Seesaw Balancing Radiation Dose and IV Contrast Dose: Evaluation of a New Abdominal CT Protocol for Reducing Age-Specific Risk

Size: px
Start display at page:

Download "Seesaw Balancing Radiation Dose and IV Contrast Dose: Evaluation of a New Abdominal CT Protocol for Reducing Age-Specific Risk"

Transcription

1 Medical Physics and Informatics Original Research CT Risk Reduction Medical Physics and Informatics Original Research Tobias Fält 1 Marcus Söderberg 2 Lisa Hörberg 1 Ingela Carlgren 1 Peter Leander 1 Fält T, Söderberg M, Hörberg L, Carlgren I, Leander P Keywords: contrast medium, CT, image quality, radiation dose DOI:1.2214/AJR Received January 8, 212; accepted after revision April 18, Department of Radiology, Skåne University Hospital, Lund University, Malmö, Södra Förstadsgatan 11, SE-25 2 Malmö, Sweden. Address correspondence to T. Fält (tobiasf@gmail.com). 2 Department of Clinical Sciences, Medical Radiation Physics Malmö, Lund University, Skåne University Hospital, Malmö, Sweden. AJR 213; 2: X/13/ American Roentgen Ray Society Seesaw Balancing Radiation Dose and IV Contrast Dose: Evaluation of a New Abdominal CT Protocol for Reducing Age-Specific Risk OBJECTIVE. The purpose of this study was to evaluate an abdominal CT protocol in which radiation dose was reduced and IV contrast dose increased for young patients and radiation dose was increased and IV medium dose decreased for elderly patients. The hypothesis was that these adjustments would result in constant image quality and a reduction in age-specific risk. MATERIALS AND METHODS. Patients were divided into four age groups of 25 patients each: group 1, years; group 2, 26 5 years; group 3, years; and group 4, older than 75 years. The quality reference tube load ranged from 1 to 3 mas, and the IV contrast dose ranged from 6 to 35 mg I/kg. Group 3 was the reference group. Signal-tonoise and contrast-to-noise ratios for a hypothetical hypovascular liver metastatic lesion were calculated. Subjective image quality was evaluated by visual grading characteristic analysis in which four readers assessed the reproduction of seven image-quality criteria. RESULTS. Radiation dose was reduced 57% in the youngest group, and the IV contrast dose was reduced 18% in elderly patients. There were no statistically significant differences between the groups with respect to signal-to-noise and contrast-to-noise ratios. Subjective image quality was graded significantly lower for four criteria in group 1 compared with group 3. No significant difference was found in comparisons of groups 2 (except for one criterion) and 4 with group 3. CONCLUSION. It is possible to balance radiation dose and contrast dose against each other and maintain signal-to-noise and contrast-to-noise ratios. Subjective image quality was affected by increased noise level on the images but was judged acceptable in all groups except the one with the lowest radiation dose. C T is commonly used, and its use is steadily increasing [1]. Contrastenhanced CT (CECT) of the abdomen is an important part of the workup for a great number of conditions in patients of all ages. It is fast, robust, and readily available. As with most other procedures, though, CECT also has possible adverse effects for the patient undergoing the examination. The most important risks are radiation-induced injuries and contrast-induced nephropathy (CIN). There is evidence that exposure to high doses of ionizing radiation is a risk factor for cancer development [2]. It has also been found that the lifetime risk of cancer is greater the younger a patient is at exposure [3]. Even though the risks imposed by lower doses of ionizing radiation associated with radiologic examinations such as CT are still under debate, the consensus is to adhere to the as low as reasonably achievable principle and keep a cautious approach to the medical use of ionizing radiation, especially for younger patients. CIN is defined as a greater than 25% or 44-µmol/L increase in serum creatinine within 3 days after administration of IV contrast medium [4]. The risk of CIN is related to the dose of contrast medium and the number of risk factors. Important risk factors are reduced renal function, congestive heart failure, diabetes, and age older than 7 years [4, 5]. CIN is uncommon in patients with normal renal function [6]. There have been reports [6, 7] that the risk of CIN after CECT may be overstated. It remains important, however, to keep to a minimum the dose of IV contrast medium administered to patients with risk factors and to elderly patients. Quantum noise is inversely related to the square root of the tube load (in milliampere seconds), and tube load is directly related to radiation absorbed dose. Reduction of tube load in a CT protocol thus leads to an increase in image noise. The signal-tonoise ratio (SNR) is a description of the relation between attenuation and image noise in AJR:2, February

2 TABLE 1: CT Parameters and Sex Distribution in Four Age Groups Group Female-to-Male Ratio Tube Load (mas) a IV Contrast Dose (mg I/kg) Maximum Iodine Dose (g) : : : > 75 16: a Quality reference. Quality Reference Tube Load (mas) IV Contrast Dose (mg IL/kg) Fig. 1 Theoretic model behind selection of examination parameters for protocol studied. A, Graph shows quality reference tube load plotted in linear relation to patient age. Dot represents formerly used parameter. B, Graph shows resulting arbitrary noise, which is inversely proportional to square root of quality reference tube load. C, Graph shows IV contrast dose plotted to match noise curve. D, Graph shows resulting theoretic constant contrast-to-noise (CNR) ratio. a specified area. Of greater implication for diagnostic imaging is the attenuation difference between adjacent structures, that is, the contrast. The lower the contrast between two structures, the greater their conspicuity is reduced by increased noise. This relation is described by the contrast-to-noise ratio (CNR). A C Quantum Noise (Arbitrary) CNR (Arbitrary) In abdominal CT, diagnosis of most pathologic entities relies on use of contrast medium. A higher contrast dose increases image contrast by a higher iodine concentration in enhancing tissues. Thus it is theoretically possible to balance noise and image contrast against each other by varying the radiation and contrast doses and maintaining a constant CNR. A similar idea was evaluated by Watanabe et al. [8]. The purpose of this study was to evaluate the idea behind an abdominal CECT protocol introduced in our department in May 29 to reduce age-specific risk while theoretically B D 384 AJR:2, February 213

