1 AirLife brand Misty Max 10 nebulizer Purpose Introduction Delivery of nebulized medication to the lungs is a complex process dependant upon a variety of clinical and device-related variables. Patient breathing patterns, the choice and dilution of drug nebulized, nebulizer flow rates, device selection and orientation all have the potential to affect the amount of drug delivered to the patient. In vitro (bench) testing can provide the end-user with useful information regarding nebulizer performance. Various test methods have been used over the years; however, the quantification of inhaled drug mass through the use of a simulated breathing patient has become an increasingly well-accepted method of evaluation. The use of a simulated breathing patient allows for a more accurate estimation of drug dose. 1 Medication delivery: Misty Max 10 nebulizer performance CareFusion conducted a study, the purpose of which was to compare the performance of four conventional small volume nebulizers (SVN) currently available on the market. The study was designed to closely simulate typical clinical conditions in the acute care setting. Conclusions The Misty Max 10 nebulizer creates finer particles with a tighter distribution than any of the other nebulizers tested. As a result, the respirable fraction performance of Misty the Max 10 nebulizer is significantly higher than that of competitive products. The Misty Max 10 nebulizer delivers significantly higher inhaled respirable mass than the other nebulizers when tested in the angled position. The Misty Max 10 and the VixOne nebulizers deliver significantly higher inhaled respirable mass than the other nebulizers when tested in the vertical position. There was no statistically significant difference between Misty Max 10 and VixOne nebulizer test results. (Continued on next page) Conventional SVN performance at 8 lpm flow AirLife brand Misty Max 10 nebulizer Westmed VixOne Hudson MicroMist Salter Labs 8900 Particle size MMAD GSD % Respirable (< 4.7 μm) % 2.08* %* 2.43* % 1.85* %* Inhaled mass performance: Treatment time (min) Total residual neb, t-piece, m.p. (μg) Inhaled respirable mass (μg) 1, , , , , * 1, , * 1, * 1, , * 1, * 1, Note: All drug masses expressed in micrograms of albuterol = Performance of nebulizer at vertical position = Performance of nebulizer at 60 degrees from vertical Comparisons were made using Student s T-test for means *Difference between MM10 statistically significant (p<.05) Difference between MM10 not statistically significant (p>.05)

2 The Misty Max 10 nebulizer delivers an averaged inhaled respirable mass approximately 14% higher than the closest competitor. The Misty Max 10 nebulizer has the lowest average drug residual volume compared to the other nebulizers tested. Performance criteria evaluated Respirable fraction This is the percent of aerosol generated by mass that falls below an aerodynamic diameter of 5μm. It has been reported that particles less than 5μm will penetrate beyond the upper airways and deposit into the tracheobronchial and pulmonary regions of the lung. 3 The closest cascade plate cutoff point of 4.7μm was utilized to quantify mass below 5μm. Respirable fraction % less than 4.7μm Misty Max 10 VixOne MicroMist 8900 This is an estimate of the amount of respirable drug inhaled by the patient under specified breathing patterns and duration. It is calculated by multiplying the respirable fraction and the total drug mass collected on the inhalation filter. Hess used a similar method of estimating respirable mass available to the patient. 4 Average inhaled respirable mass Drug mass (μg) Misty Max 10 VixOne MicroMist 8900 Treatment time This is the amount of time (minutes) to nebulize 3 ml of 0.083% albuterol solution. Average treatment time Minutes Misty Max 10 VixOne MicroMist 8900 Total residual This is the amount of drug mass found within the nebulizer, t-piece and mouthpiece after treatment. The drug mass within the nebulizer and accessories were assayed using a spectrophotometer. MMAD Mass median aerodynamic diameter (MMAD) is a measurement of central tendency of the size distribution of aerosol particles. It is the diameter, in micrometers, of which 50% of the mass of aerosol is larger and 50% is smaller. Drug mass (μg) d GSD Geometric standard deviation (GSD) is a measure of the width of the size distribution of aerosol. For log normal distributions, it is calculated as follows: GSD = (84.13% diameter/15.87% diameter). 1, Misty Max 10 VixOne MicroMist 8900

3 Test methods Particle size An Andersen eight-stage cascade impactor (ACI) with USP inlet was used and assayed with a spectrophotometer (277nm). Particle size characterization was performed using 3 ml of 0.083% albuterol solution and a nebulizer flow rate of 8 L/min. All cascade tests sampled nebulizer aerosol through the manufacturer s supplied t-piece and mouthpiece. Inhaled drug mass The test setup consisted of a filter placed between the mouthpiece of the nebulizer and a simulated breathing patient. The filter captured the inhaled aerosol exiting the mouthpiece and the drug mass deposited onto the filter was quantified with a UV spectrophotometer. A simulated breathing patient was created by using a NPB 7200 ventilator and a Michigan Test Lung. An adult breathing circuit was connected to a ventilator that drove one side of the test lung. A bar was connected to the opposing bellows, duplicating the movement of the test lung and simulating a patient breathing pattern. A BioTek flow analyzer was placed between the simulated patient and the inhalation filter to monitor actual breathing parameters. The breathing pattern was held at a tidal volume of approximately 500 ml, 12 breaths per minute and an I:E ratio of 1:2, similar to a study presented by Siafaka, Hess and Bigatello. 2 A commonly used dosing of albuterol (3 ml, 0.083% albuterol) was nebulized until aerosol production ceased for 30 seconds or after two minutes of sputtering was observed, whichever came first. All nebulizers were run at a flow rate of 8 L/min. Inhaled drug mass was collected onto the filter over the duration of treatment. Within the clinical setting, nebulizer treatments will be given to patients in a variety of incline positions. To better simulate these conditions, nebulizers were tested at both vertical and angled (60 ) positions. Products tested All nebulizers were randomly pulled from production lots. Catalog numbers chosen were of similar configuration: nebulizer, t-piece, mouthpiece, oxygen tubing and 6 flextube. A total of nine samples for each nebulizer were tested: three for particle size and six for inhaled drug mass testing. AirLife brand Misty Max 10 nebulizer (cat. no ) WestMed VixOne (cat. no. 0210) Hudson MicroMist (cat. no. 1883) Salter Labs 8900 (cat. no. 8900)

