White Paper for Rotational Stability. *smith&nephew SL-PLUS MIA. Cementless Hip Stem System

Size: px
Start display at page:

Download "White Paper for Rotational Stability. *smith&nephew SL-PLUS MIA. Cementless Hip Stem System"

Transcription

1 White Paper for Rotational Stability *smith&nephew SL-PLUS MIA Cementless Hip Stem System

2 SL-PLUS MIA Building on the long-term success of the SL-PLUS hip stem, the SL-PLUS MIA hip stem has been developed for minimally invasive surgical procedures. The design and surgical technique of this stem allow implantation with reduced lateral bone resection, preserving the tendons on the greater trochanter. The anchoring technique of the modified, dual-taper straight stem in the region of the lateral shoulder in the diaphysis is fully maintained (fig. 1, 2). The aim of this paper is to investigate the impact of the modified shoulder on the rotational stability of the SL-PLUS MIA Standard stem compared with the SL-PLUS Standard stem. Due to the complex geometric conditions and biomechanics, the finite element (FEM) method is frequently utilized (3, 4, 5, 6, 7) to calculate the load distribution, displacement and the system performance in general. This method is particularly suitable for comparative calculations to evaluate various designs. 02

3 SL-PLUS MIA Materials and methods This analysis compares the SL-PLUS Standard stem size 6 and the SL-PLUS MIA Standard stem size 6 in terms of their primary rotational stability using FEM calculations. A numerical 3D model of the femur was used to represent the bone. The cortical and cancellous bone were modeled separately (please refer to table 1 for material properties) and the hip stems to be studied were virtually implanted into the model bone. In both cases the size 6 stem achieved good cortical contact in the virtually rasped bony bed of the distal femur (fig. 1C) thereby reflecting the immediate postoperative situation. Cortical fixation of the stem A B C Fig. 1: Implanted hip stem in the bone model SL-PLUS MIA Standard (A), and SL-PLUS Standard (B). In the semi-transparent model the cancellous bone is shown in orange. Figure C shows a cross section through the femur. The SL-PLUS MIA Standard stem is less prominent in the proximal/lateral femur than the SL-PLUS Standard stem (fig. 1). In accordance with the clinical procedure, care was also taken to ensure that both stems were primarily fixed into the cortical region of the diaphysis. Due to the different implant techniques, the greater trochanter major is weakened to a greater extent with the SL-PLUS Standard stem than with the SL-PLUS MIA Standard stem (fig. 2). Fig. 2: Proximal view of the femur with the SL-PLUS MIA Standard stem (left) and the SL-PLUS Standard stem (right). The difference in the trochanter weakening caused by the implantation techniques can be clearly seen. 03

4 The 3D geometries were imported into a finite element program for further processing and loading (pre- and post-processing program MSC Patran, network with approximately tetrahedron elements). Friction at the implant/bone surfaces of the anchoring was established with a friction coefficient of 0.5. Elasticity module Poisson ration Element types Source Cortical bone N/mm Tet-4, 4 mm [1] Cancellous bone N/mm Tet-4, 2 mm [1] Titanium N/mm Tet-4, 1 mm [2] Table 1: Material properties for the FEM calculation. Based on Bergmann et al. [8] a torsional moment of approximately 16 Nm was applied as a load along the longitudinal axis of each stem. The bone was secured distally at the isthmus. The displacement of the ball head center, representing rotational stability in this model, was calculated, as was the interface load transmission in the region of the implant/bone interface. Mz = 16 Nm A B C Fig. 3: FEM models of the femora with SL-PLUS MIA Standard (A), and SL-PLUS Standard (B). A torsional moment (C) was applied along the longitudinal axis of the stem. 04

5 SL-PLUS MIA Results 1. Displacement of the overall system The maximum displacement values of the ball head center were approximately mm for the SL-PLUS MIA Standard stem and approximately mm for the SL-PLUS Standard stem (differ- ence = 0.5%). 2. Interface load transmission The stress distribution represented in the transverse levels along the prosthesis stem shows that the load in the diaphyseal cortical anchoring for both stems leads to higher distal forces (fig. 4). In the proximal region, cortical loading is reduced for both stem designs when compared to the diaphysis. Section 9 Section 9 Section 1 Section 1 Fig. 4: Stress distribution in the implant (SL-PLUS MIA Standard left, SL-PLUS Standard right) and bone, as seen in the transverse plane in the anchorage zones. High stress = red, low stress = blue. Going from distal to proximal the pattern of stress distribution is virtually identical for both stems. The maximum stress below the greater trochanter is marginally higher with the SL-PLUS Standard stem. The volume of cancellous bone in the trochanter only slightly affects the rotational stability of the SL-PLUS Standard stem in levels 6 9 in comparison to the stress in the cortical levels 1 5. SL-PLUS MIA Standard SL-PLUS Standard Distribution of friction Distal to proximal cross section SL-PLUS MIA Standard SL-PLUS Standard Fig. 5: Distribution of stress along the lateral edge for the SL-PLUS MIA stem and the SL-PLUS Standard stem. 05

