VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

Size: px
Start display at page:

Download "VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP"

Transcription

1 VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP Abstract Background: The VetScan UA urine strip reader is an analyzer designed to provide semi-quantitative and qualitative results of common urinalysis chemistries, including microalbumin (MA) and urine protein creatinine (UPC) ratios at the point of care. Objective: This study assessed the precision and accuracy of the VetScan UA semi-quantitative results against a quantitative reference laboratory analyzer. Methods: The precision study evaluated duplicate runs using a total of 126 feline and canine urine samples. Each sample was run twice on two VetScan UA analyzers using VetScan UA14 strips. Clinical Correlation of Urine Chemistries was completed on 108 of the urine samples (40 feline, 68 canine). Samples were run on a VetScan UA analyzer using VetScan UA14 strips and a Roche Automated strip analyzer using Roche Multistix urine reagent strips. Microalbumin correlation was evaluated by comparing the same 108 samples (40 feline, 68 canine) against the Beckman Coulter 2700 Chemistry analyzer, which performs a turbidimetric immunoassay for albumin. The UPC results of the same samples were evaluated against the Beckman Coulter 2700 Chemistry system. Results 1. Precision: The VetScan UA analyzer demonstrated excellent reproducibility for all 13 of the parameters examined. All results were in agreement with the replicate result for all samples tested 99% of the time or more, when defined as identical results or +/- one color block. 2. Chemistry Correlation: Direct comparisons of the VetScan UA results against an automated Roche strip reader for KET, BIL, GLU, PRO, ph, and BLD demonstrated excellent agreement for both canine and feline samples (total n=108). Agreement with the reference method for these results was: 3. Proteinuria Correlation: As proteinuria is a critical indicator of both feline and canine systemic disease, the VetScan microalbumin (MA) and urine protein creatinine ratio (UPC) were evaluated against reference methods. a. Microalbumin (MA) Evaluation: The VetScan MA parameter displayed outstanding agreement with a reference laboratory Beckman Coulter 2700 turbid metric MA assay: b. Urine Protein Creatinine Ratio (UPC) Evaluation: The UPC on the VetScan UA showed excellent accuracy compared to a Beckman Coulter 2700 UPC as follows: Conclusion: All performance indicators examined suggest that the VetScan UA provides excellent precision and accuracy for point-of-care urinalysis chemistry testing. Abaxis VetScan products are intended for veterinary use only. Abaxis and VetScan are registered trademarks of Abaxis, Inc. Abaxis 2018.

2 VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP Introduction A urinalysis is an important component of a complete clinical diagnostic workup that includes: urine chemistries, gross observations of the urine color and clarity, measurement of specific gravity and an examination of the urinary sediment. The urinary chemistries are important indicators for renal disease, diabetes mellitus, hyperadrenocorticism, hypoadrenocorticism, urinary tract infections and neoplasia. In most veterinary clinic settings, urine chemistries are tested using an inexpensive, disposable urine reagent strip. These urine strips contain a set of absorbent pads, each embedded with chemicals that react with urine components to provide colorimetric changes designed to either be interpreted by the user against a color chart or in some cases measured by an instrument. Analytes measured typically include: bilirubin (BIL), glucose (GLU), protein (PRO), specific gravity (SG), ph, and blood/hemoglobin (BLD). The analyte concentrations are derived from color changes on the pads that are induced by chemical reactions with the analytes. For traditional urine test strips, analyte concentrations are determined through a color comparison of the exposed strip to a reference color chart. The advantages of such a test method are that the testing is simple and no instrumentation is required. However, this traditional testing process contains a number of sources for potential variability in the results, including: subjective interpretation of the color shade, variability of lighting in the room, the amount of urine left on the strip during color development, potential contamination from other pads during color development, the presence of interfering substances (such as blood/hemoglobin), and the amount of time the strip is exposed to urine prior to the color chart interpretation. Each pad has a specific reaction time after which the pad should be read. On a single strip, 3-4 different optimal reaction times may be present. Reading the strip at an inappropriate time, or a delay in reading the strip, can lead to under or overdevelopment of the chemical color change on the pads and thus lead to inaccurate results. Due to the variability inherent to the traditional urine strips testing process, all urine strips report semi-quantitative results, defined as reporting results in a series of ranges or bins to represent the entire range of concentrations. Each bin represents an analyte concentration range (for example from 45 to 88 units, see Figure 1) and is named with both a symbol and numerical concentration value, such as +2 and 70 mg/dl as in the example in Figure 1. The exact color gradation on a given test pad is used to calculate the analyte concentration via an algorithm determined by the test strip manufacturer. Any test run whose calculated concentration falls within the range of a bin will have a reported result corresponding to that bin s symbol and numerical concentration value. Any two calculated results falling within a bin are reported to the user as the same value, despite the fact that the actual colors of the two pads are slightly different, as shown in the example of Pair A in Figure. Analyte Concentration (mg/dl) Figure 1. Semi-Quantitative Results. The graph illustrates how semi-quantitative results are generated for five bins. Pair A shows two samples calculated to have a concentration of 50 and 86, which will both be reported in the same bin (70 ml/dl or 2+) to the user. Pair B illustrates how two close test results can sometimes fall into separate bins. The precision of semi-quantitative results are typically evaluated by having acceptable limits of either one bin over or one bin below the actual bin. This concept is illustrated by Pair B in Figure 1, where the first calculated result is 86 units, which falls within the bin of 70, and thus reported to the user as 70 or 2+. If a replicate test is run on the same sample with a new strip, the second result may be 91 and thus reported as 125 or 3+. The actual colors of the pads in the two runs and the actual concentrations of analytes are very close to each other despite the names of the two bins, 70 and 125, appearing to be far apart. Abaxis has introduced the VetScan UA analyzer system that removes much of the variability in urine chemistry strip testing through an automated scanning of test strips at precise times to prevent under or overdevelopment and subjectivity of evaluating the color change of the 1 of 5

