1 MVCT Image Guidance and QA Robert Staton, PhD DABR MD Anderson Cancer Center Orlando ACMP Annual Meeting 2011
2 Disclosures MDACCO has received grant funding from TomoTherapy, Inc.
3 Overview TomoTherapy MVCT review Recommended QA procedures Implications for Adaptive Radiotherapy
4 Megavoltage CT Imaging 6 MV Accelerator (tuned to 3.5 MV for MVCT) Primary Collimator 6mm slice width (1mm for 4.x software) 85cm Binary MLC (all open during MVCT) 85 cm Gantry Aperture 40 cm CT FOV Approximately 50cm
5 Spectral Energy Distribution
6 MVCT Imaging 3 choices of reconstructed slice thickness Fine (2mm) Normal (4mm) Coarse (6mm) Gantry Period = 10 sec Couch speed: Fine = 4 mm/rotation Normal = 8 mm/rotation Coarse = 12 mm/rotation Typical scan times 2-4 mins
7 Imaging Time: Examples Prostate: t Fine, 34 slices, 6.8 cm => 170 seconds Head and Neck: Normal, 42 slices, 16.8 cm => 210 seconds Thorax: Coarse, 44 slices, 26.4 cm => 220 seconds
8 Prostate Diagnostic KVCT TomoTherapy MVCT
9 Head and Neck Diagnostic KVCT TomoTherapy MVCT
10 Imaging Frequency At M. D. Anderson-Orlando, we MVCT for every tx What if. we had imaged only 1st day? we had imaged only first 5 days? we had imaged weekly?
11 Imaging Frequency Zeidan et al., IJROBP, 67(3), pp , 2007 Retrospective analysis of 24 HN patients (802 tx) All 802 fx were IGRT Replay different protocols: Use the daily MVCT images Calculate positioning errors for different frequencies of IGRT
12 Imaging Frequency IG % - Protocol A: no imaging (system. corr.) - Protocol B: Image 1, 1-3 3, 1-5 5,1-7 day - Protocol C: Weekly imaging 0 % 3-21% 20% - Protocol D: first 5, then weekly 31% - Protocol E: every other day IG, running average 50%
13 Imaging Frequency Results systematic errors are reduced with minimal image workload (i.e. 9% IGRT) Random errors are not reduced for non-ig treatments
14 Imaging Frequency % of Residual Error vs. % IGRT % Image guidanced > 3mm (all-tx) > 5mm (all-tx) > 3mm (non-ig) > 5mm (non-ig)
15 MVCT Imaging QA What do I need to QA? How often?
16 TG-148 Imaging QA Geometric Image Quality Imaging Dose
17 TG-148 Daily QA
18 Daily QA Example -Test imaging, registration, alignment chain 3) Align & test automatic couch setup 1) Scan 2)R Register & compare to known offsets Tolerance: Consistency within 1 mm
19 TG-148 Monthly
20 TG-148 Quarterly
21 Monthly QA Example One MVCT scan Geometric Noise Uniformity Spatial resolution MVCT HU Dose
22 Suggested Test & Tolerances Geometric Verify dimensions of phantom are consistent with known values Alignment of fiducials TG-148: 2mm/1mm (SRS/SBRT) Noise Select a consistent ROI in a homogeneous portion of the cheese phantom TG-148: Consistency with Baseline MDACCO Std Dev of Central ROI Reference value: ~40 HU
23 Suggested Test & Tolerances Uniformity Compare central and peripheral MVCT HU in uniform phantom TG-148 Compare to Baseline Within 25 HU if using MVCT for dose calculations
24 Suggested Test & Tolerances Noise Assess low contrast visibility using phantom with various density plugs TG-148: Compare to baseline For example, can you visualize the 1.02 density plug? Tomo spec: 3-cm object of 3% difference in electron density
25 Suggested Test & Tolerances Spatial Resolution TG mm high contrast object 3 rd row of contrast detail plug
26 Image Artifacts Button/Zipper Artifact Due to sharp dropoff of detector response in the center of the detector array Possible solutions Energy spectral recalibration Repositioning of detector
27 Suggested Test & Tolerances MVCT Dose Measure MSAD in phantom with ion chamber in a consistent location Typical values 1-3 cgy TG-148: compare to baseline MDACCO Scan of entire cheese phantom (18 cm) Normal cgy
28 Suggested Test & Tolerances MVCT HU calibration Important if using dose calculations on MVCT TG-148: Near Water within 30 HU of baseline Lung/Bone within 50 HU of baseline
