Assessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler. and dobutamine stress echocardiography

Size: px
Start display at page:

Download "Assessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler. and dobutamine stress echocardiography"

Transcription

1 Assessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler and Dobutamine Stress Echocardiography Akira Sezai, MD, PhD, Yuji Kasamaki, MD, PhD, Keisuke Abe, RMS, Mitsumasa Hata, MD, PhD, Hisakuni Sekino, MD, PhD, Kazuma Shimura, MD, PhD, and Kazutomo Minami, MD, PhD Departments of Cardiovascular Surgery and Cardiology, Sekino Hospital, Nihon University School of Medicine, Tokyo, Japan ADULT CARDIAC Background. The St. Jude Medical (SJM) Regent prosthetic valve (St. Paul, MN), a recently developed mechanical valve, is an improvement on the conventional SJM valve, having a wider valve area than the SJM HP valve. We evaluated this mechanical valve by Doppler echocardiography and dobutamine stress echocardiography (DSE). Methods. The functions of the SJM Regent valve were evaluated by continuous-wave Doppler echocardiography and DES in 58 cases of aortic valve replacement during a stable postoperative period. Results. The peak pressure gradient of the replaced valves sized 17, 19, 21, and 23 mm was , , , and , mm Hg respectively, and the effective orifice area index was , , , and cm 2 /m 2, respectively, with prosthesis-patient mismatch (PPM) found in 1, 3, 2, and 0 cases for the 17-, 19-, 21- and 23-mm valves, respectively, with a total incidence of 10.3%. In 20 cases, the peak pressure gradient and the effective orifice area index were significantly increased during DSE compared with those at rest. Conclusions. Although the PPM incidence was 6.6%, it was deemed from the data of DSE and clinical symptoms that there were no clinical issues for such cases of PPM in the early and intermediate phases after operation. Particularly, the effectiveness of the 17- and 19-mm valves in patients with a small aortic annulus was demonstrated, confirming the satisfactory functions of the SJM Regent prosthetic valve. (Ann Thorac Surg 2010;89:87 92) 2010 by The Society of Thoracic Surgeons Since the St. Jude Medical (SJM; St. Jude Medical Inc, St. Paul, MN) valve was introduced in 1977 as the first bileaflet mechanical valve, many reports have appeared about its durability, antithrombotic efficacy, and low level of hemolysis. It has been highly evaluated as an excellent mechanical valve based on long-term results and is the most widely used mechanical valve in the world [1 3]. The SJM valve was also used at our hospital in Japan for the first case of valve replacement (July 1978), and good clinical data have been obtained since then [1, 4, 5]. In recent years, various mechanical valves with a wide valve area have been developed for the replacement of a small aortic annulus, with which an adequately wide valve area can be secured without a surgical procedure for aortic annulus enlargement. Patient-prosthesis mismatch (PPM) does occur, however, and development of a prosthetic valve with a wider valve area has been desired. The SJM Regent valve of the present study has a wider valve area compared with conventional mechanical valves and was implanted for the first time in Japan at our facility in October 2004 [6]. The valve was considered Accepted for publication Sept 1, Address correspondence to Dr Sezai, Department of Cardiovascular Surgery, Nihon University School of Medicine, 30-1 Oyaguchi-kamimachi Itabashi-ku Tokyo, , Japan; asezai@med.nihon-u.ac.jp. an effective mechanical valve for Japanese patients, who generally have a physical constitution smaller than that of white patients. Data on the function of the SJM Regent valve after replacement are important for Japanese patients, who often have a small physique and a small aortic annulus, but it is unclear from the results to date how many patients have PPM. The objectives of the present study were to evaluate valve function by Doppler echocardiography, to detect patients with PPM, and to assess valve function under stress by performing dobutamine stress echocardiography (DSE) in patients who had undergone aortic valve replacement with an SJM Regent valve at our hospital. Patients and Methods The Ethics Committee of Nihon University School of Medicine Itabashi Hospital approved this study. Details of the study were explained to each patient, and informed consent was obtained. This study was registered with the University Hospital Medical Information Network (UMIN; Study ID: UMIN ). The study cohort consisted of 58 patients who survived for 6 months or longer among the patients who underwent aortic valve replacement with an SJM Regent valve from October 2004 to August 2008 at the Itabashi Hospital 2010 by The Society of Thoracic Surgeons /10/$36.00 Published by Elsevier Inc doi: /j.athoracsur

2 ADULT CARDIAC 88 SEZAI ET AL Ann Thorac Surg ECHOCARDIOGRAPHIC DATA OF THE SJM REGENT 2010;89:87 92 Table 1. Preoperative Patient Characteristics Variable No. or Mean SD (Range) Patients, No 58 Age, years (51 76) Gender Male 30 Female 28 Body surface area, m Diagnosis Aortic stenosis 34 Aortic regurgitation 9 Aortic stenosis and regurgitation 15 Infective endocarditis 1 Previous operations Closed mitral commissurotomy 2 Mitral valve replacement 1 Ejection fraction New York Heart Association Class II 19 Class III 28 Class IV 11 SD standard deviation. of Nihon University. Among the 60 patients who had aortic valve replacement with the SJM Regent valve, one operative death (1.7%) occurred in a patient with chronic renal failure who went into shock during dialysis and died on postoperative day 7. One patient was discharged with no postoperative complications but died of arrhythmia during the second month postoperatively. An autopsy was performed in both cases, and no abnormalities of the artificial valves were found. The 58 remaining patients, 30 men (52%) and 28 women (48%), were a mean age of years (range, 51 to 76 years), and 18 patients (31.0%) were aged older than 70 years. Their causative diseases included aortic stenosis in 34, aortic regurgitation in 9, and aortic stenosis and regurgitation in 15. Some patients had already undergone mitral valve operations, but no patient had had a previous aortic valve intervention (Table 1). The size of the prosthetic valve used was 17 mm in 4 cases, 19 mm in 16, 21 mm in 20, and 23 mm in 18. The body surface area (BSA) in cm 2 for each valve was for the 17-mm valve, for the 19-mm valve, for the 21-mm valve, and for the 23-mm valve. Concomitant procedures consisted of mitral valve replacement in 11 patients, open mitral valve commissurotomy in 1, tricuspid annuloplasty in 2, coronary artery bypass grafting in 15, and the Maze procedure in 4 (sinus rhythm was restored in all 4 patients). Aortic valve replacement was always performed in the supra-annular position. All 58 patients in were New York Heart Association functional class I postoperatively. At 6 months postoperatively, the left ventricular enddiastolic diameter (LVEDD), left ventricular end-systolic diameter (LVESD), ejection fraction (EF), and percentage of fractional shortening (%FS) were measured by transthoracic echocardiography in the LV long-axis view. The maximum blood flow through the prosthetic valve was recorded by the continuous-wave Doppler method, after which peak pressure gradient (PPG) and mean pressure gradient (MPG) were calculated using the simplified Bernoulli formula. The effective orifice area (EOA) of the valve was determined by the continuous formula and divided by the BSA to calculate the EOA index (EOAI), with 0.65 to 0.85 cm 2 /m 2 defined as mild PPM and less than 0.65 cm 2 /m 2 defined as severe PPM [7, 8]. Dobutamine stress echocardiography was performed at 6 months postoperatively in 20 patients who gave consent. Infusion of dobutamine was started at 10 g/kg/ min after obtaining baseline data on LV function and Table 2. Left Ventricular Function in Before and After Operation Valve Size Variable 17 mm 19 mm 21 mm 23 mm LVEDD, mm Pre-op a a Post-op EF Pre-op a a Post-op LVESD, mm Pre-op a a Post-op %FS Pre-op a Post-op a p EF ejection fraction; LVEDD left ventricular end-diastolic diameter; LVESD left ventricular end-systolic diameter; %FS percentage fractional shortening.

