Use and accuracy of instruments to estimate sperm concentration: Pros, Cons & Economics
|
|
- Willa Melton
- 5 years ago
- Views:
Transcription
1 Use and accuracy of instruments to estimate sperm concentration: Pros, Cons & Economics R. V. Knox, S. L. Rodriguez-Zas, S. Roth, and Kelly Ruggiero Department of Animal Sciences, University of Illinois, Urbana Importance of Sperm Concentration for AI AI success can depend upon sperm quality at collection, sperm age when inseminated, insemination volume, and AI times relative to estrus. However, in addition to these factors, another essential component of AI effectiveness is the number of sperm inseminated. In fact, concentration of sperm inseminated, may actually compensate for many deficiencies involving the technology when using AI, when conditions are less than optimal. Therefore, properly estimating the concentration of the boar ejaculate is critical for efficient boar use and for influencing the effectiveness of the AI technology to optimize farrowing rate (FR) and litter size (LS). Two billion sperm appears to optimize conventional AI for LS and FR when semen that is less than 36 h old, is administered within 28 h prior to ovulation 1. To further illustrate this, Watson and Behan 2 reported that when using conventional AI, 1 billion sperm resulted in lower farrowing rates and smaller litter sizes when compared to 2 or 3 billion sperm, (AI at 0 and 24 h, and semen used within 48 h of collection). However, when bypassing the cervix and much of the uterus with surgical insemination, 1 versus 3 billion had no effect when insemination was performed at < 10 h before ovulation 3. Apparently, the importance of sperm concentration used in an AI dose becomes clear when it was reported that the presence and numbers of sperm found in the oviduct following AI was only observed when > million sperm /cc (1-10 billion/dose) were used 4. However, Steverink et al. 5 indicated that when semen doses contained 1-6 x 10 9 sperm/dose, and were used within 36 h, the effect on fertility and accessory sperm counts were not apparent. Yet, as dose of sperm increased, and as the insemination to ovulation interval decreased, more accessory sperm were found bound to the eggs. This, and other data, suggests that in conditions of limited fertility, reproductive efficiency may depend on sperm concentration. From these studies it is clear that the sperm used in conventional AI must supply enough sperm to allow establishment of a viable sperm reservoir. The high numbers of sperm in a dose may help to keep a reservoir functional and compensate for increased intervals from insemination to ovulation. In a critical test of the effect of low sperm numbers (or viable sperm), the low dose technique using surgical insemination into the tip of the uterine horn has shown that 1 million to 2 billion sperm inseminated in 0.5 ml 3 results in no differences in FR or LS were observed when 10 million or more sperm were used for FR and LS. However, a 50% reduction was observed when 2 million or less was used. Another test of the effects of limited viable sperm occurs following freezing and thawing. In fact, more sperm must be used (5 billion) to achieve the same reproductive rates when compared to non-frozen-thawed sperm. Additionally, insemination must occur even closer to time of ovulation (within 4 hours of ovulation 6 ). In most cases, when using frozen thawed sperm, the 5 billion sperm used for AI, are intended to compensate for loss of fertile sperm, and provides ~1.5 billion motile sperm 7. Collectively, these studies indicate that when using conventional AI, 2 billion fertile sperm 20
2 can reliably result in good fertility. However, when using compromised sperm numbers, fertility is at risk. It is clear that a quick and reliable method to evaluate sperm concentration in an ejaculate for use in AI is required. The two methods that are commonly used for evaluation of sperm concentration include cell-counting chambers used with a light microscope, and analysis by photometry. Gross Evaluation of Sperm Concentration Sperm concentration greater than 200 million cells per/ml is the minimal concentration recommended for extension. Ejaculates should initially be evaluated based on weight (volume), opacity (clarity), color, odor, and turbidity (swirling) before further processing. A visual assessment the opacity or the amount of light that passes through an ejaculate is an indication of the number of sperm cells. Less transparent samples therefore contain more sperm cells. The sperm rich fraction is creamy in appearance and allows little light pass though. The sperm poor fractions are more watery in appearance, are more transparent, and have fewer cells. Ejaculate turbidity is another indicator for sperm cell numbers. It appears when the sample is gently rotated and a swirling motion can be seen in the liquid. For an example of this, let the sample set long enough for sperm cells to settle to the bottom of the container and then gently rotate the sample. Semen Sub-Sampling for Evaluation Before any further technical evaluation is performed, sub-sampling of the ejaculate should be made with a well-mixed sample. Gentle swirling is required since sperm cells settle to the bottom with time. It should be mentioned that shaking or swirling too vigorously could cause damage to the sensitive attachment region for the sperm head and tail. Estimating Sperm Concentration Using the Hemacytometer Evaluation of a semen sample with the hemacytometer (blood cell counting chamber) provides a method for accurate determination of sperm cell concentration. An accurate estimation of concentration will allow extension rates that optimize boar utilization and fertility of extended semen. The materials needed for this routine procedure should include a good quality microscope (~$1500), and a hemacytometer (~$100/ea). The sample dilution system can include a red or white blood cell dilution system (Unopette, ~1.00/ea) or a handpipette system. The Unopette white blood cell dilution system is desirable since no additional equipment is needed, sperm cells are killed, and pipetting and dilution errors can be minimized. In this system, a sub-sample from the ejaculate is taken by mixing the ejaculate well and placing 2-3 drops on a microscope slide. The Unopette system allows a µl sample to be obtained with a Unopette capillary pipette. The semen sample in the capillary tube is then added to the Unopette dilution chamber and mixed. The diluted semen sample is then applied to both chambers of hemacytometer. After ~3-5 minutes, the sperm cells settle onto the glass and counting of sperm at magnification can begin. Since sperm are almost transparent under light microscopy, a phase-contrast scope can facilitate sperm visualization. When using a standard light microscope, the iris diaphragm should be closed to improve sperm visualization. Sperm are counted in 5 diagonal squares. Sperm heads are not 21
