Measurement of Ulnar Variance on Uncalibrated Digital Radiographic Images

Size: px
Start display at page:

Download "Measurement of Ulnar Variance on Uncalibrated Digital Radiographic Images"

Transcription

1 HAND (2010) 5: DOI /s ORIGINAL ARTICLE Measurement of Ulnar Variance on Uncalibrated Digital Radiographic Images Johannes de Vries & J. Sebastiaan Souer & David Zurakowski & David Ring Received: 27 October 2009 / Accepted: 1 December 2009 / Published online: 22 December 2009 # American Association for Hand Surgery 2009 Abstract Uncalibrated digital radiographs used in multicenter trials hinder quantitative measures such as articular step and ulnar variance. This investigation tested the reliability of alternative measures of ulnar variance that are scaled to the length of the capitate. A sample of 30 sets of radiographs from patients enrolled in a prospective study of operative treatment of fractures of the distal radius were blinded and randomized. Five observers measured the ulnar variance (UV) and longitudinal length of the capitate (CH) on two separate occasions with greater than 2 weeks between measurements. During each measurement session, the observers made the measurements on both a calibrated and a noncalibrated workstation. The ratio of the ulnar variance to the length of capitate was calculated (UV/CH ratio). Paired t tests were used to compare two rounds of measurements for both methods. Intra- and interobserver reliability was assessed by the Pearson product-moment correlation coefficients. The ratios were compared using J. de Vries : J. S. Souer : D. Ring (*) Orthopaedic Hand and Upper Extremity Service, Harvard Medical School, Massachusetts General Hospital, Yawkey Center, Suite 2100, 55 Fruit Street, Boston, MA 02114, USA dring@partners.org D. Zurakowski Department of Anesthesia, Children s Hospital Boston, Boston, USA D. Zurakowski : D. Ring Department of Orthopaedic Surgery, Children s Hospital Boston, Boston, USA D. Zurakowski Orthopaedic Surgery, Harvard Medical School, 300 Longwood Avenue, Boston, MA 02115, USA analysis of variance with a Bonferroni correction. The intraobserver reliability was excellent for each of the three variables (UV, CH, UV/CH ratio) for each workstation. The interobserver reliability of the UV/CH ratios obtained for each workstation was moderate to excellent as judged by the Pearson correlations between observers. The Bland Altman method indicated a mean difference in UV/CH between calibrated and uncalibrated measurement techniques of with limits of agreement of 0.11 to Measurements of ulnar variance that are scaled to the length of the capitate may be useful measures of deformity in studies that utilize uncalibrated digital radiographs. Keywords Ulnar variance. Uncalibrated digital images. Ratio. Capitate Introduction Ulnar variance and articular step and gap are important quantitative measures of the alignment of the articular surface of the distal radius that are used to evaluate the results of fracture treatment [9, 11, 18]. Steyers and colleagues compared various methods for measuring ulnar variance including the project-a-line technique, the method of concentric circles, and the method of perpendiculars and found that the method of perpendiculars was most reliable for both intraobserver and interobserver comparisons [16]. Whereas angular measurements do not change with scale, measures of ulnar variance and articular incongruity in millimeters require knowledge of the scale or calibration of the radiographs. Given the increasing use of digital images, and the transfer of uncalibrated or unscaled versions of those images between centers for multicenter research, there is a need for measures of ulnar variance and

2 268 HAND (2010) 5: articular incongruity that are expressed as ratios that can be used when the scale of the radiograph is not known. In this study, we evaluated calibrating ulnar variance by dividing it by the length of capitate. The hypothesis of our study was that the use of measurements scaled to the capitate length would be as accurate and reproducible as measurements on calibrated or actual-size images. Materials and Methods Thirty standard radiographs of the wrist obtained after manipulative reduction of a fracture of the distal radius were selected from an IRB approved clinical trial enrolling patients who elect open reduction and internal plate fixation. Inclusion criteria for this study were: (1) calibrated digital postreduction posteroanterior and lateral radiographs including the base of the third metacarpal. Exclusion criteria included (1) complex multifragmentary intra-articular fracture (e.g., greater than three articular fragments); (2) fracture dislocations of the wrist; (3) intercarpal injuries; (4) dislocation at the distal radio-ulnar joint determined on a true lateral radiograph (reference). According to the Comprehensive Classification for Fractures [13], there were one A2.2 fracture, two A3.2 fractures, one B2.1 fracture, one B2.2 fracture, one B2.3 fracture, one B3.1 fracture, one B3.3 fracture, four C1.1 fractures, six C1.2 fractures, one C1.3 fracture, four C2, four C3.1 fractures, two C3.2 fractures, and one C3.3 fracture. According to the Fernandez classification [6], there were three extra articular bending fractures, five shearing (type 2) fractures, and 22 compression fractures. lunate facet at the sigmoid notch. The point at which a third perpendicular line drawn midway between the volar and dorsal articular marginal lines intersects the ulnar margin of the distal radial articular surface was used as the point of reference for the radial articular surface (Fig. 1). Uncalibrated Measurements The 30 radiographs were introduced as jpeg images into an uncalibrated workstation. (efilm software, Merge Healthcare, Milwaukee, WI, USA). Using the same guidelines, the measurements of the length of the capitate and the ulnar variance were quantified with a digital ruler tool and expressed in pixels instead of millimeters. Study Design Five observers not involved in the care of the patients measured all 30 radiographs blinded and in random order using the techniques described. The measurements were scheduled into two rounds; each round consisting of two sessions one session on the calibrated and one on the uncalibrated workstation. The observers were trained by one of us using three sets of radiographs that were not used in the study. Calibrated Measurements Using commercial software (IMPAX ES DS3000, Agfa- Gevaerts N.V., Mortsel, Belgium), we made direct, calibrated measurements of length of capitate and the ulnar variance in millimeters using a digital ruler tool. Using the posteroanterior radiographs, the length of capitate was measured as the longest distance between the subchondral margin of the capitate at the capitate third metacarpal junction to the proximal aspect of the capitate, as described by Natrass et al. [14] (Fig. 1). Ulnar variance was measured as the distance between a line perpendicular to the longitudinal axis through the distal ulnar aspect of the radius and the distal cortical rim of the ulna [3, 16]. In order to correct for angulation of the distal radius fragment, we had to adjust the perpendicular line through the distal ulnar aspect of the radius. First, a line was drawn from the tip of the radial styloid to the dorsal margin lunate facet of the distal radius at the sigmoid notch. A second line perpendicular to the first was placed at the volar margin of the Figure 1 Example of a radiograph for which correction was necessary to account for rotation/angulation of the distal fracture fragments. A Length of capitate. B The distal cortical rim of the ulna. C The tip of the radial styloid to the dorsal aspect of the sigmoid notch. D The tip of the radial styloid to the volar aspect of the sigmoid notch. E The perpendicular line corrected for angulation.

