P rotoundly deaf patients benefit from multichannel cochlear implants.

Size: px
Start display at page:

Download "P rotoundly deaf patients benefit from multichannel cochlear implants."

Transcription

1 Christian Czerny1 Erich Steiner1 Wolfgang Gstoettner2 Wolf-Dieter Baumgartner2 Herwig lmhof1 Received February 20, 1997; accepted after revision May 12, Department of Radiology ) Division of Osteology), University Hospital of Vienna, Waehringer Guertel 18-20,1090Vienna, Austria. Address correspondence to C. Czerny. 2Department of Otolaryngology, University Hospital of Vienna, 1090 Vienna, Austria. AJR 1997:169: X/97/ American Roentgen Ray Society Postoperative Radiographic Assessment ofthe Combi 40 Cochlear Implant OBJECTIVE. The aims of this study were to establish a plain radiographic technique for the assessment of the postoperative appearance. position. and insertion depth of the Combi 40 cochlear implant and to correlate the radiologic findings with surgical reports. SUBJECTS AND METHODS. In an experimental study. an electrode of the Combi 40 device was inserted into the cochlea of a cadaveric skull. Digital radiographs were obtained in a modified Chausse III projection. in which the skull was placed supine on the radiography table with the infraorbitomeatal line strictly perpendicular to the film cassette. The skull was then rotated 30#{176} away from the side to be examined. and the central X-ray beam was angled I 5#{176} cephalad to the infraorbitomeatal line. On these radiographs. the point of cochleostomy was marked by a needle tip and was projected inferior to the vestibule and on a line drawn through the supenor semicircular canal and the vestibule. The appearance and position of the electrode was evaluated. An electrode was defined as completely inserted if all electrode contacts projected medial to the line drawn through the superior semicircular canal and the vestibule. We also studied cochlear implant insertion ofthe Combi 40 device in 37 patients. Postoperative digital radiographs of these patients were obtained and analyzed for the criteria as defined in the cadaveric study. In addition. the insertion depth of the electrode and the angle of insertion were measured on the radiographs. This depth was correlated with depth of insertion as estimated at surgery. RESULTS. The cadaveric study showed that the completely inserted electrode was seen on radiographs as a nonoverlapping spiral within the cochlea. All electrode contacts projected medial to the line drawn through the superior semicircular canal and the vestibule. In all 37 patients. the electrode could be seen without overlapping. According to our criteria. a completely inserted electrode was seen in 32 patients. In these patients. the insertion depth ranged from 21 to 34 mm and the angle of insertion ranged from 350#{176} to 810#{176}.In two patients. we saw a completely inserted electrode with a bend. In three patients. an incompletely inserted electrode was seen. Excellent correlation existed between the radiologic and surgical results with regard to insertion depth (r =.92). CONCLUSION. Digital radiographs obtained in the modified Chausse III projection allow clear depiction of the electrode and avoid overlapping. Such radiographs enable a reliable and accurate assessment of the position and insertion depth of the electrode of this new cochlear implant. Such images can serve as a baseline tir further radiographic examinations when extrusion or slippage of the electrode is clinically suspected. P rotoundly deaf patients benefit from multichannel cochlear implants. which considerably improve auditory performance I I The multichannel cochlear implant prosthesis consists of the receiver-stimulator. which is implanted in the postauricular region. and the electrode array. which is inserted into the scala tympani through a cochleostomy drilled ilito the cochlear promontory. Typically. the electrode is inserted into the basal turn of the cochlea. However, insertion in the apical regions of the cochlea is advantageous because it allows better stimulation of surviving neural elements in the deep-frequency range 131- Postoperative or intraoperative radiologic evaluation of intracochlear electrode position is a standard procedure in cochlear implant surgery. Postoperative assessment of cochlear implants usually is perfornied using con- AJR:169, December

2 Czerny et al. ventional radiographic techniques and CT [4-8j. Deep insertion of the electrode of the Nucleus device (Cochlear Corporation, Melbourne, Australia), as measured on radiographs, was approximately as deep as 25 mm, or 540#{176}. These data were correlated with audiologic tests, and it was suggested that deeper insertions of the electrode might be of potential benefit to the patient [3, 5]. In 1994, a new type of cochlear implant (Combi 40 device; MED-EL, Innsbruck, Austria) was introduced, designed to provide deeper insertions into the cochlea than was possible with other devices [3, 9]. For accurate radiologic assessment of deeply inserted electrodes without overlapping in one projection, the central X- ray beam should be perpendicular to the turns of the electrode with respect to the spiral turn of the cochlea. Therefore, a radiographic projection in which the central X-ray beam is angled perpendicular to the turns of the cochlea or parallel to the axis of the modiolus would be more appropriate for the accurate assessment of this new type of electrode [10-12]. To our knowledge, reports of the radiologic appearance, assessment of electrode position, and estimation of insertion depth of this new cochlear implant have not been published. The purposes of this study were to establish a plain radiographic technique for the assessment of this new cochlear implant, to present the postoperative radiologic appearance of the electrode, to assess the electrode position, to determine the depth of electrode insertion, and to correlate the radiographic findings with the surgical reports. Subjects and Methods Experimental Study To obtain a plain radiograph of the electrode of the Combi 40 cochlear implant in one projection and to assess the position of the electrode without overlapping and by which the point of cochleostomy was projected inferior to the vestibule, an original Combi 40 electrode was inserted into the cochlea of a cadaveric skull. The electrode insertion was performed by the same otolaryngologist with the cochleostomy drilled in the same location used in the patient study described later. The point of cochleostomy was marked by a needle tip. The electrode of this cochlear implant consists of eight pairs of contacts arranged in a twin-surface configuration for eight-channel stimulation. The distribution range covered by the eight pairs of contacts is 20.6 mm, and the last contact is located on the tip of the electrode. First, digital radiographs were obtained in the anteroposterior and lateral view, in the Stenver view, and in the view proposed by Marsh et al. [5]. Then, digital radiographs in a projection similar to the Chausse Ill view were obtained with the following settings [10-12]: The skull was positioned supine on the radiography table with the infraorbitomeatal line (the line connecting the infraorbital rim and the supenor aspect of the external auditory meatus) strictly perpendicular to the film cassette. The skull was then rotated 30#{176} away from the side to be examined. The central X-ray beam was angled 15#{176} cephalad to the infraorbitomeatal line and then centered to the middle ofthis line (Fig. 1). The conus was set such that the orbit and, particularly, the lens were not in the imaged area. The projection, which was the most suitable for the depiction of the intracochlear electrode without overlapping and in which the point of cochleostomy was projected inferior to the vestibule and on a line drawn through the superior semicircular canal (SSC) and the vestibule, was chosen by two radiologists in consensus and performed in the patient study. Patient Study In 37 patients, postoperative digital radiographs of the Combi 40 electrode in the modified Chausse III projection were performed using settings identical to those described in the cadaveric study. All patients (16 male patients, 2 1 female patients; mean age, years; range, 2-76 years) underwent implantation of the Combi 40 device by the same otolaryngologist. One of these patients underwent implantation of an older type of device outside our otolaryngology department. A complete insertion of the electrode was defined surgically if all electrode contacts had passed through the cochleostomy. An incomplete insertion of the electrode was defined surgically if one or more electrode contacts had not passed the Central X-ray beam Film 300 cassette Midsagittal line A cochleostomy. The depth of insertion of the electrode was estimated surgically by counting the number of contacts that passed through the cochleostomy. Deeper insertion was estimated surgically by measuring the distance from the first contact to a marker located 30.8 mm from the tip of the electrode. This marker did not prevent an insertion deeper than 30.8 mm. Image analysis of all digital radiographs was performed by two radiologists in consensus who evaluated the appearance of the electrode. the position of the electrode with respect to complete or incomplete insertion. the number of turns of the electrode within the spiral of the cochlea. the insertion depth in millimeters, and the angle of insertion. As defined in the cadaveric study, the point of cochleostomy was projected on a line drawn through the SSC and the vestibule on the radiographs obtained in the modified Chausse III projection (Figs. 2-4). An insertion was defined as complete if all electrode contacts could be depicted medial to this line. An insertion was defined as incomplete if one or more contacts were projected on this line or lateral to this line. If an electrode could not be visualized within the cochlea, it was defined as a misplaced inserted electrode. Other complications, such as a bend in the electrode, were also noted. In addition, the insertion depth of the electrode was assessed by counting the turns of the electrode within the cochlea. The depth of insertion (in millimeters) of the electrode was determined by measuring the distance from the line drawn through the SSC and the vestibule to the first contact of the electrode and adding this distance to the distance from the first contact to the last contact. The angle of insertion (in degrees) of the electrode within the cochlea was determined in the following way (Fig. 4): A line was drawn parallel to the course of the basal Central X-ray beam Film cassette Fig. 1.-Drawings show tube setting in relation to patients head, which is in supine position with infraorbitomeatal line strictly perpendicularto film cassette. A, Head is rotated 300 away from the side to be examined. B, Central X-ray beam is angled 15#{176} cephalad to infraorbitomeatal line AJR:169, December 1997

