Ultrasound Physics & Terminology

Size: px
Start display at page:

Download "Ultrasound Physics & Terminology"

Transcription

1 Ultrasound Physics & Terminology This module includes the following: Basic physics terms Basic principles of ultrasound Ultrasound terminology and terms Common artifacts seen Doppler principles Terms for labeling and scan orientation Ultrasound Physics & Terminology, Rev

2 Basic Physics Terms Ultrasound: Resolution: Axial Resolution: Lateral Resolution: Utilizes sound waves of very high frequency (2MHz or greater). It is propagated via waves of compression and rarefaction, and requires a medium (tissue) for travel. The higher the frequency, the less depth penetration, however the resolution is improved. Is the parameter of an ultrasound imaging system that characterizes its ability to detect closely spaced interfaces and displays the echoes from those interfaces as distinct and separate objects. The better the resolution, the greater the clarity of an ultrasound image. Is the minimum required reflector separation along the direction of propagation required to produce separate reflections. Good axial resolution is achieved with short spatial pulse lengths. Short spatial pulse lengths are a result of higher frequency and higher damped transducers. Therefore the higher the frequency the better the resolution. Is the minimum reflector separation perpendicular to the direction of propagation required to produce separate reflections. Good lateral resolution is achieved with narrow acoustic beams. A narrow acoustic beam is the result of a long near zone and a small angle of divergence in the far zone. Transducers: Convert one form of energy to another. Ultrasound transducers convert electric energy into ultrasound energy and vice versa. Transducers operate on piezoelectricity meaning that some Materials (ceramics, quartz) produce a voltage when deformed by an applied pressure, and reversely results in a production of pressure when these materials are deformed by an applied voltage. Pulsed Transducers: Mechanical Probes: Consists of one transducer element which functions as both the source and receiving transducers. Allows the sweeping of the ultrasound beam through the tissues rapidly and repeatedly. This is accomplished by oscillating a transducer. The oscillating component is immersed in a coupling liquid within the transducer assembly. In our case the coupling fluid is deionized water. It is important that the fluid is bubble free, so that your image is not compromised. Check the Ultrasound Physics & Terminology, Rev

3 Attenuation: water level in the transducer assembly before scanning and if you see air bubbles, make sure you fill it with the deionized water. A decrease in amplitude and intensity, as sound travels through a medium. Attenuation occurs with absorption (conversion of sound to heat), reflection (portion of sound returned from the boundary of a medium, and scattering (diffusion or redirection of sound in several directions when encountering a particle suspension or a rough surface). These different forms of attenuation are responsible for artifacts that may be in your image. Some of these artifacts are useful and some are not. Some artifacts are produced by improper transducer location or machine settings. Basic Principles of Ultrasound Sound Waves Audible sound waves lie within the range of 20 to 20,000 Hz. Clinical ultrasound systems use transducers of between 2 and 17 MHz. The Vevo VisualSonics ultrasound system uses frequencies of between 20 to 55 MHz. Sound waves do not exist in a vacuum, and propagation in gases is poor because the molecules are too widely spaced which is why lung does not image well with ultrasound. A gel couplant is used between the skin of the subject and the transducer face otherwise the sound would not be transmitted across the air-filled gap. The strength of the returning echo is directly related to the angle at which the beam strikes the acoustic interface. The more nearly perpendicular the beam is the stronger the returning echo; smooth interfaces at right angles are known as specular reflectors. This is best seen in the walls of a large blood vessel such as the aorta or the carotid artery. Transducers Ultrasound Physics & Terminology, Rev

4 The choice of which transducer should be used depends on the depth of the structure being imaged. The higher the frequency of the transducer crystal, the less penetration it has but the better the resolution. So if more penetration is required you need to use a lower frequency transducer with the sacrifice of some resolution. The shape of the beam is varied and is different for each transducer frequency. There is a fixed focused region of the ultrasound beam which is indicated on the system with a small triangle to the right of the image. This indicates the focal zone of that transducer and is where the best resolution can be achieved with that particular transducer. Effort should be taken to position the object of interest in the subject to within that focused area to obtain the best detail. This can be achieved with the use of more or less ultrasound gel and moving the transducer closer to or farther away from the subject. Ultrasound Terminology and Terms Anechoic: A-Mode Amplitude modulation: Attenuation: B-Mode Brightness modulation: Complex: Cystic: Enhancement (acoustic): Frequency: Gain: A structure that does not produce any internal echoes A single dimension display consisting of a horizontal baseline. This baseline represents time and or distance with upward (vertical) deflections (spikes depicting the acoustic interfaces) The ultrasound beam undergoes a progressive weakening as it penetrates the body due to absorption, scattering and beam spread. The amount of weakening is dependent on frequency, tissue density, and the number and types of interfaces A two-dimensional display of ultrasound. The A- mode spikes are electronically converted into dots and displayed at the correct depth from the transducer Refers to a mass that has both fluid-filed and solid areas within it This term is used to describe any fluid-filled structure, for example, the urinary bladder Sound is not weakened (attenuated) as it passes through a fluid-filled structure and therefore the structure behind appears to have more echoes than the same tissue beside it The number of complete cycles per second (Hertz) Refers to the amount of amplification of the returning echoes Ultrasound Physics & Terminology, Rev

