1 Validation of Analysis of Amphetamines, Opiates, Phencyclidine, Cocaine, and Benzoylecgonine in Oral Fluids by Liquid Chromatography Tandem Mass Spectrometry Subbarao V. Kala*, Steve E. Harris, Tom D. Freijo, and Stan Gerlich One Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, Texas Abstract The aim of the present study was to develop and validate a method for the detection and quantitation of drugs of abuse in oral fluids. Fortified oral fluid samples (made in-house) and samples from donors collected with Quantasil oral fluid collection kits from Immunalysis were screened on an Olympus 5400 using reagents purchased from Immunalysis. Amphetamines (AMPs), opiates, phencyclidine (PCP), and cocaine and its metabolite benzoylecgonine (BE) in oral fluids were quantitated by an Applied Biosystems 3200 QTRAP liquid chromatograph tandem mass spectrometer (LC MS MS). AMPs, opiates, PCP, cocaine, and BE were extracted from samples using liquid liquid or solid-phase extractions and the extracts were separated on a Shimadzu highperformance liquid chromatograph prior to the MS MS analysis. For each drug, two multiple reaction mode transitions were monitored using positive electrospray ionization coupled to an MS MS detector. Corresponding d 3, d 5, d 6, and d 11 internal standards were used to quantitate the results. The limit of detection/quantitation for AMPs, opiates, PCP, cocaine, and its metabolite BE were 10, 10, 2, 2, and 2 ng/ml of oral fluid, respectively, on a signal-to-noise ratio > 4. This corresponded to 25, 25, 5, 5, and 5 pg on column. The method was verified by participating in the North America Oral Fluid Proficiency Testing administered by Research Triangle Institute and by analyzing real samples from donors. In conclusion, LC MS MS provided a simple way to analyze and quantitate drugs of abuse in oral fluids. Introduction In recent years, oral fluid testing for drugs of abuse has become more prominent and is used in a wide variety of situations such as pre-employment, school athletics, roadside testing, prisons, and other institutions that are monitored by drug courts (1 4). Although urine is still the specimen of choice for testing of drug abuse in workplaces, oral fluid testing has become more important to detect the recent use of drugs because the drugs in oral fluids typically last between 12 to 24 h (5). A liquid chromatography tandem mass spectrometry (LC MS MS) method was developed for the confirmation and quantitation of the following drugs of abuse: amphetamine (AMP), methamphetamine (MAMP), 3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxymethamphetamine (MDMA), codeine, morphine, hydrocodone, hydromorphone, oxycodone, oxymorphone, phencyclidine (PCP), and cocaine and its metabolite benzoylecognine (BE) in oral fluids. The method was verified by testing the samples obtained from real subjects as well as by participating in the North American Oral Fluid Proficiency Testing administered by Research Triangle Institute (RTI). Oral fluid drug testing in contrast to urine is advantageous, as collection is non-invasive, and because samples are collected under direct observation, issues regarding adulteration are of little concern. The advantages and disadvantages of oral fluid testing are described in great detail by other investigators (6 9). Several investigators have initiated the testing of oral fluid samples by LC MS MS as opposed to gas chromatography MS, as the cutoffs for positive identification of drugs in oral fluids are much lower (at least 10 times) than urine. Previous studies from other investigators have shown that LC MS MS is highly sensitive for the identification of drugs in urine, oral fluids, and hair samples (10 12). Additionally, both parent and metabolite compounds can be detected with greater sensitivity. This report describes the details regarding the validation of an LC MS MS method for the detection and quantitation of different classes of drugs of abuse in oral fluids. Experimental Materials and Methods * Author to whom correspondence should be addressed. Subbarao V. Kala, Ph.D., Scientific Director, One Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, TX skala@onesourcetox.com. Chemicals and reagents All drugs of abuse (AMP, MAMP, MDA, MDMA, AMP-d 11, Reproduction (photocopying) of editorial content of this journal is prohibited without publisher s permission. 605
2 MAMP-d 11, MDA-d 5, MDMA-d 5, codeine, morphine, hydrocodone, hydromorphone, oxycodone, oxymorphone, codeine-d 3, morphine-d 3, hydrocodone-d 3, hydromorphoned 3, oxycodone-d 6, cocaine, BE, cocaine-d 3, BE-d 3, PCP, and PCP-d 5 ) were purchased from Cerilliant (Round Rock, TX). High-performance liquid chromatography (HPLC)-grade ammonium acetate, acetonitrile, formic acid, methanol, and other reagents were purchased from Fisher Scientific (Houston, TX). Calibrators and quality controls All calibrators and controls were prepared in diluted drugfree saliva in an Immunalysis collection buffer (Pomona, CA). For all drugs, a calibrator, a low control (40% of cutoff), a high control (25% above cutoff), and a negative control were used for the analysis of oral fluid samples from donors. For the determination of limits of detection (LOD) and quantitation (LOQ), diluted negative (1:3 in Immunalysis buffer) oral fluid was fortified with different concentrations of drug, and the samples were analyzed in triplicate by LC MS MS after extraction (either liquid liquid or solid-phase). Deuterated internal standards were used to quantitate the results. Several criteria, including the shape of the chromatographic peak, the retention time (within ± 2%), and qualifying ion ratios (within ± 20%) were used to monitor the quality control of the analysis of the samples. 