3 pads. The VetScan UA analyzer also provides the ability to semi-quantitatively measure microalbumin (MA) levels and calculate a urine protein creatinine ratio (UPC). Both MA and UPC are indicators of proteinuria, which is an important prognostic indicator and monitoring tool for renal disease in both dogs and cats (1) (2). MA levels above 2.5 mg/dl (25 mg/l) suggest that glomerular protein leakage is clinically significant. While a single high MA result may occur due to transient physiological changes, such as an acute inflammatory process, repeated high MA levels over a period of weeks may be suggestive of early renal disease or a chronic inflammatory condition requiring further diagnostic investigation (3). The UPC provides a good method to quantify renal proteinuria and is a key indicator for staging a dog or cat for chronic kidney disease (3). UPC values above 0.4 (feline) or 0.5 (canine) indicate leakage of protein from the kidneys at levels which may indicate renal dysfunction, and require monitoring for potential progression of renal disease (3). A UPC over 1.0 requires diagnostic and potentially therapeutic intervention, and a UPC over 2.0 is usually associated with overt glomerular proteinuria (3). This study examines the accuracy and precision of the VetScan UA analyzer against reference laboratory analyzers for traditional urine chemistries on feline and canine urine samples. As the VetScan UA tests for both MA and UPC on its UA14 urine test strip, this study also examined how the VetScan UA UPC and MA compared to a reference laboratory analyzer. Materials and Methods Calibration: Analyzers were checked for correct calibration using VetScan UA controls and VetScan UA Check strips daily prior to data collection. Precision: The precision study evaluated duplicate sample runs using a total of 126 urine samples that included both canine and feline samples. Samples were run twice on each of two VetScan UA analyzers using VetScan UA14 strips. Replicate runs were run immediately after the first run on each sample. Sample runs were considered to be repeatable if the results of the two duplicate runs were within one color block (either up or down, but not both) of the replicate sample results. The number of identical results and results +/- one block between the two replicate runs on both analyzers were tallied and calculated as a percentage of the total sample runs. Clinical Correlation of Urine Chemistries: 108 urine samples (40 feline, 68 canine) were run on a VetScan UA analyzer using VetScan UA14 strips and a Roche Automated strip analyzer using Roche Multistix urine reagent strips. Samples were run on either the Roche analyzer or VetScan UA analyzer within one hour of running on the other analyzer. The VetScan UA urine chemistry results were compared to those results for the following parameters: KET, BIL, GLU, PRO, ph, and BLD. Data for LEU, NIT, URO, ASC and Ca were collected for information purposes only and not evaluated in this study since the parameters are not relevant in veterinary medicine. The SG measurement was also not evaluated, as SG evaluation on a refractometer is recommended. The results were rated in similar bins to facilitate comparison. The VetScan UA results were then compared as to whether they were in agreement ( IA for In Agreement, defined as identical or within one bin of the Roche results). Results greater than one bin away were classified as out of range (OOR). The numbers of samples in agreement were described as a percentage of the total number of samples. Microalbumin (MA) Evaluation: To evaluate the accuracy of the microalbumin (MA) results of the VetScan UA analyzer, the results from 108 urine samples (40 feline, 68 canine) all performed on the same analyzer were compared to results from a reference laboratory Beckman Coulter 2700 Chemistry analyzer, which performs a turbidimetric immunoassay for albumin. Since the VetScan UA analyzer uses a threshold of 2.5 mg/dl (25 g/l) to indicate a positive MA result, the quantitative Beckman MA results were rated as to whether the results were greater than 2.5 mg/dl (25 g/l) and compared with the VetScan UA results. Information on the health of each patient was not available for this study. Urine Protein Creatinine Ratio (UPC) Evaluation: The UPC results of the VetScan UA analyzer were evaluated against a reference laboratory Beckman Coulter 2700 Chemistry analyzer. 108 samples (40 cat samples, 68 dog samples) were run between the VetScan UA and Beckman Coulter systems. The UPC results on the VetScan UA analyzer were categorized into 3 bins for cats and dogs, respectively: between 0.0 and 0.4/0.5, between 0.4/0.5 and 2.0, and >2.0. The 0.4/0.5 UPC threshold was selected according to the International Renal Interest Society Chronic Kidney Disease staging guidelines (3). These were matched against the results of the Beckman Coulter analyzer and then categorized as false positive or false negative to generate the sensitivity and specificity ratings. Statements relating to UPC thresholds of 0.5 are intended to include feline samples for which the threshold is 0.4. Results Precision: The VetScan UA analyzer demonstrated excellent reproducibility for all 13 of the parameters examined. All results were in agreement with the replicate result for all samples tested 99% of the time or more, when defined as identical results or +/- one color block (Table 1). This held true across all 13 parameters. Identical results were seen between the replicate runs, ranging from 75% to 100% of the samples. For example, 100.0% of the replicate PRO results were in agreement, with 91.3% of results being identical between the pairs and 8.7% being +/- one color block from the other replicate. Table 1. VetScan UA Precision on VetScan UA14 Strips Urine Samples 2 of 5

