Intensity output and effectiveness of light curing units in dental offices

Size: px
Start display at page:

Download "Intensity output and effectiveness of light curing units in dental offices"

Transcription

1 Journal section: Operative Dentistry and Endodontics Publication Types: Research doi: /jced Intensity output and effectiveness of light curing units in dental offices Baharan-Ranjbar Omidi 1, Armin Gosili 2, Mona Jaber-Ansari 3, Ailin Mahdkhah 4 1 Assistant Professor, Department of Operative Dentistry, Faculty of dentistry, Qazvin University of Medical science, Qazvin, Iran 2 Assistant Professor, Department of Orthodontics, Faculty of dentistry, Golestan University of Medical science, Gorgan, Iran 3 Dentist, Private Practice, Tehran, Iran 4 Post-graduate Student of Operative Dentistry, Faculty of Dentistry, Qazvin University of Medical Science, Qazvin, Iran Correspondence: Department of Operative Dentistry Faculty of Dentistry, Qazvin University of Medical Science Qazvin, Iran mahdkhahailin@yahoo.com Received: 20/02/2018 Accepted: 11/04/2018 Omidi BR, Gosili A, Jaber-Ansari M, Mahdkhah A. Intensity output and effectiveness of light curing units in dental offices. J Clin Exp Dent. 2018;10(6):e Article Number: Medicina Oral S. L. C.I.F. B eissn: jced@jced.es Indexed in: Pubmed Pubmed Central (PMC) Scopus DOI System Abstract Background: The aims of the study were measuring the light intensity of light curing units used in Qazvin s dental offices, determining the relationship between the clinical age of these units and their light intensity, and identifying the reasons for repairing them. Material and Methods: In this cross-sectional study, the output intensity of 95 light curing devices was evaluated using a radiometer. The average output intensity was divided up into four categories (less than 200, , , and more than 500 mw/cm 2 ). In addition, a questionnaire was designed to obtain information mainly about the type, clinical age, and frequency of maintenance of the units and the reasons for fixing them. Data were analyzed using Kolmogorov-Smirnov, chi-squared, and t-tests (p < 0.05) on SPSS 24. Results: A total of 95 light curing units were examined, with 61 (64.2%) of them being of the LED type and 34 (35.8%) of the QTH type. While average light intensity in LED units was significantly higher than in QTH devices, the two device types were not significantly different regarding desirable light intensity (i.e., 300 mw/cm 2 ). A negative correlation was observed between clinical age and light intensity. In addition, bulb replacement in QTH devices was over three times as much as in LED units. Also, repairing QTHs was more than twice as much frequent as fixing LEDs. The most common reason for repair was the breakage of the tip of the device. Conclusions: The light intensity of LED units is significantly higher than that of QTH devices, and the frequency of repairing in QTHs was significantly more than in LEDs. Furthermore, light intensity decreases with aging, and dentists should regularly monitor the conditions of light units. Key words: Light curing unit, radiometer, light intensity, dental equipment, dental offices. e555

2 Introduction Over three decades have passed since the beginning of the extensive use of composite resins in dentistry, and the demand for using esthetic restorative materials is still on the increase (1). Resins must begin polymerization in order to perform operation. During this process, monomer units bond with each other to build long and heavy polymers. Due to the increased use of optical composites, the importance of polymerization has become more prominent. The strength of these restorations depends on the degree of polymerization of composite resins. Incomplete polymerization produces adverse biological effects, increasing water absorption, composite solubility, and reducing hardness. Various factors contribute to the polymerization of the composites, and they include the wavelength and intensity of the output of light curing units, duration of radiation, dimensions and location of the dental cavity, direction and distance of the tip of the device (related to the composite), the composition of the composite, the wavelength and bandwidth of the curing light, the intensity of the curing light, the irradiation time, and color and thickness of the composite (2,3). In composite resins, camphorquinone is the light-sensitive component, which responds to irradiation by creating free radicals and initiates the polymerization process (4). An appropriate intensity of light with the maximum absorption wavelength range of camphorquinone is the main factor in the polymerization of these resins. If the light output intensity decreases, it will adversely influence the clinical and cosmetic performance. The light intensity of curing devices is defined by the International Organization for Standardization as the ISO 4049 standard, which recommends an intensity of 300 mw/cm 2 with a wavelength bandwidth of nm on the tip of the light curing device. At this standard wavelength, the minimum depth of cure is assumed to be 1.5 mm, which is 50% of the length of the composite specimen (3). The reduction in the light intensity of the device can affect the success rate of the restorative methods via reducing the degree of convergence of composites, which leads to an increase in microleakage and recurrent caries (5). The light source for polymerization of composite resins are available in four types: quartz-tungsten-halogen (QTH), light-emitting diode (LED), plasma arc curing (PAC), and argon laser. Halogen-based curing lamps have several limitations. One of the main disadvantages of these lamps is the high energy consumption. Only 1% of the consumed energy by these devices turns into light and almost all the remaining energy is converted into heat. The heat generated by these lamps should be eliminated, and this requires expensive thermal filters. Cooling fans are also loud and bulky. Also, the longevity of halogen lamps is short (between 40 and 100 hours) (6). In 1995, Mills and colleagues presented solid-state LED technology for the polymerization of dental materials capable of being activated with light. In LEDs, instead of hot strands as used in halogen lamps, semiconductor connections are employed to produce light. These lamps have a very long shelf life of about 1,000 hours and can withstand mechanical shocks and vibrations with very low error rates (7). LEDs are also capable of producing blue light at a wavelength of nm. LEDs can be cordless and are almost silent while being operated (7). In QTH and LED light curing devices, the main factors affecting the intensity of light output are: inappropriate performance of the lamp and filter, breakage and pollution of the device tip, the blurring of the bulb, the failure of electrical components, and defect in light transmitting fibers (6,7). In these devices, if maintenance is not carried out routinely, after a while, there will be some problems with the lamp, fan, or power supply (8). There are two main problems with the quality of cured resin composite in the office: 1) Composite surface hardness is not a reliable guide because even at a low light intensity, the surface can sufficiently harden while the depth of the cure is not adequate. Moreover, it is impossible for the dentist to distinguish completely-cured composite resin from the one incompletely cured using a device with a low light intensity (9). 2) The output light of the device decreases as the device is used more, but this is not detectable by the unarmed eye because sometimes a seemingly bright light is not suitable for wavelengths. Furthermore, insufficient radiation intensity is not always compensated for by prolonging exposure time (10). Therefore, a digital radiometer is needed to measure the intensity of the curing light of the units to determine when the device needs to repaired or replaced (11). The aims of the present study were four-fold: 1) measuring the light intensity of light curing units used in the offices, 2) comparing the light intensity of LED and QTH units, 3) determining the relationship between the clinical age of these devices and their light intensity, and 4) exploring the reasons for and the frequency of repairing these units. The findings of this study underscore the importance of timely fixing or replacing defective light curing devices, which can consequently ensure the continued quality of restorative treatments. This improvement can increase public health in the long run. It is believed that these findings can be used in the macro-planning of the health system. Material and Methods The following criteria were used to include private dental offices in this study: 1) the person operating the office had to be a general dentist rather than a specialist, and 2) the dentist had to routinely use composite resins for too- e556

