1 ORIGINAL ARTICLE Gauging Effective Spatial Resolution in Multirow Helical Cardiac Computed Tomography With a Dynamic Phantom Friedrich D. Knollmann, MD,* Tarkan Cangöz, MA,* Erdogan Cesmeli, PhD, Thomas Toth, Peter Edic, PhD, Johannes Müller, PhD, MD, and Roland Felix, MD, PhD* Purpose: To devise a numerical indicator of image quality for multirow helical cardiac computed tomography (CT) and its relation to temporal resolution. Materials and Methods: A pulsatile cardiac assist device was used to simulate cardiac wall motion by mechanically transmitting the device dynamics to a piece of tungsten wire. Wire motion induced by different device rates was captured with an 8-row subsecond helical CT scanner operating with various scanning parameters. Image artifacts were visually assessed and compared with the image point spread function (PSF) using the full width at half maximum (FWHM) area as a numerical estimate of spatial accuracy. Results: At rest, the FWHM area was determined as 1.3 mm 2.Ata device rate of 60 bpm, the FWHM area ranged from 1.51 mm 2 to mm 2, depending on the time of image reconstruction. Mean reproducibility of the FWHM area measurements was determined as 0.05, whereas visual estimates of motion artifact were highly variable between different readers (kappa 0.19). Visually determined image quality correlated closely with the FWHM area metric (Spearman s rank correlation, P , rho 0.841). At a device rate of 100 bpm, the minimum FWHM area was 2.00 mm 2 using a single-sector algorithm, 1.41 mm 2 using a 2-segment algorithm, and 1.37 mm 2 using a 4-segment algorithm. Conclusions: Use of a pulsatile cardiac assist device could serve as an in vitro test bed for cardiac CT imaging methods. Area FWHM of the PSF correlates well with visually determined image quality of a dynamic phantom, but provides better reproducibility than visual analysis. Received September 29, 2002 and accepted for publication, after revision, July 28, From the *Department of Radiology, Charité, Campus Virchow-Klinikum, Humboldt-University, Berlin, Germany; GE Global Research, Niskayuna, New York; GE Medical Systems, Milwaukee, Wisconsin; and the Department of Cardiothoracic and Vascular Surgery, German Heart Institute, Berlin, Germany. Reprints: Friedrich D. Knollmann, MD, Strahlenklinik, Charité, Campus Virchow-Klinikum, Augustenburger Platz 1, Berlin, Germany. Friedrich.Knollmann@charite.de Copyright 2003 by Lippincott Williams & Wilkins ISSN: /04/ DOI: /01.rli d5 Key Words: computed tomography (CT), artifact, computed tomography (CT), image quality, coronary vessels, CT, heart, CT, computed tomography (CT), experimental studies (Invest Radiol 2004;39: 13 19) The recent introduction of multirow detector computed tomography (MDCT) 1,2 has provided a potentially useful noninvasive cardiac imaging modality. 3 5 The most important potential application is noninvasive coronary angiography, but motion artifact remains an important limitation. In a direct comparison with conventional coronary angiography, all arteries were evaluable with 4-row multidetector CT in only 30% of the cases. 6 To overcome motion artifacts, temporal resolution of the CT technique can be improved by accelerating gantry speed or using segmented image reconstruction techniques. 7 To improve the diagnostic capabilities of cardiac CT, new image reconstruction algorithms need to be tested and objectively compared. Because the performance of such algorithms can vary with different heart rates, time points within the cardiac cycle, and other image acquisition parameters such as gantry speed, an in vitro test using a dynamic phantom would be desirable for a comprehensive preclinical assessment. Assessment of the full width at half maximum (FWHM) of the point spread function (PSF) using the image of an intensity point source is a well-established approach for the evaluation of imaging methods and forms the basis of acquiring the modulation transfer function (MTF) as a widely accepted way to define spatial resolution. However, the calculation of the MTF using currently applied algorithms depends on rotational symmetry of the imaging system. 8 In the case of a moving intensity point source, this precondition is violated, and the methods to describe spatial resolution are necessarily much more complex than with a stationary object. Without rotational symmetry of the PSF, it becomes necessary to take full account of the 2-dimensional properties of the PSF, which explains why the FWHM area of the 2-di- Investigative Radiology Volume 39, Number 1, January

2 Knollmann et al Investigative Radiology Volume 39, Number 1, January 2004 mensional PSF curve should be used with the presented approach, whereas a 1-dimensional analysis sufficed with stationary intensity point sources. 9 According to transfer theory of imaging systems, the point spread function of an infinitely small object can define the system quality and help predict the system output to any given input. A 50- m tungsten wire comes close to being infinitely small in relation to CT pixel dimensions, which explains its use as a point source for the characterization of CT systems. We set out to validate the use of a dynamic phantom to compare the effective spatial resolution as the FWHM PSF of segmented cardiac CT reconstruction algorithms. MATERIALS AND METHODS A pulsatile cardiac assist device (Novacor N100 LVAS; Edwards Life Sciences, Oakland, CA) was used to simulate human ventricular wall motion at device rates of 0, 60, and 100 bpm. This device consists of 2 aluminum pusher plates that compress the artificial heart chamber and is steered by a portable microcomputer. This device was set into electrocardiogram (EKG) triggering mode using a synthetic EKG signal (Heart Simulator Model 3550; St. Jude Medical, Sylmar, CA). The device motion was mechanically transmitted to a 10-mm long piece of 50- m tungsten wire, which is commonly used to assess the in-plane spatial resolution of CT systems. The tungsten wire was placed parallel to the z-axis in the isocenter of the CT scanner and moved vertically. To synchronize the CT scanner with the dynamic phantom, we used the same simulated EKG signal that also steered the assist device. To determine the motion characteristics of our phantom, the tungsten probe was replaced by a felt tip pen, and the device motion recorded on paper using a modified EKG printer (Schwarzer Cardioscript CD 6000; Picker, Wiesbaden, Germany). To compare different cardiac CT protocols, we tested a subsecond helical CT scanner incorporating an 8-row detector (LightSpeed Ultra; GE Medical Systems, Milwaukee, WI) with our phantom. A slice thickness of 1.25 mm, 0.5 seconds gantry rotation time, and 3-mm table feed per rotation (pitch: 0.3) were selected according to the clinical protocol for coronary artery CT angiography (CTA). With this protocol, images were acquired at 120 kv tube voltage and 270 ma tube current. The scan field of view was set at its minimum of 21 cm, and the display field of view at its minimum of 96 mm. With the image matrix, pixel size was mm 2. All images were reconstructed with a standard kernel. Image reconstruction was performed using a partial scan algorithm ( snapshot segment algorithm) with a temporal resolution of 330 ms (one half rotation time plus transit 14 time of 1 fan beam width). Then, the same helical raw data set was reconstructed with a segmented 2 sector algorithm ( snapshot burst algorithm) that used image data from 2 consecutive pump cycles with a data acquisition window of 125 ms and a segmented 4-sector algorithm ( snapshot burst plus algorithm), which assembled each image from 4 consecutive device cycles with a data acquisition window of 63 ms. All data sets were repeatedly reconstructed at 0% to 90% of the RR-interval in 10% increments. For device rates of 0 and 60 bpm, the console allowed for using the snapshot segment algorithm only. Thus, comparisons of different reconstruction algorithms were performed for 100 bpm only. For visual analysis, all images were visually rated on a 6-point scale using a display window level of -780 HU and a window width of 650 HU. Images were rated by 1 person after measurement system analysis of his readings had been performed using a panel of 5 radiologists and a medical student. A set of 2 reference images was created to define visual scores of 0 and 5. Score zero was defined by the image of the tungsten wire at rest without any motion artifact. Score 1 indicated the presence of a blurred wire contour and deformation of its shape but with the image having still less than twice the diameter of the wire at rest. Score 2 indicated that the wire transection could still be located, but several streak artifacts were present in its vicinity. With a score of 3, the pin location could not be clearly determined, but emerging streak artifacts were confined to less than 3 times the wire width representation at rest. Score 4 denoted more extensive streak artifacts, and with a score of 5, the wire transsection could not be seen at all. The reproducibility of visual analysis was tested with data from a helical scan at a device rate of 60 bpm. Ten consecutive images from this data set were rated twice by an experienced radiologist, and then by a medical student and 4 radiology residents with variable experience in CT diagnosis. Observer agreement was compared using kappa statistics. 10 For numeric analysis, all image data were digitally transferred to a personal computer and processed by publicly available software (ImageJ, version 1.26t, Wayne Rasband, National Institutes of Health, Bethesda, MD). In a first step, a39 39-mm 2 region of interest (ROI) was centered on the wire transsection in a fashion that the ROI contained all or most of the visible image artifacts. The remaining portions were cropped away and the ROI saved as a text file. The text file was then loaded into an Excel spreadsheet (Microsoft, Redmond, WA) and numerically processed. In a first step, the point spread function of the wire transsection was graphically displayed. Then, the FWHM attenuation of the wire point spread function was determined Lippincott Williams & Wilkins