5 Gel Couplant: Homogenous: Hypo-echoic: Hyper-echoic: Interface: M-Mode: Noise: Reverberation: Shadowing: Time-Gain Compensation: Transducer: Velocity (of sound): A trans-sonic material which eliminates the air interface between the transducer and the animal s skin Of uniform appearance and texture A relative term used to describe an area that has decreased brightness of its echoes relative to an adjacent structure Also a relative term used to describe a structure which has increased brightness of its echoes relative to an adjacent structure Strong echoes that delineate the boundary of organs, caused by the difference between the acoustic impedance of the two adjacent structures; an interface that is usually more pronounced when the transducer is perpendicular to it is the motion mode displaying moving structures along a single line in the ultrasound beam An artifact that is usually due to the gain control being too high An artifact that results from a strong echo returning from a large acoustic interface to the transducer. This echo returns to the tissues again, causing additional echoes parallel and equidistant to the first echo Failure of the sound beam to pass through an object, e.g. a bone does not allow any sound to pass through it and there is only shadowing seen behind it Compensation for attenuation is accomplished by amplifying echoes in the near field slightly and progressively increasing amplification as echoes return from greater depths A device which houses the element for transmitting and receiving ultrasound waves. Also referred to as a probe or Scanhead Is the speed at which a sound wave is traveling. In soft tissue at 37 degrees C. sound travels at 1540 m/second Common Artifacts Seen Reverberation: Multiple reflections commonly seen in the bladder or heart Ultrasound Physics & Terminology, Rev

6 Shadowing: Enhancement: Speckle: Created by strong reflectors, or attenuating structures, i.e. bone, gas, calcifications and air Increase in reflection amplitude from reflectors that lie behind a weakly attenuating structure, i.e. cysts, solid masses The granular appearance of images and spectral displays that is caused by the interference of echoes from the distribution of scatterers in tissue. Within transducers, there is a FOCUS which concentrates the sound beam into a smaller beam area than would exist otherwise. This area of focus is where you will obtain your best images. You will find the focus on the monitor (arrow), on the vertical millimeter scale. So when positioning your anatomy make sure it is in the region of the focus, so that you obtain your best images. Time Gain Compensation (TGC): Equalizes differences in received reflection amplitudes because of the reflector depth. Reflectors with equal reflector coefficients will not result in equal amplitude reflections arriving at the transducer if their travel distances are different. TGC allow you to adjust the amplitude to compensate for the path length differences. The longer the path length the higher the amplitude. The TGC is located on the right upper hand corner of the monitor, and is displayed graphically. B-MODE (brightness mode): The mode that is used for the display of echoes that return to the transducer. There is a change in spt brightness for each echo that is received by the transducer. The returning echoes are displayed on a television monitor as shades of gray. Typically the brighter gray shades represent echoes with greater intensity levels. This mode allows you to scan. Ultrasound Physics & Terminology, Rev

7 M-MODE (motion mode): Is a graphic B-mode pattern that is a single line time display that represents the motion of structures along the ultrasound beam, 1000fps. This mode allows you to trace motion i.e. heart wall motion, vessel wall motion. PW MODE (pulsed-wave mode): Frequency change of reflected sound waves as a result of reflection motion relative to the transducer used to detect the velocity and direction of blood flow. This reflection shift can be displayed graphically, as well as audibly. During Doppler operation the reflected sound has the same frequency as the transmitted sound if the blood is stationary ( we know that blood is not stationary it moves) therefore if the blood is moving away from the transducer a lower frequency is detected (negative shift) the spectrum appears below the baseline. If the blood is moving toward the transducer a higher frequency (positive shift) is detected and the spectral displays above the baseline Doppler shift: Is dependent on the insonating frequency, the velocity of moving blood and the angle between the sound beam and direction of the moving blood. If the sound beam is perpendicular to the direction of blood flow, there will be no doppler shift therefore there would be no display of flow in the vessel. The angle of the sound beam should be less than 60 degrees at all times. Sample Volume: Is the gate length which chooses the doppler shifts that will be used to produce audible sounds or spectral display. The larger the sample volume the more Doppler frequencies detected. Aliasing: Is the production of false doppler shift and blood velocity information when the Doppler shift exceeds a threshold. It appears as if the spectral display is cut off and wraps around and reappears in the opposite region of the display. Spectral Broadening: The widening of the doppler shift spectrum. Meaning the increase of the range of doppler shift frequencies present, owing to a broader range of flow speeds encountered by the sound beam. Ultrasound Physics & Terminology, Rev

8 Color Doppler: Doppler echoes are usually displayed with gray scale brightness corresponding to their intensities. In color doppler echoes are displayed with colors corresponding to the direction of flow that their positive or negative doppler shifts represent (toward or away from the transducer). The brightness of the color represents the intensity of the echoes, and sometimes other colors are added to indicate the extent of spectral broadening. Power Doppler: Depicts the amplitude or power of doppler signals rather than the frequency shift. This allows detection of a larger range of doppler shifts and therefore better visualization of the smaller vessels, but at the expense of directional and velocity information. DOPPLER PRINCIPLES Doppler Displays the change in frequency of a wave resulting in the motion of the wave source or reflector. In ultrasound the reflector is the moving red blood cell. The Doppler shift is dependent on the insonating frequency (transducer frequency), the velocity of the moving red blood cells, and the angle of the sound beam and direction of the moving red blood cells. The following is the Doppler equation: Df = 2f v cos q c Df = Doppler shift frequency (difference between transmitted and received) f = Transmitted frequency v = Velocity of the blood c = Speed of sound q = Angle of the sound beam and direction of moving blood cells The equation can be rearranged to detect the velocity of the blood flow with the following equation: v = Df c_ 2 f cos q Ultrasound Physics & Terminology, Rev

9 Two important facts can be observed from this equation. First, since the cosine of 90 degrees is zero, if the ultrasound beam is perpendicular to the direction of the blood flow a Doppler shift and potentially incorrect impression of the blood flow velocities. Second, the cosine function is steeper above 60 degrees and therefore errors are magnified with insonating angles above this. Therefore, careful consideration should be taken to obtain an angle of less than 60 degrees to the direction of the blood flow to obtain reliable and accurate results in quantifying the velocity in a certain blood vessel. Power Doppler Depicts amplitude or power of the Doppler signal rather than the frequency shift. Therefore, there is less angle dependence and a visualization of smaller vessels with a Doppler shift, however, with the sacrifice of velocity and directional information. Pulsed Wave Doppler With the use of a sample gate or volume gives a graphical display of all the velocities within the area sampled. The amplitude of the signal is proportional to the number of blood cells and is indicated as a shade of gray. Ultrasound Physics & Terminology, Rev

