Mehmet Nuri Konya1, Özgür Verim2

Size: px
Start display at page:

Download "Mehmet Nuri Konya1, Özgür Verim2"

Transcription

1 Balkan Med J 2017;34: Original Article 425 Numerical Optimization of the Position in Femoral Head of Proximal Locking Screws of Proximal Femoral Nail System; Biomechanical Study Mehmet Nuri Konya1, Özgür Verim2 1 Department of Orthopeadics and Traumatology, Afyon Kocatepe University School of Medicine, Afyon, Turkey 2 Department of Mechanics, Afyon Kocatepe University School of Engineering, Afyon, Turkey Background: Proximal femoral fracture rates are increasing due to osteoporosis and traffic accidents. Proximal femoral nails are routinely used in the treatment of these fractures in the proximal femur. Aims: To compare various combinations and to determine the ideal proximal lag screw position in pertrochanteric fractures (Arbeitsgemeinschaft für Osteosynthesefragen classification 31-A1) of the femur by using optimized finite element analysis. Study Design: Biomechanical study. Methods: Computed tomography images of patients right femurs were processed with Mimics. Afterwards a solid femur model was created with SolidWorks 2015 and transferred to ANSYS Workbench 16.0 for response surface optimization analysis which was carried out according to anterior-posterior (-10 <anteriorposterior<10 ), inferior-superior (-6 <inferiorsuperior<7 ) and tip-apex distance (10 mm<tip-apex distance<30 mm) proximal lag screw positions in the fracture region. The optimum position of the proximal lag screw was determined based on the von Mises stress values occurring on the fracture line. Initial analysis of the system was realized under the surgeon s normal positioning conditions (anterior-posterior, inferiorsuperior=0 ; tip-apex distance=12 mm). Results: The maximum and minimum (compression) von Mises stresses were found to be 438 MPa and MPa, respectively, and risky stresses for the system occurred in the regions where the proximal lag screw passes through the proximal femoral nail hole, the small diameter portion of stem joints with a large diameter and lag screw mounts to the stem. The most suitable position of the proximal lag screw was found at the middle position of the tip-apex distance (20 mm) and femoral neck (anterior-posterior, inferior-superior=0 ), according to von Mises compression stress values occurring on the fracture line. Conclusion: In our study, we couldn t find any correlation between proximal lag screw movement and tip-apex distance on stresses of the fracture surfaces, but the proximal lag screw position in the inferior (inferiorsuperior<0)-superior (inferior-superior>0) and posterioranterior directions of the femur neck significantly increased these stresses. The most suitable position of the proximal lag screw was confirmed as the middle of the femoral neck by using optimized finite element analysis. Keywords: Femoral neck fractures, biomechanics, finite element analysis Address for Correspondence: Dr. Mehmet Nuri Konya, Department of Orthopeadics and Traumatology, Afyon Kocatepe University School of Medicine, Afyon, Turkey nurikonya@hotmail.com Received:12 May 2016 Accepted: 24 March 2017 DOI: /balkanmedj Available at Cite this article as: Konya MN, Verim Ö. Numerical Optimization of the Position in Femoral Head of Proximal Locking Screws of Proximal Femoral Nail System; Biomechanical Study. Balkan Med J 2017;34: Copyright 2017 by Trakya University Faculty of Medicine / The Balkan Medical Journal published by Galenos Publishing House.

2 426 Konya and Verim. Screw Position Incidence of hip fracture is increasing continuously due to increased average age and osteoporosis (1-3). Biomechanical studies have enounced that intramedullary and cephalomedullary devices provide a stable structure for pertrochanteric fractures (4-6). Pertrochanteric fracture [Arbeitsgemeinschaft für Osteosynthesefragen (AO)-OTA 31-A1] lines (7) can start from anywhere on the greater trochanter to above or below the lesser trochanter (Figure 1). These pertrochanteric fractures may be treated with a cephalomedullary nail, sliding hip screw or platescrew fixation (8). The ideal position of the proximal lag screw (PLS) within the femoral head to provide optimal stability to the implant-bone structure has not been studied sufficiently for cephalomedullary nails (9-11). Tip-apex distance (TAD) has been shown to be an important predictor for cut-out of the PLS in cephalomedullary nails (12). Although there is general consensus that the PLS in the middle of femur neck is optimal, there is continuing debate about the ideal position of the PLS (9-13). Our study confirms the importance of the concepts of a suitable TAD, anterior-posterior angle, and inferior-posterior angle as a clinically useful way of describing the suitable position of the PLS. In our study, we aimed to compare various combinations and to determine the ideal position of the PLS in pertrochanteric fractures of the femur by using optimized finite element analysis (FEA). MATERIALS AND METHODS FEA was used to evaluate the importance of adequate positioning of the PLS within the femoral head. Initial analysis of the system was realized under the patients normal positioning conditions [anterior-posterior (AP), inferior-superior (IS)=0 ; TAD=12 mm]. Reconstruction of three-dimensional models from computed tomography This study was approved by the institutional ethics committee (Afyon Kocatepe University Local ethics committee No: 2015/11/300). In our study we used femur computed tomography (CT) images of adult males with no orthopaedic disorders selected from a stored hospital imaging archive with blind selection; for this reason there was no need for patient informed consent forms. Images with 512x512 pixel resolution were obtained using a Sensation 40 CT (Siemens, Erlangen, Germany) and device adjustments were set to 120 kv and m. CT images were processed with Mimics software FIG. 1. Simulation of AO 31-A1 fracture and description of tip-apex distance.

3 Konya and Verim. Screw Position 427 (Materialise, Leuven, Belgium). A total of 665 DICOM (Digital Imaging and Communications in Medicine) images of human femurs with 1 mm slice thickness and 0.6 mm pixel size were analysed. Models were transformed into non-uniform rational basis spline surface format with SolidWorks 2015 software using point cloud methods. Intertrochanteric femur fracture was simulated according to AO classification 31-A1 in three-dimensional (3D) modelling software (SolidWorks). Fea of the cephalomedullary nail The cephalomedullary nail on the human femur head was analysed with FEA determined under loading conditions (7,14). Material properties of all components were assigned as linear elastic isotropic. All materials of the cephalomedullary nail were made of Ti6Al4V (titanium 6, aluminium 4, vanadium alloy) with a modulus of elasticity value of 114 GPa. A value of 0.3 was assigned to all the material Poisson ratios. The modulus of elasticity of the trabecular and cortical components of the femur was taken as 0.86 GPa and 16.8 GPa, respectively (15). Our study simulated the stance phase of walking which is the most commonly used position in virtual environment studies (Figure 2). Finite element models of the proximal femoral nail (PFN) and bone system were composed of approximately elements and nodes. 3D 10-node tetrahedral structural solid elements were used to model the whole system. The element size was 3 mm for the cortical bone, and the contact size between the PFN and the trabecular bone was 1 mm. In this study, the main force (applied to the femoral head at 23 on the frontal and 6 on the sagittal planes) was taken to be 2460 N. Forces of the abductor and the iliopsoas muscles were taken as 1700 N (24 on the frontal and 15 on the sagittal planes) and 771 N (41 on the frontal and 26 on the sagittal planes), respectively (7,14). Implant design We used an uncemented 3D model of modular nail prosthesis combination [modular prosthesis nail combination (MNP ); Neologic Sağlık Hizmetleri, İzmir, Turkey] in this study. MNP is a newly designed modular hip system which combines a PFN and hip prosthesis. The femoral stem of this implant can be used as either a PFN or hip arthroplasty stem and also the can easily convert to a hip prosthesis by changing one part (Figure 3a, 3b). Optimization of the lag screw position The response surface optimization method is a combination of mathematical expressions and statistical techniques used to analyse a response that is affected by several variables and to optimize the response. For many response surface methods, there is a need to estimate the mathematical forms of the FIG. 2. Loading condition of the femoral head. FIG. 3. Modular nail prosthesis: (a) PFN, (b) hip prosthesis.

