SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World
|
|
- Chrystal Brooks
- 5 years ago
- Views:
Transcription
1 MIT OpenCourseWare SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World Spring 2009 For information about citing these materials or our Terms of Use, visit:
2 Colorimetric Diagnostic Assay for the Measurement of Blood Glucose Deepali Ravel, Yi Wang, and anonymous students AL, ML and DR Dlab Health 1
3 Background One glucometer is shared between 15 clinics in the Nicaraguan province of Estelí. Test strips are proprietary and expensive Poor distribution of test strips and glucometers leads to a mismatch in supply at hospitals National Prevalence 7.6% (2007) 9.4% (2025) Courtesy of the International Diabetes Federation. Used with permission. Nicaragua Diabetes Statistics: Deaths Attributable Males 10.4% (2007) Females 18.2% (2007) Prevalence of Diabetes in Central America 2
4 Diabetes Management When a patient has diabetes, their body cannot effectively regulate glucose levels. It is dangerous if glucose concentrations fall outside normal range Goal: Glycemic control Keep within normal values of mg/dl Poor glycemic control mg/dl over months Measure blood glucose levels 4 5 times a day Allow patients to act quickly to correct for swings in their glucose concentrations, by either injecting a dose of insulin or consuming glucose pills. 3
5 Diabetes in the Developing World Why is it a problem? Has reached epidemic proportions - Due to rapid cultural changes, increasing urbanization, dietary changes Currently affects 246 million people, 6% of world s adult population - Contributes to 19.7 million DALYs - By 2025, 80% of all new cases will appear in developing countries 6 million people develop diabetes each year - Less than half the people with diabetes in developing countries are diagnosed Diabetes causes 3.5 million deaths each year - Two-thirds of deaths caused by diabetes occur in developing countries The burden of the disease increasingly falls on developing countries 4
6 Diabetes Prevalence in 2007 Courtesy of the International Diabetes Federation. Used with permission. 5
7 Diabetes Prevalence in 2025 Courtesy of the International Diabetes Federation. Used with permission. 6
8 Problem Statement In order to control their disease, diabetes patients must monitor their blood glucose level on a regular basis. In the region of Estelí, in Nicaragua, one glucometer is shared between 15 clinics, meaning that the vast majority of patients are not receiving daily monitoring. Glucose testing is so infrequent because test strips are proprietary and expensive. Our challenge is to design a fast, cheap, safe and reliable way of measuring glucose levels that can be used widely by diabetes patients, even in poor resource settings. - Team GlucoTank 7
9 Current Glucometers Drop a prick of blood onto a test strip, which the glucometer than reads to measure glucose levels, through an electrochemical assay. Source: Wikipedia (public domain) 8
10 Design Specifications Characteristic Current Our Prototype Sample Volume 0.3ul to 10ul 10ul Accuracy 10-15% Unknown Range Be able to measure healthy glucose range: mg/dl mg/dl Processing Time 3-60 seconds seconds Testing Site Mostly fingertips Fingertip Testing Frequency Up to times per day Same Reader Size Around 8cm x 5cm x 3cm (palm size) 20cm x 6cm x 2.5cm Reader Price $25 to $100. Some companies offer free monitors, but sell test strips Very low Test Strip Price $1.00-$2.00 <$0.20 Units US preference is mg/dl. Canada and Europe preference is mmol/l (x/18) mg/dl Other Qualities Allow data storage/ monitoring None 9
11 Design Ideas I: Telemedicine A. Use Camera Phone to take image of test strip B. Transfer image to computer in offsite lab/hospital C. Use imaging software such as Photoshop to convert image to grayscale D. Read out the mean intensity and use key to translate value to Glucose level Create colorimetric test strips, and a phone (or computer) application that, when a picture of the strip is loaded, it converts it to gray scale and measures the average light intensity- translating that to a glucose level. However: - Cell phones that have computer power to do this are quite expensive (idea not able to adapt to single user level) - Similar idea being pursued by Diagnostics for All 10
12 Design Ideas II: Glucophone More expensive technology than the telemedicine idea. Uses cell phone to display, power, and store data Data can be sent to a central database for remote patient monitoring However: Expensive to develop Technologically challenging? Company already doing this Would be difficult designing component attachable to all phones Image remove due to copyright restrictions. Photos of the GlucoPhone, from 11