4 Clinical Correlation General Urine Chemistries: The VetScan UA results compared very well to the Roche automated strip analyzer for the urine chemistries. For the most critical analytes shared between the VetScan UA and the Roche analyzers(bil, GLU, PRO, ph and BLD) the percent of samples in agreement (IA) were 100%, 99%, 93%, 90% and 92%, respectively (Table 2). Only 0.9% of samples on the VetScan UA analyzer were out of range (OOR) compared to GLU results from the Roche. Out of range results ranged from 7.4%, 10.2% and 8.3% for PRO, ph and BLD, respectively. The VetScan UA demonstrated complete agreement with the Roche analyzer for all BIL results tested. results being within the healthy range, when the reference result was in a higher bin. Performance for UPC on feline and canine samples were similar in the percentage of false positives and negatives (data not shown). Table 2. Agreement between VetScan UA Chemistries and a Roche Automated Strip Reader (n=109 samples, run on two analyzers, total of 216 comparisons) The VetScan UA demonstrated excellent sensitivity (87%) as it detected 65 of 75 samples as non-proteinuric, as defined by a UPC value below 0.4 (feline) or 0.5 (canine). The VetScan UA analyzer also provided results in agreement in 22 of 28 samples in the borderline proteinuric category. The VetScan UA results were in 100% agreement for the 5 samples that showed a UPC above 2.0 on the reference analyzer. Microalbumin (MA): The VetScan UA MA results were compared to the reference analyzer for 108 samples matched the reference analyzer extremely well. 85% of the samples were in agreement between the two analyzers as either being above or below the 2.5 mg/dl (25 mg/l) MA threshold which indicates a positive test for microalbuminuria. Of the 51 MA positive samples per the reference analyzer (MA>2.5 mg/dl), MA was detected in 49 samples by the VetScan UA analyzer (sensitivity 96.23%), showing that the VetScan UA accurately indicated protein leakage (Table 3). The VetScan UA produced 12 false positives in 108 samples for a specificity of 82.61%. The VetScan UA only produced 2 false negatives in this sample set. Performances for MA on feline and canine samples were similar in the percentage of false positives and negatives (data not shown). Table 3. Evaluation of Microalbumin (MA) Results vs Reference Analyzer. FP, FN are False Positives and False Negative, respectively. Table 4. Evaluation of UPC Results vs Reference Analyzer. Clinical Use of MA and UPC The clinical utility of the microalbumin (MA) and urine protein creatinine ratio (UPC) were further examined in conjunction with each other. Microalbuminuria indicates the early leakage of micro-amounts of protein into the urine, usually preceding a rise in UPC value. Healthy animals should not have any proteinuria (MA<2.5, UPC<0.5). In early renal disease, microalbumin leakage will begin to increase before UPC reaches significant levels (MA+, UPC<0.5). As disease develops, the UPC levels will begin to rise above 0.5 and the MA readout loses its diagnostic value as renal disease has become overt and can be detected by UPC. To examine how the Vetscan UA UPC measurements performed relative to a reference analyzer regarding the correlation of UPC and MA, they were plotted against the reference analyzer MA for the same sample. As expected, samples with MA of less than 2.5 mg/dl (25 mg/l) corresponded with a UPC value lower than 0.5 (Figure 2). Four samples on the VetScan UA resulted in a UPC between 0.5 and 2.0 with MA < 0.5. Three of those samples resulted in UPC levels on the reference analyzer of 0.1 and one had a UPC of 0.2 (Table 5). Thus, the UPC reference analyzer results for all four samples were borderline to an adjacent bin and were only one bin away from the results seen on the VetScan UA. Table 5. Reference Analyzer Results (Beckman Coulter 2700 Chemistry Analyzer, BC) for Four MA-UPC Level 3 ( ) Results Urine Protein Creatinine Ratio (UPC): The VetScan UA UPC results showed very good agreement with the results of the reference analyzer. 92% of VetScan UA UPC results were in exact agreement or within one bin of the reference analyzer result. Previous studies and IRIS guidelines have determined that healthy cats and dogs have UPC ratios below 0.4 or 0.5, respectively (3). The number of results from the two analyzers falling above and below that critical UPC limit (0.4 or 0.5) (3) was examined (see Table 4). (Henceforth, references to the 0.5 UPC for dogs will assume the feline UPC threshold of 0.4 as well.) False Positives (FP) are defined as Vetscan UA results reading above 0.5 when the reference result was below 0.5 and False Negatives (FN) are defined as VetScan UA 3 of 5 As MA levels increased, the UPC typically increased, as expected. Eighteen samples displayed positive MA results with UPC<0.5 on the reference analyzers, but only at the lower end of MA concentrations. The presence of such MA positive, UPC <0.5 samples is consistent with previous studies (4). All of these observations are consistent with the expectations of the development of renal disease.

5 Examining the data from the reference analyzers in a similar manner produced matching results with the VetScan UA data, indicating good clinical agreement between high end reference analyzers and the VetScan UA when both UPC and MA are considered together (Figure 3). An alternative graphical comparison of the MA and UPC results was examined to determine if the VetScan UA produced equivalent diagnostic results when both parameters were considered simultaneously. Figure 4 shows that the VetScan UA produced results that closely matched the reference analyzer. The number of samples above the clinically significant UPC limit of 0.5 was plotted when the samples had MA>2.5 mg/dl (MA+) or MA<2.5 mg/dl (MA- ). The VetScan UA closely matched the number of samples on the reference analyzers for MA+, UPC , in addition to MA+, UPC>2.0. This data indicates that the VetScan UA agrees well with reference analyzers when the MA and UPC parameters are evaluated together for clinical diagnosis of proteinuria. The VetScan UA results labeled more samples as MA+, UPC<0.5 than the reference analyzers did (Figure 4B), but a closer examination of these showed that the VetScan UA differed from the Beckman Coulter results by only one bin in all but two of these cases and thus were considered within agreement levels. In addition, any MA+ result should be repeated in 2-3 weeks prior to making a diagnosis. Conclusions Urine chemistry testing is an important point-of-care diagnostic tool for screening and monitoring renal disease in veterinary patients. The VetScan UA was evaluated for precision and accuracy in traditional urine chemistries as well as two newly available point-of-care parameters, microalbumin and the urine protein creatinine ratio. The VetScan UA showed excellent precision on results from the UA14 strips for the most commonly tested urine chemistry analytes, with matching results in 99% of replicates. The Vetscan UA also demonstrated excellent accuracy compared to a reference automated urine strip reader for all the analytes tested, including MA and UPC. Despite the fact that the reference analyzers provide quantitative results, while the VetScan UA produces semi-quantitative results, the results correlated well. For the two novel parameters on a urine test strip, both microalbumin and urine protein creatinine ratios showed very good agreement with the reference analyzer results. Data indicates that the VetScan UA generated very few false negative MA results, and thus is an effective screening tool for low level proteinuria and early stage renal disease. When MA and UPC on the VetScan UA are used in combination, they provide a powerful screening tool for renal disease that correlates well with results from reference laboratory analyzers. References: 1. Survival of cats with naturally occurring chronic renal failure is related to severity of proteinuria. Syme HM, Markwell PJ, Pfeiffer D, Elliott J. 3, May-Jun 2006, J Vet Intern Med., Vol. 20, pp Evaluation of the association between initial proteinuria and morbidity rate or death in dogs with naturally occurring chronic renal failure. Jacob F, Polzin DJ, Osborne CA, Neaton JD, Kirk CA, Allen TA, Swanson LL. 3, Feb 1, 2005, J Am Vet Med Assoc., Vol. 226, pp IRIS CKD Staging Guidelines. [Online] Nov 17, [Cited: Nov 17, 2017.] 4. Comparison of a semiquantitative point-of-care assay for the detection of canine microalbuminuria with routine semiquantitative methods for proteinuria. Garner, BC, Wiedmeyer, CE. 3, s.l. : Vet Clin Path, 2007, Vol. 36, pp Figure 2. Combination of UPC and MA in Clinical Diagnosis for the VetScan UA. UPC results on the UA were categorized into 4 levels (Level 1, 2, 3 and 4 corresponding to <0.2, , , >2.0, respectively, and graphed versus MA on the reference analyzer (Beckman Coulter 2700). Dashed lines indicate clinical thresholds of 0.4 (feline)/0.5 (canine) UPC and 2.5 mg/dl MA for felines and canines. 4 of 5

