BS EN ISO 14971: Risk Analysis and Risk Management Report
|
|
- Hannah Eaton
- 5 years ago
- Views:
Transcription
1 BLACKHORSE ROAD, LETCHWORTH, HERTS, SG6 1HB, ENGLAND Tel: + 44 (0) Fax: + 44 (0) BS EN ISO 14971: Risk Analysis and Risk Management Report Product: Gates and Peeso Classification: Class 2a Medical Device (Multi use Device) Risk Management Report 1. PRODUCTS Gates Drills enlarge the upper portion of the canal for bulk removal of tooth structure. Peeso Reamers are used to enlarge the cervical portion of the canal and to prepare the canal entrance. It creates a parallel preparation and is also used for removing gutta-percha in preparation for post space. Gates Drills and Peeso Reamers are not end cutting, making them both by definition actually reamers. There is a nipple at the end of these instruments that prevents them from cutting at their tip. When the nipple engages the wall of a curved canal the instrument spins and does not cut apically and will not perforate a canal in an apical plane. The cutting head is smaller and the shaft is thinner on the Gates vs. the Peeso and the diameters of the heads are different for the same number instrument. Gates drills and Peeso reamers are Class 2a Medical devices and are for use in the mouth only by (or under the instruction of) a qualified dental professional. They are manufactured in accordance with the standard ISO 1797, and should only be used in conjunction with a rotary hand piece that conforms to ISO the operator should ensure that the rotary instrument is correctly installed in the hand piece prior to commencement of any procedure. 2. RISK MANAGEMENT PLAN The risk management plan is designed to identify and analyse the risks associated with the devices detailed above during their intended usage life cycle. The process verifies as far as is possible the risks and control solutions cited and establishes criteria for risk acceptability. The process includes review of risk management at appropriate stages and as a result of post market surveillance. Risk assessment has been conducted retrospectively on established products but is also conducted as an essential part of the design and development of new or revised products or product types. The process is conducted to comply with the guidance in BS EN ISO Risk management is conducted by the quality management team with guidance as necessary from other departments within the company. Page 1 of 6
2 3. RISK ANALYSIS 3.1. INTENDED USES The Dental surgeon uses Gates Drills to enlarge the upper portion of the canal for bulk removal of tooth structure. The Dental surgeon uses Peeso Reamers to enlarge the cervical portion of the canal and to prepare the canal entrance. It creates a parallel preparation and is also used for removing gutta-percha in preparation for post space CONSTRUCTION Stoddard Gates and Peesos are of one-piece construction and consists of a stainless steel fine ground mandrel with RA shanks to ISO HAZARD IDENTIFICATION 1. The materials used in construction of these devices are inert in the context of their method of use and therefore present no hazards. 2. Improper use of the rotary instrument by the action of the operator. 3. If the instrument became detached from the handpiece during use, injury to the tooth or gums or other parts of the buccal cavity or ingestion or inhalation of the instrument could result 3.4. ESTIMATION OF RISK 1. Under normal conditions of use the hazard presented by rotary instruments are limited by the action of the operator. For the purpose of this risk assessment these are presumed to be adequately controlled through the education, experience and skill of the operator. The inevitable trauma of dental treatment using rotary instruments is reduced by the use of torque limiting features on dental handpieces, and the use of anaesthetic and analgesic agents. 2. If the instrument became detached from the handpiece during use, injury to the tooth or gums or other parts of the buccal cavity or ingestion or inhalation of the instrument could result. There have been no reported incidents of these instruments becoming detached from the handpiece as a result of a defect in the instrument. Such incidents are generally caused by failure of the handpiece collet. The level of risk is considered acceptable. 4. RISK EVALUATION 1. The risks presented by these products in normal conditions are considered to be well controlled by proper use or under the supervision of a suitably qualified dental surgeon. 2. In view of the very low risk associated with the instrument becoming detached from the handpiece, the level of risk is considered acceptable. 5. RISK REDUCTION After evaluation the level of risk is considered acceptable and therefore under these circumstances risk reduction is not considered necessary OPTION ANALYSIS RISK REDUCTION BY DESIGN The use of designs with dimensions able to withstand the stresses of normal use. Where applicable these are generally in accordance with ISO designs. Page 2 of 6