13 Pugh Chart Design Goals Status Quo Design 1 Design 2 Design 3 Design 4 Current Microfluidic Current Paper glucometers are glucose test Glucometer with diagnostic- only expensive and system non-proprietary colorimetric proprietary. strips strip- no meter Design Parameters Weights (1-3) New, Local Glucometer with Locally Built Test Strips Long-term Effectiveness Speed of Use Reliability Accuracy Economy Accessibility Usability Simplicity Practicality Maintainability Aesthetic Appeal Environmental Friendliness Safety Versatility Efficiency Marketability Ease of Integration Total Weighted Total
14 Pugh Chart Design Goals Weights (1-3) Status Quo Design 1 Design 2 Design 3 Design 4 Current Microfluidic Current Paper New, Local glucometers are glucose test Glucometer with diagnostic- only Glucometer with expensive and proprietary. system non-proprietary strips colorimetric strip- no meter Locally Built Test Strips Long-term Effectiveness Reliability Accuracy Economy Maintainability Safety Total Weighted Total
15 Our Proposed Solution: Colorimetric Strips The reagents for the glucose assay are placed on a Filter Paper Test Strip. 10ul of the glucose solution (blood) are deposited on strip. Color change from clear to blue is witnessed. 14
16 Our Proposed Solution: Reader Battery-operated hand-held reader LEDs indicate blood glucose-level within set ranges 15
17 Costs Paper Test Strips: Reagents per strip : 1.5 cents Packaging: approximately 5 cents per strip -Packaging constraints: must protect the strips from light, since some reagents are light-sensitive. Solution: use aluminum coated paper. Reader: Approximately $10 or less per reader - Cost of electrical components is very low. 16
18 Testing and Validation Test strips: Tested different concentrations of glucose Fasting levels, normal levels, and high levels all give different dye intensity. Tested different reagent deposition orders Future work: test time/light sensitivity of reagents, inkjet printing of reagents 17
19 Testing and Validation Reader: Tested device with calibration strips and our strips - Sensitive to white and three increasing intensities of blue on calibration strips Tested device in normal room lighting and within device housing (dark) Future work: Test device with specific glucose concentrations, actual strips and smaller ranges of concentration 18
20 Manufacturing and Regulatory Pathways Our design is simple and cheap to produce: Local diagnostic company Local materials Distribution: Follow the distribution system currently used in Nicaragua, and spread to Polyclinics. 3-4 for use in clinic at first, but eventually 1 glucometer per Diabetes patient. Regulation: 510(k) pathway Predicate device- colorimetric glucose assay, optic technology 19
21 Medical Device Compass Chart Lab-Intensive Glucose Assay Effectiveness Our Current Prototype Urinanalysis Strips Our Goal Colorimetric Test Strips Standard Glucometer Glucophone Advanced Glucometer Colorimetric test strips were abandoned around 10 years ago for the electrochemical test strips, since there was no appropriate reader at that time. The electrochemical assay is more expensive, but also more accurate. Cost 20
22 Future Work Developing the Test Strip: Double layer strip idea, with a top layer that is able to filter the blood cells allowing just the serum to pass through to reagents. Developing the Reader: Creating a more sensitive reader with more LED indicator lights, so we can inform of patients their glucose levels more accurately. 21
23 Questions? 22
System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th
More informationMonitoring of Blood Glucose Level
University of Zagreb Faculty of Electrical Engineering and Computing Biomedical instrumentation Monitoring of Blood Glucose Level Biomedical instrumentation 1/17 Outline Introduction Principles of Blood
More informationThe Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values
University of New Hampshire University of New Hampshire Scholars' Repository Honors Theses and Capstones Student Scholarship Spring 2017 The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary
More informationDexcom Current and Future Technology Innovations. Jorge Valdes Chief Technical Officer
Dexcom Current and Future Technology Innovations 2015 Jorge Valdes Chief Technical Officer G5 App Sensor (Same as G4) Indicated for adults and children down to Age 2 years Tiny insertion needle (26Ga)
More informationAccessible Insulin Pump
Accessible Insulin Pump ECE 480 Design Issues Team 3 Miriel Garcia Hoyoung Jung Marshall Williams Anthony Iafrate 4/10/15 Introduction Insulin pumps are medical devices that deliver insulin 24 hours a
More informationOGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.
ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationAccessible Blood Glucose Monitor Interface
Accessible Blood Glucose Monitor Interface Team 2 Matthew Bularzik, David Price, Michael Rivera Sponsored by the Rehabilitation Engineering Research Center Purpose People with diabetes must check their
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess
More informationNEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP
NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP No commercial support or sponsorship was received for this project I have nothing to disclose OBJECTIVES Identify at
More informationWhy do we need SD LipidoCare?
SD Biosensor, Inc. Innovative Global Leader of In Vitro Diagnostics Total supplier of Blood Glucose Monitoring System and Lipid Analyzer System in vitro diagnostics test kits for Point of Care Testing
More informationWhite Paper. High performance resistors are key to meeting the demanding requirements of portable medical electronics designs. Issued in June 2014
White Paper High performance resistors are key to meeting the demanding requirements of portable medical electronics designs Issued in June 2014 The contents of this White Paper are protected by copyright
More informationLIGHTING LABORATORY TESTING SERVICES EFFECTIVE & EFFICIENT ENERGY SAVING SOLUTIONS V1015W
LIGHTING LABORATORY TESTING SERVICES V1015W EFFECTIVE & EFFICIENT ENERGY SAVING SOLUTIONS ABOUT THE LAB The Energywise Lighting Laboratory is offering complete photometric, electric and thermal testing
More informationFigure 2.1: Glucose meter
CHAPTER TWO: MONITORING TECHNOLOGIES 2.1 Introduction Glucose monitoring is a method of self-testing glucose (blood sugar) levels for the management of diabetes. Traditionally, it involves pricking the
More informationDiabetes initiatives gaining steam on government and stakeholder sides
Diabetes initiatives gaining steam on government and stakeholder sides ByTESSIE SANCIOCTOBER 16, 2018 Diabetes Canada published this letter to Prime Minister Justin Trudeau, asking his government to support
More informationAbout the Project Education Play Community Games Publications News Help Home
About the Project Education Play Community Games Publications News Help Home About the Project Workbook Games Workshop poems Publications Events Writers Fructosamine to a1c conversion whst does dr Petre
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationNoninvasive Glucose Monitors Devices, Technologies, Players and Prospects
Report Prospectus Noninvasive Glucose Monitors Devices, Technologies, Players and Prospects 1 Improving Diabetes Management For diabetes patients, glucose monitoring is a way of life, a several-times-per-day
More informationPHYSIOLOGICAL RESEARCH
DOMAIN STUDIES PHYSIOLOGICAL RESEARCH In order to understand the current landscape of psychophysiological evaluation methods, we conducted a survey of academic literature. We explored several different
More informationNon-Invasive Glucose Monitoring. Kevin Walls April 12, :30 PM-7:30 PM.
Non-Invasive Glucose Monitoring Kevin Walls April 12, 2018 6:30 PM-7:30 PM. Technology is evolving Technology is evolving at such a rapid rate it s hard to keep up. Imagine living a healthier life without
More informationYour Sweat Will See You Now
Your Sweat Will See You Now A new device wearable, wireless and battery free improves the ability to monitor and diagnose health problems by analyzing the sweat on your skin. By Apoorva Mandavilli Jan.