29 The image cannot be displayed. Your computer may not have enough memory to open the image, or the image may have been corrupted. Restart your computer, and then open the file again. If the red x still appears, you may have to delete the image and then insert it again. TG-148 Annual QA Verification of imaging/treatment/laser coordinate system Design a phantom end-to-end test Film or diode array TG-148 tolerance: 2mm/1mm (SRS/SBRT)
30 Component Replacements Imaging QA is recommended after major component replacements Magnetron SSM Target Linac
31 Adaptive Radiotherapy Allows assessment of the dosimetric impact of anatomical changes during the course of treatment
32 Adaptive Protocol at MDACCO Daily MVCT scans are collected for H&N patients Images are run through a research version of TomoTherapy Planned Adaptive software Automatically ti deforms contours to MVCT images and recalculates dosimetry for that fraction Cumulative and daily DVHs can be used to tract changes in the dosimetry Replans are triggered by physician review
33 Planned Adaptive Research Version
34 Thorax Phantom DVH No Deformation
35 Previous MVCT Study Thorax Phantom DVH Previous Calculation K M Langen et al, The use of megavoltage CT (MVCT) images for dose recomputations, Phys Med Biol 50, (2005).
36 Thorax Phantom Density Comparison: Original IVDT
37 MVCT Uniformity Cup of water used to calibrate IVDT 27 HU On one machine, the HU value increased near the edge of the FOV 5 HU This is illustrated by the bottle of water shown at the right 15 HU 37 HU
38 Virtual Water (Cheese) CT Calibration Phantom Water plug shown at upper-right Cheese phantom Cheese phantom configuration shown at lower-right
39 Thorax Phantom Density Comparison: New IVDT
40 New Thorax Phantom DVH Less than 1% error at D 50
41 Tomo1 Temporal MVCT Number Variability 120 Tomo1 Solid Water Variation HU MVCT Number HU -20
42 Tomo2 Temporal MVCT Number Variability Tomo2 Solid Water Variation 100 HU MVC CT Number HU 0
43 Energy variation HI*ART II S/N 002, ENERGY CHECK Target changes DEPTH DOSE AT 15 cm Dec Jun Jan-04 1-Aug Feb-05 5-Sep Mar Oct Apr Nov-07 1J 08 DATE
44 Effect of Component Changes 110 Tomo1 Solid Water Variation for Ion Chamber Magnetron/Circulator Target MVC CT Number Magnetron/Target SSM/Gun board Magnetron Magnetron/Gun board
45 IVDT Curve Comparison IVDT curve appears to maintain shape over time 1000 The slope of the curve appears to 0 increase over time (with target wear) 200 MVCT Numb ber IVDT Temporal Variation Maximum Minimum There is greater variation in MVCT number for higher density materials Therefore, dosimetric errors will likely be greater in regions containing higher proportions of these materials Density (g/cc)
46 Dosimetric Error Due to Temporal MVCT Number Variability Solid Water MVCT Number PTV1 D 50 (Gy) Error 1.7% 1.2% 1.0% 0.6% 0.3% 0.0% 0.2% 0.5% 0.8% 1.2% Rt Parotid D 50 (Gy) Error 0.7% 07% 0.5% 05% 0.5% 05% 0.2% 02% 0.1% 01% 00% 0.0% 02% 0.2% 04% 0.4% 05% 0.5% 08% 0.8% Lt Parotid D 50 (Gy) Error 0.8% 0.6% 0.5% 0.5% 0.1% 0.0% 0.1% 0.2% 0.2% 0.4% All values shown are given for dose recalculations performed on a single MVCT of a head & neck patient Total D 50 variation in the primary target was ~3% Parotid D 50 variation was ~1.5% This is consistent with estimates based on simple physics models If the IVDT or MVCT images were never recalibrated, one could expect to limit the dosimetric error due to temporal HU variations to 3% If recalibrations were performed to maintain the solid water MVCT number within 30 HU of the commissioned IVDT value, dosimetric errors should be within ±1% These errors should increase as depth increases
