated serum creatinine level (n 2) or previous allergic reaction to iodine-based contrast medium (n 1). The local ethics committee approved the study, and informed consent was obtained from all patients. Background A crucial step for performing coronary CT angiography is to relate the minimum temporal resolution available with the scanner configuration to the cardiac cycle phases in which coronary motion is minimal. Cardiac phase preselection in this study was influenced by the results obtained by Achenbach et al (10) with regard to coronary artery motion on transverse Volume 234 Number 1 16 Detector Row CT Coronary Angiography: Heart Rate 87

3 electron-beam CT images. Achenbach et al found that minimal motion occurred in the end-systolic phase, at 48% of the cardiac cycle, and in the middiastolic phase, at 84% of the cardiac cycle. We therefore preselected phases at 50% and 80% of the cardiac cycle for this study. The duration of these motion-free intervals is inversely related to heart rate: The middiastolic phase interval (at approximately 80% of the cardiac cycle) decreases with a curvilinear function against increasing heart rate (11), whereas the end-systolic phase interval (at approximately 50% of the cardiac cycle) remains relatively constant and decreases linearly with increasing heart rate (Fig 1) (12). According to the physiology of the cardiac cycle, this end-systolic phase (at approximately 50% of the cycle) comprises both reduced ejection during systole and isovolumic relaxation during early diastole. The middiastolic phase (at approximately 80% of the cycle) corresponds to diastasis, the quiescent phase of diastole, which occurs between rapid filling and atrial contraction. If both phases could be reliably detected, reconstruction could be performed by using data from only these two phases, with a total number of 600 images per patient (scan length, 12 cm; z-axis resolution, 0.8 mm; increment, 0.4 mm). This result would represent an improvement over total coverage of the cardiac cycle, which might involve reconstruction of as many as 10 phases and result in 3000 images per patient, necessitating reader-guided empiric phase evaluation. To identify a specific cardiac phase, a dynamic model or delay algorithm was used (13) that enables modeling of a compliance function according to both a linearly declining constant phase delay (systolic changes) and an exponentially declining relative phase delay (diastolic changes). The model was applied in this study by using so-called physiologic percentages to track cardiac phases (13). Physiologic percentages include a delay offset (systolic linear changes) and a percentage of delay (diastolic exponential changes). This procedure theoretically results in better tracking of phases during scanning, as the heart rate changes during the breath hold. In addition, the delay algorithm should enable better interindividual characterization of heart phases, although all studies of coronary motion during the cardiac cycle have shown large variations in phase duration among patients (10,13,14). Data Acquisition Contrast-enhanced multi detector row CT examinations were performed in the supine position, during a single breath hold, by using a 16 detector row scanner (MX 8000 IDT; Philips Medical Systems). All CT angiographic studies were preceded by a scout acquisition. For bolus tracking, transverse sections were acquired at the level of the aortic root (section position was related to the location of the carina in the scout images). Fifteen transverse sections were acquired successively, one section every 1.7 seconds (120 kv, 30 mas, 6-mm section thickness). Scanning began with a fixed delay of 10 seconds after contrast material administration. A total of 10 ml of an iodinated contrast agent containing 370 mg of iodine per milliliter (iomeprol, Solutrast 370; Altana, Konstanz, Germany) was given at a rate of 4 ml/sec, followed by a chaser bolus of 30 ml saline solution, given at 3.5 ml/sec. A region of interest (mean diameter, 11.2 mm 5.1 [standard deviation]) was positioned in the aortic root, and attenuation enhancement values (in Hounsfield units) within this region were plotted against time (F.T.S., M.H.K.H.). Coronary CT angiography was timed to coincide with peak contrast enhancement, which was derived from attenuation measurements during bolus tracking. Electrocardiography (ECG) was performed during continuous CT data acquisition. The scanning protocol included a collimation of 16 sections with an individual section thickness of 0.75 mm ( mm), table feed increment of 6.86 mm/sec, pitch of 0.24, and gantry rotation time of 420 msec). Tube voltage of 140 kv and current of 230 ma (400 mas) were typically applied. Depending on cardiac dimensions, the scanning time varied between 16 and 23 seconds. Image Reconstruction After acquisition of the raw helical CT data, retrospectively ECG-synchronized sections were reconstructed. A multicyclic reconstruction algorithm with capabilities for slight improvement in temporal resolution was used (9). The word multicyclic in this context is synonymous with multisegmental, another term used in the literature; both terms refer to the segmentation of the gating window for data acquisition over multiple heartbeats. This method of segmentation theoretically improves temporal resolution commensurately with increase in the absolute number of cardiac cycles used in image reconstruction. The algorithm used in this study differed from those used in previous studies in that it was integrated with 3D voxel-based backward projection of the cone beam (9,15,16). This difference, we hypothesized, would enable improvements in image quality beyond those possible with other multisegmental approaches of which we were aware, that were investigated prior to this study (15). A description of the reconstruction algorithm is given in the Appendix. Low-pitch helical CT is performed to image the heart with z-axis redundancy in coverage over multiple cardiac cycles. The range of angles used in acquisition of a voxel, from the first to the last acquisition angle, defines the illumination window (Fig 2). This window determines the amount of data that can be used for the reconstruction of the voxel. The algorithm adaptively uses the maximum number of cardiac cycles available in the illumination window for the reconstruction of that voxel (16). A particular phase of the cardiac cycle is chosen to be the center of image generation. Projection data from several cardiac cycles acquired around this phase are used in the reconstruction, with the number of cycles determined by the width of the illumination window (Fig 2). The best temporal resolution is obtained by using a gating window for each cardiac cycle, with a gating window width that is minimal but allows the reconstruction of a data set (15). The temporal resolution is defined separately for every voxel in the volume. The order of segmentation or number of cardiac cycles used in reconstruction of a certain subvolume depends on the pitch setting in combination with the heart rate (15). With the heart rate remaining constant, the pitch or the table feed will determine how many cardiac cycles are available for an illumination window of a given width. With pitch remaining constant, more cardiac cycles become available with increasing heart rate. There are two periodically moving objects involved in the imaging process: the gantry, which circles the patient at a maximum speed of one rotation every 0.42 second, and the heart, which beats at a variable rate. The motion of these two objects may be correlated and synchronous, in accordance with so-called frequency harmonics, or may be decorrelated and asynchronous. Segmentation 88 Radiology January 2005 Hoffmann et al

