Biomedical Instrumentation
|
|
- Owen Davis
- 5 years ago
- Views:
Transcription
1 University of Zagreb Faculty of Electrical Engineering and Computing Biomedical Instrumentation Safety of electrical medical devices prof.dr.sc. Ratko Magjarević
2 Medical technology Achievements Significant improvement of health care in all medical specialties Reducing the number of seriously ill people Reduction of mortality Consequences increased complexity of medical devices and their applications increased incidence of adverse effects of implementation of MT in the U.S. approx. 10,000 cases of patient injury due to the application of MT
3 Medical technology Causes of unintended consequences: improper handling insufficient or improper staff practice (training) in handling with MT insufficient experience example, medical staff rarely read manuals for the users (User Manual) equipment malfunction Demand for MT developers developement of safe devices (failure-safe)
4 Danger to patients In a clinical environment patient is exposed to various risks, more than a typical workplace or at home frequent invasive (blood) operations - penetration through skin or mucous membranes presence of potentially hazardous chemicals and substances - anesthetics, medicines, medical gases sources of infection - particularly "hospital infection" various sources of energy that penetrate into or through the patient: current, voltage, ionizing and non-ionizing radiation, sound and ultrasound, electric and magnetic field, UV radiation, lasers, microwave radiation, mechanical stress, etc.
5 Safe MT development Electrical MT security is considered: physiological effects of electricity possibility (risk) of failures and their consequences methods of patients and staff protection standards describing electrical safety electrical safety testing modes Understanding the possible dangers and risks Implementation in achieving security
6 Physiological effects of electricity Body (tissue) becoming a part of an electrical circuit The amount (amplitude) of electricity often depends on the ratio of voltage present and the sum of all (serially connected) impedance Usually, the highest impedance is the impedance of the skin The consequence of current flow: nerve and/or muscle tissue stimulation heating of tissues (a result of tissue resistance) burns
7 Pulsed current I t curves for sensory and motor responses
8 Physiological effects of current
9 Electricity grid Unintended consequences are usually a result of human involvement in the circuit, in some way connecting to the grid (inattention, failure) Several levels of action perception contractions (let-go) paralysis (respiratory), pain, fatigue ventricular fibrillation tetanic contraction burns and physical injuries
10 Consequences of dangerous voltage Fibrilation Convulsion Consequences depende on the voltage current relation, but also on individual susceptibility Expected values Intact skin, R K = 2-5 kω Damaged skin - injuries, surgical procedures, R O = 100 Ω 1 kω
11 Consequences of dangerous voltage Focus on the worst possible scenario, for example, a person holding two conductors at different potentials in his hands resulting muscle spasms, inability to release captured conductors In case of intact skin, R K = 2-5 kω, cramps can occur even at a voltage of few dozen ma In case of damaged skin - injuries, surgical procedures, R O = 100 Ω 1 kω, fibrillation may occur at voltages of several volts and current of few dozen ma
12 Excitability dependence of the pulse waveforms Monopolar vs. bipolar pulses, sinusoidal current Extrapolation of sinusoidal current What is the duration of the half-cycle sine current to achieve maximum excitability?
