The results of the titanium-coated RM acetabular component at 20 years

Size: px
Start display at page:

Download "The results of the titanium-coated RM acetabular component at 20 years"

Transcription

1 The results of the titanium-coated RM acetabular component at 20 years A LONG-TERM FOLLOW-UP OF AN UNCEMENTED PRIMARY TOTAL HIP REPLACEMENT M. Ihle, S. Mai, D. Pfluger, W. Siebert From the Orthopädische Klinik, Kassel, Germany In a prospective study, 93 unselected consecutive uncemented hip arthroplasties were performed in 80 patients using the titanium-coated RM acetabular component and the CLS femoral component. The mean age of the patients at operation was 52 years (28 to 81). None were lost to follow-up. In the 23 patients who had died (26 hips) only one acetabular component had been revised. In the 57 living patients (67 hips), 13 such revisions had been performed. Of the 14 revisions, seven were for osteolysis, five for loosening and two for infection. Survival analysis of this implant showed a total probability of survival of 83% (95% confidence interval 73 to 90), with all revisions as the endpoint, and a probability of 94% (95% confidence interval 87 to 98) with revision for aseptic loosening as the endpoint, indicating reliable long-term fixation of the titanium-coated RM acetabular component.! M. Ihle, MD, Clinical Research Fellow! S. Mai, MD, Orthopaedic Surgeon! W. Siebert, MD, Orthopaedic Surgeon, Medical Director Orthopädische Klinik Kassel, Wilhelmshöher Allee 345, D Kassel, Germany.! D. Pfluger, PhD, Biostatistician Datametrix AG, Technoparkstrasse 1, CH-8005 Zürich, Switzerland. Correspondence should be sent to Dr W. Siebert; wsiebert@okkassel.de 2008 British Editorial Society of Bone and Joint Surgery doi: / x.90b $2.00 J Bone Joint Surg [Br] 2008;90-B: Received 28 September 2007; Accepted after revision 14 May 2008 The main challenges in providing a durable total hip replacement (THR) relate to acetabular fixation and the bearing surface. 1 Uncemented acetabular components were developed in the hope of improving fixation. 2 Key factors for their long-term survival are their immediate stability and surface properties which promote secondary stability by bony ingrowth. 3 Porous-coated, threaded and pressfit designs have been developed, as have hemispherical versions and shells with supplementary screw fixation. Long-term results of more than 15 years are available for only a limited number of such implants. 1,2,4-6 Some series have to be interpreted with care, as many designs were modified during the period of study or too many patients were lost to follow-up. We have used the titanium-coated RM acetabular component in combination with the CLS femoral component for over 20 years. We report the long-term survival of this system in a consecutive cohort of 80 patients. Patients and Methods In this prospective study we included 93 consecutive uncemented primary THRs performed in 80 patients between January 1986 and July 1989 (Table I). Altogether, 498 THRs using the RM acetabular component were carried out during this time, mainly in combination with a cemented Müller femoral component (405 cases) generally in older patients. The uncemented RM acetabular component (Mathys Ltd, Bettlach, Switzerland) is a nonmodular, hemispherical device manufactured from ultrahigh molecular weight polyethylene (UHMWPE) and coated with a thin layer of pure titanium particles (Fig. 1). This heatpressed coating promotes secondary osseointegration but does not influence the elasticity, as occurs with metal-backed acetabular components. 7 Primary fixation is provided by two anchoring pegs on the weightbearing part supplemented by screws inserted through the periphery. 8 The polyethylene used was Chirulen (ISO 5834/2) from GUR 1120 (MediTECH, Vreden, Germany). The finished product was sterilised by gamma-irradiation in air. The components ranged in size from 50 mm to 64 mm in outer diameter. In all hips the acetabular component was paired with a straight, tapered CLS femoral component (Protek AG, Bern, Switzerland) with a caputcollum-diaphysis angle of 145. Primary fixation is obtained by metaphyseal press-fit. In 80 hips the femoral head was made of alumina (Biolox, CeramTec, Plochingen, Germany) and in 13 of CoCrMo alloy (Protasul 2, Protek AG). All heads were 32 mm in diameter. The operations were performed by four experienced orthopaedic surgeons through an anterolateral or posterior approach. No prophylaxis was given to prevent heterotopic ossification. The patients were partially weight-bearing for six weeks post-operatively, and thereafter as tolerated THE JOURNAL OF BONE AND JOINT SURGERY

2 THE RESULTS OF THE TITANIUM-COATED RM ACETABULAR COMPONENT AT 20 YEARS 1285 Table I. Demographic characteristics at operation Male Female Total/mean (range) Number of hips Number of patients Mean age (yrs) (28 to 81) Mean body mass index (kg/m 2 ) (18.7 to 36.3) Initial diagnosis (number of hips) Osteoarthritis Avascular necrosis Developmental dysplasia Femoral neck fracture Post-traumatic osteoarthritis Rheumatoid arthritis Fig. 1a Fig. 1b a) Photograph showing titanium-coated RM acetabular component with two fixation pegs, b) cross-section. Clinical evaluation included a pre-operative assessment of pain, mobility and walking ability according to Merle d Aubigné and Postel. 9 Also, the Harris Hip score (HHS) 10 was calculated at the latest follow-up. In patients with bilateral procedures each hip was assessed separately. The potential dependence resulting from the use of bilateral procedures was not considered to introduce substantial bias, and therefore the paired structure of bilateral cases was ignored with respect to statistical analysis. The radiographs were examined by two of the authors (MI, SM) and an independent, experienced orthopaedic surgeon for evidence of acetabular migration according to the criteria of Massin, Schmidt and Engh 11 and for radiolucent lines or osteolysis according to the definition proposed by Engh et al 12 in the zones described by DeLee and Charnley. 13 The most recent radiograph was compared with its immediate post-operative counterpart and any interim films. In the hips which were revised, the status of the components was assessed on the pre-operative radiograph before revision. The acetabular component was considered to be radiologically loose if it had migrated > 3 mm, or tilted > 8, or if the radiographs demonstrated circumferential radiolucent lines > 2 mm in width. 14 Heterotopic ossification was graded according to Brooker et al. 15 Statistical analysis. The probabilities of survival were estimated by the Kaplan-Meier method. 16 Failure was defined as revision for aseptic loosening or any other reason. The revision operations were differentiated into three groups, namely aseptic loosening, isolated acetabular osteolysis and infection. There was no case of recurrent dislocation. The time to revision was between the dates of implantation and revision. Patients without revision were censored at the date of the last follow-up or at death. Cox s proportional hazards model 17 was used to examine the survival rates for different risk factors for acetabular component revision which were age, the material of the femoral head, gender, the body mass index, the surgical approach, indication and the size of the acetabular component. To account for within-patient correlation, robust variances were estimated based on the marginal approach of Wei, Lin and Weissfeld. 18 These robust sandwich estimates were about 2% to 13% larger than ignoring correlation. VOL. 90-B, No. 10, OCTOBER 2008

