Radiation Dose Optimization in Cardiac CT: A Technical Review. CHENG Wai Kwong. 17 May Contents

Size: px
Start display at page:

Download "Radiation Dose Optimization in Cardiac CT: A Technical Review. CHENG Wai Kwong. 17 May Contents"

Transcription

1 Radiation Dose Optimization in Cardiac CT: A Technical Review CHENG Wai Kwong 17 May 2014 Contents 1. Introduction and background 2. Cardiac CT synchronization techniques Conclusion 1

2 Introduction and background Introduction & background Cardiac CT (mainly CCTA): promising, low-invasive Radiation dose: major concern for CCTA, esp. for young female patients. increase from 16- to 64-slice CT ECG-gated helical acquisitions: dose up 40 msv (Jean- François & Hicham, 2007) Differences in dose due to performance of CT scanners, technical dose-sparing tools. Jean-François, P & Hicham, A. (2007). Strategies for reduction of radiation dose in cardiac multislice CT, European Radiology, 17(8), pp

3 Cardiac CT synchronization techniques Cardiac CT synchronization techniques Prospective ECG-triggering Synchronization occurs when HR and sequential CT scanning in harmony x-ray tube in same position for various heart beats Trigger x-ray beam and acquire data in certain cardiac phase preferably diastolic phase (70 80 % of R-R interval) with min. cardiac motion e.g. snapshot of 100 ms in 70 % R-R interval Slow and stable HR, low radiation dose (Courtesy to S. Edyvean) 3

4 Cardiac CT synchronization techniques Retrospective ECG-gating ECG signals and helical scan data acquired simultaneously. Scan data selected for image reconstruction w.r.t. pre-defined cardiac phase with a certain temporal relation to onset of R-waves. Better temporal resolution with segmented reconstruction Redundancy of data acquisition phase-selective image reconstruction High radiation dose (Courtesy to S. Edyvean) Cardiac CT synchronization techniques Prospectively ECG-triggered high-pitch helical acquisition State-of-art dual-source scanner - a pitch of 3.4 to fill in the gaps of raw data acquired by second X-ray tube Acquire data very quickly in ~ an 1-mile-per hour table translation speed Not prospective ECG-triggering, not retrospective ECG-gating Helical cardiac data acquired by 1 st tube Helical cardiac data acquired by 2 nd tube (Courtesy to Suhny Abbara) 4

5 Strategy 1: ECG-gated ma modulation Retrospective ECG-gating ma adjusted based on predicted cardiac phases to optimal setting (optimal max. and min. ma) acceptable image quality e.g. 350 ma during mid-diastole (60 80% of R-R interval) and 200 ma for the remainder 30 35% radiation dose (Halliburton, Sola & Kuzmiak 2008; Abbara, Arbab-Zadeh & Callister et al. 2009) 350 ma 200 ma (Courtesy to J. Leipsic & B. Heibron) Halliburton, SS, Sola S & Kuzmiak S (2008). Effect of dual-source cardiac computed tomography on patient radiation dose in a clinical setting: Comparison to single-source imaging. Journal of Cardiovascular Computed Tomography, vol. 2, pp Abbara, CS, Arbab-Zadeh, A & Callister, TQ et al. (2009). SCCT Guidelines for the performance and acquisition of coronary computed tomographic angiography. Journal Cardiovascular Computed Tomography; doi: /j.jcct

6 Strategy 2: Optimization of kvp Radiation dose with CT ~ α ( kvp) 2, 120 to 100 kv 31% reduction in dose (Bischoff, Hein & Meyer et al. 2009). Intravascular contrast with tube voltage (100 kvp or 80 kvp) because iodine resorption α 1/ kvp due to higher degree of P.E. effect and lower degree of Compton scattering (Gutstein, Dey & Cheng et al 2008). kvp attenuation of vessel lumen and cardiac chambers if iodinated C.M. used kvp protocols in myocardial CT perfusion, CCTA (< 70 kg) Bischoff, B, Hein, F, Meyer, T, Hadamitzky, M, Martinoff, S, Schomig, A, Hausleiter, J (2009). Impact of a reduced tube voltage on CT angiography and radiation dose: results of the PROTECTION I study. JACC Cardiovascular Imaging, Vol. 2, pp Gutstein, A, Dey, D, Cheng, V, Wolak, A, Gransar, H, Suzuki, Y, Friedman, J, Thomson, LE, Hayes, S, Pimentel, R, Paz, W, Slomka, P, Le Meunier, L, Germano, G & Berman, DS (2008). Algorithm for radiation dose reduction with helical dual source coronary computed tomography angiography in clinical practice. Journal of Cardiovascular Computed Tomography, vol. 2, pp Strategy 2: Optimization of kvp Pflederer, Rudofsky & Ropers et al (2009) 100 kvp for patients weighing <= 90 kg or with a BMI <= 30 kg/m kvp for patients weighing > 90 kg or with a BMI > 30 kg/m 2 Higher kvp (e.g. 135 kvp) for severely obese patients (> 100 kg). Pflederer, T, Rudofsky, L, Ropers, D, Bachmann, S, Marwan, M, Daniel, WG & Achenbach, S (2009). Image quality in a low radiation exposure protocol for retrospectively ECG-gated coronary CT angiography. American Journal of Roentgenology, vol. 192, pp