4 Technical data David A. Rivera, Research and Development February 17, 2004 The purpose of this document is to provide the in vitro performance characteristics of the AirLife brand Misty Max 10 nebulizer. In vitro measurements will help guide the end user as to the likelihood of effective delivery of inhaled medication to the lower respiratory tract Cascade impaction measures aerodynamic diameter directly, which accounts for the density and irregular shape of drug particles. It is believed that aerodynamic diameter more accurately predicts the behavior of aerosol as it is delivered into the patient s lungs. 6 Performance characteristics MMAD Mass median aerodynamic diameter GSD Geometric standard deviation Respirable fraction % < 4.7μm (by mass) Output rate (g/min) Treatment time 3 ml albuterol solution (min) Residual (g) Operating pressure (psi) 3 ml dose of 0.083% albuterol (μg) 5 L/min flow 10 L/min flow 2.21 ±0.07 (μm) 1.30 ± ± ± ± ± ± ± ± ± ± ± ± ± ± ±14 2. There is more historical data on particle size measurement using cascade data than any other method. Relative comparison with historical data can be readily made. 3. Cascade impaction is one of the USP methods for characterization of particle distributions. 7 Particle size testing was conducted with components included with AirLife brand Misty Max 10 nebulizer : T-adapter, mouthpiece, 7 oxygen tubing and 6 flextube. The nebulizer was driven by compressed air (regulated to 50 psi) at 5 and 10 L/min flow. Ambient air, which flows through the open end of the T-adapter, was controlled at 50% ±10% relative humidity. For all cascade tests, the total mass of drug collected on all impactor stages and accessories were within 10% of initial dose. Definition of parameters measured 1. MMAD: Mass median aerodynamic diameter is a measure of central tendency of the size distribution of aerosol particles. It is the diameter, in micrometers, of which 50% of the mass of aerosol is larger and 50% is smaller. 2. GSD: Geometric standard deviation is a measure of the width of the size distribution of aerosol. For log normal distributions, it is calculated as follows 6 : Data presented: mean ±standard deviation Particle size An Andersen eight-stage cascade impactor (ACI) with USP inlet 6 was used and assayed with a spectrophotometer (277nm). Particle size characterization was performed using 3 ml of 0.083% albuterol in normal saline. Cascade impaction was chosen for the following reasons: GSD = (84.13% diameter/15.87% diameter). 5,6 3. Respirable fraction is the percent of aerosol generated by mass that falls below an aerodynamic diameter of 5μm MMAD. It has been reported that particles less than 5μm will penetrate beyond the upper airways and deposit into the tracheobronchial and pulmonary regions of the lung. 8 This measurement is often used to describe the quality of aerosol. 9 The closest cascade plate cutoff point of 4.7μm was utilized to quantify the mass below 5μm.

5 Output rate Nebulizers were weighed before and after aerosolizing a fill volume of 5 ml 0.083% albuterol for three min. Performance was reported in the average weight loss per minute (g/min). Testing was done with nebulizers randomly pulled from production lots. The nebulizer was operated with compressed air regulated to 50 psi, and 0.083% albuterol was the nebulized solution. Treatment time and residual The measurements of treatment time and residual will depend on the criteria used to determine the end of treatment. In clinical practice, it is common to tap the nebulizer near the end of the treatment to maximize the amount of drug nebulized. To simulate this practice, the nebulizer was periodically tapped when sputtering was observed for a maximum of two minutes or until tapping resulted in aerosol production for less than three sec. Treatment time was the total time required to nebulize 3 ml of solution. Residual was calculated by subtracting the weight of the nebulizer after the trial from the weight of the nebulizer empty. Treatment time and residual testing were done with nebulizers randomly pulled from production lots. The nebulizer was operated with compressed air and 0.083% albuterol was the nebulized solution. Operating pressure is an estimate of the quantity of respirable drug inhaled by the patient under typical clinical conditions. The concept of inhaled mass was originally introduced in 1991 by Smaldone 10 and has become a wellaccepted method for evaluating nebulizer performance. The test setup consisted of a filter placed between the mouthpiece of the nebulizer and a simulated breathing patient. The filter captured the inhaled aerosol exiting the mouthpiece and the drug mass deposited onto the filter was quantified with a UV spectrophotometer (277nm). To estimate the inhaled respirable mass, the drug mass deposited onto the filter was then multiplied by the corresponding respirable fraction. A simulated breathing patient was created by using a NPB 7200 ventilator and a Michigan Test Lung. An adult breathing circuit was connected to a ventilator that drove one side of the test lung. A bar was connected to the opposing bellows, thus duplicating the movement of the test lung, and thereby simulating a patient breathing pattern. A BioTek flow analyzer was placed between the simulated patient and the inhalation filter to monitor actual breathing parameters. The breathing pattern was held at a tidal volume of approximately 500 ml, 12 breaths per minute and an I:E ratio of 1:2, similar to a study presented by Siafaka, Hess and Biatello. 11 A commonly used dose of albuterol (3 ml, 0.083% albuterol) was nebulized until aerosol production ceased for 30 seconds or after two minutes of sputtering was observed, whichever came first. Collection of inhaled drug mass occurred over the duration of treatment. Testing was done with nebulizers randomly pulled from production lots. Product components used were from AirLife brand Misty Max 10 nebulizer : T-adapter, mouthpiece, 7 oxygen tubing and 6 flextube. Pressure was obtained from a T-connection between the compressed air flow meter and the nebulizer. Testing was done with nebulizers randomly pulled from production lots. Pressures at 5 and 10 L/min flow were evaluated while nebulizing 3 ml of 0.083% albuterol.

6 References 1 Dolovich, M. Assessing nebulizer performance. Respiratory Care, 2002; 47(11). 2 Siafake, A., Hess, D., Bigatello, L. Laboratory evaluation of new generation pneumatic nebulizers. 49th International Respiratory Congress, Abstracts 2003 Open Forum. Respiratory Care, Laube, B. In vivo measure of aerosol dose and distribution: Clinical relevance. Journal of Aerosol Medicine, 1996; 9(1). 4 Hess, D. Medication nebulizer performance. Chest, 1996: 110(2). 5 Consensus Statement: Aerosols and Delivery Devices. Respiratory Care, 2000; 45(6). 6 Hess, D. Medication nebulizer performance. Chest, 1996; 110(2): Aerosols. U.S. Pharmacopia, 23(601). 8 Laube, B. In vivo measurements of aerosol dose and distribution: Clinical relevance. Journal of Aerosol Medicine, 1996; 9(1). 9 Dolovich, M. Influences of inspiratory flow rate, particle size and airway caliber on aerosolized drug delivery to the lung. Respiratory Care, 2000; 45(6). 10 Smaldone, GC. Drug delivery via aerosol systems: Concept of aerosol inhaled. Journal of Aerosol Medicine, 1991; 4: Siafake, A., Hess, D., Bigatello, L. Laboratory evaluation of new generation pneumatic nebulizers. 49th International Respiratory Congress, Abstracts 2003 Open Forum. Respiratory Care, WARNING U.S. Federal Law restricts this device to sale by or on the order of a physician. Manufactured for: CareFusion Savi Ranch Parkway Yorba Linda, CA toll-free tel fax CareFusion France 309 S.A.S. 8 bis rue de la Renaissance F Châteaubriant, France 2011 CareFusion Corporation or one of its subsidiaries. All rights reserved. AirLife and Misty Max 10 are trademarks or registered trademarks of CareFusion Corporation or one of its subsidiaries. All other trademarks are property of their respective owners. RC2564 (0411/1000) L3384 Rev. A CareFusion Yorba Linda, CA carefusion.com

