liquicolor (AMP Buffer, IFCC) Design Verification
|
|
- Angelina Maxwell
- 5 years ago
- Views:
Transcription
1 Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction Imprecision arity and Detection Limit arity Detection Limit Comparison of Methods Stability Stability Working Reagent Real Time Stability Recovery of Control Sera Interference Traceability... 7 QM-Element: ISO9001:2000 No /7 valid of
2 1 Introduction The performance characteristics of ALKALINE PHOSPHATASE liquicolor IFCC have been tested and documented in order to verify the clinical usefulness and compliance with the essential requirements of directive 98/79/EC. 2 Imprecision The imprecision (within-run and day-to-day) of the ALKALINE PHOSPHATASE (AMP) liquicolor method was calculated from six determinations on ten consecutive days. Predetermined pool sera of low, medium and high level were employed as sample material. Hitachi 717 Analyte concentration Intra-assay Inter-assay (U/L) SD (U/l) %CV SD (U/l) %CV arity and Detection Limit 3.1 arity The linearity of ALKALINE PHOSPHATASE liquicolor method was controlled by employing a high concentration pool serum and dilutions with phys. saline. The analysed concentrations were calculated vs. the regression line. Deviation from the regression line are expressed in absolute and relative values. As an example data measured on the Hitachi 717 are reported. High Pool Analytical Content (%) U/l U/l U/l (%) Design Verification and Product for ALKALINE PHOSPHATASE liquicolor (AMP BUFFER; IFCC) 2/7
3 MEASURED ANALYTE CONCENTRATION (U 2500 LINEARITY ar Regression Error Bars = ± 2SD HIGH POOL CONTENT (%) The ALKALINE PHOSPHATASE liquicolor IFCC is linear up to about 2000 U/l. 3.2 Detection Limit A 20-fold determination of a 0 calibrator (phys. saline) on a HumaStar 100 analyser revealed an absolute mean of U/l and a SD of 0.54 U/l. From the three-fold standard deviation the detection limit can therefore be calculated on the base of mean +3 SD to 2.11U/l. 4 Comparison of Methods ALKALINE PHOSPHATASE liquicolor has been compared with a commercially available alkaline phosphatase IFCC method. Control sera as well as patient samples have been employed in the comparison. The results have been evaluated by a main component analysis. The linear regression obtained could be described as follows: ALKALINE PHOSPHATASE liquicolor (Y) = * Ref. method (X) The coefficient of correlation was r = Evaluation of the method comparison by a non-parametric model according to Bablok&Passing yielded the following results: Reference (REF) Lowest value = Highest value = Arithmetic mean = Median = Standard deviation = Standard error of the mean = HUMAN AP-IFCC Lowest value = Highest value = Arithmetic mean = Median = Standard deviation = Standard error of the mean = Design Verification and Product for ALKALINE PHOSPHATASE liquicolor (AMP BUFFER; IFCC) 3/7
4 Y = X Intercept A : ;95% CI : to Slope B : ;95% CI : to Cusum test for linearity: No significant deviation from linearity (P>0.10) 40 0 Method Comp arison Alkaline Phosphatase IFCC REF., U/l The comparison study confirmed an excellent agreement between both methods. 5 Stability 5.1 Stability Working Reagent Recovery in Control Sera Commercially available control sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer s instructions. Fivefold determinations of each control serum have been performed with ALKALINE PHOSPHATASE (AMP) liquicolor reagents. The means of the fivefold determinations have been calculated and compared with the values of fresh working reagent. Design Verification and Product for ALKALINE PHOSPHATASE liquicolor (AMP BUFFER; IFCC) 4/7
