A NEW CHAMBER FOR RAPID SPERM COUNT AND MOTILITY ESTIMATION

Size: px
Start display at page:

Download "A NEW CHAMBER FOR RAPID SPERM COUNT AND MOTILITY ESTIMATION"

Transcription

1 FERTILITY AND STERILITY Copyright ' 1978 The American Fertility Society Vol. 30, No.3, September 1978 Printed in U.S A. A NEW CHAMBER FOR RAPID SPERM COUNT AND MOTILITY ESTIMATION AMNON MAKLER, M.D. Fertility Clinic, Departments of Obstetrics and Gynecology "A" and "B," Rambam Medical Center, Technion, Faculty of Medicine, Haifa, Israel A new chamber for sperm count and motility estimation is described. This chamber, which is only 10 pm deep, enables free horizontal movement of spermatozoa in one focal plane and provides conditions for the examination of undiluted samples. Therefore, with the aid of this instrument it is possible to compare sperm motility in various samples from the same person or in different samples at different times. This can be done either by simple estimation or with any other method of motility evaluation chosen by the examiner. The sperm count can be made rapidly and directly from an undiluted, preheated sample by counting spermatozoa in the area of a grid located within the eyepiece; the count is expressed in millions per milliliter. Thirty-seven specimens were analyzed with this chamber. Statistical evaluation of the results revealed high precision, accuracy, and reliability of sperm counts when compared with the hemocytometric method. Better results were obtained when motility estimation was compared with the ordinary slide technique. Easy performance, rapid sperm counts, and improvement of motility estimation make this chamber a useful tool where sperm analysis is carried out. Fertil Steril 30:313, 1978 The most popular technique for sperm count involves the use of a hemocytometer. 1-5 The method consists of several steps: preparing 1:20 or 1:100 dilutions of the original sample, thoroughly mixing the diluent, inserting the diluted sample in a special way into the chamber, and allowing it to rest for about 20 minutes to permit the precipitation of spermatozoa to the bottom of the chamber before the counting itself is performed. These steps are necessary because the chamber, which is 100 JLm deep, does not permit undiluted semen to be examined. When such a sample is inserted into the chamber and viewed through the microscope, the spermatozoa are found to be too crowded, multilayered, cannot be seen in one focal plane, and cannot be counted. These several steps also increase inaccuracy, 6 are time-consuming, and discourage the untrained technician from using this method. The most popular method of spermatozoal motility evaluation is by microscopic examination Received February 14, 1978; revised March 23, 1978; accepted April 21, of a drop of fresh semen pressed between a slide and a cover slip. 1-3 The motility rate is estimated in one of several ways as chosen by the examiner: by estimating the percentage of motile spermatozoa, grading motility from + 1 to +4, or by more sophisticated methods of speed determination with a stopwatch 7 or with still cameras, 8 9 movie cameras, 10 or a closed-circuit TV system.11 There is one common point which devalues the results of all of these systems: since the cover slip is pressed with an uncontrolled pressure by the examiner, the thickness of the examined drop is not a standard one, since it can range from a few micrometers to several tenths of them. 12 As a result, the more the spermatozoa are pressed, the less freely they move, and vice versa. Moreover, if the drop of semen is too thick some spermatozoa will be out of focus and remain unnoticed, or may gain nonhorizontal movement. In a previous preliminary study 13 we were able to demonstrate how changes in the thickness of the examined drop can affect extensively the motility and the subsequent results. Therefore, comparing motil- 313

2 314 MAKLER ity rates of different samples when using the simple slide technique has limited value. We report here the invention of a new counting chamber which solves the above-mentioned problems of sperm count and motility evaluation. This chamber is only 10 p.m deep; this depth was chosen after preliminary experiments had shown that it fulfilled the following requirements: (1) All of the spermatozoa are located in one focal plane and all are seen, thus none is omitted by the observer. (2) The chamber provides for the free movement of the spermatozoa with minimal friction between surfaces of the chamber, so that spermatozoal speed is pradically unaffected. (3) The spermatozoa move horizontally with no vertical component that might affect the observer's interpretation when watching them from above. As a result, the semen may be analyzed without dilution, in one step and quickly. Since motility is established from a sample with standard depth, comparing various samples is possible with the highest accuracy. MATERIALS AND METHODS Basically, the new chamber* is constructed from two discs of high quality, transparent, optical flat glass; each disc is about 30 mm in diameter. The lower disc, which is 2 mm thick serves as a slide; the upper disc, which is 1 mm thick, is the cover glass. The tips of three pins located equidistant on the periphery of the lower glass are raised exactly 10 p.m above the plane, and when the cover glass is placed on those tips, a chamber of that thickness is thus obtained (Fig. 1). The accuracy of this space was checked interferometrically with a monochromatic laser beam and a distortion of less than 0.3 p.m was found. Technique of Semen Analysis. With the aid of a glass rod or a wooden stick, a drop of a well-mixed fresh seminal specimen is put.in the center ofthe lower glass and immediately covered. For observation, a microscope with a x20 objective and x 10 eyepiece is used. A grid calculated to cover an area of sq mm is inserted into the eyepiece to aid counting and motility estimation. Motility is estimated immediately after the drop is placed in the chamber. Although one can use any method for estimating motility subjectively or obj.ectively, the grid can be used for this purpose in the following way: First the nonmotile spermatozoa within the area of the grid are counted, then the motile spermatozoa are *Pending patent no , manufactured by EL-OP, Rehovoth, Israel. September 1978 FIG. 1. Schematic side view of the main part of the instrume?t. An undilute~ drop of specimen is placed between two optical flat glass shdes separated exactly 10 /Lm by the tips of three pins (one pin is omitted from the figure). counted, and their quality of movement is determined either by simple grading or by measuring the time elapsed for each spermatozoon to cross the area of the grid (about 150 x 200 p.m). After repeating this process in several fields of the examined drop to make it sufficient for statistical purposes, the percentage ofmotile spermatozoa is calculated and the quality of motility is also determined. These estimations can be repeated in several drops of the same specimen. Sperm Count. When the seminal fluid is not too concentrated and the spermatozoa are not too active, the count of live spermatozoa can be easily done directly in a fresh specimen. All spermatozoa whose heads are included within the borders of the grid (Fig. 2) are counted in the same way as blood cells are counted in a hemocytometer. This procedure is repeated in two more areas of.the drop by shifting the instrument on the stage of the microscope. The sum in the areas. counted represents the number of spermatozoa in onemillionth milliliter according to the following calculations: 0.01 mm x sq mm x 3 = cu mm = O.OOOOOlml. Therefore, adding six zeros to the total number in three different grid areas (e.g., =.72) gives their concentration (72,000,000) in 1 ml. When the spermatozoa are very active and concentrated, counting of live specimens is more difficult. In this case immobilization of the spermatozoa can be achieved by transferring part of the specimen into another test tube and inserting it into 50 to 60 C heated water for about 5 minutes (a cup containing a mixture of two parts boiling water and one part tap water is sufficient for this purpose). It is unwise to add spermicidal material such as formaldehyde or any detergent since it induces clumping, and errors result. Increasing the accuracy of the count can be achieved by counting more grid areas in the same drop or in several drops from the same specimen. In this