3 Ann Thorac Surg SEZAI ET AL 2010;89:87 92 ECHOCARDIOGRAPHIC DATA OF THE SJM REGENT Table 3. Valve Function in this Study and Data from Previous Reports on Each Size of the St. Jude Medical Regent Valve Valve Size No. PPG a (mm Hg) MPG a (mm Hg) EOAI a (cm 2 /m 2 ) PPM (%) 17 mm Sezai Gelsomino [9] Okamura [10] Takaseya [11] Minardi [12] mm Sezai Gelsomino [9] Bach [13] mm Sezai Gelsomino [9] Bach [13] Guenzinger [14] mm Sezai Gelsomino [9] Bach [13] Guenzinger [14] ADULT CARDIAC a Data are presented as the mean standard deviation. EOAI effective orifice area index; MPG mean pressure gradient; PPG peak pressure gradient; PPM patient-prosthesis mismatch. artificial valve function. The infusion rate was increased to 20, 30, and 40 g/kg/min at 15-minute intervals. Echocardiography was performed with a VIVID 7 model (E Yokokawa Medical Systems Co, Tokyo, Japan) at a frequency of 3.4 MHz by a specialist echocardiographer who was blinded to information about the size and type of the artificial valve. Statistical Analysis Data are presented as the mean standard deviation, and the echocardiography findings were compared by the unpaired t test. Comparison between the PPM and non-ppm groups was performed with the Fisher exact test or the unpaired t test. The level of significance was set at p All analyses were conducted with SPSS software (SPSS Inc, Chicago, IL). Results LV Function Preoperative and postoperative LV functions are summarized in Table 2. Postoperative LVEDD significantly decreased from the preoperative level for patients with the 21-mm valve (p 0.002) and the 23-mm valve (p 0.014). Postoperative LVDs also decreased from the preoperative level for patients with the 21-mm valve (p 0.007) and the 23-mm valve (p ). Postoperative EF significantly increased from the preoperative level in for cases with the 19-mm valve (p 0.026) and the 23-mm valve (p ), and postoperative % FS also increased from the baseline level for cases with the 23-mm valve (p ). Prosthetic Valve Functions Results of continuous-wave Doppler echocardiography for prosthetic valve functions are summarized in Table 3 [9 14]. The PPG values for patients with the 17-, 19-, 21-, and 23-mm valves were (range, 17.0 to 41.0), (range, 8.3 to 48.1), (range, 9.3 to 20.2), and (range, 4.3 to 18.9) mm Hg, respectively. The MPG values for cases with the 17-, 19-, 21- and 23-mm valves were (range, 9 to 19.0), (range, 3.5 to 19.9), (range, 9 to 12.7), and (range, 2.8 to 9.3) mm Hg, respectively. The EOA values for the 17-, 19-, 21-, and 23-mm valves were (range, 0.76 to 1.71), (range, 0.81 to 2.4), (range, 0.9 to 2.35), and (range, 1.42 to 3.00) cm 2, respectively. The EOAI values for patients with the 17-, 19-, 21-, and 23-mm valves were (0.53 to 1.23), (0.50 to 1.59), (0.61 to 1.82), and (0.86 to 2.08) cm 2 /m 2, respectively. Patient-prosthesis mismatch was found in 1 patient (25.0%) with the 17-mm valve (severe), in 3 (18.8%) with the 19-mm valve (2 mild, 1 severe), 2 (10.0%) with the 21-mm valve (both severe) and in no patients with the 23-mm valve, for with a total incidence of 10.3%. Comparison of Groups With and Without PPM An EOAI of 0.85 cm 2 /m 2 or less was defined as PPM, and the patients with and without PPM were compared (Table 4). No differences were found with respect to preoperative factors such as age, gender, BSA, and EF. In the PPM group, however, the size of the replacement

4 ADULT CARDIAC 90 SEZAI ET AL Ann Thorac Surg ECHOCARDIOGRAPHIC DATA OF THE SJM REGENT 2010;89:87 92 Table 4. Clinical Profile and Echocardiographic Data Patients With and Without Prosthesis Mismatch Variable a PPM ( ) PPM ( ) p Value Patients, No Age, years Female gender 3 (50.0) 25 (48.1) 0.99 Body surface area, m Ejection fraction Pre-op Post-op Valve size Peak pressure gradient, mm Hg Mean pressure gradient, mm Hg Effective orifice area, cm 2 EOAI, cm 2 /m a Continuous data are shown as mean standard deviation, categoric data as No. (%). EOAI effective orifice area index; PPM patient-prosthesis mismatch. valve was significantly smaller (p ) and significantly higher values were noted for PPG (p ) and MPG (p ). The EOA was significantly smaller in the PPM group (p ), as was the EOAI (p ). The postoperative EF showed no significant difference between the two groups. Dobutamine Stress Echocardiography Results of DSE are summarized in Table 5. Systolic blood pressure, heart rate, PPG, and MPG significantly increased in all 20 patients with DSE at rest and at doses of 10, 20, 30, and 40 g/kg/min. EF significantly increased at rest and at doses of 30 and 40 g/kg/min, whereas EOA and EOAI significantly increased at rest and at a dose of 40 g/kg/min. Four patients who showed an EF higher than 100 mm Hg by dobutamine stress. DSE was performed on 2 patients with PPM, showing that EF did not exceed 100 mm Hg by dobutamine stress in both cases. Comment In recent years, the HP series of the SJM valve [15], the AP series of the ATS valve [16] (ATS Medical Inc, Minneapolis, MN), and the TopHat series of the CarboMedics valve (Carbomedics, Austin, TX) [17] were developed to secure a wide valve area for replacement of a small aortic annulus without annular enlargement, and these valves have been used clinically. Patients still have PPM, however, so the development of mechanical valves with a wider effective orifice area is awaited. The SJM Regent valve used in this study showed no design changes in terms of the contact area with the blood, but it is an improved mechanical valve with a wider orifice area than the standard series or HP series of SJM valves. Walker and colleagues [18] compared the valve sizes of 17-, 19-, and 21-mm between the SJM Regent and the SJM-HP valves by an in vitro study, and reported that the PPG of the SJM Regent valve decreased by 37%, 36%, and 39%, respectively, and the EOA increased by 22%, 22%, and 21%, respectively, compared with the SJM HP series valve [18]. Although few reports on the clinical use of the SJM Regent valve are available, its efficacy has been reported in Western countries. Albes and colleagues [19] reported that the SJM Regent valve had a lower pressure gradient than the Carbo- Medics standard and the SJM HP valve, and that the 21-mm valve had a lower pressure gradient than the CarboMedics standard [19]. In a study of 32 patients with SJM Regent valves, Gelsomino and colleagues [9] found that 15.6% had PPM on discharge, but the rate decreased to 6.3% by 1 year. Bach and colleagues [13] reported the results obtained with SJM Regent valves in 361 patients at 17 hospitals. In 197 patients who underwent echocardiography at 6 months postoperatively, MPG was 10 mm Hg or less and EOAI was 1.0 cm 2 /m 2 or higher at all valve sizes from 19 mm up. No PPM was detected. However, Guenzinger and colleagues [14] studied 40 patients with SJM Regent valves and found severe PPM in 2.6% and moderate PPM in 10.3%. In the present study, the MPG exceeded mm Hg, and was 10 mm Hg for the 17-mm valve, but was 10 mm Hg or less for all sizes from 19 mm up. The prevalence of PPM was 25% at 17 mm, 18.8% at 19 mm, and 10% at 21 mm (10.3% of all patients). Because few patients with the SJM Regent valve have been reported so far, direct comparison is impossible. However, the SJM Regent valve appears to cause less PPM than conventional mechanical valves, although PPM occurs Table 5. Results of Dobutamine Stress Echocardiography a Dosage SBP DBP HR PPG MPG EOA EOAI (mm Hg) (mm Hg) (beats/min) EF (mm Hg) (mm Hg) (cm 2 ) (cm 2 /cm 2 ) Rest b c c c c c c c c c b c c c c b c c b b a Data are presented as mean standard deviation. b p c p g/kg/min; DBP diastolic blood pressure; EF ejection fraction; EOA effective orifice area; EOAI effective orifice area index; HR heart rate; MPG mean pressure gradient; PPG peak pressure gradient; SBP systolic blood pressure.