3 counted if they touch the right or bottom triple lines of the squares (Figure 1). These sperm are not included since they contribute to over-estimation of the concentration when multiplication is performed. Both chambers of the hemacytometer are counted. Sperm numbers from each side should be within 10% of each other. If not, clean the hemacytometer and repeat sample addition (it is not necessary to repeat the dilution step). When sperm counts are within 10%, average the two numbers. If a 1:100, or a 1:200 Unopette dilution was used, simply add seven zeros to get sperm concentration/ml. This is a simplified way to determine concentration (for calculation example, see Appendix). Accuracy of the Hemacytometer The accuracy of the hemacytometer has a built in measurement system, with dual counting chambers. The objective is to arrive at count that is within 10% from one side to the other. However, this only estimates the intra-sample error from counting. To evaluate error from pipetting and counting, evaluate the count from two independent (inter-assay) counting events of the same sample. An evaluation of a raw semen sample diluted 1:100 with standard procedures compared to a count of semen from the spectrophotometer dilution range (1/5 to 1/50) can be used for providing counts in the optimal ranges. In our evaluation of the hemacytometer system (n= 60), intra-assay errors were less than 10% and inter-assay error estimation of sperm concentrations indicated that values were not statistically different and were within 5% of each other (188 versus 196 ± 12 million sperm/ml). Estimating Sperm Concentration Using the Photometer or Spectrophotometer Spectrophotometry is a measure of transmitted light as a function of wavelength through a reference liquid or blank. When a beam of light (wavelength), is focused on a semen sample, light is absorbed, scattered, or transmitted, depending upon the number, size, shape and opacity of the particles. The wavelength chosen is not as critical with white suspensions, but wavelengths for sperm appear most sensitive in the range of 550 to 576 nm (Foote). The equipment measures the relative amount of light transmitted, which can be displayed (% Transmittance, T) or converted into absorbance (-log T, A), which is directly proportional to concentration (Absorbance = absorptivity of sample x wavelength x concentration) 8. For semen evaluation, spectrophotometers (variable wavelengths) or photometers (single wavelengths) can be used. These single beam light sources can read in transmittance (0-125%), absorbance ( ), or concentration (0 999) units. The equipment is designed to minimize stray light and maintain narrow bandwidths. A sample of diluted semen (range: 1:2 to 1:200) is placed in a sample container inside the machine. A tungsten lamp (which has the features of high energy and longevity) generates light in the visible light range ( nm). The radiant energy of the lamp passes through light filters and through a lens, which concentrates the light passing through the sample cuvette or tube (a high quality plastic or glass sample holder, which typically transmits >75% of the light through a blank sample). On the opposite side of the sample, a detector (photodiode) produces an electrical current proportional to the amount of light transmitted. This signal is then amplified and converted for display. 22
4 Photometers and spectrophotometers are priced between $1,500 to $6,000. The equipment may read in light absorbance units or % light transmittance. Many of the commercially available photometers have predetermined curves that calculate the concentration of sperm/cc while others provide a reading that is used to convert the reading to sperm/cc. The accuracy for each machine is based on a standard curve. Taking multiple semen samples and diluting them (1:10, 1:25, 1:50, 1:100, and 1:200) generates the standard curve. The transmittance or absorbance is then measured for each sample. A sperm cell count can then be performed on the original semen sample to determine the actual number of sperm cells/cc. A linear curve can be generated by dividing the actual known cell concentration by each of the dilutions (i.e. 5.0 x 10 8 /cc divided by 200 (e.g. 1:200 dilution) = 2.5 x 10 6 /cc. This is repeated for each dilution of each ejaculate sample. Each sample dilution is associated with an absorbance reading and establishes a relationship with sperm cell numbers (linear regression). A regression equation is then used to predict sperm cell numbers based on the absorbance or transmittance reading. There is inherent variation around any reading and only repeated sampling of different ejaculates around a photometer reading will improve confidence in the estimate of the sperm cell numbers. Unfortunately, light scattering is affected by seminal plasma within and between boar ejaculates, which may account for up to -19% to +30% errors in photometer estimations of actual sperm concentrations. Some units require dilution prior to reading while others do not. Readings near the upper and lower limits of detection (<10% and >90% for transmittance, or <0.2 and >1.8 absorbance) have high degrees of reading variation (low accuracy). This is because of the combination of too few or too many cells, and the detection of other factors that can alter the passage of light through a semen sample. This may actually become exaggerated at high semen dilution rates. Manufacturer instructions should be followed and calibration and accuracy of the equipment should be performed as recommended. Accuracy in Estimating Sperm Concentration Most of the equipment can resolve to.001 absorbance units, and have a precision of.005 absorbance units. The accuracy for spectrophotometers is reported to approximately ±.003 absorbance units and typically for photometers to.01 absorbance units (± 2% of the true value). Early research evaluating boar semen at concentrations of 54 to 287 x 10 6 sperm/ml using a photometer, indicated that the average error was ± 16.2%. Comparison of the accuracy of the spectrophotometer to the hemacytometer has been performed in many species including honeybees, buffalo, fish, turkeys, cattle, horses, sheep and swine. Generally measurements have all been performed in the nm wavelength range and units measured were transmittance or absorbance. Paulenz et al. 9 compared the spectrophotometer (1:2 dilution; SpermaCue, HemoCue AB, Angelholm, Sweden), hemacytometer (1:200 diltuion), and Coulter counter (Coulter Electronics, Luton, England). There findings indicated that the hemacytometer was the most variable (12% CV, compared to the spectrophotometer (2.9%), and the Coulter counter (2.3%). Yet all three appeared to be highly related, and in an evaluation of the accuracy of the three methods, it was clear that both the spectrophotometer and the Coulter counter over- and under estimated the values at high and low concentrations. Correlation between concentrations and transmittance in almost all cases was high and greater than In our study involving 29 boar ejaculates, we also observed a high correlation of concentration with transmittance (-0.98). However, the 23
5 primary factors limiting this relationship are the precision of pipetting, the reliability of the reagents, and the optimal transmittance/absorbance ranges for the equipment. The Twyman- Lothian curve (Spectronic 401) indicates the percent error in concentration as a function in percent transmittance (Figure 2). The curve indicates that when reading outside of the optimal rage (20 to 50 % T or A, the estimates for concentration can be less precise 10. Other sources for variation can include particulate matter in the suspension, scratched containers, precipitation of sperm cells or other dissolved substances. Generating a Standard Curve. A useful way for quickly predicting the sperm concentration based on transmittance or absorbance is through construction of a standard curve, which plots the absorbance (or transmittance) of samples on the y-axis with the known concentration on the x-axis, and determining the useful portion with the best fit. Validation for the useful range for the spectrophotometer can be determined by taking a sample of known concentration and diluting it from 1:25 to 1:250. The results of the known concentration divided by the dilution rate can be plotted against the transmittance. This figure can be used to evaluate the predictive range for the machine (Figure 3). As can be seen, the data is similar to that for the Twyman-Lothian curve and arrows indicate the optimal accuracy ranges based on the previous reports. However, if plotting the straight line regression, it appears that the best fit for the dat actually lies between 35 to 80% transmittance, suggesting that the Twyman Lothian findings for some substances, may not apply to semen. Comparison of Equipment Evaluation of different photometer equipment has been