3 CT Risk Reduction Fig. 2 Examples of CT images of abdomen in four age groups. A, 23-year-old woman with minor liver laceration, in group 1. B, 39-year-old man with suspected incarcerated inguinal hernia in group 2. C, 57-year-old woman with abdominal trauma, in group 3. D, 78-year-old woman with suspected gynecologic cancer, in group 4. maintaining constant image quality. According to this protocol, younger patients receive a reduced radiation absorbed dose and an increased contrast dose, and elderly patients receive a reduced contrast dose compensated by an increased radiation dose. Materials and Methods The institutional review board approved this retrospective study, and the requirement for informed consent was waived. Patients 16 years or older undergoing CECT of the abdomen because of acute symptoms were divided into four age groups: group 1, years; group 2, 26 5 years; group 3, years; and group 4, older than 75 years. One hundred patients (63 women, 37 men) examined between May and December 29 were consecutively selected as 25 patients per group (Table 1). Because adipose tissue is poorly perfused, it may be helpful to adapt the IV contrast dose to the A C lean body weight [9]. For this reason, the IV contrast dose in the evaluated protocol was linearly proportional to body weight only up to 75 kg. To keep the number of variables at a minimum in the evaluation of the theoretic model behind the protocol, patients with a body weight exceeding 75 kg were excluded from the study. All patients were examined with a 16-MDCT scanner (Somatom Sensation 16, Siemens Healthcare). The reconstruction kernel was B3f; collimation, 1.5 mm; pitch,.65; and tube voltage, 12 kv. By use of an automatic exposure control system (CARE Dose 4D, Siemens Healthcare) the tube load was set to a level-denominated quality reference. The IV contrast medium used was iohexol (Omnipaque 3 mg I/mL, GE Healthcare). Theoretic Model The quality reference tube load and contrast dose for the different groups were determined with the theoretic model shown in Figure 1, which we have likened to a seesaw balancing radiation dose and IV contrast dose. Radiation dose (quality reference tube load in milliampere seconds) was plotted in a linear relation to age. Quantum noise is inversely related to the square root of the radiation dose) (quality reference tube load), and arbitrary noise values were calculated with this relation. To compensate for variation in image noise, the IV contrast dose curve was plotted to match the noise curve. Theoretically, this plot would yield a constant CNR for all ages. The resulting parameters used for the different groups are shown in Table 1. The parameters used for group 3 were the same as in the formerly used abdominal CECT protocol in the department. Radiation Dose Volume CT dose index and dose-length product (DLP) were recorded for each patient. An estimation of the effective dose (E) for each group B D AJR:2, February