3 Radiographic Assessment of a Cochlear Implant Fig. 2.-Skull of cadaver, from 75-year-old man. Digital radiograph of skull with inserted electrode reveals that point of cochleostomy is marked by tip of needle (straight solid arrow) and is projected inferior to vestibule (arrowhead) on line (curvedarrow) drawn through superior semicircular canal and vestibule. All electrode contacts can be seen medial to cochleostomy. indicating completely inserted electrode (open arrow). Fig. 3.-Drawing corresponding to Figure 2 shows superior and lateral semicircular canal, vestibule, and cochlea with completely inserted electrode. Point of cochleostomy is seen inferior to vestibule and projected on line drawn through superior semicircular canal and vestibule. Fig. 4.-Digital radiograph of 2-yearold boy reveals completely inserted electrode (large straight arrow) without overlapping. Electrode contacts are delineated as small rectangular structures (open arrow). Point of cochleostomy (small straight arrow) is visualized inferior to vestibule and on line (curved arrow) drawn through superior semicircular canal and yestibule. Lines (arrowheads) for measurement of angle of insertion of electrode are drawn. Insertion depth of electrode is 28 mm and angle of insertion is 576#{176}. turn with respect to the electrode within the basal turn, which presented as a horizontal line. A second line was then drawn through the virtual apex of the cochlea perpendicular to the line, which was parallel to the basal turn. A third line was drawn perpendicular to the last contact and through the line running through the apex of the cochlea. The angle between these two lines was measured and added to 360#{176}, 540#{176},or 720#{176} if the last contact was beyond the line drawn through the apex of the cochlea. The angle was subtracted from 360#{176}. 540#{176}, or 720#{176} if the last contact was before that line (Fig. 4). The radiographic findings concerning a completely or incompletely inserted electrode were compared with the surgical reports. The radiographically measured insertion depth was compared with the surgically estimated insertion depth. For statistical evaluation. the Pearson conelation coefficient was used. Results On the radiographs obtained in the modifled Chausse III projection, the point of cochleostomy for the insertion of the electrode was projected inferior to the vestibule in the cadaveric study and in all patients. As defined previously, the point of cochleostomy could be visualized inferior to the vestibule and was projected on a line drawn through the superior semicircular canal and the vestibule (Figs. 2-4). In this projection, a completely inserted electrode was depicted in the basal turn as a horizontal line and as a spiral within the turns of the cochlea without overlapping of the electrode and with all eight contacts projected medial to the point of cochleostomy with respect to the line drawn through the SSC and the vestibule (Figs. 2-4). The contacts of the electrode could be depicted as small rectangular structures (Fig. 4). In the other projections performed in the cadaveric study we noted an overlapping of the electrode. and these projections were not performed in the patient study. A complete insertion of the electrode within the cochlea was seen on digital radiographs in 32 patients. and in five of these patients even the twin-surface configuration of the first two pairs of electrode contacts could be identified (Fig. 5). In 32 patients. the completely inserted intracochlear electrode was turned approximately one to 2.25 times. In these patients, the depth of insertion of the electrode varied between 2 1 and 34 mm, and the angle of insertion varied between 350#{176} and 810#{176} (Figs. 4-6). Point of cochleoston y Among patients with a completely inserted electrode with the same insertion depth measured on the radiographs, the angle of insertion varied (e.g., insertion depth = 21 mm, range of angle = #{176}) (Fig. 7). In one of the patients with a completely inserted Combi 40 electrode, parts of an older type of electrode from a previous operation performed outside our otolaryngology department were projected into the vestibule (Fig. 8). In two patients. the electrode was completely inserted. but a bend in the electrode outside the contact range of the electrode was depicted (Fig. 9). Three other patients showed an incompletely inserted electrode, less than 2 1 mm deep. with one electrode contact located in the cochleostomy in each patient (Fig. 10). Two of these patients had a maximum angle of insertion of 340#{176},and one of these patients had an angle of insertion of 280#{176}. All radiographic findings concerning a complete or incomplete insertion of the electrode were confirmed by the surgeon s report. We found no discrepancies between the surgical reports and the radiologic findings concerning completely or incompletely inserted electrodes. The radiologically evaluated insertion depth of the electrode showed excellent correlation with the surgically estimated insertion depth (r=.92)(fig. 11). Discussion The postoperative radiographic assessment ofthe stimulating electrode is important for all implant patients, to document the correct intracochlear electrode placement and to determine the depth of insertion [4, 5]. This assessment confirms the success of the surgical procedure and can serve as a baseline for further control in the event of complications, such as extrusion of the electrode, which would cause a worsening of function. Furthermore, it has AJR:169, December

4 ... Czerny et al. Fig. 5.-Digital radiograph of 65-year-old man reveals completely inserted electrode (arrow) without overlapping, with insertion depth of 22 mm and angle of insertion of 560#{176}. Twin configuration of first two pairs of electrode contacts (arrowheads) is shown. Depthoflnseition 35 (mm) as measured on rac5ographs i*..ti. ----,.4 : Ane otinsert,on Idegrees) as measured on radiographs Fig. 7.-Scatter diagram shows that various angles of insertion measured on radiographs correspond to one insertion depth in millimeters as measured on radiographs. 800 I,1.,... Fig. 6.-Digital radiograph of 66-year-old man reveals completely inserted electrode (arrow) without overlapping and with 2.25 turns of electrode within cochlea. Insertion depth of electrode is 34 mm and angle of insertion is 810#{176}. Fig. 8.-Digital radiograph of 44-year-old man reveals completely inserted electrode (arrow) with insertion depth of 28 mm and angle of insertion of 570#{176}. Parts of older type of electrode (arrowhead) resulting from previous operation performed outside our otolaryngology department are visualized in vestibule. Fig. 9.-Digital radiograph of 5-year-old girl reveals completely inserted electrode (arrow) with angle of insertion of 370#{176}. Bend in electrode (arrowhead) outside stimulating contacts is revealed. Fig. 10.-Digital radiograph of 37-year-old woman reveals incompletely inserted electrode (arrow) with one electrode contact (arrowhead) projected in cochleostomy. Angle of insertion of electrode is 340#{176} AJR:169, December 1997

