Clinical Data Summary: Avoid FFS Study

Size: px
Start display at page:

Download "Clinical Data Summary: Avoid FFS Study"

Transcription

1 Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead

2 Clinical Data Summary: AVOID FFS Study Introduction: Far-field sensing (FFS) is the most common cause of inappropriate mode switching episodes in dual-chamber pacing, yet FFS represents a stubborn problem for pacemaker clinicians to resolve effectively. FFS occurs when the atrial lead of a dual-chamber pacemaker detects signals from the ventricle and inappropriately senses them as intrinsic atrial events. The use of a new unique bipolar lead design that places the two distal electrodes in very close proximity to each other may mitigate this situation (see Figure 1). Figure 1 The bipolar atrial pacing lead is an active-fixation lead with a titanium-coated platinum-iridium helix (2 mm extension), tip size 7 mm 2, and tip-to-ring electrode spacing of 1.1 mm. The control lead (Tendril bipolar atrial lead) is comparable in specifications but uses a 10 mm tip-to-ring electrode spacing. Patients at the study centers indicated for an initial dualchamber pacemaker implant were randomized into one of two arms of the study ( lead group or conventional group) in a 1:1 pattern. PVAB was not optimized in the study group but programmed to the shortest possible value of 60 ms. In the control group, PVAB was optimized to a value at least 25 ms longer than the measured interval between QRS and sensed FFS at an atrial sensitivity of 0.1 mv. Atrial sensing threshold was programmed to 0.3 mv in both groups. Owing to the nature of this study, implanting physicians and following clinicians were not blinded. Large Spacing 10 mm Small Spacing 1.1 mm Larger antenna increases risk of far-field sensing. Short spacing reduces risk of far-field sensing. The AVOID FFS study is a randomized, multi-center prospective study with a sample size of over 180 patients, designed to investigate whether a new bipolar atrial lead with a very short tip-to-ring spacing shows equally low incidence of FFS with short PVAB (post-ventricular atrial blanking period) compared to optimized PVAB with standard bipolar atrial leads. A period of short-term follow-up yielded preliminary results regarding the sensing and pacing properties of two atrial pacing leads. Lead Overview: The AVOID FFS study evaluated the model 1699 atrial pacing lead with novel 1.1 mm tip-to-ring spacing of the lead s two distal electrodes. This unique electrode configuration is designed to offer more precise measurement and differentiation of atrial signals, considered to be clinically important features in terms of both avoiding far-field sensing and appropriately sensing low-amplitude atrial fibrillatory waves. This innovative lead design was compared in AVOID FFS to a conventional bipolar atrial pacing lead which relies on the established technology of 10 mm electrode spacing (the St. Jude Medical Tendril model 1388, 1688 or 1788 lead). Source: Lotze et al. Sensing and pacing properties of a new atrial lead with a 1.1 mm tip-to-ring spacing (AVOID FFS study). Study Objectives: The objective of the first interim analysis was to compare electrical performance of the two leads by measuring pacing threshold, sensing threshold, lead impedance, and contrasting atrial intracardiac electrograms (EGMs) post-discharge and at one month post-implant. The hypothesis of the study is that the use of a new lead with 1.1 mm tip-to-ring spacing and a short PVAB (programmed to 60 ms) is equally effective in suppressing FFS as a standard lead with additional FFS-test and PVAB optimization. Patient Demographics: A total of 204 patients participated in the study, all of whom had a conventional dual-chamber pacemaker indication. About 60% of patients were male. Mean ages of the patients at enrollment were 74±10 years for the control group (n=101) and 73±9 years for the study group (n=103), which is thought to be a representative age for the general pacemaker patient population (see Table 1). BASELINE CHARACTERISTICS of the PATIENT CONTROL GROUP STUDY GROUP P VALUE (n=101) (n=103) Mean age [±SD] (years) 74±10 73± Men (%) 60 (60) 61 (60) 0.97 Pacemaker indications 0.62 Sick sinus syndrome (%) 38 (37.6) 43 (41.7) AV block (%) 46 (12.5) 38 (36.9) Binodal disease (%) 15 (14.9) 20 (19.4) Other (%) 2 (2.0) 2 (1.9) Atrial lead placement 0.53 Lateral (%) 44 (43.6) 41 (39.8) Right atrial appendage (RAA) (%) 38 (37.6) 49 (47.6) Septal (%) 2 (2.0) 3 (2,9) Other (%) 17 (17.8) 10 (9.7) Fluoroscopy time [±SD] (s) 355± ± Procedure time [±SD] (min) 62±19 61± Clinical Data Summary: AVOID FFS Study Follow-up of 1 month (%) 95 (94.1) 99 (96.1) 0.50 Follow-up of 3 months (%) 94 (93.1) 94 (91.3) 0.23