4 TABLE 2: Mean Radiation Dose and Noise Group CT Dose Index (mgy) Effective Dose (msv) Noise (HU) ± ± ± ± ± ± ± ± ± ± ± ± SNR CNR Group 1 Group 2 Group 3 Group 4 was derived from the dose-length product and the conversion factor E DLP as follows: E = E DLP DLP, where E DLP is region specific and DLP is the normalized effective dose (msv / mgy cm) determined by Huda et al. [1] (abdominal region.16). IV Contrast Dose At administered dose of IV contrast was recorded for each patient. Objective Evaluation of Image Quality With the PACS (Sectra PACS, Sectra, Linöping, Sweden), the attenuation (signal) in HU was measured by placement of a circular region of interest with a diameter of approximately 3 cm in a homogeneous part of the right liver lobe, avoiding vascular and biliary structures (Fig. 2). Standard deviation (SD) was used as a measurement of image noise, and the liver SNR was calculated. For calculation of the CNR, the difference in attenuation between the measured value and a hypothetical hypovascular liver metastatic lesion with attenuation of 4 HU was used for contrast and then divided by the noise value. Liver metastatic lesions often have attenuation slightly lower than the liver parenchyma, are typically hypovascular, and are not enhancing in the portal venous phase [11, 12]. Subjective Evaluation of Image Quality Image quality was evaluated by visual grading characteristic (VGC) analysis [13]. In VGC analysis, the task of the observers is to rate their opinion about reproduction or visibility of a certain structure. In a previous study [14], the use of VGC analysis made it possible to discriminate small differences in noise levels between abdominal CT images. In the current Fig. 3 Graph shows mean liver signal-tonoise ratio (SNR) and contrast-to-noise ratio (CNR) for hypothetical hypovascular liver metastatic lesion in different groups. Error bars represent 1 SD of measurements. study, the following seven criteria from the European guidelines on quality criteria for CT (EUR 16262) [15] were used: visually sharp reproduction of the liver parenchyma and intrahepatic portal veins; visually sharp reproduction of the pancreatic contours; visually sharp reproduction of the kidneys and proximal ureters; reproduction of the gallbladder wall; visually sharp reproduction of the right adrenal gland from adjacent structures; visually sharp reproduction of the structures of the liver hilum; and reproduction of the choledochal ductus in the pancreatic parenchyma. All criteria were judged absolutely on a 5-grade scale on which grade 1 was unacceptable; 2, substandard; 3, acceptable; 4, above average; and 5, superior. A sixth alternative, not applicable, was allowed for reproduction of the gallbladder wall if the gallbladder had been removed, and the data were excluded from further analysis. The images were presented and evaluated with the free software ViewDEX [16]. The software and all images were loaded on a USB memory stick and inserted for the evaluation into one of the PACS workstations in our department. All images were viewed individually by four readers: two gastrointestinal radiologists with 15 and 2 years of experience and two residents in radiology with 3 and 4 years of experience. The information on patient identity and scanning parameters was removed, and the images were presented at random to each radiologist. The readers could scroll through the stacks, magnify the images, and alter the window width and window level settings. First an adaption series with 12 patients, three from each group, was presented to each reader to accustom the readers to the method and the range of image quality. The results of the test series were not used for evaluation. Thereafter images from the full set of 1 patients were presented individually to each reader. The rating data were analyzed with methods developed in receiver operating characteristic analysis. A VGC curve was obtained by plotting the cumulative distributions of rating data for two compared systems against each other [13]. The area under the VGC curve (AUC) was used as a measure of the difference in image quality between the two systems. An AUC of.5 corresponded to equal image quality of the two systems, an AUC <.5 indicated that the image quality was higher for the reference system, and an AUC >.5 indicated that the image quality was higher for the evaluated system. Because the parameters used for group 3 were the same as in the formerly used protocol, this group was selected as the reference for the VGC analyses. Software for multiple observers (VGC Analyzer, Båth M, Hansson J, Sahlgrenska Academy at University of Gothenburg) was used to perform calculations comparing groups 1, 2, and 4 with group 3 individually for all seven criteria and all readers combined. If the 95% confidence interval (CI) for the estimation of AUC did not cover the value.5, a statistically significant difference at the 95% level between the two evaluated systems was established. Receiver operating characteristic software (Rockit, CE Metz, University of Chicago) was used to calculate interobserver agreement comparing groups 1, 2, and 4 with group 3 individually for all four readers. In a separate analysis, data on 2 patients (age range, years) examined before May 29 with the old protocol (2 mas quality reference, 42 mg I/kg) were analyzed by three readers and compared with data on 2 weight-matched patients from group 1. This step was taken to study the effect of radiation dose reduction on this age group. The young patients examined with the formerly used protocol were also compared with group 3 to analyze whether there was a difference in image quality between age groups when the same dose parameters were used. VGC analysis gives information only about image quality in relation to another group and not about absolute image quality. It is statistically incorrect to calculate mean values of the grades because they are ordinal data. Therefore, the percentage of grades in each group rated acceptable or better was used as a measure of the absolute level of image quality. Results Radiation Dose The mean volume CT dose index and mean effective doses for the four groups are shown in Table 2. Compared with the values in group 3, representing the formerly used protocol, 386 AJR:2, February 213

5 CT Risk Reduction the mean effective dose was reduced 57% in group 1 and 22% in group 2 but increased 46% in group Group 1 Group 2 Group 4 IV Contrast Dose The mean administered IV contrast dose in each group was: group 1, 594 mg I = kg; group 2, 492 mg I = kg; group 3, 418 mg I = kg; group 4, 344 mg I = kg. This equals a reduction in IV contrast dose by 18% in group 4 and an increase in IV contrast dose by 42% in group 1 and by 18% in group 2. Objective Evaluation of Image Quality The mean liver SNR in groups 1 4 ranged from 7.3 for group 2 to 8.4 for group 4, and the mean CNR values in groups 1 4 ranged from 4.3 for group 3 to 5.1 for group 1 (Fig. 3). Analysis of variance and Tukey test results did not establish a significant difference in SNR or CNR between the groups. Subjective Evaluation of Image Quality The VGC AUCs for all readers combined are presented in Figure 4. In groups 1 and 2 the AUCs showed significantly lower subjective image quality for four and one of the seven criteria compared with the results in group 3. No significant difference between groups 3 and 4 was found. The percentages of grades acceptable or better (grades 3 5) were 71% for group 1, 8% for group 2, 85% for group 3, and 83% for group 4. The comparison between 2 patients from group 1 and 2 patients years old examined with the formerly used protocol (2 mas, 42 mg I/kg), showed significantly lower subjective image quality in group 1 for four criteria (visually sharp reproduction of the pancreatic contours, visually sharp reproduction of the kidneys and proximal ureters, visually sharp reproduction of the right adrenal gland from adjacent structures, and visually sharp reproduction of the structures of the liver hilum). No significant difference was seen between group 3 and the young patients examined with the formerly used protocol. Interobserver Agreement In a direct comparison of all grades, that is, seven criteria in each of the 1 patients, the agreement between the four readers ranged from 29% to 5%. This interobserver agreement was affected by baseline variation in grades between the readers. Analyzing the agreement of significant differences shown by the AUCs for each criterion instead avoids this baseline variation, and the agreement is better, between 69% and 95% (Table 3). AUC Criteria Discussion According to the generally accepted view, the radiation doses associated with CT examinations may increase the lifetime risk of cancer, and IV contrast medium may induce CIN. The protocol evaluated in this study reduced the radiation effective dose 57% in the youngest patient group with maintained CNR, a reduction made possible by a concurrent increase in IV contrast dose. At the other end of the age spectrum, the IV contrast dose given to elderly patients was reduced 18% at the cost of an increased radiation dose. In this way, the most important risk for each age group was reduced. In support of our theory that radiation dose and contrast dose can be balanced against each other is the lack of difference in SNR and CNR between the groups and the lack of significant difference in subjective image quality between groups 3 and 2 (except for one criterion) and between groups 3 and 4. The lack of significant differences in SNR between the groups shows that the variation in contrast enhancement of the liver can match the variation in noise level caused by as much as a threefold difference in radiation dose. Subjective image quality was significantly lower for four criteria in group 1, which was examined with a considerably reduced radiation dose compared with the reference group. Analysis of the percentage of grades acceptable or better within the groups shows similar levels in groups 2 4, whereas the level was slightly lower for group 1. From these findings Fig. 4 Graph shows values of area under the visual grading characteristic curve per criterion for all readers combined. Error bars represent the 95% CI of AUC. If the error bars do not include.5, subjective image quality is statistically different from reference group (group 3). AUC represents difference in image quality between groups and is not an absolute value of subjective image quality. TABLE 3: Interobserver Agreement (%) Reader Reader 1 Reader 2 Reader we conclude that even though the subjective image quality was rated lower for one criterion in group 2, it was still acceptable according to the general opinion of the readers, but it was slightly lower than acceptable for group 1. The discrepancy between objective and subjective image quality in this study is of interest and needs to be analyzed further. One explanation may be that the noise level simply was too high in the low radiation groups, thereby negatively affecting image quality. It may also reflect a weakness in the method in that reproduction of normal structures may not be affected by image noise in a way that is directly comparable to the conspicuity of low-contrast lesions. The theory behind the protocol is that increased noise is compensated by increased contrast of a lesion, and this may not be applicable to the normal structures rated. Yet another explanation is that the readers, from everyday practice, were not accustomed to reading high-noise-level images and rated these images lower. An assumption with VGC analysis is that there is a correlation between the reproduction of anatomic and pathologic features, but further studies are needed to determine the diagnostic performance of this protocol. Another weakness of this study was that CNR was calculated with the difference in attenuation between the measured CT number in the liver and a hypothetical hypovascular lesion with the CT number 4 HU. This simulation was used because liver lesions are AJR:2, February

