Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Size: px
Start display at page:

Download "Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods"

Transcription

1 CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory and oscillatory blood pressure measurements. Setting: Inner-city prenatal clinic. Participants: Eighty-one women in their 2nd to 9th month of pregnancy. Design: Participants were assessed for systolic and diastolic blood pressures on left and right arms using auscultatory (manual) and oscillatory (electronic) methods. A correlational study design was used. Main outcome measures: Differences in pressures related to arm and method of measurement. Results: The oscillatory method produced consistently higher readings for both systolic (F[ 1,801 = 45.9, p < 0.001) and diastolic (F[1, 801 = 25.79,p < 0.001) pressure readings. Correlations between estimates generally treated as substitutable all fell below the recommended level of 0.80 for measurement equivalence. Conclusions: Results suggest the need for caution when interpreting blood pressure estimates as interchangeable. This is particularly important when patients move from clinic settings, where auscultatory methods predominate, to inpatient settings, where oscillatory methods of measurement are used. JOGNN, 25, ; Accepted: August 1994 tween the two methods are important clinical considerations when screening or treating patients for hypertension. The goal of the current research was to determine the equivalence of the auscultatory and oscillatory methods of measuring blood pressure. Background Because an increase in blood pressure above prepregnant levels is symptomatic of hypertensive disorders in pregnancy, the accurate measurement of blood pressure is a requisite of prenatal care. Pregnancy-induced hypertension may be divided into three,categories: hypertension, preeclampsia, and eclampsia. An increase in the patient's systolic or diastolic pressure may indicate the presence of hypertension. Diagnostic blood pressure criteria are based on the auscultatory system. The popularity of noninvasive oscillatory equipment has created a variation in methods of blood pressure measurement. Researchers have begun to address the issue of the equivalence of the auscultatory versus oscillatory readings. Are auscultatory (manual) and oscillatory (electronic) blood pressure readings equivalent? he measurement of blood pressure is a critical element in detecting and managing hypertension during pregnancy. Auscultatory (manual) methods of blood pressure measurement are used in most outpatient facilities, whereas oscillatory (electronic) equipment is used in most inpatient settings. Researchers have begun to question the accuracy of estimates using oscillatory methods compared with auscultatory methods, as well as the equivalence of measurements from these devices. Possible differences in blood pressure measurement be- Weaver, Park, and Lee (1990) measured the blood pressure of 381 fifth-grade students with the auscultatory method and compared estimates with those obtained by the Dinamap monitor (Critikon, Tampa, FL), an oscillometric device. Two measurements were obtained using each of the two methods. Weaver et al. (1990) found that oscillatory systolic pressures were 5-8 mm Hg higher than the auscultatory values. Oscillatory diastolic pressures were 8-9 mm Hg lower than auscultatory measurements. Correlation co- February 1996 J O C N N 155