More informationSP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World
MIT OpenCourseWare http://ocw.mit.edu SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World Spring 2009 For information about citing these materials or our
More informationHOME DIAGNOSTIC TESTS FOR PERSONAL HEALTH PRODUCT CATALOGUE
HOME DIAGNOSTIC TESTS FOR PERSONAL HEALTH PRODUCT CATALOGUE 2 WELCOME TO CIGA HEALTHCARE AND Take Charge of Your Own Health With consumers all over the world becoming increasingly more concerned about
More informationSee Your Body BetterTM
Product Brochure 2000 See Your Body BetterTM Products to Track Your Weight and Body Fat % Know Your Weight and Body Fat % Ordinary bathroom scales measure only one thing weight. They can t distinguish
More information1 Dexcom G4 Platinum User Guide May 2012
1 Dexcom G4 Platinum User Guide May 2012 Page 1. Dexcom G4 PLATINUM Continuous Glucose Monitoring System. CE marked for use in 1Dexcom G4 User Guide, May 2012. LBL-011277 Rev04. color display. Dexcom G4
More informationEDSGN 100 DESIGN PROJECT #1 PROGRESS REPORT
EDSGN 1 DESIGN PROJECT #1 PROGRESS REPORT TOOTHBRUSH DESIGN INTRODUCTION TO ENGINEERING DESIGN EDSGN 1 Sec. 25 Team #5 Siyang Liu Raymond Richlin Nam Mi Suh Kim sxl557@psu.edu rer5211@psu.edu nos531@psu.edu
More informationSmartSensor Home OGTT Kit For Detection Of Diabetes
SmartSensor Home OGTT Kit For Detection Of Diabetes Background to Diabetes Global diabetes epidemic 50% of type 2 undiagnosed; 183m people Prediabetes affects over 350m people 18% of pregnancies affected
More informationMulti User Model. Product Overview
Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL
More informationGLUCOMETER: High Level Design
Fall 2014 High Level Design 1 GLUCOMETER: High Level Design Electrical Engineers: Michael Williams, Christopher Homa, Matthew Henne Mechanical Engineer: Ish Amagashie Software Engineer: Sofyan Saputra
More informationCustomer Bulletin. Questions/Comments? Please call Customer Service at (Direct) or (Toll-free inside USA).
Customer Bulletin Assignment of Ranges for PTS Panels Multi-Chemistry Controls (Cat. No./REF 721) and PTS Panels Controls (Cat. No./REF 722) Background: PTS Panels Multi-Chemistry Controls and Controls
More informationBeneCheck BK6-12M. Plus Multi-Monitoring Meter and Strips
BeneCheck BK6-12M Plus Multi-Monitoring Meter and Strips The BeneCheck BK6-12M multi-monitoring system is an easy to use, handheld device which allows you to check your Total Cholesterol, as well as Blood
More informationGlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System
GlucCell TM SYSTEM USER S GUIDE Ver 2.1 Intended Use The GlucCell TM Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during cell culture.
More informationBLOOD GLUCOSE METER TEST STRIP STEP THERAPY CRITERIA
BLOOD GLUCOSE METER TEST STRIP STEP THERAPY CRITERIA Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit plan.
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationTechnology Innovation Alert. N o n - i n v a s i v e D i a b e t e s M o n i t o r i n g. 16 December 31 December, Published. Granted.