6 Discussion In this investigation a torsional moment was applied along the longitudinal axis of the stems for an isolated observation of rotational stability. This does not represent the actual physiological load situation but is well suited to allow an isolated evaluation of the impact of the design differences on the rotation stability and thus on the primary stability. Both stem models, SL-PLUS MIA Standard and SL-PLUS Standard, were identically and neutrally positioned in the FEM model. In both models the lateral cortical contact ends in the region of the trochanter, representing an average clinical situation. The virtually identical values for the ball head displacement taken as a measure of the maximum total deformation of the bone implant system show that, in terms of the displacement caused by rotation, the SL-PLUS MIA Standard stem is equivalent to the SL-PLUS Standard stem. The slightly lower displacement of the SL-PLUS MIA Standard stem compared with the SL-PLUS Standard stem (0.5%) can be attributed to the reduced weakening of the greater trochanter, leading to greater rigidity in this region. Both systems show a primarily distally based load transmission to the cortical bone. The stress represented as a measure of the anchoring forces is greater in the distal cortical bone than in the trochanter region. In the SL-PLUS Standard stem these forces fall steadily above the second-lowest cross hole. Further, in the proximal region the contact is primarily between the cancellous bone and the implant, where the forces are lower due to the mechanical properties of the bone. The lateral shoulder of the SL-PLUS Standard stem, which has been eliminated in the SL-PLUS MIA Standard stem in favor of an implantation technique that preserves more tissue, has a relatively low impact on the rotational stability, as it is mainly in the cancellous bone. This explains why, without the lateral shoulder of the SL-PLUS Standard stem, the SL-PLUS MIA Standard stem is very similar to the SL-PLUS Standard stem in terms of its rotational stability under application of a torsional moment. Finally, it should be noted that, as a result of the different implantation techniques, the greater trochanter is weakened to a greater extent with the SL-PLUS Standard stem than with the SL-PLUS MIA Standard stem. As the greater trochanter is not weakened to the same extent as with the implantation of the SL-PLUS Standard stem, this seems to compensate for the anticipated loss of rotational stability resulting from the design of the SL-PLUS MIA Standard stem. Both bone-implant systems show identical values for deflection during rotational loading on its own. 06

7 SL-PLUS MIA References A. Rotem Effect of implant material properties on the performance of a hip joint replacement. J of Medical Engineering, Vol 18, Nr 6, p , Taylor&Francis 1994 M. Ungethüm, W. Winkler Metallische Werkstoffe in der Orthopädie und Unfallchirurgie Georg Thieme Verlag Stuttgart, New York, 1984 P. J. Prendergast, D. Taylor Stress analysis of the proximal-medial femur after total hip replacement Journal of Biomedical Engineering, Vol. 12, p Butterworth & Co Ltd, 1990 T. Vail, R. Richard, D. Theodosis The effect of hip stem material modulus on surface strains in human femora Journal of Biomechanics, Vol. 31, p , 1998 M. Akay, N. Aslan Numerical and experimental stress analysis of a polymeric composite hip joint prosthesis Journal of Biomedical Materials Research, Vol. 31, p John Wiley & Sons Inc E. Savidis, F. Loer, O. Werner The significance of the torque loading of the total hip prosthesis Biomechanics, Basic and Applied Research, p G. Bergmann, R. Kölbel, A. Rohlmann Martinus Nijhoff Publishers, Dordrecht, Boston, Lancaster, 1987 R. R. Tarr, I. Clarke, T. Gruen Predictions of bone cement failure criteria: Three dimensional finite element Models versus clinical reality of total hip replacement Finite Elements in Biomechanics, p P. C. Johnson, J. F. Gross, J. Wiley & Sons Inc., New York, 1986 G. Bergmann et. Al. Hip contact forces and gait patterns from routine activities Journal of Biomechanics Vol. 34, p ,

8 Manufaturer Smith & Nephew Orthopaedics AG Erlenstrasse 4a 6343 Rotkreuz Schweiz Contactntakt Für weitere Informationen kontaktieren Sie unser lokales Verkaufsbüro. Trademark of Smith & Nephew Lit. No e Ed. 04/ Ex. 05/

*smith&nephew SL-PLUS Cementless Femoral Hip System. Product Information

*smith&nephew SL-PLUS Cementless Femoral Hip System. Product Information Product Information *smith&nephew SL-PLUS Cementless Femoral Hip System First Came the Philosophy to develop a universal hip system that could be used in almost every indication, immaterial to the patient

More information

*smith&nephew SBG Anatomical Hip Stem Prosthesis. Product Information

*smith&nephew SBG Anatomical Hip Stem Prosthesis. Product Information Product Information *smith&nephew SBG Anatomical Hip Stem Prosthesis For a Perfect Fit Because Anatomy Matters On account of its anatomical shape the cementless SBG stem, which has been used successfully

More information

*smith&nephew SL-PLUS

*smith&nephew SL-PLUS Surgical Technique *smith&nephew SL-PLUS Cementless Femoral Hip System SL-PLUS Standard and Lateral Stem Table of Contents Notes from the Author s Clinic... 3 Indications... 4 Contraindications... 5 Preoperative