2 C L I N I C A L S T U D I E S efficients were r = 0.78 for the systolic blood pressure and r = 0.42 for the diastolic 4th Korotkoff phase. Although statistically significant, the correlations between diastolic readings are low for supposedly equivalent, concurrent measures, suggesting a poor relationship between the two readings. Systolic readings in the present study, though more closely correlated, were borderline for equivalent estimates (Froman & Owen, 1991). The findings of Weaver et al. (1990) suggest that the published, normative levels of auscultatory blood pressure may be inappropriate as standards when blood pressure measurement is obtained by the Dinamap monitor. Marx, Schwalbe, Cho, and Whitty (1993) studied the differences in blood pressure values using both methods and the significance of these differences during labor with epidural analgesia. They compared auscultatory blood pressure measurements for 30 women in labor, 20 full-term pregnant women not in labor, and 20 nonpregnant volunteers. In the nonlaboring parturients and the nonpregnant controls, systolic and diastolic measurements were considered to show nonsignificant differences. In the laboring parturients, the oscillatory systolic pressures were consistently and significantly higher (p < 0.02) compared with the auscultatory measurements. Diastolic pressures of the parturients were consistently and significantly lower (p < 0.002) when measured by the oscillatory method compared with the auscultatory method. In 13 women, the systolic pressure was >lo mm Hg higher when measured by the oscillatory method compared with the auscultatory method. The diastolic pressure was >10 mm Hg lower with the oscillatory method in nine subjects. Mean arterial pressures did not differ, although the mean difference in arterial pressure was described as moderate (6.2 mm Hg). The researchers hypothesized that there is a factor in the changing pulse pressure of a laboring woman that is not taken into account by the algorithm used by the oscillatory method. The authors found a discrepancy between systolic and diastolic pressures obtained by the oscillatory versus the auscultatory method of measurement and recommended that the diagnoses of hypertension and hypotension be based on the mean arterial pressure rather than on the systolic or diastolic pressure. In 1991, Marx, Sofair, and Winikoff conducted a preliminary study of 12 patients during labor. Their findings revealed that all systolic pressures were higher and diastolic pressures were lower with the Dinamap monitor than with the auscultatory method. Differences ranged from 5-18 mm Hg (M = 10.2 f 3.7 mm Hg) in systolic pressure and from 4-14 mm Hg (M= 9.9 f 3.0 mm Hg) in diastolic pressure. Differences in auscultatory pressures between two observers were 0-2 mm Hg. The same cuffs were used for both oscillatory and auscultatory measurements of blood pressure. An interval of at least 2 minutes separated the measurements, with the auscultatory method preceding the oscillatory. Subjects were in labor, and measurements were taken between contractions and repeated twice within 20 minutes. The purpose of the Marx et al. (1991) study was to determine the sensitivity of the Dinamap monitor in detecting hypotension during and after epidural analgesia. Hypotension is a critical complication of epidural anesthesia. Results demonstrated that the normal definition of hypotension as a systolic pressure below 100 mm Hg, based on auscultatory standards, must be revised when an oscillatory blood pressure device is used, because the systolic pressure was higher than 100 mm Hg with the Dinamap monitor when clinical hypotension was present. The Marx et al. (1991) study, however, was a preliminary study of only 12 patients; thus, the results may not be generalized. Ornstein, Markert, Litchfield, and Zemp (1988) evaluated the Dinamap 8100 (Critikon, Tampa, FL) using the Hawksley Random Zero Sphygmomanometer (Sommerville, NJ) as the standard. Eighty ambulatory patients were studied. Readings were compared in a two-instrument, two-arm, two-investigator study with four pairs of simultaneous measurements. The research design allowed a repeated measures analysis of variance, which assessed the contributions of observer, order, arm, and instrument to blood pressure measurement. Analysis confirmed significant differences in systolic blood pressure determinations between instruments (p < , paired t-test). Dinamap 8100 readings showed higher systolic values. No significant differences were found between observer, order, or arm for systolic measurements. For diastolic measurements, observer differences were barely statistically significant (mean difference 1.O mm Hg, p < 0.05). Order and arm differences were not statistically significant for diastolic measurements, and mean difference in diastolic measurements between devices was not statistically significant. The authors suggested that routine use of the Dinamap 8100 could lead to misclassification errors in routine screening for hypertension, and, given the low predictive values of its readings, they cautioned against using the Dinamap 8100 for the management of hypertension. The Association for the Advancement of Medical Instrumentation (1987) includes standards for the evaluation of oscillatory blood pressure monitor readings compared with standard sphygmomanometers. These standards recommend that mean differences in blood pressure be +5 mm Hg or less, with a standard deviation of 8 mm Hg. Methods A convenience sample of 81 pregnant women was secured from a prenatal clinic population in a large, urban hospital by selecting the first subjects meeting inclusion criteria. Inclusion criteria were that each woman be pregnant, 15 to 42 years of age, and under care at the prenatal clinic of the participating hospital. Subjects were asked to refrain from exercising, eating, and smoking for 1 hour and to rest for 10 minutes with an empty bladder before blood pressure measurement. Permission of the Nursing Research Committee at the study facility was obtained, and all subjects gave written, informed consent to participate in the study. The 156 J O C N N Volume 25, Number 2

3 Blood Pressure Meusurement Table 1. Descriptive Statistics (N = 81) Vartable Name Mean SD Range Age Weight (Ibs) Height (in) Gestational age (days) Arm (cm) circumference Systolic blood pressure" (manual) Diastolic blood pressure" (manual) Systolic blood pressure" (Electronic) Diastolic blood pressuren (Electronic) Ethnic status Black Latino White Asian Readings averaged across arms n = 28 n = 38 n= 12 n= purpose of the study, as well as the risks and benefits, were delineated in the permission form. Confidentiality of information was assured by the use of assigned identification numbers for collecting data, conducting analysis, and reporting results. Blood pressure was measured by one of the authors (LAG) in a quiet room with the patient seated at a table and her arm supported at heart level. Left arm measurements were taken first. Blood pressure measurements on the same subject were taken with the same size cuff. The cuff was 40-50% of the circumference of the upper arm, as recommended by the American Heart Association (1987). The 5th Korotkoff phase was used as the diastolic standard because the fourth sound was not heard consistently. Subjects' blood pressure was measured by the same author for all measurements. The auscultatory method was used first to prevent observer bias as an intervening variable because recording an electronic readout could create an expectancy about a manual reading. Two blood pressure measurements were made with each instrument on each arm at intervals of at least 2 minutes until venous congestion was resolved, as recommended by the American Heart Association (1987). The Dinamap model 1846SX (Critikon, Tampa, FL) was the oscillometric study instrument. A mercury gravity sphygmomanometer was the standard auscultatory device. Proficiency and accuracy in the use of the mercury sphygmomanometer and the Dinamap 1846SX were demonstrated by adherence to the standards of the American Heart Association (1987) and the Association for the Advancement of Medical Instrumentation (1987). Both Blood pressure readings were higher with the oscillatory method. instruments were tested for accuracy by the institution's biomedical engineering department using manufacturer's recommended guidelines. Data from systolic and diastolic readings for each arm were recorded on data collection sheets. Demographic information recorded for each subject included identification number, age, ethnicity, weight, height, arm circumference, and weeks of gestation. Analysis and Results Data were analyzed in three steps. First, descriptive statistics were calculated on the demographic variables to assess representativeness of the sample. These statistics are provided in Table 1. The second step of data analysis used repeated measures analysis of variance (R-ANOVA) to test for arm and method differences in systolic and diastolic readings. Two within-group variables, arm and method (oscillatory versus auscultatory), were used in the analysis. Separate R-ANOVAs were conducted on systolic and diastolic data. Results of the R-ANOVA are presented in Tables 2 and 3. Significant differences (F[1, 801 = 45.9, p < 0.001) were found between auscultatory and oscillatory systolic readings. Systolic pressures were significantly higher for the oscillatory method. Average systolic blood pressures, using the oscillatory method, were 115 and 112 for right and left arms, respectively. Average auscultatory systolic pressures were 108 and 105 for right and left arms, respectively. The test for differences in systolic pressures between right and left arms showed significant differences (F[l, 801 = 9.70, p < 0.005). Right arm readings were higher than left arm readings. The R-ANOVA comparing diastolic readings produced similar results. Oscillatory readings were significantly higher than auscultatory readings (F [I, 801 = 25.79, p < 0.001). Average diastolic pressures using the oscillatory method were 64 and 65 for right and left arms, Table2. Analysis of Variance on Systolic Readings Source DF Mean square F Method 1, * Arm 1, t Interaction Arm X Method 1, * p < tp< February 1996 JOG" 157