3 PROTECTION AGAINST RISK The quality control measures in place during and following the manufacture of Gates and Peesos are designed to control the incidence of failure, breakage or bending of the product. These have been detailed in the relevant parts of the QA system manuals INFORMATION The directions for use for these products specify conditions of use (intended uses, application speeds etc), which are considered adequate to allow safe and effective use of the devices RISK CONTROL MEASURES No further risk control measures other than those described above are considered necessary RESIDUAL RISK Under proper professional use in accordance with the directions for use it is believed that, apart from those controlled risks cited above, there are no significant residual risks with these products RISK/BENEFIT ANALYSIS None considered necessary 5.5. COLLATERAL HAZARDS There are collateral hazards associated with reprocessing. These are well documented and have been the subject of extensive research in recent years in response to the challenges of hepatitis, HIV and CJD. These instruments should be sterilised before first and subsequent use, the directions for use relating to these products are based on validated reprocessing (sterilisation) methods designed to control the risk from contaminated instruments. 6. RISK EVALUATION SUMMARY These instruments are generally low risk products with no uncontrollable residual risks in proper use. No further risk analysis or control is considered necessary unless there is a significant change in knowledge or legislation that may highlight unbeknown or undocumented hazards. 7. POST PRODUCTION ANALYSIS A procedure is established and maintained to review information gained about the medical device or similar devices in the postproduction phase. The Quality Management Team evaluates this information. Analysis of customer complaints indicates that there have been few incidents or defects with these products that could have harmed the patient or end-user or resulted in any clinical problem. The details of the complaints history are contained in the relevant technical file. Risk Analysis Question Answer C.2.1 What is the intended use and how is the medical device to be used? For general dental surgery. For use by a qualified dental surgeon. C.2.2 Is the Medical device intended to be implanted? The device is not intended to be implanted C.2.3 Is the medical device intended to be in contact with the patient or other persons? Yes. It comes into contact with the patient s body and also with the C.2.4 What materials or components are utilised in the medical device or are used with, or are in contact with the medical device? surgeon s hands. Refer to Product specification and Safety Data sheet. Page 3 of 6
4 C.2.5 Is energy delivered to or extracted from the patient? No. C.2.6 Are substances delivered to or extracted from the patient? No. C.2.7 Are Biological materials processed by the medical device for No. subsequent re-use, transfusion or transplantation? C.2.8 Is the medical device supplied sterile or intended to be sterilised by the user, or are any other microbiological controls applicable? use. C.2.9 Is the medical device intended to be routinely cleaned and disinfected by the user? C.2.10 Is the medical device intended to modify the patient No. environment? C.2.11 Are measurements taken? No. C.2.12 Is the medical device interpretative? No. C.2.13 Is the medical device intended for use in conjunction with other medical devices, medicines or other medical technologies? C.2.14 Are there unwanted outputs of energy or substances? No. C.2.15 Is the medical device susceptible to environmental influences? No. C.2.16 Does the medical device influence the environment? No. C.2.17 Are there essential consumables or accessories associated with the medical device? No. C.2.18 Is maintenance or calibration necessary? No. C.2.19 Does the medical device contain software? No. Yes. The instrument is intended to be sterilised before first and subsequent Yes. The instrument is intended to be sterilised before first and subsequent use. This Instrument is used with Dental Handpieces. C.2.20 Does the medical device have a restricted shelf life? Yes - 5 years. C.2.21 Are there any delayed or long-term use effects? No. C.2.22 To what mechanical forces will the medical device be Rotary Action. subjected? C.2.23 What determines the lifetime of the medical device? Level of wear. C.2.24 Is the medical device single use? No Multi-use. C.2.25 Is safe decommissioning or disposal of the medical device necessary? Yes Suitable disposal is required after use. C.2.26 Does installation or use of the medical device require special training or special skills? Yes - For use by a qualified dental surgeon. C.2.27 How will information for safe use be provided? Refer to Instructions for use. C.2.28 Will new manufacturing processes need to be established or No. introduced? C.2.29 Is the successful application of medical device critically Yes. dependent on human factors such as the user interface? C.2.30 Does the medical device use an alarm system? No. C.2.31 In what way(s) might the medical device be deliberately Use by unqualified personnel. misused? C.2.32 Does the medical device hold data critical to patient care? No. C.2.33 Is the medical device intended to be mobile or portable? No. C.2.34 Does the use of the medical device depend on essential performance? Yes. Page 4 of 6
