Imaging appearances of programmable ventricular shunt systems : What the radiologist needs to know
|
|
- Elizabeth Crawford
- 5 years ago
- Views:
Transcription
1 Imaging appearances of programmable ventricular shunt systems : What the radiologist needs to know Poster No.: C-2030 Congress: ECR 2012 Type: Educational Exhibit Authors: A. Gontsarova, S. C. Thust, J. Shand Smith, H. S. CHANDRASHEKAR; London/UK Keywords: Cerebrospinal fluid, Technical aspects, Surgery, Shunts, MR, CT, Conventional radiography, Neuroradiology brain DOI: /ecr2012/C-2030 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 25
2 Learning objectives To illustrate radiographic appearances of the ventricular shunt system with programmable valves which are currently used in Europe and to assist radiologists in the identification of programmable shunt valves to complete the assessment of skull radiographs in patients with shunts. Background Hydrocephalus affects 1-2% of the population. The prevalence of congenital and infantile hydrocephalus lies between 0.48 and 0.81 per 1000 births (including live and still births). In the United Kingdom and Ireland, the number of shunt operations is estimated to range between 3500 and 4000 per year according to the Cambridge based UK Shunt Registry. In the United States, about shunt procedures are carried out annually. Although to this date fixed-pressure shunt valves remain the most frequently inserted valve type in the UK, the adjustable (often-called programmable) valve has become an important tool in hydrocephalus treatment, particularly in the normal pressure hydrocephalus population and in pediatric patients with complex hydrocephalus. Programmable valves are more expensive, but offer an advantage in that the operating pressure of the valve can be altered by the use of an external magnet as a simple noninvasive procedure. The proportion of programmable valves used in the UK and Ireland has increased from 4.9% in 2000 to 22.4% in Page 2 of 25
3 Fig. 1: Graphic representation of the Miethke progav valve setting change. References: Reproduced with permission from Aesculap, Inc. One-third of all shunts fail within 1 year of placement, and in children 4.5% per year thereafter. Manifestations of shunt failure are variable so patients with shunts frequently undergo radiographic evaluation. The important part of the evaluations include plain radiographs of the shunt system. Most CSF shunts consist of 3 components: a ventricular catheter, a valve, and a distal catheter. A shunt is a completely internalized system, as opposed to an external ventricular drain in which a ventricular catheter drains to a collection system at the bedside. Page 3 of 25
4 Fig. 2: Typical ventricularperitonel shunt system with a programmable valve. References: Department of Radiology, National Hospital Neurology and Neurosurgery, UCLH, London/UK 2012 The catheter components of a shunt are made from Silastic (Dow Corning, Midland, Michigan), a form of rubber tubing resistant to breakdown in the body. They are frequently impregnated with radiopaque material to aid in their radiographic visualization. The ventricular catheter sits within 1 of the ventricular spaces in the brain, most commonly the right lateral ventricle. The ventricular catheter is connected to a valve that regulates flow. To counter a siphoning effect associated with upright posture, many shunt systems also include an antisiphon device or gravitational unit; this reduces overdrainage when the patient is standing. Page 4 of 25
5 The gravitational unit is a cylindrical structure and it function of the gravitational unit is to prevent postural overdrainage. Its opening pressure gradually increases as the patient moves from a supine to an upright position. Multiple models of the gravitational unit are available, each with a different maximal opening pressure. Images for this section: Fig. 1: Graphic representation of the Miethke progav valve setting change. Page 5 of 25
6 Fig. 2: Typical ventricularperitonel shunt system with a programmable valve. Page 6 of 25
7 Imaging findings OR Procedure details The Codman Hakim Programmable Valve (Codman/Johnson & Johnson, Raynham, Massachusetts) permits regulation of the opening pressure between 30 and 200 mm H2O. The valve component consists of a hyperdense disk with a notched edge and hyperdense valve. Fig. 3: Radiographic appearance of the Codman Hakim Programmable Valve. Note that a proper radiograph will be taken when the film is shot perpendicular to the plane of the valve with the non-implanted side of the patient's head resting on the plate. The film must be taken in relation to the valve and not the patient's anatomy. References: Reproduced with permission from Codman/Johnson & Johnson. The valve setting is interpreted on the basis of the position of the notch and orientation marker. Current product literature states that patients with an implanted Codman Hakim Programmable Valve can safely undergo MR imaging under the following conditions: 1) static magnetic field of #3T 2) spatial gradient of #720 G/cm and 3) limited radio-frequency energy to a whole-body-averaged specific absorption rate of 3 watts per kilogram for 15 minutes. The manufacturer advises that the setting should checked on a plain radiograph after MR imaging to ensure that no change in opening pressure has occurred. Page 7 of 25