Rush University, College of Health Sciences

Rush University, College of Health Sciences Rush University, College of Health Sciences An Evaluation of the Prototype OxyMulti Mask Prototype Compared to the Oxymask Aerosol, OxyMulti Mask and Airlife Aerosol Mask for Aerosol Delivery in Adults

More information

Latex Free. An affordable, easy to use, high density, small volume nebulizer with a breath enhanced design! Breath Enhanced High Density Jet Nebulizer

Latex Free. An affordable, easy to use, high density, small volume nebulizer with a breath enhanced design! Breath Enhanced High Density Jet Nebulizer Latex Free Breath Enhanced High Density Jet Nebulizer The NebuTech HDN nebulizer, a breath enhanced design, by Salter Labs is quickly becoming the product of choice for caregivers and patients alike. This

More information

Delivery of Iloprost Inhalation Solution With the HaloLite, Prodose, and I-neb Adaptive Aerosol Delivery Systems: An In Vitro Study

Delivery of Iloprost Inhalation Solution With the HaloLite, Prodose, and I-neb Adaptive Aerosol Delivery Systems: An In Vitro Study Delivery of Iloprost Inhalation Solution With the HaloLite, Prodose, and I-neb Adaptive Aerosol Delivery Systems: An In Vitro Study Robert E Van Dyke MSc and Kurt Nikander BACKGROUND: Iloprost (Ventavis)

More information

Understanding cascade impaction and its importance for inhaler testing

Understanding cascade impaction and its importance for inhaler testing Understanding cascade impaction and its importance for inhaler testing Mark Copley, Technical Sales Manager Inhalation product development is an important area of activity for the pharmaceutical sector.

More information

Articles. The Advantages of Nebulization in the Treatment of Mechanically Ventilated Neonates. Kristin Smith, RRT-NPS

Articles. The Advantages of Nebulization in the Treatment of Mechanically Ventilated Neonates. Kristin Smith, RRT-NPS Articles The Advantages of Nebulization in the Treatment of Mechanically Ventilated Neonates Kristin Smith, RRT-NPS A major goal in the care of premature babies is growth, and so all therapies are applied

More information

Computational Fluid Dynamics Modeling of Amsino OneMask Oxygen Mask

Computational Fluid Dynamics Modeling of Amsino OneMask Oxygen Mask Computational Fluid Dynamics Modeling of Amsino OneMask Oxygen Mask Abstract This study s objective was to model the Amsino OneMask Oxygen Mask using Computational Fluid Dynamics (CFD). A three-dimensional

More information

Medical Nebulizer Performance: Effects of Cascade Impactor Temperature

Medical Nebulizer Performance: Effects of Cascade Impactor Temperature Medical Nebulizer Performance: Effects of Cascade Impactor Temperature Yue Zhou PhD, Amitkumar Ahuja MSc, Clinton M Irvin, Dean A Kracko, Jacob D McDonald PhD, and Yung-Sung Cheng PhD BACKGROUND: During

More information

Latex Free. An affordable, easy to use, high density, small volume nebulizer with a revolutionary breath enhanced design!

Latex Free. An affordable, easy to use, high density, small volume nebulizer with a revolutionary breath enhanced design! Latex Free Breath Enhanced High Density Jet Nebulizer The NebuTech HDN nebulizer, a revolutionary breath enhanced design, by Salter Labs is quickly becoming the product of choice for caregivers and patients

More information

JAMES B. FINK, RAJIV DHAND, JERRY GRYCHOWSKI, PATRICK J. FAHEY, and MARTIN J. TOBIN

JAMES B. FINK, RAJIV DHAND, JERRY GRYCHOWSKI, PATRICK J. FAHEY, and MARTIN J. TOBIN Reconciling In Vitro and In Vivo Measurements of Aerosol Delivery from a Metered-Dose Inhaler during Mechanical Ventilation and Defining Efficiency-enhancing Factors JAMES B. FINK, RAJIV DHAND, JERRY GRYCHOWSKI,

More information

Cascade Impactor Study Comparing the Amsino OneMask Oxygen Mask and Predicate Mask from Hudson RCI on a Human Model

Cascade Impactor Study Comparing the Amsino OneMask Oxygen Mask and Predicate Mask from Hudson RCI on a Human Model (p) 93-850-6630 (f) 93-850-6635 Cascade Impactor Study Comparing the Amsino OneMask Oxygen Mask and Predicate Mask from Hudson RCI on a Human Model Abstract This in vitro study compares drug delivery efficiencies

More information

15 cm Reservoir Tubing. 210 cm Tubing

15 cm Reservoir Tubing. 210 cm Tubing 12 Aerosol Therapy nebulisers AEROSOL THERAPY MICRO MIST SMALL VOLUME NEBULISER The MICRO MIST small volume nebuliser is designed for performance and economy and can be used for hand-held or in-line treatments.

More information

Invacare Aerosol Products. Stratos Aerosol Compressors Nebulizers Asthma Management

Invacare Aerosol Products. Stratos Aerosol Compressors Nebulizers Asthma Management Invacare Aerosol Products Stratos Aerosol Compressors Nebulizers Asthma Management Invacare Aerosol Products Invacare is proud to offer an updated aerosol product line that includes the Stratos Compact,

More information

In Vitro Evaluation of Positive Expiratory Pressure Devices Attached to Nebulizers

In Vitro Evaluation of Positive Expiratory Pressure Devices Attached to Nebulizers In Vitro Evaluation of Positive Expiratory Pressure Devices Attached to Nebulizers Ariel Berlinski MD BACKGROUND: Patients with cystic fibrosis perform airway clearance techniques and receive nebulized

More information

In Vitro Comparison of Aerosol Delivery Using Different Face Masks and Flow Rates With a High-Flow Humidity System

In Vitro Comparison of Aerosol Delivery Using Different Face Masks and Flow Rates With a High-Flow Humidity System In Vitro Comparison of Aerosol Delivery Using Different Face Masks and Flow Rates With a High-Flow Humidity System Hui-Ling Lin MSc RRT RN FAARC, Robert J Harwood MSA RRT FAARC, James B Fink PhD RRT FAARC,

More information

Nebulizers. Small Volume Nebulizers

Nebulizers. Small Volume Nebulizers Small Volume MICRO MIST Nebulizer The MICRO MIST small volume nebulizer is designed for performance and economy and includes the design features needed by today s respiratory care practitioners. It can

More information

Redefining Continuous Aerosol Drug Delivery 2015 PM223

Redefining Continuous Aerosol Drug Delivery 2015 PM223 Redefining Continuous Aerosol Drug Delivery 2015 PM223 Explanation of Continuous via Traditional Small Volume (Jet) Nebulizer Traditional small volume (jet) nebulizers require compressed gas to aerosolize

