Thousands of patients die annually from acute and

Size: px
Start display at page:

Download "Thousands of patients die annually from acute and"

Transcription

1 Hemodynamic Effect of a Low-Resistance Artificial Lung in Series With the Native Lungs of Sheep William R. Lynch, MD, J. Patrick Montoya, PhD, David O. Brant, MS, Robert J. Schreiner, MD, Mark D. Iannettoni, MD, and Robert H. Bartlett, MD Department of Surgery, University of Michigan Medical Center, Ann Arbor, Michigan Background. An artificial lung with 1 to 6 month work life could act as a bridge to transplantation. A pumpless artificial lung has been developed. Methods. The artificial lung was placed in series with the native lungs of adult sheep. Hemodynamics were observed, as the right ventricle generated flow through the device. Through a left thoracotomy, two 20-mm grafts were anastomosed in an end-to-side fashion to the pulmonary artery. The grafts were externalized, and directed flow through the chest wall, to the extracorporeal lung. The animals were recovered, weaned from the ventilator, and when standing, flow was diverted through the device. Results. Five of 7 animals survived 24 hours with 75% to 100% of the cardiac output diverted through the device. All animals were active, with interest in food and water, and able to stand. Conclusions. The right ventricle perfused the artificial lung with 75% to 100% of the cardiac output for 24 hours. This device demonstrates the feasibility of a pumpless pulmonary assist device relying on the right ventricle for perfusion. (Ann Thorac Surg 2000;69:351 6) 2000 by The Society of Thoracic Surgeons Presented at the Thirty-fifth Annual Meeting of The Society of Thoracic Surgeons, San Antonio, TX, Jan 25 27, Address reprint requests to Dr Bartlett, Section of General Surgery, University of Michigan, Taubman 2920, Box 0331, 1500 E Medical Center Dr, Ann Arbor, MI 48109; robbar@umich.edu. Thousands of patients die annually from acute and chronic respiratory failure. When unsuccessfully treated, the respiratory failure becomes irreversible, and the only viable treatment is lung transplantation. Extracorporeal life support (ECLS) evolved as a treatment option in respiratory failure, providing support while offering time for diseased lungs to recover. ECLS has been accepted as treatment for severe respiratory failure in newborns with survival rates approaching 80%. In children and adults, however, ECLS survival is closer to 50% [1, 2]. Although not widely accepted for these patients, ECLS can be effective therapy for up to 1 month. When ECLS is used beyond this point, it becomes too complicated and expensive, making it an impractical treatment for prolonged support. For children and adults, those who do not recover during this 30 day window, progress to end-stage pulmonary fibrosis without return of native lung function. A possible treatment would be lung transplantation, however, these patients are unattractive candidates because of the infections and bleeding complications that so often are a part of prolonged ECLS. With no other treatment options, these patients die. Lung transplantation has played a successful role in the treatment of chronic respiratory failure such as emphysema, chronic obstructive pulmonary disease, cystic fibrosis, idiopathic pulmonary fibrosis, 1 -antitrypsin deficiency, and primary pulmonary hypertension. Transplantation is successful for patients who are not acutely ill, and has been reserved for those with an estimated life expectancy of 18 months or less. Donor availability is a major limitation with the average wait for a lung, approaching 2 years. For those who do receive an organ, 1-year survival is about 70% [3]. Some transplant candidates progress to end-stage respiratory failure while waiting for a suitable donor organ. Others acutely decompensate, requiring mechanical ventilation. Infectious complications become prohibitive for these patients, and transplantation becomes less likely. ECLS has been considered as a bridge-to-transplantation for this patient group but, again, complications limit therapy long before a suitable organ becomes available. A pulmonary assist device, capable of supporting patients for up to 6 months, could sustain both acute and chronic end-stage respiratory failure patients. Prolonged support could offer time for resolution of acute failure as well as provide a window of opportunity for lung transplantation. The concept for such a device is not new, and preliminary reports appeared in the literature in the 1970s [4 7]. Various design strategies began to evolve, as membrane technology improved, making gas exchange more efficient [8 10]. An oxygenator that did not rely on a mechanical pump was an attractive strategy, because the concept was simple. Investigators considered using the right heart or an arteriovenous shunt to generate flow to a pumpless oxygenator [11, 12]. Others tried intravascular oxygenators (IVOX), exposing gas exchange membranes to the cardiac output within the vena cava [13 15]. Some investigators included pumps, incorporating a circulatory assist device with an oxygenator as a form of cardiopulmonary support [16, 17]. Each of these strate by The Society of Thoracic Surgeons /00/$20.00 Published by Elsevier Science Inc PII S (99)

2 352 LYNCH ET AL Ann Thorac Surg LOW RESISTANCE ARTIFICIAL LUNG 2000;69:351 6 gies has demonstrated the potential for partial support of respiratory function and investigation continues. Our strategy is development of a pumpless pulmonary assist device, perfused by the right ventricle. Design goals include total respiratory support, 6-month durability, and the potential to be implanted. Fundamental to the success of this strategy is the ability of the right ventricle to tolerate perfusing an oxygenator for prolonged periods of time and under a variety of conditions. An oxygenator with minimal resistance to blood flow is a critical component of this design strategy. A prototype artificial lung (Michigan Critical Care Consultants, Ann Arbor, MI) has been designed for this purpose. Unique construction of the shell and arrangement of the membrane provides minimal resistance to blood flow. Earlier generations of this prototype were investigated in anesthetized sheep [18], with the device in series with the native lungs. This preliminary work demonstrated that the right ventricle could tolerate perfusing the low resistance device for short periods of time under controlled conditions. This study is an extension of that early work, and evaluates the current version of the artificial lung in series with the native lungs of an active animal model. Material and Methods Surgical Preparation Seven adult sheep (55 3 kg) were fasted for 24 hours before surgical preparation. Each animal was sedated with 5 to 10 mg/kg sodium thiopental (Abbott Laboratories, North Chicago, IL), and placed in a supine position upon the operating table. The animal was initially mask ventilated with Isoflurane (Abbott Laboratories, North Chicago, IL) using a gas anesthesia machine (Narkomed 2; North American Drager, Telford, PA). Fifty-thousand /kg penicillin G potassium (Marsam Pharmaceuticals, Cherry Hill, NJ) was given intravenously before skin incision. A midline tracheotomy was performed, a cuffed tracheostomy tube placed, and anesthesia continued through the tracheostomy. The left femoral artery and vein were surgically exposed and isolated. Invasive arterial pressure was measured through the femoral artery using a fluid coupled strain-gauge pressure transducer, and the signal was inverted for heart rate display (Hewlett Packard 78304A, 78901A; Hewlett Packard, Andover, MA). The lines were secured and the wounds closed in two layers. Lactated Ringer s solution was used as maintenance infusion and an orogastric tube was placed. The animal was repositioned with the right side down and paralyzed with 0.1 mg/kg pancuronium bromide (Elkins-Sinn, Cherry Hill, NJ). A muscle-sparing left lateral thoracotomy was performed through the fourth rib interspace, and the fourth rib was removed. The left lung was collapsed and the pericardium incised, exposing the heart and great vessels. The aorta was isolated, and a transit time ultrasonic flow probe was placed for continuous cardiac output monitoring (Transonic T101; Transonic Systems, Ithaca, NY). The pulmonary artery was isolated, and a 20-mm vascular flow occluder (In Vivo Metric, Healdsburg, CA) was positioned at midlength. A 20-mm Dacron (C.R. Bard, Haverhill, PA) vascular graft, bonded to a 25-cm length of three-quarter inch polyvinylchloride tubing, was used as the connecting conduits to divert blood to and from the prototype lung. The grafts were anastomosed to the pulmonary artery in an end-to-side fashion on either side of the vascular occluder. Heparin sodium (Elkins Sinn, Cherry Hill, NJ) was given as a bolus infusion of 100 /kg as the conduits were placed. The two conduits, the aortic flow probe cable and the vascular occluder tubing were tunneled through the subcutaneous tissue and externalized to the chest wall. A chest tube was placed, the lung reexpanded, and the chest was closed in three layers. The prototype lung was primed with 240 cc of heparinized saline, and attached to the connecting conduits. The prototype was secured to the lateral chest wall and stabilized with a dressing. A second flow probe was placed on the inlet conduit to the prototype lung to measure continuously the diverted flow rate. The animal was removed from the operating table and placed in a padded pen in the sternal position. The chest tube was placed on 20 cm H 2 O continuous suction. Initially, the sweep gas through the artificial lung was oxygen at a rate of 2 L/min. Vacuum was applied to the gas outlet of the lung at 20 cm H 2 O. Volume cycled mechanical ventilation (Bennett MA-1; Bennett Respiratory Products, Santa Monica, CA) supported the sheep as it recovered from anesthesia, and the sheep was weaned to room air as directed by blood gas analysis (ABL 505, OSM 3 Hemoximeter; Radiometer, Copenhagen, Denmark). The experimental protocol began once the animal was spontaneously breathing room air and had demonstrated the ability to stand. Prototype Artificial Lung The prototype membrane was a hollow fiber, microporous polypropylene mat, with 300 m outer diameter fibers and a surface area of 2.25 m 2. The membrane was wrapped around a mandrill, and housed in a cylindrical shell with two circular endcaps. The rigid housing was made of ABS plastic (Michigan Critical Care Consultants, Ann Arbor, MI). Flow entered the device through a 20-mm diameter inlet at the center of one endcap. Blood entered in the axial direction, through the core, and then proceeded through the membrane in a radial fashion (Fig 1). This manner of flow allowed multiple passes across the wound membrane, providing efficient gas exchange. The blood reached the periphery of the device, and was directed by a gutter to exit the prototype. In vitro studies of gas exchange, using bovine blood, demonstrated delivery of 350 cc O 2 /min at 4.0 L/min blood flow, with a pressure gradient across the device of only 7.5 mm Hg. The prototype lung is pictured in Figure 2. Experimental Protocols Animal hemodynamics were characterized by heart rate, blood pressure, and cardiac output. Once spontaneously breathing room air, and with the ability to stand, flow was diverted through the prototype by slowly inflating the vascular occluder with saline. With the occluder fully inflated, maximum flow was established. Acceptable initial prototype flow was defined to be at least 75% of the