3 HAND (2010) 5: Table 1 Intraobserver reliability for measurements based on each workstation. Round 1 Round 2 Mean difference SD of difference p value a Pearson r correlation b IMPAX UV (cm) CH (cm) UV/CH efilm UV (pixels) CH (pixels) UV/CH Data are based on mean values for all 5 observers pooled together a Paired t test. b Correlations indicate excellent agreement between rounds for each variable (p<0.001) In the first round, the five observers performed the measurements on the 30 radiographs using the calibrated workstation for a total of 150 observations (one observation includes length of capitate and ulnar variance). The images were then randomized and introduced into an uncalibrated workstation. After 7 days, the observers repeated all the measurements on an uncalibrated workstation for another total of 150 observations. The second round was performed more than 2 weeks after completion of the first round. Again, all observers performed the measurements on the calibrated workstation (150 observations) followed by a final session on the uncalibrated workstation (150 observations). Statistical Analysis Intra- and interobserver reliability was assessed by the Pearson product-moment correlation coefficient (r). Paired t tests were used to compare measurements of ulnar variance (UV), capitate height (CH), and the UV/CH ratio between rounds 1 and 2 for the calibrated IMPAX and noncalibrated efilm workstations. Repeated-measures mixed-model analysis of variance (ANOVA) with Bonferroni correction for multiple comparisons was used to compare the UV/CH ratios obtained by the five observers in each of two rounds based on IMPAX and efilm systems [19]. In addition, the Bland Altman method in which the average UV/CH is Table 2 Intraobserver reliability for the ulnar variance/capitate height ratio based on each workstation. Round 1 Interobserver correlation (r) Round 2 Interobserver correlation (r) Mean ± SD Mean (range) a Mean ± SD Mean (range) a IMPAX Observer ± ±0.100 Observer ± ±0.105 Observer ± ±0.095 Observer ± ±0.089 Observer ± ±0.120 Total 0.023± ( ) 0.023± ( ) efilm Observer ± ±0.087 Observer ± ±0.113 Observer ± ±0.096 Observer ±0.099 a 0.056±0.107 a Observer ± ±0.128 Total 0.024± ( ) 0.027± ( ) Repeated-measures analysis of variance was used to compare observers to account for the same 30 X-rays evaluated by each observer. No significant differences were detected between observers for the IMPAX workstation. For the efilm workstation, the UV/CH ratio was greater for observer 4 compared to observers 1, 2, and 5 in round 1, and compared to observers 1 and 2 in round 2 (p<0.05, ANOVA with Bonferroni correction). a Highly significant average correlations between observers, with ranges that indicate at least moderately high interobserver agreement).

4 270 HAND (2010) 5: Figure 2 A scatter plot of the calibrated (Impax) and uncalibrated (Efilm) UV/CH measurements shows a moderately strong positive linear relationship. plotted against the difference with 95% limits of agreement was applied to evaluate differences in the ratio between the two workstations for all five observers pooled together based on a paired analysis of the same 30 X-rays and two rounds for each observer using IMPAX and efilm (300 paired observations) [1]. One common application of the standard Bland Altman method is to compare a new measurement technique with an established one. The primary objective of comparison studies is to determine whether the two methods agree closely enough to be used interchangeably. However, use of the Bland Altman methodology, including the 95% limits of agreement, does not require that one method be a gold standard. For example, two different assays can be compared, two different devices for measuring bone density, two different workstation systems for measuring ulnar variance. The analysis does not depend on either method being a gold standard, rather that the two methods measure the same variable on the same scale in the same patients (paired analysis) and that the difference between methods should be approximately normally distributed. This is unlike correlation analysis, which does not require that the two measurements be on the same scale or even that they represent the same characteristic or quantity. The Bland Altman technique was applied in this context to graphically examine the agreement beween in UV, CH, and UV/CH between IMPAX and efilm methods in order to determine the average difference (i.e., bias) and the limits of agreement (precision) between these methods. The Bland Altman technique does not have p values associated with it but does allow comparison of two methods based on how closely the individual patients fall with respect to the 45 line of identity, in which the two methods perfectly agree (Y=X). The mean difference (bias) between the two workstations and 95% CI for the difference (precision) indicate the level of agreement between the two methods of measurement. A linear regression model strategy was applied, using the general form y=mx+b, to derive the linear relationship between the UV/CH ratio based on the IMPAX and efilm workstations. A power analysis indicated that 30 X-rays assessed twice by five different observers using each workstation (300 paired data points) for UV, CH, and UV/ CH would provide sufficient number of observations to construct a 95% confidence interval around the difference in UV/CH ratio between the IMPAX and efilm systems with a precision of 0.1 (version 7.0, nquery Advisor, Statistical Solutions, Saugus, MA). Statistical analysis was performed using the Stata software package (version 9, Stata Corporation, College Station, Texas). Two-tailed values of p<0.05 were considered statistically significant with a Bonferroni correction for multiple comparisons between observers. Results The intraobserver reliability was excellent for each of the three variables (UV, CH, UV/CH ratio) for each workstation as indicated by no significant differences between round 1 and round 2 for the five observers (all p>0.20, paired t tests). The correlations are all excellent and highly significant in terms of measurements between the two rounds (Table 1). Interobserver reliability regarding the UV/ CH ratios obtained for each workstation was moderate to excellent as judged by the average Pearson correlations between observers (Table 2). No differences were found Figure 3 The histogram of the difference between the calibrated (IMPAX) and uncalibrated (efilm) UV/CH ratios is centered very close to zero but clearly shows variation such that the ratio can be larger or smaller with either workstation.

5 HAND (2010) 5: between observers using the IMPAX system, whereas with the noncalibrated efilm workstation, only observer 4 showed significantly different results as compared to other observers. The average correlation between the five observers was moderately high on both rounds 1 and 2 using the IMPAX workstation (r=0.82 and r=0.89, respectively, both p<0.001), whereas with efilm the interobserver correlations were similar for the two rounds, on the average (r=0.86 and r=0.82, respectively, both p< 0.001). The Bland Altman method indicated a mean difference in UV/CH between calibrated and uncalibrated measurement techniques of with limits of agreement of 0.11 to Linear regression analysis indicated a strong positive relationship between IMPAX and efilm UV/CH with a prediction equation, y=0.86x+0.005, where x is the ratio as obtained using the IMPAX workstation (r 2 =0.73, p<0.001). The scatter plot of the UV/CH measurements for the two systems shows a moderately strong positive linear relationship (Fig. 2). The histogram of the difference between the IMPAX and efilm UV/CH ratios is centered very close to zero but clearly shows variation such that the ratio can be larger or smaller with either workstation (Fig. 3). In fact, the mean difference is constant across the range of possible values for UV/CH as demonstrated by a lack of a correlation (i.e., no significant slope) between the difference in UV/CH (IMPAX efilm) versus the average UV/CH based on both workstations (Pearson r= 0.01, p=0.95, Fig. 2). Discussion We realize that there is not yet an accepted standard measurement of ulnar variance. The issues relate not only to the method of measurement [10, 15, 16] but also to methods of accounting for the position of the patient and the distal radial fracture fragment [8, 17, 20] as well as the position of the radiographic beam and the machine [2, 4, 7, 20]. We could not find a reference for our technique of using the radial styloid and the volar and dorsal margins of the lunate fossa to determine a reference point on the radius for determining ulnar variance and the technique may or may not be novel. On the other hand, for the purposes of this study, the exact technique is not as important as the reliability of the measurements made on calibrated and uncalibrated digital images. We chose the capitate as a reference for standardizing measurements made on calibrated films because there is precedent for its use in this regard. One common use of the capitate is in the revised carpal height ratio. This was devised because wrist radiographs do not routinely include the entire third metacarpal as needed for the original carpal height ratio [12, 21]. The advantages of using the longitudinal length of the capitate include: consistent measurement in length [5]; the length measurement is not subject to ulnar or radial deviation of less than 18 [14], and injury to the capitate is uncommon. Furthermore, the capitate is consistently present on standard wrist radiographs. Our results suggest that observers using the capitate to scale radiographic length measure of deformity of the distal radius make consistent measurements and agree with one another. Measurements of ulnar variance that are scaled to the length of the capitate may be useful alternative measures of deformity in studies that utilize uncalibrated digital radiographs. The ratio is best used for statistical comparisons within a single study and is less intuitive when comparing data from different studies. Some aspects of our analysis may have affected the results. For instance, we rounded the ratios to two significant digits while recording the measurements. Although highly reliable in our hands, the findings should be confirmed in other settings using various accepted techniques and workstations for verification. References 1. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476): Bozentka DJ, Beredjiklian PK, Westawski D, Steinberg DR. Digital radiographs in the assessment of distal radius fracture parameters. Clin Orthop Relat Res. 2002;397: Coleman DA, Blair WF, Shurr D. Resection of the radial head for fracture of the radial head. Long-term follow-up of seventeen cases. J Bone Joint Surg Am. 1987;69(3): Epner RA, Bowers WH, Guilford WB. Ulnar variance the effect of wrist positioning and roentgen filming technique. J Hand Surg [Am]. 1982;7(3): Feipel V, Rinnen D, Rooze M. Postero-anterior radiography of the wrist. Normal database of carpal measurements. Surg Radiol Anat. 1998;20(3): Fernandez DL, Jupiter JB. Fractures of the distal radius. New York: Springer; p Johnson PG, Szabo RM. Angle measurements of the distal radius: a cadaver study. Skeletal Radiol. 1993;22(4): Jung JM, Baek GH, Kim JH, Lee YH, Chung MS. Changes in ulnar variance in relation to forearm rotation and grip. J Bone Joint Surg Br. 2001;83(7): Kreder HJ, Hanel DP, McKee M, Jupiter J, McGillivary G, Swiontkowski MF. X-ray film measurements for healed distal radius fractures. J Hand Surg [Am]. 1996;21(1): Kristensen SS, Thomassen E, Christensen F. Ulnar variance determination. J Hand Surg [Br]. 1986;11(2): Mackenney PJ, McQueen MM, Elton R. Prediction of instability in distal radial fractures. J Bone Joint Surg Am. 2006;88(9): McMurtry RY, Youm Y, Flatt AE, Gillespie TE. Kinematics of the wrist. II. Clinical applications. J Bone Joint Surg Am. 1978;60 (7):