3 1286 M. IHLE, S. MAI, D. PFLUGER, W. SIEBERT Acetabular component/stem 93/93 Follow-up 54/61 Lost to follow-up 0/0 Deceased (unrevised) 25/24 Revisions 14/8 Follow-up (combinations) 49/49 9 Acetabular components 5 Combinations 3 Femoral components Fig. 2 Flow diagram of the cohort. The model assumptions were checked graphically by comparing log-log survival curve plots (parallel lines over time) and by a formal correlation test based on the Schoenfeld residuals. 19 Risk ratios and 95% confidence intervals (CI) are presented. Scores were compared in a nonparametric manner by the Wilcoxon s signed ranks test for paired observations, with the level of significance set at α = 0.05 two-sided. For statistical analysis SAS software version was used (SAS Institute, Cary, North Carolina). Results Intra-operatively, longitudinal fractures or fissures of the femur were observed in 18 hips (19.4%). Eight healed uneventfully after cerclage wiring, five after screw fixation, and five without fixation. None showed loosening or needed revision at the latest follow-up. Three hips had an intra-operative fracture of the greater trochanter which was treated by screw fixation. During the early post-operative period there were four haematomas, three wound dehiscences, two dislocations and two infections, one of which was superficial and was successfully treated by wound revision and antibiotic therapy. The other was a deep infection requiring revision of both components. No nerve damage was reported. At the last follow-up 23 patients (26 hips) had died. Two of them had been revised, one of both components for infection, and one of the femoral components for aseptic loosening. The mean age of the deceased at the time of operation was 56.2 years (37 to 81), and at the last clinical and radiological follow-up before death there were no signs of radiological loosening or clinical failure in the unrevised cases. Of the 57 patients (67 hips) who are still alive, 13 (15 hips) underwent revision. In nine hips the acetabular component, in two hips the femoral component and in four hips both components were revised (Fig. 2). The two acetabular components corresponding to the isolated revised femoral components were stable intra-operatively and showed no radiographic signs of loosening or osteolysis at the latest follow-up. Of the 54 unrevised acetabular components, 49 were available for follow-up. Three could not be examined because of poor health, and two no longer had the original combination after revision of the femoral component. The remaining 44 unrevised patients (49 hips) had a mean follow-up of 19.3 years (17.4 to 20.9). Their mean age at the initial operation was 51.2 years (28 to 65) and at latest follow-up was 71.3 years (46 to 85). The HHS for these 49 hips was a mean of 87.5 points (54 to 100); 27 hips (55.1%) were graded excellent, 12 (24.5%) good, five (10.2%) fair and five (10.2%) poor. Musculoskeletal comorbidities had a considerable influence on these results. A total of 14 cases without comorbidity had a mean HHS of 95.5 (88 to 100), whereas 35 cases with comorbidity had a mean HHS of 84.3 (54 to 100). Hip pain was considerably relieved after surgery. At the latest follow-up 35 hips (71.4%) were pain free, ten (20.4%) had occasional mild pain not affecting daily activities, and four (8.2%) had significant pain, but considered it severe enough to warrant revision surgery. Thigh pain was not reported. The Merle d Aubigné score 9 showed significant improvements in all three parameters for the 49 unrevised hips (Table II). Of the 54 acetabular components which had not undergone revision, three radiographs could not be obtained because the patients were in nursing care, and three patients refused examination, leaving 48 complete series of radiographs for evaluation. According to the criteria of Engh et al, 5 only one acetabular component was radiologically loose, probably because of a fall. A radiolucent line adjacent to the acetabular component was seen in zone 1 (DeLee and Charnley 13 ) on two THE JOURNAL OF BONE AND JOINT SURGERY

4 THE RESULTS OF THE TITANIUM-COATED RM ACETABULAR COMPONENT AT 20 YEARS 1287 Table II. Merle d Aubigné score 9 Pre-operative Current Wilcoxon s signed ranks test Mean SD Range Mean SD Range p-value Pain to to 6 < Mobility to to 6 < Ability to walk to to 6 < Table III. Acetabular component revisions in chronological order Case Gender Diagnosis * Age at operation Time to revision Reason Osteolysis Migration Femoral head 1 M AN Infection - - Metal 2 F OA AL - Yes Ceramic 3 M OA Infection - - Ceramic 4 F OA AL Zone 1 Yes Metal 5 (a) F OA Osteolysis Zone Metal 6 (a) M AN Osteolysis Zone 1 Yes Metal 7 M RA Osteolysis Zone 1 - Ceramic 8 M AN Osteolysis Zone 1 - Ceramic 5 (b) F OA Osteolysis Zone 1 - Ceramic 9 M NF AL - Yes Metal 10 F OA Osteolysis Zone 1 - Ceramic 6 (b) M AN Osteolysis Zone Ceramic 11 F OA AL - Yes Ceramic 12 F OA AL Zone 1 Yes Metal * AN, avascular necrosis; OA, osteoarthritis; RA, rheumatoid arthritis; NF, femoral neck fracture AL, aseptic loosening radiographs, and in zone 3 on one. None had continuous radiolucencies in all three zones. Osteolysis was seen as a sharply demarcated radiolucent space on eight radiographs, six in zone 3, one in zone 2 and one in zone 1. At the latest follow-up these components did not show any clinical signs of loosening and none are awaiting revision. Grade 4 heterotopic ossification was found in two hips, grade 3 in two, grade 2 in nine and grade 1 in 12. There was no heterotopic bone in 23 hips. The 14 acetabular component revisions were carried out after a mean of 12.3 years (0.5 to 20) from the initial operation. Two patients had bilateral revisions. The 12 cases had a mean age of 49.2 years (32 to 57) at revision. Detailed information on the acetabular revisions is shown in Table III. The reasons for revision and the timing thereof are shown in Figure 3. The mean HHS in the follow-up of the 13 revised hips in 11 living patients was 78.9 points (53 to 97). In ten acetabular component revisions sufficient bone substance was available for the insertion of an equally sized or slightly larger uncemented RM component. Seven acetabular components, which were stable in spite of the osteolysis, were removed after being broken apart without additional bone loss. A reinforcement ring was needed owing to extensive bone defects, in four cases. Number of revisions Osteolysis Loosening Infection Years Fig. 3 Graph showing time course of revisions of the acetabular component. The Kaplan-Meier survival rate, with acetabular revision for aseptic loosening as the endpoint, was 96.2% at 15 years and 94.4% at 19.8 years (95% CI 87 to 98, Fig. 4). As rates based on samples < 20 are questionable, 20 we focused on the rates at 19.8 years, and did not place much emphasis on the estimated survival rate beyond 20 years when the number of patients remaining at risk fell below 20. With any acetabular VOL. 90-B, No. 10, OCTOBER 2008