5 Control serum recovery Working Reagent fresh Working Reagent 5 days 25 C Working Reagent 7 days 25 C Control LOT Result Result Deviation Result Deviation serum U/l U/l (%) U/l (%) HumaTrol N # HumaTrol P # SERODOS # SERO.plus # Mean arity The linearity of ALKALINE PHOSPHATASE liquicolor was controlled according to the procedure already described in 2.1. The working reagents were stressed 5 and 7 days at 25 C. ALKALINE PHOSPHATASE working reagent 5 days 25 C ALKALINE PHOSPHATASE working reagent 7 days 25 C High Pool Analytical Analytical Content (%) U/l U/l U/l (%) U/l U/l U/l (%) The working reagent is stable for 5 days at C and for 4 weeks at 4 C. 5.2 Real Time Stability The stability of ALKALINE PHOSPHATASE liquicolor has been tested on real-time storage conditions over a period of 16, 18 and 20 months for three independent production lots (H019, exp. date: Dec ; H017, exp. date: Oct. 2003; H016.exp. date: Aug. 2003) The measurement was made on Hitachi Recovery of Control Sera Commercially available control sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer s instructions. Twofold determinations of each control serum have been performed with ALKALINE PHOSPHATASE liquicolor method. The means of the twofold determinations have been calculated and compared with the values of the reference reagent. Design Verification and Product for ALKALINE PHOSPHATASE liquicolor (AMP BUFFER; IFCC) 5/7
6 Control serum recovery Reference Reagent Lot H019, 16 months Lot H017, 18 months Lot H016, 20 months Control LOT Result Result Deviation Result Deviation Result Deviation Serum U/l U/l (%) U/l (%) U/l (%) HumaTrol N # HumaTrol P # Serodos # Serodos plus # Serodos plus # Precinorm # Precipath # Mean arity ALKALINE PHOSPHATASE Lot H019, 16 months ALKALINE PHOSPHATASE Lot H017, 18 months High Pool Analytical Analytical Content (%) U/l U/l U/l (%) U/l U/l U/l (%) ALKALINE PHOSPHATASE Lot H016, 20 months High Pool Analytical Content (%) U/l U/l U/l (%) All results support the claimed stability of 18 months from production for the test kit. 6 Interference Interference by bilirubin, hemoglobin and lipemic sera have been studied by adding known amounts of the potentially interfering substance to a known sample. Design Verification and Product for ALKALINE PHOSPHATASE liquicolor (AMP BUFFER; IFCC) 6/7
7 Recoveries have been analyzed according to the method of Glick et al. (Clin.Chem. 1986, ). The results are summarised in the following table. Bilirubin mg/dl Result U/l Dev. % Hemoglobin mg/dl Result U/l Dev. % Intralipid mg/dl Result U/l Dev. % Glick This test is not significantly influenced by bilirubin up to 40 mg/dl, hemoglobin up to 250 mg/dl and lipemic sera up to 1000 mg/dl. 7 Traceability The alkaline phosphatase concentration has been verified with the reference material SRM 909 B. ALKALINE PHOSPHATASE liquicolor revealed recovery well within the reproducibility limits of the method. The method is traceable to SRM 909 B. Design Verification and Product for ALKALINE PHOSPHATASE liquicolor (AMP BUFFER; IFCC) 7/7
liquicolor Design Verification
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Imprecision for... 2 2.2 Imprecision for... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 4
More informationLIPASE liquicolor. Design Verification. Multipurpose Reagent
Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of
More information1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...
More information3 Linearity and Detection Limit Linearity Detection limit... 4
Design Verification liquicolor 1 Introduction... 2 2 Imprecision, Reproducibility... 2 2.1 Data from internal study... 2 2.2 Date from external study (Cholesterol Reference Method Laboratory Network)...
More informationDesign Verification IMTEC-TSH R -A C Intended Use... 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 Interferences... 4 Imprecision...
Design Verification IMTEC-TSH RECEPTOR-ANTIBODIES CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 2.1 Determination of Diagnostic Sensitivity and Diagnostic Specificity...
More informationHumaTex CRP. Design Verification. Contents
Design Verification HumaTex CRP Contents 1 Function... 2 2 Reproducibility... 2 3 Sensitivity and dynamic range... 2 Preparation of serum control panel... 2 Sensitivity test results... 2 Prozone check...
More informationTotal Thyroxine ELISA (T4)
Design Verification Total Thyroxine ELISA (T4) Contents 1 Assay Principle... 2 2 Imprecision... 2 Within-run Imprecision... 2 Between run Imprecision... 2 3 Comparison of Methods, Accuracy... 2 4 Linearity...
