Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Size: px
Start display at page:

Download "Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water"

Transcription

1 Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water Richard F. Jack, 1 Jinshui Che, 2 Lipika Basumallick, 1 and Jeffrey Rohrer 1 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, Shanghai, People s Republic of China

2 Overview Purpose: Hexavalent chromium (CrVI) refers to chemical compounds that contain the element chromium in the +6 oxidation state. All CrVI compounds are strong oxidizing agents, and are considered toxic and potentially carcinogenic. Hence, chromates are regulated in drinking water and the environment in the U.S. and many countries around the world. For example, the state of California established a public health goal (PHG) of 0.2 µg/l (ppb) for CrVI and 2.5 µg/l for total chromium in The PHG is based on an estimated one-in-a-million lifetime cancer risk level. However, U.S. EPA Method 3060A does not allow sufficient sensitivity for routine analysis at the recently proposed California PHG level of 0.02 µg/l. Modifications to this EPA method presented here result in a µg/l limit of quantitation (LOQ), which is more than sufficient for analysis at the proposed California PHG level. Water is not the only exposure source for CrVI, as it is also prevalent in soils and many other matrices. Therefore, as an alternative to the labor-intensive EPA Method 3060A, a faster, easier procedure for extracting CrVI from soil matrices by using a Thermo Scientific Dionex ASE 350 Accelerated Solvent Extractor is shown here. Introduction Improvements described here to both analysis and extraction of CrVI are based on the conditions detailed in EPA Method This includes use of the column in the 2 mm format and a smaller reaction coil to increase method sensitivity. By using Thermo Scientific Dionex IonPac AG7 Guard (2 50 mm) and Dionex IonPac AS7 Analytical (2 250 mm) columns, an eluent of 250 mm ammonium sulfate/100 mm ammonium hydroxide at a flow rate of 0.36 ml/min, a 1000 µl injection volume, and postcolumn reaction with 2 mm diphenylcarbazide/10% methanol/1 N sulfuric acid (using a 125 µl reaction coil) followed by visible absorbance detection at 530 nm, a µg/l, minimum detection limit (MDL) for chromate is achieved. Although the efficacy of an automated method to extract CrVI from soil using accelerated solvent extraction (as an alternative to EPA Method 3060A) has previously been demonstrated, 1 the removal recovery was limited. This was due to the lower ph of the sample extraction buffer than that specified in the EPA method. This lower ph was required in earlier experiments due to the stainless steel (SST) accelerated solvent extraction cells, which are less resistant to high ph buffers. But the Dionium (Zr) cells in the Dionex ASE 350 overcome this limitation. Reported here is the use of the same ph with cells able to withstand higher ph values. Methods Ion Chromatography Thermo Scientific Dionex ICS-2100, ICS-1600, ICS-1100, or ICS-5000 system including: SP Single Pump or DP Dual Pump module DC Detector/Chromatography module Injection loop, 1000 µl Reaction coil, 125 µl Sample syringe, 5 ml ICS Series VWD Variable Wavelength Absorbance Detector Postcolumn Reagent (PCR) Delivery: PC-10 pneumatic delivery or AXP pump if not using the DP AS Autosampler Thermo Scientific Dionex Chromeleon Chromatography Data System (CDS) software Reagents and Standards 2 Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

