Pseudoenhancement of Renal Cysts: Influence of Lesion Size, Lesion Location, Slice Thickness, and Number of MDCT Detectors

Size: px
Start display at page:

Download "Pseudoenhancement of Renal Cysts: Influence of Lesion Size, Lesion Location, Slice Thickness, and Number of MDCT Detectors"

Transcription

1 Genitourinary Imaging Original Research Tappouni et al. MDCT of Renal Cysts Genitourinary Imaging Original Research Rafel Tappouni 1 Jennifer Kissane 2 Nabeel Sarwani 1 Erik B. Lehman 1 Tappouni R, Kissane J, Sarwani N, Lehman EB Keywords: CT, pseudoenhancement, renal cyst DOI: /AJR Received November 3, 2010; accepted after revision May 20, Department of Radiology, Penn State Hershey Medical Center, 500 University Dr, Hershey, PA Address correspondence to R. Tappouni (raftappouni@gmail.com). 2 Penn State University College of Medicine, Hershey, PA. AJR 2012; 198: X/12/ American Roentgen Ray Society Pseudoenhancement of Renal Cysts: Influence of Lesion Size, Lesion Location, Slice Thickness, and Number of MDCT Detectors OBJECTIVE. The purpose of this study was to determine the effect of renal cyst size and location on pseudoenhancement in human subjects. MATERIALS AND METHODS. Simple renal cysts obtained with 16-, 40-, and 64- MDCT scanners were analyzed for the presence of pseudoenhancement. Cyst size, location, and attenuation in the unenhanced and nephrographic phases were recorded. Pseudoenhancement was defined as an attenuation increase of 10 HU or more on nephrographic phase compared with unenhanced images. RESULTS. The pseudoenhancement rate was 22% (51/233). There was a statistically significant increase in the pseudoenhancement rate of lesions smaller than 10 mm (38/233) compared with those 10 mm and larger (13/233) (odds ratio, 6.4; p < ). Twelve of 62 cysts measuring mm exhibited pseudoenhancement. There was a statistically significant increase in the pseudoenhancement rate of central (39/53) compared with peripheral (12/51) cysts (odds ratio, 2.7; p < ). The pseudoenhancement rates for the 16-, 40-, and 64-MDCT scanners were 20%, 19%, and 26% with no statistically significant difference between them. CONCLUSION. Pseudoenhancement of renal cysts significantly correlates with size smaller than 1 cm and central location. Although pseudoenhancement increases with larger numbers of detectors, the correlation was not statistically significant. Cysts in the 1- to 1.5-cm range have a 19% likelihood of pseudoenhancement. R enal cysts are common in adults, and the prevalence increases with age to more than 27% in patients older than 50 years [1, 2]. When a cystic lesion meets all the criteria for being a simple renal cyst, no further evaluation or intervention is required. A crucial feature in characterizing cystic renal masses at CT is the presence of contrast enhancement. Simple renal cysts have a characteristic appearance of a spheric, sharply demarcated lesion with a smooth border on CT images. Furthermore, because they are fluid filled, cysts have attenuation similar to that of water (0 20 HU) and do not greatly enhance after IV contrast administration [3, 4]. Although simple cysts are not enhancing, they may have an increase in attenuation after contrast administration because of technical factors [3, 5, 6]. A less than 10 HU increase in attenuation is considered within the technical limits of the study and is not considered to represent enhancement [3, 7, 8]. Renal cyst pseudoenhancement is an artifactual increase in the attenuation of a simple renal cyst after contrast administration, even when the effects of partial volume averaging have been removed [5, 7, 9]. Technical and anatomic variations such as the part of the cyst measured, location of the cyst within the kidney, and adjacent renal or extrarenal structures can affect the attenuation measured [6, 10, 11]. In a retrospective study of the cases of 24 patients [7], attenuation in the renal cysts increased artifactually on contrast-enhanced images, but this pseudoenhancement was not substantial when the cyst was larger than 1.0 cm in diameter. Various authors [9, 12 17] have used phantom model studies to confirm the increased incidence of renal cyst pseudoenhancement with smaller cyst size (< 1.5 cm) and central location. These studies also showed that pseudoenhancement is increasingly a problem with higher numbers of CT detectors and higher voltage. Because of the AJR:198, January

2 Tappouni et al. TABLE 1: MDCT Parameters Scanner Peak Voltage (kvp) Tube Current (mas) Slice Thickness (mm) Collimation (mm) Rotation Time (s) Sensation 16-MDCT (Siemens Healthcare) (varied) Emotion 16-MDCT (Siemens Healthcare) (varied) Sensation 40-MDCT (Siemens Healthcare) (varied) Definition 64-MDCT (Siemens Healthcare) (varied) increasing use of 16-, 40- and 64-MDCT scanners in clinical practice, confirmation of these findings and improved understanding of the means of reducing pseudoenhancement are needed. Although it has been found that size and location of a lesion, peak tube voltage settings, and number of CT detectors affect image contrast and attenuation in phantom studies, clinical studies are lacking. We undertook this study to determine the effect of lesion size and location on renal cyst pseudoenhancement. Materials and Methods Patients The institutional review board approved our study with waiver of informed consent, and the study was compliant with HIPAA. To evaluate the effect of CT detector number, we analyzed data from the first quarter of 2006 and the last quarter of Search of our radiology information system yielded 1321 abdominal CT examinations, of which 597 were triphasic renal mass CT and were included in the study. Triphasic CT scans of renal masses obtained during this period were reviewed for presence of cysts. A cyst was defined as a wellcircumscribed, uniformly low-attenuation spheric lesion without a discernable wall and content with attenuation between 10 and 20 HU in the unenhanced phase. Only scans that had tube current standardized across all phases were included. CT studies of two patients with polycystic kidney disease were excluded because of inability to differentiate individual cysts within the kidneys. Cysts smaller than 6 mm and larger than 5 cm were excluded. CT Technique All CT examinations were performed with 16-, 40-, and 64-MDCT scanners. The unenhanced, arterial, and nephrographic phases were used as part of the triphasic renal mass protocol. The same parameters were used for all phases, as was z-axis current tube modulation. The scans were reconstructed in slice thickness. The matrix was , and all images were viewed in an abdominal window (width, 400 HU; level, 40 HU). The contrast-enhanced images were obtained after IV administration of 100 ml of iohexol (Omnipaque 300, GE Healthcare) injected at 3.5 ml/s and scan delays of 30 and 100 seconds for the arterial and nephrographic phases. All scans were obtained with an abdominal reconstruction kernel and z- axis tube current modulation. The parameters are shown in Table 1. A Image Interpretation One author with 5 years of experience in genitourinary CT and a third-year medical student reviewed the unenhanced and contrast-enhanced images for the presence of renal cysts. Identified cysts with attenuation between 10 and 20 HU were measured at the approximate center of the lesion, and the diameter of the lesion in millimeters was noted. Corresponding circular regionof-interest measurements were obtained for each cyst. For each cyst, the region of interest was a minimum of 66% of the circumference of the lesion (Fig. 1). Attenuation in the unenhanced and nephrographic phases was recorded in similar regions of interest. Cysts were identified in the nephrographic phase. Manual narrowing of the image window, magnification, and lesion localization were used in some cases. Cysts not visualized on the unenhanced scan were excluded. The location of the lesion was assessed by scanning through the entire lesion. Central cysts were defined as a cyst completely surrounded by renal parenchyma. Peripheral cysts were defined as those in which a portion of the cyst caused bulging of the renal capsule or was exophytic (Fig. 1). Other data collected for each lesion included CT scanner model and number of detectors, slice thickness, and age and sex of the patient. Cysts were defined as having displayed pseudoenhancement if they met one of the following criteria: attenuation 20 HU or less in the unenhanced phase (simple cyst) and enhancement of 10 HU or greater in the nephrographic phase. Cysts with 11- to 20-HU attenuation on unenhanced scans remained stable. Stability of a cyst was defined as a 1-mm or less change in diameter at comparison CT performed 1 year or more before or after the CT date. Statistical Analysis All analyses were performed with statistical software (SAS version 9.2, SAS Institute). Descriptive statistics were computed for all variables with frequencies and percentages for categoric variables and means and SD for continuous variables. The relation between lesion size and pseudoenhancement was analyzed by two approaches. The first approach applied a repeated measures analysis with a linear mixed-effects model to compare the mean lesion sizes in the groups with and without Fig. 1 CT attenuation measurement in two patients. Regions of interest are minimum 66% of circumference of lesion for each cyst. A, 63-year-old man with renal cyst abutting renal capsule and therefore considered peripheral. B, 41-year-old man with renal cysts entire circumference of which is surrounded by renal parenchyma and is defined as central. B 134 AJR:198, January 2012