3 Vol. 30, No.3 A NEW CHAMBER FOR RAPID SPERM COUNT AND MOTILITY ESTIMATION 315 FIG. 2. Photomicrograph of an undiluted semen specimen located in the chamber, while a grid is placed within the eyepiece. The area covered by the grid is exactly sq mm when a x20 objective and a x10 eyepiece are used. By multiplying by 3 the-number of spermatozoa in the grid, or by counting them in three different grid areas, the concentration.in millions per milliliter is obtained. All spermatozoa are in one focal plane; none is blurred or omitted by the observer even in this undiluted specimen. case, calculation of the concentration is done according to the formula C = [(T x 3)/G] x 106, where T is the total number of spermatozoa within all grid a:reas and G is the number of grid areas explored in one or all drops examined. For example: 275 spermatozoa are counted in four grid areas in each of five drops from the same specimen. Then C = [(275 x 3)/20] x 10 6 = 41,250,000/ml. Difficulties. Real difficulties are extremely rare. Since samples need not be aspirated or inserted into capillary_ spaces, drops from very viscid samples can also be applied on the lower part of the chamber, being pressed uniformly into the 10~/Lm space by the cover glass. Specimens that remain persistently unliquefied induce problems which are common to any technique that requires sampling, such as the ordinary hemocytometric method. In such cases liquefying agents should be added to the specimen before counting is done. Difficulties in mixing spermatozoa homogenously in a viscid sample may lead to an uneven distribution of spermatozoa within the chamber. In this case more grid areas from the same drop or from several drops should be explored and more spermatozoa should be counted. When spermagglutination occurs, counting is done in areas devoid of agglutinated spermatozoa, with a special note about the. existence of this phenomenon and its estimated intensity. Very oligospermic or hyperspermic specimens do not present any difficulties. In the first case, one may find many "empty" grid areas among others containing one or two spermatozoa per grid. If 20 to 30 grid areas are selected randomly for exploration and those "empty" grids are intable 1. Frequency of Distribution in Fifteen Counts in Five Drops of a Normospermic Specimen Drop no millions spermatozoa/ml Mean Mean sperm count = 127 million; SD = 8.7; CV = 6.69.

4 316 MAKLER September 1978 TABLE 2. Means± Standard Deviation often to Fifteen Counts in Each of Thirty-Seven Specimens Divided into Five Subgroups According to Concentration at Bottom Range of counts ~10 1~20 3.0± ± ± ± ± ± ± ± 3.6 Counted grids Mean± SD 6.3± ± 3.4 cv a CV = (SD X 100)/x. eluded in the above formula, accurate results are obtained even in cases where concentration is below 100,000 spermatozoa/mi. It never occurred to us that any hyperspermic specimen could not be counted in its own undiluted medium. There is enough space for all sperm heads to be seen clearly within the area of the grid, with no overlapping heads. Statistical Evaluation. For this purpose, 37 specimens from normal, oligospermic, and asthenospermic males were examined. Sperm count and motility were evaluated two or three times in five drops of every specimen as described above. The results were statistically analyzed for accuracy and precision. In 5 of 37 cases, the same specimens were examined by the same technician in three different instruments of 10 p,m, and the results of the semen analyses were compared for reproducibility. In 18 cases the same specimens were also examined by a qualified technician using the ordinary method of semen analysis: in every sample the percentage of motility was estimated from 10 view fields of a regular slide according to the technique described by Eliason. 12 A count was done in four different diluted drops (1:20 or 1:100), using the Neubauer hemocytometer where spermatozoa were counted in 25 squares of0.04 sq mm, which is roughly equivalent to the area covered in our counts. All results were compared statistically. RESULTS Table 1 describes the frequency of distribution of 15 counts performed in five different drops in 2~50 5~100 10~200 millionslml 22.0± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± one of the normospermic specimens, the average of the counts of each drop, and the average of all 15 counts as well as the standard deviation and coefficient of variance. From Table 1 it can be seen that the distribution of the counts in this concentration is very small. Table 2 shows the average and standard deviation of 10 to 15 counts carried out for each of the 37 specimens divided into subgroups according to their concentration. For each subgroup, the average, standard deviation (SD), and coefficient of variance (CV) are given. It can be seen that, whereas the values of SD are increasing with the concentration, the values of CV are decreasing. This means that, practically, fewer counts are necessary in samples of 50 x 10 6 or more. Oligospermic samples need more counts to increase accuracy, although results such as 5 ± 3 million/ ml are clinically adequate. Table 3 compares results of 15 counts from 5 various specimens carried out in three different 10-p,m instruments. The difference between instruments is very small and reliability is fairly high. Figure 3 shows the correlation of 18 sperm counts each performed both with the 10-p,m TABLE 3. Results of Fifteen Counts of Five Specimens Carried Out in Three Different Chamber Instruments Chamber Mean (X) Mean Mean II III r,;- x) r,; ->Mx%

5 Vol. 30, No. 3 A NEW CHAMBER FOR RAPID SPERM COUNT AND MOTILITY ESTIMATION r , N=18 Y= x Fra. 3. Results of18 sperm counts obtained with the aid of a 10-1-'m chamber plotted against results of counts performed with an ordinary hemocytometer. Each point indicates the results of two counts for each specimen as described in the text. chamber and with an ordinary m hemocytometer. The average of differences between the instruments did not exceed 5.4% and was almost independent of sperm concentration except in very low concentrations. The results are highly correlated, and the reliability of the m chamber is evident. Table 4 reveals results of motility estimation in 50 view fields of 3 different specimens each examined by the ordinary slide method and also with the 10-J.Lm chamber. The average motility and SD of each specimen using both techniques are compared. Although the results were closely related, those of the slide method were more widely distributed, almost twice as much as those with the 10-J.Lm chamber technique. DISCUSSION The above results show that sperm counts using the new 10-J.Lm-chamber are accurate, precise, and reliable even though the technique is very TABLE 4. Percentage of Motility Estimated by Observing Fifty Fields in Each of Three Different Specimens in Ordinary Slide and 10-~ Chamber Estimated 10-JLm Chamber Estimated Ordinary slide motility SD cv motility SD cv 18% % 14% % 40% % 36% % 49% % 47% % simple and rapid. Moreover, motility estimation is more accurate than estimation done with the ordinary slide method. The constant thickness of the examined drop excludes external physical factors which may induce differences between samples of the same specimen or between various specimens that need to be compared, whatever the method of estimation chosen by the observer. The fact that the chamber is shallow enough to enable undiluted samples to be examined is very important for estimation of sperm motility in its own natural medium, unless dilution is purposely desired. Therefore, this instrument can be an important tool for every clinical or research study, when the effect of a certain factor on sperm motility needs an accurate evaluation or comparison of motility rates. When one compares the technique for use of this chamber with that for use of a hemocytometer the following advantages can be mentioned: 1. While both sperm count and motility estimation can be done in the same m chamber, only the sperm count can be done with the hemocytometer. 2. As sperm dilution is not necessary, no accessories such as pipettes, solutions, and facilities for pipette flushing and drying are needed. Nor is special training for these procedures necessary. 3. As the chamber is very thin and all spermatozoa can be seen immediately and in one focal plane, there is no need to wait for sperm to be precipitated on the bottom of the chamber, a process which takes about 20 minutes when using the hemocytometer. 4. The grid located within the eyepiece allows one to count in the same drop as many grid areas as one wishes. This can be accomplished by shifting the instrument on the stage of the microscope from one area to another. Therefore, artefacts, bubbles, and pieces of dirt or clumped spermatozoa can be ignored and do not interfere with counts as might occur with the fixed scale in the ordinary hemocytometer. All of these advantages make this instrument a helpful tool in every institute, laboratory, or office where semen analysis is done. Acknowledgments. The author wishes to thank Professor Lipson from the Haifa Technion for his interferometric analysis of the instruments, Mr. Amnon Levy for his useful advice, Mr. N. Kipperman for his skillful assistance in constructing the model of the chamber, Mrs. Oma Heidenfeld for preparing the results, and my wife, Y osepha, for her statistical analysis of the results.