10 TERMS FOR LABELING AND SCAN ORIENTATION Coronal: Transverse: Sagittal (Longitudinal): Superior, Cranial, Cephalad, Rostral: Inferior or Caudal: Anterior or Ventral: Posterior or Dorsal: Medial: Lateral: Proximal: Distal: The long axis of a scan performed from the subject s side where the slice divides the anterior from the posterior or the dorsal from the ventral in the long axis A cross-sectional view The long axis plane Interchangeable terms indicating the direction towards the head Indicating the direction towards the feet A structure lying towards the front of the subject A structure ling towards the back of the subject Towards the midline Away from the midline Towards the origin Away from the origin Ultrasound Physics & Terminology, Rev

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012 Principles of Ultrasound Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012 None Disclosures Outline Introduction Benefits and Limitations of US Ultrasound (US) Physics

More information

CONTENTS. Test Number cpd Tanya Reynolds (Nat. Dip. Diag. Rad., B. Tech. Diag. Rad., B. Tech. Ultrasound)

CONTENTS. Test Number cpd Tanya Reynolds (Nat. Dip. Diag. Rad., B. Tech. Diag. Rad., B. Tech. Ultrasound) CONTENTS page 1-15 page 16 BASIC 2-DIMENSIONAL ULTRASOUND PRINCIPLES Multiple Choice Test Test Number cpd 41640 Tanya Reynolds (Nat. Dip. Diag. Rad., B. Tech. Diag. Rad., B. Tech. Ultrasound) Tanya is

More information

1 Fundamentals. Basic Definitions and Physics Principles. Fundamentals

1 Fundamentals. Basic Definitions and Physics Principles. Fundamentals 1 To become versed in the language of ultrasonography, it is necessary to review some of the basic principles of physics. The wave physics principles of ordinary (i.e., audible) sound apply to ultrasound

More information

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics The Physics of Ultrasound Pipe Organ 10-8000 Emission Dog 452-1080 Man 85-1100 Spectrum Bat 10,000-120,000 Porpoise 7000-120,000 Claus G. Roehrborn Professor and Chairman 10 20 Cycles per second Reception

More information

Ultrasound Physics & Doppler

Ultrasound Physics & Doppler Ultrasound Physics & Doppler Endocrine University 2018 Mark Lupo, MD, FACE, ECNU Objectives Review the essential components of ultrasound physics in neck sonography Demonstrate the importance of ultrasound

More information

Ultrasound Principles cycle Frequency Wavelength Period Velocity

Ultrasound Principles cycle Frequency Wavelength Period Velocity ! Teresa S. Wu, MD, FACEP Director, EM Ultrasound Program & Fellowship Co-Director, Simulation Based Training Program & Fellowship Associate Program Director, EM Residency Program Maricopa Medical Center

More information

Terminology Tissue Appearance

Terminology Tissue Appearance By Marc Nielsen, MD Advantages/Disadvantages Generation of Image Ultrasound Machine/Transducer selection Modes of Ultrasound Terminology Tissue Appearance Scanning Technique Real-time Portable No ionizing

More information

Introduction to Ultrasound Guided Region Anesthesia

Introduction to Ultrasound Guided Region Anesthesia Introduction to Ultrasound Guided Region Anesthesia Brian D. Sites, MD Dept of Anesthesiology Dartmouth-Hitchcock Medical Center INTRODUCTION Welcome to Introduction to Ultrasound Guided Regional Anesthesia.

More information

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016 Ultrasound Principles of Medical Imaging Prof. Dr. Philippe Cattin MIAC, University of Basel Contents Abstract 1 Image Generation Echography A-Mode B-Mode M-Mode 2.5D Ultrasound 3D Ultrasound 4D Ultrasound

More information

Ultrasound Physics and Knobology Alan Macfarlane. Consultant Anaesthetist Glasgow Royal Infirmary

Ultrasound Physics and Knobology Alan Macfarlane. Consultant Anaesthetist Glasgow Royal Infirmary Ultrasound Physics and Knobology Alan Macfarlane Consultant Anaesthetist Glasgow Royal Infirmary RAPM 2009; 34: 40-46 Ultrasound Proficiency Understanding US image generation and device operation Image

More information

Physical Principles of Ultrasound

Physical Principles of Ultrasound Physical Principles of Ultrasound Grateful appreciation to Richard A. Lopchinsky, MD, FACS and Nancy H. Van Name, RDMS, RTR, and MarleneKattaron, RDMS 2000 UIC All Rights Reserved. Course Objectives Identify

More information

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D. Diagnostic Ultrasound Sutiporn Khampunnip, M.D. Definition of Ultrasound Ultrasound is simply sound waves, like audible sound. High-frequency sound and refers to mechanical vibrations above 20 khz. Human

More information

Preamble (disclaimer)

Preamble (disclaimer) Preamble (disclaimer) PHYSICS AND PRINCIPLES OF HEAD/NECK ULTRASOUND Joseph C. Sniezek, MD FACS LTC, MC, USA Otolaryngology/H&N Surgery Tripler Army Medical Center 1. I am not a physicist 2. ACS has recommended

More information

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah ULTRASOUND IMAGING EE 472 F2018 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Diagnostic Ultrasound: Physics and Equipment, 2nd ed., by Peter R. Hoskins (Editor), Kevin Martin (Editor),

More information

What is Ultrasound? What is Ultrasound? B A. Basic Principles of Ultrasound. Basic Principles of Ultrasound. Basic Principles of Ultrasound

What is Ultrasound? What is Ultrasound? B A. Basic Principles of Ultrasound. Basic Principles of Ultrasound. Basic Principles of Ultrasound Introduction to Ultrasound Principles Mani Montazemi, RDMS Baylor College of Medicine Division of Maternal-Fetal Medicine Department of Obstetrics and Gynecology Manager, Maternal Fetal Center Imaging

More information

Lesson 03: Sound Wave Propagation and Reflection. This lesson contains 15 slides plus 14 multiple-choice questions.