3 th restoration. The 2016 alphabetical listing of dentists published by the Qazvin University of Medical Sciences was used to identify dental offices located in Qazvin. All the 105 offices on the list were contacted via telephone to determine if they satisfied the criteria. Of these offices, 88 fulfilled both criteria. The dentists in charge of these latter offices were contacted by telephone to explain the rationale and methodology of the study and to obtain his or her consent. Ultimately, 70 dental offices agreed to participate. Upon this, an appointment was arranged for a visit to the office. To measure light intensity, an analog radiometer (Digi- Rate, Monitex, Taiwan) with a range of 0 to 1,000 mw/ cm 2 was used. The radiometer was sent to the Laboratory of Optics at Sharif University of Technology in Tehran, Iran to confirm its performance. Once approval was received, the offices were visited to perform the measurements. At each office, a few minutes were spared to allow the radiometer to match up with the ambient temperature. Then, three measurements of light intensity were recorded for each light curing unit, and the average was reported as the final measure. At the intervals between visits, the accuracy of the radiometer was checked against a radiometer at the restorative dentistry laboratory of the Qazvin University of Medical Sciences. Light intensity of less than 300 mw/cm 2 was considered unacceptable (3), and a device with a light intensity of less than 200 mw/cm 2 was regarded unusable (12). In addition, a questionnaire, designed by the researchers, was used to obtain information about type of light curing unit, the age of the unit, the frequency of maintenance, the reasons for repair, the date of the last repair, the number of times the bulb had been replaced and the last time this had been done, the typical duration of light irradiation at each restoration, the availability of a radiometer in the office, and the number of office hours per week. Once the data had been collected, they were analyzed using SPSS 24 (IBM Corporation, USA, 2016). The mean and standard deviation were used to determine the mean light intensity of the light curing units. The Kolmogorov-Smirnov statistic was used to determine the normality of the distribution of the light intensity scores. The clinical age of the device was calculated using the following equation: Clinical age = time in use (in terms of year) 52 (number of weeks in a year) number of working days of the office per week average number of instances of the use of the device per day average duration of light exposure (in terms of second) at each instance of the use of the device. Pearson product-moment (r) correlation coefficient was used to determine the correlation between the clinical age and the light intensity of the device. Independent-samples t-test was used to compare the light intensity of LED and QTH units. The chi-squared test was employed to find out about the correlation between device type on the one hand and the frequency of bulb replacement, the frequency of device repair, and the reason for device repair on the other hand. Statistical significance for all tests was set at an α level of < 0.05 (2-tailed). -Ethical considerations The study was endorsed by the Research Ethics Committee of Qazvin University of Medical Sciences under the approval ID of IR.QUMS.REC Additionally, the information contained in this study does not mention the names of the honorable participating dentists. Results -Device type A total of 95 light curing units were examined: 61 (64.2%) were LED and 34 (35.8%) were QTH. -Correlation between device type and light intensity The light intensity of LED and QTH units is shown in Table 1. As can be seen, about 93% of LED devices had a desirable light intensity (i.e., 300 mw/cm 2 ), whereas this figure is around 97% for QTH units. Results from the t-test showed that the relationship between the two variables was insignificant regarding desirable light intensity. However, average light intensity in LED units was significantly higher than in QTH devices. -Relationship between clinical age and light intensity Pearson r revealed a negative relationship between clinical age and light intensity (Table 2). Table 1: Distribution of acceptable/nonacceptable light intensity for the curing devices under study. Curing Light intensity (mw/cm devices <200 nonacceptable nonacceptable acceptable acceptable QTH 0 (0%) 1 (2.9%) 15 (44.1%) 18 (52.9%) ↆ LED 0 (0%) 4 (6.6%) 8 (13.1%) 49 (80.3%) Total 0 (0%) 5 (5.3%) 23 (24.2%) 67 (70.5%) : quartz-tungsten-halogen. 1 : light emitting diode. 2 ) e557

4 Table 2: Clinical age and light intensity of light curing devices. Clinical age Light intensity (mw/cm Total (hour) > (0%) 1 (1.1%) 25 (26.3%) 26 (27.4%) (2.1%) 5 (5.3%) 19 (20.0%) 26 (27.4%) (3.2%) 10 (10.5%) 14 (14.7%) 27 (28.4%) (0%) 5 (5.3%) 5 (5.3%) 10 (10.4%) (0%) 1 (1.1%) 2 (2.1%) 3 (3.2%) >230 0 (0%) 1 (1.1%) 2 (2.1%) 3 (3.2%) 2 ) -Relationship between device type and the frequency of replacing device bulbs Table 3 shows the relationship between device type and the frequency of replacing the bulb. It is evident that bulb replacement in QTH devices was over three times as much as in LED units. The chi-squared test showed a significant relationship between the two variables. -Relationship between device type and the frequency of device repair The relationship between device type and the frequency of device repair is given in Table 4. According to the chi-squared test results (Table 4), the relationship between these two variables was statistically insignificant. -Relationship between device type and the reasons for device repair According to Table 5, the most common reason for repair was the breakage of the tip of the device. More specifically, this was the cause in 78% of all instances of repair in the case of LED devices, and 40% in the case of QTH units. The chi-squared test revealed a significant relationship between the two variables. The devices studied are based on the model presented in Table 6. The most common model used in dentistry offices in Qazvin is the Woodpecker model, which has a good radiation intensity in all cases. It was also found that the Gnatus light cure device had the lowest percentage of light-intensity. Table 3: Frequency of replacing the bulb in LED and QTH devices. N LED QTH p-value 0 51 (83.6%) 14 (41.2%) 1 5 (8.2%) 14 (41.2%) 2 4 (6.5%) 6 (17.6%) 3 1 (1.7%) 0 (0%) Total 61 (100%) 34 (100%) Bulb replacing Table 4: Relationship between device type and the frequency of device repair. N LED QTH p-value 0 52 (85.2%) 24 (70.5%) 1 8 (13.1%) 7 (20.5%) 2 0 (0%) 3 (9.0%) 3 1 (1.7%) 0 (0%) Total 61 (100%) 34 (100%) Device repair Discussion This study found that most of the devices were LED-type. Despite the popularity of halogen devices in the past, there are some problems with QTHs. As noted above, these include the short life span of halogen lamps and the fact that the reflector and filter can undergo a reduction in efficiency over time. Overheating is one of the other disadvantages of such devices (6). However, these issues are less frequently seen in LEDs, which has resulted in a rapid growth of the use of these units. Another upside of LEDs is that their light output wavelength peak of 456 nm matches the absorption peak of camphorquinone (13). However, initiators other than camphorquinone have slightly different absorption spectra. Thus, it is better to use halogen devices with these ma- Table 5: Relationship between device type and the reasons for device repair. Device Reason for device repair Device tip Blown Fan Cord Power switch Total p-value breakage fuse failure change failure LED 7 (11.4%) 0 (0%) 0 (0%) 0 (0%) 2 (3.3%) 9 (14.7%) QTH 4 (11.7%) 2 (5.8%) 3 (8.8%) 1 (2.9%) 0 (0%) 10 (29.4%) e558

