Improving Endograft Durability with EndoAnchors

Size: px
Start display at page:

Download "Improving Endograft Durability with EndoAnchors"

Transcription

1 Improving Endograft Durability with EndoAnchors William D. Jordan, Jr., M.D. John E. Skandalakis Chair in Surgery Professor and Chief Division of Vascular Surgery and Endovascular Therapy Emory University Atlanta, Georgia

2 DISCLOSURES CME standards NONE Clinical Investigator paid to Emory Gore, Medtronic, Endologix, Cook Consultant paid to Emory Medtronic Equity Shareholder None 2

3 PROBLEM PERSISTS DESPITE EVAR : DURABILITY Unibody endograft Aneurysm wall Aneurysm wall 3

4 LONG TERM OUTCOMES FOR AAA REPAIR: MEDICARRE PATIENTS NEJM ,966 matched patients; day mortality: 5.2% vs 1.6% 3 year survival better survival after open repair Conversion rate declines from 2.2% to 0.3% Rupture occurred 5.4 % after EVAR; 1.4% after open Reinterventions have improved 10.4% 9.1% 4

5 Legacy Studies Highlight Need for Lifelong Surveillance and Re-intervention after EVAR DREAM De Bruin et al. NEJM 2010 EVAR-1 Greenhalgh et al. NEJM 2010 ACE Becquemin et al. JVS 2011 Late ruptures in EVAR, none in open surgery With EVAR, predictors for rupture (endoleaks and migration) increase with time In ACE, 16% reinterventions in EVAR vs. 2.4% for open repair at 3 yr median f/u 5

6 Hostile Proximal Neck Predicts Challenges diameter, shape, length, thrombus, calcium Type I endoleaks 4.5x more likely at 1-year after EVAR in hostile proximal neck anatomy (P =.010) 4.5x 9x Aneurysm-related mortality risk 9x greater in hostile neck anatomy at 1-year (P=.013) Meta-Analysis of 7 major studies in EVAR by Antoniou et al 1 compared outcomes in hostile vs. friendly neck anatomies (total pts N = 1559) Study Sample Size Endografts Torsello et al, Endurant AbuRahma et al, AneuRx, Excluder, Zenith, Talent Hoshina et al, Excluder, Zenith Abbruzzese et al, AneuRx, Excluder, Zenith Choke et al, Talent, Zenith, Excluder, AneuRx Fulton et al, AneuRx Fairman et al, Talent 1 Antoniou GA et al. JVS. 2013;57(2):

7 Influence of Multiple Hostile Neck Parameters Neck Hostility Intra-op Adjunctive Procedures Intra-op Endoleaks All Cause Mortality On Label 9.9% 0.5% 1.1% 2 Hostile Neck Parameters Straight Tapered Reverse Angulate d d* Tapered >2 Hostile Neck Parameters Bulge Short + Speziale et al. shows greater proximal seal complication risks as the number of hostile neck parameters increases 26.7% 6.7% 13.3% 50% 16.7% 16.7% >1 Greater than 1 hostile neck parameter substantially increases: Mortality MAEs Endoleaks Adjunctive procedures Speziale et al, Annals VS

8 Displacement Force (Newtons) Recreating the Stability of Surgical Anastomosis EndoAnchor implants establish surgical anastomosis strength in EVAR & TEVAR Surgical Anastomosis EndoAnchor implant fixation Case images from John Aruny MD, Bart Edward Muhs, MD, PhD. No EndoAnchor implants Melas et al. JVS 2012;55(6):

9 Heli-FX System: Applier + Guide + 10 EndoAnchor Implants 3 mm Cross Bar 1.0 mm 3.5 mm 9

10 ANCHOR Registry: Capturing Real-World Usage Registry Design Registry Principal Investigators Treatment Arms* Enrollment & Duration Follow-up Prospective & Observational, International & Multi-Center, Dual-arm Registry with Core Lab Analysis Europe: Dr Jean-Paul de Vries Chief of Vascular Surgery, St. Antonius Hospital US: Dr William Jordan Chief of Vascular Surgery/Endovascular Therapy, Emory University School of Medicine Primary Revision Enrollment began 2012 and patients will be followed for 5 years Per Standard of Care at each center & discretion of Investigator *Expanded registry to include Thoracic and Advanced Disease arms. ~ 780 Patients Enrolled 10

11 ANCHOR Registry* Primary Arm represents 72.4% of pts Stent Grafts - Primary Arm 604 ANCHOR REGISTRY 604 Subjects Medtronic Endurant Gore Excluder Cook Zenith Jotec Other *Data cut June 15, PRIMARY ARM 167 REVISION ARM Stent Grafts - Revision Arm Metronic Endurant Medtronic Talent Medtronic AneuRx Gore Excluder Cook Zenith Jotec Other 11

12 ANCHOR Registry Primary Arm EndoAnchor implants without evidence of type 1a endoleak, but concern for late failure and/or prevention of neck dilatation ANCHOR REGISTRY PRIMARY ARM N=437 REVISION ARM N=167 PROPHYLACTIC USE (72%, N=314) INTRA-OP TYPE IA ENDOLEAKS (28%, N=123) 1 Data cut June 15,

13 72 YO, 6.3 CM AAA, CREAT cm 13

14 ANCHOR Registry Prophylactic Use (N=314) Indications for EndoAnchor Implants Hostile Necks: 91.2% Per the SVS definition Per treating physician, patients at risk of late failure Short, Wide, Angulated, Conical, Young (SWACY) Infrarenal Diameter: 25.6 mm Infrarenal Angulation: 24.6 Conical Neck: (>10%/10mm): 41.4% Neck Length: 11.5 mm (median) Aneurysm Diameter: 55.7 mm Avg Neck Calcium Thickness: 1.2 mm Male: 78% Female: 22% Mean Age: 72.4 Years * Mean Core Lab measurements based on 251 pts with baseline CTs 14

15 ANCHOR Registry Prophylactic Use Technical Success Successful deployment of EndoAnchor implants with adequate penetration into aortic wall Procedural Success Technical success without type Ia endoleak at completion arteriography 94.9% Prophylactic 94.6% Prophylactic Avg. duration Avg. time to Avg. number of EndoAnchor implants of Procedure (min) EndoAnchor implants (min) At 12 months N=182 At 24 months N=85 Core Lab 12 months 24 months Type 1a Endoleak 0.6% (1/181) 0.0% (0/86) Endograft Migration 0.0% (0/129) 0.0% (0/43) Decre ase 45. Incre ase 1.6% Stabl e 52. Decre ase 61.2% Increa se 1.2% Stable 37.6% 15