More information

*smith&nephew SLR-PLUS

*smith&nephew SLR-PLUS Surgical Technique *smith&nephew SLR-PLUS Cementless Revision Stem SLR-PLUS Table of Contents Comments from the Author s Clinic... 3 Indications... 4 Contraindications... 5 Preoperative Planning... 5

More information

Surgical Technique VPLWK QHSKHZ 1$126 1HFN 3UHVHUYLQJ +LS 6WHP 1716-e_NANOS_OPT.indd :27

Surgical Technique VPLWK QHSKHZ 1$126 1HFN 3UHVHUYLQJ +LS 6WHP 1716-e_NANOS_OPT.indd :27 Surgical Technique NANOS Neck Preserving Hip Stem Table of Contents Introduction... 3 Development/Concept... 4 Indications/Contraindications... 5 Preoperative Planning... 5 Surgical Technique... 6 Prosthesis

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Surgical Technique. SL-PLUS Cementless Femoral Hip System

Surgical Technique. SL-PLUS Cementless Femoral Hip System Surgical Technique *smith&nephew SL-PLUS Cementless Femoral Hip System SL-PLUS Standard and Lateral Stem Table of Contents Comment from the Author s Clinic... 3 Indications... 4 Contraindications... 5

More information

28 Surgical Technique

28 Surgical Technique Surgical Technique 10 12 14 16 18 20 22 24 28 26 Technique described by James L. Guyton, MD Campbell Clinic Memphis, Tennessee James W. Harkess, MD Campbell Clinic Memphis, Tennessee David G. LaVelle,

More information

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY E C H E L O N P R I M A R Y H I P S Y S T E M P R O D U C T R A T I O N A L E ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY Echelon Primary Total Hip System HIGH OFFSET STANDARD OFFSET Cementless

More information

Design Rationale. ECHELON Primary Hip System

Design Rationale. ECHELON Primary Hip System Design Rationale ECHELON Primary Hip System ECHELON Primary Total Hip System Addressing clinical issues of cementless hip arthroplasty Cementless total hip arthroplasty has provided a proven method of

More information

Featuring. Technology. Product Rationale

Featuring. Technology. Product Rationale Featuring Technology Product Rationale 2 Optimum implant geometry Extending proven TRI-LOCK heritage The original TRI-LOCK Stem was introduced in 1981. This implant was the first proximally coated tapered-wedge

More information

Surgical Technique Final Trial Reduction and Component Implantation of

Surgical Technique Final Trial Reduction and Component Implantation of Surgical Technique Final Trial Reduction and Component Implantation of TC *smith&nephew TC-PLUS PRIMARY Mobile Bearing TC-PLUS PRIMARY Mobile Bearing Final Trial Reduction and Component Implantation of

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Bone Preservation Stem

Bone Preservation Stem TRI-LOCK Bone Preservation Stem Featuring GRIPTION Coating Surgical Technique Implant Geometry Extending the TRI-LOCK Stem heritage The original TRI-LOCK Stem was introduced in 1981. This implant was

More information

INTACT AND IMPLANTED FEMUR BEHAVIOUR DURING WALKING AND JOGGING

INTACT AND IMPLANTED FEMUR BEHAVIOUR DURING WALKING AND JOGGING 1 from 4 INTACT AND IMPLANTED FEMUR BEHAVIOUR DURING WALKING AND JOGGING V. A. Papathanasopoulou 1, D. I. Fotiadis 2 and C.V. Massalas 3 1 Dept. of Medical Physics, Medical School, University of Ioannina,

More information

*smith&nephew TOEFIT-PLUS Modular Replacement of the First Metatarsophalangeal Joint. Product Information

*smith&nephew TOEFIT-PLUS Modular Replacement of the First Metatarsophalangeal Joint. Product Information Product Information *smith&nephew TOEFIT-PLUS Modular Replacement of the First Metatarsophalangeal Joint The moving solution for hallux rigidus An alternative to the fusion of the first metatarsophalangeal

More information

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA 94065 info@stinsonortho.com www.stinsonortho.com Table of Contents Introduction 3 Features 4 Surgical Technique 5 Preoperative

More information

Comparative Study of Fixation Devices for Intertrochanteric Fractures

Comparative Study of Fixation Devices for Intertrochanteric Fractures Comparative Study of Fixation Devices for Intertrochanteric Fractures C. Sticlaru * A. Davidescu Politehnica University of Timişoara Politehnica University of Timişoara Timişoara, România Timişoara, România

More information

Specifications guide ECHELON. Hip System

Specifications guide ECHELON. Hip System Specifications guide ECHELON Hip System Primary and revision porous implants Porous coating ROUGHCOAT porous coating increases the friction between the implant and bone, improving implant stability and

More information

Metha Short Hip Stem System

Metha Short Hip Stem System Metha Short Hip Stem System Accuracy That Stands Alone Aesculap Orthopaedics Metha Short Hip Stem System Designed For Anatomic Accuracy The Metha Short Hip Stem is designed for anatomic accuracy to restore

More information

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION SINCE 1983 PREOPERATIVE PLANNING EXPLANTATION OPTIONS the cement from inside the cement canal until the bone/ cement bond