7 Strategy 2: Optimization of kvp Hoe (2013) < 100 kg kvp (Siemens DSCT FLASH) < 85 kg kvp (Siemens DSCT) < 25 kg/m 2 BMI 100 kvp (GE 64-slice MDCT) < 30 kg/m 2 BMI 100 kvp (GE 750 HD) < 80 kg 100 kvp, 400 ma > 80 kg 120 kvp, 800 ma (GE 64-slice MDCT) < 60 kg 80 kvp dose by 80 % What body weight or BMI for lower kvp CCTA scan? Remark: (i) < 70 kg 100 kvp, 350 ma, 0.35 s (ii) 120 kvp for post-stenting CCTA (Toshiba Aquilion 64) Hoe, J. (2013). Low dose coronary CTA: What you need to know about medical radiation and how to reduce it, presented in the 9 th CT Coronary Angiography Teaching Course Intermediate-Advanced Level, at HKEC Training Centre for Healthcare Management & Clinical Technology on 30 March Strategy 2: Optimization of kvp Automatic kvp selection AEC selecting tube potential according to diagnostic task and patient attenuation or size kvp constant for same patient, anatomical region and exam. (Courtesy to Katharine Grant & Bernhard Schmidt) 7

8 Strategy 3: Optimal scan length Absorbed dose α DLP = CTDIvol x scan length Optimal scan range: minimum scan length clinically necessary With reference to calcium score scan, CCTA scan range from level just above LM to heart base Margins of 10 mm sup. and inf. to most cranial and caudal slices allowed account for breath-hold inconsistencies scan manually stopped if desired anatomy covered Optimal scan length planning each breath-hold at same inspiration depth Strategy 3: Optimal scan length Retrospectively gated helical acquisitions: a scan length reduction of 1 cm approx. dose saving of 1 msv (Hausleiter, Meyer, Hermann et al, 2009) A scan length reduction of 1 cm dose by 10 % (Hoe, 2013) Hausleiter, J, Meyer, T, Hermann, F. et al. (2009). Estimated radiation dose associated with cardiac CT angiography. The Journal of the American Medical Association, 301(5), pp Hoe, J. (2013). Low dose coronary CTA: What you need to know about medical radiation and how to reduce it, presented in the 9 th CT Coronary Angiography Teaching Course Intermediate-Advanced Level, at HKEC Training Centre for Healthcare Management & Clinical Technology on 30 March

9 Strategy 4: Reconstruction of optimal slice thickness Noise α 1 / (reconstructed slice thickness) radiation dose (through a reduction in kvp or ma) with in slice thickness while maintaining same image noise. 3-mm thick image has 73% less noise than 1-mm thick image (Halliburton, Abbara, Chen, Gentry, Mahesh, Raff, Shaw & Hausleiter, 2011). Images reconstructed with greatest possible slice thickness for given cardiovascular CT indications (e.g. 0.5 mm slices for evaluating coronary artery stenosis, 3.0 mm for calcium scoring or evaluating larger blood vessels such as aorta and PA). Halliburton, SS, Abbara, S, Chen, MY, Gentry, R, Mahesh, M, Raff, GL, Shaw, LJ & Hausleiter, J (2011). SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT. Journal of Cardiovascular Computed Tomography, vol. 5, no. 4, pp Strategy 5: Helical scanning with higher pitch Latest generation of dual-source CT technology: ECG-triggered helical scanning at very high pitch values (High helical pitch flash acquisition) Interleaving data measured from 2 detector systems separated by ~ 90 degrees, pitch 3.4 Near complete elimination of overlap in cone beam amount of redundant data collected (Courtesy to J. Leipsic & B. Heibron) (Courtesy to D.W. Entrikin, J.A. Leipsic & J.J. Carr) (Courtesy to U.J. Schoepf) 9

