Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage

Size: px
Start display at page:

Download "Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage"

Transcription

1 JIOS /jp-journals RESEARCH ARTICLE Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage 1 PS Vijayalakshmi, 2 AS Veereshi, 3 Vijay P Jayade, 4 MR Dinesh, 5 Mukesh Kumar ABSTRACT Biomechanical influences on bone structure play an important role in the longevity of bone around an implant. The quantity and direction of applied force influence the implant and cause deformation of the bone. FEA was used to analyze the changes in the bone on loading the implant with orthodontic force in oblique and vertical directions and orthopedic force. The Mini-implants used in the present study efficiently resisted the oblique loading. But their use, for the purpose of orthopedic loading is questionable. FE models showed the area with the highest stress and strain to be around the neck of the implant and the surrounding bone at the cervical margin. Keywords: Finite element analysis, Mini-implant, Stress, Strain, Cortical bone, Trabecular bone. How to cite this article: Vijayalakshmi PS, Veereshi AS, Jayade VP, Dinesh MR, Kumar M. Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage. J Ind Orthod Soc 2012;46(4): INTRODUCTION Efficient attainment and control of anchorage is fundamental to successful orthodontic and dentofacial orthopedic treatment. Adequate anchorage to correct dental or skeletal malocclusions is often a critical consideration in treatment planning. Because of anchorage limitations, we may have to settle for a compromised treatment alternative, or more complicated treatment alternatives like extraoral traction devices (which heavily depend on patient s compliance) or orthognathic surgery. 1 Implants are an excellent alternative to traditional orthodontic anchorage methodologies, and they are a necessity when dental elements lack quantity or quality, when extraoral devices are impractical, or when noncompliance during treatment is likely. In the last few years, implants of smaller sizes have been specially designed for orthodontic anchorage. These miniimplants are small enough for placement at any surface of the 1 Reader, 2 Associate Professor, 3 Former Professor and Head, 4 Professor and Head, 5 Professor 1,4 Department of Orthodontics, DAPM RV Dental College, Bengaluru Karnataka, India 2,5 Department of Orthodontics, Rungta College of Dental Sciences Bhilai, Chhattisgarh, India 3 Department of Orthodontics, SDM College of Dental Sciences, Dharwad Karnataka, India Corresponding Author: PS Vijayalakshmi, Reader, Department of Orthodontics, DAPM RV Dental College, Bengaluru, Karnataka, India drvijaya_ps@yahoo.com Received on: 20/12/11 Accepted after Revision: 20/6/12 alveolar process, even in the interdental areas. They are relatively inexpensive, and techniques for their placement and retrieval are simple. 2 Several researchers have reported studies related to the use of implants for orthodontic anchorage When using any of the load bearing implants in the bone, the stress and strain distribution in the bone around the implant must be understood. Biomechanical influences on bone structure play an important role in the longevity of bone around an implant. The quantity and direction of applied force influence the implants, and cause deformation of the bone. Stress analyses of endosseous implants are necessary for the investigation of bone turnover and maximum anchorage possible. Incorrect loading or overloading may lead to disturbed bone turnover and consequent implant loss. 26 Since, clinical determination of stress and strain distribution in the bone is not possible, an alternative technique should be used. The finite element method (FEM), which has been successfully applied to the mechanical study of stresses and strains in the field of engineering, makes it practicable to elucidate stresses in the living structures caused by various internal and external forces. Finite element analysis (FEA) has become an increasingly useful tool for the prediction of the effects of stress on the implant and its surrounding bone. The key factor for the success or failure of dental implant is the manner in which the stresses are transferred to the surrounding bone. FEA allows predicting stress distribution in the contact area of the implant with the cortical bone, and around the apex of the implant in the trabecular bone. 27 Numerous investigations 28,29 have been conducted to assess the stress and strain distribution that occurs around larger/bulkier prosthodontic implants, but limited literature is available regarding the same about mini-implants used for The Journal of Indian Orthodontic Society, October-December 2012;46(4):

2 PS Vijayalakshmi et al orthodontic anchorage. Also, the nature of loading of prosthodontic implants would be different compared to those used for orthodontic purposes. Hence, it was felt that there is a need to explore the changes in the bone adjacent to the implant following orthodontic loading. Therefore, the present study was undertaken with the following aims and objectives: 1. To analyze, by applying the FEM the stress and strain distribution patterns in the bone surrounding a miniimplant. 2. To investigate the deformation of the bone around an implant in response to forces of different magnitudes and applied in different directions. 3. To derive clinical implications, which may aid in optimization of implant design and its usage. MATERIALS AND METHODS The study was done in two parts as follows: 1. Data collection 2. FEM analysis Data Collection Data collection involved four steps: a. In the first step, a case was selected for the placement of the mini-implants. A patient, who presented with severe bimaxillary dentoalveolar protrusion and had critical anchorage requirements (needing complete utilization of the extraction spaces for retraction) was selected- an ideal case indicated for placing mini-implants for anchorage. 29 The implants were fixed using the following surgical procedure (as recommended by the authors) 30 in the interdental region between the 2nd premolar and the 1st molar (Figs 1A and B) and the insertion site chosen was measured from a guide bar on the bite-wing X-ray. To avoid any damage to the roots, the screws were implanted at a 60 angle between the teeth to the occlusal plane. 29 At the same time, the implants maintained an angle of approximately 30 to 40 to the long axis of the maxillary teeth and 10 to 20 to the mandibular teeth, to ensure optimum retention by augmenting the area of contact between the implant and adjacent bone (Figs 1A and B). b. The second step of the study was to evaluate the implant to bone integration. After 3 weeks of healing period, before loading the implants, the Dentascan (Figs 2A and B) was taken to ascertain the implant to bone integration. The data was also used to measure the thickness and depth of the cortical and trabecular bone around the implant. c. The third step was to determine the dimensions of the implant, so that these measurements could be used for modeling the implant. The dimensions of the implant were measured directly using a micrometer (to measure the pitch of the screw) and Vernier calipers (to measure the vertical and horizontal dimensions) Using the micrometer (0.25 range), the diameter at the tip, diameter at the first pitch, and the diameter at the tip of the implant were measured. The Vernier calipers of 8" were used to measure the vertical and horizontal dimensions. Along with the micrometer, a threading pitch gauge was also used to measure the pitch of the screw. d. Finally, the thickness of the bone surrounding the implant was measured using the implant assessment algorithm provided in the resident Dentascan software. As per the Dentascan, cortical bone thickness was 2.6 mm and cancellous bone thickness was 2.4 mm. Figs 1A and B: Mini-implants placed in the interdental region between the second premolar and the first molar region in a patient having critical anchorage requirement 176 JAYPEE

3 JIOS Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage Figs 3A and B: Models of implant and implant-cortical-trabecular bone in hypermesh FE modeling package Assigning the material properties: In this study, the assumption was made that the materials were homogenous and linear and that they had elastic material behavior characterized by 2 material constants viz. Young s modulus and Poisson s ratio. Figs 2A and B: Dentascan of the same patient, which was used to evaluate the implant to bone integration and to measure the thickness of the cortical and trabecular bone Finite Element Analysis Software The initial modeling was done using hypermesh FE modeling package and AutoCAD mechanical. The finite element analysis was done using MSC Patran version 3 as the pre- and postprocessor and MSC Nastran solver for Windows XP professional. Modeling the Mini-implant and the Surrounding Bone The modeling was done using the software, Hypermesh FE modeling package. Once the dimensions of the bone and the implant were obtained, these values were fed as input in both X and Y dimensions into the modeling software. These points were joined by lines to create the 2D crossection of the implant and the bone. Then, this crossection was revolved 360 to get 3D model (Figs 3A and B). The completed 3D model was then converted into a MSC Nastran input file (dat format) and imported into MSC Patran. The final model had hexahedral elements and nodes. Boundary conditions: Since the implant was assumed to be rigidly anchored in the bone, the entire outer surface of the bone was restrained from translation along all three axes, and restrained from rotations around all three axes. The FE Model Two types of FE models of the implant bone complex were generated (Figs 3A and B) Type I: Implant being surrounding completely by cortical bone. Type II: Implant being surrounded 3/4th by cortical bone and the tip being embedded in the trabecular bone. Implant Loading The effects of 2 different static loading conditions were planned to be examined. These loads were to be applied to simulate the following clinical situations: Orthodontic loading: Intrusion and retraction/oblique Vertical force Orthopedic loading ( as in protraction of maxilla). Magnitudes of Force Following force values were applied: Orthodontic loading: Intrusion and retraction/oblique: 150, 200 and 250 gm. Vertical force: 45, 60, 75 and 90 gm Orthopedic loading (as in protraction of maxilla): 400, 500 and 600 gm. The displacements produced by each of the force sets, and the stresses generated in the implant and in the surrounding The Journal of Indian Orthodontic Society, October-December 2012;46(4):