5 1288 M. IHLE, S. MAI, D. PFLUGER, W. SIEBERT Survival (%) All revisions Aseptic loosening Follow-up (yrs) Patients at risk Fig. 4 Survival (%) Age: > 52 yrs Age: < 52 yrs Follow-up (yrs) Patients at risk > 52 yrs < 52 yrs Fig. 5 Kaplan-Meier survival curve for the titanium-coated RM acetabular component. The number of patients at risk is identical for both outcomes. Plot symbols indicate revisions and censored observations. Kaplan-Meier survival curve stratified by age group, with all revisions as the endpoint. Plot symbols indicate revisions and censored observations. revision as the endpoint, the survival rate was 87.2% at 15 years and 82.7% at 19.8 years (95% CI 73 to 90). The cumulative probability of survival of both components at 19.8 years was 79.7% (95% CI 70 to 88) with any revision as the endpoint, and 90.7% (95% CI 83 to 96) with revision for aseptic loosening as the endpoint. The Cox regression analysis 17 showed that younger patients had a fourfold risk of acetabular revision compared to older patients (Fig. 5, Risk Ratio (RR) = 4.0, 95% CI 1.2 to 13.8). The main factor was the material of the head with metal heads having an eight times greater risk of acetabular revision than with ceramic heads (RR = 8.0, 95% CI 2.4 to 26.5). The effect of the size and inclination of the component were not significant (Wald chi-square p = 0.99 and 0.85, respectively) and were not included in the final model. Neither age nor gender violated the assumption of proportional hazards over time (correlations between Schoenfeld residuals 19 and follow-up time were not significant; p > 0.85, Pearson correlation test following the approach outlined by Ng andu 21 ). The other variables in the model, namely gender, surgical approach and body mass index respectively, did not appear to directly affect survival of the acetabular component, but were retained to account for potential confounding. The effect of the primary diagnosis appeared to be substantial, but could not be evaluated with the Cox model owing to scarcity of data. In particular, patients with dysplasia appeared to have a more favourable prognosis, as none needed revision (log-rank test dysplasia versus remaining indications, p = 0.027). Discussion Long-term follow-up is necessary in order to evaluate the survival of any primary THR. The strength of this study is the minimum follow-up of 17 years with complete records, the combination of only two components and the absence of any significant design changes to either component. The low rate of aseptic loosening demonstrates the fixation potential of the acetabular component. Survival rates after ten years of 98%, as calculated by Flückiger and Bamert 22 and 99.1% by Diks et al, 23 confirm good anchorage. The coating of pure titanium particles produces excellent osseointegration and osseoconduction. In contrast to metal-backed modular systems, the structural behaviour of the polyethylene is not affected. Direct contact between uncoated polyethylene components and bone has shown comparatively poor results, with a loosening rate of 9% after seven, 17% after eight and 28% after nine years. 24 The RM concept of a monoblock acetabular component is unique. It provides minimal relative movement at the bone-implant interface, along with optimal osseointegration. 25 Backside wear, which occurs between a hard metal shell and the polyethylene liner, can be avoided. The design has remained unchanged and the material properties have been altered only slightly, gamma sterilisation is done under nitrogen and the polyethylene is stearate free. Complete information was available on the deceased or revised patients and all the patients were reviewed regularly. With smaller cohorts even a few cases lost to followup reduce the reliability of the results. 26 A cohort restricted to a fixed combination of components minimises the influence of different implant designs. All revisions were considered for the calculation of the survival curves. Long-term results beyond 15 years are available for only a few uncemented systems. The Zweymüller threaded component demonstrated a survival rate of 94% with revision for aseptic loosening (respectively a rate of 85% for all acetabular revisions) after 15 years, 4 and the AML THE JOURNAL OF BONE AND JOINT SURGERY

6 THE RESULTS OF THE TITANIUM-COATED RM ACETABULAR COMPONENT AT 20 YEARS 1289 acetabular component, a non-modular press-fit system, gave a rate of 95.3%. 5 The survival rates of uncemented modular components with all revisions as the endpoint have been reported for the PCA acetabular component as 79% after 20 years in young patients (mean age 48 years) 6 and 86% after 15 years for older patients (mean 58 years). 27 For the Harris-Galante I acetabular component, Della Valle et al 28 reported a rate of 82% after 18 years for younger patients (mean 52 years), and Gaffey et al 2 a rate of 81% after 15 years for older patients (mean 62 years). In many studies, however, bias due to incomplete follow-up, exclusion of liner exchange from the revision rate, or the use of various types of femoral component must be considered. In order to reduce wear and particulate debris, highly cross-linked polyethylene has been used in THR over the past decade, with Dorr et al 29 reporting a marked reduction of penetration of the metal femoral head at five years and Bragdon et al 30 confirming low rates of wear at six years. With our current understanding of the complex biological process through which particulate debris leads to periprosthetic osteolysis, a marked reduction of wear using a contemporary bearing surface may lead to a marked reduction in osteolysis. 31 The CLS femoral component is frequently used in Europe, and good long-term fixation has been reported A revision rate of 4.3% due to aseptic loosening after 20 years in our study confirms this experience. The high rate of femoral fractures, occurring mainly in the first two years, reflects the learning curve after the introduction of the new design of cementless femoral component in our hospital in Our current rate of femoral component fracture and fissures in cementless THR is 3%. The limitations of this study are the relatively small number of patients and the fact that it is a single centre study. Initially, the indications for completely uncemented THR were strictly defined. Furthermore, the Merle d Aubigné score, used initially, shows less differentiation than the HHS. Because of the relatively young age of the patients in this study almost half of the cases had developmental dysplasia of the hip, avascular necrosis or post-traumatic osteoarthritis as the diagnosis leading to THR. These diagnoses commonly reflected some deficiency in bone stock, particularly in the acetabulum. Along with their presumed greater levels of activity, caution was advised in the treatment of these young patients. Considering the high failure rates reported with a variety of cemented acetabular components, 35,36 it was hoped that uncemented acetabular components would be more durable in younger patients. Therefore, we decided to use a completely uncemented combination for young patients and this later became common practice. In this overall young cohort Cox regression analysis revealed an approximately fourfold higher risk for acetabular component revision in the under 52-year-old group, compared to the older patients. Metal heads had approximately eight times greater risk for acetabular component revision than ceramic heads. Patients with dysplastic hips (19%) showed a significantly better outcome, with none needing revision. Although we can offer no definite explanation for this finding, some patients may be less active, and metal heads were inserted in only two cases. Furthermore, this acetabular component does not require circumferential bone contact, and half of them were implanted without complete cover in order to achieve a good inclination angle. The outcome strongly suggests that the anchoring principle and component design are suitable for these cases. Between 13.7 and 15.7 years after the initial operation there was a notable number of revisions due to osteolysis, probably due to polyethylene wear. We evaluated these cases separately (Fig. 3). In spite of good anchorage, as seen in a low number of revisions because of aseptic loosening, the total survival rate is negatively biased owing to the influence of stable components revised for isolated acetabular osteolysis. Therefore, reduction in wear is a future challenge for uncemented hip prostheses. 37 Supplementary material A further figure and two additional tables are available with the electronic version of this article on our website at The authors wish to thank Professor W. Krause, Dr K. Vogt, Dr P. Rode, Dr R. Träger and Dr J. Richter for their help wih this study. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. References 1. Kearns SR, Jamal B, Rorabeck CH, Bourne RB. Factors affecting survival of uncemented total hip arthroplasty in patients 50 years or younger. Clin Orthop 2006;453: Gaffey JL, Callaghan JJ, Pedersen DR, et al. Cementless acetabular fixation at 15 years: a comparison with the same surgeon s results following acetabular fixation with cement. J Bone Joint Surg [Am] 2004;86-A: Morscher EW. Current status of acetabular fixation in primary total hip arthroplasty. Clin Orthop 1992;274: Grübl A, Chiari C, Giurea A, et al. Cementless total hip arthroplasty with the rectangular titanium Zweymüller stem: a concise follow-up, at a minimum of fifteen years, of a previous report. J Bone Joint Surg [Am] 2006;88-A: Engh CA Jr, Claus AM, Hopper RH Jr, Engh CA. Long-term results using the anatomic medullary locking hip prosthesis. Clin Orthop 2001;393: Kim YH. Long-term results of the cementless porous-coated anatomic total hip prosthesis. J Bone Joint Surg [Br] 2005;87-B: Mathys R Jr, Müller W, Mathys R, Herzig P. The coating of hip joint cups on their outer surface. In: Willert HG, Buchhorn GH, Eyerer P, eds. Ultra-high molecular weight polyethylene as biomaterial in orthopaedic surgery. Toronto: Hogrefe & Huber Publishers, 1991: Hierholzer G, Jukema GN. Die zementfreie Robert Mathys Hüftgelenkpfanne. In: Bergmann EF, ed. Die RM-Pfanne: Monographie eines beschichteten Hüftgelenkimplantates. Eichhorn-Presse Verlag, 1997: Merle d Aubigné R, Postel M. Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg [Am] 1954;36-A: Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty: an end result study using a new method of result evaluation. J Bone Joint Surg [Am] 1969;51-A: Massin P, Schmidt L, Engh CA. Evaluation of cementless acetabular component migration: an experimental study. J Arthroplasty 1989;4: Engh CA Jr, Claus AM, Hopper RH Jr, Engh CA. Long-term results using the anatomic medullary locking hip prosthesis. Clin Orthop 2001;393: DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop 1976;121: VOL. 90-B, No. 10, OCTOBER 2008