3 Reagents and Standards Prepare all solutions from analytical reagent-grade chemicals (when commercially available). Note, there is a possibility of the presence of trace levels of chromate in some commercially available chemicals. Deionized (DI) water, 18 MΩ-cm or better Ammonium sulfate (Mallinckrodt General P/N AR 7725) Ammonium hydroxide (Sigma P/N A6899) Sulfuric acid, 95-98% (JT Baker Instra-Analyzed P/N 9673) Methanol, HPLC grade (Fisher Optima P/N A454-4) Potassium dichromate (JT Baker P/N ) Sodium and potassium salts, ACS reagent-grade, for preparing the anion standards 1,5-diphenylcarbazide (JT Baker K620-03) Method Columns: Dionex IonPac AG7 Guard 2 50 mm (PN ), Dionex IonPac AS7 Analytical mm (PN ) Eluent: 250 mm Ammonium sulfate and 100 mm ammonium hydroxide Flow Rate: 0.36 ml/min Inj. Volume: 1000 µl (Full loop) Temperature: 30 C Back Pressure: psi PCR: 2 mm diphenylcarbazide, 10% methanol, 1 N sulfuric acid Flow Rate: 0.12 ml/min Detection: Visible absorbance, 530 nm Noise: 6 8 µau Run Time: 25 min Eluent 250 mm Ammonium sulfate 100 mm Ammonium hydroxide Dissolve 66 g of ammonium sulfate in ~1 L of DI water and add 13 ml of 29% ammonium hydroxide solution. Dilute to 2.0 L with DI water. PCR 2 mm Diphenylcarbazide 10% Methanol 1 N Sulfuric acid Sample Adjustment Buffer 250 mm Ammonium sulfate 1000 mm Ammonium hydroxide Dissolve 3.3 g of ammonium sulfate in ~75 ml of DI water and add 6.5 ml of 29% ammonium hydroxide. Dilute to 100 ml with DI water. Accelerated Solvent Extraction Conditions Solvent: 10 mm NaOH, 4 g/l NaCl (ph > 11.5) Temperature: 100 C Preheat Time: 5 min Static Time: 5 min Cycle Time: 2 Flush Volume: 60% Purge Time: 90 s Cell: 66 ml Zr cell Total Time: Approximately 20 min Sample Pretreatment: Dry soil samples at 50 C and pulverize. Place 10 g of soil sample mixed with quartz sand into a 66 ml Zr extraction cell, and insert into the Dionex ASE 350 system with an extract constant volume of 100 ml. Filter through a 0.45 µm membrane and inject into the Dionex ICS system. Results Thermo Scientific Poster Note PN70094_e 05/12S 3

4 Results Blank of SST Cell and Zr Cell By filling the 66 ml blank sample in both SST and Zr cells with quartz sand, five parallel extracts yielded a background of approximately 19.1 µg/l. The background of the blank in the 66 ml Zr cell was approximately 0.9 µg/l, lower than the LOQ of 1.6 µg/l. The limit of detection (LOD) was 0.5 µg/l. 2.5 mau Minutes FIGURE 1. Chromatograms of blank extractions using either an SST cell (2:66 ml) or a Zr cell (1:66 ml) Recovery and Reproducibility The recovery experiment was carried out by filling the 66 ml Zr cell with the quartz sand, adding the standard solution of CrVI, then performing the accelerated solvent extraction. The results are shown in Table 1. The average recoveries were 105.2% and 105.1%, with the relative standard deviation (RSD) 2.2% and 0.4% (n = 3) for the different added levels. TABLE 1. Recoveries and their RSDs (n = 3) at different added amounts of CrVI Amount Added (µg/l) Recovery % RSD % % Sample Detection Only one soil sample was used to evaluate this method. The chromatogram below in Figure 2 shows a CrVI content of 39.4 ng/g..9 Cr(VI) mau Minutes FIGURE 2. Chromatogram of CrVI from a soil sample known to be contaminated. 4 Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

5 Conclusion The preliminary results shown here for the extraction of CrVI from soil using accelerated solvent extraction demonstrates the potential for a faster extraction that is easy to use with minimal background contamination. The use of the Dionex ASE 350 is potentially a viable alternative to the labor-intensive EPA Method 3060A for compliance monitoring. Data showing improvements to the detection limits have been previously reported. 2,3 References 1.Giuriati, C.; Abballe, F.; Cristoforia. M.C.; and Gorni, A. Accelerated and Automated Extraction of Hexavalent Chromium from Solid Waste Materials Coupled with Ion Chromatography Determination. LC/GC Eur. 2005, 18 (4) Dionex (now part of Thermo Scientific) Application Note 144: Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography, 2002 [Online]. (accessed Apr. 9, 2012). 3.Dionex (now part of Thermo Scientific) Application Note 179: Carbohydrate and Amino Acid Analysis Using 3-D Amperometry, 2007 [Online]. webdocs/56256-an179_released pdf (accessed Apr. 9, 2012) Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. U.S./Canada (847) Brazil (55) Austria (43) Benelux (31) (32) Denmark (45) France (33) Germany (49) Ireland (353) Italy (39) Sweden (46) Switzerland (41) United Kingdom (44) Australia (61) China (852) India (91) Japan (81) Korea (82) Singapore (65) Taiwan (886) PN70094_E 08/16S

Mercury Speciation Determinations in Asian Dietary Supplements

Mercury Speciation Determinations in Asian Dietary Supplements Mercury Speciation Determinations in Asian Dietary Supplements Terri Christison, Deepali Mohindra, Frank Hoefler, and Linda Lopez, Thermo Fisher Scientific, Sunnyvale, California, USA Overview Purpose:

More information

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection Hua Yang, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

More information

Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography

Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography Application Note 8 Determination of Ethyl Sulfate Impurity in INTRODUCTION Indinavir sulfate is a specific and potent inhibitor of HIV- protease and is widely used in the treatment of AIDS. Indinavir sulfate

More information

Separation of Chromium (III) and Chromium (VI) by Ion Chromatography

Separation of Chromium (III) and Chromium (VI) by Ion Chromatography Application Update 65 Separation of Chromium (III) and Chromium (VI) by Ion Chromatography Introduction Chromium in the environment exists primarily in two oxidation states: Cr(III) and Cr(VI). While the

More information

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1.

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1. Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods Deanne Hurum and Jeffrey Rohrer, Thermo Fisher Scientific, Sunnyvale, CA, USA Introduction Sialic acids are critical

More information

Optimizing Sample. Chromium Analyses in Waters. Jane Timm, James Lovick Jr, Raymond Siery, ato 2011 NEMC, Bellevue, Washington

Optimizing Sample. Chromium Analyses in Waters. Jane Timm, James Lovick Jr, Raymond Siery, ato 2011 NEMC, Bellevue, Washington Optimizing Sample Preservation for Hexavalent Chromium Analyses in Waters Jane Timm, James Lovick Jr, Raymond Siery, and Yongtao Li Underwriters Laboratories ato 2011 NEMC, Bellevue, Washington 2011 Underwriters

More information

A HPAE-PAD Chromatographic Assay for Carbohydrates in Urine as a Measure of Intestinal Permeability

A HPAE-PAD Chromatographic Assay for Carbohydrates in Urine as a Measure of Intestinal Permeability A HPAE-PAD Chromatographic Assay for Carbohydrates in Urine as a Measure of Intestinal Permeability Deanna Hurum and Jeffrey Rohrer, Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Purpose: A high-performance

More information

Rapid determination of phosphate and citrate in carbonated soft drinks using ion chromatography

Rapid determination of phosphate and citrate in carbonated soft drinks using ion chromatography APPLICATION NOTE 7250 Rapid determination of phosphate and citrate in carbonated soft drinks using ion chromatography Authors Jingli Hu and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords

More information

Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS

Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS Application Note 269 Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS INTRODUCTION Ion chromatography (IC) has been extensively used as the preferred separation technique

More information

Determination of Plant-Derived Neutral Oligo- and Polysaccharides Using the CarboPac PA200

Determination of Plant-Derived Neutral Oligo- and Polysaccharides Using the CarboPac PA200 Application Update 15 Determination of Plant-Derived Neutral Oligo- and Polysaccharides Using the CarboPac PA2 INTRODUCTION High-performance anion-exchange chromatography with pulsed amperometric detection

More information

Determination of Phenolic Compounds in Apple Orchard Soil

Determination of Phenolic Compounds in Apple Orchard Soil Determination of Phenolic Compounds in Apple Orchard Soil Che Jinshui, Xu Qun, Liang Lina, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo Fisher Scientific, Sunnyvale,

More information

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography Application Note 65 Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography INTRODUCTION Preservatives are commonly added to many food products, such as soda, fruit juice,

More information

Determination of Carbohydrates in Kombucha Using HPAE-PAD

Determination of Carbohydrates in Kombucha Using HPAE-PAD Determination of Carbohydrates in Kombucha Using HPAE-PAD Beibei Huang, Jingli Hu, and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA USA Application Note 1152 Key Words Glucose, Fructose, Sucrose,

More information

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Fast Determination of Lactose and Lactulose in Dairy Products Using a 4 μm Particle Column and High-Performance Anion-Exchange Chromatography with Carl Fisher, Terri Christison, Hua Yang, Monika Verma,

More information

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages Application Note Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages INTRODUCTION The determination of inorganic anions and cations and organic acids in non-alcoholic

More information

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection Application Note 9 The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection INTRODUCTION The accurate measurement of the amount of sugar in final molasses

More information

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Measuring Phytosterols in Health Supplements by LC/MS Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Overview Purpose: Develop a method for the extraction of phytosterols

More information

Determination of Trace Sodium in Cranberry Powder

Determination of Trace Sodium in Cranberry Powder Determination of Trace Sodium in Cranberry Powder Brian De Borba and Jeffrey Rohrer; Thermo Fisher Scientific Inc., Sunnyvale, CA, USA Application Note 00 Key Words Dietary Supplements, Ion Chromatography,