6 Figure 3. Combination of UPC and MA in Clinical Diagnosis for the reference laboratory analyzer (Beckman Coulter 2700). Dashed lines indicate the clinical thresholds of 0.4 (feline)/0.5 (canine) UPC and 2.5 mg/dl MA for felines and canines.. (A) No. of Samples < >2.0 UPC UA MA- Ref MA- (B) No. of Samples < >2.0 UPC UA MA+ Ref MA+ Figure 4. The Diagnostic Combination of UPC and MA on the VetScan UA. UPC and MA closely matches those on reference lab chemistry analyzer. (A) Plot of number of MA negative samples versus UPC results on the VetScan UA and reference analyzer (Beckman Counter 2700) shows similar numbers of samples in each category for this set of samples. (B) Plot of number of MA positive samples versus UPC results on the VetScan UA and reference analyzers (Beckman Counter 2700 and Roche analyzers) shows similar numbers of samples in each category for this set of samples. 5 of 5 Abaxis VetScan products are intended for veterinary use only. Abaxis and VetScan are registered trademarks of Abaxis, Inc. Abaxis Rev. B

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k063276 B. Purpose for Submission: New device C. Measurand: Urinary Glucose, Blood,

More information

Evaluation of VACUETTE Urine CCM Tubes for Urine Strip Testing

Evaluation of VACUETTE Urine CCM Tubes for Urine Strip Testing Evaluation of VACUETTE Urine CCM Tubes for Urine Strip Testing Background The VACUETTE Urine is for the collection, transport and storage of urine samples for urine culture and urinalysis in the laboratory.

More information

1. Staging of CKD based on blood creatinine concentration

1. Staging of CKD based on blood creatinine concentration 1. Staging of CKD based on blood creatinine concentration Staging is undertaken following diagnosis of chronic kidney disease (CKD) in order to facilitate appropriate treatment and monitoring of the patient.

More information

Role of Quantitative MA Testing in Everyday Practice

Role of Quantitative MA Testing in Everyday Practice Print PDF Save PDF Sign Up for Newsletter Past Editions August/September 2012 Focus: Renal Disease Diagnostics Role of Quantitative MA Testing in Everyday Practice Interview with Brenda Harai, DVM, MS,

More information

Role of Quantitative MA Testing in Everyday Practice

Role of Quantitative MA Testing in Everyday Practice Print PDF Save PDF Sign Up for Newsletter Past Editions August/September 2012 Focus: Renal Disease Diagnostics Role of Quantitative MA Testing in Everyday Practice Interview with Brenda Harai, DVM, MS,

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

Taking a dip into urinalysis

Taking a dip into urinalysis Vet Times The website for the veterinary profession https://www.vettimes.co.uk Taking a dip into urinalysis Author : Christine Jameison Categories : RVNs Date : July 1, 2009 Christine Jameison RVN, probes

More information

Association of microalbuminuria and the urine albumin-to-creatinine ratio with systemic disease in cats

Association of microalbuminuria and the urine albumin-to-creatinine ratio with systemic disease in cats Association of microalbuminuria and the urine albumin-to-creatinine ratio with systemic disease in cats Jacqueline C. Whittemore, dvm, dacvim; Zona Miyoshi, dvm; Wayne A. Jensen, dvm, phd, mba; Steven

More information

Urinalysis (Macroscopic Physical Tests) Urinalysis. Basic Urinalysis 12/22/2012. Mohammad Reza Bakhtiari DCLS, PhD

Urinalysis (Macroscopic Physical Tests) Urinalysis. Basic Urinalysis 12/22/2012. Mohammad Reza Bakhtiari DCLS, PhD Urinalysis (Macroscopic Physical Tests) Mohammad Reza Bakhtiari DCLS, PhD Urinalysis Basic Urinalysis CLSI: U/A is the testing of urine with procedures commonly performed in an expeditious, reliable, accurate,

More information

The Minimum Diagnostic Database: Urinalysis

The Minimum Diagnostic Database: Urinalysis The Minimum Diagnostic Database: Urinalysis Jeff Niziolek, DVM Professional Services Veterinarian IDEXX Laboratories, Inc. 208 Bay Meadows Drive Holland, MI 49424 The minimum database includes three types

More information

cobas u 601 urine analyzer Providing precise and secure test results

cobas u 601 urine analyzer Providing precise and secure test results Providing precise and secure test results 1 of 6 Providing precise and secure test results cobas u 601 urine analyzer is a modular, analytical system platform for urine strip testing and features a fully

More information

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000 v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination

More information

CLINITEK Novus Automated Urine Chemistry Analyzer

CLINITEK Novus Automated Urine Chemistry Analyzer CLINITEK, Atlas, Novus, Status+, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks are properties of their respective owners. Product

More information

Semi-Automatedvs. Visual Readingof UrinalysisDipsticks

Semi-Automatedvs. Visual Readingof UrinalysisDipsticks CLIN. CHEM. 2/2, 222-22 () Semi-Automatedvs. Readingof UrinalysisDipsticks James D. Peeie, Jr., Richard H. Gadsden, and Rachel Crews Semi-automated reflectance readings of urinalysis dipsticks showed better

More information

HM5. Hematology Analyzer BETTER. ACTUALLY.

HM5. Hematology Analyzer BETTER. ACTUALLY. HM5 Hematology Analyzer BETTER. ACTUALLY. Advanced Hematology Five-Part Differential The VetScan HM5 is a fully automated five-part differential hematology analyzer displaying a comprehensive 24-parameter

More information

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Gentian Canine CRP Immunoassay Application Note for scil VitroVet* v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of

More information

A. SAP is the D-Lab's name for a specific set of serum biochemical tests.

A. SAP is the D-Lab's name for a specific set of serum biochemical tests. Understanding CBC, SAP, UA/Laura J. Steadman, DVM I. CBC - Complete Blood Count A. Three major types of cells are counted 1. Red Blood Cells 2. White Blood Cells 3. Platelets B. Cells are counted at the

More information

It s not just water! What is Urinalysis?

It s not just water! What is Urinalysis? It s not just water! An introduction to Urinalysis What is Urinalysis? Urinalysis or the analysis of urine is one of the oldest laboratory procedures in the practice of medicine. It is a good test for

More information

Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity and Effect of Interferent

Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity and Effect of Interferent ORIGINAL ARTICLE Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity

More information

Clinical Test Report. of DUS10 (Urine Reagent Strips) Effective Date: April DFI Lab. Dong-Ai Hospital Medical Center: Clinical Pathology

Clinical Test Report. of DUS10 (Urine Reagent Strips) Effective Date: April DFI Lab. Dong-Ai Hospital Medical Center: Clinical Pathology Page 1 / 8 APPROVED Clinical Test Report of DUS10 (Urine Reagent Strips) Effective Date: April. 13. 2010 DFI Lab Dong-Ai Hospital Medical Center: Clinical Pathology Page 2 / 8 The evaluation followed Method

More information

Advanced Hematology Five-Part Differential. Simple Operation

Advanced Hematology Five-Part Differential. Simple Operation HematologyAnalyzer Advanced Hematology Five-Part Differential analyzer displaying a comprehensive 22-parameter complete blood count (CBC) with cellular histograms on an easy-to-read touch-screen. Its superior

More information

Diabetic Nephropathy

Diabetic Nephropathy Diabetic Nephropathy Outline Introduction of diabetic nephropathy Manifestations of diabetic nephropathy Staging of diabetic nephropathy Microalbuminuria Diagnosis of diabetic nephropathy Treatment of

More information

Evaluation of the Clinitek 200 urinary test-strip reader in the analysis of dog and rat urines in pre-clinical toxicology studies

Evaluation of the Clinitek 200 urinary test-strip reader in the analysis of dog and rat urines in pre-clinical toxicology studies 240 Laboratory Animals (1993) 27, 240-246 Evaluation of the Clinitek 200 urinary test-strip reader in the analysis of dog and rat urines in pre-clinical toxicology studies A. PAQUIGNON, G. TRAN & J. P.