13 Safe voltages - AC Security measures in designing defining voltage values that may not appear on the conductive parts of instruments or equipment coming into contact with the patients skin or user for alternating current: safety extra low voltage, SELV of 50V, 50/60Hz medical safety extra low voltage, MSELV of 25V, 50/60Hz
14 Safe voltages - DC Security measures in designing defining voltage values that may not appear on the conductive parts of instruments or equipment coming into contact with the patients skin or user for direct current: safety extra low voltage, SELV of 120V, medical safety extra low voltage, MSELV of 60V, 50/60Hz
15 Ventricular fibrillation In cases of shock due to unwanted connection to power grid voltage, current passing through the heart is considered the most dangerous effect and can cause ventricular fibrillation (ventricles) with fatal consequences Ventricular fibrillation is completely irregular, asynchronous heartbeat; no synchronism in ventricular contractions, blood flow stops and if such heart work does not stop, it leads to death within a few minutes
16 Microshock vs. macroshock The consequences of current passage through the tissue depend on the contact point with the source of voltage Even greater power will not necessarily cause fatal consequences, eg. fibrillation, when going through the limbs
17 Microshock vs. macroshock Macroshock is caused by current passing through the body through the skin, and the current that can cause harmful effects is relatively high and significantly depends on the point of contact Microshock is caused by the passage of relatively low current, but the source of electricity was brought directly into contact with the heart, for example, during cardiac catheterization (catheter in the heart is a diagnostic procedure). Electricity sufficient to induce fibrillation is of n x 10μA. Another point of contact can be anywhere on the body (eg limbs). Current of 10 μa is considered safe limit to prevent microshock
18 Protective grounding Generally, patients are only occasionally and accidentally exposed to risk of getting in touching with devices whose conductive parts may be energized In hospitals, especially in intensive care units, patients are deliberately connected to the diagnostic and/or therapeutic electrical devices/equipment particularly careful with the isolation of any conductive parts connected to the heart or its vicinity from all other conductive parts all conductive parts in the vicinity of the patient must be connected to a single point grounding (eg, metal bed, cupboard, etc.) periodic testing of the grounding impedance must be provided Tolerable difference in potential between the grounded conductive parts in clinical areas: in the general - of 500 mv intensive care and other critical cases- of 40 mv
19 Power in clinical environments Under normal conditions, increasing patient safety is provided by: formation of isolated parts of the grid installation of automatic monitoring of impedance grounding (line isolation monitor)
20 Power failure In case of grid power failure, in critical areas in hospitals power is automatically provided within 10s after stoppage Critical components are: lighting alarms equipment in intensive care
21 Equipment safe during the first failure The principle of redundancy can be applied to the safety of medical devices and equipment (KEU) For example, it is required that electrical medical equipment is safe even in case of a breakdown on the conductive contact with patient Grounding impedance must be sufficiently low so that the patient may not get in touch with any dangerous voltage
22 Leakage currents Occur between conductors that are not in direct contact (touch) and are at different potentials For devices that are powered from the grid, leakage currents are: capacitive character (spreading capacity) operative character (current through the insulation, dust, humidity...)
23 Leakage currents pathways Pathways of leakage current through the patient a) Intact ground lead b) Broken protective ground lead, patien connected to ground through a catheter c) Broken protective ground lead, patien connected to ground by an electrode or unintentianally.
24 Leakage currents
25 Conductive paths to / from the heart In clinical settings there are current flow pathways that cause heart microshock: The external pacemaker connected to epicardial (heart area) or endocardial (inside heart) electrodes Intracardial ECG measuring electrodes Catheters for invasive heart pressure measuring taking blood samples injection of contrast or medication into the heart
26 Regulations, standards and recommendations Regulations are subordinate legislation which are obligatory and are related to the characteristics of devices and equipment that must be assured Standards stipulate features that must be assured, but their appliance is voluntary. Contain instructions on how to accomplish / verify compliance with the standard Recommendations (usually manufacturers) stipulate how to ensure the safety and prescribe manner and frequency of testing safety of devices
27 Standards Various standards institutes prescribe standards for electromedical devices and equipment These standards are mostly aligned, due to the global market The fundamental norm in Europe, IEC , Medical Electrical Equipment, Part 1:General requirements for basic safety and essential performances These codes are followed by the standard xxx, Particular requirements for...