5 Table 6: Models of light curing units used in Qazvin s dental offices (2017). Light intensity Arialux Astralis Blue dent Blue light Coltolux.2.5 Coltolux.50 Dentine Dentmate > Total arazdentine Firozdental Gnatus Litex Monitex Regenesis SANY Ski Woodpecker Total terials, but they may not fully cured with LEDs because the light output wavelength range of halogen is wider than that of LEDs. On the other hand, spectral purity in LED devices allows better polymerization of composites with camphorquinone. In addition, LEDs do not emit too much heat (14). The first aim of the present study was measuring the light intensity of light curing units in use in the dental offices of the city of Qazvin. It was found that none of the devices had a light intensity of less than 200 mw/ cm 2 and that 6.6% of LEDs, 2.9% of QTHs, and 5.3% of all total devices had a light intensity of less than 300 mw/cm 2, the desirable level. Unlike this study, Miyazaki et al. (10), Barghi et al. (1994), (6) and Martin (15) reported high percentages: 41.9%, 29.7% and 27%, respectively. In 2009, Javaheri and Ashreghi (16) concluded that the light intensity of 27.4% of the devices in their study was less than desirable. Since that study was also carried out in Qazvin, the much lower percentage reported in the present study can be attributed to the increased awareness among the dentists of the necessity of continued maintenance of light curing units over recent years. On the other hand, our findings are similar to those of Barghi et al., (11) and Savadi Oskoee (8), who reported an undesirable percentage of 10.4% and 10%, respectively. Overall, the wide variety in the results reported in the literature can be ascribed to the variety in the devices in terms of type, maintenance, and clinical age. Indeed, the clinical age of 94% of devices in the present study did not exceed 170 hours. Another possible cause is the cleanliness of the tip of the devices as research has shown that the gradual buildup of the debris of composite resins on the curing tip can significantly decrease light intensity (12,17). This was not measured in the present study. Another observation was that the light intensity of LEDs is significantly higher than that of QTHs, which may be due to the difference in the output light spectrum and clinical age of devices. This finding is consistent with a study by Hegde et al. (18). It was also observed that the clinical age of the devices is negatively correlated with the intensity of light. A similar finding was reported by Barghi et al. (6), Martin (15), Poulos and Styner (19) and Friedman (20). However, Javaheri and Ashreghi (16) concluded that there is no significant correlation between clinical age and light intensity. Two possible reasons for the difference in the results can be the use of different models of light curing devices in different studies and different levels of device consciousness on the part of participating dentists. Furthermore, a significant relationship was found between the frequency of bulb replacement and the type of device, with greater frequency in the case of QTHs (58.8% as opposed to 16.4% for LEDs). This is attributable to the longer life span of an LED lamp, which is over 1,000 hours as opposed to 100 hours for a QTH lamp (7). Also, the relationship between the frequency of lamp replacement and light intensity did not teach significance. However, it should be noted that only a small number of devices in our study had their lamp replaced. It is worth noting at this point that more frequent bulb replacement is an indicator of the aging of the device, which in turn signals the aging of the other parts of the device as the potential factors contributing to light intensity. The study also found that the main reason for repair was the breakage of the tip of device. This clearly shows the vulnerability of the tip to breakage. Finally, the most common model used in the dental offices of Qazvin was Woodpecker (Woodpecker Med. Instrument, Guilin, china), which was found to have the favorable light intensity of all models. It was also found that the Gnatus model (Gnatus Medical-Dental Equipments, Brazil) has the lowest light intensity, but it should be noted that there were only a small number of Gnatus devices, which were also very old. However, the dentist s attention to the periodic measurement of the intensity of the device and doing the necessary repairs can increase the efficiency of the device. Conclusions This study concluded the following: The light intensity of the light curing devices in the dental offices of Qazvin was acceptable (i.e., over 300 mw/ cm 2 ). 1. The light intensity of LEDs was significantly higher than that of QTH devices. 2. Light intensity decreases with the aging of the device. 3. The frequency of bulb replacement in QTHs was significantly higher than in LEDs. e559

6 4. There was a significant relationship between device model and light intensity. 5. For dentists who do not carry out a regular maintenance of light curing devices, it is better to use LEDs because these devices have a higher light intensity and a longer bulb life span. 6. Regarding the fact that reduction in light intensity can affect the success rate of restorative methods, light intensity measurement should be carried out regularly. 20. Friedman J, Hassan R. Comparison study of visible curing lights and hardness of light-cured restorative materials. The Journal of prosthetic dentistry. 1984;52: Conflict of interest The authors have declared that no conflict of interest exist. References 1. Dunne S, Davies B, Millar B. A survey of the effectiveness of dental light-curing units and a comparison of light testing devices. British dental journal. 1996;180: El-Mowafy O, El-Badrawy W, Lewis DW, Shokati B, Kermalli J, Soliman O, et al. Intensity of quartz-tungsten-halogen light-curing units used in private practice in Toronto. Journal of the American Dental Association (1939). 2005;136:766-73; quiz Fan PL, Schumacher RM, Azzolin K, Geary R, Eichmiller FC. Curing-light intensity and depth of cure of resin-based composites tested according to international standards. Journal of the American Dental Association (1939). 2002;133:429-34; quiz Willems G, Lambrechts P, Braem M, Celis JP, Vanherle G. A classification of dental composites according to their morphological and mechanical characteristics. Dent Mater. 1992;8: Aguiar FHB, Ajudarte KF, Lovadino JR. Effect of light curing modes and filling techniques on microleakage of posterior resin composite restorations. Operative dentistry. 2002;27: Barghi N, Berry T, Hatton C. Evaluating intensity output of curing lights in private dental offices. Journal of the American Dental Association (1939). 1994;125: Goyal A, Jyothikiran H, Shivalinga BM. Use of Light Curing Units in Orthodontics: A Review. Annals of Dental Research. 2011;1: Savadi Oskoee S, Poor Abbas R, Hafezehquran A. Evaluation of light curing units effectiveness used in clinics and private dental offices of Tabriz, Journal of Dental School Shahid Beheshti University of Medical Sciences. 2004;22: Hansen EK, Asmussen E. Reliay of three dental radiometers. European Journal of Oral Sciences. 1993;101: Miyazaki M, Hattori T, Ichiishi Y, Kondo M, Onose H, Moore BK. Evaluation of curing units used in private dental offices. Operative dentistry. 1998;23: Barghi N, Fischer DE, Pham T. Revisiting the intensity output of curing lights in private dental offices. Compendium of continuing education in dentistry (Jamesburg, NJ : 1995). 2007;28:380-4; quiz Rueggeberg FA, Caughman WF, Curtis JW Jr, Davis HC. Factors affecting cure at depths within light-activated resin composites. American journal of dentistry. 1993;6: Chen YC, Ferracane JL, Prahl SA. Quantum yield of conversion of the photoinitiator camphorquinone. Dental materials : official publication of the Academy of Dental Materials. 2007;23: Althoff O, Hartung M. Advances in light curing. American journal of dentistry. 2000;13(Spec No):77d-81d. 15. Martin FE. A survey of the efficiency of visible light curing units. Journal of dentistry. 1998;26: Javaheri M, Ashreghi M. Evaluation of curing light intensity in private dental offices (2005). The Journal of Qazvin University of Medical Sciences. 2009;12: Duke ES. Light-emitting diodes in composite resin photopolymerization. Compendium of continuing education in dentistry (Jamesburg, NJ : 1995). 2001;22: Hegde V, Jadhav S, Aher GB. A clinical survey of the output intensity of 200 light curing units in dental offices across Maharashtra. Journal of conservative dentistry : JCD. 2009;12: Poulos JG, Styner DL. Curing lights: changes in intensity output with use over time. General dentistry. 1997;45:70-3. e560

Curious Curing. How do you know if you re getting the right output from your LCU? 50 JULY 2016 // dentaltown.com

Curious Curing. How do you know if you re getting the right output from your LCU? 50 JULY 2016 // dentaltown.com Curious Curing How do you know if you re getting the right output from your LCU? by Richard Price, BDS, DDS, MS, FRCD(C), PhD Dr. Richard Price is the division head for fixed prosthodontics at Dalhousie

More information

WHY PLASMA TECHNOLOGY?