performed and for two instruments using the same bandwidth, the calibration curves were not different. In other studies, triplicate assessment of sample counting errors was estimated at 12.3% (CV) for the hemacytometer and 2.9% for the spectrophotometer. In our lab, we evaluated a spectrophotometer (Spectronic 401, Spectronic Instruments, Inc. Rochester, NY) versus a photometer (Hyperion Micro-Reader I, Hyperion, Inc., Miami, FL) by dilution levels from 1:5 to 1:500 using 29 boar ejaculates. The data suggested that for each machine, the dilution range could influence the optimal reading range and for the Micro-reader I, the optimal range based on previously published reports, was 1:20 (0.3 to 0.7 A0), and for the Spectronic 401 was 1:15. At the same wavelength of 576 nm, the patterns were very similar and gave similar predictive values (Figure 4). The absorbance values resulting from the actual hemacytometer counts of the semen samples indicated that the values were influenced by boar, but not by equipment. For all samples, the readings, although nearly identical in pattern, produced different absolute levels. In fact the Spectronic 401 always produced higher transmittance readings. In our study, the average concentration evaluated was x 10 6 sperm/ml (± 83.0, SD), and all samples ranged from 47.5 to x 10 6 sperm/ml In this study, the correlation between the transmittance and absorbance was 0.98 and transmittance was defined as = e (-abs). It is clear then, that regardless of the equipment used, the same degree of reliability could be expected between machines. However, absolute values are expected to be different. 24
6 Optimal Dilution Range The ability to perform a single dilution that would provide optimal readings for any given ejaculate within the limits of the spectrophotometer would be ideal. The dilution range from sperm concentrations that produced the steepest slope, with the least variation, and accounts for most of the variation in absorbance (or transmittance) would be the best for predicting the concentration from the absorbance reading. For evaluation of the repeatability of certain dilution ranges, an evaluation of 8 separate boar ejaculates at dilutions of 1/50, 1/25 and 1/10 was performed in triplicate. The average readings for the Micro-reader was 0.17, 0.34, and 0.76 O.D. units, respectively. The standard errors as a percentage of the mean for each dilution were similar, and for the 1:10 was 5.1%, for the 1/25 was 6.6%, and for the 1/50 was 6.8%. This indicates, that the lower dilution rates may be expected to be less variable. Our evaluation of different dilution ranges using a spectrophotometer (Figure 5) and a photometer (Figure 6) and how they relate to % transmittance is shown. Both pieces of equipment produced nearly identical results, and the dilution ranges at 1:50 or greater (150 to 1:500) produced slopes that did not significantly differ from 0 (P >.05), and therefore would make poor estimates for concentration. The slopes were consistently different from 0 when semen was diluted less than 1:50 (P<.05). The dilution range that explained the greatest amount of variation in transmittance by concentration was the 1:5 dilution for both the Micro-reader I and the Spectronic 401 (P<.0001) compared to all other dilutions. This dilution had the steepest slope, the smallest standard error of the slope and was most significant. In fact a, 1 million change in sperm cell concentration would alter the transmittance.0006%. In the range of concentrations tested, 50 to 360 million sperm/cc, the transmittance (45% to 20%) or absorbance, was a good predictor for sperm concentration. In an assessment of the accuracy of certain dilutions over the others, at each dilution we compared the proportion of the rankings that were similar between the photometer and the spectrophotometer (samples arbitrarily ranked high, medium, or low based on transmittance) for all boars. The percent correct rankings were 1:5, 86%, 1:10 (80%), 1:15 (80%), 1/20 (90%), and 1/25 (66%). Although not statistically tested, this does indicate high repeatability and confidence at the lower dilution rates, and matches the high predicting equation accuracy at the 1:5 dilution. It should again be mentioned, that Foote 8 and the Twyman-Lothian curve (Spectronic 401) 10 have indicated that the optimal range for transmittance and absorbance are narrow and accurate only in certain ranges (30-50% T, and 0.3 to 0.7 A). Since from Figure 5 and 6, values out of the optimal range provide the best estimators for concentration, and values lying within the suggested optimal range are poor estimators, an overall conclusion that should be drawn is that each operation, machine, and dilution series should be evaluated by a standard curve which was generated within lab, since not all conditions can be expected to be optimal from one lab to the next. Support for this comes from the fact that the Spectronic 401 consistently reads at a lower absorbance than the Micro-reader. Perhaps this is not surprising, since the Spectronic would be expected to reduce the level of outside light interference. However, perhaps another concern should address the expected or actual levels of boar sperm concentration observed in the filed today compared to past decades. It is well noted that semen samples evaluated over the last four decades have ejaculates in the range of
7 billion/ejaculate and a concentration of million/ml. However, newer data 11 indicates an average ejaculate output of 82 billion when boars are collected about once per week. Based on estimates then, the expected volume might be ~220 ml and average sperm concentration of ~370 million sperm/cc. It is possible, that lower collection frequencies, selection for increased testes size, greater maturity, better health and nutrition, have all lead to a dramatic increase in average sperm output from boars in stud. If this is the case, perhaps new standard curves (and higher dilution rates) will be needed to evaluate the increased sperm concentration. One example of this comes from Paulenz et al. 9 who observed that when using the SpermaCue at 1:2, the optimal reading range was x 10 6 sperm/ml However, in their study, more than 83% of all boar ejaculate samples were of greater concentration than the maximum recommended reading value of 259 x 10 6 sperm/ml In fact, the average concentration for the 1-3 year old boars was ~ 410 x 10 6 sperm/ml This lends further support to the notion that conditions for sperm production have been greatly improved over the last decades. Semen Extension In gross semen extension, criteria such as volume, opacity, turbidity, and color are used to evaluate whether to extend the sample. In this system, no microscope or photometer is used to determine sperm concentration. Extension rates instead, are based on previous knowledge of the normal ranges for total sperm output in an ejaculate. Since most boars that are healthy and have not been overused, ejaculate in the range of billion sperm, dilution rates between 1:4 and 1:10 can be used. For example, a 1:4 dilution rate, would in theory, provide semen doses with 5-15 billion cells per dose, while a 1:10 dilution would produce doses containing 2-6 billion sperm. Obviously, there are inherent flaws in this system, but "blind" dilutions within this range can generally produce acceptable fertility results. However Rozeboom et al. 12 reported that seminal plasma should be at least 10% of the extended semen to take advantage of its multiple effects. This typically limits the extension range based on the original collection volume ( ml) and the total motile sperm in the ejaculate ( x 10 9 sperm) and the need to keep seminal plasma between 10-20%. REFERENCES Nissen, A. K., N. M. Soede, P. Hyttel, M. Schmidt, and L. D Hoore The influence of time of insemination relative to time of ovulation on farrowing frequency and litter size in sows, as investigated by ultrasonography. Theriogenology 47: Watson, P., and J. Behan Deep insemination of sows with reduced sperm numbers does not compromise fertility: A commercially based field trial. IMV International Swine Reproduction Seminar, Minneapolis, MN. Rath, D., C. Krueger, and L.A. Johnson Low dose insemination in the pig.. In: Boar Semen Preservation IV. L.A. Johnson and H.D. Guthrie (eds.). Allen Press, Inc., Lawrence, KS. 26
