Despite its benefit over balloon angioplasty in patients

Size: px
Start display at page:

Download "Despite its benefit over balloon angioplasty in patients"

Transcription

1 Coronary Heart Disease Quantitative Assessment of Angiographic Restenosis After Sirolimus-Eluting Stent Implantation in Native Coronary Arteries Jeffrey J. Popma, MD; Martin B. Leon, MD; Jeffrey W. Moses, MD; David R. Holmes, Jr, MD; Nicholas Cox, MBBS; Michelle Fitzpatrick, RN; John Douglas, MD; Charles Lambert, MD; Michael Mooney, MD; Steven Yakubov, MD; Richard E. Kuntz, MD, MSc; for the SIRIUS Investigators Background Sirolimus-eluting stents (SESs) reduce angiographic restenosis in patients with focal, native coronary artery stenoses. This study evaluated the usefulness of SESs in complex native-vessel lesions at high risk for restenosis. Methods and Results Angiographic follow-up at 240 days was obtained in 701 patients with long (15- to 25-mm) lesions in small-diameter (2.5- to 3.5-mm) native vessels who were randomly assigned to treatment with SESs or bare-metal stents (BMSs) in the SIRIUS trial. Quantitative angiographic measurements of minimal lumen diameter and percent diameter stenosis were obtained within the treated segment, within the stent, and within its 5-mm proximal and distal edges. Patients treated with SESs had lower rates of binary ( 50% diameter stenosis) angiographic restenosis within the segment (8.9% versus 36.3% with the BMS; P 0.001) and within the stent (3.2% versus 35.4% with the BMS; P 0.001). SESs were associated with significantly less late lumen loss within the treated segment, within the stent, and within its 5-mm proximal and distal edges (all P 0.001). The reduction of restenosis with the SES was consistent in patients at risk for restenosis, including those with small vessels, long lesions, and diabetes mellitus. The frequency of late aneurysms was similar in the 2 groups. Conclusions Compared with BMSs, SESs reduced angiographic late lumen loss within the stent and its adjacent 5-mm margins in patients with complex native-vessel lesions. (Circulation. 2004;110: ) Key Words: stents sirolimus restenosis coronary disease Despite its benefit over balloon angioplasty in patients with native-vessel coronary artery disease, 1,2 coronary stenting using balloon-expandable stents is limited by the occurrence of angiographic restenosis in 10% to 35% of patients, highest in patients with smaller vessels, longer lesions, and diabetes mellitus. 3 5 The stent-versus-stent equivalency trials have shown similarities in angiographic restenosis rates with a number of balloon-expandable baremetal stent (BMS) designs with a consistent degree of late lumen loss (0.80 to 1.05 mm), presumably because of intimal hyperplasia within the stent and its margins. 6 8 Only 1 coiled stent had more restenosis than the Palmaz-Schatz stent, owing to both immediate recoil and late lumen loss within the coiled stent. 8 Excessive late lumen loss occurs most often within the central portion of the stent, 9 although, in some cases, a diffuse pattern of stent develops within the stents and extends into its nonstented margins. 10 Isolated margin renarrowing, particularly involving only the proximal or distal edge of the stent, is an uncommon cause of stent restenosis with the BMS. Marked reductions in angiographic restenosis have been shown in patients with focal lesions undergoing stent implantation by use of a single balloon-expandable sirolimus-eluting stent (SES). 11,12 One randomized trial of 238 patients found no cases of binary angiographic restenosis ( 50% diameter stenosis) 6 months after the procedure in patients treated with the SES, compared with a 26.6% restenosis rate in patients undergoing BMS implantation. 11 This benefit was because of a near obliteration of late lumen loss ( 0.01 mm) caused by intimal hyperplasia within the SES by use of both angiographic and intravascular ultrasound measurements. 11 These effects have persisted up to 2 years after the index procedure. 12 The purpose of this study was to compare the angiographic outcomes between SESs and BMSs in patients with complex Received September 24, 2003; revision received August 2, 2004; accepted August 11, From the Departments of Internal Medicine (Cardiovascular Division) of Brigham and Women s Hospital, Boston, Mass (J.J.P., N.C., M.F., R.E.K.); Lenox Hill Hospital, New York, NY (M.B.L., J.W.M.); Mayo Clinic, Rochester, Minn (D.R.H.); Emory University Hospital, Atlanta, Ga (J.D.); Health First Institute, Melbourne, Fla (C.L.); Abbott Northwestern Hospital, Minneapolis, Minn (M.M.); Riverside Methodist Hospital, Columbus, Ohio (S.Y.); and the Harvard Clinical Research Institute, Boston, Mass (R.E.K.). The online-only Data Supplement, which contains an Appendix, can be found with this article at Correspondence to Jeffrey J. Popma, MD, Director, Interventional Cardiology, Brigham and Women s Hospital, 75 Francis St, Boston, MA jpopma@partners.org 2004 American Heart Association, Inc. Circulation is available at DOI: /01.CIR B 3773

2 3774 Circulation December 21/28, 2004 lesions at high risk of restenosis, with regard to (1) the lumen changes within the stent and its 5-mm nonstented edges; (2) the degree and incidence of angiographic renarrowing in patients with smaller vessels and longer lesions; (3) the pattern of angiographic restenosis; and (4) the occurrence of late untoward angiographic findings, including the frequency of aneurysms. Methods Patient Selection A substudy of 850 patients was prespecified as the angiographic follow-up cohort within the 1101-patient SIRIUS trial, a 53-center, randomized, double-blind study comparing the angiographic and clinical outcomes of patients treated with an SES or BMS for complex native-vessel coronary disease 13 (see Data Supplement Appendix). Lesions were suitable for inclusion in the study if they were between 15 and 30 mm in length and located in a visually estimated reference vessel between 2.5 and 3.5 mm in diameter. Patients were excluded with evidence of a recent ( 24 hours) myocardial infarction, markedly reduced left ventricular function (left ventricular ejection fraction 25%), a totally occluded target lesion, or a target lesion in an ostial or bifurcation location. All investigational sites received approval from their local hospital Institutional Review Boards. The primary objective of this substudy was to contrast angiographic restenosis parameters in patients treated with an SES or BMS 8 months after the procedure. Coronary Stent Procedure After pretreatment with oral aspirin, clopidogrel, and intravenous heparin, patients were randomized to undergo implantation of either a bare-metal (uncoated) Bx velocity Balloon-Expandable BMS or SES (Cordis Corp, Johnson & Johnson). Stents were available in 8- and 18-mm lengths and in 2.5-, 3.0-, and 3.5-mm diameters. The concentration of sirolimus on the SES was 140 g/cm 2. Controlled release of sirolimus was facilitated by a polymeric matrix coating contained on the stainless steel stent that allowed elution of the drug up to 45 days after implantation. Predilatation with a conventional balloon was performed before stent implantation and postdilatation after the procedure was used to obtain a 0% visual residual stenosis within the stent. Angiographic success was defined as a final vessel diameter stenosis 50%. Angiographic Analysis Standard image acquisition was performed at the clinical sites using 2 or more angiographic projections of the stenosis, intracoronary nitroglycerin to provide maximum coronary vasodilation, and repetition of identical angiographic projections of the lesion at the time of follow-up angiography. Cineangiograms were then forwarded to the Brigham and Women s Hospital Angiographic Core Laboratory for review by observers who were blinded to the treatment assignment. Repeat angiography was planned at days in the first 850 randomized patients. All procedural and follow-up angiograms were reviewed by use of standard morphological criteria. 14,15 Lesion length was defined as the axial extent of the lesion that contained a shoulder-to-shoulder lumen reduction by 20%. Coronary dissections were assessed using the National Heart Lung and Blood Institute criteria. 16 Restenosis patterns were qualitatively assessed using the Mehran classification system. 10 Coronary aneurysms were defined as a maximum lumen diameter within the treatment zone that was 1.2 times larger than the average reference diameter of the vessel. With the contrast-filled injection catheter as the calibration source, quantitative angiographic analysis was performed by use of a validated automated edge-detection algorithm (Medis CMS). 17 Selected images for analysis were identified by use of angiographic projections that demonstrated the stenosis in an unforeshortened view, minimized the degree of vessel overlap, and displayed the stenosis in its sharpest and tightest view. A 5-mm segment of reference diameter proximal and distal to the stenosis was used to calculate the average reference vessel diameter at baseline, after stent implantation, and at follow-up; side branches and other anatomic landmarks were used to identify and maintain the consistency of the measurement length during the follow-up period. Minimal lumen diameters (MLDs) were measured at these same time points within the stent (in-stent analysis), within the 5-mm proximal and distal edges of the stent, and within the segment between the proximal and distal reference vessel (in-segment analysis) (Figure 1, A and B). Total occlusions were assigned an MLD 0 mm and a 100% diameter stenosis. Angiographic follow-up was performed 6 to 9 months after the index procedure unless earlier angiography was required clinically. Binary angiographic restenosis was defined as the incidence of percent diameter stenosis 50% at the qualifying angiographic follow-up, performed from as early as 1 month for symptomatic patients to up to 9 months for asymptomatic patients. Angiographic percent diameter stenosis was defined as [1 (MLD reference vessel diameter)] 100. Acute gain was defined as the MLD immediately after the procedure minus the MLD before the procedure, and late loss was defined as the MLD immediately after the procedure minus the MLD at 8-month follow-up. The loss index was defined as the slope of the relationship between late loss and acute gain. Statistical Analysis A sample size of 850 patients provided 80% statistical power to detect a 0.2-mm difference in the follow-up MLD between the SESand BMS-treated patients with a 5% (2-sided) false-positive rate. After blinded randomization, a total of 26 patients (3% of the angiographic substudy) were deregistered and did not receive the assigned treatment. Deregistration from the study was because of either unavailability of the assigned stent at the site or discovery of exclusion criteria prohibiting further trial participation apparent only after the pretreatment angiograms. The final patient cohort included 824 patients and 824 lesions; a per-lesion analysis was performed for all angiographic parameters in this report. Angiographic restudy was obtained in 701 patients (85.1%), including 350 (85.6%) in SES-treated patients and 351 (84.6%) in BMS-treated patients. All analyses were performed on a per-lesion basis. Categorical data are expressed as rates or proportions, and continuous variables are expressed as mean SD. Binary variables were compared by use of 2 and continuous variables were compared by use of the Student s t test. The angiographic data set was used to build a multivariable model of 8-month angiographic restenosis. All analyses were performed by use of SAS (Version 8.0). A 2-sided probability value of P 0.05 was considered significant. Results Baseline Clinical Characteristics The average age was 61.7 years in patients assigned to treatment with an SES and 62.6 years in patients assigned to treatment with a BMS. Diabetes was present in 25.7% of patients assigned to treatment with the SES and 29.5% of patients assigned to treatment with a BMS. In 27.7% of patients, 2 or more stents were used, with an average of 1.4 stents per patient. The average stent length was 21.4 mm, and the stent length to lesion length ratio was 1.6 to 1. Angiographic Findings Baseline angiographic findings were similar in the 2 groups (see Data Supplement Table I). The left anterior descending coronary artery was the most commonly treated vessel (42.5%), and the majority of lesions (89.4%) were found in the proximal and mid segments of the vessel. The mean lesion length was 14.4 mm, and the American College of Cardiology/American Heart Association lesion complexity was Class