4 C L I N I C A L STUDIES Table3. Analysis of Variance on Diastolic Readings Source DF Mean square F Method 1, * Arm 1, Interaction Method X Arm 1, p < Discussion and Implications Readings from an oscillatory device may be misleading if measurements are used to diagnose and treat pregnancyinduced hypertension. Existing standards for screening and referral are based on data collected using the manual, auscultatory method of blood pressure estimation. These standards may not be appropriate cut-off values for oscillatory methods of blood pressure measurement. respectively. Average auscultatory diastolic pressures were 60 and 61 for right and left arms, respectively. The test for differences between right and left arms was nonsignificant. The final analysis used correlation to estimate the equivalence of readings across the two methods. The R- ANOVAs were useful in noting gross differences between readings by method but could not specify whether rank ordering within each method was equivalent. Correlational analysis was used to determine whether application of a regression formula or addition of a constant would allow auscultatory readings to be transformed into equivalent oscillatory estimates. If a minimum correlation of 0.80 was found between any pair of readings, development of a regression formula would be indicated (Froman & Owen, 1991). Table 4 presents the correlations between all estimates. Correlations between auscultatory and oscillatory measurements assumed to be equivalent ranged from a low of 0.58 to a high of These correlations are less than the recommended 0.80 alternate methods reliability generally accepted for the substitution of scores (Nunnally, 1978). This implies that there are differences between oscillatory and auscultatory readings, as shown by the R-ANOVA, and that these differences cannot be adequately adjusted with the use of a regression formula or the addition of a constant. Caution must be taken in interpreting auscultatory and oscillatory blood pressure readings as equivalent. If women are screened for hypertension at an outpatient clinic with the auscultatory method and subsequently are treated using oscillatory readings, discrepancies between readings may occur resulting from measurement differences inherent in the two devices. These differences may interfere with interpreting the effectiveness of treatment. If patients are diagnosed with hypertension in the prenatal clinic and are sent to the labor and delivery unit for further evaluation, placing them in a side-lying position for awhile may affect their blood pressure even though the measurements are taken in a sitting position. In such a situation, differences in.position and activity level rather than in equipment may be affecting the readings. Another danger lies in not detecting hypertension because of possible falsely low readings from use of the auscultatory method. If oscillatory methods are more objective, accurate estimates of blood pressure, borderline hypertension may be missed at early screenings if the auscultatory method is used. Table 4. Correlation Matrix for Arm and Method Equivalence Manual Manual Manual Manual Electronic Electronic Electronic Diastolic Diastolic Systolic Systolic Dtastoltc Diastoltc Systolic Electronic Systolic Left Arm Right Arm Left Arm Right Arm Left Arm Right Arm Left Arm Right Arm Manual diastolic right arm Manual diastolic left arm 0.61 Manual systolic right arm Manual systolic left arm Electronic diastolic right arm 0.61" Electronic diastolic left arm " Electronic systolic right arm " Electronic systolic left arm a a Alternate methods equivalence estimates. Note. All values reoresent uroduct moment correlations 158 JOGNN Volume 25, Number 2

