CoaguChek XS Plus System Whole Blood Protime/INR
|
|
- Elmer Shaw
- 5 years ago
- Views:
Transcription
1 1
2 CoaguChek XS Plus System Whole Blood Protime/INR A. PURPOSE The CoaguChek XS System is intended for use by professional healthcare providers for quantitative protime testing (PT) and calculated INR for monitoring Warfarin therapy. Results are considered definitive for patient care and diagnosis. This CLIA waived system for point of care testing allows for immediate results and subsequent timely patient education throughout the Bassett Healthcare Network. B. PRINCIPLE OF OPERATION The CoaguChek XS Plus PT Test will provide an electrochemical measurement of prothrombin time following activation of blood coagulation with human recombinant thromboplastin. Blood interacts with the chemicals in the test strip to make a small electric current in the test strip that measures blood-clotting time. The INR is calculated based on ISI of the thromboplastin in the test strip. C. SPECIMEN REQUIREMENTS 1. Sample type: CoaguChek XS Plus PT tests may be performed with fresh capillary whole blood from a fingerstick or fresh venous whole blood drawn in a nonanticoagulant plastic syringe. Sample Size: Minimum sample size is 20 µl of blood. 2. Procedure for Capillary Sample Collection Seat the patient comfortably a) Warm the hand with a hand warmer or warm water. b) Have the patient let their arm hang down by their side. c) Massage the finger from its base. Clean the selected area with rubbing alcohol. Allow the puncture site to air dry. Prepare lancet (depth adjust if needed). Prepare the Capillary Blood Collection System if needed by firmly inserting the capillary tube into the bulb. Lance the fingertip. Use the side of the fingertip and massage the lanced finger until a drop of blood is formed. Gently squeeze the finger until a ladybug size of hanging blood forms. If using the Capillary Blood Collection System, collect the sample into the capillary tube by holding the tube horizontal to the drop. Fill tube with 2-3 drops. Cover the opening of the bulb. Gently squeeze the bulb with your hand to dispense the sample on the test strip. 3. Procedure for venous sample collection: For sample collection use a standard blood collection device. Do not use anti-coagulants (e.g. EDTA, citrate, fluoride, oxalate, or heparin) to collect the blood sample. Venous blood samples may be collected from a venous line, after an appropriate line volume is discarded. Collect into a nonanticoagulated syringe, 2
3 If venous samples are collected by venipuncture, note the following: a) Use a needle at least 23 G (approx mm) or larger in size. b) Discard the first four drops of blood collected (within the first 10 seconds). Then immediately apply one drop of blood (at least 8 µl) directly onto the target area of the test strip. Ensure that no air bubbles are introduced into the sample. Unacceptable Specimens Plasma or serum CANNOT be used as a testing sample. Sample size CANNOT be less than 10 µl Venous sample CANNOT be collected in a syringe containing anticoagulant, Sample must be USED IMMEDIATELY after collection. Glass tubes or glass syringes CANNOT be used (only use plastic). D. REAGENTS/MATERIALS/INSTRUMENTATION 1. Materials a) Roche Diagnostics CoaguChek XS Plus meter. Store and use meter between 15 and 32 C and 10 to 85% humidity. 3
4 b) CoaguChek XS test strips and code chip (Cat. No ) - Warehouse # c) CoaguChek XS Pro PT Controls (Cat No ). Store controls at 2 to 8 C until the expiration date. DO NOT FREEZE. Obtain from Point of Care. d) CoaguChek XS Systems Capillary Blood Collection Systems (Cat No. # ). Obtain from Point of Care. e) AccuChek safe T-Pro plus lancet single use disposable (Roche) Warehouse # f) Alcohol swab g) Gauze or cotton ball h) Clorox germicidal wipes EPA# req no i) 70% alcohol i) Scissors 2. Storage and Handling a) CoaguChek PT test strips Store CoaguChek XS test strips in their original vial tightly capped at room temperature (18-32 C) until ready to use. b) When stored at the recommended room temperature, test strips are stable until the expiration date that is printed on the vial. c) The test strip must be used within 10 minutes once removed from the vial. Recap vial immediately after opening. d) Do not handle CoaguChek XS PT test strips with damp or moist hands. E. QUALITY CONTROL REQUIREMENTS Built in Quality Control A check of the electronic components and functions every time the meter is turned on. A check of the test strip temperature while a test is in progress. A check of the expiration date and lot information on the test strip carried out by the code chip. A two-level, on-board quality control test and patient result determination within a single test chamber. Liquid Quality Control Level 1 and Level 2 CoaguChek XS Pro PT Controls. Frequency Level 1 and Level 2 with each new lot reagent strips prior to patient testing. Level 1 and Level 2 each month for each lot number of reagent strips in use. QC is rotated among operators and locations. F. PROCEDURE 1. Quality Control a) Insert the Quality Control code chip into the meter. This tells the meter the acceptable ranges for this box of controls. b) Remove the QC vials from refrigeration, they can be reconstituted immediately. Remove the screw-cap and rubber stopper from the quality control bottle. Label the bottle with the date and time that you reconstitute it. 4