3 Popma et al Angiographic Findings in SIRIUS 3775 Figure 1. Quantitative angiographic analysis. A, A type II pattern of central restenosis is shown, with axial stent length and its 5-mm edge measurement obtained by use of CMS automated edge detection system. B, A type Ib pattern of stent restenosis with preserved lumen within the axial length of the stent and a restenosis within the proximal margin of the stent. B or C in 93.4% of lesions. Angiographic success rates were not different between the 2 arms (99%). There were no differences between the 2 arms in the incidence of abrupt closure (0.7%), dissection (1%), or occurrence of no reflow ( 1%). Quantitative changes in the lumen dimension within the treated arterial segment, within the stent, and within its 5-mm margins are found in Table 1. The mean reference vessel diameter was mm in the SES group and in the BMS group. SESs were associated with significantly less late lumen loss within the segment ( versus mm with BMS; P 0.001), within the stent ( versus mm with BMS; P 0.001), and within its 5-mm proximal ( versus mm with BMS; P 0.001) and distal ( versus mm with BMS; P 0.001) edges 8 months after stent placement (Figure 2). Relative changes in lumen dimensions are found in Table 2. Patients treated with SESs had lower rates of binary ( 50% diameter stenosis) angiographic restenosis within the segment (8.9% versus 36.3% with BMS; P 0.001), within the stent (3.2% versus 35.4% with BMS; P 0.001), and at its

4 3776 Circulation December 21/28, 2004 TABLE 1. Quantitative Changes In Lumen Dimensions in the SIRIUS Angiographic Substudy Variable Sirolimus Control P RVD, mm Baseline Final Follow-up Minimum lumen diameter, mm Baseline MLD Within the segment Final MLD Acute lumen gain Follow-up MLD Late lumen loss Proximal 5-mm edge Final MLD Follow-up MLD Late lumen loss Within the stent Final MLD Acute lumen gain Follow-up MLD Late lumen loss Distal 5-mm edge Final MLD Follow-up MLD Late lumen loss RVD indicates reference vessel diameter; MLD, minimal lumen diameter. Baseline and final angiographic measurements include 409 patients in the sirolimus group and 415 patients in the bare-metal stent group. Follow-up measurements included 350 sirolimus-treated patients and 351 patients treated with the bare metal stent. distal edge (2.0% versus 7.5% with BMS; P 0.001). The in-segment binary restenosis rates were significantly reduced in patients treated with the SES in all clinical and angiographic subsets (see Data Supplement Figure). Figure 2. Late lumen loss within stent and within its 5-mm proximal and distal nonstent margins. Dark shade indicates SES and light shade indicates BMS. TABLE 2. Relative Changes in Lumen Dimensions in the SIRIUS Angiographic Substudy Variable, % Diameter Stenosis Sirolimus (n 409) Control (n 415) P Baseline Within the segment Final % stenosis Follow-up % stenosis Restenosis rate Proximal 5-mm edge Final % stenosis Follow-up % stenosis Restenosis rate Within the stent Final % stenosis Follow-up % stenosis Restenosis rate Distal 5-mm edge Final % stenosis Follow-up % stenosis Restenosis rate Baseline and final angiographic measurements include 409 patients in the sirolimus group and 415 patients in the bare-metal stent group. Follow-up measurements included 350 sirolimus-treated patients and 351 patients treated with the bare metal stent. Different patterns of lumen renarrowing were found in patients who developed restenosis after treatment with the SES (n 31) compared with those treated with the BMS (n 128) (Table 3). In patients who developed restenosis, a pattern of focal margin restenosis (type 1B) was found in 61.3% of restenotic patients treated with the SES but was found in only 14.1% of restenotic patients treated with the BMS (P 0.001). Conversely, a diffuse pattern of restenosis (type II IV) was found more often in patients treated with the BMS (57.0%) than in those treated with the SES (16.2%; P 0.001). The in-stent restenosis length was also signifi- TABLE 3. Angiographic Patterns of Restenosis in the SIRIUS Angiographic Substudy Variable Sirolimus, n (%) Control, n (%) P In-stent restenosis pattern n 31 n 128 Focal 1A 1 (3.2) 0 (0) B 19 (61.3) 18 (14.1) C 3 (9.7) 33 (25.8) D 3 (9.7) 4 (3.1) Diffuse II 1 (3.2) 46 (35.9) III 2 (6.5%) 17 (13.3) IV (total occlusion) 2 (6.5) 10 (7.8) 1.00 ISR length, mm Aneurysm 2 (0.6) 4 (1.1) In-stent restenosis (ISR) patterns classified using the Mehran criteria. 10

5 Popma et al Angiographic Findings in SIRIUS 3777 TABLE 4. Quantitative Angiographic Findings Categorized by Vessel Size Small Vessel Medium Vessel Large Vessel Sirolimus Control P Sirolimus Control P Sirolimus Control P RVD, mm Lesion length, mm Within segment Final MLD, mm Follow-up MLD, mm Late lumen loss, mm Restenosis rate, % 17.6% 42.7% % 36.0% % 30.5% Within stent Final MLD, mm Follow-up MLD, mm Late lumen loss, mm Restenosis rate, % 5.0% 41.8% % 35.1% % 29.7% RVD indicates reference vessel diameter; MLD, minimal lumen diameter. Baseline and final angiographic measurements include 409 patients in the sirolimus group and 415 patients in the bare-metal stent group. Follow-up measurements included 350 sirolimus-treated patients and 351 patients treated with the bare metal stent. cantly shorter in patients treated with a SES (9.1 mm) than in those treated with the BMS (14.8; P 0.001). Aneurysms were uncommon ( 1%) and not different between the 2 groups. Relationship to Vessel Size and Lesion Length Late angiographic findings categorized by the baseline reference vessel terciles are found in Table 4. In patients treated with the BMS, in-segment binary restenosis rate was lowest in the largest vessel size tercile (30.5% versus 36.0% in the middle tercile versus 42.7% in the smallest vessel tercile). In patients treated with the SES, the in-segment binary angiographic restenosis rate was also lowest in patients in the largest vessel tercile (1.9% versus 6.6% in the middle vessel tercile and 17.6% in the smallest vessel tercile). For each tercile of reference vessel diameter, there was a significant (P 0.001) reduction in binary restenosis within the segment and within the stent. In patients treated with the SES, late lumen loss within the axial length of the stent was similar with all vessel sizes, but more late lumen loss occurred within the segment in patients treated with smaller vessels. Lesion length was a less important determinant of angiographic restenosis (see Data Supplement Table II). In patients treated with the BMS, the in-segment binary restenosis rate was 37.9% in the longest lesion length tercile, whereas it was 35.0% in the middle tercile versus 36.7% in the shortest lesion length tercile. In patients treated with the SES, the in-segment binary angiographic restenosis rate was 12.1% in patients in the longest lesion tercile, compared with 5.8% in the middle vessel tercile and 8.5% in the shortest lesion length tercile. For each tercile of lesion length, however, there was a 68% to 83% reduction in binary angiographic restenosis associated with the use of the SES within the treated segment (P 0.001) and an 81% to 97% reduction in binary restenosis associated with the use of the SES within the stent (P 0.001). Multivariable Predictors of Angiographic Restenosis Multivariable predictors of higher binary in-segment restenosis rates included treatment with the BMS (P 0.001), the presence of diabetes mellitus (P 0.001), a smaller baseline (P 0.036) or final in-stent (P 0.001) MLD, and longer total stent lengths (P 0.039). The effect of these and other known risk factors for restenosis in patients assigned to either the SES or BMS is demonstrated in the nomogram of angiographic binary restenosis (Table 5), which estimates resteno- TABLE 5. Predicted In-Segment Restenosis Rates in the SIRIUS Angiographic Substudy Lesion Length, mm Reference Vessel Size, mm Standard stent Nondiabetics Diabetics Sirolimus-eluting stent Nondiabetics Diabetics

