Routine low-radiation-dose coronary computed tomography angiography

Size: px
Start display at page:

Download "Routine low-radiation-dose coronary computed tomography angiography"

Transcription

1 European Heart Journal Supplements (2014) 16 (Supplement B), B12 B16 The Heart of the Matter doi: /eurheartj/suu024 Routine low-radiation-dose coronary computed tomography angiography Mouaz H. Al-Mallah 1,2 *, Ahmed Aljizeeri 1, Mohsen Alharthi 1,2, and Ahmed Alsaileek 1,2 1 Division of Advanced Cardiac Imaging, King Abdulaziz Cardiac Center, King Abdulaziz Medical City, Ministry of National Guard - Health Affairs, PO BOX 22490, Riyadh 11426, Kingdom of Saudi Arabia 2 King Saud bin Abdulaziz University for Health Sciences, Riyadh, Kingdom of Saudi Arabia KEYWORDS Radiation dose; Coronary computed tomography angiography Introduction In recent years, coronary computed tomography angiography (CCTA) has emerged as a novel tool for the noninvasive detection of coronary artery disease (CAD). 1,2 Multiple studies have demonstrated its high sensitivity and a negative predictive value for the detection and exclusion of CAD. 3,4 However, CCTA was limited by the use of high-radiation doses, which may lead to downstream malignancies in the future. 5,6 In addition, a multicenter study demonstrated significant variability in the radiation * Corresponding author. Tel: , Ext , Fax: , Ext , mouaz74@gmail.com Recent advances in coronary computed tomography angiography (CCTA) increased its utilization as a tool for non-invasive detection of coronary artery disease (CAD). The aim of this analysis is to determine the impact of adopting new radiation dose-reducing techniques on the radiation exposure in consecutive patients undergoing CCTA. We included 1341 consecutive patients who underwent CCTA to rule out CAD between January 2007 and December New dose-reducing techniques were adopted in July 2010 in 1034 patients. These included high pitch scanning, 100 KVP (Kilovoltage Peak) imaging, and iterative reconstruction. The total radiation dose was calculated for each scan from the dose length product multiplied by conversion factor (0.014). The annual median radiation doses were compared over the study period. After the adoption of the new scanning techniques (n ¼ 578), 53% of the scans were done with high pitch scanning, 46% with prospective gating, and 1% with retrospective gating. This was associated with.90% reduction in the radiation doses with a median radiation dose of 2.7, 1.5, 1.7, and 1.5 msv (Millisievert) in 2010, 2011, 2012, and 2013, respectively. A total of 15 and 66% of the CCTA scans had a radiation dose of,1 and 2 msv, respectively. There was no difference in the frequency of non-diagnostic studies or imaging quality before and after July Our analysis demonstrates that, in the current era, low-radiation CCTA can be routinely done in clinical practice. exposure associated with CCTA, even among users of the same scanners. 6 In the past few years, all vendors developed new tools that allow to minimize the radiation exposure associated with CCTA, like prospective gating, 100 KVP (Kilovoltage Peak) imaging, iterative reconstruction, high pitch scanning and volumetric imaging. 7,8 Two single-center studies suggested that the utilization of these tools would result in a significant reduction in the associated radiation dose in a highly selected subgroup of patients. 9,10 However, there are limited data on whether these tools can be utilized in all consecutive patients undergoing CCTA. Thus, the aim of this analysis is to determine whether lowradiation doses can be routinely achieved in all consecutive patients referred for CCTA. Published on behalf of the European Society of Cardiology. All rights reserved. & The Author For permissions please journals.permissions@oup.com