WHY PLASMA TECHNOLOGY? WHY PLASMA TECHNOLOGY? Why Plasma Technology? With the increased number of curing lights in the market today, a dental professional is faced with a number of choices. In past years, the quartz-tungsten-halogen

More information

Assistant Professor and Consultant, Department of Restorative Dental Science, College of Dentistry, Taibah University, Al Madinah Al

Assistant Professor and Consultant, Department of Restorative Dental Science, College of Dentistry, Taibah University, Al Madinah Al Light Intensity Decay in Quartz-Tungsten-Halogen Polymerization Units Khalid H Al-Samadani 1, Ayman Al-Dharrab 2, Mohammed A Wahbi 3, Loay I Algizani 4 1 Assistant Professor and Consultant, Department

More information

Efficacy of New LED Light-Curing Units in Hardening of Class II Composite Restorations

Efficacy of New LED Light-Curing Units in Hardening of Class II Composite Restorations Applied RESEARCH Efficacy of New LED Light-Curing Units in Hardening of Class II Composite Restorations Omar El-Mowafy, BDS, PhD, FADM; Wafa El-Badrawy, BDS, MSc; Mira Wasef; Hanan Omar, BDS, MSc; Sanaz

More information

Effect of light curing units on physical properties of composites

Effect of light curing units on physical properties of composites VARIA Effect of light curing units on physical properties of composites Özcan Çakmakçioglu 1, Pinar Yilmaz 1, Bülent Topbaºi 2 Istanbul, Turkey Summary Objective. The objective of this study is to determine

More information

Influence of the light curing unit and thickness of residual dentin on generation of heat during composite photoactivation

Influence of the light curing unit and thickness of residual dentin on generation of heat during composite photoactivation 137 Journal of Oral Science, Vol. 50, No. 2, 137-142, 2008 Original Influence of the light curing unit and thickness of residual dentin on generation of heat during composite photoactivation Ricardo D.

More information

The use and maintenance of visible light activating units in general practice

The use and maintenance of visible light activating units in general practice The use and maintenance of visible light activating units in general practice B. A. Mitton, 1 and N. H. F. Wilson, 2 Aim The present study to investigate the use, care and maintenance of light units in

More information

Evaluation of a Second-Generation LED Curing Light

Evaluation of a Second-Generation LED Curing Light A P P L I E D R E S E A R C H Evaluation of a Second-Generation LED Curing Light Richard B.T. Price, BDS, DDS, MS, FDS RCS (Edin), FRCD(C), PhD Corey A. Felix, BSc, MSc Pantelis Andreou, PhD A b s t r

More information

Evaluation of Curing Light Distance on Resin Composite Microhardness and Polymerization

Evaluation of Curing Light Distance on Resin Composite Microhardness and Polymerization Operative Dentistry, 2007, 32-6, 571-578 Evaluation of Curing Light Distance on Resin Composite Microhardness and Polymerization KM Rode Y Kawano ML Turbino Clinical Relevance Different light curing units,

More information

Effectiveness of Second-generation Light-emitting Diode (LED) Light Curing Units

Effectiveness of Second-generation Light-emitting Diode (LED) Light Curing Units Effectiveness of Second-generation Light-emitting Diode (LED) Light Curing Units Aim: The purpose of this study was to evaluate the effectiveness of three commercially available light emitting diode (LED)

More information

Vickers Hardness of Composite Resins Cured with LED and QTH Units

Vickers Hardness of Composite Resins Cured with LED and QTH Units Journal of Dental Biomaterials. 2016;3(1) Original Article Vickers Hardness of Composite Resins Cured with LED and QTH Units Alaghemand H a, Ramezani M b, Abedi H c, Gholamrezaee Saravi M c, Zarenejad

More information

Comparison of Quartz-Tungsten-Halogen, Light-emitting Diode, and Plasma Arc Curing Lights

Comparison of Quartz-Tungsten-Halogen, Light-emitting Diode, and Plasma Arc Curing Lights Comparison of Quartz-Tungsten-Halogen, Light-emitting Diode, and Plasma Arc Curing Lights Richard B.T. Price a /Lars Ehrnford b /Pantelis Andreou c /Corey A. Felix d Purpose: This study determined which

More information

Effect of. Conversion Degree and Hardness of Dual-cured Resin Cement

Effect of. Conversion Degree and Hardness of Dual-cured Resin Cement Operative Dentistry, 2010, 30-5, 120-124 Effect of Light Energy Density on Conversion Degree and Hardness of Dual-cured Resin Cement PCP Komori AB Paula AA Martin RN Tango MAC Sinhoreti L Correr-Sobrinho

More information

Comparison of the Heat Generation of Light Curing Units

Comparison of the Heat Generation of Light Curing Units Comparison of the Heat Generation of Light Curing Units Abstract Aim: The aim of this study was to evaluate the heat generation of three different types of light curing units. Methods and Materials: Temperature

More information

The Influence of Temperature on the Efficacy of Polymerization of Composite Resin

The Influence of Temperature on the Efficacy of Polymerization of Composite Resin The Influence of Temperature on the Efficacy of Polymerization of Composite Resin Abstract Aim: The purpose of this study was to investigate the effect of different temperatures on the efficacy of polymerization

More information

Comparison of Temperature Changes in the Pulp Chamber Induced by Various Light Curing Units, In Vitro

Comparison of Temperature Changes in the Pulp Chamber Induced by Various Light Curing Units, In Vitro Operative Dentistry, 2006, 31-2, 261-265 Comparison of Temperature Changes in the Pulp Chamber Induced by Various Light Curing Units, In Vitro AR Yazici A Müftü G Kugel RD Perry Clinical Relevance The

More information

Degree of conversion and temperature rise during polymerization of composite resin samples with blue diodes

Degree of conversion and temperature rise during polymerization of composite resin samples with blue diodes Journal of Oral Rehabilitation 2001 28; 586±591 Degree of conversion and temperature rise during polymerization of composite resin samples with blue diodes A. KNEZÏ EVICÂ *, Z. TARLE*, A. MENIGA*, J. SÏ

More information

Opacity and Color Changes of Light-Cured Ideal Makoo (IDM)

Opacity and Color Changes of Light-Cured Ideal Makoo (IDM) Ghavam M, Goodarzy A. Journal of Dentistry, Tehran University of Medical Sciences Opacity and Color Changes of Light-Cured Ideal Makoo () Ghavam M 1, Goodarzy A 1 Assistant Professor, Dept of Operative

More information

Department of Prosthodontics, Faculty of Dentistry, Mahidol University, Bangkok, Thailand

Department of Prosthodontics, Faculty of Dentistry, Mahidol University, Bangkok, Thailand Dental Journal Mahidol Dental Journal Original Article Dimensional stability of compression and injection molding denture bases in long and short curing procedures Noppavan Nagaviroj, Krantarat Wanitchanont,

More information

Introduction. Device Description. Dual-Band Spectral Output

Introduction. Device Description. Dual-Band Spectral Output Optimum Spectrum and Pulse Shape for Vascular Lesion Treatment: The Science Behind MaxG James Childs, Ph.D.; Andrei Erofeev, Ph.D.; Mikhail Smirnov, Ph.D.; and Gregory Altshuler, Ph.D., Sc.D. Introduction

More information

Light curing in dentistry and clinical implications: a literature review

Light curing in dentistry and clinical implications: a literature review Critical Review Dental Materials/Dentistry Light curing in dentistry and clinical implications: a literature review Frederick Allen RUEGGEBERG (a) Marcelo GIANNINI (b) Cesar Augusto Galvão ARRAIS (c) Richard

More information

Light Conduction Capability of Different Light-Transmitting FRC Posts

Light Conduction Capability of Different Light-Transmitting FRC Posts Original Research Light Conduction Capability of Different Light-Transmitting FRC Posts Saied Mostafa Moazzami 1, Reza Kazemi 2, Maliheh Alami 1, Ebrahim Attaran 3, Michael Mehhary 4, Majid Sarmad 5, Heydar

More information

2011 Dental Materials Clinical Dentistry Survey

2011 Dental Materials Clinical Dentistry Survey 2011 Dental Materials Clinical Dentistry Survey Nov 2011 Dental Evaluation and Consultation Service (DECS) 2011 Federal Dental Service Dental Materials Survey Because of advances in technology, dental

More information

Degree of conversion of two lingual retainer adhesives cured with different light sources

Degree of conversion of two lingual retainer adhesives cured with different light sources European Journal of Orthodontics 27 (2005) 173 179 doi: 10.1093/ejo/cjh085 The Author 2005. Published by Oxford University Press on behalf of the European Orthodontic Society. Αll rights reserved. For