5 Ann Thorac Surg SEZAI ET AL 2010;89:87 92 ECHOCARDIOGRAPHIC DATA OF THE SJM REGENT with the smaller valve sizes. However, there have been reports stating that no clinical problems occur in patients with PPM, and we also found no clinical problems in any of the patients, even those with PPM. Okamura and colleagues [10] examined 23 cases of replacement with the 17-mm SJM Regent valve in Japanese patients, who generally have a smaller physical constitution than that of whites, and reported that the PPG was mm Hg, the EOA was cm 2 /m 2, and that PPM was found in 8 patients (34.8%). Takaseya and colleagues [11] reported replacement with the 17-mm SJM-Regent valve in 11 elderly patients, in whom the PPG was mm Hg, MPG was mm Hg, EOAI was cm 2 /m 2, and that incidence of PPM was 18.2%. Both articles concluded that the SJM Regent valve was effective for patients with a small aortic root and that there were no clinical issues, despite the occurrence of PPM. Although reports have indicated that PPM is strongly related to midterm and long-term mortality rates or cardiac events [20, 21], they have also indicated PPM is not significant for elderly patients or patients with a small physical constitution, although it may become a negative factor in young patients or patients with a larger physical constitution [22], and there have been no issues even in cases of PPM using modern mechanical valves [23]. Therefore, there is no clear consensus with respect to PPM. The incidence of PPM with the SJM Regent valve is lower than for conventional valves, and it gives better short-term outcomes, being considered as the most effective mechanical valve currently available for cardiac operations in patients with a narrow valve ring, although assessment through long-term follow-up is necessary in the future. Dobutamine stress echocardiography of the SJM Regent valve has only been reported previously by Minardi and colleagues [12], who examined 19 cases of replacement with the 17-mm SJM Regent valve and reported significant increases during DSE for heart rate ( to beat/min), EF ( to ), PPG ( to mm Hg), and MPG ( to mm Hg) compared with values at rest, whereas EOA and EOAI were unchanged. In the present study, systolic blood pressure, heart rate, PPG, and MPG were significantly increased during DSE at all time points at rest and at doses of 10, 20, 30, and 40 g/kg/min. Ejection fraction was significantly increased at rest and at doses of 30 and 40 g/kg/min, whereas EOA and EOAI were significantly increased at rest and at the dose of 40 g/kg/min. Four patients showed blood pressure higher than 100 mm Hg by dobutamine stress. Dobutamine stress echocardiography was performed on 2 patients with PPM, whose blood pressure did not exceed 100 mm Hg. However, DSE was performed only on 20 patients in the present study, from whom consent was first obtained, and no comparisons by size of mechanical valve and the presence or absence of PPM were made; therefore, further examination with an increased number of cases is necessary. In addition, we also consider it necessary to examine patients showing PPM data at rest. The debut of the SJM Regent valve in the Japanese medical field is very interesting, because there are more patients with a small aortic annulus in the Japanese population compared with the white population. In the present study, echocardiographic functional assessment on the SJM Regent valve proved that its functions as a prosthetic valve are excellent; thus, the acceptance of the prosthetic valve is expected in the future. Hereafter, we will accumulate the number of cases using this mechanical valve to continue further examination. In particular, we consider it necessary to make observations focusing on the point at which changes are found through long-term follow-up for cases of PPM. This study evaluated the function of the SJM Regent valve by Doppler echocardiography and DSE to clarify potential future problems with the valve. When performing the evaluation of artificial valves, however, it is important to examine valve-related complications during long-term follow-up. One limitation of our study is that such an examination could not be performed because of the small number of patients and short follow-up period. Therefore, more patients must be recruited and longer-term results need to be obtained in the future. By performing a randomized controlled trial for comparison with other mechanical valves and biologic prostheses, it will be possible to further clarify the efficacy and potential problems of the SJM Regent valve. Such a randomized controlled trial should be planned in the future. References 1. Sezai Y. Experience with the St. Jude Medical prosthesis. In: DeBakey ME (ed). Advances in cardiac valves: clinical perspectives. DeBakey ME (ed). New York: Yorke Medical Books; 1983: Emery RW, Krough CC, Arom KV, Emery AM, et al. The St. Jude Medical cardiac valve prosthesis: a 25-year experience with single valve replacement. Ann Thorac Surg 2005;79: Yoshikawa K, Fukunaga S, Arinaga K, et al. Long-term results of aortic valve replacement with a small St. Jude Medical valve in Japanese patients. Ann Thorac Surg 2008;85: Sezai A, Shiono M, Akiyama K, et al. Evaluation of St. Jude Medical valve s long-term function by Doppler echocardiography. Ann Thorac Cardiovasc Surg 2001;7: Sezai A, Shiono M, Hata M, Iida M, Negishi N, Sezai Y. 40 years experience in mitral valve replacement using Starr- Edwards, St. Jude Medical and ATS valves. Ann Thorac Cardiovasc Surg 2006;12: Sezai A, Shiono M, Akiyama K, et al. A case of aortic valve replacement with St. Jude Medical Regent valve (first implant in Japan). Ann Thorac Cardiovasc Surg 2005;11: Blais C, Dumesnil JG, Baillot R, Simard S, Doyle D, Pibator P. Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation 2003;108: Pibarot P, Dumesnil JG. Prosthesis-patient mismatch: definition, clinical impact, and prevention. Heart 2006;92: Gelsomino S, Morocutti G, Da Col P, et al. Early in vivo hemodynamic results after aortic valve replacement with the St. Jude Medical Regent mechanical heart valve in patients with pure aortic stenosis. J Card Surg 2003;18: Okamura H, Yamaguchi A, Tanaka M, et al. The 17-mm St. Jude Medical Regent valve is a valid option for patients with a small aortic annulus. Ann Thorac Surg 2009;87: Takaseya T, Kawara T, Tokunaga S, et al. Aortic valve replacement with 17-mm St. Jude Medical prostheses for a small aortic root in elderly patients. Ann Thorac Surg 2007;83: Minardi G, Manzara C, Creazzo V, et al. Evaluation of 17-mm St. Jude Medical Regent prosthetic aortic heart valves by rest and dobutamine stress echocardiography. J Cardiothorac Surg 2006;1: ADULT CARDIAC