3 Study Results: n Capture Thresholds: Atrial pacing thresholds were evaluated in both groups at discharge and again one month post-implant. Threshold measurements were performed at a pulse width of 0.4 ms and pulse amplitude thresholds (in voltage) were compared. leads had a lower but not significant (P=0.065) atrial pacing threshold at implant, but this lower threshold became significant (P=0.006) at one month post-implant (see Figure 2). This improvement is consistent with the electrical threshold performance observed with the initial FFS IDE clinical study performed in Figure n Bipolar Atrial Sensing Thresholds: Atrial sensing thresholds did show a slight difference, with atrial bipolar leads having lower sensing thresholds at implant (P=0.006) and at one month post-implant (P<0.001) (see Figure 3); however, the ability of this lead to be programmed at more sensitive settings enables it to reliably sense the P waves. Figure , , Minimal P Wave (mv) at discharge At discharge OptiSense P (n=92) (n=98) value P wave [mv] 2.57± ± Pacing Threshold 0.4 ms) at discharge At discharge OptiSense P P wave [mv] (n=90) (n=94) value Pacing threshold [V] 0.65± ± Minimal P Wave (mv) at 1 month post-implant At 1 month OptiSense P post-implant (n=88) (n=92) value P wave [mv] 2.69± ±1.12 <0.001 Pacing Threshold 0.4 ms) at 1 month post-implant At 1 month OptiSense P post-implant (n=81) (n=86) value Pacing threshold [V] 0.72± ±

4 n Lead Pacing Impedance: Lead impedance, an indicator of how current travels through the lead, is a combination of the resistance within the lead itself and the impedance at the lead-tissue interface. leads showed higher impedance values at discharge (P<0.001) and one month post-implant (P=0.007) compared to conventional atrial bipolar leads (see Figure 4). The lead s higher impedance and equivalent or lower pacing thresholds compared with standard leads means less current drain which can positively impact device longevity. Figure 4 1 n Far-field Interference: Particularly meaningful for clinicians, intracardiac electrograms were obtained from both types of atrial bipolar pacing leads. Far-field sensing occurs because signals originating from the ventricle seem disproportionately large to the device s atrial sensing circuitry, creating a sort of electrical version of an optical illusion (see Figure 5). DeVoogt et al. reported that the ratio of intrinsic atrial signal amplitude to intrinsic ventricular signal amplitude for a conventional bipolar atrial pacing lead was approximately 5 to 1. 1 On the other hand, the P-to-R ratio for the pacing lead is approximately 40 to 1. Figure Impedance at discharge At discharge OptiSense P (n=93) (n=99) value Impedance [ohms] 341±73 405±109 < Study Conclusions: The novel model 1699 atrial bipolar pacing lead with its unique 1.1 mm tip-to-ring electrode spacing was shown in the AVOID FFS randomized, controlled, prospective clinical trial to offer good and stable electrical performance characteristics compared to a conventional atrial bipolar lead with 10 mm tip-to-ring electrode spacing. In terms of atrial pacing thresholds, lead impedance, and sensing characteristics, this novel lead offered superior and often significantly superior performance characteristics. The authors conclude that initial results show great promise. Further recent study of this lead by de Groot et al. confirmed, FFRW can be suppressed without compromising P-wave sensing by using a pacing electrode with a short tip-ring distance. 2 1 De Voogt WG, Van Hemel NM, Willems AR, Visser J, Chitre Y, Bornzin G, Helland J. Far-field R-wave reduction with a novel lead design: Experimental and human results. PACE 2005; 28: De Groot et al. Clinical results of far-field R-wave reduction with a short tip-ring electrode. PACE 2008; 31: Impedance at 1 month post-implant At 1 month OptiSense P post-implant (n=88) (n+93) value Impedance [ohms] 333±67 361±