3 Investigative Radiology Volume 39, Number 1, January 2004 Gauging Effective Spatial Resolution To assess the reproducibility of numeric measurements, the relative variability of repeated measurements was determined as Var Abs((FWHM 1 -FWHM 2 )/((FWHM 1 FWHM 2 )/2)). with FWHM1 being the FWHM of the PSF from the original measurement and FWHM2 the FWHM of the wire PSF from a repeated measurement. The reproducibility is reported as the mean variability (Var) between 2 repeated measurements for the same imaging parameters as with the assessment of visual reproducibility (device rate 60 bpm, images reconstructed at 10 times throughout the cardiac cycle with 10% of the RR-interval interscan spacing). The numeric surrogate of spatial resolution (FWHM of the wire PSF) was compared with visual analysis by Spearman s rank correlation test. The level of statistical significance was set to 5% in all instances. To compare visual and numeric measurements, measurement system analysis 11 was performed (Minitab Statistical Software, release 13 for Windows; Minitab Inc., State College, PA). With this approach, the better measurement system is defined by a lower contribution of the measurement system to overall data variability. It is important to note that the attribute data type of visual assessment alone severely limits all further analysis of visual assessments. RESULTS The wire dynamics remained almost unchanged in systole for all device rates, whereas the diastolic interval decreased with increasing device rates. Reviewing the paper tracings of probe dynamics, 4 different sections of the device cycle could be determined with rapid systolic motion, slow early diastolic filling, almost no motion in the late diastolic phase, and rapid end-diastolic filling (Fig. 1). According to these recordings, the maximum velocity occurred with rapid diastolic filling (1500 mm/s) and ejection (380 mm/s), whereas slow diastolic filling correlated with a pin speed of 51 mm/s. The late diastolic period lasted for 470 ms at a device rate of 60 bpm and for 70 ms at 100 bpm. Visual Analysis At rest, the tungsten wire was depicted as a point in the cross-sectional plane. On visual analysis of the moving phantom, the image of the tungsten wire degenerated into linear patterns as illustrated in Figure 2. At a device rate of 60 bpm, visual assessment of image artifact severity varied between 0 (no motion artifact) at 0% of the RR-interval to 5 (most severe) at 60% of the RRinterval. There was general agreement that the wire was reproduced most accurately at 0% and 90% of the RR- FIGURE 1. (A) Experimental setup of the dynamic phantom with the Novacor circulatory assist device seen from above, and mechanical transmission of device motion to a piece of tungsten wire. (B) Paper recording of pin motion at a device rate of 100 bpm. At this rate, the diastolic interval lasts only 70 ms. Device phases comprise slow diastolic filling, diastolic standstill, rapid end-diastolic filling, and rapid contraction (device systole). The markings on the upper edge of the paper recording indicate a 1-second period. Minor lines are separated by 1 mm Lippincott Williams & Wilkins 15

4 Knollmann et al Investigative Radiology Volume 39, Number 1, January 2004 FIGURE 2. (A) Axial computed tomography (CT) image of the 50- m tungsten wire at rest with a display field of view of 96 mm. (B) Axial CT image of the tungsten wire at 100 bpm with moderate motion artifact at 60% of the RR-interval using the single-segment snapshot reconstruction algorithm. (C) Axial CT image of the tungsten wire at 100 bpm with severe motion artifact at 30% of the RR-interval using the single-segment snapshot reconstruction algorithm. (D) Point spread function of the tungsten wire CT image at rest. Attenuation is given as the difference above air attenuation. (E) Point spread function of the tungsten wire at 100 bpm with maximum motion artifact at 30% of the RR-interval using the single-segment snapshot reconstruction algorithm. Attenuation is given as the difference above air attenuation. Note that the attenuation scale has been adjusted to its maximum. 16 interval, whereas the timing of the most severe artifact was more variable (Fig. 3). However, intraobserver agreement using repeated readings was moderate with a kappa of 0.51, and interobserver agreement was poor with a kappa value of 0.19 on average (range, ) in our 6 readers. At a device rate of 100 bpm, the 1-segment snapshot algorithm resulted in visual scores in the same observer between 2 at 60% of the RR-interval and 5 at 10 and 30% of the RR-interval (Fig. 3B). Using the 2-segment snapshot burst algorithm, the score varied between 1 at 10% of the RR-interval and 5 at 30% of the RR-interval (Fig. 3B). Using the 4-segment snapshot burst plus algorithm, the score varied between 0 at 10% of the RR-interval and 5 at 30% of the RR-interval. Comparing the visual ratings of the different reconstruction algorithms, the relative merit of each technique was highly dependent on the time point of image reconstruction within the cardiac cycle. However, the 2-segment snapshot burst algorithm provided a better representation of the wire phantom than the 1-segment snapshot algorithm at most times, and the 4-segment snapshot burst plus algorithm performed better than both the 1-segment snapshot algorithm at most times, and also better than the 2-segment snapshot burst algorithm at some times. Numeric Analysis On numeric analysis, the tungsten wire at rest was reproduced with a 2% MTF of 7.8 line pairs per mm and an FWHM area of the image point spread function (PSF) of 1.30 mm 2. The FWHM area at a device rate of 60 bpm ranged from 1.51 mm 2 at 10% of the RR-interval to mm 2 at 60% of the RR-interval (Fig. 4). At a device rate of 100 bpm, the 1-segment snapshot algorithm resulted in FWHM areas between 2.00 mm 2 at 90% 2003 Lippincott Williams & Wilkins