4 PS Vijayalakshmi et al bone were studied. Maximum and minimum principal stresses produced in each of the said structures were tabulated. Along with these, the maximum and minimum principal strain was also tabulated. A current measure for small strain values is strain µeps, where a strain of 1 means a deformation of 100%. Thus, for example, a strain of 100 µeps coincides with a strain of or 0.001% ( = 10 6 ). These values obtained were then correlated with Frost s Mechanostat theory (1994) 30 given in the table below. The compressive stresses were greater than the tensile stresses. These compressive and tensile stresses gradually reduced toward the tip of the implant. b. Distribution of stresses in the bone: Bone experienced a similar pattern of stress distribution, tensile stresses on RESULTS Interpretation of Results The results obtained are discussed under three headings: 1. Distribution of maximum and minimum principal stresses in the implant and bone. 2. Displacement of the implant and bone 3. Distribution of maximum and minimum principal strain in the implant-bone junction and bone. These parameters are analyzed for both types I and II FE models, under 3 different loading conditions. Figures 6 to 9 are graphical representation of the results obtained. Fig. 5: Distribution of maximum principal (compressive) strain in the cortical bone, on application of an orthopedic load of 400 gm to the type I FE model OBLIQUE LOADING OF THE IMPLANT FOR INTRUSION AND RETRACTION Implant in Cortical Bone: Type I FE Model The oblique force, when applied to the implant, caused the following stress distribution pattern: a. Distribution of stresses in the implant: Compressive stresses were seen on the mesial surface of the implant, with the maximum compressive stress being concentrated at the neck of the implant (Fig. 4).The tensile stresses were distributed on the opposite side of force application, i.e. on the distal surface. The maximum tensile stresses were concentrated at the point of force application. Fig. 6: Maximum and minimum principal stresses at the implantcortical bone junction for both type I and type II FE models Fig. 4: Distribution of maximum principal stresses in the implant, on application of an oblique force of 250 gm to the type I FE model Fig. 7: Maximum and minimum principal stresses experienced by the implant for both type I and type II FE models 178 JAYPEE

5 JIOS Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage The maximum principal strain further increased as the magnitude of force was increased. These strain values were correlated with the Frost mechanostat theory (1994). 29 On comparison, it was found that the strain values at implant-bone interface were in the range required for physiologic bone maintenance. In the deeper layers of the bone, these strain values decreased significantly. Implant in Cortical and Trabecular Bone: Type II FE Model Fig. 8: Maximum and minimum principal strains at the implant-cortical bone junction for both type I and type II FE models Fig. 9: Maximum and minimum principal strains in the cortical bone for both type I and type II FE models the distal aspect and compressive stresses on the mesial surface. The cortical bone around the implant revealed that the cervical margin and the bone around the first thread of the implant were the most stressed areas. c. Distribution of stresses at the implant-bone interface: At the junction of the implant and bone, maximum compressive stresses were seen. The 1st thread of the implant had maximum compressive stress. The stresses gradually reduced from the 1st ring to the 5th ring. The pattern of stress distribution was similar for application of 150, 200 and 250 gm loads. Displacement of the Implant and the Bone On application of oblique force to the implant, the implant showed a very negligible amount of displacement. The implant tended to tip toward the direction of force applied. This displacement was in the range of to mm. Distribution of Strain On application of force on the implant, it created an area of maximum strain at the junction of the implant and the bone. On oblique loading, the implant in the type II model experienced the same magnitude of stress as the type I model. The concentration of the stresses was seen at the neck of the implant, with the maximum tensile stress at the point of force application. The cortical bone experienced lower compressive and tensile stresses as compared to the type I model. However, unlike in model I, there were no significant differences between the tensile and the compressive stresses in the bone. The trabecular bone experienced lower stresses when compared to the cortical bone. But in contrast to the cortical bone, compressive stresses were slightly greater than the tensile stresses. Within the trabecular bone, maximum stress concentration of both tensile and compressive stresses was seen at the corticotrabecular junction. Implant-cortical bone junction experienced greater stresses when compared to the implant-trabecular bone junction. At both the interfaces, tensile stresses were greater than the compressive stresses. The implant in the type II model experienced a greater displacement when compared to the type I model. The magnitude of displacement increased in all the areas with the increase in the load applied. On loading the implant, the maximum strain was found at the implant-cortical bone junction, near the point of force application. Overall, the bone experienced greater tensile strain than compressive strain, which reduced gradually to the last layer of bone to a negligible amount. These strain values were correlated with the Frost s mechanostat Table 1, similar to the type I model. The correlation revealed that the strain values at implant-cortical bone interface and in the cortical bone were well within the optimum range for bone maintenance. The strain values in the trabecular bone and the implant-trabecular bone interface were significantly lower. According to the Frost s theory, such low strain values are not conducive for bone maintenance. Factor Table 1: Frost s mechanostat values Mechanical Disuse atrophy <200 Bone maintenance 200 to 2500 Physiologic hypertrophy 2500 to 4000 Pathologic overload > micron ( ) = 10 6 Peak load in microstrain ( E) The Journal of Indian Orthodontic Society, October-December 2012;46(4):

6 PS Vijayalakshmi et al VERTICAL LOADING OF THE IMPLANT FOR INTRUSION Implant in Cortical Bone: Type I Model The vertical force, applied to the implant to cause intrusion, created areas of greater tensile stresses than compressive stresses. The implant, cortical bone and the junction between implant and bone experienced greater tensile than compressive stresses,unlike the oblique loading, where the implant experienced greater compressive stresses than tensile stresses. On application of vertical force to the implant, the implant showed a very negligible amount of displacement. This displacement was very minimal in the range of to mm. Vertical loading of the implant created an area of maximum strain at the junction of the implant and the bone, concentrating at the point of force application. The maximum principal strain further increased as the magnitude of force was increased. The strain values gradually decreased in the deeper layers of the bone (Table 1). Implant in Cortical and Trabecular Bone: Type II Model The implant experienced the same magnitude of stresses as the type I model. The concentration of the stresses was seen at the neck of the implant, with maximum tensile stress at the point of force application. At the junction of implant and cortical bone, the stresses were slightly lower than the type I model. The cortical bone experienced similar compressive and tensile stresses when compared to the type I model, with tensile stresses were greater than the compressive stresses. Highest tensile stresses were concentrated at the implantcortical bone junction. The trabecular bone experienced lower stresses when compared to the cortical bone. Both the tensile and the compressive stresses were in the same range. Greater concentration of both tensile and compressive stresses was seen at the corticotrabecular junction. The implant in the Type II model displaced to the same extent as the Type I model. The cortical bone showed displacement slightly greater than the Type I model in the range of to mm with maximum displacement at the implant-cortical junction.the trabecular bone displaced to a negligible amount in the range of to mm. On loading the implant, the maximum strain was found at the implant-cortical bone interface, similar to the type I model. Overall, the bone experienced greater tensile strain than compressive strain, which reduced gradually to the last layer of bone. ORTHOPEDIC LOADING OF THE IMPLANT Relatively heavy forces (400, 500 and 600 gm) were applied on the implant in a mesial (sagittal) direction simulating the orthopedic force in maxillary protraction. Implant in Cortical Bone: Type I FE Model On loading the implant with orthopedic forces (Fig. 5), the implant experienced greater tensile and compressive stresses when compared to the previous two orthodontic loadings. These stresses were concentrated at the point of force application. In contrast, the minimum principal compressive stresses were greater than the maximum principal tensile stresses. The bone experienced greater compressive and tensile stresses compared to the orthodontic loading. The maximum principal tensile stress was greater than the minimum principal compressive stress.the interface experienced greater stresses compared to the previous two orthodontic loadings. Here, the compressive stresses were greater than the tensile stresses in contrast to the overall stress distribution in the bone. Displacement of the implant: On application of orthopedic force to the implant, the implant showed a greater amount of displacement when compared to the previous two orthodontic loadings. This displacement was in the range of to mm. On application of orthopedic force to the implant, the bone experienced maximum compressive and tensile strain. The junction between the implant and the cortical bone experienced greater strain compared to the deeper layers of bone. Similar to all the previous loads, these strain values were compared with the Frost s mechanostat Table 1, which revealed that all the three orthopedic loads resulted in strain in the range of physiologic bone maintenance. Implant in Cortical and Trabecular Bone: Type II FE Model The implant experienced greater tensile and compressive stresses compared to the type I implant. i. The cortical bone also experienced greater tensile and compressive stresses when compared to the type I model. Similar to the previous model, the tensile stresses were greater than the compressive stresses. ii. The trabecular bone experienced lower stresses compared to the cortical bone. Implant-cortical bone junction experienced greater stresses when compared to the implant-trabecular bone junction. Displacement of the implant: The implant displaced to a greater extent than the type I implant. This displacement was in the range of to mm. The strain values were comparatively higher than the type I model. Similar to all the previous loads, the values obtained were compared with the Frost s mechanostat Table 1. Strain values in the cortical bone and implant-cortical bone interface were sufficient for bone maintenance, while the strain values in trabecular bone and implant-trabecular interface were significantly less, which would probably result in atrophy of bone according to Frost s mechanostat theory. 180 JAYPEE