6 ADULT CARDIAC 92 SEZAI ET AL Ann Thorac Surg ECHOCARDIOGRAPHIC DATA OF THE SJM REGENT 2010;89: Bach DS, Sakwa MP, Goldbach M, Petracek MR, Emery RW, Mohr FW. Hemodynamics and early clinical performance of the St. Jude Medical Regent mechanical aortic valve. Ann Thorac Surg 2002;74: Guenzinger R, Eichinger WB, Hettich I, et al. A prospective randomized comparison of the Medtronic Advantage Supra and St. Jude Medical Regent mechanical heart valves in the aortic position: is there an additional benefit of supra-annular valve positioning? J Thorac Cardiovasc Surg 2008;136: Vitale N, Caldarera I, Muneretto C, et al. Clinical evaluation of St. Jude Medical Hemodynamic Plus versus standard aortic valve prostheses: the Italian multicenter prospective, randomized study. J Thorac Cardiovasc Surg 2001;122: Labrousse LM, Choukroun E, Serena D. Prospective study of hemodynamic performances of standard ATS and AP-ATS valves. J Heart Valve Dis 2003;12: Gillinov AM, Blackstone EH, Alster JM, et al. The CarboMedics Top Hat supraannular aortic valve: a multicenter study. Ann Thorac Surg 2003;75: Walker DK, Brendzel AM, Scotten LN. The new St. Jude Medical Regent mechanical heart valve: laboratory measurements of hydrodynamic performance. J Heart Valve Dis 1999;8: Albes JM, Hartrumpf M, Rudolph V, et al. Are mechanical valves with enhanced inner diameter advantageous in the small sized aortic annulus? Ann Thorac Surg 2003;76: Mohty-Echahidi D, Malouf JF, Girard SG, et al. Impact of prosthesis-patient mismatch on long-term survival in patients with small St. Jude Medical mechanical prostheses in the aortic position. Circulation 2006;113: Tasca G, Mhagna Z, Perotti S, et al. Impact of prosthesispatient mismatch on cardiac events and midterm mortality after aortic valve replacement in patients with pure aortic stenosis. Circulation 2006;113: Moon MR, Pasque MK, Munfakh NA, et al. Prosthesis-patient mismatch after aortic valve replacement: impact of age and body size on late survival. Ann Thorac Surg 2006;81: Izzat MB, Kadir I, Reeves B, Wilde P, Bryan AJ, Angelini GD. Patient-prosthesis mismatch is negligible with modern small-size aortic valve prostheses. Ann Thorac Surg 1999;68: INVITED COMMENTARY The St. Jude Medical Regent valve (St. Jude Medical, St. Paul, MN) was designed to maximize the effective flow orifice area for any given tissue annulus diameter. It was the first mechanical valve prosthesis to demonstrate left ventricular mass regression similar to stentless biologic valves, yet with unquestioned durability. Sezai and colleagues [1] are well known for their work with prosthetic heart valves and have documented the hemodynamics of smaller sized St. Jude Medical Regent valves implanted in the Japanese population. They found low peak to peak and low mean gradients across the prostheses, and a minimal incidence of patient prosthesis mismatch (PPM). Although the numbers are small, if PPM was present, there was no impact on the postoperative morbidity or mortality or on the follow-up survival. These data are similar to our report that was made to The Society of Thoracic Surgeons in 2005 (unpublished data), in which we presented a review of PPM after aortic valve replacement, using the St. Jude Medical valve and follow-up for as long as 25 years. The PPM occurred the least with the St. Jude Medical Regent valve compared with the HP and standard model valves. In addition, PPM (whether moderate or severe, geometric, or echocardiographically determined) did not impact mortality or survival for the operative course or for the long-term follow-up. However, age, ventricular function, and patient activity level were unable to be evaluated. Importantly, this may be an advantage for using a mechanical valve prosthesis. In the early postoperative period, even though a mismatch may occur, the transvalvar gradients are substantially reduced from that of the aortic stenosis of the preoperative period. This allows for ventricular recovery and ventricular mass regression. Gradients across the mechanical prosthesis do not change with time, as Dr Emery discloses a financial relationship with St. Jude Medical Inc. opposed to biological valves, which most commonly become more rigid as structural valve degeneration occurs, recreating an environment of aortic stenosis within the biologic valve. As with Sezai and colleagues [1], our group has not found aortic annular enlargement procedures necessary since the approval of the St. Jude Medical Regent valve. However, it is interesting that the highest incidence of severe PPM in this report consisted of only two cases having occurred with the size 21 St. Jude Medical Regent valve, which had a reported mean valve area of cm 2. This is approximately equivalent to the flow orifice area of a St. Jude Medical standard size 24 prosthesis; therefore, it must have been implanted in substantially larger individuals. In summary, we concur with the conclusion of the article that finds the following: Echocardiographic functional assessment was done on the SJM Regent valve and proved that its functions as a prosthetic valve are excellent. Robert W. Emery, MD Cardiovascular Surgery St. Joseph s Hospital 69 W Exchange St St. Paul, MN rwemery@healtheast.org Goya V. Raikar, MD Regions Hospital 655 Jackson St St. Paul, MN Reference 1. Sezai A, Kasamaki Y, Abe K, et al. Assessment of the St. Jude Medical Regent prosthetic valve by continuous-wave Doppler and dobutamine stress echocardiography. Ann Thorac Surg 2010;89: by The Society of Thoracic Surgeons /10/$36.00 Published by Elsevier Inc doi: /j.athoracsur

A Case of Aortic Valve Replacement with St. Jude Medical Regent Valve (First Implant in Japan)

A Case of Aortic Valve Replacement with St. Jude Medical Regent Valve (First Implant in Japan) Case Report A Case of Aortic Valve Replacement with St. Jude Medical Regent Valve (First Implant in Japan) Akira Sezai, MD, 1 Motomi Shiono, MD, 1 Kenji Akiyama, MD, 1 Yukihiko Orime, MD, 1 Mitsumasa Hata,

More information

Evaluation of St. Jude Medical Valve s Long-term Function by Doppler Echocardiography

Evaluation of St. Jude Medical Valve s Long-term Function by Doppler Echocardiography Evaluation of St. Jude Medical Valve s Long-term Function by Doppler Echocardiography Akira Sezai, MD, Motomi Shiono, MD, Kenji Akiyama, MD, Yukihiko Orime, MD, Hiroaki Hata, MD, Shinya Yagi, MD, Tomonori

More information

Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement?

Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Original Article Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Hiroaki Sakamoto, MD, PhD, and Yasunori Watanabe, MD, PhD Background: Recently, some articles

More information

Since Hufnagel and associates implanted a plastic ball

Since Hufnagel and associates implanted a plastic ball Evaluation of Valve Sound and Its Effects on ATS Prosthetic Valves in Patients Quality of Life Akira Sezai, MD, Motomi Shiono, MD, Yukihiko Orime, MD, Hiroaki Hata, MD, Shinya Yagi, MD, Nanao Negishi,

More information

An Evaluation Study on the Quietness of the ATS Valve

An Evaluation Study on the Quietness of the ATS Valve Original Article An Evaluation Study on the Quietness of the ATS Valve Tomofumi Umeda, MD Background: The closure sound of the ATS bileaflet mechanical valve is said to be quieter than that of the other

More information

How to Avoid Prosthesis-Patient Mismatch

How to Avoid Prosthesis-Patient Mismatch How to Avoid Prosthesis-Patient Mismatch Philippe Pibarot, DVM, PhD, FACC, FAHA, FASE, FESC Canada Research Chair in Valvular Heart Diseases INSTITUT UNIVERSITAIRE DE CARDIOLOGIE ET DE PNEUMOLOGIE DE QUÉBEC

More information

Prosthetic valve dysfunction: stenosis or regurgitation

Prosthetic valve dysfunction: stenosis or regurgitation Prosthetic valve dysfunction: stenosis or regurgitation Jean G. Dumesnil MD, FRCP(C), FACC, FASE(Hon) Quebec Heart and Lung Institute, Québec, Québec No disclosures Possible Causes of High Gradients in

More information

Patient/prosthesis mismatch: how to evaluate and when to act?

Patient/prosthesis mismatch: how to evaluate and when to act? Patient/prosthesis mismatch: how to evaluate and when to act? Svend Aakhus, MD, PhD Oslo University Hospital, Norway Disclosures: No conflict of interest Types of aortic valve prostheses (AVR) Mechanical

More information

Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient Prosthesis Mismatch a Relevant Phenomenon?

Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient Prosthesis Mismatch a Relevant Phenomenon? Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient Prosthesis Mismatch a Relevant Phenomenon? Andrea Garatti, MD, Francesca Mori, MD, Francesco Innocente, MD, Alberto Canziani, MD, Piervincenzo

More information

SOLO SMART. The smart way to return to life. Native-like performance now with stented-like implantability

SOLO SMART. The smart way to return to life. Native-like performance now with stented-like implantability SOLO SMART TM The smart way to return to life Native-like performance now with stented-like implantability MANY PATIENTS NEED SUPERIOR HEMODYNAMIC PERFORMANCE TO RETURN TO THEIR NORMAL LIFESTYLE. 2 SOLO

More information

The impact of prosthesis patient mismatch after aortic valve replacement varies according to age at operation

The impact of prosthesis patient mismatch after aortic valve replacement varies according to age at operation Editor s choice Scan to access more free content 1 Division of Cardiac Surgery, The Johns Hopkins Hospital, Baltimore, Maryland, USA 2 Division of Cardiac Surgery, University of Ottawa Heart Institute,

More information

Valve prosthesis-patient mismatch (PPM) was first defined

Valve prosthesis-patient mismatch (PPM) was first defined Impact of Valve Prosthesis-Patient Mismatch on Short-Term Mortality After Aortic Valve Replacement Claudia Blais, BSc; Jean G. Dumesnil, MD; Richard Baillot, MD; Serge Simard, MS; Daniel Doyle, MD; Philippe

More information

Management of Difficult Aortic Root, Old and New solutions

Management of Difficult Aortic Root, Old and New solutions Management of Difficult Aortic Root, Old and New solutions Hani K. Najm MD, Msc, FRCSC,, FACC, FESC Chairman, Pediatric and Congenital Heart Surgery Cleveland Clinic Conflict of Interest None Difficult

More information

Favorable Results in Patients with Small Size CarboMedics Heart Valves in the Aortic Position

Favorable Results in Patients with Small Size CarboMedics Heart Valves in the Aortic Position Favorable Results in Patients with Small Size CarboMedics Heart Valves in the Aortic Position Kazuhito Imanaka, MD, Shinichi Takamoto, MD, and Akira Furuse, MD 2 Hemodynamic performance of the CarboMedics

More information

Impact of Prosthesis-Patient Mismatch on Long-Term Survival After Aortic Valve Replacement

Impact of Prosthesis-Patient Mismatch on Long-Term Survival After Aortic Valve Replacement Journal of the American College of Cardiology Vol. 53, No. 1, 9 9 by the American College of Cardiology Foundation ISSN 735-197/9/$36. Published by Elsevier Inc. doi:1.116/j.jacc.8.9.22 Valvular Heart

More information

Effect of Valve Suture Technique on Incidence of Paraprosthetic Regurgitation and 10-Year Survival

Effect of Valve Suture Technique on Incidence of Paraprosthetic Regurgitation and 10-Year Survival Effect of Valve Suture Technique on Incidence of Paraprosthetic Regurgitation and 10-Year Survival Sukumaran K. Nair, FRCS (C Th), Gauraang Bhatnagar, MBBS, Oswaldo Valencia, MD, and Venkatachalam Chandrasekaran,

More information

Incidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves

Incidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves INTERVENTION/VALVULAR HEART DISEASE ORIGINAL ARTICLE Cardiology Journal 2016, Vol. 23, No. 2, 178 183 DOI: 10.5603/CJ.a2016.0011 Copyright 2016 Via Medica ISSN 1897 5593 Incidence of prosthesis-patient

More information

PPM: How to fit a big valve in a small heart

PPM: How to fit a big valve in a small heart PPM: How to fit a big valve in a small heart Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC King Abdulaziz Cardiac Centre National Guard Health Affairs Riyadh, Saudi Arabia GHA meeting Muscat

More information

Hemodynamics and Early Clinical Performance of the St. Jude Medical Regent Mechanical Aortic Valve

Hemodynamics and Early Clinical Performance of the St. Jude Medical Regent Mechanical Aortic Valve Hemodynamics and Early Clinical Performance of the St. Jude Medical Regent Mechanical Aortic Valve David S. Bach, MD, Marc P. Sakwa, MD, Martin Goldbach, MD, Michael R. Petracek, MD, Robert W. Emery, MD,

More information

Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis

Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis CLINICS 2012;67(1):55-60 DOI:10.6061/clinics/2012(01)09 CLINICAL SCIENCE Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis Luis M. Astudillo, I Orlando

More information

Experience with 500 Stentless Aortic Valve Replacements

Experience with 500 Stentless Aortic Valve Replacements Experience with 500 Stentless Aortic Valve Replacements Dimitrios C. Iliopoulos, MD Cardiac Surgeon Ass. Professor of Surgery University of Athens, School of Medicine I declare no conflict of interest

More information

Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis?

Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis? EuroValves 2015, Nice Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis? Philippe Pibarot, DVM, PhD, FACC, FAHA, FASE FESC Canada Research Chair in Valvular Heart Diseases Université LAVAL Disclosure

More information

Prosthesis-Patient Mismatch in High Risk Patients with Severe Aortic Stenosis in a Randomized Trial of a Self-Expanding Prosthesis

Prosthesis-Patient Mismatch in High Risk Patients with Severe Aortic Stenosis in a Randomized Trial of a Self-Expanding Prosthesis Prosthesis-Patient Mismatch in High Risk Patients with Severe Aortic Stenosis in a Randomized Trial of a Self-Expanding Prosthesis George L. Zorn, III On Behalf of the CoreValve US Clinical Investigators

More information

Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz

Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia Decision process for

More information

40 Years Experience in Mitral Valve Replacement Using Starr-Edwards, St. Jude Medical and ATS Valves

40 Years Experience in Mitral Valve Replacement Using Starr-Edwards, St. Jude Medical and ATS Valves Original Article 40 Experience in Mitral Valve Replacement Using Starr-Edwards, St. Jude Medical and ATS Valves Akira Sezai, MD, Motomi Shiono, MD, Mitsumasa Hata, MD, Mitsuru Iida, MD, Masataka Yoda,

More information

CARDIACSURGERY TODAY. Commentary and Analysis on Advances in the Surgical Treatment of Cardiac Disease

CARDIACSURGERY TODAY. Commentary and Analysis on Advances in the Surgical Treatment of Cardiac Disease VOLUME 1 NUMBER 2 23 CARDIACSURGERY TODAY Commentary and Analysis on Advances in the Surgical Treatment of Cardiac Disease EDITORS-IN-CHIEF Robert W Emery, St Paul, MN, USA Francesco Musumeci, Rome, Italy

More information

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision Prof. Pino Fundarò, MD Niguarda Hospital Milan, Italy Introduction

More information

Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves

Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Speaker's name: I have the following potential conflicts of interest to report: Proctorship for Medtronic Agenda Failure modes

More information

TSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD

TSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD TSDA Boot Camp September 13-16, 2018 Introduction to Aortic Valve Surgery George L. Hicks, Jr., MD Aortic Valve Pathology and Treatment Valvular Aortic Stenosis in Adults Average Course (Post mortem data)

More information

2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD

2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD 2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD David L Saint M.D. Tallahassee Memorial Hospital Southern Medical Group Division of Cardiothoracic

More information

The implantation of bioprostheses is the preferred. Influence of Prosthesis Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement

The implantation of bioprostheses is the preferred. Influence of Prosthesis Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement Influence of Prosthesis Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement Shahab Nozohoor, MD, Johan Nilsson, MD, PhD, Carsten Lührs, MD, Anders Roijer, MD, PhD, and Johan Sjögren,

More information

Patient prosthesis mismatch after mitral valve replacement: Myth or reality?

Patient prosthesis mismatch after mitral valve replacement: Myth or reality? Patient prosthesis mismatch after mitral valve replacement: Myth or reality? Pasquale Totaro, MD, a and Vincenzo Argano, MD b Objective: Determining the risk of patient prosthesis mismatch after mitral

More information

Mitral Gradients and Frequency of Recurrence of Mitral Regurgitation After Ring Annuloplasty for Ischemic Mitral Regurgitation

Mitral Gradients and Frequency of Recurrence of Mitral Regurgitation After Ring Annuloplasty for Ischemic Mitral Regurgitation Mitral Gradients and Frequency of Recurrence of Mitral Regurgitation After Ring Annuloplasty for Ischemic Mitral Regurgitation Matthew L. Williams, MD, Mani A. Daneshmand, MD, James G. Jollis, MD, John

More information

QUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH

QUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH QUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH 1,2 Radu A. SASCĂU 3 Cristina OLARIU 1,2 Cristian STĂTESCU 1 Internal Medicine Department, Gr.T.Popa University of Medicine and Pharmacy,

More information

Seventeen-millimeter St. Jude Medical Regent valve in patients with small aortic annulus: dose moderate prosthesis-patient mismatch matter?