5 Risk Evaluation Hazard Type Energy Biological Environmental Incorrect use Functional failure Contributing Factors Rotary Power see incorrect use Microbial contamination Metallic component parts Improper use of the rotary instrument by the action of the operator. Instrument becomes detached from the handpiece Possible effect Severity Probability Score Risk control mechanism Injury to the tooth or gums or other parts of the oral cavity due to excessive pressure. Viral infection of patient Inappropriate waste disposal Injury to the tooth or gums or other parts of the oral cavity Injury to the tooth or gums or other parts of the oral cavity or ingestion or inhalation of the instrument could result. S2 S1 3 Instruments are only for use by or under the supervision of a suitably qualified person. Product sterilised prior to use Incineration and recycling is recommended. S2 S1 3 Misuse of rotary instruments is limited by the action of the operator. Presumed to be adequately controlled through the education, experience and skill of the operator. S2 S1 3 In view of the very low risk associated with the instrument becoming detached from the handpiece, the level of risk is considered acceptable. Connection to hand piece is made or supervised by a suitably qualified person. Severity after applying control Probability after applying control Score after applying control Page 5 of 6
6 Explanation of calculation of Risk Evaluation of Products: For the calculation of Risk evaluation the level of acceptable risk has been set by the Company at a score of 10. The Risk Score is calculated by multiplying the score factors from the Severity and Probability, tables listed below: - Risk of Harm Table (Severity level) Common Term Description Rating Score Significant Death or loss of function or structure S3 5 Moderate Reversible or minor injury S2 3 Negligible Will not cause injury or will injure slightly S1 1 Risk of Harm Table (Probability) Common Term Description Rating Score High Likely to happen, often, frequent S3 5 Medium Can happen, but not frequently S2 3 Low Unlikely to happen, rare, remote S1 1 Example: - A product, which is deemed to be a significant risk of harm, but is unlikely to occur would be given a score of 5 X 1 = 5, which is taken from the tables above. i.e. Significant severity resulting in death or loss of function or structure (Score 5) but its probability is low, unlikely to happen, rare, remote (Score 1) Conclusion of Risk Evaluation. From the calculation of risk shown above and with the severity of potential injury assessed after the application of the risk control mechanism the score is reduced from 11 points to 5 points, which is deemed perfectly acceptable and is well within the target acceptable risk of 10 points. Page 6 of 6
BS EN ISO 14971: Risk Analysis and Risk Management Report
BLACKHORSE ROAD, LETCHWORTH, HERTS, SG6 1HB, ENGLAND Tel: + 44 (0) 1462 686221 www.stoddard.co.uk Fax: + 44 (0) 1462 480711 BS EN ISO 14971: Risk Analysis and Risk Management Report Product: Tungsten Carbide
More informationBS EN ISO 14971: Risk Analysis and Risk Management Report
BLACKHORSE ROAD, LETCHWORTH, HERTS, SG6 1HB, ENGLAND Tel: + 44 (0) 1462 686221 www.stoddard.co.uk Fax: + 44 (0) 1462 480711 BS EN ISO 14971: Risk Analysis and Risk Management Report Product: Maxflex Pop
More informationProUltra ULTRASONIC SURGICAL TIPS
ProUltra ULTRASONIC SURGICAL TIPS EN FOR DENTAL USE ONLY DIRECTIONS FOR USE (ULTRASONIC SURGICAL TIPS) A0640 A0650 Satelec M3 x 0,6 EMS M3 x 0,5 ProUltra Surgical tips Ref. A0640 Ref. A0650 1) INDICATIONS
More informationINSTRUCTIONS FOR USE AND CARE
DIAMONDS INSTRUCTIONS FOR USE AND CARE DIAMONDS, MULTI-USE, FRICTION GRIP Quala Diamonds are a rotary cutting device made of stainless steel and coated with diamond particles on the working end. It is
More information1) INDICATIONS FOR USE These products have to be used only in hospital environments, clinics or dental offices by qualified dental personnel.
ProFile 04/06 & O.S. EN DENTAL USE ONLY DIRECTIONS FOR USE PROFILE 0) COMPOSITION The cutting part of these instruments is made of a nickel-titanium alloy. 1) INDICATIONS FOR USE These products have to
More informationD.T. LIGHT-POST. Bisco. Instructions for Use. Radiopaque Translucent Fiber Post System. Double Taper
Bisco D.T. LIGHT-POST Double Taper 0459 Radiopaque Translucent Fiber Post System Instructions for Use IN-083R7 Rev. 8/16 BISCO, Inc. 1100 W. Irving Park Rd. Schaumburg, IL 60193 U.S.A. 847-534-6000 1-800-247-3368
More informationAS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany
Before using the intramedullary. please ensure that it is functioning properly and free of any defects. In addition, please follow these instructions, including the information regarding repair as wellas
More informationUNICLIP DIRECTIONS FOR USE REF C226U - 0,8 MM / 1 MM FOR DENTAL USE ONLY. FISDR / F EN / 11 / updated 06/2016 1/7
UNICLIP FOR DENTAL USE ONLY DIRECTIONS FOR USE REF C226U - 0,8 MM / 1 MM EN FISDR / F19 02 31.EN / 11 / 2001 - updated 06/2016 1/7 1) Indications for use These products have to be used only in hospital
More informationProUltra Ultrasonic Non Surgical Endo Tips
ProUltra Ultrasonic Non Surgical Endo Tips EN FOR DENTAL USE ONLY DIRECTIONS FOR USE (ULTRASONIC NON SURGICAL TIPS) A0620 A0621 A0630 A0631 Satelec M3 x 0,6 EMS M3 x 0,5 ProUltra Endo tips, Zirconium Ref.