7 JIOS Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage DISCUSSION Biomechanical influences on bone structure play an important role in the longevity of bone around implants. 3 Bone tissue is known to remodel its structure in response to mechanical stress. Low stress levels around an implant may result in poor connection with bone or bone atrophy. On the other hand, abnormally high stress concentrations in the supporting tissues can result in pressure necrosis and subsequently in implant failure. 3 The present study was aimed at analyzing the deformation of bone surrounding a mini-implant in response to different force magnitudes. COMPARISON WITH OTHER STUDIES Melsen and Verna (2005) 21 evaluated the load transfer from the Aarhus miniscrew on the bone, and the influence of different cortical bone thicknesses and the underlying trabecular bone density, on applying a mesially directed force of 50 gm. The results obtained from our study matched with their study. The primary component of the load transfer was seen at the first revolution of the miniscrew thread within the cortex. Authors found that on decreasing the thickness of the cortical bone, the peak strain values reached the pathological overload window (Frost s mechanostat theory). But in our study, the thickness of the cortical bone was kept constant with the larger part of the implant being surrounded by the cortical bone. Hence, the cortical bone experienced greater strain values than the underlying trabecular bone. At this thickness of the cortical bone, the peak strain values were in the optimal range of bone maintenance. Chen, Esterle and Roberts (1999) 31 investigated the mechanical environment of cortical bone adjacent to retromolar endosseous implants used for orthodontic anchorage in a coordinated histomorphometric and finite element study. The authors studied the change in the stress pattern in the bone before and after placement of implants. They noticed a strong change in the stress pattern in the bone after implantation, wherein the stresses increased by 2 to 4 times on orthodontic loading. It was observed that on loading the implant by an anterior force, compressive stresses were found on the buccal and distal sides, and tensile stresses were found on the lingual and mesial sides. The results obtained in our study were similar to the above study. Their coordinated histomorphometric study revealed that the high areas of stress were associated with high remodeling activity of the bone. Our study also showed similar areas of high stresses distribution around the first few threads of implant. CLINICAL IMPLICATIONS From the results of the present study, following clinical implications could be derived: 1. The mini-implants would efficiently resist oblique loading, simulating anterior intrusion and retraction. The loads of 200 and 250 gm produced strain in the optimal range of bone maintenance. These devices would serve as a reliable means of absolute anchorage as per the initial stress pattern. 2. The use of these mini-implants for the purpose of orthopedic loading is questionable, since the stress values experienced by the implant are quite high. Even though the strain values experienced by the bone were in the range of optimal maintenance, the site of placement of the implants and the timing become crucial factors in deciding the usage of these implants for orthopedic loading. The ideal time for orthopedic protraction of maxilla is during the mixed dentition phase. The miniscrew can loosen, even after having been initially fixed, if an adjacent deciduous tooth is exfoliating. There can also be a risk of injuring the underlying permanent tooth bud. 3. In the implant, the most critical area is its neck, where there is maximum stress concentration, and the marginal bone (cervical margin) which surrounds it. Thus, this area should be preserved clinically in order to maintain the bone-implant interface structurally and functionally. 4. It was seen in our study that the implant tipped to a very negligible amount in the direction of the load applied, like a tooth tipping on application of load. But the displacement seen was very negligible and clinically insignificant. 5. Based on the experience from our study, the following suggestions can be made for optimization of the implant design: a. The neck of the implant must be long enough to project away from the soft tissues, so that any attachments placed on the implants do not impinge the mucosa. The inflammation of the overlying soft tissue and/or the marginal bone resorption can jeopardize the stability of the implant. b. The neck of the implant must be sturdy enough, since the maximum stress concentration occurs at the neck of the implant. If the implant is not strong in this region, it may affect the integrity of the implant. c. When using these mini-implants for orthodontic loading, it is advisable to take all the necessary precautions to place the implant as much in the cortical bone as possible. The reason is that the stress and strain values in the trabecular bone were very low, which would result in atrophy of the surrounding bone (as postulated in Frost s mechanostat principle). CONCLUSION From this study, it can be concluded that: 1. The mini-implants used in the present study efficiently resisted the oblique loading, which simulated intrusion and retraction of the anterior teeth. The loads of 200 and 250 gm produced strains in the optimal range of bone maintenance. The Journal of Indian Orthodontic Society, October-December 2012;46(4):

8 PS Vijayalakshmi et al 2. The use of these mini-implants for the purpose of orthopedic loading is questionable. 3. The implant tipped on loading in all the three clinical situations simulated in the present study. But this displacement was to a negligible amount and was clinically insignificant. 4. Both the types of FE models showed the area with the highest stress and strain to be around the neck of the implant and the surrounding bone at the cervical margin. This finding is clinically important in order to preserve the bone-implant interface functionally and structurally in this area. 5. To obtain an optimal biomechanical response, the implant should preferably be placed entirely in the cortical bone (however, this may not always be feasible clinically). The neck of the implant should be sturdy enough, and the head of the implant should not produce any kind of irritation to the overlying mucosa. REFERENCES 1. Roberts WE, Helm FR, Marshall KJ, Gongloff RK. Rigid endosseous implants for orthodontic and orthopedic anchorage. Angle Orthod 1989;59(4): Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants 2004;19: Gainesforth BL, Higley LB. A study of orthodontic anchorage possibilities in basal bone. Am J Orthod Oral Surg 1945;31: Linkow L. Implanto-orthodontics. J Clin Orthod 1970;4: Creekmore TD, Eklund MK. The possibility of skeletal anchorage. J Clin Orthod 1983;17(4): Roberts WE, Marshall KJ, Mozsary PG. Rigid endosseous implant utilized as anchorage to protract molars and close an atropic extraction site. Angle Orthod 1990;60: Roberts WE, Nelson CL, Goodacre CJ. Rigid implant anchorage to close a mandibular first molar extraction site. J Clin Orthod 1994;38: Southard TE, Buckley MJ, Spivey JD, Krizan KE, Casko JS. Intrusion anchorage potential of teeth versus rigid endosseous implants: A clinical and radiographic evaluation. Am J Orthod Dentofacial Orthop 1995;107: Block MS, Hoffman DR. A new device for absolute anchorage for orthodontics. Am J Orthod Dentofacial Orthop 1995;107(3): Hoffman DR. Implants and orthodontics. In: Block MS, Kent JN (Eds). Endosseous implants for maxillofacial reconstruction (1st ed). Philadelphia: WB Saunders Co 1995: Block MS. Orthodontic and orthopedic anchorage using subperiosteal bone anchors. In: Higuchi KW (Ed). Orthodontic applications of osseointegrated implants. Illinois, Quintessence Publishing Co Inc 2000; Janssens F, Swennen G, Dujardin T, Glineur R, Malevez C. Use of an onplant as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2002;122(5): Wehrbein H, Merz BR, Diedrich P, Glatzmaier J. The use of palatal implants for orthodontic anchorage. Design and clinical application of the orthosystem. Clin Oral Implants Res 1996;7(4): Umemori M, Sugawara J, Mitani H, Nagasaka H, Kawamura H. Skeletal anchorage system for open bite correction. Am J Orthod Dentofacial Orthop 1999;115(2): Byloff FK, Karcher H, Clar E, Stoff F. An implant to eliminate anchorage loss during molar distalization: A case report involving the Graz implant-supported pendulum. Int J Adult Orthodon Orthognath Surg 2000;15(2): De Clerck H, Geerinckx V, Siciliano S. The zygoma anchorage system. J Clin Orthod 2002;36(8): Kanomi R. Mini-implant for orthodontic anchorage. J Clin Orthod 1997;31(11): Bousquet F, Bousquet P, Mauran G, Parguel P. Use of an impacted post for anchorage. J Clin Orthod 1996;30(5): Costa A, Raffaini M, Melsen B. Miniscrews as orthodontic anchorage: A preliminary report. Int J Adult Orthod Orthognath Surg 1998;13: Melsen B, Verna C. Miniscrew implants: The Aarhus anchorage system. Semin Orthod 2005;11: Park HS, Bae SM, Kyung HM, Sung JH. Micro-implant anchorage for treatment of skeletal class I bialveolar protrusion. J Clin Orthod 2001;35(7): Bae SM, Park HS, Kyung HM, Kwon OW, Sung JH. Clinical application of micro-implant anchorage. J Clin Orthod 2002;36(5): Park HS, Kyung HM, Kwon OW, Sung JH. A simple method of molar uprighting with micro-implant anchorage. J Clin Orthod 2002;36(10): Maino BG, Bednar J, Pagin P, Paola Mura. The spider screw for skeletal anchorage. J Clin Orthod 2003;37(2): Gedrange T, Bourauel C, Kobel C, Harzer W. Three-dimensional analysis of endosseous palatal implants and bones after vertical, horizontal, and diagonal force application. Eur J Orthod 2003;25: Geng JP, Tan KBC, Liu GR. Application of finite element analysis in implant dentistry: A review of literature. J Prosthet Dent 2001;85: Rieger MR, Fareed K, Adams WK, Tanquist RA. Bone stress distribution for three endosseous implants. J Prosthet Dent 1989;61: Canay S, Hersek N, Irfan Akpinar, Asik Z. Comparison of stress distribution around vertical and angled implants with finite-element analysis. Quintessence Int 1996;27: Roberts EW. Bone physiology, metabolism and biomechanics in orthodontic practice. In: Graber TM, Vanarsdall RL (Eds). Orthodontics current principles and techniques (3rd ed). St Louis: Mosby 2000: Chen J, Esterle M, Roberts WE. Mechanical response to functional loading around the threads of retromolar endosseous implants utilized for orthodontic anchorage: Coordinated histomorphometric and finite element analysis. Int J Oral Maxillofac Implants 1999;14: JAYPEE