3 Ann Thorac Surg LYNCH ET AL 2000;69:351 6 LOW RESISTANCE ARTIFICIAL LUNG 353 cardiac output. Hemodynamic characteristics were recorded before and after inflating the vascular occluder. Animals were observed for 24 hours, while maximum flow was diverted to the prototype. Hemodynamic characteristics were recorded hourly. Activated clotting time (Hemochron 800; Technidyne Corporation, Edison, NJ) was measured every 1 to 3 hours and heparin sodium was given to keep clotting times between 200 to 250 seconds. Fifty-thousand /kg penicillin G potassium was given every 6 hours, as was 5.0 g/kg buprenorphine hydrochloride (Reckitt & Colman Pharmaceuticals, Richmond, VA) for analgesia. Hourly measurements of the artificial lung included flow rate, inlet pressure, and outlet pressure. Blood gas analysis of systemic blood, and blood entering and exiting the artificial lung, was done hourly. Sweep gas was adjusted to keep the tension of carbon dioxide exiting the artificial lung within the normal range for mixed venous blood of sheep (pco 2 35 to 40 mm Hg). Inotropic and vasoactive drugs were not used. Animals that appeared to be in distress were sacrificed. At the conclusion of the experiment, the sheep were euthanized with 300 mg/kg sodium pentobarbitol (Schering Plough, Kenilworth, NJ). The sheep used in these experiments were cared for according to guidelines established by the University of Michigan Unit for Laboratory Animal Medicine and the National Institutes of Health, Guidelines for the Care and Use of Laboratory Animals. Fig 1. Prototype artificial lung. Cross-section through the long axis of the prototype artificial lung depicting membrane, blood path, and gas path. Fig 2. The prototype artificial lung and connecting conduits. Prototype artificial lung with connecting conduits and graft material. Dimensions of housing are 170 mm in length with an 80 mm diameter. The 2.25 m 2 membrane is hollow fiber microporous polypropylene with fibers of 300 m outer diameter. Oxygen transfer rate is 350 cc/min at 4 L/min blood flow rate. The pressure gradient generated at that flow rate is 7.5 mm Hg. Results Seven sheep underwent and survived surgical placement of the prototype lung. Each survived the immediate postoperative period, and was weaned from the ventilator to room air. Five of these animals tolerated full occlusion of the pulmonary artery to establish flow through the prototype lung of at least 75% of the cardiac output. The hemodynamics are reported as preocclusion and postocclusion of the pulmonary artery, and appear in Table 1. The percentage change of each parameter is based on the preocclusion value. Two animals did not tolerate occlusion of the pulmonary artery. One became hemodynamically unstable as the occluder was inflated. This animal did not recover with deflation of the occluder and was sacrificed. The second animal did not tolerate total occlusion, becoming hypotensive and with a falling cardiac output. Partial occlusion was tolerated, but resulted in less than 50% flow through the prototype. This animal was also sacrificed. Five animals tolerated total inflation of the vascular occluder, and were observed for 24 hours. The hemodynamic parameters, and the mixed venous saturation for each of these animals are presented in Table 2. The data are the mean standard deviation over the 24 hour experiment. Table 3 displays oxygen transfer and the pressure gradient of the prototype. Again, these data are presented as means standard deviation for the 24 hours. Figure 3 displays the oxygen transferred as a function of prototype flow rate. The data used for this plot are pooled from the five animal trials. Comment In this study using awake and active sheep, the initial diversion of cardiac output through the prototype lung was tolerated with little hemodynamic change. The continuous diversion of flow through the prototype for 24 hours was also tolerated. Hemodynamics were varied during the 24 hours for each animal, as well as when compared between animals. Mixed venous saturation indicated adequate perfusion for all animals. All animals remained active throughout the 24 hours, indicating adequate cardiopulmonary function. The initial and 24 hour results, along with these observations, demonstrate that the right ventricle was able to perfuse the prototype lung for 24 hours, without disrupting cardiovascular performance. In addition, the performance of the prototype artificial lung closely matched in vitro measures for oxygen delivery and pressure gradient. The first question addressed was if an awake and active sheep would hemodynamically tolerate diversion of the cardiac output through the prototype. Prior to diverting flow, the animal had to be spontaneously breathing room air and able to stand. These requirements reflected anesthetic recovery. Flow through the device was established by slowly inflating the vascular occluder. Preliminary investigations demonstrated that

4 354 LYNCH ET AL Ann Thorac Surg LOW RESISTANCE ARTIFICIAL LUNG 2000;69:351 6 Table 1. Initial Hemodynamic Parameters as Flow Established Through Device a Animal Trial Mean Heart Rate (beat/min) Change Blood Pressure (mm Hg) Change Cardiac Pre Post Pre Post Pre Post Change Device Flow Rate % Cardiac SvO Error Error Error 3.5 Error a Hemodynamic parameters as occluder is inflated, diverting blood from the pulmonary artery to the prototype lung. Data are hemodynamics before occluder is inflated (pre) and after occlusion (post) has diverted blood flow to the prototype. Error flow probe acoustical error. flows of 75% to 100% could be reproducibly achieved, if the anastomoses were patent and the pulmonary artery adequately occluded. The ability to establish at least 75% flow through the prototype was arbitrarily chosen to reflect successful surgical preparation. Diverting 75% of cardiac output through the prototype was felt to be a sufficient challenge to the right ventricle for 24 hours. Two of the 7 animals failed to meet this standard. These animals were sacrificed, and at necropsy the anastomoses were noted to be widely patent. It was observed that each animal had relatively short pulmonary arteries, which resulted in close graft positioning. With inflation of the vascular occluder, the back wall of the artery was drawn within close proximity to each arteriotomy. It is possible that this resulted in obstruction to flow, increasing resistance, and acute right heart failure. Five animals survived diversion of flow through the prototype lung, and the results appear in Table 1. The percentage change in these parameters as flow was established is listed. The cardiac output for animal 4 is not reported because an acoustical error was occurring as flow was diverted. The animal remained in the study because flow through the prototype equaled expected cardiac output for a sheep this size. The acoustical error subsequently resolved, revealing prototype flow to be 95% of cardiac output. Only the heart rate in animal 4, and the mean blood pressure in animal 3 changed by greater than 10% as flow was established (12.3% and 11.5% respectively). All other changes were less than 10%, with most less than 5%. Hemodynamically, these 5 animals tolerated diversion of flow through the prototype lung. The 5 animals were observed for 24 hours, with flow diverted through the prototype. All animals survived and remained awake, active, with interest in food and water for the duration of the study. Table 2 displays hemodynamic data summarized as the mean standard deviation over 24 hours. Mixed venous saturation is also included in this table. The data in Table 1 represent baseline hemodynamics for the 24 hour study. Reviewing Table 2, the standard deviation for most entries reflects wide variation in each characteristic. There is also a large range of means when comparing the 5 animals. Comparing mean 24 hour heart rate, blood pressure, and cardiac output to baseline values from Table 1, some observations can be made. The mean 24 hour heart rate is elevated in each animal, when compared to baseline values. All rhythms were sinus tachycardias. Mean 24 hour blood pressure was similar to baseline for each animals. Mean cardiac output was increased from baseline in animals 2, 3, and 5, but decreased in animal 1 (baseline not available for 4). Continuing review of Table 2, animal 2 is noticeable for a persistent hypotension comparable to the baseline pressure. This animal remained awake and active with good urine output for the 24 hour study. Cardiac output was generous, and mixed venous saturation indicated adequate perfusion. Drawing attention to animal 2, the standard deviation for percent cardiac output (Table 2) is Table 2. Hemodynamic Variables Averaged Over 24 Hours for Each Animal a Animal Trial Heart Rate (beats/min) Mean Blood Pressure (mm Hg) Cardiac Prototype Flow Rate % Cardiac SvO a Hemodynamic parameters represent the 24 hours flow was diverted through the prototype artificial lung. Mixed venous saturations are included for each animal. Data are means standard deviation for the duration of the study. % Cardiac (prototype flow rate/cardiac output) 100; SvO 2 mixed venous saturation.