More information

Aerosol Delivery and Modern Mechanical Ventilation In Vitro/In Vivo Evaluation

Aerosol Delivery and Modern Mechanical Ventilation In Vitro/In Vivo Evaluation Aerosol Delivery and Modern Mechanical Ventilation In Vitro/In Vivo Evaluation Dorisanne D. Miller, Mohammad M. Amin, Lucy B. Palmer, Akbar R. Shah, and Gerald C. Smaldone Department of Respiratory Care,

More information

21/03/2011 AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY. Fundamentals of aerosols

21/03/2011 AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY. Fundamentals of aerosols AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY Steve Newman Scientific Consultant Norfolk, UK steve.newman@physics.org

More information

Effect of Interval Between Actuations of Albuterol Hydrofluoroalkane Pressurized Metered-Dose Inhalers on Their Aerosol Characteristics

Effect of Interval Between Actuations of Albuterol Hydrofluoroalkane Pressurized Metered-Dose Inhalers on Their Aerosol Characteristics Effect of Interval Between Actuations of Albuterol Hydrofluoroalkane Pressurized Metered-Dose Inhalers on Their Aerosol Characteristics Ariel Berlinski MD and David Pennington BACKGROUND: Albuterol hydrofluoroalkane

More information

Revision 2. Improving Aerosol Drug Delivery During Invasive Mechanical Ventilation with Redesigned Components

Revision 2. Improving Aerosol Drug Delivery During Invasive Mechanical Ventilation with Redesigned Components Revision 2 Improving Aerosol Drug Delivery During Invasive Mechanical Ventilation with Redesigned Components P. Worth Longest, Ph.D., 1,2* Mandana Azimi, Pharm.D., 2 Laleh Golshahi, Ph.D., 2 and Michael

More information

NebuTech nebulizer from Salter. Striving to be #1 IN PATIENT SATISFACTION. To help you and your patients. easy. breathe

NebuTech nebulizer from Salter. Striving to be #1 IN PATIENT SATISFACTION. To help you and your patients. easy. breathe nebulizer from Salter To help you and your patients breathe easy Striving to be #1 IN PATIENT SATISFACTION nebulizer from Salter Let go to work for you and your patients today. Performance Versatility

More information

Redefining Continuous Aerosol Drug Delivery PM223

Redefining Continuous Aerosol Drug Delivery PM223 Redefining Continuous Aerosol Drug Delivery PM223 Explanation of Continuous via Traditional Small Volume (Jet) Nebulizer Traditional small volume (jet) nebulizers require compressed gas to aerosolize medication

More information

Caption: The equipment required for testing Fluticasone Propionate (FP) Inhalation Powder in line with a new product-specific monograph (USP36-NF31).

Caption: The equipment required for testing Fluticasone Propionate (FP) Inhalation Powder in line with a new product-specific monograph (USP36-NF31). Product-specific FDA guidance, and product-specific pharmacopeial monographs, point to the use of test equipment, some of which isn t included in the general USP/Ph. Eur. chapters for orally inhaled products

More information

Aerosolized Antibiotics in Mechanically Ventilated Patients

Aerosolized Antibiotics in Mechanically Ventilated Patients Aerosolized Antibiotics in Mechanically Ventilated Patients Gerald C Smaldone MD PhD Introduction Topical Delivery of Antibiotics to the Lung Tracheobronchitis Aerosolized Antibiotic Delivery in the Medical

More information

MEET THE DRIVING FORCE IN NEBULIZER THERAPY... Matched Nebulizer Compressor System

MEET THE DRIVING FORCE IN NEBULIZER THERAPY... Matched Nebulizer Compressor System MEET THE DRIVING FORCE IN NEBULIZER THERAPY... Matched Nebulizer Compressor System Matched Nebulizer Compressor System 4 1 Air is passed through an inlet filter. 2 Clean air is then driven through the

More information

600 RESPIRATORY CARE MAY 2016 VOL 61 NO 5

600 RESPIRATORY CARE MAY 2016 VOL 61 NO 5 Quantifying Aerosol Delivery in Simulated Spontaneously Breathing Patients With Tracheostomy Using Different Humidification Systems With or Without Exhaled Humidity Arzu Ari PhD RRT PT CPFT FAARC, Robert

More information

...Porta-neb matched systems from Respironics (UK)

...Porta-neb matched systems from Respironics (UK) ...Porta-neb matched systems from Respironics (UK) It is important that the system you choose consists of a compressor and nebulizer designed to work together. The matching of a nebulizer and compressor

More information

Respiratory Care in PICU Aerosol Therapy ส พ ชชา ชา แสงโขต โรงพยาบาลสมเด จพระป นเกล า

Respiratory Care in PICU Aerosol Therapy ส พ ชชา ชา แสงโขต โรงพยาบาลสมเด จพระป นเกล า Respiratory Care in PICU Aerosol Therapy น.อ.หญ ง ส พ ชชา ชา แสงโขต โรงพยาบาลสมเด จพระป นเกล า Aerosol liquid droplets or solid particles suspended in a gas(air) visible like fog Aerosol vs Humidity Humidity

More information

Metered Dose Inhalers with Valved Holding Chamber: A Pediatric Hospital Experience

Metered Dose Inhalers with Valved Holding Chamber: A Pediatric Hospital Experience Metered Dose Inhalers with Valved Holding Chamber: A Pediatric Hospital Experience 8th Annual North Regional Respiratory Care Conference Minnesota & Wisconsin Societies for Respiratory Care Mayo Civic

More information

Small Volume Nebulizer Treatment (Hand-Held)

Small Volume Nebulizer Treatment (Hand-Held) Small Volume Aerosol Treatment Page 1 of 6 Purpose Policy Physician's Order Small Volume Nebulizer Treatment To standardize the delivery of inhalation aerosol drug therapy via small volume (hand-held)

More information

University of Groningen. Technology in practice Lexmond, Anne

University of Groningen. Technology in practice Lexmond, Anne University of Groningen Technology in practice Lexmond, Anne IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document

More information

NEBULIZERS, METERED DOSE INHALERS, AND DRY POWDER INHALERS

NEBULIZERS, METERED DOSE INHALERS, AND DRY POWDER INHALERS NEBULIZERS, METERED DOSE INHALERS, AND DRY POWDER INHALERS Douglas S. Gardenhire, Ed.D, RRT-NPS MODULE 1 Manipulate Small Volume Nebulizers by Order or Protocol 1 Objectives for Module 1 At the end of

More information

Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pmdi)

Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pmdi) Respiratory Drug Delivery Europe 2017 Nagel and Suggett Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pmdi) Mark W. Nagel and Jason A. Suggett

More information

MasterScope Body. Powered by SentrySuite

MasterScope Body. Powered by SentrySuite MasterScope Body Powered by SentrySuite For over half a century the JAEGER bodybox generations have set the standard for excellence in bodyplethysmography. Today, powered by the new SentrySuite platform,

More information

Blow-by as Potential Therapy for Uncooperative Children: An In-Vitro Study. Mohamed Mohsen Mansour MD and Gerald C Smaldone MD PhD

Blow-by as Potential Therapy for Uncooperative Children: An In-Vitro Study. Mohamed Mohsen Mansour MD and Gerald C Smaldone MD PhD Original Research Blow-by as Potential Therapy for Uncooperative Children: An In-Vitro Study Mohamed Mohsen Mansour MD and Gerald C Smaldone MD PhD BACKGROUND: Blow-by, a common form of nebulizer therapy,

More information

Go With the Flow REGULATORY LANDSCAPE. Mark Copley at Copley Scientific

Go With the Flow REGULATORY LANDSCAPE. Mark Copley at Copley Scientific Go With the Flow Image: Guzel Studio shutterstock.com Increasing global requirements for efficacious, inexpensive products to treat respiratory illnesses are driving the development of inhaled generics.