5 Radiographic Assessment of a Cochlear Implant Depth of insertion (mm) as measured at surgery , Depth of insertion (mm) as measured on radiographs Fig. 11.-Scatter diagram shows correlation between insertion depth of electrodes measured on radiographs and at surgery (r=.92). been suggested that deeply inserted electrodes might influence postoperative hearing success because more deeply inserted electrodes result in lower electrical thresholds and better auditory performance [2, 5]. Previous studies assessed the Nucleus multichannel cochlear prosthesis by using conventional radiography either in the anteroposterior and lateral view, in the Stenver view, or in a so-called modified Stenver view [4, 5]. A plain radiographic technique for the assessment of the Combi 40. which has been shown to provide deep intracochlear insertion, has not yet been established [3]. In addition, to our knowledge, detailed postoperative evaluation of the radiographic appearance and of the position of electrodes of the Combi 40 device has not yet been reported in the literature, and a correlation between the radiologically measured insertion depth of this electrode and surgically estimated insertion depth has not been performed. We performed digital radiography in our study because it offers several advantages compared with conventional radiography. First, the dose-response characteristics are perfectly linear and thus subject contrast and density are captured with the same efficiency from low to high doses. Second, the stability of contrast and density in this dose range provide an improvement in image uniformity, and repeated examination because ofexposure errors is thus eliminated. Third, postprocessing of the images can enhance structures of diagnostic importance, such as the superior semicircular canal, the vestibule, and the electrode [13-15]. These advantages enabled an easier assessment of the Combi 40 electrode and of the electrode position. In addition, the use of a small conus combined with digital radiography and only one projection has the advantage of dose reduction as well as the exclusion of the lens from the radiated field. All of these advantages proved to be beneficial, particularly when children were examined. The Combi 40 electrode seems to offer the possibility of a deeper insertion compared with the Nucleus implant on the basis of our results, as compared with the results of a previous study [5]. As mentioned, deeper insertion of the electrode might result in better auditory performance [2, 3, 5]. In our patients with completely inserted electrodes, a maximum insertion of 34 mm and 8 10#{176} was found, although in a previous study deeply inserted electrodes with an insertion depth of 25 mm or 540#{176} were described [5]. In our opinion, this cannot be explained by our different imaging technique but may be the result of the mechanical properties of the electrode and of the surgical technique of our otolaryngologist who performed endoscopy of the cochlea before insertion of the electrode [9]. For the radiographic evaluation of the position and insertion depth of the Nucleus implant, a Stenver view or a modified Stenver view was used, as described in previous studies [4, 5]. However, these projections were not suitable for the radiographic assessment of the Combi 40 electrode because the electrode was overlapping, which made the accurate evaluation of electrode position and estimation of insertion depth difficult. The overlapping of the electrode in those projections might be explained by the fact that the spiral turns of the cochlea with respect to the electrode were not perpendicular to the central X-ray beam in those projections, and therefore other authors also proposed a so-called Chausse III projection for the best depiction of the spiral turns of the cochlea [10-12]. However, by obtaining digital radiographs in the cadaveric and patient study in our modified projection, we found the electrode of the Combi 40 device was not overlapping and was best depicted in all patients. In this projection, the twin-surface configuration of the first two pairs of electrode contacts could be identified in five patients. An overlapping of the electrode as noted in the other projections might have made a bend in the electrode invisible. Bends were noted in two patients in our study. The clear depiction of the intracochlear electrode without overlapping confirms the fact that the central X-ray beam might have been perpendicular to the intracochlear electrode in our proposed projection. In addition, the point of cochleostomy was always projected inferior to the vestibule and on the line drawn through the SSC and the vestibule in this projection, which offered a consistent landmark for determining the electrode position and insertion depth. This point was also confirmed by the fact that we found no discrepancies between the surgical reports and our radiologic findings concerning a complete or incomplete insertion of the electrode. Therefore, this projection seems to be reliable for the evaluation of the position and insertion depth of the electrode and for the depiction of a bend in the electrode, which was reported to be accurately revealed on CT only [4]. The visibility of the entire electrode without overlapping and the possibility of counting the contacts medial to the point of cochleostomy with respect to the line drawn through the SSC and the vestibule offered a simple method for evaluating complete or incomplete electrode insertion and for counting the number of turns of the electrode within the cochlea. The excellent correlation between the radiographically measured insertion depth and the surgically estimated insertion depth might result in the consistent projection of the cochleostomy inferior to the vestibule on all radiographs. In contrast to our results, Marsh et al. [5] found a discrepancy between the radiologically and surgically measured insertion depths. This discrepancy might be explained either by our projection, on which the point of cochleostomy was always projected inferior to the vestibule, or by the fact that our surgeon implanted the electrode keeping the insertion site in the same location in all patients. However, measuring the angle of insertion might be more useful for the prediction of auditory performance than determining the insertion depth in millimeters [5]. This angle could be determined by drawing lines as proposed in AJR:169, December

6 Czerny et al. this study. As shown on the radiographs, the basal turn with respect to the electrode in the basal turn was depicted as a horizontal line. In our opinion, this line could be taken as a reference line or baseline for measuring the angle of insertion because evaluating the angle from this line reflects precisely how many turns of the cochlea were covered by the electrode. The angle of insertion reflects more precisely the position of the electrode tip and the proximity of the electrode tip to the apex of the cochlea than the inserted distance measured in millimeters. In addition, we have found that, in patients with a completely inserted electrode of 21 mm, for example, the angle of insertion varied between 350#{176} and 450#{176}, which might be because of the fact that the electrode may not always lie at the outer wall of the scala tympani [3]. Therefore, it might be more useful to conelate the angle of insertion with audiologic tests than with the insertion depth in millimeters [5]. In conclusion, the findings of our study show that the appearance, position, and insertion depth of the electrode with no overlapping could be reliably and accurately assessed using our plain radiographic technique. Our proposed projection and method for the assessment of electrode position and for the evaluation of the depth and angle of insertion was precise and simple. It may serve as a baseline for further control should complications arise, such as extrusion or slippage of the electrode. CT evaluation of electrode position, location, and insertion depth might be unnecessary, which would decrease the radiation dose and the cost of examination. Therefore, the postoperative assessment of cochlear implants with this projection has become a standard procedure in our institution and is now performed intraoperatively on request of the surgeon, to offer the possibility of changing the insertion depth of the electrode. References 1. Burian K. Hochmair-Desoyer IJ. Eisenwort B. The Vienna cochlear implant program. Otolaryngol Clin NorthAm 1986;19: Clark AO. The University of Melboume/Cochlear Corporation (Nucleus) Program. Otolarvngol Clipi NorthAin 1986:19: Gstoettner W. Plenk H, Franz P. et al. Cochlear implant deep electrode insertion: extent of insertional trauma. Ada Otolarvngol 1997;l 17: Shpizner BA, Holliday RA, Roland JT, Cohen NL, Waltzman SB. Shapiro WH. Postoperative imaging of the multichannel cochlear implant. AJNR 1995:16: Marsh MA, Xu J, Blamey PJ, et al. Radiologic evaluation of multichannel intracochlear implant insertion depth. Am J Otol 1993:14: Rosenberg RA. Cohen NL, Reede DL. Radiographic imaging for the cochlear implant. Ann Otol Rhino! Laryngo! 1987;96: Ball JB Jr. Miller GW, Hepfner ST. Computed tomography of single-channel cochlear implants. AJNR 1986:7: Mukherji 5K, Mancuso AA, Kotzur IM, et al. CT of the temporal bone: findings after mastoidectomy, ossicular reconstruction, and cochlear implantation. AiR 1994:163: Gstoeuner W, Baumgartner WD, Franz P. Hamzavi J. Cochlear deep insertion surgery. Laryngoscope 1997;107(in press) 10. Brusis T, Moedder U. HNO Roentgen-A ujhahmetechnik und Norma!befunde. Berlin: Springer- Verlag. 1984: Zonnefeld FW. Computed tomography ofthe tempora! bone and orbit: technique ofdirect muftip!anar high-resolution CT and correlative cryosectional anatomy. Baltimore: Urban and Schwarzenbezg, 1987: Dimopoulos P. Muren C. Anatomic variations of the cochlea and relations to other temporal bone structures. Acta Radio! 1990;3 l(fasc. 5): Sonoda M, Tanako M, Miyahara J. Kato H. Computed radiography utilizing scanning laser stimulated luminescence. Radiology 1983;148: Merrin CRB, Tutton RH, Bell KA, et al. Clinical application of digital radiography: computed radiographic imaging. RadioGraphics 1985:5: BaIter S. On the work of the radiologist: separation of image capture from image display. Acta Radio! 1988:29: AJR:169, December 1997

ORIGINAL ARTICLE. new implantation method, the combined electric-acoustic stimulation

ORIGINAL ARTICLE. new implantation method, the combined electric-acoustic stimulation ORIGINAL ARTICLE Predicting Basal Cochlear Length for Electric-Acoustic Stimulation Oliver Adunka, MD; Marc H. Unkelbach, MD; Martin G. Mack, MD, PhD; Andreas Radeloff, MD; Wolfgang Gstoettner, MD, PhD