4 over multiple cardiac cycles is of maximum benefit when motion is fully asynchronous, but it provides little advantage for temporal resolution when motion occurs in harmonic cycles. With multicyclic segmentation, therefore, the curve for mean temporal resolution plotted against increasing heart rate has a wavelike appearance (Fig 1). Depending on the heart rate, temporal resolution varies between 95 and 210 msec (Fig 1). Because data were acquired continuously and retrospective gating was applied, reconstruction could be carried out by using data from any point within the cardiac cycle. Routinely, two data sets were reconstructed, with gating windows centered at 80% and 50% of the R-R cycle. These positions are referred to in this article as middiastolic (approximately 80% of the cycle) and end-systolic (approximately 50% of the cycle) reconstruction windows. Adjacent positions ( 5%) were used if no satisfactory results were achieved with these initial window positions. The transverse sections and volumerendered images were compared side by side at a dedicated workstation for CT image reconstruction and display (Mx- View, version 4.1; Philips Medical Systems, Cleveland, Ohio), and the image data set with the fewest motion artifacts was selected for further analysis. If satisfactory results could not be obtained for all coronary artery segments in the same cardiac phase, multiple phases were used for the analysis of the different branches or segments. Postprocessing and Analysis The CT data sets were postprocessed at the workstation described earlier. Image analysis was performed by using multiplanar reformation (MPR), maximum intensity projection, volume rendering, and curved MPR techniques, in addition to transverse source image data. Two investigators (M.H.K.H., with 3 years of experience, and H.S., with 1 year of experience, in coronary CT angiography) assessed each coronary artery segment according to guidelines of the American College of Cardiology and the American Heart Association (17), and a decision was achieved by means of direct consensus reading. A first evaluation was performed to assess the possibility of obtaining motion-free images. Transverse image stacks were controlled for continuous depiction (successively matched contours) of the coronary arteries by scrolling through the transverse image data sets at a Figure 2. Schematic of adaptive multicyclic postprocessing of CT image data. Illumination window and number of available phase points, which depend on variable patient heart rate and acquisition parameters, were determined for sample voxels (A C) in reconstructed volume. Trapezoids represent three illumination windows (corresponding to sample voxels A C) inside which available phase points (dotted vertical lines) determined from the ECG tracing, and corresponding cardiac gating windows (squared sinusoid shapes centered around the phase points), are indicated. Illumination weighting and cardiac weighting functions were combined on a per-voxel basis by using normalization (16). Width of illumination windows and number of phase points available within each window varied according to acquisition geometry and heart rate variations. viewing workstation. The assessment was repeated by using thin-slab maximum intensity projections (mean slab thickness, 3 mm; range, 1 5 mm) in coronal and sagittal imaging planes. Subsequently, curved MPR images were assessed for motion-free imaging (ie, no tattered or ragged delineation) of the coronary branches. Images were generated as follows: Volume-rendered images, with the volume of interest defined to exclude the rib cage and pulmonary structures, were oriented along the z-axis. A line was electronically drawn in the center of the lumen of all major coronary artery branches to generate curved MPR images. At least two curved MPR images were generated for each center line by using perpendicular viewing angles and planes. All three image postprocessing modalities (transverse image stacks, thin-slab maximum intensity projections, and curved MPRs) were used for subjective evaluation. Images were graded for quality of depiction of each vessel, with a five-point scale from 1 (best image quality) to 5 (worst image quality). The scores were defined as shown in Table 1. Curved MPR images and volume-rendered images with various quality scores are shown in Figures 3 and 4. Images with a score between 1 and 3 were considered adequately diagnostic for coronary assessment; those with a score of 4 or 5 were not. Scores 1 3 indicated motion-free delineation of the coronary arteries. ECG Data ECG tracings acquired during scanning were digitized and stored with the unprocessed CT data set. Each ECG tracing, from the start of scanning to the end, was retrospectively assessed for arrhythmic or premature beats (M.H.K.H.). Arrhythmic or premature beats were correlated with the corresponding z-axis imaging plane for assessment of motion artifact induction. An R-wave peak trigger was generated according to a voltage threshold that was automatically adjusted on the basis of the patient s ECG data. The distance in milli- Volume 234 Number 1 16 Detector Row CT Coronary Angiography: Heart Rate 89

5 TABLE 1 Subjective Scoring of Image Quality Score seconds between two R-wave peaks was calculated and plotted against scanning time. Mean R-R interval and standard deviation were calculated for the total scanning duration. A -blocker was not applied in any of the examinations. (Patients already receiving a -blocker [n 19] continued with their established regimen.) Conventional Angiographic Data Images obtained at conventional angiography were postprocessed at the workstation by using software (efilm, version 1.8.3; Merge efilm, Milwaukee, Wis). Image display was calibrated on the basis of the dimensions of the catheter tip (4 6 F) in the coronary artery ostium. Angiograms were evaluated in consensus by a radiologist (M.H.K.H., with 8 years of experience in coronary angiogram reading) and a consulting physician from the cardiology service. All segments of the coronary artery tree with a reference diameter of 1.5 mm or greater were included in the study. Reference diameter was measured in the midsection of each segment, or, in the presence of stenosis in the midsection of the segment, proximal to the stenosis. Comparisons of CT and conventional angiographic data were performed separately for the subgroup of images that were motion artifact free according to primary subjective analysis. Images in this subgroup were evaluated for completeness of coverage of distal segments and side branches with a reference diameter of at least 1.5 mm. This was done by quantifying the segment diameter on the conventional angiogram and by verifying that all segments with a diameter of 1.5 mm or greater were contrast enhanced on the CT angiogram. In the subgroup of images with quality scores of 4 or 5 (ie, those on which motion artifacts were seen), all arterial segments for which depiction was affected by motion artifacts were reviewed and Characteristics 1 (best) No artifacts; excellent vessel delineation 2 Minimal increase of noise or minor artifacts not compromising diagnostic image quality 3 High noise or compromised distal coverage, but adequate (motion-free) access to the proximal vessels 4 Motion artifacts impeding assessment of single segments 5 (worst) Motion artifacts impeding assessment of multiple segments Note. Scores 1 3 encompass motion-free imaging. evaluated for the presence of stenotic lesions or occlusions that may mimic artifacts. Patients with no available conventional angiographic data (n 8) were not included in the secondary evaluation. To avoid verification bias, we did not include the results of stenosis evaluation in the study analysis. Statistical Analysis Continuous variables were expressed as mean standard deviation. A P value of less than.05 indicated a statistically significant difference. Descriptive statistics were stratified according to image quality score in five groups. The Fisher exact test was used to evaluate categorical data (distribution of the sexes among image quality groups). The Kruskal-Wallis test was used to evaluate quantitative parameters (age, heart rate) stratified across the five image quality scores. The Wilcoxon two-sample test was used to compare the selection of phases according to heart rate. Pearson correlation analysis was used to compare the average image quality score for all segments with the heart rate. A linear regression equation was used to calculate heart rate threshold values for future -blocker application. Retrospective power analysis was performed to test the correlation coefficient obtained against no correlation (r 0) and low correlation (r 0.30) presumed for the general population. Power analysis was conducted for the sample size of 50 patients. All statistical analyses were performed with software (SAS, version 8.02; SAS Institute, Cary, NC). RESULTS There were no statistically significant variations in patient age (P.3) among the five image quality groups. There were fewer women in the group with an image quality score of 1 (P.012) than was expected on the basis of assumed equal distribution of the sexes. The mean CT acquisition time was 19.5 seconds (range, seconds). CT Image Reconstruction The two cardiac phases (50% and 80% of the R-R interval) initially used for reconstruction of image data for all patients were sufficient to generate motionfree images for 40 (80%) of 50 patients. Additional phases had to be reconstructed for 10 (20%) patients. For one patient, an additional middiastolic reconstruction was performed at 75% of the R-R interval to achieve best image quality (mean heart rate, 67.3 beats per minute). For nine patients, additional reconstructions were performed in the endsystolic phase, at 45% 60% of the R-R interval, which resulted in motion-free delineation of the coronary arteries in three patients but not in the other six patients. The mean heart rate in patients for whom additional reconstruction phases were required was 76.4 beats per minute 5.4. All end-systolic phase frames were summarized at 50% of the R-R interval, and all middiastolic frames, at 80%. A plot of heart rates against phase selection (Fig 5) showed that the middiastolic phase (80% of the R-R interval) yielded the best results at lower heart rates (mean, 63.2 beats per minute 8.2) and the end-systolic phase (50% of the R-R interval) yielded the best results at higher heart rates (mean, 82.4 beats per minute 8.8; P.05). Image Evaluation A negative linear correlation was found between image quality and heart rate (Pearson correlation coefficient r 0.80, P.001) (Fig 6). Image quality scores of 1 3 (which indicate motion artifact free delineation of the coronary arterial tree) were achieved for all 39 patients (100%) with heart rates of less than 80 beats per minute. A separate comparison of CT and conventional angiographic data for this subgroup was performed. Conventional angiograms were available for 31 (80%) of these 39 patients. For all patients (n 11) with a subjective image quality score of 1, CT and conventional angiograms showed matching results for depiction of all coronary artery segments with a diameter of 1.5 mm or greater. Images from 11 other patients with a subjective image quality score of 2 showed adequate coverage of all coronary artery segments with a diameter of at least 1.5 mm. CT 90 Radiology January 2005 Hoffmann et al