Technology Innovation Alert N o n - i n v a s i v e D i a b e t e s M o n i t o r i n g 16 December 31 December, 2018 3 10 1 1 Published 10 Granted 5 Inactive 0 Start ups Companies Universities Individuals
More informationDiagnostics: Bench to Bedside Pathway
Diagnostics: Bench to Bedside Pathway Rosanna W Peeling Professor and Chair, Diagnostic Research Director, International Diagnostics Centre London School of Hygiene & Tropical Medicine www.idx-dx.org The
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k063276 B. Purpose for Submission: New device C. Measurand: Urinary Glucose, Blood,
More informationTruSignal SpO2 Sensors
Product Overview TruSignal SpO2 Sensors Photo Here Delivering exceptional oximetry performance Page 1 About GE TruSignal SpO2 sensors Page 2 About TruSignal SpO2 sensors Delivering exceptional oximetry
More informationDiabetes Devices Workshop Angela Aldrich, PharmD, PhC April Mott, PharmD, PhC, BCPS Presbyterian Medical Group 28 January 2018
Diabetes Devices Workshop Angela Aldrich, PharmD, PhC April Mott, PharmD, PhC, BCPS Presbyterian Medical Group 28 January 2018 Pumps & Sensors & Meters, Oh My! A Tale of Two Meters Technology for glucometers
More informationNoninvasive Glucose Monitors to 2022
Noninvasive Glucose Monitors to 2022 Devices, Technologies, Players and Prospects Report Prospectus Greystone Research associates N o n i n v a s i v e G l u c o s e M o n i t o r s t o 2 0 2 2 Greystone
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationUse Tomorrow s Technology Today
Use Tomorrow s Technology Today Company Profile (MES) is a leading edge company specializing in rapid, automated semen analysis for both the human and the veterinary markets. The Company s comprehensive
More informationThis is a licensed product of Ken Research and should not be copied
1 TABLE OF CONTENTS 1. The US Diabetes Care Devices Market Introduction 1.1. What is Diabetes and it s Types? 2. The US Diabetes Care Devices Market Size, 2007-2013 3. The US Diabetes Care Devices Market
More informationYour pet s diabetes is manageable
BLOOD GLUCOSE MONITORING SYSTEM Your pet s diabetes is manageable Treatment and Monitoring Plan for: Pet s Name: Clinic Name: Primary Vet: Clinic Phone No.: Clinic Email: Emergency Phone No: Today s Date:
More information1. INDENTIFICATION OF THE SUBSTANCE/PREPARATION AND OF THE COMPANY UNDERTAKING Trade name
DN GM-0227 Version 1 Revision 0 Effective Date: 2016.1.20 Page 1 of 3 SAFETY DATA SHEET 1. INDENTIFICATION OF THE SUBSTANCE/PREPARATION AND OF THE COMPANY UNDERTAKING Trade name On Call Pro Blood Glucose
More informationNHS Greater Glasgow and Clyde Adult Blood Glucose and Combined Ketone Meter and Test Strip Formulary
NHS Greater Glasgow and Clyde Adult Blood Glucose and Combined Ketone Meter and Test Strip Formulary 2018-20 Primary Care and Acute Joint Formulary Blood Glucose Meter formulary developed by the Diabetes
More informationEfficient Measurement of Large Light Source Near-field Color and Luminance Distributions for Optical Design and Simulation
Efficient Measurement of Large Light Source Near-field Color and Luminance Distributions for Optical Design and Simulation Hubert Kostal*, Douglas Kreysar, Ronald Rykowski Radiant Imaging, Inc., 22908
More informationSummary of Significant Changes at this Revision. Items Required
Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of
More informationthe best of care Managing diabetes with the FORA Diamond MINI and tools from Discovery Health Medical Scheme
the best of care 2014 Managing diabetes with the FORA Diamond MINI and tools from Discovery Health Medical Scheme contents What this document is about This document gives an overview of the FORA Diamond
More informationQ.E.D. Saliva Alcohol Test
Q.E.D. Saliva Alcohol Test A Quantitative Test for the Determination of Equivalent Blood Alcohol Content (BAC) Using a Saliva Sample. Approved by the Federal Department of Transportation (DOT) for Commercial
More informationLOGIQ S8 XDclear 2.0. Simply Amazing
LOGIQ S8 XDclear 2.0 Simply Amazing Available with FibroScan technology to aid in the diagnosis, staging, and monitoring of chronic liver disease Enhanced B-Flow Imaging a GE exclusive with exquisite sensitivity
More informationGuess: Correct or Incorrect. Trial (perform in random order)
AP Biology Senses Lab Names Per. Our senses are constantly bombarded with various stimuli from the environment, which are relayed to the central nervous system where the information is interpreted. In
More informationWELL CARD HEALTH TOOL BOX
WELL CARD HEALTH TOOL BOX 1 ADVOCATE Brand Advantage Well Card Health Members Will Save 52-70% On Diabetic Supplies Free Talking Meter Or Dash Meter Programs To Well Card Health Members 3 to 5 day UPS
More informationIcon specifications. Handpiece compatibility
Icon specifications Voltage Frequency Max. power Speed Weight 90-264 VAC 50-60 Hz 1600 VA Up to 3 Hz (pulse rate) 104 lb. (47.1kg) Handpiece compatibility Handpiece Technologies Hair removal Pigmented
More informationPractical Alternatives for Remote Neurotherapy: How Do We Reach the Home, School, and Office? Thomas F. Collura, Ph.D. BrainMaster Technologies, Inc.