More information

Controls (positive and negative) for quality assessment-located in central laboratory Note: See central lab UA procedure for QC. Read strips visually.

Controls (positive and negative) for quality assessment-located in central laboratory Note: See central lab UA procedure for QC. Read strips visually. ACTION DATE INITIAL Adopted Reviewed 2/8/16 KLW Reviewed 12/12/16 KLW Reviewed Reviewed Revised 5/11/10 MAS Supersedes 2/6/08 name change from Bayer to Siemens TITLE: URINALYSIS TESTING USING Siemens MULTISTIX

More information

Urine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator

Urine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator December 2014 PRODUCT ANNOUNCEMENT AU Clinical Chemistry Systems AU400/400 e, AU480, AU640/640 e, AU680, AU2700, AU5400, and AU5800 Product Urine/CSF Albumin Reagent Urine/CSF Albumin Reagent Urine/CSF

More information

URINE DIPSTICK AND SULPHOSALICYLIC ACID TEST. Špela Borštnar UREX 2015, Ljubljana, Slovenia

URINE DIPSTICK AND SULPHOSALICYLIC ACID TEST. Špela Borštnar UREX 2015, Ljubljana, Slovenia URINE DIPSTICK AND SULPHOSALICYLIC ACID TEST Špela Borštnar UREX 2015, Ljubljana, Slovenia KIDNEY DISEASE? severity of kidney disease = estimating GFR cause of kidney disease = urinalysis URINE EXAMINATION

More information

Clinical Laboratory Science: Urinalysis

Clinical Laboratory Science: Urinalysis Clinical Laboratory Science: Urinalysis Urine is produced by the kidney to maintain constant plasma osmotic concentration; to regulate ph, electrolyte and fluid balances and to excrete some 50 grams of

More information

Urinary albumin creatinine ratio (ACR) A comparative study on two different instruments

Urinary albumin creatinine ratio (ACR) A comparative study on two different instruments Original Research Article Urinary albumin creatinine ratio (ACR) A comparative study on two different instruments Luca Van Rooy 1, Vasudha K C 2*, Vanitha Gowda M N 3, Radhika K 4, Leen Puimege 5 1 Undergraduate

More information

Year 1 MBChB Clinical Skills Session Urinalysis

Year 1 MBChB Clinical Skills Session Urinalysis Year 1 MBChB Clinical Skills Session Urinalysis Reviewed & ratified by: Dr V Taylor-Jones & Ms C Tierney. Urinalysis Aims and Objectives Aim: For the student to be able to safely conduct a urinalysis on

More information

Urinalysis (UA) provides information about the urinary

Urinalysis (UA) provides information about the urinary Today s TeChniCian PEER REVIEWED URINALYSIS IN COMPANION ANIMALS Part 1: Collection, Sample Handling, & Initial Evaluation Theresa E Rizzi, DVM, Diplomate ACVP (Clinical Pathology) Oklahoma State University

More information

Pars Azmon (Automation) Pars Azmon (Manual) Biosystem (Automation) Biosystem (Manual) Ziest Chemi (Manual) 62

Pars Azmon (Automation) Pars Azmon (Manual) Biosystem (Automation) Biosystem (Manual) Ziest Chemi (Manual) 62 STANDARDIZATION OF URNIE PROTEIN MEASURMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty EQAP 7 : URINE PROTEIN Method No. Mean (mg/dl) CV% Pars Azmon (Automation) 45 25 33 Pars Azmon

More information

URINE PROTEIN In Normal State, Maximally There Is 150 mg/24h or 10 mg/dl Protein in Urine More Than 200 Protein Have Been Found in Urine Including 1/3

URINE PROTEIN In Normal State, Maximally There Is 150 mg/24h or 10 mg/dl Protein in Urine More Than 200 Protein Have Been Found in Urine Including 1/3 URNIE PROTEIN R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty URINE PROTEIN In Normal State, Maximally There Is 150 mg/24h or 10 mg/dl Protein in Urine More Than 200 Protein Have Been

More information

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A53727 The QMS Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole

More information

The Hill s Healthy Weight Protocol. Using the BFI Risk Chart

The Hill s Healthy Weight Protocol. Using the BFI Risk Chart The Hill s Healthy Weight Protocol Using the BFI Risk Chart Effective 12.05.2011 Welcome! The Hill s BFI Risk Chart is a new way to assess patient body composition and communicate that assessment with

More information

i-stat Alinity v Utilization Guide

i-stat Alinity v Utilization Guide istat Alinity v Utilization Guide The istat Alinity v delivers blood gas, acidbase, electrolyte, chemistry, and hematology measurements in a completely portable, handheld package. Accuracy is ensured by

More information

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007 Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007 Hosted by: Australian Small Animal Veterinary Association (ASAVA) Australian Small Animal Veterinary Association (ASAVA)

More information

40 CHRONIC KIDNEY DISEASE IN DOGS & CATS

40 CHRONIC KIDNEY DISEASE IN DOGS & CATS 40 CHRONIC KIDNEY DISEASE IN DOGS & CATS CONTINUING EDUCATION CONTINUING EDUCATION TREATMENT GUIDELINES FOR CHRONIC KIDNEY DISEASE IN DOGS & CATS: International Renal Interest Society Recommendations Gregory

More information

Detection and Estimation of Some Abnormal Constituents. Amal Alamri

Detection and Estimation of Some Abnormal Constituents. Amal Alamri Detection and Estimation of Some Abnormal Constituents Amal Alamri Lecture Over view Abnormal constituent of urine Urine analysis Experiments Physical Chemical Micro/Macro 1-Detection of some abnormal

More information

i-stat Alinity v Utilization Guide

i-stat Alinity v Utilization Guide istat Alinity v Utilization Guide The istat Alinity v delivers blood gas, acidbase, electrolyte, chemistry, and hematology results in a completely portable, handheld package. Accuracy is ensured by extensive

More information

LABORATORY 5: The Complete Urinalysis

LABORATORY 5: The Complete Urinalysis LABORATORY 5: The Complete Urinalysis Notes 1. This lab combines the objectives and activities of the macroscopic and microscopic lab activities. Students are expected to review those labs for reference.