10 Strategy 6: Filters and breast shields (Rajiah, Halliburton & Flamm, 2012) Filters placed beneath the x-ray tube selectively attenuate low-energy x-rays not contributing to image formation Flat and cardiac bow-tie filters used with small SFOV radiation exposure to lungs (Courtesy to S. Edyvean) Rajiah, P, Halliburton, SS & Flamm, SD (2012). Strategies for Dose Reduction in Cardiovascular Computed Tomography, Applied Radiology, vol. 41, no. 7-8, pp Strategy 6: Filters and breast shields (Rajiah, Halliburton & Flamm, 2012) Bismuth-impregnated latex shield placed between x-ray beam and breasts (i.e. in AP projection) specifically radiation dose by over 25 % (Courtesy to M. Gunn) degradation of image quality by increased pixel attenuation values Rajiah, P, Halliburton, SS & Flamm, SD (2012). Strategies for Dose Reduction in Cardiovascular Computed Tomography, Applied Radiology, vol. 41, no. 7-8, pp

11 Strategy 7: Organ-based dose modulation ma modulation avoiding direct X-ray in radiosensitive organs (AAPM Workgroup) (Courtesy to Siemens) Strategy 7: Organ-based dose modulation ma over radiosensitive organs on periphery dose to other organs (Dual et al, 2011) 11

12 Strategy 8: Adaptive collimation Over-ranging in MDCT generally with collimation unnoticed exposure to breasts Adaptive collimation elimination of radiation exposure unnecessary for image reconstruction Helical scanning (Courtesy to Toshiba Medical Systems) (Courtesy to P.D. Deak) Strategy 9: Iterative reconstruction Filtered back-projection (FBP) fast, images prone to high noise in low dose situations clinical situations in which image quality tends to be less acceptable include: (1) CT scans in large patients, (2) images with small voxels, such as the targeted FOV used for cardiac CT, (3) images through the bony pelvis. 12

13 Strategy 9: Iterative reconstruction Iterative reconstruction algorithm assumes initial μx,y,z for all voxels to predict projection data. predicted projection data compared with actual, measured projection data voxel attenuations modified and compared with measured data until the error between estimated and measured projection data acceptable. significant computational power and additional time improves noise properties (Leipsic, Labounty & Heilbron 2010) (Courtesy to Brian Hutton) (Leipsic, J, Labounty, TM, Min, JK, et al (2010). Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. American Journal Roentgenology, 195, ) Strategy 9: Iterative reconstruction of images IR performed from raw (projection) data or image (slice) data or both (i) IRIS (Iterative Reconstruction in Image Space) (Siemens) IR from image data only (ii) Veo (Model-based Iterative Reconstruction)(GE) IR from raw data only (iii) ASIR (Adaptive Statistical Iterative Reconstruction) (GE) SAFIRE (Sinogram Affirmed Iterative Reconstruction) (Siemens) idose 4 (Philips) AIDR (Adaptive Iterative Dose Reduction) (Toshiba) IR from both raw data and image data (Courtesy to Toshiba Medical) 13

14 Strategy 10: Garnet gemstone-based detector Garnet gemstone-based scintillator (CT750 HD) configured to emit fluorescence when irradiated with x-rays decay time of 30 ns, 100 times faster than conventional scintillators (e.g. GOS, Bismuth Germinate (Bi4Ge3012) and Cadmium Tungstate (CdWO4)) afterglow level only 25 % of those of conventional scintillator materials spatial resolution: 18.2 lp / cm high-definition imaging at up to 230-μm resolution and delineation of stents enhances S/N improves low contrast resolution optimizes ma and radiation dose without significant loss of image quality (Yanagawa, Tomiyama & Honda et al, 2010) Yanagawa, M, Tomiyama, N & Honda, O et al. (2010). Multidetector CT of the Lung: Image Quality with Garnet-based Detectors. Radiology, 255(3), pp Strategy 11: Patient preparation Proper patient preparation (i) Careful explanation of procedure warm sensation to pelvis during CM injection breathing instructions (esp. for 16- or 64- MDCT) (ii) Administration of oral (50 mg + 50 mg) / IV (5 mg + 5 mg) beta-blockers (iii) TNG (1 puff SL just prior to CCTA scan) High HR, arrhythmia and breathing artefacts degrade image quality Proper HR control prospectively ECG-triggered axial sequential scanning if HR < 60 & stable. 14