5 Ann Thorac Surg LYNCH ET AL 2000;69:351 6 LOW RESISTANCE ARTIFICIAL LUNG 355 Table 3. Prototype Artificial Lung Performance Measures a Animal Trial Oxygen Transfer (cc/min) Pressure Gradient (mm Hg) a Prototype artificial lung performance was reflected by oxygen transfer and pressure gradient. Oxygen transfer was calculated by calculating oxygen content of blood entering and exiting the artificial lung. The difference was multiplied by the flow rate through the prototype. The pressure gradient was the pressure drop across the prototype. This was the difference of the mean pressure of blood entering the prototype minus the mean pressure exiting the prototype. Mean pressure was measured using fluid coupled pressure transduction at the inlet and outlet ports of the prototype. Oxygen Content as Volume % [(Hb g/dl %sat 1.36 cc/g) (Po cc O 2 /mm Hg/dL)]; Oxygen Transfer [(Oxygen Content as Volume %) 10 dl/l] prototype flow rate. large. A leak in the vascular occluder, allowing intermittent deflation, was responsible for this spread. Mixed venous saturation data are included in both Tables 1 and 2. Mean values for the 24 hour preparation are comparable to the baseline values listed in Table 1. The mixed venous saturations were stable for the duration of the experiment, without drifting far from baseline values. In active, awake animals, the stable mixed venous saturations remained within the normal limits for sheep, suggesting the cardiopulmonary performance was satisfactory to support metabolic demands. The wide variability is difficult to interpret. There are many factors in the immediate postoperative period that could explain such variation. A partial list includes normal physiology, pneumothorax, hemorrhage, and sepsis. At necropsy, there was no evidence of significant hemorrhage or pneumothorax to the side opposite the thoracotomy. Chest tubes appeared to function during the 24 hour preparation, but partial or total collapse on the Fig 3. Oxygen transfer rate versus flow rate of blood through the prototype artificial lung. surgical side could not be ruled out. Hemodynamic performance did not suggest sepsis. Another cause of hemodynamic variation could have been right ventricular dysfunction related to the work of perfusing the prototype. Specific measures of right ventricular function were not made, and this too, could not be ruled out. Whatever the cause, mixed venous saturations, as an indicator of integrated cardiopulmonary performance, suggested hemodynamics compensated adequately to support metabolism and activity. Although there was hemodynamic variability, none of the animals suffered catastrophic hemodynamic collapse. Anticipated causes of hemodynamic collapse from this surgical preparation could be exsanguination, pulmonary embolus, tension pneumothorax, and sepsis. Another reasonable cause could be right ventricular failure related to the prototype lung, with an active animal model, exaggerating this cardiovascular stress. Because hemodynamic collapse did not occur during the 24 hours for any of the animals, we concluded the prototype lung did not cause right ventricular failure. Table 3 presents data related to performance of the prototype lung. Oxygen transfer and the pressure gradient are presented for each trial as the mean standard deviation over the 24 hour duration. Carbon dioxide transfer is not reported because the sheep continued to breathe, regulating ventilation. Sweep gas was adjusted to maintain mixed venous pco 2 in normal ranges (35 to 40 mm Hg). The ability of earlier versions of this prototype to clear CO 2 has been previously demonstrated and reported [18]. Oxygen transfer was sufficient to support oxygen demands of animals this size. Oxygen transfer data were pooled from all 5 animal trials, and plotted against device blood flow in Figure 3. The best linear fit yields the expected oxygen transfer rate for this prototype. Compared to in vitro testing, oxygen transfer was approximately 66% predicted. This difference is because the hemoglobin content of sheep blood is normally 8 to 10 g/dl. Bovine blood was used for the in vitro testing with normal hemoglobin values of 12 to 15 g/dl. Hemodynamic collapse, as a marker of right ventricular failure, although crude, is easy to recognize in an active animal. While the many hemodynamic variations in this experiment are unexplained, hemodynamic collapse did not occur. The hemodynamic variations could be the result of the prototype device, but the absence of hemodynamic collapse suggests the right ventricle can tolerate perfusing the prototype in an active animal. The flow rate generated through the prototype was sufficient for oxygen exchange capable of total support. A better understanding of the cardiovascular impact of the prototype will result from specific measures of right ventricular function. Echocardiography and invasive pressure monitoring would be useful adjuncts when evaluating right ventricular performance as it perfuses this artificial lung. This knowledge will become important as the design iteration continues. In summary, the prototype artificial lung was tolerated as an addition to the pulmonary circulation for 24 hours in an active animal model. These results suggest that the right ventricle has the ability to perfuse this prototype for