3 MDCT of Renal Cysts TABLE 2: Cysts by Size Category Total No. Percentage of Pseudoenhancing Percentage of All Size Category of Cysts No. of Cysts With Pseudoenhancement Cysts in Size Category Pseudoenhancing Cysts (n = 51) < 10 mm mm TABLE 3: Breakdown of Cysts by Scanner Type (n = 233) MDCT Scanner Model Total No. of Cysts pseudoenhancement. Repeated measures analysis also was used to determine significant differences in degree of pseudoenhancement between lesion size recoded as category and between lesion locations. Tukey adjustment for multiple comparisons was used when means in more than two groups were being compared to maintain a family-wise error rate of Box plots were used to visualize the differences in means and distributions. The second approach was a generalized estimating equations model, which is an extension of logistic regression for repeated measures with a binary outcome to examine the association between lesion size recoded as category and rate of pseudoenhancement. The same approach was used to analyze the association between rate of pseudoenhancement and other variables such as lesion location, slice thickness, and scanner model. Odds ratios were used to quantify the magnitude and direction of any significant associations. Results Effect of Variables on Pseudoenhancement Rate From the 597 triphasic renal mass CT examinations, a total 233 cysts were isolated. The incidence was 21% for cysts in the 6- to 50-mm range, 128 of 597 triphasic CT scans viewed showing cysts, some multiple. Of the 233 cysts included in the study, 51 cysts exhibited pseudoenhancement, for an overall 22% rate of pseudoenhancement. The mean lesion size was significantly (p < 0.01) lower in the group with pseudoenhancement (mean, 8.3 mm) than in the group without pseudoenhancement (mean, 16.1 mm). None of the cysts exhibited hypervascular enhancement in the arterial phase, defined as enhancement equal or greater than that of the renal tissue. Renal cyst size ranged from 6 to 48 mm in diameter (mean, 14.4 [SD, 9.2] mm). The size categories evaluated were smaller than No. of Cysts with Pseudoenhancement 10 mm and 10 mm in diameter or larger. Table 2 shows the breakdown of cyst size category and pseudoenhancement. The 10 mm and larger category was further divided into mm (62 cysts), mm (25 cysts), and mm (50 cysts). Twelve of the 10- to 14-mm cysts (19%) were pseudoenhancing, constituting 24% of the study total that was pseudoenhancing; 0.04% of the 15- to 19-mm cysts were pseudoenhancing, comprising only 0.02% of the study total that was pseudoenhancing. None of the 20- to 49- mm cysts was pseudoenhancing. Only one cyst 1.5 cm or larger was pseudoenhancing compared with 50 cysts smaller than 1.5 cm. Therefore, a cyst smaller than 1.5 cm was 30.4 times (95% CI, ) as likely as a cyst 1.5 cm or larger to be pseudoenhancing. A significant association was found between pseudoenhancement and lesion size (p < 0.001). A lesion smaller than 10 mm was likelier than a lesion 10 mm or larger to be pseudoenhancing (odds ratio, 6.4; 95% CI, ). There were 109 peripheral cysts. In 12 of 109 cysts (11%) pseudoenhancement was seen and accounted for 24% of all pseudoenhancement cases. There were 124 central cysts defined as a cyst completely surrounded by the renal parenchyma. In 39 of 124 cysts (32%) pseudoenhancement was seen and accounted for 76% of all pseudoenhancement cases. There was a statistically significant association between pseudoenhancement and peripheral or central group (p < ). Central cysts were likelier than peripheral cysts to be pseudoenhancing (odds ratio, 2.7; 95% CI, ). Table 3 shows the breakdown of cysts by scanner type. Although there was more pseudoenhancement with the 64-MDCT scanner, there was no statistically significant difference Percentage of Cysts Showing Pseudoenhancement with Scanner Model Percentage of All Pseudoenhancing Cysts (n = 51) 16 Sensation or Emotion (Siemens Healthcare) Sensation (Siemens Healthcare) Definition (Siemens Healthcare) in the rate of pseudoenhancement between the detector number categories (p = 0.74). Degree of Pseudoenhancement Versus Size and Location of Cysts Degree of pseudoenhancement varied between 10 and 42 HU. Table 4 shows the breakdown of cysts by degree of enhancement. Fifty of 51 of the pseudoenhancing cysts (98%) were in the HU attenuation category. The one cyst with HU attenuation was stable at 1- and 2-year followup evaluations. The largest cyst that exhibited pseudoenhancement is shown in Figure 2. The cyst was stable at follow-up evaluation, although when measured on subsequent scans in the same phases, the cyst enhancement decreased from 10 to 6 HU. Figure 3 shows the variations in degree of pseudoenhancement by size criteria. When cyst size was divided into smaller than 1 cm and 1 cm and larger, there was no statistically significant difference between the groups (mean pseudoenhancement, 17 and 12 HU; p = 0.22). The mean degree of pseudoenhancement was 13 HU in the peripheral cysts (n = 12) and 16 HU in the central cysts (n = 39). There was no was no statistically significant difference between the two groups with respect to degree of pseudoenhancement (p = 0.194). Discussion Benign cyst is the most common focal renal lesion seen at abdominal CT. It occurs in as much as 40% of the general population [18]. Although not nearly as common, renal cell carcinoma has a mortality of 25%, making it imperative to accurately differentiate benign cyst from carcinoma when a renal lesion is identified. Since 1986, the accepted method of classifying renal lesions has been the Bosniak system, in which a renal cyst is AJR:198, January