More information

Proceeding of the ACVP Annual Meeting

Proceeding of the ACVP Annual Meeting http://www.ivis.org Proceeding of the ACVP Annual Meeting Oct.17-21, 2015 Minneapolis, MN, USA Next Meeting: Dec. 3-7, 2016 - New Orleans, LA, USA Reprinted in the IVIS website with the permission of the

More information

Coping with Analytical Interferences

Coping with Analytical Interferences Coping with Analytical Interferences (Handling Icteric, Hemolytic and Lipaemic Samples) Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney Surabaya Indonesia 2016 Acknowledgements

More information

Larry G. Adams, DVM, PhD, David J. Polzin, DVM, PhD, Carl A. Osborne, DVM, PhD, and Timothy D. O'Brien, DVM, PhD

Larry G. Adams, DVM, PhD, David J. Polzin, DVM, PhD, Carl A. Osborne, DVM, PhD, and Timothy D. O'Brien, DVM, PhD Correlation of Urine Protein/Creatinine Ratio and Twenty-Four-Hour Urinary Protein Excretion in Normal Cats and Cats with Surgically Induced Chronic Renal Failure Larry G. Adams, DVM, PhD, David J. Polzin,

More information

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153 S150 KEEP 2009 Analytical Methods American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153 S151 The Kidney Early Evaluation program (KEEP) is a free, communitybased health screening

More information

Color: Gray/Yellow. 5/7/2018 L Hematology results from IDEXX VetLab In-clinic Laboratory Requisition ID: 0 Posted Final Test Result Reference Range

Color: Gray/Yellow. 5/7/2018 L Hematology results from IDEXX VetLab In-clinic Laboratory Requisition ID: 0 Posted Final Test Result Reference Range 5/7/2018 L LD AA-Urinalysis results from IDEXX VetLab In-clinic COLLECTION = free catch COLOR = yellow CLARITY = clear SP GR = 1.020 GLUCOSE = neg BILIRUBIN = neg KETONE = neg BLOOD = neg PH = 6.0 PROTEIN

More information

CJ Shuster A&P2 Lab Addenum Urinanalysis 1. Urinanalysis

CJ Shuster A&P2 Lab Addenum Urinanalysis 1. Urinanalysis CJ Shuster A&P2 Lab Addenum Urinanalysis 1 Urinanalysis PLEASE NOTE: The actual urinanalysis goes rather quickly. You may want to skip directly to the Lab Exercise, collect data, and read the Introduction

More information

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats IDEXX Catalyst SDMA Test for in-house measurement of SDMA in serum from dogs and cats By Graham Bilbrough, Barbara Evert, Karen Hathaway, Gina Panagakos, Jane Robertson, and Maha Yerramilli Introduction

More information

Grading of acute kidney injury(2013)

Grading of acute kidney injury(2013) Acute kidney disease represents a spectrum of disease associated with a sudden onset of renal parenchymal injury most typically characterized by generalized failure of the kidneys to meet the excretory,

More information

Performance of three portable blood glucose monitoring devices used in a veterinary application

Performance of three portable blood glucose monitoring devices used in a veterinary application Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

Urinalysis and Body Fluids CRg. Unit Outline. Chemical Exam of Urine. Unit 3. Reagent strip manufactures

Urinalysis and Body Fluids CRg. Unit Outline. Chemical Exam of Urine. Unit 3. Reagent strip manufactures Urinalysis and Body Fluids CRg Unit 3 Chemical Examination of Urine Part 1 Overview of Procedure & ph Unit Outline Introduction to reagent strip testing ph Glucose & Carbohydrates Ketones Protein Bilirubin

More information

Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry

Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry Background The VACUETTE Urine CCM tube is for the collection, transport and storage of urine samples for urine culture and urinalysis in the

More information

Urinalysis (Macroscopic( Chemical Tests) ) Background. Ishihara Color Blindness Tests 12/22/2012. Mohammad Reza Bakhtiari DCLS, PhD

Urinalysis (Macroscopic( Chemical Tests) ) Background. Ishihara Color Blindness Tests 12/22/2012. Mohammad Reza Bakhtiari DCLS, PhD Urinalysis (Macroscopic( Chemical Tests) ) Mohammad Reza Bakhtiari DCLS, PhD Background Routine chemical examination of urine has changed dramatically since the early days of urine testing, owing to the

More information

PHYSICAL PROPERTIES AND DETECTION OF NORMAL CONSTITUENTS OF URINE

PHYSICAL PROPERTIES AND DETECTION OF NORMAL CONSTITUENTS OF URINE PHYSICAL PROPERTIES AND DETECTION OF NORMAL CONSTITUENTS OF URINE - OBJECTIVES: 1- The simple examination of urine. 2- To detect some of the normal organic constituents of urine. 3- To detect some of the

More information

The analytical phase

The analytical phase The analytical phase Result interpretation Test request Result Sampling Black box: the lab ANALYTICAL PHASE The CASE Uncle Pete, 67 years old Marked abdominal pain 8 pm, ED Acute abdomen? Assessment (+

More information

Urinary Collagen IV EIA

Urinary Collagen IV EIA Urinary Collagen IV EIA Enzyme Immunoassay Instructions for Use FOR RESEARCH USE ONLY Not for use in Diagnostic Procedures ACTIF 3-27-3 Kyodo, Setagaya-ku, Tokyo 1560052 JAPAN TEL&FAX: +81(0)3-6413-1860

More information

Andrea Tessari Microbiology Unit, Hospital of Rovigo, ULSS 18 Rovigo (Italy)

Andrea Tessari Microbiology Unit, Hospital of Rovigo, ULSS 18 Rovigo (Italy) URINARY SEDIMENT: STILL AN IMPORTANT DIAGNOSTIC TOOL International Symposium Rome, December 4 5 2014 The Application in Microbiology Andrea Tessari Microbiology Unit, Hospital of Rovigo, ULSS 18 Rovigo

More information

1 PROTOCOL. Comparison Study Summary

1 PROTOCOL. Comparison Study Summary Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015

More information

Proteinuria. Louisiana State University

Proteinuria. Louisiana State University Proteinuria W S A V A W C P, 2005 David F. Senior Louisiana State University The normal glomerulus is a highly selective barrier for filtration based on size (and on charge in the case of larger molecules).