7 1290 M. IHLE, S. MAI, D. PFLUGER, W. SIEBERT 14. Engh CA, Griffin WL, Marx CL. Cementless acetabular components. J Bone Joint Surg [Br] 1990;72-B: Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr. Ectopic ossification following total hip replacement: incidence and a method of classification. J Bone Joint Surg [Br] 1973;55-B: Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Stat Assoc 1958;53: Cox DR. Regression models and life tables. J Roy Stat Soc 1972;34: Wei LJ, Lin DY, Weissfeld L. Regression analysis of multivariate incomplete failure time data by modeling marginal distributions. JASA 1989;84: Schoenfield DA. Residuals for the proportional hazards regression model. Biometrika 1982;69: Dorey F, Amstutz HC. Survivorship analysis in the evaluation of joint replacement. J Arthroplasty 1986;1: Ng andu NH. An empirical comparison of statistical tests for assessing the proportional hazards assumption of Cox s model. Stat Med 1997;16: Flückiger G, Bamert P. Mittelfristige Ergebnisse von Hydroxlapatit (Ceros 80)-beschichteten RM-Pfannen. In: Bergmann EG, ed. Die RM-Pfanne. Monographie eines beschichteten Hüftgelenkimplantates. Eichhorn-Presse Verlag, 1997: Diks MJF, van den Broek CM, Anderson PG, van Limbeek J, Spruit M. The uncemented, titanium-coated cup: survival and analyses of failures. Hip Int 2005;15: Wilson MacDonald J, Morscher E, Masar Z. Cementless uncoated polyethylene acetabular components in total hip replacement: review of five- to 10-year results. J Bone Joint Surg [Br] 1990;72-B: Witzel U. Distribution of stress in a hemispherical RM cup and its bony bed. In: Bergmann EG, ed. Die RM-Pfanne. Monographie eines beschichteten Hüftgelenkimplantates. Eichhorn-Presse Verlag, 1997: Murray DW, Britton AR, Bulstrode CJ. Loss to follow-up matters. J Bone Joint Surg [Br] 1997;79-B: Bojescul JA, Zenos JS, Callaghan JJ, Davory CG. Results of porous-coated anatomic total hip arthroplasty without cement at fifteen years: a concise follow-up of a previous report. J Bone Joint Surg [Am] 2003;85-A: Della Valle CJ, Berger RA, Shott S, et al. Primary total hip arthroplasty with a porous-coated acetabular component: a concise follow-up of a previous report. J Bone Joint Surg [Am] 2004;86-A: Dorr LD, Wan Z, Shahrdar C, et al. Clinical performance of a Durasul highly crosslinked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg [Am] 2005;87-A: Bragdon CR, Kwon YM, Geller JA, et al. Minimum 6-year followup of highly cross-linked polyethylene in THA. Clin Orthop 2007;465: Harris WH. An integrated solution to acetabular revision surgery. Clin Orthop 2006;453: Jung A, Parsch D, Ewerbeck V, Aldinger PR. Langzeitergebnisse (15-20 Jahre) des zementfreien CLS-Schaftes (abstract). Deutscher Kongress für Orthopädie und Unfallchirurgie, Berlin Aldinger PR, Breusch SJ, Lukoschek M, et al. A ten- to 15-year follow-up of the cementless spotorno stem. J Bone Joint Surg [Br] 2003;85-B: Siebold R, Scheller G, Schreiner U, Jani L. Long-term results with the cementfree spotorno CLS shaft. Orthopäde 2001;30: (in German). 35. Sarmiento A, Ebramzadeh E, Gogan WJ, McKellop HA. Total hip arthroplasty with cement: a long-term radiographic analysis in patients who are older than fifty and younger than fifty years. J Bone Joint Surg [Am] 1990;72-A: Garcia-Cimbrelo E, Cruz-Pardos A, Cordero J, Sanchez-Sotelo J. Low-friction arthroplasty in patients younger than 40 years old: 20- to 25-year results. J Arthroplasty 2000;15: Harris WH. Conquest of a worldwide human disease: particle-induced periprosthetic osteolysis. Clin Orthop 2004;429: THE JOURNAL OF BONE AND JOINT SURGERY

RM Philosophy Innovation based on clinical evidence.

RM Philosophy Innovation based on clinical evidence. RM Philosophy Innovation based on clinical evidence RM Philosophy Innovation based on clinical evidence Design 1973 RM Classic Cup First uncoated RM Classic Cup out of Polyacetal was inserted by E. Morscher

More information

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).

More information

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists : 31 35 doi 10.1308/1478708051450 Audit Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists Jac Ciampolini, Matthew JW Hubble Princess Elizabeth Orthopaedic Centre,

More information

Klinische und radiologische Ergebnisse einer «isoelastischen» Pfanne

Klinische und radiologische Ergebnisse einer «isoelastischen» Pfanne Klinische und radiologische Ergebnisse einer «isoelastischen» Pfanne Martin Beck, Prof. Dr.med. Klinik für Orthopädie und Unfallchirurgie Luzerner Kantonsspital resorption of acetabular bone can be observed

More information

Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A PREVIOUS REPORT*

Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A PREVIOUS REPORT* 1066 COPYRIGHT 2003 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A

More information

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement National Joint Replacement Registry Lay Summary 2015 Annual Report Hip and Knee Replacement SUPPLEMENTARY REPORT 2015 TABLE OF CONTENTS Introduction... 1 A brief history of the Registry origins... 1 The

More information

Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision

Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision Clin Orthop Relat Res (2009) 467:1501 1506 DOI 10.1007/s11999-008-0499-6 ORIGINAL ARTICLE Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision Byung-Woo Min MD, Kwang-Soon Song

More information

Principles of acetabular fixation in primary and revision hip arthroplasty Piotr Wojciechowski, Damian Kusz, Anna WAGNER

Principles of acetabular fixation in primary and revision hip arthroplasty Piotr Wojciechowski, Damian Kusz, Anna WAGNER Principles of acetabular fixation in primary and revision hip arthroplasty Piotr Wojciechowski, Damian Kusz, Anna WAGNER Department of Orthopedics and Traumatology Medical University of Silesia Head of

More information

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem Original Article Clinics in Orthopedic Surgery 2009;1:105-109 doi:10.4055/cios.2009.1.2.105 Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem Kyoung Ho Moon, MD, Joon Soon

More information

Featuring. Technology. Product Rationale

Featuring. Technology. Product Rationale Featuring Technology Product Rationale 2 Optimum implant geometry Extending proven TRI-LOCK heritage The original TRI-LOCK Stem was introduced in 1981. This implant was the first proximally coated tapered-wedge