606 Extraction procedures The Immunalysis oral fluid collection device (Quantisal) allows the collection of 1 ml of oral fluid, and this is collected into 3 ml of buffer solution. During this process, the drug in the oral fluid is diluted 4 times. All drugs were extracted using 0.5 ml of diluted oral fluid (4 times) sample. For AMP and PCP extractions, deuterated internal standard (25 µl of 250 ng/ml of amphetamine mix or 25 µl of 500 ng/ml PCP), 2.0 ml of 100 mm phosphate buffer (ph 9.3), and 2.0 ml of hexane/ethyl acetate (50:50) were added to controls and samples. Samples were vortex mixed and placed on a mechanical rotator for 4 min to allow the extraction of the drugs. The extraction batch was centrifuged at 4000 rpm for 4 min and decanted into mm test tubes. Mobile phase (125 µl of 5 mm ammonium acetate for AMPs or 0.1% formic acid for PCP) was added to each test tube and evaporated at 45 C under nitrogen stream until only the aqueous phase remained. The samples were then transferred to vials for the analysis by LC MS MS. For opiates, cocaine, and BE extractions, deuterated internal standard (25 µl of 200 ng/ml opiate mix or 25 µl of 400 ng/ml cocaine or BE) and 1.0 ml of 100 mm phosphate buffer (ph 6.0) were added to controls and samples. UCT Zsdau005 columns were prepped with 500 µl methanol, followed by 500 µl of DI water, and finally with 500 µl of 100 mm phosphate buffer (ph 6.0). The samples and controls were loaded onto the columns. The columns were then washed with 500 µl of DI water, followed by 500 µl of 100 mm acetic acid (for opiates) or 0.1 N HCl (for cocaine and BE), and finally with 500 µl of methanol. The columns were dried under vacuum and the drugs were eluted using 800 µl of MeCl/IPA/NH 4 OH (78:20:2). The samples were evaporated completely at 45 C under a nitrogen stream and 100 µl of mobile phase were added (0.1% formic acid and water) and transferred to the vials for the analysis by LC MS MS. The final concentrations of internal standards for AMPs, opiates, PCP, cocaine, and BE in diluted oral fluid were 12.5, 10, 25, 20, and 20 ng/ml, respectively, which corresponded to 50, 40, 100, 80, and 80 ng/ml of undiluted oral fluid. LOD/LOQ, precision, and accuracy studies For the precision and accuracy studies, 10 samples of low controls (40% of cutoff; AMPs at 20 ng/ml, opiates at 16 ng/ml, PCP at 4 ng/ml, cocaine at 8 ng/ml, and BE at 8 ng/ml) for each drug were analyzed by LC MS MS and the coefficient of variation and percentage of accuracy were calculated. LOD/LOQ were evaluated in triplicate. LOD is defined as the concentration producing a peak eluting within ± 0.05 min of the analyte s retention time for the lowest concentration of the drug with a signal-to-noise ratio of at least 3:1 and the qualifier ion ratios within ± 20% of the calibration standard. For LOQ, the quantitative results also should meet the criteria of being within 20% of target value in addition to satisfying all the previously mentioned criteria of LOD. Screening Samples from donors and the proficiency testing program were screened prior to confirmation by LC MS MS. Screening of oral fluids was performed using an Olympus AU5400 with reagents and controls from Immunalysis. The screening cutoffs for AMPs, opiates, PCP, cocaine, and BE were 50, 40, 10, 20, and 20 ng/ml, respectively. HPLC All HPLC work was carried out using Shimadzu HPLC system equipped with dual pumps, autosampler, and a column heater. The separation of AMPs was carried out using a Pinnacle II C18 column ( mm, 5 µm, Restek, Austin, TX), whereas opiates, PCP, cocaine, and BE were separated on an Allure PFP propyl column ( mm, 5 µm). (For the details of the mobile phases, please see the captions of Figures 1 4.) MS MS analysis The detection and quantitation of the drugs was carried out using an Applied Biosystems 3200 QTRAP system. The turbo spray conditions were identical for analysis of all the drugs (curtain gas: 45 psi; collision gas: 8 psi; ion spray voltage: 5500 V; temperature: 65 C; ion source gas: 65 psi; ion source gas2: 70 psi; dwell time: 100 ms), and two multiple reaction monitoring (MRM) transitions were monitored for each drug. The details of MRM transitions, retention times, and collision energy used for each drug are given in Table I. The cutoff concentrations for AMPs, opiates, PCP, cocaine, and BE were set at 50, 40, 10, 20, and 20 ng/ml oral fluid according to the suggestions from RTI. Data analysis was carried out using Applied Biosystems Analyst Software (version 1.4.2). Ion-suppressive effects of oral fluid matrix were studied by analyzing water and oral fluid samples spiked with different drugs by LC MS MS. No ion suppressive effects of oral fluid matrix were found on any of the analytes studied.