6 rare in patients in groups 1 and 2 and it would not have been possible to accrue enough patients with liver lesions for the study. The four readers had different baselines in grading subjective image quality. Our method included an adaption series of 12 patients viewed individually by each reader to calibrate the grading. The study showed that 12 was not enough. A joint adaptation session may be a more effective way of avoiding baseline differences between readers in future studies. It my also be wise to standardize the individual reading sessions regarding, for example. session length and time of day, thereby further reducing variations in the study. Only patients with a body weight up to 75 kg were included in the study, whereas in clinical practice the protocol is used for all patients independently of body weight. The assumption is that in most individuals, the additional body weight over 75 kg consists mainly of adipose tissue with limited vascular supply and small interstitial space. The amount of adipose tissue therefore has only a minor effect on the concentration of contrast medium in blood and enhancing organs [9]. A more exact way to determine the contrast dose would be to calculate the lean body weight for all patients, but this is not suitable in emergency care. The effect of body size on the noise level of the images is compensated by the automatic exposure control system. Therefore, we believe the findings of this study are applicable to patients with body weights over, as well as under, 75 kg. To our knowledge, there have been no previous studies exploring the concept of balancing radiation dose against contrast dose for abdominal CT. Watanabe et al. [8] evaluated the idea of increasing the IV contrast dose to allow a reduced radiation dose. Those authors, however, used a slightly different method of rating the depiction of vessels when different radiation and contrast doses were used. They found preserved qualitative image quality at 3% reduced radiation dose compensated by a 15% increase in contrast dose. They used a noise index 12 HU as the definition for a standard radiation dose and 15 HU as a low dose, values similar to the noise levels in groups 2 and 3 in our study (Table 2). Previous studies by Kalra et al. [17] and Wessling et al. [18] showed possible radiation dose reduction up to approximately 4% for abdominal CT without a negative effect on subjective image quality or diagnostic acceptability, this without compensation by an increased contrast dose. Hence there seems to be support for a reduction in radiation dose for abdominal CT in general. The finding that it is possible to compensate a reduction in radiation dose by an increase in IV contrast dose may represent a further step in the pursuit of minimizing radiation dose to young patients. Increasing radiation dose to be able to lower the IV contrast dose administered to older patients, leaving other parameters unchanged, has not, to our knowledge, been proposed before. Our results support this concept, and this may be an important finding for reducing the risk of CIN in examinations of older patients with reduced renal function. Future development of the protocol may be to lower the kilovoltage for smaller patients to achieve a higher CNR for the same amount of iodine and to add iterative reconstruction to reduce the noise level on low-radiation-dose images. Another finding of this study is that the noise level affects subjective image quality more than age-related anatomic differences between patients, such as the amount of intraabdominal fat. This knowledge may be useful for future studies comparing image quality between patients of different ages. Conclusion This study showed that increasing the dose of IV contrast medium can compensate for a reduced radiation dose and vice versa while SNR and CNR are maintained. Subjective image quality was affected by increased noise level on the images but was judged acceptable in all groups except the one with the lowest radiation dose. The discrepancy between objective and subjective image quality may, at least in part, be caused by weaknesses in the method of rating normal anatomic structures. In this protocol the radiation dose was reduced 57% in young patients and the dose of IV contrast medium was reduced 18% in elderly patients. References 1. National Council on Radiation Protection and Measurements (NCRP). Ionizing radiation exposure of the population of the United States. Report no. 16. Bethesda, MD: National Council on Radiation Protection, Brenner DJ, Doll R, Goodhead DT, et al. Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. Proc Natl Acad Sci USA 23; 1: Brenner DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med 27; 357: Morcos SK, Thomsen HS, Webb JA. Contrastmedia-induced nephrotoxicity: a consensus report. Contrast Media Safety Committee, European Society of Urogenital Radiology (ESUR). Eur Radiol 1999; 9: Stacul F, van der Molen AJ, Reimer P, et al. Contrast induced nephropathy: updated ESUR contrast media safety committee guidelines. Eur Radiol 211; 21: Katzberg RW, Newhouse JH. Intravenous contrast medium-induced nephrotoxicity: is the medical risk really as great as we have come to believe? Radiology 21; 256: Bruce RJ, Djamali A, Shinki K, Michel SJ, Fine JP, Pozniak MA. Background fluctuation of kidney function versus contrast-induced nephrotoxicity. AJR 29; 192: Watanabe H, Kanematsu M, Miyoshi T, et al. Improvement of image quality of low radiation dose abdominal CT by increasing contrast enhancement. AJR 21; 195: Kondo H, Kanematsu M, Goshima S, et al. Body size indexes for optimizing iodine dose for aortic and hepatic enhancement at multidetector CT: comparison of total body weight, lean body weight, and blood volume. Radiology 21; 254: Huda W, Ogden KM, Khorasani MR. Converting dose-length product to effective dose at CT. Radiology 28; 248: Kanematsu M, Kondo H, Goshima S, et al. Imaging liver metastases: review and update. Eur J Radiol 26; 58: Heiken JP, Menias CO, Elsayes K. Liver. In: Lee JK, Sagel SS, Stanley RJ, Heiken JP, eds. Computed body tomography with MRI correlation, 4th ed., vol. 1. Philadelphia, PA: Lippincott Williams & Wilkins, 26: Båth M, Månsson LG. Visual grading characteristics (VGC) analysis: a non-parametric rank-invariant statistical method for image quality evaluation. Br J Radiol 27; 8: Leander P, Söderberg M, Fält T, Gunnarsson M, Albertsson I. Post-processing image filtration enabling dose reduction in standard abdominal CT. Radiat Prot Dosimetry 21; 139: Bongartz G, Golding SJ, Jurik AG, et al. European guidelines on quality criteria for computed tomography. Report EUR Luxembourg, Luxembourg: European Commission, Börjesson S, Håkansson M, Båth M, et al. A software tool for increased efficiency in observer performance studies in radiology. Radiat Prot Dosimetry 25; 114: Kalra MK, Maher MM, Kamath RS, et al. Sixteen-detector row CT of abdomen and pelvis: study for optimization of Z-axis modulation technique performed in 153 patients. Radiology 24; 233: Wessling J, Esseling R, Raupach R, et al. The effect of dose reduction and feasibility of edge-preserving noise reduction on the detection of liver lesions using MSCT. Eur Radiol 27; 17: AJR:2, February 213