16 ANCHOR Registry Prophylactic Use Hostile Necks: 91.2% Per the SVS definition Kaplan-Meier Estimates 1 Year (N=301) 2 Year (N=214) Freedom from ACM 94.5% 90.6% Freedom from ARM 98.4% 98.4% Freedom from 2 nd Procedures 95.9% 92.1% Adverse Events through 2 Years Patients with Events Procedure-Related SAE % EndoAnchor implant-related SAE 0 0.0% Aneurysm Rupture 0 0.0% AAA-Related Mortality 5 1.6% Open Surgical Conversion 1 0.3% 16

17 In the absence of randomized prospective clinical trial (RCT), how can we really compare the results? Do EndoAnchors Have Value in Preventing Proximal Neck Complications in Patients Undergoing EVAR? 17

18 METHODOLOGY Pre-EVAR baseline CT scans evaluated by a Core Laboratory for both groups 19 baseline variables entered into a propensity matching algorithm (SPSS v22; binary logistic regression with group as the independent variable) Match: 198 patients matched (99 patients in EndoAnchor; 99 in Control) Well-matched by the 19 baseline variables Analysis: Primary endpoints indicative of proximal neck failure Including Type 1a endoleak, sac enlargement, sac regression, endograft migration and neck dilation Mean follow-up Control (26.2 months; range months) EndoAnchor (13.7 months; range months) 18

19 BASELINE ANATOMY IN PROPENSITY-MATCHED COHORTS Anatomic Measures for Propensity Matching Controls N =91 EndoAnchor implants N = 87 P Value Max AAA Diameter 56 ± 13 mm 56 ± 10 mm.674 Suprarenal Diameter 27 ± 4 mm 27 ± 3 mm.999 Diameter at Lowest Renal 25 ± 4 mm 26 ± 4 mm.458 Proximal Neck Length 23 ± 14 mm 20 ± 13 mm.093 Suprarenal Angulation 16 ± ± Infrarenal Angulation 37 ± ± Neck Thrombus 23± ± Neck Calcium 20± ± Necks <10mm Length 18.4% 26.5%

20 NECK DILATION EndoAnchor FF 98.4% ±1.6% Improving Durability with EndoAnchors 1-YEAR RESULTS TYPE 1A ENDOLEAK EndoAnchor FF 97.0% ±2.1% Control FF 94.9% ±2.5% Control FF 94.1% ±2.5% P-value =.725 P-value =.350 Observations: there was a trend in reduction of Neck Dilation 3.0% Risk at 2 years in the EndoAnchor Group 5.9% risk in the control group 20

21 Transmural Fixation of EndoAnchor TM Implants Has Unique Benefits Propensity matched comparison of those treated without EndoAnchors Promotes increased rate of sac regression: In a propensity-matched study design, significantly greater AAA sac regression with EndoAnchor fixation at 2 years post-evar (81.1% vs. 48.7% without EndoAnchor fixation, p=.01) 1 Methodology Pre-EVAR CTs core lab evaluated Propensity matching on 19 variables 2 cohorts: 99pts EVAR 99pts EVAR+EndoAnchor P-value = 0.01 EndoAnchor +EVAR 81.1% ± 9.5% 1. Muhs, BE et al. J Vasc Surg. 2017, Article in press EVAR 48.7% ± 5.9% 21

22 SUMMARY OF RESULTS FROM PROPENSITY ANALYSIS Sac regression better the EndoAnchor group compared to anatomically match control cohorts 28.6% in EndoAnchor implant group vs. 20.3% in the non- EndoAnchor implant group (p=.017) Type 1A endoleak better in EndoAnchor group 3.0% vs 5.9% Early results not statistically significant Less proximal neck dilation in EndoAnchor group 2.6% in the EndoAnchor test group vs. 5.1% in the control group Early results not statistically significant 22 Do not distribute, for internal educational purposes only

23 July 2017 What are predictors for Aortic Neck Dilatation? Is Aortic Neck Dilatation impacted by EndoAnchors? 23 Do not distribute, for internal educational purposes only

24 METHODS 267 patients prospectively enrolled in ANCHOR Indication for EndoAnchor Use Repair of intraoperative Type Ia endoleak Prophylactic in patients with hostile neck anatomy Aortic diameter measured by independent core lab at: Suprarenal level (20mm proximal to lowest main renal artery) Three (3) levels within proximal neck (distal to lowest main renal artery) 0mm 5mm 10mm 24 Do not distribute, for internal educational purposes only

25 METHODS Baseline of aortic neck diameter measured on preoperative CT angiogram Same level measured at 1-month and 12-month postoperative CT scan Dilatation expressed as the difference between 1-month and 12- month post-evar Multivariable analyses (backward stepwise linear regression) performed 25

26 RESULTS NECK DILATATION Aortic Level Predictors of Dilatation at Specified Level Coefficient* Effect P Value Lowest renal Endograft type 0.62 (0.12, 1.04) *.006 Aortic aneurysm sac diameter (-0.07, 0.00) Protective.020 Aortic diameter at lowest renal 0.16 (0.08, 0.24) Risk factor <.001 Aortic neck length (-.04, 0.00) Protective.021 Infrarenal Angulation 0.02 (0.00, 0.04) Risk factor.016 Endograft oversizing 5.37 (2.34, 8.39) Risk factor.001 5mm distally Aortic diameter at lowest renal 0.17 (0.07, 0.26) Risk factor.001 Endograft oversizing 6.00 (2.68, 9.31) Risk factor mm distally Aortic diameter at lowest renal 0.17 (0.04, 0.29) Risk factor.003 Endograft oversizing 4.86 (0.13, 9.58) Risk factor.032 Number of EndoAnchors placed (-0.55, -0.04) Protective.037 Suprarenal level Suprarenal aortic diameter 0.08 (0.01, 0.16) Risk factor

27 SUMMARY OF NECK DILATATION ASSESSMENT Aggressive endograft oversizing and larger baseline aortic neck diameter at the lowest renal artery appear to be independent predictors of Early Proximal Aortic Neck Dilatation (1-month to 12-months to post EVAR) EndoAnchors appear to offer a PROTECTIVE EFFECT towards aortic neck dilatation Long-term follow-up studies necessary for more conclusive results 27

