Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models

Size: px
Start display at page:

Download "Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models"

Transcription

1 Clinical Chemistry 60: (2014) Endocrinology and Metabolism Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models James C. Boyd 1 and David E. Bruns 1* BACKGROUND: Total error allowances have been proposed for glucose meters used in tight-glucose-control (TGC) protocols. It is unclear whether these proposed quality specifications are appropriate for continuous glucose monitoring (CGM). METHODS: We performed Monte Carlo simulations of patients on TGC protocols. To simulate use of glucose meters, measurements were made hourly. To simulate CGM, glucose measurements were made every 5 min. Glucose was measured with defined bias (varied from 20% to 20%) and imprecision (0% to 20% CV). The measured glucose concentrations were used to alter insulin infusion rates according to established treatment protocols. Changes in true glucose were calculated hourly on the basis of the insulin infusion rate, the modeled patient s insulin sensitivity, and a model of glucose homeostasis. We modeled patients, equally divided between the hourly and every-5-min measurement schemas and distributed among 45 combinations of bias and imprecision and 2 treatment protocols. RESULTS: With both treatment protocols and both measurement frequencies, higher measurement imprecision increased the rates of hypoglycemia and hyperglycemia and increased glycemic variability (SD). These adverse effects of measurement imprecision were lower at the higher measurement frequency. The rate of hypoglycemia at an imprecision (CV) of 5% with hourly measurements was similar to the rate of hypoglycemia at 10% CV when measurements were made every 5 min. With measurements every 5 min, imprecision up to 10% had minimal effects on hyperglycemia or glycemic variability. Effects of simulated analytical bias on glycemia were unaffected by measurement frequency. CONCLUSIONS: Quality specifications for imprecision of glucose meters are not transferable to CGM American Association for Clinical Chemistry Control of blood glucose is a major goal of therapy in diabetes and other conditions associated with hyperglycemia (1, 2). In the management of stress-induced hyperglycemia, which is commonly observed in patients with or without diabetes in intensive care units (ICUs), 2 tight glucose control (TGC) reportedly decreased morbidity or mortality or both in surgical, medical, and pediatric ICUs (3 5). More recently, however, a large multicenter study (6) and 2 metaanalyses of published studies (7, 8) have reported a failure of TGC to decrease mortality. A factor in the success of TGC may be the performance of the devices used to measure glucose (9). The successful TGC outcomes (3 5) were achieved with a highly accurate glucose measurement system and blood samples collected from central lines. By contrast, most subsequent studies that failed to achieve reductions in mortality used less accurate measurement systems (9). Thus glucose measurement systems may play a role in the reported failures of TGC, but it is difficult to test the relationship of measurement error and glycemic control in clinical trials. To explore the effects of measurement error on control of glycemia, we previously used in silico simulations of patients on TGC (10). In these in silico simulations (10), each modeled patient s true glucose was measured hourly by simulated devices with defined bias and imprecision. The measured glucose concentrations were used to select insulin infusion rates according to the Yale (11) and University of Washington (UW) (12) treatment regimens. Regardless of the treatment regimen used, increases in bias and imprecision of the measurements were found to be associated with higher rates of hypo- or hyperglycemia or both in the modeled patients (10). Moreover, increased imprecision was associated with higher variability in the patients true glucose. We concluded that to avoid degradation of glycemic control, total error in the hourly measurements needed to be 10% 15%. Similarly, a 1 Department of Pathology, University of Virginia, Charlottesville, VA. * Address correspondence to this author at: Department of Pathology, Box , University of Virginia School of Medicine and Health System, Charlottesville, VA dbruns@virginia.edu. Received September 17, 2013; accepted December 18, Previously published online at DOI: /clinchem Nonstandard abbreviations: ICU, intensive care unit; TGC, tight glucose control; UW, University of Washington; CGM, continuous glucose monitoring. 644

2 Effects of Measurement Frequency on Quality Goals current guideline on glucose meters for use in hospitals concludes that the quality specification for total error of glucose meters used in acute care facilities should be 12.5% at most concentrations of glucose and may need to be 10% for TGC in ICUs (13). Recent efforts in TGC have focused on the possible use of continuous glucose monitoring (CGM) to replace glucose meters (14). More frequent measurements of glucose provided by CGM devices are expected to improve the control of glycemia and decrease occurrences of hypoglycemia (15). However, it is not currently known to what extent the relationship between measurement error and glycemic control may be affected by increasing the frequency of glucose measurement. Here we report studies aimed at using our earlier modeling approach to address this question by comparing the control of glycemia when glucose measurements were made hourly (as in our earlier studies) vs 12 times per hour, as may be seen with CGM. Methods The modeling scheme presented previously (10) was modified and extended to simulate CGM with glucose measurements made more frequently than once per hour. In brief, our approach, as before, used simulated glucose measurements by assays with specified amounts of inaccuracy and imprecision. Simulated assay bias was varied from 20% to 20% in increments of 5% to assess the effects of inaccuracy; simulated imprecision (expressed as CV) was varied from 0% to 20% in increments of 5%. We used measured glucose concentrations to select insulin infusion rates on the basis of 2 published insulin administration protocols (11, 12), referred to here as the Yale and UW regimens, respectively. As in the previous study (10) modeled patients had various insulin sensitivities, starting glucose concentrations, and rates of gluconeogenesis and intravenous glucose administration, all set with use of random number generators. Starting glucose concentrations, for example, were randomly set with equal probabilities of occurrence in the range of mg/dl ( mmol/l). For each modeled patient, we measured glucose concentrations by a simulated assay having a preselected combination of bias and imprecision. We used the measured glucose concentrations to make hourly adjustments to insulin infusion rates as specified by either the Yale or UW regimen. One hundred patients were modeled at each combination of bias and imprecision for 100 h with each treatment regimen. We used a model of glucose homeostasis, patterned after the minimal model of glucose regulation (16), to determine glucose concentrations after each hour of insulin infusion at the chosen rate. Within-patient variability in rates of gluconeogenesis or glucose infusion was modeled by adding normally distributed random hourly mean (SD) increments of 50 (10) mg/dl to glucose concentrations predicted from the minimal model of glucose regulation. Compared with our earlier study, the simulations using the UW regimen were enhanced to better reflect the specific steps suggested in that protocol for treatment of hypoglycemia. For this, the increment in actual plasma glucose 1 h after giving 12.5 g glucose intravenously was estimated to be 44 mg/dl, the concentration that would remain after the expected 141 mg/dl initial increment in a person with a typical 7.25-L volume of distribution (17, 18). To simplify calculations, we treated these hypoglycemic patients according to the UW protocol for conscious patients only. MODELING OF CGM Each modeled patient s true glucose was calculated hourly based on the change from the current glucose concentration predicted by the physiologic model for each hour s rate of insulin infusion specified by the TGC treatment regimen. For the simulations in which glucose was measured every 5 min, we calculated the hourly true glucose concentrations every 5 min from an assumed linear relationship between the initial glucose concentrations and the concentration predicted by the physiologic model (and the then-current insulin infusion rate) to occur at 60 min. Each of the 12 true glucose concentrations per hour was measured by a simulated glucose assay with a defined imprecision and bias. The results of the 12 measurements over each hour were fitted by linear least-squares regression, and the glucose concentration predicted from the regression line at 60 min (rather than the single measured glucose at 60 min) was used to determine the insulin infusion rate according to the chosen treatment regimen (UW or Yale) for that patient. Thus, 100 patients were modeled for 100 h with each regimen with both the hourly and every-5-min measurement schemas under 45 combinations of simulated bias and imprecision. This modeling generated simulated glucose measurements for the 2 regimens combined. To investigate the effect of a CGM failing to provide a reading at a given time point, we conducted additional simulations, changing the model to randomly drop a mean of 10% of simulated CGM results. Because each result was assigned a 10% probability of being suppressed, and unavailable for use, in a given sequence of 12 between-hour results, it was possible to observe 0, 1, 2, or more missing results; by probability calculations, only 28% of sequences in this experiment would have all 12 results. Clinical Chemistry 60:4 (2014) 645