4 428 Konya and Verim. Screw Position functions between the response and the independent variables in the problems (16). If there is a curvature in the response surface of the system, a second order equation is needed, as follows: A linear relationship of parameters was developed using the product moment correlation coefficient of Spearman and Pearson. A correlation-sensitivity matrix was generated to show the correlation between input and output parameters and the sensitivity of output parameters according to input parameters. The lower, initial and upper values for the input parameters used in the DOE (design of experiment) workspace are given in Table 1. TABLE 1. Input parameters of the proximal locking screw position Lower value (mm) RESULTS Initial value (mm) Upper value (mm) AP ( ) IS ( ) TAD (mm) AP: anterior-posterior; IS: inferior-superior; TAD: tip-apex distance; mm: millimetre In this study, the most suitable position of the lag screw and FEA of the cephalomedullary nail implanted on the human femur were examined. According to the results of the analysis of the current system, the maximum and minimum (compression) von Mises stresses were found to be 438 MPa and MPa, respectively, and risky stresses for the system occurred in the regions where the PLS passes through the PFN hole, the small diameter portion of stem joints with a large diameter and lag screw mounts to the stem (Figure 3a, b). The maximum stresses on the cephalomedullary nail occurred at the start of the stem slope (Figure 4) and stress on the cephalomedullary nail was obtained as 438 MPa. The maximum stress on the PLS was MPa in the region where the PLS passes through the PFN hole (Figure 4), but none of the stresses occurring on the components exceeded the yield strength of the materials (Figure 4c-4e). When the stresses on the fracture line were taken into consideration according to the results of the optimization, 3D diagrams of the response surface of the parameters were reported (Figure 5). The most suitable position of the lag screw was found at the middle position of the TAD (20 mm) and femoral neck (AP, IS=0 ), according to von Mises compression stress values occurring on the fracture line. In our study, we couldn t find any correlation with PLS movement and TAD on fracture line stress distribution, but PLS positions in the inferior (IS<0)-superior (IS>0) and anterior (AP<0)-posterior (AP>0) regions of the femur neck significantly increased these stresses. The most suitable position of the PLS, confirmed by clinical experience in the literature, is in the middle of the femoral neck and TAD (9-11). The worst position of the PLS is in the anterior, posterior, superior region of the femoral neck. But in this context, the risk of damage of the components was not observed. We also determined screw displacement in every direction. According to these results screw displacement in the AP direction was mm, mediolateral direction 0.3 mm, and SI direction 0.2 mm (Figure 4b). DISCUSSION Proximal femur fractures are generally treated with internal fixation or arthroplasty (3). Both treatment regimes have advantages and disadvantages. Internal fixation consists of intramedullary techniques such as PFN and extramedullary techniques such as dynamic hip screw or plate and screw fixation techniques. In either intramedullary or extramedullary techniques, the femoral head must be stabilized with screws. The proximal femoral screw position in both techniques is important for fracture healing. In our literature review for the best screw position, some previous studies were found. Some authors reported suitable screw positions correlated with clinical experience and reported the failure rate as 6-16% (17). Screw positioning and lag screw cut-out reasons are discussed in many studies. Most of these studies are based on clinical experience and studies done retrospectively. In this study, our FEA was totally blinded and aimed to determine different location combinations such as AP, medial-lateral, and inferiorposterior positions of the lag screw. Kane et al. (18) discussed the importance of TAD in a cadaveric study and found that the most suitable position and found no significant differences between low-centre and centre-centre screw position treatment groups but they didn t evaluate various screw configurations in the coronal and sagittal planes and their correlation. We evaluated nearly every point of the femoral head in different directions and load bearing either of lag screw or fracture lines. Goffin et al. (10) tried to find the most suitable PLS position in a finite element study by using nine different positions manually. Screw positions were evaluated with the lag screw and proximal bone fragment (head and neck) and the safest seemed to be inferior middle and inferior posterior positions. A navigation system improved the lag screw position according to Regling

5 Konya and Verim. Screw Position 429 FIG. 4. (a, b) Equivalent von Mises stress and total deformation of the cephalomedullary nail and femur, (c) stresses on the cephalomedullary nail, (d) stresses on the locking screw, (e) stresses on the fracture line of the pertrochanteric fracture.

6 430 Konya and Verim. Screw Position FIG. 5. Von Mises stress distribution on the fracture line according to screw position in three planes, tip-apex distance (mm), AP ( ) and IS ( ). et al. (19) cadaveric study in operations carried out by less experienced and experienced surgeons and the median TAD was 12 mm and 13 mm, respectively. In another study, Hrubina et al. (9) reported 297 patients with 308 hips treated with a dynamic hip screw; the complication rate was reported as 10% and the re-operation rate as 3.9%. In our clinical experience, there are two malpositions of the PFN which can cause pull-out of the nail but no delay or non-union in the fracture line. Munemoto et al. (20) reported hip multidetector row CT in 10 elderly patients and found the highest bone microstructure quality, in the weight-bearing part, in the superior site of the femoral head. In the femoral neck the greatest bone quality was found in the inferior region. But this study was conducted on elderly patients only and didn t evaluate healthy people, which was reported as a limitation. As reported above, bone imaging of elderly people was used in our study; we evaluated CT images from healthy persons and the position of the proximal femoral screw was determined automatically from the program. AP, IS, and TAD measurements were evaluated. In our study the safest position is at the middle position of the TAD (20 mm) and femoral neck (AP, IS=0 ), according to von Mises compression stress values occurring on the fracture line. If the lag screw is not in an adequate position, the load can cause excess stress and this can lead to the failure of components. But, according to results of our study, the stresses on all components did not exceed the yield stress. Our study cannot perform prospectively any exact prediction of complications. In our study, according to optimized FEA, the most suitable position was found to be the middle position of the TAD (20 mm) and femoral neck (AP, IS=0 ), according to von Mises compression stress values occurring on the fracture line. When mechanical properties of the bone and implant of each patient are entered into the FEA, the position of the lag screw can be estimated. Future studies should consider the results of the developed FEA compared to in vivo and mechanical tests with different kinds of implant and, as a clinical study, the impact of the screw position on fracture healing. There are several limitations to note in this study. This research was a biomechanical study so we couldn t simulate the fracture