2 Routine low-radiation-dose coronary computed tomography angiography B13 Methods We included consecutive patients who underwent coronary CT angiography for the exclusion of CAD between September 2006 and December Patients younger than 18 years old were excluded from this analysis. In addition, patients with known CAD (prior myocardial infarction, prior angioplasty, and coronary artery bypass grafting), atrial fibrillation, and patients undergoing non-coronary cardiac CTwere excluded from this analysis. Coronary computed tomography angiography image acquisition All CCTA scans were performed on either a 64-multidetector row scanner (High Definition Scanner, General Electric, Milwaukee, WI, USA) or a dual source cardiac CTsystem with high pitch scanning capability (Siemens Healthcare, Erlangen, Germany). Patients were in normal sinus rhythm and were capable of the breath-hold needed for CCTA. For patients presenting with baseline heart rates.65 b.p.m., beta-blockers were administered to slow the heart rate. Following a scout radiograph, contrast timing was determined using a test bolus (15 ml contrast) to detect optimal time for peak contrast opacification in the ascending aorta. Nitroglycerine sublingually was administered immediately before contrastinjection. During CCTA acquisition, ml of iodinated contrast was injected followed by a ml saline flush. All CCTA scans were attended and sometimes performed by a consultant cardiologist with special training in CCTA. Radiation dose reduction program In July 2010, the new radiation reduction tools became available in our center and were strictly utilized for every CCTA. A radiation reduction protocol was developed and included minimizing longitudinal scan range; use of sufficient beta-blockers to control heart rate and heart rate variability; using 100 KVP for patients with body mass index,30 kg/m 2 ; use of high pitch scanning when possible; and minimizing the milliamperage used and applying iterative reconstruction in all cases in addition to direct supervision of the CCTA by a consultant physician. 5 Estimation of radiation dose Radiation doses were estimated from the scanner provided protocol summary that contain the dose length product for each image series, which integrated estimated absorbedradiation in the x-, y-, and z-directions based on the CT dose index volume. The effective radiation dose was derived from the summed dose length product Table 1 Baseline characteristics of the study population multiplied by the European Working Group for Guidelines on Quality Criteria in Computed Tomography conversion coefficient (k ¼ msv (Millisievert)/mGy cm). 6 Image quality assessment Physicians rated the quality of each image on a per-patient basis at the time of scan interpretation. This is adopted from the prior published criteria. 5 Excellent (score ¼ 1) was defined as the complete absence of motion artefacts, excellent signal-to-noise ratio, and clear delineation of vessel walls, with the ability to assess luminal stenosis as well as plaque characteristics. Good (score ¼ 2) was defined as non-limiting motion artefacts, reduced signal-to-noise ratio, and/or calcifications are present, with preserved ability to assess luminal stenosis as well as plaque characteristics. Fair (score ¼ 3) was defined as reduced image quality due to any combination of noise, motion, poor contrast enhancement, or calcium that significantly impairs ease of interpretation, but image quality is sufficient to rule out significant stenosis. Non-diagnostic (score ¼ 4) was defined as reduced image quality that precludes adequate assessment of stenosis in the majority of vessels. Statistical analysis Patients were grouped into two groups, before and after July Group comparisons were conducted between the two groups using x 2 testing or Students t-testing when appropriate. Results A total of 1341 patients underwent CCTA in the study period with 307 patients (23%) before July The baseline characteristics of the included patients are summarized in Table 1. After July 2010, patients undergoing CCTA had a lower risk profile with less frequent hypertension, diabetes, and dyslipidaemia. However, there was no difference in the body mass index of the patients before or after the implementation of the low-radiation CCTA program (Table 1). In accordance to the low-radiation protocol, the higher radiation retrospective gating protocol was rarely used after July 2010 as presented in Table 2 (,1%). Most patients were scanned using the lower radiation protocols (prospective gating 46% and high pitch scanning 52%) and more often with 100 KVP imaging (Figure 1) Before July 2010 After July 2010 P-value Mean age (years) Male gender 69% 58%, Hypertension 66% 52%, Diabetes 44% 36% Dyslipidaemia 70% 56%, Smoking 28% 18% Mean body mass index (kg/m 2 ) Group comparisons are made before and after the implementation of the low-radiation imaging protocols.

3 B14 M.H. Al-Mallah et al. Table 2 Scanning parameters before and after the implementation of the low-radiation imaging protocols Before July 2010 After July 2010 P-value Mean heart rate (b.p.m.) Mode of image acquisition Prospective gating 10% 46%, Retrospective gating 90% 1% High pitch scanning 0% 53% Figure 1 Utilization of the low tube voltage before and after the implementation of a low-radiation-dose imaging protocol. As per the protocol, more patients are imaged with the lower voltage despite no change in the body mass index of the patients before and after July Figure 2 Median annual radiation dose. There was.90% drop in the radiation dose after the implementation of low-radiation-dose imaging protocol (P, ). It was also sustained for 4 years. Adherence to the low-radiation protocol resulted in a significant drop in the radiation dose from a median of msv after July 2010 (Figure 2). Most patients after July 2010 had a radiation dose of,2 msv as shown in Figure 3. Only 6.2% of the study population had a radiation dose of.5 msv after July 2010, compared with 91% prior to July 2010 (P, ). Most importantly, there were no differences in the image quality before or after the implementation of the low-radiation protocol or in the rate of poor quality scans (Figure 4).

4 Routine low-radiation-dose coronary computed tomography angiography B15 Figure 3 Percentage of patients according to the radiation dose (P, ). Coronary computed tomography angiography with a radiation dose of,5 msv was rarely possible before July Figure 4 Image quality before and after the implementation of low-radiation-dose imaging protocol. There were no differences in the image quality before or after the implementation of the low-radiation protocol or in the rate of poor quality scans. Discussion Our analysis shows that, in the current era, low-radiation CCTA can be routinely used in consecutive patients without prior CAD. This is possible with the utilization of the new imaging radiation lowering techniques and tools that allow the minimization of radiation exposure without impacting the image quality or the diagnostic accuracy of the study. Our analysis is in agreement with prior analysis from the advanced consortium of cardiovascular imaging analysis from the state of Michigan, which showed that utilizing radiation exposure lowering measures would allow for routine low-dose CCTA. The median radiation dose in the consortium was 4.8 msv at the end of the study. 11 In our analysis, we took this further and showed that the radiation dose can be lowered further to,2 msv. This is achieved primarily by applying iterative reconstruction in every patient and selecting the imaging protocol that is associated with the least radiation exposure according to the patient s clinical status. In addition, the presence of the physician by the scanner console in every case allowed for adjusting the scanning parameters and achieving this result. In conclusion, we have shown that in the current CCTA can be and should be routinely done in consecutive patients with low-radiation doses. Conflict of interest: none declared. References 1. Achenbach S, Friedrich MG, Nagel E, Kramer CM, Kaufmann PA, Farkhooy A, Dilsizian V, Flachskampf FA. CV imaging: what was new in 2012? JACC Cardiovasc Imaging 2013;6:

5 B16 M.H. Al-Mallah et al. 2. Perrone-Filardi P, Achenbach S, Mohlenkamp S, Reiner Z, Sambuceti G, Schuijf JD, Van der Wall E, Kaufmann PA, Knuuti J, Schroeder S, Zellweger MJ. Cardiac computed tomography and myocardial perfusion scintigraphy for risk stratification in asymptomatic individuals without known cardiovascular disease: a position statement of the working group on nuclear cardiology and cardiac CTof the European Society of Cardiology. Eur Heart J 2011;32: a, b. 3. Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E, Scherer M, Bellinger R, Martin A, Benton R, Delago A, Min JK. Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter accuracy (assessment by coronary computed tomographic angiography of individuals undergoing invasive coronary angiography) trial. J Am Coll Cardiol 2008;52: Raff GL, Gallagher MJ, O Neill WW, Goldstein JA. Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 2005;46: RaffGL, ChinnaiyanKM, ShareDA, Goraya TY, KazerooniEA, MoscucciM, Gentry RE, Abidov A. Radiation dose from cardiac computed tomography before and after implementation of radiation dose-reduction techniques. JAMA 2009;301: Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC, McCollough C, Martinoff S, Kastrati A, Schomig A, Achenbach S. Estimated radiation dose associated with cardiac CT angiography. JAMA 2009;301: Leipsic J, Labounty TM, Heilbron B, Min JK, Mancini GB, Lin FY, Taylor C, Dunning A, Earls JP. Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: The ERASIR study. AJR Am J Roentgenol 2010;195: Leipsic J, Labounty TM, Heilbron B, Min JK, Mancini GB, Lin FY, Taylor C, Dunning A, Earls JP. Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CTangiography. AJR Am J Roentgenol 2010;195: Fuchs TA, Stehli J, Bull S, Dougoud S, Clerc OF, Herzog BA, Buechel RR, Gaemperli O, Kaufmann PA. Coronary computed tomography angiography with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination. Eur Heart J 2014;35: Achenbach S, Marwan M, Ropers D, Schepis T, Pflederer T, Anders K, Kuettner A, Daniel WG, Uder M, Lell MM. Coronary computed tomography angiography with a consistent dose below 1 msv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J 2010;31: Chinnaiyan KM, Boura JA, DePetris A, Gentry R, Abidov A, Share DA, Raff GL. Progressive radiation dose reduction from coronary computed tomography angiography in a statewide collaborative quality improvement program: results from the advanced cardiovascular imaging consortium. Circ Cardiovasc Imaging 2013;6:

Comparison of 64 slice multi detector CT vs 128 slice multi detector CT for acute chest pain evaluation in the chest pain unit

Comparison of 64 slice multi detector CT vs 128 slice multi detector CT for acute chest pain evaluation in the chest pain unit Comparison of 64 slice multi detector CT vs 128 slice multi detector CT for acute chest pain evaluation in the chest pain unit Poster No.: C-0642 Congress: ECR 2010 Type: Topic: Authors: Keywords: Keywords:

More information

Organization and metrics of the Division of Advanced Cardiac Imaging, King Abdulaziz Cardiac Center

Organization and metrics of the Division of Advanced Cardiac Imaging, King Abdulaziz Cardiac Center European Heart Journal Supplements (2014) 16 (Supplement B), B7 B11 The Heart of the Matter doi:10.1093/eurheartj/suu016 Organization and metrics of the Division of Advanced Cardiac Imaging, King Abdulaziz

More information

Computed tomography in coronary imaging: current status

Computed tomography in coronary imaging: current status 7 Computed tomography in coronary imaging: current status ARJUN NAIR AND ANAND DEVARAJ Recent technological advances have led to improvements in the use of computerised tomography for coronary imaging.

More information

Cardiopulmonary Imaging Original Research

Cardiopulmonary Imaging Original Research Cardiopulmonary Imaging Original Research Bischoff et al. Versus Helical Mode for Coronary CTA Cardiopulmonary Imaging Original Research Bernhard Bischoff 1 Franziska Hein 1 Tanja Meyer 1 Markus Krebs

More information

Purpose. Methods and Materials

Purpose. Methods and Materials Comparison of iterative and filtered back-projection image reconstruction techniques: evaluation of heavily calcified vessels with coronary CT angiography Poster No.: C-1644 Congress: ECR 2011 Type: Scientific

More information

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Low-Dose Computed Tomography Coronary Angiography With Prospective Electrocardiogram Triggering

Low-Dose Computed Tomography Coronary Angiography With Prospective Electrocardiogram Triggering Journal of the American College of Cardiology Vol. 57, No. 3, 2011 2011 by the American College of Cardiology Foundation ISSN 0735-1097/$36.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2010.08.634

More information

Computed Tomography of the Coronary Arteries

Computed Tomography of the Coronary Arteries Cardiology Update DAVOS 2011 Computed Tomography of the Coronary Arteries Anders Persson M.D., Ph.D Director, Assoc. Professor Center for Medical Image Science and Visualization Linköping University SWEDEN

More information

Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography

Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography Review Article Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography Zhonghua Sun Discipline of Medical Imaging, Department of Imaging and Applied

More information

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Junfu Liang 1,2, Hui Wang 1, Lei Xu 1, Li Dong 1, Zhanming

More information

CT Imaging of Atherosclerotic Plaque. William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA

CT Imaging of Atherosclerotic Plaque. William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA CT Imaging of Atherosclerotic Plaque William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA PREVALENCE OF CARDIOVASCULAR DISEASE In 2006 there were 80 million

More information

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography Journal of the American College of Cardiology Vol. 52, No. 18, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.07.048

More information

Is computed tomography angiography really useful in. of coronary artery disease?