More information

Electrode Trauma Assessed by Microdissection

Electrode Trauma Assessed by Microdissection Electrode Trauma Assessed by Microdissection Peter S Roland MD C Gary Wright PhD University of Texas Medical Center Dallas, Texas Why worry about electrode trauma? Bilateral Implants Electro-acoustic implants

More information

Cochlear Implant Failure: Imaging Evaluation of the Electrode Course

Cochlear Implant Failure: Imaging Evaluation of the Electrode Course Clinical Radiology (2003) 58: 288 293 doi:10.1016/s0009-9260(02)00523-8, available online at www.sciencedirect.com Pictorial Review Cochlear Implant Failure: Imaging Evaluation of the Electrode Course

More information

trauma? Bilateral Implants Electro-acoustic implants Preservation of ganglion cells?

trauma? Bilateral Implants Electro-acoustic implants Preservation of ganglion cells? Electrode Trauma Assessed by Microdissection Peter S Roland MD C Gary Wright PhD University of Texas Medical Center Dallas, Texas Why worry about electrode trauma? Bilateral Implants Electro-acoustic implants

More information

During the past decade, combined electric acoustic stimulation

During the past decade, combined electric acoustic stimulation ORIGINAL RESEARCH S.E.J. Connor D.J. Bell R. O Gorman A. Fitzgerald- O Connor CT and MR Imaging Cochlear Distance Measurements May Predict Cochlear Implant Length Required for a 360 Insertion BACKGROUND

More information

RADIOLOGIC EVALUATION OF MULTICHANNEL INTRACOCHLEAR IMPLANT INSERTION DEPTH

RADIOLOGIC EVALUATION OF MULTICHANNEL INTRACOCHLEAR IMPLANT INSERTION DEPTH THE AMERICAN JOURNAL OF OTOLOCY!VOLUME 14, NUMBER 4 July 1993 469 ORIGINAL MANUSCRIPTS RADIOLOGIC EVALUATION OF MULTICHANNEL INTRACOCHLEAR IMPLANT INSERTION DEPTH Michael A. Marsh, MD.,** JinXu, MD.,*

More information

A surgical approach for a cochlear implant: An anatomical study

A surgical approach for a cochlear implant: An anatomical study A surgical approach for a cochlear implant: An anatomical study By GRAEME M. CLARK (Melbourne) Introduction THERE is now increased interest in the possibility of restoring brain and nerve function by applying

More information

An optical coherence tomography study for imaging the round window niche and the promontorium tympani

An optical coherence tomography study for imaging the round window niche and the promontorium tympani An optical coherence tomography study for imaging the round window niche and the promontorium tympani T. Just *a, E. Lankenau b, G. Hüttmann b, H.W. Pau a a Department of Otorhinolaryngology, University

More information

Our Experience with AB HiFocus MSE Cochlear Implant from Laboratory to OR: Electrode and Access Variables

Our Experience with AB HiFocus MSE Cochlear Implant from Laboratory to OR: Electrode and Access Variables Our Experience with AB HiFocus MSE Cochlear Implant from Laboratory to OR: Electrode and Access Variables Sean O McMenomey, David R Friedmann MD, Ling Zhou MD, Stephen Rebscher,* J Thomas Roland, Jr. MD

More information

ORIGINAL ARTICLE. The Cochlear-Carotid Interval: Preoperative Assessment for Cochlear Implant

ORIGINAL ARTICLE. The Cochlear-Carotid Interval: Preoperative Assessment for Cochlear Implant ORIGINAL ARTICLE The Cochlear-Carotid Interval: Preoperative Assessment for Cochlear Implant Eshrak Hassanein MD; Eman Geneidi MD; and Mohamed Taha MD From the Departments of Radiology (E.Hassanein), Otorhinolaryngology

More information

For Professionals. Electrode Arrays. Designed for Atraumatic Implantation Providing Superior Hearing Performance

For Professionals. Electrode Arrays. Designed for Atraumatic Implantation Providing Superior Hearing Performance For Professionals Electrode Arrays Designed for Atraumatic Implantation Providing Superior Hearing Performance Electrode Arrays Designed for Atraumatic Implantation Providing Superior Hearing Performance,,,

More information

A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND NRTs FOR THE NUCLEUS 422 AND CONTOUR ELECTRODE ARRAYS

A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND NRTs FOR THE NUCLEUS 422 AND CONTOUR ELECTRODE ARRAYS A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND NRTs FOR THE NUCLEUS 422 AND CONTOUR ELECTRODE ARRAYS December 13, 2014! Sandra Velandia, AuD! Ear Institute, Miller School of Medicine, University of Miami!

More information

Improved and Simplified Methods for Specifying Positions of the Electrode Bands of a Cochlear Implant Array

Improved and Simplified Methods for Specifying Positions of the Electrode Bands of a Cochlear Implant Array 902 The American Journal ojorologv 17:859-865 1996, The American Journal of Otology, Inc Improved and Simplified Methods for Specifying Positions of the Electrode Bands of a Cochlear Implant Array Lawrence

More information

Scala Tympani Cochleostomy II: Topography and Histology

Scala Tympani Cochleostomy II: Topography and Histology The Laryngoscope Lippincott Williams & Wilkins 2007 The American Laryngological, Rhinological and Otological Society, Inc. Scala Tympani Cochleostomy II: Topography and Histology Oliver F. Adunka, MD;

More information

INTRACOCHLEAR FACTORS CONTRIBUTING TO PSYCHOPHYSICAL PERCEPTS FOLLOWING COCHLEAR IMPLANTATION: A CASE STUDY

INTRACOCHLEAR FACTORS CONTRIBUTING TO PSYCHOPHYSICAL PERCEPTS FOLLOWING COCHLEAR IMPLANTATION: A CASE STUDY INTRACOCHLEAR FACTORS CONTRIBUTING TO PSYCHOPHYSICAL PERCEPTS FOLLOWING COCHLEAR IMPLANTATION: A CASE STUDY A. KAWANO, MD, PHD; H. L. SELDON, MD, PHD; B. PYMAN, MD, FRACS; G. M. CLARK, PHD, FRACS From

More information

Investigating the Effect of Cochlear Size in Insertion of Electrode Depth in Patients with Cochlear Implantation Evaluated by CT- Scan

Investigating the Effect of Cochlear Size in Insertion of Electrode Depth in Patients with Cochlear Implantation Evaluated by CT- Scan http:// ijp.mums.ac.ir Original Article (Pages: 9207-9213) Investigating the Effect of Cochlear Size in Insertion of Electrode Depth in Patients with Cochlear Implantation Evaluated by CT- Scan *Mohammad

More information

PRECURVED ELECTRODE ARRAY AND INSERTION TOOL

PRECURVED ELECTRODE ARRAY AND INSERTION TOOL PRECURVED ELECTRODE ARRAY AND INSERTION TOOL C. G. TREABA, MENGSC; J. Xu, MD; S.-A. Xu, MD; G. M. CLARK, PHD, FRACS FromCochlearPty limiled, Sydney (l'reaba), the Department ofotolaryngology (J. Xu, S.-A.

More information

The number of cochlear implantations is increasing rapidly,

The number of cochlear implantations is increasing rapidly, ORIGINAL RESEARCH HEAD & NECK Radiologic and Functional Evaluation of Electrode Dislocation from the Scala Tympani to the Scala Vestibuli in Patients with Cochlear Implants N. Fischer, L. Pinggera, V.