47 Dosimetric Error Due to Temporal MVCT Number Variability Solid Water MVCT Number PTV1 D 50 (Gy) Error 1.7% 1.2% 1.0% 0.6% 0.3% 0.0% 0.2% 0.5% 0.8% 1.2% Rt Parotid D 50 (Gy) Error 0.7% 07% 0.5% 05% 0.5% 05% 0.2% 02% 0.1% 01% 00% 0.0% 02% 0.2% 04% 0.4% 05% 0.5% 08% 0.8% Lt Parotid D 50 (Gy) Error 0.8% 0.6% 0.5% 0.5% 0.1% 0.0% 0.1% 0.2% 0.2% 0.4% All values shown are given for dose recalculations performed on a single MVCT of a head & neck patient Total D 50 variation in the primary target was ~3% Parotid D 50 variation was ~1.5% This is consistent with estimates based on simple physics models If the IVDT or MVCT images were never recalibrated, one could expect to limit the dosimetric error due to temporal HU variations to 3% If recalibrations were performed to maintain the solid water MVCT number within 30 HU of the commissioned IVDT value, dosimetric errors should be within ±1% These errors should increase as depth increases
48 Dosimetric Error Due to Temporal MVCT Number Variability Solid Water MVCT Number PTV1 D 50 (Gy) Error 1.7% 1.2% 1.0% 0.6% 0.3% 0.0% 0.2% 0.5% 0.8% 1.2% Rt Parotid D 50 (Gy) Error 0.7% 07% 0.5% 05% 0.5% 05% 0.2% 02% 0.1% 01% 00% 0.0% 02% 0.2% 04% 0.4% 05% 0.5% 08% 0.8% Lt Parotid D 50 (Gy) Error 0.8% 0.6% 0.5% 0.5% 0.1% 0.0% 0.1% 0.2% 0.2% 0.4% All values shown are given for dose recalculations performed on a single MVCT of a head & neck patient Total D 50 variation in the primary target was ~3% Parotid D 50 variation was ~1.5% This is consistent with estimates based on simple physics models If the IVDT or MVCT images were never recalibrated, one could expect to limit the dosimetric error due to temporal HU variations to 3% If recalibrations were performed to maintain the solid water MVCT number within 30 HU of the commissioned IVDT value, dosimetric errors should be within ±1% These errors should increase as depth increases
49 Dosimetric Error Due to Temporal MVCT Number Variability Solid Water MVCT Number PTV1 D 50 (Gy) Error 1.7% 1.2% 1.0% 0.6% 0.3% 0.0% 0.2% 0.5% 0.8% 1.2% Rt Parotid D 50 (Gy) Error 0.7% 07% 0.5% 05% 0.5% 05% 0.2% 02% 0.1% 01% 00% 0.0% 02% 0.2% 04% 0.4% 05% 0.5% 08% 0.8% Lt Parotid D 50 (Gy) Error 0.8% 0.6% 0.5% 0.5% 0.1% 0.0% 0.1% 0.2% 0.2% 0.4% All values shown are given for dose recalculations performed on a single MVCT of a head & neck patient Total D 50 variation in the primary target was ~3% Parotid D 50 variation was ~1.5% This is consistent with estimates based on simple physics models If the IVDT or MVCT images were never recalibrated, one could expect to limit the dosimetric error due to temporal HU variations to 3% If recalibrations were performed to maintain the solid water MVCT number within 30 HU of the commissioned IVDT value, dosimetric errors should be within ±1% These errors should increase as depth increases
50 Dosimetric Error Due to Temporal MVCT Number Variability Solid Water MVCT Number PTV1 D 50 (Gy) Error 1.7% 1.2% 1.0% 0.6% 0.3% 0.0% 0.2% 0.5% 0.8% 1.2% Rt Parotid D 50 (Gy) Error 0.7% 07% 0.5% 05% 0.5% 05% 0.2% 02% 0.1% 01% 00% 0.0% 02% 0.2% 04% 0.4% 05% 0.5% 08% 0.8% Lt Parotid D 50 (Gy) Error 0.8% 0.6% 0.5% 0.5% 0.1% 0.0% 0.1% 0.2% 0.2% 0.4% All values shown are given for dose recalculations performed on a single MVCT of a head & neck patient Total D 50 variation in the primary target was ~3% Parotid D 50 variation was ~1.5% This is consistent with estimates based on simple physics models If the IVDT or MVCT images were never recalibrated, one could expect to limit the dosimetric error due to temporal HU variations to 3% If recalibrations were performed to maintain the solid water MVCT number within 30 HU of the commissioned IVDT value, dosimetric errors should be within ±1% These errors should increase as depth increases
51 Patient Example: Linac Change Linac Change Solid water: 36 HU to 8 HU 1.2% increase in dose to PTV1