7 Konya and Verim. Screw Position 431 healing stages. In internal fixation devices for intertrochanteric fracture, whichever is used, lag screws pull the head to the shaft of the femur for compression which can cause femoral neck shortening and lag screws slides laterally (21). Another limitation is that we couldn t compare the results for the femurs of patients of different gender. Strengths of this study are comparing multiple screw positions in different directions in the femoral head with optimized FEA of a newly developed PFN. In our study, we couldn t find any correlation between PLS movement and TAD on stresses of the fracture surfaces, but the PLS position in the inferior (IS<0)-superior (IS>0) and posterior-anterior directions of the femur neck significantly increased these stresses. The most suitable position of the PLS was confirmed as the middle of the femoral neck and TAD by using optimized FEA. Financial Disclosure: This study was supported by Afyon Kocatepe University Scientific Research Project Coordination Unit 16 KARIYER 40 Project number date This study supported by Turkish Society of Orthopeadics and Traumatology Research Councile Project number 23/date Conflict of Interest: No conflict of interest was declared by the authors. REFERENCES 1. Mereddy P, Kamath S, Ramakrishnan M, Malik H, Donnachie N. The AO/ ASIF proximal femoral nail antirotation (PFNA): a new design for the treatment of unstable proximal femoral fractures. Injury 2009;40: Verheyden AP, Josten C. Intramedullary Fixation of Intertrochanteric Fractures with the Proximal Femoral Nail (PFN). Operat Orthop Traumatol 2003;15: Sener M, Onar V, Kazlmoğlu C, Yağdi S. [Mortality and morbidity in elderly patients who underwent partial prosthesis replacement for proximal femoral fractures]. Eklem Hastalik Cerrahisi 2009;20: Curtis MJ, Jinnah RH, Wilson V, Cunningham BW. Proximal femoral fractures: a biomechanical study to compare intramedullary and extramedullary fixation. Injury 1994;25: Kuzyk PR, Zdero R, Shah S, Olsen M, Waddell JP, Schemitsch EH. Femoral head lag screw position for cephalomedullary nails: a biomechanical analysis. J Orthop Trauma 2012;26: Wu X, Yang M, Wu L, Niu W. A Biomechanical Comparison of Two Intramedullary Implants for Subtrochanteric Fracture in Two Healing Stages: A Finite Element Analysis. Appl Bionics Biomech 2015: Oken OF, Soydan Z, Yildirim AO, Gulcek M, Ozlu K, Ucaner A. Performance of modified anatomic plates is comparable to proximal femoral nail, dynamic hip screw and anatomic plates: finite element and biomechanical testing. Injury 2011;42: Azboy I, Demirtaş A, Gem M, Cakır IA, Tutak Y. A comparison of proximal femoral locking plate versus 95-degree angled blade plate in the treatment of reverse intertrochanteric fractures. Eklem Hastalik Cerrahisi 2014;25: Hrubina M, Horák Z, Bartoška R, Navrátil L, Rosina J. Computational modeling in the prediction of Dynamic Hip Screw failure in proximal femoral fractures. J Appl Biomed 2013;11: Goffin JM, Pankaj P, Simpson AH. The importance of lag screw position for the stabilization of trochanteric fractures with a sliding hip screw: a subject-specific finite element study. J Orthop Res 2013;31: Kuzyk P, Higgins G, Zedero R, Shah S, Olsen M, Wadell J, et al. Femoral head lag screw position for cephalomedullary nails: a biomechanical analysis. J Orthop Trauma 2012;26: Kashigar A, Vincent A, Gunton MJ, Backstein D, Safir O, Kuzyk PR. Predictors of failure for cephalomedullary nailing of proximal femoral fractures. Bone Joint J 2014;96: Baumgaertner MR, Solberg BD. Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg Br 1997;79: Seral B, García JM, Cegoñino J, Doblaré M, Seral F. Finite element study of intramedullary osteosynthesis in the treatment of trochanteric fractures of the hip: Gamma and PFN. Injury 2004;35: Mei J, Liu S, Jia G, Cui X, Jiang C, Ou Y. Finite element analysis of the effect of cannulated screw placement and drilling frequency on femoral neck fracture fixation. Injury 2014;45: Kilickap, E. and Huseyinoglu M. Optimization and modelling of burrheight by using response surface methodology and genetic algorithm in drilling AISI 316. J Eng Fac Eng Dicle Univ 2010;1: Gundle R, Gargan MF, Simpson AH. How to minimize failures of fixation of unstable intertrochanteric fractures. Injury 1995;26: Kane P, Vopat B, Heard W, Thakur N, Paller D, Koruprolu S, et al. Is tip apex distance as important as we think? A biomechanical study examining optimal lag screw placement. Clin Orthop Relat Res 2014;472: Regling M, Blau A, Probe RA, Maxey JW, Solberg BD. Improved lag screw positioning in the treatment of proximal femur fractures using a novel computer assisted surgery method: a cadaveric study. BMC Musculoskelet Disord 2014;15: Munemoto M, Kido A, Sakamoto Y, Inoue K, Yokoi K, Shinohara Y, et al. Analysis of trabecular bone microstructure in osteoporotic femoral heads in human patients: in vivo study using multidetector row computed tomography. BMC Musculoskelet Disord 2016;17: Min BW, Lee KJ, Bae KC, Lee SW, Lee SJ, Choi JH. Result of Internal Fixation for Stable Femoral Neck Fractures in Elderly Patients. Hip Pelvis 2016;28:43-8.

Comparative Study of Fixation Devices for Intertrochanteric Fractures

Comparative Study of Fixation Devices for Intertrochanteric Fractures Comparative Study of Fixation Devices for Intertrochanteric Fractures C. Sticlaru * A. Davidescu Politehnica University of Timişoara Politehnica University of Timişoara Timişoara, România Timişoara, România

More information

Technique Guide. DHS Blade. For osteoporotic bone.

Technique Guide. DHS Blade. For osteoporotic bone. Technique Guide DHS Blade. For osteoporotic bone. Table of Contents Introduction Features and Benefits 2 Indications and Contraindications 4 Clinical Cases 5 Surgical Technique Implantation 6 Implant

More information

Failed Subtrochanteric Fracture How I Decide What to Do?

Failed Subtrochanteric Fracture How I Decide What to Do? Failed Subtrochanteric Fracture How I Decide What to Do? Gerald E. Wozasek Thomas M. Tiefenboeck 5 October 2016, Washington Medical University of Vienna, Department of Trauma Surgery ordination @wozasek.at

More information

9/24/2015. When Can I Use a SHS? When CAN T I Use a SHS? Sliding Hip Screw. Time proven. Technically simple. Cheap. Quick

9/24/2015. When Can I Use a SHS? When CAN T I Use a SHS? Sliding Hip Screw. Time proven. Technically simple. Cheap. Quick When Can I Use a SHS? Frank A. Liporace, MD Associate Professor Director of Orthopaedic Trauma Research Director of Orthopaedic Trauma Jersey City Medical Center New York University / Hospital for Joint

More information

Effect of different load conditions on a DHS implanted human femur

Effect of different load conditions on a DHS implanted human femur Effect of different load conditions on a DHS implanted human femur Nooshin S. Taheri, Aaron S. Blicblau, and Manmohan Singh Abstract As a result of improving life expectancy, the number of elderly people

More information

Finite Element Analysis for the Treatment of Proximal Femoral Fracture

Finite Element Analysis for the Treatment of Proximal Femoral Fracture Copyright 2009 Tech Science Press CMC, vol.11, no.1, pp.1-13, 2009 Finite Element Analysis for the Treatment of Proximal Femoral Fracture Ching-Chi Hsu 1, Jinn Lin 2, Yongyut Amaritsakul 3, Takalamesar

More information

Comparitive Study between Proximal Femoral Nailing and Dynamic Hip Screw in Intertrochanteric Fracture of Femur *

Comparitive Study between Proximal Femoral Nailing and Dynamic Hip Screw in Intertrochanteric Fracture of Femur * Open Journal of Orthopedics, 2013, 3, 291-295 Published Online November 2013 (http://www.scirp.org/journal/ojo) http://dx.doi.org/10.4236/ojo.2013.37053 291 Comparitive Study between Proximal Femoral Nailing

More information

PATENTED A-PFN. Antirotator Proximal Femoral Nail. Medical Devices

PATENTED A-PFN. Antirotator Proximal Femoral Nail. Medical Devices PATENTED A-PFN Antirotator Proximal Femoral Nail Medical Devices Introductions Intertrochanteric femoral fractures constitute 0% of all the bone fractures. They are frequently seen in elderly patients

More information

ORIGINAL ARTICLE. INTER TROCHANTERIC # NECK FEMUR FIXATION WITH TFN 250 CASES. Prasad Vijaykumar Joshi, Chandrashekar Yadav.