6 356 LYNCH ET AL Ann Thorac Surg LOW RESISTANCE ARTIFICIAL LUNG 2000;69:351 6 extended periods of time, while accommodating the additional stress of activity. The flow rates generated through the prototype were sufficient to maintain metabolic needs during activity. This pilot study of this prototype artificial lung suggests the concept of a pumpless pulmonary assist device could be a practical method of prolonged respiratory support. We thank Edward Salah and Shigeki Sawada for their surgical and technical assistance. We also appreciate the financial support of the National Institutes of Health (HL ), making this work possible. References 1. Bartlett RH. Extracorporeal life support for cardiopulmonary failure. Curr Prob Surg 1990;27: Tracy TF, DeLosh T, Stolar CJH. The registry of the extracorporeal life support organization. In: Zwischenberger JB, Bartlett RH, eds. ECMO: Extracorporeal cardiopulmonary support in critical care. Ann Arbor: ELSO, 1995: Davis RD, Pasque MK. Pulmonary transplantation. Ann Surg 1995;221: Bodell BR, Head JM, Head LR, Formolo AJ. An implantable artificial lung. JAMA 1965;191: Shah-Mirany J, Head LR, Ghetzler R, Formolo AJ, Palmer AS, Bodell BR. An implantable artificial lung. Ann Thorac Surg 1972;13: Palmer AS, Collins J, Head LR. Development of an implantable artificial lung. J Thorac Cardiovasc Surg 1973;66: Mortensen JD. Feasibility of total lung replacement by a pumpless prosthetic lung. Artif Organs 1978;2: Clowes GHA, Jr, Hopkins AL. Further studies with plastic films and their use in oxygenating blood. Trans ASAIO 1956; 2: Bodell BR, Head JM, Head LR, Formolo AJ, Head JR. A capillary membrane oxygenator. J Thorac Cardiovasc Surg 1963;46: Keller KH, Shultis KL. Oxygen permeability in ultrathin and microporous membranes during gas-liquid transfer. Trans ASAIO 1979;25: Ohtake S, Kawashima Y, Hirose H, et al. Experimental evaluation of pumpless arteriovenous ECMO with polypropylene hollow fiber membrane oxygenator for partial respiratory support. Trans ASAIO 1983;29: Kostopoulos C, Rasidakis A, Vardas P, et al. Pumpless extracorporeal oxygenation of the blood through a hollow fiber oxygenator. Trans ASAIO 1985;31: Mortensen JD. An intravenacaval blood gas exchange (IVCBGE) device: preliminary report. ASAIO J 1987;33: High KM, Snider MT, Richard R, et al. Clinical trials of an intravenous oxygenator in patients with adult respiratory distress syndrome. Anesthesiology 1992;77: Hattler BG, Johnson PC, Sawzik PJ, et al. Respiratory dialysis : a new concept in pulmonary support. ASAIO J 1992;38:M Tatsumi E, Takano H, Taenaka Y, et al. A novel method for long-term respiratory support. Trans ASAIO 1989;35: Tatsumi E, Taenaka Y, Nakatani T, et al. A VAD and novel high performance compact oxygenator for long-term ECMO with local anticoagulation. Trans ASAIO 1990;36:M Fazzalari FL, Montoya JP, Bonnell MR, Bliss DW, Hirschl RB, Bartlett RH. The development of an implantable artificial lung. ASAIO J 1994;40:M DISCUSSION DR RICHARD N. PIERSON III (Nashville, TN): That was an excellent paper, and this would appear to be very exciting technology. Two points of clarification about the model. Were your cannulae proximal and distal on the main pulmonary artery, and is that how you are able to divert full flow through the device? And secondly, is any anticoagulation required to maintain perfusion through this device? DR LYNCH: Thank you for your comments. To your first question, the grafts used were 20-mm vascular grafts. The pulmonary artery of a sheep is quite long, and that lends itself well to this particular type of study. The grafts were end-to-side anastomoses, and allowed upwards of 3 L, and at times 6 L, of flow through the device. DR PIERSON: So both grafts were sewn to the main pulmonary artery, and the entire cardiac output was diverted through the device? DR LYNCH: This was the main pulmonary artery. The series flow was established by shunting the flow through the device, from the proximal pulmonary artery, and then returning it to the main pulmonary artery, which then went through the pulmonary vasculature. In answer to your second question, yes, we heparinized the animals with a bolus infusion at grafting and then kept ACTs of 200 to 220. DR CAROLYN E. REED (Charleston, SC): Doctor Lynch, could you share with us what your next project will be using the model, what your next series of experiments are that you plan? DR LYNCH: The ultimate goal of this kind of technology is to have a device that could hopefully perfuse from pulmonary artery (PA) and return to left atrium. Two chronic studies that are underway: one PA-PA for 5 days, and one PA-LA for 5 days. DR SI M. PHAM (Miami, FL): Did the resistance across the oxygenator increase with time? DR LYNCH: No, it did not. And at the conclusion of the study, there is no minimal thrombus and no evidence or reason for the resistance to change. DR FLORIAN M. WAGNER (Dresden, Germany): Congratulations, a very interesting study. Obviously, if I understood it correctly, this device is very similar to the membrane oxygenator as we use on cardiopulmonary bypass during cardiac operation. Therefore the most important concern would be that it causes hemolysis and stimulation of all sorts of inflammatory responses, particularly if you keep any kind of circulation going through this device for more than, let us say, 12 hours. So I am wondering, did you get a chance to look at any of these inflammatory response parameters, or hemolysis? DR LYNCH: The hemolysis and inflammation with cardiopulmonary bypass is caused by coronary suction. In these experiments hemolysis and inflammation was negligible.

Hemodynamic Monitoring and Circulatory Assist Devices

Hemodynamic Monitoring and Circulatory Assist Devices Hemodynamic Monitoring and Circulatory Assist Devices Speaker: Jana Ogden Learning Unit 2: Hemodynamic Monitoring and Circulatory Assist Devices Hemodynamic monitoring refers to the measurement of pressure,

More information

Extracorporeal Life Support Organization (ELSO) Guidelines for Pediatric Respiratory Failure

Extracorporeal Life Support Organization (ELSO) Guidelines for Pediatric Respiratory Failure Extracorporeal Life Support Organization (ELSO) Guidelines for Pediatric Respiratory Failure Introduction This pediatric respiratory failure guideline is a supplement to ELSO s General Guidelines for all

More information

Extracorporeal life support (ECLS) supplements the

Extracorporeal life support (ECLS) supplements the CARDIOPULMONARY SUPPORT AND PHYSIOLOGY A PROSPECTIVE COMPARISON OF ATRIO-FEMORAL AND FEMORO-ATRIAL FLOW IN ADULT VENOVENOUS EXTRACORPOREAL LIFE SUPPORT Preston B. Rich, MD Samir S. Awad, MD Stefania Crotti,

More information

ECMO as a bridge to durable LVAD therapy. Jonathan Haft, MD Department of Cardiac Surgery University of Michigan

ECMO as a bridge to durable LVAD therapy. Jonathan Haft, MD Department of Cardiac Surgery University of Michigan ECMO as a bridge to durable LVAD therapy Jonathan Haft, MD Department of Cardiac Surgery University of Michigan Systolic Heart Failure Prevalence 4.8 million U.S. 287,000 deaths per year $39 billion spent

More information

ECLS as Bridge to Transplant

ECLS as Bridge to Transplant ECLS as Bridge to Transplant Marcelo Cypel MD, MSc Assistant Professor of Surgery Division of Thoracic Surgery Toronto General Hospital University of Toronto Application of ECLS Bridge to lung recovery

More information

Case scenario V AV ECMO. Dr Pranay Oza

Case scenario V AV ECMO. Dr Pranay Oza Case scenario V AV ECMO Dr Pranay Oza Case Summary 53 y/m, k/c/o MVP with myxomatous mitral valve with severe Mitral regurgitation underwent Mitral valve replacement with mini thoracotomy Pump time nearly

More information

Extracorporeal Membrane Oxygenation (ECMO)

Extracorporeal Membrane Oxygenation (ECMO) Extracorporeal Membrane Oxygenation (ECMO) Policy Number: Original Effective Date: MM.12.006 05/16/2006 Line(s) of Business: Current Effective Date: HMO; PPO; QUEST Integration 01/01/2017 Section: Other/Miscellaneous

More information

Hemodynamic Monitoring

Hemodynamic Monitoring Perform Procedure And Interpret Results Hemodynamic Monitoring Tracheal Tube Cuff Pressure Dean R. Hess PhD RRT FAARC Hemodynamic Monitoring Cardiac Rate and Rhythm Arterial Blood Pressure Central Venous

More information

Extracorporeal Membrane Oxygenation (ECMO)

Extracorporeal Membrane Oxygenation (ECMO) Extracorporeal Membrane Oxygenation (ECMO) Policy Number: Original Effective Date: MM.12.006 05/16/2006 Line(s) of Business: Current Effective Date: HMO; PPO; QUEST Integration 11/01/2014 Section: Other/Miscellaneous

More information

University of Florida Department of Surgery. CardioThoracic Surgery VA Learning Objectives

University of Florida Department of Surgery. CardioThoracic Surgery VA Learning Objectives University of Florida Department of Surgery CardioThoracic Surgery VA Learning Objectives This service performs coronary revascularization, valve replacement and lung cancer resections. There are 2 faculty

More information

Management of Cardiogenic Shock. Dr Stephen Pettit, Consultant Cardiologist

Management of Cardiogenic Shock. Dr Stephen Pettit, Consultant Cardiologist Dr Stephen Pettit, Consultant Cardiologist Cardiogenic shock Management of Cardiogenic Shock Outline Definition, INTERMACS classification Medical management of cardiogenic shock PA catheters and haemodynamic

More information

Innovative ECMO Configurations in Adults

Innovative ECMO Configurations in Adults Innovative ECMO Configurations in Adults Practice at a Single Center with Platinum Level ELSO Award for Excellence in Life Support Monika Tukacs, BSN, RN, CCRN Columbia University Irving Medical Center,

More information

In-Parallel Artificial Lung Attachment at High Flows in Normal and Pulmonary Hypertension Models