More information

Dolci U 1, Sidler-Moix AL 1, Di Paolo ER 1,Berger- Gryllaki 1 M, Pannatier A 1, Cotting J 2

Dolci U 1, Sidler-Moix AL 1, Di Paolo ER 1,Berger- Gryllaki 1 M, Pannatier A 1, Cotting J 2 SALBUTAMOL DELIVERY IN AN IN VITRO PAEDIATRIC VENTILATOR-LUNG MODEL: COMPARISON OF JET, ULTRASONIC AND MESH NEBULISERS Dolci U 1, Sidler-Moix AL 1, Di Paolo ER 1,Berger- Gryllaki 1 M, Pannatier A 1, Cotting

More information

Pediatrics in mechanical ventilation

Pediatrics in mechanical ventilation Pediatrics Optimization Intitulé du cours of aerosol therapy in mechanical ventilation Thèmes donnés Ermindo Di Paolo, PhD Departments of Pharmacy and Pediatrics Lausanne University Hospital Switzerland

More information

Draft Guidance on Fluticasone Propionate; Salmeterol Xinafoate. Fluticasone Propionate; Salmeterol Xinafoate. Powder/Inhalation

Draft Guidance on Fluticasone Propionate; Salmeterol Xinafoate. Fluticasone Propionate; Salmeterol Xinafoate. Powder/Inhalation Reprinted from FDA s website by EAS Consulting Group, LLC Contains Nonbinding Recommendations Draft Guidance on Fluticasone Propionate; Salmeterol Xinafoate This draft guidance, once finalized, will represent

More information

Introduc)on & Background

Introduc)on & Background Doug Gardenhire, EdD RRT- NPS FAARC Georgia State University Atlanta, GA January 2013 Introduc)on & Background The standard Circulaire II high- efficiency aerosol drug delivery system has been shown to

More information

IVIVC in Pediatric OIPs

IVIVC in Pediatric OIPs IPAC-RS/UF Orlando Conference 2014 March 20, 2014 IVIVC in Pediatric OIPs Herbert Wachtel Declaration of Conflicts of Interest H. Wachtel is employee of Boehringer Ingelheim Pharma GmbH & Co. KG, Germany.

More information

Anaesthetic and respiratory equipment Nebulizing systems and components

Anaesthetic and respiratory equipment Nebulizing systems and components Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 27427 Third edition 2013-12-15 Anaesthetic and respiratory equipment Nebulizing systems and components Matériel d anesthésie et de réanimation

More information

What Do Cascade Impaction Measurements Tell Us: In Vitro Aspects. Context of Presentation

What Do Cascade Impaction Measurements Tell Us: In Vitro Aspects. Context of Presentation What Do Cascade Impaction Measurements Tell Us: In Vitro Aspects Jolyon P. Mitchell, Ph.D. Trudell Medical International London, Canada IPAC-RS Conference November 2006 1 Context of Presentation This talk

More information

Effect of particle size of bronchodilator aerosols on lung distribution and pulmonary function in patients

Effect of particle size of bronchodilator aerosols on lung distribution and pulmonary function in patients Thorax 1987;42:457-461 Effect of particle size of bronchodilator aerosols on lung distribution and pulmonary function in patients with chronic asthma D M MITCHELL, M A SOLOMON, S E J TOLFREE, M SHORT,

More information

Effects of Heat and Moisture Exchangers and Exhaled Humidity on Aerosol Deposition in a Simulated Ventilator-Dependent Adult Lung Model

Effects of Heat and Moisture Exchangers and Exhaled Humidity on Aerosol Deposition in a Simulated Ventilator-Dependent Adult Lung Model Effects of Heat and Moisture Exchangers and Exhaled Humidity on Aerosol Deposition in a Simulated Ventilator-Dependent Adult Lung Model Arzu Ari PhD RRT PT CPFT FAARC, Khalid S Alwadeai MSc RRT-NPS, and

More information

AEROSOL THERAPY: THE PRACTICALITIES

AEROSOL THERAPY: THE PRACTICALITIES AEROSOL THERAPY: THE PRACTICALITIES Lester I. Harrison, PhD Section Head, Clinical Pharmacokinetics, 3M Pharmaceuticals, 3M Center 270-3S-05, St. Paul, MN, USA 55144 liharrison@mmm.com Introduction: Horses,

More information

Unit 5 Humidity/Aerosol Generators

Unit 5 Humidity/Aerosol Generators 5-1 Unit 5 Humidity/Aerosol Generators GOAL On completion of this unit, the student should have an understanding of the principles of operation and proper use of different humidity/aerosol generators.

More information

Performing a Methacholine Challenge Test

Performing a Methacholine Challenge Test powder for solution, for inhalation Performing a Methacholine Challenge Test Provocholine is a registered trademark of Methapharm Inc. Copyright Methapharm Inc. 2016. All rights reserved. Healthcare professionals

More information

Assessing Quality of Inhaled Products And Links to Efficacy and Safety

Assessing Quality of Inhaled Products And Links to Efficacy and Safety Assessing Quality of Inhaled Products And Links to Efficacy and Safety Prasad Peri, PhD ONDQA 2011 IPAC-RS Conference Bringing Value To The Patient In A Changing World March 30, 2011 1 Outline of the Presentation

More information

ISO INTERNATIONAL STANDARD. Anaesthetic and respiratory equipment Nebulizing systems and components

ISO INTERNATIONAL STANDARD. Anaesthetic and respiratory equipment Nebulizing systems and components INTERNATIONAL STANDARD ISO 27427 First edition 2009-05-01 Anaesthetic and respiratory equipment Nebulizing systems and components Matériel d'anesthésie et de réanimation respiratoire Systèmes de nébulisation

More information

Pulmo Waves Device for respiratory physiotherapy

Pulmo Waves Device for respiratory physiotherapy MADE IN ITALY Pulmo Waves Device for respiratory physiotherapy Respiratory physiotherapy Accumulation of secretions in certain parts of the human respiratory system can take place for different reasons.