More informationD.T. LIGHT-POST. Instructions for Use. Bisco 0459 D.T. LIGHT-POST X-RO. Radiopaque Translucent Fiber Post System ILLUSION
Bisco 0459 D.T. LIGHT-POST -RO D.T. LIGHT-POST ILLUSION Radiopaque Translucent Fiber Post System Instructions for Use RTD Patents: US8298973, US7726971. IN-165R6 Rev. 10/16 BISCO, Inc. 1100 W. Irving Park
More informationCURVES. Anatomically Adaptable IPR strips. Brochure Instructions for use Processing instructions
CURVES Anatomically Adaptable IPR strips Brochure Instructions for use Processing instructions Visit us online! To learn more about all our orthodontic products plus many helpful user videos Innovative
More informationConvex Array Transducer User Guide. Types 8567-S and 8667
Convex Array Transducer User Guide Types 8567-S and 8667 English BB0889-D August 2006 WORLD HEADQUARTERS Mileparken 34 DK-2730 Herlev Denmark Tel.:+45 44528100 / Fax:+45 44528199 www.bkmed.com Email: info@bkmed.dk
More informationTHE. Curve THE CURVE STRIPPING TOOL
THE Curve THE CURVE STRIPPING TOOL Innovative Product Made in Germany THE Curve Mastering the perfect curve Curvable polishing strips for interproximal reduction Sterilizable, double-sided diamond instruments
More informationTools for Endodontic Success
CLEANING & SHAPING Tools for Endodontic Success A Look Inside: C-Files Gates Glidden Drills K-Files Mid-Sized K Files To Dye For Flexicut Files Sodium Hypochlorite Hedstrom Files C.L. Canal Lubricant Reamers
More informationMANUAL FOR USE NO. 5 FLEXIBLE INTRAMEDULLARY REAMER REV ( GB ) non - sterile
MANUAL FOR USE NO. 5 FLEXIBLE INTRAMEDULLARY REAMER REV. 10.10.2013 ( GB ) non - sterile Before using the intramedullary reamer. please ensure that it is functioning properly and free of any defects. In
More informationInnovative revolutionary approach to root canal preparation
Innovative revolutionary approach to root canal preparation The promise Maintaining the anatomic structure of the canal Symmetric removal of the dentine layer Filing the canal walls rather then drilling
More informationS5 Rotary Files. only. Manufacturer. Packing unit. Batch code. For professional use only (0)
EN S5 Rotary Files AR Manufacturer 100 Packing unit www.sendoline.com +46 (0)8 445 88 30 Batch code only For professional use only Catalogue number Consult instructions for use Sendoline AB Box 7037, Tillverkarvägen
More informationLinear Array Transducer
User Guide Type 8809 Linear Array Transducer English BB 0911-H August 2012 For Professional Users Only WORLD HEADQUARTERS Mileparken 34 DK-2730 Herlev Denmark Tel.:+45 44528100 / Fax:+45 44528199 www.bkmed.com
More informationConvex Array Transducer
User Guide Type 8802 Convex Array Transducer English BB0334-I August 2012 For Professional Users Only WORLD HEADQUARTERS Mileparken 34 DK-2730 Herlev Denmark Tel.:+45 44528100 / Fax:+45 44528199 www.bkmed.com
More informationA Guide to Food Hygiene Regulations in the UK
A Guide to Food Hygiene Regulations in the UK Business Information Factsheet BIF193 February 2016 Introduction UK food hygiene regulations (the Regulations) consist of four separate sets of national legislation:
More informationConvex Array Transducer
User Guide Type 8823 Convex Array Transducer English BB1740-C June 2012 For Professional Users Only BK MEDICAL Mileparken 34 2730 Herlev Denmark Tel.:+45 4452 8100 / Fax:+45 4452 8199 www.bkmed.com Email:
More informationRace family Endodontic NiTi rotary files
EN Race family Endodontic NiTi rotary files Foreword Race (Reamer with Alternating Cutting Edges) With the appearance of nickel-titanium (NiTi) alloy, the domain of endodontic instrument production discovered
More informationTable of Contents. Introduction Indications For Use Contraindications Warnings Precautions...5
User Manual 3 Table of Contents Introduction....4 1. Indications For Use...4 2. Contraindications...4 3. Warnings...5 4. Precautions...5 5. Adverse Reactions...5 6. Step-By-Step Instructions...6 A. Contents...6
More informationPreparation Instructions for the CAMLOG /CONELOG Implant System
J8000.0032 Rev.6 EN 06/2015 Preparation Instructions for the CAMLOG /CONELOG Implant System ENGLISH The following descriptions contain detailed instructions on cleaning, disinfection and sterilization
More informationYou succeed your root canal treatments! Characteristics...page 4. Instruments...page 6. Protocol...page 10. Questions & Answers...
We invent HERO Shaper. You succeed your root canal treatments! Summary Characteristics...page 4 Instruments...page 6 Protocol.....page 10 Questions & Answers...page 16 Clinical cases...page 19 2 Introduction
More informationHigh Frequency Linear Array Transducer
User Guide Type 8870 High Frequency Linear Array Transducer English BB1980-C May 2016 For Professional Users Only BK MEDICAL Mileparken 34 2730 Herlev Denmark Tel.:+45 4452 8100 / Fax:+45 4452 8199 www.bkultrasound.com
More informationIntegra Miltex. Endodontic Files. Limit uncertainty with the brand you trust
Limit uncertainty with the brand you trust Liberator Niti Rotary Files The Rotary File that Will Not Self-Thread Liberator rotary nickel-titanium endodontic files are the result of years of research and
More informationINTRAMEDULLARY NAILING TIBIA AND FEMUR 1.3 TIBIAL AND FEMORAL NAILING FEMORAL NAILING WITH RETROGRADE INSERTION IMPLANTS AND OPERATING MANUAL
1.3 TIBIA AND FEMORA NAIING FEMORA NAIING WITH RETROGRADE INSERTION IMPANTS AND OPERATING MANUA INTRAMEDUARY NAIING TIBIA AND FEMUR Medical Products Manufacturing and Trading td. General informations
More information4766 Research Dr. San Antonio, TX insightdentalsystems.com
OVERVIEW OF THE INSIGHT DENTAL IMPLANT DELIVERY SYSTEM The IDS system comes in a unit dose implant system where its advantage provides sterile instrumentation in one single-use kit. It is organized to
More informationInstructions for Use Reprocessed Arthroscopic Shavers
Reprocessed by Instructions for Use Reprocessed Arthroscopic Shavers Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. STERILE
More informationPolicy Title: Clinical Asepsis Policy Policy Number :19. Effective Date: 6/10/2013 Review Date: 6/10/2016
Policy Title: Clinical Asepsis Policy Policy Number :19 6.4.9. Take/send instruments and handpieces to the decontamination/sterilization area. 6.4.10. Remove and dispose of the disposable gown (if used)
More informationEndo Easy Efficient. The efficient NiTi system. Power in Control.