6 272 HAND (2010) 5: Müller ME, Nazarian S, Koch P, Schatzker J. The comprehensive classification of fractures of long bones. New York: Springer; p Nattrass GR, King GJ, McMurtry RY, Brant RF. An alternative method for determination of the carpal height ratio. J Bone Joint Surg Am. 1994;76(1): Palmer AK, Glisson RR, Werner FW. Ulnar variance determination. J Hand Surg [Am]. 1982;7(4): Steyers CM, Blair WF. Measuring ulnar variance: a comparison of techniques. J Hand Surg [Am]. 1989;14(4): Tomaino MM. The importance of the pronated grip X-ray view in evaluating ulnar variance. J Hand Surg [Am]. 2000;25(2): Trumble TE, Schmitt SR, Vedder NB. Factors affecting functional outcome of displaced intra-articular distal radius fractures. J Hand Surg [Am]. 1994;19(2): Vittinghoff E, Glidden DV, Shiboski SC, McCulloch CE. Regression methods in biostatistics: linear, logistic, survival, and repeated measures models. New York: Springer; p Yeh GL, Beredjiklian PK, Katz MA, Steinberg DR, Bozentka DJ. Effects of forearm rotation on the clinical evaluation of ulnar variance. J Hand Surg [Am]. 2001;26(6): Youm Y, McMurthy RY, Flatt AE, Gillespie TE. Kinematics of the wrist. I. An experimental study of radial-ulnar deviation and flexion-extension. J Bone Joint Surg Am. 1978;60(4):

journal ORIGINAL RESEARCH

journal ORIGINAL RESEARCH texas orthopaedic journal ORIGINAL RESEARCH Assessment of Volar Tilt Measurements with Variations in X-Ray Beam Centralization Along the Longitudinal Axis of the Radius Russell A. Wagner, MD; Will Junius,

More information

)451( COPYRIGHT 2017 BY THE ARCHIVES OF BONE AND JOINT SURGERY. Evaluation of Normal Ranges of Wrist Radiologic Indexes in Mashhad Population

)451( COPYRIGHT 2017 BY THE ARCHIVES OF BONE AND JOINT SURGERY. Evaluation of Normal Ranges of Wrist Radiologic Indexes in Mashhad Population )451( COPYRIGHT 2017 BY THE ARCHIVES OF BONE AND JOINT SURGERY SHORT COMMUNICATION Evaluation of Normal Ranges of Wrist Radiologic Indexes in Mashhad Population Tohid Vaezi, MD; Golnaz Ghayyem Hassankhani,

More information

Essential Radiographic Evaluation for Distal Radius Fractures

Essential Radiographic Evaluation for Distal Radius Fractures Hand Clin 21 (2005) 279 288 Essential Radiographic Evaluation for Distal Radius Fractures Robert J. Medoff, MD Department of Orthopaedic Surgery, University of Hawaii, 30 Aulike Street #506, Kailua, HI

More information

University of Groningen. Fracture of the distal radius Oskam, Jacob

University of Groningen. Fracture of the distal radius Oskam, Jacob University of Groningen Fracture of the distal radius Oskam, Jacob IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document

More information

Radiographic Evaluation and Classification of Distal Radius Fractures

Radiographic Evaluation and Classification of Distal Radius Fractures Radiographic Evaluation and Classification of Distal Radius Fractures Robert J Medoff, MD Introduction X-rays are essential to the treatment of distal radius fractures. When combined with the age and baseline

More information

Factors Predicting Late Collapse of Distal Radius Fractures

Factors Predicting Late Collapse of Distal Radius Fractures http://dx.doi.org/10.5704/moj.1111.006 Factors Predicting Late Collapse of Distal Radius Fractures, MD Regional Hospital Durres, Orthopaedic Clinic, Durres, Albania ABSTRACT Background: Although fractures

More information

ORIGINAL ARTICLE. possible. Accurate assessment of standard radiographs is essential for appropriate 3. management. And includes true posterior- 4

ORIGINAL ARTICLE. possible. Accurate assessment of standard radiographs is essential for appropriate 3. management. And includes true posterior- 4 ORIGINAL ARTICLE Treatment of Colle's Fracture with Wrist Immobilization in Palmar flexed & Dorsiflexed Position Sohail Iqbal Shaikh, Abdul Basit, Javed Iqbal, Saba Sohail Shaikh, Imran Sohail Shaikh 26

More information

Radiographic indices in one hundred fifty normal Iranian wrists

Radiographic indices in one hundred fifty normal Iranian wrists Original Research Medical Journal of Islamic Republic of Iran, Vol. 26, No. 3, Aug. 2012, pp.132-139 Radiographic indices in one hundred fifty normal Iranian wrists Davood Jafari 1, Hamid Taheri 2, Hooman

More information

Hand and wrist emergencies

Hand and wrist emergencies Chapter1 Hand and wrist emergencies Carl A. Germann Distal radius and ulnar injuries PEARL: Fractures of the distal radius and ulna are the most common type of fractures in patients younger than 75 years.

More information

Department of Orthopedic Surgery, Konyang University Hospital, Daejeon, Korea

Department of Orthopedic Surgery, Konyang University Hospital, Daejeon, Korea Original Article Clinics in Orthopedic Surgery 2012;4:83-90 http://dx.doi.org/10.4055/cios.2012.4.1.83 Evaluation of the Sigmoid Notch Involvement in the Intra-Articular Distal Radius Fractures: The Efficacy

More information

Interesting Case Series. Ulnolunate Impaction Syndrome

Interesting Case Series. Ulnolunate Impaction Syndrome Interesting Case Series Ulnolunate Impaction Syndrome Saptarshi Biswas, MD, FRCS Westchester University Medical Center, Valhalla, NY Keywords: ulnar impaction, ulnar impaction syndrome, ulnar wrist pain,

More information

Introduction. The wrist contains eight small carpal bones, which as a group act as a flexible spacer between the forearm and hand.