In-Parallel Artificial Lung Attachment at High Flows in Normal and Pulmonary Hypertension Models In-Parallel Artificial Lung Attachment at High Flows in Normal and Pulmonary Hypertension Models Begum Akay, MD, Junewai L. Reoma, MD, Daniele Camboni, MD, Joshua R. Pohlmann, MS, John M. Albert, BS, Ayushi

More information

Kevin K. Nunnink Extracorporeal Membrane Oxygenation Program

Kevin K. Nunnink Extracorporeal Membrane Oxygenation Program PATIE NT I N FO R M ATI O N Kevin K. Nunnink Extracorporeal Membrane Oxygenation Program A family s guide to understanding this specialized treatment for cardiac and pulmonary complications Saint Luke

More information

Ambulatory Oxygenator Right Ventricular Assist Device for Total Right Heart and Respiratory Support

Ambulatory Oxygenator Right Ventricular Assist Device for Total Right Heart and Respiratory Support Ambulatory Oxygenator Right Ventricular Assist Device for Total Right Heart and Respiratory Support Dongfang Wang, MD, Scott D. Lick, MD, Xiaoqin Zhou, MD, Xiaojun Liu, MD, Robert J. Benkowski, MD, and

More information

Artificial Lung: A New Inspiration

Artificial Lung: A New Inspiration Artificial Lung: A New Inspiration Joseph B. Zwischenberger MD Johnston-Wright Professor and Chairman: Department of Surgery j.zwische@uky.edu The University of Kentucky Lexington, Kentucky Presenter Disclosure

More information

37 1 The Circulatory System

37 1 The Circulatory System H T H E E A R T 37 1 The Circulatory System The circulatory system and respiratory system work together to supply cells with the nutrients and oxygen they need to stay alive. a) The respiratory system:

More information

FOCUS CONFERENCE 2018

FOCUS CONFERENCE 2018 FOCUS CONFERENCE 2018 Current Practice in Pediatric and Neonatal Extracorporeal Life Support Daniel W. Chipman, RRT Assistant Director of Respiratory Care Massachusetts General Hospital Boston, Massachusetts

More information

Perfusion for Repair of Aneurysms of the Transverse Aortic Arch

Perfusion for Repair of Aneurysms of the Transverse Aortic Arch technique This new section is open for technicians to explore the unusual, the difficult, the innovative methods by which perfusion meets the challenge of the hour and produces the ultimate goal - a life

More information

Comparison of Flow Differences amoiig Venous Cannulas

Comparison of Flow Differences amoiig Venous Cannulas Comparison of Flow Differences amoiig Venous Cannulas Edward V. Bennett, Jr., MD., John G. Fewel, M.S., Jose Ybarra, B.S., Frederick L. Grover, M.D., and J. Kent Trinkle, M.D. ABSTRACT The efficiency of

More information

Extra Corporeal Life Support for Acute Heart failure

Extra Corporeal Life Support for Acute Heart failure Extra Corporeal Life Support for Acute Heart failure Benjamin Medalion, MD Director Heart and Lung Transplantation Department of Cardiothoracic Surgery Rabin Medical Center, Beilinson Campus, Israel Mechanical

More information

Extracorporeal support in acute respiratory failure. Dr Anthony Bastin Consultant in critical care Royal Brompton Hospital, London

Extracorporeal support in acute respiratory failure. Dr Anthony Bastin Consultant in critical care Royal Brompton Hospital, London Extracorporeal support in acute respiratory failure Dr Anthony Bastin Consultant in critical care Royal Brompton Hospital, London Objectives By the end of this session, you will be able to: Describe different

More information

To ECMO Or Not To ECMO Challenges of venous arterial ECMO. Dr Emily Granger St Vincent s Hospital Darlinghurst NSW

To ECMO Or Not To ECMO Challenges of venous arterial ECMO. Dr Emily Granger St Vincent s Hospital Darlinghurst NSW To ECMO Or Not To ECMO Challenges of venous arterial ECMO Dr Emily Granger St Vincent s Hospital Darlinghurst NSW The Start: 1972 St Vincent s Hospital The Turning Point ECMO program restarted in 2004

More information

ECLS Registry Form Extracorporeal Life Support Organization (ELSO)

ECLS Registry Form Extracorporeal Life Support Organization (ELSO) ECLS Registry Form Extracorporeal Life Support Organization (ELSO) Center ID: Center name: Run No (for this patient) Unique ID: Birth Date/Time Sex: (M, F) Race: (Asian, Black, Hispanic, White, Other)

More information

Initial Experience With Single Cannulation for Venovenous Extracorporeal Oxygenation in Adults

Initial Experience With Single Cannulation for Venovenous Extracorporeal Oxygenation in Adults Initial Experience With Single Cannulation for Venovenous Extracorporeal Oxygenation in Adults Christian A. Bermudez, MD, Rodolfo V. Rocha, MD, Penny L. Sappington, MD, Yoshiya Toyoda, MD, PhD, Holt N.

More information

Implantable Ventricular Assist Devices and Total Artificial Hearts. Policy Specific Section: June 13, 1997 March 29, 2013

Implantable Ventricular Assist Devices and Total Artificial Hearts. Policy Specific Section: June 13, 1997 March 29, 2013 Medical Policy Implantable Ventricular Assist Devices and Total Artificial Hearts Type: Medical Necessity and Investigational / Experimental Policy Specific Section: Surgery Original Policy Date: Effective

More information

Minimally invasive left ventricular assist device placement

Minimally invasive left ventricular assist device placement Original Article on Cardiac Surgery Minimally invasive left ventricular assist device placement Allen Cheng Department of Cardiovascular and Thoracic Surgery, University of Louisville, Louisville, USA

More information

Use of the Total Artificial Heart in the Failing Fontan Circulation J William Gaynor, M.D.

Use of the Total Artificial Heart in the Failing Fontan Circulation J William Gaynor, M.D. Use of the Total Artificial Heart in the Failing Fontan Circulation J William Gaynor, M.D. Daniel M. Tabas Endowed Chair in Pediatric Cardiothoracic Surgery at The Children s Hospital of Philadelphia The

More information

Omar Sami. Mustafa Khader. Yanal Shafaqouj

Omar Sami. Mustafa Khader. Yanal Shafaqouj 8 Omar Sami Mustafa Khader Yanal Shafaqouj Let us retrieve our discussion about the ventilation-perfusion ratio (V/Q). - When (V/Q) is Zero this means that no ventilation is taking place, V is Zero; bronchial

More information

Use of EKOS Catheter in the management of Venous Mr. Manoj Niverthi, Mr. Sarang Pujari, and Ms. Nupur Dandavate, The GTF Group

Use of EKOS Catheter in the management of Venous Mr. Manoj Niverthi, Mr. Sarang Pujari, and Ms. Nupur Dandavate, The GTF Group Use of EKOS Catheter in the management of Venous Thromboembolism @ Mr. Manoj Niverthi, Mr. Sarang Pujari, and Ms. Nupur Dandavate, The GTF Group Introduction Georgia Thrombosis Forum (GTF, www.gtfonline.net)

More information

Cardiovascular Institute

Cardiovascular Institute Allegheny Health Network Cardiovascular Institute Extracorporeal Membrane Oxygenation (ECMO) Program Our patient survival rate is higher than the national average. ECMO experts. Multidisciplinary team.