Endo Easy Efficient The efficient NiTi system Power in Control www.vdw-dental.com Intelligent Design Superior Cutting Efficiency and Flexibility For an efficient root canal preparation Each instrument
More informationOPERATING INSTRUCTIONS. Orascoptic DK
User Manual Orascoptic DK is a versatile, 3-in-1 dental device that employs a handheld LED light source and three unique, interchangeable diagnostic instruments. Table of Contents 1. Orascoptic DK Kit
More informationORIFICE FLOW METERS WITH CHAMBERS MEASURING RINGS
Bulletin E79 rev0 02/03/03 ORIFICE FLOW METERS WITH CHAMBERS MEASURING RINGS POP-U-S SERIES FEATURES UNI PN6 flanges: iron Chamber measuring rings: iron Disc: AISI30 Max. operating temperature: 00 C Max.
More informationSafety and Simplicity!
Safety and Simplicity! A safe and precise system for your daily practice A safe and precise system for your daily practice CMA system consists of 4 nickel-titanium rotary endodontic instruments for root
More informationInstructions for Use. Contra-angle handpieces. Endea Endo Cursor EB-62 Endea EB-75 / EB-79. Contra-angle handpieces. Endo NiTi WD-73 M / WD-74 M
Instructions for Use Contra-angle handpieces Endea Endo Cursor EB-62 Endea EB-75 / EB-79 Contra-angle handpieces Endo NiTi WD-73 M / WD-74 M Contents W&H Symbols... 3 1. Introduction... 4 5 2. Before use
More informationNARROW DIAMETER implant
ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball
More informationPropex IQ Apex Locator. Getting Started Guide Just a few images to help you set up your apex locator easily and get the most out of it.
Propex IQ Apex Locator Getting Started Guide Just a few images to help you set up your apex locator easily and get the most out of it. User manual 1 Package contents Propex IQ Apex Locator Propex IQ Clamp
More informationInstructions for use: FlexMaster. the rotary NiTi-System for every case. Endo Easy Efficient
Instructions for use: FlexMaster the rotary NiTi-System for every case Endo Easy Efficient 2 Content 1. Why change to the FlexMaster system? 3 2. Why nickel-titanium? 3 3. Taper, what does it mean? 4 4.
More informationPolicy Title: Single-Use (Disposable) Devices Policy Number: 13. Effective Date: 6/10/2013 Review Date: 6/10/2016
Policy Title: Single-Use (Disposable) Devices Policy Number: 13 5. ROLES AND RESPONSIBILITIES: 5.1. All dental healthcare personnel have responsibility to conform and respect all aspects of this policy.
More informationProstate Biplane Transducer
User Guide Type 8808e Prostate Biplane Transducer English BB1856-D June 2012 For Professional Users Only BK MEDICAL Mileparken 34 2730 Herlev Denmark Tel.:+45 4452 8100 / Fax:+45 4452 8199 www.bkmed.com
More informationEndo Easy Efficient one file endo
Endo Easy Efficient one file endo www.vdw-dental.com one file endo Contents I Reciprocation Reciprocating History Reciprocation by VDW 04 04 05 II The RECIPROC System 06 RECIPROC Instruments Instrument
More informationEndo Easy Efficient. Single File Root Canal Preparation.
Endo Easy Efficient one file endo Single File Root Canal Preparation www.vdw-dental.com INSPIRED to find a simpler and safer way to successfully prepare a root canal, whilst complying with the highest
More informationSt. Jude Medical 8-Channel Adapter. Clinician's Manual
St. Jude Medical 8-Channel Adapter Clinician's Manual CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. For U.S. California Only: Proposition 65, a State of California
More informationSWISS INSTRUMENTS PM
SWISS INSTRUMENTS PM Precision by design PS Instrument [>magnified 8 x ] PIEZON method PIEZON Method} SWISS INSTRUMENTS PM Piezon ultrasound instruments for a complete range of treatments } Endodontics
More informationSWISS INSTRUMENTS PM
SWISS INSTRUMENTS PM Precision by design PIEZON method PIEZON Method} SWISS INSTRUMENTS PM Piezon ultrasound instruments for a complete range of treatments } Endodontics } Scaling } Perio } Cavity preparation
More informationInstructions For Use
Instructions For Use Self Adjusting File Instructions for Use A. Product description: SAF, Self-Adjusting File, a mechanically-operated endodontic file. B. Indication for use: The Self Adjusting file (SAF)
More informationA great file. Even better.