6 318 MAKLER REFERENCES 1. Freund M, Peterson RN: In Human Semen and Fertility Regulation in Men, Edited by ESE Hafez. St. Louis, CV Mosby Co, 1976, p Eliasson R: Standards for investigation of human semen. Andrologia 3:49, Amelar RD: Infertility in Men. Philadelphia, FH Davis Co, 1966, p Schirren C: Practical Andrology. Berlin, Verlag Bruder, 1972, p Bauer JD, Ackerman PG, Toro G: Clinical Laboratory Methods. St. Louis, CV Mosby Co, 1974, p Freund M, Carol B: Factors affecting hemocytometer counts of sperm concentration in human semen. J Reprod Fertil 8:149, Harvey C: The speed of human spermatozoa and the effect on it of various diluents. J Reprod Fertil1:84, 1 }60 September Rothschild L: The movement of spermatozoa. In Mammalian Germ Cells, Edited by GEW Wolstenholme. Boston, Little, Brown and Co., 1953, p Janick J, MacLeod J: Measurement of human spermatozoa motility. Fertil Steril 21:140, Rikmenspoel R, Harren GV: A microscopic film of the movement of bull sperm cells in dark-field illumination. Experimenter Xll:160, Jecht EW, Russo JJ: A system for the quantitative analysis ofhuman sperm motility. Andrologia 5:215, Eliasson R: Analysis of semen. In Progress in Infertility, Edited by SJ Behrman, RW Kistner. Boston, Little, Brown and Co, 1975, p Makler A: The thickness of the microscopically examined sample and its relationship to motility estimation. Int J Androl 1:213, 1978

A NEW MULTIPLE EXPOSURE PHOTOGRAPHY METHOD FOR OBJECTIVE HUMAN SPERMATOZOAL MOTILITY DETERMINATION

A NEW MULTIPLE EXPOSURE PHOTOGRAPHY METHOD FOR OBJECTIVE HUMAN SPERMATOZOAL MOTILITY DETERMINATION FERTILITY AND STERILITY Copyright < 1978 The American Fertility Society Vol. 30, No.2, August 1978 Printed in U.S A. A NEW MULTIPLE EXPOSURE PHOTOGRAPHY METHOD FOR OBJECTIVE HUMAN SPERMATOZOAL MOTILITY

More information

SEMEN ANALYSIS. A minimum of 2 days and a maximum of 7 days of sexual abstinence.

SEMEN ANALYSIS. A minimum of 2 days and a maximum of 7 days of sexual abstinence. 2nd Symposium of the Croatian Society of Clinical Embryologists and Andrology Workshop Workstation 1 SEMEN ANALYSIS Evaluation of male fertility: Testis function and male genital tract; accessory sex glands

More information

Counting Spermatozoa - Part 1 By B. W. Pickett, Ph.D. Professor Emeritus, Colorado State University

Counting Spermatozoa - Part 1 By B. W. Pickett, Ph.D. Professor Emeritus, Colorado State University Counting Spermatozoa - Part 1 By B. W. Pickett, Ph.D. Professor Emeritus, Colorado State University A hemacytometer, originally designed to count blood cells, has been used for many years to count other

More information

Microscope Requirements

Microscope Requirements SEMEN EVALUATION EQUIPMENT Microscope Requirements Good quality lenses Phase-contrast preferred for % progressive motility evaluations Objectives 10X, 20X*, 40X*, 100X, minimum Heated stage preferred *Preferably

More information

Spectrophotometric Quantitation of Mammalian Spermatozoon Motility I. Human

Spectrophotometric Quantitation of Mammalian Spermatozoon Motility I. Human BIOLOGY OF REPRODUCTION 18, 624-628 (1978) Spectrophotometric Quantitation of Mammalian Spermatozoon Motility I. Human R.W. ATI-IERTON, E.W. RADANY and K.L. POLAKOSKI2 Department of Zoology and Physiology,

More information

FACTORS AFFECTING SPERM MOTILITY. II. HUMAN SPERM VELOCITY AND PERCENTAGE OF MOTILITY AS INFLUENCED BY SEMEN DILUTION*

FACTORS AFFECTING SPERM MOTILITY. II. HUMAN SPERM VELOCITY AND PERCENTAGE OF MOTILITY AS INFLUENCED BY SEMEN DILUTION* FERTlJTY AND SrERLTY Copyright 1979 The American Fertility Society Vol. 32. No.4. October 1979 Printed in U.SA. FACTORS AFFECTNG SPERM MOTLTY.. HUMAN SPERM VELOCTY AND PERCENTAGE OF MOTLTY AS NFLUENCED

More information

FACTORS AFFECTING SPERM MOTILITY. V. WASHING AND RESUSPENSION OF HUMAN SPERMATOZOA IN VARIOUS ARTIFICIAL MEDIA

FACTORS AFFECTING SPERM MOTILITY. V. WASHING AND RESUSPENSION OF HUMAN SPERMATOZOA IN VARIOUS ARTIFICIAL MEDIA FERTILITY AND STERILITY Copyright c 1981 The American Fertility Society Vol. 35, No.4, April 1981 Printed in U.s.A. FACTORS AFFECTING SPERM MOTILITY. V. WASHING AND RESUSPENSION OF HUMAN SPERMATOZOA IN

More information

New Technic for Sperm-Mucus Penetration Tests Using a Hemocytometer. H. R. Guard, M.D. *

New Technic for Sperm-Mucus Penetration Tests Using a Hemocytometer. H. R. Guard, M.D. * New Technic for Sperm-Mucus Penetration Tests Using a Hemocytometer H. R. Guard, M.D. * LIE CERVICAL MU~US and its penetration by spermatozoa is the first and the most important barrier in the process

More information

isperm Technical Document

isperm Technical Document isperm Technical Document Content 1.Operating Procedure........2 Set up isperm Set up isperm App Semen preparation Sampling Using isperm App 2. The principle of isperm......4 Concentration Motility 3.Photometer,

More information

FACTORS AFFECTING SPERM MOTILITY. III. INFLUENCE OF VISIBLE LIGHT AND OTHER ELECTROMAGNETIC RADIATIONS ON HUMAN SPERM VELOCITY AND SURVIVAL*