More information

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction SM Rahmat Ullah, Kannan Srinivasan, Chris Pohl, Pranathi Perati and Alexander Zhang Thermo

More information

Application Note 178 INTRODUCTION

Application Note 178 INTRODUCTION Application Note 78 Improved Determination of Trace Concentrations of Perchlorate in Drinking Water Using Preconcentration with Two-Dimensional Ion Chromatography and Suppressed Conductivity Detection

More information

Application Note 178. Brian DeBorba and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA. 10 Typically, this requires the use of sample

Application Note 178. Brian DeBorba and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA. 10 Typically, this requires the use of sample Improved Determination of Trace Concentrations of Perchlorate in Drinking Water Using Preconcentration with Two-Dimensional Ion Chromatography and Suppressed Conductivity Detection Brian DeBorba and Jeff

More information

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages APPLICATION NOTE 21671 Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages Authors Jon Bardsley, Thermo Fisher Scientific, Runcorn, UK Keywords Vanquish Flex, Acclaim PolarAdvantage

More information

IC-MS Environmental Applications - Water Testing. Application Notebook

IC-MS Environmental Applications - Water Testing. Application Notebook IC-MS Environmental Applications - Water Testing Application Notebook Table of Contents Environmental Sample Using IC-MS 3 Environmental Applications: Water Testing Acrylamide in Water 4 Trace Urea in

More information

Determination of Carbachol In Ophthalmic Solutions Using a Reagent-Free Ion Chromatography System

Determination of Carbachol In Ophthalmic Solutions Using a Reagent-Free Ion Chromatography System Application Note 9 Determination of Carbachol In Ophthalmic s Using a Reagent-Free Ion Chromatography System INTRODUCTION Carbachol is a choline ester and a positively charged quaternary ammonium compound

More information

Determination of Tigecycline in a Cell Lysate

Determination of Tigecycline in a Cell Lysate Determination of Tigecycline in a Cell Lysate Chen Jing, Dai Zhenyu, Xu Qun, Lina Liang, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo Fisher Scientific, Sunnyvale,

More information

Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS Application Note 242 Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS INTRODUCTION Tobacco-specific nitrosamines (TSNA) are a group of carcinogens found only in tobacco products. They

More information

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column APPLICATION NOTE 21673 Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column Authors Aaron Lamb and Brian King, Thermo Fisher Scientific, Runcorn, UK

More information

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode Deteration of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode Bruce Bailey, Marc Plante, David Thomas, Qi Zhang and Ian Acworth Thermo Fisher

More information

Determination of Water- and Fat-Soluble Vitamins in Nutritional Supplements by HPLC with UV Detection

Determination of Water- and Fat-Soluble Vitamins in Nutritional Supplements by HPLC with UV Detection Application Note 5 Determination of Water- and Fat- Vitamins in Nutritional Supplements by HPLC with UV Detection Introduction Vitamins are a well-known a group of compounds that are essential for human

More information

HPAE-PAD Determination of Infant Formula Sialic Acid

HPAE-PAD Determination of Infant Formula Sialic Acid HPAE-PAD Determination of Infant Formula Sialic Acid Deanna Hurum and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 53 Introduction Dietary sialic acids are important for infant

More information

Thermo Scientific Dionex CarboPac MA1 Column

Thermo Scientific Dionex CarboPac MA1 Column CHROMATOGRAPHY Thermo Scientific Dionex CarboPac MA Column Product Specifications HPLC columns for the analysis of weakly ionizable carbohydrates. Simple methods with pulsed amperometric detection Reduced

More information

Determination of Organic Acids in Kombucha Using a High-Pressure Ion Chromatography System

Determination of Organic Acids in Kombucha Using a High-Pressure Ion Chromatography System APPLICATION NOTE Determination of Organic Acids in Kombucha Using a High-Pressure Ion Chromatography System No. 1157 Beibei Huang, Jingli Hu, and Jeffrey Rohrer Keywords Dionex IonPac AS11-HC-4µm column,

More information

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice Application Note Author Food Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India 8 6 4 2

More information

Separation of 2AA-Labeled N-Linked Glycans from Glycoproteins on a High Resolution Mixed-Mode Column

Separation of 2AA-Labeled N-Linked Glycans from Glycoproteins on a High Resolution Mixed-Mode Column Separation of 2AA-Labeled N-Linked Glycans from Glycoproteins on a High Resolution Mixed-Mode Column Udayanath Aich, Julian Saba, Xiaodong Liu, Jeff Rohrer, Jim Thayer and Chris Pohl, Thermo Fisher Scientific,