15 Strategy 11: Patient preparation Breath hold practice Patients instructed to take comfortable breath in about 75 % of full inspiration Scan performed in end inspiration not Valsalva Place hand under costal margin to monitor if patient really holding breath Count 1 sec, 2 sec, 3 sec, 4 sec, etc. for each second Strategy 12: Monitoring and recording dose information Two primary purposes: 1. documenting individual dose burden 2. quality control Recording dose descriptors in: 1. A DICOM image with radiation information in a PACS 2. A paper-based logbook 3. epr or RIS 4. A dedicated database or local registry periodic review of radiation doses local data compared with international standards and other references (e.g. Diagnostic Reference Levels) 15

16 Conclusion Conclusion 1. Cardiac imaging - demanding application of MDCT - improving with technological advances - better temporal resolution 2. Prospective ECG-triggering, retrospective ECG-gating & scan parameters optimize cardiac imaging protocols & radiation dose 3. Recommended radiation dose optimization strategies applied clinical indications characteristics of patient hardware and software constraints on CT scanner lowest possible dose with diagnostic image quality 4. Radiation dose recording & monitoring dose awareness of radiographers 16

17 Thank you 17

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could

More information

Cardiac CT Lowering the Dose Dramatically

Cardiac CT Lowering the Dose Dramatically Cardiac CT Lowering the Dose Dramatically U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant for / research

More information

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Cardiac CT - Coronary Calcium Basics Workshop II (Basic) Cardiac CT - Coronary Calcium Basics Workshop II (Basic) J. Jeffrey Carr, MD, MSCE Dept. of Radiology & Public Health Sciences Wake Forest University School of Medicine Winston-Salem, NC USA No significant

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Introduction. Original Article

Introduction. Original Article Original Article Superior CT coronary angiography image quality at lower radiation exposure with second generation 320-detector row CT in patients with elevated heart rate: a comparison with first generation

More information

Radiation Dose Reduction Strategies in Coronary CT Angiography

Radiation Dose Reduction Strategies in Coronary CT Angiography Radiation Dose Reduction Strategies in Coronary CT Angiography Noor Diyana Osman, PhD noordiyana@usm.my Contents: Introduction Radiation dosimetry in CT Radiation risk associated with coronary CT angiography

More information

Comparison of 64 slice multi detector CT vs 128 slice multi detector CT for acute chest pain evaluation in the chest pain unit

Comparison of 64 slice multi detector CT vs 128 slice multi detector CT for acute chest pain evaluation in the chest pain unit Comparison of 64 slice multi detector CT vs 128 slice multi detector CT for acute chest pain evaluation in the chest pain unit Poster No.: C-0642 Congress: ECR 2010 Type: Topic: Authors: Keywords: Keywords:

More information

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design 2 Cardiovascular Computed Tomography: Current and Future Scanning System Design Wm. Guy Weigold Introduction The heart can be visualized in gross form on any standard chest computed tomography (CT), but

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

Technology Assessment Institute: Summit on CT Dose Cardiac CT - Optimal Use of Evolving Scanner Technologies

Technology Assessment Institute: Summit on CT Dose Cardiac CT - Optimal Use of Evolving Scanner Technologies Cardiac CT - Optimal Use of Evolving Scanner Technologies P. Rogalla, M.D. Dept. of Medical Imaging University of Toronto Special thanks to Dr. Lembcke, Dr. Hein Charité, Berlin Disclosures No salaries

More information

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School OUTLINE 1 st Presentation: Dose optimization strategies

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

A Snapshot on Nuclear Cardiac Imaging

A Snapshot on Nuclear Cardiac Imaging Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the

More information

Routine low-radiation-dose coronary computed tomography angiography

Routine low-radiation-dose coronary computed tomography angiography European Heart Journal Supplements (2014) 16 (Supplement B), B12 B16 The Heart of the Matter doi:10.1093/eurheartj/suu024 Routine low-radiation-dose coronary computed tomography angiography Mouaz H. Al-Mallah

More information

WHAT WE CAN LEARN FROM CURRENT CARDIAC CT TECHNOLOGY AND WHAT S NEXT FOR CARDIAC?

WHAT WE CAN LEARN FROM CURRENT CARDIAC CT TECHNOLOGY AND WHAT S NEXT FOR CARDIAC? WHAT WE CAN LEARN FROM CURRENT CARDIAC CT TECHNOLOGY AND WHAT S NEXT FOR CARDIAC? Gilbert Raff, MD Director of Advanced Cardiovascular Imaging Oakland University William Beaumont Medical School DISCLOSURES

More information

A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering

A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering DOI 10.1007/s10554-015-0802-z ORIGINAL PAPER A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering J. D. van Dijk 1,4 E. D. Huizing 1,4 P.