Is computed tomography angiography really useful in. of coronary artery disease? Is computed tomography angiography really useful in screening patients with high risk of coronary artery disease? Myeong-Ki Hong, M.D. Ph D Professor of Medicine Division of Cardiology, Severance Cardiovascular

More information

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population

Improved Noninvasive Assessment of Coronary Artery Bypass Grafts With 64-Slice Computed Tomographic Angiography in an Unselected Patient Population Journal of the American College of Cardiology Vol. 49, No. 9, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2006.10.066

More information

Disclosure Information

Disclosure Information Coronary CTA Pearls and Pitfalls Ricardo C. Cury, MD, FSCCT, FAHA, FACC Chairman of Radiology Radiology Associates of South Florida Director of Cardiac Imaging Miami Cardiac and Vascular Institute Past-President

More information

Cardiac CT imaging in coronary artery disease: Current status and future directions

Cardiac CT imaging in coronary artery disease: Current status and future directions Research Highlight Cardiac CT imaging in coronary artery disease: Current status and future directions Zhonghua Sun Discipline of Medical Imaging, Department of Imaging and Applied Physics, Curtin University,

More information

Radiation dose of cardiac CT what is the evidence?

Radiation dose of cardiac CT what is the evidence? Eur Radiol (2009) 19: 1311 1315 DOI 10.1007/s00330-009-1312-y CARDIAC Hatem Alkadhi Radiation dose of cardiac CT what is the evidence? Received: 10 December 2008 Revised: 2 January 2009 Accepted: 12 January

More information

Research Article. Open Access. Jin-Da WANG, Hua-Wei ZHANG, Qian XIN, Jun-Jie YANG, Zhi-Jun SUN, Hong-Bin LIU, Lian CHEN, Luo-Shan DU, Yun-Dai CHEN

Research Article. Open Access. Jin-Da WANG, Hua-Wei ZHANG, Qian XIN, Jun-Jie YANG, Zhi-Jun SUN, Hong-Bin LIU, Lian CHEN, Luo-Shan DU, Yun-Dai CHEN Journal of Geriatric Cardiology (2014) 11: 39 43 2014 JGC All rights reserved; www.jgc301.com Research Article Open Access Safety and efficacy of intravenous esmolol before prospective electrocardiogram-triggered

More information

CT Coronary Angiography - Indications: From the guidelines to clinical practice

CT Coronary Angiography - Indications: From the guidelines to clinical practice CT Coronary Angiography - Indications: From the guidelines to clinical practice Multimodality Working Group of Cardiovascular Imaging (Nuc C, CCT CMR) Hellenic Cardiology Society Seminars, Thessaloniki,

More information

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology,

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology, Chapter 4 Impact of Coronary Calcium Score on Diagnostic Accuracy of Multislice Computed Tomography Coronary Angiography for Detection of Coronary Artery Disease Gabija Pundziute, 1,3 Joanne D. Schuijf,

More information

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography Journal of the American College of Cardiology Vol. 46, No. 3, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.05.056

More information

The radiation dose in retrospective

The radiation dose in retrospective The radiation dose in retrospective gated tdcoronary computed td tomography (CCT) Saeed AL Ahmari, Ghormallah AL Zahrani, Sumiah AL Helali, Samir AL Dulikan, Abdullah Bafagih, HibaKhashojji Prince Sultan

More information

Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich

Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse CH-0 Zurich www.zora.uzh.ch Year: 01 Image quality and radiation dose comparison of prospectively triggered low-dose

More information

Physiologic Assessment by Cardiac CT

Physiologic Assessment by Cardiac CT Physiologic Assessment by Cardiac CT Mouaz Al-Mallah, MD MSc FACC Associate Professor of Medicine Division Head, Cardiac Imaging King Abdul-Aziz Cardiac Center National Guard Health Affairs Riyadh, Saudia

More information

Radiation Dose Optimization in Cardiac CT: A Technical Review. CHENG Wai Kwong. 17 May Contents

Radiation Dose Optimization in Cardiac CT: A Technical Review. CHENG Wai Kwong. 17 May Contents Radiation Dose Optimization in Cardiac CT: A Technical Review CHENG Wai Kwong 17 May 2014 Contents 1. Introduction and background 2. Cardiac CT synchronization techniques 3. 4. Conclusion 1 Introduction

More information

A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering

A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering DOI 10.1007/s10554-015-0802-z ORIGINAL PAPER A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering J. D. van Dijk 1,4 E. D. Huizing 1,4 P.

More information

Diagnostic value of coronary CT angiography with prospective ECG-gating in the

Diagnostic value of coronary CT angiography with prospective ECG-gating in the Diagnostic value of coronary CT angiography with prospective ECG-gating in the diagnosis of coronary artery disease: A systematic review and meta-analysis Zhonghua Sun PhD 1, Kwan-Hoong Ng PhD 2 1. Discipline

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

Cardiac CT: Your chest pain patient CAN be discharged from the Emergency Department

Cardiac CT: Your chest pain patient CAN be discharged from the Emergency Department Cardiac CT: Your chest pain patient CAN be discharged from the Emergency Department Asif Serajian, DO FACC FSCAI Co-director of Cardiac CT Elmhurst Memorial Hospital 1 No financial disclosures relevant