More information

The NEPHROCHECK Test System Training Program. The NEPHROCHECK Training Program US: Astute Medical, Inc PN Rev D 2014/09/16

The NEPHROCHECK Test System Training Program. The NEPHROCHECK Training Program US: Astute Medical, Inc PN Rev D 2014/09/16 The NEPHROCHECK Test System Training Program The NEPHROCHECK Training Program US: Astute Medical, Inc. 2014 PN 700001 Rev D 2014/09/16 Learning Objectives At the conclusion of this training, participants

More information

INSTRUCTIONS FOR USE

INSTRUCTIONS FOR USE INSTRUCTIONS FOR USE VITROS Chemistry Products malb Reagent malb 680 1740 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products malb Reagent is used on the VITROS 5,1 FS Chemistry

More information

Lab Values Explained. working at full strength. Other possible causes of an elevated BUN include dehydration and heart failure.

Lab Values Explained. working at full strength. Other possible causes of an elevated BUN include dehydration and heart failure. Patient Education Lab Values Explained Common Tests to Help Diagnose Kidney Disease Lab work, urine samples and other tests may be given as you undergo diagnosis and treatment for renal failure. The test

More information

URINANLYSIS. Pre-Lab Guide

URINANLYSIS. Pre-Lab Guide URINANLYSIS Pre-Lab Guide NOTE: A very useful Study Guide! This Pre-lab guide takes you through the important concepts that where discussed in the lab videos. There will be some conceptual questions on

More information

Hexagon PSA. Design Verification. Contents

Hexagon PSA. Design Verification. Contents Design Verification Hexagon PSA Contents 1. Function...2 2. Sensitivity, Dynamic Range and Traceability...2 Description of Control Materials...2 Analytical Sensitivity...2 3. Clinical Evaluation...3 Evaluation

More information

David Bruyette, DVM, DACVIM

David Bruyette, DVM, DACVIM VCAwestlaspecialty.com David Bruyette, DVM, DACVIM Disorders of calcium metabolism are common endocrine disorders in both dogs and cats. In this article we present a logical diagnostic approach to patients

More information

Sensatest 1 PARAMETER: BLOOD, GLUCOSE AND PROTEIN URINE REAGENT STRIPS (1 PARAMETER) Test for Urine Occult Blood

Sensatest 1 PARAMETER: BLOOD, GLUCOSE AND PROTEIN URINE REAGENT STRIPS (1 PARAMETER) Test for Urine Occult Blood Sensatest 1 PARAMETER: BLOOD, GLUCOSE AND PROTEIN URINE REAGENT STRIPS (1 PARAMETER) Test for Urine Occult Blood INTENDED USE: Urine Reagent Strips-1B provide a semi-quantitative determination for the

More information

PET CARE VETERINARY CARE CENTER 2009 W SLAUSON AVE ACCOUNT #: ATTENDING VET: ANDERSON, DVM, JOY

PET CARE VETERINARY CARE CENTER 2009 W SLAUSON AVE ACCOUNT #: ATTENDING VET: ANDERSON, DVM, JOY Text KISMET EVENTOFF PET OWNER: EVENTOFF SPECIES: Feline BREED: GENDER: Female AGE: 2 Months PATIENT ID: PET CARE VETERINARY CARE CENTER 2009 W SLAUSON AVE 323-294-4030 ACCOUNT #: 93530 ATTENDING VET:

More information

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced

More information

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction Clin Chem Lab Med 2015; 53(11): 1737 1743 Beryl E. Jacobson, David W. Seccombe*, Alex Katayev and Adeera Levin A study examining the bias of albumin and albumin/creatinine ratio measurements in urine DOI

More information

Senior Wellness Screening Protocol & Guidance Notes

Senior Wellness Screening Protocol & Guidance Notes Senior Wellness Screening Protocol & Guidance Notes Early detection and prevention are the most important reasons why you should screen every pet, every year especially where statistics show 10% of normal

More information

THE PREVALENCE OF HEMATURIA IN DOGS AND CATS

THE PREVALENCE OF HEMATURIA IN DOGS AND CATS Scientific Works. Series C. Veterinary Medicine. Vol. LXIII (2) ISSN 2065-1295; ISSN 2343-9394 (CD-ROM); ISSN 2067-3663 (Online); ISSN-L 2065-1295 THE PREVALENCE OF HEMATURIA IN DOGS AND CATS Alexandra

More information

Quantitative protein estimation of Urine

Quantitative protein estimation of Urine Quantitative protein estimation of Urine 1 In a healthy renal and urinary tract system, the urine contains no protein or only trace amounts. The presence of increased amounts of protein in the urine can

More information

Rapid Laboratories In House Tests

Rapid Laboratories In House Tests Electrolytes CL CL (CHLORIDE) Electrolytes CO2 CO2 (BICARBONATE) Electrolytes K K (POTASSIUM) Electrolytes NA NA (SODIUM) Basic Metabolic Panel (BMP) GLU GLU (GLUCOSE) Basic Metabolic Panel (BMP) CA CA

More information

General Practice Service Willows Information Sheets. Diabetes Mellitus

General Practice Service Willows Information Sheets. Diabetes Mellitus General Practice Service Willows Information Sheets Diabetes Mellitus Diabetes Mellitus What is diabetes mellitus? As in humans, dogs and cats with diabetes mellitus are unable to control the level of

More information

Quantitative estimation of protein in urine

Quantitative estimation of protein in urine Quantitative estimation of protein in urine By sulphosalicalic acid Method BCH 472 In a healthy renal and urinary tract system, the urine contains no protein or only trace amounts. The presence of increased

More information

GIMA ŠLAPIMO ANALIZATORIAI

GIMA ŠLAPIMO ANALIZATORIAI GIMA ŠLAPIMO ANALIZATORIAI Eil. Nr. 1. Aprašymas 24051 URILYZER 100 URINE ANALYZER - with printer Urilyzer 100 is a small urine test strip reader that offers a broad range of optimized features. The intuitive

More information

Your pet s diabetes is manageable

Your pet s diabetes is manageable BLOOD GLUCOSE MONITORING SYSTEM Your pet s diabetes is manageable Treatment and Monitoring Plan for: Pet s Name: Clinic Name: Primary Vet: Clinic Phone No.: Clinic Email: Emergency Phone No: Today s Date:

More information

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. 2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846

More information

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626 Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi

More information

Elevated Urine Zinc Concentration Reduces the Detection of Methamphetamine, Cocaine, THC and Opiates in Urine by EMIT