5 Investigative Radiology Volume 39, Number 1, January 2004 Gauging Effective Spatial Resolution FIGURE 3. (A) Visual assessment of multirow detector computed tomography (MDCT) artifact severity at a device rate of 60 bpm. Image numbers 1 to 10 correspond to reconstruction times of zero to 90% of the RR-interval. Interobserver variability at all times indicates that visual assessment entails significant observer bias. (B) Visual assessment of MDCT artifact severity at a device rate of 100 bpm for single- (SSEG), 2- (SB2), and 4-segement (SB4) reconstruction algorithms. Image numbers 1 to 10 correspond to reconstruction times of zero to 90% of the RR-interval. of the RR-interval and mm 2 at 80% of the RR-interval (Fig. 4B). Using the 2-segment snapshot burst algorithm, the FWHM varied between 1.41 mm 2 at 20% of the RR-interval and mm 2 at 30% of the RR-interval. Using the 4-segment snapshot burst plus algorithm, FWHM varied between 1.37 mm 2 at 20% of the RR-interval and mm 2 at 30% of the RR-interval. Comparing the FWHM values of the different reconstruction algorithms, the relative merit of each technique was highly dependent on the time point of image reconstruction within the cardiac cycle. The 2-segment snapshot burst algorithm provided a better representation of the wire phantom than the 1-segment snapshot algorithm at only 3 of the 10 time points tested, and the 4-segment snapshot burst plus algorithm performed better than both the 1-segment snapshot algorithm at 4 times, and also better than the 2-segment snapshot burst algorithm at 4 times. Both segmented algorithms resulted in much worse image quality than the single-segment snapshot technique at 30% of the RRinterval, however. FIGURE 4. (A) Effective spatial accuracy of multirow detector computed tomography as the full width at half maximum area of the point spread function at a device rate of 60 bpm. Image numbers 1 to 10 correspond to reconstruction times of 0 to 90% of the RR-interval. (B) Effective spatial resolution of helical computed tomography protocols at a device rate of 100 bpm for single- (SSEG), 2- (SB2), and 4-segment (SB4) reconstruction algorithms. Image numbers 1 to 10 correspond to reconstruction times of 0 to 90% of the RR-interval. The reproducibility of numerical effective spatial accuracy as per the FWHM area of the tungsten wire PSF was determined as a mean variability of 0.05 (range, ). Correlation of Visual With Numeric Estimates of Image Quality On Spearman s rank correlation, the visual estimates closely correlated with the numeric surrogate FWHM (P , rho 0.841), confirming that the numeric surrogate for image quality increases with visually more severe artifacts and decreases with better visual estimates of image quality (Fig. 5). On gauging repeatability and reproducibility (Gage R&R), it was found that total Gage R&R of the numeric analysis method contributed only 5.9% of overall study variability, which indicates excellent performance of the numeric measurement system. With this method, 24 distinct categories could have been discerned, whereas more than 5 distinct categories would indicate that the test is acceptable. For the visual analysis method, using Attribute Gage R&R analysis, it was found that only 3.3% of all assessments agreed with each other Lippincott Williams & Wilkins 17

6 Knollmann et al Investigative Radiology Volume 39, Number 1, January 2004 FIGURE 5. Correlation of effective spatial resolution as expressed by full width at half maximum area of the wire point spread function and the visual estimate of image quality. DISCUSSION Assessment of cardiac CT image quality by calculating the effective spatial resolution as the FWHM area of a CT PSF correlated well with visual estimates of image quality, albeit at a higher level of reproducibility. Using this method, distinct advantages of multisegment image reconstruction algorithms for coronary artery CTA could be demonstrated. Because the time of image reconstruction can be adjusted retrospectively with EKG-gated helical acquisition techniques, the reconstruction time with the best possible image quality can be chosen even after the patient has completed his examination. Thus, we suggest citing the best effective spatial resolution achieved with a specific acquisition technique for comparison. For the reconstruction algorithms used in this investigation, the best effective spatial resolution was 1.51 mm 2 at a device rate of 60 bpm, compared with 2.00 mm 2 at 100 bpm with the single-segment snapshot algorithm. The consistency of the lowest artifact severity throughout the cardiac cycle is less significant for retrospectively gated helical acquisition methods, because the time of best image quality can be retrospectively determined form the initial data set. It is also clinically routine for helical cardiac examinations to perform tentative image reconstructions at many different time points throughout the cardiac cycle to determine which time point offers the optimal image quality. Like in clinical practice, the best image quality that can be derived from a scan is dependent on the reconstruction time with least image artifacts. We therefore compared the best image quality that could be possibly achieved with each algorithm, and thus came to the conclusion that the multisegment algorithms offer an image quality benefit. 18 Our experiments imply that increased temporal resolution could cause more severe motion artifacts at selected systolic time points, which can be explained by the inability of a low-resolution technique to capture the entire motion. This effect is illustrated by the deterioration of spatial resolution at 30% of the RR-interval at 100 bpm with the 2- and 4-segment snapshot burst algorithms. The single-segment snapshot algorithm yielded a clearly better spatial resolution under these conditions, although its nominal temporal resolution was only a half and one fourth of the other methods temporal resolution, respectively. Comparisons of the FWHM area with visual estimates of image quality reveal that severe artifacts are consistently noted once the FWHM area exceeds 3 mm 2.WithaFWHM area of less than 2 mm 2, image artifacts could be visible, but the location of the object remains well defined. Using multisegmented image reconstruction algorithms allows us to improve the FWHM area even at a device rate of 100 bpm from 2.0 mm 2 with the single-segment snapshot technique to 1.41 mm 2 with the 2-segment snapshot burst algorithm or to 1.37 mm 2 with the 4-segment snapshot burst plus algorithm. However, the success of multisegmented reconstruction algorithms depends on consistently stable RR-intervals and constant cardiac preload, because small differences of the reconstruction time could cause severe degradation of image quality. 12 With the present setup of our dynamic phantom, device motion was perfectly regular as a result of its microprocessor-controlled rhythm. Yet, the method can potentially reproduce any degree of sinus arrhythmia if the source EKG signal is modified accordingly. After clinical observations, the use of a 2-segment reconstruction algorithm could not prevent the degeneration of image quality with increasing heart rates. 13 This observation indicates that sinus arrhythmia and variations in cardiac preload play an important role when image data are collected from more than a single heartbeat. The validity of our dynamic phantom is confirmed by the multiphasic phantom dynamics, similar to human coronary artery dynamics. The device velocity during slow diastolic filling is in the range of what others have found as the mean velocity of human coronary arteries by electron beam CT, MRI, and cineangiography. 14 Similarly, the systolic velocity in our phantom is within the range of the maximum human coronary artery velocity. The late diastolic filling boost exceeded the normal velocity of human coronary arteries and provides an opportunity to investigate the effect of higher velocities with the same setup. However, this single 10-ms moment of supernatural velocity should not be used to predict the extent of motion artifact in human coronary imaging. Like in the human cardiac cycle, systolic dynamics are mostly unaffected by heart rate, and the shortening of the RR-interval is achieved mostly by decreasing the diastolic 2003 Lippincott Williams & Wilkins

