Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model

Size: px
Start display at page:

Download "Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model"

Transcription

1 Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model Ahyoung Jeon*, Sooyoung Ye**, Gyerok Jeon* *Department of Biomedical-engineering Pusan National University, 207, School of Medicine, Pusan National University, Yangsan **Department of Radiological Science Catholic University of Pusan, 57 Oryundae-ro, Geumjeong-gu, Busan SOUTH KOREA Abstract: - In this study, the artery s compliance model and the pulsation waveform model was proposed to estimate blood pressure without applying HPF (high pass filter) on signal measured by the oscillometric method. The method proposed in the study considered two ways of estimating blood pressure. The first method of estimating blood pressure is by comparing and analyzing changes in pulsation waveform s dicrotic notch region during each cardiac period. The second method is by comparing and analyzing morphological changes in the pulsation waveform during each cardiac period, which occur in response to the change in pressure applied on the cuff. To implement these methods, we proposed the compliance model and the pulsation waveform model of the artery based on hemodynamic theory, and then conducted various simulations. The artery model presented in this study only took artery s compliance into account. Then, a pulsation waveform model was suggested, which uses characteristic changes in the pulsation waveform to estimate blood pressure. In addition, characteristic changes were observed in arterial volume by applying artery s pulsation waveform to the compliance model. Furthermore, characteristic changes were also observed in the pulsation waveform by mapping the arterial volume change in accordance with applied cuff s pressure change to the pulsation waveform s change according to applied pressure changes by cuff. Then the systolic and diastolic blood pressures were estimated by analyzing locational changes in pulsation waveform s dicrotic notch region during each cardiac periods, represent by the arterial volume change according to applied cuff s pressure change. Key-Words: - Blood pressure estimation, Artery compliance model, Pulsation waveform model. 1 Introduction As population aging is taking place all over the world, there is a growing interest and demand for development of medical devices that can monitor health of a patient using convenient and noninvasive methods. Currently there are on-going researches on home/portable medical devices as well as medical instruments that use non-invasive methods to minimize patient s pain. In particular, the blood pressure of a patient is used as an important vital sign for determining the patient s health status [1]-[3]. One of the non-invasive blood pressure measurement methods is the oscillometric method [4]. Forster [5] proposed oscillometric model considering the pressure and volume of the blood vessels, and analyzed MAP and various parameters that determine the systolic and diastolic pressure. Then he reported that characteristic ratio between the blood pressures is influenced by arterial pressure s waveform and compliance of the arteries. Mauck [6] presented an artery model and found out that cuff pressure at maximum oscillation point was equal to MAP. Gizdulich[7] modeled relationship between pressure and volume of the arteries in the form of an arc-tangent function. Ursino [8], [9] proposed a model that took various factors that affect the blood pressure when applying oscillometric method into consideration. The proposed model considered the characteristics of biological tissue and cuff. Then simulation was executed for the proposed model and analyzed. However this method has several disadvantages as follows. The raw signal from the oscillometric method is a mixed signal due to cuff pressure and the oscillation waveform, which is the arterial pulsation signal in response to applied cuff pressure. Thus, to extract the oscillation waveform, cuff pressure is subtracted from the original signal. For ISBN:

2 this, usually a high pass filter (HPF) is used, but different oscillation waveforms are displayed depending on frequency bandwidth of HPF. That is, the oscillation signal may be distorted or attenuated. Next, maximum amplitude of the oscillation waveform is set as the mean arterial pressure (MAP) [10]. And finally, the systolic and diastolic pressure is estimated by applying characteristic ratio (CR) on the MAP. But the characteristic ratio does not take into account patient s arterial condition, gender, age, disease, etc., and instead applies a constant CR. Therefore the oscillometric method may measure inaccurate blood pressure. In this study, we tried to come up with a method to estimate the blood pressure without applying HPF on signal measured by the oscillometric method. To do this, we tried to estimate the blood pressure by analyzing and comparing locational changes in pulsation waveform s dicrotic notch region during each cardiac period and morphological changes that occur in pulsation waveform during each cardiac period, in response to change in applied pressure on the cuff. And we suggested a model for estimating the blood pressure based on hemodynamic theory, and then executed simulations. The suggested model and simulation execution process are as follows. The artery model suggested in this study took only the compliance of the arteries into account, hence it was named the compliance model [11], [12],[13], which represents changes in the arterial volume according(in response to) to change in the transmural pressure. To estimate the blood pressure using characteristic changes in the pulsation waveform, the pulsation waveform model was presented. The pulsation waveform in the artery consists of the sum of sine wave and a Fourier s series in combination form up to the 10th harmonics components of the sinusoidal waveform is deployed. Also, we observed characteristic change in blood volume in arterial vessel by inputting pulsation waveform to the compliance model. In other words, change in arterial blood volume in response to change in cuff pressure by externally applied force was observed. Then we observed characteristic changes in the pulsation waveform by mapping the arterial volume change in accordance with applied cuff s pressure change to the pulsation waveform s change in response to change in applied cuff pressure. And finally, we estimated the systolic and diastolic blood pressures by analyzing locational changes in pulsation waveform s dicrotic notch region during each cardiac periods, represented by the arterial volume change in applied cuff pressure. 2 Experimental Methods 2.1 Pulsation Waveform Model When including a subsection you must use, for its heading, small letters, 12pt, left justified, bold, In this study, we created compliance model by considering only one of characteristics of the artery model and then executed simulation using Matlab (Matlab2009b, Mathworks Co., USA). Since it reflects characteristics of the arterial volume change due to changes in internal pressure of the arteries, it was named the compliance model. Transmural pressure is the arterial pressure when the blood from heart perfuse with artery, subtracted by the external pressure applied on the cuff. In this study, we estimated blood pressure using characteristics changes in pulsation waveform of the arteries. To do this, the arterial pulsation waveform from Jeon s [14] study was applied. The pulsation waveform in the artery consists of the sum of sine waves and a Fourier s series in combination form up to 10th harmonics components of the sinusoidal waveform is utilized, and the resulting pulsation waveform looks like Fig. 1. Fig. 1. Pulsation waveform applied to the artery s pulsation waveform model 2.2 Blood Volume Change Considered Compliance of the Artery Footnotes are permitted for notes pertaining to the text only. For biographies and funding acknowledgements, please refer to Section Based on the compliance model of the artery, we presented the volume-change model of the artery during each cardiac period according to change in pressure applied on the cuff. Simulation process, which took compliance of the artery into consideration, on the arterial blood volume change is as follows. The transmural pressure was found and then change in arterial blood volume in response to change in trasmural pressure was observed. Next, we observed change in the pulsation waveform while pressure applied on the cuff was increase from 1 mmhg to 180 mmhg. The ISBN:

3 observation results are shown in Fig. 2. Fig. 2 indicates change in the pulsation waveform in response to the blood volume change due to the cuff pressure change. Fig. 2 is result of the simulation, which was conducted to observe change in pulsation waveform. The x-axis of Fig. 2 represents time series data of the pulsation waveform. The y-axis represents change in pulsation waveform in response to change in applied cuff pressure. As shown in Fig.2, the pulsation waveform changes in the direction from top to bottom. This phenomenon means that the pulsation waveform should be change according to the blood volume of the artery change in response to the applied cuff pressure increase which is come down more in direction from top to bottom in y-axis. Fig. 3 shows change in pulsation waveform in response to change in arterial blood volume according to externally applied pressure on the cuff. That is, change in the pulsation waveform is represented by change in arterial blood volume according to applied cuff s pressure change while cuff pressure decreased from 180 mmhg to 1 mmhg. In Fig. 3, the pulsation waveform s amplitude was smaller when pressure applied on the cuff was higher. On the other hand, the pulsation waveform s amplitude was greater when applied the cuff s pressure decreased. We confirmed this phenomenon by observing decrease in amplitude of the pulsation waveform in response to decrease in pressure applied on the cuff. Also, the pulsation waveform s dicrotic notch region, which is the inflection portion generated by reflected wave, represent changing phenomenon according to applied the cuff s pressure. This dicrotic notch portion included in each pulsation waveform is represented by the change in arterial blood volume according to change in pressure applied on the cuff. Fig. 2. The pulsation waveform s changing aspect according to change of the cuff s pressure. Figure 3. A mapped result waveform that the arterial volume change in accordance with applied cuff s pressure change to the pulsation waveform s change according to applied the cuff s pressure change. To conveniently compare and analyze, and observe distinctive locational changes in the pulsation waveform s dicrotic notch region, normalization process was applied to this study. (The normalization result is shown in Fig. 4) Fig. 4. Change in pulsation waveform in response to change in cuff pressure, after applying normalization. 3 Simulations 3.1 Pulsation Waveform Model Simulation was carried out to estimate the systolic and diastolic blood pressure from the positional change of pulsation waveform s dicrotic notch region in artery during each cardiac period according to applied the cuff s pressure. Simulation process is as follows. First, according to Hemodynamic theory, the blood vessel volumes change due to the cuff pressure degree. The compliance model considered only artery s compliance, and it showed arterial volume change according to the transmural pressure. The pulsation waveform model was suggested to estimate blood pressure using characteristic changes of the pulsation waveform in the arteries. This model was composed of the sum of sine waves and ISBN:

4 a Fourier s series in combination form up to 10th harmonics components of the sinusoidal waveform. Then characteristic of arterial volume change was observed by inputting pulsation waveform into the compliance model. In other words, change in pulsation waveform was observed which results from arterial volume change caused by change in external pressure applied on the cuff. The characteristic changes were also observed in the pulsation waveform by mapping the arterial volume change in accordance with applied cuff s pressure change to the pulsation waveform s change according to applied pressure changes by cuff. The systolic and diastolic blood pressures were estimated by applying positional change of pulsation waveform s dicrotic notch region. By slowly decreasing cuff pressure 5 mmhg at a time, we decreased the externally applied cuff pressure from 180 mmhg to 60 mmhg, and estimated systolic and diastolic blood pressure. When the 12th pulsation waveform s pressure is calculated, the resulting value is 60 mmhg (12th x 5mmHg). And the pressure of the 20th pulsation waveform is 100 mmhg (20th x 5mmHg). By subtracting the 20th waveform s pressure from the maximum pressure, we estimated that the diastolic pressure is 80 mmhg. By using methods mentioned above, we were able to estimate systolic and diastolic blood pressures. 3.2 Simulation Result As a result of simulation that in the pulsation waveform by mapping the arterial volume change in accordance with applied cuff s pressure change to the pulsation waveform s change according to applied pressure changes by cuff, one can observe the following phenomenon. When externally applied pressure is high, arterial volume decreases, resulting in very small pulsation waveform amplitude. When pressure applied is decreased, amplitude of the pulsation waveform continues to increase. However, when pressure applied continues to decrease, amplitude of the pulsation waveform once again decreases exponentially. This phenomenon is in accordance with the compliance model, which shows change in blood vessel volume in response to change in applied pressure. And in Fig. 4, one can observe that depending on externally applied pressure, position of the pulsation waveform s dicrotic notch region undergoes transition. Fig. 5 shows result of simulation, which was conducted to estimate systolic and diastolic blood pressure by observing positional change in dicrotic notch region. One can see from Fig. 5 that when X is 1300, position of periodic pulsation waveform s dicrotic notch region rises. Since periodic pulsation waveform consists of 100 samples, it corresponds to 13th pulsation waveform. Thus, the starting position of change in pulsation waveform s dicrotic notch region is the previous waveform the 12th pulsation waveform. We estimated that this point s pressure corresponds to systolic blood pressure. In addition, when X is 2000, the position of pulsation waveform s dicrotic notch region does not change and remain constant. This point corresponds to the 20th pulsation waveform and this point s pressure corresponds to diastolic blood pressure. Figure 5. Estimated systolic and diastolic blood pressure by applying change in pulsation waveform s dicrotic notch region in response to change in cuff pressure. 4 Conclusion In this study, we made models and conducted simulations to estimate systolic and diastolic pressure by observing positional change in dicrotic notch region of the periodically pulsation waveform. To do this, we presented artery s compliance model and pulsation waveform model. We obtained the following results by applying the models to observe change in the area of pulsation waveform. When pressure applied on the cuff decreased, the area of pulsation waveform was increased in response to change in arterial volume. And when pressure applied on the cuff continued to decrease, the area of pulsation waveform was decreased. In addition, we observed that dicrotic notch region (inflection region formed by the reflective waves) of the pulsation waveform, changed position in response to pressure applied on the cuff. Using the two models presented in the study, we conducted simulation on estimating blood pressure using positional change in pulsation waveform s dicrotic notch region. The ISBN:

5 pulsation waveform model s systolic and diastolic blood pressure was set to 120 mmhg and 80 mmhg, respectively. Then positional change in the pulsation waveform s dicrotic notch region was observed when externally applied cuff pressure was reduced from 180 mmhg to 1 mmhg, 5 mmhg at a time. Simulation result showed that pulsation waveform s dicrotic notch region first began to rise in the 12th pulsation waveform, and this point was estimated to be the systolic blood pressure. The systolic blood pressure was estimated by subtracting 60 mmhg (5 mmhg x 12) from 180 mmhg, which is 120 mmhg. Simulation result also showed that from the 20th pulsation waveform, dicrotic notch region s position became constant (unchanged), and this point was estimated to be the diastolic blood pressure. Diastolic blood pressure was calculated by subtracting 100 mmhg (5mmHg x 20) from 180 mmhg, which is 80 mmhg. From the simulation results we were able to observe that pulsation waveform changes because arterial volume changes in response to change in cuff pressure. Acknowledgment : This research was supported by Basic Science Research Program through the National Research Foundation of Korea(NRF) funded by the Ministry of Education, Science and Technology( ). References: [1] A. C. Guyton, Textbook of medical physiology (W.B. Saunders Company, 2000). [2] D. H. Kang, Physiology (Shin Kwang Publishing Inc., 1998). [3] Ho Kyung Sung, Ki Hwan Kim, Physiology (Ui Hak Moon Hwa Sa Inc., 1996). [4] E. J. Marey, Pression et vitesse du sang, Physiologic Experimentable. Masson, 2(7), Paris, 1876, [5] F. K. Forster and D. Turney. Oscillometric determination of diastolic, mean, and systolic blood pressure. A numerical model, ASME J. Biomech. Eng., 108, 1986, [6] G. W. Mauck, C. R. Smith, L. A. Geddes, and J. D. Bourland, The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure: Part II, ASME J. Biomech. Eng., 102, 1980, [7] P. Gizdulich and K. H. Wesseling, Forearm arterial pressure-volume relationships in man, Clin. Phys. Physiol. Meas., 9, 1988, [8] M. Ursino and C. Ccristalli, Mathematical modelingn of noninvasive blood pressure estimation techniques-part I : Pressure transmissionacross the arm tissue, J. Biomech. Eng., 117, 1995, [9] M. Ursino and C. Cristalli, A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurement, IEEE Trans. Biomed. Eng.,43, 1996, [10] J. A. Posey, L. A., H. Williams and A. G. Moore, The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure: Part I, Cardiovasc. Res. Center Bull., Vol 8, 1969, [11] F. K. Forster and D. Turney, Oscillometric determination of diastolic, mean, and systolic blood pressure, A numerical model. ASME J. Biomech. Eng., vol. 108, 1986, [12] H. H. Hardy and R. E. Collins, On the pressurevolume relationship in circulatory elements, Med. Biol. Eng. Comput., vol. 20, 1982, [13] S. Sun, New approasches to the noninvasive determination of arterial blood pressure and compliance profile, PH. D. dissertation, University of Utah, Salt Lake city, USA, [14] G. R. Jeon, J. H. Jung, I. C. Kim, A. Y. Jeon, S. H. Yoon, J. M. Son, J. H. Kim, S. Y. Ye, J. H. Ro, D. H. Kim, C. H. Kim, A simulation for estimation of the blood pressure using arterial pressure-volume model, International Journal of Biological and Life Science 4:1, 2008, ISBN:

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model Gye-rok Jeon, Jae-hee Jung, In-cheol Kim, Ah-young Jeon, Sang-hwa Yoon, Jung-man Son, Jae-hyung Kim, Soo-young Ye,

More information

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Soo-young Ye, Gi-Ryon Kim, Dong-Keun Jung, Seong-wan Baik, and Gye-rok Jeon Abstract In this paper, algorithm estimating the blood

More information

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1982 Characterization of the Oscillometric Method for Measuring Indirect

More information

Blood Pressure Estimation Using Photoplethysmography (PPG)

Blood Pressure Estimation Using Photoplethysmography (PPG) Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4

More information

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation J. JILEK*, M. STORK** *Carditech, Culver City, California, USA **Department of Applied Electronics

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

Oscillometric measurement of systolic and diastolic blood pressures validated in a physiologic mathematical model

Oscillometric measurement of systolic and diastolic blood pressures validated in a physiologic mathematical model Babbs BioMedical Engineering OnLine 2012, 11:56 RESEARCH Open Access Oscillometric measurement of systolic and diastolic blood pressures validated in a physiologic mathematical model Charles F Babbs Correspondence:

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (  The story so far. Medical Electronics Dr Neil Townsend Michaelmas Term 2001 (wwwrobotsoxacuk/~neil/teaching/lectures/med_elec) Blood Pressure has been measured (in some way) for centuries However, clinical interpretation

More information

Blood Pressure Determination Using Analysis of Biosignals

Blood Pressure Determination Using Analysis of Biosignals Blood Pressure Determination Using Analysis of Biosignals RADIM ČÍŽ 1, MILAN CHMELAŘ 2 1 Department of Telecommunications Brno University of Technology Purkyňova 118, 612 00 Brno CZECH REPUBLIC cizr@feec.vutbr.cz,

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

Special Lecture 10/28/2012

Special Lecture 10/28/2012 Special Lecture 10/28/2012 HYPERTENSION Dr. HN Mayrovitz Special Lecture 10/28/2012 Arterial Blood Pressure (ABP) - Definitions ABP Review Indirect Oscillographic Method Resistance (R), Compliance (C)

More information

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 64 CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 4.1 GENERAL This chapter presents the methodologies that are usually adopted for the measurement of blood pressure, heart rate and pulse

More information

Non- invasive vascular testing. Pros and Cons of ABIs and Alternative Physiologic Assessments

Non- invasive vascular testing. Pros and Cons of ABIs and Alternative Physiologic Assessments Non- invasive vascular testing Pros and Cons of ABIs and Alternative Physiologic Assessments Non- Invasive Physiologic Arterial Studies Segmental Systolic Pressure Measurements ABIs, TBIs, and full segmentals

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE

CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE VASILE MANOLIU Electrical Engineering Faculty, POLITEHNICA University of Bucharest, Splaiul

More information

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062

More information

Keun-Ho Kim and Sang-Suk Lee* Department of Oriental Biomedical Engineering, Sangji University, Wonju , Korea

Keun-Ho Kim and Sang-Suk Lee* Department of Oriental Biomedical Engineering, Sangji University, Wonju , Korea Journal of Magnetics 20(1), 47-51 (2015) ISSN (Print) 1226-1750 ISSN (Online) 2233-6656 http://dx.doi.org/10.4283/jmag.2015.20.1.047 Peripheral Blood Flow Velocity and Peripheral Pulse Wave Velocity Measured

More information

Invasive Measurement of Blood Pressure

Invasive Measurement of Blood Pressure Invasive Measurement of Blood Pressure Arterial Lines & Pulmonary Artery Lines Key Points Invasive Measurement of Blood Pressure: - understand blood flow through the body - understand Pulmonary Artery

More information

MRI MEASUREMENTS OF CRANIOSPINAL AND INTRACRANIAL VOLUME CHANGE IN HEALTHY AND HEAD TRAUMA CASES

MRI MEASUREMENTS OF CRANIOSPINAL AND INTRACRANIAL VOLUME CHANGE IN HEALTHY AND HEAD TRAUMA CASES 1of 4 MRI MEASUREMENTS OF CRANIOSPINAL AND INTRACRANIAL VOLUME CHANGE IN HEALTHY AND HEAD TRAUMA CASES N. Alperin, Y. Kadkhodayan, B. Varadarajalu, C. Fisher, B. Roitberg Department of Radiology and Neurosurgery,

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND D. Špulák 1, R. Čmejla 1, V. Fabián 2 Czech Technical University in Prague, Faculty of Electrical Engineering, 1 Department of Circuit Theory 2

More information

Identification of heart disease-prone personality using oscillometric blood pressure measurements

Identification of heart disease-prone personality using oscillometric blood pressure measurements Technology and Health Care 23 (2015) S211 S222 DOI 10.3233/THC-150956 IOS Press S211 Identification of heart disease-prone personality using oscillometric blood pressure measurements Seungah Jung a and

More information

A Comparative Study of Methods of Measurement of Peripheral Pulse Waveform

A Comparative Study of Methods of Measurement of Peripheral Pulse Waveform 2009. Vol.30. No.3. 98-105 The Journal of Korean Oriental Medicine Original Article A Comparative Study of Methods of Measurement of Peripheral Pulse Waveform Hee-Jung Kang 1, Yong-Heum Lee 2, Kyung-Chul

More information

Glabella. Continuously Sensing Blood Pressure Behavior using an Unobtrusive Wearable Device. Christian Holz Edward Wang Microsoft Research

Glabella. Continuously Sensing Blood Pressure Behavior using an Unobtrusive Wearable Device. Christian Holz Edward Wang Microsoft Research Glabella Continuously Sensing Blood Pressure Behavior using an Unobtrusive Wearable Device Christian Holz Edward Wang Microsoft Research Glabella continuously records the pulse at 3 locations, pulse transit

More information

Blood Pressure Monitor PASESA

Blood Pressure Monitor PASESA Blood Pressure Monitor PASESA new oscillometric device with AVI&API PASESA Prevent ArterioSclerosis and Enjoy Successful Aging PASESA, digital blood pressure monitor for medical use Sit still for two minutes

More information

Efficacy and safety of the reciprocal pulse defibrillator current waveform

Efficacy and safety of the reciprocal pulse defibrillator current waveform Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1984 Efficacy and safety of the reciprocal pulse defibrillator current

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,900 116,000 120M Open access books available International authors and editors Downloads Our

More information

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device 2016 3 rd International Conference on Engineering Technology and Application (ICETA 2016) ISBN: 978-1-60595-383-0 Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable

More information

Temporal Analysis and Remote Monitoring of ECG Signal

Temporal Analysis and Remote Monitoring of ECG Signal Temporal Analysis and Remote Monitoring of ECG Signal Amruta Mhatre Assistant Professor, EXTC Dept. Fr.C.R.I.T. Vashi Amruta.pabarekar@gmail.com Sadhana Pai Associate Professor, EXTC Dept. Fr.C.R.I.T.