Seventeen-millimeter St. Jude Medical Regent valve in patients with small aortic annulus: dose moderate prosthesis-patient mismatch matter? Hu et al. Journal of Cardiothoracic Surgery 2014, 9:17 RESEARCH ARTICLE Open Access Seventeen-millimeter St. Jude Medical Regent valve in patients with small aortic annulus: dose moderate prosthesis-patient

More information

Echocardiographic changes after aortic valve replacement: Does the failure rate of mitral valve change? Original Article

Echocardiographic changes after aortic valve replacement: Does the failure rate of mitral valve change? Original Article Echocardiographic changes after aortic valve replacement: Does the failure rate of mitral valve change? Abstract Arezoo Khosravi (1), Hadi Sheykhloo (2), Reza Karbasi-Afshar (1), Amin Saburi (3) Original

More information

Comparison of eight prosthetic aortic valves in a cadaver model

Comparison of eight prosthetic aortic valves in a cadaver model Thomas Jefferson University Jefferson Digital Commons Department of Surgery Faculty Papers Department of Surgery July 2007 Comparison of eight prosthetic aortic valves in a cadaver model Benjamin A. Youdelman

More information

Echocardiographic Evaluation of Mitral Valve Prostheses

Echocardiographic Evaluation of Mitral Valve Prostheses Echocardiographic Evaluation of Mitral Valve Prostheses Dennis A. Tighe, M.D., FACC, FACP, FASE Cardiovascular Medicine University of Massachusetts Medical School Worcester, MA www.asecho.org 1 Nishimura

More information

Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia.

Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Decision process for Management of any valve Timing Feasibility

More information

Prof. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM

Prof. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM The Patient with Aortic Stenosis and Mitral Regurgitation Prof. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM Aortic Stenosis + Mitral Regurgitation?

More information

Influence of Prosthetic Heart Valve Sound on a Patient s Quality of Life

Influence of Prosthetic Heart Valve Sound on a Patient s Quality of Life Original Article Influence of Prosthetic Heart Valve Sound on a Patient s Quality of Life Katsuo Nishi, MD, 1 Kiyoyuki Eishi, MD, 2 Yoshisada Shibata, PhD, 3 Jun Amano, MD, PhD, 4 Tatsuo Kaneko, MD, 5

More information

Doppler echocardiography is currently the

Doppler echocardiography is currently the Doppler Echocardiography of 119 Normal-functioning St Jude Medical Mitral Valve Prostheses: A Comprehensive Assessment Including Time-velocity Integral Ratio and Prosthesis Performance Index* Joseph F.

More information

Copyright by ICR Publishers 2014

Copyright by ICR Publishers 2014 Comprehensive Hemodynamic Performance and Frequency of Patient-Prosthesis Mismatch of the St. Jude Medical Trifecta Bioprosthetic Aortic Valve Ajay Yadlapati 1, Jimmy Diep 3, Mary-Jo Barnes 2, Tristan

More information

Assessment of Hemodynamics Properties of a New-Type Artificial Heart Valve Prosthesis Using Catheterization and Echocardiography

Assessment of Hemodynamics Properties of a New-Type Artificial Heart Valve Prosthesis Using Catheterization and Echocardiography American Journal of Hematology 81:563 567 (2006) Assessment of Hemodynamics Properties of a New-Type Artificial Heart Valve Prosthesis Using Catheterization and Echocardiography Y.J. Zeng, 1,4 * S.W. Xu,

More information

Clinical material and methods. Fukui Cardiovascular Center, Fukui, Japan

Clinical material and methods. Fukui Cardiovascular Center, Fukui, Japan Mitral Valve Regurgitation after Atrial Septal Defect Repair in Adults Shohei Yoshida, Satoshi Numata, Yasushi Tsutsumi, Osamu Monta, Sachiko Yamazaki, Hiroyuki Seo, Takaaki Samura, Hirokazu Ohashi Fukui

More information

Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease

Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease TIRONE E. DAVID, MD ; SEMIN THORAC CARDIOVASC SURG 19:116-120c 2007 ELSEVIER INC. PRESENTED BY INTERN 許士盟 Mitral valve

More information

Influence of patient gender on mortality after aortic valve replacement for aortic stenosis

Influence of patient gender on mortality after aortic valve replacement for aortic stenosis Influence of patient gender on mortality after aortic valve replacement for aortic stenosis Jennifer Higgins, MD, W. R. Eric Jamieson, MD, Osama Benhameid, MD, Jian Ye, MD, Anson Cheung, MD, Peter Skarsgard,

More information

Echocardiographic variables associated with mitral regurgitation after aortic valve replacement for aortic valve stenosis

Echocardiographic variables associated with mitral regurgitation after aortic valve replacement for aortic valve stenosis The Egyptian Heart Journal (2013) 65, 135 139 Egyptian Society of Cardiology The Egyptian Heart Journal www.elsevier.com/locate/ehj www.sciencedirect.com ORIGINAL ARTICLE Echocardiographic variables associated

More information

The Tricuspid Valve: The Not So Forgotten Valve. Manuel J Antunes Cardiothoracic Surgery Coimbra, Portugal

The Tricuspid Valve: The Not So Forgotten Valve. Manuel J Antunes Cardiothoracic Surgery Coimbra, Portugal The Tricuspid Valve: The Not So Forgotten Valve Manuel J Antunes Cardiothoracic Surgery Coimbra, Portugal No Conflicts of Interest to declare with regards to this subject 2 INCIDENCE OF TRICUSPID REGURGITATION

More information

ECHOCARDIOGRAPHY DATA REPORT FORM

ECHOCARDIOGRAPHY DATA REPORT FORM Patient ID Patient Study ID AVM - - Date of form completion / / 20 Initials of person completing the form mm dd yyyy Study period Preoperative Postoperative Operative 6-month f/u 1-year f/u 2-year f/u

More information

Indication, Timing, Assessment and Update on TAVI

Indication, Timing, Assessment and Update on TAVI Indication, Timing, Assessment and Update on TAVI Swedish Heart and Vascular Institute Ming Zhang MD PhD Interventional Cardiology Structure Heart Disease Conflict of Interest None Starr- Edwards Mechanical

More information

HOW IMPORTANT ARE THESE ECHO MEASUREMENTS ANYWAY?

HOW IMPORTANT ARE THESE ECHO MEASUREMENTS ANYWAY? HOW IMPORTANT ARE THESE ECHO MEASUREMENTS ANYWAY? John D. Carroll, MD Professor, Director of Interventional Cardiology and Co-Medical Director of the Cardiac and Vascular Center, University of Colorado

More information

Reverse left atrium and left ventricle remodeling after aortic valve interventions

Reverse left atrium and left ventricle remodeling after aortic valve interventions Reverse left atrium and left ventricle remodeling after aortic valve interventions Alexandra Gonçalves, Cristina Gavina, Carlos Almeria, Pedro Marcos-Alberca, Gisela Feltes, Rosanna Hernández-Antolín,

More information

Late incidence and predictors of persistent or recurrent heart failure in patients with aortic prosthetic valves

Late incidence and predictors of persistent or recurrent heart failure in patients with aortic prosthetic valves Late incidence and predictors of persistent or recurrent heart failure in patients with aortic prosthetic valves Marc Ruel, MD, MPH a,b Fraser D. Rubens, MD a Roy G. Masters, MD a Andrew L. Pipe, MD a

More information

EDITORIAL TX ET CSP ACD GTS ACD

EDITORIAL TX ET CSP ACD GTS ACD Vitale et al Clinical evaluation of St Jude Medical Hemodynamic Plus versus standard aortic valve prostheses: The Italian multicenter, prospective, randomized study Nicola Vitale, MD, PhD Ilaria Caldarera,

More information

Michigan Society of Echocardiography 30 th Year Jubilee

Michigan Society of Echocardiography 30 th Year Jubilee Michigan Society of Echocardiography 30 th Year Jubilee Stress Echocardiography in Valvular Heart Disease Moving Beyond CAD Karthik Ananthasubramaniam, MD FRCP (Glas) FACC FASE FASNC Associate Professor

More information

Journal of the American College of Cardiology Vol. 44, No. 9, by the American College of Cardiology Foundation ISSN /04/$30.