ORIGINAL ARTICLE. INTER TROCHANTERIC # NECK FEMUR FIXATION WITH TFN 250 CASES. Prasad Vijaykumar Joshi, Chandrashekar Yadav. INTER TROCHANTERIC # NECK FEMUR FIXATION WITH TFN 250 CASES. Prasad Vijaykumar Joshi, Chandrashekar Yadav. 1. Assistant Professor. Department of Orthopaedics, Joshi Hospital Pvt. Ltd. Phaltan, Maharashtra.

More information

Provision of Rotational Stability: Prevention of Collapse: Closed Fracture Reduction: Minimally Invasive Surgery with no Exposure of the Fracture:

Provision of Rotational Stability: Prevention of Collapse: Closed Fracture Reduction: Minimally Invasive Surgery with no Exposure of the Fracture: INTRODUCTION Percutaneous Compression Plating was developed by considering each of the stages in the surgical procedure for pertrochanteric fractures and the ways in which these might be improved. Primary

More information

Randomized comparative study to evaluate the role of proximal femoral nail and dynamic hip screw in unstable trochanteric fractures

Randomized comparative study to evaluate the role of proximal femoral nail and dynamic hip screw in unstable trochanteric fractures International Journal of Research in Orthopaedics Mayi SC et al. Int J Res Orthop. 2016 Sep;2(3):75-79 http://www.ijoro.org Research Article DOI: http://dx.doi.org/10.18203/issn.2455-4510.intjresorthop20162618

More information

A comparative study of 30 cases of trochanteric fracture femur treated with dynamic hip screw and proximal femoral nailing

A comparative study of 30 cases of trochanteric fracture femur treated with dynamic hip screw and proximal femoral nailing Original Article A comparative study of 30 cases of trochanteric fracture femur treated with dynamic hip screw and proximal femoral nailing Jaswinder Pal Singh Walia *, Himanshu Tailor**, H S Mann ***,

More information

Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw

Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw Orthopaedics & Traumatology: Surgery & Research (2010) 96, 21 27 ORIGINAL ARTICLE Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw M. Güven a,, U. Yavuz

More information

A Comparison of the Proximal Femoral Nail Antirotation Device and Dynamic Hip Screw in the Treatment of Unstable Pertrochanteric Fracture

A Comparison of the Proximal Femoral Nail Antirotation Device and Dynamic Hip Screw in the Treatment of Unstable Pertrochanteric Fracture The Journal of International Medical Research 2010; 38: 1266 1275 A Comparison of the Proximal Femoral Nail Antirotation Device and Dynamic Hip Screw in the Treatment of Unstable Pertrochanteric Fracture

More information

Effect of the femoral stem size on femur bone quality towards THR

Effect of the femoral stem size on femur bone quality towards THR Original article: Effect of the femoral stem size on femur bone quality towards THR Palash Kumar Maji a,*, Amit Roy Chowdhury b, Debasis Datta b, S Karmakar a, Subhomoy Chatterjee b and A K Prasad a a

More information

Case Report Late Occurring Medial Migration of a Lag Screw in Gamma Nailing

Case Report Late Occurring Medial Migration of a Lag Screw in Gamma Nailing Case Reports in Orthopedics Volume 2016, Article ID 5201674, 4 pages http://dx.doi.org/10.1155/2016/5201674 Case Report Late Occurring Medial Migration of a Lag Screw in Gamma Nailing S. van Hoef, M. C.

More information

Journal of Orthopaedic Surgery and Research 2010, 5:62

Journal of Orthopaedic Surgery and Research 2010, 5:62 Journal of Orthopaedic Surgery and Research This Provisional PDF corresponds to the article as it appeared upon acceptance. Fully formatted PDF and full text (HTML) versions will be made available soon.

More information

PFNA-II. Proximal Femoral Nail Antirotation.

PFNA-II. Proximal Femoral Nail Antirotation. PFNA-II. Proximal Femoral Nail Antirotation. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. PFNA-II. Proximal Femoral

More information

Functional evaluation of proximal femoral fractures managed with cephalomedullary nailing by oxford hip score - A prospective study

Functional evaluation of proximal femoral fractures managed with cephalomedullary nailing by oxford hip score - A prospective study 2017; 3(3): 980-985 ISSN: 2395-1958 IJOS 2017; 3(3): 980-985 2017 IJOS www.orthopaper.com Received: 13-05-2017 Accepted: 14-06-2017 Pranav Kothiyal Senior Resident, Department of Orthopaedics, SGRRIMHS,

More information

INTERTAN Nails Geared for Stability

INTERTAN Nails Geared for Stability Geared for stability The TRIGEN INTERTAN nail brings advanced TRIGEN nail technology to hip fractures. With a unique integrated, interlocking screw construct, TRIGEN INTERTAN nail provides all the benefits

More information

The Journal of the Korean Society of Fractures Vol.16, No.1, January, 2003

The Journal of the Korean Society of Fractures Vol.16, No.1, January, 2003 The Journal of the Korean Society of Fractures Vol16, No1, January, 2003 : 351 ( )463-712 TEL: (031) 780-5270/5271 FAX : (031) 708-3578 E-mail: bskima@netsgocom 16,, ( > 20mm ) 5, ) 20 % 1 ), 6,, 3 8 8

More information

Standard intramedullary (IM) nails are usually used

Standard intramedullary (IM) nails are usually used 67 Effect of Piriformis Versus Trochanteric Starting Point on Fixation Stability of Short Intramedullary Reconstruction Nails Edward T. Su MD Hargovind DeWal MD Roy Sanders MD Frederick J. Kummer PhD Mohammed

More information

Review of Proximal Nail Antirotation (PFNA) and PFNA-2 Our Local Experience

Review of Proximal Nail Antirotation (PFNA) and PFNA-2 Our Local Experience doi: 10.5704/MOJ.1107.001 Review of Proximal Nail Antirotation (PFNA) and PFNA-2 Our Local Experience WL Loo, M Med Orth, SYJ Loh, FRCS (Edin), HC Lee, FRCS (Edin) Department of Orthopaedic Surgery, Changi

More information

Int J Clin Exp Med 2017;10(5): /ISSN: /IJCEM Xiaowei Huang 2*, Baoqing Yu 1*, Yong Gu 2, Zexiang Li 1

Int J Clin Exp Med 2017;10(5): /ISSN: /IJCEM Xiaowei Huang 2*, Baoqing Yu 1*, Yong Gu 2, Zexiang Li 1 Int J Clin Exp Med 2017;10(5):7867-7874 www.ijcem.com /ISSN:1940-5901/IJCEM0047129 Original Article Biomechanical comparison of dynamic hip screw and Gamma nail for the treatment of unstable trochanteric