6 3778 Circulation December 21/28, 2004 sis rates on the basis of the multivariable models and levels of the risk factor chosen by tercile distribution cutoff points. Discussion This study demonstrates that compared with treatment using a BMS, treatment with the SES reduced all angiographic parameters of restenosis in patients with complex nativevessel disease at high risk for restenosis. A 91% reduction of restenosis was found within the stent, and a 75% reduction was found within the treated arterial segment that included the 5-mm proximal and distal nonstented margins. The effect on restenosis was presumably a result of a profound inhibition of intimal hyperplasia with the SES (0.17 mm) compared with the BMS (1.0 mm) (P 0.001). The reduction in intimal hyperplasia with the SES was also shown in the smallest vessels and longest lesions. Focal restenosis at the stent margin was the most common cause of failure in patients treated with the SES, whereas central restenosis was the most common cause of failure in patients treated with the BMS. There were no late deleterious effects associated with SES placement, such as the occurrence of late total occlusion or the formation of late aneurysms within the vessel. Restenosis After BMS Implantation Conventional angiography cannot adequately assess the relative degree of late lumen loss because of intimal hyperplasia or arterial remodeling after balloon angioplasty or directional atherectomy, 18 but angiography can estimate the magnitude of intimal hyperplasia after stent implantation, because one of the benefits of stenting is the near elimination of late arterial remodeling. A number of angiographic studies have demonstrated a relatively fixed degree of late lumen loss in patients undergoing BMS implantation, ranging from 0.80 to 1.05 mm. 6 8 A fixed amount of late lumen loss has an even larger effect on restenosis in smaller vessels. 19 Angiographic studies in patients treated with BMS have reported binary restenosis rates within the stent and within the treated arterial segment that included its nonstented margins. 6,7 Although previous BMS stent studies did not examine the specific lumen changes at the proximal or distal nonstent margins, intravascular ultrasound studies have shown that balloon injury proximal and distal to the stent may result in lumen renarrowing resulting from arterial remodeling at the edges of the stent. 9 Studies evaluating angiographic outcomes in patients undergoing radiation brachytherapy or using radioactive stents have also shown that proliferation at the stent edges may be associated with higher-than-expected restenosis rates at the stent margins ( candy wrapper effect ) because of an untoward effect of low-dose radiation at the stent margins. 20,21 No previous studies have evaluated these potential effects at the margins of the SES. Previous Restenosis Studies Using SES Sirolimus is a naturally occurring macrolide antibiotic and antiinflammatory agent that inhibits smooth muscle cell replication by interrupting the G 1 phase in the cell cycle that prevents the downregulation of p27 before DNA replication, in addition to its antiinflammatory effects. 22 A first-inhumans study of 45 patients treated with the SES showed a profound reduction in late lumen loss within the stent, an effect that persisted up to 24 months after the procedure. 12,23 A subsequent randomized trial confirmed the effect of SESs in a larger series of patients with focal lesions (average lesion length, 9.6 mm), with a reported 0.01-mm late lumen loss; there were no cases of angiographic restenosis in patients treated with the SES compared with 26.6% in patients treated with a BMS. 11 These results were confirmed with serial intravascular ultrasound, which showed a marked reduction in neointimal volume in patients treated with the SES. 24 Present Study The results of this study demonstrate that angiographic restenosis was markedly reduced in patients with complex native-vessel disease by use of the SES. A 91% reduction in binary angiographic restenosis was shown within the stent, and a 75% reduction in binary angiographic restenosis was demonstrated within the treated segment, which included the nonstented margins. The beneficial effect of the SES on binary angiographic restenosis was attributed to reductions in late lumen loss within the SES (0.17 mm) compared with the BMS (1.00 mm), a favorable effect that extended to its proximal and distal 5-mm nonstented margins (Figure 2). These angiographic effects on late lumen loss were presumably attributable to reductions in intimal hyperplasia because of the sustained release of sirolimus into the vessel wall after stent implantation. This study also showed that the angiographic benefit with the SES occurred in the most complex patients: those with the smallest vessels and the longest lesions and in patients with diabetes mellitus. The beneficial effects of the SES were most pronounced within the stent in smaller vessels and longer lesions but were also highly significant within the segment that included the nonstented margins. The higher restenosis rates within the segment that included the nonstented margins were attributable to lumen changes that occurred in regions that were not fully treated with sirolimus. These non stent margin lumen changes were also found in patients treated with the BMS, but it appears that the beneficial effect of the SES on late lumen loss did extend to a lesser extent into the 5-mm proximal and distal margins of the stent, potentially because of upstream or downstream elution of the drug or because of the effect on an overall reduction in stent restenosis on the adjacent nonstented margins. No deleterious effects of the SES were found at the margins of the stent. Patterns of Restenosis The study also demonstrated a novel shift in the pattern of restenosis with the SES. Intimal hyperplasia after BMS implantation most commonly involves the central portion of the stent. In patients treated with a BMS in this study, 50% of patients had a diffuse pattern ( 10 mm) of restenosis that was confined within the stent (35.9%) or extended from within the stent to include its proximal or distal margins (13.3%). In contrast, 61.3% of the SES patients who developed restenosis had the intimal hyperplasia localized to the margin of the stent, often sparing the axial region of the stent. The restenosis length was also shorter in patients treated with the SES. Given the finding that the edge lumen dimensions were

7 Popma et al Angiographic Findings in SIRIUS 3779 actually larger at follow-up in patients treated with the SES, it is not certain what precise factors contributed to the margin restenosis in patients treated with the SES. The most likely cause of the margin restenosis pattern was the occurrence of injury at the margins of the stent that was not adequately covered with sirolimus. Alterations in operator technique, such as complete coverage of the lesion with adequate margins, avoidance of injury at the edges of the stent, and careful postdilatation within the stent may reduce the pattern of margin restenosis. Coronary Aneurysms Coronary aneurysms are rare findings after coronary intervention, with an incidence of 0.3% to 6%, depending on the definition used for vessel expansion Pseudoaneurysms are the most common form of trauma-induced coronary aneurysms, because of abnormal healing of an intimal dissection that occurs at the time of intervention. True aneurysms involving all 3 layers are uncommon after coronary intervention. Stents that elute antiproliferative agents have the potential to cause both forms of aneurysms, similar to potent systemic antiproliferative agents. 30 With a conservative definition of 1.2 times the reference diameter, there was no difference in the incidence of late aneurysms in patients treated with the SES. Limitations There were several limitations to this study. Although angiographic recording of all balloon inflations before and after stent implantation was recommended, it was often difficult to determine whether a significant degree of arterial injury occurred outside the margin of the stent. The margin restenosis that occurred in patients treated with the SES, although less than in those treated with the BMS, may have been a result of margin injury that was not matched with drug elution. Further studies are ongoing to determine the precise distribution of vessel injury and the degree of late lumen loss with the SES and BMS. Another limitation of this study was that the smallest stent size was 2.5 mm, and vessels as small as 2.0 mm by quantitative angiography were treated. It is not clear whether smaller stents (2.25 mm) more appropriately matched to the reference diameter would have been associated with less margin restenosis in the smallest vessels. Finally, repeat angiography was performed 8 months after stent implantation, and although there was no evidence of a delayed catch-up in restenosis in a pilot series of sirolimustreated patients 2 years after the procedure, 12 the very late occurrence of angiographic events will require additional follow-up studies. The results of this study demonstrate that treatment with the SES in patients with native-vessel coronary artery disease resulted in marked reductions of all angiographic indices of restenosis. When restenosis occurred with the SES, it developed most often at the margins of the stent. The beneficial effect was also present in the smallest vessel and longest lesions. There were no differences in late aneurysms between the 2 groups. We conclude that the use of the SES is safe and effective for the treatment of complex native-vessel coronary artery disease. Acknowledgment This study was supported by a grant from Cordis, a Johnson & Johnson Company. Disclosure Dr Moses has served a consultant to and speaker for Cordis, a Johnson & Johnson Company. Dr Popma received research grants from Cardis for the angiographic analysis of this study. References 1. Fishman D, Leon M, Baim D, et al. A randomized comparison of coronary stent placement and balloon angioplasty in the treatment of coronary artery disease. N Engl J Med. 1994;331: Serruys PW, de Jaegere P, Kiemeneij F, et al. A comparison of balloonexpandable-stent implantation with balloon angioplasty in patients with coronary artery disease. Benestent Study Group. N Engl J Med. 1994; 331: Cutlip DE, Chauhan MS, Baim DS, et al. Clinical restenosis after coronary stenting: perspectives from multicenter clinical trials. J Am Coll Cardiol 2002;40: Holmes D, Savage MPS, Lablanche J-M, et al. Results of Prevention of REStenosis with Tranilast and its Outcomes (PRESTO) Trial. Circulation 2002;106: O Malley A, Norman S-L, Kuntz R. Models for multivariate mixed outcomes applications to the estimation of objective performance: coronary-artery stenting. Stat Med. 2003;22: Baim D, Cutlip D, Midei M, et al. Final results of a randomized trial comparing the MULTI-LINK stent with the Palmaz-Schatz stent for narrowings in native coronary arteries. Am J Cardiol. 2001;87: Baim D, Cutlip D, O Shaughnessy C, et al. Final results of a randomized trial comparing the NIR stent to the Palmaz-Schatz stent for narrowings in native coronary arteries. Am J Cardiol. 2001;87: Holmes D, Lansky A, Kuntz R, et al. The Paragon stent study: a randomized trial of a new martinsitic nitinol stent versus the Palmaz-Schatz (PS) TM stent for the treatment of complex native coronary arterial lesions. Am J Cardiol. 2000;86: Hoffmann R, Mintz GS, Dussaillant GR, et al. Patterns and mechanisms of in-stent restenosis: a serial intravascular ultrasound study. Circulation. 1996;94: Mehran R, Dangas G, Abizaid A, et al. Angiographic patterns of in-stent restenosis: classification and implications for long-term outcome. Circulation. 1999;100: Morice M, Serruys P, Sousa J, et al. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002;346: Sousa J, Costa M, Sousa A, et al. Two-year angiographic and intravascular ultrasound follow-up after implantation of sirolimus-eluting stents in human coronary arteries. Circulation. 2003;107: Moses JW, Leon MB, Popma JJ, et al. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. N Engl J Med. 2003;349: Ellis S, Vandormael M, Cowley M, et al. Coronary morphologic and clinical determinants of procedural outcome with angioplasty for multivessel coronary disease. Circulation. 1990;82: Ryan T, Faxon D, Gunnar R, et al. Guidelines for percutaneous transluminal coronary angioplasty: a report of the American College of Cardiology/American Heart Association Task Force on Assessment of Diagnostic and Therapeutic Cardiovascular Procedures (Subcommittee on Percutaneous Transluminal Coronary Angioplasty). J Am Coll Cardiol. 1988;12: Huber MS, Mooney JF, Madison J, et al. Use of a morphologic classification to predict clinical outcome after dissection from coronary angioplasty. Am J Cardiol. 1991;68: van der Zwet PM, Reiber JH. A new approach for the quantification of complex lesion morphology: the gradient field transform: basic principles and validation results. J Am Coll Cardiol. 1994;24: Mintz GS, Popma JJ, Pichard AD, et al. Arterial remodeling after coronary angioplasty: a serial intravascular ultrasound study. Circulation. 1996;94: Dussaillant G, Mintz G, Pichard A, et al. Small stent size and intimal hyperplasia contribute to restenosis: a volumetric intravascular ultrasound analysis. J Am Coll Cardiol. 1995;26:

8 3780 Circulation December 21/28, Albiero R, Nishida T, Adamian M, et al. Edge restenosis after implantation of high activity 32 P radioactive -emitting stents. Circulation. 2000;101: Albiero R, Adamian M, Kobayashi N, et al. Short- and intermediate-term results of 32 P radioactive -emitting stent implantation in patients with coronary artery disease: the Milan Dose-Response Study. Circulation. 2000;101: Marx S, Marks A. Bench to bedside: the development of rapamycin and its application to stent restenosis. Circulation. 2001;104: Sousa J, Costa M, Abizaid A, et al. Sustained suppression of neointimal proliferation by sirolimus-eluting stents: one-year angiographic and intravascular ultrasound follow-up. Circulation. 2001;104: Degertekin M, Serruys P, Foley D, et al. Persistent inhibition of neointimal hyperplasia after sirolimus-eluting stent implantation: long-term (up to 2 years) clinical, angiographic, and intravascular ultrasound follow-up. Circulation. 2002;106: Baumbach A, Bittl JA, Fleck E, et al. Acute complications of excimer laser coronary angioplasty: a detailed analysis of multicenter results. Coinvestigators of the U.S. and European Percutaneous Excimer Laser Coronary Angioplasty (PELCA) Registries. J Am Coll Cardiol. 1994;23: Condado JA, Waksman R, Gurdiel O, et al. Long-term angiographic and clinical outcome after percutaneous transluminal coronary angioplasty and intracoronary radiation therapy in humans. Circulation. 1997;96: Slota PA, Fischman DL, Savage MP, et al. Frequency and outcome of development of coronary artery aneurysm after intracoronary stent placement and angioplasty. STRESS Trial Investigators. Am J Cardiol. 1997;79: Vassanelli C, Turri M, Morando G, et al. Coronary arterial aneurysms after percutaneous transluminal coronary angioplasty: a not uncommon finding at elective follow-up angiography. Int J Cardiol. 1989;22: Walford GD, Midei MG, Aversano TR, et al. Coronary artery aneurysm formation following percutaneous transluminal coronary angioplasty: treatment of associated restenosis with repeat percutaneous transluminal coronary angioplasty. Cathet Cardiovasc Diagn. 1990; 20: Rab ST, King SD, Roubin GS, et al. Coronary aneurysms after stent placement: a suggestion of altered vessel wall healing in the presence of anti-inflammatory agents. J Am Coll Cardiol. 1991;18:

Lesions at coronary bifurcations represent a challenging

Lesions at coronary bifurcations represent a challenging Randomized Study to Evaluate Sirolimus-Eluting Stents Implanted at Coronary Bifurcation Lesions Antonio Colombo, MD; Jeffrey W. Moses, MD; Marie Claude Morice, MD; Josef Ludwig, MD; David R. Holmes, Jr,

More information

Angiographic and Intravascular Ultrasound Predictors of In-Stent Restenosis

Angiographic and Intravascular Ultrasound Predictors of In-Stent Restenosis 1630 JACC Vol. 32, No. 6 Angiographic and Intravascular Ultrasound Predictors of In-Stent Restenosis SHUNJI KASAOKA, MD, JONATHAN M. TOBIS, MD, FACC, TATSURO AKIYAMA, MD,* BERNHARD REIMERS, MD,* CARLO

More information

Sirolimus-Eluting Stents for Treatment of In-Stent Restenosis

Sirolimus-Eluting Stents for Treatment of In-Stent Restenosis Clinical Investigation Alfonso Medina, MD José Suárez de Lezo, MD Manuel Pan, MD Antonio Delgado, MD José Segura, MD Djordje Pavlovic, MD Francisco Melián, MD Miguel Romero, MD Federico Segura, MD Enrique

More information

Journal of the American College of Cardiology Vol. 39, No. 8, by the American College of Cardiology Foundation ISSN /02/$22.

Journal of the American College of Cardiology Vol. 39, No. 8, by the American College of Cardiology Foundation ISSN /02/$22. Journal of the American College of Cardiology Vol. 39, No. 8, 2002 2002 by the American College of Cardiology Foundation ISSN 0735-1097/02/$22.00 Published by Elsevier Science Inc. PII S0735-1097(02)01774-6

More information

Predictors of 6-Month Angiographic Restenosis inside Bare-Metal Stent in Chinese Patients with Coronary Artery Disease

Predictors of 6-Month Angiographic Restenosis inside Bare-Metal Stent in Chinese Patients with Coronary Artery Disease Original Article Acta Cardiol Sin 2009;25:1 6 Coronary Artery Disease Predictors of 6-Month Angiographic Restenosis inside Bare-Metal Stent in Chinese Patients with Coronary Artery Disease Yung-Lung Chen,

More information

In-Stent Restenosis. Can we kill it?

In-Stent Restenosis. Can we kill it? In-Stent Restenosis Can we kill it? However, In-stent Restenosis is the most serious problem (2-25%) More than 15, lesions will need treatment because of in-stent restenosis. Varying Prevalence Rates of

More information

Results of the Washington Radiation for In-Stent Restenosis Trial for Long Lesions (Long WRIST) Studies

Results of the Washington Radiation for In-Stent Restenosis Trial for Long Lesions (Long WRIST) Studies Intracoronary Radiation Therapy Improves the Clinical and Angiographic Outcomes of Diffuse In-Stent Restenotic Lesions Results of the Washington Radiation for In-Stent Restenosis Trial for Long Lesions

More information

Coronary Heart Disease. Treatment of Left Anterior Descending Coronary Artery Disease With Sirolimus-Eluting Stents

Coronary Heart Disease. Treatment of Left Anterior Descending Coronary Artery Disease With Sirolimus-Eluting Stents Coronary Heart Disease Treatment of Left Anterior Descending Coronary Artery Disease With Sirolimus-Eluting Stents Neil Sawhney, MD; Jeffrey W. Moses, MD; Martin B. Leon, MD; Richard E. Kuntz, MD; Jeffrey

More information

Endovascular beta-irradiation with a liquid 188 Re-filled balloon to reduce restenosis after coronary angioplasty.