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760 Can Weight-Adapted IV Contrast Media Protocols Reduce Iodine Dose and Still Produce a Diagnostic Contrast Enhancement Level in Abdomino-Pelvic CT Scans? Poster No.: C-1760 Congress: ECR 2015 Type: Authors:

More information

Diagnostic imaging evaluating image quality using visual grading characteristic (VGC) analysis

Diagnostic imaging evaluating image quality using visual grading characteristic (VGC) analysis Vet Res Commun (2010) 34:473 479 DOI 10.1007/s11259-010-9413-2 SHORT COMMUNICATION Diagnostic imaging evaluating image quality using visual grading characteristic (VGC) analysis Eberhard Ludewig & Andreas

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

Improved image quality of low-dose thoracic CT examinations with a new postprocessing software*

Improved image quality of low-dose thoracic CT examinations with a new postprocessing software* JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 3, Summer 2010 Improved image quality of low-dose thoracic CT examinations with a new postprocessing software* Anne Catrine Traegde Martinsen,

More information

CT Dose Optimization for Whole- Body PET/CT Examinations

CT Dose Optimization for Whole- Body PET/CT Examinations Nuclear Medicine and Molecular Imaging Original Research Tonkopi et al. CT Dose Optimization for PET/CT Nuclear Medicine and Molecular Imaging Original Research FOCUS ON: Elena Tonkopi 1,2 Andrew A. Ross

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation

More information

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Debra Pennington, MD Director of Imaging Dell Children s Medical Center Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

Optimizing radiation dose by varying age at pediatric temporal bone CT

Optimizing radiation dose by varying age at pediatric temporal bone CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Optimizing radiation dose by varying age at pediatric temporal bone CT Daichi Noto, 1 Yoshinori Funama, 2a Mika Kitajima, 3 Daisuke

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

Dual Energy CT Aortography: Can We Reduce Iodine Dose??

Dual Energy CT Aortography: Can We Reduce Iodine Dose?? Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Sagara et al. Abdominal CT With Low Versus Routine Dose Medical Physics and Informatics Original Research Yoshiko Sagara 1 Amy K. Hara 2 William Pavlicek

More information

CT Dose Reduction in Pediatric Patients

CT Dose Reduction in Pediatric Patients CT Dose Reduction in Pediatric Patients By Kelly Firestine, RT(R)(CT)(M) Executive Summary CT is an incredibly valuable imaging tool, but there are unique concerns with pediatric patients, including the

More information

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters Medical Physics and Informatics Review Maldjian and Goldman Reducing Radiation Dose in Body CT Medical Physics and Informatics Review FOCUS ON: Pierre D. Maldjian 1 Alice R. Goldman Maldjian PD, Goldman

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could

More information

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Mannudeep K. Kalra, MD, DNB Michael M. Maher, MD, FRCR, FFR (RCSI) Srinivasa R.