7 Investigative Radiology Volume 39, Number 1, January 2004 Gauging Effective Spatial Resolution period. Although the device dynamics differ from human coronary artery motion patterns in some details, the range of velocities encountered in this setup should represent the clinical setting sufficiently to allow for its use as an in vitro test bed. However, the exact timing of the highest spatial accuracy relative to the cardiac cycle does not correspond to clinical observations, which is the result of the 150-ms start delay of the assist device after the trigger EKG R-wave. This delay can be freely selected in future investigations. To apply the exact timing sequence to clinical practice, clinical confirmation of the timing would be required. In earlier investigations, the ideal time for reconstruction was highly variable, and the authors recommended to perform image reconstructions at multiple times to determine the best time point on an individual basis. 15 The use of dynamic phantoms for the assessment of cardiac CT techniques has been described by several investigators before. 16,17 However, these studies relied on a visual assessment of image quality, which implies significant observer bias. In our experiments, interobserver agreement on visual assessment was generally poor, and the extent of interobserver disagreement was, in some cases, as large as the differences observed by a single reader between different time points or acquisition methods. Thus, the extent of observer bias has the potential to obscure resolution differences that can be detected by numeric analysis. Also, visual estimates remain semiquantitative by necessity, whereas the numeric surrogate marker presented here can be used as a true interval scale variable. In 1 other investigation, a cardiac assist device has been used as a dynamic phantom for the quantification of coronary calcium. 18 Although a comparison of visual estimates of artifact severity with the presented numeric method might seem challenging at first, understanding of measurement system analysis still allows for a valid comparison. With this approach, the observed variation of both visual and numeric indicators of image quality can be analyzed by determining the reproducibility of each indicator, which addresses that portion of the observed variability that is the result of the measurement system. Reviewing Figures 3 and 4, it is obvious that the portion of image quality variability ascribable to the measurement system is much larger for visual evaluation than for the presented numeric method with its mean variability of 5% on repeated testing. Thus, the numeric method is better suited to detect smaller differences of image quality, which would be obscured by the relatively poor reproducibility of the visual score method. We conclude that the use of moving tungsten wire to determine the FWHM area of the wire PSF could be used as an objective and valid measure of effective spatial accuracy for the evaluation of cardiac CT protocols. This approach extends the concept of using an intensity point source for the comparison of CT spatial resolution to moving objects. The method has the potential to compare different cardiac CT protocols in vitro to foster the development of new CT techniques for noninvasive coronary imaging. REFERENCES 1. Klingenbeck-Regn K, Schaller S, Flohr T, et al. Subsecond multi-slice computed tomography: basics and applications. Eur J Radiol. 1999;31: Hu H. Multi-slice helical CT: scan and reconstruction. Med Phys. 1999;26: Achenbach S, Ulzheimer S, Baum U, et al. Noninvasive coronary angiography by retrospectively ECG-gated multislice CT. Circulation. 2000;102: Ohnesorge B, Flohr T, Becker C, et al. Cardiac imaging by means of electrocardiographically gated multislice spiral CT: initial experiences. Radiology. 2000; 217: Nieman K, Oudkerk M, Rensing BJ, et al. Coronary angiography with multi-slice computed tomography. Lancet. 2001;357: Achenbach S, Giesler T, Ropers D, et al. Detection of coronary artery stenoses by contrast-enhanced, retrospectively electrocardiographicallygated, multislice spiral computed tomography. Circulation. 2001;103: Ohnesorge B, Flohr T, Fischbach R, et al. Reproducibility of coronary calcium quantification in repeat examinations with retrospectively ECGgated multisection spiral CT. Eur Radiol. 2002;12: Rossmann K. Point spread function and modulation transfer function. Radiology. 1969;93: Assimakopulos PA, Boyd DP, Jaschke W, et al. Spatial resolution of computed tomographic images. Invest Radiol. 1986;21: Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33: Breyfogle FW III. Implementing Six Sigma Smarter Solutions Using Statistical Methods. New York: John Wiley and Sons; 1999: Hong C, Becker CR, Huber A, et al. ECG-gated reconstructed multidetector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology. 2001;220: Vogl TJ, Abolmaali ND, Diebold T, et al. Techniques for the detection of coronary atherosclerosis: multi-detector row CT coronary angiography. Radiology. 2002;223: Achenbach S, Ropers D, Holle J, et al. In-plane coronary arterial motion velocity: measurement with electron-beam CT. Radiology. 2000;216: Kopp AF, Schroeder S, Kuettner A, et al. Coronary arteries: retrospectively ECG-gated multi-detector row CT angiography with selective optimization of the image reconstruction window. Radiology. 2001;221: Knollmann F, Halfmann R, Regn J, et al. Motion artifacts in cardiac CT. Acta Radiol. 1999;40: Wicky S, Rosol M, Hamberg LM, et al. Evaluation of retrospective multisector and half scan ECG-gated multi-detector cardiac CT protocols with moving phantom.j Comput Assist Tomogr. 2003;27: Hopper KD, Strollo DC, Mauger DT. Comparison of electron-beam and ungated helical CT in detecting coronary arterial calcification by using a working heart phantom and artificial coronary arteries. Radiology. 2002; 222: Lippincott Williams & Wilkins 19

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Coronary CT Angiography

Coronary CT Angiography Coronary CT Angiography Byoung Wook Choi, M.D. Department of Diagnostic Radiology Yonsei University College of Medicine, Severance Hospital E mail : bchoi@yumc.yonsei.ac.kr Abstract With the advent of

More information

Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure. Coronary Artery Calcium Volume?

Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure. Coronary Artery Calcium Volume? Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure Coronary Artery Calcium Volume? 1 Objective: The purpose of this study is to investigate how accurately we can measure CAC volume

More information

Sang Ho Lee, Byoung Wook Choi, Hee-Joung Kim*, Member, IEEE, Haijo Jung, Hye-Kyung Son, Won-Suk Kang, Sun Kook Yoo, Kyu Ok Choe, Hyung Sik Yoo

Sang Ho Lee, Byoung Wook Choi, Hee-Joung Kim*, Member, IEEE, Haijo Jung, Hye-Kyung Son, Won-Suk Kang, Sun Kook Yoo, Kyu Ok Choe, Hyung Sik Yoo Analysis of Heart Rate and its Variation Affecting Image Quality and Optimized Reconstruction Window in Retrospective ECG-gated Coronary Angiography Using Multi-detector Row CT Sang Ho Lee, Byoung Wook

More information

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Cardiac CT - Coronary Calcium Basics Workshop II (Basic) Cardiac CT - Coronary Calcium Basics Workshop II (Basic) J. Jeffrey Carr, MD, MSCE Dept. of Radiology & Public Health Sciences Wake Forest University School of Medicine Winston-Salem, NC USA No significant

More information

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Int J Cardiovasc Imaging (2008) 24:547 556 DOI 10.1007/s10554-007-9282-0 ORIGINAL PAPER Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Jaap M. Groen Æ Marcel J.

More information

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 1, FEBRUARY

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 1, FEBRUARY IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 1, FEBRUARY 2004 225 Analysis of the Heart Rate and Its Variation Affecting Image Quality and Optimized Reconstruction Window in Retrospective ECG-Gated

More information

Studies with electron beam computed tomography (EBCT) Imaging

Studies with electron beam computed tomography (EBCT) Imaging Imaging Predictive Value of 16-Slice Multidetector Spiral Computed Tomography to Detect Significant Obstructive Coronary Artery Disease in Patients at High Risk for Coronary Artery Disease Patient- Versus

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial Poster No.: C-0595 Congress: ECR 2013 Type: Authors:

More information

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing Cademartiri et al. Heart Rhythm Irregularities on MDCT Angiography Cardiac Imaging Original Research A C M E D E N T U R I C A L I M A G I N G AJR 2006; 186:634 638 0361 803X/06/1863 634 American Roentgen

More information

General Cardiovascular Magnetic Resonance Imaging

General Cardiovascular Magnetic Resonance Imaging 2 General Cardiovascular Magnetic Resonance Imaging 19 Peter G. Danias, Cardiovascular MRI: 150 Multiple-Choice Questions and Answers Humana Press 2008 20 Cardiovascular MRI: 150 Multiple-Choice Questions

More information

Purpose. Methods and Materials

Purpose. Methods and Materials Comparison of iterative and filtered back-projection image reconstruction techniques: evaluation of heavily calcified vessels with coronary CT angiography Poster No.: C-1644 Congress: ECR 2011 Type: Scientific

More information

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT Eur Radiol (2006) 16: 1964 1972 DOI 10.1007/s00330-006-0262-x CARDIAC Sebastian Leschka Lars Husmann Lotus M. Desbiolles Oliver Gaemperli Tiziano Schepis Pascal Koepfli Thomas Boehm Borut Marincek Philipp

More information

Department of Cardiology, Grosshadern Clinic, University of Munich, Marchioninistrasse 15, Munich, Germany. Department of Cardiology,

Department of Cardiology, Grosshadern Clinic, University of Munich, Marchioninistrasse 15, Munich, Germany. Department of Cardiology, Eur Radiol (2002) 12:1532 1540 DOI 10.1007/s00330-002-1394-2 CARDIAC B. Ohnesorge T. Flohr R. Fischbach A. F. Kopp A. Knez S. Schröder U. J. Schöpf A. Crispin E. Klotz M. F. Reiser C. R. Becker Reproducibility

More information

Non-invasive coronary angiography with high resolution multidetector-row computed tomography

Non-invasive coronary angiography with high resolution multidetector-row computed tomography European Heart Journal (2002) 23, 1714 1725 doi:10.1053/euhj.2002.3264, available online at http://www.idealibrary.com on Non-invasive coronary angiography with high resolution multidetector-row computed

More information

Electron Beam CT of the Heart

Electron Beam CT of the Heart CHAPTER 2 / ELECTRON BEAM TOMOGRAPHY 15 2 Electron Beam CT of the Heart DAVID G. HILL, PhD INTRODUCTION Electron beam tomography (EBT)* was developed by Douglas Boyd, PhD, and his associates at Imatron,

More information

Angio-CT: heart and coronary arteries

Angio-CT: heart and coronary arteries European Journal of Radiology 45 (2003) S32/S36 www.elsevier.com/locate/ejrad Angio-CT: heart and coronary arteries Andreas F. Kopp * Tübingen University Hospital, Tübingen, Germany Received 22 November

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

Coronary Calcium Screening Using Low-Dose Lung Cancer Screening: Effectiveness of MDCT with Retrospective Reconstruction

Coronary Calcium Screening Using Low-Dose Lung Cancer Screening: Effectiveness of MDCT with Retrospective Reconstruction Cardiac Imaging Original Research Kim et al. Coronary Calcium Screening Using Lung Cancer Screening Cardiac Imaging Original Research Sung Mok Kim 1 Myung Jin Chung 1 Kyung Soo Lee 1 Yeon Hyun Choe 1 Chin

More information

Variability of Repeated Coronary Artery Calcium Scoring and Radiation. Dose on 64-slice and 16-slice CT by Prospective

Variability of Repeated Coronary Artery Calcium Scoring and Radiation. Dose on 64-slice and 16-slice CT by Prospective Title Page: Variability of Repeated Coronary Artery Calcium Scoring and Radiation Dose on 64-slice and 16-slice CT by Prospective Electrocardiograph-triggered Axial and Retrospective Electrocardiograph-gated

More information

ECG-Gated 16-MDCT of the Coronary Arteries: Assessment of Image Quality and Accuracy in Detecting Stenoses

ECG-Gated 16-MDCT of the Coronary Arteries: Assessment of Image Quality and Accuracy in Detecting Stenoses ECG-Gated 16-MDCT of the Coronary Arteries ECG-Gated 16-MDCT of the Coronary Arteries: Assessment of Image Quality and Accuracy in Detecting Stenoses Martin Heuschmid 1 Axel Kuettner 1 Stephen Schroeder

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Junfu Liang 1,2, Hui Wang 1, Lei Xu 1, Li Dong 1, Zhanming

More information

Coronary angiography is the standard way of visualizing

Coronary angiography is the standard way of visualizing Clinical Investigation and Reports Coronary Artery Fly-Through Using Electron Beam Computed Tomography Peter M.A. van Ooijen, MSc; Matthijs Oudkerk, MD, PhD; Robert J.M. van Geuns, MD; Benno J. Rensing,

More information

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design 2 Cardiovascular Computed Tomography: Current and Future Scanning System Design Wm. Guy Weigold Introduction The heart can be visualized in gross form on any standard chest computed tomography (CT), but

