Effect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant

Size: px
Start display at page:

Download "Effect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant"

Transcription

1 Effect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant Agung Setyo Darmawan, Waluyo Adi Siswanto, Haryanti Samekto Faculty of Mechanical and Manufacturing Engineering Universiti Tun Hussein Onn Malaysia, Parit Raja - Batu Pahat E-mai :agungsetyod@yahoo.com ABSTRACT Titanium alloys are commonly used in the application of biomechanics parts. One of the applications for hip joint prostheses. When wearing is one parameter that mostly considered, the Titanium alloy is much more superior compared to that from polyethylene derivatives. Although the wear of existing Ti- 6Al-4V Titanium Alloy on Ti-6Al-4V Titanium Alloy hip joint prostheses is much lower than the more widely used polyethylene on metal bearings, there are some concerns about the toxicity of Al and V wear debris in the human body that might negative side effects to the human body. For this reason, the use of pure Titanium is a potential metal and safer to replace Ti-6Al-4V Titanium Alloy bearings. An investigation has been conducted to see the effect of femoral head size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant. The radius of femoral heads are 27 mm, 28 mm, and 30 mm. Numerical analysis based on Finite Element Method is used to simulate and calculate the contact stresses of the bearings. A commercial package CATIA V5 R14 is employed to analyse the contact stresses. For the radius of 27mm, 28mm, and 30 mm the largest contact pressures of each radius are 6.33 x N/m 2, 4.92 x N/m 2, and 2.9 x N/m 2, respectively. These values illustrates that the larger the radius, the smaller the contact stress. Keywords: Hip Joint, Pure Titanium, Contact Stress. 1. Introduction Hip joint is an important component of the human skeletal system. It is located at the junctions between the pelvis and the upper leg bone (femur). The primary disease of the hip is osteoarthritis, and is usually described as the degeneration and progressive loss of normal structure and function of articular cartilage. Osteoarthritis affects 90 % of the human population by the age of 70 years. thus increasing joint friction [1]. Hence, hip joint needs to be replaced. A schematic diagram of the artificial hip joint is presented in Fig. 1a; Fig. 1b shows an X- ray of a total hip replacement. Hip joint replacement is described as the greatest achievement in orthopedic surgery in the twentieth century. The field has been dominated for some forty years by implants based upon metallic femoral heads and stems and polymeric acetabular cups [2]. Other hip disease is rheumatoid arthritis. Osteoarthritis and rheumatoid arthritis result in physical wear of the articular surface, leading to joint paint, restriction of motion and deformity. Under these conditions, damage to the synovial fluid is inevitable, which causes a decrease in viscosity under all possible shear rates,

2 plasma treatment or by appropriate coating films [10-11]. Metal on metal bearings have been shown to result in much lower wear volume [12]. Their higher stiffness allows metal on metal components to be manufactured with larger diameters. The purpose of this study was therefore to investigate effect of femoral head size to contact stresses at outer surface of femoral ball head. 2. Methods Figure 1 (a) Schematic diagram and (b) x-ray of an artificial total hip replacement [3]. The use of polymer for total hip replacement has increase significantly in recent years, mainly as a consequence of their good combination of mechanical and surface properties together with recent improvements in biocompatibility and bioactivity [4]. The wear of polymer cup in hip implants lead to loosening of the implant and osteolysis. Mass loss as a result of wear leading to osteolysis may cause dislocation and subsequently revision and/or replacement of the implant [5]. A significant number of total joint replacements fail long before their expected life causing severe traumas to the patients. Hence, ceramic on ceramic, and metal on metal bearing pairs are an interesting alternative [6-9]. One of the high potential metals on metal is made from Pure Titanium, which is appreciated for its light weight, good biocompatibility, and elastic properties similar to natural bone. Several attempts have been undertaken to improve the surface properties of titanium alloys either by The material used for this work is pure titanium. The properties of the material are young modulus of 1.14 x N/m 2, Poisson ratio of 0.34, density of 4460 kg/m 3, and yield strength of 8.25 x 10 8 N/m 2. Three different diameters of femoral ball head are considered: 54 mm, 56 mm, and 60 mm, respectively. The femoral ball head, stem, and acetabular cup are shown in Figure 2. To determine the stress distribution in the titanium femoral ball head for the 5 kn, the load of 5 kn is applied as a pressure of 1.5 x 10 7 N/m 2 as shown in Figure 3. Figure 2 The components of an artificial total hip replacement.

3 Figure 3. A pressure of 1.5 x 10 7 N/m 2 is applied at black colour part. 3. RESULT AND ANALYSIS Numerical simulation results for various radius obtained by CATIA are shown from Figure 3 to Figure 5. Figure 3. A numerical simulation result of 27 mm radius femoral ball head

4 Figure 4. A numerical simulation result of 28 mm radius femoral ball head Figure 5. A numerical simulation result of 30 mm diameter femoral ball head The largest contact stresses on each radius of the femoral head outer surfaces are listed in table 1. This table demonstrates that the larger femoral ball head may induce larger contact stress. For the radius of 27mm, 28mm, and 30 mm the largest contact pressures of each radius are 6.33 x N/m 2, 4.92 x N/m 2, and 2.9 x N/m 2, respectively. Curve for contact pressure versus radius is shown in Fig. 6. The largest contact stress of the femoral head outer surface is reduced because of the increased area.