More information

Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner

Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner ORIGINAL ARTICLE Hip Pelvis 27(4): 216-222, 2015 http://dx.doi.org/10.5371/hp.2015.27.4.216 Print ISSN 2287-3260 Online ISSN 2287-3279 Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene

More information

Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty

Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty M Nomura, S The Journal et al. of International Medical Research Endovascular 2000; 28: Embolization 307 312 of Unruptured Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty

More information

PINNACLE Acetabular Cup System

PINNACLE Acetabular Cup System PINNACLE Acetabular Cup System Clinical Summary A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up Engh CA Jr., Hopper RH

More information

Smith & Nephew. R3 Cementless Cup

Smith & Nephew. R3 Cementless Cup Implant Smith & Nephew Comprising PRIMARY hips implanted up to: 09 September 2018 NJR Database extract: 08 November 2018 Produced on: Licenced for use until: 18 November 2018 18 March 2019 Contents Recorded

More information

Tri-Lock Bone Preservation Stem

Tri-Lock Bone Preservation Stem Tri-Lock Bone Preservation Stem Clinical Results of the Tri-lock BPS Femoral Stem at One Year Follow-Up William L. Healy, M.D. Lahey Clinic Medical Center Burlington, Massachusetts Introduction Cementless

More information

Furlong H-A.C. THR System

Furlong H-A.C. THR System Important Information Please read prior to use in a clinical setting. The Surgeon should be familiar with the operative technique. Caution Federal (U.S.A) law restricts this device to sale by or on the

More information

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Acta Orthop. Belg., 2008, 74, 337-342 ORIGINAL STUDY Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Adrian SKWARA, Simone STRACKE, Carsten O. TIBESKU, Susanne FUCHS-WINKELMANN

More information

The Journal of Arthroplasty

The Journal of Arthroplasty The Journal of Arthroplasty 29 (2014) 2256 2261 Contents lists available at ScienceDirect The Journal of Arthroplasty journal homepage: www.arthroplastyjournal.org The 27 to 29-Year Outcomes of the PCA

More information

Optimum implant geometry

Optimum implant geometry Surgical Technique Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Cementless Acetabular Fixation With and Without Screws

Cementless Acetabular Fixation With and Without Screws Cementless Acetabular Fixation With and Without Screws Analysis of Stability and Migration The Journal of Arthroplasty Vol. 25 No. 2 2010 Richard Iorio, MD,* Brian Puskas, MD,y William L. Healy, MD,* John

More information

Acetabular Cup System. Clinical Summary

Acetabular Cup System. Clinical Summary Acetabular Cup System Clinical Summary A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up Engh CA Jr., Hopper RH Jr., Huynh

More information

Bone Bangalore

Bone Bangalore Dr Suresh Annamalai MBBS, MRCS(Edn), FRCS( Tr & Orth)(Edn), FEBOT(European Board), Young Hip and Knee Fellowship(Harrogate, UK) Consultant Arthroplasty and Arthroscopic Surgeon Manipal Hospital, Whitefield,

More information

Optimizing function Maximizing survivorship Accelerating recovery

Optimizing function Maximizing survivorship Accelerating recovery Surgical Technique Optimizing Function Maximizing Survivorship Accelerating Recovery The company believes in an approach to patient treatment that places equal importance on: Optimizing function Maximizing

More information

Stability without compromise. Epsilon Durasul Constrained Insert

Stability without compromise. Epsilon Durasul Constrained Insert Stability without compromise Epsilon Durasul Constrained Insert History of Dislocation and Constrained Acetabular Inserts The Need for a Better Solution The Problem Dislocation is the second most common

More information

Revision hip arthroplasty with S-ROM prosthesis: a study of clinical outcomes and implant stability

Revision hip arthroplasty with S-ROM prosthesis: a study of clinical outcomes and implant stability J Orthopaed Traumatol () 7:1 1 DOI 1.17/s1195--15- ORIGINAL M. El-Deen S. Zahid D.T. Miller A. Nargol R. Logishetty Revision hip arthroplasty with S-ROM prosthesis: a study of clinical outcomes and implant

More information

ORIGINAL PAPER. Department of Orthopedic Surgery, Hamamatsu Medical Center ABSTRACT

ORIGINAL PAPER. Department of Orthopedic Surgery, Hamamatsu Medical Center ABSTRACT Nagoya J. Med. Sci. 71. 145 ~ 150, 2009 ORIGINAL PAPER AUTOGENOUS BULK STRUCTURAL BONE GRAFTING FOR RECONSTRUCTION OF THE ACETABLUM IN PRIMARY TOTAL HIP ARTHROPLASTY: AVERAGE 12-YEAR FOLLOW-UP TETSUO MASUI,

More information

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574)

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574) References 1. Sanchez-Sotelo J, Haidukewych GJ, Boberg CJ. Hospital Cost of Dislocation After Primary Total Hip Arthroplasty. J Bone Joint Surg. 2006;88A:290-294. 2. Brodner W, Grübl A, Jankovsky R, Meisinger

More information

Hedrocel trabecular metal monoblock acetabular cups : mid-term results

Hedrocel trabecular metal monoblock acetabular cups : mid-term results Acta Orthop. Belg., 2006, 72, 326-331 ORIGINAL STUDY Hedrocel trabecular metal monoblock acetabular cups : mid-term results Michiel MULIER, Bart RYS, Lieven MOKE From the University Hospital Pellenberg,

More information

A further enhanced classic. Wagner SL Revision Hip Stem

A further enhanced classic. Wagner SL Revision Hip Stem A further enhanced classic Wagner SL Revision Hip Stem The original Wagner SL Revision Stem offers a time-proven solution in the treatment of revision hips. While its underlying anchorage philosophy and

More information

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter Design Philosophy Cementless Stems of 12/14 Taper Stem Length Neck shaft Angle STEM-N3 STEM-N4 STEM-N5 STEM-N6 STEM-N7 STEM-N8 STEM-N9 123 mm 125 mm 130 mm 135 mm 140 mm 145 mm 150 mm 132 Cement Polished

More information

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing Journal of Orthopaedic Surgery 2001, 9(1): 45 50 Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing KY Chiu, TP Ng, WM Tang and P Lam Department of Orthopaedic Surgery, The University

More information

Produced on: Licenced for use until: Corail Stem (Standard Offset Collared)

Produced on: Licenced for use until: Corail Stem (Standard Offset Collared) Implant Bespoke Report for: DePuy Comprising PRIMARY hips implanted up to: 09 October 2017 NJR Database extract: 08 December 2017 Produced on: Licenced for use until: 29 December 2017 29 April 2018 Contents

More information

A Proven Concept. Alloclassic Zweymüller Stem

A Proven Concept. Alloclassic Zweymüller Stem A Proven Concept Alloclassic Zweymüller Stem Disclaimer This document is intended exclusively for experts in the field, i. e. physicians in particular, and is expressly not for the information of laypersons.

