Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Size: px
Start display at page:

Download "Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D"

Transcription

1 Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D Julie A. Horner 1, Marta Kozak 1, Subodh Nimkar 1, and August A. Specht 1 1 Thermo Fisher Scientific, San Jose, CA

2 Overview Purpose: Comparison of ion trap full scan methods for quantitation of 25-OH Vitamin D2 and D3 in human plasma samples. Methods: Liquid-liquid extraction followed by LC-MS2 or LC-MS3 analysis of fresh plasma samples. Results: Both methods are acceptable for Vitamin D quantitation. Full scan MS2 provides superior sensitivity whereas full scan MS3 provides superior selectivity. Introduction There has been a growing interest in Vitamin D and its role in human metabolism. Vitamin D is metabolized to 25-OH D in the liver. Total vitamin D is best determined by measuring total 25-hydroxyvitamin D (D 2 and D 3 ) in serum because the half-life of 25- OH D is about three weeks with serum concentrations of 1 5. Vitamin D supplementation in both food and tablets comes in both the D 2 and D 3 forms, making it imperative to measure 25-OH D 2 and 25-OH D 3. Although optimal serum concentrations of total 25-OH D are generally agreed to be 3, there is considerable discussion on the serum concentration of 25-OH D considered to be inadequate for bone and overall health, but < is generally regarded as deficient. Serum concentrations >1 are generally regarded as potentially toxic. We have developed dual pressure linear ion trap methods based on Full Scan MS2 and Full Scan MS3 for maximum sensitivity and selectivity and therefore precision and accuracy in measurement of 25-OH and D3. Methods Sample Preparation Plasma samples were prepared at a university clinical laboratory using the liquid-liquid extraction procedure described below: ul each sample is treated with ul of.1 M NaOH, vortexed and left at room temperature for 15 minutes ul of internal standard is added and vortexed 3. 2 ml of n-heptane is added and vortexed 1 minute. 4. The mixture is centrifuged at 15 rpm for 4 minutes. 5. The tubes are placed in a dry ice:acetone bath until the aqueous layer freezes; the top layer is then discarded. 6. The remainder is dried under nitrogen at 37 C and reconstituted with 1 ul of : mobile phase. Liquid Chromatography Liquid-liquid extracted plasma samples were injected using a Thermo Scientific Accela Open AS and separated using a reversed-phase gradient delivered by a Thermo Scientific Accela 125 uhplc pump. The 5 minute method uses a 2.1x5 mm Phenomenex Luna c18 column and a flow rate of ul/min. Both gradient and column were duplicated from a clinical laboratory. Mass Spectrometry Atmospheric pressure chemical ionization source conditions were optimized for best signal to noise. Mass spectrometric detection on a dual pressure linear ion trap was achieved using 3 Full Scan MS3 or MS2 scan events for the water loss of 25-OH, 25-OH and the internal standard 25- OH -d6.. Data Analysis All data were processed using Thermo Scientific LC Quan Quantitation software in plasma 1.5 in plasma 2 Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

3 Results Challenges in Vitamin D Analysis Vitamin D is typically not found free in serum samples due to a high probability of protein binding and thus poses a challenge for sensitive and reproducible high performance liquid chromatography (HPLC). Additionally, various serum sample matrix constituents are found to cause ion suppression that reduces accuracy and reproducibility during real sample analyses. This ion suppression can be especially troublesome with rapid LC mass spectrometry (MS) methods where insufficient resolution can be observed between matrix contaminants and analyte peaks. 25-OH 25-OH C:\j.horner\...\Raw files\p-46 8/4/11 9:51:3 Relative Abundance m/z Vitamin D Quantitation Using Full Scan MS2 In addition, Vitamins D2 and D3 are fairly linear structures resulting in multiple fragmentation channels and distributed ion current in MS2 experiments. The full scan MS2 total and base peak extracted ion currents for 25-OH, 25-OH Vitamin D3 and the internal standard shown in Figures 1 a and b respectively show an improvement in S/N with approximately five-fold difference in intensity. FIGURE 1a. Full Scan MS2 Total Ion Current Chromatograms for 25-OH (top), 25-OH - d6 (top) and 25-OH (bottom) for Sample in plasma in 2.5 plasma NL: 1.57E7 TIC NL: 1.57E7 Full MS @cid29. [.-28.] NL: 2.96E6 Full MS @cid29. [.-28.] NL: 1.35E5 Full MS @cid29. [.-28.] FIGURE 1b. Full Scan MS2 Base Peak Extracted Ion Chromatograms for 25-OH (top), 25-OH -d6 (middle) and 25-OH (bottom) for Sample RT: SN: RT: SN: 8813RMS RT: 2.65 SN: NL: 1.57E7 TIC NL: 3.88E5 m/z= Full MS @cid29. [.-28.] NL: 4.22E5 m/z= Full MS @cid29. [.-28.] NL: 2.E4 m/z= Full MS @cid29. [.-28.] However, the increase in selectivity of the base peak extracted in chromatograms is necessary for accurate peak identification and quantitation of target levels. Using the total ion current of the full scan MS2 experiment, even with a restricted product ion mass scan range, results in significant interference FIGURE 3. Standard Curves(right) for obtained using Full Scan MS2. Accura tables (left) The peak top full scan MS2 mass spectra shown in Figure 2 demonstrate the issue surrounding multiple fragmentation channels, particularly for 25-OH and the internal standard. These additional channels result in noise in the chromatogram as well as reduced sensitivity for the extracted ion chromatograms. Vitamin D Quantitation Using Full Sca The majority of LC-MS methods for Vitam types. The sensitivity and simplicity of a desirable for any analysis; however, as s chromatograms, selectivity is also require there can be interference in a given trans such as plasma. Thermo Scientific Poster Note PN63565_E 6/12S 3