5 c) Using scissors cut off the tip of the dropper at the end of the stem. Hold the dropper a safe distance from your face. CAUTION: To avoid loss of diluent, hold the dropper by the stem; do not squeeze the bulb of the dropper while cutting the tip. d) Invert the dropper and place the tip into the bottle. Be careful to avoid losing any diluent. e) Gently squeeze the bulb to dispense all of the contents of the dropper over the dried material. Do not allow the dropper to touch the dried material. IMPORTANT: Make sure you dispense ALL the diluent. f) Remove the dropper from the bottle. Do not discard the dropper. Replace the cap first and gently swirl the bottle to dissolve the dried material. Do not shake or invert the bottle. Let the bottle sit undisturbed for at least a minute to ensure that the dried material dissolves completely. Use the reconstituted quality control solution within 30 minutes from the time the diluent is added. 2. Performing the Test a) When you are ready to test, remove 1 strip from the container and immediately close the container. Make sure it seals tightly. IMPORTANT: Do not open a container of test strips or touch a test strip with wet hands or wet gloves. This may damage the test strips. You must use the test strip within 10 minutes of removing it from the container. Otherwise, you may get an error message and you will have to repeat the test. b) Insert test strip as far as you can into the meter. This powers the meter ON. c) Press QC TEST. The meter warms the test strip for about 30 seconds. Then the meter begins a countdown. You have 180 seconds to apply a liquid quality control solution to the test strip. d) The meter automatically checks to see if you have the right test strip code chip. The three-digit code on the test strip container must match the number on the test strip code chip before the test can be run. If you are using test strips from a new, unopened container, you will need to change the test strip code chip. e) When the meter is ready for the sample, gently swirl the control bottle once or twice to mix the control solution. Do not mix the solution with the dropper. Draw control solution into the dropper and put one drop of the liquid on top of the target area (clear area of the test strip). Do not add more control. f) Do not touch or remove the test strip while the test is in progress. The flashing dropper symbol changes to an hourglass symbol when the meter detects sufficient sample. If the meter's beeper is enabled, a beep sounds as well. If any control remains in the dropper after you dose the test strip, return the remaining control material to the control bottle. Save extra control until after the test result is obtained, in case the control test needs to be repeated. The result appears in approximately one minute. g) Record the result on the CoaguChek XS Plus QC Log. After you verify the validity of the control result, discard the test strip, dropper and the reconstituted bottle of quality control solution. h) Remove the Quality Control code chip and store it with the opened box of controls. Re-insert the test strip code chip. i) Power the meter OFF. 5
6 j) Retrieve QC test information by selecting REVIEW RESULTS, QC RESULT. The meter stores 500 patient results, the most recent is displayed first. NOTE: If you need to repeat a test, use a new test strip. Expected results The meter displays the acceptable control range and the result. The reading is automatically saved in the memory of the meter. The system is working properly and all handling has been done correctly when the test results obtained are within the acceptable control range. Unacceptable results Repeat the sample within 30 minutes of reconstitution. Contact Point of Care for technical assistance. An out-of-range result is indicated by an arrow. An arrow pointing up means the result is too high. An arrow pointing down means the result is too low. Coding the Meter: A code chip is included with each new container of test strips. Use the code chip only with the tests in that container. Each code chip is specific to a particular lot of test strips. The code chip contains programming information about the test type, lot number, expiration date, and function curve. The code chip must not be exposed to moisture or devices that produce magnetic fields. A new code chip is inserted into the CoaguChek XS Plus monitor each time a new vial of strips is opened. The code chip remains in the monitor until the lot of test strips in use is changed. The old code chip is then discarded. The meter can store information for 60 code chips. Turn off monitor before inserting or removing code chip. Remove old code chip, if one is installed, and discard. Insert new code chip with code number facing up until it snaps into place. Turn monitor on. New information is then programmed into the monitor from the code chip. Meter is ready for patient testing. Patient Test Check the meter for the correct lot code chip. Place meter on a flat surface. Turn the meter on by pressing the On/Off Button and holding until the meter turns on. Check the battery level. If there are no bars left in the battery symbol, you cannot perform anymore tests. Check that the date and time are correct. Take a test strip out of the container. Close the container tightly. Do not touch a test strip with wet hands or gloves. This may damage the test strips. Use the test strips within 10 minutes. Enter operator ID (operator bar code number if available). Touch ( ) OK to log on. Select PATIENT TEST. Enter Patient ID, using the visit number. The test strip symbol prompts you to insert a test strip. Hold the test strip so the lettering CoaguChek XS PT is facing upward. 6
7 Slide the test strip into the test strip guide in the direction indicated by the arrows. Slide the test strip in as far as it will go. Listen for a beep. An hourglass symbol shows that the test strip is pre-warming. At the completion of the pre-warming period you hear a beep tone. You will see an icon of a falling drop of blood and a 180 second countdown begins. DO NOT puncture finger until the flashing test strip and blood drop symbols appear on the meter screen. Perform the finger puncture, collect the blood using one of the methods as described in section C-Specimen Requirements. Apply the first drop of blood (8 µl) to the semicircular, transparent sample application area of the test strip. Dose the meter using top dosing or side dosing. Dose the meter within 15 seconds of the puncture. Hold the blood drop to the test strip until you hear a beep. The flashing blood drop symbol disappears. The sample must be delivered to the test strip all at once. Deliver the sample in a single drop for the instrument to test accurately without errors. The meter performs a two-level on-board quality control test on the test strip before it displays the test result. QC appears in the display. Following a successful outcome of the quality control test, a check mark appears after QC. The PT/INR results are displayed. The results are displayed in less than one minute. Record results in EPIC/Hemoglobin and PT/INR Patient Result Log or the downtime log. After the test and results are displayed touch the comment icon. It is recommended to document that the meter has been cleaned per policy after each