8 Baker, R.D., P.J. Dziuk, and H. W. Norton Effect of volume of semen, number of sperm and drugs on transport of sperm in artificially inseminated gilts. J. Anim. Sci. 27: Steverink, D. W., N. M. Soede, E. G. Bouwman, and B. Kemp Influence of insemination-ovulation interval and sperm cell dose on fertilization in sows. J. Reprod. Fertil. 111: Waberski D., K. F. Weitze, T. Gleumes, M. Schwartz, T. Willmen and R. Petzoldt Effect of time of insemination relative to ovulation on fertility with liquid and frozen boar semen. Theriogenology 42: Hofmo, P.O., and I. S. Grevle Development and commercial use of frozen boar semen in Norway. In: Boar Semen Preservation IV. L.A. Johnson and H.D. Guthrie (eds.). Allen Press, Inc., Lawrence, KS. Foote, R.H How to measure sperm cell concentration by turbidity (optical density). Proceedings of the Fourth Technical Conference on Artificial Insemination and Reproduction. N.A.A.B. Paulenz, H., I. S. Grevle, A. Tverdal, P. O. Hofmo, and K. Andersen Berg Precision of the Coulter counter for routine assessment of boar-sperm concentration in comparison with the haemocytometer and spectrophotometer. Reprod. Dom. Anim. 30: Spectronic 401, Spectronic Instruments, Inc. Rochester, NY. Rutten, S. C., R. B. Morrison, and D. Reicks Boar stud production analysis. Swine Health and Prod. 8: Rozeboom K. F., M. H. T. Troedsson, G. C. Shurson, J. D. Hawton, and B. G. Crabo Late estrus or metestrus insemination after estrual inseminations decreases farrowing rate and litter size in swine. J. Anim. Sci. 75:
9 APPENDIX Quality Control Procedures for Evaluation of Semen Concentration Exercise 1 Determine sperm concentration of sample A. The actual formula is ((sperm in 5 squares x 5; (or total sperm in 25 squares)) x dilution rate (100 or 200) x hemacytometer chamber volume (10,000 or 10 4 ) = sperm/cc) ). Example: (25 sperm average (in 5 squares) x 5) x 200 (dilution factor) x 10 4 = 250,000,000 sperm/cc. The total time required for this procedure is about 10 minutes per sample. Errors typically occur with this procedure due to pipetting error and improper chamber filling (under or over-filling). Exercise 2 Calculate repeatability of hemacytometer reading. Intra-dilution (both sides of a single chamber should be within 10%). Inter-dilution error between samples, of sample C at 1:100 (Unopette) versus 1:20 of the diluted sample (typical of a spectrophotometer sample). Exercise 3 Determine spec repeatability (boar A and B) at a single dilution in triplicate Boar A Boar B The intra-assay can be determined. The CV is a measure of variation of the triplicates and should be less than 20%. The CV is determined by dividing the SD by the mean. Exercise 4 A standard curve can be generated within a single day or over a period of time. What is needed is for a variety of samples that differ in concentration, the range covers the typical or expected range for samples that the lab should expect to evaluate over time. The blank should be 3% sodium citrate, but 0.9% sodium chloride can be used. It is possible to find the optimal value for the equipment, boar ejaculate concentration, that different ranges (two) dilution levels may need to be evaluated (for example 1:5 compared to 1:20). Test for dilution range (1:5 and 1:20) Boars A-E A regression line can be generated to predict the concentration based on the values of transmittance or absorbance (or optical density). In an EXCEL spreadsheet, create two columns, the 1 st containing the concentration and the second containing the spectrophotometer values (in this case, transmittance). Select the function f x icon to select the regression function to determine the slope and intercept and use the x and y coordinates to create the estimates. The regression equation can be generated by the formula y = b 1 x + b 0 where b 1 is the slope, b 0 the intercept, and x the spectrophotometer value. One equation is generated for 1:5 (boars A-E) and another at 1:20 (Boars A-E). A correlation coefficient can be generated to determine which regression equation accounts for the most variation ( r = ) in absorbance (or transmittance) using the sperm concentration. 28
10 Column 1 Column 2 b 1 b 0 Y % Transmittance Conc. (x10 6 ) slope Intercept Predicted Value 29
11 Exercise 5 Determine Extension Rate 1. Sperm cells/cc x % motile x % normal x ejaculate volume = total sperm. 2. e.g. 60 billion sperm x 90% motile x 95% normal x 150 cc = 51 billion sperm. 3. To get a desired 3 billion sperm/80cc insemination dose: 340 x 10 6 sperm/cc (determined sperm concentration) x 150 cc (ejaculate volume) = 51 x 10 9 total sperm x 10 9 sperm 3.0 x 10 9 sperm/dose = 17 doses 5. Multiply doses (17) x volume of semen dose (80 cc) = 1360 cc final volume cc cc = 1210 cc 7. Add 1210 cc of extender to 150 cc semen sample. 8. Gently aliquot 80 cc of extended semen into 17 bottles, tubes, or bags for 3.0 x 10 9 sperm/dose. 9. Test with counting under microscope (usually dilution needed at 1:5-1:50) 30
12 Economics of concentration errors In this scenario, a boar is collected and produces 200 ml of semen. The actual sperm concentration is 250 x 10 6 sperm/ml. However, in five different scenarios, errors ± 25-50% are made when determining the sperm concentration. Based on the concentration determination, dilution rates and number of semen doses are finalized. As ca be seen, at error rates within the range estimated, less than optimal dilution rates. It is clear from the literature that AI with 2 to 10 billion sperm can produce optimal fertility results in females when used within 2 days. However, what is not clear is the effect of the different concentrations and interaction with sperm storage, and how this might effect semen fertility and fertility in breeding herds. From the data it is clear that under estimation of sperm numbers leads to excessive sperm concentration and reduced value due to lower numbers of doses obtained. In the case of overestimation of sperm concentration, the optimal sperm numbers are not obtained, and with storage over the next 3-4 days, fertility level in the extended sperm is expected to decline. What is not clear is how the reduced fertility will ultimately impact the breeding herds that utilize multiple doses to compensate for the expected spread in time from insemination to fertilization. Regardless of the original ejaculate concentration (higher or lower than the example of 250 million/ml), the errors in final sperm concentration are similar. Scenario Error Conc.(x 10 6 ) Vol Dil SP(%) sperm.(x 10 9 ) doses vol(f) actual.(x 10 9 ) c-doses Cost/collect A -50% : $41.00 B -25% : , $21.00 C : , D +25% : , $20.50 E +50% : , $42.50 Semen dose valued at $5.00/dose Desired 3 billion sperm /80 cc Error = Percentage error in estimated sperm concentration from actual Vol = volume of ejaculate Dil = final dilution SP(%)= final seminal plasma percentage Doses = number of doses to be produced vol(f) = final extension volume actual.(x 10 9 ) = sperm concentration in doses for final use c-doses = change in doses produced Cost/collect = cost relative to doses above or below optimal 31
Counting Spermatozoa - Part 1 By B. W. Pickett, Ph.D. Professor Emeritus, Colorado State University
Counting Spermatozoa - Part 1 By B. W. Pickett, Ph.D. Professor Emeritus, Colorado State University A hemacytometer, originally designed to count blood cells, has been used for many years to count other
More informationEffect of Frequency of Boar Exposure on Estrus and Ovulation in Weaned Sows as Determined by Real-Time Ultrasound.