More information

Non-invasive Coronary Angiography: the Role, Limitations and Future of 64-Slice Spiral Computed Tomography Coronary Angiography

Non-invasive Coronary Angiography: the Role, Limitations and Future of 64-Slice Spiral Computed Tomography Coronary Angiography HOSPITAL CHRONICLES 2009, 4(3): 105 109 Review Non-invasive Coronary Angiography: the Role, Limitations and Future of 64-Slice Spiral Computed Tomography Coronary Angiography Arkadios C. Roussakis, MD

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

M Marwan, D Ropers, T Pflederer, W G Daniel, S Achenbach

M Marwan, D Ropers, T Pflederer, W G Daniel, S Achenbach Department of Cardiology, University of Erlangen, Erlangen, Germany Correspondence to: Dr M Marwan, Innere Medizin II, Ulmenweg 18, 91054 Erlangen, Germany; mohamed.marwan@ uk-erlangen.de Accepted 17 November

More information

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011

Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease. Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011 Multidetector CT Angiography for the Detection of Left Main Coronary Artery Disease Rani K. Hasan, M.D. Intro to Clinical Research July 22 nd, 2011 Outline Background Hypothesis Study Population Methodology

More information

Low-dose CT coronary angiography in the step-andshoot mode: diagnostic performance

Low-dose CT coronary angiography in the step-andshoot mode: diagnostic performance 1 Institute of Diagnostic Radiology, University Hospital Zurich, Zurich, Switzerland; 2 Clinic for Cardiovascular Surgery, University Hospital Zurich, Zurich, Switzerland; 3 Cardiovascular Centre, University

More information

Εξελίξεις και νέες προοπτικές στην καρδιαγγειακή απεικόνιση CT. Σταμάτης Κυρζόπουλος Ωνάσειο Καρδιοχειρουργικό Κέντρο

Εξελίξεις και νέες προοπτικές στην καρδιαγγειακή απεικόνιση CT. Σταμάτης Κυρζόπουλος Ωνάσειο Καρδιοχειρουργικό Κέντρο Εξελίξεις και νέες προοπτικές στην καρδιαγγειακή απεικόνιση CT Σταμάτης Κυρζόπουλος Ωνάσειο Καρδιοχειρουργικό Κέντρο No conflict of interest to disclose Noninvasive Cardiac Imaging Unresolved Issues-Future

More information

The Emerging Role of Cardiac CT in Cardiovascular Imaging. Anthony Gemignani, MD Vermont Cardiac Network April 28, 2016

The Emerging Role of Cardiac CT in Cardiovascular Imaging. Anthony Gemignani, MD Vermont Cardiac Network April 28, 2016 The Emerging Role of Cardiac CT in Cardiovascular Imaging Anthony Gemignani, MD Vermont Cardiac Network April 28, 2016 Conflict Disclosures I have no significant financial relationship with any companies

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015 2384 Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients GUIRU HE, XIAOPEI LIU, YAN LIU, WEI

More information

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006). Feasibility, image quality and radiation dose of coronary CT angiography (CCTA) in patients with atrial fibrillation using a new generation 256 multi-detector CT (MDCT) Poster No.: C-2378 Congress: ECR

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6

The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6 The Final 10-Year Follow-up Results from the Bari Randomized Trial J Am Coll Cardiol (2007) 49;1600-6 n&list_uids=17433949 64-Multislice Detector Computed Tomography Coronary Angiography as Potential Alternative

More information

Multislice CT angiography in cardiac imaging. Part III: radiation risk and dose reduction

Multislice CT angiography in cardiac imaging. Part III: radiation risk and dose reduction R e v i e w A r t i c l e Singapore Med J 2010; 51(5) : 374 Multislice CT angiography in cardiac imaging. Part III: radiation risk and dose reduction Sun Z, Ng K H Discipline of Medical Imaging, Department

More information

During the past several years, the need to reduce radiation

During the past several years, the need to reduce radiation Advances in Cardiovascular Imaging Achieving Very-Low-Dose Radiation Exposure in Cardiac Computed Tomography, Single-Photon Emission Computed Tomography, and Positron Emission Tomography Damini Dey, PhD;

More information

Sixty four slice Computed Tomography Scan (64-slice

Sixty four slice Computed Tomography Scan (64-slice Kathmandu University Medical Journal (2008), Vol. 6, No. 2, Issue 22, 257-261 Review Article 64-Slice CT Scan in Kathmandu Medical College Teaching Hospital Karki D 1, Neopane A 2, Regmi S 3, Acharya S

More information

Since the introduction of multidetector. Original Research

Since the introduction of multidetector. Original Research Original Research Hellenic J Cardiol 2014; 55: 184-190 Radiation Dose and Image Noise Evaluation in Coronary Computed Tomography Angiography (CCTA) Using an Iterative Reconstruction Algorithm Sofia D.