28 Can EndoAnchors Improve Endograft Durability? Consider the Long Term Outlook... Select Subset of Endovascular Patients Existing Seal Complications Highly Challenging Anatomies ANCHOR Registry A B C Improves Durability Acute & late type I endoleaks 2 Type I endoleaks in urgent or ruptured EVAR Augmenting stability in migrated grafts 1 Irregularly shaped necks (short, wide, highly angulated, conical) 1 Difficult landing zones 2 Few re-interventions Better sac regression Less neck dilatation Consider for younger patients with long remaining life expectancy 3 Primary/Secondary Primary Primary 1 Jordan et al. J Vasc Surg. 2015:61(6): de Vries. JEVT. 2013;20(4): Schanzer et al. Circulation 2011;123:

29 THANK YOU

Use of Aptus Heli-FX EndoAnchor implants with standard endografts to strengthen seal in hostile anatomies:

Use of Aptus Heli-FX EndoAnchor implants with standard endografts to strengthen seal in hostile anatomies: Use of Aptus Heli-FX EndoAnchor implants with standard endografts to strengthen seal in hostile anatomies: technical tips, case experience and clinical results CACVS Medtronic Symposium Paris, France January

More information

Mid-term results from ANCHOR: How does this data influence the treatment algorithm for hostile EVAR anatomies

Mid-term results from ANCHOR: How does this data influence the treatment algorithm for hostile EVAR anatomies Mid-term results from ANCHOR: How does this data influence the treatment algorithm for hostile EVAR anatomies Jean-Paul P.M. de Vries Head Department of Surgery, University Medical Centre Groningen The

More information

Hostile Neck During EVAR, The Role Of Endoanchores

Hostile Neck During EVAR, The Role Of Endoanchores Hostile Neck During EVAR, The Role Of Endoanchores Samer Koussayer, MD, FACS, RVT Prof, Al Faisal University Section Head and consultant Vascular & Endovascular Surgery Division King Faisal Specialist

More information

EVAR Revision Setting - How can Heli-FX EndoAnchors improve the outcomes?

EVAR Revision Setting - How can Heli-FX EndoAnchors improve the outcomes? D. Böckler Department of Vascular and Endovascular Surgery University Hospital Heidelberg EVAR Revision Setting - How can Heli-FX EndoAnchors improve the outcomes? Disclosures Speaker name: Dittmar Böckler

More information

Type 1a Endoleak in hostile neck anatomies: Endoanchor can fix it! D. Böckler University Hospital Heidelberg, Germany

Type 1a Endoleak in hostile neck anatomies: Endoanchor can fix it! D. Böckler University Hospital Heidelberg, Germany Type 1a Endoleak in hostile neck anatomies: Endoanchor can fix it! D. Böckler University Hospital Heidelberg, Germany Disclosures Speaker name: Dittmar Böckler I have the following potential conflicts

More information

How to select FEVAR versus EVAR + endoanchors in short-necked AAAs

How to select FEVAR versus EVAR + endoanchors in short-necked AAAs How to select FEVAR versus EVAR + endoanchors in short-necked AAAs Jean-Paul P.M. de Vries, Richte C.L. Schuurmann St. Antonius Hospital Nieuwegein, The Netherlands 21st Critical Issues Congress Nuernberg,

More information

Technique and Tips for Complicated AAA Cases with Stent Graft

Technique and Tips for Complicated AAA Cases with Stent Graft Technique and Tips for Complicated AAA Cases with Stent Graft Seung-Woon Rha, MD, PhD FACC, FAHA, FESC, FSCAI, FAPSIC Cardiovascular Center, Korea University Guro Hospital Mar 15, 2018 LINC AP 2018 Endoleak;

More information

Ralf R. Kolvenbach. Verbund Katholischer Kliniken Gefäßzentrum Augusta Krankenhaus Düsseldorf

Ralf R. Kolvenbach. Verbund Katholischer Kliniken Gefäßzentrum Augusta Krankenhaus Düsseldorf Ralf R. Kolvenbach Verbund Katholischer Kliniken Gefäßzentrum Augusta Krankenhaus Düsseldorf Endograft migration Stent no longer above renal artery 15 mm migration 6 Mos Do we really need any adjuncts

More information

When to use standard EVAR with EndoAnchors or CHEVAR in short-neck AAAs LINC ASIA 18

When to use standard EVAR with EndoAnchors or CHEVAR in short-neck AAAs LINC ASIA 18 When to use standard EVAR with EndoAnchors or CHEVAR in short-neck AAAs JEAN-PAUL P.M. DE VRIES, DIRECTOR OF VASCULAR SURGERY ST.ANTONIUS HOSPITAL NIEUWEGEIN, THE NETHERLANDS. LINC ASIA-PACIFIC HongKong,

More information

Patient selection in Hostile Necks and how. to prevent endoleaks a word of caution

Patient selection in Hostile Necks and how. to prevent endoleaks a word of caution Patient selection in Hostile Necks and how to prevent endoleaks a word of caution P M Kasprzak, K. Pfister Department of Vascular Surgery Endovascular Surgery University Hospital Regensburg, Germany Conflict

More information

Aortic Neck Issues Associated Clinical Sequelae/Implications for Graft Choice

Aortic Neck Issues Associated Clinical Sequelae/Implications for Graft Choice Aortic Neck Issues Associated Clinical Sequelae/Implications for Graft Choice Eric Verhoeven, MD, PhD, A. Katsargyris, MD Department of Vascular and Endovascular Surgery, Paracelsus Medical University,

More information

How to Categorize the Infrarenal Neck Properly? I Van Herzeele Dept. Thoracic and Vascular Surgery, Ghent University, Belgium

How to Categorize the Infrarenal Neck Properly? I Van Herzeele Dept. Thoracic and Vascular Surgery, Ghent University, Belgium How to Categorize the Infrarenal Neck Properly? I Van Herzeele Dept. Thoracic and Vascular Surgery, Ghent University, Belgium Disclosure Speaker name: Isabelle Van Herzeele I have the following potential

More information

Nellix Endovascular System: Clinical Outcomes and Device Overview

Nellix Endovascular System: Clinical Outcomes and Device Overview Nellix Endovascular System: Clinical Outcomes and Device Overview Jeffrey P. Carpenter, MD Professor and Chief, Department of Surgery CAUTION: Investigational device. This product is under clinical investigation

More information

Considerations for a Durable Repair

Considerations for a Durable Repair Considerations for a Durable Repair Eric Verhoeven, MD, PhD, A. Katsargyris, MD Department of Vascular and Endovascular Surgery, Paracelsus Medical University, Nuremberg, Germany Disclosures William Cook

More information

Now that Endoanchors are Approved (and paid for) We have a Durable Solution to Short Necks That s so Easy!