More information

Primary stability of a shoulderless Zweymüller hip stem: a comparative in vitro micromotion study

Primary stability of a shoulderless Zweymüller hip stem: a comparative in vitro micromotion study Bieger et al. Journal of Orthopaedic Surgery and Research (2016) 11:73 DOI 10.1186/s13018-016-0410-1 RESEARCH ARTICLE Open Access Primary stability of a shoulderless Zweymüller hip stem: a comparative

More information

*smith&nephew POLARCUP Dual Mobility System. Product Information

*smith&nephew POLARCUP Dual Mobility System. Product Information Product Information *smith&nephew POLARCUP Dual Mobility System Dual Mobility 30 Years of Clinical Experience Concept combining an extended ROM, a low rate of dislocation and unexpectedly limited wear.

More information

Topology optimisation of hip prosthesis to reduce stress shielding

Topology optimisation of hip prosthesis to reduce stress shielding Computer Aided Optimum Design in Engineering IX 257 Topology optimisation of hip prosthesis to reduce stress shielding S. Shuib 1, M. I. Z. Ridzwan 1, A. Y. Hassan 1 & M. N. M. Ibrahim 2 1 School of Mechanical

More information

PLR. Proximal Loading Revision Hip System

PLR. Proximal Loading Revision Hip System PLR Proximal Loading Revision Hip System The PLR splined revision stem is designed to recreate the natural stresses in the revised femur, where proximal bone may be compromised. PLR Hip System Design Considerations

More information

Preoperative Planning. The primary objectives of preoperative planning are to:

Preoperative Planning. The primary objectives of preoperative planning are to: Preoperative Planning The primary objectives of preoperative planning are to: - Determine preoperative leg length discrepancy. - Assess acetabular component size and placement. - Determine femoral component

More information

Proven technology for customized solutions. Revitan Curved Revision Hip System

Proven technology for customized solutions. Revitan Curved Revision Hip System Proven technology for customized solutions Revitan Curved Revision Hip System Disclaimer This document is intended exclusively for physicians and is not intended for laypersons. Information on the products

More information

AML Hip System. Design Rationale/ Surgical Technique

AML Hip System. Design Rationale/ Surgical Technique AML Hip System Design Rationale/ Surgical Technique Design Rationale Evolution In 1977, DePuy Synthes Companies introduced the original cementless total hip. The AML Hip launched in order to solve one

More information

INTERTAN Nails Geared for Stability

INTERTAN Nails Geared for Stability Geared for stability The TRIGEN INTERTAN nail brings advanced TRIGEN nail technology to hip fractures. With a unique integrated, interlocking screw construct, TRIGEN INTERTAN nail provides all the benefits

More information

Section of Modular Hip Prostheses cemented. TMC-3 Modular Hip Prosthesis, cemented. TMC-3 Modular Hüftprothese, zementiert

Section of Modular Hip Prostheses cemented. TMC-3 Modular Hip Prosthesis, cemented. TMC-3 Modular Hüftprothese, zementiert Section of Modular Hip Prostheses cemented TMC-3 Modular Hip Prosthesis, cemented TMC-3 Modular Hüftprothese, zementiert Prothèse de hanche modulaire TMC-3, cimentée Indication : The TMC-3 Modular hip

More information

ACOR cementless anatomical femoral stem with a modular neck

ACOR cementless anatomical femoral stem with a modular neck ACOR cementless anatomical femoral stem with a modular neck ACOR stem: important components of intra-medullary part Stem-neck modular connection via an elliptical Morse taper. Threaded bottom of the oblong

More information

Surgical Technique Final Trial Reduction and Component Implantation of

Surgical Technique Final Trial Reduction and Component Implantation of Surgical Technique Final Trial Reduction and Component Implantation of TC *smith&nephew TC-PLUS PRIMARY Fixed Bearing TC-PLUS PRIMARY Fixed Bearing Final Trial Reduction and Component Implantation of

More information

Optimum implant geometry

Optimum implant geometry Surgical Technique Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Titel. *smith&nephew MODULAR-PLUS. Text. Cementless Modular Revision System. Surgical Technique

Titel. *smith&nephew MODULAR-PLUS. Text. Cementless Modular Revision System. Surgical Technique Titel Surgical Technique Text *smith&nephew MODULAR-PLUS Cementless Modular Revision System MODULAR-PLUS Table of Contents Introduction... 3 Concept... 4 Indications...7 Preoperative Planning...7 Surgical

More information

operative technique Kent Hip

operative technique Kent Hip operative technique Kent Hip The Kent Hip Operative Technique The Kent Hip was developed by Mr Cliff Stossel, FRCS in Maidstone, Kent, UK and first implanted in 1986. It was designed to deal with problems

More information

Avenir Femoral Hip System

Avenir Femoral Hip System Avenir Femoral Hip System Avenir Hip System Intuitive and Reproducible During its decade of experience in the bone compaction stem market, the Avenir Hip System has helped surgeons experience flexible,