More information

Effect of Shade and Thickness of Porcelain Veneers on Depth of Cure of Light-Cured Resin-Based Luting Cement

Effect of Shade and Thickness of Porcelain Veneers on Depth of Cure of Light-Cured Resin-Based Luting Cement BALKAN JOURNAL OF STOMATOLOGY ISSN 1107-1141 TUPNBUPMPHJDBM!!TPDJFUZ Effect of Shade and Thickness of Porcelain Veneers on Depth of Cure of Light-Cured Resin-Based Luting Cement SUMMARY Objective: The

More information

Hardening of a dual-cure resin cement using QTH and LED curing units

Hardening of a dual-cure resin cement using QTH and LED curing units www.scielo.br/jaos Hardening of a dual-cure resin cement using QTH and LED curing units Maria Jacinta Moraes Coelho SANTOS 1, Sheila Pestana PASSOS 2, Monalisa Olga Lessa da ENCARNAÇÃO 3, Gildo Coelho

More information

Evaluation of Ocular Hazards from 4 Types of Curing Lights

Evaluation of Ocular Hazards from 4 Types of Curing Lights ca Applied Research Cite this article as: J Can Dent Assoc 2010;76:aXXX 2011;77:b116 Evaluation of Ocular Hazards from 4 Types of Curing Lights Daniel Labrie, PhD; Justine Moe, DDS; Richard B.T. Price,

More information

Fluorescent Properties of Porcelain-Restored Teeth and Their Discrimination

Fluorescent Properties of Porcelain-Restored Teeth and Their Discrimination Materials Transactions, Vol. 45, No. 4 (24) pp. 11 to 114 Special Issue on Frontiers of Smart Biomaterials #24 The Japan Institute of Metals Fluorescent Properties of Porcelain-Restored Teeth and Their

More information

Charisma Topaz PRODUCT INFORMATION. The three - in - one composite solution. Heraeus Kulzer Nordic AB

Charisma Topaz PRODUCT INFORMATION. The three - in - one composite solution. Heraeus Kulzer Nordic AB Charisma Topaz PRODUCT INFORMATION The three - in - one composite solution is a light-curing, radiopaque nano-hybrid composite based on the patented TCD-matrix. It offers the same clinical indications

More information

QSP Mode Characteristics of 3rd Generation ASP Powered Er:YAG Dental Lasers

QSP Mode Characteristics of 3rd Generation ASP Powered Er:YAG Dental Lasers ISSN 1855-9913 Journal of the Laser and Health Academy Vol. 2016, No.1; www.laserandhealth.com QSP Mode Characteristics of 3rd Generation ASP Powered Er:YAG Dental Lasers Nejc Lukac 1, Matjaz Lukac 2,

More information

Near-UV light detection

Near-UV light detection C L I N I C A L Near-UV light detection Javier Tapia Guadix 1 Near-UV light induced fluorescence has already proven to be very useful as an alternative to classic caries-detector dyes. However its potential

More information

Quantitative Radiometric Assessment of Neonatal Phototherapy Units used in the Treatment of Hyperbilirubinaemia

Quantitative Radiometric Assessment of Neonatal Phototherapy Units used in the Treatment of Hyperbilirubinaemia Quantitative Radiometric Assessment of Neonatal Phototherapy Units used in the Treatment of Hyperbilirubinaemia Michael Towey, Tom Cantwell* Dept. of Medical Physics and Bioengineering, Portiuncula Hospital

More information

Evaluation the Effect of Different Surface Treatment on Shear Bond Strength between Composite Increments (An in vitro study)

Evaluation the Effect of Different Surface Treatment on Shear Bond Strength between Composite Increments (An in vitro study) International Journal of Scientific and Research Publications, Volume 7, Issue 11, November 2017 676 Evaluation the Effect of Different Surface Treatment on Shear Bond Strength between Composite Increments

More information

Effect of Diode Laser 810nm in Hardness of Dental Ceramic

Effect of Diode Laser 810nm in Hardness of Dental Ceramic Optics and Photonics Journal, 2014, 4, 104-112 Published Online May 2014 in SciRes. http://www.scirp.org/journal/opj http://dx.doi.org/10.4236/opj.2014.45011 Effect of Diode Laser 810nm in Hardness of

More information

The Sof-Lex Finishing and Polishing System 3ESPE

The Sof-Lex Finishing and Polishing System 3ESPE The Sof-Lex Finishing and Polishing System 3ESPE 3M ESPE is a recognized leader in dental innovation. Introduced in the mid-1970s, the Sof-Lex Finishing and Polishing System was a revolutionary breakthrough

More information

A Posterior Composite Case Utilizing the Incremental and Stratified Layering Technique

A Posterior Composite Case Utilizing the Incremental and Stratified Layering Technique Operative Dentistry, 2006, 31-4, 512-516 Clinical Technique/Case Report A Posterior Composite Case Utilizing the Incremental and Stratified Layering Technique HH Chi Clinical Relevance This paper illustrates

More information

Curing Effectiveness of Bulkfill Composites

Curing Effectiveness of Bulkfill Composites ORIGINAL ARTICLE Curing 10.5005/jp-journals-10015-1563 Effectiveness Of Bulkfill Composites Curing Effectiveness of Bulkfill Composites 1 Ali AR Al-Shekhli, 2 Isra a Al-Aubi ABSTRACT Aim: The aim of this

More information

Up to date literature at your disposal to discover the benefits of laser dentistry

Up to date literature at your disposal to discover the benefits of laser dentistry Up to date literature at your disposal to discover the benefits of laser dentistry For further information contact info@doctor-smile.com +39-0444-349165 Dipartimento di Scienze e Tecnologie Biofisiche

More information

TRANSMITTANCE CHARACTERISTICS OF DIMETHACRYLATE RESIN BASED DENTAL COMPOSITES

TRANSMITTANCE CHARACTERISTICS OF DIMETHACRYLATE RESIN BASED DENTAL COMPOSITES Mars 2019 Vol. 8 No. 1 Postgraduate Medical Journal of Ghana TRANSMITTANCE CHARACTERISTICS OF DIMETHACRYLATE RESIN BASED DENTAL COMPOSITES Abstract Quartey-Papafio N. Biomaterials Science Department, University

More information

Influence of Curing Rate of Resin Composite on the Bond Strength to Dentin

Influence of Curing Rate of Resin Composite on the Bond Strength to Dentin Operative Dentistry, 2007, 32-2, 144-148 Influence of Curing Rate of Resin Composite on the Bond Strength to Dentin AR Benetti E Asmussen A Peutzfeldt Clinical Relevance When deciding on a specific curing

More information

Effect of light irradiation distance during intermediate. polymerization on depth of cure and hardness of. indirect composite

Effect of light irradiation distance during intermediate. polymerization on depth of cure and hardness of. indirect composite Effect of light irradiation distance during intermediate polymerization on depth of cure and hardness of indirect composite Kota Okamura Nihon University Graduate School of Dentistry, Major in Fixed Prosthodontics

More information

INTERCHANGEABLE HEAD SYSTEM. Easily and quickly change between caries diagnosis and intraoral imaging. T r i t o n H D

INTERCHANGEABLE HEAD SYSTEM. Easily and quickly change between caries diagnosis and intraoral imaging. T r i t o n H D INTERCHANGEABLE HEAD SYSTEM. Easily and quickly change between caries diagnosis and intraoral imaging. T r i t o n H D Whether its patient education, caries detection or accurate clinical documentation,

More information

Effect of pre-heating composite resin on gap formation at three different temperatures

Effect of pre-heating composite resin on gap formation at three different temperatures 1 of 6 1/23/2012 9:39 AM J Conserv Dent. 2011 Apr-Jun; 14(2): 191 195. doi: 10.4103/0972-0707.82618 PMCID: PMC3146116 Copyright Journal of Conservative Dentistry Effect of pre-heating composite resin on