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Disclosure Information

Disclosure Information Coronary CTA Pearls and Pitfalls Ricardo C. Cury, MD, FSCCT, FAHA, FACC Chairman of Radiology Radiology Associates of South Florida Director of Cardiac Imaging Miami Cardiac and Vascular Institute Past-President

More information

An Introduction to Dual Energy Computed Tomography

An Introduction to Dual Energy Computed Tomography An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the

More information

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation Poster No.: C-2014 Congress: ECR 2014 Type: Educational Exhibit Authors: M. A. Ottesen; Oslo/NO Keywords: Cardiac, Arteries / Aorta, CT, CT-Angiography,

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research

More information

Purpose. Methods and Materials

Purpose. Methods and Materials Comparison of iterative and filtered back-projection image reconstruction techniques: evaluation of heavily calcified vessels with coronary CT angiography Poster No.: C-1644 Congress: ECR 2011 Type: Scientific

More information

Les Outils Cliniques de Demain en Scanner Cardiaque. Cardiaque Status en ECR 2018 From Diagnosis to Prognosis

Les Outils Cliniques de Demain en Scanner Cardiaque. Cardiaque Status en ECR 2018 From Diagnosis to Prognosis ECR 2018 From Diagnosis to Prognosis ECR 2018 From Diagnosis to Prognosis Thursday, March 1, 2018/08:30-10:00/Room N Les Outils Cliniques de Demain en Scanner Cardiaque Cardiaque Status en 2018 Rodrigo

More information

Ultralow Dose Chest CT with MBIR

Ultralow Dose Chest CT with MBIR Ultralow Dose Chest CT with MBIR Ella A. Kazerooni, M.D. Professor & Director Cardiothoracic Radiology Associate Chair for Clinical Affairs University of Michigan Disclosures Consultant: GE Healthcare

More information

Since the introduction of multidetector. Original Research

Since the introduction of multidetector. Original Research Original Research Hellenic J Cardiol 2014; 55: 184-190 Radiation Dose and Image Noise Evaluation in Coronary Computed Tomography Angiography (CCTA) Using an Iterative Reconstruction Algorithm Sofia D.

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

Computed Tomography of the Coronary Arteries

Computed Tomography of the Coronary Arteries Cardiology Update DAVOS 2011 Computed Tomography of the Coronary Arteries Anders Persson M.D., Ph.D Director, Assoc. Professor Center for Medical Image Science and Visualization Linköping University SWEDEN

More information

Cardiac CTA Prospective Gating Broad Beam

Cardiac CTA Prospective Gating Broad Beam Cardiac CTA Prospective Gating Broad Beam ACQUISITION- Broad Beam Gating: Prospective Non Contrast Scan-Calcium Score Patient Position Supine Feet First into Gantry Heart Isocenter Scanogram AP and Lateral

More information

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre Pushing the limits of cardiac CT Steven Dymarkowski Radiology / Medical Imaging Research Centre 5 X 2013 Introduction Rapid technological advances and new clinical applications in cardiovascular imaging

More information

Cardiopulmonary Imaging Original Research

Cardiopulmonary Imaging Original Research Cardiopulmonary Imaging Original Research Bischoff et al. Versus Helical Mode for Coronary CTA Cardiopulmonary Imaging Original Research Bernhard Bischoff 1 Franziska Hein 1 Tanja Meyer 1 Markus Krebs

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Dual Energy CT of the Heart: Perfusion and Beyond

Dual Energy CT of the Heart: Perfusion and Beyond Dual Energy CT of the Heart: Perfusion and Beyond U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant

More information

Gender differences in CT calcium scoring: A phantom study

Gender differences in CT calcium scoring: A phantom study Gender differences in CT calcium scoring: A phantom study Nicholas Petrick, Qin Li, Benjamin Berman, Marios A Gavrielides, Rongping Zeng, Berkman Sahiner CDRH/OSEL/DIDSR U.S. Food and Drug Administration

More information

MEDICAL UNIVERSITY of SOUTH CAROLINA

MEDICAL UNIVERSITY of SOUTH CAROLINA U. Joseph Schoepf, MD Prof. (h.c.), FAHA, FSCBT-MR, FNASCI, FSCCT Professor of Radiology, Medicine, and Pediatrics Director, Division of Cardiovascular Imaging MEDICAL UNIVERSITY of SOUTH CAROLINA Disclosures