Now that Endoanchors are Approved (and paid for) We have a Durable Solution to Short Necks That s so Easy! Now that Endoanchors are Approved (and paid for) We have a Durable Solution to Short Necks That s so Easy! David H. Deaton, MD Vascular Surgery / University of South Carolina / Palmetto Health Chief Medical

More information

Abdominal and thoracic aneurysm repair

Abdominal and thoracic aneurysm repair Abdominal and thoracic aneurysm repair William A. Gray MD Director, Endovascular Intervention Cardiovascular Research Foundation Columbia University Medical Center Abdominal Aortic Aneurysm Endografts

More information

Durability of The Endurant Stent-Graft through 5 Years

Durability of The Endurant Stent-Graft through 5 Years Durability of The Endurant Stent-Graft through 5 Years Michel S. Makaroun MD Co-Director, UPMC Heart and Vascular Institute Professor and Chair, Division of Vascular Surgery University of Pittsburgh School

More information

Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC

Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC Interventional Cardiologist/Endovascular Specialist Bradenton Cardiology Center Bradenton, FL, USA Disclosures

More information

Current Status of EVAR for Infrarenal AAA. 31 st Annual Florida Vascular Society. PENN Surgery

Current Status of EVAR for Infrarenal AAA. 31 st Annual Florida Vascular Society. PENN Surgery Current Status of EVAR for Infrarenal AAA 31 st Annual Florida Vascular Society PENN Surgery No Disclosures Stent Grafts Design Related Differences What really matters? Modular Unibody Supported Unsupported

More information

Standardization of the CHEVAR procedure: How a standard approach has improved outcomes. Prof Peter Holt St George s, London

Standardization of the CHEVAR procedure: How a standard approach has improved outcomes. Prof Peter Holt St George s, London Standardization of the CHEVAR procedure: How a standard approach has improved outcomes Prof Peter Holt St George s, London Disclosure Speaker name: Prof Peter Holt I have the following potential conflicts

More information

ENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients

ENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients ENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients The Ovation System is approved to treat infrarenal abdominal aortic aneurysms and is not

More information

From 1996 to 1999, a total of 1,193 patients with

From 1996 to 1999, a total of 1,193 patients with THE ANEURX CLINICAL TRIAL AT 8 YEARS Lessons learned following the US AneuRx clinical trial from 1996 to 2004. BY CHRISTOPHER K. ZARINS, MD From 1996 to 1999, a total of 1,193 patients with infrarenal

More information

Anatomical challenges in EVAR

Anatomical challenges in EVAR Anatomical challenges in EVAR M.H. EL DESSOKI, MD,FRCS PROFESSOR OF VASCULAR SURGERY CAIRO UNIVERSITY Disclosure Speaker name:... I have the following potential conflicts of interest to report: Consulting

More information

Ovation. Sean Lyden, MD Department Chair, Vascular Surgery Cleveland Clinic

Ovation. Sean Lyden, MD Department Chair, Vascular Surgery Cleveland Clinic Ovation Sean Lyden, MD Department Chair, Vascular Surgery Cleveland Clinic Disclosure Statement of Financial Interest Within the past 12 months, I or my spouse/partner have had a financial interest/arrangement

More information

Predictors of abdominal aortic aneurysm sac enlargement after EVAR Longterm results from the ENGAGE Registry

Predictors of abdominal aortic aneurysm sac enlargement after EVAR Longterm results from the ENGAGE Registry Predictors of abdominal aortic aneurysm sac enlargement after EVAR Longterm results from the ENGAGE Registry D. Böckler Department of Vascular and Endovascular Surgery University Hospital Heidelberg Disclosure

More information

Hostile Proximal Neck: A New Conformable EVAR Device

Hostile Proximal Neck: A New Conformable EVAR Device Hostile Proximal Neck: A New Conformable EVAR Device Young-Guk Ko, M.D., Seoul, Korea Currently Available Devices for EVAR in Korea, 2018 Zenith Flex, Cook Endurant IIs, Medtronic INCRAFT, Cordis AFX2,

More information

TriVascular Ovation Prime Abdominal Stent Graft System

TriVascular Ovation Prime Abdominal Stent Graft System TriVascular Ovation Prime Abdominal Stent Graft System Science of the Seal O-Ring Sealing Technology O-Ring Sealing in Proven Engineering Solutions O-rings are designed to seal by blocking the flow of

More information

Endovascular Repair o Abdominal. Aortic Aneurysms. Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida

Endovascular Repair o Abdominal. Aortic Aneurysms. Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida Endovascular Repair o Abdominal Aortic Aneurysms Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida Disclosure Nothing to disclose. 2 Mr. X AAA Mr. X. Is a 70 year old male who presented to

More information

Analysis of Type IIIb Endoleaks Encountered with Endologix Endografts

Analysis of Type IIIb Endoleaks Encountered with Endologix Endografts Analysis of Type IIIb Endoleaks Encountered with Endologix Endografts Alan R. Wladis, MD, FACS, David Varnagy, MD, FACS, Manuel R. Perez-Izquierdo, MD, Mark Ranson, MD FACS, Delos Clift, MD FACS, Rebecca

More information

Endoanchor-assisted TEVAR

Endoanchor-assisted TEVAR Endoanchor-assisted TEVAR May 29, 2015 NCVH2015 Grayson H. Wheatley III, MD Director of Aortic and Endovascular Surgery Associate Professor of Surgery Temple University School of Medicine Disclosures Consultant

More information

Young-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System,

Young-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System, Young-Guk Ko, M.D., Dangas G, J Am Coll Cardiol Intv 2012;5:1071 All-cause Mortality Dangas Severance G, J Am Coll Cardiovascular Cardiol Intv Hospital, 2012;5:1071 Yonsei University Health System Aneurysm-related

More information

EVAS using Nellix in my practice Where are we today?