28 Design Recommendations for Safety Reliable grounding for equipment Reduction of leakage current Double insulated equipment Operation at low voltages Driven right leg circuit Current limiters Electrical isolationisolated heart connectors
29 Safety checking Electrical Safety Analysis: general for certain types of electrical medical devices and equipment Demo practice: "General" analyzer according to Defibrillator analyzing operation
30 Determination of device categorie
31 Mind the costs, but satisfy safety requirements! Price/performance
32 Literature J. Webster: Medical Instrumentation, Chapter 14, Electrical Safety
Biomedical Instrumentation
University of Zagreb Faculty of Electrical Engineering and Computing Biomedical Instrumentation Electrical stimulation prof.dr.sc. Ratko Magjarević December 2015 Electrical stimulation The effect of electric
More informationImplantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević
Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular
More informationAPPLICATIONS OBJECTIVES PRE-LECTURE
93 E8 APPLICATIONS OBJECTIVES Aims From this chapter you should gain a basic understanding of two important applications of electrical principles to the human body - the study of the function of the heart
More informationDEFIBRILLATORS. Prof. Yasser Mostafa Kadah
DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic
More informationHigh Voltage Safety. Dangers of Electricity. W. Maes. Department of Marine Engineering Antwerp Maritime Academy HV, 2015
Dangers of Electricity W. Maes Department of Marine Engineering Antwerp Maritime Academy HV, 2015 Outline 1 The physical effect of an electric current on the human body Heating Source of power Electro-magnetical
More informationVALLIAMMAI ENGINEERING COLLEGE
VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur 603 203 DEPARTMENT OF ELECTRONICS AND COMMUNICATION ENGINEERING QUESTION BANK VI SEMESTER EC6001 MEDICAL ELECTRONICS Regulation 2013 Academic Year
More informationDry body skin has a ( higher / lower ) resistance to the flow of the electric current than the internal organs of the body.
RESISTANCE OF INTERNAL ORGANS (SUCH AS HEART, LUNGS, LIVER, ETC.) Dry body skin has a ( higher / lower ) resistance to the flow of the electric current than the internal organs of the body. 19. Electricity
More information08-Electrical Safety. EE570 Energy Utilization & Conservation Professor Henry Louie
08-Electrical Safety EE570 Energy Utilization & Conservation Professor Henry Louie 1 Overview Statistics Electric Shock Effect of Voltage Effect of Current Effect of Resistance Effect of Frequency 2 Dr.
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa INTENSIVE CARE POLICY & PROCEDURES Policy Number: R-25 Subject: Policy: Purpose: Cardioversion - Synchronized An RN may perform a synchronized
More informationPrinciples of Electrical Currents. HuP 272
Principles of Electrical Currents HuP 272 Electricity is an element of PT modalities most frightening and least understood. Understanding the basis principles will later aid you in establishing treatment
More informationDEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION
1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,
More informationOutline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work
Electrical Activity of the Human Heart Oguz Poroy, PhD Assistant Professor Department of Biomedical Engineering The University of Iowa Outline Basic Facts about the Heart Heart Chambers and Heart s The
More informationManual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev
Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value
More informationOBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.
Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe
More informationISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers
INTERNATIONAL STANDARD ISO 14708-2 Second edition 2012-08-15 Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers Implants chirurgicaux Dispositifs médicaux implantables actifs
More informationBiomedical Instrumentation
Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous
More informationEC6001 MEDICAL ELECTRONICS
EC6001 MEDICAL ELECTRONICS UNIT I ELECTRO-PHYSIOLOGY AND BIO-POTENTIAL RECORDING The origin of Bio-potentials; bio-potential electrodes, biological amplifiers, ECG, EEG, EMG, PCG, lead systems and recording
More informationMedical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) The story so far
Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) The story so far The heart pumps blood around the body. It has four chambers which contact
More informationManual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing
Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:
More informationImplants for surgery Active implantable medical devices. Part 7: Particular requirements for cochlear implant systems
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 14708-7 First edition 2013-01-15 Implants for surgery Active implantable medical devices Part 7: Particular requirements for cochlear implant systems
More informationPacing Lead Implant Testing. Document Identifier
Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant
More informationEE 4BD4 Lecture 20. Therapeutic Stimulation
EE 4BD4 Lecture 20 Therapeutic Stimulation 1 2 Extracellular Stimulation (at cathode) 3 4 Design of FES (cont.): Example stimulus waveform shapes: monophasic, biphasic, chopped, triphasic, and asymmetric,
More informationFigure 2. Normal ECG tracing. Table 1.
Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are
More informationElectrodes (for TransAeris System)
Surgical Procedures for Implantation of TransLoc Electrodes (for TransAeris System) 1 R L 1 TransAeris System TransLoc Electrode Kit 3 3 TransAeris Patient Kit (FrictionLoc) TransAeris Patient Kit (TransAeris
More information8/28/18. Anesthetic considerations for patients with implanted devices for treating chronic pain and more. Objectives:
Anesthetic considerations for patients with implanted devices for treating chronic pain and more Alaa Abd-Elsayed, MD, MPH Medical Director, UW Pain Services Medical Director, Pain Clinic Section Head,
More informationEI2311 BIOMEDICAL INSTRUMENTATION
66 EI2311 BIOMEDICAL INSTRUMENTATION 1. What is meant by cell? UNIT I PHYSIOLOGY AND TRANSDUCERS The basic living unit of the body is cell. The function of organs and other structure of the body is understood
More informationThe heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.
PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in
More informationISO INTERNATIONAL STANDARD. Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers
INTERNATIONAL STANDARD ISO 14708-2 Second edition 2012-08-15 Implants for surgery Active implantable medical devices Part 2: Cardiac pacemakers Implants chirurgicaux Dispositifs médicaux implantables actifs
More informationMini Pulse Electronic Stimulator
Mini Pulse Electronic Stimulator Model: PM-180 Operating Manual IMPORTANT: Please read all instructions before using this product. Retain this manual for future reference. www.santamedical.com IMPORTANT
More informationProSim 3 and 2. Vital Signs Simulators
ProSim 3 and 2 Vital Signs Simulators Technical Data Don t need a comprehensive patient monitor tester? The 6-in-1 ProSim 3 and 4-in-1 ProSim 2 Vital Signs Simulators are clear choices for biomedical engineers
More informationProSim 3 and 2 Vital Signs Simulators
ProSim 3 and 2 Vital Signs Simulators Technical Data Don t need a comprehensive patient monitor tester? The 6-in-1 ProSim 3 and 4-in-1 ProSim 2 Vital Signs Simulators are clear choices for biomedical engineers
More informationThe HeartCheck TM PEN Handheld ECG Is your heartbeat slow, fast, or irregular? Are you at risk? Put your heart health in your own hands
The HeartCheck TM PEN Handheld ECG Is your heartbeat slow, fast, or irregular? our hands Are you at risk? Put your heart health in your own hands The first FDA-Cleared device that can be unlocked to allow
More informationMultiStim SWITCH and MultiStim SENSOR Setting the trend in nerve stimulation
MultiStim SWITCH and MultiStim SENSOR Setting the trend in nerve stimulation Plexus and Epidural Anaesthesia The essential advantage regarding safety MultiStim SENSOR and MultiStim SWITCH With MultiStim
More informationEXTERNAL TEMPORARY PACEMAKERS
Czech Republic EXTERNAL TEMPORARY PACEMAKERS Lukas Peter Engineering and Business Coordinator Pavel Taborsky Chief Executive Officer MEDIATRADE s.r.o. PRAGUE Valasske Mezirici MEDIATRADE s.r.o. More than
More informationThe Normal Electrocardiogram
C H A P T E R 1 1 The Normal Electrocardiogram When the cardiac impulse passes through the heart, electrical current also spreads from the heart into the adjacent tissues surrounding the heart. A small
More informationELECTRICAL INJURY 9/21/2015 I HAVE NO DISCLOSURES WE HAVE OBTAINED APPROVAL FOR USE OF IDENTIFIABLE PATIENT PHOTOS
ELECTRICAL INJURY SAMUEL P. MANDELL, MD, MPH ASSISTANT PROFESSOR OF SURGERY UNIVERSITY OF WASHINGTON SEPTEMBER 28, 2015 I HAVE NO DISCLOSURES WE HAVE OBTAINED APPROVAL FOR USE OF IDENTIFIABLE PATIENT PHOTOS
More informationCSA Z32 TESTING GUIDELINE AND PROCEDURES
CSA-Z32 Testing CSA Z32 TESTING GUIDELINE AND PROCEDURES Table of Contents CSA Z32 TESTING GUIDELINE AND PROCEDURES...1 Table of Contents...1 FACILITIES REQUIRED TO MEET CSA Z32:...1 TEST #1: INSULATION
More informationCORONARY ARTERIES HEART
CARDIAC/ECG MODULE THE HEART CORONARY ARTERIES FIBRILLATING HEART CORONARY ARTERIES HEART PRACTICE RHYTHMS PRACTICE RHYTHMS ELECTRICAL CONDUCTION SA Node (60 100) Primary pacemaker AV Node (40 60) ***Creates
More informationThis leaflet is intended for patients with permanent pacemakers and their parents.