Effect of Frequency of Boar Exposure on Estrus and Ovulation in Weaned Sows as Determined by Real-Time Ultrasound. Robert V. Knox, Kilby Willenburg, Gina Miller, Sandra Rodriguez-Zas Department of Animal
More informationTECH EXTENSION. How low can we go? Finding a practical path to produce more pigs from fewer and better boars.
TECH EXTENSION How low can we go? Finding a practical path to produce more pigs from fewer and better boars www.fastgenetics.com Artificial insemination (AI) has become the standard breeding method for
More informationPreservation of Liquid Boar Semen: Effect of Genotype, Boar and Sperm Parameters on Motility and Acrosome Integrity
VETERINARY RESEARCH INTERNATIONAL Journal homepage: www.jakraya.com/journal/vri ORIGINAL ARTICLE Preservation of Liquid Boar Semen: Effect of Genotype, Boar and Sperm Parameters on Motility and Acrosome
More informationEliKine Free Thyroxine (ft4) ELISA Kit
EliKine Free Thyroxine (ft4) ELISA Kit Booklet Item NO. KET0005 Product Name EliKine Free Thyroxine (ft4) ELISA Kit ATTENTION For laboratory research use only. Not for clinical or diagnostic use TABLE
More informationand W. Singleton, Department of Animal Science, Purdue University.
Introduction Determining the initial quality of a boar ejaculate is the first step in semen processing and should ensure that prior to further processing, a high quality artificial insemination dose of
More informationMicroscope Requirements
SEMEN EVALUATION EQUIPMENT Microscope Requirements Good quality lenses Phase-contrast preferred for % progressive motility evaluations Objectives 10X, 20X*, 40X*, 100X, minimum Heated stage preferred *Preferably
More informationMouse C-Peptide ELISA Kit
Mouse C-Peptide ELISA Kit Cat.No: DEIA4507 Lot. No. (See product label) Size 96T Intended Use The Mouse C-Peptide ELISA kit is for the quantitative determination of c-peptide in mouse serum, plasma, and
More informationSmall Ruminant Reproductive Management Workshop
Small Ruminant Reproductive Management Workshop Animal Nutrition and Physiology Center, North Dakota State University Sponsors: American Sheep and Goat Center, North Dakota State University, University
More informationMorinaga Mouse C-peptide ELISA Kit
Morinaga Mouse C-peptide ELISA Kit For the quantitative determination of C-peptide in mouse serum, plasma, and fluid 96wells For Laboratory Use Only, not for use in diagnostic procedure Please read full
More informationREPRODUCTIVE INNOVATIONS: CONTROL OF SOW ESTRUS AND BREEDING
REPRODUCTIVE INNOVATIONS: CONTROL OF SOW ESTRUS AND BREEDING Roy Kirkwood Department of Large Animal Clinical Sciences Michigan State University Lansing, Michigan 48824-1314 E-mail: kirkwood@cvm.msu.edu
More informationRat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit
Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit Catalog No. E0228r (96 tests) Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS!
More informationProthrombin (Human) ELISA Kit
Prothrombin (Human) ELISA Kit Catalog Number KA0496 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay... 3 General
More informationNext Generation AI - New Developments to Maximize Efficiency Christianne E Glossop, Malmesbury, England
Next Generation AI - New Developments to Maximize Efficiency Christianne E Glossop, Malmesbury, England AI has been in use worldwide for almost 50 years. As it reaches maturity, the industry is demanding
More informationEvaluating Boar Semen Quality 1 Kevin J. Rozeboom, Ph.D.
Evaluating Boar Semen Quality 1 Kevin J. Rozeboom, Ph.D. Introduction Determining the initial quality of a boar ejaculate is the first step in semen processing and should ensure that prior to further processing,
More informationLDL (Human) ELISA Kit
LDL (Human) ELISA Kit Cat. No.:DEIA3864 Pkg.Size:96T Intended use This immunoassay kit allows for the specific measurement of human low density lipoprotein, LDL concentrations in cell culture supernates,
More informationXVII Congresso Internazionale SIVE
SOCIETÀ ITALIANA VETERINARI PER EQUINI SOCIETÀ FEDERATA ANMVI XVII Congresso Internazionale SIVE XVII SIVE International Congress Palacongressi d Abruzzo Montesilvano (PE) - ITALY 4-6 Febbraio 2011 February
More informationProcine sphingomyelin ELISA Kit
Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH
More informationHbA1c (Human) ELISA Kit
HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,
More informationEvaluation and Economic Impact of Boar Fertility
Evaluation and Economic Impact of Boar Fertility A. Minton 1, A. Johnson 2, T. Werner 2, E. Triemert 2, N. Holden 2, G. Foxcroft 3, J. Patterson 3 and M. Dyck 3 1 PIC North America100 Bluegrass Commons
More informationCanine Thyroid Stimulating Hormone, TSH ELISA Kit
Canine Thyroid Stimulating Hormone, TSH ELISA Kit Catalog No: E0463c 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
More informationRat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.