More information

The effect of a plastic-wrapped LED light curing unit and curing distance variances on diametral tensile strength of composite resin

The effect of a plastic-wrapped LED light curing unit and curing distance variances on diametral tensile strength of composite resin Journal of Physics: Conference Series PAPER OPEN ACCESS The effect of a plastic-wrapped LED light curing unit and curing distance variances on diametral tensile strength of composite resin To cite this

More information

Curing efficiency of various resin-based materials polymerized through different ceramic thicknesses and curing time

Curing efficiency of various resin-based materials polymerized through different ceramic thicknesses and curing time ORIGINAL ARTICLE http://dx.doi.org/10.4047/jap.2011.3.3.126 Curing efficiency of various resin-based materials polymerized through different ceramic thicknesses and curing time Jung-Won Lee, DDS, MSD,

More information

Research Article Surface Hardness of Resin Cement Polymerized under Different Ceramic Materials

Research Article Surface Hardness of Resin Cement Polymerized under Different Ceramic Materials International Dentistry Volume 2012, Article ID 317509, 5 pages doi:10.1155/2012/317509 Research Article Surface Hardness of Resin Cement Polymerized under Different Ceramic Materials Pimmada Kesrak 1

More information

The effect of light curing units, curing time, and veneering materials on resin cement microhardness y

The effect of light curing units, curing time, and veneering materials on resin cement microhardness y Journal of Dental Sciences (2013) 8, 141e146 Available online at www.sciencedirect.com journal homepage: www.e-jds.com ORIGINAL ARTICLE The effect of light curing units, curing time, and veneering materials

More information

The Effect of Soft-start Polymerization by Second Generation LEDs on the Degree of Conversion of Resin Composite

The Effect of Soft-start Polymerization by Second Generation LEDs on the Degree of Conversion of Resin Composite Operative Dentistry, 2007, 32-2, 160-165 The Effect of Soft-start Polymerization by Second Generation LEDs on the Degree of Conversion of Resin Composite LES Soares PCS Liporoni AA Martin Clinical Relevance

More information

Comparative efficiency of plasma and halogen light sources on composite micro-hardness in different curing conditions

Comparative efficiency of plasma and halogen light sources on composite micro-hardness in different curing conditions Article Comparative efficiency of plasma and halogen light sources on composite micro-hardness in different curing conditions DIETSCHI, Didier, MARRET, N, KREJCI, Ivo Abstract Recent developments have

More information

Self-Adhesive (SA) Cement VS Dual-Cured (DC) Resin Cement

Self-Adhesive (SA) Cement VS Dual-Cured (DC) Resin Cement EsteCem II Self-Adhesive (SA) Cement VS Dual-Cured (DC) Resin Cement May-June 2017 Vol. 24, No.3 Self-adhesive resin cements are easy to use and provide low to medium bond strengths to tooth structure.

More information

how to technique Issue How to improve the longevity of porcelain veneers.

how to technique Issue How to improve the longevity of porcelain veneers. Issue 4 2013 How to improve the longevity of porcelain veneers. 1. Avoid finishing on a resin composite restoration. longer term weak spot Although tempting in the interests of minimal tooth removal, finishing

More information

dental fillings facts About the brochure:

dental fillings facts About the brochure: dental fillings facts About the brochure: Your dentist is dedicated to protecting and improving oral health while providing safe dental treatment. This fact sheet provides information you need to discuss

More information

Effect of Reduced Exposure Times on the Microhardness of Nanocomposites Polymerized by QTH and Second-generation LED Curing Lights

Effect of Reduced Exposure Times on the Microhardness of Nanocomposites Polymerized by QTH and Second-generation LED Curing Lights Operative Dentistry, 2011, 36-1, 98-103 Effect of Reduced Exposure Times on the Microhardness of Nanocomposites Polymerized by QTH and Second-generation LED Curing Lights SM Marchan D White WA Smith V

More information

TriAuto mini Compact, Cordless Endodontic Handpiece

TriAuto mini Compact, Cordless Endodontic Handpiece TriAuto mini Compact, Cordless Endodontic Handpiece Thinking ahead. Focused on life. Lightweight with Compact Head and Body The refined, compact design of TriAuto mini s head and body is very small and

More information

Key words: Finite element analysis, composite resin, ceramic, temperature distribution, light polymerization unit.

Key words: Finite element analysis, composite resin, ceramic, temperature distribution, light polymerization unit. Application of Finite Element Method for Thermal Analysis of Esthetic Restorative Materials in Dentistry Elif Pak Tunç*, K.M.Murat Tunç ** &, O.Derya Tunç(***),Sedat Şisbot(****) (*) Istanbul University,

More information

POWER BLEACHING A SOLUTION FOR DISCOLOURED TEETH

POWER BLEACHING A SOLUTION FOR DISCOLOURED TEETH doi:10.5368/aedj.2011.3.1.3.5 POWER BLEACHING A SOLUTION FOR DISCOLOURED TEETH 1 Chandrasekhar M 2 Jaya prakash D patil 3 Ramesh T 1 Director, Professor and Head, Department of Conservative Dentistry 2

More information

EMDOLA Education Program: First Year Details. History of lasers: Optics data : The quantum nature of the light:

EMDOLA Education Program: First Year Details. History of lasers: Optics data : The quantum nature of the light: EMDOLA Education Program Module 1: Optics Module 2: Physics of lasers Module 3:Interaction laser-tissues Module 4: Laser Safety Properties of lasers Module 5:Lasers conservative dentistry Laser in caries

More information

EFFECT OF LIGHT CURING UNIT ON RESIN-MODIFIED GLASS-IONOMER CEMENTS: A MICROHARDNESS ASSESSMENT

EFFECT OF LIGHT CURING UNIT ON RESIN-MODIFIED GLASS-IONOMER CEMENTS: A MICROHARDNESS ASSESSMENT www.fob.usp.br/jaos or www.scielo.br/jaos J Appl Oral Sci. 2009;17(3):150-4 EFFECT OF LIGHT CURING UNIT ON RESIN-MODIFIED GLASS-IONOMER CEMENTS: A MICROHARDNESS ASSESSMENT Daniela Francisca Gigo CEFALY

More information

how to technique How to treat a cracked, but still inact, cusp. Disadvantages. 1 Issue Full coverage crown. >>

how to technique How to treat a cracked, but still inact, cusp. Disadvantages. 1 Issue Full coverage crown. >> Issue 37 2013 How to treat a cracked, but still inact, cusp. 1 Full coverage crown. >> Advantages. A full crown can splint all sections of the tooth together. 1 Disadvantages. 1 Encircles crown and acts

More information

Quantifying Light Energy Delivered to a Class I Restoration

Quantifying Light Energy Delivered to a Class I Restoration Applied R e s e a r c h Quantifying Light Energy Delivered to a Class I Restoration Richard B.T. Price, BDS, DDS, MS, PhD, FDS RCS (Edin), FRCD(C); Melanie E. McLeod, BSc, DDS; Christopher M. Felix, BSc

More information

Comparison of Composite Curing Parameters: Effects of Light Source and Curing Mode on Conversion, Temperature Rise and Polymerization Shrinkage

Comparison of Composite Curing Parameters: Effects of Light Source and Curing Mode on Conversion, Temperature Rise and Polymerization Shrinkage Operative Dentistry, 2006, 31-2, 219-226 Comparison of Composite Curing Parameters: Effects of Light Source and Curing Mode on Conversion, Temperature Rise and Polymerization Shrinkage Z Tarle A Knezevic

More information

A real leader takes you further.