5

6 St. Jude Medical is focused on reducing risk by continuously finding ways to put more control into the hands of those who save and enhance lives. AtriAl FibrillAtion CArdiAC rhythm MAnAgeMent CArdiAC Surgery CArdiology neuromodulation Cardiac Rhythm Management Division Valley View Court Sylmar, California usa Fax St. Jude Medical Sweden AB Veddestavägen Järfälla Sweden Fax St. Jude Medical Coordination Center BVBA the Corporate Village da Vincilaan 11 box F Zaventem belgium Fax Global Headquarters one lillehei Plaza St. Paul, Minnesota usa Fax sjm.com CAUTION: FEDERAL LAW (USA) RESTRICTS THIS DEVICE TO SALE, DISTRIBUTION AND USE BY OR ON THE ORDER OF A PHYSICIAN. Consult the User s Manual for information on indications, contraindications, warnings and precautions. Unless otherwise noted, or indicates that the name is a trademark of, or licensed to, St. Jude Medical, or one of its subsidiaries. ST. JUDE MEDICAL, the nine-squares symbol and MORE CONTROL. LESS RISK. are trademarks and service marks of St. Jude Medical, Inc. and its related companies St. Jude Medical. All Rights Reserved. Item No. P0135

How to Record & Send Information

How to Record & Send Information Patient Handbook How to Record & Send Information with your SJM Confirm Implantable Cardiac Monitor The Patient Activator is part of the SJM Confirm Implantable Cardiac Monitor (ICM). It is, in essence,

More information

Make the Connection A clinical compendium on the relationship between AF, stroke, and early intervention

Make the Connection A clinical compendium on the relationship between AF, stroke, and early intervention Make the Connection A clinical compendium on the relationship between AF, stroke, and early intervention Data that date back to the 1970 s have illustrated the strong relationship between atrial fibrillation

More information

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only. SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000 1 APPROVED MRI SCAN CONDITIONS 1.5 TESLA FULL BODY 2 SETUP SJM MRI ACTIVATOR HANDHELD DEVICE The SJM MRI Activator handheld device must be enabled

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. WARNING: This

More information

ViewFlex Xtra ICE Catheter. Positioning Reference Manual

ViewFlex Xtra ICE Catheter. Positioning Reference Manual ViewFlex Xtra ICE Catheter Positioning Reference Manual ViewFlex Xtra ICE Catheter Index The ViewFlex Xtra ICE Catheter, which is compatible with the ViewMate Z and ViewMate II ultrasound consoles, provides

More information

DRG THERAPY FOR CHRONIC PAIN ACCURATE CLINICAL STUDY FACT SHEET FOR PATIENTS

DRG THERAPY FOR CHRONIC PAIN ACCURATE CLINICAL STUDY FACT SHEET FOR PATIENTS ACCURATE CLINICAL STUDY DRG THERAPY FOR CHRONIC PAIN FACT SHEET FOR PATIENTS It was the kind of pain where you couldn t push through it now I am back to living life again and not having any pain. Jenifer,

More information

GDT1000 SENSING ACUTE STUDY

GDT1000 SENSING ACUTE STUDY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY GDT1000 SENSING ACUTE STUDY Boston

More information

PAIN INTER- DRG THERAPY FOR CHRONIC PAIN RUPTED LIFE TRANSFORMED

PAIN INTER- DRG THERAPY FOR CHRONIC PAIN RUPTED LIFE TRANSFORMED PAIN INTER- RUPTED DRG THERAPY FOR CHRONIC PAIN LIFE TRANSFORMED A DIFFERENT TYPE OF PAIN CALLS FOR A DIFFERENT APPROACH. Chronic pain* is a common condition, affecting almost one in five adults globally.