More information

Clinical Study Imaging of Electrode Position after Cochlear Implantation with Flat Panel CT

Clinical Study Imaging of Electrode Position after Cochlear Implantation with Flat Panel CT International Scholarly Research Network ISRN Otolaryngology Volume 2012, Article ID 728205, 5 pages doi:10.5402/2012/728205 Clinical Study Imaging of Electrode Position after Cochlear Implantation with

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/37612 holds various files of this Leiden University dissertation Author: Marel, Kim van der Title: Unraveling the implanted cochlea : radiological evaluation

More information

Complete Cochlear Coverage WITH MED-EL S DEEP INSERTION ELECTRODE

Complete Cochlear Coverage WITH MED-EL S DEEP INSERTION ELECTRODE Complete Cochlear Coverage WITH MED-EL S DEEP INSERTION ELECTRODE hearlife CONTENTS A Factor To Consider... 3 The Cochlea In the Normal Hearing Process... 5 The Cochlea In the Cochlear Implant Hearing

More information

Philipp Mittmann, 1 Grit Rademacher, 2 Sven Mutze, 2 Frederike Hassepass, 3 Arneborg Ernst, 1 and Ingo Todt Introduction

Philipp Mittmann, 1 Grit Rademacher, 2 Sven Mutze, 2 Frederike Hassepass, 3 Arneborg Ernst, 1 and Ingo Todt Introduction BioMed Research International Volume 2015, Article ID 706253, 6 pages http://dx.doi.org/10.1155/2015/706253 Research Article Evaluation of the Relationship between the NRT-Ratio, Cochlear Anatomy, and

More information

Difficult Cases: Controversies in Cochlear Implantation

Difficult Cases: Controversies in Cochlear Implantation Difficult Cases: Controversies in Cochlear Implantation David S Haynes, MD FACS Fred F Telischi, MD MEE FACS Lawrence R. Lustig, MD Robert F Labadie, PhD MD Nikolas H Blevins, MD Matthew L. Carlson, MD

More information

Research Article MRI-Based Estimation of Scalar Cochlear-Implant Electrode Position

Research Article MRI-Based Estimation of Scalar Cochlear-Implant Electrode Position Hindawi BioMed Research International Volume 2017, Article ID 6372704, 5 pages https://doi.org/10.1155/2017/6372704 Research Article MRI-Based Estimation of Scalar Cochlear-Implant Electrode Position A.

More information

ORIGINAL ARTICLE. Perimodiolar electrode position: Effects on thresholds, comfort levels, impedance measurements, and neural response telemetry

ORIGINAL ARTICLE. Perimodiolar electrode position: Effects on thresholds, comfort levels, impedance measurements, and neural response telemetry The Mediterranean Journal of Otology ORIGINAL ARTICLE Perimodiolar electrode position: Effects on thresholds, comfort levels, impedance measurements, and neural response telemetry Angel Ramos Macias, MD;

More information

1. Axial view, left temporal bone. Plane through the upper antrum (A), superior semicircular canal (SSC) and IAC.

1. Axial view, left temporal bone. Plane through the upper antrum (A), superior semicircular canal (SSC) and IAC. PA IAC SSC A 1. Axial view, left temporal bone. Plane through the upper antrum (A), superior semicircular canal (SSC) and IAC. IAC VII M I LSC Plane through the IAC, malleus head and incus and the lateral

More information

Clinical Workflow. White Paper No. 35 / 2016

Clinical Workflow. White Paper No. 35 / 2016 01 The mobile 3D C-arm Ziehm Vision RFD 3D A well-suited tool for intraoperative imaging in cochlear implantation More than 5 % of the world s population 360 million people suffer from hearing loss in

More information

Electrode Impedance among Children Using the Combi-40+ Medel Cochlear Implant

Electrode Impedance among Children Using the Combi-40+ Medel Cochlear Implant Electrode Impedance among Children Using the Combi-40+ Medel Cochlear Implant Khalid Tubishi MD*, Hussein Al-Qasem PhD*, Abed Rabu Qubilat MD*, Salman Assaf MD*, Mefleh Sarhan MD*, Sufian Roud MD*, Ahmed

More information

HOW TO IMPROVE COCHLEAR IMPLANT IN ADULT

HOW TO IMPROVE COCHLEAR IMPLANT IN ADULT HOW TO IMPROVE COCHLEAR IMPLANT IN ADULT Poor performances in adult CI patients and its remediation B. FRAYSSE XXXVI Congreso LIMA November 14-17, 2018 GOAL OF THE STUDY To propose a predictive model during

More information

An Evidence-Based Algorithm for Intraoperative Monitoring During Cochlear Implantation

An Evidence-Based Algorithm for Intraoperative Monitoring During Cochlear Implantation Otology & Neurotology 33:169Y176 Ó 2012, Otology & Neurotology, Inc. An Evidence-Based Algorithm for Intraoperative Monitoring During Cochlear Implantation Maura K. Cosetti, Scott H. Troob, Jonathan M.

More information

HOW TO IMPROVE COCHLEAR IMPLANT IN ADULT

HOW TO IMPROVE COCHLEAR IMPLANT IN ADULT HOW TO IMPROVE COCHLEAR IMPLANT IN ADULT Poor performances in adult CI patients and its remediation B. FRAYSSE IFOS WORLD MASTER COURSE ON HEARING REHABILITATION DUBAI March 2019, 28-29-30 INTRODUCTION

More information

CONFLICTS OF INTEREST

CONFLICTS OF INTEREST COCHLEAR IMPLANTATION: A SURGEON S PERSPECTIVE Ravi N. Samy, M.D., F.A.C.S. Ravi.Samy@UC.edu Director, Adult Cochlear Implant Program Program Director, Neurotology Fellowship CONFLICTS OF INTEREST RESEARCH

More information

External carotid blood supply to acoustic neurinomas

External carotid blood supply to acoustic neurinomas External carotid blood supply to acoustic neurinomas Report of two cases HARVEY L. LEVINE, M.D., ERNEST J. FERmS, M.D., AND EDWARD L. SPATZ, M.D. Departments of Radiology, Neurology, and Neurosurgery,

More information

Fitting of the Hearing System Affects Partial Deafness Cochlear Implant Performance

Fitting of the Hearing System Affects Partial Deafness Cochlear Implant Performance 1 Fitting of the Hearing System Affects Partial Deafness Cochlear Implant Performance Marek Polak 1, Artur Lorens 2, Silke Helbig 3, Sonelle McDonald 4, Sheena McDonald 4 ; Katrien Vermeire 5 1 MED-EL,

More information

NIDCD NATIONAL TEMPORAL BONE, HEARING AND BALANCE PATHOLOGY RESOURCE REGISTRY

NIDCD NATIONAL TEMPORAL BONE, HEARING AND BALANCE PATHOLOGY RESOURCE REGISTRY NIDCD NATIONAL TEMPORAL BONE, HEARING AND BALANCE PATHOLOGY RESOURCE REGISTRY Guidelines for Removal of Temporal Bones for Pathological Study The temporal bones should be removed as soon as possible. If

More information

Residual Hearing Preservation After Cochlear Implantation via Round Window or Cochleostomy Approach

Residual Hearing Preservation After Cochlear Implantation via Round Window or Cochleostomy Approach The Laryngoscope VC 2014 The American Laryngological, Rhinological and Otological Society, Inc. Residual Hearing reservation After Cochlear Implantation via Round Window or Cochleostomy Approach Chuan-Hung

More information

Lamb Temporal Bone as a Surgical Training Model of Round Window Cochlear Implant Electrode Insertion

Lamb Temporal Bone as a Surgical Training Model of Round Window Cochlear Implant Electrode Insertion Otology & Neurotology 37:52 56 ß 2015, Otology & Neurotology, Inc. Lamb Temporal Bone as a Surgical Training Model of Round Window Cochlear Implant Electrode Insertion Georgios Mantokoudis, Markus E. Huth,

More information

Visualisation of the Cochlea Implant/Tissue Interface

Visualisation of the Cochlea Implant/Tissue Interface Visualisation of the Cochlea Implant/Tissue Interface Christopher Wong 1, 4, Ian S. Curthoys 2, Stephen O Leary 3,4, Allan S. Jones 1 1 Australian Centre for Microscopy and Microanalysis, University of

More information

Insertion of electrode array using percutaneous cochlear implantation technique: a cadaveric study

Insertion of electrode array using percutaneous cochlear implantation technique: a cadaveric study Insertion of electrode array using percutaneous cochlear implantation technique: a cadaveric study Ramya Balachandran 1, Jason E. Mitchell 2, Jack Noble 3, Daniel Schurzig 1, Grégoire Blachon 1, Theodore