3 Results and Discussion Linearity studies were initially conducted to verify the range at which AMPs (AMP, MAMP, MDA, MDMA), opiates (codeine, morphine, hydrocodone, hydromorphone, oxycodone, oxymorphone), PCP, cocaine, and BE can be detected using LC MS MS. Saliva or oral fluid is generally collected into a collection device containing a buffer solution. Immunalysis collection devices Figure 1. Ion chromatograms and linear regression curves of AMP, MAMP, MDA, and MDMA. Extracted ion chromatogram of 10 ng/ml amphetamines mixture in oral fluids generated by LC MS MS (A). Note that the signal-to-noise ratio for each of the compounds is > 4. The deuterated internal standard concentrations were 50 ng/ml. Linear regression curves were generated by plotting the data obtained from LC MS MS analysis of negative oral fluids spiked with different concentrations of AMPs (10 to 2000 ng/ml) (B). The linear regression coefficient (r) for AMPs in oral fluids was found to be > over the range of 10 to 2000 ng/ml. AMPs were separated on a Pinnacle II C18 column, using gradient of 5 mm ammonium acetate and methanol containing 0.1% formic acid. The flow rate was set at 1.2 ml/min and the run time was 5 min. Figure 2. Ion chromatograms and linear regression curves of opiates (codeine, morphine, hydrocodone, hydromorphone, oxycodone, and oxymorphone). Extracted ion chromatogram of 10 ng/ml opiate mixture in oral fluids generated by LC MS MS (A). Note that the signal-to-noise ratio for each of the compounds is > 4. The deuterated internal standard concentrations were 40 ng/ml. Linear regression curves were generated by plotting the data obtained from LC MS MS analysis of negative oral fluids spiked with different concentrations of opiates (1 to 500 ng/ml) (B). The linear regression coefficient (r) of opiates in oral fluids was found to be > over the range of 10 to 500 ng/ml. Opiates were separated on an Allure PFP propyl column using gradient of 0.1% formic acid in water and 80:20 methanol/acetonitrile containing 0.1% formic acid. The flow rate was set at 1.0 ml/min and the run time was 6 min. 607
4 were used for the validation of this method as well as to analyze the oral fluid samples from real donors. The Immunalysis oral fluid collection device (Quantisal) allows the collection of 1 ml of oral fluid and this is collected into 3 ml of buffer solution. During this process, the drug in the oral fluid is diluted 4 times. A In order to maintain the same dilution, diluted oral fluid samples were prepared and fortified with concentrations that would give a final concentration of the drug in oral fluid at near and around cutoff levels. The results are expressed as concentration of the drug in the original (undiluted) oral fluid. B Figure 3. Ion chromatograms and linear regression curves of cocaine and BE. Extracted ion chromatogram of 2 ng/ml cocaine and BE mixture in oral fluids generated by LC MS MS (A). Note that the signal-to-noise ratio for each of the compounds is > 4. The deuterated internal standard concentrations were 80 ng/ml. Linear regression curves were generated by plotting the data obtained from LC MS MS analysis of negative oral fluids spiked with different concentrations of cocaine and BE (1 to 1000 ng/ml) (B). The linear regression coefficient (r) of cocaine and BE in oral fluids was found to be > over the range of 1 to 1000 ng/ml. Cocaine and BE were separated on an Allure PFP propyl column using a gradient of 0.1% formic acid in water and 80:20 methanol/acetonitrile containing 0.1% formic acid. The flow rate was set at 1.0 ml/min, and the run time was 6 min. Figure 4. Ion chromatograms and linear regression curves of PCP. Extracted ion chromatogram of 2 ng/ml PCP in oral fluids generated by LC MS MS (A). Note the signal-to-noise ratio for each of the compounds is > 4. The deuterated internal standard concentration was 100 ng/ml. Linear regression curves were generated by plotting the data obtained from LC MS MS analysis of negative oral fluids spiked with different concentrations of PCP (1 to 500 ng/ml) (B). The linear regression coefficient (r) of PCP in oral fluids was found to be > over the range of 1 to 500 ng/ml. PCP was separated on an Allure PFP propyl column using gradient of 0.1% formic acid in water and 80:20 methanol/acetonitrile containing 0.1% formic acid. The flow rate was set at 1.2 ml/min, and the run time was 6 min. 608
5 Table I. MRM Transitions and Retention Times of Drugs of Abuse* Drug RT (min) MRM CE (V) Amphetamine ; , 13 Amphetamine-d ; , 13 Methamphetamine ; , 15 Methamphetamine-d ; , 17 MDA ; , 31 MDA-d ; , 31 MDMA ; , 33 MDMA-d , 35 Codeine ; , 79 Codeine-d ; , 93 Morphine ; , 51 Morphine-d ; , 79 Hydrocodone ; , 73 Hydrocodone-d ; , 97 Hydromorphone ; , 57 Hydromorphone-d ; ,75 Oxycodone ; , 39 Oxycodone-d ; , 41 Oxymorphone ; , 41 Oxycodone-d ; , 41 Cocaine ; , 41 Cocaine-d ; , 43 Benzoylecgonine ; , 41 Benzoylecgonine-d ; , 43 Phencyclidine ; , 17 Phencyclidine-d ; , 17 * MRM: multiple reaction monitoring; CE: collision energy; internal standards used in this study are shown in italics. For oxymorphone, oxycodone-d 6 was used as internal standard. Electrospray positive ionization has been used for the detection of all drugs of abuse in oral fluids in this study. Initially standard drug solutions of 200 ng/ml were infused into the MS MS system to optimize the conditions for each drug. Two product ions were selected for each drug and an MRM method was set up to monitor the MRM transitions of each drug during LC MS MS analysis (Table I). Oral fluids containing various concentrations of AMPs (amphetamine, MAMP, MDA, and MDMA) ranging from 10 to 2000 ng/ml final oral fluid concentration were prepared and extracted by liquid liquid extraction and the final extracts were analyzed by LC MS MS. Amphetamine-d 11, MAMP-d 11, MDAd 5, and MDMA-d 5 at a concentration of 50 ng/ml were used as the internal standards. The extracted ion chromatogram of 10 ng/ml AMP mixture is presented in Figure 1A. The linear regression for AMPs was found to be > The LOD/LOQ for all the AMPs was determined to be 10 ng/ml oral fluid. The details of HPLC separation of AMPs are given in the legend for Figure 1. Similarly, oral fluid samples containing various concentrations of opiates (codeine, morphine, hydrocodone, hydromorphone, oxycodone, oxymorphone) ranging from 10 to 500 ng/ml were extracted by solid-phase extraction and analyzed by LC MS MS. Codeine-d 3, morphine-d 3, hydrocodoned 3, hydromorphone-d 3, and oxycodone-d 6 at a concentration of 40 ng/ml were used as the internal standards. The extracted ion chromatogram and linear regression curves for each opiate is presented in Figure 2. The linear regression for opiates was > and