Permanent PACEMAKER This leaflet is intended for patients with permanent pacemakers and their parents. It serves as an information tool and a guide that offers some advice in order to live safely. General
More informationFlowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields
Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and
More informationThe Automated Defibrillator: A Biomedical Engineering Success Story. Dr. James A. Smith
The Automated Defibrillator: A Biomedical Engineering Success Story Dr. James A. Smith A Aorta VCS Superior Vena Cava RA Right Atrium RV Right Ventricle LV Left Ventricle The Heart Beating Heart: Video
More informationAdvanced ICD Concepts
1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation
More informationAutomatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms
Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable
More informationNeurostimulation Systems
Neurostimulation Systems Expanding the Array of Pain Control Solutions PROVEN EFFECTIVE FOR SIMPLE PAIN ITREL 3 SYSTEM For simple neuropathic pain Totally implantable for enhanced quality of life Easy
More informationELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG) The heart is a muscular organ, which pumps blood through the blood vessels of the circulatory system. Blood provides the body with oxygen and nutrients, as well as assists in
More informationSinus rhythm with premature atrial beats 2 and 6 (see Lead II).
Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess
More informationPediatric pacemakers & ICDs:
Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components
More informationCompatible with the following products TENS. Replacement Pads. User Manual. Last revised: V Last revised: V
Compatible with the following products TENS Replacement Pads User Manual Last revised: V11-160923 Last revised: V4-170619 Contents Introduction 2-3 Parts 3 General Warnings and Safety 4-10 Using your TENS
More informationWHOLE BODY VIBRATION THERAPY
Body Vibe WHOLE BODY VIBRATION THERAPY Instruction manual INCREASES CIRCULATION STIMULATES LYMPHATICS REDUCES CELLULITE LIQUIFIES FAT IMPROVES BALANCE DECREASES CORTISOL INCREASES ENDURANCE NON-IMPACT
More informationThe Electrocardiogram
The Electrocardiogram Chapters 11 and 13 AUTUMN WEDAN AND NATASHA MCDOUGAL The Normal Electrocardiogram P-wave Generated when the atria depolarizes QRS-Complex Ventricles depolarizing before a contraction
More informationFunction of Cell Membrane
Function of Cell Membrane Electrical Potential Concept Electrical Circuit Human Cell electrical circuit Is Human Cell rechargeable? Why Negative Potential Is Important. Why Positive Potential Can Be Harmful
More informationModel 5392 EPG Temporary Pacer
Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short
More informationC J Mieny, U Mennen: Principles of Surgical Patient Care - Volume I. Chapter 2: Intraoperative Care
C J Mieny, U Mennen: Principles of Surgical Patient Care - Volume I Chapter 2: Intraoperative Care J A L van Wyk: 2.1 Aseptic and Antiseptic Techniques An accidental break in skin or mucosal surface as
More informationEnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can
EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds
More informationTIP CARD. Model Temporary External Pacemaker
TIP CARD Model 5392 Temporary External Pacemaker PRE-USE What to do... What NOT to do... Battery Replace the battery for each new patient. Do NOT reuse battery. Physical Check case for cracks/damage.**
More informationMAGPRO. Versatility in Magnetic Stimulation. For clinical and research use
MAGPRO Versatility in Magnetic Stimulation For clinical and research use Magnetic Stimulation From A World Leader MagPro is a complete line of non-invasive magnetic stimulation systems, including both
More informationSTUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES
STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES March, 2007 Ministry of Internal Affairs and Communications (MIC) Introduction We are moving toward the creation of a ubiquitous network society
More informationCardiac Cycle. Each heartbeat is called a cardiac cycle. First the two atria contract at the same time.