More information

International Journal of Research and Review E-ISSN: ; P-ISSN:

International Journal of Research and Review   E-ISSN: ; P-ISSN: International Journal of Research and Review www.ijrrjournal.com E-ISSN: 2349-9788; P-ISSN: 2454-2237 Original Research Article A Prospective Observational Study of Clinical, Functional and Radiological

More information

DISLOCATION AND FRACTURES OF THE HIP. Dr Károly Fekete

DISLOCATION AND FRACTURES OF THE HIP. Dr Károly Fekete DISLOCATION AND FRACTURES OF THE HIP Dr Károly Fekete 1 OUTLINE Epidemiology Incidence Anatomy Patient s examination, clinical symptons Diagnosis Classification Management Special complications 2 EPIDEMIOLOGY,

More information

JMSCR Vol. 03 Issue 08 Page August 2015

JMSCR Vol. 03 Issue 08 Page August 2015 www.jmscr.igmpublication.org Impact Factor 3.79 Index Copernicus Value: 5.88 ISSN (e)-2347-176x ISSN (p) 2455-0450 DOI: http://dx.doi.org/10.18535/jmscr/v3i8.08 Study of Functional and Radiological Outcome

More information

The Mechanical Properties Of Fixing Greater Trochanter Or Lesser Trochanter In Complex Four Part Intertrochanteric Fractures

The Mechanical Properties Of Fixing Greater Trochanter Or Lesser Trochanter In Complex Four Part Intertrochanteric Fractures The Mechanical Properties Of Fixing Greater Trochanter Or Lesser Trochanter In Complex Four Part Intertrochanteric Fractures Chih-Hui Chen 1, Kui-Chou Huang 2, Cheng-Kung Cheng, PhD 3, Hung-Chan Kao 4.

More information

Mechanical simulation study of postoperative displacement of trochanteric fractures using the finite element method

Mechanical simulation study of postoperative displacement of trochanteric fractures using the finite element method Furui et al. Journal of Orthopaedic Surgery and Research (2018) 13:300 https://doi.org/10.1186/s13018-018-1011-y RESEARCH ARTICLE Open Access Mechanical simulation study of postoperative displacement of

More information

Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur

Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur Sowmianarayanan.S Assistant Consultant, Tata Consultancy Services Limited Prof. A.Chandrasekaran Professor, Dept. of Orthopaedic

More information

Angular Malalignment in Subtrochanteric and Proximal Shaft Femur Fractures after Intramedullary Nailing using SIGN Nails

Angular Malalignment in Subtrochanteric and Proximal Shaft Femur Fractures after Intramedullary Nailing using SIGN Nails Angular Malalignment in Subtrochanteric and Proximal Shaft Femur Fractures after Intramedullary Nailing using SIGN Nails Rolando Junior L. Torres, MD Jeremiah R. Morales, MD, FPOA Subtrochanteric Femur

More information

Changes in Tip-Apex Distance by Position and Film Distance Measured by Picture Archiving and Communication System (PACS)

Changes in Tip-Apex Distance by Position and Film Distance Measured by Picture Archiving and Communication System (PACS) ORIGINAL ARTICLE Hip Pelvis 27(1): 36-42, 2015 http://dx.doi.org/10.5371/hp.2015.27.1.36 Print ISSN 2287-3260 Online ISSN 2287-3279 Changes in Tip-Apex Distance by Position and Film Distance Measured by

More information

The Lateral Trochanteric Wall A Key Element in the Reconstruction of Unstable Pertrochanteric Hip Fractures

The Lateral Trochanteric Wall A Key Element in the Reconstruction of Unstable Pertrochanteric Hip Fractures CLINICAL ORTHOPAEDICS AND RELATED RESEARCH Number 425, pp. 82 86 2004 Lippincott Williams & Wilkins The Lateral Trochanteric Wall A Key Element in the Reconstruction of Unstable Pertrochanteric Hip Fractures

More information

Long anatomical femoral stem reduces stress concentrations on femurs with lateral bowing deformity: A finite element analysis

Long anatomical femoral stem reduces stress concentrations on femurs with lateral bowing deformity: A finite element analysis Annals of Medical Research DOI: 10.5455/annalsmedres.2018.09.205 2019;26(1):106-12 Original Article Long anatomical femoral stem reduces stress concentrations on femurs with lateral bowing deformity: A

More information

Page Proof 1 of 5. Fig. E1-A The INTERTAN nail was short or long.

Page Proof 1 of 5. Fig. E1-A The INTERTAN nail was short or long. Page 1 of 5 Fig. E1-A The INTERTAN nail was short or long. Fig. E1-B The sliding hip screw comes in different lengths, and is used with or without a trochanteric stabilizing plate. Page Proof 1 of 5 Page

More information

Assessment of Prognosis of Patients with Intertrochanteric Fractures Undergoing Treatment with PFN: An Observational Study

Assessment of Prognosis of Patients with Intertrochanteric Fractures Undergoing Treatment with PFN: An Observational Study Original article: Assessment of Prognosis of Patients with Intertrochanteric Fractures Undergoing Treatment with PFN: An Observational Study Gajraj Singh 1, Sandhya Gautam 2 1Assistant Professor, Department

More information

GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE

GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE FEATURES AND BENEFITS The One Shot Device is a new component of the Gamma Locking Nail instrumentation system determining the correct position of the

More information

INTERTROCHANTERIC FEMORAL FRACTURES TREATED BY DYNAMIC HIP SCREW

INTERTROCHANTERIC FEMORAL FRACTURES TREATED BY DYNAMIC HIP SCREW 29 INTERTROCHANTERIC FEMORAL FRACTURES TREATED BY DYNAMIC HIP SCREW Muhammad Ayoub Laghari, Asadullah Makhdoom, Pir Abdul latif Qureshi, Abbass Memon, Faheem Ahmed Memon, Professor Khaleeque Ahmed Siddiqui

More information

Treatment of Femoral Intertrochanteric Fracture with Proximal Femoral Nail

Treatment of Femoral Intertrochanteric Fracture with Proximal Femoral Nail 17, 1, 20041 Journal of the Korean Fracture Society Vol. 17, No. 1, January, 2004 Treatment of Femoral Intertrochanteric Fracture with Proximal Femoral Nail Bum-Soo Kim, M.D., Sogu Lew, M.D., Sang-Hun

More information

Lower risk of implant failure and reoperation

Lower risk of implant failure and reoperation The evidence is in... risk of implant failure and reoperation Faster time to fracture union High return to pre-fracture status How satisfied are you with current hip fracture outcomes? Around 1 in 4 hip

More information

The choice of internal fixator for fractures around the femoral trochanter depends on area classification

The choice of internal fixator for fractures around the femoral trochanter depends on area classification DOI 10.1186/s40064-016-3206-1 RESEARCH Open Access The choice of internal fixator for fractures around the femoral trochanter depends on area classification Hiroaki Kijima 1,2*, Shin Yamada 1,2, Natsuo

More information

Biomet Large Cannulated Screw System

Biomet Large Cannulated Screw System Biomet Large Cannulated Screw System s u r g i c a l t e c h n i q u e A Complete System for Simplified Fracture Fixation 6.5mm & 7.3mm The Titanium, Self-drilling, Self-tapping Large Cannulated Screw

More information

Long-stem revision prosthesis for salvage of failed fixation of extracapsular proximal femoral fractures

Long-stem revision prosthesis for salvage of failed fixation of extracapsular proximal femoral fractures Acta Orthop. Belg., 2009, 75, 340-345 ORIGINAL STUDY Long-stem revision prosthesis for salvage of failed fixation of extracapsular proximal femoral fractures Rory J. SHARVILL, Nicholas A. FERRAN, Huw G.