FACTORS AFFECTING SPERM MOTILITY. III. INFLUENCE OF VISIBLE LIGHT AND OTHER ELECTROMAGNETIC RADIATIONS ON HUMAN SPERM VELOCITY AND SURVIVAL* FRTILITY AND STRILITY Copyright " 198 The American Fertility Society Vol. 33, No.4, April 198 Printed in U.SA. FACTORS AFFCTING SPRM MOTILITY. III. INFLUNC OF VISIBL LIGHT AND OTHR LCTROMAGNTIC RADIATIONS

More information

The design of the Micro-Cell chamber was based on the physical properties of capillary flow, MATERIALS AND METHODS The Micro-Cell Chamber

The design of the Micro-Cell chamber was based on the physical properties of capillary flow, MATERIALS AND METHODS The Micro-Cell Chamber FERTILITY AND STERILITY Copyright " 1990 The American Fertility Society Vol. 53, No.5, May 1990 Printed on ocid-free poper in U.S.A. The influence of chamber characteristics on the reliability of sperm

More information

Computer-aided evaluation of assessment of grade a spermatozoa by experienced technicians

Computer-aided evaluation of assessment of grade a spermatozoa by experienced technicians Computer-aided evaluation of assessment of grade a spermatozoa by experienced technicians Trevor G. Cooper, Ph.D., and Ching-Hei Yeung, Ph.D. Institute of Reproductive Medicine of the University of Münster,

More information

FACTORS AFFECTING SPERM MOTILITY. VII. SPERM VIABILITY AS AFFECTED BY CHANGE OF ph AND OSMOLARITY OF SEMEN AND URINE SPECIMENS

FACTORS AFFECTING SPERM MOTILITY. VII. SPERM VIABILITY AS AFFECTED BY CHANGE OF ph AND OSMOLARITY OF SEMEN AND URINE SPECIMENS FERTLTY AND STERLTY Copyright c 1981 The American Fertility Society Vol. 36, No.4, October 1981 Printed in U.SA. FACTORS AFFECTNG SPERM MOTLTY. V. SPERM VABLTY AS AFFECTED BY CHANGE OF ph AND OSMOLARTY

More information

FACTORS AFFECTING SPERM MOTILITY. VI. SPERM VIABILITY UNDER THE INFLUENCE OF BACTERIAL GROWTH IN HUMAN EJACULATES

FACTORS AFFECTING SPERM MOTILITY. VI. SPERM VIABILITY UNDER THE INFLUENCE OF BACTERIAL GROWTH IN HUMAN EJACULATES FERTILITY AND STERILITY Copyright c 191 The American Fertility Society Vol. 3, No.6, June 191 Printed in U.SA. FACTORS AFFECTING SPERM MOTILITY. VI. SPERM VIABILITY UNDER THE INFLUENCE OF BACTERIAL GROWTH

More information

Manual versus computer-automated semen analyses. Part I. Comparison of counting chambers

Manual versus computer-automated semen analyses. Part I. Comparison of counting chambers FERTILITY AND STERILITY Vol 65, No I, January 1996 Copyright c 1996 American Society for Reproductive Medicine Printed on acid free paper in U S A Manual versus computer-automated semen analyses Part I

More information

Abnormalities of Spermatogenesis

Abnormalities of Spermatogenesis Abnormalities of Spermatogenesis Male Factor 40% of the cause for infertility Sperm is constantly produced by the germinal epithelium of the testicle Sperm generation time 73 days Sperm production is thermoregulated

More information

**Florida licensees, please note: This exercise will appear in CE Broker under the specialty of Andrology.

**Florida licensees, please note: This exercise will appear in CE Broker under the specialty of Andrology. EDUCATIONAL COMMENTARY SEMEN ANALYSIS AND FERTILITY Commentary provided by: Henry C. Lee, CLS, MASCP, CPHQ Laboratory Manager Arrowhead Regional Medical Center Colton, CA Educational commentary is provided

More information

In Vitro Speeds of Bovine Spermatozoa

In Vitro Speeds of Bovine Spermatozoa In Vitro Speeds of Bovine Spermatozoa A. N. Moeller, M.S., and N. l. VanDemark, Ph.D. THE RATE OF progressive movement of the spermatozoa has been used as one criterion in physiologic studies for evaluation

More information

Evaluation of a New Contraceptive

Evaluation of a New Contraceptive ,. Evaluation of a New Contraceptive Aquiles J. Sobrero, M.D. To EVALUATE a new chemical contraceptive, a series of tests are employed to measure its spermicidal efficacy, harmlessness, and acceptability.5

More information

Evaluation of the CeliSoft* automated semen analysis system in a routine laboratory settingt

Evaluation of the CeliSoft* automated semen analysis system in a routine laboratory settingt r FERTILITY AND STERILITY Copyright e 1988 The American Fertility Society Printed in U.S.A. Evaluation of the CeliSoft* automated semen analysis system in a routine laboratory settingt David Mortimer,

More information

although work THE TOXICITY OF VARIOUS NON-ELECTROLYTES TO HUMAN SPERMATOZOA AND THEIR PROTECTIVE EFFECTS DURING FREEZING

although work THE TOXICITY OF VARIOUS NON-ELECTROLYTES TO HUMAN SPERMATOZOA AND THEIR PROTECTIVE EFFECTS DURING FREEZING THE TOXICITY OF VARIOUS NON-ELECTROLYTES TO HUMAN SPERMATOZOA AND THEIR PROTECTIVE EFFECTS DURING FREEZING D. W. RICHARDSON and R. M. F. S. SADLEIR Endocrine Unit, University College Hospital, London,

More information

SQA-V Gold Semiannual Validation/Proficiency/QC Recommendations

SQA-V Gold Semiannual Validation/Proficiency/QC Recommendations MEDICAL ELECTRONIC SYSTEMS 5757 W. Century Blvd. Suite 805 Los Angeles, CA. 90045 Remember, it ALL Started with a Sperm! www.mes global.com SQA-V Gold Semiannual Validation/Proficiency/QC Recommendations

More information

Accuracy and precision of computer-aided sperm analysis in multicenter studies*t

Accuracy and precision of computer-aided sperm analysis in multicenter studies*t FERTILITY AND STERILITY Copyright 1992 The American Fertility Society Printed on acid-free paper in U.S.A. Accuracy and precision of computer-aided sperm analysis in multicenter studies*t Russell O. Davis,

More information

DEPENDENCE OF SPERM MOTILITY AND RESPIRATION ON OXYGEN CONCENTRATION

DEPENDENCE OF SPERM MOTILITY AND RESPIRATION ON OXYGEN CONCENTRATION DEPENDENCE OF SPERM MOTILITY AND RESPIRATION ON OXYGEN CONCENTRATION ABRAHAM C. NEVO A.R.C. Unit of Reproductive Physiology and Biochemistry, Cambridge, England {Received 22nd June 1964) Summary. Motility

More information

Journal of Assisted Reproduction and Genetics Plenum Publishing Corporation All Rights Reserved.