A real leader takes you further. A real leader takes you further. 2017 Dentsply Sirona. All rights reserved. The unmatched SDR bulk fill technology In 2009, SureFil SDR flow with SDR technology was the first product that allowed 4mm bulk

More information

Semiconductor Diode Lasers in Periodontics : An Innovative and Safe

Semiconductor Diode Lasers in Periodontics : An Innovative and Safe Semiconductor Diode Lasers in Periodontics : An Innovative and Safe Dr. Sushma S. Lagdive 1, Dr. Pramod P. Marawar 2, Dr. Amit M. Mani 3 1 Senior Lecturer, Dept. of Periodontics, Rural Dental College,

More information

Glass Ionomers. Reputable, Durable, Long Lasting

Glass Ionomers. Reputable, Durable, Long Lasting Glass Ionomers Reputable, Durable, Long Lasting Conventional Glass Ionomer Restoratives Glass Ionomeres Technology more up-to-date than ever before! Successfull well proven material Exhibit biocompatibility

More information

CAD/CAM PREPARATION GUIDELINES & TISSUE MANAGEMENT TECHNIQUES RECOMMENDATIONS FOR OPTIMAL SCANNING, DESIGNING, AND MILLING

CAD/CAM PREPARATION GUIDELINES & TISSUE MANAGEMENT TECHNIQUES RECOMMENDATIONS FOR OPTIMAL SCANNING, DESIGNING, AND MILLING CAD/CAM PREPARATION GUIDELINES & TISSUE MANAGEMENT TECHNIQUES RECOMMENDATIONS FOR OPTIMAL SCANNING, DESIGNING, AND MILLING CROWN PREPARATION GUIDELINES IDEAL CROWN PREPARATIONS POSTERIOR RESTORATIONS Rounded

More information

Dr. Mohammed T Omar Associate professor CAMS Mobile: Office: 2074

Dr. Mohammed T Omar Associate professor CAMS  Mobile: Office: 2074 INFRARED Therapy Dr. Mohammed T Omar Associate professor CAMS E-mail-momarar@ksu.edi.sa Dr.taher-M@yahoo.com Mobile: 0542115404 Office: 2074 Objectives Following completion of this lecture the student

More information

Influence of different light curing units on the bond strength of indirect resin composite restorations

Influence of different light curing units on the bond strength of indirect resin composite restorations Restorative Dentistry Restorative Dentistry Influence of different light curing units on the bond strength of indirect resin composite restorations Veridiana Camilotti (a) Patricia Grau Grullón (a) Márcio

More information

EXPOSURE SCHEDULES FOR SUNTANNING PRODUCTS

EXPOSURE SCHEDULES FOR SUNTANNING PRODUCTS EXPOSURE SCHEDULES FOR SUNTANNING PRODUCTS It is generally accepted that proper use of indoor tanning methods may be safer for the tanner than the uncontrolled environment of outdoor tanning. However,

More information

Adper Easy Bond. Self-Etch Adhesive. Technical Product Profile

Adper Easy Bond. Self-Etch Adhesive. Technical Product Profile Adper Easy Bond Self-Etch Adhesive Technical Product Profile Table of Contents Table of Contents Introduction... 4 Product Description... 4 Composition...5-8 Background... 5 Mechanism of Adhesion to Enamel

More information

IvoBase Material. The innovative denture base material. Tissue friendly, tough, predosed

IvoBase Material. The innovative denture base material. Tissue friendly, tough, predosed Material The innovative denture base material Tissue friendly, tough, predosed Denture base material of The innovative denture base material Material The fabrication of every quality prosthetic appliance

More information

What are the advantages and disadvantages of an inlay or onlay?

What are the advantages and disadvantages of an inlay or onlay? Problems Solvers 50 Inlays, Onlays and ¾ crowns Oh My! Synonyms: Inlays, Onlays, ¾ Crowns, 7/8 crowns When people have a cavity or a large filling there are times when a full crown is not the best choice

More information

Diffuse reflectance ultraviolet spectroscopic studies of paper

Diffuse reflectance ultraviolet spectroscopic studies of paper Diffuse reflectance ultraviolet spectroscopic studies of paper Application Note Author M. Archard Agilent Technologies, Inc. Mulgrave, Victoria 3170, Australia A. J. Michell CSIRO Division of Forestry

More information

Downloaded from journal.qums.ac.ir at 11: on Thursday June 7th 2018

Downloaded from journal.qums.ac.ir at 11: on Thursday June 7th 2018 23 JQUMS, Vol.13, No.3, Autumn 2009 Effect of flowable composite on Microleakage of packable resin composites in class II cavities AR Daneshkazemi* AR Davari** J Modaresi*** F Dastjerdi**** M Darezereshki*****

More information

A real leader takes you further.

A real leader takes you further. A real leader takes you further. The unmatched SDR bulk fill technology In 2009, SureFil SDR flow material with SDR technology was the first product that allowed 4mm bulk placement in flowable consistency.

More information

Shear bond strength of dentin bonding agents cured with a plasma arc curing light

Shear bond strength of dentin bonding agents cured with a plasma arc curing light Shear bond strength of dentin bonding agents cured with a plasma arc curing light Youngchul Kwon, Sun-Young Kim, Sae-Joon Chung, Young-Chul Han, In-Bog Lee, Ho-Hyun Son, Chung-Moon Um, Byeong-Hoon Cho*

More information

A Study to Analyze the Paramount Way of Shade Selection among Restorative Dentists in South Canara District, Karnataka

A Study to Analyze the Paramount Way of Shade Selection among Restorative Dentists in South Canara District, Karnataka Original Article Nitte University Journal of Health Science A Study to Analyze the Paramount Way of Shade Selection among Restorative Dentists in South Canara District, Karnataka 1 2 Avni Jain, Vinaya

More information

Effect of curing mode on the hardness of dual-cured composite resin core build-up materials

Effect of curing mode on the hardness of dual-cured composite resin core build-up materials Restorative Dentistry Restorative Dentistry Effect of curing mode on the hardness of dual-cured composite resin core build-up materials César Augusto Galvão Arrais (a) Aline De Cerqueira Kasaz (b) Luís

More information

ÅngströmBond Advanced Polymers for High-Tech Applications

ÅngströmBond Advanced Polymers for High-Tech Applications Epoxy & Adhesives Advanced Polymers for High-Tech Applications Choosing the correct adhesive for an application is critical and premium quality adhesives offer a wide range of solutions for any requirement.

More information

Effect of cross infection control barriers used on the light-curing device tips on the cure depth of a resin composite

Effect of cross infection control barriers used on the light-curing device tips on the cure depth of a resin composite Original Article Effect of cross infection control barriers used on the light-curing device tips on the cure depth of a resin composite Isabel Cristina Celerino de Moraes Porto 1,2, Andréia Cristina Ramos

More information

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah ULTRASOUND IMAGING EE 472 F2018 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Diagnostic Ultrasound: Physics and Equipment, 2nd ed., by Peter R. Hoskins (Editor), Kevin Martin (Editor),

More information

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA DESCRIPTION GAFCHROMIC MD-55 radiochromic dosimetry film is designed for the measurement

More information

Shear Bond Strength of Orthodontic Brackets Cured with Different Light Sources under Thermocycling

Shear Bond Strength of Orthodontic Brackets Cured with Different Light Sources under Thermocycling Shear Bond Strength of Orthodontic Brackets Cured with Different Light Sources under Thermocycling Oral Sokucu a Seyda Herguner Siso b Firat Ozturk c Ruhi Nalcaci d ABSTRACT Objectives: The aim of this

More information

Development of Low-Cost Photodynamic Therapy Device

Development of Low-Cost Photodynamic Therapy Device Vol. 112 (2007) ACTA PHYSICA POLONICA A No. 5 Proceedings of the International School and Conference on Optics and Optical Materials, ISCOM07, Belgrade, Serbia, September 3 7, 2007 Development of Low-Cost