More information

The radiation dose in retrospective

The radiation dose in retrospective The radiation dose in retrospective gated tdcoronary computed td tomography (CCT) Saeed AL Ahmari, Ghormallah AL Zahrani, Sumiah AL Helali, Samir AL Dulikan, Abdullah Bafagih, HibaKhashojji Prince Sultan

More information

SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT

SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT Journal of Cardiovascular Computed Tomography (2011) 5, 198 224 Guidelines SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT Sandra S. Halliburton, PhD a, *, Suhny

More information

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population Journal of the American College of Cardiology Vol. 49, No. 9, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.10.066

More information

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3 Coronary Artery Bypass Graft Imaging with CT Angiography and Iterative Reconstruction: Quantitave Evaluation of Radiation Dose Reduction and Image Quality Simon Nepveu 1, Irina Boldeanu 1, Yves Provost

More information

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Junfu Liang 1,2, Hui Wang 1, Lei Xu 1, Li Dong 1, Zhanming

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Cardiac CT Techniques in Neonates (and infants)

Cardiac CT Techniques in Neonates (and infants) Cardiac CT Techniques in Neonates (and infants) Siddharth P. Jadhav, MD Director, Body CT and MRI Edward B. Singleton Department of Pediatric Radiology Texas Children s Hospital Disclosures None Objectives

More information

Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography

Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography Review Article Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography Zhonghua Sun Discipline of Medical Imaging, Department of Imaging and Applied

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

CT Perfusion. U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging

CT Perfusion. U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging CT Perfusion U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant for / research support from Bayer Bracco

More information

Chief Radiographer TEI Clinical Associate 2016

Chief Radiographer TEI Clinical Associate 2016 MDCT Principles i and Applications Ε ΑGADAKOS MSc Ε. ΑGADAKOS MSc Chief Radiographer TEI Clinical Associate 2016 Aim To understand d recent technological advances in MSCT and how they can be effectively

More information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

Validation of CT Perfusion Imaging Against Invasive Angiography and FFR on a 320-MDCT Scanner

Validation of CT Perfusion Imaging Against Invasive Angiography and FFR on a 320-MDCT Scanner Validation of CT Perfusion Imaging Against Invasive Angiography and FFR on a 320-MDCT Scanner Zhen Qian, Gustavo Vasquez, Sarah Rinehart, Parag Joshi, Eric Krivitsky, Anna Kalynych, Dimitri Karmpaliotis,

More information

CT Imaging of Atherosclerotic Plaque. William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA

CT Imaging of Atherosclerotic Plaque. William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA CT Imaging of Atherosclerotic Plaque William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA PREVALENCE OF CARDIOVASCULAR DISEASE In 2006 there were 80 million

More information

Computed tomography in coronary imaging: current status

Computed tomography in coronary imaging: current status 7 Computed tomography in coronary imaging: current status ARJUN NAIR AND ANAND DEVARAJ Recent technological advances have led to improvements in the use of computerised tomography for coronary imaging.

More information

Cardiac Imaging Tests

Cardiac Imaging Tests Cardiac Imaging Tests http://www.medpagetoday.com/upload/2010/11/15/23347.jpg Standard imaging tests include echocardiography, chest x-ray, CT, MRI, and various radionuclide techniques. Standard CT and

More information

Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography

Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography Int J Cardiovasc Imaging (2013) 29:453 461 DOI 10.1007/s10554-012-0096-3 ORIGINAL PAPER Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Patient preparation and coronary CTA techniques. Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging

Patient preparation and coronary CTA techniques. Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging Patient preparation and coronary CTA techniques Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging Overview 1. Patient preparation 2. Scanning techniques Patient preparation Preparation

More information

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Cardiac CT David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Courtesy Tom Kracus Courtesy Kettering Tom Medical Kracus Cente Kettering Medical Center 2003-2006, David Stultz, MD Courtesy Tom Kracus

More information

The Emerging Role of Cardiac CT in Cardiovascular Imaging. Anthony Gemignani, MD Vermont Cardiac Network April 28, 2016

The Emerging Role of Cardiac CT in Cardiovascular Imaging. Anthony Gemignani, MD Vermont Cardiac Network April 28, 2016 The Emerging Role of Cardiac CT in Cardiovascular Imaging Anthony Gemignani, MD Vermont Cardiac Network April 28, 2016 Conflict Disclosures I have no significant financial relationship with any companies