3 In a separate set of simulation runs, the variability of the glucose measurements at concentrations 100 mg/dl was expressed in terms of constant SDs (in mg/ dl) rather than as a CV. For each selected CV, the corresponding constant SD was calculated by multiplying 100 mg/dl times the modeled CV expressed as a fraction. We conducted another set of simulation runs to assess the effect of immediate intervention on the basis of any finding of a single measured glucose in the hypoglycemic range between adjacent 1-hour time points. Such a hypoglycemic result caused an early dropout from the process of collecting 12 glucose measurements each hour, and the insulin treatment regimen was invoked immediately to perform treatment that was appropriate for the measured (low) glucose concentration. In all experiments, the observed rates of hypoglycemia and hyperglycemia, the percentage of time that true glucose was in the target range, and the variability (SD) of hourly true glucose measurements were monitored as outcomes. Results RATES OF HYPOGLYCEMIA Fig. 1 shows the relationship between the frequency of true hypoglycemia and the imprecision (expressed as CV) of the glucose measurements. In this figure, the bias was held at zero. Compared with hourly measurements, measurements made every 5 min were associated with lower frequencies of (true) hypoglycemia in the modeled patients (Fig. 1). This reduced frequency of hypoglycemia with more frequent glucose measurements was seen at analytical imprecision (CVs) of 5% through 20% when using either the Yale or UW regimen for selecting insulin infusion rates (Fig. 1). With the Yale regimen, the rate of hypoglycemic episodes with measurements every 5 min at 10% CV was similar to that seen with hourly measurements at 5% CV. When biases (constant errors) of 20% to 20% were introduced in the glucose measurement along with imprecision, the more frequent measurements again produced similar decreases of hypoglycemia at any level of bias (contour plots, Supplemental Fig. 1, which accompanies the online version of this article at Increased frequency of measurement had no perceptible ability to decrease the effects of bias (see online Supplemental Fig. 1). RATES OF HYPERGLYCEMIA AND TIME IN TARGET RANGE Fig. 2 shows the effects of glucose measurement imprecision on the rates of hyperglycemia when the measurement system has zero bias. As with hypoglycemia, the Fig. 1. Effects of imprecision, in the absence of bias, on the frequency of hypoglycemia in modeled patients. The frequency of true glucose concentrations 60 mg/dl is expressed as the number of hypoglycemic results at the 1-h time points divided by the number of hourly measurements (10 000) in the 100 patients modeled for each CV. q 5 min, 5-min intervals. frequency of hyperglycemia was lower, at any analytical imprecision, when measurements were made every 5 min rather than every hour. Similar effects of imprecision and frequency of measurement were seen when bias was added (see online Supplemental Fig. 2). The time that modeled patients true glucose was in the target range decreased with increasing imprecision (Fig. 3). Again, this adverse effect of imprecision was markedly blunted when glucose was measured every 5 min rather than every hour (Fig. 3). GLYCEMIC VARIABILITY The variability of the modeled patients true glucose increased with increasing imprecision of the glucose measurements (Fig. 4). As before, this adverse effect of imprecision was blunted with more frequent measurements of glucose. The effect of increased frequency of measurements was seen with both the Yale and UW regimens. Even at a measurement imprecision (CV) as high as 10% (UW) or 15% (Yale), the glycemic variability in the modeled patients with measurements every 5 min was comparable to the glycemic variability seen with no imprecision when measurements were 646 Clinical Chemistry 60:4 (2014)

4 Effects of Measurement Frequency on Quality Goals Fig. 2. Effects of imprecision, in the absence of bias, on frequency of hyperglycemia (>160 mg/dl) in modeled patients. The frequency of the indicated true glucose concentrations was calculated as in Fig. 1. q 5 min, 5-min intervals. Fig. 3. Effects of imprecision, in the absence of bias, on the frequency of true glucose concentrations in the target range of mg/dl at the 1-h time points. The frequency of the indicated glucose concentrations was calculated as in Fig. 1. q 5 min, 5-min intervals. made hourly (Fig. 4). Introduction of bias with the imprecision had minimal or no detectable additional effect on variability of the true glucose (not shown). The individual within-patient SDs for the 100 modeled patients at each glucose measurement imprecision (and zero bias) value are presented in online Supplemental Fig. 3. The detailed view illustrates that some modeled patients had extremely high glycemic variability (SDs). On investigation, these patients uniformly had starting glucose concentrations that were in the upper portion ( 450 mg/dl) of the range of allowed values. When the studies were repeated with the range of starting glucose concentrations limited to mg/dl, the within-patient variability of glucose was reduced at each frequency of measurement, but the findings of the study remained unchanged: Increasing the frequency of measurements decreased the glycemic variability and decreased the rates of hypoglycemia and hyperglycemia in the modeled patients. EFFECTS OF OTHER VARIABLES When the model was changed to randomly drop 10% of results (i.e., each result had a 10% chance of being suppressed and unavailable for use), no perceptible effect was observed on the measured outcomes (not shown). The regression fitting of the remaining results when approximately 10% of overall results were missing appeared sufficient to compensate for these missing values. Changing the imprecisions at low concentrations ( 100 mg/dl) from a proportion of the measured value (CV) to a fixed concentration (fixed at the SD equivalent to the CV% at a glucose concentration of 100 mg/dl) produced no evident effect on glycemic control in the modeled patients (not shown). When the model was changed to allow immediate intervention on the basis of a single hypoglycemic result between the 1-hour time points, the frequencies of hypoglycemia were slightly decreased in some trials (not shown). Discussion This simulation study demonstrated that increasing the frequency of measurement of glucose reduced the adverse impact of glucose measurement imprecision on the control of glycemia in modeled patients. This effect was seen at measurement imprecisions (CVs) of 5% to 20% and was seen with each of the 2 treatment regimens tested. More frequent measurements increased the time that modeled patients true plasma Clinical Chemistry 60:4 (2014) 647