More information

Understanding the Cardiopulmonary Bypass Machine and Its Tubing

Understanding the Cardiopulmonary Bypass Machine and Its Tubing Understanding the Cardiopulmonary Bypass Machine and Its Tubing Robert S. Leckie, MD Division of Cardiac Anesthesia, Beth Israel Deaconess Medical Center ABL 1/09 Reservoir Bucket This is a cartoon of

More information

Further Studies on the Effect of Arteriovenous Fistulas and Elevations of Sinus Pressure

Further Studies on the Effect of Arteriovenous Fistulas and Elevations of Sinus Pressure Further Studies on the Effect of Arteriovenous Fistulas and Elevations of Sinus Pressure on Mortality Rates Following Acute Coronary Occlusions By GEORGE SMITH, F.R.C.S., JAMES DEMMING, MORTON ELEFF, AND

More information

ISPUB.COM. Concepts Of Neonatal ECMO. D Thakar, A Sinha, O Wenker HISTORY PATIENT SELECTION AND ECMO CRITERIA

ISPUB.COM. Concepts Of Neonatal ECMO. D Thakar, A Sinha, O Wenker HISTORY PATIENT SELECTION AND ECMO CRITERIA ISPUB.COM The Internet Journal of Emergency and Intensive Care Medicine Volume 5 Number 2 D Thakar, A Sinha, O Wenker Citation D Thakar, A Sinha, O Wenker.. The Internet Journal of Emergency and Intensive

More information

Clinical Performance of Microporous Polypropylene Hollow-Fiber Oxygenator

Clinical Performance of Microporous Polypropylene Hollow-Fiber Oxygenator Clinical Performance of Microporous Polypropylene Hollow-Fiber Oxygenator Kozo Suma, M.D., Takayuki Tsuji, M.D., Yasuo Takeuchi, M.D., Kenji Inoue, M.D., Kenji Shiroma, M.D., Tetsuo Yoshikawa, M.D., and

More information

The management of chronic thromboembolic pulmonary

The management of chronic thromboembolic pulmonary Technique of Pulmonary Thromboendarterectomy Isabelle Opitz, MD, and Marc de Perrot, MD, MSc, FRCSC Toronto Pulmonary Endarterectomy Program, Toronto General Hospital, Ontario, Canada. Address reprint

More information

ECMO CPR. Ravi R. Thiagarajan MBBS, MPH. Cardiac Intensive Care Unit

ECMO CPR. Ravi R. Thiagarajan MBBS, MPH. Cardiac Intensive Care Unit ECMO CPR Ravi R. Thiagarajan MBBS, MPH Staff Intensivist Cardiac Intensive Care Unit Children s Hospital Boston PCICS 2008, Miami, FL No disclosures Disclosures Outline Outcomes for Pediatric in-hospital

More information

Artificial Lungs: A New Inspiration

Artificial Lungs: A New Inspiration Artificial Lungs: A New Inspiration Joseph B. Zwischenberger MD Johnston-Wright Professor and Chairman: Department of Surgery j.zwische@uky.edu The University of Kentucky Lexington, Kentucky Presenter

More information

WELCOME. Welcome to the Children s Hospital PICU (Pediatric Intensive Care Unit). We consider it a privilege to care for your child and your family.

WELCOME. Welcome to the Children s Hospital PICU (Pediatric Intensive Care Unit). We consider it a privilege to care for your child and your family. ECMO FAMILY GUIDE WELCOME Welcome to the Children s Hospital PICU (Pediatric Intensive Care Unit). We consider it a privilege to care for your child and your family. This book was created to give you the

More information

Exam 3 Study Guide. 4) The process whereby the binding of antibodies to antigens causes RBCs to clump is called:

Exam 3 Study Guide. 4) The process whereby the binding of antibodies to antigens causes RBCs to clump is called: Exam 3 Study Guide 1) Where does hematopoiesis produce new red blood cells: 2) Which of the following is a blood clotting disorder: 3) Treatment of hemophilia often involves: 4) The process whereby the

More information

Mechanical Cardiac Support in Acute Heart Failure. Michael Felker, MD, MHS Associate Professor of Medicine Director of Heart Failure Research

Mechanical Cardiac Support in Acute Heart Failure. Michael Felker, MD, MHS Associate Professor of Medicine Director of Heart Failure Research Mechanical Cardiac Support in Acute Heart Failure Michael Felker, MD, MHS Associate Professor of Medicine Director of Heart Failure Research Disclosures Research Support and/or Consulting NHLBI Amgen Cytokinetics

More information

Extracorporeal Life Support Organization (ELSO) General Guidelines for all ECLS Cases

Extracorporeal Life Support Organization (ELSO) General Guidelines for all ECLS Cases Extracorporeal Life Support Organization (ELSO) General Guidelines for all ECLS Cases Introduction This guideline describes prolonged extracorporeal life support (ECLS, ECMO). Related guidelines with more

More information

Admission of patient CVICU and hemodynamic monitoring

Admission of patient CVICU and hemodynamic monitoring Admission of patient CVICU and hemodynamic monitoring Prepared by: Rami AL-Khatib King Fahad Medical City Pi Prince Salman Heart tcentre CVICU-RN Admission patient to CVICU Introduction All the patients

More information

Lung transplantation is the only treatment for most patients with endstage

Lung transplantation is the only treatment for most patients with endstage An artificial lung reduces pulmonary impedance and improves right ventricular efficiency in pulmonary hypertension Jonathan W. Haft, MD Patrick Montoya, PhD Osamma Alnajjar, MD Steven R. Posner, MD Joseph

More information

Airway Management in a Patient with Klippel-Feil Syndrome Using Extracorporeal Membrane Oxygenator

Airway Management in a Patient with Klippel-Feil Syndrome Using Extracorporeal Membrane Oxygenator Airway Management in a Patient with Klippel-Feil Syndrome Using Extracorporeal Membrane Oxygenator Beckerman Z*, Cohen O, Adler Z, Segal D, Mishali D and Bolotin G Department of Cardiac Surgery, Rambam

More information

Mary Lou Garey MSN EMT-P MedFlight of Ohio

Mary Lou Garey MSN EMT-P MedFlight of Ohio Mary Lou Garey MSN EMT-P MedFlight of Ohio Function Prolonged and frequent access to venous circulation Allows for patient to carry on normal life; decrease number of needle sticks Medications, parenteral

More information

Extracorporeal Life Support Organization (ELSO) General Guidelines for all ECLS Cases August, 2017

Extracorporeal Life Support Organization (ELSO) General Guidelines for all ECLS Cases August, 2017 Extracorporeal Life Support Organization (ELSO) Introduction General Guidelines for all ECLS Cases August, 2017 This guideline describes prolonged extracorporeal life support (ECLS, ECMO), applicable to

More information

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations Edwards FloTrac Sensor & Edwards Vigileo Monitor FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations 1 Topics System Configuration FloTrac Sensor and PreSep Catheter Thoracotomy

More information

Extracorporeal membrane oxygenators (ECMO) provide

Extracorporeal membrane oxygenators (ECMO) provide Case Report Interhospital Transport of the ECMO Patients in Bangkok Hospital Abstract An extracorporeal membrane oxygenator (ECMO) is used to support the heart and lungs in patients with severe cardiogenic

More information

ECMO BASICS CHLOE STEINSHOUER, MD PULMONARY AND SLEEP CONSULTANTS OF KANSAS

ECMO BASICS CHLOE STEINSHOUER, MD PULMONARY AND SLEEP CONSULTANTS OF KANSAS ECMO BASICS CHLOE STEINSHOUER, MD PULMONARY AND SLEEP CONSULTANTS OF KANSAS DISCLOSURES No financial disclosures or conflicts of interest OBJECTIVES Define ECMO/ECLS and be able to identify the main types

More information

Prolonged Extracorporeal Membrane Oxygenation Support for Acute Respiratory Distress Syndrome

Prolonged Extracorporeal Membrane Oxygenation Support for Acute Respiratory Distress Syndrome CASE REPORT Prolonged Extracorporeal Membrane Oxygenation Support for Acute Respiratory Distress Syndrome Wen-Je Ko,* Hsao-Hsun Hsu, Pi-Ru Tsai When all conventional treatments for respiratory failure

More information

Heart transplantation is the gold standard treatment for

Heart transplantation is the gold standard treatment for Organ Care System for Heart Procurement and Strategies to Reduce Primary Graft Failure After Heart Transplant Masaki Tsukashita, MD, PhD, and Yoshifumi Naka, MD, PhD Primary graft failure is a rare, but

More information

Intra-operative Echocardiography: When to Go Back on Pump

Intra-operative Echocardiography: When to Go Back on Pump Intra-operative Echocardiography: When to Go Back on Pump GREGORIO G. ROGELIO, MD., F.P.C.C. OUTLINE A. Indications for Intraoperative Echocardiography B. Role of Intraoperative Echocardiography C. Criteria

More information

CHANGING THE WAY HEART FAILURE IS TREATED. VAD Therapy

CHANGING THE WAY HEART FAILURE IS TREATED. VAD Therapy CHANGING THE WAY HEART FAILURE IS TREATED VAD Therapy VAD THERAPY IS BECOMING AN ESSENTIAL PART OF HEART FAILURE PROGRAMS AROUND THE WORLD. Patients with advanced heart failure experience an impaired quality

More information

Index. Note: Page numbers of article titles are in boldface type

Index. Note: Page numbers of article titles are in boldface type Index Note: Page numbers of article titles are in boldface type A Acute coronary syndrome, perioperative oxygen in, 599 600 Acute lung injury (ALI). See Lung injury and Acute respiratory distress syndrome.