Temporary Orthodontic Anchorage Devices

Temporary Orthodontic Anchorage Devices REVIEW ARTICLE 1 Veereshi AS, 2 Vijayalakshmi PS 1 Associate Professor, Department of Orthodontics, Rungta College of Dental Sciences, Bhilai, Chhattisgarh, India 2 Reader, Department of Orthodontics,

More information

Finite Element Analysis of Dental Implant as Orthodontic Anchorage

Finite Element Analysis of Dental Implant as Orthodontic Anchorage JCDP 10.5005/jp-journals-10024-1044 ORIGINAL RESEARCH Finite Element Analysis of Dental Implant as Orthodontic Anchorage Finite Element Analysis of Dental Implant as Orthodontic Anchorage Anirban Sarmah,

More information

ORIGINAL ARTICLE , 1,2 onplant, 3,4, . 12, 13,14 , 7,8. 9, ,,. 16,17

ORIGINAL ARTICLE , 1,2 onplant, 3,4, . 12, 13,14 , 7,8. 9, ,,. 16,17 ORIGINAL ARTICLE 8 mm 14 mm 16 mm 14 mm 16 mm 14 mm 16 mm 4 20 16 mm 14 mm 14 mm 16 mm 2006;36(4):275-83) 1 11 56 12 onplant 34 12 1314 78 910 15 5 1617 36 4 2006 Table 1 Chemical composition and mechanical

More information

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. Chandrasenan.P 1, Vishnu.G 2, Akshay K Nair 3 1M Tech student, Department of Mechanical

More information

Orthodontic Microimplants and Its

Orthodontic Microimplants and Its CASE REPORT Orthodontic Microimplants and Its Applications Rajesh Patil 1, Girish Karandikar 2, Manish Sonawane 3 Abstract Microimplants usage has revolutionized the clinical orthodontic practice over

More information

Non-osseointegrated. What type of mini-implants? 3/27/2008. Require a tight fit to be effective Stability depends on the quality and.

Non-osseointegrated. What type of mini-implants? 3/27/2008. Require a tight fit to be effective Stability depends on the quality and. Non-osseointegrated What type of mini-implants? Require a tight fit to be effective Stability depends on the quality and quantity of cortical and trabecular bone. Osseointegrated Non-osseointegrated AbsoAnchor

More information

6. Timing for orthodontic force

6. Timing for orthodontic force 6. Timing for orthodontic force Orthodontic force is generally less than 300gm, so early mechanical stability is enough for immediate orthodontic force. There is no actually difference in success rate

More information

The use of mini-implants in en masse retraction for the treatment of bimaxillary dentoalveolar protrusion

The use of mini-implants in en masse retraction for the treatment of bimaxillary dentoalveolar protrusion The Saudi Dental Journal (2010) 22, 35 39 King Saud University The Saudi Dental Journal www.ksu.edu.sa www.sciencedirect.com CASE REPORT The use of mini-implants in en masse retraction for the treatment

More information

THE USE OF TEMPORARY ANCHORAGE DEVICES FOR MOLAR INTRUSION & TREATMENT OF ANTERIOR OPEN BITE By Eduardo Nicolaievsky D.D.S.

THE USE OF TEMPORARY ANCHORAGE DEVICES FOR MOLAR INTRUSION & TREATMENT OF ANTERIOR OPEN BITE By Eduardo Nicolaievsky D.D.S. THE USE OF TEMPORARY ANCHORAGE DEVICES FOR MOLAR INTRUSION & TREATMENT OF ANTERIOR OPEN BITE By Eduardo Nicolaievsky D.D.S. Skeletal anchorage, the concept of using the facial skeleton to control tooth

More information

MOLAR DISTALIZATION WITH MODIFIED GRAZ IMPLANT SUPPORTED PENDULUM SPRINGS. A CASE REPORT.

MOLAR DISTALIZATION WITH MODIFIED GRAZ IMPLANT SUPPORTED PENDULUM SPRINGS. A CASE REPORT. MOLAR DISTALIZATION WITH MODIFIED GRAZ IMPLANT SUPPORTED PENDULUM SPRINGS. A CASE REPORT. Author : Dr. Arundhati P. Tandur MDS, MDORCPS [Glasgow], IMORTHRCS [England], Diplomate of Indian Board Vice-Principal

More information

Ideal treatment of the impaired

Ideal treatment of the impaired RESEARCH IMPLANTS AS ANCHORAGE IN ORTHODONTICS: ACLINICAL CASE REPORT Dale B. Herrero, DDS KEY WORDS External anchorage Pneumatized Often, in dental reconstruction, orthodontics is required for either

More information

The International Journal of Periodontics & Restorative Dentistry

The International Journal of Periodontics & Restorative Dentistry The International Journal of Periodontics & Restorative Dentistry 559 Failure of Orthodontic Mini-implants by Age, Sex, and Arch; Number of Primary Insertions; and Frequency of Reinsertions After Failure:

More information

Skeletal Anchorage for Orthodontic Correction of Severe Maxillary Protrusion after Previous Orthodontic Treatment

Skeletal Anchorage for Orthodontic Correction of Severe Maxillary Protrusion after Previous Orthodontic Treatment The Angle Orthodontist: Vol. 78, No. 1, pp. 181 188. Skeletal Anchorage for Orthodontic Correction of Severe Maxillary Protrusion after Previous Orthodontic Treatment Eiji Tanaka; a Akiko Nishi-Sasaki;

More information

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 DOI 10.1007/s13191-012-0189-7 ORIGINAL ARTICLE Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A

More information

Bone density assessment for mini-implants position

Bone density assessment for mini-implants position O r i g i n a l A r t i c l e Bone density assessment for mini-implants position Marlon Sampaio Borges*, José Nelson Mucha** Abstract Introduction: Cortical thickness, interradicular space width and bone

More information

The growing demand for orthodontic treatment

The growing demand for orthodontic treatment ORIGINAL ARTICLE Do miniscrews remain stationary under orthodontic forces? Eric J. W. Liou, DDS, MS, a Betty C. J. Pai, DDS, b and James C. Y. Lin, DDS c Taipei, Taiwan Miniscrews have been used in recent