5 Table 1. The largest contact stresses on outer surface of femoral ball head No Radius (mm) Largest Contact Stress (N/m 2 ) x x x Radius (mm) Figure 6. Contact stress versus Diameter curve CONCLUSION From this study, it can be concluded that the larger the radius, the smaller the contact stress ACKNOWLEDGEMENT This work was supported by MOSTI under research grant Science fund Vot S015. Thanks to Agus Dwi Anggono, Bana Handaga, and Bambang Waluyo Febriyantoko for the CATIA discussion. REFERENCES [1] Bowsher, J.G., 2001, Accelerated Wear Testing Methodologies for Total Hip Replacements, Ph.D Thesis, Queen Mary University of London [2] Dowson, D., 2001, New Joints for the Millennium: Wear Control in Total Replacement Hip Joints, Proc Institute of Mechanical Engineers Vol. 215 Part II, page [3] Callister, W.D., 2007, Materials Science and Engineering: An Introduction 7ed, John Wiley and Son, New York. [4] Wilches, L.V., Uribe, J.A., Toro, A., 2008, Wear of Materials Used for Artificial Joints in Total Hip replacement, Wear 265, page [5] Alhassan, S., Goswami, T., 2008, Wear Rate Model for UHMWPE in Total Joint Application, Wear 265, page 8-13 [6] Mak, M.M., Jin, Z.M., 2002, Analysis of Contact Mechanics in Ceramic on Ceramic Hip Joint Replacement, Proc Institute of Mechanical Engineers Vol. 216 Part H, page [7] Nevelos, J.E., Ingham, E., Doyle, C., Nevelos, A.B., Fisher, J., 2001, Wear of HIPed and non HIPed Alumina- Alumina Hip Joint Under Standard and Severe Simulator Testing Condition, Biomaterials Vol.22, page [8] Zhou, Y.S., Ikeuchi, K., Ohashi, M., 1997, Comparison of The Friction Properties of Four Ceramic Materials for Joint Replacements, Wear Vol.210, page [9] Jagatia, M. Jin, Z.M., 2001, Elastohydrodynamic Lubrication Analysis of Metal on Metal Hip Prostheses Under Steady State Entraining Motion, Proc Institute of Mechanical Engineers Vol. 215 Part H, page [10] Osterle, W., Klaffke, D., Griepentrog, M., Gross, U., Kranz, I., Knabe, Ch., 2008, Potential of Wear Resistant Coatings on Ti-6Al-4V for Artificial Hip Joint Bearing Surface, Wear Vol.264, page [11] Fisher, J., Hu, X.Q., Tipper, J.L., Steward, T.D., Williams, S., Stone, M.H., Davies, C., Hatto, P., Bolton, J.,

6 Relay, M., Hardaker, C., Isaac, G. H., Berry, G., Ingham, E., 2002, An in vitro Study of the Reduction in Wear of Metal on Metal Hip Prostheses using Surface-engineered Femoral Heads, Proc Institute of Mechanical Engineers Vol. 216 Part H, page [12] Bishop, N.E., Waldow, F., Morlock, M.M., 2008, Friction moments of large metal-on-metal hip joint bearings and other modern designs, Medical Engineering & Physics

NEW DESIGN OF PURE TITANIUM METAL ON METAL TOTAL HIP REPLACEMENT WITH ION IMPLANTED WEAR RESISTANCE SURFACE

NEW DESIGN OF PURE TITANIUM METAL ON METAL TOTAL HIP REPLACEMENT WITH ION IMPLANTED WEAR RESISTANCE SURFACE NEW DESIGN OF PURE TITANIUM METAL ON METAL TOTAL HIP REPLACEMENT WITH ION IMPLANTED WEAR RESISTANCE SURFACE WALUYO AD1 SISWANTO GERAN PENYELIDIKAN SCIENCEFUND NO. VOT. SO15 UNIVERSITI TUN HUSSEIN ONN MALAYSIA

More information

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 6, Issue 12 (May 2013), PP. 119-123 A Mathematical Model to Calculate Contact Stresses

More information

Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements.

Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements. Developing techniques for the dimensional assessment of true clinical wear in retrieved total joint replacements. P. Bills*, L. Blunt*, G. Chakrabarty** *University of Huddersfield, Huddersfield, UK **West

More information

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code Patrick Chaboche, MD patrick.chaboche@wanadoo.fr Former Head of Department of Orthopaedic Surgery. Centre

More information

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head Enhancing stability and increasing range of motion Metasul LDH Large Diameter Head The Metasul large diameter head technology is the result of in-depth research, development and clinical experience that

More information

[ICESTM-2018] ISSN Impact Factor

[ICESTM-2018] ISSN Impact Factor GLOBAL JOURNAL OF ENGINEERING SCIENCE AND RESEARCHES MODELING AND FINITE ELEMENT ANALYSIS OF KNEE JOINT PROSTHESIS U.D.S.Prathap varma *1,S.Rajesh 2, B.Suresh Kumar 3 & P.Rama Murthy Raju 4 *1 M.TechScholar,

More information

Understanding Hip Implant Options

Understanding Hip Implant Options Understanding Hip Implant Options Cup (Socket) Bearing (Liner) Head Stem (Femur) Modern Hip Implants (Ball) Metal Femoral Head on Marathon Cross-linked Polyethylene Liner Proven materials Polyethylene

More information

CERAMIC-ON-METAL. References: Cat No: version 2

CERAMIC-ON-METAL. References: Cat No: version 2 CERAMIC-ON-METAL References: 1. Beaule, P. E., et al. Jumbo femoral head for the treatment of recurrent dislocation following total hip replacement. J.Bone Joint Surg (2002). 84-A: 256-63. 2. Jolles, B.

More information

Development and Analysis of Composite Material for In-Situ Hip Prosthesis

Development and Analysis of Composite Material for In-Situ Hip Prosthesis Development and Analysis of Composite Material for In-Situ Hip Prosthesis R.Gowthaman P.G. Student, Manufacturing Engineering, Department of Mechanical Engineering, Thanthai Periyar Govt. Institute of

More information

CeramTec Medical Products. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion

CeramTec Medical Products. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion CeramTec Medical Products BIOLOX DUO Ceramic Bipolar System Bone-sparing joint reconstruction with a maximum range of motion 2 BIOLOX DUO Ceramic Bipolar System Less is more BIOLOX DUO is an alternative

More information

Medical Products Division. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion

Medical Products Division. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion Medical Products Division BIOLOX DUO Ceramic Bipolar System Bone-sparing joint reconstruction with a maximum range of motion 2 BIOLOX DUO Keramisches Bipolarsystem BIOLOX DUO Ceramic Bipolar System Less

More information

Vasu Pai D orth, MS, National Board [orth],mch, FRACS, FICMR Total Hip Arthroplasty

Vasu Pai D orth, MS, National Board [orth],mch, FRACS, FICMR Total Hip Arthroplasty Vasu Pai D orth, MS, National Board [orth],mch, FRACS, FICMR Total Hip Arthroplasty Introduction Hip arthritis is a common problem, most often due to osteoarthritis. In hip arthritis affects a patient,

More information

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter Design Philosophy Cementless Stems of 12/14 Taper Stem Length Neck shaft Angle STEM-N3 STEM-N4 STEM-N5 STEM-N6 STEM-N7 STEM-N8 STEM-N9 123 mm 125 mm 130 mm 135 mm 140 mm 145 mm 150 mm 132 Cement Polished

More information

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY A PATH FOR HORIZING YOUR INNOVATIVE WORK STUDY AND ANALYSIS OF KNEE IMPLANT IN HUMAN BODY GAJANAN D. MANDAVGADE Assistant

More information

Design of a Hip Joint Simulator According to the ISO 14242

Design of a Hip Joint Simulator According to the ISO 14242 , July 6-8, 2011, London, U.K. Design of a Hip Joint Simulator According to the ISO 14242 Nikolaos I. Galanis, Member, IAENG and Dimitrios E. Manolakos Abstract Wear and simulator testing are complicated

More information

PyroCarbon Implants Durability + Biocompatibility

PyroCarbon Implants Durability + Biocompatibility PyroCarbon Implants Durability + Biocompatibility What is PyroCarbon? PyroCarbon is a specific form of carbon that has been tailored for durability and biocompatibility. PyroCarbon exhibits favorable wear

More information

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear The Journal of Arthroplasty Vol. 00 No. 0 2008 Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction Effects on Polyethylene Wear William L. Jaffe, MD,* Eric J. Strauss,

More information

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).

More information

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo?

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? Journal of Orthopaedic Surgery 2005;13(1):46-51 Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? H Tsumura, N Kaku, T Torisu Department of Orthopedic Surgery, Oita University,

More information

FINITE ELEMENT ANALYSIS OF KNEE IMPLANT AND PROTOTYPE FABRICATION USING RPT PROCESS

FINITE ELEMENT ANALYSIS OF KNEE IMPLANT AND PROTOTYPE FABRICATION USING RPT PROCESS International Journal of Mechanical Engineering and Technology (IJMET) Volume 8, Issue 12, December 2017, pp. 197 205, Article ID: IJMET_08_12_020 Available online at http://www.iaeme.com/ijmet/issues.asp?jtype=ijmet&vtype=8&itype=12

More information

Identify the role of the skeletal system particularly in relation to maintaining an upright stance and protecting vital organs.

Identify the role of the skeletal system particularly in relation to maintaining an upright stance and protecting vital organs. 9.3.3 The wide range of movements, continual absorption of shocks and diseases make the skeletal system vulnerable to damage but new technologies are allowing the replacement of some damaged structures.