More information

INTRODUCTION. Application Note 176

INTRODUCTION. Application Note 176 Application Note 76 Determining Sub-ppb Perchlorate in Drinking Water Using Preconcentration/Matrix Elimination Ion Chromatography with Suppressed Conductivity Detection by U.S. EPA Method 34. INTRODUCTION

More information

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA The Utilization of Novel Platform in a LC-MS/MS Workflow for the Analysis of Vitamin D, Testosterone, Immunosuppressants, Chemotherapeutics and Cortisol Dayana Argoti, Kerry Hassell, Sarah J. Fair, and

More information

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring and in Urine Xiaolei Xie, Joe DiBussolo, Marta Kozak; Thermo Fisher Scientific, San Jose, CA Application Note 627 Key Words, norbuprenorphine,

More information

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase Subhra Bhattacharya and Stephen C. Roemer Thermo Fisher Scientific Global Chemicals, Fair Lawn, NJ, USA AbstrAct Liquid

More information

Determination of Iodide and Iodate in Seawater and Iodized Table Salt by HPLC with UV Detection

Determination of Iodide and Iodate in Seawater and Iodized Table Salt by HPLC with UV Detection Application Note 36 Determination of Iodide and Iodate in Seawater and Iodized Table Salt by HPLC with UV Detection INTRODUCTION Iodine is an essential element naturally present in seawater and seafood.

More information

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection Bruce Bailey, Marc Plante, David Thomas, Qi Zhang and Ian Acworth Thermo Fisher Scientific, Chelmsford,

More information

columns CarboPac PA20 Column

columns CarboPac PA20 Column columns CarboPac PA Column Anion-exchange columns for the analysis of mono- and disaccharides in glycoprotein therapeutics, foods, and beverages: Simple, direct approach using pulsed electrochemical detection

More information

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples A Rapid UHPLC Method for the Analysis of Biogenic Aes and Metabolites in Microdialysis Samples Bruce Bailey and Ian Acworth; Thermo Fisher Scientific, Chelmsford, MA Overview Purpose: To develop an Ultra

More information

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants Jeremy Matthews, 1 Alex Chen, 2 and Flavio Bedini 1 1 Thermo Fisher Scientific, Singapore; 2 Alpha Analytical Pte Ltd, Singapore Overview

More information

Determination of Methacholine Chloride and Potential Impurities Using a Reagent-Free Ion Chromatography System

Determination of Methacholine Chloride and Potential Impurities Using a Reagent-Free Ion Chromatography System Application Note 49 Determination of Methacholine Chloride and Potential Impurities Using a Reagent-Free Ion Chromatography System INTRODUCTION Methacholine chloride [-(acetyloxy)-n,n,ntrimethyl-1-propanaminium

More information

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine J. Jones, S. Westwood, T. Liddicoat, L. Pereira, T. Edge Thermo Fisher Scientific, Manor Park, Runcorn, UK Overview Purpose:

More information

Impurity Profiling of Carbamazepine by HPLC/UV

Impurity Profiling of Carbamazepine by HPLC/UV Application Note: 52049 Impurity Profiling of Carbamazepine by HPLC/UV Terry Zhang, Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela Hypersil GOLD Carbamazepine Drug Analysis

More information

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Application Note: 346 MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Gargi Choudhary and Diane Cho, Thermo Fisher Scientific, San Jose, CA Wayne Skinner and

More information

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine High-Throughput Quantitative LC-MS/MS Analysis of and 14 Benzodiazepines in Urine Bill Yu, Kristine Van Natta, Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 588 Key Words Opiates,

More information

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis X. Liu, C. Pohl, Dionex Corporation, Sunnyvale, CA, USA ABSTRACT Although reversed-phase (RP) silica columns (e.g., C18 and

More information

Determination of Polydextrose in Foods by AOAC Method

Determination of Polydextrose in Foods by AOAC Method Application Note 147 Determination of Polydextrose in Foods by AOAC Method 2000.11 INTRODUCTION Polydextrose is a synthetic polymer of glucose having mainly α1,6 linkages and an average degree of polymerization

More information

Determination of perchlorate in environmental waters using a compact ion chromatography system coupled with a single quadrupole mass spectrometer