More information

2015 ST. JUDE MEDICAL THERAPY CODING GUIDE Cardiac Pacemakers

2015 ST. JUDE MEDICAL THERAPY CODING GUIDE Cardiac Pacemakers 2015 ST. JUDE MEDICAL THERAPY CODING GUIDE Cardiac Pacemakers This guide provides physician and hospital coding and reimbursement information for cardiac pacemaker procedures. In addition, St. Jude Medical

More information

St. Jude Medical 8-Channel Adapter. Clinician's Manual

St. Jude Medical 8-Channel Adapter. Clinician's Manual St. Jude Medical 8-Channel Adapter Clinician's Manual CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. For U.S. California Only: Proposition 65, a State of California

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Journal of Arrhythmia

Journal of Arrhythmia Journal of Arrhythmia 29 (2013) 204 210 Contents lists available at ScienceDirect Journal of Arrhythmia journal homepage: www.elsevier.com/locate/joa Original Article Factors involved in correct analysis

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

BEFORE AND AFTER YOUR MRI SCAN. For patients with St. Jude Medical neurostimulation systems for chronic pain

BEFORE AND AFTER YOUR MRI SCAN. For patients with St. Jude Medical neurostimulation systems for chronic pain BEFORE AND AFTER YOUR MRI SCAN For patients with St. Jude Medical neurostimulation systems for chronic pain Preparing for your MRI Depending on what type of neurostimulation system you have, you may be

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

2016 ST. JUDE MEDICAL THERAPY CODING GUIDE Cardiac Pacemakers

2016 ST. JUDE MEDICAL THERAPY CODING GUIDE Cardiac Pacemakers 2016 ST. JUDE MEDICAL THERAPY CODING GUIDE Cardiac Pacemakers This guide provides physician and hospital coding and reimbursement information for cardiac pacemaker procedures. In addition, St. Jude Medical

More information

AMPLATZER Amulet Left Atrial Appendage Occluder

AMPLATZER Amulet Left Atrial Appendage Occluder AMPLATZER Amulet Left Atrial Appendage Occluder EXPECT THE BEST OF ALL WORLDS FROM YOUR LEFT ATRIAL APPENDAGE OCCLUDER: Ease of use, the broadest range of sizes and a design allowing for a complete ostial

More information

Portico Transcatheter Aortic Valve Implantation System. portico PRECISION

Portico Transcatheter Aortic Valve Implantation System. portico PRECISION Portico Transcatheter Aortic Valve Implantation System portico PRECISION on target with more control for transcatheter valve implantation A comprehensive approach to aortic valve implantation, designed

More information

PATIENT CONTROLLER ST. JUDE MEDICAL UNDERSTANDING IMPORTANT FEATURES ON YOUR

PATIENT CONTROLLER ST. JUDE MEDICAL UNDERSTANDING IMPORTANT FEATURES ON YOUR UNDERSTANDING IMPORTANT FEATURES ON YOUR ST. JUDE MEDICAL PATIENT CONTROLLER THINGS TO KNOW BEFORE AND AFTER AN MRI OR SURGICAL PROCEDURE WHEN YOU HAVE A ST. JUDE MEDICAL PROCLAIM NEUROSTIMULATION SYSTEM

More information

DRG THERAPY FOR CHRONIC PAIN PAIN INTER- RUPTED LIFE TRANSFORMED

DRG THERAPY FOR CHRONIC PAIN PAIN INTER- RUPTED LIFE TRANSFORMED PAIN INTER- RUPTED DRG THERAPY FOR CHRONIC PAIN LIFE TRANSFORMED A DIFFERENT TYPE OF PAIN CALLS FOR A DIFFERENT APPROACH. Chronic pain* is a common condition, affecting almost one in five adults globally.

More information

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV.