More informationHexagon PSA. Design Verification. Contents
Design Verification Hexagon PSA Contents 1. Function...2 2. Sensitivity, Dynamic Range and Traceability...2 Description of Control Materials...2 Analytical Sensitivity...2 3. Clinical Evaluation...3 Evaluation
More informationGentian Canine CRP Immunoassay Application Note for scil VitroVet*
v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationGentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000
v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination
More informationDesign Verification. Form: /5 Rev. 007 valid of
Design Verification g 1 Determination of Diagnostic Sensitivity and Diagnostic Specificity... 2 Crossreactivity... 3 Potentially interfering substances... 3 2 Imprecision... 4 3 Standardisation... 4 4
More informationAccuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method
Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the
More informationSerodos and Serodos plus
Design Verification Serodos and Serodos plus Contents 1 Value Adjustment... 2 2 Target Determination... 2 3 Stability... 2 Real-Time Stability... 3 Stability after Reconstitution... 4 Stability after Reconstitution
More informationDesign Verification. Form:
Design Verification SYPHILIS TPHA liquid CONTENTS 1 Function... 2 2 Sensitivity... 2 Preparation of Serum Control Panel... 2 Kit Controls Positive and Negative... 3 3 Traceability... 3 4 Specificity and
More informationIMTEC-ANA-LIA MAXX. Design Verification
Design Verification IMTEC-ANA-LIA MAXX CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Specificity... 2 3 Interferences... 4 4 Imprecision... 6 4.1 Within-Run Imprecision... 6 4.2 Between-Run
More informationGUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS
GUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS Ferruccio Ceriotti Servizio di Medicina di Laboratorio, Ospedale San Raffaele, Milano F. Ceriotti, Milano, 27-11-2015 2 Preamble Most of the
More informationHumaTex Febrile Antigens
Design Verification HumaTex Febrile Antigens Contents 1 Function... 2 2 Product Line... 2 3 Sensitivity... 2 Description of Control Materials... 2 Kit controls Positive and Negative... 3 4 Interferences...
More informationForum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007
Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease
More informationMETHOD VALIDATION CASE
METHOD VALIDATION CASE METHOD VALIDATION PROTOCOL CLIA Regulation 493.1253 (2) 1.Accuracy (closeness to true/comparative method) 2.Precision (reproducibility) 3.Reference Interval 4.Reportable range (linearity,
More informationEvaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)
Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November
More informationMETHOD VALIDATION: WHY, HOW AND WHEN?
METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test
More informationStandard Operating Procedure
Subject Alkaline Phosphatase C111 Index Number Lab-1504 Section Laboratory Subsection Regional Clinic/Affiliate Hospital Laboratories Category Departmental Contact Kamprud, Elizabeth Last Revised 4/12/2017
More informationBÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:
BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationEvaluation Report. Eurolyser Phenobarbital test kit (VT0211) on. solo Analyser
Evaluation Report Eurolyser Phenobarbital test kit (VT0211) on solo Analyser Location Location: Eurolyser Diagnostica GmbH Operator: Bettina Schönauer Date: February 2016 - March 2016 Specimens The specimens
More information(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationMETHOD VALIDATION CASE
METHOD VALIDATION CASE METHOD VALIDATION PROTOCOL CLIA Regulation 493.1253 (2) 1.Accuracy (closeness to true/comparative method) 2.Precision (reproducibility) 3.Reference Interval 4.Reportable range (linearity,
More informationEvaluation Report. Eurolyser PT (INR) test kit (ST0180) on CUBE and smart analysers
Evaluation Report Eurolyser PT test kit (ST0180) on CUBE and smart analysers Locations Location 1: Eurolyser Diagnostica GmbH Operator: Simone Wieser Date: 30.06.2015 Specimens The specimens used for analysis
More informationhc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine
Design Verification hc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine 1 Function... 2 2 Sensitivity and Dynamic Range... 2 2.1 Analytical Sensitivity... 2 3 Validation of Efficacy... 3 3.1
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationHexagon OBTI. Design Verification. Contents
Design Verification Hexagon OBTI Contents 1 Function... 2 2 Analytical sensitivity... 2 2.1 Description of control materials... 2 2.2 Results with human haemoglobin... 2 3 Specificity, cross-reactivity
More informationUrine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator
December 2014 PRODUCT ANNOUNCEMENT AU Clinical Chemistry Systems AU400/400 e, AU480, AU640/640 e, AU680, AU2700, AU5400, and AU5800 Product Urine/CSF Albumin Reagent Urine/CSF Albumin Reagent Urine/CSF
More informationEvaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser
Evaluation Report Eurolyser HbA1c Test Kit on Smart and Cube Analyser Locations Location 1: Klinisk kemi University Hospital MAS Malmö Operator: Dr. Benny Larsson Date: June/July 2007 Location 2: Eurolyser
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma
Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationOriginal Article Interfering effect of bilirubin on the determination of alkaline phosphatase
Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo,
More informationTurbidos. Design Verification. Contents
Design Verification Turbidos Contents 1 adjustment... 2 2 Target value determination... 2 3 Stability... 2 3.1 Accelerated stress data... 2 3.2 Open vial stability... 3 3.3 Real Time Stability... 4 4 Result
More informationDesign Verification Function... 2 Analytical sensitivity... 2 Specificity, cross-reactivity and interferences... 3
Design Verification 1 Function... 2 2 Analytical sensitivity... 2 2.1 Description of control materials... 2 2.2 Results with human haemoglobin... 2 3 Specificity, cross-reactivity and interferences...