5 Blood Pressure Measuremerit A difference of 7 mm Hg in systolic readings between the two methods exceeds the standard of +5 mm Hg established by the Association for the Advancement of Medical Instrumentation (1987). No quality control method exists for routine comparison of oscillatory and auscultatory blood pressure readings. Current findings reiterate discrepancies shown in other studies comparing the oscillatory and auscultatory methods of blood pressure measurement. Some studies have found that systolic pressures are significantly higher with the oscillatory compared with the auscultatory method (Marx et al., 1993). However, other reports show diastolic readings are significantly lower with the oscillatory devices than with the auscultatory methods (Ornstein et al., 1987; Weaver et al., 1990). The current study found systolic and diastolic pressures were significantly higher when measured with the oscillometric method. Differences in blood pressure measurements found in the current study and other studies may be linked to the differences in devices or the differences in clinical profiles of the different populations. Additionally, the current study was limited by the assessment of only one instrument, the Dinamap 1846 SX. That instrument may have idiosyncratic problems not detected at the time of calibration. The Dinamap is available in several models, and other manufacturers also produce oscillatory blood pressure recorders. Data obtained in the current study are applicable only to a gravid population. The current study did not provide information as to which method more accurately reflects directly measured arterial pressure. Marx et al. (1993) recommend the use of mean arterial pressure to diagnose hypertension rather than systolic or diastolic blood pressure. To assess the related criteria and predictive validity of each method of blood pressure measurement, a longitudinal study would help to determine in which subjects hypertension actually developed during pregnancy. Auscultatory and oscillatory methods measure blood pressure in different ways. The 5th Korotkoff phase was used in the current study for auscultatory diastolic readings. Dinamap measures blood pressure through an oscillometric technique. A cuff is inflated to a pressure above the systolic pressure. As this cuff deflates, oscillations in the pressure within the cuff are sensed, stored, and analyzed electronically. Systolic pressure is the point of rapid increase in the amplitude of oscillations. Diastolic pressure is the point of rapid decrease in oscillation after maximum oscillation (Marx et al., 1993). Therefore, using the fifth sound for the auscultatory method and the fourth sound for the oscillatory method may have contributed to the differences in readings. Readings from oscillatory devices may be misleading in the clinical area when they are used to diagnose and treat hypertension of pregnancy. Existing standards for hypertension are based on data collected using the auscultatory method of blood pressure measurement. These manual standards of measurement may not be appropriate cut-off values for electronic blood pressure measurements, as indicated by results from the ANOVAs and correlational analysis conducted for the current study. Correlational analysis demonstrated that the differences between the oscillatory and auscultatory readings could not be considered directly equivalent, nor could they be corrected by a formula or constant. The ANOVAs documented significant differences within individual readings across methods. To diagnose hypertension, the current study supports the importance of evaluating the patient within the context of her clinical symptoms and not by blood pressure changes alone. References American Heart Association. (1987). Recommendations for hzcman blood pressure determination by sphygmomanometers. Dallas: Author. Association for the Advancement of Medical Instrumentation. (1987). American nationalstandard for electronic sphygmomanometers. Arlington, VA: Author. Froman, R., & Owen, S. (1991). Teaching reliabilityandvalidity: Fun with classroom applicdtion.journal of Continuing Education in Nursing, 22, Marx, G., Schwalbe, S., Cho, E., & Whitty, J. (1993). Automated BP measurement in laboring women: Are they reliable? American Journal of Obstetrics and Gjinecology, 168, Marx, G., Sofair, D., & Winikoff, S. (1991). BP values in parturients: Auscultatory vs. oscillatory values. Anestbesiu Anulgesia, 72, Nunnally, J. (1978). Psychometric theory. New York: McGraw- Hill. Ornstein, S., Markert, G., Litchfield, L., & Zemp, L. (1988). Evaluation of the DINAMAP blood pressure monitor in an ambulatory primary setting. Journal of Family Practice, 26, Weaver, M., Park, M., & Lee, D. (1990). Differences in blood pressure levels obtained by auscultatory and oscillometric methods. American Journal of Disease Control, 144, Address for correspondence: Linda A. Green, RN, MS, 26 Yeomans Road, Columbia, CT Linda A. Green ts a staff nurse at St. Francis Hospttal and Medical Center, Hartford, CT. Robtn D. Froman is a professor and director of the Center for Nurstng Research at the School of Nursing, Untversity of Connecticut, Storrs. Februuy 199G J O C N N 159

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04) SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: MEASURING BLOOD PRESSURE - MANUAL Nursing DATE: REVIEWED: PAGES: 2/80 7/17 1 of 5 RESPONSIBILITY: RN, LPN, Patient Care Technician Multi-skilled

More information

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE Pamela B. Morris, MD, FACC, FAHA, FASPC, FNLA Professor Medicine and Cardiovascular Diseases Director, Seinsheimer Cardiovascular Health

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1982 Characterization of the Oscillometric Method for Measuring Indirect

More information

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the A comparison of blood pressure measurement over a sleeved arm versus a bare arm Grace Ma MD, Norman Sabin MD, Martin Dawes MBBS MD @ See related article page 591 DOI:10.1503/cmaj.070975 Abstract Background:

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator MSc Thesis in Clinical Science and Technology Sara Rose Newell Department of Health Science

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Original article http://dx.doi.org/10.3345/kjp.2011.54.11.463 Korean J Pediatr 2011;54(11):463-469 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

How To Measure Vital Signs

How To Measure Vital Signs How To Measure Vital Signs How to Use This Tutorial This tutorial is intended for healthcare providers or students to teach basic vital signs skills Use the navigation buttons below to move through the

More information

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER Naser KA. Teaching Hospital Peradeniya, Peradeniya, Sri Lanka Zawahir S

More information

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System!