4 FIGURE 2. Peak Top Full Scan MS2 Spectra for 25-OH (top), Internal Standard (middle), and 25-OH (bottom) for Sample 46. C:\j.horner\...\Raw files\p-46 Relative Abundance 8/4/11 9:51:33 PM NL: 2.37E5 1 P-46# RT: @cid29. [.-28.] NL: 2.91E5 1 P-46# RT: AV: 5 F: ITMS + c APCI @cid29. [.-28.] NL: 1.64E4 1 P-46# RT: @cid29. [.-28.] m/z The Vitamin D analysis is done for a set of real samples. There is enough material for each quality control (4 total), standard (5 total) and real sample (9 total) to be injected just once. Our clinical laboratory collaborators carry out the analysis and send the same sample set to us for comparison. Figure 3 shows the standard curve linearity and accuracy for both 25-OH and 25-OH done using our full scan MS2 method. The analysis of Vitamin D in fresh human plasma samples was successfully carried out using ion trap based full scan MS2; both accuracy and limits of quantification are within desired specifications. FIGURE 3. Standard Curves(right) for 25-OH and 25-OH obtained using Full Scan MS2. Accuracies for each Standard are listed in the tables (left) Y =.2+.2*X R^2 =.9953 W: 1/X^ FIGURE 4a. Full Scan MS3 Total Ion Current Chromatograms for 25-OH, 25-OH and 25-OH -d6 f or Sample in plasma 2.58 d in plasma TIC MS p-45_82211 Full MS @cid @cid24. [1.-28 p-45_82211 Full MS @cid @cid26. [1.-28 p-45_82211 NL: 1.38E4 Full MS @cid @cid26. [1.-28 p-45_82211 In this study, we use a full scan MS3 ap most intense product ion from the MS2 experiment which is inherently selective also sensitive. The total ion current chromatograms an experiments are shown in Figures 4a an the MS3 experiment is sufficiently selec level of detection, 1 (bottom trace the base peak extracted ion chromatogr noise ratios for all components are with scan MS2 experiment. The combination of sensitivity and selec and D3 (Figure 5) and accu scan MS2 experiment (Figure 2) Y = *X R^2 =.9911 W: 1/X^2 chromatograms is et levels. Using the ricted product ion nstrate the issue H and the well as reduced Vitamin D Quantitation Using Full Scan MS3 The majority of LC-MS methods for Vitamin D are based on MS2 (including SRM) scan types. The sensitivity and simplicity of a total ion current measurement is highly desirable for any analysis; however, as seen in the TIC versus base peak extracted ion chromatograms, selectivity is also required. Even when the SRM experiment is used, there can be interference in a given transition particularly in a complex sample matrix such as plasma. 4 Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