6 Figure 3. Curved MPR images of right coronary artery. (a) Image with quality score of 1 shows minimal noise, well-defined side branches (short arrow), and reduced blooming artifacts (long arrow) around metallic stent surrounded by calcified plaque in right coronary artery ostium. (b) Image with quality score of 2 shows minimal banding artifacts that do not compromise diagnostic interpretation, as well as clearly discernible plaque (arrow) with outer layer of calcium and inner layer of fibrous tissue. (c) Image with quality score of 3 shows more noise than images scored 1 and 2, as well as compromised distal segment coverage (arrow). (d) Image with quality score of 4 shows single arterial segment with motion- and stentrelated artifacts (arrow) that prevent complete diagnostic evaluation of lumen. (e) Image with quality score of 5 shows motion-related artifact that impedes assessment of more than one arterial segment. images from one patient with a subjective image quality score of 2 showed incomplete coverage of distal artery segments compared with the reference image data from conventional angiography. Images from four (50%) of eight patients received a quality score of 3 because of inadequate opacification and delineation of the distal coronary artery segments. Segments with a total diameter of more than 1.5 mm that were inadequately depicted on the reference angiograms were the right coronary artery (n 6), left circumflex artery (n 5), and side branches of the left anterior descending artery (n 4). Thus, the rate of assessability of coronary artery segments with a diameter greater than 1.5 mm in this subgroup was 97% (468 segments delineated on the CT angiogram, compared with 483 segments identified on the conventional angiogram) (Table 2). Among the 11 patients with a mean heart rate of more than 80 beats per minute, motion-free delineation of the coronary artery tree was possible in six (55%). None of the motion-related artifacts classified during primary evaluation of CT images from the other five (45%) of 11 patients were due to occlusion or stenosis on the corresponding conventional angiograms. Table 2 shows patient demographic data and lists the segments that were scored as assessable at subjective image quality evaluation and the results of secondary evaluation of CT image data against conventional angiographic data. Overall, 581 (88%) of 658 artery segments were assessable. There was no significant variation in assessability over the three major vessel territories: the right Volume 234 Number 1 16 Detector Row CT Coronary Angiography: Heart Rate 91

7 Figure 4. Volume-rendered images. (a) Image with quality score of 1 (right anterior oblique view) shows good delineation of descending branch of right coronary artery (long arrow), without motion artifacts, and adequate depiction of distal left anterior descending artery (short arrow). (b) Image with quality score of 3 (oblique caudal view) provides inadequate depiction of distal one-third of right coronary artery (posterior descending artery) (arrow) along diaphragmatic surface of heart. (c) Image with quality score of 5 (right anterior oblique view), obtained in patient with bigeminal rhythm, shows banding artifacts (arrows) that resulted from premature beats in alternate cardiac cycles. coronary artery (167 [86.5%] of 193), left main and left anterior descending artery (199 [86.5%] of 230), and left circumflex artery (138 [87.3%] of 158) (P.5). The cause of nonassessability was cardiac motion related artifact in 55 (8.4%) of 658 coronary artery segments. Blooming artifacts due to calcifications, and streak artifacts induced by coronary stents or intracardiac pacing leads, caused nonassessability in 18 (2.7%) and four (0.6%) of 658 segments, respectively. Artifacts that interfered with coronary artery segment assessment in one patient were caused by intracardiac pacing leads. In all other postsurgical patients (n 6), artifacts either were not present or did not overlie coronary artery segments. The two readers who performed image quality assessment (M.H.K.H., H.S.) achieved immediate agreement in 36 (62%) of the 50 cases. The consensus discussion about the remaining 14 (28%) was primarily concerned with the discrimination of image quality scores 1 and 2. To establish heart rate thresholds for noninvasive coronary angiography, a regression analysis was performed (Fig 7). According to the linear regression equation, motion-free images should be possible in patients with a maximum heart rate of 100 beats per minute. Application of the linear equation to images with a quality score of 2 yields a heart rate threshold of 74.2 beats per minute. In most patients with heart rates at or below this level, good to excellent image quality is achievable by using the protocol used in this study. Retrospective power analysis was conducted by using the correlation coefficient obtained during linear regression, r 0.67 (95% confidence interval: 0.48, 0.80). With a sample size of n 50, the study achieved a power of 99% (no correlation [r 0] assumed for the general population) and 93% (low correlation [r 0.30] assumed for the general population). Arrhythmia In the study group, nine patients had premature ventricular beats during the image acquisition. No apparent motion artifacts could be related to these arrhythmic beats in eight (89%) of these patients. One patient with atrial fibrillation and a heart rate within the range of 60 to 90 beats per minute was included in the study. His overall image quality score was 2. Only slight banding artifacts were discernible (Fig 8). These did not compromise the diagnostic assessment of the coronary tree. One patient developed a bigeminal heart rhythm pattern during scanning. This patient s image quality score was 5; motion artifacts did not al- 92 Radiology January 2005 Hoffmann et al

8 low assessment of the coronary arteries (Fig 4c). Radiology DISCUSSION Residual cardiac motion artifacts remain a major cause of image degradation in multi detector row CT coronary angiography (7,8,18). The right coronary artery appears particularly vulnerable because of its extensive motion radius and short motion-free period (10,14). An inverse relation between heart rate and image quality was reported previously (8,18). Sixty-five beats per minute was defined as the upper heart rate threshold with which it is possible to achieve motionfree image quality (7). It is therefore common practice to use -blocking agents to maintain heart rates at a level below that threshold during image acquisition (3,19 21). These study results and practices are based on data acquired with four detector row CT scanners. Previously reported study results of high sensitivity and specificity for noninvasive coronary angiography with a 12 detector row scanner were based on the same thresholds as those established for the generation of four detector row scanners (3,22). These studies relied on two-dimensional section-oriented reconstruction algorithms. Results of the current study, obtained with a 16 detector row scanner and true 3D voxel-based reconstruction algorithms, indicate that a slightly higher threshold of 80 beats per minute can be safely applied. Motionfree images were available in 100% of patients with a heart rate of less than 80 beats per minute. Segmental reconstruction and analysis of CT image data, referenced against invasive coronary angiograms, provided motion-free depiction and assessability of 97% of coronary artery segments in all patients with a heart rate of less than 80 beats per minute. Good image quality is achievable at low heart rates by using a mid- to end-diastolic phase frame and shifting to end-systolic frames as heart rate increases. Figure 1 shows the temporal resolution achieved by using a 3D multicyclic reconstruction algorithm (9,16) versus diastolic and systolic reconstruction windows. According to theoretical considerations, both the middiastolic (80% of R-R interval) and the end-systolic (50% of R-R interval) reconstruction windows should coincide with low-motion periods in the coronary artery tree to allow motion-free imaging. The plot shown in Figure 1 was constructed according to the approximation Figure 5. Plot shows mean heart rate (beats per minute) during scanning, according to selection of optimal cardiac phase (percentage of R-R interval) for reconstruction. The phase that provided best overall image quality and motion-free delineation of coronary arteries was chosen for every patient. Reconstruction of image data from additional phases was necessary to achieve motion-free images in 10 patients. Additional data were combined with data from either the middiastolic phase selection (80% of R-R interval) or the end-systolic phase selection (50% of R-R interval), according to which phase was most proximate. Only one patient required a combination of data from both phases to achieve motion-free imaging (50% of R-R interval, for visualization of right coronary artery, and 80% of R-R interval, for depiction of left coronary artery). Mean heart rate was 75 beats per minute both for patients in whom the middiastolic phase was selected and in those for whom the end-systolic phase was selected. Whiskers show overlap in selection between the two phases at beats per minute. Figure 6. Plot shows mean heart rate (beats per minute) during scanning, according to reconstructed image quality score. Image quality was related to heart rate, with lower heart rates correlating with better image quality. Images obtained in all patients with a mean heart rate of 80 or fewer beats per minute during scanning were motion free. Six (12%) of 50 CT examinations in patients with a mean heart rate of more than 80 beats per minute resulted in motion-compromised images. Dotted line represents the threshold between acceptable and unacceptable image quality. mean, median, box first to third quartile, whisker minimummaximum value, outlier. Volume 234 Number 1 16 Detector Row CT Coronary Angiography: Heart Rate 93