Practical Alternatives for Remote Neurotherapy: How Do We Reach the Home, School, and Office? Thomas F. Collura, Ph.D. BrainMaster Technologies, Inc. Conventional Wisdom There is no reason for anyone to
More informationP.Venkatesan Chief Operating Officer
Diabetopaedia. com Meeting the challenges of Providing Information for Doctors, Patients other and Health-Carers in Diabetes. by P.Venkatesan Chief Operating Officer DIABETOPAEDIA DOT COM PVT. LTD. venkat@diabetopaedia.com
More informationThermo Scientific Finnpipette Manual Pipetting Systems. you can handle. anything
Thermo Scientific Manual Pipetting Systems you can handle anything Leader in Pipetting Technology Since the introduction of the first variable volume micropipette in 1971, more than four million Thermo
More informationData Management System (DMS) User Guide
Data Management System (DMS) User Guide Eversense and the Eversense logo are trademarks of Senseonics, Incorporated. Other brands and their products are trademarks or registered trademarks of their respective
More informationThe Small-Size Electrode Type Blood Glucose Meter Antsense III and Point of Care Testing
Feature Article The Small-Size Electrode Type Blood Glucose Meter Antsense III and Point of Care Testing Yoichi Ohmori Antsense III is a light, compact and portable electrode type blood glucose meter which
More informationApolipoprotein A1 (Apo A1) ELISA
Package Insert Apolipoprotein A1 (Apo A1) ELISA 1 x 96 Wells For Research Use Only (RUO). Not for use in clinical, diagnostic or therapeutic procedures. v. 1.0 Eagle Biosciences, Inc. 20A Northwest Blvd.,
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 15197 First edition 2003-05-01 In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus Systèmes d'essais
More informationRobustness Assessment of Safety Devices for Prefilled Syringes
Robustness Assessment of Safety Devices for Prefilled Syringes INTRODUCTION Prefilled syringes are becoming a preferred way of administration with a market growth over 10% per year (1). Preferences of
More informationGE Healthcare. LOGIQ Book XP. Innovation in compact ultrasound
GE Healthcare LOGIQ Book XP Innovation in compact ultrasound At the leading edge of healthcare High performance compact ultrasound... anywhere Clarity Confidence Simplicity Speed For more than a century,
More informationUser Guide. December_2018
User Guide December_2018 CONTENTS Contents 03 04 06 07 10 1 1 1 2 1 3 1 4 15 16 1 7 18 Download and Install Create an account Main Screen New Log My Logbook Charts Menu Profile and configuration Carbs
More informationThe Danish Medicines Agency s availability strategy
låst DECEMBER 2016 The Danish Medicines Agency s availability strategy 2017 2021 Danish Medicines Agency, 2016 This publication may be freely quoted with appropriate acknowledgement of the source. Images
More informationOneTouch Reveal Web Application. User Manual for Healthcare Professionals Instructions for Use
OneTouch Reveal Web Application User Manual for Healthcare Professionals Instructions for Use Contents 2 Contents Chapter 1: Introduction...4 Product Overview...4 Intended Use...4 System Requirements...