More informationUREA CE Insert. 01 English - Ref.: 27. Ref.:27
UREA CE Insert Ref.:27 Intended use. Colorimetric and enzymatic system for urea determination in samples of blood and urine, by end point reaction. Professional use. [Only for in vitro diagnostic use.]
More informationG. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626
Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma.
Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,
More informationINTERNAL QUALITY CONTROL ( Q I C) QC)
EXTERNAL QUALITY ASSESSMENT PROGRAM (EQAP) BIOCHEMISTRY DEPARTMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty TWO COMPLEMENTARY COMPONENETS OF TQM ARE Internal Quality Control (IQC)
More informationSee external label 2 C-8 C Σ=96 tests Cat # 3171Z. Free Estriol. Cat # 3171Z. Enzyme Linked Immunosorbent Assay
DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external
More informationManual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay
TECHNICAL NOTE GSP Neonatal Phenylalanine kit 3308-0010/3308-001B GSP Neonatal Phenylalanine kit (3308-0010/3308-001B) is a fully Introduction automated enzymatic assay for the GSP system. Optimized for
More informationKinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase
Clinical Chemistry 43:10 1932 1936 (1997) Automation and Analytical Techniques Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Yoshitaka Morishita, 1 * Kiyoshi Nakane,
More information25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL
25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL I. PRINCIPLE The role of vitamin D in bone and mineral metabolism was recognized from its first identification as a factor that could cure rickets. However,
More informationGLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE
More informationEvaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers
Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationDKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum
DKK-1 ELISA, Cat.No. BI-20413 For the quantitative determination of DKK-1 in human serum ASSAY CHARACTERISTICS Method Standard range Conversion factor Sample volume / well Incubation time, temp. Sensitivity
More informationREAD HIGHLIGHTED CHANGES
BIO-FLASH anti-hbs 3000-8581 100 tests The BIO-FLASH anti-hbs is a fully automated chemiluminescent simultaneous immunoassay for quantitative measurement of antibodies to Hepatitis B surface antigen (anti-hbs)
More informationIDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats
IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,
More informationInstructions for Use. Tg Antibody ELISA
Instructions for Use Tg Antibody ELISA Enzyme Immunoassay for the Quantitative Determination of Autoantibodies to Thyroglobulin (Tg) in Serum and Plasma I V D REF EA618/96 12 x 8 2 8 C DLD Gesellschaft
More informationCertificate of Analysis
Siemens Healthcare Diagnostics Products GmbH Emil-von-Behring-Str. 76 35041 Marburg Germany Customer To whom it may concern Certificate of Analysis Material: 10446446 Kit-: 47066 Total lot size: 445 PCK
More informationMethod Comparison Report Semi-Annual 1/5/2018
Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community
More informationDelta Check Calculation Guide
Delta Check Calculation Guide National Technology 2017, All Rights Reserved By Senior Scientific Researcher, Asmaa Taher Table of Contents Definition... 2 Purpose... 2 Delta Check Research Studies... 2
More informationGlutamine Kit. For the determination of glutamine and glutamate in human EDTA plasma and serum. For research use only K 7732.