5 Fig. 4. Effects of imprecision, in the absence of bias, on glycemic variability expressed as standard deviation (SD) of true glucose concentrations at the 1-h time points. Each plotted point represents the median SD for 100 patients. q 5 min, 5-min intervals. glucose concentrations were in the target zone. Improvements were also seen in each of the 3 aspects of glycemic control that are most predictive of patient outcomes, namely glycemic variability and the rates of hypoglycemia and hyperglycemia. The effects of bias were not altered by increasing the frequency of glucose measurements. The approach taken in these studies was to use our previously published modeling scheme (10) to examine the effect of changing the frequency of glucose measurement. We compared hourly measurements, which we had used previously to model TGC with glucose meters, with measurements made once every 5 min, as may be achieved with CGM. The previous simulation suggested that a total error of up to 10% 15% in hourly glucose measurements was compatible with control of glycemia when the measurements were used with the Yale or UW regimens (10). With both regimens, total errors of 10% or less produced little change of the control of glycemia from that seen with a perfect meter (10). Modeling studies by others have suggested similar or identical allowable total errors both for TGC (19) and for glucose meters used in self-monitoring of glucose (20). The latter study concluded that a threshold effect appeared to exist between 10% and 15% total error. A recent guideline from the Clinical Laboratory Standards Institute identified a goal of 12.5% total error for glucose meters used in acute and chronic healthcare facilities (13). The findings of the present study can be used to evaluate whether these conclusions and the Clinical Laboratory Standards Institute goal are appropriate when measurements are made more frequently than hourly. The results suggest that the above criteria for glucose meters in acute care hospitals may not apply to CGM devices that make glucose measurements more frequently than hourly. The improvement in glycemic control that was seen with increased frequency of measurements was a result of limiting the adverse impact of measurement imprecision. Theoretically, as measurement frequency increases, the confidence in the combined information from the measurements increases in a fashion similar to that expected if the measurements were expressed as a moving average. The effect of random errors would be expected to decrease with the square root of the number of determinations used for the moving average. By use of regression analysis for prediction of the measured glucose at 60 min based on up to 12 glucose measurements over a 1-h interval, we have performed an analogous calculation, reducing the information from the multiple measurements into a single extrapolated measurement with improved precision. Such effects of increased frequency of measurement are not limited to CGM devices but would also be expected when more frequent measurements are made with glucose meters or other measuring systems. Glycemic control in our studies was not noticeably improved by the use of the frequent measurements to detect hypoglycemia that occurred between the hourly measurements. The lack of a major positive effect may reflect offsetting, adverse effects that occur when the imprecision of single measurements generates false signals of hypoglycemia and thus infusion of glucose that is not warranted. An advanced algorithm, perhaps evaluating trends in glucose concentrations over short time periods, may be better able than the regimens used here to exploit the potential advantage of earlier detection of hypoglycemia while largely avoiding the penalty from imprecision of single measurements. This study has several limitations. The model of glucose homeostasis is a simplified model and may not reflect true physiologic behaviors accurately in all circumstances, nor does it allow for lag in the sensing of changes of glucose concentrations. The aim of this study, however, was to explore how our previously published results, using this model, would be affected when the frequency of measurements was increased. Moreover, for glucose meters, our conclusions using 648 Clinical Chemistry 60:4 (2014)

6 Effects of Measurement Frequency on Quality Goals this model (10) were similar to those of Breton and Kovatchev (20), who used a highly sophisticated model of glucose control, albeit in a different setting (selfmonitoring of blood glucose). A limitation of the model is that it assumes a linear relationship of true plasma glucose concentrations and time between hourly time points. Thus a linear least-squares fit was adequate for the regression analysis of measured glucose concentrations vs time. In reality, the time courses will, in some time segments, be nonlinear, which will require fitting with nonlinear equations. We could not evaluate that approach with our model. A related limitation of our study is that the modeling scheme may not have adequately accounted for factors such as oral feeding, steroid injections, and infections that occur in clinical practice and alter insulin sensitivity and other variables related to glycemia, nor did it address effects of changes in analytical bias (or imprecision) that may occur during the 100 h of monitoring of each modeled patient. The simulation did, however, introduce considerable variability in the hour-to-hour increments in glucose attributed to gluconeogenesis/glucose infusion as described in Methods. Moreover, this same limitation applied to the 2 frequency-of-glucose-measurement schemas that were being compared. Finally, these studies addressed only the imprecision and bias components of analytical error; the studies did not address extrinsic random and nonrandom variation generated by the clinical context. The data reported therefore represent a best-case scenario for glycemic control at the indicated levels of bias and imprecision. This is a clear limitation, but the model nonetheless allows exploration of the effects of changes in the selected inputs for the model, as was the intent of the present experiments. This study has several strengths. The simulation used a model of analytical error that distinguished effects of imprecision and bias rather than relying solely on the concepts of total error or mean absolute relative difference. Such a modeling scheme appears to be important for at least 2 reasons: a) Imprecision and bias produced different effects on the modeled patients glycemic control, as can be seen in the contour plots in online Supplemental Figs. 1 and 2; and b) each of these components of analytical error may be more or less problematic for a given CGM device. A second strength is that the same modeling conditions were applied to both frequencies of measurement, allowing direct comparison of the two approaches. Other variables were unchanged, including the frequency of changing insulin infusion rates. The use of the Yale and UW regimens to determine the magnitude, direction, and frequency of changes of insulin and glucose infusions seems reasonable because such regimens are representative of those likely to be used in the initial implementations of CGM devices. Studies like ours may provide insights that inform decisions about analytical quality requirements and design goals for CGM devices. Finally, Finfer et al. (21) recently called for randomized controlled trials to compare the abilities of CGM and intermittent monitoring to achieve and maintain glycemic control. The relationships found in modeling studies like ours may provide insights that are useful in the design and interpretation of such clinical trials. In the future, closed-loop algorithms likely will be used to control insulin infusion rates on the basis of measured glucose concentrations; the effect of glucose measurement error on each such closed-loop algorithm will need to be assessed individually. Instinctively, one may expect that the power of advanced algorithms for these devices, combined with high measurement frequencies, will mitigate the effect of measurement error. If these algorithms make changes in insulin infusion rates on the basis of small changes in glucose concentrations, however, the effect of small measurement errors could be magnified. In summary, the studies described here show that the predicted quality requirements for glucose measurements depend on the frequency of measurement. We conclude that analytical quality requirements proposed for glucose meters with hourly measurements for use in ICUs are inappropriate for CGMs with more frequent measurements. Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article. Authors Disclosures or Potential Conflicts of Interest: Upon manuscript submission, all authors completed the author disclosure form. Disclosures and/or potential conflicts of interest: Employment or Leadership: J.C. Boyd, Clinical Chemistry, AACC. Consultant or Advisory Role: None declared. Stock Ownership: None declared. Honoraria: None declared. Research Funding: This study was supported by funding to the University of Virginia from Edwards Life Sciences. Expert Testimony: None declared. Patents: None declared. Role of Sponsor: The funding organizations played no role in the design of study, choice of enrolled patients, review and interpretation of data, or preparation or approval of manuscript. Clinical Chemistry 60:4 (2014) 649