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System! LAB: Blood Measurable Indicator of the Health of the Circulatory System! Lab Objectives. At the completion of the lab, you should be able to: measure pulse or heart rate (HR) and respiratory rate (RR);

More information

Data Collection Worksheet

Data Collection Worksheet Data Collection Worksheet 1. Has a doctor or nurse ever said that you have: High blood pressure or hypertension? [ ] No [ ] Yes [ ] Not Sure 1.a. If Yes, then at what age were you first told this? Age

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

Managing Blood Pressure: It Takes a Team

Managing Blood Pressure: It Takes a Team Managing Blood Pressure: It Takes a Team Module Two How to Take Blood Pressure October 2011 MODULE TWO: How to Take Blood Pressure Module Two: HOW TO TAKE BLOOD PRESSURE Objectives Demonstrate proper technique

More information

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

2. Measure a subject's blood pressure and heart rate both at rest and during exercise. Lab Activity 11 The Cardiovascular System Student Learning Objectives After completing this lab, you should be able to: 1. Define, explain and correctly use the key terms. 2. Measure a subject's blood

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

1 Non-invasive measurement of arterial pressure

1 Non-invasive measurement of arterial pressure Non-invasive measurement of arterial pressure I. Background A. Circulatory systems Human circulation takes place in a closed system that consists of two subsystems, pulmonary circulation and systemic circulation,

More information

Comparison of blood pressure measured at the arm, ankle and calf

Comparison of blood pressure measured at the arm, ankle and calf doi:10.1111/j.1365-2044.2008.05633.x Comparison of blood pressure measured at the arm, ankle and calf C. Moore, 1 A. Dobson, 2 M. Kinagi 2 and B. Dillon 3 1 SpR Anaesthetics, 2 Consultant Anaesthetist,

More information

Please do not hesitate to contact our dedicated service unit for more details regarding this program at

Please do not hesitate to contact our dedicated service unit for more details regarding this program at The Blue Cross and Blue Shield Federal Employee Program (FEP) and the American Medical Association (AMA) have come together in a collaborative effort to provide physicians with resources designed to improve

More information

Thesis for doctoral degree (Ph.D.) 2008 ROLE OF AMBULATORY BLOOD PRESSURE MONITORING AFTER PEDIATRIC RENAL TRANSPLANTATION Rafael Tomás Krmar

Thesis for doctoral degree (Ph.D.) 2008 ROLE OF AMBULATORY BLOOD PRESSURE MONITORING AFTER PEDIATRIC RENAL TRANSPLANTATION Rafael Tomás Krmar Thesis for doctoral degree (Ph.D.) 2008 Thesis for doctoral degree (Ph.D.) 2008 ROLE OF AMBULATORY BLOOD PRESSURE MONITORING AFTER PEDIATRIC RENAL TRANSPLANTATION Rafael Tomás Krmar ROLE OF AMBULATORY

More information

Lab #10 Physiology of the Circulatory System

Lab #10 Physiology of the Circulatory System Lab #10 Physiology of the Circulatory System Introduction The circulatory system functions to deliver oxygen an nutrients to tissues for growth and metabolism, and to remove metabolic wastes. The heart

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

An Automated System for Blood Pressure Determination During Exercise

An Automated System for Blood Pressure Determination During Exercise An Automated System for Blood Pressure Determination During Exercise SEPHEN P. GLASSER, M.D., AND MAYNARD R. RAMSEY III, M.D., PH.D. with the technical assistance ofblanche Price SUMMARY Automation of

More information

High Blood Pressure Blood Pressure Monitoring at Home -- F...

High Blood Pressure Blood Pressure Monitoring at Home -- F... Return to Web version High Blood Pressure Blood Pressure Monitoring at Home Why should I measure my blood pressure at home and keep a record of it? Measuring your blood pressure at home and keeping a record

More information

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS LAB 8: CARDIOVASCULAR PHYSIOLOGY PART 1. HEART SOUNDS AND PULSE DETERMINATIONS Introduction Two distinct sounds can be heard during each cardiac cycle. These sounds are commonly described as lub and dup

More information

Automatic oscillometric blood pressure devices are frequently

Automatic oscillometric blood pressure devices are frequently Arterial Stiffness as Underlying Mechanism of Disagreement Between an Oscillometric Blood Pressure Monitor and a Sphygmomanometer Nicole M. van Popele, Willem Jan W. Bos, Nicole A.M. de Beer, Deirdre A.M.