The Heartbeat Cardiac Cycle Each heartbeat is called a cardiac cycle. First the two atria contract at the same time. Next the two ventricles contract at the same time. Then all the chambers relax. http://www.youtube.com/watch?v=frd3k6lkhws
More informationChapter 23 Unit 28. Therapeutic Modalities
Chapter 23 Unit 28 Therapeutic Modalities Chapter Objectives Discuss the purpose of therapeutic modalities Explain the legal implications associated with the use of therapeutic modalities List the different
More informationThis document is a preview generated by EVS
INTERNATIONAL STANDARD ISO 14708-7 First edition 2013-01-15 Implants for surgery Active implantable medical devices Part 7: Particular requirements for cochlear implant systems Implants chirurgicaux Dispositifs
More informationHow Do We Sense, Think, and Move? -- Lab #11 Bioelectronics Measuring Electrical Properties of the Body
How Do We Sense, Think, and Move? -- Lab #11 Bioelectronics Measuring Electrical Properties of the Body Experiment #1 Your Body's Resistance Equipment: Digital multimeter, Banana leads Important Equipment
More informationM Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use
DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.
More informationOWNER'S MANUAL OxyTrainer
M Model # 7102 OWNER'S MANUAL OxyTrainer WE WE ARE ARE HERE HERE TO TO HELP HELP YOU YOU 1-866-924-1688 Email: service@paradigmhw.com Address: Paradigm Health & Wellness, Inc. 1189 Jellick Ave. City of
More informationLotus transducer Instructions for use
Lotus transducer Instructions for use Instructions for use transducer Before using the product, please read the following information thoroughly Important This document provides instructions for using
More informationNewer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.
Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded
More informationChapter 3 Biological measurement 3.1 Nerve conduction
Chapter 3 Biological measurement 3.1 Nerve conduction Learning objectives: What is in a nerve fibre? How does a nerve fibre transmit an electrical impulse? What do we mean by action potential? Nerve cells
More informationThe Do s and Don ts of Pressure Transducers
The Do s and Don ts of Pressure Transducers ABSTRACT When specifying a pressure transducer for a process measurement, a number of items have to be considered. Some of the more important ones are discussed
More informationContents. Helpline No. UK/ North Ireland Rep. Ireland Web Support Model Number: 12005
Contents 1 02 03 04 06 07 11 12 13 15 17 18 19 19 Contents Welcome Safety Instructions Box Contents Notes on Measurements Getting Started Measuring Weight Saving Personal Data Measuring Body Fat, Water
More informationYes, out of sight out of mind.