7 1. American Diabetes Association. Standards of medical care in diabetes Diabetes Care 2012;35 Suppl 1:S Jacobi J, Bircher N, Krinsley J, Agus M, Braithwaite SS, Deutschman C, et al. Guidelines for the use of an insulin infusion for the management of hyperglycemia in critically ill patients. Crit Care Med 2012;40: Van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, et al. Intensive insulin therapy in critically ill patients. N Engl J Med 2001;345: Van den Berghe G, Wilmer A, Hermans G, Meersseman W, Wouters PJ, Milants I et al. Intensive insulin therapy in the medical ICU. N Engl J Med 2006;354: Vlasselaers D, Milants I, Desmet L, Wouters PJ, Vanhorebeek I, van den Heuvel I et al. (2009) Intensive insulin therapy for patients in paediatric intensive care: a prospective, randomised controlled study. Lancet 2009;373: NICE-SUGAR Study Investigators, Finfer S, Chittock DR, Su SY, Blair D, Foster D, Dhingra V et al. Intensive versus conventional glucose control in critically ill patients. N Engl J Med 2009;360: Wiener RS, Wiener DC, Larson RJ. Benefits and risks of tight glucose control in critically ill adults: a meta-analysis. JAMA 2008;300: Griesdale DE, de Souza RJ, van Dam RM, Heyland DK, Cook DJ, Malhotra A et al. Intensive insulin References therapy and mortality among critically ill patients: a meta-analysis including NICE-SUGAR study data. CMAJ 2009;180: Scott MG, Bruns DE, Boyd JC, Sacks DB. Tight glucose control in the intensive care unit: are glucose meters up to the task? Clin Chem 2009; 55: Boyd JC, Bruns DE. Monte Carlo simulation in establishing analytical quality requirements for clinical laboratory tests: meeting clinical needs. Methods Enzymol 2009;467: Goldberg PA, Siegel MD, Sherwin RS, Halickman JI, Lee M, Bailey VA et al. Implementation of a safe and effective insulin infusion protocol in a medical intensive care unit. Diabetes Care 2004; 27: Trence DL, Kelly JL, Hirsch IB. The rationale and management of hyperglycemia for in-patients with cardiovascular disease: time for change. J Clin Endocrinol Metab 2003;88: Sacks DB, Bruns DE, Horton J, Lindberg S, Mahoney JJ, Manzella S et al. Point-of-care blood glucose testing in acute and chronic care facilities: approved guideline. Clinical Laboratory Standards Institute 2013;33: Damiano ER, El-Khatib FH, Zheng H, Nathan DM, Russell SJ. A comparative effectiveness analysis of three continuous glucose monitors. Diabetes Care 2013;36: Pickup JC, Freeman SC, Sutton AJ. Glycaemic control in type 1 diabetes during real time continuous glucose monitoring compared with self monitoring of blood glucose: meta-analysis of randomised controlled trials using individual patient data. BMJ 2011;343:d Toffolo G, Bergman RN, Finegood DT, Bowden CR, Cobelli C. Quantitative estimation of beta cell sensitivity to glucose in the intact organism: a minimal model of insulin kinetics in the dog. Diabetes 1980;29: Cunningham VJ, Heath DF. An interpretation of the intravenous glucose tolerance test in the light of recent findings on the kinetics of glucose and insulin in man. Clin Sci Mol Med 1978;54: Duncan LJ. The intravenous glucose tolerance test. Exp Physiol 1956;41: Karon BS, Boyd JC, Klee GG. Glucose meter performance criteria for tight glycemic control estimated by simulation modeling. Clin Chem 2010; 56: Breton MD, Kovatchev BP. Impact of blood glucose self-monitoring errors on glucose variability, risk for hypoglycemia, and average glucose control in type 1 diabetes: an in silico study. J Diabetes Sci Technol 2010;4: Finfer S, Wernerman J, Preiser JC, Cass T, Desaive T, Hovorka R et al. Clinical review: consensus recommendations on measurement of blood glucose and reporting glycemic control in critically ill adults. Crit Care 2013;17: Clinical Chemistry 60:4 (2014)

10.4.a. Optimal glucose control: Insulin therapy March 2013

10.4.a. Optimal glucose control: Insulin therapy March 2013 10.4.a. Optimal glucose control: Insulin therapy March 2013 2013 Recommendation: Based on 26 level 2 studies, we recommend that hyperglycemia (blood sugars > 10 mmol/l) be avoided in all critically ill

More information

The Impact of Measurement Frequency on the Domains of Glycemic Control in the Critically Ill-A Monte Carlo Simulation

The Impact of Measurement Frequency on the Domains of Glycemic Control in the Critically Ill-A Monte Carlo Simulation 566507DSTXXX10.1177/1932296814566507Journal of Diabetes Science and TechnologyKrinsley et al research-article2015 Original Article The Impact of Measurement Frequency on the Domains of Glycemic Control

More information

Relationship between glucose meter error and glycemic control efficacy

Relationship between glucose meter error and glycemic control efficacy Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic

More information

What Should Be the Therapeutic Glycemic Target in Intensive Care Units?

What Should Be the Therapeutic Glycemic Target in Intensive Care Units? What Should Be the Therapeutic Glycemic Target in Intensive Care Units? Irl B. Hirsch, M.D. Professor of Medicine University of Washington School of Medicine Disclosures Research/Grants: Sanofi, Halozyme

More information

Effects of two different levels of computerized decision support on blood glucose regulation in critically ill patients

Effects of two different levels of computerized decision support on blood glucose regulation in critically ill patients i n t e r n a t i o n a l j o u r n a l o f m e d i c a l i n f o r m a t i c s 8 1 ( 2 0 1 2 ) 53 60 j ourna l homepage: www.ijmijournal.com Effects of two different levels of computerized decision support

More information

Normal glucose values are associated with a lower risk of mortality in hospitalized patients

Normal glucose values are associated with a lower risk of mortality in hospitalized patients Diabetes Care Publish Ahead of Print, published online August Hyperglycemia 20, 2008 in hospital Normal glucose values are associated with a lower risk of mortality in hospitalized patients Alberto Bruno

More information

Control of Blood Glucose in the ICU: Reconciling the Conflicting Data

Control of Blood Glucose in the ICU: Reconciling the Conflicting Data Control of Blood Glucose in the ICU: Reconciling the Conflicting Data Steven E. Nissen MD Disclosure Consulting: Many pharmaceutical companies Clinical Trials: AbbVie, Amgen, Astra Zeneca, Esperion, Eli

More information

VANDERBILT UNIVERSITY MEDICAL CENTER MULTIDISCIPLINARY SURGICAL CRITICAL CARE TRAUMA INTENSIVE CARE UNIT GLYCEMIC CONTROL PROTOCOL

VANDERBILT UNIVERSITY MEDICAL CENTER MULTIDISCIPLINARY SURGICAL CRITICAL CARE TRAUMA INTENSIVE CARE UNIT GLYCEMIC CONTROL PROTOCOL VANDERBILT UNIVERSITY MEDICAL CENTER MULTIDISCIPLINARY SURGICAL CRITICAL CARE TRAUMA INTENSIVE CARE UNIT GLYCEMIC CONTROL PROTOCOL Background: For some time, the presence of diabetes and hyperglycemia

More information

Management of Inpatient Hyperglycemia: 2011 Endocrine Society Meeting Hyperglycemia in Critically ill patients in ICU Settings.

Management of Inpatient Hyperglycemia: 2011 Endocrine Society Meeting Hyperglycemia in Critically ill patients in ICU Settings. Management of Inpatient Hyperglycemia: 2011 Endocrine Society Meeting Guillermo E. Umpierrez, MD, Emory University School of Medicine and Jack Leahy, MD, University of Connecticut Hyperglycemia in Critically

More information

Glucose control in the ICU using continuous glucose monitoring: What level of the measurement error is acceptable?

Glucose control in the ICU using continuous glucose monitoring: What level of the measurement error is acceptable? Glucose control in the ICU using continuous glucose monitoring: What level of the measurement error is acceptable? Malgorzata E Wilinska, PhD and Roman Hovorka, PhD Wellcome Trust-MRC Institute of Metabolic

More information

Journal Club ICU

Journal Club ICU Journal Club 2018.9.4 ICU Crit Care Med. 2018 Aug;46(8):1224-1229 Introduction Stress hyperglycemia SH SH >124 mg/dl or >200 mg/dl Lancet 2009;373:1798. stress hyperglycemia Lancet 2009;373:1798. SH Critical

More information

Vijayaprasad Gopichandran, Shriraam Mahadevan, Latha Ravikumar, Gomathy Parasuraman, Anjali Sathya, Bhuma Srinivasan, Usha Sriram

Vijayaprasad Gopichandran, Shriraam Mahadevan, Latha Ravikumar, Gomathy Parasuraman, Anjali Sathya, Bhuma Srinivasan, Usha Sriram Original Article Assessment of knowledge, attitudes and practices about tight glycemic control in the critically ill among endocrinologists and intensivists practicing in Chennai Vijayaprasad Gopichandran,

More information

Hypoglycemia Detection in Critical Care Using Continuous Glucose Monitors: An in Silico Proof of Concept Analysis

Hypoglycemia Detection in Critical Care Using Continuous Glucose Monitors: An in Silico Proof of Concept Analysis Journal of Diabetes Science and Technology Volume 4, Issue, January 200 Diabetes Technology Society SYMPOSIUM Hypoglycemia Detection in Critical Care Using Continuous Glucose Monitors: An in Silico Proof

More information

A Children s Bedtime Story

A Children s Bedtime Story A Children s Bedtime Story Setting: University Medical Center, Big Town, USA Scenario: 0500, last admission of the night, 10 previous admissions, all tucked in for the night Patient: 75 year old male with

More information

Hypoglycemia Detection in Critical Care Using Continuous Glucose Monitors: An in silico proof of concept analysis

Hypoglycemia Detection in Critical Care Using Continuous Glucose Monitors: An in silico proof of concept analysis Hypoglycemia Detection in Critical Care Using Continuous Glucose Monitors: An in silico proof of concept analysis Christopher G. Pretty 1, J Geoffrey CHASE 2, Aaron LE COMPTE 2, Geoffrey M. SHAW 3 and

More information

NIH Public Access Author Manuscript J Hosp Med. Author manuscript; available in PMC 2013 August 05.