More information

Revitan Curved. Surgical Technique. Modular Femoral Revision Prosthesis

Revitan Curved. Surgical Technique. Modular Femoral Revision Prosthesis Revitan Curved Surgical Technique Modular Femoral Revision Prosthesis The Revitan System 4 Goals and Methods of the Revision 6 Preoperative Planning 7 Preparation of the Preoperative Planning Sketch 14

More information

Predicting the Position of the Femoral Head Center

Predicting the Position of the Femoral Head Center The Journal of Arthroplasty Vol. 14 No. 1 1999 Predicting the Position of the Femoral Head Center Nobuhiko Sugano, MD, Philip C. Noble, PhD, and Emir Kamaric, MS Abstract: To find an accurate method to

More information

Two Years Outcomes Of Total Hip Replacement Using A Short Stem With Femoral Neck Anchoring

Two Years Outcomes Of Total Hip Replacement Using A Short Stem With Femoral Neck Anchoring Two Years Outcomes Of Total Hip Replacement Using A Short Stem With Femoral Neck Anchoring Elias I. 1, Krieger M 1, Rinaldi G. 2, Christian L. 1, Mancini N. 2 1 Klinik Rotes Kreutz, Frankfurt Am Main,

More information

5A* ODEP. Conserving bone. Preserving soft tissue. Restoring biomechanics. rating

5A* ODEP. Conserving bone. Preserving soft tissue. Restoring biomechanics. rating MiniHip is designed to give patients the advantages of a traditional hip replacement whilst also preserving the natural anatomy. MiniHip potentially allows for future revisions and so provides a pre-primary

More information

S U R G I C A L T E C H N I Q U E

S U R G I C A L T E C H N I Q U E SURGICAL TECHNIQUE RECOVERY FUNCTION SURVIVORSHIP DePuy believes in an approach to total hip replacement that places equal importance on recovery, function and survivorship. The DePuy PROXIMA Hip System

More information

Finite Element Analysis of Cemented Hip Arthroplasty: Influence of Stem Tapers

Finite Element Analysis of Cemented Hip Arthroplasty: Influence of Stem Tapers Finite Element Analysis of Cemented Hip Arthroplasty: Influence of Stem Tapers Abdul H. Abdullah, Mohd N. Mohd Asri, Mohd S. Alias and Tardan Giha, Member, IAENG Abstract Design of hip prosthesis is believed

More information

Approach Patients with Confidence

Approach Patients with Confidence Approach Patients with Confidence The is the first stem specifically designed to be utilized with tissue sparing approaches, such as the anterior approach, as well as traditional approaches. The implant

More information

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 6, Issue 12 (May 2013), PP. 119-123 A Mathematical Model to Calculate Contact Stresses

More information

A further enhanced classic. Wagner SL Revision Hip Stem

A further enhanced classic. Wagner SL Revision Hip Stem A further enhanced classic Wagner SL Revision Hip Stem The original Wagner SL Revision Stem offers a time-proven solution in the treatment of revision hips. While its underlying anchorage philosophy and

More information

Lumbo-sacral destruction fixation biomechanics

Lumbo-sacral destruction fixation biomechanics Lumbo-sacral destruction fixation biomechanics Amin Joukhar MSc * Jwalant S. Mehta FRCS Orth ^ David Marks FRCS Orth ^ Prof Vijay Goel PhD * ^ Birmingham Spine Centre, England *Engineering Centre for Orthopaedic

More information

HELIOS h i p s y s t e m

HELIOS h i p s y s t e m HELIOS h i p s y s t e m Design The Helios stem is a highly polished, High Nitrogen Stainless Steel (ISO5832-9) dual tapered cemented stem. The design of the stem is based on the clinically lly successful

More information

[ICESTM-2018] ISSN Impact Factor

[ICESTM-2018] ISSN Impact Factor GLOBAL JOURNAL OF ENGINEERING SCIENCE AND RESEARCHES MODELING AND FINITE ELEMENT ANALYSIS OF KNEE JOINT PROSTHESIS U.D.S.Prathap varma *1,S.Rajesh 2, B.Suresh Kumar 3 & P.Rama Murthy Raju 4 *1 M.TechScholar,

More information

Surgical Technique. *smith&nephew POLARSTEM Cementless Stem System

Surgical Technique. *smith&nephew POLARSTEM Cementless Stem System Surgical Technique *smith&nephew POLARSTEM Cementless Stem System POLARSTEM Cementless Stem System Contents Introduction... 3 Indications... 4 Contraindications... 4 Case Studies... 5 Preoperative Planning...