52 Baseline MVCT Recalculation Error Water Head Thorax MVCT kvct Error MVCT kvct Error MVCT kvct Error D 95 (Gy) % % % D 50 (Gy) % % % D 05 (Gy) % % % MVCT images were obtained for three phantoms The cheese phantom was scanned at the same time to obtain an accurate IVDT and eliminate any temporal effects The greatest baseline error occurred for the head phantom at -1.3%
53 Recommendations MVCT recalibration frequency depends on the dosimetric error that one is willing to accept Based on this data, if the MVCT were never recalibrated, a maximum dosimetric error of ~4% would be possible Including baseline error and temporal variations We recommend recalibrating the MVCT to maintain the MVCT number of water within ±30 HU of the commissioned value This would result in an overall dosimetric error of not greater than ±2.5% for the typical head & neck patient This is consistent with TG-148 recommendations We recommend checking the IVDT after major component changes affecting the beamline There appears to be a consistent reduction in MVCT number after linac or target changes Magnetron changes are less consistent, but some large variations were observed Apart from major component changes, the IVDT should be checked on a monthly basis
54 Conclusions Imaging QA is important for maintaining an IGRT program More frequent QA is needed when using the images for dose calculations TG-148 is a good source of information for TomoTherapyTh users
55 Acknowledgements Katja Langen, PhD MDACCO, chair of TG-148 Jason Pukala, MS UF graduate student
The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist
The Radixact System Experience at Miami Cancer Institute Tino Romaguera, D.Sc. Senior Physicist Disclosure An honorarium is provided by Accuray for this presentation The views expressed in this presentation
More informationVarian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System
Varian Edge Experience Jinkoo Kim, Ph.D Henry Ford Health System Disclosures I participate in research funded by Varian Medical Systems. Outline of Presentation Review advanced imaging in Varian Edge Linear
More informationQuality Assurance of Helical Tomotherapy Machines
Quality Assurance of Helical Tomotherapy Machines Wolfgang Tomé 1,2, Hazim Jaradat 2, David Westerly 1, John Fenwick 3, Nigel Orton 2, Rock Mackie 1,2, and Bhudatt Paliwal 1,2 1 Human Oncology and 2 Medical
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationWhat do we Know About Adaptive Radiotherapy? Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, California
What do we Know About Adaptive Radiotherapy? Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, California AAMD Region II Meeting Houston, Texas September 13-14, 2012 Learning Objectives To gain
More information7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis
Acknowledgments Thanks to Saiful Huq for an illuminating conversation about the application of TG-100 Jennifer O Daniel, Ph.D. & Fang-Fang Yin, Ph.D. Duke University Medical Center Annual AAPM Meeting,
More informationSTEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS
PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic
More informationImage Registration for Radiation Therapy Applications: Part 2: In-room Volumetric Imaging
Image Registration for Radiation Therapy Applications: Part 2: In-room Volumetric Imaging Peter Balter Ph.D University of Texas M.D. Anderson Cancer Center Houston, TX, USA Disclosure Information Peter
More informationHelical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System
The Challenges Associated with Differential Dose Delivery using IMRT Chester Ramsey, Ph.D. Director of Medical Physics Thompson Cancer Center Knoxville, Tennessee, U.S.A Collaborators Chester Ramsey, Ph.D.