More information

Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique

Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique www.jbpe.org Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique Original 1 Department of Biomedical Engineering, Amirkabir University of technology, Tehran, Iran Abbaspour

More information

1 Non-invasive measurement of arterial pressure

1 Non-invasive measurement of arterial pressure Non-invasive measurement of arterial pressure I. Background A. Circulatory systems Human circulation takes place in a closed system that consists of two subsystems, pulmonary circulation and systemic circulation,

More information

Critical Care Nursing Program August to November, 2015 Full-time Lesson A13 Pumping and Perfusion III Basic Hemodynamic Monitoring

Critical Care Nursing Program August to November, 2015 Full-time Lesson A13 Pumping and Perfusion III Basic Hemodynamic Monitoring Critical Care Nursing Program August to November, 2015 Full-time Lesson A13 Pumping and Perfusion III Basic Hemodynamic Monitoring August 2015 RN Professional Development Centre Page 1 Lesson Thirteen

More information

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection Proceedings of the 2012 International Conference on Industrial Engineering and Operations Management Istanbul, Turkey, July 3 6, 2012 Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter

More information

Blood Pressure Laboratory

Blood Pressure Laboratory Introduction The blood that circulates throughout the body maintains a flow and pressure. The nervous system can change the flow and pressure based on the particular needs at a given time. For example,

More information

Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference

Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference ORIGINAL RESEARCH Physiological Reports ISSN 2051-817X Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference

More information

Various Methods To Detect Respiration Rate From ECG Using LabVIEW

Various Methods To Detect Respiration Rate From ECG Using LabVIEW Various Methods To Detect Respiration Rate From ECG Using LabVIEW 1 Poorti M. Vyas, 2 Dr. M. S. Panse 1 Student, M.Tech. Electronics 2.Professor Department of Electrical Engineering, Veermata Jijabai Technological

More information

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS R. Mukkamala, R. G. Mark, R. J. Cohen Haard-MIT Division of Health Sciences and Technology, Cambridge, MA, USA Abstract We

More information

PSD Analysis of Neural Spectrum During Transition from Awake Stage to Sleep Stage

PSD Analysis of Neural Spectrum During Transition from Awake Stage to Sleep Stage PSD Analysis of Neural Spectrum During Transition from Stage to Stage Chintan Joshi #1 ; Dipesh Kamdar #2 #1 Student,; #2 Research Guide, #1,#2 Electronics and Communication Department, Vyavasayi Vidya

More information

Name of Policy: Computerized Pulse Waveform Analysis

Name of Policy: Computerized Pulse Waveform Analysis Name of Policy: Computerized Pulse Waveform Analysis Policy #: 020 Latest Review Date: September 2012 Category: Medical Policy Grade: Active Policy but no longer scheduled for regular literature reviews

More information

5.2. MEASUREMENT OF BLOOD PRESSURE

5.2. MEASUREMENT OF BLOOD PRESSURE 5.2. MEASUREMENT OF BLOOD PRESSURE Purpose of experiment To examine a person s blood pressure. Tasks of experiment To record the pressure changes in the vascular system: using the traditional method measure

More information

A New Approach on Digital Blood Pressure Measurement Method for u-healthcare Systems

A New Approach on Digital Blood Pressure Measurement Method for u-healthcare Systems Vol.7, No.1 (2015), pp.169-178 http://dx.doi.org/10.14257/ijbsbt.2015.7.1.17 A New Approach on Digital Blood Pressure Measurement Method for u-healthcare Systems Boyeon Kim 1, Wenhai Jin 2, Sung Hoon Woo

More information

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator MSc Thesis in Clinical Science and Technology Sara Rose Newell Department of Health Science

More information

Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms

Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms J. Jilek*, M. Stork** *Carditech, Culver City, California, USA **Department of Applied Electronics and Telecommunications University

More information

Heart Abnormality Detection Technique using PPG Signal

Heart Abnormality Detection Technique using PPG Signal Heart Abnormality Detection Technique using PPG Signal L.F. Umadi, S.N.A.M. Azam and K.A. Sidek Department of Electrical and Computer Engineering, Faculty of Engineering, International Islamic University

More information

The Investigation of the Effect of Aging through Photoplethysmogram Signal Analysis of Erectile Dysfunction Subjects

The Investigation of the Effect of Aging through Photoplethysmogram Signal Analysis of Erectile Dysfunction Subjects The Investigation of the Effect of Aging through Photoplethysmogram Signal Analysis of Erectile Dysfunction Subjects Y. K. QAWQZEH 1, M. B. I. REAZ 1, O. MASKON 2, KALAIVANI CHELLAPPAN 1, M. T. ISLAM 3,

More information

Optical sensor for Indian Siddha Diagnosis System

Optical sensor for Indian Siddha Diagnosis System Available online at www.sciencedirect.com Procedia Engineering 38 (2012 ) 1126 1131 International conference on modeling, optimization and computing Optical sensor for Indian Siddha Diagnosis System N.Deepa*,

More information

DETECTION OF HEART ABNORMALITIES USING LABVIEW

DETECTION OF HEART ABNORMALITIES USING LABVIEW IASET: International Journal of Electronics and Communication Engineering (IJECE) ISSN (P): 2278-9901; ISSN (E): 2278-991X Vol. 5, Issue 4, Jun Jul 2016; 15-22 IASET DETECTION OF HEART ABNORMALITIES USING

More information

The reproducibility of central aortic blood pressure measurements in healthy subjects using applanation tonometry and sphygmocardiography