More information

Revision. Hip Stem. Surgical Protocol

Revision. Hip Stem. Surgical Protocol U2 TM Revision Hip Stem Surgical Protocol U2 Revision Hip Stem Table of Contents Introduction... 1 Preoperative Planning... 2 Femoral Preparation... 3 Trial Reduction... 5 Implant Insertion... 6 Ordering

More information

Effect of the femoral stem size on femur bone quality towards THR

Effect of the femoral stem size on femur bone quality towards THR Original article: Effect of the femoral stem size on femur bone quality towards THR Palash Kumar Maji a,*, Amit Roy Chowdhury b, Debasis Datta b, S Karmakar a, Subhomoy Chatterjee b and A K Prasad a a

More information

Cementless Tapered Femoral Stem Surgical technique

Cementless Tapered Femoral Stem Surgical technique Cementless Tapered Femoral Stem Surgical technique Contents Operative summary 4 Pre-operative planning 5 Femoral neck osteotomy 5 Femoral canal preparation 5 Intra-medullary (IM) reamer 6 Sequential rasping

More information

The Mechanical Behavior of Bone Cement in THR in the Presence of Cavities

The Mechanical Behavior of Bone Cement in THR in the Presence of Cavities Engineering, Technology & Applied Science Research Vol. 4, No. 3, 2014, 625-630 625 The Mechanical Behavior of Bone Cement in THR in the Presence of Cavities A. Benouis Mechanical Engineering Department

More information

TKA Gap Planning. Supporting healthcare professionals

TKA Gap Planning. Supporting healthcare professionals TKA Gap Planning The NAVIO TKA Gap Planning stage helps you adjust the plan based on gap information between the femur and tibia implants. Supporting healthcare professionals Interactive Views Four interactive

More information

Shunsaku Nishihara 1, Nobuhiko Sugano 1, Takashi Nishii 1, Hisashi Tanaka 2, Hideki Yoshikawa 1, and Takahiro Ochi 1

Shunsaku Nishihara 1, Nobuhiko Sugano 1, Takashi Nishii 1, Hisashi Tanaka 2, Hideki Yoshikawa 1, and Takahiro Ochi 1 J Orthop Sci (2003) 8:352 360 Comparison of the fit and fill between the Anatomic Hip femoral component and the VerSys Taper femoral component using virtual implantation on the ORTHODOC workstation Shunsaku

More information

Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA 94065 info@stinsonortho.com www.stinsonortho.com Encina HA Stem Table of Contents Introduction 3 Encina HA Stem Features 4 Surgical

More information

10-Year Experience With Short Stem Total Hip Arthroplasty

10-Year Experience With Short Stem Total Hip Arthroplasty 10-Year Experience With Short Stem Total Hip Arthroplasty GABRIELA VON LEWINSKI, MD; THILO FLOERKEMEIER, MD abstract Since 1998, short stem total hip arthroplasty (THA) has been performed at the authors

More information

9800 Metric Blvd. Austin, Texas

9800 Metric Blvd. Austin, Texas rev. A Encore Orthopedics, Inc. 1998 www.encoremed.com 9800 Metric Blvd. Austin, Texas 78758 512-832-9500 1 contents How to use the Foundation Hip 1 2 Plan your approach Select your hardware Preparing

More information

One system for all revision types

One system for all revision types Design Rationale 1 Introduction Revisions today are anything but routine. As more patients have their first total hip at a younger age, it s becoming common for many to have multiple revision surgeries

More information

Design Team Vipul Nanavati, MD James M. Paci, MD Frederick W. Werner, MME Levi G. Sutton, MS

Design Team Vipul Nanavati, MD James M. Paci, MD Frederick W. Werner, MME Levi G. Sutton, MS Design Team Vipul Nanavati, MD James M. Paci, MD Frederick W. Werner, MME Levi G. Sutton, MS SUNY Upstate Medical University Scott Macfarlane Office of Technology Transfer 315-464-7613 Anatomically shaped

More information

Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur

Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur Sowmianarayanan.S Assistant Consultant, Tata Consultancy Services Limited Prof. A.Chandrasekaran Professor, Dept. of Orthopaedic

More information

ACTIS. TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE

ACTIS. TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE ACTIS TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE APPROACH ACTIVE PATIENTS WITH CONFIDENCE The ACTIS Total Hip System is the first DePuy Synthes Companies of Johnson & Johnson

More information

Approach Patients with Confidence

Approach Patients with Confidence Surgical Technique Approach Patients with Confidence The ACTIS Total Hip System is the first DePuy Synthes stem specifically designed to be utilized with tissue sparing approaches, such as the anterior

More information

Absolut TM Cemented Stem. Surgical Technique

Absolut TM Cemented Stem. Surgical Technique Absolut TM Cemented Stem Surgical Technique Contents ABSOLUT Cemented Stem 2 Absolut Confidence 2 Absolut Reproducibility 2 Absolut Choice 2 Pre-Operative Planning 3 Suggested Templating Method 3 Surgical

More information

Clinical Evaluation Surgical Technique

Clinical Evaluation Surgical Technique Clinical Evaluation Surgical Technique Table of Contents EMPERION Specifications 3 EMPERION Surgical Technique 9 EMPERION Catalog 18 Nota Bene: This technique description herein is made available to the

More information

MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES

MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES Prof.dr.eng. Mirela TOTH-TAŞCĂU, Lecturer.dr.eng. Lucian RUSU, PhD.eng. Flavia BĂLĂNEAN, PhD.eng. Cristian TOADER-PASTI, Politehnica

More information

Cementless Tapered Femoral Stem Surgical technique

Cementless Tapered Femoral Stem Surgical technique Cementless Tapered Femoral Stem Surgical technique Contents Operative summary 4 Pre-operative planning 5 Femoral neck osteotomy 5 Femoral canal preparation 5 Intra-medullary (IM) reamer 6 Sequential rasping