LOD/LOQ for opiates was determined to be 10 ng/ml oral fluid. (Please see the caption for Figure 2 for the details of separation of opiates by HPLC). Oral fluid samples containing cocaine, BE ( ng/ml), and PCP (1 500 ng/ml) were analyzed by LC MS MS, and the data are presented in Figures 3 and 4. Cocaine-d 3 and BE-d 3 at a concentration of 80 ng/ml and PCP-d 5 at a concentration of 100 ng/ml were used Table II. LOD/LOQ, Precision, and Accuracy Study of Drugs of Abuse in Oral Fluids by LC MS MS* Name LOD/LOQ Precision (% CV) Accuracy (%) Amphetamine Methamphetamine MDA MDMA Codeine Morphine Hydrocodone Hydromorphone Oxycodone Oxymorphone Cocaine Benzoylecgonine Phencyclidine * LOD/LOQ values are expressed as ng/ml of oral fluid. Precision and accuracy studies were carried out by analyzing 10 samples for each drug at 40% of cutoff (20 ng/ml for AMPs, 16 ng/ml for opiates, 8 ng/ml for cocaine and BE, and 4 ng/ml for PCP). Table III. Verification of the Study by Participation in North American Oral Fluid Proficiency Testing* Target Mean of Conc. Participating Sample ID Analyte (ng/ml) Laboratories Range Result OFU-1 Amphetamine OFU-1 Methamphetamine OFU-2 Codeine OFU-2 Morphine OFU-4 Benzoylecgonine OFU-5 Cocaine OFU-5 PCP * North American Oral fluid Proficiency Testing was administered by RTI. The mean value represents the average of results from 17 to 22 laboratories. The results from this study were within 10% of the mean of the all laboratories participated in the proficiency program except for cocaine, where the results came within the 20% of the mean. 609
6 Figure 5. Analysis of oral fluid samples from real donors and proficiency samples. The ion chromatograms of AMP, cocaine, and BE were from real donor samples whereas opiates and PCP are from proficiency samples. as the internal standards. The linear regressions for these drugs were found to be > and the LOD/LOQ for cocaine, BE, and PCP were determined to be 2 ng/ml oral fluid (see details of HPLC separation in the corresponding figure legends). The precision and accuracy studies were conducted using analysis of 10 individual oral fluid samples spiked with 40% of respective cutoffs of drugs of interest, and the data are presented in Table II along with summary of LOD/LOQ. The percentage variation of precision for the drugs ranged between 3 and 10%, and accuracy was found to be within 20% of the target value. Figure 5 represents ion chromatograms of the analysis of either real samples or proficiency samples obtained from RTI. The ion chromatograms of positive samples for amphetamines, cocaine, and BE are presented as Figures 5A and B. For PCP and opiates, the ion chromatograms of the RTI proficiency samples are presented in Figures 5C and D. In this study, to date there were no positive oral fluid samples for PCP and opiates from real donors. The authors have participated in the voluntary proficiency program offered by RTI, and the results are presented in Table III. All the drugs in the proficiency samples were identified by this method, and the quantitative results were found to be within 20% of the target value, suggesting that the methods developed in our laboratory are valid and accurate for the detection and quantitation of drugs of abuse in oral fluids. In conclusion, the authors have developed an LC MS MS method for the detection and quantitation of AMPs, opiates, PCP, cocaine, and BE in oral fluids. This study represents the analysis of different classes of drugs in oral fluids by LC MS MS. Additionally, the analysis of donor samples along with the proficiency samples supplied by RTI confirms and verifies the validity of this method. Presently, studies of identification of other drugs of abuse including barbiturates and benzodiazapines by LC MS MS are in progress. Acknowledgments The authors thank Mr. Jeorge Reyna for his technical assistance. References 1. E.J. Cone, J. Clarke, and L. Tsanaclis. Prevalence and disposition of drugs and opioid treatment drugs in oral fluid. J. Anal. Toxicol. 31(8): (2007). 2. S. George. A snapshot of workplace drug testing in the UK. Occup. Med. (Lond.) 55: (2005). 3. E.J. Cone and A.J. Jenkins. Saliva drug analysis. In Handbook of Analytical Therapeutic Drug Monitoring and Toxicology. S.H.Y. Wong and I. Sunshine, Eds. CRC Press, Boca Raton, FL, 1997, pp M. Concheiro, A. de Castro, O. Quintela, and C.M. Lopez-Rivadulla. Confirmation by LC MS of drugs in oral fluid obtained from roadside testing. Forensic Sci. Int. 170: (2007). 610
7 5. E.J. Cone. Saliva testing of drugs of abuse. Ann. N.Y. Acad. Sci. 694: (1993). 6. E.J. Cone and M.A. Huestis. Interpretation of oral fluid tests for drugs of abuse. Ann. N.Y. Acad. Sci. 1098: (2007). 7. O.H. Drummer. Drug testing in oral fluid. Clin. Biochem. Rev. 27(3): (2006). 8. L. Kadehjian. Legal issues in oral fluid testing. Forensic Sci. Int. 150: (2005). 9. Y.H. Caplan and B.A. Goldberger. Alternative specimens for workplace drug testing. J. Anal. Toxicol. 25(5): (2001). 10. F.M. Wylie, H. Torrance, R.A. Anderson, and J.S. Oliver. Drugs in oral fluid Part I. Validation of an analytical procedure for licit and illicit drugs in oral fluid. Forensic Sci. Int. 150: (2005). 11. M. Mueller, F.T. Peters, G.A. Ricaurte, and H.H. Maurer. Validated liquid chromatographic electrospray ionization mass spectrometric assay for simultaneous determination of 3,4-methylenedioxymethamphetamine and its metabolites 3,4-methylenedioxyamphetamine, 3,4-dihydroxymethamphetamine and 4-hydroxy-3-methoxymethamphetamine in squirrel monkey plasma. J. Chromatogr. 855: (2007). 12. K.A. Mortier, K.E. Maudens, W.E. Lambert, K.M. Clauwaert, J.F. Van Bocxlaer, D.L. Deforce, C.H. Van Peteghem, and A.P. Leenheer. Simultaneous, quantitative determination of opiates, amphetamines, cocaine and benzoylecgonine in oral fluid by liquid chromatography quadrupole-time-of-flight mass spectrometry. J. Chromatogr. 779: (2002). 13. M. Wood, M. Laloup, M. del Mar Ramirez Fernandez, K.M. Jenkins, M.S. Young, J.G. Ramaekers, G. De Boeck, and N. Samyn. Quantitative analysis of multiple illicit drugs in preserved oral fluid by solid-phase extraction and liquid chromatography tandem mass spectrometry. Forensic Sci. Int. 150: (2005). 611
Modernizing the Forensic Lab with LC-MS/MS Technology
Modernizing the Lab with LC-MS/MS Technology Innovative SCIEX Analytical Tools for the Rapid Identification of Drugs of Abuse in Samples Oscar G. Cabrices 1, Holly McCall 1, Xiang He 1, Alexandre Wang
More informationA Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography
Application Note LCMS-109 A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography Time of Flight (LC-TOF) Mass Spectrometry Introduction Many clinical
More informationQuantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics
Quantitative Analysis of piates in Urine Using RRHT LC/MS/MS Application Forensics Authors Sheher Mohsin Agilent Technologies, Inc. 10 N. Martingale Rd., Suite 550 Schaumburg, IL 60173 USA Yanan Yang Agilent
More informationAmphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS
Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 643 LC/MS/MS Application Note Forensics Authors Jason Hudson, Ph.D., James Hutchings, Ph.D., and Rebecca Wagner, Ph.D. Virginia
More informationOverview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis
A Fast and Sensitive LC/MS/MS Method for the Quantification and Confirmation of 3 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine Samples Cliquid Software for Routine LC/MS/MS Analysis
More informationQuantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer
application Note Liquid Chromatography/ Mass Spectrometry Authors Sharanya Reddy Blas Cerda PerkinElmer, Inc. Shelton, CT USA Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate
More informationA Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS
A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS Introduction Drug analysis from whole blood is gaining popularity due to a more complete measurement
More informationDETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS
FORENSICS AND TOXICOLOGY ANALYSIS DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS Solutions for Your Analytical Business Markets and Applications
More informationHigh-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine
High-Throughput Quantitative LC-MS/MS Analysis of and 14 Benzodiazepines in Urine Bill Yu, Kristine Van Natta, Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 588 Key Words Opiates,
More informationA Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer
A Robustness Study for the Agilent 7 LC-MS/MS Mass Spectrometer Application Note Clinical Research Authors Linda Côté, Siji Joseph, Sreelakshmy Menon, and Kevin McCann Agilent Technologies, Inc. Abstract
More informationWorkflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Urine Using UHPLC-TOF
APPLICATIO OTE Liquid Chromatography/ Mass Spectrometry Authors: Avinash Dalmia Joanne Mather PerkinElmer, Inc. Shelton, CT Workflow for Screening and Quantification of the SAMHSA (IDA) Panel in Urine
More informationDetermination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.
Page 1 of 10 Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection EPL-BAS Method No. 205G881B Method Summary: Residues of 6-CPA are
More informationExtraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis
Application Note AN784 Extraction of Multiple Mycotoxins from Nuts Using ISOLUTE Myco Page 1 Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis This application note
More informationWorkflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Serum Using UHPLC-TOF
APPLICATIO OTE Liquid Chromatography/ Mass Spectrometry Authors: Avinash Dalmia Bonnie Marmor PerkinElmer, Inc. Shelton, CT Workflow for Screening and Quantification of the SAMHSA (IDA) Panel in Serum
More informationDetermination of Selected Illegal Drugs and its Important Metabolites in Waste Water by Large Volume Direct Injection HPLC-MS/MS
Determination of Selected Illegal Drugs and its Important Metabolites in Waste Water by Large Volume Direct Injection HPLC-MS/MS Jean-Daniel Berset 1 and André Schreiber 2 1 Water and Soil Protection Laboratory,
More informationDetermination of Amantadine Residues in Chicken by LCMS-8040
Liquid Chromatography Mass Spectrometry Determination of Amantadine Residues in Chicken by LCMS-8040 A method for the determination of amantadine in chicken was established using Shimadzu Triple Quadrupole
More informationApplication. Detection of Cannabinoids in Oral Fluid Using Inert Source GC/MS. Introduction. Authors. Abstract. Forensic Toxicology
Detection of Cannabinoids in Oral Fluid Using Inert Source GC/MS Application Forensic Toxicology Authors Christine Moore, Sumandeep Rana, and Cynthia Coulter Immunalysis Corporation 829 Towne Center Drive
More informationDirect Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology
Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation,
More information4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column
Clinical, Forensic & Toxicology Applications 4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column By Amanda Rigdon Abstract A rapid, sensitive method
More informationHigh Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)
High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS) Michael Rummel, Matthew Trass, Michael Campognone, and Sky Countryman Phenomenex, Inc., 411 Madrid Avenue, Torrance,
More informationAnalyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18
Analyze Barbiturates in Urine with Agilent 6 LC/MS/MS and Poroshell EC-C8 Application ote Forensic Toxicology Authors Elijah Steinbauer and Pat Friel Toxicology Laboratory at the Veterans Administration
More informationRapid and Robust Detection of THC and Its Metabolites in Blood
Rapid and Robust Detection of THC and Its Metabolites in Blood Application Note Forensics/Doping Control Author Stephan Baumann Agilent Technologies, Inc. Santa Clara CA 95051 USA Abstract A robust method
More informationLC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine
LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine J. Jones, S. Westwood, T. Liddicoat, L. Pereira, T. Edge Thermo Fisher Scientific, Manor Park, Runcorn, UK Overview Purpose:
More informationFast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS
TECHNICAL NOTE 21883 Fast and easy separation of 23 drugs of abuse including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS Authors Kean Woodmansey 1, Jon Bardsley 1 and Stacy
More informationExtraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis
Application Note AN782 Extraction of Multiple Mycotoxins from Grain Using ISOLUTE Myco Page 1 Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis This application
More informationAnalysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ
Application Note Clinical Research Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ Authors Jennifer Hitchcock 1, Lauren Frick 2, Peter Stone 1, and Vaughn Miller 2 1 Agilent
More informationHigh-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine
High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring and in Urine Xiaolei Xie, Joe DiBussolo, Marta Kozak; Thermo Fisher Scientific, San Jose, CA Application Note 627 Key Words, norbuprenorphine,
More informationQuantitative Analysis of Amphetamine-Type Drugs by Extractive Benzoylation and LC/MS/MS. Application. Introduction. Authors. Abstract.