4 Tappouni et al. TABLE 4: Degree of Enhancement No. of Cysts by Location Degree of Pseudoenhancement (HU) Total No. of Cysts (n = 51) Average Size (mm) Peripheral Central NA NA NA Note NA = not applicable. A B Fig year-old man with renal cysts. A, Transverse unenhanced CT scan shows 1.5-cm area of low attenuation (0.41 HU) in lower pole of right kidney. B, Transverse contrast-enhanced nephrographic phase CT scan shows cyst measures HU. C, Follow-up CT scan obtained 18 months after B shows cyst is stable. defined as a lesion with fluid attenuation (0 20 HU) and enhancement of less than 10 HU with administration of contrast material [3]. Enhancement of at least 10 HU after contrast administration is the currently accepted minimum increase indicative of a vascular, and therefore potentially malignant, mass [7, 8, 19, 20]. Most renal cell carcinomas are the clear cell type, which is a heterogeneous, highly vascular lesion with marked enhancement after contrast administration. The next most common renal cell carcinoma is papillary, which is distinct from the clear cell type in that it is closer to homogeneous and less vascular. The attenuation values for papillary renal cell carcinoma on unenhanced images are significantly higher than fluid attenuation. The mean attenuation of papillary renal cell carcinomas on unenhanced CT scans has been found to range from 37 to 38 HU, making it unlikely to be confused with a pseudoenhancing benign simple cyst [21, 22]. A number of technical variables can affect the ability to rely on absolute attenuation values of any given tissue: CT scanner type, peak voltage, scanning time, and positional effects [23, 24]. In addition, studies with various phantom models [9, 12 17] have shown an increased incidence of renal cyst pseudoenhancement with smaller cyst size (< 1.5 cm) and intraparenchymal location. To our knowledge, however, no patient data exist on the influence of these variables on pseudoenhancement. The results of our study confirm the increased incidence of pseudoenhancement in lesions of smaller size. Cysts smaller than 10 mm in diameter were six times as likely as those 10 mm in diameter or larger to be pseudoenhancing. Furthermore, only one cyst in the 1.5- to 1.9- cm group was pseudoenhancing, and none of the cysts 2 cm or larger was pseudoenhancing. Our data also showed that central cysts were 2.7 times as likely as peripheral cysts to be pseudoenhancing. Interestingly, contrary to previous findings [12 14, 17], there was not a statistically significant difference in incidence of pseudoenhancement between CT scanner types or slice thicknesses. All lesions measured in our study were at least twice the minimal slice thickness of the CT image, to prevent undersampling. Although the types of scanners were not equally represented in our study, each CT scanner type yielded a remarkably similar percentage of pseudoenhancement (19 26%). The limitations of our study were a lack of histopathologic confirmation of the cysts imaged and a disproportionate number of cases among scanner types. Future variables to be studied for their effect on the incidence of pseudoenhancement are degree of background enhancement and alterations in tube voltage and Degree of Pseudoenhancement (HU) Size (mm) 10+ Fig. 3 Box plot shows size of lesion in relation to degree of pseudoenhancement, which is 24, 16, and 13 HU for three categories. C 136 AJR:198, January 2012

5 MDCT of Renal Cysts current-time settings in addition to effect of habitus on pseudoenhancement. In addition, dual-energy CT, which can be used to obtain unenhanced images is a promising tool that may facilitate detection of enhancement of renal lesions incidentally found during contrastenhanced CT [25]. Conclusion Pseudoenhancement of renal cysts significantly correlates with size smaller than 1 cm and central location. Although pseudoenhancement increased with higher detector number, the difference was not significant. Pseudoenhancement is exceedingly rare when the renal cyst size is 1.5 cm or larger. Although lesions smaller than 1.0 cm have traditionally been classified as too small to characterize, areas of low attenuation measuring cm have a relatively high likelihood of pseudoenhancement and therefore can fall into a similar indeterminate category. References 1. Tada S, Yamagishi J, Kobayashi H, Hata Y, Kobari T. The incidence of simple renal cyst by computed tomography. Clin Radiol 1983; 34: Laucks SP Jr, McLachlan MS. Aging and simple cysts of the kidney. Br J Radiol 1981; 54: Bosniak MA. The current radiological approach to renal cysts. Radiology 1986; 158: Silverman SG, Lee BY, Seltzer SE, Bloom DA, Corles CL, Adams DF. Small (< or = 3cm) renal masses: correlation of spiral CT features and pathologic findings. AJR 1994; 163: Maki DD, Birnbaum BA, Chakraborty DP, Jacobs JE, Carvalho BM, Herman GT. Renal cyst pseudoenhancement: beam hardening effects on CT numbers. Radiology 1999; 213: Hopper KD, Holabinko JN, Have T, Hartman DS. Variability of Hounsfield unit measurements based on CT of a renal cyst (abstr). Radiology 1995; 197(P): Bae KT, Heiken JP, Siegel CL, Bennett HF. Renal cysts: is attenuation artifactually increased on contrast-enhanced CT images? Radiology 2000; 216: Chung EP, Herts BR, Linnell G, et al. Analysis of changes in attenuation of proven renal cysts on different scanning phases of triphasic MDCT. AJR 2004; 182: Coulam CH, Sheafor DH, Leder RA, et al. Evaluation of pseudoenhancement of renal cysts during contrastenhanced CT. AJR 2000; 174: Davidson AJ, Hartman DS, Choyke PL, Wagner BJ. Radiologic assessment of renal masses: implications for patient care. Radiology 1997; 202: Siegel CL, Fisher AJ, Bennett HF. Interobserver variability in determining enhancement of renal masses on helical CT. AJR 1999; 172: Wang ZJ, Coakley FV, Fu Y, et al. Renal cyst pseudoenhancement at multidetector CT: what are the effects of number of detectors and peak tube voltage? Radiology 2008; 248: Birnbaum BA, Hindman N, Lee J, Babb JS. Renal cyst pseudoenhancement: influence of multidetector CT reconstruction algorithm and scanner type in phantom model. Radiology 2007; 244: Abdulla C, Kaira MK, Saini S, et al. Pseudoenhancement of simulated renal cysts in a phantom using different multidetector CT scanners. AJR 2002; 179: Birnbaum BA, Maki DD, Chakraborty DP, et al. Renal cyst pseudoenhancement: evaluation with an anthropomorphic body CT phantom. Radiology 2002; 225: Gokan T, Ohgiya Y, Munechika H, et al. Renal cyst pseudoenhancement with beam hardening effect on CT attenuation. Radiat Med 2002; 20: Heneghan JP, Spielmann AL, Sheafor DH, et al. Pseudoenhancement of simple renal cysts: a comparison of single and multidetector helical CT. J Comput Assist Tomogr 2002; 26: Carrim ZI, Murchison JT. The prevalence of simple renal and hepatic cysts detected by spiral computed tomography. Clin Radiol 2003; 58: Israel GM, Bosniak MA. How I do it: evaluating renal masses. Radiology 2005; 236: Bosniak MA. The small ( 3.0cm) renal parenchymal tumor: detection, diagnosis, and controversies. Radiology 1991; 179: Tsuda K, Kinouchi T, Tanikawa G, et al. Imaging characteristics of papillary renal cell carcinoma by computed tomography scan and magnetic resonance imaging. Int J Urol 2005; 12: Kim JK, Kim TK, Ahn HJ, et al. Differentiation of subtypes of renal cell carcinoma on helical CT scans. AJR 2002; 178: Levi C, Gray JE, McCullough EX, et al. The unreliability of CT numbers as absolute values. AJR 1982; 139: Bosniak MA, Rofsky NM. Problems in the detections and characterization of small renal masses. Radiology 1996; 198: Neville AM, Gupta RT, Miller CM, Merkle EM, Paulson EK, Boll DT. Detection of renal lesion enhancement with dual-energy multidetector CT. Radiology 2011; 259: AJR:198, January