More information

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

PDF hosted at the Radboud Repository of the Radboud University Nijmegen PDF hosted at the Radboud Repository of the Radboud University Nijmegen The following full text is a publisher's version. For additional information about this publication click this link. http://hdl.handle.net/2066/80687

More information

FINITE ELEMENT ANALYSIS : AN EFFECTIVE TOOL FOR PROSTHESES DESIGN

FINITE ELEMENT ANALYSIS : AN EFFECTIVE TOOL FOR PROSTHESES DESIGN Trends Biomater. Artif. Organs. Vol. 17(2) pp 141-148 (2004) http//www.sbaoi.org FINITE ELEMENT ANALYSIS : AN EFFECTIVE TOOL FOR PROSTHESES DESIGN S K Senapati & S.Pal School of Biosc, & Engg, Jadavpur

More information

THE NATURAL FIT. Surgical Technique. Hip Knee Spine Navigation

THE NATURAL FIT. Surgical Technique. Hip Knee Spine Navigation THE NATURAL FIT Surgical Technique Hip Knee Spine Navigation MiniMAX Surgical Technique Hip Knee Spine Navigation INTRODUCTION The MiniMAX TM is a cementless anatomic stem available in 9 right sizes and

More information

EVOS MINI with IM Nailing

EVOS MINI with IM Nailing Case Series Dr. John A. Scolaro EVOS MINI with IM Nailing A series of studies Introduction Intramedullary nailing has become the standard for many long bone fractures. Fracture reduction prior to nail

More information

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter Design Philosophy Cementless Stems of 12/14 Taper Stem Length Neck shaft Angle STEM-N3 STEM-N4 STEM-N5 STEM-N6 STEM-N7 STEM-N8 STEM-N9 123 mm 125 mm 130 mm 135 mm 140 mm 145 mm 150 mm 132 Cement Polished

More information

MetaFix. Cementless Total Hip Replacement Surgical technique

MetaFix. Cementless Total Hip Replacement Surgical technique Cementless Total Hip Replacement Surgical technique Contents Operative summary Acetabular preparation Pre-operative templating Femoral neck osteotomy Femoral canal preparation Femoral punch Tapered IM

More information

Surgical Technique. Hip System

Surgical Technique. Hip System Surgical Technique Hip System INDICATIONS FOR USE The TaperSet Hip System is designed for total or partial hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System.

More information

CHARNLEY Wires. Surgical Techniques

CHARNLEY Wires. Surgical Techniques CHARNLEY Wires Surgical Techniques BACKGROUND CHARNLEY Trochanteric Wires are indicated for use in re-attachment of the greater trochanter following a trochanteric osteotomy. The use of a double wire

More information

ROAR, the University of East London Institutional Repository:

ROAR, the University of East London Institutional Repository: ROAR, the University of East London Institutional Repository: http://roar.uel.ac.uk This paper is made available online in accordance with publisher policies. Please scroll down to view the document itself.

More information

Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4

Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4 Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4 Abstract Background Centre of Orthopaedic biomechanics, Bath

More information

POLARSTEM. Surgical Technique - POLARSTEM Cementless and Cemented Stem System 3

POLARSTEM. Surgical Technique - POLARSTEM Cementless and Cemented Stem System 3 Surgical Technique POLARSTEM Contents Introduction... 4 Indications... 5 Contraindications... 5 Warnings and precautions... 6 Case Studies... 8 Preoperative Planning... 9 Surgical Technique...10 Postoperative

More information

RECLAIM REVISION HIP SYSTEM

RECLAIM REVISION HIP SYSTEM RECLAIM REVISION HIP SYSTEM Where Strength and Modularity Connect DESIGN RATIONALE System Overview RECLAIM Modular Revision Hip System WHERE STRENGTH & MODULARITY CONNECT 2 Offset Options Proximal Body

More information

ME/BioE C176 Final Exam, Fall 2001

ME/BioE C176 Final Exam, Fall 2001 Saturday, December 15, 8:00 11:00 AM, 2001. Answer all questions for a maximum of 100 points. Please write all answers in the space provided. If you need additional space, write on the back sides. Indicate

More information

Alloclassic Zweymüller Stem

Alloclassic Zweymüller Stem Alloclassic Zweymüller Stem Surgical Technique A proven concept Disclaimer This document is intended exclusively for physicians and is not intended for laypersons. Information on the products and procedures

More information

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants Hip Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Hip Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All hip implants of Smith &

More information

POLARSTEM. Surgical Technique - POLARSTEM Cementless and Cemented Stem System 3

POLARSTEM. Surgical Technique - POLARSTEM Cementless and Cemented Stem System 3 Surgical Technique POLARSTEM Contents Introduction... 4 Indications... 5 Contraindications... 5 Warnings and precautions... 6 Case Studies... 8 Preoperative Planning... 9 Surgical Technique...10 Postoperative

More information

Static and Dynamic Load on Hip Contact of Hip Prosthesis and Thai Femoral Bones K. Chalernpon, P. Aroonjarattham, K.