5 NL: 2.37E5 P-46# RT: [.-28.] NL: 2.91E5 P-46# RT: AV: 5 F: ITMS + c APCI 389.5@cid29. [.-28.] NL: 1.64E4 P-46# RT: @cid29. [.-28.] FIGURE 4a. Full Scan MS3 Total Ion Current Chromatograms for 25-OH, 25-OH and 25-OH -d6 f or Sample in plasma 2.58 d in plasma TIC MS p-45_82211 Full MS @cid @cid24. [1.-28.] p-45_82211 Full MS @cid @cid26. [1.-28.] p-45_82211 NL: 1.38E4 Full MS @cid @cid26. [1.-28.] p-45_82211 FIGURE 4b. Full Scan MS3 Total Ion Chromatograms for 25-OH vitamin D2, 25-OH and Full Scan MS3 Base Peak Extracted Ion Chromatogram for 25-OH -d6 for Sample RT: SN: 1123RMS RT: 2.58 SN: 3319RMS RT: 2.63 SN: INF TIC MS p-45_82211 Full MS @cid @cid24. [1.-28.] p-45_82211 Full MS @cid @cid26. [1.-28.] p-45_82211 NL: 6.59E3 m/z= Full MS @cid @cid26. [1.-28.] p-45_ Vitamin D Anaylsis in Real Samples. We analyzed several sets of samples a shown) and MS3 (Figure 6). Good corr and (bottom). The exceptio horizontal axis were reported as below real values by Thermo In this study, we use a full scan MS3 approach for data acquisition. We chose the most intense product ion from the MS2 spectrum for each target to give an MSn experiment which is inherently selective and, through reduction of noise contributions, also sensitive. 5 5-OH are listed in the The total ion current chromatograms and signal to noise ratios for the full scan MS3 experiments are shown in Figures 4a and 4b. Immediately of note is that the TIC in the MS3 experiment is sufficiently selective to use for quantitation, even at the lowest level of detection, 1 (bottom trace Figure 4a) although for this study we use the base peak extracted ion chromatogram (bottom trace Figure 4b). Also, signall to noise ratios for all components are within a factor of two of those obtained in the full scan MS2 experiment. The combination of sensitivity and selectivity gives quantification results for 25-OH and D3 (Figure 5) and accuracies comparable to those seen in the full scan MS2 experiment (Figure 2) FIGURE 5. Standard Curves (right) for 25-OH and 25-OH obtained using Full Scan MS3. Accuracies for each Standard Level are Listed in the Tables (left). 4. Y = *X R^2 =.9988 W: 1/X^2 =.9911 W: 1/X^ including SRM) scan ment is highly e peak extracted ion experiment is used, mplex sample matrix Y = *X R^2 =.9944 W: 1/X^ Thermo Scientific Poster Note PN63565_E 6/12S 5

6 3. TIC MS p-45_82211 Full MS3 [1.-28.] p-45_82211 Full MS3 [1.-28.] p-45_82211 NL: 6.59E3 m/z= Full MS3 [1.-28.] p-45_ We chose the o give an MSn noise contributions, the full scan MS3 is that the TIC in, even at the lowest this study we use b). Also, signall to btained in the full esults for 25-OH seen in the full FIGURE 6. Correlation Plot of Full Scan MS3 Data Obtained on Velos Pro versus SRM Data Obtained at the Collaborator Site. Vitamin D Anaylsis in Real Samples. We analyzed several sets of samples against collaborator data using both MS2 (not shown) and MS3 (Figure 6). Good correlation is observed for both (top) and (bottom). The exceptions are several points at the origin of the horizontal axis were reported as below limit of quantification by the collaborator but as real values by Thermo Collaborator SRM Results Conclusion Collaborator SRM Results Vitamin D2 Vitamin D3 The full scan MSn analysis of Vitamin D in plasma results indicate the following: Both full scan MS2 and full scan MS3 provide quantitation and accuracies suitable for analysis of Vitamin D in samples.. Full Scan MS2 provides higher sensitivity for this analysis, full scan MS3 provides better selectivity. Ion trap based methods can be used for quantitation of a small number of compounds in a complex matrix. These methods will be used to determine suitability for quantitation in singly crashed plasma rather than LLE plasma samples. Acknowledgements We would like to thank Kara Lynch and Shannon Kastner from the University of California - San Francisco for supplying fresh plasma sample sets. These methods are For Research use Only. Not for use in Diagnostic Procedures. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. 6 Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

7 12 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, San Jose, CA USA is ISO Certified. Africa-Other Australia Austria Belgium Canada China Denmark Europe-Other Finland/Norway/Sweden France Germany India Italy Japan Latin America Middle East Netherlands New Zealand Russia/CIS South Africa Spain Switzerland UK USA PN63565_E 6/12S

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Valérie Thibert, Bénédicte Duretz Thermo Fisher Scientific, Courtaboeuf,

More information

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Jenny Chen, Hongxia Wang, Zhiqi Hao, Patrick Bennett, and Greg Kilby Thermo Fisher

More information

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring and in Urine Xiaolei Xie, Joe DiBussolo, Marta Kozak; Thermo Fisher Scientific, San Jose, CA Application Note 627 Key Words, norbuprenorphine,

More information

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis The Investigation of Factors Contributing to Immunosuppressant Drugs Variability in LC-MS/MS Analysis Joseph Herman, Dayana Argoti, and Sarah Fair Thermo Fisher Scientific, Franklin, MA, USA Overview Purpose:

More information

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine High-Throughput Quantitative LC-MS/MS Analysis of and 14 Benzodiazepines in Urine Bill Yu, Kristine Van Natta, Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 588 Key Words Opiates,