NIH Public Access Author Manuscript J Hosp Med. Author manuscript; available in PMC 2013 August 05. NIH Public Access Author Manuscript Published in final edited form as: J Hosp Med. 2010 October ; 5(8): 432 437. doi:10.1002/jhm.816. A Comparison Study of Continuous Insulin Infusion Protocols in the

More information

Artificial Pancreas Device System (APDS)

Artificial Pancreas Device System (APDS) Medical Policy Manual Durable Medical Equipment, Policy No. 77 Artificial Pancreas Device System (APDS) Next Review: October 2019 Last Review: October 2018 Effective: November 1, 2018 IMPORTANT REMINDER

More information

Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from Miscoded Meters

Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from Miscoded Meters Journal of Diabetes Science and Technology ORIGINAL ARTICLES Volume 1, Issue 2, March 2007 Diabetes Technology Society Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from

More information

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic

More information

Glucose control positively influences patient outcome: a retrospective study

Glucose control positively influences patient outcome: a retrospective study Glucose control positively influences patient outcome: a retrospective study Penning, Sophie (MSc) 1 : Sophie.Penning@ulg.ac.be (FRS-FNRS research fellow) Pretty, Chris (PhD) 2 : chris.pretty@canterbury.ac.nz

More information

April Dear (Editor):

April Dear (Editor): April 2014 Dear (Editor): Registered Dietitians (RD) play an integral role in patient care in the medical intensive care unit. RD s have increased knowledge in blood glucose control and reducing adverse

More information

Insulin sensitivity, its variability and glycaemic outcome:

Insulin sensitivity, its variability and glycaemic outcome: MCBMS 2009 Insulin sensitivity, its variability and glycaemic outcome: A model-based analysis of the difficulty in achieving tight glycaemic control in critical care JG Chase et al Dept of Mechanical Engineering

More information

ESPEN Congress Madrid 2018

ESPEN Congress Madrid 2018 ESPEN Congress Madrid 2018 Dysglycaemia In Acute Patients With Nutritional Therapy Mechanisms And Consequences Of Dysglycaemia In Patients Receiving Nutritional Therapy M. León- Sanz (ES) Mechanisms and

More information

and ICU - an update Michal Horácek

and ICU - an update Michal Horácek Glycemic control in perioperative period and ICU - an update Michal Horácek 73 The three Leuwen studies (1-3), especially the first one published in 2001 (1), caused a revolution in the approach to glucose

More information

VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS - A PROSPECTIVE COHORT STUDY

VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS - A PROSPECTIVE COHORT STUDY Diabetes Care Publish Ahead of Print, published online March 11, 2009 Validation of CGM in Children with CF VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS

More information

Computing the Surveillance Error Grid Analysis: Procedure and Examples

Computing the Surveillance Error Grid Analysis: Procedure and Examples 5959DSTXXX.77/99685959Journal of Diabetes Science and TechnologyKovatchev et al research-article Original Article Computing the Surveillance Error Grid Analysis: Procedure and Examples Journal of Diabetes

More information

Therapeutic Effect of Insulin in Reducing Critical Illness; Polyneuropathy and Myopathy in the Pediatric Intensive Care Unit

Therapeutic Effect of Insulin in Reducing Critical Illness; Polyneuropathy and Myopathy in the Pediatric Intensive Care Unit original ARTICLE Therapeutic Effect of Insulin in Reducing Critical Illness; Polyneuropathy and Myopathy in the Pediatric Intensive Care Unit How to Cite this Article: Bilan N, Sadegvand, Ranjbar Sh. Therapeutic

More information

ORIGINAL ARTICLE INTRODUCTION

ORIGINAL ARTICLE INTRODUCTION ORIGINAL ARTICLE Comparison of intensive insulin therapy versus conventional glucose control in traumatic brain injury patients on parenteral nutrition: A pilot randomized clinical trial Seyedeh Neda Mousavi,

More information

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES Accuracy of Roche Accu-Chek Inform Whole Blood Capillary, Arterial, and Venous Glucose Values in Patients Receiving Intensive Intravenous

More information

C CONFERENCIAS MAGISTRALES Vol. 36. Supl. 1 Abril-Junio 2013 pp S61-S68 Management of hyperglycemia in the perioperative patient. 39 th Annual Refresher Course on Anesthesiology and Perioperative Medicine,

More information

Patient safety has become a major global social health problem.

Patient safety has become a major global social health problem. REVIEWS IMAJ VOL 14 april 2012 Glycemic Control in the Intensive Care Unit: Between Safety and Benefit Alexander Samokhvalov MD 1, Raymond Farah MD 2 and Nicola Makhoul MD 1 1 Intensive Care Unit, Western

More information

ACUTE ABDOMEN IN DIABETIC PATIENTS ANALYSIS OF COMPLICATIONS AND MORTALITY

ACUTE ABDOMEN IN DIABETIC PATIENTS ANALYSIS OF COMPLICATIONS AND MORTALITY 2014 ILEX PUBLISHING HOUSE, Bucharest, Roumania http://www.jrdiabet.ro Rom J Diabetes Nutr Metab Dis. 21(4):277-284 doi: 10.2478/rjdnmd-2014-0034 ACUTE ABDOMEN IN DIABETIC PATIENTS ANALYSIS OF COMPLICATIONS

More information

Disclosures. Glycemic Control in the Intensive Care Unit. Objectives. Hyperglycemia. Hyperglycemia. History. No disclosures

Disclosures. Glycemic Control in the Intensive Care Unit. Objectives. Hyperglycemia. Hyperglycemia. History. No disclosures Disclosures Glycemic Control in the Intensive Care Unit No disclosures Jorie Frasiolas, Pharm.D., BCPS Clinical Pharmacy Manager, CTICU NewYork-Presbyterian Hospital Columbia University Medical Center

More information

Harm by hyperglycemia, early (parenteral) nutrition or both? from bed to bench and back

Harm by hyperglycemia, early (parenteral) nutrition or both? from bed to bench and back Harm by hyperglycemia, early (parenteral) nutrition or both? from bed to bench and back G. Van den Berghe MD, PhD Department of Intensive Care Medicine University of Leuven (KU Leuven) Leuven, Belgium

More information

The Realities of Technology in Type 1 Diabetes

The Realities of Technology in Type 1 Diabetes The Realities of Technology in Type 1 Diabetes May 6, 2017 Rosanna Fiallo-scharer, MD Margaret Frederick, RN Disclosures I have no conflicts of interest to disclose I will discuss some unapproved treatments