9 TABLE 2 Patient Data, Relevant Coronary Artery Segments, and Causes of Nonassessability A: Patient Data Radiology Parameter Total Image Quality Score No. of patients Heart rate (beats per minute) Mean standard deviation Range Age (y) (mean standard deviation) Sex Men Women B: Relevant Segments* Parameter Total Image Quality Score All arterial segments 581/658 (88.3) RCA segments 1 4,6 193/219 (88.1) LAD artery segments /261 (88.1) LCX artery segments /178 (88.8) C: Causes of Nonassessability at CT Parameter Total Image Quality Score Calcification 18 (2.7) (1.5) 8 (1.2) Cardiac motion 55 (8.4) (3.8) 30 (4.6) Other causes 4 (0.6) (0.6)... D: Conventional Angiographic Data Parameter Total Image Quality Score Available angiograms Assessable segments 468/483 (96.9) 157/157 (100) 153/156 (98.0) 158/170 (92.9) True motion artifacts /5 (100) 22/22 (100) Note. LAD left anterior descending, LCX left circumflex, RCA right coronary artery. * LAD, LCX, and RCA segment numbers were assigned according to the current American College of Cardiology American Heart Association consensus standard (17). Data are number of segments available for primary subjective image quality scoring. Numbers in parentheses are percentages. Data are number of nonassessable segments. Numbers in parentheses are percentages. Numbers are fraction of segments with a diameter 1.5 mm assessable at secondary analysis (comparison of multi detector row CT images against angiograms). Numbers in parentheses are percentages. Numbers are fraction of motion artifacts observed on CT images that were found to be normal (no stenosis or wall irregularities) on conventional angiograms. Numbers in parentheses are percentages. of phase duration by Herzog et al (23). The point at which a shift from the diastolic frame to the systolic frame becomes necessary occurs at approximately 65 beats per minute. The results of phase selection in our patient group, however, suggest that the shift from the middiastolic phase (at low heart rates) to the endsystolic phase (at higher heart rates) should take place at beats per minute. To explain this discrepancy, we can only postulate that the diastolic phase duration according to the theoretical considerations of Herzog et al was underestimated (23). Besides motion artifacts, other factors, such as the presence of high-attenuation structures adjacent to the coronary arteries, may impede accurate lumen assessment. Calcified plaque and extensive calcification of coronary artery walls appear oversized on CT images (24). This socalled blooming artifact is supposedly related to partial-volume averaging effects caused by insufficient spatial resolution. With 16 detector row CT, extensive calcifications prevented segment assessment to a lesser extent than with four detector row CT (in 2.7% vs 5.0% of segments) (8). The validity of this comparison, however, is limited by the lack of available data with regard to calcium prevalence stratification in the compared groups. Nevertheless, the comparison of these results indicates that the slight improvement in spatial resolution achievable with the 16 detector row scanner may reduce blooming artifacts to some extent. Compared with voxel size for four detector row scanners, which is slightly greater than 1 mm in all dimensions, effective voxel size for 16 detector row scanners is reduced to mm. High-attenuation objects inside the cardiac cavities (eg, pacing lead tips) or adjacent to the cardiac structures (eg, sternal wires after surgery or metallic 94 Radiology January 2005 Hoffmann et al

10 clips adjacent to mammary pedicles after coronary artery bypass grafting) are a further potential source of image artifacts. These artifacts, however, were sufficiently mitigated with cone-beam correction included in the reconstruction algorithm (9). Comparison with Other Studies A regression plot similar to that shown in Figure 7, relating qualitative image scores to heart rate, is available in a published report of a four detector row scanner study. Hong et al (18) calculated an upper heart rate limit of 74.5 beats per minute for achievement of motion-free images with the four detector row scanner. If the same method of calculation used by Hong et al were applied to the data set in this study (ie, if the worst image score associated with motion-free image quality were entered into the linear regression equation), we would come up with a threshold of 100 beats per minute. However, a motion-free image at this quality level would not offer the best diagnostic value, as it would include either compromised distal coverage or blurred lumen definition because of higher noise levels. To obtain optimal diagnostic results, images must have a quality score of 2 or higher. If this image quality level is factored into the regression equation, then we find that optimal diagnostic results are achievable at heart rates of up to 75 beats per minute. In other words, regarding the maximization of diagnostic image quality, the results with 16 detector row scanners do not seem to differ from those with four detector row scanners. As we do not have access to the image database of Hong et al, however, we cannot determine whether overall image quality in our group was improved. Other studies performed with four detector row CT scanners indicate that patients must have a heart rate of no more than 65 beats per minute to obtain motion-free images (7). This was the threshold extensively accepted by others (1,3,20,25). The approach recommended by Kopp et al (26), the reconstruction of multiple phases for every patient, does not seem necessary with the protocol used in this study. Although reconstruction time has been tremendously reduced, the amount of data generated may be difficult to handle, both for the radiologist and for the computerized network server or storage hardware. The data obtained in this study suggest that motion-free imaging is reliably possible in patients with a heart rate of less Figure 7. Linear regression plot of mean image quality score per patient (y-axis) against mean heart rate during CT scanning (x-axis). Dashed lines represent 95% confidence limits. Linear correlation (equation given) indicates that good to excellent image quality can be achieved at heart rates of 74.2 or fewer beats per minute (bpm) (rectangle for image quality score of 2) and that motion-free images with a quality score of 3 or better should be obtainable at 100 beats per minute (rectangle for image quality score of 3). single patient represented by mean image quality over all segments (y-axis) plotted against mean heart rate during scan acquisition (x-axis). Figure 8. Curved MPR image (left anterior oblique view; window level, 130 HU; window width, 490 HU) shows graft of right internal thoracic artery to middle left anterior descending segment of left coronary artery. Region of anastomosis (thin arrow) is well assessable. Clip (arrowhead) on mammary pedicle distal to anastomosis, and calcified plaque (thick arrow) in proximal one-third of left anterior descending artery, near branching point of left circumflex artery ( ), also are well depicted. ECG tracing showed atrial fibrillation and variation of heart rate from 60 to 90 beats per minute during scanning. Inset: Banding artifacts in pedicle and anastomotic region compromise image quality only slightly. Volume 234 Number 1 16 Detector Row CT Coronary Angiography: Heart Rate 95