Journal of the American College of Cardiology Vol. 44, No. 9, by the American College of Cardiology Foundation ISSN /04/$30. Journal of the American College of Cardiology Vol. 44, 9, 2004 2004 by the American College of Cardiology Foundation ISSN 0735-1097/04/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2004.04.062 Relation

More information

Persistent Tricuspid Regurgitation After Tricuspid Annuloplasty During Redo Valve Surgery Affects Late Survival and Valve-Related Events

Persistent Tricuspid Regurgitation After Tricuspid Annuloplasty During Redo Valve Surgery Affects Late Survival and Valve-Related Events Circulation Journal Official Journal of the Japanese Circulation Society http://www.j-circ.or.jp Advance Publication by-j-stage Persistent Tricuspid Regurgitation After Tricuspid Annuloplasty During Redo

More information

We present the case of an asymptomatic, 75-year-old

We present the case of an asymptomatic, 75-year-old Images in Cardiovascular Medicine Asymptomatic Rupture of the Left Ventricle Lech Paluszkiewicz, MD; Stefan Ożegowski, MD; Mohammad Amin Parsa, MD; Jan Gummert, PhD, MD We present the case of an asymptomatic,

More information

25 different brand names >44 different models Sizes mm

25 different brand names >44 different models Sizes mm Types of Prosthetic Valves BIOLOGIC STENTED Porcine xenograft Pericardial xenograft STENTLESS Porcine xenograft Pericardial xenograft Homograft (allograft) Autograft PERCUTANEOUS MECHANICAL Bileaflet Single

More information

Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment

Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment W.R.E. Jamieson, MD; L.H. Burr, MD; R.T. Miyagishima, MD; M.T. Janusz, MD; G.J. Fradet, MD; S.V. Lichtenstein, MD; H. Ling, MD Background

More information

Echocardiographic Evaluation of Aortic Valve Prosthesis

Echocardiographic Evaluation of Aortic Valve Prosthesis Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Co-Director, Echocardiography, Director, Interventional Cardiology Research, Beaumont Health System

More information

Surgery for Acquired Cardiovascular Disease

Surgery for Acquired Cardiovascular Disease Bovine pericardial versus porcine stented replacement aortic valves: Early results of a randomized comparison of the Perimount and the Mosaic valves John B. Chambers, MD, FACC, Ronak Rajani, MD, MRCP,

More information

Predicting functional mitral stenosis after restrictive annuloplasty for ischemic mitral regurgitation

Predicting functional mitral stenosis after restrictive annuloplasty for ischemic mitral regurgitation Powered by TCPDF (www.tcpdf.org) This is a provisional PDF only. Copyedited and fully formatted version will be made available soon. ONLINE FIRST ISSN: 1897-5593 e-issn: 1898-018X Predicting functional

More information

Restrictive Annuloplasty for Ischemic Mitral Regurgitation May Induce Functional Mitral Stenosis

Restrictive Annuloplasty for Ischemic Mitral Regurgitation May Induce Functional Mitral Stenosis Journal of the American College of Cardiology Vol. 51, No. 17, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2007.11.082

More information

Although mitral valve replacement (MVR) is no longer the surgical

Although mitral valve replacement (MVR) is no longer the surgical Surgery for Acquired Cardiovascular Disease Ruel et al Late incidence and predictors of persistent or recurrent heart failure in patients with mitral prosthetic valves Marc Ruel, MD, MPH a,b Fraser D.

More information

Valve Prosthesis Patient Mismatch (VP PM)

Valve Prosthesis Patient Mismatch (VP PM) Journal of the American College of Cardiology Vol. 60, No. 13, 2012 2012 by the American College of Cardiology Foundation ISSN 0735-1097/$36.00 Published by Elsevier Inc. http://dx.doi.org/10.1016/j.jacc.2012.05.035

More information

Impact of prosthesis-patient mismatch on early and late mortality after aortic valve replacement

Impact of prosthesis-patient mismatch on early and late mortality after aortic valve replacement Koene et al. Journal of Cardiothoracic Surgery 2013, 8:96 RESEARCH ARTICLE Open Access Impact of prosthesis-patient mismatch on early and late mortality after aortic valve replacement Bart M Koene 2*,

More information

The Edge-to-Edge Technique f For Barlow's Disease

The Edge-to-Edge Technique f For Barlow's Disease The Edge-to-Edge Technique f For Barlow's Disease Ottavio Alfieri, Michele De Bonis, Elisabetta Lapenna, Francesco Maisano, Lucia Torracca, Giovanni La Canna. Department of Cardiac Surgery, San Raffaele

More information

Doppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position

Doppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position Doppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position Chee-Siong Soo, MRCP, Mestres Ca, MD, Monica Tay, Joon-Kuan Yeoh, MRCP, Eugene Sim, FRCS, and M. Choo, FRCP,

More information

PROSTHETIC VALVE BOARD REVIEW

PROSTHETIC VALVE BOARD REVIEW PROSTHETIC VALVE BOARD REVIEW The correct answer D This two chamber view shows a porcine mitral prosthesis with the typical appearance of the struts although the leaflets are not well seen. The valve

More information

Which Type of Secondary Tricuspid Regurgitation Accompanying Mitral Valve Disease Should Be Surgically Treated?

Which Type of Secondary Tricuspid Regurgitation Accompanying Mitral Valve Disease Should Be Surgically Treated? Ann Thorac Cardiovasc Surg 2013; 19: 428 434 Online January 31, 2013 doi: 10.5761/atcs.oa.12.01929 Original Article Which Type of Secondary Tricuspid Regurgitation Accompanying Mitral Valve Disease Should

More information

Mechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute

Mechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute Mechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute Assistant Professor, Georgetown School of Medicine

More information

Reoperation for a Patient 25 Years after a Starr-Edwards Ball Mitral Valve was Installed

Reoperation for a Patient 25 Years after a Starr-Edwards Ball Mitral Valve was Installed Case Report Reoperation for a Patient 25 Years after a Starr-Edwards Ball Valve was Installed Zhidong Ye, MD, Motomi Shiono, MD, Akira Sezai, MD, Tatsuya Inoue, MD, Mitsumasa Hata, MD, Tetsuya Niino, MD,

More information

Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance

Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance W. R. Eric Jamieson, MD, a Friedrich-Christian Riess, MD, b Peter J. Raudkivi, MD, c Jacques Metras, MD, d Edward F. G. Busse,

More information

Prosthesis-Patient Mismatch After Aortic Valve Replacement: Impact of Age and Body Size on Late Survival

Prosthesis-Patient Mismatch After Aortic Valve Replacement: Impact of Age and Body Size on Late Survival ORIGINAL ARTICLES: ADULT CARDIAC SURGERY: To participate in The Annals of Thoracic Surgery CME Program, please visit http://cme.ctsnetjournals.org. Prosthesis-Patient Mismatch After Aortic Valve Replacement:

More information

Intensity of oral anticoagulation after implantation of St. Jude Medical mitral or multiple valve replacement: lessons learned from GELIA (GELIA 5)

Intensity of oral anticoagulation after implantation of St. Jude Medical mitral or multiple valve replacement: lessons learned from GELIA (GELIA 5) European Heart Journal Supplements () 3 (Supplement Q), Q39 Q43 Intensity of oral anticoagulation after implantation of St. Jude Medical mitral or multiple valve replacement: lessons learned from GELIA

More information

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses ORIGINAL CONTRIBUTION 15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses WR Eric Jamieson, MD, Eva Germann, MSc, Michel R Aupart, MD 1, Paul H Neville, MD 1, Michel A Marchand,