Advantages of Reciprocation and the RECIPROC blue System III A great file. Even better. User guide vdw-dental.com 1 Content I Reciprocation 4 II The RECIPROC blue system 8 III Advantages of reciprocation
More informationGoldspeed EVO S1 / S1-L / M5-L /
97050865 Rev. 00 15.05 Goldspeed EVO S1 / S1-L / M5-L / D1 / D1-L / R20-L / E16 ENGLISH..2 EN ENGLISH 1. INTENDED USE... 3 2. IMPORTANT WARNINGS... 3 3. SYMBOLS... 4 4. TECHNICAL SPECIFICATIONS... 4 5.
More informationDentistry Endodontic instruments. Part 3:
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 3630-3 Second edition 2015-09-01 Dentistry Endodontic instruments Part 3: Compactors: pluggers and spreaders Médecine bucco-dentaire Instruments
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 4074 First edition 2002-02-15 Natural latex rubber condoms Requirements and test methods Préservatifs masculins en latex de caoutchouc naturel Exigences et méthodes d'essai Reference
More informationThis document is a preview generated by EVS
INTERNATIONAL STANDARD ISO 16635-2 First edition 2014-09-15 Dentistry Dental rubber dam instruments Part 2: Clamp forceps Médecine bucco-dentaire Instruments de digue dentaire en caoutchouc Partie 2: Pinces
More informationWhy Single Use of Endodontic Files Makes Sense For You and Your Patients
Why Single Use of Endodontic Files Makes Sense For You and Your Patients The cost of doing business increases every year. That fact is certainly not lost on dental practitioners, who are constantly seeking
More informationfassier DUVal TelescoPIc IM system s RESCUE SYSTEM 2.0
fassier DUVal TelescoPIc IM system s RESCUE SYSTEM 2.0 The Rescue System was conceived to retrieve components of the Fassier- Duval telescopic IM System after the treatment has been completed or in case
More informationSyllabus for International Dental Assistants. Prepare and maintain the dental surgery, instruments, and equipment for clinical dental procedures
Syllabus for International Dental Assistants UNIT1 Prepare and maintain the dental surgery, instruments, and equipment for clinical dental procedures This section concerns the preparation and maintenance
More informationThermoplastic Obturation Systems. Our innovation Your success
Our innovation Your success Introduction CMS Dental produces the largest range of endodontic obturators for the final filling of the root canal. More than 20 million root canals have already been filled
More informationSINGLE USE MEDICAL DEVICES POLICY
SINGLE USE MEDICAL DEVICES POLICY First Issued by/date BWW /BKW March 2001 Issue Version Purpose of Issue/Description of Change Planned Review Date 2 Update May 2011 Named Responsible Officer:- Approved
More informationTechnique Guide Small Bone Fusion System
Technique Guide Small Bone Fusion System The Pinit Plate Small Bone Fusion System is a super low profile, modular bone plate and screw system designed to stabilize a bunionectomy with a medial to lateral
More informationAlpha-Bio Tec specializes in the development and production of advanced dental solutions, focusing on the fields of implants and prosthetics. Over the
Alpha-Bio Tec specializes in the development and production of advanced dental solutions, focusing on the fields of implants and prosthetics. Over the past few years the world of implantology has greatly
More informationTHE MULTI-FUNCTION ULTRASONIC TIPS FOR ALL APPLICATIONS TIP BOOK. Expect Perfection. PR-D265E Ver A. Powerful Partners
TH MULTI-FUNCTION ULTRASONIC TIS FOR ALL ALICATIONS TI BOOK xpect erfection R-D265 Ver.1 05.04.10.A owerful artners Offers A Wide Choice of VARIOS Ultrasonic Scaler Tips For Clinical Applications. rosthetics
More informationSINGLE FILE SYSTEM FOR MODERN ENDODONTICS SMART CONTRA-ANGLE HANDPIECE WITH RECIPROCATING MOTION
SINGLE FILE SYSTEM FOR MODERN ENDODONTICS SMART CONTRA-ANGLE HANDPIECE WITH RECIPROCATING MOTION MOVEMENT BY SWEDEN The S1 System reciprocating, smooth and flexible. Like a dance. FIVE REASONS FOR CHOOSING
More informationTable of Contents. Flexi-Overdenture Characteristics
A s with Flexi-Post Flexi-Overdenture has the unique threaded split shank that creates maximum retention with minimum stress to the root. It is manufactured in stainless steel for the direct/non-coping
More informationIntelligent design is apparent from the detail
W&H oral surgery Intelligent design is apparent from the detail 2 W&H Implantmed The intelligent drive unit from W&H What does it take to make an intelligent piece of equipment like the comprehensively
More informationEnglish. Standard. StandardCylindrical
English Standard StandardCylindrical ATID - Standard Implant The ATID is a standard cylindrical implant with a double 2 X 1.2 mm thread step. It is a simple and easy implant to use, with good control during
More informationone file endo Endo Easy Efficient
one file endo Endo Easy Efficient one file endo Contents I Reciprocation Reciprocating History Reciprocation by VDW 04 04 05 II The RECIPROC System 06 RECIPROC Instruments Instrument Design Instrument
More informationa perfect fit INSTRUCTION MANUAL CAMLOG TORQUE WRENCH WITH CONTINUOUS TORQUE ADJUSTMENT Handling Sterilization Care
a perfect fit INSTRUCTION MANUAL CAMLOG TORQUE WRENCH WITH CONTINUOUS TORQUE ADJUSTMENT Handling Sterilization Care CAMLOG TORQUE WRENCH WITH CONTINUOUS TORQUE ADJUSTMENT INTRODUCTION The torque wrench
More information2. STRAUMANN PRODUCTS COVERED BY THE STRAUMANN GUARANTEE. Abutment attached to an implant* Replacement with equivalent ceramic abutment**
Straumann Guarantee Straumann Guarantee* 1. GUARANTEE BENEFICIARY AND SCOPE This guarantee (the Straumann Guarantee as defined below) from the Institut Straumann AG, Basel, Switzerland ( Straumann ) applies
More informationSterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 11137-3 Second edition 2017-06 Sterilization of health care products Radiation Part 3: Guidance on dosimetric aspects of development, validation
More informationEndo Easy Efficient. one file endo. The RECIPROC Technique Steps to Success.