More information

Introduction. Original Article

Introduction. Original Article Original Article Superior CT coronary angiography image quality at lower radiation exposure with second generation 320-detector row CT in patients with elevated heart rate: a comparison with first generation

More information

Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography

Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography Int J Cardiovasc Imaging (2013) 29:453 461 DOI 10.1007/s10554-012-0096-3 ORIGINAL PAPER Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography

More information

MILITARY MEDICINE, Vol. 175, July MILITARY MEDICINE, 175, 7:529, 2010

MILITARY MEDICINE, Vol. 175, July MILITARY MEDICINE, 175, 7:529, 2010 MILITARY MEDICINE, 175, 7:529, 2010 Decreasing Outpatient Cardiac Catheterization Rates Associated With Cardiology Clinic Volume but Not With Increasing Cardiac Computed Tomography Utilization MAJ Eddie

More information

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses

The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Original Article Open Access The diagnostic evaluation of dual-source CT (DSCT) in the diagnosis of coronary artery stenoses Ziqiao Lei 1, Jin Gu 2, Qing Fu 3, Heshui Shi 4, Haibo Xu 5, Ping Han 6, Jianming

More information

MEDICAL POLICY SUBJECT: CORONARY CALCIUM SCORING

MEDICAL POLICY SUBJECT: CORONARY CALCIUM SCORING MEDICAL POLICY PAGE: 1 OF: 5 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical policy criteria are not applied.

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

Low-dose and High-resolution Cardiac Imaging with Revolution CT

Low-dose and High-resolution Cardiac Imaging with Revolution CT GE Healthcare Case study Low-dose and High-resolution Cardiac Imaging with Revolution CT Prof. Philipp A. Kaufmann, M.D. Ronny R. Buechel, M.D. Fran Mikulicic, M.D. Dominik C. Benz, M.D. University of

More information

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association MEDICAL POLICY SUBJECT: CARDIAC COMPUTED TOMOGRAPHIC PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical

More information

Keywords Coronary CT angiography Low dose Iterative reconstruction

Keywords Coronary CT angiography Low dose Iterative reconstruction European Heart Journal Cardiovascular Imaging (2014) 15, 1238 1245 doi:10.1093/ehjci/jeu113 Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction

More information

A Noninvasive Assessment of CAD

A Noninvasive Assessment of CAD : A Noninvasive Assessment of CAD In this article, Dr. Heilbron and Dr. Forster look at the noninvasive assessment of coronary artery disease (CAD), by means of coronary computed tomography angiography

More information

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Mathieu Sabbagh, R3 Michigan State University Radiology Garden City Hospital

More information

Contrast-Enhanced Computed Tomography Angiography (CTA) for Coronary Artery Evaluation

Contrast-Enhanced Computed Tomography Angiography (CTA) for Coronary Artery Evaluation Contrast-Enhanced Computed Tomography Angiography (CTA) for Coronary Artery Evaluation Applies to all products administered or underwritten by Blue Cross and Blue Shield of Louisiana and its subsidiary,

More information

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design 2 Cardiovascular Computed Tomography: Current and Future Scanning System Design Wm. Guy Weigold Introduction The heart can be visualized in gross form on any standard chest computed tomography (CT), but

More information

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography

Diagnostic accuracy of dual-source computed tomography in the detection of coronary chronic total occlusion: Comparison with invasive angiography African Journal of Biotechnology Vol. 10(19), pp. 3854-3858, 9 May, 2011 Available online at http://www.academicjournals.org/ajb DOI: 10.5897/AJB10.983 ISSN 1684 5315 2011 Academic Journals Full Length

More information

Impact of 64-Slice Multidetector Computed Tomography on Other Diagnostic Studies for Coronary Artery Disease

Impact of 64-Slice Multidetector Computed Tomography on Other Diagnostic Studies for Coronary Artery Disease CLINICAL RESEARCH STUDY Impact of 64-Slice Multidetector Computed Tomography on Other Diagnostic Studies for Coronary Artery Disease Alex J. Auseon, DO, Sunil S. Advani, MD, Charles A. Bush, MD, Subha

More information

What every radiologist should know about cardiac CT: A case-based pictorial review

What every radiologist should know about cardiac CT: A case-based pictorial review What every radiologist should know about cardiac CT: A case-based pictorial review Poster No.: C-0555 Congress: ECR 2010 Type: Educational Exhibit Topic: Cardiac Authors: C. M. Capuñay, P. Carrascosa,

More information

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial Poster No.: C-0595 Congress: ECR 2013 Type: Authors:

More information

Cardiac CT Angiography

Cardiac CT Angiography Cardiac CT Angiography Dr James Chafey, Radiologist Why do we need a better test for C.A.D? 1. CAD is the leading cause of death in the US CAD 31% Cancer 23% Stroke 7% 2. The prevalence of atherosclerosis

More information

Diagnostic value of 320-slice coronary CT angiography in coronary artery disease: A

Diagnostic value of 320-slice coronary CT angiography in coronary artery disease: A Diagnostic value of 320-slice coronary CT angiography in coronary artery disease: A systematic review and meta-analysis Zhonghua Sun PhD 1, Chenghsun Lin PhD 2 1. Discipline of Medical Imaging, Department

More information

Low-dose CT coronary angiography: Role of adaptive statistical iterative reconstruction (ASIR)

Low-dose CT coronary angiography: Role of adaptive statistical iterative reconstruction (ASIR) Low-dose CT coronary angiography: Role of adaptive statistical iterative reconstruction (ASIR) Poster No.: 456 Congress: ESCR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit M. A. Glazkova, I. Arkhipova,

More information

The Role of Nuclear Imaging in Heart Failure

The Role of Nuclear Imaging in Heart Failure The Role of Nuclear Imaging in Heart Failure Mouaz H. Al-Mallah, MD MSc FESC Associate Professor of Medicine Consultant Cardiologist and Division Head, Cardiac Imaging King Abdulaziz Cardiac Centre, National

More information

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography Supalerk Pattanaprichakul, MD 1, Sutipong Jongjirasiri, MD 2, Sukit Yamwong, MD 1, Jiraporn Laothammatas,

More information

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference.