More information

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Application Note: 346 MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Gargi Choudhary and Diane Cho, Thermo Fisher Scientific, San Jose, CA Wayne Skinner and

More information

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants Jeremy Matthews, 1 Alex Chen, 2 and Flavio Bedini 1 1 Thermo Fisher Scientific, Singapore; 2 Alpha Analytical Pte Ltd, Singapore Overview

More information

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA The Utilization of Novel Platform in a LC-MS/MS Workflow for the Analysis of Vitamin D, Testosterone, Immunosuppressants, Chemotherapeutics and Cortisol Dayana Argoti, Kerry Hassell, Sarah J. Fair, and

More information

Impurity Profiling of Carbamazepine by HPLC/UV

Impurity Profiling of Carbamazepine by HPLC/UV Application Note: 52049 Impurity Profiling of Carbamazepine by HPLC/UV Terry Zhang, Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela Hypersil GOLD Carbamazepine Drug Analysis

More information

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites Valérie Thibert 1, Norbert Dirsch 2, Johannes Engl 2, Martin Knirsch 2 1 Thermo Fisher Scientific,

More information

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs Eric Niederkofler, 1 Dobrin Nedelkov, 1 Urban Kiernan, 1 David Phillips, 1 Kemmons Tubbs, 1 Scott Peterman, 2 Bryan Krastins,

More information

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 709 Key Words SPE, SOLA

More information

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Measuring Phytosterols in Health Supplements by LC/MS Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Overview Purpose: Develop a method for the extraction of phytosterols

More information

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method Liliana Krotz and Guido Giazzi Thermo Fisher Scientific, Milan, Italy Overview

More information

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes Dominic Baeumlisberger 2, Christopher Kurz 3, Tabiwang N. Arrey, Marion Rohmer 2, Carola Schiller 3, Thomas

More information

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase Subhra Bhattacharya and Stephen C. Roemer Thermo Fisher Scientific Global Chemicals, Fair Lawn, NJ, USA AbstrAct Liquid

More information

PIF: Precursor Ion Fingerprinting Searching for a Structurally Diagnostic Fragment Using Combined Targeted and Data Dependent MS n

PIF: Precursor Ion Fingerprinting Searching for a Structurally Diagnostic Fragment Using Combined Targeted and Data Dependent MS n Application Note: 441 PIF: Precursor Ion Fingerprinting Searching for a Structurally Diagnostic Fragment Using Combined Targeted and Data Dependent MS n Julie A. Horner, Rohan A. Thakur, Thermo Fisher

More information

Dynamic Analysis of HIV-Human Protein-Protein Interactions During Infection

Dynamic Analysis of HIV-Human Protein-Protein Interactions During Infection Dynamic Analysis of HIV-Human Protein-Protein Interactions During Infection Jeffrey Johnson, 1 Shannon Eliuk, 2 Amnon Golan, 1 Tasha Johnson, 1 Vlad Zabrouskov, 2 Nevan Krogan 1 1 UCSF, San Francisco,

More information

LC-MS/MS Method for the Determination of Tenofovir from Plasma

LC-MS/MS Method for the Determination of Tenofovir from Plasma LC-MS/MS Method for the Determination of Tenofovir from Plasma Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 687 Key Words SPE, SOLA CX, Hypersil GOLD, tenofovir Abstract

More information

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode Deteration of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode Bruce Bailey, Marc Plante, David Thomas, Qi Zhang and Ian Acworth Thermo Fisher

More information

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine J. Jones, S. Westwood, T. Liddicoat, L. Pereira, T. Edge Thermo Fisher Scientific, Manor Park, Runcorn, UK Overview Purpose:

More information

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries FT-Raman Surface Mapping of Remineralized Artificial Dental Caries I. Stangel, 1 R. Rubinovitz, 2 G.D. Arndt, 3 D. Byerly, 3 C. Theriot, 3 J.E. Kerr, 4 1 BioMat Sciences, Bethesda, MD; 2 Thermo Fisher

More information

Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS

Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS Terry Zhang, Reiko Kiyonami, Leo Wang and Guifeng Jiang Thermo Fisher Scientific, San Jose, CA, USA Application Note 569 Key

More information

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Marta Kozak Clinical Research Applications Group Thermo Fisher Scientific San Jose CA Clinical Research use only, Not for

More information

Thermo Scientific LipidSearch Software for Lipidomics Workflows. Automated Identification and Relative. Quantitation of Lipids by LC/MS

Thermo Scientific LipidSearch Software for Lipidomics Workflows. Automated Identification and Relative. Quantitation of Lipids by LC/MS Thermo Scientific LipidSearch Software for Lipidomics Workflows Automated Identification and Relative of Lipids by LC/MS The promise of lipidomics Lipidomics is a new field of study crucial for understanding