More information

Keywords: finite element methods, digital image correlation, deformation, acetabular cup

Keywords: finite element methods, digital image correlation, deformation, acetabular cup Applied Mechanics and Materials Online: 2010-06-30 ISSN: 1662-7482, Vols. 24-25, pp 275-280 doi:10.4028/www.scientific.net/amm.24-25.275 2010 Trans Tech Publications, Switzerland Acetabular Component Deformation

More information

CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician.

CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician. CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician. ENGLISH Mpact 3D Metal Implants and Augments 3D Metal INSTRUCTION FOR USE Important notice: the device(s) can

More information

SIMULATOR FOR TESTING THE FRICTION TORQUE OF THE FEMORAL HEAD OF A HIP PROSTHESIS

SIMULATOR FOR TESTING THE FRICTION TORQUE OF THE FEMORAL HEAD OF A HIP PROSTHESIS Bulletin of the Transilvania University of Braşov Vol. 8 (57) No. 2-2015 Series I: Engineering Sciences SIMULATOR FOR TESTING THE FRICTION TORQUE OF THE FEMORAL HEAD OF A HIP PROSTHESIS S. MIHAI 1 V. FILIP

More information

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY by Timothy McTighe, Exec. Dir., JISRF, Chagrin Falls, Ohio Ying Ko, Ph.D., Cincinnati, Ohio Russell B. Bennett, Ph.D., Cincinnati, Ohio James

More information

What Are Hip and Knee Replacement Implants Made Of?

What Are Hip and Knee Replacement Implants Made Of? What Are Hip and Knee Replacement Implants Made Of? Hip and knee replacement surgery involve replacing the worn-out bone and cartilage lining your hip or knee joint with new implants that are composed

More information

Robotic-Arm Assisted Surgery

Robotic-Arm Assisted Surgery Mako TM Robotic-Arm Assisted Surgery for Total Hip Replacement A Patient s Guide Causes of Your Hip Pain Your joints are involved in almost every activity you do. Movements such as walking, bending and

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 4,000 116,000 120M Open access books available International authors and editors Downloads Our

More information

Anterior Approach to Hip Replacement Surgery

Anterior Approach to Hip Replacement Surgery Anterior Approach to Hip Replacement Surgery Introduction When debilitating pain and stiffness in your hip limits your daily activities, you may need a total hip replacement. The development of total hip

More information

Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements

Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements Daniel Kluess, Sarah Bierbaum, Wolfram Mittelmeier, Rainer Bader,

More information

Index. Gainslen prosthesis 125. Debonding 22-3 Degradation products 40-1 Diffusion theory of adhesive wear 137 Dual energy X-ray absorption 95-6, 98

Index. Gainslen prosthesis 125. Debonding 22-3 Degradation products 40-1 Diffusion theory of adhesive wear 137 Dual energy X-ray absorption 95-6, 98 Index Abrasion 40-1 mechanism of 137 Acetabular components 59-63 advantages 59-60 disadvantages 60-3 Harris Galante Porous cup 59, 60 Porous Coated Anatomic cup 60 Adhesins 155-6 Adhesive wear 137 Alloys

More information

L01:Name and locate the major bones within the skeletal system.

L01:Name and locate the major bones within the skeletal system. L01:Name and locate the major bones within the skeletal system. All physical activity requires movement using bones and muscles. Name three major bones which are located in the leg. Bone 1 - Bone 2- Bone

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

OPTIMIZATION OF SKELETAL HIP IMPLANT CROSS-SECTIONS

OPTIMIZATION OF SKELETAL HIP IMPLANT CROSS-SECTIONS OPTIMIZATION OF SKELETAL HIP IMPLANT CROSS-SECTIONS Sudesh Sivarasu, Pearline Beulah, Lazar Mathew School of Bio Sciences & Technology, VIT University, Vellore sudeshsivarasu@gmail.com Abstract: Aseptic

More information

Furlong H-A.C. THR System

Furlong H-A.C. THR System Important Information Please read prior to use in a clinical setting. The Surgeon should be familiar with the operative technique. Caution Federal (U.S.A) law restricts this device to sale by or on the

More information

Finite Element Study of Acetabular Cup Contact Region for Total Hip Replacement (THR)

Finite Element Study of Acetabular Cup Contact Region for Total Hip Replacement (THR) Journal of Mechanical Engineering Vol SI (), 7-8, 07 Finite Element Study of Acetabular Cup Contact Region for Total Hip Replacement (THR) Muhammad Faris Abd Manap Solehuddin Shuib Faculty of Mechanical

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. Trident Hemispherical Acetabular Shell Trident PSL HA Acetabular Shell Featuring: The Trident Acetabular System

More information

Total Hip Replacement. Find out why the Anterior Approach may be right for you.

Total Hip Replacement. Find out why the Anterior Approach may be right for you. Total Hip Replacement Find out why the Anterior Approach may be right for you. UNDERSTANDING TOTAL HIP REPLACEMENT This brochure offers a brief overview of the Direct Anterior Approach to total hip arthroplasty.