More information

Noncoronary Cardiac MDCT

Noncoronary Cardiac MDCT Noncoronary Cardiac MDCT David A. Bluemke, M.D., Ph.D. Professor, of Radiology and Medicine Johns Hopkins University School of Medicine Baltimore, Maryland Toshiba Disclosures Grant support Noncoronary

More information

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling Investigations and research Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling J. Wang Z. Ying V. Yao L. Ciancibello S. Premraj S. Pohlman

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Uozu city/jp, Minatoku, Tokyo/JP Bones, Extremities, CT, Surgery, Physics, Artifacts, Image verification /ecr2014/C-0462

Uozu city/jp, Minatoku, Tokyo/JP Bones, Extremities, CT, Surgery, Physics, Artifacts, Image verification /ecr2014/C-0462 Metal Artifact Reduction Algorithm enablesreduce metal artifacts and improvement of diagnosis in the postoperative Pedicle screws implant for Spinal Fusion: A Phantom Study Poster No.: C-0462 Congress:

More information

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16.

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16. NIH Public Access Author Manuscript Published in final edited form as: Acad Radiol. 2008 January ; 15(1): 40 48. doi:10.1016/j.acra.2007.07.023. Analysis of Heart Rate and Heart Rate Variation During Cardiac

More information

Cardiac Imaging Tests

Cardiac Imaging Tests Cardiac Imaging Tests http://www.medpagetoday.com/upload/2010/11/15/23347.jpg Standard imaging tests include echocardiography, chest x-ray, CT, MRI, and various radionuclide techniques. Standard CT and

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Cardiac CTA Prospective Gating Broad Beam

Cardiac CTA Prospective Gating Broad Beam Cardiac CTA Prospective Gating Broad Beam ACQUISITION- Broad Beam Gating: Prospective Non Contrast Scan-Calcium Score Patient Position Supine Feet First into Gantry Heart Isocenter Scanogram AP and Lateral

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

2004;77:800 4 MSCT OF CORONARY ARTERY BYPASS GRAFTS. Results. CABG With Adequate Diagnostic Quality

2004;77:800 4 MSCT OF CORONARY ARTERY BYPASS GRAFTS. Results. CABG With Adequate Diagnostic Quality Isotropic Half-Millimeter Angiography of Coronary Artery Bypass Grafts With 16-Slice Computed Tomography Marc Dewey, MD, Alexander Lembcke, MD, Christian Enzweiler, MD, Bernd Hamm, MD, and Patrik Rogalla,

More information

Soft and Intermediate Plaques in Coronary Arteries: How Accurately Can We Measure CT Attenuation Using 64-MDCT?

Soft and Intermediate Plaques in Coronary Arteries: How Accurately Can We Measure CT Attenuation Using 64-MDCT? 64-MDCT Measurement of Coronary Artery Plaques Cardiac Imaging Original Research Jun Horiguchi 1 Chikako Fujioka 1 Masao Kiguchi 1 Yun Shen 2 Christian E. Althoff 3,4 Hideya Yamamoto 5 Katsuhide Ito 3

More information

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography Supalerk Pattanaprichakul, MD 1, Sutipong Jongjirasiri, MD 2, Sukit Yamwong, MD 1, Jiraporn Laothammatas,

More information

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography 633 CARDIOVASCULAR MEDICINE Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography A W Leber, A Knez, C Becker, A Becker, C White, C Thilo, M Reiser,

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Radiation dose of cardiac CT what is the evidence?

Radiation dose of cardiac CT what is the evidence? Eur Radiol (2009) 19: 1311 1315 DOI 10.1007/s00330-009-1312-y CARDIAC Hatem Alkadhi Radiation dose of cardiac CT what is the evidence? Received: 10 December 2008 Revised: 2 January 2009 Accepted: 12 January

More information

MR Advance Techniques. Cardiac Imaging. Class IV

MR Advance Techniques. Cardiac Imaging. Class IV MR Advance Techniques Cardiac Imaging Class IV Heart The heart is a muscular organ responsible for pumping blood through the blood vessels by repeated, rhythmic contractions. Layers of the heart Endocardium

More information

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology,

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology, Chapter 4 Impact of Coronary Calcium Score on Diagnostic Accuracy of Multislice Computed Tomography Coronary Angiography for Detection of Coronary Artery Disease Gabija Pundziute, 1,3 Joanne D. Schuijf,

More information

Measurement of Respiratory and Cardiac Motion Using a Multi Antenna Continuous Wave Radar Operating in the Near Field

Measurement of Respiratory and Cardiac Motion Using a Multi Antenna Continuous Wave Radar Operating in the Near Field Measurement of Respiratory and Cardiac Motion Using a Multi Antenna Continuous Wave Radar Operating in the Near Field Florian Pfanner 1,2, Thomas Allmendinger 2, Thomas Flohr 2, and Marc Kachelrieß 1,3

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Towards Cardiac C-Arm Computed Tomography Günter Lauritsch*, Jan Boese, Lars Wigström, Herbert Kemeth, and Rebecca Fahrig

Towards Cardiac C-Arm Computed Tomography Günter Lauritsch*, Jan Boese, Lars Wigström, Herbert Kemeth, and Rebecca Fahrig 922 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 7, JULY 2006 Towards Cardiac C-Arm Computed Tomography Günter Lauritsch*, Jan Boese, Lars Wigström, Herbert Kemeth, and Rebecca Fahrig Abstract Cardiac

More information

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Original Article Open Access The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Ziqiao Lei 1, Jin Gu 2, Qing Fu 3, Heshui Shi 4, Haibo Xu 5, Ping Han 6, Jianming

More information

X-ray coronary angiography is considered the diagnostic

X-ray coronary angiography is considered the diagnostic Noninvasive Coronary Angiography by Retrospectively ECG-Gated Multislice Spiral CT Stephan Achenbach, MD; Stefan Ulzheimer, MS; Ulrich Baum, MD; Marc Kachelrieß, PhD; Dieter Ropers, MD; Tom Giesler, MD;

More information

Aims and objectives. Page 2 of 13

Aims and objectives. Page 2 of 13 Evaluation of right heart function with use of retrospectively ECG- gated 64-slice low dose computed tomography in subjects with and without pulmonary hypertension Poster No.: C-2562 Congress: ECR 2015

More information

On the feasibility of speckle reduction in echocardiography using strain compounding

On the feasibility of speckle reduction in echocardiography using strain compounding Title On the feasibility of speckle reduction in echocardiography using strain compounding Author(s) Guo, Y; Lee, W Citation The 2014 IEEE International Ultrasonics Symposium (IUS 2014), Chicago, IL.,

More information

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography Journal of the American College of Cardiology Vol. 46, No. 3, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.05.056