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV. Contents I. Introduction II. Website and Relevant Documents III. Background Information IV. Pacemaker System V. i. Components ii. Operating Modes iii. Parameters Example: Pacemaker in DDD Mode Pacemaker

More information

Active Fixation Models: 7740, 7741 and 7742 Passive Fixation Models: 7731, 7732, 7735 and 7736

Active Fixation Models: 7740, 7741 and 7742 Passive Fixation Models: 7731, 7732, 7735 and 7736 The INGEVITY MRI pacing leads are 6F (2.0 mm) steroid-eluting, endocardial pace / sense leads designed for permanent implantation for either atrial or ventricular applications. INGEVITY MRI is the only

More information

New generations pacemakers and ICDs: an update

New generations pacemakers and ICDs: an update Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology

More information

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants EKG Pop Quiz Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants 1 Disclosures No financial relationships to disclose. EKG #1 75 y/o woman with a dual-chamber pacemaker

More information

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or

More information

DRG STIMULATION INTERVENTIONAL PAIN CENTER

DRG STIMULATION INTERVENTIONAL PAIN CENTER DRG STIMULATION Shawn M. Sills, MD at Touchstone Interventional Pain Center has been selected by St Jude Medical to be first in Southern Oregon delivering this new FDA approved therapy to patients. You

More information

Lead and Extension Insertion Tool. Model Clinician's Manual

Lead and Extension Insertion Tool. Model Clinician's Manual Lead and Extension Insertion Tool Model 1803 Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or one of its subsidiaries. ST. JUDE

More information

Size (mm) (H W D) Mass (g) Volume (cc)

Size (mm) (H W D) Mass (g) Volume (cc) The FORMIO pacing system from Boston Scientific provides an ImageReady * MR conditional pacing system. FORMIO offers RightRate the newest generation of MV sensor technology, and the only MV sensor clinically

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

Automatic assessment of atrial pacing threshold in current medical practice

Automatic assessment of atrial pacing threshold in current medical practice Europace (212) 14, 1615 1619 doi:1.193/europace/eus76 CLINICAL RESEARCH Leads and lead extraction Automatic assessment of atrial pacing threshold in current medical practice Jean Luc Rey 1, Serge Quenum

More information

Single-Lead DDD System: A Comparative Evaluation of Unipolar, Bipolar and Overlapping Biphasic Stimulation

Single-Lead DDD System: A Comparative Evaluation of Unipolar, Bipolar and Overlapping Biphasic Stimulation 82 September 1997 Single-Lead DDD System: A Comparative Evaluation of Unipolar, Bipolar and Overlapping Biphasic Stimulation H.-F. TSE, C. P. LAU, Z. LEUNG Division of Cardiology, Department of Medicine,

More information

Advisa MRI and Ensura MRI SureScan pacing systems

Advisa MRI and Ensura MRI SureScan pacing systems Advisa MRI and Ensura MRI SureScan pacing systems MR Conditional dual chamber and single chamber SureScan pacing systems with SureScan leads MRI Technical Manual The following list includes trademarks

More information

CONSULTA CRT-P Model C4TR01

CONSULTA CRT-P Model C4TR01 CONSULTA CRT-P Model C4TR01 Physical characteristics Volume a 15 cm 3 Mass H x W x D Surface area of titanium device can Surface area of each titanium nitride LECG electrode Radiopaque ID Materials in

More information

Long Term Monitoring of the Intrinsic Ventricular Monophasic Action Potential with an Implantable DDD Pacemaker

Long Term Monitoring of the Intrinsic Ventricular Monophasic Action Potential with an Implantable DDD Pacemaker May 1998 79 Long Term Monitoring of the Intrinsic Ventricular Monophasic Action Potential with an Implantable DDD Pacemaker T. LAWO, J. BARMEYER Abteilung für Kardiologie, Universitätsklinik Bergmannsheil,

More information

A Closer Look Product Education at a glance

A Closer Look Product Education at a glance A Closer Look Product Education at a glance an ICD or CRT-D When Atrial Information is Not Used SUMMARY Boston Scientific dual-chamber ICDs and multi-chamber CRT-Ds will respond to atrial sensing regardless

More information

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

The heart's natural pacemaker is called the sinoatrial (SA) node or sinus node. PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in

More information

AMPLATZER Septal Occluder Structural Heart Therapy. Over 15 years of. Demonstrated. Clinical Experience. We ll show you our data. Ask to see theirs.