More information

TriboFit Hip System. Issue 3 - August 2013 English

TriboFit Hip System. Issue 3 - August 2013 English Important Information: Please read prior to use in a clinical setting. The surgeon should be familiar with the operative technique and all the information in this insert. Description: The TriboFit Hip

More information

Experimental Prediction of Contact Area in Hip Replacement and Hemi- Arthroplasty

Experimental Prediction of Contact Area in Hip Replacement and Hemi- Arthroplasty Experimental Prediction of Contact Area in Hip Replacement and Hemi- Arthroplasty Qianqian Wang, John Fisher, Sophie Williams. Institute of Medical and Biological Engineering, School of Mechanical Engineering,

More information

PLR. Proximal Loading Revision Hip System

PLR. Proximal Loading Revision Hip System PLR Proximal Loading Revision Hip System The PLR splined revision stem is designed to recreate the natural stresses in the revised femur, where proximal bone may be compromised. PLR Hip System Design Considerations

More information

Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method

Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method Research Article Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method Mitsugu Todo * Research Institute for Applied Mechanics, Kyushu University, Kasuga, Japan

More information

Hard Implants. Lecture #13

Hard Implants. Lecture #13 Hard Implants Lecture #13 Refs: 1. Park, J. B. and Bronzino, J. D., eds., Biomaterials: Principles and Applications. CRC Press, Boca Raton, FL, 2003. 2. Park, J. B., Biomaterials Science and Engineering.

More information

Metasul LDH Large Diameter Head

Metasul LDH Large Diameter Head Metasul LDH Large Diameter Head Surgical Technique Metasul LDH Large Diameter Head Surgical Technique Enhancing Stability and Increasing Range of Motion Metasul LDH Large Diameter Head Surgical Technique

More information

Total Hip Replacement Surgery

Total Hip Replacement Surgery Total Hip Replacement Surgery On this page: Overview What is Total hip replacement (THA)? Anatomy Preparing for surgery The Operation Post-Operative Possible Complications Exercise Restrictions Overview

More information

Surface evaluation of orthopedic hip implants marketed in Brazil

Surface evaluation of orthopedic hip implants marketed in Brazil Surface evaluation of orthopedic hip implants marketed in Brazil M.M.Souza 1, R.M. Trommer 2, M.M.Maru 2, C.R.M.Roesler 3, W.S.Barros 1, M.S. Dutra 3 1 Dimensional Metrology Laboratory, Inmetro. Brazil;

More information

The Investigation into Modes of Failure of Total Hip Replacements

The Investigation into Modes of Failure of Total Hip Replacements University of Southern Queensland Faculty of Engineering and Surveying The Investigation into Modes of Failure of Total Hip Replacements A dissertation submitted by Miss Vivienne Joyce French In fulfilment

More information

Curriculum Vitae. Zhong Min Jin

Curriculum Vitae. Zhong Min Jin Curriculum Vitae Zhong Min Jin School of Mechanical Engineering, University of Leeds, Leeds, LS2 9JT, England. (Tel: 44 (0)113 3432154. Fax: 0113 (0)113 2424611. Email: Z.Jin@leeds.ac.uk) Sex: Male Birth

More information

LING AND ANALYSIS PROSTHESIS WITH MODULAR STEM

LING AND ANALYSIS PROSTHESIS WITH MODULAR STEM International Journal of Mechanical Engineering and Technology (IJMET) Volume 8, Issue 6, June 2017, pp. 789 795, Article ID: IJMET_08_06_082 Available online at http://www.ia aeme.com/ijm MET/issues.as

More information

Biomechanics of. Knee Replacement. Mujda Hakime, Paul Malcolm

Biomechanics of. Knee Replacement. Mujda Hakime, Paul Malcolm Biomechanics of Knee Replacement Mujda Hakime, Paul Malcolm 1 Table of contents Knee Anatomy Movements of the Knee Knee conditions leading to knee replacement Materials Alignment and Joint Loading Knee

More information

PMMA Spacer. X-ray Marker

PMMA Spacer. X-ray Marker Full TM Cup 2.5 mm cement thickness PMMA Spacer Uniform 2.5 mm cement thickness; the cement mantle is built within the outer dimension of acetabular cup. Cup O.D. 20 o lipped Prevent joint dislocation

More information

Peggers Super Summaries: THR and Revision Concepts

Peggers Super Summaries: THR and Revision Concepts Total Hip replacement: ARTICUALTIONS Ceramic on polyethylene Ceramic on ceramic for young Metal on metal (resurfacing) Metal (cobalt chrome) on polyethylene WEAR Wear debris similar to bacteria o < 10

More information

Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report

Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report Journal of Long-Term Effects of Medical Implants, 18(1): 69 74 (2008) Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report W. Siebert, 1 S. Mai, 1 & S. Kurtz 2 1 Orthopädische Klinik

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. Trident Poly Acetabular System Trident PSL HA Acetabular Shell with X3 Poly insert Trident Hemispherical Acetabular