Endovascular beta-irradiation with a liquid 188 Re-filled balloon to reduce restenosis after coronary angioplasty. Endovascular beta-irradiation with a liquid 188 Re-filled balloon to reduce restenosis after coronary angioplasty. Peix A., Llerena L., Ponce F., López A., López L., Guerrero I., Cabrera L.O., Maltas A.M.,

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013 840 Identification of independent risk factors for restenosis following bare metal stent implantation: Role of bare metal stents in the era of drug eluting stents CHANG BUM PARK 1 and HOON KI PARK 2 1

More information

Coronary artery disease (CAD) is the REVIEW PROJECTED HEALTH AND ECONOMIC BENEFITS OF THE USE OF SIROLIMUS-ELUTING CORONARY STENTS

Coronary artery disease (CAD) is the REVIEW PROJECTED HEALTH AND ECONOMIC BENEFITS OF THE USE OF SIROLIMUS-ELUTING CORONARY STENTS PROJECTED HEALTH AND ECONOMIC BENEFITS OF THE USE OF SIROLIMUS-ELUTING CORONARY STENTS Rodolphe Ruffy, MD, FACC,* and Raymond J. Kaden, MBA, CPA ABSTRACT Despite remarkable technological progress in interventional

More information

Journal of the American College of Cardiology Vol. 46, No. 5, by the American College of Cardiology Foundation ISSN /05/$30.

Journal of the American College of Cardiology Vol. 46, No. 5, by the American College of Cardiology Foundation ISSN /05/$30. Journal of the American College of Cardiology Vol. 46, No. 5, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.06.009

More information

PCI for In-Stent Restenosis. CardioVascular Research Foundation

PCI for In-Stent Restenosis. CardioVascular Research Foundation PCI for In-Stent Restenosis ISR of BMS Patterns of In-Stent Restenosis Pattern I : Focal Type IA: Articulation / Gap Type IB: Marginal Type IC: Focal body Type ID: Multifocal Pattern II,III,IV : Diffuse

More information

Clinical Investigation and Reports. Two-Year Angiographic Follow-Up of Intracoronary Sr90 Therapy for Restenosis Prevention After Balloon Angioplasty

Clinical Investigation and Reports. Two-Year Angiographic Follow-Up of Intracoronary Sr90 Therapy for Restenosis Prevention After Balloon Angioplasty Clinical Investigation and Reports Two-Year Angiographic Follow-Up of Intracoronary Sr90 Therapy for Restenosis Prevention After Balloon Angioplasty David Meerkin, MBBS; Michel Joyal, MD; Jean-Claude Tardif,

More information

Journal of the American College of Cardiology Vol. 34, No. 4, by the American College of Cardiology ISSN /99/$20.

Journal of the American College of Cardiology Vol. 34, No. 4, by the American College of Cardiology ISSN /99/$20. Journal of the American College of Cardiology Vol. 34, No. 4, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00324-1 Final

More information

The leading cause of death for both men

The leading cause of death for both men A REVIEW OF DRUG-ELUTING STENTS: WHY ALL THE EXCITEMENT? Martin B. Leon, MD,* and Jeffrey W. Moses, MD* ABSTRACT Cardiovascular disease is the leading cause of death in the United States, and atherosclerosis

More information

Brachytherapy for In-Stent Restenosis: Is the Concept Still Alive? Matthew T. Menard, M.D. Brigham and Women s Hospital Boston, Massachussetts

Brachytherapy for In-Stent Restenosis: Is the Concept Still Alive? Matthew T. Menard, M.D. Brigham and Women s Hospital Boston, Massachussetts Brachytherapy for In-Stent Restenosis: Is the Concept Still Alive? Matthew T. Menard, M.D. Brigham and Women s Hospital Boston, Massachussetts Disclosure Speaker name: Matthew T. Menard... x I do not have

More information

Treating In-Stent Restenosis with Brachytherapy: Does it Actually Work?

Treating In-Stent Restenosis with Brachytherapy: Does it Actually Work? Treating In-Stent Restenosis with Brachytherapy: Does it Actually Work? Matthew T. Menard, M.D. Brigham and Women s Hospital Pacific Northwest Endovascular Conference June 15, 2018 DISCLOSURE Matthew Menard,

More information

Small Stent Size and Intimal Hyperplasia Contribute to Restenosis: A Volumetric Intravascular Ultrasound Analysis

Small Stent Size and Intimal Hyperplasia Contribute to Restenosis: A Volumetric Intravascular Ultrasound Analysis 720 JACC Vol. 26, No. 3 Small Stent Size and Intimal Hyperplasia Contribute to Restenosis: A Volumetric Intravascular Ultrasound Analysis GASTON R. DUSSAILLANT, MD, GARY S. MINTZ, MD, FACC, AUGUSTO KENNETH

More information

One-year Outcome of Stenting for Long Coronary Lesions, a Prospective Clinical Trial

One-year Outcome of Stenting for Long Coronary Lesions, a Prospective Clinical Trial Tabriz University of Medical Sciences Original Article One-year Outcome of Stenting for Long Coronary Lesions, a Prospective Clinical Trial Samad Ghaffari MD, Mohammad Reza Hasanian MD, Leili Pourafkari

More information

Superiority of sirolimus eluting stent compared with intracoronary b radiation for treatment of in-stent restenosis: a matched comparison

Superiority of sirolimus eluting stent compared with intracoronary b radiation for treatment of in-stent restenosis: a matched comparison 1584 INTERVENTIONAL CARDIOLOGY AND SURGERY Superiority of sirolimus eluting stent compared with intracoronary b radiation for treatment of in-stent restenosis: a matched comparison E Iofina, P W Radke,

More information

For Personal Use. Copyright HMP 2013

For Personal Use. Copyright HMP 2013 Case Report J INVASIVE CARDIOL 2013;25(2):E39-E41 A Case With Successful Retrograde Stent Delivery via AC Branch for Tortuous Right Coronary Artery Yoshiki Uehara, MD, PhD, Mitsuyuki Shimizu, MD, PhD,

More information

LM stenting - Cypher

LM stenting - Cypher LM stenting - Cypher Left main stenting with BMS Since 1995 Issues in BMS era AMC Restenosis and TLR (%) 3 27 TLR P=.282 Restenosis P=.71 28 2 1 15 12 Ostium 5 4 Shaft Bifurcation Left main stenting with

More information

Clinical Investigations

Clinical Investigations Clinical Investigations Clinical Outcomes for Single Stent and Multiple Stents in Contemporary Practice Qiao Shu Bin, MD; Liu Sheng Wen, MD; Xu Bo, BS; Chen Jue, MD; Liu Hai Bo, MD; Yang Yue Jin, MD; Chen

More information

Intracoronary Brachytherapy

Intracoronary Brachytherapy Page 1 of 19 Search Search Health Technology Advisory Committee Home Publications Links Search Intracoronary Brachytherapy June 2001 Executive Summary Percutaneous transluminal coronary angioplasty (PTCA)

More information

Journal of the American College of Cardiology Vol. 47, No. 7, by the American College of Cardiology Foundation ISSN /06/$32.

Journal of the American College of Cardiology Vol. 47, No. 7, by the American College of Cardiology Foundation ISSN /06/$32. Journal of the American College of Cardiology Vol. 47, No. 7, 2006 2006 by the American College of Cardiology Foundation ISSN 0735-1097/06/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.05.102

More information

Management of In-stent Restenosis after Lower Extremity Endovascular Procedures

Management of In-stent Restenosis after Lower Extremity Endovascular Procedures Management of In-stent Restenosis after Lower Extremity Endovascular Procedures Piotr Sobieszczyk, MD Associate Director, Cardiac Catheterization Laboratory Cardiovascular Division and Vascular Medicine

More information

PCI for Left Anterior Descending Artery Ostial Stenosis

PCI for Left Anterior Descending Artery Ostial Stenosis PCI for Left Anterior Descending Artery Ostial Stenosis Why do you hesitate PCI for LAD ostial stenosis? LAD Ostial Lesion Limitations of PCI High elastic recoil Involvement of the distal left main coronary

More information

Intracoronary stents reduce restenosis compared with balloon

Intracoronary stents reduce restenosis compared with balloon Treatment of In-Stent Restenosis With Excimer Laser Coronary Angioplasty Versus Rotational Atherectomy Comparative Mechanisms and Results Roxana Mehran, MD; George Dangas, MD, PhD; Gary S. Mintz, MD; Ron

More information

Non Polymer-Based Paclitaxel-Coated Coronary Stents for the Treatment of Patients With De Novo Coronary Lesions

Non Polymer-Based Paclitaxel-Coated Coronary Stents for the Treatment of Patients With De Novo Coronary Lesions Non Polymer-Based Paclitaxel-Coated Coronary Stents for the Treatment of Patients With De Novo Coronary Lesions Angiographic Follow-Up of the DELIVER Clinical Trial Alexandra J. Lansky, MD; Ricardo A.

More information

The time period between the first

The time period between the first Review Article Hellenic J Cardiol 2010; 51: 512-517 The Ever Increasing Role of Percutaneous Interventions in Coronary Revascularisation Vassilis N. Spanos, Pavlos K. Toutouzas Cardiology Department, Euroclinic

More information

Journal of the American College of Cardiology Vol. 34, No. 1, by the American College of Cardiology ISSN /99/$20.