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

Technology Assessment Institute: Summit on CT Dose Cardiac CT - Optimal Use of Evolving Scanner Technologies

Technology Assessment Institute: Summit on CT Dose Cardiac CT - Optimal Use of Evolving Scanner Technologies Cardiac CT - Optimal Use of Evolving Scanner Technologies P. Rogalla, M.D. Dept. of Medical Imaging University of Toronto Special thanks to Dr. Lembcke, Dr. Hein Charité, Berlin Disclosures No salaries

More information

CT SCAN PROTOCOL. Shoulder

CT SCAN PROTOCOL. Shoulder CT SCAN PROTOCOL Shoulder Purpose and Summary CT images made with this protocol are used to provide the orthopedic surgeon with a detailed 3D anatomical reconstruction of the patient s scapula and proximal

More information

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research

More information

Cardiac CT Techniques in Neonates (and infants)

Cardiac CT Techniques in Neonates (and infants) Cardiac CT Techniques in Neonates (and infants) Siddharth P. Jadhav, MD Director, Body CT and MRI Edward B. Singleton Department of Pediatric Radiology Texas Children s Hospital Disclosures None Objectives

More information

Radiation doses, noise properties, and in-plane spatial resolutions for CT coronary angiography using different CT scanners: Phantom study

Radiation doses, noise properties, and in-plane spatial resolutions for CT coronary angiography using different CT scanners: Phantom study ORIGINAL ARTICLES Radiation doses, noise properties, and in-plane spatial resolutions for CT coronary angiography using different CT scanners: Phantom study Kosuke Matsubara 1, Tadanori Takata 2, Hiroshige

More information

European Journal of Radiology

European Journal of Radiology European Journal of Radiology 82 (2013) e58 e63 Contents lists available at SciVerse ScienceDirect European Journal of Radiology journa l h o me pa ge: www.elsevier.com/locate/ejrad Calcium score of small

More information

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease Horizon Scanning Technology Summary National Horizon Scanning Centre Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease April 2007 This technology summary is based

More information

Research Article. Open Access. Jin-Da WANG, Hua-Wei ZHANG, Qian XIN, Jun-Jie YANG, Zhi-Jun SUN, Hong-Bin LIU, Lian CHEN, Luo-Shan DU, Yun-Dai CHEN

Research Article. Open Access. Jin-Da WANG, Hua-Wei ZHANG, Qian XIN, Jun-Jie YANG, Zhi-Jun SUN, Hong-Bin LIU, Lian CHEN, Luo-Shan DU, Yun-Dai CHEN Journal of Geriatric Cardiology (2014) 11: 39 43 2014 JGC All rights reserved; www.jgc301.com Research Article Open Access Safety and efficacy of intravenous esmolol before prospective electrocardiogram-triggered

More information

Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1

Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1 Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1 Ethan J. Halpern, MD, David J. Halpern, Jeffrey H. Yanof, PhD, Sigal Amin-Spector, PhD, David Fischman, MD, Galit Aviram,

More information

B. CT protocols for the spine

B. CT protocols for the spine B. CT protocols for the spine Poster No.: A-003 Congress: ECR 2010 Type: Invited Speaker Topic: Neuro Authors: B. Tins; Oswestry/UK Keywords: CT, spine, diagnostic imaging protocol DOI: 10.1594/ecr2010/A-003

More information

Adapted Transfer Function Design for Coronary Artery Evaluation

Adapted Transfer Function Design for Coronary Artery Evaluation Adapted Transfer Function Design for Coronary Artery Evaluation Sylvia Glaßer 1, Steffen Oeltze 1, Anja Hennemuth 2, Skadi Wilhelmsen 3, Bernhard Preim 1 1 Department of Simulation and Graphics, University

More information

Low-Dose Computed Tomography Coronary Angiography With Prospective Electrocardiogram Triggering

Low-Dose Computed Tomography Coronary Angiography With Prospective Electrocardiogram Triggering Journal of the American College of Cardiology Vol. 57, No. 3, 2011 2011 by the American College of Cardiology Foundation ISSN 0735-1097/$36.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2010.08.634

More information

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T. The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume

More information

Biases affecting tumor uptake measurements in FDG-PET

Biases affecting tumor uptake measurements in FDG-PET Biases affecting tumor uptake measurements in FDG-PET M. Soret, C. Riddell, S. Hapdey, and I. Buvat Abstract-- The influence of tumor diameter, tumor-tobackground activity ratio, attenuation, spatial resolution,

More information

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006). Feasibility, image quality and radiation dose of coronary CT angiography (CCTA) in patients with atrial fibrillation using a new generation 256 multi-detector CT (MDCT) Poster No.: C-2378 Congress: ECR

More information

Computed Tomography Imaging of the Coronary Arteries

Computed Tomography Imaging of the Coronary Arteries Chapter 6 Computed Tomography Imaging of the Coronary Arteries G.J. Pelgrim, M. Oudkerk and R. Vliegenthart Additional information is available at the end of the chapter http://dx.doi.org/10.5772/54044

More information

CT Scanning Protocol For V2R Guided Surgery Solutions

CT Scanning Protocol For V2R Guided Surgery Solutions CT Scanning Protocol For V2R Guided Surgery Solutions 2 V2R CT Scanning Protocol \\ Contents Contents General requirements... 3 V2R Dual Scan Protocol... 5 V2R Single Scan Protocol... 8 Overview... 10

More information

Gender differences in CT calcium scoring: A phantom study

Gender differences in CT calcium scoring: A phantom study Gender differences in CT calcium scoring: A phantom study Nicholas Petrick, Qin Li, Benjamin Berman, Marios A Gavrielides, Rongping Zeng, Berkman Sahiner CDRH/OSEL/DIDSR U.S. Food and Drug Administration

More information

Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control

Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control Eur Radiol (2006) 16: 2739 2747 DOI 10.1007/s00330-006-0474-0 CARDIAC Hans Scheffel Hatem Alkadhi André Plass Robert Vachenauer Lotus Desbiolles Oliver Gaemperli Tiziano Schepis Thomas Frauenfelder Thomas

More information

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography Journal of the American College of Cardiology Vol. 52, No. 18, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.07.048

More information

Computed tomography. Department of Radiology, University Medical School, Szeged

Computed tomography. Department of Radiology, University Medical School, Szeged Computed tomography Department of Radiology, University Medical School, Szeged voxel +1-4 +2 +5 +3 +1 0-2 pixel -2 0 +1-4 -6 +5 +2 +1 Department of Radiology, University Medical School, Szeged

More information

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Poster No.: B-0742 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Paper

More information

Multislice CT angiography in cardiac imaging. Part III: radiation risk and dose reduction

Multislice CT angiography in cardiac imaging. Part III: radiation risk and dose reduction R e v i e w A r t i c l e Singapore Med J 2010; 51(5) : 374 Multislice CT angiography in cardiac imaging. Part III: radiation risk and dose reduction Sun Z, Ng K H Discipline of Medical Imaging, Department

More information

Recent developments in cardiac CT

Recent developments in cardiac CT REVIEW Recent developments in cardiac CT With the introduction of 64-multidetector row CT, coronary CT angiography has become a clinical tool, owing to improved image quality and reduced breath-hold time,

More information

Institutional review board approval was obtained prior to the start of this study.