More information

Upper Limb Imaging Requirements

Upper Limb Imaging Requirements Imaging Requirements Upper Limb Imaging Requirements Instructions for Measurement Radiography and CT Scans Please read before commencing radiography Stanmore Implants 210 Centennial Avenue Centennial Park

More information

Enduring Long-Term Speech Comprehension Benefits of Apical Stimulation in Cochlear Implantation

Enduring Long-Term Speech Comprehension Benefits of Apical Stimulation in Cochlear Implantation Enduring Long-Term Speech Comprehension Benefits of Apical Stimulation in Cochlear Implantation Kevin D. Brown, Meg Dillon, Meredith Anderson, Ellen Pearce, English King, Craig Buchman and Harold C. Pillsbury

More information

CLINICAL PAPER / ORTHOPEDIC

CLINICAL PAPER / ORTHOPEDIC HIP LEG LENGTH AND OFFSET Kelley T.C. and Swank M.L. (2009) Using CAS leads to more accurate positioning within the safe zone (inclination between 30 and 50, anteversion between 5 and 25 ) CAS improves

More information

Original Policy Date

Original Policy Date MP 7.01.66 Auditory Brainstem Implant Medical Policy Section Surgery Issue 12/2013 Original Policy Date 12/2013 Last Review Status/Date Reviewed with literature search/12/2013 Return to Medical Policy

More information

Preoperative Planning. The primary objectives of preoperative planning are to:

Preoperative Planning. The primary objectives of preoperative planning are to: Preoperative Planning The primary objectives of preoperative planning are to: - Determine preoperative leg length discrepancy. - Assess acetabular component size and placement. - Determine femoral component

More information

TWO MEDICAL IMAGING PROCESSING TECHNIQUES FOR USE IN ACOUSTIC NEUROMA REMOVAL AND COCHLEAR PROSTHETIC IMPLANTATION. Kepra L McBrayer.

TWO MEDICAL IMAGING PROCESSING TECHNIQUES FOR USE IN ACOUSTIC NEUROMA REMOVAL AND COCHLEAR PROSTHETIC IMPLANTATION. Kepra L McBrayer. TWO MEDICAL IMAGING PROCESSING TECHNIQUES FOR USE IN ACOUSTIC NEUROMA REMOVAL AND COCHLEAR PROSTHETIC IMPLANTATION By Kepra L McBrayer Thesis Submitted to the Faculty of the Graduate School of Vanderbilt

More information

Active positioning device for a perimodiolar cochlear electrode array

Active positioning device for a perimodiolar cochlear electrode array Microsystem Technologies 10 (2004) 478 483 Ó Springer-Verlag 2004 DOI 10.1007/s00542-004-0376-5 Active positioning device for a perimodiolar cochlear electrode array B. Y. Arcand, P. T. Bhatti, N. V. Butala,

More information

Ear. Utricle & saccule in the vestibule Connected to each other and to the endolymphatic sac by a utriculosaccular duct

Ear. Utricle & saccule in the vestibule Connected to each other and to the endolymphatic sac by a utriculosaccular duct Rahaf Jreisat *You don t have to go back to the slides. Ear Inner Ear Membranous Labyrinth It is a reflection of bony labyrinth but inside. Membranous labyrinth = set of membranous tubes containing sensory

More information

Extraoral Radiology October 10th, 2008

Extraoral Radiology October 10th, 2008 Extraoral Radiology October 10th, 2008 Steven R. Singer, DDS srs2@columbia.edu 212.305.5674 November 8 th, 1895 Extraoral Projections Images can be produced in the dental office X-ray source can be Intraoral

More information

CT-Derived Estimation of Cochlear Morphology and Electrode Array Position in Relation to Word Recognition in Nucleus-22 Recipients

CT-Derived Estimation of Cochlear Morphology and Electrode Array Position in Relation to Word Recognition in Nucleus-22 Recipients JARO 03: 332±350 2002) DOI: 10.1007/s101620020013 JARO Journal of the Association for Research in Otolaryngology CT-Derived Estimation of Cochlear Morphology and Electrode Array Position in Relation to

More information

Institutional review board approval was obtained prior to the start of this study.

Institutional review board approval was obtained prior to the start of this study. Lower Limb Alignment and Length Measurements - Comparison of Computed Tomography, Upright Full-Length Conventional Radiography and Upright Biplanar Linear-Low Dose X-ray Scanner Poster No.: C-1382 Congress:

More information

CT of Postmeningitic Deafness: Observations and Predictive Value for Cochlear Implants in Children

CT of Postmeningitic Deafness: Observations and Predictive Value for Cochlear Implants in Children CT of Postmeningitic Deafness: Observations and Predictive Value for Cochlear Implants in Children Michele H. Johnson, M. Suzanne Hasenstab, Michael A. Seicshnaydre, and George H. Williams PURPOSE: To

More information

WINSTA-C. Clavicle Plating System

WINSTA-C. Clavicle Plating System Clavicle Plating System Clinical Advisor Michael Kurer FRCS FRCS (Orth) Consultant Orthopaedic and Shoulder Surgeon North Middlesex University Hospital NHS Trust Table of Contents Introduction Indication

More information

UP Bioengineering Our people

UP Bioengineering Our people UP Bioengineering Our people Design and application of user-specific models of cochlear implants Tania Hanekom Tiaan K Malherbe, Liezl Gross, Rene Baron, Riaze Asvat, Werner Badenhorst & Johan J Hanekom

More information

Neurophysiological effects of simulated auditory prosthesis stimulation

Neurophysiological effects of simulated auditory prosthesis stimulation Neurophysiological effects of simulated auditory prosthesis stimulation 2 th Quarterly Progress Report Neural Prosthesis Program Contract N0-DC-9-207 (no-cost extension period) April 2003 C.A. Miller,

More information

Definition of Anatomy. Anatomy is the science of the structure of the body and the relation of its parts.

Definition of Anatomy. Anatomy is the science of the structure of the body and the relation of its parts. Definition of Anatomy Anatomy is the science of the structure of the body and the relation of its parts. Basic Anatomical Terms Anatomical terms for describing positions: Anatomical position: Supine position:

More information

Predicting the Position of the Femoral Head Center

Predicting the Position of the Femoral Head Center The Journal of Arthroplasty Vol. 14 No. 1 1999 Predicting the Position of the Femoral Head Center Nobuhiko Sugano, MD, Philip C. Noble, PhD, and Emir Kamaric, MS Abstract: To find an accurate method to

More information

original article Comparison of cochlear duct length between the Saudi and non-saudi populations Alaa Alanazi, a Farid Alzhrani b

original article Comparison of cochlear duct length between the Saudi and non-saudi populations Alaa Alanazi, a Farid Alzhrani b Comparison of cochlear duct length between the Saudi and non-saudi populations Alaa Alanazi, a Farid Alzhrani b From the a College of Medicine, King Saud University, Riyadh, Saudi Arabia; b King Abdullah

More information

Speech perception and speech intelligibility in children after cochlear implantation

Speech perception and speech intelligibility in children after cochlear implantation International Journal of Pediatric Otorhinolaryngology (2004) 68, 347 351 Speech perception and speech intelligibility in children after cochlear implantation Marie-Noëlle Calmels*, Issam Saliba, Georges

More information

A Psychophysics experimental software to evaluate electrical pitch discrimination in Nucleus cochlear implanted patients

A Psychophysics experimental software to evaluate electrical pitch discrimination in Nucleus cochlear implanted patients A Psychophysics experimental software to evaluate electrical pitch discrimination in Nucleus cochlear implanted patients M T Pérez Zaballos 1, A Ramos de Miguel 2, M Killian 3 and A Ramos Macías 1 1 Departamento

More information

Three-Dimensional Modeling and Visualization of the Cochlea on the Internet

Three-Dimensional Modeling and Visualization of the Cochlea on the Internet 144 IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, VOL. 4, NO. 2, JUNE 2000 Three-Dimensional Modeling and Visualization of the Cochlea on the Internet Sun K. Yoo, Ge Wang, Jay T. Rubinstein,

More information

Cochlear View: Postoperative Radiography for Cochlear Implantation

Cochlear View: Postoperative Radiography for Cochlear Implantation ,-- OTOPUBS#~1=1=98~ The American Joumal of Otology 21:49-56 2000, The American Journal of Otology, Inc. Cochlear View: Postoperative Radiography for Cochlear Implantation *Jin Xu, tshi-ang Xu, *Lawrence

More information

STATE OF THE ART IN COCHLEAR IMPLANTATION: CONCEPTS IN MINIMALLY TRAUMATIC SURGERY!