Lesson 03: Sound Wave Propagation and Reflection. This lesson contains 15 slides plus 14 multiple-choice questions. Lesson 03: Sound Wave Propagation and Reflection This lesson contains 15 slides plus 14 multiple-choice questions. Accompanying text for the slides in this lesson can be found on pages 8 through 14 in

More information

Basic of Ultrasound Physics E FAST & Renal Examination. Dr Muhammad Umer Ihsan MBBS,MD, DCH CCPU,DDU1,FACEM

Basic of Ultrasound Physics E FAST & Renal Examination. Dr Muhammad Umer Ihsan MBBS,MD, DCH CCPU,DDU1,FACEM Basic of Ultrasound Physics E FAST & Renal Examination Dr Muhammad Umer Ihsan MBBS,MD, DCH CCPU,DDU1,FACEM What is Sound? Sound is Mechanical pressure waves What is Ultrasound? Ultrasounds are sound waves

More information

Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System

Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System January 2007 Objectives: After completion of this module, the participant will be able to accomplish the following: Understand

More information

Introduction & Physics of ED Ultrasound. Objectives. What? - Limited Studies. Who? - ED Docs

Introduction & Physics of ED Ultrasound. Objectives. What? - Limited Studies. Who? - ED Docs Introduction & Physics of ED Ultrasound Martine Sargent, MD Ultrasound Director, Assistant Professor UCSF Department of Emergency Medicine San Francisco General Hospital & Trauma Center Objectives Who?

More information

Ultrasound in Anesthesia: Applying Scientific Principles to Clinical Practice

Ultrasound in Anesthesia: Applying Scientific Principles to Clinical Practice AANA Journal Course Update for Nurse Anesthetists 3 6 CE Credits* Ultrasound in Anesthesia: Applying Scientific Principles to Clinical Practice Christian R. Falyar, CRNA, DNAP The use of ultrasound as

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu Basic principles. Comparison to X-rays Ultrasound > 20kHz

More information

ULTRASOUND. OB/Gyn (Core) Ultrasound PIEZOELECTRIC EFFECT. Principles of Ultrasound Physics and Instrumentation. Nathan Pinkney, BS, CDOS

ULTRASOUND. OB/Gyn (Core) Ultrasound PIEZOELECTRIC EFFECT. Principles of Ultrasound Physics and Instrumentation. Nathan Pinkney, BS, CDOS 1 OB/Gyn (Core) Ultrasound Principles of Ultrasound Physics and Instrumentation Nathan Pinkney, BS, CDOS Philadelphia College of Osteopathic Medicine 2016 ULTRASOUND CATEGORIES OF SOUND INFRASOUND = below

More information

ULTRASOUND NOMENCLATURE

ULTRASOUND NOMENCLATURE Chapter 1: Ultrasound Nomenclature, Image Orientation, and Basic Instrumentation CYNTHIA SIKOWSKI Ultrasound waves are sound waves that have a frequency exceeding 20,000 Hz. When sound waves are transmitted

More information

DIGITAL IMAGE PROCESSING IN ULTRASOUND IMAGES

DIGITAL IMAGE PROCESSING IN ULTRASOUND IMAGES DIGITAL IMAGE PROCESSING IN ULTRASOUND IMAGES Kamaljeet Kaur Computer Science & Engineering Department Guru Nanak Dev Engg. College, Ludhiana. Punjab-India meetk.89@gmail.com ABSTRACT-- Image processing

More information

Dr Emma Chung. Safety first - Physical principles for excellent imaging

Dr Emma Chung. Safety first - Physical principles for excellent imaging Safety first - Physical principles for excellent imaging Dr Emma Chung Lecturer in Medical Physics, University of Leicester Clinical Scientist, University Hospitals of Leicester NHS Trust Thanks to Caroline

More information

Basic Physics of Ultrasound and Knobology

Basic Physics of Ultrasound and Knobology WELCOME TO UTMB Basic Physics of Ultrasound and Knobology By Daneshvari Solanki, FRCA Laura B. McDaniel Distinguished Professor Anesthesiology and Pain Medicine University of Texas Medical Branch Galveston,

More information

Ultrasonic Testing Level I:

Ultrasonic Testing Level I: Ultrasonic Testing Level I: 1- Sound Wave - Introduction - ASNT Level I - Sound Wave Propagation - Velocity / Frequency / Wave Length - Acoustic Impedance - Energy / Intensity 2- Ultrasound Wave Modes

More information

Point-of-Care Ultrasound: An Introduction

Point-of-Care Ultrasound: An Introduction Point-of-Care Ultrasound: An Introduction Delegation Teaching Package for Registered Respiratory Therapists and Anesthesia Assistants Developed by: Rob Bryan RRT, AA Edited by: Kelly Hassall RRT, FCSRT,

More information

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus Page 1 of 7 11/18 Candidates are expected to cover all of the content of this syllabus when preparing

More information

Chapter 14. Imaging Artifacts

Chapter 14. Imaging Artifacts Chapter 14 Image Artifacts The complex physical interactions that occur between an ultrasound beam and human anatomy and the intricate and sophisticated technological components of a sonographic imaging

More information

What is Ultrasound? Resolution Image production Attenuation Imaging modes Ultrasound artifacts... 7