More information

How Should Glucose Meters. Be Evaluated. For Critical Care

How Should Glucose Meters. Be Evaluated. For Critical Care How Should Glucose Meters Be Evaluated For Critical Care IFCC Working Group GMECC Terms of reference 1 and 2 Submitted to IFCC 10 December 2017 Authors: Cynthia Bowman, WG Chair and editor Laboratory Director,

More information

Use of a Novel Fluorescent Glucose Sensor in Volunteer Subjects with Type 1 Diabetes Mellitus

Use of a Novel Fluorescent Glucose Sensor in Volunteer Subjects with Type 1 Diabetes Mellitus Journal of Diabetes Science and Technology Volume 5, Issue 3, May 2011 Diabetes Technology Society ORIGINAL ARTICLE Use of a Novel Fluorescent Glucose Sensor in Volunteer Subjects with Type 1 Diabetes

More information

Citation for published version (APA): van Hooijdonk, R. T. M. (2015). Blood glucose control and monitoring in the critically ill

Citation for published version (APA): van Hooijdonk, R. T. M. (2015). Blood glucose control and monitoring in the critically ill UvA-DARE (Digital Academic Repository) Blood glucose control and monitoring in the critically ill van Hooijdonk, R.T.M. Link to publication Citation for published version (APA): van Hooijdonk, R. T. M.

More information

Transition of Care in Hospitalized Patients with Hyperglycemia and Diabetes

Transition of Care in Hospitalized Patients with Hyperglycemia and Diabetes Transition of Care in Hospitalized Patients with Hyperglycemia and Diabetes Critically ill patients in the ICU Hospital Non-ICU Settings Home Guillermo E Umpierrez, MD, FACP, FACE Professor of Medicine

More information

What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes?

What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes? What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes? Dr. Fiona Wotherspoon Consultant in Diabetes and Endocrinology Dorset County Hospital Fiona.Wotherspoon@dchft.nhs.uk

More information

US Endocrinology. Volume 5 Extract. Glucose Control in the Critically Ill Patient Utilizing Computerized Intravenous Insulin Dosing

US Endocrinology. Volume 5 Extract. Glucose Control in the Critically Ill Patient Utilizing Computerized Intravenous Insulin Dosing US Endocrinology Volume 5 Extract Glucose Control in the Critically Ill Patient Utilizing Computerized Intravenous Insulin Dosing Samuel E Crockett, MD Associate Professor of Medicine, Department of Medical

More information

QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3?

QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3? 500 1 QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3? WHAT ADDITIONAL DATA ARE NEEDED? AACE/ACE CGM Consensus Conference:

More information

Parenteral Nutrition The Sweet and Sour Truth. From: Division of Endocrinology, Diabetes and Bone Disease Icahn School of Medicine at Mount Sinai

Parenteral Nutrition The Sweet and Sour Truth. From: Division of Endocrinology, Diabetes and Bone Disease Icahn School of Medicine at Mount Sinai ENDOCRINE PRACTICE Rapid Electronic Article in Press Rapid Electronic Articles in Press are preprinted manuscripts that have been reviewed and accepted for publication, but have yet to be edited, typeset

More information

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS)

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS) PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS) Pathways for Future Treatment and Management of Diabetes H. Peter Chase, MD Carousel of Hope Symposium Beverly Hilton, Beverly

More information

Management of arterial lines and blood sampling in intensive care: a threat to patient safety

Management of arterial lines and blood sampling in intensive care: a threat to patient safety Anaesthesia 2013, 68, 1114 1119 Original Article doi:10.1111/anae.12389 Management of arterial lines and blood sampling in intensive care: a threat to patient safety R. A. Leslie, 1 S. Gouldson, 2 N. Habib,

More information

Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial Pancreas System

Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial Pancreas System Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society SYMPOSIUM Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial

More information

In - Hospital Diabetes Care. A review and personal experience

In - Hospital Diabetes Care. A review and personal experience In - Hospital Diabetes Care A review and personal experience Hyperglycemia in the Hospital The Problem Hospitalizations with Diabetes http://www.cdc.gov/diabetes/statistics/dmany/fig1.htm Prevalence of

More information

Benchtop to Bedside to Worldwide: Implementing Model-Based Glycemic Control in Intensive Care

Benchtop to Bedside to Worldwide: Implementing Model-Based Glycemic Control in Intensive Care Benchtop to Bedside to Worldwide: Implementing Model-Based Glycemic Control in Intensive Care Aaron J. Le Compte *, Geoffrey M. Shaw **, J. Geoffrey Chase * * Department of Mechanical Engineering, University

More information

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump Protocol Artificial Pancreas Device Systems (10130) Medical Benefit Effective Date: 04/01/18 Next Review Date: 01/19 Preauthorization Yes Review Dates: 03/15, 03/16, 03/17, 01/18 Preauthorization is required.

More information

Put Your Best Figure Forward: Line Graphs and Scattergrams

Put Your Best Figure Forward: Line Graphs and Scattergrams 56:8 1229 1233 (2010) Clinical Chemistry Put Your Best Figure Forward: Line Graphs and Scattergrams Thomas M. Annesley * There is an old saying that a picture is worth a thousand words. In truth, only

More information

Insulin Pump Therapy in children. Prof. Abdulmoein Al-Agha, FRCPCH(UK)

Insulin Pump Therapy in children. Prof. Abdulmoein Al-Agha, FRCPCH(UK) Insulin Pump Therapy in children Prof. Abdulmoein Al-Agha, FRCPCH(UK) aagha@kau.edu.sa Highlights Evolution of insulin pump Pumps mimics Pancreas Goals of diabetes care What lowers HbA1c Criteria for selection

More information

Making Sense of Glucose Monitoring. My Journey with Glucose Monitoring Over the Last 37 Years 8/7/2017

Making Sense of Glucose Monitoring. My Journey with Glucose Monitoring Over the Last 37 Years 8/7/2017 HSW1 Disclosure to Participants Making Sense of Glucose Monitoring Alison B. Evert, MS, RD, CDE UW Neighborhood Clinics UW Medicine Seattle, WA Conflict of Interest (COI) and Financial Relationship Disclosures:

More information

Intensive Control Insulin-Nutrition-Glucose Model Validated in Critically Ill Patients

Intensive Control Insulin-Nutrition-Glucose Model Validated in Critically Ill Patients Intensive Control Insulin-Nutrition-Glucose Model Validated in Critically Ill Patients Jessica Lin Normy N. Razak Christopher G. Pretty Aaron Le Compte Paul Docherty Jacquelyn D. Parente Geoffrey M. Shaw

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

Postoperative Glucose Control and SCIP Measures. Gorav Ailawadi, MD Chief, Adult Cardiac Surgery University of Virginia April 25, 2015

Postoperative Glucose Control and SCIP Measures. Gorav Ailawadi, MD Chief, Adult Cardiac Surgery University of Virginia April 25, 2015 Postoperative Glucose Control and SCIP Measures Gorav Ailawadi, MD Chief, Adult Cardiac Surgery University of Virginia April 25, 2015 Diabetes in CABG Incidence of Diabetes in cardiac surgery increased

More information

Hypoglycemia a barrier to normoglycemia Are long acting analogues and pumps the answer to the barrier??