More information

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right Anthony J. Viera, MD, MPH, FAHA Professor and Chair Objectives Review limitations of office BP in making a correct diagnosis of hypertension

More information

Webinar Logistics. Now What Do I Need To Do? Blood Pressure in Kids January NC Department of Health and Human Services Division of Public Health

Webinar Logistics. Now What Do I Need To Do? Blood Pressure in Kids January NC Department of Health and Human Services Division of Public Health NC Department of Health and Human Services Division of Public Health Now What Do I Need To Do? Blood Pressure in Kids January 2018 Gerri Mattson, MD, MSPH, FAAP Pediatric Medical Consultant Children and

More information

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Soo-young Ye, Gi-Ryon Kim, Dong-Keun Jung, Seong-wan Baik, and Gye-rok Jeon Abstract In this paper, algorithm estimating the blood

More information

Physiology of the Circulatory System modified from

Physiology of the Circulatory System modified from Physiology of the Circulatory System modified from http://www.ekcsk12.org/science/aplabreview/aplab10.htm Introduction The circulatory system functions to deliver oxygen and nutrients to tissues for growth

More information

Section 03: Pre Exercise Evaluations and Risk Factor Assessment

Section 03: Pre Exercise Evaluations and Risk Factor Assessment Section 03: Pre Exercise Evaluations and Risk Factor Assessment ACSM Guidelines: Chapter 3 Pre Exercise Evaluations ACSM Manual: Chapter 3 Risk Factor Assessments HPHE 4450 Dr. Cheatham Purpose The extent

More information

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine Techniques of Vital Signs John Gazewood, MD, MSPH Department of Family Medicine Objectives Define normal in several ways. Describe correct technique of measuring vital signs. Describe correct technique

More information

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

!#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 !"#$ Blood Pressure Monitor Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 Introduction

More information

Blood Pressure Measurement in SPRINT

Blood Pressure Measurement in SPRINT Blood Pressure Measurement in SPRINT Karen C. Johnson, MD, MPH, FAHA Vice Chair, SPRINT Steering Committee University of Tennessee Health Science Center, Department of Preventive Medicine For the SPRINT

More information

Blood Pressure Protocol Training

Blood Pressure Protocol Training Blood Pressure Protocol Training May 31, 2018 Pat Spier, RN-BC, PCMH-CCE Barb Rice, RN-BC BCBS of North Dakota Fargo, ND 2013 Template and icons provided by The Advisory Board Company. Hypertension Education

More information

SECTION 16: BLOOD PRESSURE MEASUREMENT

SECTION 16: BLOOD PRESSURE MEASUREMENT CHRONIC KIDNEY DISEASE IN CHILDREN COHORT STUDY SECTION 16: BLOOD PRESSURE MEASUREMENT 16.1 OVERVIEW OF BLOOD PRESSURE MEASUREMENT This section describes the procedures that are to be performed to obtain

More information

INTRAVENOUS HYDRALAZINE POLICY

INTRAVENOUS HYDRALAZINE POLICY PURPOSE INTRAVENOUS HYDRALAZINE POLICY The purpose of this policy is to: provide safe and effective care for women establish a local approach to care that is evidence based and consistent inform good decision

More information

Clinical Medical Policy Department Clinical Affairs Division DESCRIPTION

Clinical Medical Policy Department Clinical Affairs Division DESCRIPTION Ambulatory Blood Pressure Monitoring (ABPM) [For the list of services and procedures that need preauthorization, please refer to www.mcs.com.pr. go to Comunicados a Proveedores, and click Cartas Circulares.]

More information

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Anthony J. Viera, MD, MPH, FAHA Department of Family Medicine Hypertension Research Program UNC School of Medicine Objectives Review limitations

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

Home Blood Pressure Log

Home Blood Pressure Log Home Blood Pressure Log 2 You or your health care professional may want to monitor your blood pressure at home twice a day for a week or more before appointments to accurately confirm your average blood

More information

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms Chapter 29 Vital Signs Key Terms Blood pressure Body temperature Diastolic pressure Hypertension Hypotension Pulse Pulse rate Respiration Sphygmomanometer Stethoscope Systolic pressure Vital signs Measuring

More information

LAB 9: Metabolic Rates

LAB 9: Metabolic Rates LAB 9: Metabolic Rates Introduction: The cardiovascular (circulatory) system functions to deliver oxygen and nutrients to tissues for growth and metabolism, and to remove metabolic wastes. The heart pumps

More information

BLOOD PRESSURE ASSESSMENT

BLOOD PRESSURE ASSESSMENT BLOOD PRESSURE ASSESSMENT Course Principles of Health Science Unit X Vital Signs Course Health Science Unit VI CPR / AED & First Aid Essential Question How do health care skills help to promote health

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines TITLE: Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines DATE: 25 September 2012 CONTEXT AND POLICY ISSUES A mercury manual sphygmomanometer

More information

THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM

THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM Cader FA, Das A, Kahhar MA Dhaka Medical College Hospital March 2015 Dr Aaysha Cader MBBS,

More information

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust.

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust. AN INTRODUCTION TO DOPPLER Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust. THE DOPPLER EFFECT The Doppler Principle was described by Physicist and mathematician

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

Recommendations for Blood Pressure Measurement in Humans and Experimental Ani...

Recommendations for Blood Pressure Measurement in Humans and Experimental Ani... Recommendations for Blood Pressure Measurement in Humans and Experimental Ani... Página 1 de 41 Published online before print December 20, 2004, doi:10.1161/01.hyp.0000150859.47929.8e (Hypertension. 2005;45:142.)