Analysis of Changes in Attenuation of Proven Renal Cysts on Different Scanning Phases of Triphasic MDCT

Analysis of Changes in Attenuation of Proven Renal Cysts on Different Scanning Phases of Triphasic MDCT Eugene P. Chung 1 Brian R. Herts 1,2 Grant Linnell 1 Andrew C. Novick 2 Nancy Obuchowski 1,3 Deirdre M. Coll 1,4 Mark E. Baker 1 Received June 24, 2003; accepted after revision August 28, 2003. Presented

More information

Renal Cell Carcinoma: Attenuation Values on Unenhanced CT

Renal Cell Carcinoma: Attenuation Values on Unenhanced CT Genitourinary Imaging Original Research Pooler et al. Attenuation Values of Unenhanced CT of Renal Cell Carcinoma Genitourinary Imaging Original Research B. Dustin Pooler 1 Perry J. Pickhardt 1 Stacy D.

More information

Renal Masses With Equivocal Enhancement at CT: Characterization With Contrast- Enhanced Ultrasound

Renal Masses With Equivocal Enhancement at CT: Characterization With Contrast- Enhanced Ultrasound Genitourinary Imaging Original Research Bertolotto et al. Use of Contrast-Enhanced Ultrasound to Characterize Renal Masses Genitourinary Imaging Original Research Michele Bertolotto 1 Calogero Cicero 2

More information

Chromophobe Renal Cell Carcinoma: Multiphase MDCT Enhancement Patterns and Morphologic Features

Chromophobe Renal Cell Carcinoma: Multiphase MDCT Enhancement Patterns and Morphologic Features Genitourinary Imaging Original Research Raman et al. Enhancement Patterns and Morphologic Features of Chromophobe Renal Cell Carcinoma Genitourinary Imaging Original Research Siva P. Raman 1 Pamela T.

More information

The Incidental Renal lesion

The Incidental Renal lesion The Incidental Renal lesion BACKGROUND Increase in abdominal CT/US in last 15 years Resulted in detection of many (small) renal lesions 50% > 50yrs has at least 1 lesion majority simple cysts Renal lesions

More information

CME Article Clinics in diagnostic imaging (135)

CME Article Clinics in diagnostic imaging (135) Medical Education Singapore Med J 2011; 52(5) : 384 CME Article Clinics in diagnostic imaging (135) Pojchamarnwiputh S, Muttarak M, Sriplakich S H 1a 1b 1c 1d Fig. 1 (a) Axial unenhanced; (b & c) delayed

More information

Chapter 6. Hester Gietema Cornelia Schaefer-Prokop Willem Mali Gerard Groenewegen Mathias Prokop. Accepted for publication in Radiology

Chapter 6. Hester Gietema Cornelia Schaefer-Prokop Willem Mali Gerard Groenewegen Mathias Prokop. Accepted for publication in Radiology Chapter 6 Interscan variability of semiautomated volume measurements in intraparenchymal pulmonary nodules using multidetector-row computed tomography: Influence of inspirational level, nodule size and

More information

INTERDISCIPLINARY DISCUSSIONS IN LOCALISED RCC DIAGNOSIS AND SURGICAL STRATEGIES FOR ATYPICAL RENAL CYSTIC LESIONS. Maria Cova

INTERDISCIPLINARY DISCUSSIONS IN LOCALISED RCC DIAGNOSIS AND SURGICAL STRATEGIES FOR ATYPICAL RENAL CYSTIC LESIONS. Maria Cova INTERDISCIPLINARY DISCUSSIONS IN LOCALISED RCC DIAGNOSIS AND SURGICAL STRATEGIES FOR ATYPICAL RENAL CYSTIC LESIONS Maria Cova Radiology Department University of Trieste (IT) Eleventh European International

More information

Enhancement Characteristics of Papillary Renal Neoplasms Revealed on Triphasic Helical CT of the Kidneys

Enhancement Characteristics of Papillary Renal Neoplasms Revealed on Triphasic Helical CT of the Kidneys Brian R. Herts 1 Deirdre M. Coll 1,2 Andrew C. Novick 3 Nancy Obuchowski 1,4 Grant Linnell 1 Susan L. Wirth 1 Mark E. Baker 1 Received June 29, 2001; accepted after revision August 23, 2001. 1 Department

More information

Patient-Tailored Scan Delay for Multiphase Liver CT: Improved Scan Quality and Lesion Conspicuity With a Novel Timing Bolus Method

Patient-Tailored Scan Delay for Multiphase Liver CT: Improved Scan Quality and Lesion Conspicuity With a Novel Timing Bolus Method Gastrointestinal Imaging Original Research Schneider et al. Multiphase Liver CT Gastrointestinal Imaging Original Research J. Gabriel Schneider 1 Zhen J. Wang Wilbur Wang Judy Yee Yanjun Fu Benjamin M.

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Cystic Renal Cell Carcinomas: Do They Grow, Metastasize, or Recur?