patient use. To do so, select CLEANED METER comment. Select the MENU icon until the REMOVE menu will appear on the screen. Remove the test strip from the measurement chamber. When you are done testing, turn the meter off or allow it to time out (5 minutes). Clean the exterior of the meter with Clorox Germicidal wipes after each use. See cleaning and maintenance section J. Retrieve patient test information by selecting REVIEW RESULTS, PATIENT RESULT. The meter stores 2,000 patient results, the most recent is displayed first. G. REFERENCE RANGE CoaguChek XS Plus test results are displayed as PT in seconds and INR. The physician determines the best INR level for the patient depending on the reason for anticoagulant treatment and each individual s response to treatment (based on INR). Each physician should establish expected values for his or her patient. Normal Range INR PT seconds Therapeutic Range INR INR Routine Oral Anticoagulation Mechanical Valves 7
8 H. MEASURING RANGE Reportable Range INR PT seconds Results that are above or below the measuring range are indicated by the symbols > (above) or < (below). I. CRITICAL VALUE RANGE Critical Value INR >4.4 J. RESULT DOCUMENTATION - Test Order Code POC 115 a) PT/INR results are documented in the enter/edit function of Unity (EPIC). Report the PT seconds and the INR. Document the internal QC as acceptable. Complete all required fields with the RED STOP sign. b) Complete additional optional fields listed below with patient s responses. Some fields are free text and others have default answers. Warfarin/Coumadin/Jantoven Heparin/Low Molecular Weight Heparin Direct Acting Oral Anticoagulant Medication in dose mg Medication Frequency Bleeding Bruising Bleeding/Bruising Location New Medication Prescriptions/Instructions Missed Doses in Past Week 8
9 a,b. c. c) Inhibitor Comment Append POC INR result inhibitor comment under narrative tab for ALL results. d) INR results that are greater than 6.0 are reported in the comment section. Report PT as Cannot report seconds when INR > 6.0. Report INR as > 6.0. d. 9
10 e. e) Critical Values For INR results greater than 4.4, append your critical value called documentation under the narrative tab. Search for canned comment by typing POC in the smart text field. Select POC critical value. The following statement will appear (Provider *** notified of critical value and read back on *** at *** by ***). Use the F2 key to advance to each of the *** spaces and enter the appropriate information. f) Confirmation Samples INR values that are greater than 6.0 must be confirmed at the appropriate Bassett Network Lab (MIB, CRH, OCH, and TRH). A venous sample is collected for PT/INR. A Print Screen copy of the Unity documented 10
11 results is sent with the sample (see example). A provider order in Unity is not required. g) Downtime In the event of Unity or a network downtime, testing is ordered and results are documented on the Warfarin/anticoagulant record (H7349). This record is scanned into Unity for a permanent report. K. LIMITATIONS 1. Test result accompanied by C. This usually means one of three things: The patient has a hematocrit value right around or lower than the CoaguChek XS Plus system lower limit allowance of 25%. The patient s hands were not completely dry before the blood sample was taken. The patient s finger was excessively squeezed to obtain the sample, diluting the blood and resulting in a low hematocrit. Retest on a different finger, ensuring the patient s hands are both completely dry and the finger is not excessively squeezed after the finger stick. If you still get a C with the test result, we recommend you perform a hematocrit check on that patient and send a venous specimen to the lab to check the patient s PT/INR value, if you confirm their hematocrit is less than 25%. 2. Hematocrit ranges between 25-55% do not significantly affect test results. 3. Results with blood samples (triglycerides) indicated no significant effect on test results: Bilirubin up to 30 mg/dl Lipemic samples containing up to 500 mg/dl of triglycerides Hemolysis up to 1000 mg/dl Heparin concentrations up to 0.8 U/mL 11
12 Low molecular weight hepartins (LMWH) up to 2 IU anti-factor Xa activity/ml Clopidrogrel up to 20 mg/dl Fondaparinux up to 5 mg/l 4. The CoaguChek XS Plus System should not be used for patients being treated with any direct thrombin inhibitors including Hirudin, Lepirudin, Bivalirudin and Argatroban. 5. The presence of anti-phospholipid antibodies (APAs) such as Lupus antibodies (LA) can potentially lead to prolonged clotting times, i.e., elevated INR values. L. MAINTENANCE/CLEANING Document cleaning on the Point of Care Maintenance Record. 1. Cleaning/disinfecting the exterior Ensure that the blue test strip guide cover remains tightly closed while cleaning the meter housing. With the meter powered off, wipe the meter s exterior clean using Clorox Germicidal Wipes. Do not let liquid accumulate near any opening. Ensure that no liquid enters the meter. With a fresh dry cloth or lint free tissue, wipe away residual moisture and fluids after cleaning the housing. Allow wiped area to dry for at least 10 minutes before performing a test. 2. Cleaning/disinfecting the test strip guide - Recommended cleaning schedule: once per box of strips or as needed. Remove the test strip guide cover to clean it. (Use your thumbnail to open the cover of the test strip guide by pressing its front edge upward,) Move the cover safely away from the meter. Then rinse the cover with warm water or wipe it clean using the solutions recommended. Let the test strip guide cover dry for at least 10 minutes before reattaching it. Hold the meter upright with the test strip guide facing down. Clean the easily accessible white areas with a moistened cotton swab. Ensure the swab is only damp, not wet. Wipe away residual moisture and fluids. 3. Installing or replacing batteries page Operator s Manual M. TROUBLESHOOTING See pages Operator s Manual, call Point of Care x6728 or Technical Support N. REFERENCES 1. Roche Diagnostics Corporation Policy and Procedure 2. CoaguChek XS System Package inserts for Test Strips. 3. MIBH Laboratory Hematology Policy and Procedure for coagulation. 4. CoaguChek XS Plus Operator Manual, Roche. 12
13 13
Applicable To NOTE: CAPILLARY PUNCTURE-DO NOT WIPE AWAY THE FIRST DROP OF BLOOD.