The reproducibility of central aortic blood pressure measurements in healthy subjects using applanation tonometry and sphygmocardiography Journal of Human Hypertension (1999) 13, 625 629 1999 Stockton Press. All rights reserved 0950-9240/99 $15.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE The reproducibility of central aortic

More information

Intra-Arterial Blood Pressure Monitoring

Intra-Arterial Blood Pressure Monitoring Intra-Arterial Blood Pressure Monitoring Verghese Cherian Associate Professor of Anesthesiology Hershey Medical Center Intra-Arterial Blood Pressure (IABP) measurement has several advantages over a non-invasive

More information

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013 ECG Processing &Arrhythmia Detection: An Attempt M.R. Mhetre 1, Advait Vaishampayan 2, Madhav Raskar 3 Instrumentation Engineering Department 1, 2, 3, Vishwakarma Institute of Technology, Pune, India Abstract

More information

Carotid and radial pulse feature analysis with EMFi sensor

Carotid and radial pulse feature analysis with EMFi sensor Carotid and radial pulse feature analysis with EMFi sensor Jarmo Alametsä 1, Ari Palomäki 2, Jari Viik 1 1 Tampere University of Technology, Tampere, Finland, 2 Kanta Häme Central Hospital, Hämeenlinna,

More information

Which method is better to measure arterial stiffness; augmentation index, pulse wave velocity, carotid distensibility? 전북의대내과 김원호

Which method is better to measure arterial stiffness; augmentation index, pulse wave velocity, carotid distensibility? 전북의대내과 김원호 Which method is better to measure arterial stiffness; augmentation index, pulse wave velocity, carotid distensibility? 전북의대내과 김원호 Arterial stiffness Arterial stiffness is inversely related to arterial

More information

ACUTE LEUKEMIA CLASSIFICATION USING CONVOLUTION NEURAL NETWORK IN CLINICAL DECISION SUPPORT SYSTEM

ACUTE LEUKEMIA CLASSIFICATION USING CONVOLUTION NEURAL NETWORK IN CLINICAL DECISION SUPPORT SYSTEM ACUTE LEUKEMIA CLASSIFICATION USING CONVOLUTION NEURAL NETWORK IN CLINICAL DECISION SUPPORT SYSTEM Thanh.TTP 1, Giao N. Pham 1, Jin-Hyeok Park 1, Kwang-Seok Moon 2, Suk-Hwan Lee 3, and Ki-Ryong Kwon 1

More information

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018 WAVES CATCH A WAVE.. W I S C O N S I N P A R A M E D I C S E M I N A R A P R I L 2 0 1 8 K E R I W Y D N E R K R A U S E R N, C C R N, E M T - P Have you considered that if you don't make waves, nobody

More information

Estimation of pressure pulse amplification between aorta and brachial artery using stepwise multiple regression models

Estimation of pressure pulse amplification between aorta and brachial artery using stepwise multiple regression models INSTITUTE OF PHYSICS PUBLISHING Physiol. Meas. 25 (2004) 879 889 PHYSIOLOGICAL MEASUREMENT PII: S0967-3334(04)78628-1 Estimation of pressure pulse amplification between aorta and brachial artery using

More information

Simulation Based R-peak and QRS complex detection in ECG Signal

Simulation Based R-peak and QRS complex detection in ECG Signal Simulation Based R-peak and QRS complex detection in ECG Signal Name: Bishweshwar Pratap Tasa Designation: Student, Organization: College: DBCET, Azara, Guwahati, Email ID: bish94004@gmail.com Name: Pompy

More information

On the relationship between competitive flow and FFT analysis of the flow waves in the left internal mammary artery graft in the process of CABG

On the relationship between competitive flow and FFT analysis of the flow waves in the left internal mammary artery graft in the process of CABG DOI 10.1186/s12938-016-0260-4 BioMedical Engineering OnLine RESEARCH Open Access On the relationship between competitive flow and FFT analysis of the flow waves in the left internal mammary artery graft

More information

NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION. D. Barschdorff*, M. Erig*, E. Trowitzsch**

NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION. D. Barschdorff*, M. Erig*, E. Trowitzsch** NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION D. Barschdorff*, M. Erig*, E. Trowitzsch** *Institute of Electrical Measurement University of Paderborn, D3398 Paderborn, Germany **Sleep Laboratories

More information

Biomedical Signal Processing

Biomedical Signal Processing DSP : Biomedical Signal Processing What is it? Biomedical Signal Processing: Application of signal processing methods, such as filtering, Fourier transform, spectral estimation and wavelet transform, to

More information

NONINVASIVE DETERMINATION OF ARTERIAL PRESSURE-DEPENDENT COMPLIANCE IN YOUNG SUBJECTS USING AN ARTERIAL TONOMETER

NONINVASIVE DETERMINATION OF ARTERIAL PRESSURE-DEPENDENT COMPLIANCE IN YOUNG SUBJECTS USING AN ARTERIAL TONOMETER 111 Vol. 18 No. 3 June 2006 NONINVASIVE DETERMINATION OF ARTERIAL PRESSURE-DEPENDENT COMPLIANCE IN YOUNG SUBJECTS USING AN ARTERIAL TONOMETER JIA-JUNG WANG 1, SHING-HONG LIU 2, TSIAR KAO 3, WEI-CHIH HU

More information

Monitoring Cardiac Stress Using Features Extracted From S1 Heart Sounds

Monitoring Cardiac Stress Using Features Extracted From S1 Heart Sounds e-issn 2455 1392 Volume 2 Issue 4, April 2016 pp. 271-275 Scientific Journal Impact Factor : 3.468 http://www.ijcter.com Monitoring Cardiac Stress Using Features Extracted From S1 Heart Sounds Biju V.