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More informationApplications of Modern Radiotherapy Systems
Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie Professor University of Wisconsin Co-Founder and Chairman of the Board TomoTherapy Inc. Financial Disclosure I am a founder and Chairman
More informationAAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction
AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall
More informationNormal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT
Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using
More informationVerification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,
More informationIMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor
IMRT/IGRT Patient Treatment: A Community Hospital Experience Charles M. Able, Assistant Professor Disclosures I have no research support or financial interest to disclose. Learning Objectives 1. Review
More informationUnrivaled, End-to-End
PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by
More informationTraceability and absorbed dose standards for small fields, IMRT and helical tomotherapy
Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationIGRT Solution for the Living Patient and the Dynamic Treatment Problem
IGRT Solution for the Living Patient and the Dynamic Treatment Problem Lei Dong, Ph.D. Associate Professor Dept. of Radiation Physics University of Texas M. D. Anderson Cancer Center Houston, Texas Learning
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationIn-Room Radiographic Imaging for Localization
In-Room Radiographic Imaging for Localization Fang-Fang Yin, Zhiheng Wang, Sua Yoo, Devon Godfrey, Q.-R. Jackie Wu Department of Radiation Oncology Duke University Medical Center Durham, North Carolina
More informationLearning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1
1 2 KV CBCT Imaging Part I Rabih Hammoud, MS, DABR Henry Ford Health System Detroit, Michigan Acknowledgements Indrin Chetty, PhD Teamour Nurushev, PhD Harrison Guan, PhD Jinkoo Kim, PhD JianYue Jin, PhD
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationLearning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy
3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy Sofie Ceberg, PhD Medical Physicist and Lund University Lund, Sweden Learning objectives 1: 3D Dosimetry in the Clinic: Background
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationPre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system
Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system E. Mezzenga, * E. Cagni, A. Botti, M. Orlandi and M. Iori Medical Physics Unit, IRCCS Arcispedale
More informationEvaluation of Dosimetry Check software for IMRT patient-specific quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,
More informationFeasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT
220 research article Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT Poonam Yadav 1,2,3, Velayudham Ramasubramanian 3, Bhudatt R. Paliwal 1,2 1 Department
More informationLearning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan
Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams
More informationIn-Room Radiographic Imaging for Localization
In-Room Radiographic Imaging for Localization Fang-Fang Yin, Zhiheng Wang, Sua Yoo, Devon Godfrey, Q.-R. Jackie Wu Department of Radiation Oncology Duke University Medical Center Durham, North Carolina
More informationVolumetric Modulated Arc Therapy - Patient Specific QA
Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck
More informationA VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI
More informationAre Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?
Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or
More informationWork partially supported by VisionRT
Work partially supported by VisionRT Background of frameless intracranial stereotactic radiosurgery UCSD SRS/SRT procedure Clinical Results Summary Total prescribed doses : order of 10 50 Gy Planning targets
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationAutomated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist
Automated Plan Quality Check with Scripting Rajesh Gutti, Ph.D. Clinical Medical Physicist Veera.Gutti@BSWHealth.org Outline Introduction - BSW Automation in Treatment planning Eclipse Scripting API Script
More informationPage 1. Helical (Spiral) Tomotherapy. UW Helical Tomotherapy Unit. Helical (Spiral) Tomotherapy. MVCT of an Anesthetized Dog with a Sinus Tumor
Helical (Spiral) Tomotherapy Novel Clinical Applications of IMRT Linac Ring Gantry CT Detector X-Ray Fan Beam Binary Multileaf Collimator Binary MLC Leaves James S Welsh, MS, MD Department of Human Oncology
More informationHelical Tomotherapy Research in London, Ontario. A Cancer Care Ontario Partner
Helical Tomotherapy Research in London, Ontario Objectives of this presentation To define tomotherapy To review the history of its development To review our experience with early tomotherapy installations
More informationSRS Uncertainty: Linac and CyberKnife Uncertainties
SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationLinac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR
Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,
More informationDisclosures. Clinical Implementation of SRS/SBRT. Overview. Anil Sethi, PhD. Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch
Clinical Implementation of SRS/SBRT Anil Sethi, PhD Loyola University Medical Center August 3, 2017 Disclosures Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch 2 Overview Physics Considerations
More informationIntroduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy
X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction
More informationReena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery
Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery Reena Phurailatpam Tejpal Gupta, Rakesh Jalali, Zubin Master, Bhooshan Zade,
More informationClinical Implementation of SRS/SBRT
Clinical Implementation of SRS/SBRT Anil Sethi, PhD, FAAPM Loyola University Medical Center November 4, 2017 Disclosures Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch 2 Learning
More informationEffective and efficient radiotherapy dosimetry audit: Where to next?
Effective and efficient radiotherapy dosimetry audit: Where to next? Catharine Clark, Eduard Gershkevitsh, Milan Tomsej and Nuria Jornet National Physical Laboratory Royal Surrey County Hospital Radiotherapy
More informationCan we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationImplementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL
Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau
More informationAnatomical-Based Adaptive RT: It Begins Here!
Anatomical-Based Adaptive RT: It Begins Here! Kristy K Brock, Ph.D., DABR Physicist, Radiation Medicine Program, Princess Margaret Hospital Associate Professor, Depts of Radiation Oncology & Medical Biophysics,
More informationCBCT of the patient in the treatment position has gained wider applications for setup verification during radiotherapy.
Gülcihan CÖDEL Introduction The aim of this study is to evaluate the changes in bladder doses during the volumetric modulated arc therapy (VMAT) treatment of prostate cancer patients using weekly cone
More informationIntensity Modulated RadioTherapy
Intensity Modulated RadioTherapy A clinical application of a small accelerator University Medical Center Groningen A.A. van t Veld PhD I. Hoveijn PhD part 1 part2 CERN/KVI accelerator school, Zeegse, June
More informationSpatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.
Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation
More informationQuality assurance and credentialing requirements for sites using inverse planned IMRT Techniques
TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery
More informationDELIVERED PATIENT DOSE
SOFTWARE DELIVERED PATIENT DOSE The only software that provides a true picture of patient and dose at each stage of treatment. Adaptivo patient dosimetry software does something other patient dosimetry
More informationph fax
This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY
More informationMAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:
SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,
More informationAgenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry
Controlli di qualità preclinici in IMRT statica e dinamica Michele Stasi S.C. Fisica Sanitaria, A.O. Ordine Mauriziano D.O. Fisica Sanitaria, IRCCS Candiolo TORINO mstasi@mauriziano.it Agenda 1. Introduction
More informationDosimetric Analysis Report
RT-safe 48, Artotinis str 116 33, Athens Greece RT-safe +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ------ Irradiation system: ------
More informationIROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015
IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During
More informationPresentation Outline. Patient Setup Imaging in RT: Getting the Most Bang for your Buck. Main Errors in RT. Hypothetical Patient Examples
Patient Setup Imaging in RT: Getting the Most Bang for your Buck Olivier Morin, PhD UC SF Comprehensive Cancer Center San Francisco, CA Presentation Outline Errors in RT. Radiographic films in RT. On-board
More informationVMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT
VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT Poster No.: C-0520 Congress: ECR 2011 Type: Scientific Exhibit Authors: Y. Kawasaki, S. Tadokoro,
More information8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation
Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research
More informationA new geometric and mechanical verification device for medical LINACs
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein
More informationObjectives. Image-Guided RT: Process. kv CBCT Novel Clinical Applications and Future Directions. Standard Clinical Uses. Novel Applications.
kv Novel Clinical Applications and Future Directions Douglas J. Moseley PhD, DABR Department of Radiation Oncology University of Toronto Radiation Medicine Program Princess Margaret Hospital Objectives
More informationL IGRT IN TOMOTERAPIA: Aspetti Clinici e Fisici. N. Di Muzio Istituto Scientifico S. Raffaele- Milano
L IGRT IN TOMOTERAPIA: Aspetti Clinici e Fisici N. Di Muzio Istituto Scientifico S. Raffaele- Milano Napoli, 8-9 Giugno 2007 Reduce the uncertainty of microscopic extension Precise localization of the
More informationTreatment Efficiency and Optimization of Patient Care with IBA ProteusOne
Treatment Efficiency and Optimization of Patient Care with IBA ProteusOne Terry Wu, Ph.D. Chief Physicist, Radiation Oncology Department Willis-Knighton Cancer Center/Proton Therapy Center Shreveport,
More informationAn Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center
An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More information8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit
MR-Linac is a research programme. It is not available for sale and its future availability cannot be guaranteed 8/2/2017 Clinical Implementation of an MR-Guided Treatment Unit Geoffrey S. Ibbott, Ph.D.
More informationUNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies
UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies A SINGLE INSTITUTION S EXPERIENCE IN DEVELOPING A PURPOSEFUL AND EFFICIENT OFF-LINE TECHNIQUE FOR ADAPTIVE RADIOTHERAPY IN A CLINICAL ENVIRONMENT A Research
More informationOverview of MLC-based Linac Radiosurgery
SRT I: Comparison of SRT Techniques 1 Overview of MLC-based Linac Radiosurgery Grace Gwe-Ya Kim, Ph.D. DABR 2 MLC based Linac SRS Better conformity for irregular target Improved dose homogeneity inside
More informationContributors to this Talk
Tomotherapy Thomas Rockwell Mackie Tomotherapy Research Group Depts. Of Medical Physics, Human Oncology and Biomedical Engineering University of Wisconsin Madison WI 53706 Phone: (608) 262-7358 Email:
More informationComparison of TLD dose and reconstructed dose for post-mastectomy radiation therapy with TomoTherapy
Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2011 Comparison of TLD dose and reconstructed dose for post-mastectomy radiation therapy with TomoTherapy Matthew Roberts
More informationRadiotherapy In-vivo Treatment Verification: Multi-center analysis for more than 6 million fractions delivered
Radiotherapy In-vivo Treatment Verification: Multi-center analysis for more than 6 million fractions delivered Gustavo Olivera, Daniel Galmarini, Xiaohu Mo, Donald Parnell, Stephanie Key, Yu Chen, Daniel
More informationbiij Initial experience in treating lung cancer with helical tomotherapy
Available online at http://www.biij.org/2007/1/e2 doi: 10.2349/biij.3.1.e2 biij Biomedical Imaging and Intervention Journal CASE REPORT Initial experience in treating lung cancer with helical tomotherapy
More informationImplementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy
Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out
More informationAccuray Radixact at University of WI: The next chapter in TomoTherapy innovation
Accuray Radixact at University of WI: The next chapter in TomoTherapy innovation Jennifer Smilowitz, Ph.D. Clinical Associate Professor Departments of Human Oncology and Medical Physics University of Wisconsin,
More informationQuality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose
Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel
More informationToshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery
Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,
More informationTHE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS
THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationImage Fusion, Contouring, and Margins in SRS
Image Fusion, Contouring, and Margins in SRS Sarah Geneser, Ph.D. Department of Radiation Oncology University of California, San Francisco Overview Review SRS uncertainties due to: image registration contouring
More informationImage Guided Stereotactic Radiotherapy of the Lung
Image Guided Stereotactic Radiotherapy of the Lung Jamie Marie Harris, MS DABR Avera McKennan Radiation Oncology September 25, 2015 Stereotactic Body Radiotherapy - Clinical Dose/Fractionation - Normal
More informationCURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT
CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT Purpose The purpose of this curriculum outline is to provide a framework for multidisciplinary training for radiation oncologists, medical
More informationA quality assurance tool for helical tomotherapy using a step-wedge phantom and the on-board MVCT detector
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 1, 2012 A quality assurance tool for helical tomotherapy using a step-wedge phantom and the on-board MVCT detector Vincent Althof, 1a Paul
More informationFrom position verification and correction to adaptive RT Adaptive RT and dose accumulation
From position verification and correction to adaptive RT Adaptive RT and dose accumulation Hans de Boer Move away from Single pre-treatment scan Single treatment plan Treatment corrections by couch shifts
More informationMany vendors are beginning to allow couch motion during radiation delivery.
Dynamic Couch Motion Many vendors are beginning to allow couch motion during radiation delivery. Varian developer mode allows institutions to perform research using these types of treatments. Tomotherapy
More informationHot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis
Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline
More informationAdditional Questions for Review 2D & 3D
Additional Questions for Review 2D & 3D 1. For a 4-field box technique, which of the following will deliver the lowest dose to the femoral heads? a. 100 SSD, equal dmax dose to all fields b. 100 SSD, equal
More informationAdvanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*
Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark* Purpose: To evaluate an institution s 3D treatment planning process and the institution
More informationImaging e tecnologia: cosa c è dietro l angolo?
Reggio Emilia, 17 Aprile 2010 Imaging e tecnologia: cosa c è dietro l angolo? Claudio Fiorino Fisica Sanitaria Istituto Scientifico San Raffaele, Milano Premessa Imaging e RT.una lunga storia.. La RT attuale
More informationImaging Rotation. University of Michigan Department of Radiation Oncology Division of Radiation Physics. Resident:
University of Michigan Department of Radiation Oncology Division of Radiation Physics Imaging Rotation Resident: Rotation staff mentor/ advisor: James Balter, supplemental mentors: Dale Litzenberg, Don
More information8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient
Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,
More informationQuality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.
Quality Assurance of Ultrasound Imaging in Radiation Therapy Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Louis, Missouri Typical Applications of Ultrasound Imaging in Radiation Therapy
More informationSBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session
Stereotactic Body Radiation Therapy Quality Assurance Educational Session J Perks PhD, UC Davis Medical Center, Sacramento CA SBRT fundamentals Extra-cranial treatments Single or small number (2-5) of
More informationCredentialing for the Use of IGRT in Clinical Trials
Credentialing for the Use of IGRT in Clinical Trials James M. Galvin, DSc Thomas Jefferson University Hospital Jefferson Medical College Philadelphia, PA and The Radiation Therapy Oncology Group RADIATION
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationCyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017
CyberKnife Technology in Ablative Radiation Therapy Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 Objectives Components and work flow of CyberKnife Motion management of CyberKnife
More informationMeasurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry
Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Conflict of interest None Acknowledgement: Jim
More information