8 Fig. 4: Plain film appearance of the Codman Hakim Programmable Valve References: Department of Radiology, National Hospital Neurology and Neurosurgery, UCLH, London/UK 2012 The PS Medical Strata valve (Medtronic, Minneapolis, Minnesota) is an adjustable flowcontrol valve. The Strata valve has 5 settings or performance level (P/L), ranging from 0.5 to 2.5. Each level corresponds to a range of opening pressures and flow rates. Page 8 of 25
9 Fig. 5: Radiographic appearance of the Cranial Strata valve. References: Department of Radiology, National Hospital Neurology and Neurosurgery, UCLH, London/UK 2012 Generally, a lower performance level corresponds to a lower pressure. The range of opening pressures is between 15 and 170 mm H2O. For example, level 1.5 corresponds to the opening pressure 35 to 95 mm H20 at 0 hyrostatic pressure (patient in the supine position). Multiple models of the Strata valve have been introduced, including the Strata II valve and the Strata small valve. All make use of the same radiographic scheme for setting assessment; the position of a notched disk relative to 2 small dots defines the P/ L setting. Current product literature states that patients with Strata valves may undergo MR imaging by using a static field of #3T but that inadvertent changes of the setting are possible. It advises that the setting be checked after MR imaging to ensure that this has not occurred. Page 9 of 25
10 Fig. 6: Plain Film appearance of the Strata Valve used in lumbar peritoneal shunt. References: Department of Radiology, National Hospital Neurology and Neurosurgery, UCLH, London/UK 2012 Sophysa (Orsay,France) has developed 2 programmable valves, the Sophy and the Polaris. The Polaris valve is a newer valve and it setting is determined by the position of the rotating central rectangular structure relative to a fixed peripheral dot. Page 10 of 25
11 Fig. 7: Polaris valve (Low, Standart, and 2 High Pressure valves) as seen on the plain radiograph. References: Reproduced with permission from Sophysa. Fig. 8: Plain film appearance of the Polaris valve. Note the gravitation unit connected in line with a shunt. Page 11 of 25
12 References: Department of Radiology, National Hospital Neurology and Neurosurgery, UCLH, London/UK 2012 The Polaris valve permits a total of 5 positions; in the most common SPV model (opening pressures between 30 and 200 mm H2O). Product literature states that the Polaris has been tested in MR imaging fields #3T and that no inadvertent setting changes have been seen. The Polaris is also less susceptible to inadvertent setting changes by household magnets. Nevertheless, product literature advises that patients with either a Sophy or Polaris valve have their setting checked immediately after MR imaging. The Miethke progav Programmable Shunt System (Aesculap, Tuttlingen, Germany) is a posture-dependent valve. This means that the opening pressure varies depending on the body position of the patient and consists of 2 components in series, the adjustable unit and the gravitational unit. In order to configure the progav according to the patient s individual needs, the surgeon determines the opening pressure required for both the horizontal and vertical positions. The adjustable unit is a circular structure with a rotating central pointer. Page 12 of 25
13 Fig. 9: Scheme to demonstrate appearance of the progav valve on the plain film. References: Reproduced with permission from Aesculap, Inc. The adjustable unit can be changed to a pressure setting between 0 and 20 cm H2O. The gravitational unit is a cylindrical structure and it function of the gravitational unit is to prevent postural overdrainage. Its opening pressure gradually increases as the patient moves from a supine to an upright position. Multiple models of the gravitational unit are available, each with a different maximal opening pressure. This opening pressure of the gravitational component is determined by the number of rings at the end of the cylinder (0-4 rings) and the size of the component itself (small versus large) In its current product literature, Aesculap recommends checking the setting of the valve after each MR imaging. Page 13 of 25
14 Fig. 10: Plain film appearance of the Sprung reservoir, progav valve and Gradient unit shunt system. References: Department of Radiology, National Hospital Neurology and Neurosurgery, UCLH, London/UK 2012 Some times an additional componet called Sprung Reservoir is connected in line with a valve. The Sprung Reservoir is flushing reservoir for control of the ventricular catheter's patency and the distal share of drainage. Page 14 of 25
15 The Miethke Programmable Shunt Assistant (ProSA) (Aesculap (Tuttlingen,Germany), shunt system has recently been introduced into clinical use. The valve appears simmilar to the progav valve on the plain film. Fig. 11: Scheme to demonstrate appearance of the prosa valve on the plain film. References: Department of Radiology, National Hospital Neurology and Neurosurgery, UCLH, London/UK 2012 Its adjustability is supposed to affect CSF drainage only in the vertical body position. The prosa is an adjustable gravitational valve whose opening pressure automatically adapts to the patient's body position. In the supine position, the opening pressure of the prosa is 0 cmh In this mode, the shunt opening pressure is completely defined by the differential pressure unit. When the patient is in an upright body position, the gravitational unit and Page 15 of 25