Introduction. The wrist contains eight small carpal bones, which as a group act as a flexible spacer between the forearm and hand. Wrist Introduction The wrist contains eight small carpal bones, which as a group act as a flexible spacer between the forearm and hand. Distal forearm Distal forearm 4 Distal end of the radius A. anterior

More information

Scapholunate Ligament Lesions Imaging Which and when?

Scapholunate Ligament Lesions Imaging Which and when? Scapholunate Ligament Lesions Imaging Which and when? Kolo Frank Lesions to scapholunate ligament(sl) Most frequent cause of carpal instability Traumatic tears of SL ligament = most common ligament injury

More information

Dynamic CT Assessment of Distal Radioulnar Instability

Dynamic CT Assessment of Distal Radioulnar Instability Dynamic CT Assessment of Distal Radioulnar Instability Poster No.: P-0114 Congress: ESSR 2016 Type: Educational Poster Authors: S. Dumonteil, M. A. Shah, A. Srikanthan, V. Ejindu, N. Papadakos; London/UK

More information

Long-term Outcomes of Ulnar Shortening Osteotomy for Idiopathic Ulnar Impaction Syndrome: At Least 5-Years Follow-up

Long-term Outcomes of Ulnar Shortening Osteotomy for Idiopathic Ulnar Impaction Syndrome: At Least 5-Years Follow-up Original Article Clinics in Orthopedic Surgery 2011;3:295-301 http://dx.doi.org/10.4055/cios.2011.3.4.295 Long-term Outcomes of Ulnar Shortening Osteotomy for Idiopathic Ulnar Impaction Syndrome: At Least

More information

The functional importance of malunion in distal radius fractures

The functional importance of malunion in distal radius fractures The functional importance of malunion in distal radius fractures Nadine HOLLEVOET, René VERDONK The purpose of our study was to investigate which radiological parameter had the maximal detrimental effect

More information

RADIOGRAPHIC MORPHOMETRY OF DISTAL END OF RADIUS, A RETROSPECTIVE, OBSERVATIONAL STUDY IN INDIAN POPULATION

RADIOGRAPHIC MORPHOMETRY OF DISTAL END OF RADIUS, A RETROSPECTIVE, OBSERVATIONAL STUDY IN INDIAN POPULATION RADIOGRAPHIC MORPHOMETRY OF DISTAL END OF RADIUS, A RETROSPECTIVE, OBSERVATIONAL STUDY IN INDIAN POPULATION Research article Orthopaedics Deepak Tambe 1, Arvind Goregaonkar 2, Binoti Sheth 3, Jayesh Pawar

More information

Operative Indications

Operative Indications Distal Radius Fractures: Closed Treatment? Paul Tornetta III, MD Professor Boston Medical Center Boston Medical Center Operative Indications Age Initial films Dorsal comminution Intraarticular involvement

More information

NIH Public Access Author Manuscript J Hand Surg Am. Author manuscript; available in PMC 2014 January 01.

NIH Public Access Author Manuscript J Hand Surg Am. Author manuscript; available in PMC 2014 January 01. NIH Public Access Author Manuscript Published in final edited form as: J Hand Surg Am. 2013 January ; 38(1): 40 48. doi:10.1016/j.jhsa.2012.10.007. Radiographic Outcomes of Volar Locked Plating for Distal

More information

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology. Technique Guide 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology. Table of Contents Introduction 2.4 mm Variable Angle LCP

More information

An intra-bone axial load transducer: development and validation in an in-vitro radius model

An intra-bone axial load transducer: development and validation in an in-vitro radius model Knowles et al. Journal of Experimental Orthopaedics (2015) 2:19 DOI 10.1186/s40634-015-0035-z SHORT REPORT Open Access An intra-bone axial load transducer: development and validation in an in-vitro radius

More information

Effects of Malunited Fractured Distal End of Radius on the Morphometric Parameters of Distal Radioulnar Joint in Old Age Group

Effects of Malunited Fractured Distal End of Radius on the Morphometric Parameters of Distal Radioulnar Joint in Old Age Group Website: www.aijournals.com/journals/aanat/ DOI: 10.21276/aanat.2016.2.1.3 ORIGINAL ARTICLE ISSN: 2455-5274 Effects of Malunited Fractured Distal End of Radius on the Morphometric Parameters of Distal

More information

Distal radius fractures

Distal radius fractures Distal radius fractures One of the most common fractures seen by orthopaedic surgeons Incidence 195.2/100,000 persons per year Significant financial burden Distal radius fractures Improved implants ORIF

More information

Volar fixed-angle plating for distal radius fractures 刀圭会協立病院 津村敬

Volar fixed-angle plating for distal radius fractures 刀圭会協立病院 津村敬 Volar fixed-angle plating for distal radius fractures 刀圭会協立病院 津村敬 Normal distal radius anatomy Radial styloid process EPL Volar lip Lister s s tubercle Radial styloid process Sigmoid notch palmar view

More information

Ultrasound-guided reduction of distal radius fractures

Ultrasound-guided reduction of distal radius fractures American Journal of Emergency Medicine (2010) 28, 1002 1008 www.elsevier.com/locate/ajem Original Contribution Ultrasound-guided reduction of distal radius fractures Shiang-Hu Ang, Shu-Woan Lee, Kai-Yet

More information

Historical Perspective of Distal Radius Fracture Classifications in the Twentieth Century

Historical Perspective of Distal Radius Fracture Classifications in the Twentieth Century Short Commentary imedpub Journals http://www.imedpub.com/ Journal of Clinical & Experimental Orthopaedics DO: 10.4172/2471-8416.100026 Historical Perspective of Distal Radius Fracture Classifications in

More information

Indian Journal of Orthopaedics Surgery

Indian Journal of Orthopaedics Surgery 4 Indian Journal of Orthopaedics Surgery ISSN 2395-1354(Print) e-issn 2395-1362(Online) CLASSIFYING DISTAL RADIUS FRACTURES AND NEW YARDSTICK TO ORTHOPAEDIC RESIDENTS *Corresponding Author: Younis kamal

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle  holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/35777 holds various files of this Leiden University dissertation. Author: Wijffels, Mathieu Mathilde Eugene Title: The clinical and non-clinical aspects

More information

Fractures of the distal end of the radius should be

Fractures of the distal end of the radius should be FRACTURES OF THE DISTAL END OF THE RADIUS TREATED BY INTERNAL FIXATION AND EARLY FUNCTION A PRELIMINARY REPORT OF 20 CASES DANIEL A. RIKLI, PIETRO REGAZZONI From the University Hospital, Basel, Switzerland

More information

Kousuke Iba 1*, Yasuhiro Ozasa 1, Takuro Wada 1, Tomoaki Kamiya 1, Toshihiko Yamashita 1, Mitsuhiro Aoki 2. Abstract

Kousuke Iba 1*, Yasuhiro Ozasa 1, Takuro Wada 1, Tomoaki Kamiya 1, Toshihiko Yamashita 1, Mitsuhiro Aoki 2. Abstract RESEARCH ARTICLE Open Access Efficacy of radial styloid targeting screws in volar plate fixation of intra-articular distal radial fractures: a biomechanical study in a cadaver fracture model Kousuke Iba