Cystic Renal Cell Carcinomas: Do They Grow, Metastasize, or Recur? Genitourinary Imaging Original Research Jhaveri et al. Growth Patterns of Cystic Renal Cell Carcinomas Genitourinary Imaging Original Research Kartik Jhaveri 1 Priya Gupta 1 Azadeh Elmi 2 Lior Flor 1 Hadas

More information

Small (< 4 cm) Renal Masses: Differentiation of Angiomyolipoma Without Visible Fat From Renal Cell Carcinoma Using Unenhanced and Contrast-Enhanced CT

Small (< 4 cm) Renal Masses: Differentiation of Angiomyolipoma Without Visible Fat From Renal Cell Carcinoma Using Unenhanced and Contrast-Enhanced CT Genitourinary Imaging Original Research Takahashi et al. CT of Small Renal Masses Genitourinary Imaging Original Research Naoki Takahashi 1 Shuai Leng 1 Kazuhiro Kitajima 1,2 Daniel Gomez-Cardona 1,3 Prabin

More information

Measurement error of spiral CT Volumetry:

Measurement error of spiral CT Volumetry: Measurement error of spiral CT Volumetry: Influence of Low Dose CT Technique 1 Tae Gyu Lee, M.D. 2, Myung Jin Chung, M.D., Sung Bum Cho, M.D. 2, Jae Min Cho, M.D., Seog Joon Kim, M.D. 2, Sang Hyun Baik,

More information

Pulmonary Nodules: Contrast- Enhanced Volumetric Variation at Different CT Scan Delays

Pulmonary Nodules: Contrast- Enhanced Volumetric Variation at Different CT Scan Delays Cardiopulmonary Imaging Original Research Rampinelli et al. Contrast-Enhanced CT of Pulmonary Nodules Cardiopulmonary Imaging Original Research Cristiano Rampinelli 1 Sara Raimondi 2 Mauro Padrenostro

More information

Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1

Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1 Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1 Ethan J. Halpern, MD, David J. Halpern, Jeffrey H. Yanof, PhD, Sigal Amin-Spector, PhD, David Fischman, MD, Galit Aviram,

More information

An Introduction to Dual Energy Computed Tomography

An Introduction to Dual Energy Computed Tomography An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Original Article Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Shun-Yu Gao, Xiao-Peng Zhang, Yong Cui, Ying-Shi

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

Histogram Analysis of Small Solid Renal Masses: Differentiating Minimal Fat Angiomyolipoma From Renal Cell Carcinoma

Histogram Analysis of Small Solid Renal Masses: Differentiating Minimal Fat Angiomyolipoma From Renal Cell Carcinoma Genitourinary Imaging Original Research Chaudhry et al. Histogram Analysis of Small Solid Renal Masses Genitourinary Imaging Original Research Humaira S. Chaudhry 1,2 Matthew S. Davenport 1,3 Christopher

More information

Do Incidental Hyperechoic Renal Lesions Measuring Up to 1 cm Warrant Further Imaging? Outcomes of 161 Lesions

Do Incidental Hyperechoic Renal Lesions Measuring Up to 1 cm Warrant Further Imaging? Outcomes of 161 Lesions Genitourinary Imaging Original Research Genitourinary Imaging Original Research Ankur M. Doshi 1 Abimbola Ayoola Andrew B. Rosenkrantz Doshi AM, Ayoola A, Rosenkrantz AB Keywords: angiomyolipoma, hyperechoic

More information

Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis

Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis GE Healthcare Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis CT clinical case study lesion characterization Desiree Morgan, MD Vice Chair of Clinical Research Professor

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

354 Korean J Radiol 9(4), August 2008

354 Korean J Radiol 9(4), August 2008 Review of Failed CT Phantom Image Evaluations in 2005 and 2006 by the CT Accreditation Program of the Korean Institute for Accreditation of Medical Image Hye Jung Park, MD 1 Seung Eun Jung, MD 1, 2 Young

More information

Pulmonary Nodule Volumetric Measurement Variability as a Function of CT Slice Thickness and Nodule Morphology

Pulmonary Nodule Volumetric Measurement Variability as a Function of CT Slice Thickness and Nodule Morphology CT of Pulmonary Nodules Chest Imaging Original Research Myria Petrou 1 Leslie E. Quint 1 in Nan 2 Laurence H. aker 3 Petrou M, Quint LE, Nan, aker LH Keywords: chest, lung disease, MDCT, oncologic imaging,

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

Dual Energy CT of the Liver

Dual Energy CT of the Liver 34th Annual Course October 2011 Washington, DC Dual Energy CT of the Liver Vassilios Raptopoulos, MD Beth Israel Deaconess Medical Center Harvard Medical School Dual Energy CT (DECT) Different materials

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation Medical Physics and Informatics Original Research Angel et al. Radiation Dose During Chest CT Medical Physics and Informatics Original Research Erin Angel 1,2 Nazanin Yaghmai 1 Cecilia Matilda Jude 1 John

More information

Dual-Energy CT Applications in the Abdomen

Dual-Energy CT Applications in the Abdomen Dual-Energy CT Review Dual-Energy CT Review Tobias Heye 1 Rendon C. Nelson Lisa M. Ho Daniele Marin Daniel T. Boll Heye T, Nelson RC, Ho LM, Marin D, Boll DT Keywords: abdomen, dual-energy CT, iodine extraction,

More information

Bayesian Classifier for Predicting Malignant Renal Cysts on MDCT: Early Clinical Experience

Bayesian Classifier for Predicting Malignant Renal Cysts on MDCT: Early Clinical Experience Genitourinary Imaging Original Research Lee et al. Bayesian Classifier for Diagnosing Renal Cysts on MDCT Genitourinary Imaging Original Research Bayesian Classifier for Predicting Malignant Renal Cysts

More information

Multidetector Computed Tomography Evaluation of Subtypes of Renal Cell Carcinoma

Multidetector Computed Tomography Evaluation of Subtypes of Renal Cell Carcinoma Original Article Print ISSN: 2321-6379 Online ISSN: 2321-595X DOI: 10.17354/ijss/2017/390 Multidetector Computed Tomography Evaluation of Subtypes of Renal Cell Carcinoma A Rohini 1, Vidya Bhargavi 2,

More information

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T. The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume

More information

The Frequency and Significance of Small (15 mm) Hepatic Lesions Detected by CT

The Frequency and Significance of Small (15 mm) Hepatic Lesions Detected by CT 535 Elizabeth C. Jones1 Judith L. Chezmar Rendon C. Nelson Michael E. Bernardino Received July 22, 1991 ; accepted after revision October 16, 1991. Presented atthe annual meeting ofthe American Aoentgen

More information

Role of imaging in RCC. Ultrasonography. Solid lesion. Cystic RCC. Solid RCC 31/08/60. From Diagnosis to Treatment: the Radiologist Perspective

Role of imaging in RCC. Ultrasonography. Solid lesion. Cystic RCC. Solid RCC 31/08/60. From Diagnosis to Treatment: the Radiologist Perspective Role of imaging in RCC From Diagnosis to Treatment: the Radiologist Perspective Diagnosis Staging Follow up Imaging modalities Limitations and pitfalls Duangkamon Prapruttam, MD Department of Therapeutic