Subject PT-INR Whole Blood using CoaguChek XS Plus or Pro Test System Index Number Lab-1017 Section Laboratory Subsection Point of Care Testing Category Departmental Contact Rachel Blum Last Revised 9/26/2017
More informationThis is a CLIA-waived system. Policies and Procedures Manual
This is a CLIA-waived system. Policies and Procedures Manual Approval date Approved by Reviewed/Revised: Date By whom Title Roche Diagnostics 9115 Hague Road Indianapolis, IN 46256 www.poc.roche.com COAGUCHEK
More informationCoaguChek S System. Quick Reference Guide. This is a CLIA-waived System. Coagulation Testing with Fresh Whole Blood
CoaguChek S System This is a CLIA-waived System Quick Reference Guide Coagulation Testing with Fresh Whole Blood May not print or view at 100% All unmarked prints black. Coag Swoosh prints gradients of
More informationRoche CoaguChek XS Plus INR
Page 1 of 8 Roche CoaguChek XS Plus INR Edition No: 6.0 Author: Paul Simpson Operative Date: October 2016 Review Date: October 2017 Location: Country Health SA Integrated Cardiovascular Clinical Network
More informationAlere INRatio 2 METHOD AND SAMPLE COLLECTION
Alere INRatio 2 METHOD AND SAMPLE COLLECTION 1. PURPOSE AND SCOPE The purpose of this document is to describe the procedure for performing an INR test using the Alere INRatio2 analyser. The INRatio2 meter
More informationAccu-Chek Inform II: Point of Care Glucose Testing. Sharp Healthcare 2014
Accu-Chek Inform II: Point of Care Glucose Testing Sharp Healthcare 2014 OBJECTIVES At the completion of this module the participant will be able to: Learn the proper technique of performing a finger stick
More informationItems in the package:
Intended Use: The EasyLife Hb Monitoring System is designed for in vitro diagnostic use only (external use only), and is suitable for self-testing. The system is for healthcare professionals and persons
More informationSystem. User Manual for Self-Testing
CoaguChek System XS User Manual for Self-Testing Table of Contents CoaguChek XS System User Manual For investigational use only. Not for use in diagnostic procedures. Aspects of the CoaguChek XS System
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationPROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:
PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:
More informationUser Manual. This is a CLIA waived system.
User Manual This is a CLIA waived system. The contents of this manual, including all graphics and photographs, are the property of Roche Diagnostics. No part of this document may be reproduced or transmitted
More informationPROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.
Department Of Pathology POC.514.06- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 6 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.05 LBH 1 OF 7 Printed copies are
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationTips for Training New Customers
Tips for Training New Customers Good Habits: Test Strips Recap vial immediately and tightly after removing a single test strip. Never remove the desiccant from the vial. Do not lay test strips out ahead
More informationManagement of Central Venous Access Devices. Blood Glucose Monitoring
Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including
More informationLeadCare BLOOD LEAD ANALYZER. Quick Reference Guide
LeadCare II BLOOD LEAD ANALYZER Quick Reference Guide Precautions Precautions Caution The LeadCare II Blood Lead Analyzer is a CLIA-waived device. Facilities that perform tests with the LeadCare II System
More informationPoint of Care Testing. Bedside Glucose Meter
Point of Care Testing Bedside Glucose Meter 11/2017 Regulatory Overview Bedside glucose is a point of care lab test and YOU are performing this lab test under the Lab s CLIA license. All lab tests are
More informationPoint of Care Testing for INR using CoaguChek XS Plus
Point of Care Testing, Pathology Page 1 of 18 Point of Care Testing for INR using CoaguChek XS Plus EDITION No 1.4 DATE OF ISSUE Feb 2014 REVIEW INTERVAL AUTHOR LOCATION OF COPIES 3 YEARS D O Neill 1.
More informationCoaguChek Outpatient Clinic: CoaguChek XS - Quick Reference Guide
CoaguChek Outpatient Clinic: CoaguChek XS - Quick Reference Guide CoaguChek XS Quick Reference Guide Page 1 Page 1 of 18 Coaguchek XS Monitor A DISPLAY B - M (memory button) o To recall results from memory
More informationBlood Glucose Monitoring System. User Guide
Blood Glucose Monitoring System User Guide Table of Contents Introduction...2 Important Safety Instructions...2 About ipet PRO Blood Glucose Monitoring System...3 About ipet PRO Meter...4 About the ipet
More informationPoint of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient.