AMPLATZER Septal Occluder Structural Heart Therapy. Over 15 years of. Demonstrated. Clinical Experience. We ll show you our data. Ask to see theirs. AMPLATZER Septal Occluder Structural Heart Therapy Over 15 years of Demonstrated Clinical Experience We ll show you our data. Ask to see theirs. Leading the Standard of Care 1,2 The AMPLATZER Septal Occluder

More information

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Francesco Zanon, MD, FESC, FHRS Arrhythmia and Electrophysiology Unit, Cardiology Dept. Santa Maria della Misericordia General

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter

Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter Instructions for Use Reprocessed LASSO Circular Mapping Diagnostic Electrophysiology (EP) Catheter Caution: Federal (USA) law restricts this device to sale by or on the order of a physician. DEVICE DESCRIPTION

More information

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or

More information

A predictable pattern in renal denervation. EnligHTN. Multi-Electrode Renal Denervation System. For the Treatment of Hypertension

A predictable pattern in renal denervation. EnligHTN. Multi-Electrode Renal Denervation System. For the Treatment of Hypertension A predictable pattern in renal denervation Introducing EnligHTN Multi-Electrode Renal Denervation System For the Treatment of Hypertension An Enlightened Approach to Renal Denervation The EnligHTN Multi-Electrode

More information

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.

More information

Original Article Clinical observation of atrial threshold monitoring algorithm: a single center experience

Original Article Clinical observation of atrial threshold monitoring algorithm: a single center experience Int J Clin Exp Med 2015;8(4):6079-6084 www.ijcem.com /ISSN:1940-5901/IJCEM0005972 Original Article Clinical observation of atrial threshold monitoring algorithm: a single center experience Jianqing She

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

PACING SYSTEMS.

PACING SYSTEMS. References 1 Ensura SR MRI SureScan EN1SR01 Clinician Manual. 2 Medtronic Adapta ADSR01/03/06 Implant Manual, 2005. 3 Advisa MRI SureScan pacing system included on the ARTG 22/8/2009, with full body MRI

More information

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features Definition of SVT Discrimination Identify which patient populations might benefit from these features Understand the 4 types of SVT discriminators used by St Jude Medical Be aware of programmable parameters

More information

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Review of Pacemakers and ICD Therapy: Overview and Patient Management Review of Pacemakers and ICD Therapy: Overview and Patient Management Pacing Systems Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services OSU Division

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Riata Lead Summit Jan 20, SJM Perspective Dr. Mark Carlson

Riata Lead Summit Jan 20, SJM Perspective Dr. Mark Carlson Riata Lead Summit Jan 20, 2012 SJM Perspective Dr. Mark Carlson 1 1 Presentation Outline Background, Characterization of Externalized Conductors Modeling and Testing Diagnostics and Therapeutics Recommendations

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Is This Thing Working?

Is This Thing Working? Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu Disclosures I currently receive research support from Medtronic, Inc. I have been compensated

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY

CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

introducing Quantien Integrated FFR Platform Free Your Cath Lab with Wireless FFR

introducing Quantien Integrated FFR Platform Free Your Cath Lab with Wireless FFR introducing Quantien Integrated FFR Platform Free Your Cath Lab with Wireless FFR QUANTIEN Free Your Cath Lab with Wireless F Free from Cables The QUANTIEN integrated FFR solution gives you a wireless

More information

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? CASE REPORTS Arrhythmia 2015;16(3):173-177 doi: http://dx.doi.org/10.18501/arrhythmia.2015.029 Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism? Eun-Sun Jin, MD, PhD Cardiovascular

More information

ICD Diagnostics: Making the most of it

ICD Diagnostics: Making the most of it ICD Basics and Beyond ICD Diagnostics: Making the most of it Dulce Obias-Manno RN, BSN, MHSA, FHRS, CCDS, CEPS Cardiac Arrhythmia Center, Device Clinic Washington Hospital Center, Washington DC Objectives

More information

MRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual

MRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual MRI Procedure Information For St. Jude Medical MR Conditional Neurostimulation Systems Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical

More information

Project Title Temporary Pacemaker Training Simulator

Project Title Temporary Pacemaker Training Simulator Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially

More information

DRG STIMULATION FOR CHRONIC PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN

DRG STIMULATION FOR CHRONIC PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN FOCUS ON YOUR LIFE NOT YOUR PAIN DRG STIMULATION FOR CHRONIC PAIN Focus on your friendships, not on your pain. FOCUS ON YOUR LIFE, NOT ON YOUR PAIN. Chronic pain is personal, and it can be difficult to

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients

Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients Automatic Atrial Threshold Measurement and Adjustment in Pediatric Patients ANITA HIIPPALA, M.D.,* GERALD A. SERWER, M.D., EVA CLAUSSON, M.Sc., LYNN DAVENPORT, M.S., TRINA BRAND, PH.D., and JUHA-MATTI

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish

More information

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children ERALD A. SERWER, M.D., MACDONALD DICK 11, M.D., KAREN UZARK, R.N., Ph.D., WILLIAM A. SCOTT, M.D., and EDWARD L.

More information

PORTICO CE TRIAL ASSESSMENT OF THE ST. JUDE MEDICAL PORTICO TRANSCATHETER AORTIC VALVE IMPLANT AND THE TRANSFEMORAL DELIVERY SYSTEM

PORTICO CE TRIAL ASSESSMENT OF THE ST. JUDE MEDICAL PORTICO TRANSCATHETER AORTIC VALVE IMPLANT AND THE TRANSFEMORAL DELIVERY SYSTEM PORTICO CE TRIAL ASSESSMENT OF THE ST. JUDE MEDICAL PORTICO TRANSCATHETER AORTIC VALVE IMPLANT AND THE TRANSFEMORAL DELIVERY SYSTEM Preliminary Results: Acute and 1-Year Outcomes Ganesh Manoharan, MBBCh,

More information

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Patient Scheduling and Screening Q: Is there any special (other than regular MRI consent) consent form that should be signed by the patient?

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value

More information

Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs

Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs Faruk Cingoz, MD, George Yanulis, MS, Elizabeth Ching, RN, Kiyotaka Fukamachi,

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

58 Marc-Alexander Ohlow, MD 1 ; Bernward Lauer, PhD 1 ; Michele Brunelli, PhD 2, Yunis Daralammouri, MD 1 ; J Christoph Geller, PhD 2 ;

58 Marc-Alexander Ohlow, MD 1 ; Bernward Lauer, PhD 1 ; Michele Brunelli, PhD 2, Yunis Daralammouri, MD 1 ; J Christoph Geller, PhD 2 ; www.ipej.org 58 Original Article The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing

More information

Device Diagnostics and Stroke Prevention: State of the Art

Device Diagnostics and Stroke Prevention: State of the Art Device Diagnostics and Stroke Prevention: State of the Art Ass. Prof. Dr. Carsten W. Israel Dept. of Cardiology Evangelical Hospital Bielefeld Germany C.W.Israel@em.uni-frankfurt.de Disclosure of Relationships

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

Solara CRT-P MRI SureScan W1TR06

Solara CRT-P MRI SureScan W1TR06 Solara CRT-P MRI SureScan W1TR06 MR Conditional pacemaker with cardiac resynchronization therapy, SureScan technology, and Bluetooth wireless telemetry (OAE-DDDR) Device Manual The following list includes

More information

Pacemaker Concepts and Terminology*

Pacemaker Concepts and Terminology* Pacemaker Concepts and Terminology* BAROUH V BERKOVITS, E.E., Ing. Associate in Surgery, Harvard Medical School, Boston, Mass., Associate in Electrophysiology, Miami University, Miami, Fla., Senior Research

More information

ENABLE MRI CLINICAL STUDY SUMMARY

ENABLE MRI CLINICAL STUDY SUMMARY CLINICAL STUDY SUMMARY ENABLE MRI CAUTION: Federal Law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. TABLE OF CONTENTS