More information

Mark VanDer Kaag 1, Ajmal Ikram 2. Hand Unit, Tygerberg Hospital University of Stellenbosch

Mark VanDer Kaag 1, Ajmal Ikram 2. Hand Unit, Tygerberg Hospital University of Stellenbosch A Prospective, Randomized Controlled Study To Determine The Radiological And Functional Outcomes Of IMN Fixation Of Distal Radius Fractures Using A Novel Device The Sonoma Wrx Distal Radius Nail Compared

More information

A Slightly Dorsally Tilted Lunate on MRI can be Considered Normal

A Slightly Dorsally Tilted Lunate on MRI can be Considered Normal )48( COPYRIGHT 016 BY THE ARCHIVES OF BONE AND JOINT SURGERY RESEARCH ARTICLE A Slightly Dorsally Tilted Lunate on MRI can be Considered Normal Anne-Carolin Döring, MD; Celeste L. Overbeek, MD; Teun Teunis,

More information

Management of intra-articular fractures of distal end radius in adults

Management of intra-articular fractures of distal end radius in adults International Journal of Research in Orthopaedics Gawali SR et al. Int J Res Orthop. 2016 Dec;2(4):220-228 http://www.ijoro.org Original Research Article DOI: http://dx.doi.org/10.18203/issn.2455-4510.intjresorthop20163148

More information

UNFAVORABLE RESULTS AND degenerative arthritis. Arthroscopically Guided Osteotomy for Management of Intra-Articular Distal Radius Malunions

UNFAVORABLE RESULTS AND degenerative arthritis. Arthroscopically Guided Osteotomy for Management of Intra-Articular Distal Radius Malunions SCIENTIFIC ARTICLE Arthroscopically Guided Osteotomy for Management of Intra-Articular Distal Radius Malunions Francisco del Piñal, MD, PhD, Leopoldo Cagigal, MD, Francisco J. García-Bernal, MD, PhD, Alexis

More information

Arthroplasty for advanced Kienböck s disease using a radial bone flap with a vascularised wrapping of pronator quadratus

Arthroplasty for advanced Kienböck s disease using a radial bone flap with a vascularised wrapping of pronator quadratus Arthroplasty for advanced Kienböck s disease using a radial bone flap with a vascularised wrapping of pronator quadratus H. S. Gong, M. S. Chung, Y. H. Lee, S. Lee, J. O. Lee, G. H. Baek From Seoul National

More information

Incidence 643,000 per year United States 15% of all fractures, peak occurrence age 60 70, Classification - AO/OTA (Murray 2013)

Incidence 643,000 per year United States 15% of all fractures, peak occurrence age 60 70, Classification - AO/OTA (Murray 2013) Treatment of Distal Radial Fractures John M. Bednar, M.D. Associate Clinical Professor Orthopaedic Surgery, Thomas Jefferson University School of Medicine and The Philadelphia Hand Center Incidence 643,000

More information

Clinically Relevant. Intra-articular Fractures of the Distal End of the Radius in

Clinically Relevant. Intra-articular Fractures of the Distal End of the Radius in Intra-Articular Fractures of the Distal End of the Radius in Young Adults : Scientifically Flawed but Clinically Relevant Brian M. Haus, M.D., Jesse B. Jupiter, M.D. INTRODUCTION Knirk and Jupiter s 1986

More information

Small External Fixator Wrist Spanning Frame. For the treatment of wrist fractures.

Small External Fixator Wrist Spanning Frame. For the treatment of wrist fractures. Small External Fixator Wrist Spanning Frame. For the treatment of wrist fractures. Technique Guide Part of the Small External Fixation System Small External Fixator Wrist Spanning Frame When to use The

More information

Disclosures. Distal Radius Fractures 5/16/2017. Distal Radius Fractures: Complications & Limitations of the Volar Approach

Disclosures. Distal Radius Fractures 5/16/2017. Distal Radius Fractures: Complications & Limitations of the Volar Approach Distal Radius Fractures: Complications & Limitations of the Volar Approach Frank R. Avilucea, MD Assistant Professor Department of Orthopaedic Surgery University of Cincinnati Medical Center Disclosures

More information

6/5/2018. DISCLOSURES Hassan R. Mir, MD, MBA, FACS. Evolution of Distal Radius Fracture Treatment [Chung Hand Clinics 2012]

6/5/2018. DISCLOSURES Hassan R. Mir, MD, MBA, FACS. Evolution of Distal Radius Fracture Treatment [Chung Hand Clinics 2012] DISCLOSURES Hassan R. Mir, MD, MBA, FACS Paid Consultant for a Company or Supplier Smith & Nephew Zimmer Biomet Trice Medical Stock or Stock Options Core Orthopaedics OrthoGrid Systems Research Support

More information

RADIOGRAPHY OF THE HAND, FINGERS & THUMB

RADIOGRAPHY OF THE HAND, FINGERS & THUMB RADIOGRAPHY OF THE HAND, FINGERS & THUMB FINGERS (2nd 5th) - PA Projection Patient Position: Seated; hand ; elbow on IR table top Part Position: Fingers centered to IR unless protocol is Central Ray: Perpendicular

More information

Complications of Distal Radius Fractures. How to Treat a Distal Radius Fx 11/13/2017. Michael S. Bednar, M.D. Loyola University Chicago

Complications of Distal Radius Fractures. How to Treat a Distal Radius Fx 11/13/2017. Michael S. Bednar, M.D. Loyola University Chicago Complications of Distal Radius Fractures Michael S. Bednar, M.D. Loyola University Chicago How to Treat a Distal Radius Fx Need to restore motion, begin with uninvolved parts Need to reduce an unreduced

More information

Fracture of the Distal Radius: Epidemiology and Premanagement Radiographic Characterization

Fracture of the Distal Radius: Epidemiology and Premanagement Radiographic Characterization Musculoskeletal Imaging Review Porrino et al. Imaging Fractures of the Distal Radius Musculoskeletal Imaging Review Downloaded from www.ajronline.org by 46.3.193.83 on 01/17/18 from IP address 46.3.193.83.

More information

Index. Note: Page numbers of article titles are in boldface type. Hand Clin 21 (2005)

Index. Note: Page numbers of article titles are in boldface type. Hand Clin 21 (2005) Hand Clin 21 (2005) 501 505 Index Note: Page numbers of article titles are in boldface type. A Antibiotics, following distal radius fracture treatment, 295, 296 Arthritis, following malunion of distal

More information

RADIOGRAPHY OF THE WRIST

RADIOGRAPHY OF THE WRIST RADIOGRAPHY OF THE WRIST Patient Position: WRIST PA Projection, elbow in same plane Part Position: Hand ; fingers centered to IR Central Ray: Structures Shown: NOTE: Optional AP projection best demonstrates

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle   holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/35777 holds various files of this Leiden University dissertation. Author: Wijffels, Mathieu Mathilde Eugene Title: The clinical and non-clinical aspects

More information

ORIGINAL PAPER. Department of Hand Surgery, Nagoya University School of Medicine ABSTRACT

ORIGINAL PAPER. Department of Hand Surgery, Nagoya University School of Medicine ABSTRACT Nagoya J. Med. Sci. 74. 167 ~ 171 2012 ORIGINAL PAPER TILT OF THE RADIUS FROM FOREARM ROTATIONAL AXIS RELIABLY PREDICTS ROTATIONAL IMPROVEMENT AFTER CORRECTIVE OSTEOTOMY FOR MALUNITED FOREARM FRACTURES