More information

International Journal of Medical and Health Sciences

International Journal of Medical and Health Sciences International Journal of Medical and Health Sciences Journal Home Page: http://www.ijmhs.net ISSN:2277-4505 Case Report Mini Screw Implant Taking Orthodontics to New Dimensions A Case Report Deepak Victor

More information

AAO 115th Annual Session San Francisco, CA May 17 (Sunday), 1:15-2:00 pm, 2015

AAO 115th Annual Session San Francisco, CA May 17 (Sunday), 1:15-2:00 pm, 2015 AAO 115th Annual Session San Francisco, CA May 17 (Sunday), 1:15-2:00 pm, 2015 Title: Clinical and iomechanical Considerations of TADs in Challenging Cases: Sagittal Correction beyond Orthodontic oundaries

More information

ABSTRACT INTRODUCTION

ABSTRACT INTRODUCTION ABSTRACT Purpose: Evaluation of stress distribution maxilla under parafunctional loading on short dental implant using 3-dimensional (3D) finite element models. Material and method: 3D maxillary and mandibular

More information

Clinical indices for orthodontic mini-implants

Clinical indices for orthodontic mini-implants 407 Journal of Oral Science, Vol. 53, No. 4, 407-412, 2011 Review Clinical indices for orthodontic mini-implants Mitsuru Motoyoshi 1,2) 1) Department of Orthodontics, Nihon University School of Dentistry,

More information

KJLO. A Sequential Approach for an Asymmetric Extraction Case in. Lingual Orthodontics. Case Report INTRODUCTION DIAGNOSIS

KJLO. A Sequential Approach for an Asymmetric Extraction Case in. Lingual Orthodontics. Case Report INTRODUCTION DIAGNOSIS KJLO Korean Journal of Lingual Orthodontics Case Report A Sequential Approach for an Asymmetric Extraction Case in Lingual Orthodontics Ji-Sung Jang 1, Kee-Joon Lee 2 1 Dream Orthodontic Clinic, Gimhae,

More information

Ortho-surgical Management of Severe Vertical Dysplasia: A Case Report

Ortho-surgical Management of Severe Vertical Dysplasia: A Case Report Case Report Ortho-surgical Management of Severe Vertical Dysplasia: A Case Report 1 Vinni Arora, 2 Rekha Sharma, 3 Sachin Parashar 1 Senior Resident, 2 Professor and Head of Department, 3 Former Resident

More information

The LOMAS System. Eric J. W. Liou, James C. Y. Lin

The LOMAS System. Eric J. W. Liou, James C. Y. Lin 15 CHAPTER The LOMAS System Eric J. W. Liou, James C. Y. Lin THE LOMAS SYSTEM An ideal orthodontic miniscrew should be applicable to all orthodontic anchorage requirements, capable of withstanding the

More information

Mini-implants in Orthodontics: A Review

Mini-implants in Orthodontics: A Review Annals of Dental Research (2013) Vol 3(1): 6-10 HATAM Publishers: All Rights Reserved Annals of Dental Research www.hgpub.com www.adres.yolasite.com Review Mini-implants in Orthodontics: A Review Singla

More information

INDIAN JOURNAL OF DENTAL ADVANCEMENTS

INDIAN JOURNAL OF DENTAL ADVANCEMENTS INDIAN JOURNAL OF DENTAL ADVANCEMENTS Journal homepage: www.nacd.in REVIEW Micro-Implants: Innovative Anchorage Concepts in Orthodontics Vishal Seth 1, Prasanth Kamath 2, Venkatesh M J 3, Renu Prasad 4,

More information

ORIGINAL ARTICLE. Key words: Osseointegration, Mini-implant, Finite element analysis, Von-Mises stress INTRODUCTION

ORIGINAL ARTICLE. Key words: Osseointegration, Mini-implant, Finite element analysis, Von-Mises stress INTRODUCTION ORIGINAL ARTICLE Three-dimensional finite element analysis for determining the stress distribution after loading the bone surface with two-component mini-implants of varying length Bohm Choi, DDS, MSD,

More information

Prosthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor

Prosthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor Prosthetic Options in Dentistry Hakimeh Siadat, DDS, MSc Associate Professor Dental Research Center, Department of Prosthodontics & Dental s Faculty of Dentistry, Tehran University of Medical Sciences

More information

Gentle-Jumper- Non-compliance Class II corrector

Gentle-Jumper- Non-compliance Class II corrector 15 CASE REPORT Gentle-Jumper- Non-compliance Class II corrector Amit Prakash 1,O.P.Mehta 2, Kshitij Gupta 3 Swapnil Pandey 4 Deep Kumar Suryawanshi 4 1 Senior lecturer Bhopal - INDIA 2 Professor Bhopal

More information

Maxillary Growth Control with High Pull Headgear- A Case Report

Maxillary Growth Control with High Pull Headgear- A Case Report IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 17, Issue 01 Ver. X January. (2018), PP 09-13 www.iosrjournals.org Maxillary Growth Control with High

More information

The Clinical Success of Self-tapping and Self-drilling Orthodontic Miniscrews

The Clinical Success of Self-tapping and Self-drilling Orthodontic Miniscrews Biological Forum An International Journal 7(2): 33-37(25) ISSN No. (Print): 975-3 ISSN No. (Online): 2249-3239 The Clinical Success of Self-tapping and Self-drilling Orthodontic Miniscrews Abbas Salehi

More information

Research Article Success Rate of Microimplants in a University Orthodontic Clinic

Research Article Success Rate of Microimplants in a University Orthodontic Clinic International Scholarly Research Network ISRN Surgery Volume 2011, Article ID 982671, 6 pages doi:10.5402/2011/982671 Research Article Success Rate of Microimplants in a University Orthodontic Clinic P.

More information

Sliding Mechanics with Microscrew Implant Anchorage

Sliding Mechanics with Microscrew Implant Anchorage Clinical Report Sliding Mechanics with Microscrew Implant Anchorage Hyo-Sang Park, DDS, MSD, PhD a ; Tae-Geon Kwon, DDS, MSD, PhD b Abstract: Three cases are illustrated. One was treated with maxillary

More information

In recent years, orthodontic mini-implants have been

In recent years, orthodontic mini-implants have been ONLINE ONLY Factors affecting the long-term stability of orthodontic mini-implants Mitsuru Motoyoshi, a Miwa Uemura, b Akiko Ono, c Kumiko Okazaki, d Toru Shigeeda, d and Noriyoshi Shimizu e Tokyo, Japan

More information

3D Cortical Bone Anatomy of the Mandibular Buccal Shelf: a CBCT study to define sites for extra-alveolar bone screws to treat Class III malocclusion

3D Cortical Bone Anatomy of the Mandibular Buccal Shelf: a CBCT study to define sites for extra-alveolar bone screws to treat Class III malocclusion 3D Cortical Bone Anatomy of the Mandibular Buccal Shelf: a CBCT study to define sites for extra-alveolar bone screws to treat Class III malocclusion Abstract Objective: Assess the feasibility of a proposed

More information

Intraoral molar-distalization appliances that

Intraoral molar-distalization appliances that 2014 JCO, Inc. May not be distributed without permission. www.jco-online.com Distalization with the Miniscrew- Supported EZ Slider Auxiliary ENIS GÜRAY, DDS, PHD FARUK IZZET UCAR, DDS, PHD NISA GUL, DDS

More information

Influence of Surface Characteristics on Survival Rates of Mini-Implants

Influence of Surface Characteristics on Survival Rates of Mini-Implants Original Article Influence of Surface Characteristics on Survival Rates of Mini-Implants Karim Chaddad a ; André F.H. Ferreira b ; Nico Geurs c ; Michael S. Reddy d ABSTRACT Objective: To compare the clinical

More information

An Innovative Treatment Approach with Atypical Orthodontic Extraction Pattern in Bimaxillary Protrusion Case

An Innovative Treatment Approach with Atypical Orthodontic Extraction Pattern in Bimaxillary Protrusion Case 10.5005/jp-journals-10021-1127 CASE REPORT An Innovative Treatment Approach with Atypical Orthodontic Extraction Pattern in Bimaxillary Protrusion Case 1 Anil Miglani, 2 Reena R Kumar, 3 Ashish Chopra,