More information

Percutaneous Therapy for Calcific Mitral Valve Disease

Percutaneous Therapy for Calcific Mitral Valve Disease 31 st Annual State of the Art Echocardiography San Diego, CA February 18, 2018 5:00 5:15 PM 15 min Percutaneous Therapy for Calcific Mitral Valve Disease Muhamed Sarić MD, PhD, MPA Director of Noninvasive

More information

Presenter Disclosure. Patrick O. Myers, M.D. No Relationships to Disclose

Presenter Disclosure. Patrick O. Myers, M.D. No Relationships to Disclose Presenter Disclosure Patrick O. Myers, M.D. No Relationships to Disclose Aortic Valve Repair by Cusp Extension for Rheumatic Aortic Insufficiency in Children Long term Results and Impact of Extension Material

More information

Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic,

Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic, Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic, transplantation and Vascular Surgery Hannover Medical School, Germany

More information

Echocardiographic Evaluation of Aortic Valve Prosthesis

Echocardiographic Evaluation of Aortic Valve Prosthesis Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Co Director, Echocardiography, Director, Interventional Cardiology Research, Beaumont Health System

More information

Effect of Patient-Prosthesis Mismatch in Aortic Position on Late-Onset Tricuspid Regurgitation and Clinical Outcomes after Double Valve Replacement

Effect of Patient-Prosthesis Mismatch in Aortic Position on Late-Onset Tricuspid Regurgitation and Clinical Outcomes after Double Valve Replacement Original Article Yonsei Med J 17 Sep;58(5):968-974 pissn: 0513-5796 eissn: 1976-2437 Effect of Patient-Prosthesis Mismatch in Aortic Position on Late-Onset Tricuspid Regurgitation and Clinical Outcomes

More information

The operative mortality rate after redo valvular operations

The operative mortality rate after redo valvular operations Clinical Outcomes of Redo Valvular Operations: A 20-Year Experience Naoto Fukunaga, MD, Yukikatsu Okada, MD, Yasunobu Konishi, MD, Takashi Murashita, MD, Mitsuru Yuzaki, MD, Yu Shomura, MD, Hiroshi Fujiwara,

More information

Late secondary TR after left sided heart disease correction: is it predictibale and preventable

Late secondary TR after left sided heart disease correction: is it predictibale and preventable Late secondary TR after left sided heart disease correction: is it predictibale and preventable Gilles D. Dreyfus Professor of Cardiothoracic surgery Nath J, et al. JACC 2004 PREDICT Incidence of secondary

More information

Clinical Outcome of Tricuspid Regurgitation. David Messika-Zeitoun

Clinical Outcome of Tricuspid Regurgitation. David Messika-Zeitoun Clinical Outcome of Tricuspid Regurgitation David Messika-Zeitoun I have financial relationships to disclose Consultant for: Edwards, Symetis and Valtech Tricuspid Regurgitation is a Common Finding Tricuspid

More information

Anticoagulation Therapy and Valve Surgery. Dr Pau Kiew Kong Consultant Cardiothoracic Surgeon

Anticoagulation Therapy and Valve Surgery. Dr Pau Kiew Kong Consultant Cardiothoracic Surgeon Anticoagulation Therapy and Valve Surgery Dr Pau Kiew Kong Consultant Cardiothoracic Surgeon Outline of lecture 1. Type of Valve Surgery 2. Anticoagulation requirements 3. Mechanical (Metallic) prosthetic

More information

Aortic Stenosis and Perioperative Risk With Non-cardiac Surgery

Aortic Stenosis and Perioperative Risk With Non-cardiac Surgery Aortic Stenosis and Perioperative Risk With Non-cardiac Surgery Aortic stenosis (AS) is characterized as a high-risk index for cardiac complications during non-cardiac surgery. A critical analysis of old

More information

Incidence And Predictors Of Left Bundle Branch Block After Transcatheter Aortic Valve Implantation

Incidence And Predictors Of Left Bundle Branch Block After Transcatheter Aortic Valve Implantation Incidence And Predictors Of Left Bundle Branch Block After Transcatheter Aortic Valve Implantation Ömer Aktug 1, MD; Guido Dohmen 2, MD; Kathrin Brehmer 1, MD; Verena Deserno 1 ; Ralf Herpertz 1 ; Rüdiger

More information

The St. Jude Medical mechanical valve is a low-profile,

The St. Jude Medical mechanical valve is a low-profile, Twenty-Five Year Experience With the St. Jude Medical Mechanical Valve Prosthesis J. Matthew Toole, MD, Martha R. Stroud, MS, John M. Kratz, MD, Arthur J. Crumbley III, MD, Scott M. Bradley, MD, Fred A.

More information

Bicuspid aortic root spared during ascending aorta surgery: an update of long-term results

Bicuspid aortic root spared during ascending aorta surgery: an update of long-term results Short Communication Bicuspid aortic root spared during ascending aorta surgery: an update of long-term results Marco Russo, Guglielmo Saitto, Paolo Nardi, Fabio Bertoldo, Carlo Bassano, Antonio Scafuri,

More information

Aortic valve replacement in predominant aortic stenosis: What is an appropriate size valve?

Aortic valve replacement in predominant aortic stenosis: What is an appropriate size valve? IJTCVS Joshi et al 141 Aortic valve replacement in predominant aortic stenosis: What is an appropriate size valve? Kishore Joshi, M.Ch., Sachin Talwar, M.Ch., Devagourou Velayoudham, M.Ch., Arkalgud Sampath

More information

Regression of Hypertrophy After Carpentier-Edwards Pericardial Aortic Valve Replacement

Regression of Hypertrophy After Carpentier-Edwards Pericardial Aortic Valve Replacement Regression of Hypertrophy After Carpentier-Edwards Pericardial Aortic Valve Replacement Steven S. Khan, MD, Robert J. Siegel, MD, Michele A. DeRobertis, RN, Carlos E. Blanche, MD, Robert M. Kass, MD, Wen

More information

THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT

THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT Gideon Cohen, MD Tirone E. David, MD Joan Ivanov, MSc Sue Armstrong, MSc

More information

Nineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years Follow-up

Nineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years Follow-up Nineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years Follow-up Dilip Sawant, FRCS, Arun K. Singh, MD, William C. Feng, MD, Arthur A. Bert, MD, and Fred Rotenberg, MD

More information

Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated?

Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated? Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated? RM Suri, V Sharma, JA Dearani, HM Burkhart, RC Daly, LD Joyce, HV Schaff Division of Cardiovascular Surgery, Mayo Clinic, Rochester,

More information

Prifti et al. Journal of Cardiothoracic Surgery (2015) 10:154 DOI /s

Prifti et al. Journal of Cardiothoracic Surgery (2015) 10:154 DOI /s Prifti et al. Journal of Cardiothoracic Surgery (2015) 10:154 DOI 10.1186/s13019-015-0361-3 RESEARCH ARTICLE Open Access Early and mid-term outcome in terms of functional and hemodynamic performance of

More information

Dysfunction of transcatheter mitral valve prosthesis. Early valve degeneration or thrombosis - that is the question.

Dysfunction of transcatheter mitral valve prosthesis. Early valve degeneration or thrombosis - that is the question. Dysfunction of transcatheter mitral valve prosthesis. Early valve degeneration or thrombosis - that is the question. Böhm A., Hricak V., Tomasovic B., Bena M., Postulka J. The National Institute of, Department

More information

Echocardiographic Evaluation of Aortic Valve Prosthesis

Echocardiographic Evaluation of Aortic Valve Prosthesis Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Director, Interventional Cardiology Research, Beaumont Health System Associate Professor of Medicine,

More information

TAVR: Echo Measurements Pre, Post And Intra Procedure

TAVR: Echo Measurements Pre, Post And Intra Procedure 2017 ASE Florida, Orlando, FL October 10, 2017 8:00 8:25 AM 25 min TAVR: Echo Measurements Pre, Post And Intra Procedure Muhamed Sarić MD, PhD, MPA Director of Noninvasive Cardiology Echo Lab Associate

More information