16 the differential pressure unit work together, i.e. the opening pressure of the shunt system as a whole is the sum of the differential pressure level and the pressure level set at the gravitational unit. Fig. 12: ProSa valve appearance on the skull radiograph. References: Department of Radiology, National Hospital Neurology and Neurosurgery, UCLH, London/UK 2012 Images for this section: Page 16 of 25
17 Fig. 2: Typical ventricularperitonel shunt system with a programmable valve. Page 17 of 25
18 Fig. 4: Plain film appearance of the Codman Hakim Programmable Valve Page 18 of 25
19 Fig. 5: Radiographic appearance of the Cranial Strata valve. Page 19 of 25
20 Fig. 8: Plain film appearance of the Polaris valve. Note the gravitation unit connected in line with a shunt. Page 20 of 25
21 Fig. 9: Scheme to demonstrate appearance of the progav valve on the plain film. Page 21 of 25
22 Fig. 10: Plain film appearance of the Sprung reservoir, progav valve and Gradient unit shunt system. Page 22 of 25
23 Fig. 12: ProSa valve appearance on the skull radiograph. Page 23 of 25
24 Conclusion Patients who undergo an insertion of the programmable valve often undergo repeated adjustment of the shunt valve to optimize shunt function. Changes in ventricular caliber can be the result of shunt dysfunction, shunt exchange or simply a change in the valve setting. For example, enlargement of the ventricles on serial cross-sectional imaging may be the result of obstruction in the shunt or an intentional increase in the setting of the shunt. Distinguishing between these requires a working knowledge of adjustable valves in current use and an ability to interpret the valve setting on a plain radiograph. The setting codes can be found in the each manufacturer manual which are readily available on-line. This poster provides radiologists and radiographers with a guide to programmable valves so that they may better assist clinicians in evaluating shunt function. It also provides brief MR imaging safety information. All adjustable valves described in this poster are MR imaging#compatible up to 3T. They require prompt setting assessment/readjustment after each MR imaging to correct potential changes in opening pressure induced by magnetic fields to avoid an undetected dengerous change in opening pressure. Personal Information References Greenberg MS. Handbook of Neurosurgery. 6th ed.new York: Thieme Medical Publishers; 2006 Richards H et al. Are adjustable valves effective? Data from the UK Shunt Registry. Cerebrospinal Fluid Research 2007, 4(Suppl 1):S30 doi: / s1-s30 Stein SC, Guo W. Have we made progress in preventing shunt failure? A critical analysis. J Neurosurg Pediatr 2008;1:40-47 GalliaGL, Rigamonti D, Williams MA. The diagnosis and treatment of idiopathic normal pressure hydrocephalus. Nat Clin Pract Neurol 2006;2: Lollis SS, Mamourian AC, Vaccaro TJ et al Programmable CSF Shunt Valves: Radiographic. Identificationand Interpretation. AJNR : Codman. Procedure Guide. Codman Hakim Programmable Valve system for hydrocephalus. VAL _CHPV%20Procedure%20Guide.pdf Accessed January 30,2012 Page 24 of 25
25 7. Sophysa. Polaris adjustable valve: The MRI compatible valve. Accessed January 30, Shellock FG, Habibi R, Knebel J. Programmable CSF shunt valve: in vitro assessment of MR imaging safety at 3T. AJNR Am J Neuroradiol 2006;27: prosa instructions for use. Tuttlingen, Germany: Aesculap english/3_produkte/3_3_prosa/3_3_1_beschreibung/3_3_1_prosa.html Accessed January 30, progav- product description english/3_produkte/3_2_progav/3_2_1_beschreibung/3_2_1_progav.html Accessed January 30, 2012 Page 25 of 25
Radiological evaluation of ventriculoperitoneal shunt systems
Radiological evaluation of ventriculoperitoneal shunt systems Poster No.: C-702 Congress: ECR 2009 Type: Educational Exhibit Topic: Neuro Authors: T. Rodt, C. von Falck, C. Tschan, M. Diensthuber, J. Zajaczek,
More information"Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods."
"Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods." Poster No.: C-1557 Congress: ECR 2014 Type: Authors: Keywords: DOI: Educational Exhibit I.
More informationIdentification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine
Identification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine Poster No.: C-2125 Congress: ECR 2015 Type: Authors: Scientific Exhibit S. patil 1, A. M.
More information64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes
64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes Poster No.: C-051 Congress: ECR 2009 Type: Scientific Exhibit Topic: Abdominal and Gastrointestinal Authors:
More informationCierny-Mader classification of chronic osteomyelitis: Preoperative evaluation with cross-sectional imaging
Cierny-Mader classification of chronic osteomyelitis: Preoperative evaluation with cross-sectional imaging Poster No.: C-590 Congress: ECR 2009 Type: Topic: Educational Exhibit Musculoskeletal Authors:
More informationHigh density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous?
High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous? Poster No.: C-1753 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit B. Y. Lee, H. R. KIM, J. I. Jung,
More informationLong bones manifestations of congenital syphilis
Long bones manifestations of congenital syphilis Poster No.: C-0139 Congress: ECR 2011 Type: Educational Exhibit Authors: T. F. de Souza 1, P. P. Collier 1, E. J. M. Bronzatto 1, G. L. P. Keywords: DOI:
More informationComputed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis
Computed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis Poster No.: C-1887 Congress: ECR 2016 Type: Authors: Keywords: DOI: Educational Exhibit M.
More informationCognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions
Cognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions Poster No.: B-0704 Congress: ECR 2015 Type: Scientific Paper Authors: P. P. van Westerveld, J. Vriesema, J. H. W. van den
More informationDiffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine
Diffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine Poster No.: C-0894 Congress: ECR 2012 Type: Scientific Exhibit
More informationInfluence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions.
Influence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions. Poster No.: C-0599 Congress: ECR 2013 Type: Authors: Keywords:
More informationSingle cold nodule in Graves' disease: benign vs malignant
Single cold nodule in Graves' disease: benign vs malignant Poster No.: C-0073 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper L. I. Sonoda, M. Halim, K. Balan; Cambridge/UK Head and neck,
More informationSeemingly isolated greater trochanter fractures do not exist
Seemingly isolated greater trochanter fractures do not exist Poster No.: B-0950 Congress: ECR 2012 Type: Scientific Paper Authors: D. Dunker, J. H. Göthlin, M. Geijer ; Gothenburg/SE, Lund/SE Keywords:
More informationSlowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time
Slowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time Poster No.: C-208 Congress: ECR 2009 Type: Educational Exhibit Topic: Chest Authors:
More informationArtifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil
Artifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil Poster No.: C-0671 Congress: ECR 2016 Type: Educational Exhibit Authors: H. Patel 1, W. Liu 2, J. DeSanto 2, S. Meagher 2, M. Zagardo 2,
More informationRadiological features of Legionella Pneumophila Pneumonia
Radiological features of Legionella Pneumophila Pneumonia Poster No.: E-0048 Congress: ESTI 2012 Type: Scientific Exhibit Authors: M. Vinciguerra, L. Stefanetti, E. Teti, G. Argentieri, L. G. 1 1 1 1 1
More informationBasic low - field MR imaging of meniscal injuries in children.
Basic low - field MR imaging of meniscal injuries in children. Poster No.: C-2365 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit A. Yakimov, M. Nikonova, E. Prokhorova, D. Vybornov,
More informationValsalva-manoeuvre or prone belly position for computed tomography (CT) scan when an orbita varix is suspected: a single-case study.
Valsalva-manoeuvre or prone belly position for computed tomography (CT) scan when an orbita varix is suspected: a single-case study. Poster No.: C-0512 Congress: ECR 2012 Type: Authors: Keywords: DOI:
More informationImaging the post-operative spine - are we united in where we stand?
Imaging the post-operative spine - are we united in where we stand? Poster No.: C-2424 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit J. Kumaraguru, N. Kandasamy, D. A. Elias, J. M.
More informationSpinal injury is very common in Ireland: 19 per 100,000 (1). It poses a significant disease burden.
MRI in traumatic spinal cord injury: a single national spinal centre experience and study of imaging features with clinical correlation with ASIA score and outcome Poster No.: C-1235 Congress: ECR 2011
More informationUsing diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning
Using diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning Poster No.: C-0016 Congress: ECR 2014 Type: Authors: Keywords:
More informationBI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation
BI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation e-poster: C-118 Congress: ECR 2008 Type: Educational Exhibit Topic: Breast / Ultrasound Authors:
More informationAFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).
Feasibility, image quality and radiation dose of coronary CT angiography (CCTA) in patients with atrial fibrillation using a new generation 256 multi-detector CT (MDCT) Poster No.: C-2378 Congress: ECR
More informationAetiologies of normal CT main pulmonary arterial (PA) measurements in patients with right heart catheter (RHC) confirmed pulmonary hypertension (PH)
Aetiologies of normal CT main pulmonary arterial (PA) measurements in patients with right heart catheter (RHC) confirmed pulmonary hypertension (PH) Poster No.: C-0964 Congress: ECR 2010 Type: Scientific
More informationChronology of normal brain myelination in newborns with MR imaging
Chronology of normal brain myelination in newborns with MR imaging Poster No.: C-0577 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit F. Fernandez Usagre; Sevilla/ES Neuroradiology
More informationIdiopathic dilatation of the pulmonary artery : radiographic and MDCT features in 6 cases
Idiopathic dilatation of the pulmonary artery : radiographic and MDCT features in 6 cases Poster No.: P-0075 Congress: ESTI 2014 Type: Authors: Educational Poster J. J. Woo 1, K. Y. Lee 2, Y. Cho 1, J.
More informationShear Wave Elastography in diagnostics of supraspinatus tendon.
Shear Wave Elastography in diagnostics of supraspinatus tendon. Poster No.: C-2168 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit V. Saltykova; Moscow/RU Musculoskeletal joint, Musculoskeletal
More informationNovel hybrid nitinol ring stent (TIGRIS Gore)placement for atherosclerotic disease in the popliteal artery in 50 limbs
Novel hybrid nitinol ring stent (TIGRIS Gore)placement for atherosclerotic disease in the popliteal artery in 50 limbs Poster No.: C-2437 Congress: ECR 2015 Type: Scientific Exhibit Authors: A. Parthipun,
More informationPopliteal pterygium syndrome
Popliteal pterygium syndrome Poster No.: C-1816 Congress: ECR 2011 Type: Educational Exhibit Authors: L. B. S. Santos, J. L. D. O. Schiavon, O. O. Guimaraes Neto, 1 1 2 3 1 1 C. A. P. Braga, R. S. LEMOS,
More informationComputed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma
Computed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma Poster No.: C-0729 Congress: ECR 2013 Type: Scientific Exhibit Authors: A. Marin, I. Pozek,
More informationPGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification.
PGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification. Poster No.: C-0673 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific
More informationReliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures
Reliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures Poster No.: C-0669 Congress: ECR 2014 Type: Scientific Exhibit Authors: J. Tonak, I. Wobbe, R. L. Duschka,
More informationThe Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities
The Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities Poster No.: C-1229 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit M. Osher 1, A. Pallas
More informationScientific Exhibit Authors: V. Moustakas, E. Karallas, K. Koutsopoulos ; Rodos/GR, 2
Diagnosis of Acute Appendicitis: the role of Color Doppler Ultrasound as first-line imaging method and evaluation of the higher diagnostic performances of CT against its disadvantages. Poster No.: C-0708
More informationDiffusion-weighted imaging and ADC mapping in the differentiation of intraventricular brain tumors
Diffusion-weighted imaging and ADC mapping in the differentiation of intraventricular brain tumors Poster No.: C-2652 Congress: ECR 2010 Type: Educational Exhibit Topic: Neuro Authors: M. Gavrilov, T.
More informationSingle ventricle on cardiac MRI
Single ventricle on cardiac MRI Poster No.: C-0414 Congress: ECR 2011 Type: Scientific Exhibit Authors: F. Secchi, V. G. Nardella, A. Giardino, G. Di Leo, F. 1 2 1 1 1 1 2 Sardanelli ; Milano/IT, Milan/IT
More informationIntratendinous tears of the Achilles tendon - a new pathology? Analysis of a large 4 year cohort.
Intratendinous tears of the Achilles tendon - a new pathology? Analysis of a large 4 year cohort. Poster No.: C-1680 Congress: ECR 2014 Type: Scientific Exhibit Authors: S. Morton, T. Parkes, O. Chan,
More informationThe "whirl sign". Diagnostic accuracy for intestinal volvulus.
The "whirl sign". Diagnostic accuracy for intestinal volvulus. Poster No.: C-0670 Congress: ECR 2014 Type: Scientific Exhibit Authors: M. Pire, M. Marti, A. Borobia, A. Verón; Madrid/ES Keywords: Abdomen,
More informationBreast ultrasound appearances after Mammotome vacuumassisted
Breast ultrasound appearances after Mammotome vacuumassisted biopsy. Poster No.: C-1924 Congress: ECR 2011 Type: Educational Exhibit Authors: R. Patel 1, G. R. Kaplan 2 ; 1 London/UK, 2 Herts/UK Keywords:
More informationScientific Exhibit Authors:
Clinical Audit on Optimization of Radiation Dose from MDCT: Effect on Diagnostic Reference Levels for Brain, Sinus, Cervical Spine, Chest, Abdomen-Pelvis, and Lumbar Spine Examinations and on Nationwide
More informationOligodendroglioma: imaging findings, radio-pathological correlation and evolution
Oligodendroglioma: imaging findings, radio-pathological correlation and evolution Poster No.: C-2104 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit A. Hernandez Castro, M. D. Monedero
More informationComparison of Image quality in temporal bone MRI at 3T using 2D selective RF excitation versus a routine SPACE sequence
Comparison of Image quality in temporal bone MRI at 3T using 2D selective RF excitation versus a routine SPACE sequence Poster No.: C-1065 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit
More informationMarked changes in intervertebral disc morphology in ochronosis of alkaptonuria (AKU): a quantitative analysis of lumbar spine MR images
Marked changes in intervertebral disc morphology in ochronosis of alkaptonuria (AKU): a quantitative analysis of lumbar spine MR images Poster No.: C-2314 Congress: ECR 2011 Type: Authors: Keywords: DOI:
More informationDiffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy
Diffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy Poster No.: C-1785 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific
More informationFeasibility of magnetic resonance elastography using myofascial phantom model
Feasibility of magnetic resonance elastography using myofascial phantom model Poster No.: C-0971 Congress: ECR 2013 Type: Scientific Exhibit Authors: H. J. Kang, J.-S. Yoon, S.-J. Hong, C.-H. Oh, S. H.
More informationA Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT
A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT Poster No.: C-0623 Congress: ECR 2017 Type: Authors: Keywords: DOI:
More informationMedical device adverse incident reporting in interventional radiology
Medical device adverse incident reporting in interventional radiology Poster No.: C-0226 Congress: ECR 2014 Type: Authors: Scientific Exhibit N. PARVIZI 1, I. Robertson 2, R. G. McWilliams 3 ; 1 Oxford/UK,
More informationDoppler ultrasound in the evaluation of chronic venous insufficiency: A step-by-step morphological and hemodynamic review
Doppler ultrasound in the evaluation of chronic venous insufficiency: A step-by-step morphological and hemodynamic review Poster No.: C-3206 Congress: ECR 2010 Type: Educational Exhibit Topic: Vascular
More informationHyperechoic breast lesions can be malignant.