Rat Insulin ELISA For the quantitative determination of insulin in rat serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-INSRT-E01, E10 96 wells, 10
More informationAbraxis Progesterone (bovine) ELISA Kit
Abraxis Progesterone (bovine) ELISA Kit Enzyme immunoassay for the quantitative determination of progesterone in bovine milk/serum/plasma samples PN5081M 96 Tests For Research Use Only. Not for use in
More informationHuman Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit
Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Catalog No: E0442h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
More informationBovine Insulin ELISA
Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:
More informationInstructions for Use. IA2 Antibody ELISA. Enzyme Immuno Assay for the Quantitative Determination of Antibodies against IA-2 in Serum.
Instructions for Use IA2 Antibody ELISA Enzyme Immuno Assay for the Quantitative Determination of Antibodies against IA-2 in Serum I V D REF EA114/96 12 x 8 2 8 C DLD Gesellschaft für Diagnostika und medizinische
More informationAnti-TSH-Receptor ELISA
Instructions for Use Anti-TSH-Receptor ELISA Enzyme Immuno Assay for the Quantitative Determination of TSH Receptor Autoantibodies (TRAb) in Serum I V D REF EA101/96 12 x 8 2 8 C DLD Gesellschaft für Diagnostika
More informationRat cholesterol ELISA Kit
Rat cholesterol ELISA Kit Catalog No. CSB-E11706r (96T) This immunoassay kit allows for the in vitro quantitative determination of rat Cholesterol concentrations in serum, plasma and other biological fluids.
More informationPorcine/Canine Insulin ELISA
Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research
More informationRat Hemoglobin A1c (HbA1c) Kit Instructions
V.3 Crystal Chem Rat Hemoglobin A1c (HbA1c) Kit Instructions For the quantitative determination of hemoglobin A1c (HbA1c) in rat whole blood Catalog #80300 96 Assays For research use only. Not for use
More informationAnimal Science 434. Semen Collection. Effect of Age on Sperm Output. Age When Semen Can Be Collected. Text: Ch. 10 and 11. Sexual Behavior (cont.
Animal Science 434 Age When Semen Can Be Collected Sexual Behavior (cont.) B. Applied Reproductive Behavior of the Male: Semen Collection and Processing Text: Ch. 10 and 11 Bull Boar Ram Stallion Dog 12
More informationBiodiversity Study & Biomass Analysis
PR072 G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name Biodiversity Study & Biomass Analysis Teacher s Guidebook (Cat. # BE-403) think proteins!
More informationHuman LDL ELISA Kit. Innovative Research, Inc.
Human LDL ELISA Kit Catalog No: IRKTAH2582 Lot No: SAMPLE INTRODUCTION Human low-density lipoprotein (LDL) transports cholesterol from the liver to tissues where it is incorporated into cell membranes.
More informationTiming of A.I. Swine AI 9/6/12
Breeding Herd Education Series 20 Timely, relevant & convenient learning Thank you for participating in SowBridge 20. To start this presentation, advance one slide by pressing enter or the down or right
More informationInsulin (Porcine/Canine) ELISA
Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01
More informationArtificial Insemination in Swine
Artificial Insemination in Swine Artificial Insemination in Swine Routinely practiced on swine farms More efficient and effective use of boars AI centers distribute semen from genetically valuable boars
More informationCoQ10(Coenzyme Q10) ELISA Kit
CoQ10(Coenzyme Q10) ELISA Kit Catalogue No.: EU0196 Size: 48T/96T Reactivity: Universal Detection Range: 0.781-50ng/ml Sensitivity:
More informationHBeAg and HBeAg Ab ELISA Kit
HBeAg and HBeAg Ab ELISA Kit Catalog Number KA0290 96 assays Version: 17 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3
More informationAnimal Science 434" Semen Collection" Effect of Age on Sperm Output" Age When Semen Can Be Collected" Text: Ch. 10 and 11"
Animal Science 434" Age When Semen Can Be Collected" Lecture 15b: Sexual Behavior (cont.) B. Applied Reproductive Behavior of the Male: Semen Collection and Processing Text: Ch. 10 and 11" "Bull "Boar
More informationMouse Ultrasensitive Insulin ELISA
Mouse Ultrasensitive Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use
More informationMouse C-peptide ELISA
Mouse C-peptide ELISA For the quantitative determination of C-peptide in mouse serum. For Research Use Only. Not for use in Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Calculation
More informationWATER AND SOLUTE MOVEMENT THROUGH RED BLOOD CELLS
WATER AND SOLUTE MOVEMENT THROUGH RED BLOOD CELLS Purpose This exercise is designed to demonstrate the properties of cellular membranes and the movement of water and solutes across them. In this lab, you
More informationCortisol (Sheep) ELISA Kit
Cortisol (Sheep) ELISA Kit Catalog Number KA0919 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the
More informationActivity # 4. Measurement of Enzyme Activity
Activity # 4. Measurement of Enzyme Activity Adapted from Experimental Molecular Biology Lab Manual, Stephen H. Munroe, Marquette University Week 1 Protein Extraction, Concentration Measurement & Instrument
More informationONTARIO PORK RESEARCH PROPOSAL FINAL REPORT
PROPOSAL # 04/16 ONTARIO PORK RESEARCH PROPOSAL FINAL REPORT Project Leader: R.N. Kirkwood/ R.M. Friendship Project Title: Effect of addition of seminal plasma to thawed semen on sow fertility Objectives
More informationRayBio Human Granzyme B ELISA Kit
RayBio Human Granzyme B ELISA Kit Catalog #: ELH-GZMB User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross,
More informationSelect the Sex of Your Next Calf Prior to Mating: Using Sexed Semen 1
AN163 Select the Sex of Your Next Calf Prior to Mating: Using Sexed Semen 1 Gary R. Hansen 2 Introduction Through recent advances in reproductive technology, scientists have the ability to pre-select the
More informationTNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures.
TNF-alpha ELISA For the quantitative determination of TNF-alpha in serum, plasma, buffered solution or cell culture medium. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number:
More informationFish follicle-stimulating hormone, FSH ELISA Kit
Fish follicle-stimulating hormone, FSH ELISA Kit Catalog No: E0830f 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
More informationHuman Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA Kit
Human Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA Kit Catalog No: E0228h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS!