More information

With increasing use of computed tomography (CT) in modern medicine, concerns have arisen regarding increasing radiation dose to the community from med

With increasing use of computed tomography (CT) in modern medicine, concerns have arisen regarding increasing radiation dose to the community from med Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. ORIGINAL RESEARCH

More information

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Original Article Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Shun-Yu Gao, Xiao-Peng Zhang, Yong Cui, Ying-Shi

More information

Radiation Dose Reduction Strategies in Coronary CT Angiography

Radiation Dose Reduction Strategies in Coronary CT Angiography Radiation Dose Reduction Strategies in Coronary CT Angiography Noor Diyana Osman, PhD noordiyana@usm.my Contents: Introduction Radiation dosimetry in CT Radiation risk associated with coronary CT angiography

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax

Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax Poster No.: C-0013 Congress: ECR 2013 Type: Authors: Scientific Exhibit Z. Dejian 1, X. Zhuodong 2, J. Hui 2, L. Xiao 3 ;

More information

Anatomical and Functional MRI of the Pancreas

Anatomical and Functional MRI of the Pancreas Anatomical and Functional MRI of the Pancreas MA Bali, MD, T Metens, PhD Erasme Hospital Free University of Brussels Belgium mbali@ulb.ac.be Introduction The use of MRI to investigate the pancreas has

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

Ultralow Dose Chest CT with MBIR

Ultralow Dose Chest CT with MBIR Ultralow Dose Chest CT with MBIR Ella A. Kazerooni, M.D. Professor & Director Cardiothoracic Radiology Associate Chair for Clinical Affairs University of Michigan Disclosures Consultant: GE Healthcare

More information

CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings

CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings Poster No.: C-3289 Congress: ECR 2010 Type: Scientific Exhibit Topic: Chest - Your latest results Authors: M. K. Gill,

More information

Pseudoenhancement of Renal Cysts: Influence of Lesion Size, Lesion Location, Slice Thickness, and Number of MDCT Detectors

Pseudoenhancement of Renal Cysts: Influence of Lesion Size, Lesion Location, Slice Thickness, and Number of MDCT Detectors Genitourinary Imaging Original Research Tappouni et al. MDCT of Renal Cysts Genitourinary Imaging Original Research Rafel Tappouni 1 Jennifer Kissane 2 Nabeel Sarwani 1 Erik B. Lehman 1 Tappouni R, Kissane

More information

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School OUTLINE 1 st Presentation: Dose optimization strategies

More information

Radiation dose optimization in CT planning of corrective scoliosis surgery. A phantom study.

Radiation dose optimization in CT planning of corrective scoliosis surgery. A phantom study. Radiation dose optimization in CT planning of corrective scoliosis surgery. A phantom study. Abul-Kasim, Kasim; Gunnarsson, Mikael; Maly, Pavel; Ohlin, Acke; Maly Sundgren, Pia Published in: The Neuroradiology

More information

Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis

Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis GE Healthcare Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis CT clinical case study lesion characterization Desiree Morgan, MD Vice Chair of Clinical Research Professor

More information

Purpose. Methods and Materials

Purpose. Methods and Materials The dose of contrast medium should be determined on the basis of estimated lean body weight (LBW) instead of the body weight--excessive amounts are currently given to females. Poster No.: C-2641 Congress:

More information

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference?

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference? Pediatric Imaging Original Research Reid et al. Parameter Selection for Pediatric Abdominal CT Downloaded from www.ajronline.org by 46.3.3.24 on 2/3/18 from IP address 46.3.3.24. Copyright ARRS. For personal

More information

CT image guidance ensures accurate needle placement to

CT image guidance ensures accurate needle placement to Published September 15, 2011 as 10.3174/ajnr.A2634 PATIENT SAFETY T.M. Shepherd C.P. Hess C.T. Chin R. Gould W.P. Dillon Reducing Patient Radiation Dose during CT-Guided Procedures: Demonstration in Spinal

More information

Patient-Tailored Scan Delay for Multiphase Liver CT: Improved Scan Quality and Lesion Conspicuity With a Novel Timing Bolus Method

Patient-Tailored Scan Delay for Multiphase Liver CT: Improved Scan Quality and Lesion Conspicuity With a Novel Timing Bolus Method Gastrointestinal Imaging Original Research Schneider et al. Multiphase Liver CT Gastrointestinal Imaging Original Research J. Gabriel Schneider 1 Zhen J. Wang Wilbur Wang Judy Yee Yanjun Fu Benjamin M.

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial Medical Physics and Informatics Original Research Larke et al. Estimated Radiation Dose for Low-Dose Chest CT Medical Physics and Informatics Original Research Frederick J. Larke 1 Randell L. Kruger 2

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

Assessing the standard dose to standard patients for x-ray investigations

Assessing the standard dose to standard patients for x-ray investigations Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The

More information

BODY TOMOSYNTHESIS A. INTRODUCTION/CURRENT PRACTICE. Conventional radiography (X-ray)

BODY TOMOSYNTHESIS A. INTRODUCTION/CURRENT PRACTICE. Conventional radiography (X-ray) BODY TOMOSYNTHESIS A. INTRODUCTION/CURRENT PRACTICE MAGNUS BÅTH, UNIVERSITY OF GOTHENBURG AND SAHLGRENSKA UNIVERSITY HOSPITAL, SWEDEN Conventional radiography (X-ray) Still fundamental examination in radiology

More information

Observer variation for radiography, computed tomography, and magnetic resonance imaging of occult hip fractures

Observer variation for radiography, computed tomography, and magnetic resonance imaging of occult hip fractures Observer variation for radiography, computed tomography, and magnetic resonance imaging of occult hip fractures Collin, David; Dunker, Dennis; Gothlin, Jan H.; Geijer, Mats Published in: Acta Radiologica