More information

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection Bruce Bailey, Marc Plante, David Thomas, Qi Zhang and Ian Acworth Thermo Fisher Scientific, Chelmsford,

More information

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection Hua Yang, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

More information

Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots Xiaolei Xie and Marta Kozak Thermo Fisher Scientific, San Jose, CA, USA Application

More information

Using Multiple Mass Defect Filters and Higher Energy Collisional Dissociation on an LTQ Orbitrap XL for Fast, Sensitive and Accurate Metabolite ID

Using Multiple Mass Defect Filters and Higher Energy Collisional Dissociation on an LTQ Orbitrap XL for Fast, Sensitive and Accurate Metabolite ID Application ote: 417 Using Multiple Mass Defect Filters and Higher Energy Collisional Dissociation on an LTQ rbitrap XL for Fast, Sensitive and Accurate Metabolite ID Yingying Huang 1, Shirley Liu 2, Shichang

More information

Mercury Speciation Determinations in Asian Dietary Supplements

Mercury Speciation Determinations in Asian Dietary Supplements Mercury Speciation Determinations in Asian Dietary Supplements Terri Christison, Deepali Mohindra, Frank Hoefler, and Linda Lopez, Thermo Fisher Scientific, Sunnyvale, California, USA Overview Purpose:

More information

Vitamin D3 and related compounds by ESI and APCI

Vitamin D3 and related compounds by ESI and APCI Liquid Chromatography Mass Spectrometry SSI-LCMS-9 Vitamin D and related compounds by ESI and APCI LCMS-8 Summary Vitamin D and related compounds were measured by LC-ESI/APCI-MS-MS. Background Accurate

More information

Increased Identification Coverage and Throughput for Complex Lipidomes

Increased Identification Coverage and Throughput for Complex Lipidomes Increased Identification Coverage and Throughput for Complex Lipidomes Reiko Kiyonami, David Peake, Yingying Huang, Thermo Fisher Scientific, San Jose, CA USA Application Note 607 Key Words Q Exactive

More information

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis Providing a Universal, ne-step Alternative to Liquid-Liquid Extraction in Bioanalysis Jessalynn P. Wheaton, Erin E. Chambers, and Kenneth J. Fountain APPLICATIN BENEFITS n Simple, one-step sample preparation

More information

Results of Using Re-washed Vials and Closures

Results of Using Re-washed Vials and Closures Results of Using Re-washed Vials and Closures Loy Shick, Thermo Fisher Scientific, Rockwood, TN, USA Brian King, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20670 Key Words Autosampler,

More information

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS New Solvent Grade Targeted for Trace Analysis by UHPLC-MS Subhra Bhattacharya, Deva H. Puranam, and Stephen C. Roemer Thermo Fisher Scientific Fisher Chemical, One Reagent Lane, Fair Lawn, NJ Material

More information

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS New Solvent Grade Targeted for Trace Analysis by UHPLC-MS Subhra Bhattacharya, Deva H. Puranam, and Stephen C. Roemer Thermo Fisher Scientific Fisher Chemical, One Reagent Lane, Fair Lawn, NJ Material

More information

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples A Rapid UHPLC Method for the Analysis of Biogenic Aes and Metabolites in Microdialysis Samples Bruce Bailey and Ian Acworth; Thermo Fisher Scientific, Chelmsford, MA Overview Purpose: To develop an Ultra

More information

Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research

Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research TEHNIL NOTE 64969 Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research uthors Kristine Van Natta, Neloni Wijeratne, laudia Martins Thermo Fisher Scientific,

More information

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS TECHNICAL NOTE 21882 Robust extraction, separation, and quantitation of structural isomer steroids human plasma by SPE-UHPLC-MS/MS Authors Jon Bardsley 1, Kean Woodmansey 1, and Stacy Tremintin 2 1 Thermo

More information

Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range

Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range Marika Raitio and Jorma Lampinen Thermo Fisher Scientific, Vantaa, Finland Key Words

More information

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA An Ultra High Resolution Glycan Column for Isomeric Separation and the Structural Identification of Labeled N-Glycans from Proteins Including Antibodies Udayanath Aich, 1 Julian Saba, 2 Rosa Viner, 2 Shanhua

More information

LC-MS/MS quantitative analysis of Polyunsaturated Omega 3, 6,7 and 9 Fatty Acids in Serum for

LC-MS/MS quantitative analysis of Polyunsaturated Omega 3, 6,7 and 9 Fatty Acids in Serum for POSTER NOTE 64921 LC-MS/MS quantitative analysis of Polyunsaturated Omega 3, 6,7 and 9 Fatty Acids in Serum for Research Use LC-MS/MS quantitative analysis of Poly Rory M Doyle*, Thermo Scientific, Inc,