More informationAI and its Influence on Production Efficiency
AI and its Influence on Production Efficiency Christianne E. Glossop - Malmesbury, Wiltshire, England 2007 Introduction AI in pigs is by no means a new technique. Methods for semen collection and processing,
More informationINFLUENCE OF DIFFERENT EXTENDERS, DILUTION RATE AND STORAGE TIME ON BOAR SPERM PROGRESSIVE MOTILITY
INFLUENCE OF DIFFERENT EXTENDERS, DILUTION RATE AND STORAGE TIME ON BOAR SPERM PROGRESSIVE MOTILITY Lipenský J., Lustyková A., Frydrychová S., Rozkot M., Václavková E. Institute of Animal Science, Prague,
More informationACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man Coulter HIV-1 p24 ELISA May 21, 2004
Coulter HIV p24 1. PRINCIPLE The Human Immunodeficiency Virus Type 1 (HIV-1) is recognized as the etiologic agent of acquired immunodeficiency syndrome (AIDS). The virus is transmitted by sexual contact,
More informationSemen evaluation in domestic animals I
Reproductive Biotechnologies Andrology I I Semen evaluation in domestic animals I Prof. Alberto Contri Different aims 1. Diagnosis 2. Handling and preservation Related to the male - Breeding soundness
More informationO. Repeat the measurement in all relevant modes used in your experiments (e.g. settings for orbital averaging).
Before You Begin Read through this entire protocol sheet carefully before you start your experiment and prepare any materials you may need. This year, in order to improve reproducibility, we are requiring
More informationTriiodothyronine (T3) ELISA
For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. Triiodothyronine (T3) ELISA Kit is intended for the detection of total T3 in human serum or plasma. For research
More informationisperm Technical Document
isperm Technical Document Content 1.Operating Procedure........2 Set up isperm Set up isperm App Semen preparation Sampling Using isperm App 2. The principle of isperm......4 Concentration Motility 3.Photometer,
More informationMouse C-peptide ELISA
Mouse C-peptide ELISA For the quantitative determination of C-peptide in mouse serum. For Research Use Only. Not for use in Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Preparation
More informationab Creatinine Assay Kit (Colorimetric)
ab204537 Creatinine Assay Kit (Colorimetric) Instructions for Use For the quantitative determination of Creatinine in urine samples. This product is for research use only and is not intended for diagnostic
More informationRayBio Acid Phosphatase Activity Colorimetric Assay. Kit. User Manual Version 1.0 May 5, RayBiotech, Inc. Kit Protocol. (Cat#: 68CL-AcPh-S500)
RayBio Acid Phosphatase Activity Colorimetric Assay Kit User Manual Version 1.0 May 5, 2014 RayBio Acid Phosphatase Activity Colorimetric Assay (Cat#: 68CL-AcPh-S500) RayBiotech, Inc. We Provide You With
More informationab65344 Uric Acid Assay Kit (Colorimetric/Fluorometric)
ab65344 Uric Acid Assay Kit (Colorimetric/Fluorometric) Instructions for Use For the rapid, sensitive and accurate measurement of uric acid in various samples. This product is for research use only and
More informationHuman TSH ELISA Kit. User Manual
Human TSH ELISA Kit User Manual Catalog number: GTX15585 GeneTex Table of Contents A. Product Description... 2 B. Kit Components... 3 C. Additional Required Materials (not included)... 3 D. Reagent Preparation...
More informationHuman Alpha 1 microglobulin ELISA Kit
Human Alpha 1 microglobulin ELISA Kit Catalogue No.: EH4144 Size: 48T/96T Reactivity: Human Range:0.625-40ng/ml Sensitivity:
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma.
Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationRat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum
Rat C-peptide ELISA For the quantitative determination of C-peptide in rat serum Please read carefully due to Critical Changes, e.g., see Calculation of Results. For Research Use Only. Not For Use In Diagnostic
More informationMouse Leptin ELISA Kit Instructions
V.6/Mar/2008 CRYSTAL CHEM INC. Mouse Leptin ELISA Kit Instructions For the quantitative determination of leptin in mouse serum or plasma and fluid Catalog #: 90030 96 Assays For Research Use Only. Not
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma
Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationHuman HIV (1+2) antigen&antibody ELISA Kit
Human HIV (1+2) antigen&antibody ELISA Kit Catalog Number. CSB-E18042h For the qualitative determination of human HIV (1+2) antibody and P24 antigen concentrations in serum, plasma. This package insert
More informationRayBio Human ENA-78 ELISA Kit
RayBio Human ENA-78 ELISA Kit Catalog #: ELH-ENA78 User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross,
More informationWhat Is the Relationship Between the Amount of Transmitted Light Through a Solution and Its Concentration?
What Is the Relationship Between the Amount of Transmitted Light Through a and Its Concentration? Blue Food Dye Treats Spine Injury in Rats HTTP://WWW.WIRED.COM/WIREDSCIENCE/2009/07/BLUERATS/ BY HADLEY
More information02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2
For detection and measurement of human interleukin 2 Catalog #02006 Catalog #02007 2 Plates 10 Plates Product Description The Human Interleukin 2 (IL-2) ELISA Kit is designed for the quantitative detection
More informationDetectX. Urinary Creatinine Detection Kit. Catalog Number K002-H1. Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine
World s Only One Component Assay Simple & Easy to Use DetectX Urinary Creatinine Detection Kit Catalog Number K002-H1 Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine Please read this insert
More informationBlood Urea Nitrogen Enzymatic Kit Manual Catalog #:
Blood Urea Nitrogen Enzymatic Kit Manual Catalog #: 5602-01 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 3 Required
More informationGalactose and Lactose Assay Kit
Galactose and Lactose Assay Kit Catalog Number KA0842 100 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials
More informationMicroscopic evaluation. Dr Hany Lotfi Faculty of Veterinary Medicine Zagazig University
Microscopic evaluation Faculty of Veterinary Medicine Zagazig University 1- Examination of non stained film 1- Evaluation of motility 2- Examination of stained film 1- Evaluation of live/dead percent 2-
More informationIntroduction. Cell Biology OLM
1 of 21 8/3/2011 1:46 PM Cell Biology OLM Introduction Anthocyanins are natural plant pigments that give various fruits, vegetables and flowers red, blue and purple color. Blueberries, blackberries, raspberries
More informationThyroid Stimulating Hormone (TSH) ELISA Catalog No. GWB , legacy id (96 Tests)
For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. TSH ELISA Kit is intended for the quantitative measurement of TSH in human serum or plasma. For research use only.