Institutional review board approval was obtained prior to the start of this study. Lower Limb Alignment and Length Measurements - Comparison of Computed Tomography, Upright Full-Length Conventional Radiography and Upright Biplanar Linear-Low Dose X-ray Scanner Poster No.: C-1382 Congress:

More information

Multidetector-row cardiac CT: diagnostic value of calcium scoring and CT coronary angiography in patients with symptomatic, but atypical, chest pain

Multidetector-row cardiac CT: diagnostic value of calcium scoring and CT coronary angiography in patients with symptomatic, but atypical, chest pain Eur Radiol (2004) 14:169 177 DOI 10.1007/s00330-003-2197-9 CARDIAC Christopher Herzog Martina Britten Joern O. Balzer M. G. Mack Stefan Zangos Hanns Ackermann Volker Schaechinger Stefan Schaller Thomas

More information

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography African Journal of Biotechnology Vol. 10(19), pp. 3854-3858, 9 May, 2011 Available online at http://www.academicjournals.org/ajb DOI: 10.5897/AJB10.983 ISSN 1684 5315 2011 Academic Journals Full Length

More information

SYNTAX III REVOLUTION Trial Press briefing conference. Prof. Patrick W. Serruys MD, PhD Principal Investigator Imperial College of London

SYNTAX III REVOLUTION Trial Press briefing conference. Prof. Patrick W. Serruys MD, PhD Principal Investigator Imperial College of London SYNTAX III REVOLUTION Trial Press briefing conference Prof. Patrick W. Serruys MD, PhD Principal Investigator Imperial College of London Title: Coronary Computed Tomography Angiography for Heart Team Decision-making

More information

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Poster No.: C-1207 Congress: ECR 2013 Type: Scientific Exhibit Authors: Y. Shimomiya, M. Kondo, M.

More information

Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT

Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT The Harvard community has made this article openly available. Please share how this access benefits you. Your story

More information

Spiral Multislice Computed Tomography Coronary Angiography: A Current Status Report

Spiral Multislice Computed Tomography Coronary Angiography: A Current Status Report Clin. Cardiol. 30, 437 442 (2007) Spiral Multislice Computed Tomography Coronary Angiography: A Current Status Report P. J. De Feyter, M.D., PH.D., W. B. Meijboom, M.D., A. Weustink, M.D., C. Van Mieghem,

More information

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT Ralf Koos 1 Andreas Horst Mahnken 2 Anil Martin Sinha 1 Joachim Ernst Wildberger 2 Rainer Hoffmann 1 Harald Peter Kühl 1 Received March 3, 2004; accepted after revision May 18, 2004. 1 Department of Cardiology,

More information

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA SYMPOSIA Image Acquisition and Interpretation Josef Matthias Kerl, MD,*w Lars K. Hofmann, MD,z Christian Thilo, MD,*y Thomas J. Vogl, MD,w Philip Costello, MD,* and U. Joseph Schoepf, MD*y Abstract: Computed

More information

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference.

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference. Rotation: Imaging 1 Imaging 1 provides COCATS Level 1 experience for nuclear cardiology (including SPECT and PET) and cardiac CT. Fellows will administer, process, and read cardiac nuclear studies with

More information

Image quality and, hence, the diagnostic value of cardiac. Imaging

Image quality and, hence, the diagnostic value of cardiac. Imaging Imaging Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily Practice Impact of Different Scanning Protocols on Effective Dose Estimates Jörg Hausleiter, MD; Tanja Meyer, MD; Martin

More information

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03 Introduction Cardiac Imaging Modalities Arnaud Bistoquet 12/19/03 Coronary heart disease: the vessels that supply oxygen-carrying blood to the heart, become narrowed and unable to carry a normal amount

More information

Dr Felix Keng. Imaging of the heart is technically difficult because: Role of Cardiac MSCT. Current: Cardiac Motion Respiratory Motion

Dr Felix Keng. Imaging of the heart is technically difficult because: Role of Cardiac MSCT. Current: Cardiac Motion Respiratory Motion Siemens Philips Dr Felix Keng GE Toshiba Role of Cardiac MSCT Current: Structural / congenital heart imaging Extra-cardiac / Great vessel imaging Volumes and ejection fractions (cine + gating) Calcium

More information

Fulfilling the Promise

Fulfilling the Promise Fulfilling the Promise of Cardiac MR Non-contrast, free-breathing technique generates comprehensive evaluation of the coronary arteries By Maggie Fung, MR Cardiovascular Clinical Development Manager; Wei

More information

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions.

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions. Lesson 07: Ultrasound Transducers This lesson contains 73 slides plus 16 multiple-choice questions. This lesson was derived from pages 33 through 42 in the textbook: Ultrasound Transducers Ultrasound Transducers

More information

Feasibility of a New C-arm angiography (DYNA-CT) based three-dimensional algorithm in combination with myocardial perfusion assessment

Feasibility of a New C-arm angiography (DYNA-CT) based three-dimensional algorithm in combination with myocardial perfusion assessment Feasibility of a New C-arm angiography (DYNA-CT) based three-dimensional algorithm in combination with myocardial perfusion assessment H. Rittger*, A.M.Sinha, J. Rieber±, G. Lauritsch+, J. Brachmann *Universitätsklinik

More information

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006

, David Stultz, MD. Cardiac CT. David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Cardiac CT David Stultz, MD Cardiology Fellow, PGY 6 March 28, 2006 Courtesy Tom Kracus Courtesy Kettering Tom Medical Kracus Cente Kettering Medical Center 2003-2006, David Stultz, MD Courtesy Tom Kracus

More information

Importance of early diagnosis of cardiovascular diseases

Importance of early diagnosis of cardiovascular diseases ORIGINAL ARTICLE Knowledge-Based Dynamic Volumetric Cardiac Computed Tomography With Saddle Curve Trajectory Deepak Bharkhada, MS,*Þ Hengyong Yu, PhD,þ and Ge Wang, PhD*Þþ Abstract: Motion artifact is

More information

Ischemic heart disease is the leading cause of morbidity and

Ischemic heart disease is the leading cause of morbidity and REVIEW ARTICLE Assessment of Cardiac Function Using Multidetector Row Computed Tomography Sarwar H. Orakzai, MD,* Raza H. Orakzai, MD,* Khurram Nasir, MD, MPH,*Þ and Matthew J. Budoff, MDþ Abstract: In

More information

A Snapshot on Nuclear Cardiac Imaging

A Snapshot on Nuclear Cardiac Imaging Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the

More information