More information

Incidental findings: A retrospective analysis of management

Incidental findings: A retrospective analysis of management Incidental findings: A retrospective analysis of management Authors and disclosures Authors: Steven Boe, Dana Boe, Jeffrey Kaye, Anu Bansal, Marc Glickstein Disclosures: None Purpose Determine if appropriate

More information

Outcomes in the NLST. Health system infrastructure needs to implement screening

Outcomes in the NLST. Health system infrastructure needs to implement screening Outcomes in the NLST Health system infrastructure needs to implement screening Denise R. Aberle, MD Professor of Radiology and Bioengineering David Geffen School of Medicine at UCLA 1 Disclosures I have

More information

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3 Coronary Artery Bypass Graft Imaging with CT Angiography and Iterative Reconstruction: Quantitave Evaluation of Radiation Dose Reduction and Image Quality Simon Nepveu 1, Irina Boldeanu 1, Yves Provost

More information

Radiation Dose Reduction with Hybrid Iterative Reconstruction for Pediatric CT 1

Radiation Dose Reduction with Hybrid Iterative Reconstruction for Pediatric CT 1 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Sarabjeet Singh,

More information

Minimising Ionising Radia-on Exposure in Paediatric Radiology

Minimising Ionising Radia-on Exposure in Paediatric Radiology Minimising Ionising Radia-on Exposure in Paediatric Radiology Alexandra Tavares Ferreira Hospital Dona Estefânia CHLC Coordenadora Radiologia- Dra. Eugénia Soares 19 Setembro 2013 Imaging Exams Ultrasound

More information

Any imbalance in the production or resorption of CSF results

Any imbalance in the production or resorption of CSF results ORIGINAL RESEARCH U.K. Udayasankar K. Braithwaite M. Arvaniti D. Tudorascu W.C. Small S. Little S. Palasis Low-Dose Nonenhanced Head CT Protocol for Follow-Up Evaluation of Children with Ventriculoperitoneal

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

B. CT protocols for the spine

B. CT protocols for the spine B. CT protocols for the spine Poster No.: A-003 Congress: ECR 2010 Type: Invited Speaker Topic: Neuro Authors: B. Tins; Oswestry/UK Keywords: CT, spine, diagnostic imaging protocol DOI: 10.1594/ecr2010/A-003

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Christner et al. Estimating Effective Dose for CT Medical Physics and Informatics Original Research FOCUS ON: Jodie A. Christner 1 James M. Kofler Cynthia

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

Lung Perfusion Analysis New Pathways in Lung Imaging. Case Study Brochure PLA 309 Hospital

Lung Perfusion Analysis New Pathways in Lung Imaging. Case Study Brochure PLA 309 Hospital Lung Perfusion Analysis New Pathways in Lung Imaging Case Study Brochure PLA 309 Hospital http://www.toshibamedicalsystems.com Toshiba Medical Systems Corporation 2012 all rights reserved. Design and specifications

More information

Estimating Iodine Concentration from CT Number Enhancement

Estimating Iodine Concentration from CT Number Enhancement Estimating Iodine Concentration from CT Number Enhancement Rosemary Eaton, Andrew Shah, Jane Shekhdar Medical Physics, Mount Vernon Hospital CT Users Group 4 th October 212, Edinburgh Summary Background

More information

Ryan Niederkohr, M.D. Slides are not to be reproduced without permission of author

Ryan Niederkohr, M.D. Slides are not to be reproduced without permission of author Ryan Niederkohr, M.D. CMS: PET/CT CPT CODES 78814 Limited Area (e.g., head/neck only; chest only) 78815 78816 Regional (skull base to mid-thighs) True Whole Body (skull vertex to feet) SELECTING FIELD

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation Medical Physics and Informatics Original Research Angel et al. Radiation Dose During Chest CT Medical Physics and Informatics Original Research Erin Angel 1,2 Nazanin Yaghmai 1 Cecilia Matilda Jude 1 John

More information

Gastrointestinal Imaging Original Research

Gastrointestinal Imaging Original Research Gastrointestinal Imaging Original Research MDCT Enterography in Crohn s Disease Gastrointestinal Imaging Original Research FOCUS ON: Brian C. Allen 1 Mark E. Baker 1 David M. Einstein 1 Erick M. Remer

More information

Vascular Enhancement and Image Quality of CT Venography: Comparison of Standard and Low Kilovoltage Settings

Vascular Enhancement and Image Quality of CT Venography: Comparison of Standard and Low Kilovoltage Settings Cardiopulmonary Imaging Original Research Fujikawa et al. Vascular Enhancement and Image Quality of CTV Cardiopulmonary Imaging Original Research Atsuko Fujikawa 1 Shin Matsuoka Kenji Kuramochi Tatsuo

More information

Acute renal colic Radiological investigation in patients with renal colic

Acute renal colic Radiological investigation in patients with renal colic Acute renal colic Radiological investigation in patients with renal colic Mikael Hellström Professor Department of Radiology Sahlgrenska University Hospital Göteborg University 0.9-1.8/1.000 inhabitants

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling Investigations and research Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling J. Wang Z. Ying V. Yao L. Ciancibello S. Premraj S. Pohlman

More information

Estimation of the number of total and pediatric CT procedures based. on a nationwide survey in Japan

Estimation of the number of total and pediatric CT procedures based. on a nationwide survey in Japan Estimation of the number of total and pediatric CT procedures based on a nationwide survey in Japan Koji Ono 1, Nobuhiko Ban 2, Kai Michiaki 1. 1 Oita University of Nursing and Health Sciences 2944-9,

More information

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose?