Micra TRANSCATHETER PACING SYSTEM 25.9 mm Yes, out of sight out of mind. cardiocapsule Meet Micra The Invisible Pacemaker Micra is the world s smallest pacemaker 1. It leaves no bump under the skin, no
More informationSLEEP IMPROVING WRISTBAND. Item No Owner s Guide
SLEEP IMPROVING WRISTBAND Item No. 205350 Owner s Guide Thank you for purchasing the Sharper Image Sleep Improving Wristband. Based on ancient Chinese acupuncture principles, this biofeedback device uses
More informationLABORATORY INVESTIGATION
LABORATORY INVESTIGATION Recording Electrocardiograms The taking of an electrocardiogram is an almost universal part of any complete physical examination. From the ECG record of the electrical activity
More informationECG SENSOR ML84M USER S GUIDE. CENTRE FOR MICROCOMPUTER APPLICATIONS
ECG SENSOR ML84M USER S GUIDE CENTRE FOR MICROCOMPUTER APPLICATIONS http://www.cma-science.nl Short description The ECG sensor measures electrical potentials produced by the heart (Electro- cardiogram).
More informationPACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I
1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL
More informationINTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS
INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS
More informationBiomedical Engineering in Root Cause Analysis Example: Assessing Infant Apnea-Related Deaths
Bruce H. Barkalow, Ph.D., PE, CCE, William E. Grant, M.A., M.L.I.S., and Farrah J. Curran, B.S. Biomedical Engineering in Root Cause Analysis Example: Assessing Infant Apnea-Related Deaths Biomedical/Clinical
More informationNOVIN medical engineering Co. Designer and manufacturer of physiotherapy equipment
NOVIN medical engineering Co. Designer and manufacturer of physiotherapy equipment NOVIN medical engineering Co. started its journey in the field of physiotherapy equipment in 1998. Using senior and experienced
More informationSaeboStim Micro. Saebo INSTRUCTION MANUAL
SaeboStim Micro Saebo INSTRUCTION MANUAL Introduction Saebo is pleased to provide you with the latest in evidence-based upper limb rehabilitation. The SaeboStim Micro provides sensory electrical stimulation
More informationPeri-operative management of pacemakers and implantable cardiac defibrillators
Guideline Title: Peri-operative management of patients fitted with Permanent Pacemakers (PPMs) and Implantable Cardioverter Defibrillators (ICDs). Author(s): Jessica Osman (Chief Pacing Physiologist),
More informationAdvisa MRI and Ensura MRI SureScan pacing systems
Advisa MRI and Ensura MRI SureScan pacing systems MR Conditional dual chamber and single chamber SureScan pacing systems with SureScan leads MRI Technical Manual The following list includes trademarks
More informationAfter Your Pacemaker Generator Change
UW MEDICINE PATIENT EDUCATION After Your Pacemaker Generator Change What you need to know Who to Call If you have questions about your appointments, call the Cardiology Clinic at 206.598.4300. If you have
More informationPatient Safety During Cardiac Catheterization
119 CIRCULATION VOL 58, No 6, DECEMBER 1978 32. Neff MS, Kim KE, Persoff M, Onesti G, Swartz C: Hemodynamics of uremic anemia. Circulation 43: 876, 1971 33. Cangiano JL, Ramirez-Mux, Ramirez-Gonzalez R,
More informationIntroduction to Physical Agents Part II: Principles of Heat for Thermotherapy
Introduction to Physical Agents Part II: Principles of Heat for Thermotherapy Mohammed TA, Omar momarar@ksu.edu.sa Dr.taher_m@yahoo.com Mobile : 542115404 Office number: 2074 Objectives After studying
More informationModule 1: Fundamentals of Patient Monitoring. Contents. Introduction. Why is monitoring in MR different? Exercise 1_1
MRES 7003 - MR Safety and Monitoring Module 1: Fundamentals of Patient Monitoring Contents Introduction Why is monitoring in MR different? Exercise 1_1 What parameters should we monitor? Selection of monitoring
More informationContra-Indications, Warnings, Cautions & Precautions