More information

Zimmer Segmental System

Zimmer Segmental System Zimmer Segmental System Simple solutions for solving complex salvage cases A Step Forward The Zimmer Segmental System is designed to address patients with severe bone loss associated with disease, trauma

More information

FINITE ELEMENT ANALYSIS : AN EFFECTIVE TOOL FOR PROSTHESES DESIGN

FINITE ELEMENT ANALYSIS : AN EFFECTIVE TOOL FOR PROSTHESES DESIGN Trends Biomater. Artif. Organs. Vol. 17(2) pp 141-148 (2004) http//www.sbaoi.org FINITE ELEMENT ANALYSIS : AN EFFECTIVE TOOL FOR PROSTHESES DESIGN S K Senapati & S.Pal School of Biosc, & Engg, Jadavpur

More information

CAUTION: Ceramic liners are not approved for use in the United States.

CAUTION: Ceramic liners are not approved for use in the United States. Total Hip Prostheses, Self-Centering Hip Prostheses and Hemi-Hip Prostheses IMPORTANT: This essential product information sheet does not include all of the information necessary for selection and use of

More information

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS INTRODUCTION The Summit Tapered Hip System s comprehensive set of implants and instruments

More information

6 th International Conference on Trends in Agricultural Engineering 7-9 September 2016, Prague, Czech Republic

6 th International Conference on Trends in Agricultural Engineering 7-9 September 2016, Prague, Czech Republic MEASUREMENT OF PRESSURE DISTRIBUTION IN KNEE JOINT REPLACEMENT J. Volf 1, V. Novák 1, V. Ryzhenko 1, D. Novák 2 1 Faculty of Engineering, Czech University of Life Sciences Prague, Czech Republic 2 Faculty

More information

DEPUY SYNTHES JOINT RECONSTRUCTION PATIENT EDUCATION SEMINAR

DEPUY SYNTHES JOINT RECONSTRUCTION PATIENT EDUCATION SEMINAR DEPUY SYNTHES JOINT RECONSTRUCTION PATIENT EDUCATION SEMINAR Welcome! Thank you for joining us! We hope today s information will help you on a path to reducing your pain and gaining mobility. This informational

More information

Catherine Bladen, John Fisher, Eileen Ingham, Joanne Tipper University of Leeds, UK

Catherine Bladen, John Fisher, Eileen Ingham, Joanne Tipper University of Leeds, UK Institute of Medical & Biological Engineering Reduced inflammatory potential of Antioxidant Ultra High Molecular Weight Polyethylene containing Vitamin E May Lead to Improved Longevity in Total Joint Replacement

More information

PRODUCT RATIONALE & SURGICAL TECHNIQUE

PRODUCT RATIONALE & SURGICAL TECHNIQUE This publication is not intended for distribution in the USA. PRODUCT RATIONALE & SURGICAL TECHNIQUE THE PRODUCT OF LONG-TERM CLINICAL EXPERIENCE The TRILOC cemented UHMWPE cup is a direct descendant of

More information

Impingement in THA. Georgi I. Wassilew. Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C.

Impingement in THA. Georgi I. Wassilew. Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C. Georgi I. Wassilew Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C. Perka Charité - Universitätsmedizin Berlin Orthoload club, Berlin Reasons for revision in THA Reason

More information

RM Philosophy Innovation based on clinical evidence.

RM Philosophy Innovation based on clinical evidence. RM Philosophy Innovation based on clinical evidence RM Philosophy Innovation based on clinical evidence Design 1973 RM Classic Cup First uncoated RM Classic Cup out of Polyacetal was inserted by E. Morscher

More information

Biomechanical Analysis of Different Knee Prosthesis Biomaterials Using Fem

Biomechanical Analysis of Different Knee Prosthesis Biomaterials Using Fem IOSR Journal of Mechanical and Civil Engineering (IOSR-JMCE) e-issn: 2278-1684,p-ISSN: 2320-334X, Volume 11, Issue 3 Ver. IV (May- Jun. 2014), PP 120-128 Biomechanical Analysis of Different Knee Prosthesis

More information

CEC ARTICLE: Special Medical Conditions Part 3: Hip and Knee Replacement C. Eggers

CEC ARTICLE: Special Medical Conditions Part 3: Hip and Knee Replacement C. Eggers CEC ARTICLE: Special Medical Conditions Part 3: Hip and Knee Replacement C. Eggers Joint replacement surgery removes a damaged joint and replaces it with a prosthesis or artificial joint. The purpose of

More information

TOTAL HIP REPLACEMENT:

TOTAL HIP REPLACEMENT: THR Prosthesis Design TOTAL HIP REPLACEMENT: PROSTHESIS DESIGN FEATURES JESS JOHNSTON & MELINDA ZIETH History of Hip Prosthesis Joint Replacement Registry Implant Design Technology & Future History and