STATE OF THE ART IN COCHLEAR IMPLANTATION: CONCEPTS IN MINIMALLY TRAUMATIC SURGERY! STATE OF THE ART IN COCHLEAR IMPLANTATION: CONCEPTS IN MINIMALLY TRAUMATIC SURGERY! Brendan P. O Connell MD! Assistant Professor! Otology and Neurotology- UNC Chapel Hill! EXPANDING INDICATIONS FOR CONVENTIONAL

More information

Current Standards in Cochlear Implantation. Preliminary Results of a Survey Among Politzer Society

Current Standards in Cochlear Implantation. Preliminary Results of a Survey Among Politzer Society Int. Adv. Otol. 2012; 8:(3) 379-384 ORIGINAL ARTICLE Current Standards in Cochlear Implantation. Preliminary Results of a Survey Among Politzer Society Franco Trabalzini, Enrico Muzzi, Eva Orzan Department

More information

PSY 215 Lecture 10 Topic: Hearing Chapter 7, pages

PSY 215 Lecture 10 Topic: Hearing Chapter 7, pages PSY 215 Lecture 10 Topic: Hearing Chapter 7, pages 189-197 Corrections: NTC 09-1, page 3, the Superior Colliculus is in the midbrain (Mesencephalon). Announcements: Movie next Monday: Case of the frozen

More information

Periapical Radiography

Periapical Radiography Periapical Radiography BARBARA E. DIXON B.D.S., M.Sc., D.P.D.S. Main Indications Detection of Apical infection/inflammation Assessment of the periodontal status After trauma Assessment of Unerupted teeth

More information

Hip ultrasound for developmental dysplasia: the 50% rule

Hip ultrasound for developmental dysplasia: the 50% rule Pediatr Radiol (2017) 47:817 821 DOI 10.1007/s00247-017-3802-4 COMMENTARY Hip ultrasound for developmental dysplasia: the 50% rule H. Theodore Harcke 1 & B. Pruszczynski 2 Received: 27 October 2016 /Revised:

More information

For this lab you will use parts of Exercise #18 in your Wise lab manual. Please be sure to read those sections before coming to lab

For this lab you will use parts of Exercise #18 in your Wise lab manual. Please be sure to read those sections before coming to lab Bio 322 Human Anatomy Objectives for the laboratory exercise The Eye and Ear Required reading before beginning this lab: Saladin, KS: Human Anatomy 5 th ed (2017) Chapter 17 For this lab you will use parts

More information

Cochlear Implant Innovations

Cochlear Implant Innovations HOSPITAL CHRONICLES 2011, 6(4): 195 199 Special Topic Matilda Chroni, MD, PhD, Spyros Papaspyrou, MD, PhD ENT Department, Evagelismos General Hospital of Athens, Athens, Greece Key words: cochlear implants;

More information

Considerations and Rationale for Cochlear Implant Electrode Design - Past, Present and Future

Considerations and Rationale for Cochlear Implant Electrode Design - Past, Present and Future J Int Adv Otol 2018; 14(3): 382-91 DOI: 10.5152/iao.2018.6372 Invited Review Considerations and Rationale for Cochlear Implant Electrode Design - Past, Present and Future Frank Risi Group Product Manager,

More information

Surgeon Oriented Preoperative Radiologic Evaluation in Cochlear Implantation - Our experience with a Proposed Checklist

Surgeon Oriented Preoperative Radiologic Evaluation in Cochlear Implantation - Our experience with a Proposed Checklist THIEME Original Research Surgeon Oriented Preoperative Radiologic Evaluation in Cochlear Implantation - Our experience with a Proposed Checklist Mahmoud Mandour 1 Mohammed Tomoum 1 Saad El Zayat 2 Hisham

More information

Added Value of MDCT in Cochlear Parameters Assessment in Patients with Sensory Neural Hearing Loss Candidates for Cochlear Implantation

Added Value of MDCT in Cochlear Parameters Assessment in Patients with Sensory Neural Hearing Loss Candidates for Cochlear Implantation International Journal of Medical Imaging 2018; 6(4): 33-39 http://www.sciencepublishinggroup.com/j/ijmi doi: 10.11648/j.ijmi.20180604.11 ISSN: 2330-8303 (Print); ISSN: 2330-832X (Online) Added Value of

More information

Cochlear Implantation through the Middle Fossa Approach: A Review of Related Temporal Bone Studies and Reported Cases

Cochlear Implantation through the Middle Fossa Approach: A Review of Related Temporal Bone Studies and Reported Cases 102 Systematic Review THIEME Cochlear Implantation through the Middle Fossa Approach: A Review of Related Temporal Bone Studies and Reported Cases Juan Carlos Cisneros Lesser 1,2 Rubens Vuono de Brito

More information

THE EAR AND HEARING Be sure you have read and understand Chapter 16 before beginning this lab. INTRODUCTION: hair cells outer ear tympanic membrane

THE EAR AND HEARING Be sure you have read and understand Chapter 16 before beginning this lab. INTRODUCTION: hair cells outer ear tympanic membrane BIOLOGY 211: HUMAN ANATOMY & PHYSIOLOGY ****************************************************************************************************** THE EAR AND HEARING ******************************************************************************************************

More information

ORIGINAL ARTICLE. Paul W. Bauer, MD; Franz J. Wippold II, MD; Jenifer Goldin, MS, CCC-A; Rodney P. Lusk, MD

ORIGINAL ARTICLE. Paul W. Bauer, MD; Franz J. Wippold II, MD; Jenifer Goldin, MS, CCC-A; Rodney P. Lusk, MD ORIGINAL ARTICLE Cochlear Implantation in Children With CHARGE Association Paul W. Bauer, MD; Franz J. Wippold II, MD; Jenifer Goldin, MS, CCC-A; Rodney P. Lusk, MD Objective: To explore the anomalies

More information

Across-Site Variation in Detection Thresholds and Maximum Comfortable Loudness Levels for Cochlear Implants

Across-Site Variation in Detection Thresholds and Maximum Comfortable Loudness Levels for Cochlear Implants JARO 5: 11 24 (2004) DOI: 10.1007/s10162-003-3051-0 JARO Journal of the Association for Research in Otolaryngology Across-Site Variation in Detection Thresholds and Maximum Comfortable Loudness Levels

More information

Hearing Research 221 (2006) Research paper. Representation of acoustic signals in the human cochlea in presence of a cochlear implant electrode

Hearing Research 221 (2006) Research paper. Representation of acoustic signals in the human cochlea in presence of a cochlear implant electrode Hearing Research 221 (2006) 36 43 Research paper Representation of acoustic signals in the human cochlea in presence of a cochlear implant electrode Jan Kiefer a, *, Frank Böhnke b, Oliver Adunka c, Wolfgang

More information

New Methods of Deafness and Partial Deafness Treatment

New Methods of Deafness and Partial Deafness Treatment Biocybernetics and Biomedical Engineering 2006, Volume 26, Number 1, pp. 75 83 New Methods of Deafness and Partial Deafness Treatment ARTUR LORENS*, HENRYK SKAR YÑSKI, ANNA PIOTROWSKA International Center

More information

A study of the anatomy of the caudal space using magnetic resonance imaging

A study of the anatomy of the caudal space using magnetic resonance imaging British Journal of Anaesthesia 1997; 78: 391 395 A study of the anatomy of the caudal space using magnetic resonance imaging I. M. CRIGHTON, B. P. BARRY AND G. J. HOBBS Summary We have studied, in 37 adult