More information

Lever-arm and Mini-implant System for Anterior Torque Control during Retraction in Lingual Orthodontic Treatment

Lever-arm and Mini-implant System for Anterior Torque Control during Retraction in Lingual Orthodontic Treatment Clinical Report Lever-arm and Mini-implant System for Anterior Torque Control during Retraction in Lingual Orthodontic Treatment Ryoon-Ki Hong, DDS, PHD a ; Jung-Min Heo, DDS b ; Young-Ki Ha, DDS b Abstract:

More information

International Journal of Therapeutic Applications ISSN X

International Journal of Therapeutic Applications ISSN X International Journal of Therapeutic Applications ISSN 2320-138X HEAD GEAR EFFECT USING MINI-IMPLANT (DISTALIZATION AND INTRUSION) A CASE REPORT Mohammadi Begum 1*, Arjun Karra 2 1 DRs. Sudha and Nageswara

More information

ANTERIOR AND CANINE RETRACTION: BIOMECHANIC CONSIDERATIONS. Part One

ANTERIOR AND CANINE RETRACTION: BIOMECHANIC CONSIDERATIONS. Part One In italiano, per favore ANTERIOR AND CANINE RETRACTION: BIOMECHANIC CONSIDERATIONS Part One Gabriele Floria DDS, Lorenzo Franchi DDS, Turi Bassarelli MD English Translation by Dr. Susan Eslambolchi & Dr.

More information

TADs Supported Tongue Crib: A New Minimalistic Design

TADs Supported Tongue Crib: A New Minimalistic Design Clinical Pearl TADs Supported Tongue Crib: A New Minimalistic Design Patni V 1, Kolge NE 2 To cite: Patni V, Kolge NE. TADs Supported Tongue Crib: A New Minimalistic Design. Journal of Contemporary Orthodontics,

More information

Australian Dental Journal

Australian Dental Journal Australian Dental Journal The official journal of the Australian Dental Association CASE REPORT Australian Dental Journal 2010; 55: 457 461 doi: 10.1111/j.1834-7819.2010.01270.x Intrusion of an overerupted

More information

Anchorage control is a critical consideration when

Anchorage control is a critical consideration when ORIGINAL ARTICLE Soft-tissue and cortical-bone thickness at orthodontic implant sites Hee-Jin Kim, a Hee-Sun Yun, b Hyun-Do Park, c Doo-Hyung Kim, b and Young-Chel Park d Seoul, South Korea Introduction:

More information

AAO / AAPD Scottsdale 2018

AAO / AAPD Scottsdale 2018 AAO / AAPD Scottsdale 2018 Missing Premolars : What are the Options? David Kennedy Clinical Professor UBC Vancouver Canada drdavidkennedy@yahoo.ca At what age can you know second premolars are absent?

More information

Nonextraction Treatment of Upper Canine Premolar Transposition in an Adult Patient

Nonextraction Treatment of Upper Canine Premolar Transposition in an Adult Patient Case Report Nonextraction Treatment of Upper Canine Premolar Transposition in an Adult Patient Shingo Kuroda a ; Yasuko Kuroda b Abstract: This article reports the successful treatment of a unilateral

More information

ORTHOdontics SLIDING MECHANICS

ORTHOdontics SLIDING MECHANICS ORTHOdontics PGI/II SLIDING MECHANICS FOCUS ON TARGETED SPACE GAINING AND ITS APPLICATIONS, INCLUDING WITH RAPID PALATAL EXPANDIONS. ALSO INCLUDES RETENTION AND CLINICAL PEARLS FACULTY: Joseph Ghafari,

More information

Case Report Unilateral Molar Distalization: A Nonextraction Therapy

Case Report Unilateral Molar Distalization: A Nonextraction Therapy Case Reports in Dentistry Volume 2012, Article ID 846319, 4 pages doi:10.1155/2012/846319 Case Report Unilateral Molar Distalization: A Nonextraction Therapy M. Bhanu Prasad and S. Sreevalli Department

More information

Influence of Different Angulations of Force in Stress Distribution in Implant Retained Finger Prosthesis: A Finite Element Study

Influence of Different Angulations of Force in Stress Distribution in Implant Retained Finger Prosthesis: A Finite Element Study World Journal of Medical Sciences 10 (2): 217-221, 2014 ISSN 1817-3055 IDOSI Publications, 2014 DOI: 10.5829/idosi.wjms.2014.10.2.82275 Influence of Different Angulations of Force in Stress Distribution

More information

IS THE PALATE AN OPTIMAL SITE FOR MINISCREW PLACEMENT?

IS THE PALATE AN OPTIMAL SITE FOR MINISCREW PLACEMENT? doi:10.5368/aedj.2011.3.2.1.8 IS THE PALATE AN OPTIMAL SITE FOR MINISCREW PLACEMENT? 1 Vinaya S. Pai 2 Abraham Thomas 3 Swetha. M 4 Vishal Anil Nalawade 1 Principal,Professor and Head 2 Reader 3 Senior

More information

Treatment of Class II, Division 2 Malocclusion with Miniscrew Supported En-Masse Retraction: Is Deepbite Really an Obstacle for Extraction Treatment?

Treatment of Class II, Division 2 Malocclusion with Miniscrew Supported En-Masse Retraction: Is Deepbite Really an Obstacle for Extraction Treatment? TURKISH JOURNAL of DOI: 10.5152/TurkJOrthod.2017.17034 CASE REPORT Treatment of Class II, Division 2 Malocclusion with Miniscrew Supported En-Masse Retraction: Is Deepbite Really an Obstacle for Extraction

More information

ORTHODONTIC CORRECTION Of OCCLUSAL CANT USING MINI IMPLANTS:A CASE REPORT. Gupta J*, Makhija P.G.**, Jain V***

ORTHODONTIC CORRECTION Of OCCLUSAL CANT USING MINI IMPLANTS:A CASE REPORT. Gupta J*, Makhija P.G.**, Jain V*** ORTHODONTIC CORRECTION Of OCCLUSAL CANT USING MINI IMPLANTS:A CASE REPORT Gupta J*, Makhija P.G.**, Jain V*** Abstract: The inability of orthodontists to change the cant of the maxillary occlusal plane

More information

A SIMPLE METHOD FOR CORRECTION OF BUCCAL CROSSBITE OF MAXILLARY SECOND MOLAR

A SIMPLE METHOD FOR CORRECTION OF BUCCAL CROSSBITE OF MAXILLARY SECOND MOLAR Short Communication International Journal of Dental and Health Sciences Volume 01,Issue 03 A SIMPLE METHOD FOR CORRECTION OF BUCCAL CROSSBITE OF MAXILLARY SECOND MOLAR Sumit Yadav 1,Davender Kumar 2,Achla

More information

Intrusion of the Overerupted Upper Left First and Second Molars by Mini-implants with Partial-Fixed Orthodontic Appliances: A Case Report

Intrusion of the Overerupted Upper Left First and Second Molars by Mini-implants with Partial-Fixed Orthodontic Appliances: A Case Report Case Report Intrusion of the Overerupted Upper Left First and Second Molars by Mini-implants with Partial-Fixed Orthodontic Appliances: A Case Report Chung-Chen Jane Yao, DDS, PhD a,b ; Chou-Bing Wu, DDS,

More information

Soft Tissue Thickness for Placement of an Orthodontic Miniscrew Using an Ultrasonic Device

Soft Tissue Thickness for Placement of an Orthodontic Miniscrew Using an Ultrasonic Device Original Article Soft Tissue Thickness for Placement of an Orthodontic Miniscrew Using an Ultrasonic Device Bong-Kuen Cha a ; Yeon-Hee Lee b ; Nam-Ki Lee c ; Dong-Soon Choi c ; Seung-Hak Baek d ABSTRACT

More information

Rehabilitating a Compromised Site for Restoring Form, Function and Esthetics- A Case Report

Rehabilitating a Compromised Site for Restoring Form, Function and Esthetics- A Case Report Research & Reviews: Journal of Dental Sciences Rehabilitating a Compromised Site for Restoring Form, Function and Esthetics- A Case Report Priyanka Prakash* Division of Periodontology, Department of Dental

More information

Miniscrew-supported coil spring for molar uprighting: Description

Miniscrew-supported coil spring for molar uprighting: Description Miniscrew-supported coil spring for molar uprighting: Description ntônio Carlos de Oliveira Ruellas 1, Matheus Melo Pithon 2, Rogério Lacerda dos Santos 3 Introduction: Since the beginning of miniscrews

More information

Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws

Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws Abstract Simplified mechanics are reported for uprighting horizontally impacted mandibular molars with ramus bone screws. A 27-year-old

More information

Assessment of palatal bone thickness in adults with cone beam computerised tomography

Assessment of palatal bone thickness in adults with cone beam computerised tomography Assessment of palatal bone thickness in adults with cone beam computerised tomography Antonio Gracco, Lombardo Luca, Mauro Cozzani and Giuseppe Siciliani Department of Orthodontics, University of Ferrara,

More information

Treatment of an open bite case with 3M Clarity ADVANCED Ceramic Brackets and miniscrews.