More information

Technique Guide. DHS/DCS System. Including LCP DHS and DHS Blade.

Technique Guide. DHS/DCS System. Including LCP DHS and DHS Blade. Technique Guide DHS/DCS System. Including LCP DHS and DHS Blade. Table of Contents Introduction System Overview 2 Features and Benefits 4 Indications and Contraindications 6 Clinical Cases 8 Surgical

More information

Pelvis injuries Fractures of the femur (proximal,shaft) Dr Tamás Bodzay

Pelvis injuries Fractures of the femur (proximal,shaft) Dr Tamás Bodzay Pelvis injuries Fractures of the femur (proximal,shaft) Dr Tamás Bodzay Pelvis anatomy Pelvis function - axial load bearing - protection: abdominal, pelvic structures Pelvic injury mechanism Falling from

More information

Trochanter Stabilization Plate for DHS Implants

Trochanter Stabilization Plate for DHS Implants Extends DHS Plate Construct to Help Stabilize Greater Trochanter Trochanter Stabilization Plate for DHS Implants Surgical Technique Table of Contents Introduction Trochanter Stabilization Plate for DHS

More information

Internal fixation of femoral neck fractures

Internal fixation of femoral neck fractures Acta Orthop Scand 55, 423-429, 1984 Internal fixation of femoral neck fractures Compression screw compared with nail plate fixation In a prospective, randomized study of femoral neck fracture operations,

More information

Proximal Femoral Shortening after Operation with Compression Hip Screws for Intertrochanteric Fracture in Patients under the Age of 60 Years

Proximal Femoral Shortening after Operation with Compression Hip Screws for Intertrochanteric Fracture in Patients under the Age of 60 Years ORIGINAL ARTICLE Hip Pelvis 27(2): 98-103, 2015 http://dx.doi.org/10.5371/hp.2015.27.2.98 Print ISSN 2287-3260 Online ISSN 2287-3279 Proximal Femoral Shortening after Operation with Compression Hip Screws

More information

Topology optimisation of hip prosthesis to reduce stress shielding

Topology optimisation of hip prosthesis to reduce stress shielding Computer Aided Optimum Design in Engineering IX 257 Topology optimisation of hip prosthesis to reduce stress shielding S. Shuib 1, M. I. Z. Ridzwan 1, A. Y. Hassan 1 & M. N. M. Ibrahim 2 1 School of Mechanical

More information

Intramedullary Nailing: History & Rationale

Intramedullary Nailing: History & Rationale Intramedullary Nailing: History & Rationale Overview 1. What is IM Nailing? 2. History 3. Design Rationale & Evolution 4. Modern IM Nails 5. The Future What is IM Nailing? Method of internal fixation in

More information

Use Of A Long Femoral Stem In The Treatment Of Proximal Femoral Fractures: A Report Of Four Cases

Use Of A Long Femoral Stem In The Treatment Of Proximal Femoral Fractures: A Report Of Four Cases ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 5 Number 1 Use Of A Long Femoral Stem In The Treatment Of Proximal Femoral Fractures: A Report Of Four Cases C Yu, V Singh Citation C Yu, V Singh..

More information

Micro-CT imaging of surgical screw tightening in human trabecular bone

Micro-CT imaging of surgical screw tightening in human trabecular bone Micro-CT imaging of surgical screw tightening in human trabecular bone E. Perilli 1, M. Ryan 1, R. Ab-Lazid 1, J.J. Costi 1, K.J. Reynolds 1 1 Medical Device Research Institute, School of Computer Science,

More information

Gamma3 TM Fragment Control Clip

Gamma3 TM Fragment Control Clip Osteosynthesis Gamma3 TM Fragment Control Clip Operative Technique Hip Fracture Introduction Indication The Fragment Control Clip is designed to stabilize rotationally unstable femoral head-neck fragments

More information

An effort is made to analyse the stresses experienced by the human femur. In order

An effort is made to analyse the stresses experienced by the human femur. In order Finite Element Analysis of Human Femur Bone U N Mughal 1, H A Khawaja 2*, M Moatamedi 1, M Souli 3 1. Narvik University College, Norway 2. UiT-The Arctic University of Norway, Norway 3. University of Lille,

More information

HOW TO CITE THIS ARTICLE:

HOW TO CITE THIS ARTICLE: A COMPARATIVE STUDY OF FUNCTIONAL OUTCOME BETWEEN DYNAMIC HIP SCREW AND PROXIMAL FEMORAL NAIL IN SURGICAL MANAGEMENT OF PER-TROCHANTERIC FRACTURES Umesh M. Shivanna 1, Girish H. Rudrappa 2 HOW TO CITE

More information

Salvage of failed dynamic hip screw fixation of intertrochanteric fractures

Salvage of failed dynamic hip screw fixation of intertrochanteric fractures Injury, Int. J. Care Injured (2005) xxx, xxx xxx 1 www.elsevier.com/locate/injury 2 3 4 5 6 7 8 8 9 10 11 12 13 14 15 Salvage of failed dynamic hip screw fixation of intertrochanteric s G.Z. Said, O. Farouk

More information

Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study

Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study Nikoloski et al. Journal of Orthopaedic Surgery and Research 2013, 8:35 RESEARCH ARTICLE Open Access Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)?

More information

Outcomes of Cephalomedullary Nailing in Basicervical Fracture

Outcomes of Cephalomedullary Nailing in Basicervical Fracture ORIGINAL ARTICLE Hip Pelvis 29(4): 270-276, 2017 http://dx.doi.org/10.5371/hp.2017.29.4.270 Print ISSN 2287-3260 Online ISSN 2287-3279 Outcomes of Cephalomedullary Nailing in Basicervical Fracture Seok

More information

Nailing Stability during Tibia Fracture Early Healing Process: A Biomechanical Study

Nailing Stability during Tibia Fracture Early Healing Process: A Biomechanical Study Nailing Stability during Tibia Fracture Early Healing Process: A Biomechanical Study Natacha Rosa, Fernão D. Magalhães, Ricardo Simões and António Torres Marques Enhanced Bone Healing in intramedullary

More information

Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method

Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method Research Article Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method Mitsugu Todo * Research Institute for Applied Mechanics, Kyushu University, Kasuga, Japan

More information

Distal Femur Fractures in The Elderly The Ideal Construct

Distal Femur Fractures in The Elderly The Ideal Construct Distal Femur Fractures in The Elderly The Ideal Construct Tak-Wing Lau Department of Orthopaedics and Traumatology Queen Mary Hospital The University of Hong Kong Singapore Trauma 2015 Trauma Through the

More information

PFNA. With Augmentation Option.

PFNA. With Augmentation Option. PFNA. With Augmentation Option. Superior anchor in osteoporotic bone Simple and reproducible procedure through standardized technique Intraoperative decision for augmentation PFNA. Proximal Femoral Nail

More information

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

Surgical Technique. CONQUEST FN Femoral Neck Fracture System Surgical Technique CONQUEST FN Femoral Neck Fracture System Table of Contents Introduction... 3 Indications... 3 Product Overview... 4 Surgical Technique... 5 Patient Positioning... 5 Reduce the Fracture...