More information

ECMO & Renal Failure Epidemeology Renal failure & effect on out come

ECMO & Renal Failure Epidemeology Renal failure & effect on out come ECMO Induced Renal Issues Transient renal dysfunction Improvement in renal function ECMO & Renal Failure Epidemeology Renal failure & effect on out come With or Without RRT Renal replacement Therapy Utilizes

More information

A Study of Prior Cases

A Study of Prior Cases A Study of Prior Cases Clinical theme Sub theme Clinical situation/problem Clinical approach Outcome/Lesson Searchable Key word(s) 1 Cannulation Cannulae insertion The surgeon was trying to cannulate for

More information

Bridging With Percutaneous Devices: Tandem Heart and Impella

Bridging With Percutaneous Devices: Tandem Heart and Impella Bridging With Percutaneous Devices: Tandem Heart and Impella DAVID A. BARAN, MD, FACC, FSCAI SYSTEM DIRECTOR, ADVANCED HEART FAILURE, TX AND MCS SENTARA HEART HOSPITAL NORFOLK, VA PROFESSOR OF MEDICINE

More information

CORONARY ARTERY BYPASS GRAFTING (CABG) (Part 1) Mark Shikhman, MD, Ph.D., CSA Andrea Scott, CST

CORONARY ARTERY BYPASS GRAFTING (CABG) (Part 1) Mark Shikhman, MD, Ph.D., CSA Andrea Scott, CST CORONARY ARTERY BYPASS GRAFTING (CABG) (Part 1) Mark Shikhman, MD, Ph.D., CSA Andrea Scott, CST I have constructed this lecture based on publications by leading cardiothoracic American surgeons: Timothy

More information

Bellwork Define: hemostasis anticoagulation hemophilia (Then write the underline portion of the two state standards in your notes).

Bellwork Define: hemostasis anticoagulation hemophilia (Then write the underline portion of the two state standards in your notes). Bellwork Define: hemostasis anticoagulation hemophilia (Then write the underline portion of the two state standards in your notes). A&P Standards 31) Identify the liquid and cellular components of blood

More information

Blood and Heart. Student Learning Objectives:

Blood and Heart. Student Learning Objectives: Blood and Heart Student Learning Objectives: Identify the major components of the blood. Identify the primary structures associated with the heart Follow the blood through the path of the circulation.

More information

ACUTE RESPIRATORY DISTRESS SYNDROME (ARDS) Rv

ACUTE RESPIRATORY DISTRESS SYNDROME (ARDS) Rv ACUTE RESPIRATORY DISTRESS SYNDROME (ARDS) Rv.8.18.18 ACUTE RESPIRATORY DISTRESS SYNDROME (ARDS) SUDDEN PROGRESSIVE FORM OF ACUTE RESPIRATORY FAILURE ALVEOLAR CAPILLARY MEMBRANE BECOMES DAMAGED AND MORE

More information

12/1/2009. Chapter 19: Hemorrhage. Hemorrhage and Shock Occurs when there is a disruption or leak in the vascular system Internal hemorrhage

12/1/2009. Chapter 19: Hemorrhage. Hemorrhage and Shock Occurs when there is a disruption or leak in the vascular system Internal hemorrhage Chapter 19: Hemorrhage Hemorrhage and Shock Occurs when there is a disruption or leak in the vascular system External hemorrhage Internal hemorrhage Associated with higher morbidity and mortality than

More information

Evolution of ECLS. 04/22/2016 Updated. AllinaHealthSystem. Minneapolis ECMO Early history. ELSO Member Centers

Evolution of ECLS. 04/22/2016 Updated. AllinaHealthSystem. Minneapolis ECMO Early history. ELSO Member Centers Evolution of ECLS Minneapolis 2016 Boston Children s Hospital Surgical Staff 1965 ECMO Early history Lab, 4 days ECC First cardiac case First newborn case(esperanza) 1965 1971 1972 1975 1980 First sucessful

More information

ECMO as a Bridge to Heart Transplant in the Era of LVAD s.

ECMO as a Bridge to Heart Transplant in the Era of LVAD s. Christian Bermudez MD. Associate Professor Director Thoracic Transplantation Division Cardiac Surgery Department of Surgery University of Pennsylvania ECMO as a Bridge to Heart Transplant in the Era of

More information

Chapter 13 Worksheet Code It

Chapter 13 Worksheet Code It Class: Date: Chapter 13 Worksheet 3 2 1 Code It True/False Indicate whether the statement is true or false. 1. A cardiac catheterization diverts blood from the heart to the aorta. 2. Selective vascular

More information

Adult Extracorporeal Life Support (ECLS)

Adult Extracorporeal Life Support (ECLS) Adult Extracorporeal Life Support (ECLS) Steven Scott, M.D., F.A.C.S. Piedmont Heart Institute Cardiothoracic Surgery Disclosures None ECMO = ECLS A technique of life support that involves a continuous

More information

Saphenous Vein Autograft Replacement

Saphenous Vein Autograft Replacement Saphenous Vein Autograft Replacement of Severe Segmental Coronary Artery Occlusion Operative Technique Rene G. Favaloro, M.D. D irect operation on the coronary artery has been performed in 180 patients

More information

ECMO Extracorporeal Membrane Oxygenation

ECMO Extracorporeal Membrane Oxygenation ECMO Extracorporeal Membrane Oxygenation patienteducation.osumc.edu ECMO Table of Contents ECMO: Extracorporeal Membrane Oxygenation... 3 ECMO Treatment... 5 Care Team... 7 Discontinuing ECMO... 8 Notes,

More information

Adult Extracorporeal Life Support (ECLS) Family Guide

Adult Extracorporeal Life Support (ECLS) Family Guide Adult Extracorporeal Life Support (ECLS) Family Guide Notice: This information is considered for general use only. It should not be used instead of medical advice and care. If you have a specific health

More information

Jennifer A. Brown The Cleveland Clinic School of Perfusion Cleveland, Ohio

Jennifer A. Brown The Cleveland Clinic School of Perfusion Cleveland, Ohio Biventricular Heart Failure Advanced Treatment Options at The Cleveland Clinic Jennifer A. Brown The Cleveland Clinic School of Perfusion Cleveland, Ohio I have no disclosures. Examine respiratory and

More information

Images have been removed from the PowerPoint slides in this handout due to copyright restrictions.

Images have been removed from the PowerPoint slides in this handout due to copyright restrictions. Percutaneous Coronary Intervention https://www.youtube.com/watch?v=bssqnhylvma Types of PCI Procedures Balloon Angioplasty Rotational Atherectomy Coronary Stent Balloon Inflation Rotational Atherectomy

More information

Temporary circulatory support in the form of extracorporeal

Temporary circulatory support in the form of extracorporeal Preliminary In Vivo Testing of a Novel Pump for Short-Term Extracorporeal Life Support David J. Spurlock, MD, Kelly Koch, BS, Daniel E. Mazur, BS, Emilia M. Fracz, ME, Robert H. Bartlett, MD, and Jonathan

More information

Please check appropriate box: Right Left Bilateral Not Applicable

Please check appropriate box: Right Left Bilateral Not Applicable Patient Label Here UNIVERSITY MEDICAL CENTER DISCLOSURE AND CONSENT - MEDICAL AND SURGICAL PROCEDURES TO THE PATIENT: You have the right as a patient to be informed about your condition and the recommended

More information

The Role of Mechanical Circulatory Support in Cardiogenic Shock: When to Utilize

The Role of Mechanical Circulatory Support in Cardiogenic Shock: When to Utilize The Role of Mechanical Circulatory Support in Cardiogenic Shock: Presented by Nancy Scroggins ACNP, CNS-CC CV Surgery ACNP Bayshore Medical Center The Role of Mechanical Circulatory Support in Cardiogenic

More information

Pulmonary circulation. Lung Blood supply : lungs have a unique blood supply system :