Journal of Assisted Reproduction and Genetics Plenum Publishing Corporation All Rights Reserved. Journal of Assisted Reproduction and Genetics Plenum Publishing Corporation 2003. All Rights Reserved. Volume 20(1) January 2003 p 38 45 Can Westgard Quality Control Rules Determine the Suitability of

More information

[388] COUNTING SPERMATOZOA

[388] COUNTING SPERMATOZOA [388] COUNTING SPERMATOZOA BY LORD ROTHSCHILD From the Department of Zoology, University of Cambridge, and the Marine Station, Millport (Received 13 July 1949) (With Three Text-figures) INTRODUCTION The

More information

Differences in Human Spermatozoa

Differences in Human Spermatozoa Differences in Human Spermatozoa Landrum B. Shettles, M.D. RECENT OBSERVATIONS show that thin, dried, unstained smears of human spermatozoa viewed through the dry objectives of the phase-contrast microscope

More information

Sergey I. Moskovtsev and Clifford L. Librach

Sergey I. Moskovtsev and Clifford L. Librach Chapter 2 Methods of Sperm Vitality Assessment Sergey I. Moskovtsev and Clifford L. Librach Abstract Sperm vitality is a re fl ection of the proportion of live, membrane-intact spermatozoa determined by

More information

Evaluation of a disposable plastic Neubauer counting chamber for semen analysis

Evaluation of a disposable plastic Neubauer counting chamber for semen analysis TECHNIQUES AND INSTRUMENTATION Evaluation of a disposable plastic Neubauer counting chamber for semen analysis Jackson Kirkman-Brown, Ph.D., a and Lars Bj orndahl, M.D., Ph.D. a,b a Centre for Human Reproductive

More information

THAT SEMINAL FLUID has functions in the process of conception other than

THAT SEMINAL FLUID has functions in the process of conception other than Influence of Seminal Fluid on Sperm Motility William Marden and N. T. Werthessen, Ph.D. THAT SEMINAL FLUID has functions in the process of conception other than to serve as a mere transport medium for

More information

System overview Installation System Description System Default Settings and Loading I-Button Tests Components/Kits /Accessories Maintenance & Cleaning

System overview Installation System Description System Default Settings and Loading I-Button Tests Components/Kits /Accessories Maintenance & Cleaning SQA-VISION TRAINING Content of Presentation System overview Installation System Description System Default Settings and Loading I-Button Tests Components/Kits /Accessories Maintenance & Cleaning SQA-Vision

More information

TURBIDIMETRIC ANALYSIS OF HUMAN SPERM MOTILITY*

TURBIDIMETRIC ANALYSIS OF HUMAN SPERM MOTILITY* FERTILITY AND STERILITY Copyright' 1977 The American Fertility Society Vol. 28, No. 12, December 1977 Printed in V.SA. TURBIDIMETRIC ANALYSIS OF HUMAN SPERM MOTILITY* JOSEPH E. SOKOLOSKI, B.A. LUIS BLASCO,

More information

Issued by: LABORATORY MANAGER Original Date: March 20, 2000 Approved by: Laboratory Director Revision Date: July 26, 2000 CRYPTOCOCCAL ANTIGEN

Issued by: LABORATORY MANAGER Original Date: March 20, 2000 Approved by: Laboratory Director Revision Date: July 26, 2000 CRYPTOCOCCAL ANTIGEN Policy # MI\TECH\11\v01 Page 1 of 6 Issued by: LABORATORY MANAGER Original Date: March 20, 2000 Approved by: Laboratory Director Revision Date: July 26, 2000 CRYPTOCOCCAL ANTIGEN Latex particles coated

More information

Scientific Inquiry Review

Scientific Inquiry Review Scientific Inquiry Review Adapted from Regentsprep.org Be able to USE AND APPLY the steps of the scientific method to solve a problem and design an experiment: Scientific Method: 1. Make observations about

More information

CRYOTOP SAFETY KIT Protocol. Cryotop Method

CRYOTOP SAFETY KIT Protocol. Cryotop Method CRYOTOP SAFETY KIT Protocol Cryotop Method R Vitrification PART Materials Required Cryotop Safety Kit-Vitrification No.0 Basic Solution (BS): 1 X 1.5ml vial (Only for Oocyte Vitrification) No.1 Equilibration

More information

THE SPERM-CERVICAL MUCUS CONTACT TEST: A PRELIMINARY REPORT

THE SPERM-CERVICAL MUCUS CONTACT TEST: A PRELIMINARY REPORT FERTILITY AND STERILITY Copyright" 97 The American Fertility Society Vol. 27, No.3, March 97 Printed in U.SA. THE SPERM-CERVICAL MUCUS CONTACT TEST: A PRELIMINARY REPORT JAN KREMER, M.D., AND SIEMEN JAGER,

More information

Semen. Dr. Mohamed Saad Daoud

Semen. Dr. Mohamed Saad Daoud Semen Dr. Mohamed Saad Daoud 1 Reference Books: Urinanalysis and body fluids (Susan King Strasinger- Marjorie Schaub De Lorenzo) Fifth edition Dr. Mohamed Saad Daoud 2 Diagram of the male genitalia Dr.

More information

Investigation of sickle cell disease

Investigation of sickle cell disease Investigation of sickle cell disease Sickle cell disease Sickle cell disease is an inherited disorder which affects people from Africa, India, Middle East and the Mediterranean region The main forms of

More information

See external label 2 C-8 C = C-REACTIVE PROTEIN (CRP) LATEX SLIDE TEST

See external label 2 C-8 C = C-REACTIVE PROTEIN (CRP) LATEX SLIDE TEST CORTEZ DIAGNOSTICS, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

Methods Using Limits of Agreement

Methods Using Limits of Agreement Journal of Andrology, Vol. 28, No. 3, May/June 2007 Copyright E American Society of Andrology Validation of Sperm Counting Methods Using Limits of Agreement Andrology Lab Corner* ELEANOR BAILEY, NICK FENNING,

More information

Sperm motility index: a new parameter for human sperm evaluation

Sperm motility index: a new parameter for human sperm evaluation r FERTILITY AND STERILITY Copyright 1991 The American Fertility Society Yol. 56, No.1, July 1991 Printed on acid-free paper in U.S.A. Sperm motility index: a ne parameter for human sperm evaluation Benjamin

More information

1.0 Purpose - This procedure specifies the method for conducting analysis for semen and sperm in forensic casework.

1.0 Purpose - This procedure specifies the method for conducting analysis for semen and sperm in forensic casework. Procedure for Semen and Sperm Analysis 1.0 Purpose - This procedure specifies the method for conducting analysis for semen and sperm in forensic casework. 2.0 Scope - This procedure applies to those Forensic

More information

Intracytoplasmic Sperm Injection (ICSI) with the Eppendorf micromanipulator TransferMan 4m

Intracytoplasmic Sperm Injection (ICSI) with the Eppendorf micromanipulator TransferMan 4m APPLICATION NOTE No. 009 I June 2013 Intracytoplasmic Sperm Injection (ICSI) with the Eppendorf micromanipulator TransferMan 4m Verena Nordhoff, Centre of Reproductive Medicine and Andrology, University

More information

TECHNICAL GUIDANCE FOR THE ACCREDITATION OF ANDROLOGY LABORATORIES

TECHNICAL GUIDANCE FOR THE ACCREDITATION OF ANDROLOGY LABORATORIES TECHNICAL GUIDANCE FOR THE ACCREDITATION OF ANDROLOGY LABORATORIES Approved By: Chief Executive Officer: Ron Josias Accreditation Executive: Mpho Phaloane Revised By: Medical Specialist Technical Committee