Hyperechoic breast lesions can be malignant. Poster No.: C-0041 Congress: ECR 2015 Type: Educational Exhibit Authors: G. Babu, R. bradley; Edinburgh/UK Keywords: Breast, Ultrasound, Biopsy, Cancer DOI:
More informationComputed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography.
Computed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography. Poster No.: C-1981 Congress: ECR 2012 Type: Scientific Exhibit Authors: M. S. Saltzherr, J. W. van
More informationImaging characterization of renal clear cell carcinoma
Imaging characterization of renal clear cell carcinoma Poster No.: C-0327 Congress: ECR 2011 Type: Educational Exhibit Authors: S. Ballester 1, A. Gaser 2, M. Dotta 1, M. F. CAPPA 1, F. Hammar 1 ; 1 2
More informationEssure Permanent Birth Control Device: Radiological followup results at our center
Essure Permanent Birth Control Device: Radiological followup results at our center Poster No.: C-0212 Congress: ECR 2013 Type: Scientific Exhibit Authors: R. Díaz Aguilera, A. M. Higuera Higuera, V. Palomo
More information3D ultrasound applied to abdominal aortic aneurysm: preliminary evaluation of diameter measurement accuracy
3D ultrasound applied to abdominal aortic aneurysm: preliminary evaluation of diameter measurement accuracy Poster No.: C-0493 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper A. LONG
More informationPurpose. Methods and Materials. Results
Prevalence and significance of hypoattenuating hepatic lesions deemed too small to characterise: How are we following up these lesions and what are the outcomes? Poster No.: C-014 Congress: ECR 2009 Type:
More informationBolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography
Bolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography Poster No.: C-1342 Congress: ECR 2013 Type: Scientific Exhibit Authors:
More informationHow to plan a Zenith AAA stent-graft from a CTA: Basic measurements and concepts explained
How to plan a Zenith AAA stent-graft from a CTA: Basic measurements and concepts explained Poster No.: C-3077 Congress: ECR 2010 Type: Educational Exhibit Topic: Vascular Authors: D. V. Thomas; Northampton/UK
More informationReporting of Spinal Fractures
Reporting of Spinal Fractures Poster No.: P-0026 Congress: ESSR 2014 Type: Authors: Keywords: DOI: Scientific Poster H. Al-Chalabi, C. Groves; Bradford/UK Osteoporosis, Structured reporting, Observer performance,
More informationDynamic CT Assessment of Distal Radioulnar Instability
Dynamic CT Assessment of Distal Radioulnar Instability Poster No.: P-0114 Congress: ESSR 2016 Type: Educational Poster Authors: S. Dumonteil, M. A. Shah, A. Srikanthan, V. Ejindu, N. Papadakos; London/UK
More informationA time-honored but almost forgotten sign of COPD: sabersheath trachea as a marker of severe airflow obstruction
A time-honored but almost forgotten sign of COPD: sabersheath trachea as a marker of severe airflow obstruction Poster No.: C-0958 Congress: ECR 2013 Type: Scientific Exhibit Authors: F. Ciccarese, A.
More informationCT-guided percutaneous intraspinal needle aspiration for the diagnosis and treatment of epidural collections
CT-guided percutaneous intraspinal needle aspiration for the diagnosis and treatment of epidural collections Poster No.: P-0064 Congress: ESSR 2013 Type: Scientific Exhibit Authors: G. Petrocheilou, I.
More informationThe imaging evaluation of breast implants
The imaging evaluation of breast implants Poster No.: C-0654 Congress: ECR 2014 Type: Authors: Keywords: DOI: Educational Exhibit N. Rotaru; Chisinau/MD Neoplasia, Education and training, Cancer, Imaging
More informationUser Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)
User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT) Poster No.: C-0756 Congress: ECR 2014 Type: Educational Exhibit Authors: J. Ukkonen, J. Asp; Helsinki/FI Keywords: Education
More informationTreatment options for endoleaks: stents, embolizations and conversions
Treatment options for endoleaks: stents, embolizations and conversions Poster No.: C-0861 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit G. Lombardi; napoli/it Arteries / Aorta, Abdomen,
More informationMagic angle artifact in MRI of the patellar ligament: preliminary comparison between conventional and weightbearing
Magic angle artifact in MRI of the patellar ligament: preliminary comparison between conventional and weightbearing MRI Poster No.: C-1017 Congress: ECR 2012 Type: Scientific Exhibit Authors: R. Piccazzo
More informationA pictorial essay depicting CT and MR characteristic of adrenal pathologies: Indian study
A pictorial essay depicting CT and MR characteristic of adrenal pathologies: Indian study Poster No.: C-0703 Congress: ECR 2011 Type: Educational Exhibit Authors: A. J. B. Baxi, K. L. Tourani, N. R. Thanugonda,
More informationContrast enhancement of the right ventricle during coronary CTA: is it necessary?