EVAS using Nellix in my practice Where are we today? EVAS using Nellix in my practice Where are we today? Prof. dr. Michel MPJ Reijnen Department of Vascular Surgery, Rijnstate Hospital Arnhem Faculty of Science and Technology, University of Twente The Netherlands

More information

LOWERING THE PROFILE RAISING THE BAR

LOWERING THE PROFILE RAISING THE BAR LOWERING THE PROFILE RAISING THE BAR INNOVATIVE LOW PROFILE. ADVANCED CLINICAL PERFORMANCE. The AFX TM Endovascular AAA System integrates anatomical fixation with an advanced delivery system and graft

More information

EndoAnchors: Endovascular Stitching During EVAR and TEVAR

EndoAnchors: Endovascular Stitching During EVAR and TEVAR EndoAnchors: Endovascular Stitching During EVAR and TEVAR BY JEAN-PAUL P. M. de VRIES, MD, PhD THE EVOLUTION OF ENDOVAS- CULAR REPAIR IS CHARACTER- IZED BY INNOVATION Endovascular aneurysm repair (EVAR)

More information

14F OD Ovation Abdominal Stent Graft System

14F OD Ovation Abdominal Stent Graft System 14F OD Ovation Abdominal Stent Graft System 2 Year Results from the European Study Dierk Scheinert, MD Center for Vascular Medicine Angiology and Vascular Surgery Park Hospital Leipzig, Germany Disclosure

More information

Degeneration of the Neck Post Implementation - a New Era of AAA Stent

Degeneration of the Neck Post Implementation - a New Era of AAA Stent Degeneration of the Neck Post Implementation - a New Era of AAA Stent New Mexico Heart Institute Albuquerque, New Mexico USA - Gore Current FDA-Approved EVAR Devices Sealing Mechanism Endologix Lombard

More information

EVAS How does this Impact EVAR Therapy

EVAS How does this Impact EVAR Therapy EVAS How does this Impact EVAR Therapy John S. Lane III, MD, FACS Professor and Acting Chief of Vascular Surgery University of California, San Diego PNEC, 2017 Endologix products and associated components

More information

Conflicts of Interest. When and Why Complex EVAR in Tx of juxta/suprarenal AAA? Summary. Infrarenal EVAR for short necks 2y postop

Conflicts of Interest. When and Why Complex EVAR in Tx of juxta/suprarenal AAA? Summary. Infrarenal EVAR for short necks 2y postop When and Why Complex EVAR in Tx of juxta/suprarenal AAA? Tim Resch MD Vascular Center Skane University Hospital Conflicts of Interest COOK Medical - Consulting, Speakers Bureau, IP, Research support Medtronic

More information

Mid-term results of 300+ patients treated by endovascular aortic sealing (EVAS)

Mid-term results of 300+ patients treated by endovascular aortic sealing (EVAS) Mid-term results of 300+ patients treated by endovascular aortic sealing (EVAS) Jean-Paul P.M. de Vries Dept Vascular Surgery St. Antonius Hospital, Nieuwegein,The Netherlands On behalf of the DEVASS study

More information

Insights from the PROTAGORAS/PERICLES Registries: impact on ChEVAR results

Insights from the PROTAGORAS/PERICLES Registries: impact on ChEVAR results Insights from the PROTAGORAS/PERICLES Registries: impact on ChEVAR results Konstantinos P. Donas, Giovanni Torsello Department of Vascular Surgery St. Franziskus Hospital Münster, Germany Disclosure Speaker

More information

Challenges with Complex Anatomies Advancing Care in Endovascular Aortic Treatment

Challenges with Complex Anatomies Advancing Care in Endovascular Aortic Treatment Challenges with Complex Anatomies Advancing Care in Endovascular Aortic Treatment Robert Y. Rhee, MD Chief, Vascular and Endovascular Surgery Director, Aortic Center Maimonides Medical Center Brooklyn,

More information

The Auckland Experience with the Nellix EVAS System. Andrew Holden, MBChB, FRANZCR

The Auckland Experience with the Nellix EVAS System. Andrew Holden, MBChB, FRANZCR The Auckland Experience with the Nellix EVAS System Andrew Holden, MBChB, FRANZCR Disclosure Speaker name: Associate Professor Andrew Holden I have the following potential conflicts of interest to report:

More information

Disclosures. EVAR follow-up: actual recommendation. EVAR follow-up: critical issues

Disclosures. EVAR follow-up: actual recommendation. EVAR follow-up: critical issues Disclosures is it time to discuss individualized follow-up schemes based on preoperative anatomy and high quality completion angiography? Consultant / Speaker / Proctor Cook Cordis Medtronic Invatec W.L.

More information

The Ventana Off-the-Shelf Graft for Pararenal AAA. Andrew Holden Associate Professor of Radiology Auckland Hospital

The Ventana Off-the-Shelf Graft for Pararenal AAA. Andrew Holden Associate Professor of Radiology Auckland Hospital The Ventana Off-the-Shelf Graft for Pararenal AAA Andrew Holden Associate Professor of Radiology Auckland Hospital Disclosures Andrew Holden, MBChB, FRANZCR Investigator in Nellix and Ventana Trials Clinical

More information

A New EVAR Device for Infrarenal AAAs

A New EVAR Device for Infrarenal AAAs A New EVAR Device for Infrarenal AAAs Peter Nelson, MD, MS Assistant Professor of Surgery MM0203 Rev. 01 Current U.S. EVAR Devices Anatomical Fixation Proximal Fixation Powerlink - Endologix Excluder WL

More information

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ History of the Powerlink System Design and Clinical Results Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ Powerlink System: Unibody-Bifurcated Design Long Main Body Low-Porosity Proprietary eptfe

More information

My personal experience with INCRAFT in standard and challenging cases

My personal experience with INCRAFT in standard and challenging cases My personal experience with INCRAFT in standard and challenging cases G Pratesi, MD Vascular Surgery University of Rome Tor Vergata giovanni.pratesi@uniroma2.it Disclosure Speaker name: Giovanni Pratesi,

More information

Lessons learned from Ch-EVAR for the treatment of. Miltos Matsagkas MD, PhD, FEBVS Professor of Vascular Surgery University of Thessaly

Lessons learned from Ch-EVAR for the treatment of. Miltos Matsagkas MD, PhD, FEBVS Professor of Vascular Surgery University of Thessaly Lessons learned from Ch-EVAR for the treatment of pararenal AAAs Miltos Matsagkas MD, PhD, FEBVS Professor of Vascular Surgery University of Thessaly Ch-EVAR Ch-EVAR = Chimney-EVAR Placement of single

More information

CASE BOOK: ENDOANCHOR TM FIXATION UC EN

CASE BOOK: ENDOANCHOR TM FIXATION UC EN CASE BOOK: ENDOANCHOR TM FIXATION UC201700368 EN PURPOSE OF THIS MATERIAL The Aptus TM Heli-FX TM EndoAnchor TM systems have been used in over 5,000 patients worldwide, with over 30,000 EndoAnchor implants

More information

No Neck Is Needed! Treat the Aneurysm Instead! Andrew Holden, MD Associate Professor of Radiology Auckland City Hospital