More informationNHS Greater Glasgow and Clyde Adult Blood Glucose Meter and Test Strip Formulary
NHS Greater Glasgow and Clyde Adult Blood Glucose and Test Strip Formulary 2016-17 Primary Care and Acute Joint Formulary Blood Glucose formulary developed by the Diabetes MCN Blood Glucose Formulary Review
More informationThe PATH to better cardiac care starts here
The PATH to better cardiac care starts here #1 for precision among point-of-care analyzers * Affordable, core-lab quality results in under 17 minutes Test menu: Troponin I, NTproBNP, CK-MB, D-Dimer, hscrp
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationUser Manual second language
User Manual second language GlucoTel Blood Glucose Monitoring and Diabetes Management System must be used with cell phones that have: Table of contents 2 3 Introduction 4 Bluetooth Wireless Technology
More informationBlood Glucose Test Strip Market Research Report- Global Forecast till 2020
Report Information More information from: https://www.marketresearchfuture.com/reports/976 Blood Glucose Test Strip Market Research Report- Global Forecast till 2020 Report / Search Code: MRFR/MED/0470-CRR
More informationBBB_company Brief_2018
BBB_company Brief_2018 BBB Tech creates affordable blood-testing technologies with mobile-enabled tools and data insights. BBB History 2017. 11 Launched elemark TM dual check in Korea 2017. 08 Signed OEM
More informationManagement of Central Venous Access Devices. Blood Glucose Monitoring
Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationTitle: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated.
COPY Summary of Significant Changes at this Revision SOP updated Purpose and Scope Items Required 1. The Abbott Optium Xceed meter is a battery-powered device designed for the measurement of blood glucose
More informationBlood Glucose/Sugar Monitoring
What is a glucose meter and do I need one? A glucose meter is a small battery powered machine. Meters measure your blood sugar and your blood glucose level is then shown on a small screen. There are a
More informationGlucCell TM SYSTEM USER S GUIDE ver 2.3 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System
GlucCell TM SYSTEM USER S GUIDE ver 2.3 Intended Use The GlucCell TM Cell Culture Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during
More informationBasic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009
Basic knowledge of POCT instuments The Tide Resort Bangsan Beach 18 June 2009 1 POCT instruments concepts Point of care testing Diagnostic medical devices used at or near the site of patient care Bed site,
More informationPOINT OF CARE DIAGNOSTICS
ENSURING ACCESS TO POINT OF CARE DIAGNOSTICS IN DECENTRALISED HIV AIDS PROGRAMMES REQUEST FOR EXPRESSION OF INTEREST Technology : Point of Care HIV diagnostic and treatment monitoring devices for Early
More informationOmnitest : 2013
ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample
More informationExplain how the normal mice prevented their blood glucose concentration falling when they had not eaten for 48 hours
Q1.Scientists investigated the control of blood glucose concentration in mice. They kept a group of normal mice without food for 48 hours. After 48 hours, the blood glucose concentrations of the mice were
More informationRe: ENSC 440 Project Functional Specification for a Non-invasive Glucose Measuring Device
Lumen Medical Technologies Inc. 1154 Harold Rd. North Vancouver, B.C. V7K 1G3 February 20, 2006 Dr. Andrew Rawicz Simon Fraser University 8888 University Drive Burnaby, BC V5A 1S6 Re: ENSC 440 Project
More informationSAMPLE. Glucose Monitoring in Settings Without Laboratory Support
POCT13c Glucose Monitoring in Settings Without Laboratory Support This guideline focuses on performance of point-of-care glucose monitoring systems, with an emphasis on safety practices, quality control,
More informationA1c conversion to average blood sugar
A1c conversion to average blood sugar The Borg System is 100 % A1c conversion to average blood sugar Convert HbA1c (%) to estimated Average Glucose (mg/dl). eag for plasma calibrated meters. HbA1c, 4.0,
More informationW h a t i s i d e a l b l s u g a r l e
W h a t i s i d e a l b l s u g a r l e Discuss blood glucose ( sugar ) targets with your healthcare team when creating your diabetes management plan. People who have diabetes should be testing their blood.