More information

The all-rounder for hard tissue and soft tissue The new elexxion delos 4.0

The all-rounder for hard tissue and soft tissue The new elexxion delos 4.0 elexxion delos 4.0 The all-rounder for hard tissue and soft tissue The new elexxion delos 4.0 World premier: 20-watt diode-pumped Er:YAG laser together with digital-pulsed 50-watt diode laser Diode 20.000

More information

Stress Wave Focusing Transducers

Stress Wave Focusing Transducers UCRL-K-130697 PREPRINT Stress Wave Focusing Transducers Steven R. Visuri, Richard A. London, Luiz Da Silva This paper was prepared for submittal to Optical Society of America, Spring Topical Meetings Orlando,

More information

Behaviour of general dental practitioners in Germany regarding posterior restorations with flowable composites

Behaviour of general dental practitioners in Germany regarding posterior restorations with flowable composites O R I G I N A L A R T I C L E International Dental Journal 2011; 61: 252 256 doi: 10.1111/j.1875-595X.2011.00068.x Behaviour of general dental practitioners in Germany regarding posterior restorations

More information

Evaluation of X-ray protective shielding used in dental offices in Kerman, Iran, in 2014

Evaluation of X-ray protective shielding used in dental offices in Kerman, Iran, in 2014 Received: 17 May 2016 Accepted: 3 Dec. 2016 Evaluation of X-ray protective shielding used in dental offices in Kerman, Iran, in 2014 Abstract Jahangir Haghani DDS, MSc 1, Maryam Raoof DDS, MSc 2, Maryam

More information

The Effect of Shade and Curing Time on Depth of Cure (DOC) in Two Types of Composites, Polymerized with a Halogen Light Cure System

The Effect of Shade and Curing Time on Depth of Cure (DOC) in Two Types of Composites, Polymerized with a Halogen Light Cure System The Effect of Shade and Curing Time on Depth of Cure (DOC) in Two Types of Composites, Polymerized with a Halogen Light Cure System Sa'di Shirshab Thiab, Qasim AbduI-Kareem Mohammad, Ahmed Ghanim Mahdi

More information

IN VITRO STUDY OF LIGHT RADIATION PENETRATION THROUGH DENTIN, ACCORDING TO THE WAVELENGTH

IN VITRO STUDY OF LIGHT RADIATION PENETRATION THROUGH DENTIN, ACCORDING TO THE WAVELENGTH 10.1515/amb-2015-0013 IN VITRO STUDY OF LIGHT RADIATION PENETRATION THROUGH DENTIN, ACCORDING TO THE WAVELENGTH V. Dogandzhiyska 1, I. Angelov 2, Sl. Dimitrov 3 and Ts. Uzunov 1 1 Department of Conservative

More information

J Lasers Med Sci 2016 Summer;7(3):

J Lasers Med Sci 2016 Summer;7(3): Journal of Lasers in Medical Sciences J Lasers Med Sci 2016 Summer;7(3):184-191 http://www.journals.sbmu.ac.ir/jlms Original Article doi 10.15171/jlms.2016.32 Comparison of the Amount of Temperature Rise

More information

Superior Fluorescent Labeling Dyes Spanning the Full Visible Spectrum...1. Trademarks: HiLyte Fluor (AnaSpec, Inc.)

Superior Fluorescent Labeling Dyes Spanning the Full Visible Spectrum...1. Trademarks: HiLyte Fluor (AnaSpec, Inc.) Table of Contents Fluor TM Labeling Dyes Superior Fluorescent Labeling Dyes Spanning the Full Visible Spectrum....1 Fluor TM 405 Dye, an Excellent Alternative to Alexa Fluor 405 & DyLight 405....2 Fluor

More information

Effects of Different Curing Lights on Microleakage in Resin- Based Composite Restorations

Effects of Different Curing Lights on Microleakage in Resin- Based Composite Restorations ORIGINAL RESEARCH Effects of Different Curing Lights on Microleakage in Resin- Based Composite Restorations Deeksha Kiran Shetty 1, Sheetal Mali 2, Sarika Shetty 3, Neha Patil 4, Agraja Patil 5, Ankita

More information

Effect of Time and Polymerization Cycle on the Degree of Conversion of a Resin Composite

Effect of Time and Polymerization Cycle on the Degree of Conversion of a Resin Composite Operative Dentistry, 2006, 31-4, 489-495 Effect of Time and Polymerization Cycle on the Degree of Conversion of a Resin Composite LFJ Schneider S Consani F Ogliari AB Correr LC Sobrinho MAC Sinhoreti Clinical

More information

EFFECT OF CURING LIGHT DISTANCE ON MICROLEAKAGE AND COMPRESSIVE STRENGTH OF PEDIATRIC COMPOSITE RESTORATIONS

EFFECT OF CURING LIGHT DISTANCE ON MICROLEAKAGE AND COMPRESSIVE STRENGTH OF PEDIATRIC COMPOSITE RESTORATIONS Indian J.L.Sci. 4 (1) : 25-29, 2014 EFFECT OF CURING LIGHT DISTANCE ON MICROLEAKAGE AND COMPRESSIVE STRENGTH OF PEDIATRIC COMPOSITE RESTORATIONS a b c1 A. R. PRABHAKAR, CHANDRASHEKHARYAVAGAL AND VAIBHAV

More information

Polymerization of dual cured composites by different thickness

Polymerization of dual cured composites by different thickness Polymerization of dual cured composites by different thickness ABSTRACT Yun Ju Kim, Myoung Uk Jin, Sung Kyo Kim, Tae-Yub Kwon 2, Young Kyung Kim * Department of Conservative Dentistry, School of Dentistry,

More information

Evaluation of Bond Strength of Silicone and Acrylic Resin Based Resilient Denture Liners Over A Period of Storage in Water.

Evaluation of Bond Strength of Silicone and Acrylic Resin Based Resilient Denture Liners Over A Period of Storage in Water. DOI: 0.2276/aimdr.208.4..DE4 Original Article ISSN (O):2395-2822; ISSN (P):2395-284 Evaluation of Bond Strength of Silicone and Acrylic Resin Based Resilient Denture Liners Over A Period of Storage in

More information

Handling efficiency of autopolymerized resin applied using the brush-on technique

Handling efficiency of autopolymerized resin applied using the brush-on technique Handling efficiency of autopolymerized resin applied using the brush-on technique Yasunori Suzuki, DMD, PhD, Hikari Chiba, DMD, PhD, Natsuko Kamada, DMD, PhD, Daisuke Kurihara, DMD, PhD, Yukari Kanki,

More information

Effect of Water on HEMA Conversion by FT-IR Spectroscopy

Effect of Water on HEMA Conversion by FT-IR Spectroscopy Original Article Effect of Water on HEMA Conversion by FT-IR Spectroscopy TS. Jafarzadeh Kashi 1, 2, 3, M. Erfan 3, DC. Watts 4 1 Assistant Professor, Department of Dental Materials, School of Dentistry,

More information

Semi-Direct Composite Restorations- A Clinical Report

Semi-Direct Composite Restorations- A Clinical Report MALAYSIAN DENTAL JOURNAL Semi-Direct Composite Restorations- A Clinical Report Berhanuddin Muhamad Asyraf, Tong Wah Lim Faculty of Dentistry, University Teknologi MARA, Malaysia ABSTRACT In nowadays world,

More information

Acknowledgments Introduction p. 1 Objectives p. 1 Goals p. 2 History of Dental Materials p. 3 The Oral Environment p. 4 Characteristics of the Ideal

Acknowledgments Introduction p. 1 Objectives p. 1 Goals p. 2 History of Dental Materials p. 3 The Oral Environment p. 4 Characteristics of the Ideal Preface p. v Acknowledgments p. vii Introduction p. 1 Objectives p. 1 Goals p. 2 History of Dental Materials p. 3 The Oral Environment p. 4 Characteristics of the Ideal Dental Material p. 5 Quality Assurance

More information