Determination of perchlorate in environmental waters using a compact ion chromatography system coupled with a single quadrupole mass spectrometer APPLICATION UPDATE 72507 Determination of perchlorate in environmental waters using a compact ion chromatography system coupled with a single quadrupole mass spectrometer Authors Beibei Huang and Jeffrey

More information

Title Revision n date

Title Revision n date A. THIN LAYER CHROMATOGRAPHIC TECHNIQUE (TLC) 1. SCOPE The method describes the identification of hydrocortisone acetate, dexamethasone, betamethasone, betamethasone 17-valerate and triamcinolone acetonide

More information

[ Care and Use Manual ]

[ Care and Use Manual ] MALDI Calibration Kit I. Introduction The MALDI Calibration Kit is a conveniently packaged selection of MALDI matrices and calibration standards (includes high-purity Neg Ion Mode Calibrant for accurate

More information

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis The Investigation of Factors Contributing to Immunosuppressant Drugs Variability in LC-MS/MS Analysis Joseph Herman, Dayana Argoti, and Sarah Fair Thermo Fisher Scientific, Franklin, MA, USA Overview Purpose:

More information

Analysis of Chloride, Phosphate, Malate, Sulfite, Tartrate, Sulfate, and Oxalate in Red and White Wine

Analysis of Chloride, Phosphate, Malate, Sulfite, Tartrate, Sulfate, and Oxalate in Red and White Wine Application Work AW US6-0249-062017 Analysis of Chloride, Phosphate, Malate, Sulfite, Tartrate, Sulfate, Branch Food, stimulants, beverages, flavors Keywords IC; 940; Conductivity; Chloride; Phosphate;

More information

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 709 Key Words SPE, SOLA

More information

Rapid Determination of Phosphate and Citrate in Carbonated Soft Drinks Using a Reagent-Free Ion Chromatography System

Rapid Determination of Phosphate and Citrate in Carbonated Soft Drinks Using a Reagent-Free Ion Chromatography System Application Note 69 Rapid Determination of Phosphate and Citrate in Carbonated Soft Drinks Using a Reagent-Free Ion Chromatography System INTRODUCTION Soft drinks are complex mixtures containing a variety

More information

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method Liliana Krotz and Guido Giazzi Thermo Fisher Scientific, Milan, Italy Overview

More information

Determination of Lactose in Lactose-Free Milk Products by High-Performance. anion-exchange chromatography with. Pulsed Amperometric Detection

Determination of Lactose in Lactose-Free Milk Products by High-Performance. anion-exchange chromatography with. Pulsed Amperometric Detection Determination of Lactose in Lactose-Free Milk Products by High-Performance Anion-Exchange Chromatography with Pulsed Amperometric Detection Application Note 48 Introduction Lactose is the major disaccharide

More information

LC-MS/MS Method for the Determination of Tenofovir from Plasma

LC-MS/MS Method for the Determination of Tenofovir from Plasma LC-MS/MS Method for the Determination of Tenofovir from Plasma Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 687 Key Words SPE, SOLA CX, Hypersil GOLD, tenofovir Abstract

More information

A New Era in SFC Column Technology

A New Era in SFC Column Technology B TM A ew Era in SFC Column Technology Combining state-of-the-art media manufacturing with the industry leading column technology, Viridis Supercritical Fluid Chromatography (SFC) columns bring a new level

More information

Determination of Sucralose Using HPAE-PAD

Determination of Sucralose Using HPAE-PAD Application Note 59 Determination of Sucralose Using HPAE-PAD INTRODUCTION Sucralose is used in the manufacture of nonnutritive sweetened food and beverage products. Sucralose, like most carbohydrates,

More information

Determination of Steviol Glycosides by HPLC with UV and ELS Detections

Determination of Steviol Glycosides by HPLC with UV and ELS Detections Determination of Steviol Glycosides by HPLC with UV and ELS Detections Application Note 4 Introduction The stevia plant and extracts from stevia leaves have long been used as sweeteners in Asia and Latin

More information

Improving Food Process Control Using the

Improving Food Process Control Using the Improving Food Process Control Using the UPLC Amino Acid Analysis Solution Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION Food products have different desirability and value based in

More information

Sample Concentration and Analysis of Human Hormones in Drinking Water

Sample Concentration and Analysis of Human Hormones in Drinking Water Sample Concentration and Analysis of Human Hormones in Drinking Water Carl Fisher Applications Chemist Ion Chromatography/Sample Preparation Thermo Fisher Scientific March 1, 215 1 The world leader in

More information

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA An Ultra High Resolution Glycan Column for Isomeric Separation and the Structural Identification of Labeled N-Glycans from Proteins Including Antibodies Udayanath Aich, 1 Julian Saba, 2 Rosa Viner, 2 Shanhua

More information

autosamplers ASI-100 and ASI-100T Autosamplers

autosamplers ASI-100 and ASI-100T Autosamplers autosamplers ASI-100 and ASI-100T Autosamplers The ASI-100 and ASI-100T Automated Sample Injectors are highly flexible autosamplers for automating high performance liquid chromatography (HPLC) methods.