More information

An Investigation of the Pump Operating Characteristics as a Novel Control Index for LVAD Control

An Investigation of the Pump Operating Characteristics as a Novel Control Index for LVAD Control International Seongjin Journal of Choi, Control, J. Robert Automation, Boston, and and Systems, James F. vol. Antaki 3, no. 1, pp. -8, March 5 An Investigation of the Pump Operating Characteristics as

More information

INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS

INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS A.J. Scheen, J.C. Philips, M. Marchand Division of Diabetes, Nutrition

More information

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS V.S. Murthy 1, Sripad Ramamoorthy 1, Narayanan Srinivasan 2, Sriram Rajagopal 2, M. Mukunda Rao 3 1 Department of Physics, Indian Institute

More information

Monitoring Cardiac Health Using a Laser Doppler Approach to Measure Pulse Wave Velocity

Monitoring Cardiac Health Using a Laser Doppler Approach to Measure Pulse Wave Velocity Monitoring Cardiac Health Using a Laser Doppler Approach to Measure Pulse Wave Velocity Madison Boston - Team Leader Michal Adamski - Communicator Haley Knapp - BSAC Katie Barlow - BWIG Lizzie Krasteva

More information

QUANTIFICATION OF EMOTIONAL FEATURES OF PHOTOPLETHYSOMOGRAPHIC WAVEFORMS USING BOX-COUNTING METHOD OF FRACTAL DIMENSION

QUANTIFICATION OF EMOTIONAL FEATURES OF PHOTOPLETHYSOMOGRAPHIC WAVEFORMS USING BOX-COUNTING METHOD OF FRACTAL DIMENSION QUANTIFICATION OF EMOTIONAL FEATURES OF PHOTOPLETHYSOMOGRAPHIC WAVEFORMS USING BOX-COUNTING METHOD OF FRACTAL DIMENSION Andrews Samraj*, Nasir G. Noma*, Shohel Sayeed* and Nikos E. Mastorakis** *Faculty

More information

Special Lecture 11/08/2013. Hypertension Dr. HN Mayrovitz

Special Lecture 11/08/2013. Hypertension Dr. HN Mayrovitz Special Lecture 11/08/2013 Hypertension Dr. HN Mayrovitz Arterial Blood Pressure (ABP) Major Factors Summarized Sympathetic Hormones Arteriole MAP ~ Q x TPR + f (V / C) SV x HR Renal SBP Hypertension =

More information

Neural Network based Heart Arrhythmia Detection and Classification from ECG Signal

Neural Network based Heart Arrhythmia Detection and Classification from ECG Signal Neural Network based Heart Arrhythmia Detection and Classification from ECG Signal 1 M. S. Aware, 2 V. V. Shete *Dept. of Electronics and Telecommunication, *MIT College Of Engineering, Pune Email: 1 mrunal_swapnil@yahoo.com,

More information

VENTILATOR GRAPHICS ver.2.0. Charles S. Williams RRT, AE-C

VENTILATOR GRAPHICS ver.2.0. Charles S. Williams RRT, AE-C VENTILATOR GRAPHICS ver.2.0 Charles S. Williams RRT, AE-C Purpose Graphics are waveforms that reflect the patientventilator system and their interaction. Purposes of monitoring graphics: Allow users to

More information

Texture Analysis of Supraspinatus Ultrasound Image for Computer Aided Diagnostic System

Texture Analysis of Supraspinatus Ultrasound Image for Computer Aided Diagnostic System Original Article Healthc Inform Res. 2016 October;22(4):299-304. pissn 2093-3681 eissn 2093-369X Texture Analysis of Supraspinatus Ultrasound Image for Computer Aided Diagnostic System Byung Eun Park,

More information

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

The Cardiac Cycle Clive M. Baumgarten, Ph.D. The Cardiac Cycle Clive M. Baumgarten, Ph.D. OBJECTIVES: 1. Describe periods comprising cardiac cycle and events within each period 2. Describe the temporal relationships between pressure, blood flow,

More information

Clinical application of Arterial stiffness. pulse wave analysis pulse wave velocity

Clinical application of Arterial stiffness. pulse wave analysis pulse wave velocity Clinical application of Arterial stiffness pulse wave analysis pulse wave velocity Arterial system 1. Large arteries: elastic arteries Aorta, carotid, iliac, Buffering reserve: store blood during systole

More information

Optimal Duration of Clopidogrel Therapy with DES to Reduce Late Coronary Arterial Thrombotic Event. The DES LATE Trial

Optimal Duration of Clopidogrel Therapy with DES to Reduce Late Coronary Arterial Thrombotic Event. The DES LATE Trial Optimal Duration of Clopidogrel Therapy with DES to Reduce Late Coronary Arterial Thrombotic Event The DES LATE Trial Cheol Whan Lee, MD, Seung-Jung Park, MD, PhD, On Behalf of the DES LATE Investigators

More information

A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks

A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks Younessi Heravi Mohammad Amin Department of Basic Sciences, School of Medicine, North Khorasan

More information

Heart Rate Calculation by Detection of R Peak

Heart Rate Calculation by Detection of R Peak Heart Rate Calculation by Detection of R Peak Aditi Sengupta Department of Electronics & Communication Engineering, Siliguri Institute of Technology Abstract- Electrocardiogram (ECG) is one of the most

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

CVS Hemodynamics. Faisal I. Mohammed, MD,PhD.