Endo Easy Efficient one file endo The RECIPROC Technique Steps to Success www.vdw-dental.com 1. Selecting the Correct Instrument Pre-Operative Radiograph DECISION canal is completely visible wide or medium
More informationISO/TS TECHNICAL SPECIFICATION. Dental materials Testing of adhesion to tooth structure
TECHNICAL SPECIFICATION ISO/TS 11405 Second edition 2003-02-01 Dental materials Testing of adhesion to tooth structure Produits dentaires Essai d'adhésion à la structure de la dent Reference number ISO/TS
More informationEndoWave. Enhanced safety for root canal preparation. Thinking ahead. Focused on life.
EndoWave Enhanced safety for root canal preparation Thinking ahead. Focused on life. Features that make a difference The triangular crosssection of the file minimizes the risk of fracture 3-point contact
More informationAlphaKite System. Mechanical preparation of root canals with nickel-titanium files
AlphaKite System Mechanical preparation of root canals with nickel-titanium files All rights of distribution, including photocopying, reprinting (including extracts) or storage and recovery using any type
More informationbook Seite 1 Freitag, 21. April : fjmi^kq=omwn. Operating instructions. båöäáëü. lééê~íáåö=fåëíêìåíáçåë
6278761.book Seite 1 Freitag, 21. April 2017 3:54 15 fjmi^kq=omwn Operating instructions båöäáëü lééê~íáåö=fåëíêìåíáçåë 6278761.book Seite 2 Freitag, 21. April 2017 3:54 15 Table of contents Operating
More informationDisinfection and Decontamination Core Subject. The Process of Instrument Decontamination
Disinfection and Decontamination Core Subject The Process of Instrument Decontamination Aims: To give an overview of the processes involved in dental instrument decontamination using essential quality
More informationYour Dental Benefits. The Local Choice Dental Benefits Program
Your Dental Benefits The Local Choice Dental Benefits Program WELCOME to Delta Dental of Virginia In addition to the largest network of dentists in Virginia* and valuable benefits that help keep out-of-pocket
More informationMini External Fixator.
Mini External Fixator. Assembly and Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Image intensifier control Warning
More informationGewässer Umwelt Schutz
Gewässer Umwelt Schutz GmbH CUW Oil protector Assembly instructions 2 Foreword Always study this assembly instruction manual before Installation and initial operation. This assembly instruction manual
More informationEVER Pharma D-mine Pen Pen injector for Apomorphine 10 mg/ml
EVER Pharma D-mine Pen Pen injector for Apomorphine 10 mg/ml Instructions for Use with Dacepton 3 ml Cartridges Apomorphine hydrochloride hemihydrate solution for injection Subcutaneous Use TABLE OF CONTENTS
More information» Guided Surgery. by MEDENTiKA « MEDENTIGUIDE SURGERY MANUAL MICROCONE. IPS Implant systems
» Guided Surgery by MEDENTiKA «IPS Implant systems MEDENTIGUIDE MICROCONE SURGERY MANUAL » MedentiGuide «MedentiGuide drill sleeves support the surgeon in preparing the implant bed for MEDENTiKA implants.
More informationSmarter Thinking. Simpler Design. Prima Plus. Surgical Manual
Smarter Thinking. Simpler Design. Prima Plus Surgical Manual TABLE OF CONTENTS PRIMA PLUS IMPLANT SURGICAL MANUAL SURGERY Prima Plus Characteristics 4 Surgical Considerations 5 Prima Plus Surgical Sequence
More informationManufacturer s information
WARNINGS: Observe the standard German accident prevention regulations (UVV) We are not aware of any warnings if the instructions for the devices and disinfection and cleaning agents to be used are followed.