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose? Journal of Physics: Conference Series PAPER OPEN ACCESS Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose? To cite this article: M Smarda et

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

Guidelines for the Management of Pulmonary Nodules Detected by Low-dose CT Lung Cancer Screening

Guidelines for the Management of Pulmonary Nodules Detected by Low-dose CT Lung Cancer Screening Guidelines for the Management of Pulmonary Nodules Detected by Low-dose CT Lung Cancer Screening 1. Introduction In January 2005, the Committee for Preparation of Clinical Practice Guidelines for the Management

More information

Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit detector dual-source computed tomography

Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit detector dual-source computed tomography Acta Radiol OnlineFirst, published on June 11, 2014 as doi:10.1177/0284185114535391 Original Article Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit

More information

Comparison of multidetector-row computed tomography findings of IgG4-related sclerosing cholangitis and cholangiocarcinoma

Comparison of multidetector-row computed tomography findings of IgG4-related sclerosing cholangitis and cholangiocarcinoma Comparison of multidetector-row computed tomography findings of IgG4-related sclerosing cholangitis and cholangiocarcinoma Poster No.: C-0245 Congress: ECR 2014 Type: Scientific Exhibit Authors: M. Yata,

More information

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Omer Osman,Saeed Medical Physics Department, ALNeelain University, Sudan Presentation outlines Introduction Objectives Materials and

More information

Comparison of CT findings between MDR-TB and XDR-TB

Comparison of CT findings between MDR-TB and XDR-TB Comparison of CT findings between MDR-TB and XDR-TB Poster No.: C-0757 Congress: ECR 2017 Type: Authors: Keywords: DOI: Scientific Exhibit K. Yoon, H. Soohee; Changwon-si/KR Thorax, Lung, Respiratory system,

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Minimizing CT Radiation Dose CT examinations improve health care and are an essential part

More information

Reducing Radiation Dose in Body CT: A Practical Approach to Optimizing CT Protocols

Reducing Radiation Dose in Body CT: A Practical Approach to Optimizing CT Protocols Medical Physics and Informatics Review Goldman and Maldjian Reducing Radiation Dose in ody CT Medical Physics and Informatics Review Downloaded from www.ajronline.org by lice Goldman on 03/26/13 from IP

More information

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination Int J Clin Exp Med 2015;8(11):21898-21905 www.ijcem.com /ISSN:1940-5901/IJCEM0015005 Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out

More information

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate)

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate) Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate) Overview Ga-68 dotatate binds to somatostatin receptors, with highest affinity for subtype 2 receptors (sstr2). It binds to cells that express

More information

Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT

Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Hoang et al. Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Jenny K. Hoang 1,2 Terry T. Yoshizumi 1 Kingshuk Roy

More information

Computed Tomography of Normal Adrenal Glands in Indian Population

Computed Tomography of Normal Adrenal Glands in Indian Population IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 17, Issue 01 Ver. V January. (2018), PP 26-30 www.iosrjournals.org Computed Tomography of Normal Adrenal

More information

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation Petter Bugge Askeland Project thesis at the Faculty of Medicine UNIVERSITETET

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Initial Assessment of Dual-Energy CT in Patients With Gallstones or Bile Duct Stones: Can Virtual Nonenhanced Images Replace True Nonenhanced Images?

Initial Assessment of Dual-Energy CT in Patients With Gallstones or Bile Duct Stones: Can Virtual Nonenhanced Images Replace True Nonenhanced Images? Gastrointestinal Imaging Original Research Kim et al. Evaluation of Biliary Stones Gastrointestinal Imaging Original Research Ji Eun Kim 1 Jeong Min Lee 2 Jee Hyun Baek 2 Joon Koo Han 2 Byung Ihn Choi

More information

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Alice B. Smith, MD

More information

Vascular and Interventional Radiology Original Research

Vascular and Interventional Radiology Original Research Vascular and Interventional Radiology Original Research Apfaltrer et al. High-Pitch Versus Standard-Pitch CTA of the Aorta Vascular and Interventional Radiology Original Research Paul Apfaltrer 1,2 E.

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

8/3/2016. Consultant for / research support from: Astellas Bayer Bracco GE Healthcare Guerbet Medrad Siemens Healthcare. Single Energy.

8/3/2016. Consultant for / research support from: Astellas Bayer Bracco GE Healthcare Guerbet Medrad Siemens Healthcare. Single Energy. U. Joseph Schoepf, MD Prof. (h.c.), FAHA, FSCBT-MR, FNASCI, FSCCT Professor of Radiology, Medicine, and Pediatrics Director, Division of Cardiovascular Imaging Consultant for / research support from: Astellas

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

Abdominal CT protocol s influence on postoperative follow-up of lesions detection associated with gastrointestinal tumours

Abdominal CT protocol s influence on postoperative follow-up of lesions detection associated with gastrointestinal tumours Original Article on Translational Imaging in Cancer Patient Care Abdominal CT protocol s influence on postoperative follow-up of lesions detection associated with gastrointestinal tumours Jing-Feng Zhang

More information

Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT

Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT The Harvard community has made this article openly available. Please share how this access benefits you. Your story

More information

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens DEfinitely Siemens True Dual Energy Dr. Stefan Ulzheimer, Siemens Healthcare GmbH International version. Not for distribution in the US. Unrestricted Siemens AG 2015 All rights reserved. The products/features

More information

A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction.

A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction. Biomedical Research 2017; 28 (17): 7683-7687 ISSN 0970-938X www.biomedres.info A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction. Yanhe

More information