More information

Unit B: Anatomy and Physiology of Poultry. Lesson 4: Artificial Poultry Reproduction

Unit B: Anatomy and Physiology of Poultry. Lesson 4: Artificial Poultry Reproduction Unit B: Anatomy and Physiology of Poultry Lesson 4: Artificial Poultry Reproduction 1 1 Terms Artificial insemination Sexed semen Standing heat 2 2 I. Artificial insemination is the placing of semen in

More information

Semen LEARNING OBJECTIVES KEY TERMS

Semen LEARNING OBJECTIVES KEY TERMS 13044C11.PGS 7/11/02 4:30 PM Page 169 C H A P T E R 11 Semen LEARNING OBJECTIVES Upon completion of this chapter, the reader will be able to: 1 Describe the four components of semen with regard to source

More information

COUNTING LIVE AND DEAD BULL SPERMATOZOA

COUNTING LIVE AND DEAD BULL SPERMATOZOA [] COUNTING LIVE AND DEAD BULL SPERMATOZOA BY R. C. CAMPBELL, J. L. HANCOCK AND LORD ROTHSCHILD The A.R.C. Unit of Animal Reproduction and the Department of Zoology, University of Cambridge (Received May

More information

A FLAT CAPILLARY TUBE SYSTEM FOR ASSESSMENT OF SPERM MOVEMENT IN CERVICAL MUCUS*

A FLAT CAPILLARY TUBE SYSTEM FOR ASSESSMENT OF SPERM MOVEMENT IN CERVICAL MUCUS* FERTILITY AND STERILITY Copyright 1978 The American Fertility Society Vol. 29, No. I, January 1978 Printed in U.S.A. A FLAT CAPILLARY TUBE SYSTEM FOR ASSESSMENT OF SPERM MOVEMENT IN CERVICAL MUCUS* ROSS

More information

A MICRO TIME METHOD FOR DETERMINATION OF REDUCING SUGARS, AND ITS APPLICATION TO ANALYSIS OF BLOOD AND URINE.

A MICRO TIME METHOD FOR DETERMINATION OF REDUCING SUGARS, AND ITS APPLICATION TO ANALYSIS OF BLOOD AND URINE. A MICRO TIME METHOD FOR DETERMINATION OF REDUCING SUGARS, AND ITS APPLICATION TO ANALYSIS OF BLOOD AND URINE. BY JAMES A. HAWKINS. (From Ike Hospital of The Rockefeller Institute for Medical Research,

More information

Reducing Inter-Laboratory Differences between Semen Analyses Using Z Score and Regression Transformations

Reducing Inter-Laboratory Differences between Semen Analyses Using Z Score and Regression Transformations Article Reducing Inter-Laboratory Differences between Semen Analyses Using Z Score and Regression Transformations Esther Leushuis, M.D., PhD., 2 *, Alex Wetzels, Ph.D. 3, Jan Willem van der Steeg, M.D.,

More information

The Male Factor in Fertility and Infertility

The Male Factor in Fertility and Infertility The Male Factor in Fertility and Infertility III. An Analysis of Motile Activity in the Spermatozoa of 1000 Fertile Men and 1000 Men in Infertile Marriage John Macleod, Ph.D., and Ruth Z. Gold, M.A. IN

More information

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2 For detection and measurement of human interleukin 2 Catalog #02006 Catalog #02007 2 Plates 10 Plates Product Description The Human Interleukin 2 (IL-2) ELISA Kit is designed for the quantitative detection

More information

Experimental Procedure

Experimental Procedure 1 of 10 9/13/2018, 3:52 PM https://www.sciencebuddies.org/science-fair-projects/project-ideas/foodsci_p074/cooking-food-science/juice-balls-science-of-spherification (http://www.sciencebuddies.org/science-fair-projects/project-ideas/foodsci_p074/cooking-food-science/juice-balls-science-of-spherification)

More information

Intracytoplasmic Sperm Injection (ICSI) with the Eppendorf TransferMan 4m and CellTram 4m

Intracytoplasmic Sperm Injection (ICSI) with the Eppendorf TransferMan 4m and CellTram 4m APPLICATION NOTE No. 009 Intracytoplasmic Sperm Injection (ICSI) with the Eppendorf TransferMan 4m and CellTram 4m Verena Nordhoff, Centre of Reproductive Medicine and Andrology, University Hospital of

More information

SPERM ANTIBODIES IN INFERTILE COUPLES*,t

SPERM ANTIBODIES IN INFERTILE COUPLES*,t F'ERTH.lTY AND STERn.ITY Copyright' 1971 by The Williams & Wilkins Co. Vol. 22, No.5, May 1971 Printed in U.S.A. SPERM ANTIBODIES IN INFERTILE COUPLES*,t Rum ANSBACHER, M.D., M.S.,:j:SOL MANARANG-PANGAN,

More information

WAVE DYNAMYCS OF HUMAN SPERM MOTILITY

WAVE DYNAMYCS OF HUMAN SPERM MOTILITY Page 1 of 14 WAVE PARAMETERS OF THE SPERM FLAGELLUM AS PREDICTORS OF HUMAN SPERMATOZOA MOTILITY Andrologia 30: 153-157, 1998 O. Vera 1, M.G. Muñoz 2 and K. Jaffe 2 1 Facultad de Ciencias Veterinarias,

More information

General Description Measurement Procedure.. 5. Measuring Tubes Filling the Measuring Tube Reading the SR-6 Vernier 6

General Description Measurement Procedure.. 5. Measuring Tubes Filling the Measuring Tube Reading the SR-6 Vernier 6 110-065 1/04 2 General Description..... 4 Measurement Procedure.. 5 Measuring Tubes... 5 Filling the Measuring Tube... 5 Reading the SR-6 Vernier 6 Rotation Determinations 6 Examples and Solutions to Polarimetric

More information

Pollen Slide Mounting Protocol

Pollen Slide Mounting Protocol Pollen Slide Mounting Protocol Materials: Syn-Matrix mounting medium Microcentrifuge Microscope slides Slide coverslips (18mm x 18mm) Coverslip podium (see Figure 1) Capillary tubes Dissecting microscope

More information

LECTURE 13. Dr. Teresa D. Golden University of North Texas Department of Chemistry

LECTURE 13. Dr. Teresa D. Golden University of North Texas Department of Chemistry LECTURE 13 Dr. Teresa D. Golden University of North Texas Department of Chemistry Goniometer circle - centered at the sample, with the x-ray source and detector on the circumference of the circle. Focusing

More information

Gambro Renal Products Technical Assistance Services

Gambro Renal Products Technical Assistance Services Gambro Renal Products Technical Assistance Services http://tech.usa-gambro.com 800/525-2623 303/232-6800 Technical Update PH,TR,11060 12/2008 TROUBLESHOOTING Equipment Type: Phoenix Dialysis System Subject:

More information

ARTIFICIAL INSEMINATION HOMOLOGOUS WITH OLIGOSPERMIC SEMEN SEPARATED ON ALBUMIN COLUMNS*

ARTIFICIAL INSEMINATION HOMOLOGOUS WITH OLIGOSPERMIC SEMEN SEPARATED ON ALBUMIN COLUMNS* FERTILITY AND STERILITY Copyright 1979 The American Fertility SOciety Vol. 31, No. I, January 1979 Printed in V.SA. ARTIFICIAL INSEMINATION HOMOLOGOUS WITH OLIGOSPERMIC SEMEN SEPARATED ON ALBUMIN COLUMNS*

More information

Mouse C-Peptide ELISA Kit

Mouse C-Peptide ELISA Kit Mouse C-Peptide ELISA Kit Cat.No: DEIA4507 Lot. No. (See product label) Size 96T Intended Use The Mouse C-Peptide ELISA kit is for the quantitative determination of c-peptide in mouse serum, plasma, and

More information

LESSON ASSIGMMENT. After completing this lesson, you should be able to:

LESSON ASSIGMMENT. After completing this lesson, you should be able to: LESSON ASSIGMMENT LESSON 5 Manual Cell Counts. TEXT ASSIGNMENT Paragraphs 5-1 through 5-6. LESSON OBJECTIVES After completing this lesson, you should be able to: 5-1. Select the statement that best describes

More information

A MICROLITER OXYGEN ELECTRODE SYSTEM FOR SPERM SUSPENSIONS*

A MICROLITER OXYGEN ELECTRODE SYSTEM FOR SPERM SUSPENSIONS* FERTILITY AND STERILITY Copyright 1978 The American Fertility Society Vol. 30, No. 6, December 1978 Printed in U.S.A. A MICROLITER OXYGEN ELECTRODE SYSTEM FOR SPERM SUSPENSIONS* DAVID F. KATZ, PH.D.t EDWARD

More information

Determination of the ultrastructural pathology of human sperm by atomic force microscopy

Determination of the ultrastructural pathology of human sperm by atomic force microscopy FERTILITY AND STERILITY VOL. 75, NO. 5, MAY 2001 Copyright 2001 American Society for Reproductive Medicine Published by Elsevier Science Inc. Printed on acid-free paper in U.S.A. Determination of the ultrastructural

More information

Procine sphingomyelin ELISA Kit

Procine sphingomyelin ELISA Kit Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

Protn-Latex. For Determination of Protein in Latex. (Cat. # Latex, Latex) think proteins! think G-Biosciences

Protn-Latex. For Determination of Protein in Latex. (Cat. # Latex, Latex) think proteins! think G-Biosciences 415PR-02 G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name Protn-Latex For Determination of Protein in Latex (Cat. # 786-20Latex, 786-21Latex)

More information

Research Method Validation

Research Method Validation Evaluation of the FECPAK Sheep Method (Moredun Institute, Scotland) General Observations The microscope included in the kit had a "floating stage"; samples would rapidly fade out of focus making counting

More information

THE INDICATIONS FOR, advantages and disadvantages of insemination have

THE INDICATIONS FOR, advantages and disadvantages of insemination have Fertility as Evaluated by Artificial Insemination Sheldon Payne, M.D., and Robert F. Skeels, M.D. THE INDICATIONS FOR, advantages and disadvantages of insemination have been reviewed and presented before

More information

potassium is included in the dialysis fluid. Unless the phosphate buffer

potassium is included in the dialysis fluid. Unless the phosphate buffer EFFECT OF POTASSIUM ON RAM SPERMATOZOA STUDIED BY A FLOW DIALYSIS TECHNIQUE H. M. DOTT and I. G. WHITE A.R.C. Unit of Reproductive Physiology and Biochemistry, University of Cambridge {Received 23rd July

More information

2017 SPERM QUALITY ANALYSIS MES CATALOGUE

2017 SPERM QUALITY ANALYSIS MES CATALOGUE 2017 SPERM QUALITY ANALYSIS MES CATALOGUE [ 2 ] CONTENTS Page Section 4 7 Sperm quality analyzers 8 9 Digital systems for sperm analysis 15 19 Additional information [ 3 ] SPERM ANALYSIS SQA-Vision Sperm

More information

III Detection methods for animal derived proteins

III Detection methods for animal derived proteins III Detection methods for animal derived proteins [Methods listed in the Feed Analysis Standards] 1. Sampling, storage and preparation methods for analysis samples 1 Sampling method for analysis samples

More information

Department of Prosthodontics, Faculty of Dentistry, Mahidol University, Bangkok, Thailand

Department of Prosthodontics, Faculty of Dentistry, Mahidol University, Bangkok, Thailand Dental Journal Mahidol Dental Journal Original Article Dimensional stability of compression and injection molding denture bases in long and short curing procedures Noppavan Nagaviroj, Krantarat Wanitchanont,

More information

FERTILIZATION IN THE SEA-URCHIN AS A FUNCTION OF SPERM-TO-EGG RATIO

FERTILIZATION IN THE SEA-URCHIN AS A FUNCTION OF SPERM-TO-EGG RATIO FERTILIZATION IN THE SEA-URCHIN AS A FUNCTION OF SPERM-TO-EGG RATIO H. TIMOURIAN, C. E. HUBERT and R. N. STUART Bio-Medical Division, Lawrence Livermore Laboratory, University of California, Livermore,

More information

CHAPTER MATERIAL AND METHODS

CHAPTER MATERIAL AND METHODS CHAPTER 3 3.1 MATERIAL AND METHODS The experimental study was conducted at the Agricultural Research Council in Pretoria (25 55' S; 28 12' E), South Africa. The area is situated in a typical highveld climate

More information

The following protocol describes the isolation of nuclei from tissue. Item. Catalog No Manufacturer

The following protocol describes the isolation of nuclei from tissue. Item. Catalog No Manufacturer SOP: Nuclei isolation from tissue and DNaseI treatment Date modified: 090923 Modified by: P. Sabo. (UW) The following protocol describes the isolation of nuclei from tissue. Ordering Information Item.

More information

6.02 Uniformity of Dosage Units

6.02 Uniformity of Dosage Units 6.02 Uniformity of Dosage Units Change 1. Content Uniformity, 3. Criteria and Table 6.02-2 as follows: 1. Content Uniformity Select not less than 30 units, and proceed as follows for the dosage form designated.

More information

9700 BIOLOGY. Mark schemes should be read in conjunction with the question paper and the Principal Examiner Report for Teachers.