Journal of the American College of Cardiology Vol. 34, No. 1, by the American College of Cardiology ISSN /99/$20. Journal of the American College of Cardiology Vol. 34, No. 1, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00151-5 Mechanisms

More information

PCI for Long Coronary Lesion

PCI for Long Coronary Lesion PCI for Long Coronary Lesion Shift of a General Idea with the Introduction of DES In the Bare Metal Stent Era Higher Restenosis Rate With Increasing Stent Length and Decreasing Stent Area Restenosis.6.4.2

More information

Advances in Interventional Cardiology II: A Case Study

Advances in Interventional Cardiology II: A Case Study CARDIOLOGY BOARD REVIEW MANUAL PUBLISHING STAFF PRESIDENT, PUBLISHER Bruce M. White EXECUTIVE EDITOR Debra Dreger SENIOR EDITOR Miranda J. Hughes, PhD ASSISTANT EDITOR Melissa Frederick EDITORIAL ASSISTANT

More information

Late Loss Is The Single Best Parameter For Estimating Stent-Based Restenosis Resistance

Late Loss Is The Single Best Parameter For Estimating Stent-Based Restenosis Resistance Late Loss Is The Single Best Parameter For Estimating Stent-Based Restenosis Resistance Richard Kuntz Brigham and Women s s Hospital Harvard Medical School Late Loss and DES Brief history of Late Loss

More information

DES In-stent Restenosis

DES In-stent Restenosis DES In-stent Restenosis Roxana Mehran, MD Columbia University Medical Center The Cardiovascular Research Foundation DES Restenosis Mechanisms Predictors Morphological patterns Therapy approach Mechanisms

More information

Intravascular Ultrasound in the Drug-Eluting Stent Era

Intravascular Ultrasound in the Drug-Eluting Stent Era Journal of the American College of Cardiology Vol. 48, No. 3, 2006 2006 by the American College of Cardiology Foundation ISSN 0735-1097/06/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.04.068

More information

Are We Making Progress With Percutaneous Saphenous Vein Graft Treatment? A Comparison of 1990 to 1994 and 1995 to 1998 Results

Are We Making Progress With Percutaneous Saphenous Vein Graft Treatment? A Comparison of 1990 to 1994 and 1995 to 1998 Results Journal of the American College of Cardiology Vol. 38, No. 1, 2001 2001 by the American College of Cardiology ISSN 0735-1097/01/$20.00 Published by Elsevier Science Inc. PII S0735-1097(01)01324-9 Are We

More information

Journal of the American College of Cardiology Vol. 38, No. 3, by the American College of Cardiology ISSN /01/$20.

Journal of the American College of Cardiology Vol. 38, No. 3, by the American College of Cardiology ISSN /01/$20. Journal of the American College of Cardiology Vol. 38, No. 3, 2001 2001 by the American College of Cardiology ISSN 0735-1097/01/$20.00 Published by Elsevier Science Inc. PII S0735-1097(01)01458-9 Cutting

More information

Neuestes aus der Therapie der pavk. beschichtete Stents + Ballons. Karls-University. Eberhard-Karls. of Tubingen Department of Diagnostic Radiology

Neuestes aus der Therapie der pavk. beschichtete Stents + Ballons. Karls-University. Eberhard-Karls. of Tubingen Department of Diagnostic Radiology Eberhard-Karls Karls-University of Tubingen Department of Diagnostic Radiology Neuestes aus der Therapie der pavk Berlin Dezember 08 beschichtete Stents + Ballons Gunnar Tepe 1 Local Drug Delivery Basic

More information

Prior randomized trials have shown that the Palmaz-Schatz

Prior randomized trials have shown that the Palmaz-Schatz Randomized Comparison of and almaz-schatz Stent for Elective Treatment of Coronary Stenoses Alexandra J. Lansky, MD; Gary S. Roubin, MD, hd; Charles D. O Shaughnessy; aul B. Moore, MD; Larry S. Dean, MD;

More information

The Stenting Coronary Arteries in Non-stress/benestent Disease (SCANDSTENT) Trial

The Stenting Coronary Arteries in Non-stress/benestent Disease (SCANDSTENT) Trial Journal of the American College of Cardiology Vol. 47, No. 2, 2006 2006 by the American College of Cardiology Foundation ISSN 0735-1097/06/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.10.045

More information

Journal of the American College of Cardiology Vol. 48, No. 2, by the American College of Cardiology Foundation ISSN /06/$32.

Journal of the American College of Cardiology Vol. 48, No. 2, by the American College of Cardiology Foundation ISSN /06/$32. Journal of the American College of Cardiology Vol. 48, No. 2, 2006 2006 by the American College of Cardiology Foundation ISSN 0735-1097/06/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.03.039

More information

Original Research Article. S. Praveen 1, Baiju R. 1 *, V. V. Radhakrishnan 1, Sarosh Kumar K. K. 2

Original Research Article. S. Praveen 1, Baiju R. 1 *, V. V. Radhakrishnan 1, Sarosh Kumar K. K. 2 International Journal of Research in Medical Sciences Praveen S et al. Int J Res Med Sci. 2018 Feb;6(2):448-454 www.msjonline.org pissn 2320-6071 eissn 2320-6012 Original Research Article DOI: http://dx.doi.org/10.18203/2320-6012.ijrms20180029

More information

Complex PCI. Your partner in complex PCI: In-stent restenosis (ISR)

Complex PCI. Your partner in complex PCI: In-stent restenosis (ISR) Comple PCI Your partner in comple PCI: Your partner in comple PCI: Philips provides a portfolio of specialty coronary diagnostic and therapy devices that enable safe and effective treatment of a wide variety

More information

Safety and Efficacy of Coronary Stent Implantation. Acute and Six Month Outcomes of 1,126 Consecutive Patients Treated in 1996 and 1997

Safety and Efficacy of Coronary Stent Implantation. Acute and Six Month Outcomes of 1,126 Consecutive Patients Treated in 1996 and 1997 Original Article Safety and Efficacy of Coronary Stent Implantation. Acute and Six Month Outcomes of 1,126 Consecutive Patients Treated in 1996 and 1997 Luiz Alberto Mattos, Ibraim Pinto, Alexandre Abizaid,

More information

Primary and mid-term outcome of sirolimus-eluting stent implantation with angiographic guidance alone

Primary and mid-term outcome of sirolimus-eluting stent implantation with angiographic guidance alone Journal of Cardiology (2008) 51, 18 24 Primary and mid-term outcome of sirolimus-eluting stent implantation with angiographic guidance alone Hajime Fujimoto, Susumu Tao, Tomotaka Dohi, Sachiko Ito, Jun

More information

Basics of Angiographic Interpretation Analysis of Angiography

Basics of Angiographic Interpretation Analysis of Angiography Basics of Angiographic Interpretation Analysis of Angiography Young-Hak Kim, MD, PhD Cardiac Center, University of Ulsan College of Medicine, Seoul, Korea What made us nervous Supervisors Stent Contrast

More information

Effect of Intravascular Ultrasound- Guided vs. Angiography-Guided Everolimus-Eluting Stent Implantation: the IVUS-XPL Randomized Clinical Trial

Effect of Intravascular Ultrasound- Guided vs. Angiography-Guided Everolimus-Eluting Stent Implantation: the IVUS-XPL Randomized Clinical Trial Effect of Intravascular Ultrasound- Guided vs. Angiography-Guided Everolimus-Eluting Stent Implantation: the IVUS-XPL Randomized Clinical Trial Myeong-Ki Hong, MD. PhD on behalf of the IVUS-XPL trial investigators

More information

Titan versus TAXUS Stents at 1 Year Clinical Outcome

Titan versus TAXUS Stents at 1 Year Clinical Outcome Original Article Titan versus TAXUS Stents at 1 Year Clinical Outcome Acta Cardiol Sin 2011;27:94 100 Interventional Cardiology One-Year Follow-Up after Percutaneous Coronary Intervention with Titanium-Nitride-Oxide-Coated

More information

The New England Journal of Medicine INTRAVASCULAR GAMMA RADIATION FOR IN-STENT RESTENOSIS IN SAPHENOUS-VEIN BYPASS GRAFTS

The New England Journal of Medicine INTRAVASCULAR GAMMA RADIATION FOR IN-STENT RESTENOSIS IN SAPHENOUS-VEIN BYPASS GRAFTS INTRAVASCULAR GAMMA RADIATION FOR IN-STENT RESTENOSIS IN SAPHENOUS-VEIN BYPASS GRAFTS RON WAKSMAN, M.D., ANDREW E. AJANI, M.D., R. LARRY WHITE, M.D., ROSANNA C. CHAN, M.D., LOWELL F. SATLER, M.D., KENNETH

More information

Key Words Angioplasty Coronary artery disease Revascularization Stent drug eluting stent

Key Words Angioplasty Coronary artery disease Revascularization Stent drug eluting stent J Cardiol 26 Dec; 48 6 : 325 331 3 mm : Initial and Mid-Term Effects of 3 mm Long Sirolimus-Eluting Stents in Patients With Diffuse Long Coronary Lesions: Comparison With Bare Metal Stents Abstract Yosuke

More information

A Synergistic Approach to Optimal Stenting Directional Coronary Atherectomy Prior to Coronary Artery Stent Implantation the AtheroLink Registry

A Synergistic Approach to Optimal Stenting Directional Coronary Atherectomy Prior to Coronary Artery Stent Implantation the AtheroLink Registry Journal of the American College of Cardiology Vol. 36, No. 6, 2000 2000 by the American College of Cardiology ISSN 0735-1097/00/$20.00 Published by Elsevier Science Inc. PII S0735-1097(00)00967-0 A Synergistic

More information

Current PTCA practice and clinical outcomes in the Netherlands: the real world in the pre-drug-eluting stent era