More informationDetermining the Concentration of Iron in Vitamin Supplements
Teacher Guide Determining the Concentration of Iron in Vitamin Supplements Background Information Colorimetry is the science of measuring color. Colorimetry is a useful technique for determining the concentration
More informationRat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures.
Rat C-peptide ELISA For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-CPTRT-E01 96 wells Version: May 26,
More informationEXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids
DATA SHEET EXOTESTTM ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids INTRODUCTION Exosomes are small endosome-derived lipid nanoparticles
More informationHuman Creatinine Urinary Detection Kit
Human Creatinine Urinary CATALOG NO: IRAAKT2509 Detection Kit LOT NO: SAMPLE INTENDED USE The Urinary Creatinine kit is designed to quantitatively measure creatinine present in urine samples. BACKGROUND
More informationUltra-sensitive Human TSH ELISA Kit. MyBioSource.com
Ultra-sensitive Human TSH ELISA Kit For the quantitative determination of human thyroid stimulating hormone (TSH) concentrations in serum Catalog Number: MBS590037 96 tests FOR LABORATORY RESEARCH USE
More informationDetermination of the Diffusion Coefficient for Sucrose in Aqueous Solutions
CHEM 332L Physical Chemistry Laboratory II Revision 1.1 Determination of the Diffusion Coefficient for Sucrose in Aqueous Solutions In this laboratory exercise we will measure the diffusion coefficient
More informationGAD 65 Antibody ELISA
Instructions for Use GAD 65 Antibody ELISA Enzyme Immuno Assay for the Quantitative Determination of Antibodies against Glutamic Acid Decarboxylase in Serum I V D REF EA104/96 12 x 8 2 8 C DLD Gesellschaft
More informationFinTest IgG4 Screen 20 ELISA KIT
FinTest IgG4 Screen 20 ELISA KIT Cat. No.:DEIA6196 Pkg.Size:96T Intended use Enzyme immunoassay (microtiter strips) for the detection and the quantitative determination of IgG4 antibodies against 20 Food
More informationUrea Nitrogen (BUN) detection Kit
K-ASSAY KAMIYA BIOMEDICAL COMPANY KAMIYA BIOMEDICAL COMPANY Urea Nitrogen (BUN) detection Kit For the quantitative determination of urea nitrogen in saliva and TCM Cat. No. KT-747 For Research Use Only.
More informationGSI Canine IL-5 ELISA Kit-2 Plates DataSheet
Interleukin5 (IL5) is a secreted glycoprotein that belongs to the α-helical group of cytokines (1, 2, 3). Unlike other family members, it is present as a covalently linked antiparallel dimer (4, 5). IL-5
More informationInfluenza A H1N1 HA ELISA Pair Set
Influenza A H1N1 HA ELISA Pair Set for H1N1 ( A/Puerto Rico/8/1934 ) HA Catalog Number : SEK11684 To achieve the best assay results, this manual must be read carefully before using this product and the
More informationHuman cholesterol, CH ELISA Kit
Human cholesterol, CH ELISA Kit Catalog No: E1701h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE
More informationFor the rapid, sensitive and accurate detection of long-chain Free Fatty Acids in various samples.
ab65341 Free Fatty Acid Quantification Kit Instructions for Use For the rapid, sensitive and accurate detection of long-chain Free Fatty Acids in various samples. This product is for research use only
More informationRat IL-2 ELISA Kit. Instructions for use. For research use only. Fast Track Your Research.
Rat IL-2 ELISA Kit Instructions for use Catalogue numbers: 1x96 tests: 670.100.096 2x96 tests: 670.100.192 For research use only Fast Track Your Research. Table of Contents 1. Intended use 2 2. Introduction
More informationQ: How do I get the protein concentration in mg/ml from the standard curve if the X-axis is in units of µg.
Photometry Frequently Asked Questions Q: How do I get the protein concentration in mg/ml from the standard curve if the X-axis is in units of µg. Protein standard curves are traditionally presented as
More informationUnit B: Anatomy and Physiology of Poultry. Lesson 4: Artificial Poultry Reproduction
Unit B: Anatomy and Physiology of Poultry Lesson 4: Artificial Poultry Reproduction 1 1 Terms Artificial insemination Sexed semen Standing heat 2 2 I. Artificial insemination is the placing of semen in
More informationFOR LABORATORY USE ONLY
YK060 Insulin ELISA FOR LABORATORY USE ONLY Kasumigaseki Place, 3-6-7, Kasumigaseki, Chiyoda-ku Tokyo 100-0013 Japan URL: http://www.sceti.co.jp/export/ e-mail: medical@sceti.co.jp Contents
More informationEffect of straw size and thawing time on quality of cryopreserved buffalo (Bubalus bubalis) semen
Vol. 11, No. 1 49 SHORT COMMUNICATION Effect of straw size and thawing time on quality of cryopreserved buffalo (Bubalus bubalis) semen Muhammad S Ansari, Bushra A. Rakha, Syed M. H. Andrabi, Shamim Akhter
More informationFor the quantitative measurement of Human Prothrombin concentrations in cell culture supernatants, milk, plasma and urine
ab108889 Prothrombin Human ELISA Kit Instructions for Use For the quantitative measurement of Human Prothrombin concentrations in cell culture supernatants, milk, plasma and urine This product is for research
More informationab Prothrombin (Factor II) Human ELISA Kit
ab108889 Prothrombin (Factor II) Human ELISA Kit Instructions for Use For the quantitative measurement of Human Prothrombin (Factor II) in plasma, milk, urine, CSF and cell culture samples. This product
More informationDRG International, Inc., USA Fax: (908)
Please use only the valid version of the package insert provided with the kit. 1 INTENDED USE The human Annexin V ELISA is an enzyme-linked immunosorbent assay for determination of human Annexin V. The
More informationChymotrypsin ELISA Kit
Chymotrypsin ELISA Kit Cat. No.:DEIA10041 Pkg.Size:96T Intended use The Chymotrypsin ELISA Kit is a sandwich Enzyme Immuno Assay intended for the quantitative determination of Chymotrypsin in stool. General
More informationRat Proinsulin ELISA
Rat Proinsulin ELISA For the quantitative determination of proinsulin in rat serum For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-PINRT-E01 Size: 96 wells Version: June
More information