9700 BIOLOGY. Mark schemes should be read in conjunction with the question paper and the Principal Examiner Report for Teachers. CAMBRIDGE INTERNATIONAL EXAMINATIONS GCE Advanced Subsidiary Level and GCE Advanced Level MARK SCHEME for the May/June 2013 series 9700 BIOLOGY 9700/53 Paper 5 (Planning, Analysis and Evaluation), maximum

More information

epithelial cell excretion are expressed always as twelve hour rates." Withholding fluid during the afternoon and night except for 200 cc.

epithelial cell excretion are expressed always as twelve hour rates. Withholding fluid during the afternoon and night except for 200 cc. THE ADDIS SEDIMENT COUNT IN By JOHN D. LYTTLE NORMAL CHILDREN (From the Babies Hospit and the Department of Pediatrics, Colege of Physicians and Surgeons, Columbia University, New York City) (Received

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 37-013 GARD Users Manual March 2005 Manual No. 37-013-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

Catalase Lab - A Bio ENZYME ACTIVITY Investigation Created by Gen Nelson, modified by Dr G

Catalase Lab - A Bio ENZYME ACTIVITY Investigation Created by Gen Nelson, modified by Dr G Catalase Lab - A Bio ENZYME ACTIVITY Investigation Created by Gen Nelson, modified by Dr G INTRODUCTION Hydrogen peroxide (H 2O 2) is a poisonous byproduct of metabolism that can damage cells if it is

More information

Cambridge International Examinations Cambridge International General Certificate of Secondary Education

Cambridge International Examinations Cambridge International General Certificate of Secondary Education *1669880040* Cambridge International Examinations Cambridge International General Certificate of Secondary Education CO-ORDINATED SCIENCES 0654/61 Paper 6 Alternative to Practical May/June 2014 1 hour

More information

A. Macroscopic and Physical tests:- 1. Volume 2. Colour 3. Consistency and cloudiness 4. Osmotic pressure 5. Specific Gravity 6. Electro Conductivity

A. Macroscopic and Physical tests:- 1. Volume 2. Colour 3. Consistency and cloudiness 4. Osmotic pressure 5. Specific Gravity 6. Electro Conductivity EXAMINATION AND EVALUATION OF SEMEN. Evaluation of Semen. A. Macroscopic and Physical tests:- 1. Volume 2. Colour 3. Consistency and cloudiness 4. Osmotic pressure 5. Specific Gravity 6. Electro Conductivity

More information

Semen evaluation in domestic animals I

Semen evaluation in domestic animals I Reproductive Biotechnologies Andrology I I Semen evaluation in domestic animals I Prof. Alberto Contri Different aims 1. Diagnosis 2. Handling and preservation Related to the male - Breeding soundness

More information

Microscopic evaluation. Dr Hany Lotfi Faculty of Veterinary Medicine Zagazig University

Microscopic evaluation. Dr Hany Lotfi Faculty of Veterinary Medicine Zagazig University Microscopic evaluation Faculty of Veterinary Medicine Zagazig University 1- Examination of non stained film 1- Evaluation of motility 2- Examination of stained film 1- Evaluation of live/dead percent 2-

More information

How do you use the AiM Vascular Mapping Kit?

How do you use the AiM Vascular Mapping Kit? How do you use the AiM Vascular Mapping Kit? Protocol for Successful Skin Mapping Evaluate vessel anatomy, perform measurements, etc. according to your laboratory s protocol prior to applying the AiM device

More information

Rodriguez The Semen Analysis

Rodriguez The Semen Analysis RODRIGUEZ 2016 Rodriguez 2016 The Semen Analysis DISCLOSURE "Yo divulgo que no tengo relación financiera con intereses comerciales". RODRIGUEZ 2016 Question1 Fill the Blanks with the Correct Statement

More information

MODERN TRENDS. Associate Editor. Edward Wallach, M.D. ARTIFICIAL INSEMINATION WITH FROZEN SPERMATOZOA*

MODERN TRENDS. Associate Editor. Edward Wallach, M.D. ARTIFICIAL INSEMINATION WITH FROZEN SPERMATOZOA* MODERN TRENDS Edward Wallach, M.D. Associate Editor FERTILITY AND STERILITY Copyright 1978 The American Fertility Society VoL 29, No.4, April 1978 Printed in U.S.A. ARTIFICIAL INSEMINATION WITH FROZEN

More information

3 To gain experience monitoring a titration with a ph electrode and determining the equivalence point.

3 To gain experience monitoring a titration with a ph electrode and determining the equivalence point. Titrations PURPOSE To determine the concentration of acetic acid in vinegar. GOALS 1 To perform an acid-base titration. 2 To gain experience titrating carefully to a visible endpoint. 3 To gain experience

More information

INSTRUCTIONS Pierce Primary Cardiomyocyte Isolation Kit

INSTRUCTIONS Pierce Primary Cardiomyocyte Isolation Kit INSTRUCTIONS Pierce Primary Cardiomyocyte Isolation Kit 88281 Number Description 88281 Pierce Primary Cardiomyocyte Isolation Kit, contains sufficient reagents to isolate cardiomyocytes from 50 neonatal

More information

ANDROVISION - MORE THAN CASA

ANDROVISION - MORE THAN CASA ANDROVISION - MORE THAN CASA AndroVision CASA system with Zeiss AxioScope optics and automated ScanStage Computerized semen analysis AndroVision is a highly precise CASA* system for standardised, interactive

More information

A closer look at BLOOD Lab

A closer look at BLOOD Lab A closer look at BLOOD Lab Name Synthetic by Part 1 a drop of blood White Cell Plasma Platelet White Cell Background: All blood will contain Red Cells (RBC), White Cells (WBC), and Platelets. The number

More information

Characteristics of donor semen and cervical mucus at the time of conception

Characteristics of donor semen and cervical mucus at the time of conception FERTLTY AND STERLTY Copyright 1983 The American Fertility Society Printed in U.8A. Characteristics of donor semen and cervical mucus at the time of conception Ar Edvinsson, M.D. * Per Bergman, M.D.* Yvonne

More information

Influence of Oxidative Stress on Functional Integrity of Human Spermatozoal Membrane.

Influence of Oxidative Stress on Functional Integrity of Human Spermatozoal Membrane. The Journal of Obstetrics and Gynecology of India, March / April 2011 pg 195-199 Original Article Influence of Oxidative Stress on Functional Integrity of Human Spermatozoal Membrane. Chaudhari Ajay 1,

More information

Lab 2: Investigating Variation Across Spatial Scales

Lab 2: Investigating Variation Across Spatial Scales Lab 2: Investigating Variation Across Spatial Scales What are scales and variation in a biological context? The world around us displays incredible diversity across many scales. Today s lab investigates

More information

Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL

Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL Table of Contents Bryan-DUmon Series II rigid bronchoscope 1. 2. 3. 4. 5. Diagram and Overview Universal Barrel Bronchial and

More information

Simplifying Microfluidic Separation Devices towards Field-Detection of Blood Parasites

Simplifying Microfluidic Separation Devices towards Field-Detection of Blood Parasites Electronic Supplementary Material (ESI) for Analytical Methods. Supplementary Material (ESI) for Analytical Methods Simplifying Microfluidic Separation Devices towards Field-Detection of Blood Parasites

More information

Use Tomorrow s Technology Today

Use Tomorrow s Technology Today Use Tomorrow s Technology Today Company Profile (MES) is a leading edge company specializing in rapid, automated semen analysis for both the human and the veterinary markets. The Company s comprehensive

More information

2017 SPERM QUALITY ANALYZERS FOR VETERINARY / BREEDING MES CATALOGUE

2017 SPERM QUALITY ANALYZERS FOR VETERINARY / BREEDING MES CATALOGUE 2017 SPERM QUALITY ANALYZERS FOR VETERINARY / BREEDING MES CATALOGUE COMPANY PROFILE Dear Partners! West Medica company specializes in manufacturing and distribution of veterinary equipment. The company

More information