More information

Intracheal antibiotics administration

Intracheal antibiotics administration Intracheal antibiotics administration Jean Chastre, M.D. www.reamedpitie.com Disclosure Conflicts of interest: Consulting or Lecture fees: Bayer, Pfizer, Cubist/Merck, Basilea, Kenta/Aridis, Roche, AstraZeneca/Medimmune

More information

Product Guide. MSP Corporation Rice Creek Parkway, Suite 300. Shoreview, Minnesota 55126, U.S.A. Phone: Fax:

Product Guide. MSP Corporation Rice Creek Parkway, Suite 300. Shoreview, Minnesota 55126, U.S.A. Phone: Fax: Product Guide inhalertesting MSP Corporation 5910 Rice Creek Parkway, Suite 300 Shoreview, Minnesota 55126, U.S.A. Phone: 651.287.8100 Fax: 651.287.8140 sales@mspcorp.com www.mspcorp.com About This Guide

More information

NIV and Aerosoltherapy

NIV and Aerosoltherapy NIV and Aerosoltherapy Workshop Jean-Bernard Michotte (Lausanne-CH) Simone Gambazza (Milano-IT) Jean-Bernard Michotte Haute Ecole de Santé Vaud, 1011 Lausanne - Suisse Cliniques Universitaires Saint-Luc,

More information

8/13/11. RSPT 1410 Humidity & Aerosol Therapy Part 3. Humidification Equipment. Aerosol Therapy

8/13/11. RSPT 1410 Humidity & Aerosol Therapy Part 3. Humidification Equipment. Aerosol Therapy 1 RSPT 1410 Humidity & Aerosol Therapy Part 3 Wilkins: Chapter 35, p. 775-799 Cairo: Chapter 4, p. 88-143 2 Humidification Equipment A humidifier is a device that adds molecular liquid (e.g. water vapor)

More information

Respiratory Drug Delivery. Be inspired. Nebulizer and compressor catalog

Respiratory Drug Delivery. Be inspired. Nebulizer and compressor catalog Respiratory Drug Delivery Be inspired Nebulizer and compressor catalog Introduction Inspiring innovation Philips is committed to enhancing patients lives by delivering cost-effective solutions that drive

More information

Assessing New Technologies: Patient-Device Interactions and Deposition

Assessing New Technologies: Patient-Device Interactions and Deposition Assessing New Technologies: Patient-Device Interactions and Deposition Gerald C Smaldone MD PhD Introduction: Principles The Inhaled Mass The Aerosol Deposition Measurement of Inhaled Mass Pediatric in

More information

Assessing the role of breathing simulators in OIP testing

Assessing the role of breathing simulators in OIP testing As first AppeAred in Inhalation April 2014 www.inhalationmag.com Assessing the role of breathing simulators in OIP testing Exploring how the application of patient-representative inhalation profiles can

More information

Pulmonary deposition of inhaled drugs

Pulmonary deposition of inhaled drugs Pulmonary deposition of inhaled drugs Federico Lavorini Dept. Experimental and Clinical Medicine Careggi University Hospital Florence - Italy Presenter Disclosures F.L. has received in the last 5 years

More information

In-Vitro Comparison of Six Continuous Nebulizers*

In-Vitro Comparison of Six Continuous Nebulizers* .. In-Vitro Comparison of Six Continuous Nebulizers* S. David Piper, PE** BACKGROUND: Continuous nebulization has become popular for treating patients with severe reversible airway disease. Continuous

More information

Characterizing the Performance of Metered Dose Inhalers with Add-On Devices: New Methods For Clinically Relevant Testing

Characterizing the Performance of Metered Dose Inhalers with Add-On Devices: New Methods For Clinically Relevant Testing INSTRUMENTATION» Characterizing the Performance of Metered Dose Inhalers with Add-On Devices: New Methods For Clinically Relevant Testing Mark Copley Director Copley Scientific The recent introduction

More information

Inhalation Product Research at FDA

Inhalation Product Research at FDA Inhalation Product Research at FDA Changning Guo Ph. D., Chemist Division of Pharmaceutical Analysis FDA/CDER/OPS/OTR 2016 GPhA CMC workshop, May 17, 2016 Disclaimer: This presentation reflects the views

More information

Comparison of patient spirometry and ventilator spirometry

Comparison of patient spirometry and ventilator spirometry GE Healthcare Comparison of patient spirometry and ventilator spirometry Test results are based on the Master s thesis, Comparison between patient spirometry and ventilator spirometry by Saana Jenu, 2011

More information

The Effects of Heliox on the Output and Particle-Size Distribution of Salbutamol using. Jet and Vibrating Mesh Nebulisers

The Effects of Heliox on the Output and Particle-Size Distribution of Salbutamol using. Jet and Vibrating Mesh Nebulisers The Effects of Heliox on the Output and Particle-Size Distribution of Salbutamol using Jet and Vibrating Mesh Nebulisers C O Callaghan 1 FRCPCH, DM, PhD., J White 1 C.Biol. M.I.Biol., J Jackson 1 B.Sc.,

More information

CLINICAL RELEVANCE OF IN VITRO PARTICLE SIZING DATA. Steve Newman, PhD Nottingham, UK November 2006

CLINICAL RELEVANCE OF IN VITRO PARTICLE SIZING DATA. Steve Newman, PhD Nottingham, UK November 2006 IPAC-RS Conference 6-8 November 2006 CLINICAL RELEVANCE OF IN VITRO PARTICLE SIZING DATA Steve Newman, PhD Nottingham, UK November 2006 IPAC-RS Conference 6-8 November 2006 1 EFFECT OF PARTICLE SIZE ON

More information

acapella vibratory PEP Therapy System Maximizing Therapy Effectiveness, Empowering Patient Compliance

acapella vibratory PEP Therapy System Maximizing Therapy Effectiveness, Empowering Patient Compliance acapella vibratory PEP Therapy System Maximizing Therapy Effectiveness, Empowering Patient Compliance Investigating Questions Each acapella vibratory PEP therapy system uniquely provides PEP therapy by

More information

Effect of mask dead space and occlusion of mask holes on delivery of nebulized albuterol. Ariel Berlinski, MD 1,2. Little Rock, Arkansas.