More information

DOWNLOAD PDF PRINCIPLES OF COCHLEAR IMPLANT IMAGING

DOWNLOAD PDF PRINCIPLES OF COCHLEAR IMPLANT IMAGING Chapter 1 : Imaging in cochlear implant patients 7 Principles of Cochlear Implant Imaging. Andrew J. Fishman and Roy A. Holliday. Radiographic imaging plays a major role in cochlear implantation with regard

More information

For Professionals. SYNCHRONY System. In Sync with Natural Hearing

For Professionals. SYNCHRONY System. In Sync with Natural Hearing For Professionals SYNCHRONY System In Sync with Natural Hearing For more natural hearing in any listening environment SONNET Audio Processor RONDO Single-Unit Processor Designed for exceptional hearing

More information

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION Donald L. Renfrew, MD Radiology Associates of the Fox Valley, 333 N. Commercial Street, Suite 100, Neenah, WI 54956 11/24/2012 Radiology Quiz of the Week # 100 Page 1 CLINICAL PRESENTATION AND RADIOLOGY

More information

Cochlear implantations in Northern Ireland: An overview of the first five years

Cochlear implantations in Northern Ireland: An overview of the first five years The Ulster Medical Journal, Volume 68, No. 1, pp. 12-16, May 1999. Cochlear implantations in Northern Ireland: An overview of the first five years G John, J G Toner Accepted 12 March 1999 SUMMARY During

More information

Cochlear Implants. Medical Procedures. for MED EL Implant Systems. AW33290_2.0 (English US)

Cochlear Implants. Medical Procedures. for MED EL Implant Systems. AW33290_2.0 (English US) Cochlear Implants Medical Procedures for MED EL Implant Systems AW33290_2.0 (English US) This manual provides important instructions and safety information for MED EL Implant System users who have to undergo

More information

POTENTIAL AND LIMITATIONS OF COCHLEAR IMPLANTS IN CHILDREN

POTENTIAL AND LIMITATIONS OF COCHLEAR IMPLANTS IN CHILDREN . _ POTENTIAL AND LIMITATIONS OF COCHLEAR IMPLANTS IN CHILDREN R. C. DOWELL, PHD, DAuo; P. J. BLAMEY, PHD; G. M. CLARK, PHD, FRACS From the Human Communication Research Centre (Dowell. Clark) and the Department

More information

Physiologic Consequences of Intracochlear Electrode Placement. Oliver F. Adunka, MD, FACS Craig A. Buchman, MD, FACS Douglas C.

Physiologic Consequences of Intracochlear Electrode Placement. Oliver F. Adunka, MD, FACS Craig A. Buchman, MD, FACS Douglas C. Physiologic Consequences of Intracochlear Electrode Placement Oliver F. Adunka, MD, FACS Craig A. Buchman, MD, FACS Douglas C. Fitzpatrick, PhD Disclosures Advisory Board» MED-EL North America» Advanced

More information

JMSCR Vol 06 Issue 10 Page October 2018

JMSCR Vol 06 Issue 10 Page October 2018 www.jmscr.igmpublication.org Impact Factor (SJIF): 6.379 Index Copernicus Value: 79.54 ISSN (e)-2347-176x ISSN (p) 2455-45 DOI: https://dx.doi.org/1.18535/jmscr/v6i1.145 Role of VR 3D Imaging in temporal

More information

BONEBRIDGE Bibliography

BONEBRIDGE Bibliography BONEBRIDGE Bibliography With abstracts 06/2013 1 A.M. Huber; J.H. Sim; Y.Z. Xie; M. Chatzimichalis; O. Ullrich, and C. Roosli. The Bonebridge: Preclinical evaluation of a new transcutaneously-activated

More information

Dental Cone Beam CT. What is Dental Cone Beam CT?

Dental Cone Beam CT. What is Dental Cone Beam CT? Scan for mobile link. Dental Cone Beam CT Dental cone beam computed tomography (CT) is a special type of x-ray equipment used when regular dental or facial x-rays are not sufficient. Your doctor may use

More information

Transcanal Endoscopic Approach to the Sinus Tympani: A Clinical Report

Transcanal Endoscopic Approach to the Sinus Tympani: A Clinical Report Otology & Neurotology 30:758Y765 Ó 2009, Otology & Neurotology, Inc. Transcanal Endoscopic Approach to the Sinus Tympani: A Clinical Report Daniele Marchioni, Francesco Mattioli, Matteo Alicandri-Ciufelli,

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005 Advantages of inflatable multichannel endorectal applicator in the neo-adjuvant treatment of patients with locally advanced

More information

COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION

COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION C H A P T E R 3 0 COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION Scott R. Roman, DPM INTRODUCTION The most common fixation for proximal interphalangeal fusion

More information

HEARING AND COCHLEAR IMPLANTS

HEARING AND COCHLEAR IMPLANTS HEARING AND COCHLEAR IMPLANTS FRANCIS CREIGHTON, MD NEUROTOLOGY & SKULL BASE SURGERY FELLOW JOHNS HOPKINS SCHOOL OF MEDICINE NOV 9 TH, 2017 THANKS TO: CHARLIE DELLA SANTINA, HEIDI NAKAJIMA AND DOUG MATTOX

More information

Your ear consists of three parts that play a vital role in hearing-the external ear, middle ear, and inner ear.

Your ear consists of three parts that play a vital role in hearing-the external ear, middle ear, and inner ear. What is a Cochlear Implant? A cochlear implant is an electronic device that restores partial hearing to the deaf. It is surgically implanted in the inner ear and activated by a device worn outside the

More information

SOLUTIONS Homework #3. Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03

SOLUTIONS Homework #3. Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03 SOLUTIONS Homework #3 Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03 Problem 1: a) Where in the cochlea would you say the process of "fourier decomposition" of the incoming

More information

Unit VIII Problem 9 Physiology: Hearing

Unit VIII Problem 9 Physiology: Hearing Unit VIII Problem 9 Physiology: Hearing - We can hear a limited range of frequency between 20 Hz 20,000 Hz (human hearing acuity is between 1000 Hz 4000 Hz). - The ear is divided into 3 parts. Those are:

More information

LAMINA SPREADER SURGICAL TECHNIQUE

LAMINA SPREADER SURGICAL TECHNIQUE LAMINA SPREADER SURGICAL TECHNIQUE Balanced and appropriate external rotation of the femoral component is important for tibio-femoral stability in flexion and patello-femoral tracking/function. Depending

More information

PETER]. BLAMEY,PHD. ALISON M. BROWN, BSc, DIPAuD RICHARD C. DOWELL, MSc, DIPAuD

PETER]. BLAMEY,PHD. ALISON M. BROWN, BSc, DIPAuD RICHARD C. DOWELL, MSc, DIPAuD ... Ann Otol Rhinol LaryngollOl:1992 REPRINTED FROM ANNALS OF OTOLOGY, RHINOLOGY & LARYNGOLOGY APRIL 1992 Volume 101 Number 4 C:=.O~PY.:..:R:..::I:.::G:..:H.:..:T~ 1992, ANNALS PUBLISHING COMPANY FACTORS

More information

THE COCHLEA AND AUDITORY PATHWAY

THE COCHLEA AND AUDITORY PATHWAY Dental Neuroanatomy Suzanne S. Stensaas, PhD February 23, 2012 Reading: Waxman, Chapter 16, Review pictures in a Histology book Computer Resources: http://www.cochlea.org/ - Promenade around the Cochlea

More information

Radiographic Positioning for Dogs

Radiographic Positioning for Dogs Radiographic Positioning for Dogs Elbow Radiographs: Lateral View A routine elbow exam consists of a lateral, flexed lateral and craniocaudal view. When performing elbow radiographs, a quality control

More information

ALESSANDRA RUSSO MD GRUPPO OTOLOGICO

ALESSANDRA RUSSO MD GRUPPO OTOLOGICO Gruppo Otologico Simultaneous Labyrinthectomy with Cochlear Implantation Cenacolo di Audiovestibologia, Chieti, 24-25/06/2016 ALESSANDRA RUSSO MD GRUPPO OTOLOGICO MENIERE DISEASE Therapeutic Steps 1 step

More information