11 than 80 beats per minute; the results of regression analysis and theoretical considerations suggest much higher thresholds. One factor that might explain this discrepancy between our image data and the results of our statistical analysis is table feed. In this study, we used a fixed table feed of 6.86 mm/sec, whereas the approach currently used to optimize temporal resolution is to vary the table feed settings. The adaptive multicyclic reconstruction algorithm is not discrete; it uses as many data as possible, depending on factors such as heart rate and table feed (9). With heart rate adapted table feed settings, we might therefore in the future be able to achieve motion-free imaging at higher heart rates, as indicated by the results of regression analysis. Limitations Figure A1. Diagram of actual cone-beam geometry at helical CT (left) and virtual beam geometry after parallel rebinning (right). The approach we used in reconstructing image data of the coronary arteries was an adaptation of the simple approach suggested by Achenbach et al (10), who investigated the in-plane motion of the coronary arteries by using high-resolution electron-beam CT (50- msec temporal resolution); the pattern of motion in the coronary arteries, however, is truly 3D (13,14). True 3D motion was not accounted for in the phase selection method used in this study. The simple two-dimensional approach of Achenbach et al proved sufficient in 80% of cases, and phase selection had to be empirically modified only in the remaining 20%. We were not able to apply an absolute or relative phase location algorithm for comparison with the physiologic phase selection algorithm. The overall performance in patients with rate-contained arrhythmia (ie, atrial fibrillation not exceeding a heart rate of beats per minute in any single cardiac cycle), and the good results with regard to sporadic premature heartbeats, indicate a potential advantage with use of the physiologic phase-delay algorithm. Additional studies are warranted to assess this effect more accurately. Another limitation of the current study was the incomplete conventional angiographic data set. Eight patients who were scheduled to undergo conventional coronary angiography after multi detector row CT refused the invasive procedure because of negative findings at CT. They were informed that the literature shows a 97% negative predictive value for CT angiographic findings (3,22). They subsequently decided to undertake no further diagnostic studies. Stenosis detection assessment based on this incomplete data set could be biased and was therefore omitted from the study design. Implications for Coronary Angiographic Imaging The temporal resolution of images obtained with 16 detector row CT and reconstruction based on a single cardiac cycle is only slightly improved compared with that of images obtained with four detector row CT, with improvement attributable solely to an increase in gantry rotation speed from 0.50 second to 0.42 second. Further improvements gained with multicyclic reconstruction are translated into substantial increases in resolution only at certain heart rates (9). Because of highly variable heart rates even during the short breath-hold duration applied in the current study, any strategy to optimize temporal resolution on the basis of a single fixed heart rate will fail. Variability in temporal resolution due to inconsistent heart rate during the breath hold may explain the only slight total increase in the maximum heart rate threshold for motion-free imaging with 16 detector row scanners compared with four detector row scanners. Relative to a normal resting heart rate range of beats per minute (27), however, a maximum heart rate threshold of 80 beats per minute will allow motion-free imaging without -blocker application in the majority of patients with a normal sinus rhythm. The preselection of heart phases as suggested by the data of Achenbach et al proved a very effective means of optimizing the postprocessing strategy. The use of the dynamic delay model and socalled physiologic phases enabled the accommodation of premature heartbeats and rate-contained arrhythmia without user interference. Whether this method offers advantages over currently applied simple absolute or relative phase-delay algorithms remains to be shown in future studies. The data from this study suggest that -blocker induced reduction of the resting heart rate to 65 or fewer beats per minute may not be necessary for coronary CT angiography performed with 16 detector row scanners and multicyclic reconstruction, which provide consistent motion-free image quality at heart rates up to 80 beats per minute. Because the heart rate increases slightly during scanning in most patients, however, it is suggested that the initial heart rate be reduced to less than 75 beats per minute to include a safety margin and to optimize image quality. APPENDIX With the increasing number of detector rows, dedicated 3D cone-beam reconstruction algorithms have become required instead of adapted two-dimensional reconstruction schemes. In contrast to twodimensional section-based reconstruction schemes, the acquisition geometry is accounted for by using 3D backward projection along the true paths of the rays. With the extended cardiac reconstruction framework (15), an approximate 3D backwardprojection algorithm has become available for ECG-gated helical cone-beam CT. The principles of this reconstruction technique are outlined here. As a first step, the helical projection data are re-sorted to create a virtual parallelbeam geometry (see Figure A1). This process is referred to as parallel rebinning (15). Afterward, the projection data are weighted with the cosine of the cone-angle and filtered with a one-dimensional band-limited 96 Radiology January 2005 Hoffmann et al

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Cardiac CT - Coronary Calcium Basics Workshop II (Basic) Cardiac CT - Coronary Calcium Basics Workshop II (Basic) J. Jeffrey Carr, MD, MSCE Dept. of Radiology & Public Health Sciences Wake Forest University School of Medicine Winston-Salem, NC USA No significant

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial Poster No.: C-0595 Congress: ECR 2013 Type: Authors:

More information

Coronary angiography is the standard way of visualizing

Coronary angiography is the standard way of visualizing Clinical Investigation and Reports Coronary Artery Fly-Through Using Electron Beam Computed Tomography Peter M.A. van Ooijen, MSc; Matthijs Oudkerk, MD, PhD; Robert J.M. van Geuns, MD; Benno J. Rensing,

More information

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing Cademartiri et al. Heart Rhythm Irregularities on MDCT Angiography Cardiac Imaging Original Research A C M E D E N T U R I C A L I M A G I N G AJR 2006; 186:634 638 0361 803X/06/1863 634 American Roentgen

More information

CT angiography techniques. Boot camp

CT angiography techniques. Boot camp CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary

More information

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital Coronary Artery Imaging Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital Larger array : cover scan area Detector size : spatial resolution Rotation speed : scan time Retrospective

More information

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population Journal of the American College of Cardiology Vol. 49, No. 9, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.10.066

More information

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology,

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology, Chapter 4 Impact of Coronary Calcium Score on Diagnostic Accuracy of Multislice Computed Tomography Coronary Angiography for Detection of Coronary Artery Disease Gabija Pundziute, 1,3 Joanne D. Schuijf,

More information

2004;77:800 4 MSCT OF CORONARY ARTERY BYPASS GRAFTS. Results. CABG With Adequate Diagnostic Quality

2004;77:800 4 MSCT OF CORONARY ARTERY BYPASS GRAFTS. Results. CABG With Adequate Diagnostic Quality Isotropic Half-Millimeter Angiography of Coronary Artery Bypass Grafts With 16-Slice Computed Tomography Marc Dewey, MD, Alexander Lembcke, MD, Christian Enzweiler, MD, Bernd Hamm, MD, and Patrik Rogalla,

More information

Adapted Transfer Function Design for Coronary Artery Evaluation

Adapted Transfer Function Design for Coronary Artery Evaluation Adapted Transfer Function Design for Coronary Artery Evaluation Sylvia Glaßer 1, Steffen Oeltze 1, Anja Hennemuth 2, Skadi Wilhelmsen 3, Bernhard Preim 1 1 Department of Simulation and Graphics, University

More information

Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats

Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats Amolak Singh 1*, Yash Sethi 1, Sonya Watkins 1, Angela Youtsey 1, Angie Thomas 1 1. Department of Radiology, University

More information

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography Journal of the American College of Cardiology Vol. 46, No. 3, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.05.056

More information

Detection of Coronary Artery Stenoses with Thin Slice Multi Detector Row Spiral Computed Tomography Angiography

Detection of Coronary Artery Stenoses with Thin Slice Multi Detector Row Spiral Computed Tomography Angiography Med. J. Cairo Univ., Vol. 79, No. 1, Sep. 389-406, 2011 www.medicaljournalofcairouniversity.com Detection of Coronary Artery Stenoses with Thin Slice Multi Detector Row Spiral Computed Tomography Angiography