Current PTCA practice and clinical outcomes in the Netherlands: the real world in the pre-drug-eluting stent era 3 Current PTCA practice and clinical outcomes in the Netherlands: the real world in the pre-drug-eluting stent era Pascalle S. Monraats, Willem R.P. Agema, Aeilko H. Zwinderman, Robbert J. de Winter, René

More information

Comparison of Sirolimus Versus Paclitaxel Eluting Stents for Treatment of Coronary in-stent Restenosis F. Airoldi, et al. Am J Cardiol (2006)

Comparison of Sirolimus Versus Paclitaxel Eluting Stents for Treatment of Coronary in-stent Restenosis F. Airoldi, et al. Am J Cardiol (2006) Comparison of Sirolimus Versus Paclitaxel Eluting Stents for Treatment of Coronary in-stent Restenosis F. Airoldi, et al. Am J Cardiol (2006) 97;1182-7 n&list_uids=16616023 Value of the American College

More information

CPT Code Details

CPT Code Details CPT Code 93572 Details Code Descriptor Intravascular Doppler velocity and/or pressure derived coronary flow reserve measurement (coronary vessel or graft) during coronary angiography including pharmacologically

More information

journal of medicine The new england Sirolimus-Eluting and Paclitaxel-Eluting Stents for Coronary Revascularization abstract

journal of medicine The new england Sirolimus-Eluting and Paclitaxel-Eluting Stents for Coronary Revascularization abstract The new england journal of medicine established in 1812 august 18, 2005 vol. 353 no. 7 Sirolimus-Eluting and Paclitaxel-Eluting Stents for Coronary Revascularization Stephan Windecker, M.D., Andrea Remondino,

More information

The Influence of Diabetes Mellitus on Acute and Late Clinical Outcomes Following Coronary Stent Implantation

The Influence of Diabetes Mellitus on Acute and Late Clinical Outcomes Following Coronary Stent Implantation 584 JACC Vol. 32, No. 3 The Influence of Diabetes Mellitus on Acute and Late Clinical Outcomes Following Coronary Stent Implantation ALEXANDRE ABIZAID, MD, RAN KORNOWSKI, MD, GARY S. MINTZ, MD, FACC, MUN

More information

Prevention of Coronary Stent Thrombosis and Restenosis

Prevention of Coronary Stent Thrombosis and Restenosis Prevention of Coronary Stent Thrombosis and Restenosis Seong-Wook Park, MD, PhD, FACC Division of Cardiology, Asan Medical Center University of Ulsan College of Medicine, Seoul, Korea 9/12/03 Coronary

More information

In-stent Restenosis Diagnostic and Therapeutic Challenges. Kostis Raisakis General Hospital of Athens «G. Gennimatas»

In-stent Restenosis Diagnostic and Therapeutic Challenges. Kostis Raisakis General Hospital of Athens «G. Gennimatas» In-stent Restenosis Diagnostic and Therapeutic Challenges Kostis Raisakis General Hospital of Athens «G. Gennimatas» Introduction With POBA, rates of acute and chronic vessel occlusion at 30% to 60%, secondary

More information

JAMA. 2005;293:

JAMA. 2005;293: ORIGINAL CONTRIBUTION -Eluting Stent or -Eluting Stent vs Angioplasty for Prevention of Recurrences in Patients With Coronary In-Stent Restenosis A Randomized Controlled Trial Adnan Kastrati, MD Julinda

More information

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS Journal of the American College of Cardiology Vol. 37, No. 5, 2001 2001 by the American College of Cardiology ISSN 0735-1097/01/$20.00 Published by Elsevier Science Inc. PII S0735-1097(01)01162-7 Long-Term

More information

Plaque Removal Prior to Stent Implantation in Native Coronary Arteries: Why? When? and How?

Plaque Removal Prior to Stent Implantation in Native Coronary Arteries: Why? When? and How? IAGS 1998 Proceedings Plaque Removal Prior to Stent Implantation in Native Coronary Arteries: Why? When? and How? Issam Moussa, MD, Carlo Di Mario, PhD, MD, Antonio Colombo, MD From the Centro Cuore Columbus,

More information

The New England Journal of Medicine LOCALIZED INTRACORONARY GAMMA-RADIATION THERAPY TO INHIBIT THE RECURRENCE OF RESTENOSIS AFTER STENTING

The New England Journal of Medicine LOCALIZED INTRACORONARY GAMMA-RADIATION THERAPY TO INHIBIT THE RECURRENCE OF RESTENOSIS AFTER STENTING LOCALIZED INTRACORONARY GAMMA-RADIATION THERAPY TO INHIBIT THE RECURRENCE OF RESTENOSIS AFTER STENTING MARTIN B. LEON, M.D., PAUL S. TEIRSTEIN, M.D., JEFFREY W. MOSES, M.D., PRABHAKAR TRIPURANENI, M.D.,

More information

C h a p t e r 5 8 Drug Eluting Stents : A New Revolution in the Treatment of Coronary Artery Disease

C h a p t e r 5 8 Drug Eluting Stents : A New Revolution in the Treatment of Coronary Artery Disease C h a p t e r 5 8 Drug Eluting Stents : A New Revolution in the Treatment of Coronary Artery Disease Gunasekaran Sengottuvel, MD Marie-Claude Morice, MD Institut Cardiovasculaire Paris Sud, Institut Hospitalier

More information

Late and Very Late Drug-Eluting Stent Malapposition Serial 2-Year Quantitative IVUS Analysis

Late and Very Late Drug-Eluting Stent Malapposition Serial 2-Year Quantitative IVUS Analysis Late and Very Late Drug-Eluting Stent Malapposition Serial 2-Year Quantitative IVUS Analysis Soo-Jin Kang, MD; Gary S. Mintz, MD; Duk-Woo Park, MD; Seung-Whan Lee, MD; Young-Hak Kim, MD; Cheol Whan Lee,

More information

Effectiveness of Drug-Eluting Stents in Patients With Bare-Metal In-Stent Restenosis

Effectiveness of Drug-Eluting Stents in Patients With Bare-Metal In-Stent Restenosis Journal of the American College of Cardiology Vol. 49, No. 5, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.10.049

More information

The Use of Intravascular Ultrasound and Spot Stenting for the Treatment of Long Lesions and Small Vessels

The Use of Intravascular Ultrasound and Spot Stenting for the Treatment of Long Lesions and Small Vessels IAGS 1998 Proceedings The Use of Intravascular Ultrasound and Spot Stenting for the Treatment of Long Lesions and Small Vessels Issam Moussa, MD, Joseph De Gregorio, MD, Carlo Di Mario, PhD, MD, Antonio

More information

Coronary stenting is an established form of treatment for

Coronary stenting is an established form of treatment for Intracoronary Stenting and Angiographic Results Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO) Trial Adnan Kastrati, MD; Julinda Mehilli, MD; Josef Dirschinger, MD; Franz Dotzer, MD; Helmut

More information

Innovation by design. Technology that sets a new standard

Innovation by design. Technology that sets a new standard Innovation by design Technology that sets a new standard Flexible nitinol scoring element with three rectangular spiral struts works in tandem with a semi-compliant balloon to score the target lesion Balloon

More information

Nine-year clinical outcomes of drug-eluting stents vs. bare metal stents for large coronary vessel lesions

Nine-year clinical outcomes of drug-eluting stents vs. bare metal stents for large coronary vessel lesions Journal of Geriatric Cardiology (2017) 14: 35 41 2017 JGC All rights reserved; www.jgc301.com Research Article Open Access Nine-year clinical outcomes of drug-eluting stents vs. bare metal stents for large

More information

Very late thrombosis of sirolimus-eluting stent due to late malapposition: Serial observations with optical coherence tomography

Very late thrombosis of sirolimus-eluting stent due to late malapposition: Serial observations with optical coherence tomography Journal of Cardiology (2008) 52, 290 295 CASE REPORT Very late thrombosis of sirolimus-eluting stent due to late malapposition: Serial observations with optical coherence tomography Takahiro Sawada (MD),

More information

Protection of side branch is essential in treating bifurcation lesions: overview

Protection of side branch is essential in treating bifurcation lesions: overview Angioplasty Summit TCT Asia Pacific Seoul, April 26-28, 2006 Protection of side branch is essential in treating bifurcation lesions: overview Alfredo R Galassi, MD, FACC, FSCAI, FESC Head of the Catetherization

More information

MULTIVESSEL PCI. IN DRUG-ELUTING STENT RESTENOSIS DUE TO STENT FRACTURE, TREATED WITH REPEAT DES IMPLANTATION

MULTIVESSEL PCI. IN DRUG-ELUTING STENT RESTENOSIS DUE TO STENT FRACTURE, TREATED WITH REPEAT DES IMPLANTATION MULTIVESSEL PCI. IN DRUG-ELUTING STENT RESTENOSIS DUE TO STENT FRACTURE, TREATED WITH REPEAT DES IMPLANTATION C. Graidis, D. Dimitriadis, A. Ntatsios, V. Karasavvides Euromedica Kyanous Stavros, Thessaloniki.