Elevated Urine Zinc Concentration Reduces the Detection of Methamphetamine, Cocaine, THC and Opiates in Urine by EMIT Journal of Analytical Toxicology 2013;37:665 669 doi:10.1093/jat/bkt056 Advance Access publication July 10, 2013 Article Elevated Urine Zinc Concentration Reduces the Detection of Methamphetamine, Cocaine,

More information

Evaluation of the EUROLYSER-SOLO Total T4 assay. for veterinary samples

Evaluation of the EUROLYSER-SOLO Total T4 assay. for veterinary samples Evaluation of the EUROLYSER-SOLO Total T4 assay for veterinary samples Dr. rer.-nat. Volker Schlüter Zacherlweg 18 D-82061 Neuried Date: 2011 April 27 th Scope of the study Goal of this study was the validation

More information

Journal of Global Pharma Technology

Journal of Global Pharma Technology Journal of Global Pharma Technology ISSN: 0975-8542 Available Online at: www.jgpt.co.in RESEARCH ARTICLE Concordance and Validity of Urinalysis Using Sysmex UX-2000, Cobas U411 and Manual Microscopic Examination

More information

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date: BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

Laboratory Bulletin...

Laboratory Bulletin... Laboratory Bulletin... Updates and Information from Rex Healthcare and Rex Outreach February 1997 Issue Number 17 Urinary Microalbumin (an ideal test for real managed care) Microalbuminuria is the term

More information

Documentation, Codebook, and Frequencies

Documentation, Codebook, and Frequencies Documentation, Codebook, and Frequencies Urinary Albumin and Urinary Creatinine Laboratory Survey Years: 2005 to 2006 SAS Transport File: ALB_CR_D.XPT November 2007 NHANES 2005 2006 Data Documentation

More information

Creatinine & egfr A Clinical Perspective. Suheir Assady MD, PhD Dept. of Nephrology & Hypertension RHCC

Creatinine & egfr A Clinical Perspective. Suheir Assady MD, PhD Dept. of Nephrology & Hypertension RHCC Creatinine & egfr A Clinical Perspective Suheir Assady MD, PhD Dept. of Nephrology & Hypertension RHCC CLINICAL CONDITIONS WHERE ASSESSMENT OF GFR IS IMPORTANT Stevens et al. J Am Soc Nephrol 20: 2305

More information

Diagnosis and management of feline iatrogenic hypothyroidism

Diagnosis and management of feline iatrogenic hypothyroidism Vet Times The website for the veterinary profession https://www.vettimes.co.uk Diagnosis and management of feline iatrogenic hypothyroidism Author : Sarah Caney Categories : Companion animal, Feline, Vets

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

THE PREVALENCE OF HEMATURIA IN DOGS AND CATS

THE PREVALENCE OF HEMATURIA IN DOGS AND CATS AgroLife Scientific Journal - Volume 6, Number 2, 2017 ISSN 2285-5718; ISSN CD-ROM 2285-5726; ISSN ONLINE 2286-0126; ISSN-L 2285-5718 THE PREVALENCE OF HEMATURIA IN DOGS AND CATS Alexandra Mihaela POPA,

More information

INSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY

INSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY INTENDED USE INSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY INSTANT-VIEW H. pylori Rapid Test is a rapid qualitative immunoassay

More information

Reversal of Microalbuminuria A Causative Factor of Diabetic Nephropathy is Achieved with ACE Inhibitors than Strict Glycemic Control

Reversal of Microalbuminuria A Causative Factor of Diabetic Nephropathy is Achieved with ACE Inhibitors than Strict Glycemic Control ISSN 0976 3333 Available Online at www.ijpba.info International Journal of Pharmaceutical & Biological Archives 2013; 4(5): 923-928 ORIGINAL RESEARCH ARTICLE Reversal of Microalbuminuria A Causative Factor

More information

An unresolved issue: The relationship between spot urine protein-to-creatinine ratio and 24-hour proteinuria

An unresolved issue: The relationship between spot urine protein-to-creatinine ratio and 24-hour proteinuria Clinical Research Report An unresolved issue: The relationship between spot urine protein-to-creatinine ratio and 24-hour proteinuria Journal of International Medical Research 2019, Vol. 47(3) 1179 1184!

More information

Light yellow to dark golden yellow Clear ph range Specific gravity Sediments

Light yellow to dark golden yellow Clear ph range Specific gravity Sediments #11 Objectives: Understand specific gravity and identify normal specific gravity values for urine Learn to use a urine hydrometer to measure specific gravity Define specific gravity and identify normal

More information

PRINCIPLE OF URINALYSIS

PRINCIPLE OF URINALYSIS PRINCIPLE OF URINALYSIS Vanngarm Gonggetyai Objective Can explain : the abnormalities detected in urine Can perform : routine urinalysis Can interprete : the results of urinalysis Examination of urine

More information

Assisting in the Analysis of Urine. Copyright 2011, 2007, 2003, 1999 by Saunders, an imprint of Elsevier Inc. All rights reserved.

Assisting in the Analysis of Urine. Copyright 2011, 2007, 2003, 1999 by Saunders, an imprint of Elsevier Inc. All rights reserved. Assisting in the Analysis of Urine Urinalysis Why is Urine is analyzed? To detect extrinsic conditions those in which the kidney is functioning normally, but abnormal end-products of metabolism are excreted

More information

CHEMSTRIP 7 - URINE TEST STRIP

CHEMSTRIP 7 - URINE TEST STRIP University of California, San Francisco -- Department of Laboratory Medicine San Francisco General Hospital 1001 Potrero Avenue, San Francisco CA 94110 Clinical Laboratory- Eberhard Fiebig, MD, Director

More information

Summary of Recommendation Statements Kidney International Supplements (2013) 3, 5 14; doi: /kisup

Summary of Recommendation Statements Kidney International Supplements (2013) 3, 5 14; doi: /kisup http://www.kidney-international.org & 2013 DIGO Summary of Recommendation Statements idney International Supplements (2013) 3, 5 14; doi:10.1038/kisup.2012.77 Chapter 1: Definition and classification of

More information

Lunch and Learn Fresno Veterinary Specialty and Emergency Center March 19, Domenico Bianco, DVM, PhD, DACVIM

Lunch and Learn Fresno Veterinary Specialty and Emergency Center March 19, Domenico Bianco, DVM, PhD, DACVIM Lunch and Learn Fresno Veterinary Specialty and Emergency Center March 19, 2013 Domenico Bianco, DVM, PhD, DACVIM Outline Diagnosis of chronic kidney disease (CKD) Acute kidney injury (AKI) vs. CKD International

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information