CVS Hemodynamics. Faisal I. Mohammed, MD,PhD. CVS Hemodynamics Faisal I. Mohammed, MD,PhD. Objectives point out the physical characteristics of the circulation: distribution of blood volume total cross sectional area velocity blood pressure List the

More information

Biomedical. Measurement and Design ELEC4623. Lectures 15 and 16 Statistical Algorithms for Automated Signal Detection and Analysis

Biomedical. Measurement and Design ELEC4623. Lectures 15 and 16 Statistical Algorithms for Automated Signal Detection and Analysis Biomedical Instrumentation, Measurement and Design ELEC4623 Lectures 15 and 16 Statistical Algorithms for Automated Signal Detection and Analysis Fiducial points Fiducial point A point (or line) on a scale

More information

Raw BP and Eight Hemodynamic Measurements Displayed in Real Time

Raw BP and Eight Hemodynamic Measurements Displayed in Real Time APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com Updated 01.11.2016 Application Note 115: Hemodynamic Measurements Raw BP and Eight

More information

Blood Pressure (BP) measurement becomes a vital parameter now days due to continuous increase in chronic heart

Blood Pressure (BP) measurement becomes a vital parameter now days due to continuous increase in chronic heart ISSN: 0975-766X CODEN: IJPTFI Available Online through Research Article www.ijptonline.com FAST AND ACCURATE METHOD FOR CONTINUOUS BP MEASUREMENT Harkirat Kaur*, Dr. Sunil Agrawal, Mr. Sarvjit Singh Dept.

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

BP-600 Noninvasive Blood Pressure Sensor

BP-600 Noninvasive Blood Pressure Sensor Technical Note BP-600 Overview A person's cardiac output, peripheral vascular resistance, blood pressure, and other cardiovascular parameters change in response to the activities and events taking place

More information

Separation of fetal electrocardiography (ECG) from composite ECG using adaptive linear neural network for fetal monitoring

Separation of fetal electrocardiography (ECG) from composite ECG using adaptive linear neural network for fetal monitoring International Journal of the Physical Sciences Vol. 6(24), pp. 5871-5876, 16 October, 2011 Available online at http://www.academicjournals.org/ijps ISSN 1992-1950 2011 Academic Journals Full Length Research

More information

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License Pan F, Zheng DC, He PY, Murray A. Does the Position or Contact Pressure of the Stethoscope Make Any Difference to Clinical Blood Pressure Measurements: An Observational Study. Medicine 214, 93(29). Copyright:

More information

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG Introduction An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG data can give important information about the health of the heart and can help physicians to diagnose

More information

Novel single trial movement classification based on temporal dynamics of EEG

Novel single trial movement classification based on temporal dynamics of EEG Novel single trial movement classification based on temporal dynamics of EEG Conference or Workshop Item Accepted Version Wairagkar, M., Daly, I., Hayashi, Y. and Nasuto, S. (2014) Novel single trial movement

More information

Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement

Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement DOI 10.1007/s11517-014-1150-1 Original Article Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement Dingchang Zheng Luigi Yuri Di Marco Alan

More information

Wavelet Decomposition for Detection and Classification of Critical ECG Arrhythmias

Wavelet Decomposition for Detection and Classification of Critical ECG Arrhythmias Proceedings of the 8th WSEAS Int. Conference on Mathematics and Computers in Biology and Chemistry, Vancouver, Canada, June 19-21, 2007 80 Wavelet Decomposition for Detection and Classification of Critical

More information

A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD

A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD Zahari Taha, Lum Shirley * and Mohd Azraai Mohd Razman Innovative Manufacturing, Mechatronics and Sports (imams)

More information

Non-invasive diagnosis of coronary artery disease by exercise magnetocardiography

Non-invasive diagnosis of coronary artery disease by exercise magnetocardiography Non-invasive diagnosis of coronary artery disease by exercise magnetocardiography Jai-Wun Park, MD Klinik für Kardiologie Campus Benjamin Franklin Charité - Universitätsmedizin Berlin JCR Busan, Korea

More information

Cardiovascular Diseases Detecting via Pulse Analysis

Cardiovascular Diseases Detecting via Pulse Analysis Engineering, 2013, 5, 176-180 http://dx.doi.org/10.4236/eng.2013.510b038 Published Online October 2013 (http://www.scirp.org/journal/eng) Cardiovascular Diseases Detecting via Pulse Analysis Jingjing Xia,

More information

Keywords: Blood pressure, Blood flow, Subjective experiment, Human simulation model

Keywords: Blood pressure, Blood flow, Subjective experiment, Human simulation model F: Physical responses & physiology F.2. Health assessment (incl. Thermal comfort) MEASUREMENTS OF BLOOD FLOW AND BLOOD PRESSURE UNDER DIFFERENT INDOOR TEMPERATURE AND BODY POSTURAL CONDITIONS, AND DEVELOPMENT

More information

Design and implementation of IIR Notch filter for removal of power line interference from noisy ECG signal

Design and implementation of IIR Notch filter for removal of power line interference from noisy ECG signal 70 Design and implementation of IIR Notch filter for removal of power line interference from noisy ECG signal Ravi Kumar Chourasia 1, Ravindra Pratap Narwaria 2 Department of Electronics Engineering Madhav

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

Assessing Preload Responsiveness Using Arterial Pressure Based Technologies. Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC

Assessing Preload Responsiveness Using Arterial Pressure Based Technologies. Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC Assessing Preload Responsiveness Using Arterial Pressure Based Technologies Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC Content Description : Fluid administration is a first

More information

The Conduit Artery Functional Endpoint (CAFE) study in ASCOT

The Conduit Artery Functional Endpoint (CAFE) study in ASCOT (2001) 15, Suppl 1, S69 S73 2001 Nature Publishing Group All rights reserved 0950-9240/01 $15.00 www.nature.com/jhh A Sub-study of the ASCOT Trial The Conduit Artery Functional Endpoint (CAFE) study in

More information

Device Judgement Recommended

Device Judgement Recommended dabl Educational Trust Validation Checklist Device Judgement Recommended Manufacturer IEM Device Stabil O Graph Category Upper Arm (Self) Mode Oscillometric Protocol Reference BHS + ESH Westhoff TH, Schmidt

More information