What is Ultrasound? Resolution Image production Attenuation Imaging modes Ultrasound artifacts... 7 What is Ultrasound?... 1 Resolution... 3 Image production... 3 Attenuation... 4 Imaging modes... 5 Ultrasound artifacts... 7 0 What is Ultrasound? High frequency sound of frequencies 2-50 MHz is used in

More information

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound CH.12. Dr.Rajaa Sound in medicine أ.م.د. رجاء سهيل جنم جامعة تكريت كلية Sound : It is the audible waves of frequency between 20 Hz and 20 khz. Infrasound : refers to the sound of frequency below the normal

More information

Concepts of Imaging and Knobology

Concepts of Imaging and Knobology Concepts of Imaging and Knobology Pravin Patil, MD FACC FASE Associate Professor of Medicine Director, Cardiovascular Disease Training Program Lewis Katz School of Medicine at Temple University Disclosures

More information

Supplement (videos)

Supplement (videos) Supplement (videos) Ruben s tube (sound): http://www.youtube.com/watch?v=gpcquuwqayw Doppler US (diagnostic use): http://www.youtube.com/watch?v=fgxzg-j_hfw http://www.youtube.com/watch?v=upsmenyoju8 High

More information

Physical Principles of Ultrasound

Physical Principles of Ultrasound Physical Principles of Ultrasound Pat F. Fulgham 2 Introduction The use of ultrasound is fundamental to the practice of urology. In order for urologists to best use this technology on behalf of their patients,

More information

Breast Imaging Essentials

Breast Imaging Essentials Breast Imaging Essentials Module 9 Transcript 2016 ASRT. All rights reserved. Breast Imaging Essentials Module 9 Breast Ultrasound 1. ASRT Animation 2. Welcome Welcome to Module 9 of Breast Imaging Essentials

More information

Ultrasound guidance in regional anesthesia has

Ultrasound guidance in regional anesthesia has Ultrasound and Regional Anesthesia Artifacts and Pitfall Errors Associated With Ultrasound-Guided Regional Anesthesia. Part I: Understanding the Basic Principles of Ultrasound Physics and Machine Operations

More information

Basic Physics of Ultrasound in Transesophageal Echocardiography

Basic Physics of Ultrasound in Transesophageal Echocardiography SPECIAL ARTICLE IJUTPC Basic Physics of Ultrasound in Transesophageal Echocardiography Basic Physics of Ultrasound in Transesophageal Echocardiography 1 Mary Korula, 2 Ravi Hebballi 1 Senior Consultant,

More information

Ultrasound Knobology

Ultrasound Knobology Ultrasound Knobology Raj Dasgupta MD, FACP, FCCP, FASSM Assistant Professor of Clinical Medicine Pulmonary / Critical Care / Sleep Medicine University of Southern California (USC) Objectives Physics of

More information

WELCOME! Introduction to Bedside Ultrasound

WELCOME! Introduction to Bedside Ultrasound WELCOME! Introduction to Bedside Ultrasound TEACHERS University of California-Irvine School of Medicine Nathan Molina nathan.d.molina@gmail.com Trevor Plescia taplescia90@gmail.com Jack Silva jpsilva42@gmail.com

More information

Ultrasound Evaluation of the Posterior Segment of the Eye A Ready Reckoner

Ultrasound Evaluation of the Posterior Segment of the Eye A Ready Reckoner 180 Kerala Journal of Ophthalmology Vol. XX, No. 2 OPHTHALMIC INSTRUMENTATION Ultrasound Evaluation of the Posterior Segment of the Eye A Ready Reckoner Dr. Mahesh G. MS DO DNB FRCSEd., Dr. A. Giridhar

More information

Diagnostic approach to heart disease

Diagnostic approach to heart disease Diagnostic approach to heart disease Initial work up History Physical exam Chest radiographs ECG Special studies Echocardiography Cardiac catheterization Echocardiography principles Technique of producing

More information

Physics. Norman McDicken Tom Anderson CHAPTER ULTRASOUND. Ultrasound Propagation

Physics. Norman McDicken Tom Anderson CHAPTER ULTRASOUND. Ultrasound Propagation CHPTER 2 Physics Norman McDicken Tom nderson This chapter provides an introduction to the physics of medical ultrasound (US). Several books exist that can be consulted to extend the material presented

More information

CSB 046 Complementary Imaging Techniques

CSB 046 Complementary Imaging Techniques CSB 046 Complementary Imaging Techniques - Quizzes are only ultrasound, final includes nuc med and ultrasound Week 1 Intro to Ultrasound Physics - Uses 1 to 20 MHz frequencies, which is way above the sound

More information

Basic Ultrasound Physics Board Review Questions

Basic Ultrasound Physics Board Review Questions Basic Ultrasound Physics Board Review Questions Sidney K. Edelman, PhD ESP Ultrasound The Woodlands, TX Question 1 What is the wavelength of 2 MHz sound in soft tissue? 1. 1.54 mm 2. 0.75 mm 3. 0.75 cm

More information

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System R. K. Saha, S. Karmakar, S. Saha, M. Roy, S. Sarkar and S.K. Sen Microelectronics Division, Saha Institute of

More information

Image optimization for critical care US

Image optimization for critical care US Image optimization for critical care US 1 Although we assume you are already familiar with focused US in the ED, it might not hurt to revise the basics: Machines & transducers US appearance of normal tissues

More information

Thyroid Ultrasound Physics and Doppler

Thyroid Ultrasound Physics and Doppler Thyroid Ultrasound Physics and Doppler Advanced AACE-ACE US training course 2017 Dev Abraham MD, MRCP(UK), ECNU, FACE Professor of Medicine, University of Utah No Disclosures Natural Ability to see with