More information

A Rare Type of Pacemaker Mediated Tachycardia

A Rare Type of Pacemaker Mediated Tachycardia A Rare Type of Pacemaker Mediated Tachycardia Serkan ÇAY M.D., Ümit GÜRAY M.D., Cihan DEMİR M.D. Yuksek Ihtisas Heart-Education and Research Hospital, Department of Cardiology, Ankara, Turkey ABSTRACT

More information

Introduction to a new "X out of Y" Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers

Introduction to a new X out of Y Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers April 1999 117 Introduction to a new "X out of Y" Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers C. W. ISRAEL Department of Cardiology, St.-Josef-Hospital, Ruhr-University, Bochum,

More information

A Long-Term Clinical Experience with an Active Fixation Lead for Atrial Application

A Long-Term Clinical Experience with an Active Fixation Lead for Atrial Application June 2002 77 A Long-Term Clinical Experience with an Active Fixation Lead for Atrial Application M.J.J.A. BOKERN, R. LECHNER, J. COEBERG Waterland Hospital, Purmerend, The Netherlands J.C.J. RES, R. ABELS,

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases Case Report Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases Koji Tsutsumi MD, Tatsuru Niibori MD, Keiichiro Katsumoto MD Department

More information

Acute Evaluation of the Post Atrial Stimulation Evoked Response at Various Sites

Acute Evaluation of the Post Atrial Stimulation Evoked Response at Various Sites June 2000 311 Acute Evaluation of the Post Atrial Stimulation Evoked Response at Various Sites Y. GUYOMAR, P. GRAUX, R. CARLIOZ, C. MOULIN, A. DUTOIT Catholic University of Lille, St-Philibert Hospital,

More information

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Programming of Bradycardic Parameters C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Programming of ICD Brady Parameters Conflict of Interest Biotronik

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:

More information

LONGITUDINAL SURVEILLANCE REGISTRY OF ACUITY SPIRAL LEAD

LONGITUDINAL SURVEILLANCE REGISTRY OF ACUITY SPIRAL LEAD CLINICAL SUMMARY LONGITUDINAL SURVEILLANCE REGISTRY OF ACUITY SPIRAL LEAD CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and

More information

Atrial Near-Field and Ventricular Far-Field Analysis by Automated Signal Processing at Rest and During Exercise

Atrial Near-Field and Ventricular Far-Field Analysis by Automated Signal Processing at Rest and During Exercise Atrial Near-Field and Ventricular Far-Field Analysis by Automated Signal Processing at Rest and During Exercise Frank Eberhardt, M.D., Hendrik Bonnemeier, M.D., Martin Lipphardt, M.Sc., Ulrich G. Hofmann,

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

Azure XT DR MRI SureScan W2DR01

Azure XT DR MRI SureScan W2DR01 Azure XT DR MRI SureScan W2DR01 MR Conditional dual chamber pacemaker with SureScan technology and Bluetooth wireless telemetry (OAE-DDDR) Device Manual The following list includes trademarks or registered

More information

CONTAK RENEWAL CLINICAL SUMMARY

CONTAK RENEWAL CLINICAL SUMMARY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY CONTAK RENEWAL Boston Scientific

More information

Biventricular Pacemakers Sensing

Biventricular Pacemakers Sensing Biventricular Pacemakers Sensing Sergio Dubner Director Arrhythmias and Electrophisiology Service, Clinica y Maternidad Suizo-Argentina, Buenos Aires, Argentina Left ventricular pacemaker lead technology

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

PressureWire Agile Tip Technology

PressureWire Agile Tip Technology PressureWire Agile Tip Technology FINALLY. INTRODUCING PRESSUREWIRE WITH AGILE TIP St. Jude Medical continues to set new standards in fractional flow reserve (FFR) measurement. With the next generation

More information

FRONTIER II. User s Manual. Model 5586 Model Biventricular Cardiac Resynchronization Device with AF Suppression Algorithm

FRONTIER II. User s Manual. Model 5586 Model Biventricular Cardiac Resynchronization Device with AF Suppression Algorithm FRONTIER II Model 5586 Model 5596 Biventricular Cardiac Resynchronization Device with AF Suppression Algorithm User s Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed

More information