More information

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS New Solvent Grade Targeted for Trace Analysis by UHPLC-MS Subhra Bhattacharya, Deva H. Puranam, and Stephen C. Roemer Thermo Fisher Scientific Fisher Chemical, One Reagent Lane, Fair Lawn, NJ Material

More information

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices Application ote #LCMS-2 esquire series Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices Introduction The simple monitoring

More information

Essential Lipidomics Experiments using the LTQ Orbitrap Hybrid Mass Spectrometer

Essential Lipidomics Experiments using the LTQ Orbitrap Hybrid Mass Spectrometer Application Note: 367 Essential Lipidomics Experiments using the LTQ rbitrap Hybrid Mass Spectrometer Thomas Moehring 1, Michaela Scigelova 2, Christer S. Ejsing 3, Dominik Schwudke 3, Andrej Shevchenko

More information

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Keywords Introduction Application Note PHA0413 High Pressure Liquid Chromatography

More information

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer APPLICATION NOTE 72819 Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer Author Maria Grübner; Thermo Fisher Scientific, Germering,

More information

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis Application Note AN807 Extraction of Aflatoxin M From Infant Formula Using ISLUTE Myco Page Extraction of Aflatoxin M From Infant Formula Using ISLUTE Myco SPE Columns prior to LC-MS/MS Analysis This application

More information

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Fast Determination of Lactose and Lactulose in Dairy Products Using a 4 μm Particle Column and High-Performance Anion-Exchange Chromatography with Carl Fisher, Terri Christison, Hua Yang, Monika Verma,

More information

ab Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric)

ab Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric) Version 10b Last updated 19 December 2018 ab118970 Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric) For the measurement of Lipid Peroxidation in plasma, cell culture and tissue extracts.

More information

ab65336 Triglyceride Quantification Assay Kit (Colorimetric/ Fluorometric)

ab65336 Triglyceride Quantification Assay Kit (Colorimetric/ Fluorometric) Version 10 Last updated 19 December 2017 ab65336 Triglyceride Quantification Assay Kit (Colorimetric/ Fluorometric) For the measurement of triglycerides in various samples. This product is for research

More information

All stocks and calibration levels were prepared in water: methanol (50:50) v/v to cover range of all steroid concentrations (refer Table 1).

All stocks and calibration levels were prepared in water: methanol (50:50) v/v to cover range of all steroid concentrations (refer Table 1). Application LCMS-8040 Simultaneous determination of 11 steroids and Vitamin D2/D3 in human serum using LC/MS/MS - Introduction Quantification of endogenous hormonal steroids and their precursors is essential

More information

PosterReprint OVERVIEW

PosterReprint OVERVIEW OVERVIEW A rapid multi-residue GC/MS method was developed for pesticides in foodstuffs. Plum extract was used as an example matrix. Analysis time was reduced from minutes to minutes. pesticides were analysed

More information

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction SM Rahmat Ullah, Kannan Srinivasan, Chris Pohl, Pranathi Perati and Alexander Zhang Thermo

More information

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials D. Edwards 2, D. Lennartz 1, L.Shick 2 1 Thermo Fisher Scientific, Langerwehe, Germany 2 Thermo Fisher

More information

Designer Fentanyls Drugs that kill and how to detect them. Cyclopropylfentanyl

Designer Fentanyls Drugs that kill and how to detect them. Cyclopropylfentanyl Designer Fentanyls Drugs that kill and how to detect them Cyclopropylfentanyl Science for a safer world The in vitro metabolism of cyclopropylfentanyl Simon Hudson & Charlotte Cutler, Sport and Specialised

More information

A Fully Integrated Workflow for LC-MS/MS Analysis of Labeled and Native N-Linked Glycans Released From Proteins

A Fully Integrated Workflow for LC-MS/MS Analysis of Labeled and Native N-Linked Glycans Released From Proteins A Fully Integrated Workflow for LC-MS/MS Analysis of Labeled and Native N-Linked Glycans Released From Proteins Udayanath Aich, 1 Julian Saba, 2 Xiaodong Liu, 1 Srinivasa Rao, 1 Yury Agroskin, 1 and Chris

More information

Dienes Derivatization MaxSpec Kit

Dienes Derivatization MaxSpec Kit Dienes Derivatization MaxSpec Kit Item No. 601510 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION

More information

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold Technical Note PPS366 Extraction of 25-OH Vitamin D from Plasma Using ISOLUTE SLE+ Page 1 Comparison of Biotage vs. Sample Processing Using a Vacuum Manifold Extraction of 25-OH Vitamin D from Plasma Using

More information

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT. Objective The aim of the study presented is to determine the presence of target Gingkolides in Gingko biloba leaf extract. The data presented shows Gingkolides determined to be present in the extract analyzed.