Contrast enhancement of the right ventricle during coronary CTA: is it necessary? Poster No.: C-1545 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit M. Kok, C. Mihl, B. Kietselaer,
More informationBrainstem diffuse gliomas: radiologic findings.
Brainstem diffuse gliomas: radiologic findings. Poster No.: C-2220 Congress: ECR 2013 Type: Educational Exhibit Authors: E. GARCIA MARTINEZ 1, D. H. Jiménez 1, L. Navarro Vilar 2, C. P. Fernandez Ruiz
More informationQuantitative imaging of hepatic cirrhosis on abdominal CT images
Quantitative imaging of hepatic cirrhosis on abdominal CT images Poster No.: C-0556 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit S. Kido, A. Nakamura, Y. Hirano; Ube/JP Cirrhosis,
More informationComparison of MRI and ultrasound based liver volumetry in iron overload diseases
Comparison of MRI and ultrasound based liver volumetry in iron overload diseases Poster No.: C-0913 Congress: ECR 2015 Type: Scientific Exhibit Authors: S. Keller, B. P. Schönnagel, P. Nielsen, M. T. Brehmer,
More informationNICE Guidelines for C-Spine Imaging: Real Life Impact
NICE Guidelines for C-Spine Imaging: Real Life Impact Poster No.: C-1367 Congress: ECR 2017 Type: Scientific Exhibit Authors: D. Weinberg, I. Djoukhadar, G. Potter; Salford/UK Keywords: Trauma, Audit and
More informationMRI evaluation of TMJ condylar angulations
MRI evaluation of TMJ condylar angulations Poster No.: C-2272 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Musculoskeletal M. Pregarz 1, C. Bodin 2 ; 1 Peschiera del Garda/IT,
More informationTranslent CT hyperattenuation after intraarterial thrombolysis in stroke. Contrast extravasation or hemorrhage
Translent CT hyperattenuation after intraarterial thrombolysis in stroke. Contrast extravasation or hemorrhage Poster No.: C-0053 Congress: ECR 2013 Type: Scientific Exhibit Authors: A. Losa Palacios,
More informationPreliminary experience of phlebographic studies in patients with multiple sclerosis and chronic cerebrospinal venous insufficiency
Preliminary experience of phlebographic studies in patients with multiple sclerosis and chronic cerebrospinal venous insufficiency Poster No.: C-1715 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific
More informationMRI in Patients with Forefoot Pain Involving the Metatarsal Region
MRI in Patients with Forefoot Pain Involving the Metatarsal Region Poster No.: C-0151 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit R. Vukojevi#, M. Mustapic, D. Marjan; Zagreb/HR
More informationPresence and severity of carotid siphon calcification on computed tomography images in mild cognitive impairment
Presence and severity of carotid siphon calcification on computed tomography images in mild cognitive impairment Poster No.: C-2117 Congress: ECR 2013 Type: Scientific Exhibit Authors: S. Rutkauskas, S.
More informationPneumo-esophageal 64-MDCT technique for gastric cancer evaluation
Pneumo-esophageal 64-MDCT technique for gastric cancer evaluation Poster No.: C-1627 Congress: ECR 2010 Type: Scientific Exhibit Topic: GI Tract Authors: M. Ulla, E. Gentile, E. Levy, D. Cavadas, J. Ithurralde
More informationBiliary tree dilation - and now what?
Biliary tree dilation - and now what? Poster No.: C-1767 Congress: ECR 2012 Type: Educational Exhibit Authors: I. Ferreira, A. B. Ramos, S. Magalhães, M. Certo; Porto/PT Keywords: Pathology, Diagnostic
More informationCavitary lung lesion: Two different diagnosis with similar appearence
Cavitary lung lesion: Two different diagnosis with similar appearence Poster No.: P-0043 Congress: ESTI 2015 Type: Educational Poster Authors: M. Yesildag, H. Kalkan, K. Ödev; Konya/TR Keywords: Infection,
More informationAcute pelvic pain in female patient: Clinical and Radiological evaluation
Acute pelvic pain in female patient: Clinical and Radiological evaluation Poster No.: C-0909 Congress: ECR 2014 Type: Authors: Keywords: DOI: Educational Exhibit N. Ramesh 1, T. Simelane 2 ; 1 Portlaoise/IE,
More informationAcute pelvic pain in female patient: Clinical and Radiological evaluation
Acute pelvic pain in female patient: Clinical and Radiological evaluation Poster No.: C-0909 Congress: ECR 2014 Type: Authors: Keywords: DOI: Educational Exhibit N. Ramesh 1, T. Simelane 2 ; 1 Portlaoise/IE,
More informationA retrospective audit of General Practitioner (GP) referrals for musculoskeletal radiographs.
A retrospective audit of General Practitioner (GP) referrals for musculoskeletal radiographs. Poster No.: C-1902 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit J. Jacob 1, H. Thampy
More informationComputed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation.
Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation. Award: Poster No.: E-0004 Certificate of Merit Congress:
More informationPercutaneous treatment of lumbar disc herniation with gelified ethanol/ a preliminary study
Percutaneous treatment of lumbar disc herniation with gelified ethanol/ a preliminary study Poster No.: C-0223 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit M. Runge; Besançon/FR
More informationMonophasic versus biphasic contrast application in CT of patients with head and neck tumour
Monophasic versus biphasic contrast application in CT of patients with head and neck tumour Poster No.: C-3331 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Head and Neck G.
More informationTissue characterisation, Cancer, Quality assurance /ecr2015/B-0553
Role of DWI at prostatic lesions at 3T-MRI in the discrimination of grading: correlation of imaging, quantitative analysis and pathology at 189 MR-guided prostate biopsies Poster No.: B-0553 Congress:
More informationResuscitation lateral cervical spine X-ray (LSCX): A useful mandatory screening tool in acute trauma?
Resuscitation lateral cervical spine X-ray (LSCX): A useful mandatory screening tool in acute trauma? Poster No.: C-2397 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Musculoskeletal
More informationCryoplasty versus conventional angioplasty in peripheral arterial disease: 3 year analysis of reintervention free survival by treatment received.
Cryoplasty versus conventional angioplasty in peripheral arterial disease: 3 year analysis of reintervention free survival by treatment received. Poster No.: C-0343 Congress: ECR 2011 Type: Scientific
More informationRadiological Investigation of Renal Colic in an Emergency Department of a Teaching Hospital
Radiological Investigation of Renal Colic in an Emergency Department of a Teaching Hospital Poster No.: C-0892 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit A. Koo; Leeds, West Yorkshire/UK
More informationCT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience
CT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience Poster No.: C-0097 Congress: ECR 2016 Type: Scientific Exhibit Authors: A. Casarin, G. Rech, C. Cicero, A.
More informationLumbosacral Transitional Vertebrae
Lumbosacral Transitional Vertebrae Poster No.: C-073 Congress: ECR 206 Type: Educational Exhibit Authors: M. Mustapic, R. Vukojevi#, M. Gulin, D. Marjan, I. Boric ; 2 2 Zagreb/HR, Zabok/HR Keywords: Congenital,
More informationMR imaging the post operative spine - What to expect!
MR imaging the post operative spine - What to expect! Poster No.: C-2334 Congress: ECR 2012 Type: Educational Exhibit Authors: A. Jain, M. Paravasthu, M. Bhojak, K. Das ; Warrington/UK, 1 1 1 2 1 2 Liverpool/UK
More informationS. Inagawa, N. Yoshimura, Y. Ito; Niigata/JP spinal sacral areteriovenous fistulae, CTA, MRA /ecr2010/C-2581
Localization of sacral spinal arteriovenous fistulae in reference to the dural structure with CTA and MRA of high spatial resolution: A pictorial essay Poster No.: C-2581 Congress: ECR 2010 Type: Educational
More informationRole of ultrasound in the evaluation of the ileocecal valve
Role of ultrasound in the evaluation of the ileocecal valve Poster No.: C-1581 Congress: ECR 2010 Type: Scientific Exhibit Topic: GI Tract Authors: M. Mohammed, M. Hussain, U. Momin, S. Lakhtakia, N. D.
More informationHigh frequency US of the temporomandibualar joint (TMJ) - practical guide
High frequency US of the temporomandibualar joint (TMJ) - practical guide Poster No.: C-2352 Congress: ECR 2014 Type: Educational Exhibit Authors: M. Lasecki, C. M. Olchowy, K. Kaczorowski, J. S#onina,
More informationIs ascites a sensible predictive sign of peritoneal involvement in patients with ovarian carcinoma?: our experience with FDG-PET/CT
Is ascites a sensible predictive sign of peritoneal involvement in patients with ovarian carcinoma?: our experience with FDG-PET/CT Poster No.: C-1019 Congress: ECR 2013 Type: Scientific Exhibit Authors:
More informationInstitutional review board approval was obtained prior to the start of this study.
Lower Limb Alignment and Length Measurements - Comparison of Computed Tomography, Upright Full-Length Conventional Radiography and Upright Biplanar Linear-Low Dose X-ray Scanner Poster No.: C-1382 Congress:
More informationMR-guided prostatic biopsy at 3T: the role of PI-RADS-score: a histopahologic-radiologic correlation
MR-guided prostatic biopsy at 3T: the role of PI-RADS-score: a histopahologic-radiologic correlation Poster No.: B-0969 Congress: ECR 2015 Type: Scientific Paper Authors: A. Malich; Nordhausen/DE Keywords:
More informationUltrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head
Ultrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head Poster No.: C-1430 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E. Fisenko, N. Vetsheva, E. Pershina;
More informationA pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography.
A pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography. Poster No.: C-1787 Congress: ECR 2012 Type: Educational Exhibit Authors: N. Ahmed 1, G. Avery
More informationFrequency of positive patients to eco-fast and later CT in major abdominal trauma: our experience
Frequency of positive patients to eco-fast and later CT in major abdominal trauma: our experience Poster No.: C-0155 Congress: ECR 2012 Type: Scientific Exhibit Authors: G. Posillico Keywords: Abdomen,
More information