More information

Improve your quality of life with AMIS

Improve your quality of life with AMIS Improve your quality of life with AMIS IMPROVE YOUR QUALI The AMIS (Anterior Minimally Invasive Surgery) approach is a true intermuscular and internervous minimally invasive surgical technique. In fact,

More information

Ceramic Femoral Heads and Acetabular Cup Liners

Ceramic Femoral Heads and Acetabular Cup Liners Important Information: Please read prior to use in a clinical setting. The Surgeon should be familiar with the operative technique. Caution: Federal (U.S.A) law restricts this device to sale by or on the

More information

Total hip replacement

Total hip replacement Total hip replacement 2 Hip pain Arthritis is the leading cause of disability in the United States, and the most frequent cause of discomfort and chronic hip pain. In fact, it s estimated that 1 in 5 people

More information

Stability without compromise. Epsilon Durasul Constrained Insert

Stability without compromise. Epsilon Durasul Constrained Insert Stability without compromise Epsilon Durasul Constrained Insert History of Dislocation and Constrained Acetabular Inserts The Need for a Better Solution The Problem Dislocation is the second most common

More information

Strive to improve your quality of life with AMIS

Strive to improve your quality of life with AMIS Strive to improve your quality of life with AMIS STRIVE TO IMPROVE YOUR Q The AMIS (Anterior Minimally Invasive Surgery) approach is a true intermuscular and internervous minimally invasive surgical technique.

More information

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574)

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574) References 1. Sanchez-Sotelo J, Haidukewych GJ, Boberg CJ. Hospital Cost of Dislocation After Primary Total Hip Arthroplasty. J Bone Joint Surg. 2006;88A:290-294. 2. Brodner W, Grübl A, Jankovsky R, Meisinger

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

BIOLOX delta Option Ceramic Femoral Head System. Product Features and Instructions for Use

BIOLOX delta Option Ceramic Femoral Head System. Product Features and Instructions for Use BIOLOX delta Option Ceramic Femoral Head System Product Features and Instructions for Use Product Features BIOLOX delta ceramic heads are composed of 75% aluminum oxide and approximately 25% zirconia.

More information

Strive to improve your quality of life with AMIS

Strive to improve your quality of life with AMIS A IS Strive to improve your quality of life with AMIS STRIVE TO IMPROVE YOUR Q The AMIS (Anterior Minimally Invasive Surgery) approach is a true intermuscular and internervous minimally invasive surgical

More information

ISSN: [NCIME: October 2016] Impact Factor: 4.116

ISSN: [NCIME: October 2016] Impact Factor: 4.116 IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY STRUCTURAL ANALYSIS & FABRICATION OF FEMORAL STEM COMPONENT OF A NEWLY DESIGNED CEMENTED HIP IMPLANT THROUGH FUSED DEPOSITION

More information

Your Pre-operative Guide

Your Pre-operative Guide MAKOplasty Your Pre-operative Guide please visit makoplasty.com or call 877.411.MAKO (6256) 2555 Davie Road Fort Lauderdale, FL 33317 866.647.6256 makosurgical.com 2014 MAKO Surgical Corp. 204502 r01 05/14

More information

Total Hip Replacement in Diaphyseal Aclasis: A Case Report

Total Hip Replacement in Diaphyseal Aclasis: A Case Report ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 6 Number 1 Total Hip Replacement in Diaphyseal Aclasis: A Case Report V Singh, S Carter Citation V Singh, S Carter.. The Internet Journal of

More information

ACETABULAR CUP SURGICAL TECHNIQUE

ACETABULAR CUP SURGICAL TECHNIQUE ACETABULAR CUP SURGICAL TECHNIQUE ACETABULAR CUP DEVICE INDICATIONS FOR USE The ICONACY I-Hip total hip replacement is indicated for the following conditions: 1. A severely painful and/or disabled hip

More information

UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD

UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS 66th Annual Meeting February 4-9, 1999 Anaheim, California COMMITTEE ON BIOMEDICAL ENGINEERING COMMITTEE ON

More information

Bone Bangalore

Bone Bangalore Dr Suresh Annamalai MBBS, MRCS(Edn), FRCS( Tr & Orth)(Edn), FEBOT(European Board), Young Hip and Knee Fellowship(Harrogate, UK) Consultant Arthroplasty and Arthroscopic Surgeon Manipal Hospital, Whitefield,

More information

Total Hip Replacement. Find out why Total Hip Replacement may be right for you.

Total Hip Replacement. Find out why Total Hip Replacement may be right for you. Total Hip Replacement Find out why Total Hip Replacement may be right for you. UNDERSTANDING TOTAL HIP REPLACEMENT This brochure offers a brief overview of hip anatomy, arthritis and total hip arthroplasty.

More information

Are You Living with. Hip Pain? MAKOplasty may be the right treatment option for you.