No Neck Is Needed! Treat the Aneurysm Instead! Andrew Holden, MD Associate Professor of Radiology Auckland City Hospital No Neck Is Needed! Treat the Aneurysm Instead! Andrew Holden, MD Associate Professor of Radiology Auckland City Hospital Disclosures Andrew Holden, MBChB, FRANZCR No financial investment to disclose Investigator

More information

EndoVascular Aneurysm Sealing (EVAS) with Nellix

EndoVascular Aneurysm Sealing (EVAS) with Nellix 1 2 EndoVascular Aneurysm Sealing (EVAS) with Nellix Designed to seal entire aneurysm with contained biostable polymer Non-modular design with complete fixation Expands endovascular patient eligibility

More information

EVAS is Associated with Lower All-Cause Mortality

EVAS is Associated with Lower All-Cause Mortality EVAS is Associated with Lower All-Cause Mortality Marc L Schermerhorn, MD Chief, Division of Vascular and Endovascular Surgery Beth Israel Deaconess Medical Center Professor of Surgery Harvard Medical

More information

AAA: DEBATE THERE ARE NO LIMITS USING EVAR FOR AAA. 2 nd -3 rd June 2016.

AAA: DEBATE THERE ARE NO LIMITS USING EVAR FOR AAA. 2 nd -3 rd June 2016. AAA: DEBATE THERE ARE NO LIMITS USING EVAR FOR AAA JORGE FERNÁNDEZ NOYA ANGIOLOGY AND VASCULAR SURGERY DEPARTMENT UNIVERSITARY CLINICAL HOSPITAL SANTIAGO DE COMPOSTELA 2 nd -3 rd June 2016. 1888 ENDOANEURYSMORRHAPHY

More information

GORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability. Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands

GORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability. Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands GORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands Disclosure Speaker name: Michel Reijnen I have the following potential conflicts

More information

EVAR replaced standard repair in most cases. Why?

EVAR replaced standard repair in most cases. Why? EVAR replaced standard repair in most cases. Why? Initial major steps in endograft evolution Papazoglou O. Konstantinos M.D. The story of a major breakthrough in vascular surgery 1991 Parodi introduces

More information

Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type 1b Endoleak following an EVAR for Ruptured AAA

Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type 1b Endoleak following an EVAR for Ruptured AAA Hindawi Case Reports in Vascular Medicine Volume 2017, Article ID 4931282, 5 pages https://doi.org/10.1155/2017/4931282 Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type

More information

Talent Abdominal Stent Graft

Talent Abdominal Stent Graft Talent Abdominal with THE Xcelerant Hydro Delivery System Expanding the Indications for EVAR Treat More Patients Short Necks The Talent Abdominal is the only FDA-approved device for proximal aortic neck

More information

Management of Endoleaks

Management of Endoleaks Management of Endoleaks Murray Shames, MD Professor and Chief, Director Tampa General Hospital Aortic Program Vice Chair of Research, Dept. of Surgery Conflict of Interests: Speaker: Gore, Medtronic, Cook

More information

Optimizing Accuracy of Aortic Stent Grafts in Short Necks

Optimizing Accuracy of Aortic Stent Grafts in Short Necks Optimizing Accuracy of Aortic Stent Grafts in Short Necks Venkatesh Ramaiah, MD, FACS Medical Director Arizona Heart Hospital Director Peripheral Vascular and Endovascular Research Arizona Heart Institute

More information

Jean M. Panneton, MD, FRCSC, FACS. Professor of Surgery, Chief & Program Director Division of Vascular Surgery. Norfolk, VA

Jean M. Panneton, MD, FRCSC, FACS. Professor of Surgery, Chief & Program Director Division of Vascular Surgery. Norfolk, VA Jean M. Panneton, MD, FRCSC, FACS Professor of Surgery, Chief & Program Director Division of Vascular Surgery Eastern Virginia Medical School Norfolk, VA The Houston Aortic Symposium Houston TX, February

More information

What does the data tell us about outcomes of EVAR in challenging anatomy?

What does the data tell us about outcomes of EVAR in challenging anatomy? What does the data tell us about outcomes of EVAR in challenging anatomy? UCSF Vascular Surgery Symposium 2018 Sukgu M Han, MD, MS Assistant Professor of Clinical Surgery Co-director, Comprehensive Aortic

More information

Influence of patient selection and IFU compliance on outcomes following EVAS

Influence of patient selection and IFU compliance on outcomes following EVAS Influence of patient selection and IFU compliance on outcomes following EVAS LUNCH SYMPOSIUM LINC 2017 Jan MM Heyligers, MD, PhD, FEBVS Consultant Vascular Surgeon Elisabeth TweeSteden Hospital Tilburg,

More information

THE ENDURANT STENT GRAFT IN HOSTILE ANEURYSM NECK ANATOMY

THE ENDURANT STENT GRAFT IN HOSTILE ANEURYSM NECK ANATOMY THE ENDURANT STENT GRAFT IN HOSTILE ANEURYSM NECK ANATOMY Patrice Mwipatayi FCS (SA), MMed, FRACS Professor of Vascular surgery Royal Perth Hospital, University of Western Australia, Perth, WA Co-Authors:

More information

Access More Patients. Customize Each Seal.

Access More Patients. Customize Each Seal. Access More. Customize Each Seal. The Least Invasive Path Towards Proven Patency ULTRA LOW PROFILE TO EASE ADVANCEMENT The flexible, ultra-low 12F ID Ovation ix delivery system enables you to navigate

More information

Bilateral use of the Gore IBE device for bilateral CIA aneurysms and a first interim analysis of the prospective Iceberg registry

Bilateral use of the Gore IBE device for bilateral CIA aneurysms and a first interim analysis of the prospective Iceberg registry Bilateral use of the Gore IBE device for bilateral CIA aneurysms and a first interim analysis of the prospective Iceberg registry Michel MPJ Reijnen, MD, PhD Department of Vascular Surgery, Rijnstate Hospital

More information

NEW INNOVATIONS IN ENDOLEAK MANAGEMENT

NEW INNOVATIONS IN ENDOLEAK MANAGEMENT DISCLOSURES NEW INNOVATIONS IN ENDOLEAK MANAGEMENT Kelley HodgkissHarlow, MD, RPVI, FACS Vascular and Endovascular Surgery Kaiser Permanente, San Diego Nothing to disclose GOALS BACKGROUND Background Surveillance

More information

EndoAnchors Advance Treatment of Aortic Aneurysms: Experience in a Community Hospital

EndoAnchors Advance Treatment of Aortic Aneurysms: Experience in a Community Hospital EndoAnchors Advance Treatment of Aortic Aneurysms: Experience in a Community Hospital BY MAZIN FOTEH, MD CONFRONTING THE COMPLEX ANATOMY OF AAAs AND THE LIMITATIONS OF EVAR Even though endovascular aortic