Quick Reference Training Contra-Indications, Warnings, Cautions & Precautions BEST Devices Biofeedback Electro-Stimulation Technology 13140 Coit Road, Suite 515, Dallas, TX 75240 (T) 214.575.2820 avazzia.com
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-7 Biopotentials- 2) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/ Electromyogram (EMG) Skeletal muscles are
More informationElectrocardiography Biomedical Engineering Kaj-Åge Henneberg
Electrocardiography 31650 Biomedical Engineering Kaj-Åge Henneberg Electrocardiography Plan Function of cardiovascular system Electrical activation of the heart Recording the ECG Arrhythmia Heart Rate
More informationEssentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH
Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing
More informationECG Check Wireless 1- Lead Cardiac Monitor. Rx Use
ECG Check Wireless 1- Lead Cardiac Monitor Rx Use Dear ECG Check Owner: Thank you for selecting the ECG Check cardiac monitor. This user guide includes information and instructions about your new ECG Check
More informationAnalysis of Human Cardiovascular System using Equivalent Electronic System
Analysis of Human Cardiovascular System using Equivalent Electronic System N. Vinoth 1, S. Nagarjuna Chary 2 Dept of Electronics and Instrumentation Engineering, Annamalai University, Annamalai nagar,
More informationSafety Issue of Static Magnetic Field (1)
MRI Safety Safety Issue of Static Magnetic Field (1) Currently, typical MRI utilizes static field with strengths of 0.5 to 3 Tesla (FDA guideline is max 4 T for infants < 1 month and 8 T for others) 1
More informationEffects of myocardial infarction on catheter defibrillation threshold
Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1983 Effects of myocardial infarction on catheter defibrillation threshold
More informationPROFESSIONAL RECOVERY FOR SERIOUS ATHLETES. Vibrating 4 Speed Rechargeable Foam Roller Manual
PROFESSIONAL RECOVERY FOR SERIOUS ATHLETES Vibrating 4 Speed Rechargeable Foam Roller Manual Thank you for purchasing the TimTam Vibrating Foam Roller. Before proceeding, please read the instruction manual
More informationV4, V5 and V6 follow the 5 th intercostal space and are NOT horizontal as indicated in the image. Page 1 of 7
V4, V5 and V6 follow the 5 th intercostal space and are NOT horizontal as indicated in the image. Page 1 of 7 "Is it correct to place the electrodes under the female breasts?" Placement of electrodes directly
More informationVENTRICULAR DEFIBRILLATOR
VENTRICULAR DEFIBRILLATOR Group No: B03 Ritesh Agarwal (06004037) ritesh_agarwal@iitb.ac.in Sanket Kabra (06007017) sanketkabra@iitb.ac.in Prateek Mittal (06007021) prateekm@iitb.ac.in Supervisor: Prof.
More informationRN-BC, MS, CCRN, FAHA
Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine
More informationPATIENT PROGRAMMER 37642
PATIENT PROGRAMMER 37642 Medtronic DBS Therapy user manual Activa PC Model 37601 Activa RC Model 37612 Activa SC Model 37602 Activa SC Model 37603! USA Rx only 2008 Medtronic, Activa, and SoftStart/Stop
More informationMuscle Stimulation. Page 1
Muscle Stimulation Page 1 Page 2 Page 3 Table of Contents: Get Started with the E-Wave Pg 4 Controls and Features Pg 4 Factory Settings Pg 6 Electrodes and Skin Care Pg 7 Batteries Pg 7 Indications Pg
More informationAspireSR. Essential information about your new device
AspireSR Essential information about your new device FAQs for Patients Your doctor recommends that you receive the latest VNS Therapy device. This brochure will give you more information on the differences
More informationDefibrillator. BiomedGuy
Defibrillator BiomedGuy Medtronic Physiocontrol LifePak 10 Introduction This life-support system is used by paramedic, hospital staff, and other trained authorized healthcare providers. Provides, ECG,
More informationYour Subcutaneous Implantable Cardiac Defibrillator (ICD)
UW MEDICINE PATIENT EDUCATION Your Subcutaneous Implantable Cardiac Defibrillator (ICD) What you need to know Who to Call If you have questions about your appointments, call the Cardiology Clinic at 206-598-4300.
More information