More information

Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research

Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research TECHNICAL NOTE 64913 Valium (diazepam) structure Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research Authors Claudio De Nardi,

More information

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis Application Note AN890 Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE SLE+ Page Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE SLE+ Prior to LC/MS-MS

More information

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma Version 1 Last updated 25 May 2018 ab228565 Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma For the isolation/detection of exosomes from human plasma, urine or cell culture media. This product

More information

SCIEX Vitamin D 200M Assay for the Topaz System

SCIEX Vitamin D 200M Assay for the Topaz System The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human

More information

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition Evelyn Goh Waters Pacific, Singapore APPLICATION BENEFITS This method allows for the simultaneous analysis of 12 water-soluble

More information

ab Homocysteine Assay Kit (Fluorometric) 1

ab Homocysteine Assay Kit (Fluorometric) 1 Version 1 Last updated 25 May 2018 ab208559 Homocysteine Assay Kit (Fluorometric) For the measurement of homocysteine in mammalian plasma and serum. This product is for research use only and is not intended

More information

for the Identification of Phosphorylated Peptides

for the Identification of Phosphorylated Peptides Application of a Data Dependent Neutral-Loss Experiment on the Finnigan LTQ for the Identification of Phosphorylated Peptides Gargi Choudhary Diane Cho Thermo Electron, San Jose, CA Abstracted from posters

More information

Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection

Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection David Thomas, Ian Acworth, Bruce Bailey, Marc Plante, and Qi Zhang Thermo Fisher Scientific, Chelmsford, MA Overview

More information

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Application Note Food Safety Authors Chenhao Zhai Agilent

More information

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose:

More information

Sample Concentration and Analysis of Human Hormones in Drinking Water

Sample Concentration and Analysis of Human Hormones in Drinking Water Sample Concentration and Analysis of Human Hormones in Drinking Water Carl Fisher Applications Chemist Ion Chromatography/Sample Preparation Thermo Fisher Scientific March 1, 215 1 The world leader in

More information

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS Olga Shimelis, An Trinh, and Michael Ye Supelco, Div. of Sigma-Aldrich, Bellefonte, PA T407125 Introduction Molecularly

More information

Determination of Carbohydrates in Kombucha Using HPAE-PAD

Determination of Carbohydrates in Kombucha Using HPAE-PAD Determination of Carbohydrates in Kombucha Using HPAE-PAD Beibei Huang, Jingli Hu, and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA USA Application Note 1152 Key Words Glucose, Fructose, Sucrose,

More information

Improving Selectivity in Quantitative Analysis Using MS 3 on a Hybrid Quadrupole-Linear Ion Trap Mass Spectrometer

Improving Selectivity in Quantitative Analysis Using MS 3 on a Hybrid Quadrupole-Linear Ion Trap Mass Spectrometer Improving Selectivity in Quantitative Analysis Using MS 3 on a Hybrid Quadrupole-Linear Ion Trap Mass Spectrometer Overview Increased selectivity is achieved in quantitative analysis by using MS 3. The

More information

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis Application Note AN784 Extraction of Multiple Mycotoxins from Nuts Using ISOLUTE Myco Page 1 Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis This application note

More information

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection Application Note 9 The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection INTRODUCTION The accurate measurement of the amount of sugar in final molasses

More information

Analysis of Pesticides (II) Metribuzin & their metabolites in Rice Jun Yonekubo, Nihon Waters, JAPAN

Analysis of Pesticides (II) Metribuzin & their metabolites in Rice Jun Yonekubo, Nihon Waters, JAPAN Waters Alliance LC/MS System Analytical Conditions Chromatograms Mass Spectra Calibration Quantitation Analysis of Pesticides (II) Metribuzin & their metabolites in Rice Jun Yonekubo,, JAPAN Key Words

More information

Forensic and clinical products and services

Forensic and clinical products and services r o i d s N a t u r a l p r o d u c t s T D M Forensic and clinical products and services R e f e r e n c e s t a n d a r d s F o r e n s i c a n d c l i n i c a l t o x i c o l o g y S t a b l e i s o

More information

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water Richard F. Jack, 1 Jinshui Che, 2 Lipika Basumallick, 1 and Jeffrey Rohrer 1 1 Thermo Fisher Scientific, Sunnyvale, CA, USA;

More information

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with Boron-Doped Diamond Electrode

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with Boron-Doped Diamond Electrode Deteration of Clinically Relevant Compounds using HPLC and Electrochemical Detection with Boron-Doped Diamond Electrode Bruce Bailey, Marc Plante, David Thomas, Qi Zhang and Ian Acworth Thermo Fisher Scientific,