More information

A study of the radiographic morphometry of the distal radius in a south Indian population

A study of the radiographic morphometry of the distal radius in a south Indian population Original Article Hand Microsurg 2018;7:9-15 doi:10.5455/handmicrosurg.285986 A study of the radiographic morphometry of the distal radius in a south Indian population Supreeth Nekkanti, Jacqueline Shah,

More information

Radiographic observation of the scaphoid shift test

Radiographic observation of the scaphoid shift test Radiographic observation of the scaphoid shift test Min Jong Park From Sungkyunkwan University School of Medicine, Seoul, Korea T he movements of the carpal bones during the scaphoid shift test were evaluated

More information

Reliability of radiographic measurements for acute distal radius fractures

Reliability of radiographic measurements for acute distal radius fractures Watson et al. BMC Medical Imaging (2016) 16:44 DOI 10.1186/s12880-016-0147-7 RESEARCH ARTICLE Open Access Reliability of radiographic measurements for acute distal radius fractures Narelle J. Watson 1*,

More information

Interesting Case Series. Perilunate Dislocation

Interesting Case Series. Perilunate Dislocation Interesting Case Series Perilunate Dislocation Tom Reisler, BSc (Hons), MB ChB, MRCS (Ed), Paul J. Therattil, MD, and Edward S. Lee, MD Division of Plastic and Reconstructive Surgery, Department of Surgery,

More information

Comparison between Distractor Application on Both Radial & Ulnar Side and Radial Side Only for Fracture Distal Radius with Ulnar Styloid Fracture

Comparison between Distractor Application on Both Radial & Ulnar Side and Radial Side Only for Fracture Distal Radius with Ulnar Styloid Fracture Open Journal of Orthopedics, 2013, 3, 227-233 http://dx.doi.org/10.4236/ojo.2013.35043 Published Online September 2013 (http://www.scirp.org/journal/ojo) 227 Comparison between Distractor Application on

More information

Wrist Fixation System

Wrist Fixation System Wrist Fixation System Anatomy / Fracture Implant EXTRA & SIMPLE ARTICULAR Volar Radius Volar Fixed Angle Plate Volar Bearing Plate Radial Peg Plate Volar Hook Plate Volar Buttress Pin Volar Shear Plate

More information

E ORIGINAL ARTICLE Low extra-articular (transcondylar) fractures of the distal humerus

E ORIGINAL ARTICLE Low extra-articular (transcondylar) fractures of the distal humerus Shoulder & Elbow. ISSN 1758-5732 E ORIGINAL ARTICLE Low extra-articular (transcondylar) fractures of the distal humerus Alexander A. Weening, Kim M. Brouwer, Margaritha Adams & David Ring Orthopaedic Hand

More information

Sean Walsh Orthopaedic Surgeon Dorset County Hospital

Sean Walsh Orthopaedic Surgeon Dorset County Hospital Sean Walsh Orthopaedic Surgeon Dorset County Hospital Shapes and orientation of articular surfaces Ligaments Oblique positioning of scaphoid Tendons surrounding the joints Other soft tissues Peripheral

More information

This presentation is the intellectual property of the author. Contact them for permission to reprint and/or distribute.

This presentation is the intellectual property of the author. Contact them for permission to reprint and/or distribute. The Stiff Hand: Manual Therapy Sylvia Dávila, PT, CHT San Antonio, Texas Orthopedic Manual Therapy Common Applications Passive stretch Tensile force to tissue to increase extensibility of length & ROM

More information

Outcome of Kienböck s Disease 22 Years after Distal Radius Shortening Osteotomy

Outcome of Kienböck s Disease 22 Years after Distal Radius Shortening Osteotomy CLINICAL ORTHOPAEDICS AND RELATED RESEARCH Number 000, pp. 000 000 2007 Lippincott Williams & Wilkins Outcome of Kienböck s Disease 22 Years after Distal Radius Shortening Osteotomy E. E. J. Raven, MD

More information

Scaphoid Fractures. Mohammed Alasmari. Orthopaedic Surgery Demonstrator Majmaah University

Scaphoid Fractures. Mohammed Alasmari. Orthopaedic Surgery Demonstrator Majmaah University Scaphoid Fractures Mohammed Alasmari Orthopaedic Surgery Demonstrator Majmaah University 1 2 Scaphoid Fractures Introduction Anatomy History Clinical examination Radiographic evaluation Classification

More information

Common. Common Hand Problems in Elite Athletes

Common. Common Hand Problems in Elite Athletes Common Hand Problems in Elite Athletes Fred Corley M.D. Dept. of Orthopaedic Surgery UTHSCSA I have no disclosures concerning this talk. The University of Texas Health Science Center @ San Antonio - Orthopaedics

More information

MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS FOR DISTAL RADIUS FRACTURES: SURGICAL TECHNIQUE M. TOBE 1, K. MIZUTANI 1, Y. TSUBUKU 1, Y.

MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS FOR DISTAL RADIUS FRACTURES: SURGICAL TECHNIQUE M. TOBE 1, K. MIZUTANI 1, Y. TSUBUKU 1, Y. Riv Chir Mano - Vol. 43 (3) 2006 MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS FOR DISTAL RADIUS FRACTURES: SURGICAL TECHNIQUE M. TOBE 1, K. MIZUTANI 1, Y. TSUBUKU 1, Y. YANAGIHARA 2 1 Department of 2nd Orthopaedic

More information

Common Limb Fractures. Mr Sheraz Malik MB BS MRCS Instructor Mr Paul Ofori-Atta Mb ChB FRCS President Motc Life UK April 2009

Common Limb Fractures. Mr Sheraz Malik MB BS MRCS Instructor Mr Paul Ofori-Atta Mb ChB FRCS President Motc Life UK April 2009 Common Limb Fractures Mr Sheraz Malik MB BS MRCS Instructor Mr Paul Ofori-Atta Mb ChB FRCS President Motc Life UK April 2009 Objectives To be able to describe all characteristics of a fracture Describe

More information

Clinical Study Rate of Improvement following Volar Plate Open Reduction and Internal Fixation of Distal Radius Fractures

Clinical Study Rate of Improvement following Volar Plate Open Reduction and Internal Fixation of Distal Radius Fractures SAGE-Hindawi Access to Research Advances in Orthopedics Volume 2011, Article ID 565642, 4 pages doi:10.4061/2011/565642 Clinical Study Rate of Improvement following Volar Plate Open Reduction and Internal

More information

IS CASTING AN ACCEPTABLE TREATMENT APPROACH FOR A DISTAL RADIUS FRACTURE THAT HAS UNDERGONE A SATISFACTORY CLOSED REDUCTION?