More information

General Cardiovascular Magnetic Resonance Imaging

General Cardiovascular Magnetic Resonance Imaging 2 General Cardiovascular Magnetic Resonance Imaging 19 Peter G. Danias, Cardiovascular MRI: 150 Multiple-Choice Questions and Answers Humana Press 2008 20 Cardiovascular MRI: 150 Multiple-Choice Questions

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Junfu Liang 1,2, Hui Wang 1, Lei Xu 1, Li Dong 1, Zhanming

More information

Recent developments in cardiac CT

Recent developments in cardiac CT REVIEW Recent developments in cardiac CT With the introduction of 64-multidetector row CT, coronary CT angiography has become a clinical tool, owing to improved image quality and reduced breath-hold time,

More information

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3 Coronary Artery Bypass Graft Imaging with CT Angiography and Iterative Reconstruction: Quantitave Evaluation of Radiation Dose Reduction and Image Quality Simon Nepveu 1, Irina Boldeanu 1, Yves Provost

More information

Bottom up cardiac CT for CABG assessment to resolve breathing artefact

Bottom up cardiac CT for CABG assessment to resolve breathing artefact Bottom up cardiac CT for CABG assessment to resolve breathing artefact Poster No.: C-0589 Congress: ECR 2010 Type: Educational Exhibit Topic: Cardiac Authors: P. Glass, P. Donnelly, P. Hanley, D. Higginson,

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

Cardiac CTA without and with IV Contrast

Cardiac CTA without and with IV Contrast Cardiac CTA without and with IV Contrast IMG 12122 Cardiac Prospective 12123 Cardiac Retrospective 12119 Cardiac ED Chest Pain 12150 Calcium Score A B A B Toshiba AquilionONE Toshiba AquilionPRIME Calcium

More information

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Mathieu Sabbagh, R3 Michigan State University Radiology Garden City Hospital

More information

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine Non Contrast MRA Mayil Krishnam Director, Cardiovascular and Thoracic Imaging University of California, Irvine No disclosures Non contrast MRA-Why? Limitations of CTA Radiation exposure Iodinated contrast

More information

Noncoronary Cardiac MDCT

Noncoronary Cardiac MDCT Noncoronary Cardiac MDCT David A. Bluemke, M.D., Ph.D. Professor, of Radiology and Medicine Johns Hopkins University School of Medicine Baltimore, Maryland Toshiba Disclosures Grant support Noncoronary

More information

Interpreting CT Angiography: Three-Dimensional Reconstruction Techniques

Interpreting CT Angiography: Three-Dimensional Reconstruction Techniques 4 Interpreting CT Angiography: Three-Dimensional Reconstruction Techniques Matthew J. Budoff This chapter will take you through the performance and evaluation of the non-invasive coronary angiogram. Using

More information

Voxar 3D CardiaMetrix. Reference Guide

Voxar 3D CardiaMetrix. Reference Guide Voxar 3D CardiaMetrix Reference Guide The software described in this document is furnished under a license, and may be used or copied only according to the terms of such license. Toshiba means, Toshiba

More information

Index. radiologic.theclinics.com. Note: Page numbers of article titles are in boldface type.

Index. radiologic.theclinics.com. Note: Page numbers of article titles are in boldface type. Index Note: Page numbers of article titles are in boldface type. A ALCAPA. See Anomalous left coronary artery from the pulmonary artery. Angiosarcoma computed tomographic assessment of, 809 811 Anomalous

More information

Cardiac Imaging Tests

Cardiac Imaging Tests Cardiac Imaging Tests http://www.medpagetoday.com/upload/2010/11/15/23347.jpg Standard imaging tests include echocardiography, chest x-ray, CT, MRI, and various radionuclide techniques. Standard CT and

More information

What every radiologist should know about cardiac CT: A case-based pictorial review

What every radiologist should know about cardiac CT: A case-based pictorial review What every radiologist should know about cardiac CT: A case-based pictorial review Poster No.: C-0555 Congress: ECR 2010 Type: Educational Exhibit Topic: Cardiac Authors: C. M. Capuñay, P. Carrascosa,

More information

THE ROLE OF HIGH END MULTI DETECTOR CT IN CORONARY IMAGING ESSAY

THE ROLE OF HIGH END MULTI DETECTOR CT IN CORONARY IMAGING ESSAY THE ROLE OF HIGH END MULTI DETECTOR CT IN CORONARY IMAGING ESSAY Submitted for partial fulfillment of Master degree in Radiodiagnosis By Ahmed Yehia Ahmed (M.B.B.Ch., Cairo University) Supervisors Prof.

More information

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT Eur Radiol (2006) 16: 1964 1972 DOI 10.1007/s00330-006-0262-x CARDIAC Sebastian Leschka Lars Husmann Lotus M. Desbiolles Oliver Gaemperli Tiziano Schepis Pascal Koepfli Thomas Boehm Borut Marincek Philipp

More information

Coronary Artery Anomalies from Birth to Adulthood; the Role of CT Coronary Angiography in Sudden Cardiac Death Screening

Coronary Artery Anomalies from Birth to Adulthood; the Role of CT Coronary Angiography in Sudden Cardiac Death Screening Coronary Artery Anomalies from Birth to Adulthood; the Role of CT Coronary Angiography in Sudden Cardiac Death Screening E O Dwyer 1, C O Brien 1, B Loo 1, A Snow Hogan 1, O Buckley1 2, B 1. Department

More information

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert University of Groningen Quantitative CT myocardial perfusion Pelgrim, Gert IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Cardiac CT David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Courtesy Tom Kracus Courtesy Kettering Tom Medical Kracus Cente Kettering Medical Center 2003-2006, David Stultz, MD Courtesy Tom Kracus

More information

Computer Assisted Coronary CT Angiography Analysis

Computer Assisted Coronary CT Angiography Analysis Linköping Studies in Health Sciences, Thesis No. 96 Computer Assisted Coronary CT Angiography Analysis Disease-Centered Software Development Chunliang Wang Department of Radiology And Center for Medical

More information

Dr Felix Keng. Imaging of the heart is technically difficult because: Role of Cardiac MSCT. Current: Cardiac Motion Respiratory Motion

Dr Felix Keng. Imaging of the heart is technically difficult because: Role of Cardiac MSCT. Current: Cardiac Motion Respiratory Motion Siemens Philips Dr Felix Keng GE Toshiba Role of Cardiac MSCT Current: Structural / congenital heart imaging Extra-cardiac / Great vessel imaging Volumes and ejection fractions (cine + gating) Calcium

More information

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography African Journal of Biotechnology Vol. 10(19), pp. 3854-3858, 9 May, 2011 Available online at http://www.academicjournals.org/ajb DOI: 10.5897/AJB10.983 ISSN 1684 5315 2011 Academic Journals Full Length

More information

TITLE: Multi-Slice Computed Tomography Coronary Angiography for Coronary Artery Disease: A Review of the Clinical Effectiveness and Guidelines

TITLE: Multi-Slice Computed Tomography Coronary Angiography for Coronary Artery Disease: A Review of the Clinical Effectiveness and Guidelines TITLE: Multi-Slice Computed Tomography Coronary Angiography for Coronary Artery Disease: A Review of the Clinical Effectiveness and Guidelines DATE: 25 February 2009 CONTEXT AND POLICY ISSUES: Coronary

More information

Assessment of Coronary Arteries with Total Study Time of Less than 30 Minutes by Using Whole-Heart Coronary MR Angiography

Assessment of Coronary Arteries with Total Study Time of Less than 30 Minutes by Using Whole-Heart Coronary MR Angiography Hajime Sakuma, MD Yasutaka Ichikawa, MD Naohisa Suzawa, MD Tadanori Hirano, MD Katsutoshi Makino, MD Nozomu Koyama, RT Marc Van Cauteren, PhD Kan Takeda, MD Published online before print 10.1148/radiol.2371040830

More information

Chapter 6. Hester Gietema Cornelia Schaefer-Prokop Willem Mali Gerard Groenewegen Mathias Prokop. Accepted for publication in Radiology

Chapter 6. Hester Gietema Cornelia Schaefer-Prokop Willem Mali Gerard Groenewegen Mathias Prokop. Accepted for publication in Radiology Chapter 6 Interscan variability of semiautomated volume measurements in intraparenchymal pulmonary nodules using multidetector-row computed tomography: Influence of inspirational level, nodule size and

More information

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011 Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011 Outline Background Hypothesis Study Population Methodology

More information

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference.