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference. Rotation: Imaging 1 Imaging 1 provides COCATS Level 1 experience for nuclear cardiology (including SPECT and PET) and cardiac CT. Fellows will administer, process, and read cardiac nuclear studies with

More information

Coronary Artery Calcium Score

Coronary Artery Calcium Score Coronary Artery Calcium Score August 19, 2014 by Axel F. Sigurdsson MD 174 Comments essential for living organisms. Calcium is a chemical element that is Most of the calcium within the human body is found

More information

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT Eur Radiol (2006) 16: 1964 1972 DOI 10.1007/s00330-006-0262-x CARDIAC Sebastian Leschka Lars Husmann Lotus M. Desbiolles Oliver Gaemperli Tiziano Schepis Pascal Koepfli Thomas Boehm Borut Marincek Philipp

More information

CARDIAC IMAGING FOR SUBCLINICAL CAD

CARDIAC IMAGING FOR SUBCLINICAL CAD CARDIAC IMAGING FOR SUBCLINICAL CAD WHY DON'T YOU ADOPT MORE SMART TECHNIQUE? Whal Lee, M.D. Seoul National University Hospital Department of Radiology We are talking about Coronary artery Calcium scoring,

More information

MEDICAL POLICY. Proprietary Information of YourCare Health Plan

MEDICAL POLICY. Proprietary Information of YourCare Health Plan TOMOGRAPHIC ANGIOGRAPHY (CARDIAC CTA): CONTRAST- MEDICAL POLICY PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such

More information

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease

Horizon Scanning Technology Summary. Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease Horizon Scanning Technology Summary National Horizon Scanning Centre Magnetic resonance angiography (MRA) imaging for the detection of coronary artery disease April 2007 This technology summary is based

More information

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital Coronary Artery Imaging Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital Larger array : cover scan area Detector size : spatial resolution Rotation speed : scan time Retrospective

More information

b. To facilitate the management decision of a patient with an equivocal stress test.

b. To facilitate the management decision of a patient with an equivocal stress test. National Imaging Associates, Inc. Clinical guidelines EBCT HEART CT & HEART CT CONGENITAL CCTA CPT4 Codes: 75571 EBCT 75572, 75573 Heart CT & Heart CT Congenital 75574 - CCTA LCD ID Number: L33559 J K

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 5: , 2013

EXPERIMENTAL AND THERAPEUTIC MEDICINE 5: , 2013 636 Adaptive prospective ECG-triggered sequence coronary angiography in dual-source CT without heart rate control: Image quality and diagnostic performance CHANG-JIE PAN, NONG QIAN, TAO WANG, XIAO-QIANG

More information

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre Pushing the limits of cardiac CT Steven Dymarkowski Radiology / Medical Imaging Research Centre 5 X 2013 Introduction Rapid technological advances and new clinical applications in cardiovascular imaging

More information

Validation of CT Perfusion Imaging Against Invasive Angiography and FFR on a 320-MDCT Scanner

Validation of CT Perfusion Imaging Against Invasive Angiography and FFR on a 320-MDCT Scanner Validation of CT Perfusion Imaging Against Invasive Angiography and FFR on a 320-MDCT Scanner Zhen Qian, Gustavo Vasquez, Sarah Rinehart, Parag Joshi, Eric Krivitsky, Anna Kalynych, Dimitri Karmpaliotis,

More information

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease

Perspectives of new imaging techniques for patients with known or suspected coronary artery disease Perspectives of new imaging techniques for patients with known or suspected coronary artery disease Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands Correspondence: Jeroen

More information

Coronary CT angiography future directions

Coronary CT angiography future directions Perspective Coronary CT angiography future directions Stephan Achenbach Department of Cardiology, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany Correspondence to: Stephan Achenbach.

More information

Maximizing dose reductions with cardiac CT

Maximizing dose reductions with cardiac CT Int J Cardiovasc Imaging (2009) 25:279 287 DOI 10.1007/s10554-008-9405-2 ORIGINAL PAPER Maximizing dose reductions with cardiac CT Matthew J. Budoff Received: 16 October 2008 / Accepted: 10 December 2008

More information

Cardiac computed tomography: indications, applications, limitations, and training requirements

Cardiac computed tomography: indications, applications, limitations, and training requirements European Heart Journal (2008) 29, 531 556 doi:10.1093/eurheartj/ehm544 SPECIAL ARTICLE Cardiac computed tomography: indications, applications, limitations, and training requirements Report of a Writing

More information

Technology Assessment Institute: Summit on CT Dose Cardiac CT - Optimal Use of Evolving Scanner Technologies

Technology Assessment Institute: Summit on CT Dose Cardiac CT - Optimal Use of Evolving Scanner Technologies Cardiac CT - Optimal Use of Evolving Scanner Technologies P. Rogalla, M.D. Dept. of Medical Imaging University of Toronto Special thanks to Dr. Lembcke, Dr. Hein Charité, Berlin Disclosures No salaries

More information

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16.