More information

Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA Overview

More information

Sugars in Honey Using HPAE-PAD: What Is the Best Column?

Sugars in Honey Using HPAE-PAD: What Is the Best Column? IC COLUMNS FOR ANALYZING SUGARS IN HONEY Sugars in Honey Using HPAE-PAD: What Is the Best Column? Katrin Hostettler, 1 Robert Brogioli, 1 Silvio Arpagaus, 1 Beate Müller-Werner, 2 Detlef Jensen 2 1 Department

More information

Two-Dimensional HPLC Determination of Water-Soluble Vitamins in a Nutritional Drink

Two-Dimensional HPLC Determination of Water-Soluble Vitamins in a Nutritional Drink Two-Dimensional HPLC Determination of Water-Soluble Vitamins in a Nutritional Drink Dai Zhenyu, Chen Jing, Xu Qun, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo

More information

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer APPLICATION NOTE 72819 Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer Author Maria Grübner; Thermo Fisher Scientific, Germering,

More information

Two-dimensional HPLC determination of water-soluble vitamins in a nutritional drink

Two-dimensional HPLC determination of water-soluble vitamins in a nutritional drink PPLICTION NOTE 0 Two-dimensional HPLC determination of water-soluble vitamins in a nutritional drink uthors Dai Zhenyu, Chen Jing, Xu Qun, and Jeffrey Rohrer, Thermo Fisher Scientific, Shanghai, People

More information

Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA An Improved Global Method for the Quantitation and Characterization of Lipids by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, and Ian N. Acworth

More information

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC METHOD APPROVALS Norvin Pharma Inc. Author Analytical Laboratory Approver Analytical Laboratory Group Leader Approver Manager Quality Control Chemistry

More information

GRAC - Perchlorate in Groundwater. Enhancement of Perchlorate Analysis using Solid Phase Extraction (SPE) Cartridges

GRAC - Perchlorate in Groundwater. Enhancement of Perchlorate Analysis using Solid Phase Extraction (SPE) Cartridges Enhancement of Perchlorate Analysis using Solid Phase Extraction (SPE) Cartridges Victor F. Medina*, Steven L. Larson, Barbara Extine United States Army Corp of Engineers Engineer Research and Development

More information

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow APPLICATIN NTE 7601 Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow Authors Sylvia Grosse, Mauro De Pra, Frank Steiner Thermo Fisher Scientific, Germering, Germany

More information

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D Julie A. Horner 1, Marta Kozak 1, Subodh Nimkar 1, and August A. Specht 1 1 Thermo Fisher Scientific, San Jose,

More information

Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection

Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection David Thomas, Ian Acworth, Bruce Bailey, Marc Plante, and Qi Zhang Thermo Fisher Scientific, Chelmsford, MA Overview

More information

Application Note. Authors. Abstract. Food

Application Note. Authors. Abstract. Food Determination of Hormones in Shrimp by Agilent 129 Infinity LC with Agilent Poroshell 12 LC Column and Agilent Bond Elut QuEChERS for Sample Preparation Application Note Food Authors Rongjie Fu and Andy

More information

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes Catalin E. Doneanu, Weibin Chen, and Jeffrey R. Mazzeo Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B E N E F

More information

Determination of taurine in energy drinks by high-performance liquid chromatography

Determination of taurine in energy drinks by high-performance liquid chromatography Determination of taurine in energy drinks by high-performance liquid chromatography Brad McConn Department of Chemistry, Concordia College, 901 8 th St S, Moorhead, MN 56562 Abstract The concentration

More information

Ensuring High Sensitivity and Consistent Response in UHPLC-MS Analyses

Ensuring High Sensitivity and Consistent Response in UHPLC-MS Analyses Ensuring High ensitivity and Consistent Response in UHPLC-M Analyses Markus Martin, Michael Heidorn, Frank teiner, Tobias Fehrenbach, and Fraser McLeod, Dionex Corporation, Germering, Germany Introduction

More information

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose:

More information

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT. Objective The aim of the study presented is to determine the presence of target Gingkolides in Gingko biloba leaf extract. The data presented shows Gingkolides determined to be present in the extract analyzed.