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference. Rotation: Imaging 1 Imaging 1 provides COCATS Level 1 experience for nuclear cardiology (including SPECT and PET) and cardiac CT. Fellows will administer, process, and read cardiac nuclear studies with

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design 2 Cardiovascular Computed Tomography: Current and Future Scanning System Design Wm. Guy Weigold Introduction The heart can be visualized in gross form on any standard chest computed tomography (CT), but

More information

Imaging congestive heart failure: role of coronary computed tomography angiography (CCTA)

Imaging congestive heart failure: role of coronary computed tomography angiography (CCTA) Imaging congestive heart failure: role of coronary computed tomography angiography (CCTA) Gianluca Pontone, MD, PhD, FESC, FSCCT Director of MR Unit Deputy Director of Cardiovascul CT Unit Clinical Cardiology

More information

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Testing the Accuracy of ECG Captured by through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Data Analysis a) R-wave Comparison: The mean and standard deviation of R-wave amplitudes

More information

An Introduction to Dual Energy Computed Tomography

An Introduction to Dual Energy Computed Tomography An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the

More information

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Original Article Open Access The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Ziqiao Lei 1, Jin Gu 2, Qing Fu 3, Heshui Shi 4, Haibo Xu 5, Ping Han 6, Jianming

More information

Fulfilling the Promise

Fulfilling the Promise Fulfilling the Promise of Cardiac MR Non-contrast, free-breathing technique generates comprehensive evaluation of the coronary arteries By Maggie Fung, MR Cardiovascular Clinical Development Manager; Wei

More information

Case 47 Clinical Presentation

Case 47 Clinical Presentation 93 Case 47 C Clinical Presentation 45-year-old man presents with chest pain and new onset of a murmur. Echocardiography shows severe aortic insufficiency. 94 RadCases Cardiac Imaging Imaging Findings C

More information

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16.

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16. NIH Public Access Author Manuscript Published in final edited form as: Acad Radiol. 2008 January ; 15(1): 40 48. doi:10.1016/j.acra.2007.07.023. Analysis of Heart Rate and Heart Rate Variation During Cardiac

More information

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006). Feasibility, image quality and radiation dose of coronary CT angiography (CCTA) in patients with atrial fibrillation using a new generation 256 multi-detector CT (MDCT) Poster No.: C-2378 Congress: ECR

More information

Cardiac CT Angiography

Cardiac CT Angiography Cardiac CT Angiography Dr James Chafey, Radiologist Why do we need a better test for C.A.D? 1. CAD is the leading cause of death in the US CAD 31% Cancer 23% Stroke 7% 2. The prevalence of atherosclerosis

More information

CT SCAN PROTOCOL. Shoulder

CT SCAN PROTOCOL. Shoulder CT SCAN PROTOCOL Shoulder Purpose and Summary CT images made with this protocol are used to provide the orthopedic surgeon with a detailed 3D anatomical reconstruction of the patient s scapula and proximal

More information

b. To facilitate the management decision of a patient with an equivocal stress test.

b. To facilitate the management decision of a patient with an equivocal stress test. National Imaging Associates, Inc. Clinical guidelines EBCT HEART CT & HEART CT CONGENITAL CCTA CPT4 Codes: 75571 EBCT 75572, 75573 Heart CT & Heart CT Congenital 75574 - CCTA LCD ID Number: L33559 J K

More information

Chapter. Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography

Chapter. Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography Chapter 3 Non-Invasive Coronary Imaging and Assessment of Left Ventricular Function using 16-slice Computed Tomography Joanne D. Schuijf, Jeroen J. Bax, Liesbeth P. Salm, J. Wouter Jukema, Hildo J. Lamb,

More information

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Poster No.: C-0863 Congress: ECR 2015 Type: Scientific Exhibit Authors: A. Nakamoto,

More information

Impaired Regional Myocardial Function Detection Using the Standard Inter-Segmental Integration SINE Wave Curve On Magnetic Resonance Imaging

Impaired Regional Myocardial Function Detection Using the Standard Inter-Segmental Integration SINE Wave Curve On Magnetic Resonance Imaging Original Article Impaired Regional Myocardial Function Detection Using the Standard Inter-Segmental Integration Ngam-Maung B, RT email : chaothawee@yahoo.com Busakol Ngam-Maung, RT 1 Lertlak Chaothawee,

More information

2 Synopsis Name and Address of Company: Bracco ALTANA Pharma GmbH Max-Stromeyer-Str. 116

2 Synopsis Name and Address of Company: Bracco ALTANA Pharma GmbH Max-Stromeyer-Str. 116 (For Bracco Regulatory Affairs (For National Authority Use only) Iomeron Iomeprol Title of Study: Comparison of Iomeprol 300 and Iomeprol 400 for the Efficiency of Autosegmentation in 64-Row Multi-Detector

More information

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography Supalerk Pattanaprichakul, MD 1, Sutipong Jongjirasiri, MD 2, Sukit Yamwong, MD 1, Jiraporn Laothammatas,

More information

Magnetic Resonance Angiography

Magnetic Resonance Angiography Magnetic Resonance Angiography 1 Magnetic Resonance Angiography exploits flow enhancement of GR sequences saturation of venous flow allows arterial visualization saturation of arterial flow allows venous

More information

Cardiac CTA Prospective Gating Broad Beam

Cardiac CTA Prospective Gating Broad Beam Cardiac CTA Prospective Gating Broad Beam ACQUISITION- Broad Beam Gating: Prospective Non Contrast Scan-Calcium Score Patient Position Supine Feet First into Gantry Heart Isocenter Scanogram AP and Lateral

More information

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography 633 CARDIOVASCULAR MEDICINE Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography A W Leber, A Knez, C Becker, A Becker, C White, C Thilo, M Reiser,

More information

Patient preparation and coronary CTA techniques. Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging

Patient preparation and coronary CTA techniques. Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging Patient preparation and coronary CTA techniques Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging Overview 1. Patient preparation 2. Scanning techniques Patient preparation Preparation

More information

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection Proceedings of the 2012 International Conference on Industrial Engineering and Operations Management Istanbul, Turkey, July 3 6, 2012 Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter

More information

Towards Cardiac C-Arm Computed Tomography Günter Lauritsch*, Jan Boese, Lars Wigström, Herbert Kemeth, and Rebecca Fahrig

Towards Cardiac C-Arm Computed Tomography Günter Lauritsch*, Jan Boese, Lars Wigström, Herbert Kemeth, and Rebecca Fahrig 922 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 7, JULY 2006 Towards Cardiac C-Arm Computed Tomography Günter Lauritsch*, Jan Boese, Lars Wigström, Herbert Kemeth, and Rebecca Fahrig Abstract Cardiac

More information

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING Non-invasive coronary angiography along with multidetector computed tomography or magnetic resonance imaging is attracting increasing interest