More information

PFNA. With Augmentation Option.

PFNA. With Augmentation Option. PFNA. With Augmentation Option. Superior anchor in osteoporotic bone Simple and reproducible procedure through standardized technique Intraoperative decision for augmentation This publication is not intended

More information

Type I : At the level of lesser trochanter Type II : Less than 2.5 cm below lesser trochanter. Type III : cm below lesser trochanter

Type I : At the level of lesser trochanter Type II : Less than 2.5 cm below lesser trochanter. Type III : cm below lesser trochanter Type II : Major fracture line along the intertrochanteric line with communition in coronal plain. Type III : Fracture at the level of lesser trochanter with variable communition and extension in subtrochanteric

More information

Comparison of the curative effects of PFNA and DHS fixation in treating intertrochanteric fractures in elderly patients.

Comparison of the curative effects of PFNA and DHS fixation in treating intertrochanteric fractures in elderly patients. Biomedical Research 2017; 28 (6): 2717-2723 ISSN 0970-938X www.biomedres.info Comparison of the curative effects of PFNA and DHS fixation in treating intertrochanteric fractures in elderly patients. Wenjiang

More information

Influence of plate length on the Mechanical performance of dynamic hip screw

Influence of plate length on the Mechanical performance of dynamic hip screw 2012 International Conference on System Modeling and Optimization (ICSMO 2012) IPCSIT vol. 23 (2012) (2012) IACSIT Press, Singapore Influence of plate length on the Mechanical performance of dynamic hip

More information

MRIMS Journal of Health Sciences 2016;4(1) pissn: , eissn:

MRIMS Journal of Health Sciences 2016;4(1) pissn: , eissn: MRIMS Journal of Health Sciences 216;4(1) pissn: 2321-76, eissn: 2321-7294 http://www.mrimsjournal.com/ Original Article A comparative study of proximal femoral nailing and dynamic hip screw in the management

More information

ICUC Paper. The treatment of trochanteric fractures revisited: Pietro Regazzoni, Alberto Fernandez, Dominik Heim, Stephan M. Perren.

ICUC Paper. The treatment of trochanteric fractures revisited: Pietro Regazzoni, Alberto Fernandez, Dominik Heim, Stephan M. Perren. The treatment of trochanteric fractures revisited: Pietro Regazzoni, Alberto Fernandez, Dominik Heim, Stephan M. Perren September 2016 An optimal treatment of hip fractures is crucial because of the great

More information

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Technique Guide LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Table of Contents Introduction Features and Benefits 2 AO ASIF Principles 4 Indications 5 Surgical

More information

Dynamic hip screw fixation of intertrochanteric fractures without using traction table

Dynamic hip screw fixation of intertrochanteric fractures without using traction table Acta Orthop. Belg., 2016, 82, 346-350 ORIGINAL STUDY Dynamic hip screw fixation of intertrochanteric fractures without using traction table Satish Kumar, Gaggan Naresh Chadha From the DNB orthopaedics,

More information

designed to advance the treatment of hip fractures.

designed to advance the treatment of hip fractures. designed to advance the treatment of hip fractures. introducing the tfn-advanced proximal femoral nailing system (tfna). the tfna system is a new system designed to solve a wide range of unmet needs for

More information

Accurate and Easy Measurement of Sliding Distance of Intramedullary Nail in Trochanteric Fracture

Accurate and Easy Measurement of Sliding Distance of Intramedullary Nail in Trochanteric Fracture Original Article Clinics in Orthopedic Surgery 2015;7:152-157 http://dx.doi.org/10.4055/cios.2015.7.2.152 Accurate and Easy Measurement of Sliding Distance of Intramedullary Nail in Trochanteric Fracture

More information

The role of the calcar femorale in stress distribution in the proximal femuros4_

The role of the calcar femorale in stress distribution in the proximal femuros4_ Orthopaedic Surgery (2009), Volume 1, No. 4, 311 316 ORIGINAL ARTICLE The role of the calcar femorale in stress distribution in the proximal femuros4_53 311..316 Qi Zhang MD 1, Wei Chen MD 1, Huai-jun

More information

Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System. Abbreviated Surgical Technique

Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System. Abbreviated Surgical Technique Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System Abbreviated Surgical Technique ITST System Abbreviated Surgical Technique Indications The ITST Intramedullary Nail is indicated for use in

More information

Analysis of trabecular bone microstructure in osteoporotic femoral heads in human patients: in vivo study using multidetector row computed tomography

Analysis of trabecular bone microstructure in osteoporotic femoral heads in human patients: in vivo study using multidetector row computed tomography Munemoto et al. BMC Musculoskeletal Disorders (2016) 17:13 DOI 10.1186/s12891-015-0848-z RESEARCH ARTICLE Analysis of trabecular bone microstructure in osteoporotic femoral heads in human patients: in

More information

Clinical Results of Complex Subtrochanteric Femoral Fractures with Long Cephalomedullary Hip Nail

Clinical Results of Complex Subtrochanteric Femoral Fractures with Long Cephalomedullary Hip Nail ORIGINAL ARTICLE http://dx.doi.org/10.5371/hp.2017.29.2.113 Print ISSN 2287-3260 Online ISSN 2287-3279 Clinical Results of Complex Subtrochanteric Femoral Fractures with Long Cephalomedullary Hip Nail

More information

Prediction Of Pathological Fracture In The Proximal Femora With Bone Lesions Using A CT Scan-based Finite Element Method

Prediction Of Pathological Fracture In The Proximal Femora With Bone Lesions Using A CT Scan-based Finite Element Method Prediction Of Pathological Fracture In The Proximal Femora With Bone Lesions Using A CT Scan-based Finite Element Method Yusuke Kawabata 1, Kosuke Matsuo 2, Hiroyuki Ike, MD 2, Tomoyuki Saito, MD, PhD

More information

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia Types of Plates 1. New Dynamic Compression Plate: DCP Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia 1. Undercut adjacent to the holes low contact: less stress shield 2. Undercut at the undersurface

More information

Peritroch Hip Fractures. Robert M Harris MD. Hip Fractures. Factors Influencing Construct Strength: Uncontrolled factors 4/28/2016

Peritroch Hip Fractures. Robert M Harris MD. Hip Fractures. Factors Influencing Construct Strength: Uncontrolled factors 4/28/2016 Peritroch Hip Fractures Should be treated with an IMHS Robert M Harris MD Hip Fractures General principles Approximately 250,000 hip fractures/ year Cost approximately $8.7 billion annually The number

More information

Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4

Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4 Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4 Abstract Background Centre of Orthopaedic biomechanics, Bath

More information

Aesculap Targon FN. Head Preserving Solution for Medial Femoral Neck Fractures. Aesculap Orthopaedics

Aesculap Targon FN. Head Preserving Solution for Medial Femoral Neck Fractures. Aesculap Orthopaedics Aesculap Targon FN Head Preserving Solution for Medial Femoral Neck Fractures Aesculap Orthopaedics Targon FN Operating Technique Indications for Targon FN AO 3 B. AO 3 B.2 AO 3 B.3 Undisplaced intracapsular

More information

Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay

Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay Reversed distal femoral locking plate for failed proximal femoral nail with nonunion of proximal femoral fractures Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay International Orthopaedics

More information

operative technique Kent Hip

operative technique Kent Hip operative technique Kent Hip The Kent Hip Operative Technique The Kent Hip was developed by Mr Cliff Stossel, FRCS in Maidstone, Kent, UK and first implanted in 1986. It was designed to deal with problems

More information

Case Report. Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology

Case Report. Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology Case Report Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology S. Robert Rozbruch, MD Hospital for Special Surgery New York, NY, USA ABSTRACT This is a case illustrating a 4.5 cm

More information

Reverse LISS plate in treating femoral intertrochanteric fractures: A review of 22 cases.