More informationAlex Azar Secretary, Department of Health and Human Services
February 28, 2018 Alex Azar Secretary, Department of Health and Human Services Dear Secretary Azar, On behalf of the Endocrine Society members and leaders, I write to offer our assistance as you lead the
More informationElsevier ClinicalKey TM FAQs
Elsevier ClinicalKey FAQs Table of Contents What is ClinicalKey? Where can I access ClinicalKey? What medical specialties are covered in ClinicalKey? What information is available through ClinicalKey?
More informationSUBJECT INFORMATION AND CONSENT FORM
Title of Study: Department of Anesthesiology British Columbia s Children Hospital 4480 Oak Street Vancouver V6H 3V4 Tel 604 875 2711 Fax 604 875 3221 SUBJECT INFORMATION AND CONSENT FORM Mobile Health:
More informationAdolescents to. Traditional methods applied in the dietary assessment of adolescents. Adolescence. Food records. 24 h recall
Graduate & PhD Course Measuring Dietary Behaviour the intelligent way Fudan University, Shanghai November 23 25, 2015 Adolescents FROM 10 to 19 YEARS OLD Dietary assessment methods in adolescents traditional
More informationPROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:
PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:
More informationContinuous Glucose Monitoring System. Receiver Setup
Continuous Glucose Monitoring System Receiver Setup Overview Your Dexcom G5 Continuous Glucose Monitoring (CGM) System is made up of: Receiver What do I do with it? Review your readings. Set and receive
More informationBack Row: Brogan Dexter, Daisy Rathod, Sangeeta Bookseller, Jon Young, Jude Kieltyka. Front Row: Adeseye Adekeye, Anu Topgi, Shivani Shah, Chad
Back Row: Brogan Dexter, Daisy Rathod, Sangeeta Bookseller, Jon Young, Jude Kieltyka. Front Row: Adeseye Adekeye, Anu Topgi, Shivani Shah, Chad Nishizuka Purpose Background Design Cost Analysis Obstacles
More informationJanuary 2017 Investor Presentation. confidently live life with ease
January 2017 Investor Presentation confidently live life with ease Forward-Looking Statements This presentation contains certain forward-looking statements that involve risks and uncertainties that could
More informationDetection. personal. Now, you can optimize your personal workflow. Promega instruments and reagents integrate easily.
personal Detection Now, you can optimize your personal workflow. Promega instruments and reagents integrate easily. An ultra-sensitive, versatile, and affordable single tube luminometer for Life Science
More informationBLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR
BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR P.Daarani 1 & A.Kavithamani 2 International Journal of Latest Trends in Engineering and Technology pp. 141-147 DOI: http://dx.doi.org/10.21172/1.ires.19
More informationCareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes
CareLink Management Software for Diabetes software REPORT REFERENCE GUIDE How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used
More informationThe performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.
Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added
More informationEquivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose
More informationBlood Glucose Monitoring System. Copyright 2016 Ascensia Diabetes Care Holdings AG diabetes.ascensia.com
Viewing test results in My Readings The CONTOUR DIABETES app captures all your blood glucose readings to create personalized patterns and trends, so you can see how your daily activities impact your results.
More informationI can show you how to check my child s blood sugar
Our Journey with Diabetes Si usted desea esta información en español, por favor pídasela a su enfermero o doctor. I can show you how to check my child s blood sugar Testing Blood Sugar Blood sugar levels
More information