IS CASTING AN ACCEPTABLE TREATMENT APPROACH FOR A DISTAL RADIUS FRACTURE THAT HAS UNDERGONE A SATISFACTORY CLOSED REDUCTION? Original Article FOR A DISTAL RADIUS FRACTURE THAT HAS UNDERGONE A SATISFACTORY CLOSED REDUCTION? Khalid Masood *, Muhammad Jamil **, Farooq Ibrahim Pasha*, Khalid Zulfiqar Quresshi **, Naveed Ishaq Malik

More information

A Patient s Guide to Adult Distal Radius (Wrist) Fractures

A Patient s Guide to Adult Distal Radius (Wrist) Fractures A Patient s Guide to Adult Distal Radius (Wrist) Fractures Suite 11-13/14/15 Mount Elizabeth Medical Center 3 Mount Elizabeth Singapore, 228510 Phone: (65) 6738 2628 Fax: (65) 6738 2629 1 DISCLAIMER: The

More information

University of Groningen. Fracture of the distal radius Oskam, Jacob

University of Groningen. Fracture of the distal radius Oskam, Jacob University of Groningen Fracture of the distal radius Oskam, Jacob IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document

More information

Index. Springer International Publishing Switzerland 2016 J.N. Lawton (ed.), Distal Radius Fractures, DOI /

Index. Springer International Publishing Switzerland 2016 J.N. Lawton (ed.), Distal Radius Fractures, DOI / Index A AAOS. See American Academy of Orthopaedic Surgeons (AAOS) Abductor pollicis longus (APL) tendon, 34, 73 Acute carpal tunnel syndrome, 93 American Academy of Orthopaedic Surgeons (AAOS), 66, 238

More information

Characteristics of two different locking compression plates in the volar fixation of complex articular distal radius fractures

Characteristics of two different locking compression plates in the volar fixation of complex articular distal radius fractures Characteristics of two different locking compression plates in the volar fixation of complex articular distal radius fractures The Harvard community has made this article openly available. Please share

More information

University of Groningen. Fracture of the distal radius Oskam, Jacob

University of Groningen. Fracture of the distal radius Oskam, Jacob University of Groningen Fracture of the distal radius Oskam, Jacob IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document

More information

CASE REPORT. Distal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report

CASE REPORT. Distal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report Nagoya J. Med. Sci. 79. 551 ~ 557, 2017 doi:10.18999/nagjms.79.4.551 CASE REPORT Distal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report Takaaki Shinohara 1

More information

Reliability of Lichtman s classification for Kienböck s disease in 99 subjects

Reliability of Lichtman s classification for Kienböck s disease in 99 subjects Reliability of Lichtman s classification for Kienböck s disease in subjects Masaki Shin, M.D., Masahiro Tatebe, M.D., Hitoshi Hirata, M.D., Shukuki Koh, M.D., Takaaki Shinohara, M.D. Department of Hand

More information

Instrument and Implant for wrist fracture

Instrument and Implant for wrist fracture Instrument and Implant for wrist fracture Jansri Janpanya Product specialist The Bangkok Unitrade Co,.ltd. Objectives Type of LCP for distal radius Fx. The new LCP design for distal radius Fx. Have knowledge

More information

5/9/2017. Distal Radius Fractures: What s New? What s New? Or Maybe New to you. Single Pak Sterile distal radius kits

5/9/2017. Distal Radius Fractures: What s New? What s New? Or Maybe New to you. Single Pak Sterile distal radius kits Distal Radius Fractures: What s New? What s New? Or Maybe New to you. Single Pak Sterile distal radius kits Extension of FCR split to release tendon for retraction Pronator Quadratus elevation with BR

More information

University of Groningen. Fracture of the distal radius Oskam, Jacob

University of Groningen. Fracture of the distal radius Oskam, Jacob University of Groningen Fracture of the distal radius Oskam, Jacob IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document

More information

Characterizing scaphoid nonunion deformity using 2-D and 3-D imaging techniques ten Berg, P.W.L.

Characterizing scaphoid nonunion deformity using 2-D and 3-D imaging techniques ten Berg, P.W.L. UvA-DARE (Digital Academic Repository) Characterizing scaphoid nonunion deformity using 2-D and 3-D imaging techniques ten Berg, P.W.L. Link to publication Citation for published version (APA): ten Berg,

More information

QUICK REFERENCE GUIDE. The Pennig Dynamic Wrist Fixator. Part A: Trans-articular application

QUICK REFERENCE GUIDE. The Pennig Dynamic Wrist Fixator. Part A: Trans-articular application 10 The Pennig Dynamic Wrist Fixator Part A: Trans-articular application B1 B2 B3 III IV TRANS-ARTICULAR APPLICATION The fractures that can be treated with this technique include AO type B and C fractures,

More information

Acomparison of percutaneous and pin-and-plaster techniques in distal radius fracture

Acomparison of percutaneous and pin-and-plaster techniques in distal radius fracture Original Research Medical Journal of the Islamic Republic of Iran.Vol. 22, No. 4, February, 2009. pp. 159-163 Acomparison of percutaneous and pin-and-plaster techniques in distal radius fracture Farshid

More information

Disclosure. 85% Dorsal. 15% Volar. Distal Radius Fractures- Volar Plating for All. Skeletal Dynamics- Consultant

Disclosure. 85% Dorsal. 15% Volar. Distal Radius Fractures- Volar Plating for All. Skeletal Dynamics- Consultant Distal Radius Fractures- Volar Plating for All Frontiers in Upper Extremity Surgery 2017 Francisco Rubio, M.D. Miami Hand and Upper Extremity Institute Miami, FL Disclosure Skeletal Dynamics- Consultant

More information

Fractures of the distal radius have occupied a storied history

Fractures of the distal radius have occupied a storied history Distal Radius Fractures: What Cannot Be Fixed With a Volar Plate? The Role of Fragment-Specific Fixation in Modern Fracture Treatment Jonathan Lam, MD, PhD,* and Scott W. Wolfe, MD, The advent of volar

More information

GALEAZZI FRACTURE. Galeazzi fracture-dislocations can be difficult to recognize and are often not initially appreciated.

GALEAZZI FRACTURE. Galeazzi fracture-dislocations can be difficult to recognize and are often not initially appreciated. GALEAZZI FRACTURE Introduction In the Galeazzi fracture-dislocation there is a fracture of the distal third of the shaft of the radius in association with a subluxation or dislocation of the distal radio-ulna

More information

The Role of Lunate Morphology on Scapholunate Instability and Fracture Location in Patients Treated for Scaphoid Nonunion

The Role of Lunate Morphology on Scapholunate Instability and Fracture Location in Patients Treated for Scaphoid Nonunion Original Article Clinics in Orthopedic Surgery 2016;8:175-180 http://dx.doi.org/10.4055/cios.2016.8.2.175 The Role of Lunate Morphology on Scapholunate Instability and Fracture Location in Patients Treated

More information

COMPARATIVE STUDY OF FUNCTIONAL OUTCOME OF EXTERNAL AND INTERNAL FIXATION IN TREATMENT OF COMMINUTED DISTAL RADIUS FRACTURES

COMPARATIVE STUDY OF FUNCTIONAL OUTCOME OF EXTERNAL AND INTERNAL FIXATION IN TREATMENT OF COMMINUTED DISTAL RADIUS FRACTURES COMPARATIVE STUDY OF FUNCTIONAL OUTCOME OF EXTERNAL AND INTERNAL FIXATION IN TREATMENT OF COMMINUTED DISTAL RADIUS FRACTURES R. Sahaya Jose 1 1Assistant Professor, Department of Orthopaedics, Sree Mookambika

More information

Redisplacement of Distal Radius Fracture after Initial Closed Reduction: Analysis of Prognostic Factors

Redisplacement of Distal Radius Fracture after Initial Closed Reduction: Analysis of Prognostic Factors Original Article Clinics in Orthopedic Surgery 2015;7:377-382 http://dx.doi.org/10.4055/cios.2015.7.3.377 Redisplacement of Distal Radius Fracture after Initial Closed Reduction: Analysis of Prognostic

More information

JMSCR Vol 05 Issue 04 Page April 2017

JMSCR Vol 05 Issue 04 Page April 2017 www.jmscr.igmpublication.org Impact Factor 5. Index Copernicus Value: 3.7 ISSN (e)-37-17x ISSN (p) 55-5 DOI: https://dx.doi.org/1.1535/jmscr/v5i.1 Functional Outcome after Surgical Stabilization of Fractures

More information

Anatomical Variations of the Extensor Carpi Ulnaris Groove: A New Computed Tomography-based Evaluation

Anatomical Variations of the Extensor Carpi Ulnaris Groove: A New Computed Tomography-based Evaluation Original Article The Journal of Hand Surgery (Asian-Pacific Volume) 2017;22(3):281-285 DOI: 10.1142/S0218810417500307 Anatomical Variations of the Extensor Carpi Ulnaris Groove: A New Computed Tomography-based

More information

Small External Fixator Nonspanning Wrist Frame. For the treatment of wrist fractures.