1 2 Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient. All such tests are considered lab tests and are
More informationBlood Glucose & Ketone Monitoring System
Blood Glucose & Ketone Monitoring System Self monitoring of blood glucose is an integral part of diabetes care, but the high cost of testing can make this impossible. At ACON, our goal is to provide high
More informationDear HighQ Check System Owner :
Dear HighQ Check System Owner : Thank you for purchasing the HighQ Check Blood Glucose Monitoring System. This manual provides important information to help you to use the system properly. Before using
More informationCentriVet GK Blood Glucose & Ketone Monitoring System
CentriVet GK Blood Glucose & Ketone Monitoring System FOR ANIMAL USE. NOT FOR HUMAN USE. Welcome and thank you for choosing the CentriVet GK Blood Glucose & Ketone Monitoring System. The CentriVet GK Blood
More informationAbbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory
Abbott PXP Glucose Meters 2015 Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory Objectives After you complete this Computer-Based Learning (CBL) module, you should be able to: Explain
More informationGlucose Meter. User Guide. Veterinary Monitoring System. For dog and cat use only
Glucose Meter User Guide Veterinary Monitoring System For dog and cat use only Gpet instruction Manual 31/5/09 18:06 Page 2 Gpet instruction Manual 31/5/09 18:06 Page 3 TABLE OF CONTENTS Your g-pet system
More informationTitle: Glucose Testing: StatStrip Glucose Meter Page 1 of 6
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:
More informationTitle/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised:
Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised: PURPOSE The Abbott FreeStyle Precision Pro (FSPP) Blood Glucose
More informationLogic. Owner s Guide. Blood Glucose Monitor
Owner s Guide Logic Blood Glucose Monitor TM You will notice these symbols on your vials of test strips and control solution as well as the label on the back of your BD Logic TM Blood Glucose Monitor.
More informationPoint of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter
Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter The Nova Stat Strip Blood Glucose meter is UHB s approved Glucometer
More informationFraser Health Authority Accu-chek Inform II: Training Script
1. Introduction Ensure that each participant Signs the attendance Sheet (please PRINT first & last name, and middle initial legibly) Refer to; For Reference Only: o Quick Reference Guide o What s new document
More informationInstructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial
Instructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial How do I prepare and give an injection with Enbrel multiple-dose vial? A multiple-dose vial contains
More informationMDT2. Self-Monitoring Blood Glucose System. Quick Reference Guide
MDT2 Self-Monitoring Blood Glucose System Quick Reference Guide Inserting Batteries 1. Open the battery door on the back of the meter by pushing the tab in the direction of the arrow. 2. Insert two batteries.
More informationBLOOD COLLECTION GUIDELINES
I. Patient Identification Lee Memorial Health System Lee County, FL CLINICAL LABORATORY BLOOD COLLECTION GUIDELINES A. Inpatient / Outpatient with armband 1. When possible, ask patient to state their name
More informationBlood Ketone and Glucose Monitoring System Owner s Guide For Veterinary Animal Use Only Not For Human Use
Blood Ketone and Glucose Monitoring System Owner s Guide For Veterinary Animal Use Only Not For Human Use Nova Vet Monitor OG PN52899E EN.indd 1 Nova Vet Monitor OG PN52899E EN.indd 2 Nova Vet Blood Ketone
More informationTable of Contents Section 1 Introduction... 1 Section 2 System Components... 4 Section 4 Initial Setup... 9 Section 5 Meter Setup and Options...
Table of Contents Section 1 Introduction... 1 Intended Use... 1 About Prothrombin Time (PT) Testing... 2 About International Normalized Ratio (INR) Values... 2 Test Principle... 2 Section 2 System Components...
More informationPrecision ceedpro. Online Training for. Glucometer & Capillary Collection
Online Training for Precision ceedpro Glucometer & Capillary Collection How long will it take? This module takes approximately 30 minutes to complete. What if I have to leave before I finish? Progress
More informationTo provide you with necessary knowledge and skills to accurately perform 3 HIV rapid tests and to determine HIV status.
Module 9 Performing HIV Rapid Tests Purpose To provide you with necessary knowledge and skills to accurately perform 3 HIV rapid tests and to determine HIV status. Pre-requisite Modules Module 3: Overview
More informationHow to Use ENBREL : Vial Adapter Method
How to Use ENBREL : Vial Adapter Method SETTING UP FOR AN INJECTION Select a clean, well-lit, flat working surface, such as a table. Take the ENBREL dose tray out of the refrigerator and place it on your
More informationEME. Self-Monitoring Blood Glucose System. Quick Reference Guide
EME Self-Monitoring Blood Glucose System Quick Reference Guide Inserting Batteries 1. Open the battery door on the back of the meter by pushing the tab in the direction of the arrow and pulling the door
More informationGlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System
GlucCell TM SYSTEM USER S GUIDE Ver 2.1 Intended Use The GlucCell TM Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during cell culture.