Quantitative Analysis of Amphetamine-Type Drugs by Extractive Benzoylation and LC/MS/MS Application Forensics Authors Neil Campbell, B. Sc. Forensic Science Laboratory Chemistry Centre (WA) 5 Hay Street
More informationFast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system
Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system Xiang He 1, Adrian Taylor 2 and Alexandre Wang 1 1 SCIEX, Redwood City, USA. 2 SCIEX, Concord,
More informationComparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs
Seyed Sadjadi, Shahana Huq, Sean Orlowicz and Laura Snow Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA Comparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis
More informationExtraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis
Application Note AN807 Extraction of Aflatoxin M From Infant Formula Using ISLUTE Myco Page Extraction of Aflatoxin M From Infant Formula Using ISLUTE Myco SPE Columns prior to LC-MS/MS Analysis This application
More informationDienes Derivatization MaxSpec Kit
Dienes Derivatization MaxSpec Kit Item No. 601510 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION
More informationAchieving Proposed Federal Concentrations using Reduced Specimen Volume for the Extraction of Amphetamines from Oral Fluid
Achieving Proposed Federal Concentrations using Reduced Specimen Volume for the Extraction of Amphetamines from Oral Fluid Christine Moore, Cynthia Coulter, and Katherine Crompton Immunalysis Corporation,
More informationDetermination of Benzodiazepines in Oral Fluid using LC MS MS
Determination of Benzodiazepines in Oral Fluid using LC MS MS Technical Note Christine Moore 1,*, Cynthia Coulter 1, Katherine Crompton 1, and Michael Zumwalt 2 1 Immunalysis Corporation, 829 Towne Center
More informationDetermination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry
Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Application Note Food Safety Authors Chenhao Zhai Agilent
More informationApplication Note. Abstract. Authors. Pharmaceutical
Analysis of xycodone and Its Metabolites-oroxycodone, xymorphone, and oroxymorphone in Plasma by LC/MS with an Agilent ZRBAX StableBond SB-C18 LC Column Application ote Pharmaceutical Authors Linda L.
More informationA FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ
FORENSICS MARKET A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ Desomorphine, also known by its street name
More informationCannabinoid Quantitation Using an Agilent 6430 LC/MS/MS
Cannabinoid Quantitation Using an Agilent 643 LC/MS/MS Application Note Forensics Authors Jason Hudson, Ph.D., James Hutchings, Ph.D., and Rebecca Wagner, Ph.D. Virginia Department of Forensic Science
More informationA Novel Platform of On-line Sample Pre-treatment and LC/MS/MS Analysis for Screening and Quantitation of Illicit Drugs in Urine
PO-CON737E A Novel Platform of On-line Sample Pre-treatment and LC/MS/MS Analysis for Screening and Quantitation of Illicit Drugs in Urine ASMS 7 WP 353 Shao Hua Chia ; Zhi Wei Edwin Ting ; Daisuke Kawakami
More informationAbstract. Introduction
A Fast Dilute-And-Shoot Method for Simultaneous 5-Hydroxyindoleacetic Acid (5-HIAA), Vanillylmandelic Acid (VMA), and Homovanillic Acid (HVA) LC-MS/MS Analysis in Human Urine By Shun-Hsin Liang and Sue
More informationQuantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup
Application Note Forensic Testing Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup Author Limian Zhao Agilent Technologies, Inc.
More informationRapid Hydrolysis of Benzodiazepines in Urine. Alicia Zook 1 and Crystal Xander B.S. 2. Cedar Crest College, Allentown, PA 1
Rapid Hydrolysis of Benzodiazepines in Urine Alicia Zook 1 and Crystal Xander B.S. 2 Cedar Crest College, Allentown, PA 1 Health Network Laboratories, Allentown, PA 2 Abstract: Benzodiazepines are sedative/hypnotic
More informationComprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS
Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS SCIEX QTRAP 4500 LC-MS/MS System and Spark Holland PICO Adrian M. Taylor 1, Peter Ringeling 2, Martin Sibum 2, Stefan Sturm
More informationQuantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS
Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS Application Note Forensic Toxicology Authors Jason Hudson, Ph.D., James Hutchings, Ph.D., and Rebecca
More informationSupporting Information
Supporting Information Schlosburg et al. 10.1073/pnas.1219159110 SI Materials and Methods: Quantification of Heroin Metabolites Sample Collection. Trunk blood was collected in a 1:1 ratio with acetate
More informationDetermination of Bath Salts (Pyrovalerone Analogs) in Biological Samples
Determination of Bath Salts (Pyrovalerone Analogs) in Biological Samples Application Note Forensic Toxicology Authors Joe Crifasi Saint Louis University Forensic Toxicology Laboratory Saint Louis, Mo.