More information

Integrating CT in minimally invasive treatment of the coronary arteries

Integrating CT in minimally invasive treatment of the coronary arteries Integrating CT in minimally invasive treatment of the coronary arteries e-poster: C-200 Congress: ECR 2008 Type: Scientific Exhibit Topic: Cardiac / Miscellaneous MeSH: Keywords: Coronary Vessels [A07.231.114.269]

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

for Heart-Health Scanning

for Heart-Health Scanning Magnetocardiography for Heart-Health Scanning CardioMag Imaging, Inc. 1 Basic Principles of Magnetocardiography (MCG) The cardiac electric activity that produces a voltage difference on the body surface

More information

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview CARDIAC MRI Dr Yang Faridah A. Aziz Department of Biomedical Imaging University of Malaya Medical Centre Cardiovascular Disease Diseases of the circulatory system, also called cardiovascular disease (CVD),

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015 2384 Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients GUIRU HE, XIAOPEI LIU, YAN LIU, WEI

More information

Essentials of Clinical MR, 2 nd edition. 99. MRA Principles and Carotid MRA

Essentials of Clinical MR, 2 nd edition. 99. MRA Principles and Carotid MRA 99. MRA Principles and Carotid MRA As described in Chapter 12, time of flight (TOF) magnetic resonance angiography (MRA) is commonly utilized in the evaluation of the circle of Willis. TOF MRA allows depiction

More information

Studies with electron beam computed tomography (EBCT) Imaging

Studies with electron beam computed tomography (EBCT) Imaging Imaging Predictive Value of 16-Slice Multidetector Spiral Computed Tomography to Detect Significant Obstructive Coronary Artery Disease in Patients at High Risk for Coronary Artery Disease Patient- Versus

More information

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1

Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1 Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1 Ethan J. Halpern, MD, David J. Halpern, Jeffrey H. Yanof, PhD, Sigal Amin-Spector, PhD, David Fischman, MD, Galit Aviram,

More information

B. CT protocols for the spine

B. CT protocols for the spine B. CT protocols for the spine Poster No.: A-003 Congress: ECR 2010 Type: Invited Speaker Topic: Neuro Authors: B. Tins; Oswestry/UK Keywords: CT, spine, diagnostic imaging protocol DOI: 10.1594/ecr2010/A-003

More information

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography Journal of the American College of Cardiology Vol. 52, No. 18, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.07.048

More information

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease Perspectives of new imaging techniques for patients with known or suspected coronary artery disease Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands Correspondence: Jeroen

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Pre-procedural CT angiography for Transcatheter Aortic Valve Implantation: What a Radiologist Needs to Know?

Pre-procedural CT angiography for Transcatheter Aortic Valve Implantation: What a Radiologist Needs to Know? Pre-procedural CT angiography for Transcatheter Aortic Valve Implantation: What a Radiologist Needs to Know? E O Dwyer, C O Brien, I Murphy, C Shortt, O Buckley Department of Radiology, AMNCH, Dublin,

More information

Raja Muthupillai, PhD. Department of Diagnostic and Interventional Radiology St. Luke s Episcopal Hospital. Research Support: Philips Healthcare

Raja Muthupillai, PhD. Department of Diagnostic and Interventional Radiology St. Luke s Episcopal Hospital. Research Support: Philips Healthcare 3D Cardiac Imaging Raja Muthupillai, PhD Department of Diagnostic and Interventional Radiology St. Luke s Episcopal Hospital Houston, TX Disclosures Research Support: Philips Healthcare This presentation

More information

Imaging Strategies for Endovascular Cardiovascular Procedures and Percutaneous Aortic Valves. Roy K Greenberg, MD

Imaging Strategies for Endovascular Cardiovascular Procedures and Percutaneous Aortic Valves. Roy K Greenberg, MD Imaging Strategies for Endovascular Cardiovascular Procedures and Percutaneous Aortic Valves Roy K Greenberg, MD Disclosure Research support Cook Inc, Boston Scientific, W.L.Gore, Cordis, Vascutek, Terarecon

More information

MR coronary artery imaging with 3D motion adapted gating (MAG) in comparison to a standard prospective navigator technique

MR coronary artery imaging with 3D motion adapted gating (MAG) in comparison to a standard prospective navigator technique Journal of Cardiovascular Magnetic Resonance (2005) 7, 793 797 Copyright D 2005 Taylor & Francis Inc. ISSN: 1097-6647 print / 1532-429X online DOI: 10.1080/10976640500287547 ANGIOGRAPHY MR coronary artery

More information

Utility of CT angiography for pre-operative evaluation of robotic-assisted minimally invasive mitral valve surgery.

Utility of CT angiography for pre-operative evaluation of robotic-assisted minimally invasive mitral valve surgery. Utility of CT angiography for pre-operative evaluation of robotic-assisted minimally invasive mitral valve surgery. Poster No.: C-2214 Congress: ECR 2014 Type: Educational Exhibit Authors: M. Muthuvelu,

More information

MR Advance Techniques. Vascular Imaging. Class II

MR Advance Techniques. Vascular Imaging. Class II MR Advance Techniques Vascular Imaging Class II 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

Coronary artery bypass grafting has been a historically. Multislice CT Evaluation of Coronary Artery Bypass Graft Patients SYMPOSIA

Coronary artery bypass grafting has been a historically. Multislice CT Evaluation of Coronary Artery Bypass Graft Patients SYMPOSIA SYMPOSIA Multislice CT Evaluation of Coronary Artery Bypass Graft Patients Robert Chapman Gilkeson, MD* and Alan H. Markowitz, MDw Abstract: Continuous improvement in multislice computed tomography technology

More information

High-Resolution MR Angiography: Results in Diseased Arteries

High-Resolution MR Angiography: Results in Diseased Arteries IAGS Proceedings NEW IMAGING FOR NEW AND OLD DISEASES High-Resolution MR Angiography: Results in Diseased Arteries Peter Gonschior, M D, Ingo Pragst, M D, Gregor Valassis, M D, Claudia Vo g e l - Wiens,

More information

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA SYMPOSIA Image Acquisition and Interpretation Josef Matthias Kerl, MD,*w Lars K. Hofmann, MD,z Christian Thilo, MD,*y Thomas J. Vogl, MD,w Philip Costello, MD,* and U. Joseph Schoepf, MD*y Abstract: Computed

More information

Alessandro Albonico Philips

Alessandro Albonico Philips Alessandro Albonico Philips Alessandro.albonico@philips.com Noise (Standard Deviation in HU) Virtually noise-free Characteristic of a true knowledge-based IR 80 70 Standard Recon idose4 Level6 1 mm Slice

More information

Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease

Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease Hendrik Treede, MD, Christoph Becker, MD, Hermann Reichenspurner,

More information

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre Pushing the limits of cardiac CT Steven Dymarkowski Radiology / Medical Imaging Research Centre 5 X 2013 Introduction Rapid technological advances and new clinical applications in cardiovascular imaging

More information

On the feasibility of speckle reduction in echocardiography using strain compounding

On the feasibility of speckle reduction in echocardiography using strain compounding Title On the feasibility of speckle reduction in echocardiography using strain compounding Author(s) Guo, Y; Lee, W Citation The 2014 IEEE International Ultrasonics Symposium (IUS 2014), Chicago, IL.,

More information

Acute Aortic Syndromes

Acute Aortic Syndromes Acute Aortic Syndromes Carole J. Dennie, MD Acute Thoracic Aortic Syndromes Background Non-Traumatic Acute Thoracic Aortic Syndromes Carole Dennie MD FRCPC Associate Professor of Radiology and Cardiology

More information

The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters

The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters Iodinated Contrast Opacification Gradients in Normal Coronary Arteries Imaged With Prospectively ECG-Gated Single Heart Beat 320-Detector Row Computed Tomography The Harvard community has made this article

More information