Effect of mask dead space and occlusion of mask holes on delivery of nebulized albuterol. Ariel Berlinski, MD 1,2. Little Rock, Arkansas. Effect of mask dead space and occlusion of mask holes on delivery of nebulized albuterol Ariel Berlinski, MD 1,2 1 University of Arkansas for Medical Sciences, Department of Pediatrics, Pulmonology Section

More information

Breath Actuated Nebulizer CONFIDENCE IN AEROSOL DELIVERY

Breath Actuated Nebulizer CONFIDENCE IN AEROSOL DELIVERY S T U D Y S U M M A R Y Use of the Brand of Breath Actuated Nebulizer CONFIDENCE IN AEROSOL DELIVERY * 1 AeroEclipse* Breath Actuated Nebulizer (BAN) and AeroEclipse* II BAN Study Summary Table of Contents

More information

Nebulized Albuterol Delivery in a Model of Spontaneously Breathing Children With Tracheostomy

Nebulized Albuterol Delivery in a Model of Spontaneously Breathing Children With Tracheostomy Nebulized Albuterol Delivery in a Model of Spontaneously Breathing Children With Ariel Berlinski MD BACKGROUND: Nebulized therapy is commonly used in spontaneously breathing tracheostomized patients, despite

More information

Compressor Nebulizer. Model UN-014

Compressor Nebulizer. Model UN-014 Compressor Nebulizer Model UN-014 Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione ITALIANO ESPAÑOL FRANÇAIS ENGLISH 1WMPD4003398

More information

Important Principles of Aerosol Therapy for Your Patients

Important Principles of Aerosol Therapy for Your Patients Important Principles of Aerosol Therapy for Your Patients Disclaimer Appendix 3 Declaration of Vested Interest Form Name of presenter: David Henry RRT Name of employer: DeVilbiss Healthcare Inc. Definition:

More information

Special Problems in Aerosol Delivery: Artificial Airways

Special Problems in Aerosol Delivery: Artificial Airways Special Problems in Aerosol Delivery: Artificial Airways Rajiv Dhand MD Introduction Factors Influencing Aerosol Delivery through Endotracheal Tubes Endotracheal Tube Size Endotracheal Tube Material/Design

More information

Albuterol Delivery via Facial and Tracheostomy Route in a Model of a Spontaneously Breathing Child

Albuterol Delivery via Facial and Tracheostomy Route in a Model of a Spontaneously Breathing Child Albuterol Delivery via Facial and Tracheostomy Route in a Model of a Spontaneously Breathing Child Brandy Cooper and Ariel Berlinski MD BACKGROUND: Some pediatric patients receiving therapeutic aerosols

More information

Factors affecting bronchodilator delivery in mechanically ventilated adults

Factors affecting bronchodilator delivery in mechanically ventilated adults LITERATURE REVIEW Factors affecting bronchodilator delivery in mechanically ventilated adults Arzu Ari and James B Fink ABSTRACT Background: Bronchodilators are increasingly being used in patients undergoing

More information

Spirometry and Flow Volume Measurements

Spirometry and Flow Volume Measurements Spirometry and Flow Volume Measurements Standards & Guidelines December 1998 To serve the public and guide the medical profession Revision Dates: December 1998 Approval Date: June 1998 Originating Committee:

More information

I. Subject: Medication Delivery by Metered Dose Inhaler (MDI)

I. Subject: Medication Delivery by Metered Dose Inhaler (MDI) I. Subject: Medication Delivery by Metered Dose Inhaler (MDI) II. Policy: Aerosol medication administration by metered dose inhaler will be performed upon a physician's order by Respiratory Therapy personnel.

More information

DeVilbiss. Aerosol.

DeVilbiss. Aerosol. DeVilbiss A E R O S O L F A M I L Y B R O C H U R E Aerosol www.sunrisemedicalonline.com PulmoMate The DeVilbiss PulmoMate compressor/ nebulizer system keeps pace with the world of managed care, providing

More information

In-vitro Comparison of 4 Large-Volume Nebulizers in 8 Hours of Continuous Nebulization

In-vitro Comparison of 4 Large-Volume Nebulizers in 8 Hours of Continuous Nebulization In-vitro Comparison of 4 Large-Volume Nebulizers in 8 Hours of Continuous Nebulization Ariel Berlinski MD, J Randy Willis RRT-NPS, and Tom Leisenring CRT BACKGROUND: Continuous albuterol nebulization has

More information

IPAC-RS/UF Orlando Inhalation Conference March 20, S.T. Horhota 1, C.B. Verkleij 2, P.J.G. Cornelissen 2, L. Bour 3, A. Sharma 3, M.

IPAC-RS/UF Orlando Inhalation Conference March 20, S.T. Horhota 1, C.B. Verkleij 2, P.J.G. Cornelissen 2, L. Bour 3, A. Sharma 3, M. IPAC-RS/UF Orlando Inhalation Conference March 20, 2014 Case Study: Pharmacokinetics and Pharmacodynamics of Tiotropium and Salmeterol Following Parallel Administration in COPD Patients Using Different

More information

MEDIUM-FLOW PNEUMOTACH TRANSDUCER

MEDIUM-FLOW PNEUMOTACH TRANSDUCER MEDIUM-FLOW PNEUMOTACH TRANSDUCER SS11LA for MP3x and MP45 System TSD117 & TSD117-MRI for MP150/MP100 System RX117 Replacement Airflow Head See also: AFT series of accessories for airflow and gas analysis

More information

IMPROVING THE REALISM AND RELEVANCE OF MOUTH-THROAT MODELS FOR INHALED PRODUCT TESTING

IMPROVING THE REALISM AND RELEVANCE OF MOUTH-THROAT MODELS FOR INHALED PRODUCT TESTING IMPROVING THE REALISM AND RELEVANCE OF MOUTH-THROAT MODELS FOR INHALED PRODUCT TESTING In this piece, Mark Copley, Sales Director of Copley Scientific, provides some background on mouth-throat models for

More information

REDUCING VARIABILITY IN TEST RESULTS FOR OINDPS WITH AUTOMATED ACTUATION

REDUCING VARIABILITY IN TEST RESULTS FOR OINDPS WITH AUTOMATED ACTUATION REDUCING VARIABILITY IN TEST RESULTS FOR OINDPS WITH AUTOMATED ACTUATION In this article, Heli Chauhan, Senior Applications Chemist, and Linda (Lingzhi) Liao, Field Applications Scientist, both of Proveris

More information

EVALUATION OF THE CYCLEX IMPACTOR PHASE 2: REPORT 3

EVALUATION OF THE CYCLEX IMPACTOR PHASE 2: REPORT 3 EVALUATION OF THE CYCLEX IMPACTOR PHASE 2: COLLECTION EFFICIENCY EVALUATION WITH MICROBIAL SPORES (IMPACTOR S CONFIGURATIONS 1, 2, AND 3) REPORT 3 Submitted to: ENVIRONMENTAL MONITORING SYSTEMS, INC. Prepared

More information

Adapting Pharmacopeial Testing for Pharmaceutical Equivalence Studies. Julie D. Suman, Ph.D. March 29, 2016

Adapting Pharmacopeial Testing for Pharmaceutical Equivalence Studies. Julie D. Suman, Ph.D. March 29, 2016 Adapting Pharmacopeial Testing for Pharmaceutical Equivalence Studies Julie D. Suman, Ph.D. March 29, 2016 AGENDA Regulatory Approaches for Bioequivalence (BE) Compendial Testing Vs Pharmaceutical Equivalence