Hypoglycemia a barrier to normoglycemia Are long acting analogues and pumps the answer to the barrier?? Hypoglycemia a barrier to normoglycemia Are long acting analogues and pumps the answer to the barrier?? Moshe Phillip Institute of Endocrinology and Diabetes National Center of Childhood Diabetes Schneider

More information

123 Are You Providing Evidence-Based Diabetes Care? - Martin

123 Are You Providing Evidence-Based Diabetes Care? - Martin Donna Martin, DNP, RN, CDE, CMSRN Lewis University Learner will be able to: Identify current inpatient standards of care for patients with diabetes Describe causes of hyperglycemia / hypoglycemia in the

More information

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the

More information

Deepika Reddy MD Department of Endocrinology

Deepika Reddy MD Department of Endocrinology Deepika Reddy MD Department of Endocrinology Management of hyperglycemic crisis Review need for inpatient glycemic control Brief overview of relevant trials Case based review of diabetes management strategies/review

More information

A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults

A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults 704443DSTXXX10.1177/1932296817704443Journal of Diabetes Science and TechnologySinha et al research-article2017 Original Article A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

Validity and Reliability of a Glucometer Against Industry Reference Standards

Validity and Reliability of a Glucometer Against Industry Reference Standards 514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference

More information

The oral meal or oral glucose tolerance test. Original Article Two-Hour Seven-Sample Oral Glucose Tolerance Test and Meal Protocol

The oral meal or oral glucose tolerance test. Original Article Two-Hour Seven-Sample Oral Glucose Tolerance Test and Meal Protocol Original Article Two-Hour Seven-Sample Oral Glucose Tolerance Test and Meal Protocol Minimal Model Assessment of -Cell Responsivity and Insulin Sensitivity in Nondiabetic Individuals Chiara Dalla Man,

More information

Point accuracy and reliability of an interstitial continuous glucose-monitoring device in critically ill patients: a prospective study

Point accuracy and reliability of an interstitial continuous glucose-monitoring device in critically ill patients: a prospective study van Hooijdonk et al. Critical Care (2015) 19:34 DOI 10.1186/s13054-015-0757-4 RESEARCH Open Access Point accuracy and reliability of an interstitial continuous glucose-monitoring device in critically ill

More information

Clinical Value and Evidence of Continuous Glucose Monitoring

Clinical Value and Evidence of Continuous Glucose Monitoring Clinical Value and Evidence of Continuous Glucose Monitoring 9402313-012 Objective To review the clinical value and the recent clinical evidence for Professional and Personal CGM Key Points CGM reveals

More information

Case Discussion in Blood Glucose Variability Charles C. Reed, Randy D. Beadle, Susan D. Gerhardt, Gail L. Kongable, Ronald M.

Case Discussion in Blood Glucose Variability Charles C. Reed, Randy D. Beadle, Susan D. Gerhardt, Gail L. Kongable, Ronald M. 70 Case Discussion in Blood Glucose Variability Charles C. Reed, Randy D. Beadle, Susan D. Gerhardt, Gail L. Kongable, Ronald M. Stewart ABSTRACT Stress-induced hyperglycemia has been associated with poor

More information

Method Precision and Frequent Causes of Errors Observed in Point-of-Care Glucose Testing. A Proficiency Testing Program Perspective

Method Precision and Frequent Causes of Errors Observed in Point-of-Care Glucose Testing. A Proficiency Testing Program Perspective Method Precision and Frequent Causes of Errors Observed in Point-of-Care Testing A Proficiency Testing Program Perspective Berna Aslan, MD, 1 Julia Stemp, ART, 1 Paul Yip, PhD, 2 and Jane Gun-Munro, MLT,

More information

Diabetes: Definition Pathophysiology Treatment Goals. By Scott Magee, MD, FACE

Diabetes: Definition Pathophysiology Treatment Goals. By Scott Magee, MD, FACE Diabetes: Definition Pathophysiology Treatment Goals By Scott Magee, MD, FACE Disclosures No disclosures to report Definition of Diabetes Mellitus Diabetes Mellitus comprises a group of disorders characterized

More information

The Diamond Study: Continuous Glucose Monitoring In Patients on Mulitple Daily Insulin Injections

The Diamond Study: Continuous Glucose Monitoring In Patients on Mulitple Daily Insulin Injections 8/5/217 The Diamond Study: Continuous Glucose Monitoring In Patients on Mulitple Daily Insulin Injections Richard M. Bergenstal, MD Executive Director International Diabetes Center at Park Nicollet Minneapolis,

More information

magic DRAGONS: A Protocol for Accurate Glycaemic Control in General Wards

magic DRAGONS: A Protocol for Accurate Glycaemic Control in General Wards Preprints of the 19th World Congress The International Federation of Automatic Control magic DRAGONS: A Protocol for Accurate Glycaemic Control in General Wards Felicity Thomas,* Hamish Tomlinson,* Angus

More information

Αναγκαιότητα και τρόπος ρύθμισης του διαβήτη στους νοσηλευόμενους ασθενείς

Αναγκαιότητα και τρόπος ρύθμισης του διαβήτη στους νοσηλευόμενους ασθενείς Αναγκαιότητα και τρόπος ρύθμισης του διαβήτη στους νοσηλευόμενους ασθενείς Αναστασία Θανοπούλου Επίκουρη Καθηγήτρια Β Παθολογικής Κλινικής Πανεπιστημίου Αθηνών Διαβητολογικό Κέντρο, Ιπποκράτειο Νοσοκομείο

More information

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim Division of Endocrinology and Metabolism, Samsung Medical Center, Sungkyunkwan University School of Medicine Conflict of interest

More information

Early change in blood glucose concentration is an indicator of mortality in critically ill children

Early change in blood glucose concentration is an indicator of mortality in critically ill children Intensive Care Med (2013) 39:123 128 DOI 10.1007/s00134-012-2738-2 PEDIATRIC ORIGINAL Prabhakar P. Nayak Paul Davies Parth Narendran Simon Laker Fang Gao Stephen C. L. Gough John Stickley Kevin P. Morris

More information

Alarge, randomized, controlled trial

Alarge, randomized, controlled trial Emerging Treatments and Technologies O R I G I N A L A R T I C L E Glucommander A computer-directed intravenous insulin system shown to be safe, simple, and effective in 120,618 h of operation PAUL C.

More information

Glycemia Prediction in Critically Ill Patients Using an Adaptive Modeling Approach

Glycemia Prediction in Critically Ill Patients Using an Adaptive Modeling Approach Journal of Diabetes Science and Technology Volume 1, Issue 3, May 2007 Diabetes Technology Society SYMPOSIUM Glycemia Prediction in Critically Ill Patients Using an Adaptive Modeling Approach Tom, M.Sc.,

More information

Appendix. Lifetime extrapolation of data from the randomised controlled DiGEM trial

Appendix. Lifetime extrapolation of data from the randomised controlled DiGEM trial Appendix Lifetime extrapolation of data from the randomised controlled DiGEM trial Judit Simon, Alastair Gray, Philip Clarke, Alisha Wade, Andrew Neil, Andrew Farmer on behalf of the Diabetes Glycaemic

More information

Report Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1

Report Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1 Report Reference Guide THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1 How to use this guide Each type of CareLink report and its components are described in the following sections.

More information

Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase

Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase Washington University School of Medicine Digital Commons@Becker Open Access Publications 2013 Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase Dennis J. Dietzen

More information

Diabetes Guidelines in View of Recent Clinical Trials Are They Still Applicable?