More information

Reprint requests: Dr Brenda Dower CLINICAL ARTICLE SA ORTHOPAEDIC JOURNAL Spring 2012 Vol 11 No 3 / Page 29

Reprint requests: Dr Brenda Dower CLINICAL ARTICLE SA ORTHOPAEDIC JOURNAL Spring 2012 Vol 11 No 3 / Page 29 CLINICAL ARTICLE SA ORTHOPAEDIC JOURNAL Spring 2012 Vol 11 No 3 / Page 29 C L I N I C A L A RT I C L E The results of a cementless acetabular component combined with impaction bone grafting in patients

More information

Produced on: Licenced for use until: Corail Stem (Standard Offset Non-Collared)

Produced on: Licenced for use until: Corail Stem (Standard Offset Non-Collared) Implant Bespoke Report for: DePuy Comprising PRIMARY hips implanted up to: 09 October 2017 NJR Database extract: 08 December 2017 Produced on: Licenced for use until: 29 December 2017 29 December 2018

More information

Are there limits for a cementless short hip stem?

Are there limits for a cementless short hip stem? Are there limits for a cementless short hip stem? Results after 2 year follow up Dr. Sabine Mai Consulting Surgeon Vitos Orthopädische Klinik Kassel Germany Akademisches Lehrkrankenhaus der Universität

More information

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty The Journal of Arthroplasty Vol. 23 No. 2 2008 Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty 7135 Primary Total Hip Arthroplasties after Developmental Dysplasia of the

More information

Midterm Results of 506 Solid Trispiked Reflection Cementless Acetabular Components for Primary Total Hip Arthroplasty

Midterm Results of 506 Solid Trispiked Reflection Cementless Acetabular Components for Primary Total Hip Arthroplasty The Journal of Arthroplasty Vol. 00 No. 0 2011 Midterm Results of 506 Solid Trispiked Reflection Cementless Acetabular Components for Primary Total Hip Arthroplasty Kristoff Corten, MD, Richard W. McCalden,

More information

CAUTION: Ceramic liners are not approved for use in the United States.

CAUTION: Ceramic liners are not approved for use in the United States. Total Hip Prostheses, Self-Centering Hip Prostheses and Hemi-Hip Prostheses IMPORTANT: This essential product information sheet does not include all of the information necessary for selection and use of

More information

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS INTRODUCTION The Summit Tapered Hip System s comprehensive set of implants and instruments

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Smith & Nephew. Polarstem Cementless

Smith & Nephew. Polarstem Cementless Implant Smith & Nephew Comprising PRIMARY hips implanted up to: 09 September 2018 NJR Database extract: 08 November 2018 Produced on: Licenced for use until: 20 November 2018 20 March 2019 Contents Recorded

More information

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5.

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5. Surgical Technique Contents Templating and Pre Operative Planning 2 Preparation of the Acetabulum 4 Trial Sizing and Impaction of the Shell 5 Cup Positioning 6 Joint Stability 7 Trial sizing and Impaction

More information

SUMMIT and DURALOC. Clinical Summary

SUMMIT and DURALOC. Clinical Summary SUMMIT and DURALOC Clinical Summary Ten-year results of a press-fit, porous-coated acetabular component Grobler G.P. Learmonth I.D. Bernstein B.P. Dower B.J. (2005) The Journal of Bone and Joint Surgery;

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Charnley Total Hip Arthroplasty with Use of Improved Cementing Techniques A MINIMUM TWENTY-YEAR FOLLOW-UP STUDY

Charnley Total Hip Arthroplasty with Use of Improved Cementing Techniques A MINIMUM TWENTY-YEAR FOLLOW-UP STUDY 1840 COPYRIGHT 2001 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Charnley Total Hip Arthroplasty with Use of Improved Cementing Techniques A MINIMUM TWENTY-YEAR FOLLOW-UP STUDY BY AIMEE S. KLAPACH,

More information

CLS Brevius Stem with Kinectiv Technology

CLS Brevius Stem with Kinectiv Technology CLS Brevius Stem with Kinectiv Technology The Next Chapter in 27 Years of CLS Spotorno Stem History: Flexibility 2 CLS Brevius Stem with Kinectiv Technology Product Information Introduction Through independent

More information

Total hip arthroplasty for primary osteoarthrosis in younger patients in the Finnish arthroplasty register

Total hip arthroplasty for primary osteoarthrosis in younger patients in the Finnish arthroplasty register Acta Orthopaedica ISSN: 1745-3674 (Print) 1745-3682 (Online) Journal homepage: http://www.tandfonline.com/loi/iort20 Total hip arthroplasty for primary osteoarthrosis in younger patients in the Finnish

More information

REVISION. Zimmer Biomet All Trabecular Metal Cups vs. All non-tm cementless cups

REVISION. Zimmer Biomet All Trabecular Metal Cups vs. All non-tm cementless cups REVISION Bespoke Implant Report for: Zimmer Biomet vs. All non-tm cementless cups Comprising REVISION hips implanted up to: 01 August 2015 NJR Database extract: 30 September 2015 Produced on: 05 November

More information

Continuing the Tradition. VerSys Heritage Hip System

Continuing the Tradition. VerSys Heritage Hip System Continuing the Tradition VerSys Heritage Hip System Heritage Following the Tradition The low-friction hip prosthesis developed by Sir John Charnley has more than a 20-year history of outstanding results.

More information

Allofit /Allofit -S Alloclassic Acetabular Cup System

Allofit /Allofit -S Alloclassic Acetabular Cup System Allofit /Allofit -S Alloclassic Acetabular Cup System Product Information Convincing Primary Stability The Allofit acetabular cup was developed in 1993. The goal was to create a reliable implant that was

More information

Clinical and radiological results of the stemmed Mc Minn cup in hip revision surgery

Clinical and radiological results of the stemmed Mc Minn cup in hip revision surgery Acta Orthop. Belg., 2010, 76, 58-62 ORIGINAL STUDY Clinical and radiological results of the stemmed Mc Minn cup in hip revision surgery Pax WILLEMSE, Rene M. CASTELEIN, Paul L. P. A. BOM, Aart VERBURG,

More information

Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty

Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty Christopher Samujh, MD; Samrath Bhimani, MS; Langan Smith, BS; Arthur L. Malkani, MD abstract A major

More information

WHAT IS THE BEST BONE FIXATION TYPE? 2/11/2011

WHAT IS THE BEST BONE FIXATION TYPE? 2/11/2011 WHAT IS THE BEST BONE FIXATION TYPE? 2/11/2011 A Comparison of cement vs. bone ingrowth. Thomas P. Gross, M.D. At 2 years of follow-up cemented and uncemented femoral resurfacing is equivalent. Femoral

More information

CIRCUMFERENTIAL PROXIMAL FEMORAL ALLOGRAFTS IN REVISION SURGERY ON TOTAL HIP ARTHROPLASTY: CASE REPORTS WITH A MINIMUM FOLLOW-UP OF 20 YEARS

CIRCUMFERENTIAL PROXIMAL FEMORAL ALLOGRAFTS IN REVISION SURGERY ON TOTAL HIP ARTHROPLASTY: CASE REPORTS WITH A MINIMUM FOLLOW-UP OF 20 YEARS CASE REPORT CIRCUMFERENTIAL PROXIMAL FEMORAL ALLOGRAFTS IN REVISION SURGERY ON TOTAL HIP ARTHROPLASTY: CASE REPORTS WITH A MINIMUM FOLLOW-UP OF 20 YEARS Bruno Dutra Roos 1, Milton Valdomiro Roos 2, Antero

More information

Understanding Hip Implant Options

Understanding Hip Implant Options Understanding Hip Implant Options Cup (Socket) Bearing (Liner) Head Stem (Femur) Modern Hip Implants (Ball) Metal Femoral Head on Marathon Cross-linked Polyethylene Liner Proven materials Polyethylene