More information

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS TECHNICAL NOTE 21882 Robust extraction, separation, and quantitation of structural isomer steroids human plasma by SPE-UHPLC-MS/MS Authors Jon Bardsley 1, Kean Woodmansey 1, and Stacy Tremintin 2 1 Thermo

More information

Elemental Scientific. Automated Speciation. prepfast-ic Chromium Speciation for Toy Materials (EN 71-3) ICP ICPMS AA

Elemental Scientific. Automated Speciation. prepfast-ic Chromium Speciation for Toy Materials (EN 71-3) ICP ICPMS AA prepfast-ic Automated Speciation prepfast-ic Chromium Speciation for Toy Materials (EN 71-3) Presented at the 015 European Winter Plasma Conference Authors: Aurelien Viscardi, Kyle Uhlmeyer & Patrick Sullivan

More information

Glycoprotein monosaccharide analysis using HPAE-PAD with manually prepared eluent

Glycoprotein monosaccharide analysis using HPAE-PAD with manually prepared eluent TECHNICAL NOTE 75 Glycoprotein monosaccharide analysis using HPAE-PAD with manually prepared eluent Authors Sachin Patil and Jeffrey Rohrer, Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Glycoprotein,

More information

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials D. Edwards 2, D. Lennartz 1, L.Shick 2 1 Thermo Fisher Scientific, Langerwehe, Germany 2 Thermo Fisher

More information

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Jenny Chen, Hongxia Wang, Zhiqi Hao, Patrick Bennett, and Greg Kilby Thermo Fisher

More information

A Versatile Detector for the Sensitive and Selective Measurement of Numerous Fat-Soluble Vitamins and Antioxidants in Human Plasma and Plant Extracts

A Versatile Detector for the Sensitive and Selective Measurement of Numerous Fat-Soluble Vitamins and Antioxidants in Human Plasma and Plant Extracts A Versatile Detector for the Sensitive and Selective Measurement of Numerous Fat-Soluble Vitamins and Antioxidants in Human Plasma and Plant Extracts Ian Acworth, Bruce Bailey, Paul Gamache, and John Waraska

More information

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Valérie Thibert, Bénédicte Duretz Thermo Fisher Scientific, Courtaboeuf,

More information

Fast UHPLC Methods for Analysis of Amino Acids

Fast UHPLC Methods for Analysis of Amino Acids Fast UHPLC Methods for Analysis of Ao Acids Qi Zhang, Bruce Bailey, Marc Plante, David Thomas, and Ian Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: Methods that offer both rapid

More information

Accumulation and transformation of inorganic and organic arsenic in rice and role of

Accumulation and transformation of inorganic and organic arsenic in rice and role of Supplementary Information: Accumulation and transformation of inorganic and organic arsenic in rice and role of thiol-complexation to restrict their translocation to shoot Seema Mishra 1,2 *, Jürgen Mattusch

More information

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE Translated English of Chinese Standard: GB5009.85-2016 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB 5009.85-2016

More information

Determination of Carbohydrates in Beverages and Milk Products by HPAE-PAD and Capillary HPAE-PAD

Determination of Carbohydrates in Beverages and Milk Products by HPAE-PAD and Capillary HPAE-PAD Determination of Carbohydrates in Beverages and Milk Products by HPAE-PAD and Capillary HPAE-PAD Terri Christison, Monika Verma, Aaron Kettle, Carl Fisher, Linda Lopez; Thermo Fisher Scientific, Sunnyvale,

More information

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Application Note Food Safety Authors Chenhao Zhai Agilent

More information

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites Valérie Thibert 1, Norbert Dirsch 2, Johannes Engl 2, Martin Knirsch 2 1 Thermo Fisher Scientific,

More information

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column Application Note Pharmaceutical and Food Testing Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column Author Lu Yufei Agilent Technologies, Inc. Abstract A liquid chromatographic

More information

Determination of Amantadine Residues in Chicken by LCMS-8040

Determination of Amantadine Residues in Chicken by LCMS-8040 Liquid Chromatography Mass Spectrometry Determination of Amantadine Residues in Chicken by LCMS-8040 A method for the determination of amantadine in chicken was established using Shimadzu Triple Quadrupole

More information