Static and Dynamic Load on Hip Contact of Hip Prosthesis and Thai Femoral Bones K. Chalernpon, P. Aroonjarattham, K. Static and Dynamic Load on Hip Contact of Hip Prosthesis and Thai Femoral Bones K. Chalernpon, P. Aroonjarattham, K. Aroonjarattham International Science Index, Mechanical and Mechatronics Engineering

More information

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia Types of Plates 1. New Dynamic Compression Plate: DCP Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia 1. Undercut adjacent to the holes low contact: less stress shield 2. Undercut at the undersurface

More information

Zimmer Segmental System

Zimmer Segmental System Zimmer Segmental System Simple solutions for solving complex salvage cases A Step Forward The Zimmer Segmental System is designed to address patients with severe bone loss associated with disease, trauma

More information

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 12, June 2013

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 12, June 2013 Numerical Analysis of Human Femoral Bone in Different Phases Sherekar R.M, Pawar A.N. Department of Mechanical Engineering, Jawaharlal Darda Institute of Engineering and Technology, Yavatmal, 445001 (M.S).

More information

R3think your options. For US distribution only.

R3think your options. For US distribution only. R3think your options For US distribution only. R3think acetabular surgery It is widely accepted that total hip replacement surgery is one of the most successful procedures done today. Typically more thought

More information

Biomechanics of Fractures and Fixation

Biomechanics of Fractures and Fixation Biomechanics of Fractures and Fixation Theodore Toan Le, MD Original Author: Gary E. Benedetti, MD; March 2004 New Author: Theodore Toan Le, MD; Revised October 09 Basic Biomechanics Material Properties

More information

Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements

Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements Daniel Kluess, Sarah Bierbaum, Wolfram Mittelmeier, Rainer Bader,

More information

PROFEMUR Total Hip System P R O D U C T C A T A L O G

PROFEMUR Total Hip System P R O D U C T C A T A L O G PROFEMUR Total Hip System P R O D U C T C A T A L O G The PROFEMUR Total Hip System Superior Product Modular Neck Technology Seat stem in best possible bone stock Control leg length, offset, and version

More information

VERSYS HERITAGE REVISION HIP PROSTHESIS. Surgical Technique for Revision Hip Arthroplasty

VERSYS HERITAGE REVISION HIP PROSTHESIS. Surgical Technique for Revision Hip Arthroplasty VERSYS HERITAGE REVISION HIP PROSTHESIS Surgical Technique for Revision Hip Arthroplasty SURGICAL TECHNIQUE FOR VERSYS HERITAGE REVISION HIP PROSTHESIS CONTENTS DESIGN PHILOSOPHY................ 2 PREOPERATIVE

More information

HOLOGRAPHIC DEFORMATION ANALYSIS OF THE HUMAN FEMUR

HOLOGRAPHIC DEFORMATION ANALYSIS OF THE HUMAN FEMUR HOLOGRAPHIC DEFORATION ANALYSIS OF THE HUAN FEUR T. atsumoto 1, A. Kojima 2,N.Kato 1, T.Watanabe 1,.Tamiwa 1,.Baba 3 1 Faculty of Biology-Oriented Sci. & Tech., Kinki Univ,Iwade-Uchita, Wakayama, Japan

More information

Approach Patients with CONFIDENCE

Approach Patients with CONFIDENCE Design Rationale Approach Patients with CONFIDENCE The ACTIS Total Hip System is the first DePuy Synthes stem specifically designed to be utilized with tissue sparing approaches, such as the anterior

More information

CAUTION: Ceramic liners are not approved for use in the United States.

CAUTION: Ceramic liners are not approved for use in the United States. Total Hip Prostheses, Self-Centering Hip Prostheses and Hemi-Hip Prostheses IMPORTANT: This essential product information sheet does not include all of the information necessary for selection and use of

More information

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS INTRODUCTION The Summit Tapered Hip System s comprehensive set of implants and instruments

More information

Long anatomical femoral stem reduces stress concentrations on femurs with lateral bowing deformity: A finite element analysis

Long anatomical femoral stem reduces stress concentrations on femurs with lateral bowing deformity: A finite element analysis Annals of Medical Research DOI: 10.5455/annalsmedres.2018.09.205 2019;26(1):106-12 Original Article Long anatomical femoral stem reduces stress concentrations on femurs with lateral bowing deformity: A

More information

Operating Instruction. Müller Hip Stem

Operating Instruction. Müller Hip Stem Müller Hip Stem Ortho Select GmbH Eltastrasse 2 D 78573 Wurmlingen Germany Tel: +49-(0)7461-96632-30 Fax: +49-(0)7461-96632-35 info@ortho-select.de www.ortho-select.de 12009 Introduction and product description

More information

CLS Spotorno Hip Stem. Surgical Technique

CLS Spotorno Hip Stem. Surgical Technique CLS Spotorno Hip Stem Surgical Technique Surgical Technique CLS Spotorno Stem Table of Contents CLS Spotorno Stem 4 Indications for the CLS Spotorno Stem 5 Preoperative Planning 10 Surgical Technique

More information