More information

Blood Pressure Estimation Using Photoplethysmography (PPG)

Blood Pressure Estimation Using Photoplethysmography (PPG) Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4

More information

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure? MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents 01... Blood Pressure Monitor Intended Use 02... 1. What is blood pressure? 02... 2. Why is it useful measure blood pressure at home?... A. WHO blood pressure

More information

Magnesium Sulphate - Management of Hypertensive Disorders of Pregnancy

Magnesium Sulphate - Management of Hypertensive Disorders of Pregnancy 1. Purpose Magnesium sulphate is the anticonvulsant of choice for pre-eclampsia prophylaxis and treatment. This clinical guideline outlines the indications, contraindications, administration and monitoring

More information

Protocol 4: Measuring Blood Pressure

Protocol 4: Measuring Blood Pressure Zool 430L Protocols Page 8 of 12 Protocol 4: Measuring Blood Pressure 4.1 Via Auscultation (listening through stethoscope) The estimate of blood pressure is obtained by cutting off blood flow with a known

More information

OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension. Ontario Health Technology Advisory Committee

OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension. Ontario Health Technology Advisory Committee OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension Ontario Health Technology Advisory Committee May 2012 Background Hypertension in Canada Hypertension occurs when

More information

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods Vol. 2, Issue 1, pages 31-36 ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY By Alejandro de la Sierra, MD Luis M. Ruilope, MD Hypertension Units, Hospital Clinico, Barcelona & Hospital 12 de Octubre,

More information

Automated versus manual blood pressure measurement: A randomized crossover trial

Automated versus manual blood pressure measurement: A randomized crossover trial Automated versus manual blood pressure measurement: A randomized crossover trial Author Heinemann, Megan, Sellick, Ken, Rickard, Claire, Reynolds, Pam, McGrail, Matthew Published 2008 Journal Title International

More information

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM AP Biology Laboratory Date: Name and Period: AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM OVERVIEW In this lab you will: 1. in Exercise 10A you will learn how to measure blood pressure. 2. in

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

Lecture #3 - Blood Pressure Recording Procedure. Equipment

Lecture #3 - Blood Pressure Recording Procedure. Equipment Lecture #3 - Blood Pressure Recording Procedure CKiD Blood Pressure Training Equipment Mabis Medic-Kit aneroid sphygmomanometer Mabis Medic-Kit assortment of cuff sizes Stethoscope with bell (pediatric

More information

Comparison of Exercise Blood Pressure Measured by Technician and an Automated System

Comparison of Exercise Blood Pressure Measured by Technician and an Automated System Clin. Cardiol. 7, 315-321 (1984) Clinical Cardiology Publishing Co., nc. Practitioner's Corner Comparison of Exercise Blood Pressure Measured by Technician and an Automated System J. A. GARCA-GREGORY,

More information

Ambulatory Services Orientation & Skill Review Skill Performance Checklist: Vital Signs. Name: Date:

Ambulatory Services Orientation & Skill Review Skill Performance Checklist: Vital Signs. Name: Date: Evaluation Levels Behavior Level 3 Met Performed all skills competently. No behaviors omitted. Level 2 Satisfactory Omitted no more than three (3) non-critical behaviors. Level 1 Unsatisfactory Omitted

More information

Welcome to the S410. Blood Pressure System. Instruction Manual

Welcome to the S410. Blood Pressure System. Instruction Manual Welcome to the S410 Blood Pressure System Instruction Manual The simulator is to be used as part of an approved educational program for healthcare. It is not a substitute for traditional learning methods

More information

Training and Certification of Blood Pressure Observers

Training and Certification of Blood Pressure Observers Training and Certification of Blood Pressure Observers J. DAVID CURB, M.D., M.P.H., DARWIN R. LABARTHE, M.D., PH.D., SHARON POIZNER COOPER, PH.D., GARY R. CUTTER, PH.D., AND C. MORTON HAWKINS, SC.D. SUMMARY

More information

INDIRECT ANKLE, BRACHIAL AND DIRECT INTRA -ARTERIAL BLOOD PRESSURE MEASUREMENTS

INDIRECT ANKLE, BRACHIAL AND DIRECT INTRA -ARTERIAL BLOOD PRESSURE MEASUREMENTS VOLUME 26, NO. 3 JUNE 1985 INDIRECT ANKLE, BRACHIAL AND DIRECT INTRA -ARTERIAL BLOOD PRESSURE MEASUREMENTS C K Koay M K Chin SYNOPSIS The purpose of this study was to compare the accuracy of indirect ankle

More information

5.2 Key priorities for implementation

5.2 Key priorities for implementation 5.2 Key priorities for implementation From the full set of recommendations, the GDG selected ten key priorities for implementation. The criteria used for selecting these recommendations are listed in detail

More information

K/DOQI Clinical Practice Guidelines on Hypertension and Antihypertensive Agents in Chronic Kidney Disease