Li StarFish S.r.l. Via Cavour, 35-20063 Cernusco S/N (MI), Italy Tel. +39-02-92150794 - Fax. +39-02-92157285 info@listarfish.it -www.listarfish.it Manual Kit For the determination of glutamine and glutamate
More informationSCIEX Vitamin D 200M Assay for the Topaz System
The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human
More informationThis package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems.
en CHOLESTEROL 7D62 304796/R02 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems. Read Highlighted Changes Revised October 2012 Package insert
More informationBovine Insulin ELISA
Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:
More informationBias in clinical chemistry. Elvar Theodorsson EFLM and Linköping University
Bias in clinical chemistry Elvar Theodorsson EFLM and Linköping University Bias a controversial subject Different perspectives Reseachers Users Regulatory bodies Standardisation organisations Metrologists
More informationTotal Bilirubin Assay Kit Manual Catalog #:
Total Bilirubin Assay Kit Manual Catalog #: 3460-10 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 2 Required Materials
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationArticle in press - uncorrected proof
Clin Chem Lab Med 2006;44(6):768 773 2006 by Walter de Gruyter Berlin New York. DOI 10.1515/CCLM.2006.125 2006/60 Integration between the Tele-Cardiology Unit and the central laboratory: methodological
More informationVerification and validation of diagnostic laboratory tests in clinical virology
Journal of Clinical Virology 40 (2007) 93 98 Review Verification and validation of diagnostic laboratory tests in clinical virology Holger F. Rabenau a,, Harald H. Kessler b, Marhild Kortenbusch a, Andreas
More informationThyroid Stimulating Hormone (TSH) ELISA Catalog No. GWB , legacy id (96 Tests)
For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. TSH ELISA Kit is intended for the quantitative measurement of TSH in human serum or plasma. For research use only.
More informationCCQM-K11: Key Comparison on the Determination of Total Glucose in Human Serum. Final Report May 2003
CCQM-K11: Key Comparison on the Determination of Total Glucose in Human Serum Final Report May 2003 Michael Welch, Lorna Sniegoski, Reenie Parris, and Willie May Analytical Chemistry Division Chemical
More informationEvaluation of the Abbott CELL-DYN 4000 Hematology Analyzer
Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD
More informationDiagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.
2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846
More informationCOMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP)
European Medicines Agency Pre-authorisation Evaluation of Medicines for Human Use London, 21 September 2006 EMEA/CHMP/BWP/298390/2005 COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) GUIDELINE ON
More informationPorcine/Canine Insulin ELISA
Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research
More informationClinical Chemistry Specific Proteins Presentation
Clinical Chemistry Specific Proteins Presentation November 2004 AEROSET and c8000 are registered trademarks of Abbott Laboratories. All other trademarks, brands, trade names and product names are the property
More informationInsulin (Porcine/Canine) ELISA
Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01
More informationHbA1c enzymatic assay evaluation at AN academic
Peer reviewed ORIGINAL ARTICLE HbA1c enzymatic assay evaluation at AN academic laboratory B Phiri-Ramongane Mmed Chem Path, FCP Path(SA)-Chem Department of Chemical Pathology, Sefako Makgatho Health Sciences
More informationEXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids
DATA SHEET EXOTESTTM ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids INTRODUCTION Exosomes are small endosome-derived lipid nanoparticles
More informationThe Virtues and Pitfalls of Implementing a New Test
The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations
More information3. System Accuracy Evaluation LUX Hgb test 1. Samples
3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be
More informationWeek 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including
Week 17 and 21 Comparing two assays and Measurement of Uncertainty 2.4.1.4. Explain tools used to compare the performance of two assays, including 2.4.1.4.1. Linear regression 2.4.1.4.2. Bland-Altman plots
More informationValidity and Reliability of a Glucometer Against Industry Reference Standards
514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference
More informationFinTest IgG4 Screen 20 ELISA KIT
FinTest IgG4 Screen 20 ELISA KIT Cat. No.:DEIA6196 Pkg.Size:96T Intended use Enzyme immunoassay (microtiter strips) for the detection and the quantitative determination of IgG4 antibodies against 20 Food
More informationQMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800
QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A53727 The QMS Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole
More informationValidation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens
Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens This document describes the validation of a Neogen Fentanyl kit for the semi-quantitative analysis of
More informationThis package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System.