More informationA Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS
A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS By Shun-Hsin Liang and Frances Carroll Abstract Vitamin K₁ and K₂ analysis is typically complex and time-consuming because these
More informationVitamin D3 and related compounds by ESI and APCI
Liquid Chromatography Mass Spectrometry SSI-LCMS-9 Vitamin D and related compounds by ESI and APCI LCMS-8 Summary Vitamin D and related compounds were measured by LC-ESI/APCI-MS-MS. Background Accurate
More informationQuantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation
Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Valérie Thibert, Bénédicte Duretz Thermo Fisher Scientific, Courtaboeuf,
More informationA RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS
A RAPID AD SESITIVE AALYSIS METD OF SUDA RED I, II, III & IV I TOMATO SAUCE USIG ULTRA PERFORMACE LC MS/MS Choon Keow G, aomi TAAKA, Michelle KIM, Swee Lee YAP Waters Asia, Regional Technology Center,
More informationValidation of a Benzodiazepine and Z-Drug Method Using an Agilent 6430 LC/MS/MS
Validation of a Benzodiazepine and Z-Drug Method Using an Agilent 6430 LC/MS/MS Application Note Forensics Authors Jason Hudson, Ph.D., James Hutchings, Ph.D., and Rebecca Wagner, Ph.D. Virginia Department
More informationSUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM
SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM UNITED KINGDOM WOMEN DO NOT DECREASE OVER TWELVE MONTHS
More informationThe Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis
The Investigation of Factors Contributing to Immunosuppressant Drugs Variability in LC-MS/MS Analysis Joseph Herman, Dayana Argoti, and Sarah Fair Thermo Fisher Scientific, Franklin, MA, USA Overview Purpose:
More informationDetermination of Amphetamine and Derivatives in Urine
Determination of Amphetamine and Derivatives in Urine Using a Modified QuEChERS and Capillary Electrophoresis Tandem Mass Spectrometry Analysis Application Note Authors Vagner B. dos Santos and Claudimir
More informationQuantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system
Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 641 Triple Quadrupole LC/MS system Application Note Clinical Research Author Siji Joseph Agilent Technologies Bangalore, India
More informationAuthors. Abstract. Introduction. Forensics
Determination Of Opioids, Cocaine, and Cocaine Metabolites by Liquid Chromatography Mass Spectrometry Using ZORBAX Eclipse Plus C18 Columns Application Forensics Authors Patrick Friel B.S. and Ann Marie
More informationRapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases
Clinical, Forensic & Toxicology Applications Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases By Shun-Hsin
More informationSPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine
SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 709 Key Words SPE, SOLA
More informationDetermination of Opiates and Metabolites in Blood Using Electrospray LC/MS. Application Note
Deteration of piates and Metabolites in Blood Using Electrospray LC/MS Application Note Scott A. Schlueter and James D. Hutchison, Jr. Montana Department of Justice, Division of Forensic Science John M.
More informationUltrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System
Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System Application Note Forensic Toxicology Authors Nikunj R. Parikh, Michelle
More informationLC-MS/MS Method for the Determination of Tenofovir from Plasma
LC-MS/MS Method for the Determination of Tenofovir from Plasma Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 687 Key Words SPE, SOLA CX, Hypersil GOLD, tenofovir Abstract
More informationApplication Note. Author. Abstract. Introduction. Food Safety
Determination of β2-agonists in Pork with SPE eanup and LC-MS/MS Detection Using Agilent BondElut PCX Solid-Phase Extraction Cartridges, Agilent Poroshell 120 column and Liquid Chromatography-Tandem Mass
More information[application note] Simultaneous detection and quantification of D 9 THC, 11-OH-D 9 T H C and D 9 THC-COOH in whole blood by GC tandem quadrupole MS
Simultaneous detection and quantification of D 9 THC, 11-OH-D 9 T H C and D 9 THC-COOH in whole blood by GC tandem quadrupole MS Marie Bresson, Vincent Cirimele, Pascal Kintz, Marion Villain; Laboratoire
More informationOral Fluid Drugs of Abuse Testing
Solutions at Work Oral Fluid Drugs of Abuse Testing Volume indicator turns blue when collection is complete (average 3 minutes) Non-invasive collection test anytime, anywhere Directly observed, no gender
More informationLC/MS/MS Analysis of Metabolites of Synthetic Cannabinoids JWH-018 and JWH-073 in Urine
Abstract A liquid chromatographic method was developed to resolve a comprehensive set of metabolites of JWH-018 and JWH-073. In addition to the chromatographic analysis method, an extraction method was
More informationConfirmation by Triple Quadrupole LC/MS/MS for HHS-compliant Workplace Urine Drug Testing
Confirmation by Triple Quadrupole LC/MS/MS for HHS-compliant Workplace Urine Drug Testing A complete applications solution for forensic toxicology laboratories John Hughes Agilent Technologies Pleasanton,
More informationAdvancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations
Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations For Research Use Only. Not for use in diagnostic procedures. 1 2015 AB Sciex RUO-MKT-11-1018-A For research
More informationSimple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID
Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID James Little, Eastman Chemical Company, Kingsport, TN Overview Phospholipids and
More informationIntegration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation
PO-CON69E Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation MSACL 6 EU Stéphane Moreau, Daisuke Kawakami, Toshikazu Minohata Shimadzu Europe GmbH, Duisburg,
More informationAll stocks and calibration levels were prepared in water: methanol (50:50) v/v to cover range of all steroid concentrations (refer Table 1).
Application LCMS-8040 Simultaneous determination of 11 steroids and Vitamin D2/D3 in human serum using LC/MS/MS - Introduction Quantification of endogenous hormonal steroids and their precursors is essential
More informationRobust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS
TECHNICAL NOTE 21882 Robust extraction, separation, and quantitation of structural isomer steroids human plasma by SPE-UHPLC-MS/MS Authors Jon Bardsley 1, Kean Woodmansey 1, and Stacy Tremintin 2 1 Thermo
More informationLC/MS Method for Comprehensive Analysis of Plasma Lipids
Application Note omics LC/MS Method for Comprehensive Analysis of Plasma s Authors Tomas Cajka and Oliver Fiehn West Coast Metabolomics Center, University of California Davis, 451 Health Sciences Drive,
More informationValidation of an Automated Method to Remove
Shahana Wahab Huq 1, Richard Thomas 2, Agnes Cua 2, Seyed Sadjadi 1 1 Phenomenex, 411 Madrid Avenue, Torrance, CA 951 2 Precision Toxicology, 33 Bunker Hill Street, San Diego, CA 9219 Validation of an
More informationMaria João Mendes Mendonça Barreira Department of Health Promotion and Chronic Deseases, INSA
Quantification of four immunosuppressant drugs by Liquid Chomatography-Tandem Mass Spectrometry using Direct Injection. Maria João Mendes Mendonça Barreira Department of Health Promotion and Chronic Deseases,
More informationAuthor. Introduction. Abstract
Simultaneous Measurement of Sirolimus and Tacrolimus Concentrations in Blood by Semi-Automated Extraction and Liquid Chromatography-Electrospray Ionization Mass Spectrometry Application Author Gary L.