More information

ACETABULAR CUP SURGICAL TECHNIQUE

ACETABULAR CUP SURGICAL TECHNIQUE ACETABULAR CUP SURGICAL TECHNIQUE ACETABULAR CUP DEVICE INDICATIONS FOR USE The ICONACY I-Hip total hip replacement is indicated for the following conditions: 1. A severely painful and/or disabled hip

More information

Ceramic Femoral Heads and Acetabular Cup Liners

Ceramic Femoral Heads and Acetabular Cup Liners Important Information: Please read prior to use in a clinical setting. The Surgeon should be familiar with the operative technique. Caution: Federal (U.S.A) law restricts this device to sale by or on the

More information

Midterm Results Of Total Hip Arthroplasty (THA) In Developmental Dysplasia Of The Hip (DDH)

Midterm Results Of Total Hip Arthroplasty (THA) In Developmental Dysplasia Of The Hip (DDH) ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 23 Number 1 Midterm Results Of Total Hip Arthroplasty (THA) In Developmental Dysplasia Of The Hip (DDH) A Öner, A Köksal, O D Tunç, M B Balioğlu,

More information

Clinical Study Clinical and Radiographic Outcomes with a Hydroxyapatite and Porous Coated Cup Design

Clinical Study Clinical and Radiographic Outcomes with a Hydroxyapatite and Porous Coated Cup Design Advances in Orthopedic Surgery, Article ID 302969, 5 pages http://dx.doi.org/10.1155/2014/302969 Clinical Study Clinical and Radiographic Outcomes with a Hydroxyapatite and Porous Coated Cup Design John

More information

The shape and size of femoral components in revision total hip arthroplasty among Chinese patients

The shape and size of femoral components in revision total hip arthroplasty among Chinese patients Journal of Orthopaedic Surgery 2003: 11(1): 53 58 The shape and size of femoral components in revision total hip arthroplasty among Chinese patients KY Chiu, TP Ng, WM Tang Department of Orthopaedic Surgery,

More information

Ten-year survival of the MS-30 matt-surfaced cemented stem

Ten-year survival of the MS-30 matt-surfaced cemented stem Ten-year survival of the MS-30 matt-surfaced cemented stem B. J. Berli, D. Schäfer, E. W. Morscher From the Orthopaedic Department, University of Basel, Basel, Switzerland The Morscher-Spotorno (MS-30)

More information

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique Enhanced Stability Constrained Liners Design Rationale Surgical Technique The Pinnacle Acetabular Cup System was designed to maximize the number of options available to the surgeon, and provide those options

More information

28 Surgical Technique

28 Surgical Technique Surgical Technique 10 12 14 16 18 20 22 24 28 26 Technique described by James L. Guyton, MD Campbell Clinic Memphis, Tennessee James W. Harkess, MD Campbell Clinic Memphis, Tennessee David G. LaVelle,

More information

Operating Instruction. Müller Hip Stem

Operating Instruction. Müller Hip Stem Müller Hip Stem Ortho Select GmbH Eltastrasse 2 D 78573 Wurmlingen Germany Tel: +49-(0)7461-96632-30 Fax: +49-(0)7461-96632-35 info@ortho-select.de www.ortho-select.de 12009 Introduction and product description

More information

CONSENSUS ORTHOPEDICS INC.

CONSENSUS ORTHOPEDICS INC. CONSENSUS ORTHOPEDICS INC. CONSENSUS HIP SYSTEMS Instructions for Use (IFU) IMPORTANT INFORMATION FOR SURGEON: PLEASE READ PRIOR TO IMPLANTING THIS DEVICE IN A CLINICAL SETTING. THE SURGEON SHOULD BE FAMILIAR

More information

Total Hip Arthroplasty Performed Using Conventional and Computer-Assisted, Tissue- Preserving Techniques 6

Total Hip Arthroplasty Performed Using Conventional and Computer-Assisted, Tissue- Preserving Techniques 6 Total Hip Arthroplasty Performed Using Conventional and Computer-Assisted, Tissue- Preserving Techniques 6 Stephen B. Murphy, MD, Timo M. Ecker, MD and Moritz Tannast, MD Introduction Less invasive techniques

More information

TEN YEARS EXPERIENCE WITH NONCEMENTED REVISION SOCKETS. Lawrence D. Dorr, MD Zhinian Wan, MD

TEN YEARS EXPERIENCE WITH NONCEMENTED REVISION SOCKETS. Lawrence D. Dorr, MD Zhinian Wan, MD ""'".."c.",".~" ""ill""."""'""'" I TEN YEARS EXPERIENCE WITH NONCEMENTED REVISION SOCKETS. Lawrence D. Dorr, MD Zhinian Wan, MD f ABSTRACT A prospective study was completed on 167 noncemented revision

More information

Centralization of the femoral component in cemented hip arthroplasty using guided stem insertion

Centralization of the femoral component in cemented hip arthroplasty using guided stem insertion Arch Orthop Trauma Surg (1998) 117 : 425 429 Springer-Verlag 1998 ORIGINAL ARTICLE G. Köster H.-G. Willert T. Ernstberger H.-P. Köhler Centralization of the femoral component in cemented hip arthroplasty

More information

CLS Spotorno Hip Stem A Milestone in the History of Hip Replacements

CLS Spotorno Hip Stem A Milestone in the History of Hip Replacements CLS Spotorno Hip Stem A Milestone in the History of Hip Replacements 2 The Original CLS Cementless Spotorno Hip Stem The Creation of CLS Marked a Milestone in the History of Hip Replacements Until the

More information

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE DePuy Synthes DURALOC Surgical Technique CONTENTS Templating and Pre Operative Planning 2 Preparation of the Acetabulum

More information

CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician.

CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician. CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician. ENGLISH Mpact 3D Metal Implants and Augments 3D Metal INSTRUCTION FOR USE Important notice: the device(s) can

More information

1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe

1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe 1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe Abstract Introduction: Osteolysis generated by wear debris remains a problem

More information

Ten-year survival of the cemented MS-30 femoral stem Increased revision rate in male patients

Ten-year survival of the cemented MS-30 femoral stem Increased revision rate in male patients Acta Orthop. Belg., 9, 5, 6-5 ORIGINAL STUDY Ten-year survival of the cemented MS-3 femoral stem Increased revision rate in male patients Daniela WITTE, Michael KLIMM, Dominik Parsch, Michael CLARIUS,

More information

TOTAL HIP REPLACEMENT:

TOTAL HIP REPLACEMENT: THR Prosthesis Design TOTAL HIP REPLACEMENT: PROSTHESIS DESIGN FEATURES JESS JOHNSTON & MELINDA ZIETH History of Hip Prosthesis Joint Replacement Registry Implant Design Technology & Future History and

More information

Formosan Journal of Musculoskeletal Disorders

Formosan Journal of Musculoskeletal Disorders Formosan Journal of Musculoskeletal Disorders 3 (2012) 83e87 Contents lists available at SciVerse ScienceDirect Formosan Journal of Musculoskeletal Disorders journal homepage: www.e-fjmd.com Original Article

More information

Medical Products Division. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion

Medical Products Division. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion Medical Products Division BIOLOX DUO Ceramic Bipolar System Bone-sparing joint reconstruction with a maximum range of motion 2 BIOLOX DUO Keramisches Bipolarsystem BIOLOX DUO Ceramic Bipolar System Less

More information

Peggers Super Summaries: THR and Revision Concepts

Peggers Super Summaries: THR and Revision Concepts Total Hip replacement: ARTICUALTIONS Ceramic on polyethylene Ceramic on ceramic for young Metal on metal (resurfacing) Metal (cobalt chrome) on polyethylene WEAR Wear debris similar to bacteria o < 10

More information

Management Of Acetabular Deficiency In Total Hip Arthroplasty: A Series Of 15 Cases

Management Of Acetabular Deficiency In Total Hip Arthroplasty: A Series Of 15 Cases ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 21 Number 2 Management Of Acetabular Deficiency In Total Hip Arthroplasty: A Series Of 15 Cases G Khanna, R Sharma, D Singh, T A Chandy Citation

More information

Reaching new heights. Comprehensive. Efficient. Simple.