More informationFASSIER DUVAL TELESCOPIC IM SYSTEM S RESCUE SYSTEM 2.0
FASSIER DUVAL TELESCOPIC IM SYSTEM S RESCUE SYSTEM 2.0 The Rescue System was conceived to retrieve components of the Fassier- Duval telescopic IM System after the treatment has been completed or in case
More informationStraumann ProClean Cassette. Basic Information on the
Straumann ProClean Cassette Basic Information on the Straumann ProClean Cassette Contents 1. Straumann ProClean Cassette Uncompromised hygiene 2 2. Straumann ProClean Cassette product at a glance 3 2.1
More informationBEST PRACTICE CHANGE CONTROL : DECIDING WHEN TO SUBMIT A 510(K) FOR A DEVICE CHANGE. Yuan Xu 18- JAN- 2018
BEST PRACTICE CHANGE CONTROL : DECIDING WHEN TO SUBMIT A 510(K) FOR A DEVICE CHANGE Yuan Xu 18- JAN- 2018 Background AGENDA What is Medical Device? What is 510(k)? Principles of best practices of device
More informationpiezosteril 6 - piezolight 6
97050228 Rev. 06 15.04 piezosteril 6 - piezolight 6 ENGLISH...2 EN ENGLISH 1. INTENDED USE... 3 2. IMPORTANT WARNINGS... 3 3. SYMBOLS... 4 4. TECHNICAL SPECIFICATIONS... 4 5. CONNECTION TO THE POWER SUPPLY
More informationDescription: Prosthetic Components for dental restoration under CE-Series, CL- Series, CN-Series, CH-Series, CF-Series, CR-Series, CS-Series.
Instructions for use Components for dental restoration. Description: Components for dental restoration under CE-Series, CL- Series, CN-Series, CH-Series, CF-Series, CR-Series, CS-Series. Indications: Abutment
More informationASTM Workshop on Medical Device Cleanliness
ASTM Workshop on Medical Device Cleanliness Steve Goldstine Consultants SteveGoldstine@Yahoo.com 11/16/2010 ASTM: Cleanliness of Medical Devices ASTM Workshop on Medical Device Cleanliness Standard Practices
More informationProducts for a Fast and Reliable Fill
o bt u r at i o n Products for a Fast and Reliable Fill A Look Inside: 2Seal and 2Seal easymix Paste Carrier Gutta Percha Finger Pluggers Paper Points Finger Spreaders Now with Slim, 360º Flex Tip.......................
More informationCannulated Angled Blade Plate 3.5 and 4.5, 90.
Cannulated Angled Blade Plate 3.5 and 4.5, 90. Technique Guide This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Table of Contents Introduction
More informationNotification of a Body in the framework of a technical harmonization directive
Notification of a Body in the framework of a technical harmonization directive From : Ministry of Health ul. Miodowa 15 00-952 Warszawa Poland To : European Commission GROWTH Directorate-General 200 Rue
More information» Guided Surgery. by MEDENTiKA « MEDENTIGUIDE SURGERY MANUAL QUATTROCONE QUATTROCONE30. IPS Implant systems
» Guided Surgery by MEDENTiKA «IPS Implant systems MEDENTIGUIDE QUATTROCONE QUATTROCONE30 SURGERY MANUAL » MedentiGuide «MedentiGuide drill sleeves support the surgeon in preparing the implant bed for
More informationPinit Plate Small Bone Fusion System Bone Plate & Screw System
Pinit Plate Small Bone Fusion System Bone Plate & Screw System Description The Pinit Plate Small Bone Fusion System consists of 2-hole bone plates made available in three length options and two thickness
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 6360-7 First edition 2006-02-01 Dentistry Number coding system for rotary instruments Part 7: Specific characteristics of mandrels and special instruments Art dentaire Système
More informationYour Dental Benefits The Local Choice. Retiree Health Benefits Program for Medicare-Eligible Retirees
Your Dental Benefits The Local Choice Retiree Health Benefits Program for Medicare-Eligible Retirees Welcome to Delta Dental of Virginia In addition to the largest network of dentists in the nation* and
More informationHDM 330/500 Operating instructions Mode d emploi Manual de instrucciones Manual de instruções
HDM 330/500 Operating instructions Mode d emploi Manual de instrucciones Manual de instruções en fr es pt Printed: 16.12.2014 Doc-Nr: PUB / 5208057 / 000 / 01 1 Printed: 16.12.2014 Doc-Nr: PUB / 5208057
More informationEndo Easy Efficient.
Endo Easy Efficient www.vdw-dental.com Intelligent drive technology to meet the highest demands Optimal motor performance Regardless of the method of preparation, the RECIPROC motors ensure reliability
More informationbook Seite 1 Mittwoch, 26. Juli : qq=ifkb=l=^áê=jçíçê. Operating Instructions / T4 LINE motor kit. båöäáëü. lééê~íáåö=fåëíêìåíáçåë
6397959.book Seite 1 Mittwoch, 26. Juli 2017 2:52 14 qq=ifkb=l=^áê=jçíçê Operating Instructions / T4 LINE motor kit båöäáëü lééê~íáåö=fåëíêìåíáçåë 6397959.book Seite 2 Mittwoch, 26. Juli 2017 2:52 14 Table
More informationSafety Data Sheet Product Name: Heparin / PF4 Serum Panel - For In Vitro Diagnostic Use Only
Section 1 Identification Product Name Catalog Numbers Heparin / PF4 Assayed Antibody Serum Panel 4036027, 2 Member QC Panel 4036028, Multi-Member Qualification Panel (6 member) Company Identification AKERS
More informationSWISS INSTRUMENTS PM
SWISS INSTRUMENTS PM PRECISION BY DESIGN PS Instrument [> magn nified 8 x] PIEZON METHOD PIEZON METHOD} SWISS INSTRUMENTS PM PIEZON ULTRASOUND INSTRUMENTS FOR A COMPLETE RANGE OF TREATMENTS } ENDODONTICS
More information