CHOLESTEROL 7D62-21 304342/R1 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System. NOTE: This package insert must be read
More informationIDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats
IDEXX Catalyst SDMA Test for in-house measurement of SDMA in serum from dogs and cats By Graham Bilbrough, Barbara Evert, Karen Hathaway, Gina Panagakos, Jane Robertson, and Maha Yerramilli Introduction
More informationBIOO LIFE SCIENCE PRODUCTS
BIOO LIFE SCIENCE PRODUCTS FOR REFERENCE PURPOSES This manual is for Reference Purposes Only. DO NOT use this protocol to run your assays. Periodically, optimizations and revisions are made to the kit
More informationIGG Diluent NaCl 9 % (50 ml) System-ID English System information For cobas c 311/501 analyzers:
Order information Analyzer(s) on which cobas c pack(s) can be used 03507432 190 150 tests System-ID 07 6787 5 cobas c 311, cobas c 501/502 11355279 216 Calibrator f.a.s. Proteins (5 1 ml) Code 656 03121305
More informationHuman Neurology 3-Plex A
Human Neurology 3-Plex A SUMMARY AND EXPLANATION OF THE TEST The Human N3PA assay is a digital immunoassay for the quantitative determination of total Tau, Aβ42, and Aβ40 in human plasma and CSF. Determination
More informationPERIOSTIN ELISA CONTENTS
PERIOSTIN ELISA for the quantitative determination of Periostin in human serum, EDTA plasma, heparin plasma, and citrate plasma Cat. No. BI-20433. 12 x 8 tests CONTENTS ASSAY CHARACTERISTICS Summary...
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More information2 Screen Islet Cell Autoantibody Human ELISA
2 Screen Islet Cell Autoantibody Human ELISA Cat. No.:DEIA4453 Pkg.Size:96T Intended use The 2 Screen Islet Cell autoantibody (2 Screen) ELISA kit is intended for use by professional persons only, for
More informationACCESS hstni SCIENTIFIC LITERATURE
ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval
More informationAgenda. 15 Dec
Implementing Traceability for Heterogeneous Analytes An Industry View JCTLM Symposium on Reference Measurement Systems for Biologicals, December 15, 2004 R. Miller 15 Dec. 2004 1 Agenda Recent trends in
More informationBlood Urea Nitrogen Enzymatic Kit Manual Catalog #:
Blood Urea Nitrogen Enzymatic Kit Manual Catalog #: 5602-01 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 3 Required
More informationHigh-density Lipoprotein Cholesterol (HDL-C) Assay Kit
(FOR RESEARCH USE ONLY. DO NOT USE IT IN CLINICAL DIAGNOSIS!) High-density Lipoprotein Cholesterol (HDL-C) Assay Kit (Double reagents) Catalog No: E-BC-K221 Method: Colorimetric method Specification: 96T
More informationELEGANCE Chlamydia pneumoniae IgG ELISA KIT
INTENDED USE The ELEGANCE Chlamydia pneumoniae IgG ELISA has been designed for the in vitro diagnostic measurement of anti- Chlamydia pneumoniae IgG in the screening of human serum. PRINCIPLES OF THE ELEGANCE
More informationChapter 2. High-sensitivity C-reactive protein methods examined
Chapter High-sensitivity C-reactive protein methods examined Snježana Rothkrantz-Kos, Maria PJ Schmitz, Otto Bekers, Paul PCA Menheere, Marja P van Dieijen-Visser Clin Chem ;48:359-6 3 Chapter Abstract
More information1. Calibra - H - Store at 2-8 ºC.
CALIBRA H Insert Ref.:80 Lot Expiration Calibrator Attention. It is suggested to verify carefully if the lot number printed in this insert corresponds to the lot on the bottle label. Intended use. is a
More informationLDL LD. 01 English - Ref.: 129. Ref.:129. Insert. Intended use. Methodology. Reagents. Test principle. Summary. Precautions and warnings
LDL LD Insert Ref.:129 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. [For in vitro diagnostic use.] Test principle. The method
More information