More informationUser s Manual. Blood Glucose Meter
User s Manual Blood Glucose Meter This User s Manual features the following 3 symbols: W This symbol indicates a possible risk of injury or of damage to your own health or the health of others. H This
More informationUser s Manual Blood Glucose Meter
User s Manual Blood Glucose Meter Contents Introduction...3 Chapter 1: Your New System...5 Chapter 2: Blood Glucose Tests...13 Chapter 3: Control Tests...25 Chapter 4: Meter Settings...31 Chapter 5: Review
More informationEasyMax Voice III Self-Monitoring Blood Glucose System. Quick Reference Guide
EasyMax Voice III Self-Monitoring Blood Glucose System Quick Reference Guide 1 Installing Batteries 1. Open the battery door on the back of the meter by pushing the tab in the direction of the arrow and
More informationUser manual for healthcare professionals
Blood Glucose Management System Accu-Chek Performa User manual for healthcare professionals CONTENTS 3 Introduction 4 Staff Responsibilities 5 Protocol 6 About the Accu-Chek Performa 7 Coding/Calibrating
More informationBlood glucose monitor. MediTouch 2
Blood glucose monitor for self-testing MediTouch 2 Quick Reference Guide GB Art. 79034 0483 EXACT MDD 93/42/EWG WARNING Before using the MediTouch 2 blood glucose monitor from MEDISANA for the first time,
More informationBeneCheck BK6-12M. Plus Multi-Monitoring Meter and Strips
BeneCheck BK6-12M Plus Multi-Monitoring Meter and Strips The BeneCheck BK6-12M multi-monitoring system is an easy to use, handheld device which allows you to check your Total Cholesterol, as well as Blood
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationANTIBODY SCREENING by Uni-Gold Recombigen HIV
ANTI-HIV SPECIMEN 1 REQUIREMENTS ANTIBODY SCREENING by Uni-Gold Recombigen HIV PRINCIPLE: The Uni-Gold Recombigen HIV was designed as a rapid immunoassay and is intended to detect antibodies to HIV- 1
More informationBlood Glucose Monitoring System. Owner s Booklet
Blood Glucose Monitoring System Owner s Booklet The FreeStyle Freedom Lite Meter and How It Works System Check Screen This display always appears when the meter is turned on. You should check that your
More information2. Before Testing Monitor Checker Test...10 Inserting Lancets into Lancing Device...11 Quality Control Testing...13
Table of Contents 1. About your HemoSmart Haemoglobin Screening System Contents of Kit...4 HemoSmart Haemoglobin Meter...5 HemoSmart Haemoglobin Test Strip...7 Adjustable Lancing Device and Lancets...8
More informationCurrent Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE
Current Status: Active PolicyStat ID: 4749849 Origination: 12/2013 Last Approved: 03/2018 Last Revised: 03/2018 Next Review: 03/2020 Owner: Area: References: Karla Gonzalez Anchondo: Point of Care Coordinator
More informationTable of Contents. Section 1 Introduction... 1 Section 2 Getting Started... 2 Section 3 Components... 5
Table of Contents Section 1 Introduction... 1 Section 2 Getting Started... 2 Section 3 Components... 5 Meter 5 Test Strips 8 Test Devices 8 Control Strips or Control Devices 13 Section 4 Initial Setup...
More informationPerforma BLOOD GLUCOSE METER. Operator s Manual
Performa BLOOD GLUCOSE METER Operator s Manual The Accu Chek Performa System Intended Use The Accu Chek Performa Meter is designed to be used with the Accu Chek Inform II Test Strip to quantitatively
More informationBlood Glucose Monitoring System. Owner's Booklet. Instructions for Use
Blood Glucose Monitoring System Owner's Booklet Instructions for Use Blood Glucose Monitoring System Owner's Booklet 1 Thanks for choosing OneTouch! The OneTouch Verio Blood Glucose Monitoring System
More informationDear HemoSmart GOLD Owner
Dear HemoSmart GOLD Owner Thank you for choosing the HemoSmart GOLD. All the information that will be needed to use and maintain the HemoSmart GOLD System is included in this manual. Read it carefully
More informationBlood Glucose Monitoring System
Blood Glucose Monitoring System Contents Chapter 1 OMRON Blood Glucose Monitoring System HEA-232 1.1 Introduction 1.2 List of Kit Contents 1.3 Product and Meter Display Chapter 2 Installing and Setting
More informationTHE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES
58:03 THE VALLEY HOSPITAL Ridgewood, New Jersey PATIENT CARE SERVICES (PCS) POLICIES AND PROCEDURES SUBJECT: Whole Blood Glucose PURPOSE: The Accu-chek Inform II is intended for use to quantitatively measure
More information04 ACCU-CHEK INFORM II GLUCOSE METER
Point-of-Care Testing Procedure SYS: POC-04.V2 Page 1 of 11 04 ACCU-CHEK INFORM II GLUCOSE METER SECTION: Point-of-Care Testing WRITTEN BY: Althea Reid TYPED BY: Althea Reid RECORD MANAGEMENT DISK ID:
More informationGlucCell TM SYSTEM USER S GUIDE ver 2.3 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System
GlucCell TM SYSTEM USER S GUIDE ver 2.3 Intended Use The GlucCell TM Cell Culture Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during
More informationInstructions for Use HEMLIBRA (hem-lee-bruh) (emicizumab-kxwh) injection, for subcutaneous use
Instructions for Use HEMLIBRA (hem-lee-bruh) (emicizumab-kxwh) injection, for subcutaneous use Be sure that you read, understand, and follow the Instructions for Use before injecting HEMLIBRA. Your healthcare