More information

Emergency Medicine Interest Group (EMIG) 2016

Emergency Medicine Interest Group (EMIG) 2016 Emergency Medicine Interest Group (EMIG) 2016 Welcome to the flipped classroom (learning objectives summary) for the 2016 Emergency Medicine Interest Group (EMIG) Procedures Workshop. Overview - Tuesday

More information

POSTERIOR 1. situated behind: situated at or toward the hind part of the body :

POSTERIOR 1. situated behind: situated at or toward the hind part of the body : ANATOMICAL LOCATION Anatomy is a difficult subject with a large component of memorization. There is just no way around that, but we have made every effort to make this course diverse and fun. The first

More information

Pulse-Echo Ultrasound Imaging. Resolution in Ultrasound Imaging. Doppler Ultrasound. Resolution vs Penetration. Medical Imaging (EL582/BE620/GA4426)

Pulse-Echo Ultrasound Imaging. Resolution in Ultrasound Imaging. Doppler Ultrasound. Resolution vs Penetration. Medical Imaging (EL582/BE620/GA4426) Medical Imaging (EL582/BE620/GA4426) Pulse-Echo Ultrasound Imaging Ultrasound Imaging Lecture 2 Daniel (Dan) Turnbull, Ph.D. Skirball Institute and Dept of Radiology NYU School of Medicine (daniel.turnbull@med.nyu.edu)

More information

Chapter 3 Physical Principles of Ultrasound of the Male Genitalia

Chapter 3 Physical Principles of Ultrasound of the Male Genitalia Chapter 3 Physical Principles of Ultrasound of the Male Genitalia Bruce R. Gilbert and Pat Fox Fulgham Introduction The value of ultrasound evaluation of the male genitalia depends, in large part, on the

More information

Ultrasonic Testing. Basic Principles

Ultrasonic Testing. Basic Principles Ultrasonic Testing Ultrasonic Testing (UT) uses high frequency sound waves (typically in the range between 0.5 and 15 MHz) to conduct examinations and make measurements. Besides its wide use in engineering

More information

Ultrasonography of the Neck as an Adjunct to FNA. Nicole Massoll M.D.

Ultrasonography of the Neck as an Adjunct to FNA. Nicole Massoll M.D. Ultrasonography of the Neck as an Adjunct to FNA Nicole Massoll M.D. Basic Features of Head and Neck Ultrasound and Anatomy Nicole Massoll M.D. University of Arkansas for Medical Sciences, Little Rock

More information

High resolution ultrasound scanner for skin imaging

High resolution ultrasound scanner for skin imaging High resolution ultrasound scanner for skin imaging Christine Turlat Sales Director Atys medical 17 Parc d Arbora 69510 SOUCIEU EN JARREST Atys company Principle of ultrasound imaging DERMCUP Normal image

More information

Optimising your Doppler settings for an accurate PI. Alison McGuinness Mid Yorks Hospitals

Optimising your Doppler settings for an accurate PI. Alison McGuinness Mid Yorks Hospitals Optimising your Doppler settings for an accurate PI Alison McGuinness Mid Yorks Hospitals Applications Both maternal uterine and fetal circulations can be studied with doppler sonography Uterine arteries

More information

TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound

TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound STEVEN SUTLIEF DOUG PFEIFFER (HEATHER PIERCE, WENGZHENG FENG, JIM KOFLER) AAPM ANNUAL MEETING 2010 Educational Objectives To describe the

More information

Ultrasound Applied Physics

Ultrasound Applied Physics Ultrasound Applied Physics University of Toronto Department of Medical Imaging Applied Physics Mini-Course #3 2016 Ultrasound Laboratory Manual and Examination Booklet 1/21/2016 Ultrasound Applied Physics

More information

Principles of echocardiography for the anesthesiologist.

Principles of echocardiography for the anesthesiologist. Principles of echocardiography for the anesthesiologist. Nikolaos Skubas, MD PhD Abstract Ultrasound-based diagnostic techniques are now part of the cardiological patients chart, while echocardiography

More information

Chapter 2 Pitfalls in Musculoskeletal Ultrasound

Chapter 2 Pitfalls in Musculoskeletal Ultrasound Chapter 2 Pitfalls in Musculoskeletal Ultrasound Violeta Maria Vlad MD, PhD Introduction Taking a good ultrasound (US) picture is an art. Interpreting it is a science. This is in fact everything US is

More information

INTRODUCTION. Getting the best scan. Choosing a probe. Choosing the frequency

INTRODUCTION. Getting the best scan. Choosing a probe. Choosing the frequency Getting the best scan Choosing a probe Select the most appropriate probe for the particular scan required. s vary in their: operating frequency range higher ultrasound frequencies provide better discrimination

More information

Background & Indications Probe Selection

Background & Indications Probe Selection Teresa S. Wu, MD, FACEP Director, EM Ultrasound Program & Fellowship Co-Director, Simulation Based Training Program & Fellowship Associate Program Director, EM Residency Program Maricopa Medical Center

More information

The 2 nd Cambridge Advanced Emergency Ultrasound Course

The 2 nd Cambridge Advanced Emergency Ultrasound Course The 2 nd Cambridge Advanced Emergency Ultrasound Course Addenbrooke s Hospital Cambridge Sept 2008 1 2 Faculty! UK! USA! Australia! Toshiba! Emergency Medicine! Radiology 3 Programme! Day 1 Introduction

More information

Ultrasound: Past and Present. Lecturer: Dr. John M Hudson, PhD

Ultrasound: Past and Present. Lecturer: Dr. John M Hudson, PhD Ultrasound: Past and Present Lecturer: Dr. John M Hudson, PhD Disclosures 2 No conflicts of interest to declare Course Outline 3 1. Survey of ultrasound physics & applications 2. (Sep 21) 3. (Sep 28) 4.

More information

The Essentials Tissue Characterization and Knobology

The Essentials Tissue Characterization and Knobology The Essentials Tissue Characterization and Knobology Randy E. Moore, DC, RDMS RMSK No relevant financial relationships Ultrasound The New Standard of Care Musculoskeletal sonography has become the standard

More information

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions.