More information

UROGENITAL. Abstract Objective To assess the effect of oral hydration and contrastmedium

UROGENITAL. Abstract Objective To assess the effect of oral hydration and contrastmedium Eur Radiol DOI 10.1007/s00330-010-1785-8 UROGENITAL Dieter H. Szolar Manfred Tillich Klaus W. Preidler Multi-detector CT urography: effect of oral hydration and contrast medium volume on renal parenchymal

More information

Small Pulmonary Nodules: Our Preliminary Experience in Volumetric Analysis of Doubling Times

Small Pulmonary Nodules: Our Preliminary Experience in Volumetric Analysis of Doubling Times Small Pulmonary Nodules: Our Preliminary Experience in Volumetric Analysis of Doubling Times Andrea Borghesi, MD Davide Farina, MD Roberto Maroldi, MD Department of Radiology University of Brescia Brescia,

More information

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Poster No.: C-1207 Congress: ECR 2013 Type: Scientific Exhibit Authors: Y. Shimomiya, M. Kondo, M.

More information

Qualitative and Quantitative MDCT Features for Differentiating Clear Cell Renal Cell Carcinoma From Other Solid Renal Cortical Masses

Qualitative and Quantitative MDCT Features for Differentiating Clear Cell Renal Cell Carcinoma From Other Solid Renal Cortical Masses Genitourinary Imaging Original Research Lee-Felker et al. MDCT Differentiation of Clear Cell RCC Genitourinary Imaging Original Research Stephanie A. Lee-Felker 1 Ely R. Felker 1 Nelly Tan 1 Daniel J.

More information

Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT

Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT YAO ET AL ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 6, 252 257 p.p., 2003 Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT

More information

Measure #405: Appropriate Follow-up Imaging for Incidental Abdominal Lesions National Quality Strategy Domain: Effective Clinical Care

Measure #405: Appropriate Follow-up Imaging for Incidental Abdominal Lesions National Quality Strategy Domain: Effective Clinical Care Measure #405: Appropriate Follow-up Imaging for Incidental Abdominal Lesions National Quality Strategy Domain: Effective Clinical Care 2016 PQRS OPTIONS FOR INDIVIDUAL MEASURES: CLAIMS, REGISTRY DESCRIPTION:

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Detection of Renal Stones on Portal Venous Phase CT: Comparison of Thin Axial and Coronal Maximum- Intensity-Projection Images

Detection of Renal Stones on Portal Venous Phase CT: Comparison of Thin Axial and Coronal Maximum- Intensity-Projection Images Genitourinary Imaging Original Research Corwin et al. Detection of Renal Stones on Portal Venous Phase CT Genitourinary Imaging Original Research Michael T. Corwin 1 Justin S. Lee 1 Ghaneh Fananapazir

More information

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Alice B. Smith, MD

More information

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM 1. The CT Imaging Chain-10 major components & their functions a. The x-ray tube b. Generator c. Filter d. Pre-patient collimator e. Pre-detector collimator f. Detector system g. Analog to digital converter

More information

Dual Energy CT Aortography: Can We Reduce Iodine Dose??

Dual Energy CT Aortography: Can We Reduce Iodine Dose?? Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest

More information

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters Medical Physics and Informatics Review Maldjian and Goldman Reducing Radiation Dose in Body CT Medical Physics and Informatics Review FOCUS ON: Pierre D. Maldjian 1 Alice R. Goldman Maldjian PD, Goldman

More information

HHS Public Access Author manuscript Abdom Imaging. Author manuscript; available in PMC 2017 June 27.

HHS Public Access Author manuscript Abdom Imaging. Author manuscript; available in PMC 2017 June 27. Motion Artifacts in Kidney Stone Imaging Using Single-Source and Dual-Source Dual-Energy CT Scanners. A Phantom Study El-Sayed H. Ibrahim 1,2,*, Joseph G. Cernigliaro 1, Robert A. Pooley 1, James C. Williams

More information

Case Report Three-Dimensional Dual-Energy Computed Tomography for Enhancing Stone/Stent Contrasting and Stone Visualization in Urolithiasis

Case Report Three-Dimensional Dual-Energy Computed Tomography for Enhancing Stone/Stent Contrasting and Stone Visualization in Urolithiasis Case Reports in Urology Volume 2013, Article ID 646087, 4 pages http://dx.doi.org/10.1155/2013/646087 Case Report Three-Dimensional Dual-Energy Computed Tomography for Enhancing Stone/Stent Contrasting

More information

Management of the Incidental Renal Mass on CT: A White Paper of the ACR Incidental Findings Committee

Management of the Incidental Renal Mass on CT: A White Paper of the ACR Incidental Findings Committee ORIGINAL ARTICLE Management of the Incidental Renal Mass on CT: A White Paper of the ACR Incidental Findings Committee Brian R. Herts, MD a, Stuart G. Silverman, MD b, Nicole M. Hindman, MD c, Robert G.

More information

MDCT Findings of Renal Cell Carcinoma Associated With Xp11.2 Translocation and TFE3 Gene Fusion and Papillary Renal Cell Carcinoma

MDCT Findings of Renal Cell Carcinoma Associated With Xp11.2 Translocation and TFE3 Gene Fusion and Papillary Renal Cell Carcinoma Genitourinary Imaging Original Research Woo et al. MDCT in Renal Cell Carcinoma Genitourinary Imaging Original Research Sungmin Woo 1 Sang Youn Kim 1 Myoung Seok Lee 1 Kyung Chul Moon 2 See Hyung Kim 3

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

Optimization of scan delay for multi-phase computed tomography by using bolus tracking in normal canine kidney

Optimization of scan delay for multi-phase computed tomography by using bolus tracking in normal canine kidney Original Article J Vet Sci 2018, 19(2), 290-295 ㆍ https://doi.org/10.4142/jvs.2018.19.2.290 JVS Optimization of scan delay for multi-phase computed tomography by using bolus tracking in normal canine kidney

More information

The role of Bosniak classification in malignant tumor diagnosis: A single institution experience

The role of Bosniak classification in malignant tumor diagnosis: A single institution experience Original Article - Urological Oncology http://dx.doi.org/10.4111/icu.2016.57.2.100 pissn 2466-0493 eissn 2466-054X The role of Bosniak classification in malignant tumor diagnosis: A single institution

More information

Imaging characterization of renal clear cell carcinoma

Imaging characterization of renal clear cell carcinoma Imaging characterization of renal clear cell carcinoma Poster No.: C-0327 Congress: ECR 2011 Type: Educational Exhibit Authors: S. Ballester 1, A. Gaser 2, M. Dotta 1, M. F. CAPPA 1, F. Hammar 1 ; 1 2