Reaching new heights. Comprehensive. Efficient. Simple. Reaching new heights Comprehensive. Efficient. Simple. Various acetabular cup choices Compatible with the different head and liner options including VERILAST Technology Reach for proven OR efficient Instrumentation

More information

Metha Short Hip Stem System

Metha Short Hip Stem System Metha Short Hip Stem System Accuracy That Stands Alone Aesculap Orthopaedics Metha Short Hip Stem System Designed For Anatomic Accuracy The Metha Short Hip Stem is designed for anatomic accuracy to restore

More information

Zimmer TM. Continuum Acetabular System. Bringing together proven technologies.

Zimmer TM. Continuum Acetabular System. Bringing together proven technologies. Zimmer TM Continuum Acetabular System Bringing together proven technologies. The power to meet individual patient needs. Zimmer Continuum Acetabular System provides highly flexible solutions for orthopaedic

More information

Revision Total Hip Arthroplasty Using Tantalum Augment in Patients with Paprosky III or IV Acetabular Bone Defects: A Minimum 2-year Follow Up Study

Revision Total Hip Arthroplasty Using Tantalum Augment in Patients with Paprosky III or IV Acetabular Bone Defects: A Minimum 2-year Follow Up Study ORIGINAL ARTICLE Hip Pelvis 28(2): 98-103, 2016 http://dx.doi.org/10.5371/hp.2016.28.2.98 Print ISSN 2287-3260 Online ISSN 2287-3279 Revision Total Hip Arthroplasty Using Tantalum Augment in Patients with

More information

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA 94065 info@stinsonortho.com www.stinsonortho.com Table of Contents Introduction 3 Features 4 Surgical Technique 5 Preoperative

More information

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital)

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) 1. Introduction Total hip replacement arthroplasty (THRA) using a metal-on-polyethylene articulation has become one of the most

More information

CeramTec Medical Products. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion

CeramTec Medical Products. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion CeramTec Medical Products BIOLOX DUO Ceramic Bipolar System Bone-sparing joint reconstruction with a maximum range of motion 2 BIOLOX DUO Ceramic Bipolar System Less is more BIOLOX DUO is an alternative

More information

DePuy Corail Collared vs. Collarless (ex MoM)

DePuy Corail Collared vs. Collarless (ex MoM) Bespoke Implant Report for: DePuy Comprising PRIMARY hips implanted up to: 09 October 2017 NJR Database extract: 08 December 2017 Produced on: Licenced for use until: 29 December 2017 29 December 2018

More information

Department of Orthopaedic Surgery, University of Toledo Medical Center, Toledo, OH, US.

Department of Orthopaedic Surgery, University of Toledo Medical Center, Toledo, OH, US. ISSN: 2572-2964 Volume 2, Issue 1, 6 Pages Research Article Acetabular fractures can complicate hip arthroplasty and result in higher rates of intraoperative complications and worse long-term outcomes

More information

Informed Consent for HRA

Informed Consent for HRA Updated March 09 Thomas P Gross MD Informed Consent for HRA Dr. Gross has now performed over 5500 Hip Resurfacing Arthroplasty (HRA) procedures over the last 8 years. Most failures occur during the first

More information

Is Resurfacing Arthroplasty Appropriate for Posttraumatic Osteoarthritis?

Is Resurfacing Arthroplasty Appropriate for Posttraumatic Osteoarthritis? Clin Orthop Relat Res DOI 10.1007/s11999-010-1655-3 SYMPOSIUM: UPDATE ON HARD-ON-HARD BEARINGS IN HIP ARTHROPLASTY Is Resurfacing Arthroplasty Appropriate for Posttraumatic Osteoarthritis? Aaron J. Johnson

More information

NATIONAL INSTITUTE FOR CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR CLINICAL EXCELLENCE INTERVENTIONAL PROCEDURES PROGRAMME Interventional procedures overview of minimally invasive two-incision surgery for total hip replacement Introduction This

More information

Cotyloplasty in Cementless Total Hip Arthroplasty for an Insufficient Acetabulum

Cotyloplasty in Cementless Total Hip Arthroplasty for an Insufficient Acetabulum Original Article Clinics in Orthopedic Surgery 2010;2:148-153 doi:10.4055/cios.2010.2.3.148 Cotyloplasty in Cementless Total Hip Arthroplasty for an Insufficient Acetabulum Yong Lae Kim, MD, Kwang Woo

More information

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY Fixation First FIXATION FIRST 1958 Charnley Hip Over 1,000,000 implanted 1 96.2 percent survivorship at 32 years

More information

Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination.

Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination. ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 17 Number 1 Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination. Roberts, Garlick

More information

TOTAL HIP ARTHROPLASTY WITH BIOMEC PROSTHESIS: 20-YEAR FOLLOW-UP

TOTAL HIP ARTHROPLASTY WITH BIOMEC PROSTHESIS: 20-YEAR FOLLOW-UP ORIGINAL ARTICLE TOTAL HIP ARTHROPLASTY WITH BIOMEC PROSTHESIS: 2-YEAR FOLLOW-UP Rodrigo Benedet Scheidt 1, Ricardo Rosito 2, Carlos Alberto de Souza Macedo 3, Carlos Roberto Galia 4 ABSTRACT Objective:

More information

HOSPITAL "A" (2007) ORTHOPEDIC JOINT IMPLANT COMPONENT MATRIX

HOSPITAL A (2007) ORTHOPEDIC JOINT IMPLANT COMPONENT MATRIX HOSPITAL "A" (2007) ORTHOPEDIC JOINT IMPLANT COMPONENT MATRIX HIP IMPLANTS TOTAL HIPS Ultra High Demand (metal on metal) Femur, ingrowth Femoral head, metal Acetabular shell, ingrowth Acetabular liner,

More information

MicroSeal Total Hip Acetabular System. Like Nothing Else

MicroSeal Total Hip Acetabular System. Like Nothing Else MicroSeal Total Hip Acetabular System Like Nothing Else Like Nothing Else Unique Patented Sealing and Locking Design The MicroSeal design features: Locking/Sealing mechanisms work together to minimize

More information

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Timothy L. Tan, BS 1, Brenna C. Moeljadi, BS 2, Edward Ebramzadeh, PhD 3, Patricia Campbell, PhD 3, Sophia Nicole

More information

REVISING THE DEFICIENT PROXIMAL FEMUR

REVISING THE DEFICIENT PROXIMAL FEMUR REVISING THE DEFICIENT PROXIMAL FEMUR by David Mattingly, Boston, MA Joseph McCarthy, Boston, MA Benjamin E. Bierbaum, Boston, MA Hugh P. Chandler, Boston, MA Roderick H. Turner, Boston, MA Hugh U. Cameron,

More information