More information

NASDAQ: ELGX December Innovation that Empowers

NASDAQ: ELGX  December Innovation that Empowers NASDAQ: ELGX www.endologix.com December 2014 Innovation that Empowers Safe Harbor This presentation contains forward-looking statements within the meaning of the Private Securities Litigation Reform Act

More information

Endovascular Treatment of the Aorta with Fenestrated and Branched Grafts

Endovascular Treatment of the Aorta with Fenestrated and Branched Grafts Endovascular Treatment of the Aorta with Fenestrated and Branched Grafts Eric LG Verhoeven,MD, PhD, A. Katsargyris, MD Vascular and Endovascular Surgery, Paracelsus Medical University, Nuremberg, Germany

More information

Taming The Aorta. David Minion, MD Program Director, Vascular Surgery University of Kentucky Medical Center Lexington, Kentucky, USA

Taming The Aorta. David Minion, MD Program Director, Vascular Surgery University of Kentucky Medical Center Lexington, Kentucky, USA Taming The Aorta David Minion, MD Program Director, Vascular Surgery University of Kentucky Medical Center Lexington, Kentucky, USA Faculty Disclosure Consulting: Endologix, Cook 1 Objectives Review the

More information

Low profile TEVAR: is it an added value? Michel Bosiers, G. Torsello Münster

Low profile TEVAR: is it an added value? Michel Bosiers, G. Torsello Münster Low profile TEVAR: is it an added value? Michel Bosiers, G. Torsello Münster Disclosure Speaker name:...michel Bosiers... I have the following potential conflicts of interest to report: Consulting Employment

More information

symptomatic aneurysms or aneurysms that grow >1cm/yr

symptomatic aneurysms or aneurysms that grow >1cm/yr 1. Elective repair for aneurysm >5.5 cm, symptomatic aneurysms or aneurysms that grow >1cm/yr 2. Ruptured AAA Aneurysm Detection and Management Study (ADAM) and UK Small Aneurysm Trial early open surgery

More information

Treatment options of late failures of EVAS. Michel Reijnen Rijnstate Arnhem The Netherlands

Treatment options of late failures of EVAS. Michel Reijnen Rijnstate Arnhem The Netherlands Treatment options of late failures of EVAS Michel Reijnen Rijnstate Arnhem The Netherlands Disclosure Speaker name: Michel Reijnen I have the following potential conflicts of interest to report: Consulting

More information

Trattamento Endovascolare degli Aneurismi dell Aorta Addominale con Colletto Prossimale Ostile:

Trattamento Endovascolare degli Aneurismi dell Aorta Addominale con Colletto Prossimale Ostile: Trattamento Endovascolare degli Aneurismi dell Aorta Addominale con Colletto Prossimale Ostile: Risultati Precoci e a Medio Termine in Esperienza Monocentrica A.Lauricella - R.Silingardi Chirurgia Vascolare,

More information

EXPERIENCE AFTER 500 ENDOLUMINAL STENT GRAFTS. DEVICES SPECIFIC OUTCOME AND LESSONS LEARNED.

EXPERIENCE AFTER 500 ENDOLUMINAL STENT GRAFTS. DEVICES SPECIFIC OUTCOME AND LESSONS LEARNED. EXPERIENCE AFTER 500 ENDOLUMINAL STENT GRAFTS. DEVICES SPECIFIC OUTCOME AND LESSONS LEARNED. A/Prof. B. Patrice Mwipatayi MD, MMed (Surg), MClinEd (Melb), FCS, FRACS Department of Vascular Surgery, Royal

More information

Eight Year Experience with Type I Endoleaks at a Tertiary Care Center

Eight Year Experience with Type I Endoleaks at a Tertiary Care Center Eight Year Experience with Type I Endoleaks at a Tertiary Care Center Adam Tanious MD, Megan Carroll MD, Mathew Wooster MD, Andrew Jung BA, Marcello Giarelli MSN, Martin Back MD, Bruce Zwiebel MD, Peter

More information

From the Western Vascular Society

From the Western Vascular Society From the Western Vascular Society The role of aortic neck dilation and elongation in the etiology of stent graft migration after endovascular abdominal aortic aneurysm repair with a passive fixation device

More information

Performance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry

Performance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry University of Milan Thoracic Aortic Research Center Performance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry Santi Trimarchi, MD, PhD Associate

More information

Aorfix Pythagoras US Clinical Trial:

Aorfix Pythagoras US Clinical Trial: Overall 1 Year Results from the Aorfix Pythagoras US Clinical Trial: 1 st IDE Study in Highly Angulated 60 133 Aortic Necks Mark Fillinger MD for the PYTHAGORAS Investigators Disclosure Speaker name: Mark

More information

Santi Trimarchi, MD, PhD Vascular Surgeon Thoracic Aortic Research Center, Director IRCCS Policlinico San Donato University of Milan

Santi Trimarchi, MD, PhD Vascular Surgeon Thoracic Aortic Research Center, Director IRCCS Policlinico San Donato University of Milan The Gore GREAT Registry: Update about a real life data collection Santi Trimarchi, MD, PhD Vascular Surgeon Thoracic Aortic Research Center, Director IRCCS Policlinico San Donato University of Milan Disclosures

More information

Important Update to Field Safety Notice Nellix EndoVascular Aneurysm Sealing System Updated Instructions for Use (IFU)

Important Update to Field Safety Notice Nellix EndoVascular Aneurysm Sealing System Updated Instructions for Use (IFU) October 6, 2017 Important Update to Field Safety Notice Nellix EndoVascular Aneurysm Sealing System Updated Instructions for Use (IFU) Dear Physician, This notification is to provide you with further information

More information

INCRAFT system: Update from the Pivotal INSPIRATION Study

INCRAFT system: Update from the Pivotal INSPIRATION Study INCRAFT system: Update from the Pivotal INSPIRATION Study Michel S. Makaroun MD Co-Director, UPMC Heart and Vascular Institute Professor and Chair, Division of Vascular Surgery University of Pittsburgh

More information

Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University Hospital Sangmin Kim

Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University Hospital Sangmin Kim Endovascular Procedures for Isolated Common Iliac and Internal Iliac Aneurysm Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University

More information

Jean M Panneton, MD Professor of Surgery Program Director Vascular Surgery Chief EVMS. Arch Pathology: The Endovascular Era is here