More information

Computed Tomography of Normal Adrenal Glands in Indian Population

Computed Tomography of Normal Adrenal Glands in Indian Population IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 17, Issue 01 Ver. V January. (2018), PP 26-30 www.iosrjournals.org Computed Tomography of Normal Adrenal

More information

Taller-Than-Wide Sign of Thyroid Malignancy: Comparison Between Ultrasound and CT

Taller-Than-Wide Sign of Thyroid Malignancy: Comparison Between Ultrasound and CT Neuroradiology/Head and Neck Imaging Original Research Yoon et al. Taller-Than-Wide Sign of Thyroid Malignancy Neuroradiology/Head and Neck Imaging Original Research Soo Jeong Yoon 1 Dae Young Yoon 1,2

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

Imaging Findings of Primary Angiomyolipoma of the Pancreas: A Case Report 췌장의원발성혈관근육지방종의영상소견 1 예 : 증례보고

Imaging Findings of Primary Angiomyolipoma of the Pancreas: A Case Report 췌장의원발성혈관근육지방종의영상소견 1 예 : 증례보고 Case Report pissn 1738-2637 / eissn 2288-2928 https://doi.org/10.3348/jksr.2017.77.1.9 Imaging Findings of Primary Angiomyolipoma of the Pancreas: A Case Report 췌장의원발성혈관근육지방종의영상소견 1 예 : 증례보고 Hye Hee Kim,

More information

Comparison of CT findings of between MDR-TB and XDR-TB: A propensity score matching study

Comparison of CT findings of between MDR-TB and XDR-TB: A propensity score matching study Comparison of CT findings of between MDR-TB and XDR-TB: A propensity score matching study Purpose: The purpose of this study was to compare the CT findings between MDR-TB and XDR-TB groups using by propensity

More information

Nephrographic and Pyelographic Analysis of CT Urography: Principles, Patterns, and Pathophysiology

Nephrographic and Pyelographic Analysis of CT Urography: Principles, Patterns, and Pathophysiology Genitourinary Imaging Review Wolin et al. CT Urography Principles, Patterns, and Genitourinary Imaging Review FOCUS ON: Ely A. Wolin 1 David S. Hartman J. Ryan Olson Wolin EA, Hartman DS, Olson JR Keywords:

More information

Small solid noncalcified pulmonary nodules detected by screening chest computed tomography

Small solid noncalcified pulmonary nodules detected by screening chest computed tomography Respiratory Medicine (2007) 101, 1880 1884 Small solid noncalcified pulmonary nodules detected by screening chest computed tomography Sang-Man Jin a,b, Seung-Ho Choi c, Chul-Gyu Yoo a,b, Young-Whan Kim

More information

8/3/2016. Consultant for / research support from: Astellas Bayer Bracco GE Healthcare Guerbet Medrad Siemens Healthcare. Single Energy.

8/3/2016. Consultant for / research support from: Astellas Bayer Bracco GE Healthcare Guerbet Medrad Siemens Healthcare. Single Energy. U. Joseph Schoepf, MD Prof. (h.c.), FAHA, FSCBT-MR, FNASCI, FSCCT Professor of Radiology, Medicine, and Pediatrics Director, Division of Cardiovascular Imaging Consultant for / research support from: Astellas

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Low-Dose CT: Clinical Studies & the Radiologist Perspective

Low-Dose CT: Clinical Studies & the Radiologist Perspective Low-Dose CT: Clinical Studies & the Radiologist Perspective RD-ASiR RD-MBIR SD-FBP RD=0.35 msv (80% dose reduction) Perry J. Pickhardt, MD UW School of Medicine & Public Health Low-Dose CT: Clinical Overview

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

STANDARDIZED MANAGEMENT RECOMMENDATIONS FOR ADRENAL NODULES: EVIDENCE-BASED CONSENSUS POWERSCRIBE MACROS FROM AN ACADEMIC/PRIVATE PRACTICE

STANDARDIZED MANAGEMENT RECOMMENDATIONS FOR ADRENAL NODULES: EVIDENCE-BASED CONSENSUS POWERSCRIBE MACROS FROM AN ACADEMIC/PRIVATE PRACTICE STANDARDIZED MANAGEMENT RECOMMENDATIONS FOR ADRENAL NODULES: EVIDENCE-BASED CONSENSUS POWERSCRIBE MACROS FROM AN ACADEMIC/PRIVATE PRACTICE COLLABORATIVE Pamela Johnson 1, Darcy Wolfman 2, Upma Rawal 3,

More information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

Sonographically Identified Echogenic Renal Masses Up to 1 cm in Size Are So Rarely Malignant They Can Be Safely Ignored

Sonographically Identified Echogenic Renal Masses Up to 1 cm in Size Are So Rarely Malignant They Can Be Safely Ignored ORIGINAL RESEARCH Sonographically Identified Echogenic Renal Masses Up to 1 cm in Size Are So Rarely Malignant They Can Be Safely Ignored Malak Itani, MD, Amit Pandya, MBBS, Ronald O. Bude, MD Received

More information

Comparison of CT findings between MDR-TB and XDR-TB

Comparison of CT findings between MDR-TB and XDR-TB Comparison of CT findings between MDR-TB and XDR-TB Poster No.: C-0757 Congress: ECR 2017 Type: Authors: Keywords: DOI: Scientific Exhibit K. Yoon, H. Soohee; Changwon-si/KR Thorax, Lung, Respiratory system,

More information

MEASUREMENT OF EFFECT SOLID TUMOR EXAMPLES

MEASUREMENT OF EFFECT SOLID TUMOR EXAMPLES MEASUREMENT OF EFFECT SOLID TUMOR EXAMPLES Although response is not the primary endpoint of this trial, subjects with measurable disease will be assessed by standard criteria. For the purposes of this

More information

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Int J Cardiovasc Imaging (2008) 24:547 556 DOI 10.1007/s10554-007-9282-0 ORIGINAL PAPER Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Jaap M. Groen Æ Marcel J.

More information

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Young Hen Lee 1 Eun-tae Park 1 Pyong Kon Cho 1 Hyung Suk Seo 1 Bo-Kyung Je 1 Sang-il Suh 2 Kyung-Sook

More information

CT-imaging features of renal epithelioid angiomyolipoma

CT-imaging features of renal epithelioid angiomyolipoma Liu et al. World Journal of Surgical Oncology (2015) 13:280 DOI 10.1186/s12957-015-0700-9 WORLD JOURNAL OF SURGICAL ONCOLOGY RESEARCH Open Access CT-imaging features of renal epithelioid angiomyolipoma

More information

Low-dose CT Lung Cancer Screening Guidelines for Pulmonary Nodules Management Version 2

Low-dose CT Lung Cancer Screening Guidelines for Pulmonary Nodules Management Version 2 Low-dose CT Lung Cancer Screening Guidelines for Pulmonary Nodules Management Version 2 The Committee for Management of CT-screening-detected Pulmonary Nodules 2009-2011 The Japanese Society of CT Screening

More information

MRI IN THE CHARACTERIZATION OF SEMINOMATOUS AND NONSEMINOMATOUS GERM CELL TUMORS OF THE TESTIS

MRI IN THE CHARACTERIZATION OF SEMINOMATOUS AND NONSEMINOMATOUS GERM CELL TUMORS OF THE TESTIS MRI IN THE CHARACTERIZATION OF SEMINOMATOUS AND NONSEMINOMATOUS GERM CELL TUMORS OF THE TESTIS Ambesh Deshar *, Gyanendra KC and Zhang Lopsang *Department of Medical Imaging and Nuclear Medicine, First

More information

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Mannudeep K. Kalra, MD, DNB Michael M. Maher, MD, FRCR, FFR (RCSI) Srinivasa R.