More information

Clinical Considerations for CTO

Clinical Considerations for CTO 38 RCTs Clinical Considerations for CTO 18,000 pts Revascularization Whom to treat, Who derives benefit and What can we achieve? David E. Kandzari, MD FACC, FSCAI Director, Interventional Cardiology Research

More information

VASCULAR BRACHYTHERAPY USING A BETA EMITTER SOURCE IN DIABETIC PATIENTS WITH IN-STENT RESTENOSIS: ANGIOGRAPHIC AND CLINICAL OUTCOMES

VASCULAR BRACHYTHERAPY USING A BETA EMITTER SOURCE IN DIABETIC PATIENTS WITH IN-STENT RESTENOSIS: ANGIOGRAPHIC AND CLINICAL OUTCOMES doi:10.1016/s0360-3016(03)00537-6 Int. J. Radiation Oncology Biol. Phys., Vol. 57, No. 2, pp. 536 542, 2003 Copyright 2003 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/03/$ see front

More information

RESTENOSIS Facing up to the problem

RESTENOSIS Facing up to the problem RESTENOSIS Facing up to the problem Petr Kala University Hospital Brno Czech Republic ESC 2011, Paris Disclosure Scientific Advisory Boards or Education presentations fee Abbott, Boston Scientific, Cordis

More information

Influence of Planned Six-Month Follow-Up Angiography on Late Outcome After Percutaneous Coronary Intervention A Randomized Study

Influence of Planned Six-Month Follow-Up Angiography on Late Outcome After Percutaneous Coronary Intervention A Randomized Study Journal of the American College of Cardiology Vol. 38, No. 4, 2001 2001 by the American College of Cardiology ISSN 0735-1097/01/$20.00 Published by Elsevier Science Inc. PII S0735-1097(01)01476-0 Influence

More information

Adjunctive Stent Implantation Following Directional Coronary Atherectomy in Patients With Coronary Artery Disease

Adjunctive Stent Implantation Following Directional Coronary Atherectomy in Patients With Coronary Artery Disease JACC Vol. 32, No. 7 December 1998:1855 60 1855 Adjunctive Stent Implantation Following Directional Coronary Atherectomy in Patients With Coronary Artery Disease EZIO BRAMUCCI, MD,* LUIGI ANGOLI, MD,* PIERA

More information

The publication in January 2001 of the first-in-man

The publication in January 2001 of the first-in-man MINI-REVIEW: EXPERT OPINIONS Drug-Eluting Stents Cost Versus Clinical Benefit Pedro A. Lemos, MD; Patrick W. Serruys, MD, PhD; J. Eduardo Sousa, MD, PhD The publication in January 2001 of the first-in-man

More information

Coronary Aneurysms After Stent Placement: A Suggestion of Altered Vessel Wall Healing in the Presence of Anti-inflammatory Agents

Coronary Aneurysms After Stent Placement: A Suggestion of Altered Vessel Wall Healing in the Presence of Anti-inflammatory Agents 1524 JACC Vol. 18,. 6 Coronary Aneurysms After Stent Placement: A Suggestion of Altered Vessel Wall Healing in the Presence of Anti-inflammatory Agents S. TANVEER RAB, MD, FACC,* SPENCER B. KING III, MD,

More information

Paradigm Shifts in Cardiovascular Medicine

Paradigm Shifts in Cardiovascular Medicine Journal of the American College of Cardiology Vol. 43, No. 4, 2004 2004 by the American College of Cardiology Foundation ISSN 0735-1097/04/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2003.08.049

More information

Five-Year Follow-Up After Sirolimus-Eluting Stent Implantation

Five-Year Follow-Up After Sirolimus-Eluting Stent Implantation Journal of the American College of Cardiology Vol. 53, No. 17, 2009 2009 by the American College of Cardiology Foundation ISSN 0735-1097/09/$36.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2009.01.050

More information

Drug-eluting stents (DES) reduce the rate of restenosis and

Drug-eluting stents (DES) reduce the rate of restenosis and Original Articles Mechanisms of In-Stent Restenosis After Drug-Eluting Stent Implantation Intravascular Ultrasound Analysis Soo-Jin Kang, MD; Gary S. Mintz, MD; Duk-Woo Park, MD; Seung-Whan Lee, MD; Young-Hak

More information

FFR and intravascular imaging, which of which?

FFR and intravascular imaging, which of which? FFR and intravascular imaging, which of which? Ayman Khairy MD, PhD, FESC Associate professor of Cardiovascular Medicine Vice Director of Assiut University Hospitals Assiut, Egypt Diagnostic assessment

More information

Vascular Responses at Proximal and Distal Edges of Paclitaxel-Eluting Stents. Serial Intravascular Ultrasound Analysis From the TAXUS II Trial

Vascular Responses at Proximal and Distal Edges of Paclitaxel-Eluting Stents. Serial Intravascular Ultrasound Analysis From the TAXUS II Trial Vascular Responses at Proximal and Distal Edges of Paclitaxel-Eluting Stents Serial Intravascular Ultrasound Analysis From the TAXUS II Trial Patrick W. Serruys, MD; Muzaffer Degertekin, MD; Kengo Tanabe,

More information

DESolve NX Trial Clinical and Imaging Results

DESolve NX Trial Clinical and Imaging Results DESolve NX Trial Clinical and Imaging Results Alexandre Abizaid, MD, PhD, Instituto Dante Pazzanese, Sao Paulo, Brazil On behalf of the DESolve Nx Trial Investigators Please refer to the TCT2014 App or

More information

Mechanisms and Predictors of Carotid Artery Stent Restenosis A Serial Intravascular Ultrasound Study

Mechanisms and Predictors of Carotid Artery Stent Restenosis A Serial Intravascular Ultrasound Study Journal of the American College of Cardiology Vol. 47, No. 12, 2006 2006 by the American College of Cardiology Foundation ISSN 0735-1097/06/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.01.076

More information

Morphological changes after percutaneous transluminal coronary angioplasty of unstable plaques

Morphological changes after percutaneous transluminal coronary angioplasty of unstable plaques European Heart Journal (1996) 17, 1554-1559 orphological changes after percutaneous transluminal coronary angioplasty of unstable plaques Insights from serial angioscopic follow-up C. Bauters, J.-. Lablanche,

More information

The American College of Cardiology/American Heart

The American College of Cardiology/American Heart Prognostic Value of the Modified American College of Cardiology/American Heart Association Stenosis Morphology Classification for Long-Term Angiographic and Clinical Outcome After Coronary Stent Placement

More information

Intracoronary Stenting and Angiographic Results: Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO-2) Trial

Intracoronary Stenting and Angiographic Results: Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO-2) Trial Journal of the American College of Cardiology Vol. 41, No. 8, 2003 2003 by the American College of Cardiology Foundation ISSN 0735-1097/03/$30.00 Published by Elsevier Science Inc. doi:10.1016/s0735-1097(03)00119-0

More information

Why have interventional cardiologists salaries

Why have interventional cardiologists salaries The New 2013 Coronary Intervention Codes As of January 1, 2013, coronary intervention codes in use since 1992 were replaced by new codes with new values for complex interventions. By James C. Blankenship,

More information

Rationale and algorithm for below-the-knee acute gain optimization

Rationale and algorithm for below-the-knee acute gain optimization Leipzig, 01 Feb. 2018 Rationale and algorithm for below-the-knee acute gain optimization Flavio Airoldi, MD Flavio Airoldi Multimedica IRCCS Milan ITALY flavio.airoldi@multimedica.it Rationale Restenosis

More information

IN-STENT RESTENOSIS. K.Boerlage-van Dijk CarVasZ 2014

IN-STENT RESTENOSIS. K.Boerlage-van Dijk CarVasZ 2014 IN-STENT RESTENOSIS K.Boerlage-van Dijk CarVasZ 2014 Definition ISR Angiographic: recurrent diameter stenosis >50% at the stent segment or edges (5-mm segments adjacent to stent) Mehran system morphological

More information

IVUS Assessment of the Mechanism of In-stent Restenosis? Gary S. Mintz, MD Cardiovascular Research Foundation

IVUS Assessment of the Mechanism of In-stent Restenosis? Gary S. Mintz, MD Cardiovascular Research Foundation IVUS Assessment of the Mechanism of In-stent Restenosis? Gary S. Mintz, MD Cardiovascular Research Foundation SURE Trial: Restenosis in non-stented lesions Average of the two image slices with the smallest

More information

T he RAVEL and the SIRIUS studies, using a

T he RAVEL and the SIRIUS studies, using a 127 VIEWPOINT Drug eluting stents: maximising benefit and minimising cost J Gunn, A C Morton, C Wales, CMHNewman, D C Crossman, D C Cumberland... A policy of selective implantation of drug eluting stents,

More information

Sirolimus-Eluting Stents vs Vascular Brachytherapy for In-Stent Restenosis Within Bare-Metal Stents JAMA. 2006;295:

Sirolimus-Eluting Stents vs Vascular Brachytherapy for In-Stent Restenosis Within Bare-Metal Stents JAMA. 2006;295: ORIGINAL CONTRIBUTION JAMA-EXPRESS s vs Vascular for In- Restenosis Within Bare-Metal s The SISR Randomized Trial David R. Holmes, Jr, MD Paul Teirstein, MD Lowell Satler, MD Michael Sketch, MD James O

More information

Stent Thrombosis in Randomized Clinical Trials of Drug-Eluting Stents

Stent Thrombosis in Randomized Clinical Trials of Drug-Eluting Stents T h e n e w e ng l a nd j o u r na l o f m e dic i n e original article Thrombosis in Randomized Clinical Trials of Drug-Eluting s Laura Mauri, M.D., Wen-hua Hsieh, Ph.D., Joseph M. Massaro, Ph.D., Kalon

More information