More informationUser Guide For in vitro Diagnostic Use in Animals Featuring the AlphaTRAK 2 Meter
Distributed by Zoetis Inc. Kalamazoo, MI 49007 BLOOD GLUCOSE MONITORING SYSTEM SYSTÈME DE SURVEILLANCE DE LA GLYCÉMIE www.alphatrakmeter.com Contact Customer Care at 1-888-963-8471 User Guide For in vitro
More informationPRIMA. Blood Glucose Monitoring System. Owner s Manual
PRIMA Blood Glucose Monitoring System Owner s Manual English Dear FORA Diamond PRIMA System Owner: Thank you for purchasing the FORA Diamond PRIMA Blood Glucose Monitoring System. This manual provides
More informationDETERMINE SPECIMEN REQUIREMENTS HIV-1/2 Ag/Ab Combo TESTING PROCEDURE
DETERMINE SPECIMEN REQUIREMENTS HIV-1/2 Ag/Ab Combo TESTING PROCEDURE PRINCIPLE: The Determine HIV-1/2 Ag/Ag Combo assay is a qualitative immunoassay for the simultaneous detection of Human Immunodeficiency
More informationBlood Glucose Monitoring System. Owner's Booklet. Instructions for Use
Blood Glucose Monitoring System Owner's Booklet Instructions for Use Blood Glucose Monitoring System Owner's Booklet 1 Thanks for choosing OneTouch! The OneTouch Verio Blood Glucose Monitoring System
More informationI can show you how to check my child s blood sugar
Our Journey with Diabetes Si usted desea esta información en español, por favor pídasela a su enfermero o doctor. I can show you how to check my child s blood sugar Testing Blood Sugar Blood sugar levels
More informationWorksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide
LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON
More informationInstructions for Use Enbrel (en-brel) (etanercept) Single-use Prefilled SureClick Autoinjector
Instructions for Use Enbrel (en-brel) (etanercept) Single-use Prefilled SureClick Autoinjector How do I prepare and give an injection with Enbrel Single-use Prefilled SureClick Autoinjector? This section
More informationPLUS One. Blood Glucose Monitoring System. Owner s Manual
PLUS One Blood Glucose Monitoring System Owner s Manual Dear PLUS One System Owner: Thank you for purchasing the PLUS One Blood Glucose Monitoring System. This manual provides important information to
More informationSummary of Significant Changes at this Revision
COPY Summary of Significant Changes at this Revision CR5890 Removal of Ketone guidelines - refer to Specialist Diabetes nurse (adult wards); Children's ward has their own protocol Remove all references
More informationW IMPORTANT INFORMATION FOR PERFORMING A BLOOD GLUCOSE TEST
W IMPORTANT INFORMATION FOR PERFORMING A BLOOD GLUCOSE TEST W An incorrectly performed blood glucose test may lead to incorrect test results which can cause the wrong therapy recommendation to be made
More informationPRIMA Lab SA Via Antonio Monti Balerna Switzerland 0041 (0) primahometest.com
PRIMA Lab SA Via Antonio Monti 7 6828 Balerna Switzerland 0041 (0) 91 605 1030 info@primahometest.com primahometest.com prima self-testing kit user manual 2in1 / 3in1 Multiparameter diagnostic device cholesterol
More informationBlood Glucose Monitoring System. Owner's Booklet. Instructions for use
Blood Glucose Monitoring System Owner's Booklet Instructions for use Verio Blood Glucose Monitoring System Owner's Booklet 1 Thanks for choosing OneTouch! The OneTouch Verio Blood Glucose Monitoring System
More informationVOICE. Blood Glucose Monitoring System. Owner s Manual
VOICE Blood Glucose Monitoring System Owner s Manual English Dear FORA Diamond VOICE System Owner: Thank you for purchasing the FORA Diamond VOICE Blood Glucose Monitoring System. This manual provides
More informationMETHOD FOR THE MEASUREMENT OF BEDSIDE TROPONIN T (cobas h232)
Page 1 of 6 METHOD FOR THE MEASUREMENT OF BEDSIDE TROPONIN T (cobas h232) Edition No: 1.3.1 Author: Rosy Tirimacco / Paul Simpson Date: April 2012 Operative Date: April 2012 Review Date: November 2012
More informationOwner s Booklet. 24/7 Customer Care English or Spanish Manufactured by:
Owner s Booklet 24/7 Customer Care English or Spanish 1-800-803-6025 www.niprodiagnostics.com Manufactured by: Fort Lauderdale, FL 33309 U.S.A. 2014 Nipro Diagnostics, Inc. TRUEtrack and the Nipro Diagnostics
More informationBlood Glucose Monitoring System 12:24 PM 10/13/12. mg/dl. Before Meal. See Options. Uses only Contour Next blood glucose test strips USER GUIDE
Blood Glucose Monitoring System 10/13/12 See Options 12:24 PM mg/dl 93 Before Meal Uses only Contour Next blood glucose test strips USER GUIDE INDICATIONS FOR USE The Contour Next blood glucose monitoring
More informationThe performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.
Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added
More informationIMPORTANT POINTS TO REMEMBER:
Policies Department: PCS-HouseWide Policy #: Page 1 of Sub Dept: Owner: Dottie Ahler Issued: 8/1/2003 Next Review: 12/3/2014 Published: This document is filed electronically on the CMC Intranet>Policy/Procedures>HealthDocX.
More informationBlood Glucose Monitoring System
Blood Glucose Monitoring System Thank you for choosing the On Call Pro Blood Glucose Monitoring System. The System will provide accurate blood glucose results in just a few simple steps. The On Call Pro
More informationTitle: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated.