More information

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation Petter Bugge Askeland Project thesis at the Faculty of Medicine UNIVERSITETET

More information

Research Article The Capabilities and Limitations of Clinical Magnetic Resonance Imaging for Detecting Kidney Stones: A Retrospective Study

Research Article The Capabilities and Limitations of Clinical Magnetic Resonance Imaging for Detecting Kidney Stones: A Retrospective Study Biomedical Imaging Volume 2016, Article ID 4935656, 6 pages http://dx.doi.org/10.1155/2016/4935656 Research Article The Capabilities and Limitations of Clinical Magnetic Resonance Imaging for Detecting

More information

Dual-Energy CT for Quantification of Urinary Stone Composition in Mixed Stones: A Phantom Study

Dual-Energy CT for Quantification of Urinary Stone Composition in Mixed Stones: A Phantom Study Genitourinary Imaging Original Research Leng et al. Dual-Energy CT of Urinary Stones Genitourinary Imaging Original Research Shuai Leng 1 Alice Huang 1 Juan Montoya Cardona 1 Xinhui Duan 1,2 James C. Williams

More information

Hyperechoic renal masses

Hyperechoic renal masses Hyperechoic renal masses Jean-Yves Meuwly, MD Department of Diagnostic and Interventional Radiology, University Hospital Lausanne, Switzerland Department of Diagnostic and Interventional Radiology Renal

More information

Genitourinary Imaging Original Research

Genitourinary Imaging Original Research Genitourinary Imaging Original Research Masch et al. Genitourinary Imaging Original Research William R. Masch 1 Richard H. Cohan 1,2 James H. Ellis 1,2 Jonathan R. Dillman 1,3 Jonathan M. Rubin 1,2 Matthew

More information

Outcomes in the NLST. Health system infrastructure needs to implement screening

Outcomes in the NLST. Health system infrastructure needs to implement screening Outcomes in the NLST Health system infrastructure needs to implement screening Denise R. Aberle, MD Professor of Radiology and Bioengineering David Geffen School of Medicine at UCLA 1 Disclosures I have

More information

Fracture risk in unicameral bone cyst. Is magnetic resonance imaging a better predictor than plain radiography?

Fracture risk in unicameral bone cyst. Is magnetic resonance imaging a better predictor than plain radiography? Acta Orthop. Belg., 2011, 77, 230-238 ORIGINAL STUDY Fracture risk in unicameral bone cyst. Is magnetic resonance imaging a better predictor than plain radiography? Nathalie PiREAU, Antoine DE GHELDERE,

More information

REVIEW. Distinguishing benign from malignant adrenal masses

REVIEW. Distinguishing benign from malignant adrenal masses Cancer Imaging (2003) 3, 102 110 DOI: 10.1102/1470-7330.2003.0006 CI REVIEW Distinguishing benign from malignant adrenal masses Isaac R Francis Professor of Radiology, Department of Radiology, University

More information

Genitourinary Imaging Original Research

Genitourinary Imaging Original Research Genitourinary Imaging Original Research Patel et al. CT of Synchronous Renal Masses Genitourinary Imaging Original Research Uday Patel 1 Navin Ramachandran James Halls Aneeta Parthipun Catherine Slide

More information

Hepatocellular carcinoma (HCC) is a malignant liver neoplasm

Hepatocellular carcinoma (HCC) is a malignant liver neoplasm Diagn Interv Radiol 2011; 17:328 333 Turkish Society of Radiology 2011 ABDOMINAL IMAGING ORIGINAL ARTICLE Correlation of dynamic multidetector CT findings with pathological grades of hepatocellular carcinoma

More information

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert University of Groningen Quantitative CT myocardial perfusion Pelgrim, Gert IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check

More information

Estimating Iodine Concentration from CT Number Enhancement

Estimating Iodine Concentration from CT Number Enhancement Estimating Iodine Concentration from CT Number Enhancement Rosemary Eaton, Andrew Shah, Jane Shekhdar Medical Physics, Mount Vernon Hospital CT Users Group 4 th October 212, Edinburgh Summary Background

More information

Imaging-Guided Radiofrequency Ablation of Cystic Renal Neoplasms

Imaging-Guided Radiofrequency Ablation of Cystic Renal Neoplasms Vascular and Interventional Radiology Original Research Allen et al. Radiofrequency Ablation of Cystic Renal Neoplasms Vascular and Interventional Radiology Original Research Brian C. Allen 1 Michael Y.

More information

Pediatric Retroperitoneal Masses Radiologic-Pathologic Correlation

Pediatric Retroperitoneal Masses Radiologic-Pathologic Correlation Acta Radiológica Portuguesa, Vol.XVIII, nº 70, pág. 61-70, Abr.-Jun., 2006 Pediatric Retroperitoneal Masses Radiologic-Pathologic Correlation Marilyn J. Siegel Mallinckrodt Institute of Radiology, Washington

More information

Automated Detection of Polyps from Multi-slice CT Images using 3D Morphologic Matching Algorithm: Phantom Study

Automated Detection of Polyps from Multi-slice CT Images using 3D Morphologic Matching Algorithm: Phantom Study Automated Detection of Polyps from Multi-slice CT Images using 3D Morphologic Matching Algorithm: Phantom Study Yonghum Na, Jin Sung Kim, Bruce R Whiting, K. Ty Bae Electronic Radiology Laboratory, Mallinckrodt

More information

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Poster No.: B-0742 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Paper

More information

INTRALUMINAL GAS IN NON-PERFORATED ACUTE APPENDICITIS: A predictor of gangrenous appendicitis

INTRALUMINAL GAS IN NON-PERFORATED ACUTE APPENDICITIS: A predictor of gangrenous appendicitis INTRALUMINAL GAS IN NON-PERFORATED ACUTE APPENDICITIS: A predictor of gangrenous appendicitis DM Plata Ariza, MD; E Martínez Chamorro, MD; D Castaño Pardo, MD; M Arroyo López, MD; E Peghini Gavilanes,

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information