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions. Lesson 07: Ultrasound Transducers This lesson contains 73 slides plus 16 multiple-choice questions. This lesson was derived from pages 33 through 42 in the textbook: Ultrasound Transducers Ultrasound Transducers

More information

Roaa M.Hussein Professor, Department of Physics, College of Science, Ramadi, Iraq. Abstract:

Roaa M.Hussein Professor, Department of Physics, College of Science, Ramadi, Iraq. Abstract: Abstract: Ultrasound Waves Employment in the Medical Diagnostic for Lıver and Gallbladder Faik H. Antar Professor, Department of Physics, College of Science, AL- Anbar University, Ramadi, Iraq. Ultrasound

More information

Table of contents. Foreword. Preface. 1 Introduction Historical Perspective 00

Table of contents. Foreword. Preface. 1 Introduction Historical Perspective 00 Table of contents Foreword Preface 1 Introduction 00 1.1 Historical Perspective 00 2 Fundamentals of musculoskeletal ultrasound 00 2.1 Frequency and wavelength 00 2.2 Generating ultrasound waves 00 2.3

More information

Essentials of Anatomy and Physiology, 9e (Marieb) Chapter 1 The Human Body: An Orientation. Short Answer. Figure 1.1

Essentials of Anatomy and Physiology, 9e (Marieb) Chapter 1 The Human Body: An Orientation. Short Answer. Figure 1.1 Essentials of Anatomy and Physiology, 9e (Marieb) Chapter 1 The Human Body: An Orientation Short Answer Figure 1.1 Using Figure 1.1, identify the following: 1) Label A points to the cavity. 2) Label B

More information

Routine Quality Assurance Cookbook

Routine Quality Assurance Cookbook This Cookbook is a companion guide to the AIUM Routine Quality Assurance (QA) for Diagnostic Ultrasound Equipment document, which outlines the basic QA requirements for AIUM-accredited practices. The Guide

More information

1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material

1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material 1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material. intensity / MW m 2 thickness / mm 0.91 0.40 0.69 0.80 0.52 1.20 0.40 1.60

More information

Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification

Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification Whitepaper Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification ACUSON S2000 Ultrasound System Answers for life. Page 1 Tissue Strain Analytics: Virtual Touch Tissue Imaging and Quantification

More information

Guide to Small Animal Reproductive Imaging using the Vevo 770

Guide to Small Animal Reproductive Imaging using the Vevo 770 Guide to Small Animal Reproductive Imaging using the Vevo 770 Course Objectives: After completion of this module, the participant will be able to accomplish the following: Recognize reproductive female

More information

Special Section: Fractures

Special Section: Fractures 2609_c1_cover1.qxd 10/9/07 10:56 AM Page 1 THE LEADING EDGE September 2007, Vol. 26, No. 9 Special Section: Fractures September 2007 Vol. 26, No. 9 Pages 91081-1232 Fractures THE SOCIETY OF EXPLORATION

More information

Critical Care Ultrasound Study Notes

Critical Care Ultrasound Study Notes Critical Care Ultrasound Study Notes Compiled by David Tripp October 2014 Ultrasound Physics 2 Ultrasound in Tissue 2 Ultrasound Interaction with Tissue 2 Pulsed Ultrasound and Imaging 3 Image Formation

More information

Basic Training Programme. 16 Februrary 2018, ROTTERDAM. Pre and Post-Course Test Answers

Basic Training Programme. 16 Februrary 2018, ROTTERDAM. Pre and Post-Course Test Answers Basic Training Programme 16 Februrary 2018, ROTTERDAM Pre and Post-Course Test Answers Your details: Name: Conference registration number/ BT delegate number: Email address: Are you already performing

More information

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE 4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE Purpose of experiment Determine the main characteristics of ultrasound waves, and the distances and positions of models using an ultrasonic echoscope.

More information

Descriptions of NDT Projects Fall 2004 October 31, 2004

Descriptions of NDT Projects Fall 2004 October 31, 2004 Descriptions of NDT Projects Fall 2004 October 31, 2004 Introduction There are two separate NDT labs in Magister: ULTRA for ultrasound and EDDY for eddy current. Both labs are equipped with mechanical

More information

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE 4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE Purpose of experiment Determine the main characteristics of ultrasound waves, and the distances and positions of models using an ultrasonic echoscope.

More information

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Jason Silver August 26, 2009 Presentation Outline Introduction Thesis Objectives Mathematical Model and Principles Methods

More information

Magnetic Resonance Angiography

Magnetic Resonance Angiography Magnetic Resonance Angiography 1 Magnetic Resonance Angiography exploits flow enhancement of GR sequences saturation of venous flow allows arterial visualization saturation of arterial flow allows venous

More information

Carotid Abnormalities Coils, Kinks and Tortuosity David Lorelli M.D., RVT, FACS Michigan Vascular Association Conference Saturday, October 20, 2012

Carotid Abnormalities Coils, Kinks and Tortuosity David Lorelli M.D., RVT, FACS Michigan Vascular Association Conference Saturday, October 20, 2012 Carotid Abnormalities Coils, Kinks and Tortuosity David Lorelli M.D., RVT, FACS Michigan Vascular Association Conference Saturday, October 20, 2012 Page 1 Table of Contents Carotid Anatomy Carotid Duplex

More information

Anatomy & Physiology. An Introduction

Anatomy & Physiology. An Introduction Anatomy & Physiology An Introduction An Overview of Anatomy Anatomy - The study of the structure of the human body Physiology - The study of body function Branches of Anatomy Surface anatomy Gross anatomy

More information

Strain Assessment in Echo

Strain Assessment in Echo Strain Assessment in Echo Joe M. Moody, Jr, MD UTHSCSA and STVHCS 2010 Acknowledging many illustrations from Weyman s text and others. Echo-Doppler Basic Principles Background: Ultrasound physics (resolution,

More information

Outline. QA/QC of Ultrasound Imagers: Basic Physics, Procedures and Experiences. Frame Rate Limitation. US Imaging Range Equation.