Are You Living with. Hip Pain? MAKOplasty may be the right treatment option for you. Are You Living with Hip Pain? MAKOplasty may be the right treatment option for you. Hip pain shouldn t keep you from doing the things you love. Understanding Common Causes of Hip Pain If you are one of

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 EMPIRICAL TESTING CORP. 4628 Northpark Drive Colorado Springs, CO 80918 Dawn Lissy Phone: 719 264 9937 dlissy@empiricaltesting.com MECHANICAL Valid To: January

More information

Massachusetts Institute of Technology Harvard Medical School Brigham and Women s Hospital VA Boston Healthcare System 2.782J/3.961J/BEH.

Massachusetts Institute of Technology Harvard Medical School Brigham and Women s Hospital VA Boston Healthcare System 2.782J/3.961J/BEH. Massachusetts Institute of Technology Harvard Medical School Brigham and Women s Hospital VA Boston Healthcare System 2.782J/3.961J/BEH.451J/HST524J WEAR AND CORROSION M. Spector, Ph.D. WEAR PROCESSES

More information

TOTAL HIP REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE HIP

TOTAL HIP REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE HIP TOTAL HIP REPLACEMENT: MODERN SURGERY FOR SEVERE ARTHRITIS OF THE HIP By John T. Dearborn, M.D. Please read this pamphlet before your visit so that we can answer any questions that you have during our

More information

YOUR TOTAL HIP REPLACEMENT SURGERY STEPS TO RETURNING TO A LIFESTYLE YOU DESERVE

YOUR TOTAL HIP REPLACEMENT SURGERY STEPS TO RETURNING TO A LIFESTYLE YOU DESERVE YOUR TOTAL HIP REPLACEMENT SURGERY STEPS TO RETURNING TO A LIFESTYLE YOU DESERVE IMPORTANT. PLEASE NOTE. This brochure offers a brief overview of hip anatomy, arthritis and hip replacement surgery. The

More information

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital)

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) 1. Introduction Total hip replacement arthroplasty (THRA) using a metal-on-polyethylene articulation has become one of the most

More information

The Leader in Orthopaedic Innovation

The Leader in Orthopaedic Innovation The Leader in Orthopaedic Innovation Wright is a leading international manufacturer and distributor of superior, easy to use, and innovative orthopaedic implants and instrumentation. For over 50 years,

More information

Shoulder Joint Replacement

Shoulder Joint Replacement Shoulder Joint Replacement Although shoulder joint replacement is less common than knee or hip replacement, it is just as successful in relieving joint pain. Shoulder replacement surgery was first performed

More information

Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc

Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc Mechanics of Contact and Lubrication, MTM G230 Department of Mechanical & Industrial Enineering Northeastern University Spring 2006 Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc

More information

Document No Rev. B Pg. 1 of 5 Approvals / Date Title: INSTRUCTIONS FOR USE DJO SURGICAL MODULAR REVISION HIP SYSTEM AND ACETABULAR CAGE

Document No Rev. B Pg. 1 of 5 Approvals / Date Title: INSTRUCTIONS FOR USE DJO SURGICAL MODULAR REVISION HIP SYSTEM AND ACETABULAR CAGE Document No. 0400-0205 Rev. B Pg. 1 of 5 Approvals / Date Title: INSTRUCTIONS FOR USE DJO SURGICAL MODULAR REVISION HIP SYSTEM AND ACETABULAR CAGE Revision ECO Date ECO Summary of Changes Deann Rector

More information

Progeny Hip Stem. Surgical Protocol and Product Specifications

Progeny Hip Stem. Surgical Protocol and Product Specifications Progeny Hip Stem Surgical Protocol and Product Specifications Progeny Hip Stem Introduction With emphasis on maximum stability and ease of use, the StelKast ProgenyTM Hip System provides the surgeon with

More information

Adult Hip Dysplasia David S. Feldman, MD

Adult Hip Dysplasia David S. Feldman, MD Adult Hip Dysplasia David S. Feldman, MD Chief of Pediatric Orthopedic Surgery Professor of Orthopedic Surgery & Pediatrics NYU Langone Medical Center & NYU Hospital for Joint Diseases Overview Adult hip

More information

YOUR TOTAL HIP REPLACEMENT SURGERY

YOUR TOTAL HIP REPLACEMENT SURGERY YOUR TOTAL HIP REPLACEMENT SURGERY STEPS TO RETURNING TO A LIFESTYLE YOU DESERVE Exactech_030H Rev A_Total Hip Replacement Surgery_PRINT.indd 1 IMPORTANT. PLEASE NOTE. This brochure offers a brief overview

More information

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC Zimmer Longevity Highly Cross-linked Polyethylene Clinical Value Dossier - APAC Table of Contents 1 Executive Summary...................................................... 1 2 Burden of Illness........................................................

More information

Surface Roughness of Femoral Head Prostheses After Dislocation

Surface Roughness of Femoral Head Prostheses After Dislocation An Original Study Surface Roughness of Femoral Head Prostheses After Dislocation Kenny T. Mai, MD, Christopher Verioti, DO, Darryl D Lima, MD, PhD, Clifford W. Colwell, Jr., MD, and Kace A. Ezzet, MD Abstract

More information

1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe

1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe 1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe Abstract Introduction: Osteolysis generated by wear debris remains a problem

More information