Jean M Panneton, MD Professor of Surgery Program Director Vascular Surgery Chief EVMS. Arch Pathology: The Endovascular Era is here Jean M Panneton, MD Professor of Surgery Program Director Vascular Surgery Chief EVMS Arch Pathology: The Endovascular Era is here Disclosures Consultant: Cook Medical, Bolton Medical, Medtronic Inc, Volcano,

More information

Overcoming Challenging Aortic Anatomy: from Hostile Necks to Tortuous Iliac arteries

Overcoming Challenging Aortic Anatomy: from Hostile Necks to Tortuous Iliac arteries Overcoming Challenging Aortic Anatomy: from Hostile Necks to Tortuous Iliac arteries Robert Fisher Royal Liverpool University Hospital, UK Liverpool Vascular and Endovascular Service Introduction Late

More information

The clinical update for the Zenith AAA Endovascular Graft has included results from the Zenith AAA Endovascular Graft multi-center clinical study,

The clinical update for the Zenith AAA Endovascular Graft has included results from the Zenith AAA Endovascular Graft multi-center clinical study, The clinical update for the Zenith AAA Endovascular Graft has included results from the Zenith AAA Endovascular Graft multi-center clinical study, the 36 mm diameter Zenith Flex AAA Endovascular Graft

More information

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE GUIDANCE EXECUTIVE (GE)

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE GUIDANCE EXECUTIVE (GE) NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE GUIDANCE EXECUTIVE (GE) Review of TA167; Endovascular stent-grafts for abdominal aortic aneurysms Final recommendation post consultation The guidance should

More information

Redo treatment and open conversion after TEVAR

Redo treatment and open conversion after TEVAR Redo treatment and open conversion after TEVAR Roberto Chiesa Vascular Surgery, Vita-Salute University Scientific Institute San Raffaele Milan, Italy Number of procedures Off-Label indications for TEVAR

More information

Challenges. 1. Sizing. 2. Proximal landing zone 3. Distal landing zone 4. Access vessels 5. Spinal cord ischemia 6. Endoleak

Challenges. 1. Sizing. 2. Proximal landing zone 3. Distal landing zone 4. Access vessels 5. Spinal cord ischemia 6. Endoleak Disclosure I have the following potential conflicts of interest to report: Consulting: Medtronic, Gore Employment in industry Stockholder of a healthcare company Owner of a healthcare company Other(s)

More information

Case Report A Case of Successful Coil Embolization for a Late-Onset Type Ia Endoleak after Endovascular Aneurysm Repair with the Chimney Technique

Case Report A Case of Successful Coil Embolization for a Late-Onset Type Ia Endoleak after Endovascular Aneurysm Repair with the Chimney Technique Case Reports in Vascular Medicine Volume 2016, Article ID 5307416, 4 pages http://dx.doi.org/10.1155/2016/5307416 Case Report A Case of Successful Coil Embolization for a Late-Onset Type Ia Endoleak after

More information

Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully

Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully Physician Training Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully supported by self-expanding z-stents H&L-B

More information

FEVAR FIFTEEN YEARS OF EFFICIENCY E.DUCASSE MD PHD FEBVS CHU DE BORDEAUX

FEVAR FIFTEEN YEARS OF EFFICIENCY E.DUCASSE MD PHD FEBVS CHU DE BORDEAUX FEVAR FIFTEEN YEARS OF EFFICIENCY E.DUCASSE MD PHD FEBVS CHU DE BORDEAUX 2018 A BIT OF HISTORY First use of F-EVAR : 1990s Park et al. J Vasc Interv Radiol. 1996;7:819-823. Faruqi et al. J Endovasc Surg.

More information

Durable outcomes. Proven performance.

Durable outcomes. Proven performance. Durable outcomes. Proven performance. GORE EXCLUDER AAA Endoprosthesis GORE EXCLUDER Iliac Branch Endoprosthesis GORE EXCLUDER AAA Endoprosthesis The most-studied* EVAR stent graft designed for durable

More information

Intravascular Ultrasound in the Treatment of Complex Aortic Pathologies. Naixin Kang, M.D. Vascular Surgery Fellow April 26 th, 2018

Intravascular Ultrasound in the Treatment of Complex Aortic Pathologies. Naixin Kang, M.D. Vascular Surgery Fellow April 26 th, 2018 Intravascular Ultrasound in the Treatment of Complex Aortic Pathologies Naixin Kang, M.D. Vascular Surgery Fellow April 26 th, 2018 DISCLOSURES Nothing To Disclose 2 ENDOVASCULAR AORTIC INTERVENTION Improved

More information

Is EVAS a proper choice in women?

Is EVAS a proper choice in women? Is EVAS a proper choice in women? CACVS 2018 Jan MM Heyligers, PhD, FEBVS Consultant Vascular Surgeon Elisabeth TweeSteden Hospital Tilburg The Netherlands Disclosures Consultant for Endologix DEVASS =Dutch

More information

CAUTION: The Nellix EndoVascular Aneurysm Sealing System is an investigational device. Limited by federal (or United States) law to investigational

CAUTION: The Nellix EndoVascular Aneurysm Sealing System is an investigational device. Limited by federal (or United States) law to investigational Safety and Effectiveness Results of the Nellix System IDE Pivotal Trial for Endovascular Aneurysm Sealing PRESENTED BY: Jeffrey P. Carpenter, MD National Principal Investigator Professor and Chairman,

More information

Why Nellix? Treating Concomitant Common Iliac Aneurysms

Why Nellix? Treating Concomitant Common Iliac Aneurysms INDICATIONS FOR USE The Nellix EndoVascular Aneurysm Sealing System can be used in patients who have an infrarenal abdominal aortic (AAA) or aortoiliac aneurysm with suitable anatomy as indicated below:

More information

Fenestrated Implantation: How I do it. Tara M Mastracci, FRCSC, FACS, FRCS, MSc Royal Free London

Fenestrated Implantation: How I do it. Tara M Mastracci, FRCSC, FACS, FRCS, MSc Royal Free London Fenestrated Implantation: How I do it. Tara M Mastracci, FRCSC, FACS, FRCS, MSc Royal Free London Disclosures Cook Medical Inc -- Proctorship and Consultation All speaking and consultation fees are donated

More information

Experience of endovascular procedures on abdominal and thoracic aorta in CA region

Experience of endovascular procedures on abdominal and thoracic aorta in CA region Experience of endovascular procedures on abdominal and thoracic aorta in CA region May 14-15, 2015, Dubai Dr. Viktor Zemlyanskiy National Research Center of Emergency Care Astana, Kazakhstan Region Characteristics

More information