Diabetes Guidelines in View of Recent Clinical Trials Are They Still Applicable? Diabetes Guidelines in View of Recent Clinical Trials Are They Still Applicable? Jay S. Skyler, MD, MACP Division of Endocrinology, Diabetes, and Metabolism and Diabetes Research Institute University of

More information

Safety and Efficacy of Continuous Insulin Infusion in Noncritical Care Settings

Safety and Efficacy of Continuous Insulin Infusion in Noncritical Care Settings ORIGINAL RESEARCH Safety and Efficacy of Continuous Insulin Infusion in Noncritical Care Settings Dawn Smiley, MD 1 Mary Rhee, MD 1 Limin Peng, PhD 2 Laurian Roediger, BS 1 Patrick Mulligan, BS 1 Lewis

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

GLYCEMIC CONTROL IN CARDIAC SURGERY: IMPLEMENTING AN EVIDENCE-BASED INSULIN INFUSION PROTOCOL. Critical Care Evaluation

GLYCEMIC CONTROL IN CARDIAC SURGERY: IMPLEMENTING AN EVIDENCE-BASED INSULIN INFUSION PROTOCOL. Critical Care Evaluation Critical Care Evaluation GLYCEMIC CONTROL IN CARDIAC SURGERY: IMPLEMENTING AN EVIDENCE-BASED INSULIN INFUSION PROTOCOL By Joelle D. Hargraves, RN, DNP, APN, CCRN, CCNS 2014 American Association of Critical-Care

More information

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes Protocol Artificial Pancreas Device Systems Medical Benefit Effective Date: 07/01/18 Next Review Date: 01/20 Preauthorization Yes Review Dates: 03/15, 03/16, 03/17, 01/18, 05/18, 01/19 Preauthorization

More information

When Will CGM Replace SMBG? Roy W. Beck, MD, PhD. JAEB Center for Health Research Tampa, Florida

When Will CGM Replace SMBG? Roy W. Beck, MD, PhD. JAEB Center for Health Research Tampa, Florida When Will CGM Replace SMBG? Roy W. Beck, MD, PhD JAEB Center for Health Research Tampa, Florida Financial Disclosures Dr. Beck does not have any personal conflicts of interest His employer, the JAEB Center

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

DIABETES MANAGEMENT DISCHARGE COMMUNICATION (DM-DC) AUDIT TOOL

DIABETES MANAGEMENT DISCHARGE COMMUNICATION (DM-DC) AUDIT TOOL DIABETES MANAGEMENT DISCHARGE COMMUNICATION (DM-DC) AUDIT TOOL Facility: Date: Data Collector s name: Email/Phone: Purpose: To evaluate your facility practices regarding communication of requisite diabetes

More information

Carbohydrate Ratio Optimization and Adaptation Algorithm. Supplementary Figure 1. Schematic showing components of the closed-loop system.

Carbohydrate Ratio Optimization and Adaptation Algorithm. Supplementary Figure 1. Schematic showing components of the closed-loop system. Basal Profile Adaptation Algorithm Carbohydrate Ratio Optimization and Adaptation Algorithm Supplementary Figure 1. Schematic showing components of the closed-loop system. The insulin pump and Dexcom G4

More information

Martin J Stevens MD, FRCP, Endocrinologist and Professor of Medicine

Martin J Stevens MD, FRCP, Endocrinologist and Professor of Medicine The Approach to Inpatient Hyperglycemia Martin J Stevens MD, FRCP, Endocrinologist and Professor of Medicine Great Lakes Hospital Medical Symposium May 7th 2010 Further Increases in the Prevalence of Diabetes

More information

Cleveland Clinic Cardiovascular Intensive Care Unit Insulin Conversion Protocol

Cleveland Clinic Cardiovascular Intensive Care Unit Insulin Conversion Protocol Journal of Diabetes Science and Technology Volume 3, Issue 3, May 2009 Diabetes Technology Society ORIGINAL ARTICLES Cleveland Clinic Cardiovascular Intensive Care Unit Insulin Conversion Protocol Leann,

More information

Glargine as a Basal Insulin Supplement in Recovering Critically Ill Patients An In Silico Study

Glargine as a Basal Insulin Supplement in Recovering Critically Ill Patients An In Silico Study Glargine as a Basal Insulin Supplement in Recovering Critically Ill Patients An In Silico Study Normy N. Razak Jessica Lin J. Geoffrey Chase Geoffrey M. Shaw Christopher G. Pretty Aaron Le Compte Fatanah

More information

Endocrine Update Mary T. Korytkowski MD Division of Endocrinology University of Pittsburgh

Endocrine Update Mary T. Korytkowski MD Division of Endocrinology University of Pittsburgh Endocrine Update 2016 Mary T. Korytkowski MD Division of Endocrinology University of Pittsburgh Disclosure of Financial Relationships Mary Korytkowski MD Honoraria British Medical Journal Diabetes Research

More information

1. What s the point of a network the case for research? 2. How to use CSII effectively

1. What s the point of a network the case for research? 2. How to use CSII effectively 1. What s the point of a network the case for research? 2. How to use CSII effectively John Pickup King s College London Faculty of Medicine Guy s Hospital, London What should an insulin pump network do?

More information

Unit 1 Exploring and Understanding Data

Unit 1 Exploring and Understanding Data Unit 1 Exploring and Understanding Data Area Principle Bar Chart Boxplot Conditional Distribution Dotplot Empirical Rule Five Number Summary Frequency Distribution Frequency Polygon Histogram Interquartile

More information

Anomaly Detection in the Artificial Pancreas

Anomaly Detection in the Artificial Pancreas Anomaly Detection in the Artificial Pancreas Avila L., Martínez E. INGAR (CONICET-UTN) Abstract The integration of subcutaneous sensing and insulin delivery technologies with novel control strategies has

More information

Received: 23 September Accepted: 17 October 2009

Received: 23 September Accepted: 17 October 2009 ORIGINAL ARTICLE Comparing Effects of Continuous Insulin Infusion with or without Subcutaneous Glargine Insulin on Glycemic Control in Diabetic Patients Undergoing Coronary Artery Bypass Graft (CABG) Seyed

More information

ADAG Study Group Data Links A1C Levels with Empirically Measured Blood Glucose Values - New Treatment Guidelines Will Now be Needed

ADAG Study Group Data Links A1C Levels with Empirically Measured Blood Glucose Values - New Treatment Guidelines Will Now be Needed 529638DSTXXX10.1177/1932296814529638Journal of Diabetes Science and TechnologyKlonoff research-article2014 Editorial ADAG Study Group Data Links A1C Levels with Empirically Measured Blood Glucose Values

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Continuous Monitoring of Glucose in the Interstitial Fluid File Name: Origination: Last CAP Review: Next CAP Review: Last Review: continuous_monitoring_of_glucose_in_the_interstitial_fluid

More information

Meeting the Challenge of Inpatient Glycemic Management in the Non-Critical Care Setting

Meeting the Challenge of Inpatient Glycemic Management in the Non-Critical Care Setting Meeting the Challenge of Inpatient Glycemic Management in the Non-Critical Care Setting Jane Jeffrie Seley, DNP, MPH, GNP, BC-ADM, CDE, CDTC, FAAN, FAADE Diabetes Nurse Practitioner, Inpatient Diabetes

More information

Evaluation and treatment of hyperglycemia in critically ill patients

Evaluation and treatment of hyperglycemia in critically ill patients Acta Biomed 2016; Vol. 87, N. 3: 329-333 Mattioli 1885 Review - Emergency medicine Evaluation and treatment of hyperglycemia in critically ill patients Andrea Balloni 1, Federico Lari 2, Fabrizio Giostra

More information

What Makes Tight Glycemic Control (TGC) Tight? The impact of variability and nutrition in 2 clinical studies

What Makes Tight Glycemic Control (TGC) Tight? The impact of variability and nutrition in 2 clinical studies What Makes Tight Glycemic Control (TGC) Tight? The impact of variability and nutrition in 2 clinical studies Fatanah Suhaimi 1, BEng; Aaron Le Compte 1, PhD; Jean-Charles Preiser 2, MD, PhD; Geoffrey M

More information