More information

Thermo Scientific ART Pipette Tips

Thermo Scientific ART Pipette Tips LIQUID HANDLING Thermo Scientific ART Pipette Tips Now offering a broad selection of barrier, non-filter and specialty tips to fit virtually any pipette in your lab, Thermo Scientific ART pipette tips

More information

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis Application Note AN782 Extraction of Multiple Mycotoxins from Grain Using ISOLUTE Myco Page 1 Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis This application

More information

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Identification of Steroids in Water by Ion Trap LC/MS/MS Application Identification of Steroids in Water by Ion Trap LC/MS/MS Application Environmental Author Paul Zavitsanos and Christine Miller Agilent Technologies, Inc. Centerville Road Wilmington, DE 9- USA Abstract

More information

Determination of Amantadine Residues in Chicken by LCMS-8040

Determination of Amantadine Residues in Chicken by LCMS-8040 Liquid Chromatography Mass Spectrometry Determination of Amantadine Residues in Chicken by LCMS-8040 A method for the determination of amantadine in chicken was established using Shimadzu Triple Quadrupole

More information

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages APPLICATION NOTE 21671 Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages Authors Jon Bardsley, Thermo Fisher Scientific, Runcorn, UK Keywords Vanquish Flex, Acclaim PolarAdvantage

More information

Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology. Ultra sensitive quantitative performance

Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology. Ultra sensitive quantitative performance Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology Ultra sensitive quantitative performance Unmatched quantitative performance for the most challenging analyses The Agilent 6490 Triple

More information

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c Bio-Rad Laboratories Hemoglobin Testing VARIANT II Hemoglobin Testing System For HbA 1c Bio-Rad CDM System CDM 5.1 VII Instrument Patient Report V2_A1c_NU Patient Data Sample ID: AD Patient ID: Name: Physician:

More information

ab ORAC Assay Kit

ab ORAC Assay Kit Version 1 Last updated 10 April 2018 ab233473 ORAC Assay Kit For the measurement of ORAC activity in cell lysate, plasma, serum, tissue homogenates and food extracts. This product is for research use only

More information

[ Care and Use Manual ]

[ Care and Use Manual ] MALDI Calibration Kit I. Introduction The MALDI Calibration Kit is a conveniently packaged selection of MALDI matrices and calibration standards (includes high-purity Neg Ion Mode Calibrant for accurate

More information

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ Application Note Clinical Research Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ Authors Jennifer Hitchcock 1, Lauren Frick 2, Peter Stone 1, and Vaughn Miller 2 1 Agilent

More information

LC-Based Lipidomics Analysis on QTRAP Instruments

LC-Based Lipidomics Analysis on QTRAP Instruments LC-Based Lipidomics Analysis on QTRAP Instruments Junhua Wang and Paul RS Baker SCIEX LC-Based Lipidomics Analysis Topics Covered Lipid extraction techniques Hydrophilic Interaction Chromatography (HILIC)

More information

Don t miss a thing on your peptide mapping journey How to get full coverage peptide maps using high resolution accurate mass spectrometry

Don t miss a thing on your peptide mapping journey How to get full coverage peptide maps using high resolution accurate mass spectrometry Don t miss a thing on your peptide mapping journey How to get full coverage peptide maps using high resolution accurate mass spectrometry Kai Scheffler, PhD BioPharma Support Expert,LSMS Europe The world

More information

Protect Your Mass Spec

Protect Your Mass Spec Protect Your Mass Spec Sample Preparation Solutions for Biological Samples www.phenomenex.com/bioms Dirty Samples Could be Costing You Money and Sensitivity Biological samples contain a wide variety of

More information

ab Hemoglobin Assay Kit (Colorimetric)

ab Hemoglobin Assay Kit (Colorimetric) Version 1 Last updated 8 May 2018 ab234046 Hemoglobin Assay Kit (Colorimetric) For the measurement of hemoglobin in various biological samples. This product is for research use only and is not intended

More information

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification 22 SEPARATION SCIENCE REDEFINED MAY 2005 Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification In the drug discovery process the detection and

More information

LC-MS/MS Method for the Determination of Raloxifene and its Glucuronide Metabolites from Human Plasma Using SPE Micro Elution

LC-MS/MS Method for the Determination of Raloxifene and its Glucuronide Metabolites from Human Plasma Using SPE Micro Elution LC-MS/MS Method for the Determination of Raloxifene and its Glucuronide Metabolites from Human Plasma Using SPE Micro Elution Krishna Rao Dara, Dr. Tushar N. Mehta, Centre of Excellence for Asia Pacific

More information