K/DOQI Clinical Practice Guidelines on Hypertension and Antihypertensive Agents in Chronic Kidney Disease K/DOQI Clinical Practice Guidelines on Hypertension and Antihypertensive Agents in Chronic Kidney Disease GUIDELINE 1: GOALS OF ANTIHYPERTENSIVE THERAPY IN CKD Hypertension is common in CKD, and is a risk

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

Name of Policy: Computerized Pulse Waveform Analysis

Name of Policy: Computerized Pulse Waveform Analysis Name of Policy: Computerized Pulse Waveform Analysis Policy #: 020 Latest Review Date: September 2012 Category: Medical Policy Grade: Active Policy but no longer scheduled for regular literature reviews

More information

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM Faculty Disclosure I have no financial interest to disclose No off-label use of medications will be discussed FIFTH ANNUAL SYMPOSIUM Recognize changes between

More information

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement Direct Care Staff Trained in Accurate BP Measurement 1. 2. 3. 5. 22 to 26 cm Small adult (12X22 cm) 27 to 34 cm Adult (16X30 cm) 35 to 44 cm Large adult (16X36 cm) 45 to 52 cm Adult thigh (16X42 cm) widths

More information

Declaration of Equivalence Form

Declaration of Equivalence Form Declaration of Equivalence Form DECLARATION OF BLOOD PRESSURE MEASURING DEVICE EQUIVALENCE 2013 SECTION A - Please complete all items. I Bill Huang,, Name of a Company Director A SIGNED COPY WILL BE POSTED

More information

https://www.lucidoc.com/cgi/doc-gw.pl?ref=overlake_p:53602

https://www.lucidoc.com/cgi/doc-gw.pl?ref=overlake_p:53602 Page 1 of 6 Protocol : Severe Hypertension in Obstetrics: Emergent Treatment DocID: 53602 Revision: 7 Status: Official Department: Women's and Infants' Admin Manual(s): Labor and Delivery Mother Baby Unit

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval BPU22 with the AViTA BPM63S Devices Item 9 HARTMANN Veroval BPU22 AViTA BPM63S Pictures Display Image Validation Equivalence ESH 2010 ESH 2002 BHS AAMI Category Arm Type

More information

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 64 CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 4.1 GENERAL This chapter presents the methodologies that are usually adopted for the measurement of blood pressure, heart rate and pulse

More information

Importance Of Taking Manual Blood Pressure In Both Arms

Importance Of Taking Manual Blood Pressure In Both Arms Importance Of Taking Manual Blood Pressure In Both Arms Its importance to health is now well-recognised, in that high blood pressure can lead to Have a look at our slide show on how to take a blood pressure

More information

Blood Pressure Monitors Nissei

Blood Pressure Monitors Nissei We have made it easy for you to find a PDF Ebooks without any digging. And by having access to our ebooks online or by storing it on your computer, you have convenient answers with blood pressure monitors

More information

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients Vital signs (pulse, blood pressure, temperature, respiratory rate, pain) are physiological parameters that a healthcare professional requires when dealing with patients. Accurate measurement of vital signs

More information

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL Douglas G. Kelling Jr., MD Carmella Gismondi-Eagan, MD, FACP George C. Monroe, III, MD Revised, April 8, 2012 The information contained in this protocol

More information

M1 Classic. Digital Blood Pressure Monitor. Instruction manual page 1 SYS. mmhg DIA. mmhg. PULSE/min

M1 Classic. Digital Blood Pressure Monitor. Instruction manual page 1 SYS. mmhg DIA. mmhg. PULSE/min 1 Classic Digital Blood Pressure onitor Instruction manual page 1 1 Classic English Introduction The ORON 1 Classic measures your blood pressure and pulse simply and quickly from the arm, without the use

More information

4/4/17 HYPERTENSION TARGETS: WHAT DO WE DO NOW? SET THE STAGE BP IN CLINICAL TRIALS?

4/4/17 HYPERTENSION TARGETS: WHAT DO WE DO NOW? SET THE STAGE BP IN CLINICAL TRIALS? HYPERTENSION TARGETS: WHAT DO WE DO NOW? MICHAEL LEFEVRE, MD, MSPH PROFESSOR AND VICE CHAIR DEPARTMENT OF FAMILY AND COMMUNITY MEDICINE UNIVERSITY OF MISSOURI 4/4/17 DISCLOSURE: MEMBER OF THE JNC 8 PANEL

More information

Chapter-III METHODOLOGY. In this chapter, the procedure adopted for the selection of the

Chapter-III METHODOLOGY. In this chapter, the procedure adopted for the selection of the Chapter-III METHODOLOGY In this chapter, the procedure adopted for the selection of the subject, selection of the variables, criterion variable reliability and data, reliability of investment, competency

More information

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM 1. SCOPE Blood pressure is a vital sign that is used in many clinical situations, and assesses the pressure of blood as it flows

More information

EQUIPMENT: Nitrous Oxygen Delivery System:

EQUIPMENT: Nitrous Oxygen Delivery System: Policy: Nitrous Oxide Use in the Intrapartum and Immediate Postpartum Period for Obstetrical Patients in the Family Birth Place Approvers: CEO. CNO, Medical Staff President, Anesthesia Chair, OB Medical

More information