More information

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM 1. The CT Imaging Chain-10 major components & their functions a. The x-ray tube b. Generator c. Filter d. Pre-patient collimator e. Pre-detector collimator f. Detector system g. Analog to digital converter

More information

Computed Tomography Imaging of the Coronary Arteries

Computed Tomography Imaging of the Coronary Arteries Chapter 6 Computed Tomography Imaging of the Coronary Arteries G.J. Pelgrim, M. Oudkerk and R. Vliegenthart Additional information is available at the end of the chapter http://dx.doi.org/10.5772/54044

More information

Usefulness of New CT Technologies for Interventional Cardiovascular Procedures

Usefulness of New CT Technologies for Interventional Cardiovascular Procedures Usefulness of New CT Technologies for Interventional Cardiovascular Procedures Ronen Rubinshtein, MD FACC FESC Department of Cardiovascular Medicine Lady Davis Carmel Medical Center & Technion Israel Institute

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Standardized medical terminology for cardiac computed tomography: A report of the Society of Cardiovascular Computed Tomography

Standardized medical terminology for cardiac computed tomography: A report of the Society of Cardiovascular Computed Tomography Journal of Cardiovascular Computed Tomography (2011) -, - - Guidelines Standardized medical terminology for cardiac computed tomography: A report of the Society of Cardiovascular Computed Tomography Wm.

More information

Low-dose and High-resolution Cardiac Imaging with Revolution CT

Low-dose and High-resolution Cardiac Imaging with Revolution CT GE Healthcare Case study Low-dose and High-resolution Cardiac Imaging with Revolution CT Prof. Philipp A. Kaufmann, M.D. Ronny R. Buechel, M.D. Fran Mikulicic, M.D. Dominik C. Benz, M.D. University of

More information

Dose Reduction Options in Cardiac CT

Dose Reduction Options in Cardiac CT Dose Reduction Options in Cardiac CT Doyle P, Ball P*, Donnelly P # Radiological Sciences & Imaging, Forster Green Hospital *Department of Radiology, Ulster Hospital # Department of Cardiology, Ulster

More information

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose?

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose? Journal of Physics: Conference Series PAPER OPEN ACCESS Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose? To cite this article: M Smarda et

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Lei Zhao 1,2, Fabian Plank 2, Moritz Kummann 2, Philipp Burghard 2, Andrea Klauser 2, Wolfgang Dichtl 3, Gudrun Feuchtner 2.

Lei Zhao 1,2, Fabian Plank 2, Moritz Kummann 2, Philipp Burghard 2, Andrea Klauser 2, Wolfgang Dichtl 3, Gudrun Feuchtner 2. Original Article Improved non-calcified plaque delineation on coronary CT angiography by sonogram-affirmed iterative reconstruction with different filter strength and relationship with BMI Lei Zhao 1,2,

More information

Alessandro Albonico Philips

Alessandro Albonico Philips Alessandro Albonico Philips Alessandro.albonico@philips.com Noise (Standard Deviation in HU) Virtually noise-free Characteristic of a true knowledge-based IR 80 70 Standard Recon idose4 Level6 1 mm Slice

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015 2384 Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients GUIRU HE, XIAOPEI LIU, YAN LIU, WEI

More information

Patient / Organ Dose in CT

Patient / Organ Dose in CT Patient / Organ Dose in CT Patient specific and organ dose estimation H.D. Nagel Dr. HD Nagel, Science & Technology for Radiology Buchholz / Germany www.sascrad.com 1 Topics CTDI & patient dose SSDE Organ

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

Research Article. Open Access. Jin-Da WANG, Hua-Wei ZHANG, Qian XIN, Jun-Jie YANG, Zhi-Jun SUN, Hong-Bin LIU, Lian CHEN, Luo-Shan DU, Yun-Dai CHEN

Research Article. Open Access. Jin-Da WANG, Hua-Wei ZHANG, Qian XIN, Jun-Jie YANG, Zhi-Jun SUN, Hong-Bin LIU, Lian CHEN, Luo-Shan DU, Yun-Dai CHEN Journal of Geriatric Cardiology (2014) 11: 39 43 2014 JGC All rights reserved; www.jgc301.com Research Article Open Access Safety and efficacy of intravenous esmolol before prospective electrocardiogram-triggered

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

State of the art and future development for standardized estimation of organ doses in CT

State of the art and future development for standardized estimation of organ doses in CT State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida

More information

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program CT Low Dose Lung Cancer Screening Part I Journey to LDCT LCS Program Paul Johnson, M.S., DABHP, DABR Cleveland Clinic September 26, 2015 Lung Caner is No. 1 In Cancer Related Death In The United States

More information

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Mathieu Sabbagh, R3 Michigan State University Radiology Garden City Hospital

More information

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination Int J Clin Exp Med 2015;8(11):21898-21905 www.ijcem.com /ISSN:1940-5901/IJCEM0015005 Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out

More information

The Computed Tomography Examination

The Computed Tomography Examination CONTENT SPECIFICATIONS The Computed Tomography Examination The purpose of The American Registry of Radiologic Technologists (ARRT ) Computed Tomography Examination is to assess the knowledge and cognitive

More information

Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax

Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax Poster No.: C-0013 Congress: ECR 2013 Type: Authors: Scientific Exhibit Z. Dejian 1, X. Zhuodong 2, J. Hui 2, L. Xiao 3 ;

More information

Subtraction coronary CT angiography with iterative reconstruction: a feasibility study of coronary calcium subtraction

Subtraction coronary CT angiography with iterative reconstruction: a feasibility study of coronary calcium subtraction Subtraction coronary CT angiography with iterative reconstruction: a feasibility study of coronary calcium subtraction Poster No.: C-1756 Congress: ECR 2013 Type: Scientific Exhibit Authors: R. Tanaka,

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

During the past several years, the need to reduce radiation

During the past several years, the need to reduce radiation Advances in Cardiovascular Imaging Achieving Very-Low-Dose Radiation Exposure in Cardiac Computed Tomography, Single-Photon Emission Computed Tomography, and Positron Emission Tomography Damini Dey, PhD;

More information

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING

ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING ROLE OF MULTISLICE COMPUTED TOMOGRAPHY IN CARDIAC IMAGING Non-invasive coronary angiography along with multidetector computed tomography or magnetic resonance imaging is attracting increasing interest

More information

Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source coronary CT angiography

Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source coronary CT angiography www.nature.com/scientificreports Received: 1 March 2017 Accepted: 27 June 2017 Published online: 31 July 2017 OPEN Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source

More information

Exploration the Method of Low Dose Coronary Artery Imaging with Dual-Source CT

Exploration the Method of Low Dose Coronary Artery Imaging with Dual-Source CT Journal of Biosciences and Medicines, 2013, 1, 6-10 doi:10.4236/jbm.2013.11002 Published Online February 2013 (http://www.scirp.org/journal/jbm/) Exploration the Method of Low Dose Coronary Artery Imaging

More information

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography 633 CARDIOVASCULAR MEDICINE Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography A W Leber, A Knez, C Becker, A Becker, C White, C Thilo, M Reiser,

More information

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial Poster No.: C-0595 Congress: ECR 2013 Type: Authors:

More information

The Post-64 Era of Coronary CT Angiography: Understanding New. Technology from Physical Principles

The Post-64 Era of Coronary CT Angiography: Understanding New. Technology from Physical Principles The Post-64 Era of Coronary CT Angiography: Understanding New Technology from Physical Principles The Harvard community has made this article openly available. Please share how this access benefits you.

More information

CT Dosimetry in the Clinical Environment: Methods and Analysis

CT Dosimetry in the Clinical Environment: Methods and Analysis CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida

More information

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference.

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference. Rotation: Imaging 1 Imaging 1 provides COCATS Level 1 experience for nuclear cardiology (including SPECT and PET) and cardiac CT. Fellows will administer, process, and read cardiac nuclear studies with

More information

Electron Beam CT of the Heart

Electron Beam CT of the Heart CHAPTER 2 / ELECTRON BEAM TOMOGRAPHY 15 2 Electron Beam CT of the Heart DAVID G. HILL, PhD INTRODUCTION Electron beam tomography (EBT)* was developed by Douglas Boyd, PhD, and his associates at Imatron,

More information

Cardiac CTA without and with IV Contrast

Cardiac CTA without and with IV Contrast Cardiac CTA without and with IV Contrast IMG 12122 Cardiac Prospective 12123 Cardiac Retrospective 12119 Cardiac ED Chest Pain 12150 Calcium Score A B A B Toshiba AquilionONE Toshiba AquilionPRIME Calcium

More information

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens DEfinitely Siemens True Dual Energy Dr. Stefan Ulzheimer, Siemens Healthcare GmbH International version. Not for distribution in the US. Unrestricted Siemens AG 2015 All rights reserved. The products/features

More information