More informationNeoSal Oral Fluid Collection System Solutions for Forensic Drug Detection
PERFORMANCE OF THE NEOSAL ORAL FLUID COLLECTION SYSTEM Neogen develops and manufactures a comprehensive range of ELISA test kits and accessories for forensic drug detection. Neogen offers solutions to
More informationHsiu-Chuan Liu 1, Hsi-Tzu Lee 1, Ya-Ching Hsu 2, Mei-Han Huang 2, Ray H. Liu 3, Tai-Jui Chen 4 and Dong-Liang Lin 1 *
Journal of Analytical Toxicology 2015;39:472 480 doi:10.1093/jat/bkv041 Advance Access publication May 2, 2015 Article Direct Injection LC MS-MS Analysis of Opiates, Methamphetamine, Buprenorphine, Methadone
More informationAnalysis of drugs of abuse in biological matrix using Time of Flight technology
Analysis of drugs of abuse in biological matrix using Time of Flight technology Niclas Stephanson, PhD Department of Clinical Pharmacology Karolinska University Hospital Karolinska Institutet Stockholm,
More informationDetection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System
Application Note Forensics, Workplace Drug Testing Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System Authors Fred Feyerherm and Anthony Macherone Agilent Technologies, Inc.
More informationMS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine
Application Note: 346 MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Gargi Choudhary and Diane Cho, Thermo Fisher Scientific, San Jose, CA Wayne Skinner and
More informationScreening by immunoassay and confirmation & quantitation by GC-MS of buprenorphine and norbuprenorphine in urine, whole blood and serum
Screening by immunoassay and confirmation & quantitation by GC-MS of buprenorphine and norbuprenorphine in urine, whole blood and serum NINA KANGAS, SIRPA MYKKÄNEN, SANNA KYLLÖNEN, PÄIVI RAJALA, KARI ARINIEMI
More informationAuthors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.
SAMHSA-Compliant LC/MS/MS Analysis of 11-nor-9-carboxy-D 9 - Tetrahydrocannabinol in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120 Application Note Forensic Toxicology Authors Irina
More informationValidation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens
Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens This document describes the validation of a Neogen Fentanyl kit for the semi-quantitative analysis of
More informationExtraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS
Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS UCT Part Numbers SSHLD063 Styre Screen HLD 60 mg, 6 ml column SPHPHO6001-5 Select ph Buffer Pouches 100 mm Phosphate
More informationRapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition
Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition Evelyn Goh Waters Pacific, Singapore APPLICATION BENEFITS This method allows for the simultaneous analysis of 12 water-soluble
More informationApplication Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer
Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer Abstract This study demonstrates a sensitive, rapid and reliable research
More informationMeasuring Lipid Composition LC-MS/MS
Project: Measuring Lipid Composition LC-MS/MS Verification of expected lipid composition in nanomedical controlled release systems by liquid chromatography tandem mass spectrometry AUTHORED BY: DATE: Sven
More informationUPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes
UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes Catalin E. Doneanu, Weibin Chen, and Jeffrey R. Mazzeo Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B E N E F
More informationDetection of Cotinine and 3- hydroxycotine in Smokers Urine
Detection of Cotinine and 3- hydroxycotine in Smokers Urine Behavioural and Situational Research Group School of Medicine, University of Tasmania Version number: 2 Effective date: 01/12/2015 Review due:
More informationSurviving Matrix Effects Experiments. Grace van der Gugten St. Paul s Hospital, Vancouver, BC, Canada
Surviving Matrix Effects Experiments Grace van der Gugten St. Paul s Hospital, Vancouver, BC, Canada gvandergugten@providencehealth.bc.ca Outline What are matrix effects Post Column Infusion Phospholipid
More informationQuantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System
Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Marta Kozak Clinical Research Applications Group Thermo Fisher Scientific San Jose CA Clinical Research use only, Not for
More informationFig. 1: Chemical structure of arachidonic acid COOH CH 3
Elimination of Matrix Effects Using Mixed-mode SPE Plate for High Throughput Analysis of Free Arachidonic Acid in Plasma by LC-MS/MS Wan Wang, Suzi Qin, Linsen Li, Warren Chen, Jerry Wang 179, Southern
More informationDetermination of Gamma-Hydroxy-Butyrate (GHB) in Biological Samples
Determination of Gamma-Hydroxy-Butyrate (GHB) in Biological Samples Application Note Forensic Toxicology Authors Joe Crifasi Saint Louis University Forensic Toxicology Laboratory Saint Louis, MO, USA Ron
More informationThe Drug Testing Process. Employer or Practice
Disclosures Clinical Professor, Jefferson Medical College BOD MROCC [Medical Review Officer Certification Council] BOD National Sleep Foundation BOD POEMS [Pennsylvania Occupational & Environmental Medicine
More informationPaul D. Rainville Ph.D. Health Sciences Group Waters Corporation Waters Corporation 1
A new level of sensitivity and standardization: The promise of integrated micro fluidic devices coupled with mass spectrometry for the analysis of biofluids Paul D. Rainville Ph.D. Health Sciences Group
More informationThe Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples
WA20714 The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples Michael S. Young and Kevin M. Jenkins Waters Corporation, 34 Maple Street, Milford, MA
More informationQuantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup
Application Note Forensic Testing Quantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup Author Limian Zhao Agilent Technologies, Inc. Abstract
More informationAnalysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS
Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS Food Application Author Fadwa Al-Taher Institute for Food Safety and Health Illinois Institute of Technology
More information