Reverse LISS plate in treating femoral intertrochanteric fractures: A review of 22 cases. Biomedical Research 2017; 28 (7): 2941-2945 ISSN 0970-938X www.biomedres.info Reverse LISS plate in treating femoral intertrochanteric fractures: A review of 22 cases. Hong-Wei Chen 1*, Jun Pan 2 1 Department

More information

STUDY OF RESULTS OF ENDER NAILING AND CANNULATED CANCELLOUS SCREW IN THE TREATMENT OF INTERTROCHANTERIC FRACTURE FEMUR

STUDY OF RESULTS OF ENDER NAILING AND CANNULATED CANCELLOUS SCREW IN THE TREATMENT OF INTERTROCHANTERIC FRACTURE FEMUR RESEARCH ARTICLE STUDY OF RESULTS OF ENDER NAILING AND CANNULATED CANCELLOUS SCREW IN THE TREATMENT OF INTERTROCHANTERIC FRACTURE FEMUR Bhavik Dalal, Tarkik Amin, Archit Gandhi, Rohit Shah Smt NHL Municipal

More information

Greater trochanter of the femur (GTF) vs. proximal femoral nail anti-rotation (PFNA) for unstable intertrochanteric femoral fracture

Greater trochanter of the femur (GTF) vs. proximal femoral nail anti-rotation (PFNA) for unstable intertrochanteric femoral fracture European Review for Medical and Pharmacological Sciences 2018; 22 (1 Suppl): 8-14 Greater trochanter of the femur (GTF) vs. proximal femoral nail anti-rotation (PFNA) for unstable intertrochanteric femoral

More information

ABSTRACT INTRODUCTION. Hui Xie 1,*, Zhan Wang 2,*, Junji Zhang 3, Langhai Xu 2 and Bao Chen 1. Research Paper

ABSTRACT INTRODUCTION. Hui Xie 1,*, Zhan Wang 2,*, Junji Zhang 3, Langhai Xu 2 and Bao Chen 1. Research Paper /, 2017, Vol. 8, (No. 47), pp: 82700-82704 Clinical outcome of dynamic hip locking plates and proximal femoral nails anti-rotation-asia for treating intertrochanteric femur fracture with lateral wall fractures

More information

Comparative study between proximal femoral nail and proximal femoral nail antirotation in management of unstable trochanteric fractures

Comparative study between proximal femoral nail and proximal femoral nail antirotation in management of unstable trochanteric fractures International Journal of Research in Orthopaedics Kashid MR et al. Int J Res Orthop. 2016 Dec;2(4):354-358 http://www.ijoro.org Original Research Article DOI: http://dx.doi.org/10.18203/issn.2455-4510.intjresorthop20164168

More information

Static and Dynamic Load on Hip Contact of Hip Prosthesis and Thai Femoral Bones K. Chalernpon, P. Aroonjarattham, K.

Static and Dynamic Load on Hip Contact of Hip Prosthesis and Thai Femoral Bones K. Chalernpon, P. Aroonjarattham, K. Static and Dynamic Load on Hip Contact of Hip Prosthesis and Thai Femoral Bones K. Chalernpon, P. Aroonjarattham, K. Aroonjarattham International Science Index, Mechanical and Mechatronics Engineering

More information

Distal femoral fracture with subsequent ipsilateral proximal femoral fracture

Distal femoral fracture with subsequent ipsilateral proximal femoral fracture Distal femoral fracture with subsequent ipsilateral proximal femoral fracture by M Agarwal, MS FRCS, AA Syed, FRCSI, PV Giannoudis (!), BSc,MB,MD,EEC(Orth) Dept. of Orthopaedics and Trauma, St.James University

More information

Functional outcome of proximal femoral nailing in inter trochanteric fractures of femur: A prospective study

Functional outcome of proximal femoral nailing in inter trochanteric fractures of femur: A prospective study 2017; 3(2): 513-518 ISSN: 2395-1958 IJOS 2017; 3(2): 513-518 2017 IJOS www.orthopaper.com Received: 18-02-2017 Accepted: 19-03-2017 Boblee James Professor and Head, Ram Prasath Junior Resident, Vijayakumaran

More information

INTACT AND IMPLANTED FEMUR BEHAVIOUR DURING WALKING AND JOGGING

INTACT AND IMPLANTED FEMUR BEHAVIOUR DURING WALKING AND JOGGING 1 from 4 INTACT AND IMPLANTED FEMUR BEHAVIOUR DURING WALKING AND JOGGING V. A. Papathanasopoulou 1, D. I. Fotiadis 2 and C.V. Massalas 3 1 Dept. of Medical Physics, Medical School, University of Ioannina,

More information

Treatment of Comminuted Subtrochanteric Fractures by Dynamic Hip Screw

Treatment of Comminuted Subtrochanteric Fractures by Dynamic Hip Screw Treatment of Comminuted Subtrochanteric Fractures by Dynamic Hip Screw Pages with reference to book, From 212 To 215 Modood Ali ( Department of Surgery, College of Medicine, King Saud University, P.O.

More information

AO / Synthes Proximal Posterior Medial Tibia Plate. Tibial Plateau Fx. Osteosynthesis

AO / Synthes Proximal Posterior Medial Tibia Plate. Tibial Plateau Fx. Osteosynthesis Tibial Plateau Fx. Osteosynthesis Articular Fractures Osteosynthesis Characteristics of the LCP system LCP 3.5 system LCP 4.5 system Many different traditional plates (small, large) Lag screws Rafting

More information

HIP JOINT CONTACT PRESSURE DISTRIBUTION PRIOR AND AFTER TRIPLE OSTEOTOMY

HIP JOINT CONTACT PRESSURE DISTRIBUTION PRIOR AND AFTER TRIPLE OSTEOTOMY 148 Paper presented at Bucharest, Romania HIP JOINT CONTACT PRESSURE DISTRIBUTION PRIOR AND AFTER TRIPLE OSTEOTOMY Tiberiu LAURIAN 1), Lucian MARINCA 2), Andrei TUDOR 1), Felix PÂRVU 1) 1) POLITEHNICA

More information

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 12, June 2013

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 12, June 2013 Numerical Analysis of Human Femoral Bone in Different Phases Sherekar R.M, Pawar A.N. Department of Mechanical Engineering, Jawaharlal Darda Institute of Engineering and Technology, Yavatmal, 445001 (M.S).

More information

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia. LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia. LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

More information

Dynamic Finite Element Analysis of Impulsive Stress Waves Propagating from the Greater Trochanter of the Femur by a Sideways Fall

Dynamic Finite Element Analysis of Impulsive Stress Waves Propagating from the Greater Trochanter of the Femur by a Sideways Fall 215 9 3 15171 Dynamic Finite Element Analysis of Impulsive Stress Waves Propagating from the Greater Trochanter of the Femur by a Sideways Fall a* b a b 1 9 3 ʼ ɵ ɸ ɵ ɸ ɵ ɸ ɵ ɸ (A (B Sideways fall configuration.

More information