COPY Summary of Significant Changes at this Revision SOP updated Purpose and Scope Items Required 1. The Abbott Optium Xceed meter is a battery-powered device designed for the measurement of blood glucose
More informationUser s Manual Blood Glucose Meter
User s Manual Blood Glucose Meter This file may not print or view at 100%. Die lines and color breaks do not print. Holding Area text, box and rules do not print. Roche USA 54082 V2/1 07790953001 07790953001(01)
More informationInstructions for Use Enbrel (en-brel) (etanercept) injection, for subcutaneous use Single-dose Prefilled Syringe
Instructions for Use Enbrel (en-brel) (etanercept) injection, for subcutaneous use Single-dose Prefilled Syringe How do I prepare and give an injection with Enbrel Single-dose Prefilled Syringe? There
More informationUniversity of Virginia Health System AccuChek Inform II Glucose Meters POC C2.21
PURPOSE Whole blood glucose testing with Accu-Chek Inform II monitors is used to monitor blood glucose concentrations for adjustments of insulin dosage in diabetic patients. It is not used for definitive
More informationUnigold Recombigen HIV 1/2 Training for HIV Testing Sites. Updated: February 2018 Cicely Richard Office of HIV/AIDS
Unigold Recombigen HIV 1/2 Training for HIV Testing Sites Updated: February 2018 Cicely Richard Office of HIV/AIDS Updated: February 2018 Cicely Richard Office of HIV/AIDS UNIGOLD RECOMBIGEN HIV 1/2 TRAINING
More informationGlucoCheck BLOOD GLUCOSE MONITORING SYSTEM OWNER'S MANUAL
GlucoCheck BLOOD GLUCOSE MONITORING SYSTEM OWNER'S MANUAL GlucoCheck BLOOD GLUCOSE MONITORING SYSTEM OWNER'S MANUAL Version 1.0 January, 2010 311-4277100-001 Dear GlucoCheck XL System Owner: Thank you
More informationBlood Glucose Monitoring System
Blood Glucose Monitoring System Self monitoring of blood glucose (SMBG) is an important part of diabetes care, but the high cost of testing can make it difficult. At ACON, our goal is to provide high quality
More informationGM700S USER S MANUAL. Blood Glucose Monitoring System
GM700S BIONIME GmbH Tramstrasse 16, 9442 Berneck Switzerland E-mail: info@bionime.ch 0197 101-3GM780-011 EN Rev. Date: September/2014 BIONIME CORPORATION No. 100, Sec. 2, Daqing St., South Dist., Taichung
More informationInstructions for Use. For use with. 10 mg vial
Instructions for Use For use with 10 mg vial Table of Contents Parts of the ZOMA-Jet 10... 1 Supplies you will need to mix a ZOMACTON 10 mg Vial... 2 Mix a ZOMACTON 10 mg vial... 3 Reset the ZOMA-Jet 10...
More informationTITLE: Whole Blood Glucose Monitoring Using the Nova StatStrip Meter
Page 4 of 14 IV. Performance of Quality Control A. Purpose: To detect any errors due to system failures or operator performance B. Frequency: Levels 1 and 3 controls are run every 24 hours of use. C. Quality
More informationPrecision Xtra Training
Precision Xtra Training Objectives At the end of this module, the participant will be able to: List the steps to calibrate the machine Describe the steps for sample collection and blood glucose testing
More informationBlood Glucose Monitoring System. Link the Effects of Food to Glucose Results. Owner s Booklet
Blood Glucose Monitoring System Link the Effects of Food to Glucose Results Owner s Booklet Welcome to the OneTouch family! We know diabetes can be difficult to understand and manage. The right products
More informationUser Manual second language
User Manual second language GlucoTel Blood Glucose Monitoring and Diabetes Management System must be used with cell phones that have: Table of contents 2 3 Introduction 4 Bluetooth Wireless Technology
More informationOwner's Booklet. Blood Glucose Monitoring System. Accurate, complete blood glucose monitoring through advanced biosensor technology
Owner's Booklet Blood Glucose Monitoring System Accurate, complete blood glucose monitoring through advanced biosensor technology i-sens, Inc. 465-6 Wolgye4-dong, Nowon-gu Seoul 139-845, Korea www.i-sens.com
More informationALPROLIX Coagulation Factor IX (Recombinant), Fc Fusion Protein INSTRUCTIONS FOR USE Do not Do not YOUR KIT CONTAINS:
ALPROLIX Coagulation Factor IX (Recombinant), Fc Fusion Protein INSTRUCTIONS FOR USE Read the Instructions for Use before you start using ALPROLIX and each time you get a refill. There may be new information.
More informationCurrent Status: Active PolicyStat ID: Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019
Current Status: Active PolicyStat ID: 3690447 Implementation: 02/2012 Last Reviewed: 06/2017 Last Revised: 05/2017 Next Review: 06/2019 Owner: Policy Area: References: Applicability: Christine Landerholm:
More informationBlood Glucose Meter Owner s Manual
89115-0055-01 Meter Owner s Manual Preface Thank you for selecting the McKesson Quintet AC Monitoring System. Please read this manual thoroughly before you start testing. This User s Manual provides all
More informationSummary of Significant Changes at this Revision. Items Required. Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
COPY Summary of Significant Changes at this Revision CR 5279 Update section 10 as per re-wording by NP Purpose and Scope 1. The Abbott Precision Xceed Pro/FreeStyle Precision Pro meter is a battery-powered
More informationRoche CoaguChek XS Plus Error Codes For Firmware Version onwards
Roche CoaguChek XS Plus Error Codes For Firmware Version 03.01.00 onwards Description of Error The instrument will not turn on -The instrument has 'frozen' or not receiving power -Remove the power to the
More informationPEN USER MANUAL Byetta 5 micrograms solution for injection in pre-filled pen (exenatide)
PEN USER MANUAL Byetta 5 micrograms solution for injection in pre-filled pen (exenatide) Section 1. WHAT YOU NEED TO KNOW ABOUT YOUR BYETTA PEN Read this section completely before you begin. Then, move
More information