Small External Fixator Nonspanning Wrist Frame. For the treatment of wrist fractures. Small External Fixator Nonspanning Wrist Frame. For the treatment of wrist fractures. Technique Guide Part of the Small External Fixation System Small External Fixator Nonspanning Wrist Frame When to use

More information

Bilateral Trans-Scaphoid Perilunate Fracture Dislocation

Bilateral Trans-Scaphoid Perilunate Fracture Dislocation CASE REPORT pissn 1598-3889 eissn 2234-0998 J Korean Soc Surg Hand 2015;20(3):127-132. http://dx.doi.org/10.12790/jkssh.2015.20.3.127 JOURNAL OF THE KOREAN SOCIETY FOR SURGERY OF THE HAND Bilateral Trans-Scaphoid

More information

CLASSIFYING RADIUS FRACTURES WITH X-RAY AND TOMOGRAPHY IMAGING

CLASSIFYING RADIUS FRACTURES WITH X-RAY AND TOMOGRAPHY IMAGING Original Article CLASSIFYING RADIUS FRACTURES WITH X-RAY AND TOMOGRAPHY IMAGING Paulo Roberto Miziara Yunes Filho, Miguel Viana Pereira Filho, Fabiano Cortese Paula Gomes, Rodrigo Serikawa de Medeiros,

More information

Hands PA; Obl. Lat.; Norgaard s Thumb AP; Lat. PA. PA; Lat.: Obls.; Elongated PA with ulnar deviation

Hands PA; Obl. Lat.; Norgaard s Thumb AP; Lat. PA. PA; Lat.: Obls.; Elongated PA with ulnar deviation Projections Region Basic projections Additional / Modified projections Upper Limbs Hands PA; Obl. Lat.; Norgaard s Thumb ; Lat. PA Fingers PA; Lat. Wrist PA; Lat. Obls. Scaphoid Lunate Trapezium Triquetral

More information

University of Groningen. Osteoporosis, identification and treatment in fracture patients de Klerk, Gijsbert

University of Groningen. Osteoporosis, identification and treatment in fracture patients de Klerk, Gijsbert University of Groningen Osteoporosis, identification and treatment in fracture patients de Klerk, Gijsbert IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish

More information

Original Article Clinics in Orthopedic Surgery 2018;10:

Original Article Clinics in Orthopedic Surgery 2018;10: Original rticle Clinics in Orthopedic Surgery 2018;10:80-88 https://doi.org/10.4055/cios.2018.10.1.80 Where Is the r Styloid Process? Identification of the bsolute Location of the r Styloid Process ased

More information

SCAHPO-LUNATE DISSOCIATION

SCAHPO-LUNATE DISSOCIATION SCAHPO-LUNATE DISSOCIATION Introduction Scapho-lunate dissociation is the most common significant ligamentous injury of the wrist. The condition is also sometimes referred to as rotary subluxation of the

More information

I-A-1) Non-specific thickening of synovial membrane

I-A-1) Non-specific thickening of synovial membrane I-A-1) Non-specific thickening of synovial membrane Grayscale Metatarsal Power Doppler Dorsal aspect of metatarsophalangeal joint in right 1 st toe, longitudinal view Asterisks indicate non-specific thickening

More information

Carpal rows injuries!

Carpal rows injuries! Carpal rows injuries! Michael Papaloïzos! Center for Hand Surgery and Therapy Geneva, Switzerland no conflict of interest to declare Fractures of carpal bones! The fractured scaphoid! Fracture-dislocations

More information

Degrees Of Volar Angulation In Distal Radius Fracture Effects Distal Radioulnar Joint Stability:a Biomechanical Study

Degrees Of Volar Angulation In Distal Radius Fracture Effects Distal Radioulnar Joint Stability:a Biomechanical Study Degrees Of Volar Angulation In Distal Radius Fracture Effects Distal Radioulnar Joint Stability:a Biomechanical Study Yuki Bessho, MD, Toshiyasu Nakamura, MD PhD, Takeo Nagura, MD PhD, Yoshimori Kiriyama,

More information

Comparison of Below The Elbow Cast with Above The Elbow Cast in Treating Distal Third Forearm Fractures in Children

Comparison of Below The Elbow Cast with Above The Elbow Cast in Treating Distal Third Forearm Fractures in Children ORIGINAL ARTICLE ABSTRACT Comparison of Below The Elbow Cast with Above The Elbow Cast in Treating Distal Third Forearm Fractures in Children Noor Rahman, Wasim Anwar, Malik Javed Iqbal, Israr Ahmad, Mohammad

More information

ComparisonofUlnar-ShorteningOsteotomyWitha NewTrimedDynamicCompressionSystemVersusthe SynthesDynamicCompressionSystem:ClinicalStudy

ComparisonofUlnar-ShorteningOsteotomyWitha NewTrimedDynamicCompressionSystemVersusthe SynthesDynamicCompressionSystem:ClinicalStudy SCIENTIFICARTICLE ComparisonofUlnar-ShorteningOsteotomyWitha NewTrimedDynamicCompressionSystemVersusthe SynthesDynamicCompressionSystem:ClinicalStudy ShaiLuria,MD,AnthonyJ.Lauder,MD,ThomasE.Trumble,MD

More information

Standardised post-operative radiographs for volar radial plate fixation: the 22/11 X-ray

Standardised post-operative radiographs for volar radial plate fixation: the 22/11 X-ray Matthee W, et al. SAOJ 2017;16(3) DOI 10.17159/2309-8309/2017/v16n3a9 South African Orthopaedic Journal http://journal.saoa.org.za UPPER LIMB Standardised post-operative radiographs for volar radial plate

More information

Ulna Shortening Osteotomy Using a Compression Device

Ulna Shortening Osteotomy Using a Compression Device Ulna Shortening Osteotomy Using a Compression Device Neal C. Chen, MD, Boston, MA, Scott W. Wolfe, MD, New York, NY Purpose: To report the functional and radiographic outcomes of a cohort of patients treated

More information

We have assessed the influence of isolated and

We have assessed the influence of isolated and The effect of rotational malunion of the radius and the ulna on supination and pronation AN EXPERIMENTAL INVESTIGATION C. E. Dumont, R. Thalmann, J. C. Macy From the University of Zürich, Switzerland We

More information

ORIGINAL ARTICLE TREATMENT DISTAL RADIUS FRACTURE WITH VOLAR BUTTRESS TECHNIQUE- A CLINICAL STUDY

ORIGINAL ARTICLE TREATMENT DISTAL RADIUS FRACTURE WITH VOLAR BUTTRESS TECHNIQUE- A CLINICAL STUDY TREATMENT DISTAL RADIUS FRACTURE WITH VOLAR BUTTRESS TECHNIQUE- A CLINICAL STUDY Neelanagowda V P Patil 1, Lingaraj 2, P S Kaladagi 3, Paramanda Hospeti 4, Nizamuddin 5. 1. Assistant professor, Department

More information