More information

HCS70004P, HCS70004G HCS70004C, HCS70004BL HCS70004DP, HCS70004 AEROMIST COLORS NEBULIZER COMPRESSOR KIT. Instruction Manual

HCS70004P, HCS70004G HCS70004C, HCS70004BL HCS70004DP, HCS70004 AEROMIST COLORS NEBULIZER COMPRESSOR KIT. Instruction Manual HCS70004P, HCS70004G HCS70004C, HCS70004BL HCS70004DP, HCS70004 AEROMIST COLORS NEBULIZER COMPRESSOR KIT Instruction Manual TABLE OF CONTENTS IEC Symbols...2 Important Safeguards...2 Introduction...3 Specifications...4

More information

OPTIMISING ANALYTICAL STRATEGIES FOR THE DEMONSTRATION OF BIOEQUIVALENCE IN A GENERIC NEBULISER

OPTIMISING ANALYTICAL STRATEGIES FOR THE DEMONSTRATION OF BIOEQUIVALENCE IN A GENERIC NEBULISER OPTIMISING ANALYTICAL STRATEGIES FOR THE DEMONSTRATION OF BIOEQUIVALENCE IN A GENERIC NEBULISER US FDA guidance for the in vitro demonstration of bioequivalence in a generic nebuliser directly references

More information

OG TAL ODUCT CA PR e e HEALTH

OG TAL ODUCT CA PR e e HEALTH PRODUCT CATALOG e e HEALTH MODEL: OC 5 HEALTH OXYGEN CONCENTRATOR ACCESSORIES FILTER AIR FILTER NASAL CANNULA BUBBLE HUMIDIFIER Oxygen concentrator is a device that generates oxygen with a purity of %90-92

More information

PORTO 2 VENT CPAP OS. Operator s Manual. PORTO 2VENT CPAP OS System Operator s Manual Part Number Rev I

PORTO 2 VENT CPAP OS. Operator s Manual. PORTO 2VENT CPAP OS System Operator s Manual Part Number Rev I PORTO 2 VENT CPAP OS Operator s Manual 1 2 TABLE of CONTENTS 1. Introduction 3. Operating Instructions 1a. Definitions 3a. Setting the CPAP Level 1b. General Description 3b. Applying the Breathing Circuit

More information

Albuterol Delivery via Intrapulmonary Percussive Ventilator and Jet Nebulizer in a Pediatric Ventilator Model

Albuterol Delivery via Intrapulmonary Percussive Ventilator and Jet Nebulizer in a Pediatric Ventilator Model Albuterol Delivery via Intrapulmonary Percussive Ventilator and Jet Nebulizer in a Pediatric Ventilator Model Ariel Berlinski MD and J Randy Willis RRT-NPS BACKGROUND: In-line administration of bronchodilators

More information

Randomized Controlled Trial of a Breath-Activated Nebulizer in Patients with Exacerbation of COPD. Jeffrey M Haynes RRT RPFT

Randomized Controlled Trial of a Breath-Activated Nebulizer in Patients with Exacerbation of COPD. Jeffrey M Haynes RRT RPFT Original Research Randomized Controlled Trial of a Breath-Activated Nebulizer in Patients with Exacerbation of COPD Jeffrey M Haynes RRT RPFT BACKGROUND: Exacerbations of COPD (ECOPD) are characterized

More information

Flowmeters and. Specialty Products

Flowmeters and. Specialty Products Flowmeters and Specialty Products Chrome Flowmeter Durable, Accurate, Precise, Dependable 1MFA9005 4MFA1001 1MFA2001 1MFA1101 6MFA1001 1MFA3001 1MFA3501 1MFA1001 Specifications Model Gas Scale Increments

More information

Sleep Diagnostics. With your investment in the SomnoStar

Sleep Diagnostics. With your investment in the SomnoStar Sleep Diagnostics With your investment in the SomnoStar system, you are receiving not only the most sought after sleep system on the market today, but also a partnership with a company dedicated to serving

More information

I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device

I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device II. Policy: Continuous Positive Airway Pressure CPAP by the Down's system will be instituted by Respiratory Therapy personnel

More information

NomoLine No-moisture sampling lines for intubated and non-intubated patients in low- and high-humidity applications

NomoLine No-moisture sampling lines for intubated and non-intubated patients in low- and high-humidity applications NomoLine No-moisture sampling lines for intubated and non-intubated patients in low- and high-humidity applications Advanced technology enables cost-effective, hassle-free sidestream capnography and gas

More information

Considerations for Inhalation Safety Assessment: Approaches and Application

Considerations for Inhalation Safety Assessment: Approaches and Application Considerations for Inhalation Safety Assessment: Approaches and Application Madhuri Singal, PhD, RRT, DABT Inhalation Toxicologist, Senior Consumer Safety Associate Reckitt Benckiser, LLC CIR Expert Panel

More information

Randomized, Placebo-Controlled Clinical Trial of an Aerosolized Beta-2 Agonist for Treatment of Acute Lung Injury

Randomized, Placebo-Controlled Clinical Trial of an Aerosolized Beta-2 Agonist for Treatment of Acute Lung Injury Randomized, Placebo-Controlled Clinical Trial of an Aerosolized Beta-2 Agonist for Treatment of Acute Lung Injury The National Heart, Lung, and Blood Institute ARDS Clinical Trials Network The members

More information

Aerosol and Airway Clearance Therapies: Challenges and Opportunities. Growth of the Medicare. Aerosol and ACT Therapies Terry L. Forrette, M.H.

Aerosol and Airway Clearance Therapies: Challenges and Opportunities. Growth of the Medicare. Aerosol and ACT Therapies Terry L. Forrette, M.H. Baby Boomers: The Next Patient Population Aerosol and Airway Clearance Therapies: Challenges and Opportunities Terry L. Forrette, M.H.S., RRT The person sitting next to you may be your patient in the next

More information

Compressor Nebulizer System

Compressor Nebulizer System MedPro The Professional Choice Le choix des professionnels MedProTM MC Compressor Nebulizer System 705-445 Instruction Manual Index 1. Introduction...2 2. Product Identification...2 3. Important Safeguards...3

More information

Medescan Nebuliser Med - S600A Instructions Manual

Medescan Nebuliser Med - S600A Instructions Manual Medescan Nebuliser Med - S600A Instructions Manual Please read this guidebook carefully before operating this unit Your Nebuliser is intended for use in the treatment of asthma, COPD and other respiratory

More information

. Type of solution/medication. Amount/dose to be delivered. Frequency/duration. Mode of administration.

. Type of solution/medication. Amount/dose to be delivered. Frequency/duration. Mode of administration. Page 1 of 9 (Large, Medium Volume, and Aerogen Continuous Nebulizer) Purpose Scope Physician's Order To standardize the use of continuous aerosol therapy as a modality used for the rescue of patients with

More information