Treatment of an open bite case with 3M Clarity ADVANCED Ceramic Brackets and miniscrews. SM 3M Health Care Academy Treatment of an open bite case with 3M Clarity ADVANCED Ceramic Brackets and miniscrews. Dr. J.C. Pérez-Varela MD, DDS, MS, Ph.D. Specialist in Orthodontics. Doctor of Medicine

More information

There had been controversies concerning how to achieve maximum anchorage in the first premolar extraction cases.

There had been controversies concerning how to achieve maximum anchorage in the first premolar extraction cases. Cronicon OPEN ACCESS EC DENTAL SCIENCE Research Article Comparison Between the Stresses Induced on Vital Structures During Two Steps and En-Masse Retraction Methods of Upper Anterior Segment Depending

More information

Uprighting Movement Relative to Defects Forced Eruption Alteration of the Gingival Embrasure Interactions with Implants

Uprighting Movement Relative to Defects Forced Eruption Alteration of the Gingival Embrasure Interactions with Implants Frank Celenza DDS Periodontist and Orthodontist Past President, Northeastern Society of Periodontists Associate Clinical Professor, New York University Member, ITI (International Team of Implantologists)

More information

Use of Palatal Miniscrew Anchorage and Lingual Multi-Bracket Appliances to Enhance Efficiency of Molar Scissors-Bite Correction

Use of Palatal Miniscrew Anchorage and Lingual Multi-Bracket Appliances to Enhance Efficiency of Molar Scissors-Bite Correction Case Report Use of Palatal Miniscrew Anchorage and Lingual Multi-Bracket Appliances to Enhance Efficiency of Molar Scissors-Bite Correction Nagato Tamamura a ; Shingo Kuroda b ; Yasuyo Sugawara b ; Teruko

More information

Pendulum appliance: skeletal and dentoalveolar effects. A systematic review.

Pendulum appliance: skeletal and dentoalveolar effects. A systematic review. Article ID: ISSN 2046-1690 Pendulum appliance: skeletal and dentoalveolar effects. A systematic review. Peer review status: No Corresponding Author: Dr. Martina Mezio, Submitting Author: Dr. Denise Giovannoni,

More information

Orthodontic-prosthetic implant anchorage in a partially edentulous patient

Orthodontic-prosthetic implant anchorage in a partially edentulous patient Journal of Dental Sciences (2011) 6, 176e180 available at www.sciencedirect.com journal homepage: www.e-jds.com Clinical Report Orthodontic-prosthetic implant anchorage in a partially edentulous patient

More information

Many patients require prosthodontic rehabilitation

Many patients require prosthodontic rehabilitation CASE LETTER Case Studies on the Use of Skeletal Anchorage Orthodontic Techniques Prior to Prosthetic Rehabilitation Fernanda Faot, MSc, PhD 1 * Livia Forster Ribeiro, MSc 2 Raissa Micaella Marcello Machado

More information

Technique Guide. Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth.

Technique Guide. Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth. Technique Guide Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth. Table of Contents Introduction Orthodontic Bone Anchor (OBA) System 2 Indications and

More information

Angle s Molar Classification Revisited

Angle s Molar Classification Revisited Devanshi Yadav et al Original Article 10.5005/jp-journals-10021-1282 Angle s Molar Classification Revisited 1 Devanshi Yadav, 2 MS Rani, 3 AM Shailaja, 4 Dhruv Anand, 5 Nikhil Sood, 6 Rajat Gothi ABSTRACT

More information

Effective en-masse retraction design with orthodontic mini-implant anchorage: A finite element analysis

Effective en-masse retraction design with orthodontic mini-implant anchorage: A finite element analysis ORIGINAL ARTICLE Effective en-masse retraction design with orthodontic mini-implant anchorage: A finite element analysis Sang-Jin Sung, a Gang-Won Jang, b Youn-Sic Chun, c and Yoon-Shik Moon d Seoul and

More information

EFFICIENCY OF THE DISTAL SCREW IN THE DISTAL MOVEMENT OF MAXILLARY MOLARS

EFFICIENCY OF THE DISTAL SCREW IN THE DISTAL MOVEMENT OF MAXILLARY MOLARS Mauro Cozzani, DMD, MScD 1 Francesco Zallio, MD, DDS 2 Luca Lombardo, DDS 3 Antonio Gracco, DDS 3 EFFICIENCY OF THE DISTAL SCREW IN THE DISTAL MOVEMENT OF MAXILLARY MOLARS Aim: Conventionally, noncompliance

More information

A finite element analysis of the effects of archwire size on orthodontic tooth movement in extraction space closure with miniscrew sliding mechanics

A finite element analysis of the effects of archwire size on orthodontic tooth movement in extraction space closure with miniscrew sliding mechanics Kawamura and Tamaya Progress in Orthodontics (2019) 20:3 https://doi.org/10.1186/s40510-018-0255-8 RESEARCH Open Access A finite element analysis of the effects of archwire size on orthodontic tooth movement

More information

Pre Prosthetic Adjunct Orthodontic Tooth Movement for Prosthetic Replacement Using an Osseointegrated Implant-A Case Report

Pre Prosthetic Adjunct Orthodontic Tooth Movement for Prosthetic Replacement Using an Osseointegrated Implant-A Case Report Case Report To cite: Karan Sharma, Dhananjay Sharma, Diksha Sharma, Deepak Rai, Pre Prosthetic Adjunct Orthodontic Tooth Movement for Prosthetic Replacement Using Osseointegrated Implant- A Case Report.

More information

Introduction Subjects and methods

Introduction Subjects and methods European Journal of Orthodontics 33 (2011) 126 131 doi:10.1093/ejo/cjq047 Advance Access Publication 8 November 2010 The Author 2010. Published by Oxford University Press on behalf of the European Orthodontic

More information

Alveolar Bone Remodeling and Development after Immediate Orthodontic Root Movement

Alveolar Bone Remodeling and Development after Immediate Orthodontic Root Movement Journal of Dental Health, Oral Disorders & Therapy Alveolar Bone Remodeling and Development after Immediate Orthodontic Root Abstract Introduction: Adult orthodontics is rapidly expanding primarily due

More information

Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element analysis

Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element analysis ORIGINAL ARTICLE https://doi.org/15125/jkaoms.2018.44.2.59 pissn 224-7550 eissn 224-590 Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element

More information

A Modified Three-piece Base Arch for en masse Retraction and Intrusion in a Class II Division 1 Subdivision Case

A Modified Three-piece Base Arch for en masse Retraction and Intrusion in a Class II Division 1 Subdivision Case Dhaval Ranjitbhai Lekhadia, Gautham Hegde RESEARCH ARTICLE 10.5005/jp-journals-10029-1149 A Modified Three-piece Base Arch for en masse Retraction and Intrusion in a Class II Division 1 Subdivision Case

More information

Three-dimensional analysis of endosseous palatal implants and bones after vertical, horizontal, and diagonal force application

Three-dimensional analysis of endosseous palatal implants and bones after vertical, horizontal, and diagonal force application European Journal of Orthodontics 25 (2003) 109-115 2003 European Orthodontic Society Three-dimensional analysis of endosseous palatal implants and bones after vertical, horizontal, and diagonal force application

More information

JCAMECH Vol. 46, No. 2, July 2015, pp

JCAMECH Vol. 46, No. 2, July 2015, pp JCAMECH Vol. 46, No. 2, July 2015, pp 257-265 Thread Pitch Variant in Orthodontic Mini-screws: A 3-D Finite Element Analysis Fatemeh Motaghi Dastenaei 1 2, Mahdi Moghimi Zand *, Saeed Noorollahian 3 1.