Pulmonary circulation. Lung Blood supply : lungs have a unique blood supply system : Dr. Ali Naji Pulmonary circulation Lung Blood supply : lungs have a unique blood supply system : 1. Pulmonary circulation 2. Bronchial circulation 1- Pulmonary circulation : receives the whole cardiac

More information

ECMO for cardiac arrest patients: Update 2017

ECMO for cardiac arrest patients: Update 2017 ECMO for cardiac arrest patients: Update 2017 Lim Swee Han MBBS (NUS), FRCS Ed (A&E), FRCP (Edin), FAMS Senior Consultant, Department of Emergency Medicine, Singapore General Hospital Adjunct Associate

More information

Unit 11. Objectives. Indications for IV Therapy. Intravenous Access Devices & Common IV Fluids. 3 categories. Maintenance Replacement Restoration

Unit 11. Objectives. Indications for IV Therapy. Intravenous Access Devices & Common IV Fluids. 3 categories. Maintenance Replacement Restoration Unit 11 Fluids, Electrolytes and Acid Base Imbalances Intravenous Access Devices & Common IV Fluids Objectives Review the purpose and types of intravenous (IV) therapy. Recall the nursing care related

More information

Troubleshooting Adult ECMO

Troubleshooting Adult ECMO The Journal of ExtraCorporeal Technology Troubleshooting Adult ECMO David Sidebotham, FANZCA Departments of Anaesthesia and Intensive Care, Auckland City Hospital, Auckland, New Zealand Presented at Perfusion

More information

Mechanical Support in the Failing Fontan-Kreutzer

Mechanical Support in the Failing Fontan-Kreutzer Mechanical Support in the Failing Fontan-Kreutzer Stephanie Fuller MD, MS Thomas L. Spray Endowed Chair in Congenital Heart Surgery Associate Professor, The Perelman School of Medicine at the University

More information

CIE Biology GCSE. 9: Transport in animals. Notes.

CIE Biology GCSE. 9: Transport in animals. Notes. CIE Biology GCSE 9: Transport in animals Notes The circulatory system acts as the main transport system in animals. It is made up of blood vessels such as arteries, veins and capillaries, in which blood

More information

Patient Management Code Blue in the CT Suite

Patient Management Code Blue in the CT Suite Patient Management Code Blue in the CT Suite David Stultz, MD November 28, 2001 Case Presentation A 53-year-old woman experienced acute respiratory distress during an IV contrast enhanced CT scan of the

More information

Circulation.notebook March 07, Maintaining Dynamic Equilibrium. Oct 11 3:27 PM. Circulatory System p

Circulation.notebook March 07, Maintaining Dynamic Equilibrium. Oct 11 3:27 PM. Circulatory System p Unit 3 Maintaining Dynamic Equilibrium Circulatory system The heart Blood Blood Pressure Circulatory Disorders Oct 11 3:27 PM Circulatory System p.304 307 If an organism is small and flat it can transport

More information

Topic 6: Human Physiology

Topic 6: Human Physiology Topic 6: Human Physiology 6.2 The Blood System D.4 The Heart Essential Questions: 6.2 The blood system continuously transports substances to cells and simultaneously collects waste products. D.3 The chemical

More information

Anaesthetic considerations for laparoscopic surgery in canines

Anaesthetic considerations for laparoscopic surgery in canines Vet Times The website for the veterinary profession https://www.vettimes.co.uk Anaesthetic considerations for laparoscopic surgery in canines Author : Chris Miller Categories : Canine, Companion animal,

More information

2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD

2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD 2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD David L Saint M.D. Tallahassee Memorial Hospital Southern Medical Group Division of Cardiothoracic

More information

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS CARDIAC MECHANICAL SUPPORT PROGRAM GUIDELINES CARDIAC MECHANICAL SUPPORT: LVAD BASICS FREQUENT SCENARIOS AND TROUBLESHOOTING Review Date: July 2011

More information

Long Term Support for Respiratory Failure: VV ECMO or Oxygenated RVAD?

Long Term Support for Respiratory Failure: VV ECMO or Oxygenated RVAD? Long Term Support for Respiratory Failure: VV ECMO or Oxygenated RVAD? Mechanical Circulatory Support Symposium 2018 (Houston, TX) Charles Hoopes, MD Professor of Surgery University of Alabama (Birmingham)

More information

Patent Ductus Arteriosus (PDA)

Patent Ductus Arteriosus (PDA) Patent Ductus Arteriosus (PDA) How does the heart work? The heart is the organ responsible for pumping blood to and from all tissues of the body. The heart is divided into right and left sides. The job

More information

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER Page 1 of 5 ASPIRATION CATHETER Carefully read all instructions prior to use, observe all warnings and precautions noted throughout these instructions. Failure to do so may result in complications. STERILE.

More information

What would be the response of the sympathetic system to this patient s decrease in arterial pressure?

What would be the response of the sympathetic system to this patient s decrease in arterial pressure? CASE 51 A 62-year-old man undergoes surgery to correct a herniated disc in his spine. The patient is thought to have an uncomplicated surgery until he complains of extreme abdominal distention and pain

More information

Cardiothoracic Fellow Expectations Division of Cardiac Anesthesia, Beth Israel Deaconess Medical Center

Cardiothoracic Fellow Expectations Division of Cardiac Anesthesia, Beth Israel Deaconess Medical Center The fellowship in Cardiothoracic Anesthesia at the Beth Israel Deaconess Medical Center is intended to provide the foundation for a career as either an academic cardiothoracic anesthesiologist or clinical

More information

Section 6 Intra Aortic Balloon Pump

Section 6 Intra Aortic Balloon Pump Section 6 Intra Aortic Balloon Pump The Intra Aortic Balloon Pump (IABP) The balloon is synthetic and is made for single use only. It is threaded into the aorta, usually via a femoral approach. The balloon

More information

Diastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock

Diastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock Diastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock Nickolas Trubov, B.Sc., C.C.P. and Steven J. Phillips, M.D. Mercy Hospital, Des Moines, Iowa 50314 ABSTRACT An 80 cc

More information

Lesta Whalen, MD Medical Director, Sanford ECMO Pediatric Critical Care

Lesta Whalen, MD Medical Director, Sanford ECMO Pediatric Critical Care Lesta Whalen, MD Medical Director, Sanford ECMO Pediatric Critical Care Disclosures I have no financial disclosures. The use of certain devises for providing long-term cardiopulmonary support is investigational.

More information

Why would epidural analgesia cause these symptoms? How would increasing the blood volume change venous pressure (VP)?

Why would epidural analgesia cause these symptoms? How would increasing the blood volume change venous pressure (VP)? CASE 12 A 25-year-old pregnant woman is in labor at the hospital. She has no medical problems and has had no complications with this pregnancy. She is in the active phase of labor, feeling intense contractions,

More information

Dr. Rami M. Adil Al-Hayali Assistant Professor in Medicine

Dr. Rami M. Adil Al-Hayali Assistant Professor in Medicine Dr. Rami M. Adil Al-Hayali Assistant Professor in Medicine Venous thromboembolism: pulmonary embolism (PE) deep vein thrombosis (DVT) 1% of all patients admitted to hospital 5% of in-hospital mortality

More information

MODERATOR Felix Rapaport, other members of this

MODERATOR Felix Rapaport, other members of this The First Lung Transplant in Man (1963) and the First Heart Transplant in Man (1964) J.D. Hardy MODERATOR Felix Rapaport, other members of this distinguished panel, and members of the audience, I will

More information

Objective 2/9/2012. Blood Gas Analysis In The Univentricular Patient: The Need For A Different Perspective. VENOARTERIAL CO2 GRADIENT

Objective 2/9/2012. Blood Gas Analysis In The Univentricular Patient: The Need For A Different Perspective. VENOARTERIAL CO2 GRADIENT Blood Gas Analysis In The Univentricular Patient: The Need For A Different Perspective. Gary Grist RN CCP Chief Perfusionist The Children s Mercy Hospitals and Clinics Kansas City, Mo. Objective The participant

More information

Facing Coronary Artery Bypass Surgery? Learn about minimally invasive da Vinci Surgery

Facing Coronary Artery Bypass Surgery? Learn about minimally invasive da Vinci Surgery Facing Coronary Artery Bypass Surgery? Learn about minimally invasive da Vinci Surgery The Condition: Coronary Artery Disease Coronary artery disease is a form of heart disease that affects your arteries.

More information