Outline. QA/QC of Ultrasound Imagers: Basic Physics, Procedures and Experiences. Frame Rate Limitation. US Imaging Range Equation. QA/QC of Ultrasound Imagers: Basic Physics, Procedures and Experiences Zheng F. Lu, PhD Radiology Department Columbia University Email: zfl1@columbia.edu Outline General overview of basic ultrasound physics

More information

Underwater Acoustic Measurements in Megahertz Frequency Range.

Underwater Acoustic Measurements in Megahertz Frequency Range. Underwater Acoustic Measurements in Megahertz Frequency Range. Current State and Prospects of Development in Russia Alexander M. Enyakov,, Many medical applications of underwater acoustic measurements

More information

Ultrasound 10/1/2014. Basic Echocardiography for the Internist. Mechanical (sector) transducer Piezoelectric crystal moved through a sector sweep

Ultrasound 10/1/2014. Basic Echocardiography for the Internist. Mechanical (sector) transducer Piezoelectric crystal moved through a sector sweep Ultrasound Basic Echocardiography for the Internist Carol Gruver, MD, FACC UT Erlanger Cardiology Mechanical wave of compression and rarefaction Requires a medium for transmission Ultrasound frequency

More information

Ultrasound Measurements and Non-destructive Testing Educational Laboratory

Ultrasound Measurements and Non-destructive Testing Educational Laboratory Session 3548 Ultrasound Measurements and Non-destructive Testing Educational Laboratory Vladimir Genis, Horacio Sosa Goodwin College of Professional Studies, Drexel University, Philadelphia, 19104 Emil

More information

Abdominal Ultrasound

Abdominal Ultrasound Abdominal Ultrasound Imaging Control Buttons Depth The organ imaged should take up 3/4 of the screen Frequency = Penetration Use high frequencies (harmonics) for fluid filled and superficial structures

More information

Thyroid and Parathyroid Ultrasound Protocol

Thyroid and Parathyroid Ultrasound Protocol Thyroid and Parathyroid Ultrasound Protocol Reviewed By: Anna Ellermeier, MD Last Reviewed: December 2017 Contact: (866) 761-4200, Option 1 **NOTE for all examinations: 1. If documenting possible flow

More information

Ultrasound B-Scan for Posterior Segment Evaluation

Ultrasound B-Scan for Posterior Segment Evaluation Retina Ultrasound B-Scan for Posterior Segment Evaluation Shalini Singh MS Shalini Singh MS, Manisha Agarwal MS, Aditya Bansal DNB Dr. Shroff s Charity Eye Hospital, New Delhi B reproducible investigation

More information

Ultrasound Guided Injections

Ultrasound Guided Injections Ultrasound Guided Injection Technique More accurate injections Better Results! 1 Benefits: Increased Level of Certainty ie : really know how accurate PRP/Prolotherapy Avoid damage to articular cartilage

More information

The Evolution and Benefits of Phased Array Technology for the Every Day Inspector

The Evolution and Benefits of Phased Array Technology for the Every Day Inspector ECNDT 2006 - Poster 198 The Evolution and Benefits of Phased Array Technology for the Every Day Inspector Dan KASS, Tom NELLIGAN, and Erich HENJES Olympus NDT, Waltham, USA Abstract. Phased arrays were

More information

Body Planes & Positions

Body Planes & Positions Learning Objectives Objective 1: Identify and utilize anatomical positions, planes, and directional terms. Demonstrate what anatomical position is and how it is used to reference the body. Distinguish

More information

Chapter 3. Sonographic Image Interpretation

Chapter 3. Sonographic Image Interpretation Chapter 3 Sonographic Image Interpretation Sonograms are two-dimensional gray-scale images that allow assessment and diagnosis of many anatomic and pathologic changes that can occur in the human body.

More information

THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION REFERENCE REFLECTORS AND TRANSDUCERS FOR COMPRESSOR BLADE DOVETAILS

THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION REFERENCE REFLECTORS AND TRANSDUCERS FOR COMPRESSOR BLADE DOVETAILS International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION

More information

An Overview of Ultrasound Testing For Lesion Detection in Human Kidney

An Overview of Ultrasound Testing For Lesion Detection in Human Kidney Journal of Tomography System & Sensors Application Vol.1, Issue 1, June 2018 An Overview of Ultrasound Testing For Lesion Detection in Human Kidney Aina Fadhilah Abd Rahim 1, Zawin Najah Abd Halim 1, Jaysuman

More information

Abdominal Ultrasound

Abdominal Ultrasound Abdominal Ultrasound What is Ultrasound Imaging of the Abdomen? What are some common uses of the procedure? How should I prepare? What does the equipment look like? How does the procedure work? How is

More information

Technique of obtaining cardiac views

Technique of obtaining cardiac views Chapter 1 Technique of obtaining cardiac views Successful ultrasound diagnosis in any context depends first, on obtaining a series of defined crosssectional images and, second, on the correct interpretation

More information

3/20/2017. Disclosures. Ultrasound Fundamentals. Ultrasound Fundamentals. Bone Anatomy. Tissue Characteristics

3/20/2017. Disclosures. Ultrasound Fundamentals. Ultrasound Fundamentals. Bone Anatomy. Tissue Characteristics Disclosures Images of ultrasound equipment in this presentation are not an endorsement Fundamentals of Musculoskeletal Ultrasound Physics and Knobology Shane A. Shapiro, M.D. Assistant Professor Orthopedic

More information