More information

Since their introduction, the use of orthodontic

Since their introduction, the use of orthodontic ORIGINAL ARTICLE Survival analysis of orthodontic mini-implants Shin-Jae Lee, a Sug-Joon Ahn, a Jae Won Lee, b Seong-Hun Kim, c and Tae-Woo Kim d Seoul and Gyeonggi-Do, Korea Introduction: Survival analysis

More information

Effect of bite force on orthodontic mini-implants in the molar region: Finite element analysis

Effect of bite force on orthodontic mini-implants in the molar region: Finite element analysis Original Article THE KOREAN JOURNAL of ORTHODONTICS pissn 2234-7518 eissn 2005-372X Effect of bite force on orthodontic mini-implants in the molar region: Finite element analysis Hyeon-Jung Lee a Kyung-Sook

More information

Molar distalisation with skeletal anchorage

Molar distalisation with skeletal anchorage Molar distalisation with skeletal anchorage Antonio Gracco, Lombardo Luca and Giuseppe Siciliani Department of Orthodontics, University of Ferrara, Ferrara, Italy Background: Distalisation of the upper

More information

Case Report: Long-Term Outcome of Class II Division 1 Malocclusion Treated with Rapid Palatal Expansion and Cervical Traction

Case Report: Long-Term Outcome of Class II Division 1 Malocclusion Treated with Rapid Palatal Expansion and Cervical Traction Case Report Case Report: Long-Term Outcome of Class II Division 1 Malocclusion Treated with Rapid Palatal Expansion and Cervical Traction Roberto M. A. Lima, DDS a ; Anna Leticia Lima, DDS b Abstract:

More information

Dual Force Cuspid Retractor

Dual Force Cuspid Retractor CLINICAL INNOVATION 1 Matrishva B Vyas, 2 Neeraj Alladwar ABSTRACT The most time consuming stage of bicuspid extraction-based treatment is cuspid retraction. Cuspid retraction with both types of conventional

More information

Influence of the thickness of cortical bone on the stability of orthodontic miniscrews

Influence of the thickness of cortical bone on the stability of orthodontic miniscrews DOI: 10.1051/odfen/2013305 J Dentofacial Anom Orthod 2013;16:403 Ó RODF / EDP Sciences Influence of the thickness of cortical bone on the stability of orthodontic miniscrews J. COBO PLANA, F. DE CARLOS

More information

In the search for the ideal anchorage system, orthodontists. Clinical Application of Micro-Implant Anchorage in Initial Orthodontic Retraction

In the search for the ideal anchorage system, orthodontists. Clinical Application of Micro-Implant Anchorage in Initial Orthodontic Retraction CLINICAL Clinical Application of Micro-Implant Anchorage in Initial Orthodontic Retraction Shaji Wahabuddin, MDS Rohan Mascarenhas, MDS* Mahamad Iqbal, DK, MDS Akhter Husain, MDS Micro-implant is a device

More information

Crowded Class II Division 2 Malocclusion

Crowded Class II Division 2 Malocclusion Class II Division 2 Malocclusion Crowded Class II Division 2 Malocclusion Clinicians: Drs. Chris Chang, Hsin-Yin Yeh, Sophia Pei-Wen Shu, W. Eugene Roberts Patient: Miss Jhan Pre-treatment Diagnosis An

More information

Evaluation of cortical bone thickness of mandible with cone beam computed tomography for orthodontic mini implant installation

Evaluation of cortical bone thickness of mandible with cone beam computed tomography for orthodontic mini implant installation ISSN: 2203-1413 Vol.02 No.02 Evaluation of cortical bone thickness of mandible with cone beam computed tomography for orthodontic mini implant installation Seyed Hossein Moslemzade 1, Yusef Kananizadeh

More information

Unusual transmigration of canines report of two cases in a family

Unusual transmigration of canines report of two cases in a family ISSN: Electronic version: 1984-5685 RSBO. 2014 Jan-Mar;11(1):88-92 Case Report Article Unusual transmigration of canines report of two cases in a family Sulabha A. Narsapur 1 Sameer Choudhari 2 Shrishal

More information

B U J O D. Original Research

B U J O D. Original Research Original Research EVALUATATION OF STRESS DISTRIBUTION WITH VARIOUS ORTHODONTIC FORCES IN MAXILLARY CENTRAL INCISOR WITH VARIOUS ROOT MORPHOLOGY - A FINITE ELEMENT STUDY AUTHORS: Parikshit Rao*, Narayan

More information

Biomechanical Analysis of Jaw Bone with Cyst Using CT-Image Based Finite Element Method

Biomechanical Analysis of Jaw Bone with Cyst Using CT-Image Based Finite Element Method Research Article Biomechanical Analysis of Jaw Bone with Cyst Using CT-Image Based Finite Element Method Takaaki Arahira 1, Mitsugu Todo 2* 1 Department of Dental Engineering, Fukuoka Dental College 2-15-1

More information

ABSTRACT. are of greater importance than others, the most important of

ABSTRACT. are of greater importance than others, the most important of 10.5005/jp-journals-10015-1237 Ranjit ORIGINAL Kamble RESEARCH et al Assessment of Positional Variation of Maxillary Permanent First Molar with respect to the Infrazygomatic Crest (Key Ridge) in Skeletal

More information

Anchorage control is a fundamental prerequisite

Anchorage control is a fundamental prerequisite ONLINE ONLY Five-year experience with orthodontic miniscrew implants: A retrospective investigation of factors influencing success rates Joanna Antoszewska, a Moschos A. Papadopoulos, b Hyo-Sang Park,

More information

Maxillary Molar Distalization with micro-implants:

Maxillary Molar Distalization with micro-implants: Predictable Maxillary Molar Distalization with Micro-implant Anchorage in the correction of class II Malocclusion Dr. Ramesh Sabhlok BDS, MDS, Cert. Ortho. (USA), FDS RCS (Edinburgh), FDS RCPS (Glasgow),

More information

Prediction of optimal bending angles of a running loop to achieve bodily protraction of a molar using the finite element method

Prediction of optimal bending angles of a running loop to achieve bodily protraction of a molar using the finite element method Original Article THE KOREAN JOURNAL of ORTHODONTICS pissn 2234-718 eissn 200-372X https://doi.org/10.4041/kjod.2018.48.1.3 Prediction of optimal bending angles of a running loop to achieve bodily protraction

More information

It has been proposed that partially edentulous maxillectomy

It has been proposed that partially edentulous maxillectomy CLASSICAL ARTICLE Basic principles of obturator design for partially edentulous patients. Part II: Design principles Mohamed A. Aramany, DMD, MS* Eye and Ear Hospital of Pittsburgh and University of Pittsburgh,

More information

Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws

Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws Case Report Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws Dr Shih-Yung Lin, 1 Dr Chris Chang, 2 W. Eugene Roberts 3 1I BOI Diplomate, 2 Founder, Beethoven Orthodontic Center,

More information

Canine Extrusion Technique with SmartClip Self-Ligating Brackets

Canine Extrusion Technique with SmartClip Self-Ligating Brackets Canine Extrusion Technique with SmartClip Self-Ligating Brackets Dr. Luis Huanca Ghislanzoni Dr. Luis Huanca received his DDS in 2006 and the MS and Specialist in Orthodontics in 2009 from the University

More information

Class II Correction with Invisalign Molar rotation.

Class II Correction with Invisalign Molar rotation. Tips from your peers to help you treat with confidence. Class II Correction with Invisalign Molar rotation. Dr. Mazyar Moshiri. Class II Correction with Invisalign Molar Rotation. Dr. Mazyar Moshiri. Orthodontic

More information

Three-dimensional finite element analysis of the application of attachment for obturator framework in unilateral maxillary defect

Three-dimensional finite element analysis of the application of attachment for obturator framework in unilateral maxillary defect Journal of Oral Rehabilitation 2007 Three-dimensional finite element analysis of the application of attachment for obturator framework in unilateral maxillary defect S. JIAN*, J. TING*, T. YING &W.DONG-MEI

More information

Osseointegrated dental implant treatment generally

Osseointegrated dental implant treatment generally Placement of Dental Implants Without Flap Surgery: A Clinical Report Bader H. Al-Ansari, BDS, MScD*/Robert R. Morris, DMD** Traditionally, the procedure of implant placement requires a surgical periosteal

More information

Invisalign technique in the treatment of adults with pre-restorative concerns

Invisalign technique in the treatment of adults with pre-restorative concerns Mampieri and Giancotti Progress in Orthodontics 2013, 14:40 REVIEW Open Access Invisalign technique in the treatment of adults with pre-restorative concerns Gianluca Mampieri * and Aldo Giancotti Abstract

More information