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16. NIH Public Access Author Manuscript Published in final edited form as: Acad Radiol. 2008 January ; 15(1): 40 48. doi:10.1016/j.acra.2007.07.023. Analysis of Heart Rate and Heart Rate Variation During Cardiac

More information

Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients

Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients (2006) 30, 569 573 & 2006 Nature Publishing Group All rights reserved 0307-0565/06 $30.00 www.nature.com/ijo ORIGINAL ARTICLE Image quality and diagnostic accuracy of 16-slice multidetector computed tomography

More information

Multisclice CT in combination with functional imaging for CAD. Temporal Resolution. Spatial Resolution. Temporal resolution = ½ of the rotation time

Multisclice CT in combination with functional imaging for CAD. Temporal Resolution. Spatial Resolution. Temporal resolution = ½ of the rotation time Multisclice CT in combination with functional imaging for CAD Prof. Juhani Knuuti, MD, FESC Turku University Hospital and University of Turku Turku, Finland MSCT and functional imaging for CAD Practical

More information

Image quality and, hence, the diagnostic value of cardiac. Imaging

Image quality and, hence, the diagnostic value of cardiac. Imaging Imaging Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily Practice Impact of Different Scanning Protocols on Effective Dose Estimates Jörg Hausleiter, MD; Tanja Meyer, MD; Martin

More information

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association MEDICAL POLICY SUBJECT: CARDIAC/CORONARY COMPUTED TOMOGRAPHIC ANGIOGRAPHY PAGE: 1 OF: 6 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial

More information

Prognostic Value of Coronary Magnetic Resonance Angiography for Prediction of Cardiac Events in Patients With Suspected Coronary Artery Disease

Prognostic Value of Coronary Magnetic Resonance Angiography for Prediction of Cardiac Events in Patients With Suspected Coronary Artery Disease Journal of the American College of Cardiology Vol. 60, No. 22, 2012 2012 by the American College of Cardiology Foundation ISSN 0735-1097/$36.00 Published by Elsevier Inc. http://dx.doi.org/10.1016/j.jacc.2012.07.060

More information

Computed Tomography Imaging of the Coronary Arteries

Computed Tomography Imaging of the Coronary Arteries Chapter 6 Computed Tomography Imaging of the Coronary Arteries G.J. Pelgrim, M. Oudkerk and R. Vliegenthart Additional information is available at the end of the chapter http://dx.doi.org/10.5772/54044

More information

Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats

Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats Amolak Singh 1*, Yash Sethi 1, Sonya Watkins 1, Angela Youtsey 1, Angie Thomas 1 1. Department of Radiology, University

More information

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination Int J Clin Exp Med 2015;8(11):21898-21905 www.ijcem.com /ISSN:1940-5901/IJCEM0015005 Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out

More information

Cardiac Imaging Tests

Cardiac Imaging Tests Cardiac Imaging Tests http://www.medpagetoday.com/upload/2010/11/15/23347.jpg Standard imaging tests include echocardiography, chest x-ray, CT, MRI, and various radionuclide techniques. Standard CT and

More information

Kavitha Yaddanapudi Stony brook University New York

Kavitha Yaddanapudi Stony brook University New York Kavitha Yaddanapudi Stony brook University New York 8 million ER visits a year for chest pain 2-10% have Acute coronary syndrome (ACS) Coronary CTA(CCTA) -safe alternative to standard of care (SOC) and

More information

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications SYMPOSIA Indications, Patient Selection, and Clinical Implications Christian Thilo, MD,* Mark Auler, MD,* Peter Zwerner, MD,w Philip Costello, MD,* and U. Joseph Schoepf, MD* Abstract: Recent technical

More information

SPECT-CT: Τι πρέπει να γνωρίζει ο Καρδιολόγος

SPECT-CT: Τι πρέπει να γνωρίζει ο Καρδιολόγος SPECT-CT: Τι πρέπει να γνωρίζει ο Καρδιολόγος Δρ Αναστασία Κίτσιου Διευθύντρια, Καρδιολογική Κλινική, Σισμανόγλειο ΓΝΑ Chair, Education Committee, Section on Nuclear Cardiology & Cardiac CT, EACVI, ESC

More information

Coronary CT Angiography

Coronary CT Angiography Coronary CT Angiography Byoung Wook Choi, M.D. Department of Diagnostic Radiology Yonsei University College of Medicine, Severance Hospital E mail : bchoi@yumc.yonsei.ac.kr Abstract With the advent of

More information

The 2016 NASCI Keynote: Trends in Utilization of Cardiac Imaging: The Coronary CTA Conundrum. David C. Levin, M.D.

The 2016 NASCI Keynote: Trends in Utilization of Cardiac Imaging: The Coronary CTA Conundrum. David C. Levin, M.D. The 2016 NASCI Keynote: Trends in Utilization of Cardiac Imaging: The Coronary CTA Conundrum David C. Levin, M.D. October 16, 2016 MPI Utilization Rates/1000[includes PET] total radiologists 2014 total

More information

Highlights of the Second Annual Scientific Meeting of the Society of Cardiovascular Computed Tomography

Highlights of the Second Annual Scientific Meeting of the Society of Cardiovascular Computed Tomography Journal of the American College of Cardiology Vol. 50, No. 24, 2007 2007 by the American College of Cardiology Foundation ISSN 0735-1097/07/$32.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2007.08.046

More information

General Cardiovascular Magnetic Resonance Imaging

General Cardiovascular Magnetic Resonance Imaging 2 General Cardiovascular Magnetic Resonance Imaging 19 Peter G. Danias, Cardiovascular MRI: 150 Multiple-Choice Questions and Answers Humana Press 2008 20 Cardiovascular MRI: 150 Multiple-Choice Questions

More information