Implant options for the treatment of intertrochanteric fractures of the hip

Size: px
Start display at page:

Download "Implant options for the treatment of intertrochanteric fractures of the hip"

Transcription

1 TRAUMA Implant options for the treatment of intertrochanteric fractures of the hip RATIONALE, EVIDENCE, AND RECOMMENDATIONS A. R. Socci, N. E. Casemyr, M. P. Leslie, M. R. Baumgaertner From Yale University School of Medicine, Connecticut, United States A. R. Socci, MD, Assistant Professor N. E. Casemyr, MD, Assistant Professor, M. P. Leslie, DO, Associate Professor M. R. Baumgaertner, MD, Professor Yale University School of Medicine, Department of Orthopaedics and Rehabilitation, 800 Howard Ave, New Haven, CT 06520, USA. Correspondence should be sent to A. R. Socci; The British Editorial Society of Bone & Joint Surgery doi: / x.99b1. BJJ R1 $2.00 Bone Joint J 2017;99-B: Received 24 March 2016; Accepted after revision 17 August 2016 Aims The aim of this paper is to review the evidence relating to the anatomy of the proximal femur, the geometry of the fracture and the characteristics of implants and methods of fixation of intertrochanteric fractures of the hip. Materials and Methods Relevant papers were identified from appropriate clinical databases and a narrative review was undertaken. Results Stable, unstable, and subtrochanteric intertrochanteric fractures vary widely in their anatomical and biomechanical characteristics, as do the implants used for their fixation. The optimal choice of implant addresses the stability of the fracture and affects the outcome. Conclusion The treatment of intertrochanteric fractures of the hip has evolved along with changes in the design of the implants used to fix them, but there remains conflicting evidence to guide the choice of implant. We advocate fixation of 31A1 fractures with a sliding hip screw and all others with an intramedullary device. Cite this article: Bone Joint J 2017;99-B: Fractures of the hip are arguably the most important public health problem faced by orthopaedic surgeons. More than lowenergy fractures of the hip occur in the United States each year. 1 The International Osteoporosis Foundation 2 estimates that approximately 1.6 million such fractures occur per year worldwide, a figure which may rise to six million by Their treatment consumes a growing percentage of healthcare expenditure. 2 The goals of care are to restore function with the lowest possible rate of surgical and medical complications. Achieving stable reduction and fixation of the fracture, permitting immediate mobilisation, is key to these goals. The last 25 years has seen the introduction and increasing acceptance and use of intramedullary devices in preference to the sliding hip screw (SHS) for the treatment of intertrochanteric fractures. 3 Despite little good quality evidence to support their use, most clinicians use them for most fractures, and practice guidelines from the American Association of Orthopaedic Surgery (AAOS) 4 support their use. 5 In this paper we review the evidence and highlight the weaknesses or absence. We also present our practice with its rationale and recommendations. The pertrochanteric region has many thickenings of trabecular bone in compressive and tensile groups. The most structurally significant of these are the primary compressive trabeculae along the posteromedial femoral neck and shaft. 6 While several classification systems exist for these fractures, they are all based on the concept of stability. A stable fracture is a simple one that, once reduced and fixed, is compressed and minimally impacted by the nearly perpendicular weight-bearing force of single leg stance. 7 Unstable fractures due either to comminution, reverse oblique orientation, or both, are associated with collapse on axial loading. Both the posteromedial cortex and the lateral cortical buttress beneath the vastus ridge contribute to the stability of these fractures. The instability increases with the degree of comminution of the posteromedial cortex. Increased comminution implies less support for axial loading through cortical contact. The lateral cortex beneath the vastus ridge provides the final buttress to impaction of the fracture after fixation, further contributing to its stability and avoiding collapse. 8,9 Incompetence of either of these cortical regions therefore renders a fracture unstable. 128 THE BONE & JOINT JOURNAL

2 IMPLANT OPTIONS FOR THE TREATMENT OF INTERTROCHANTERIC FRACTURES OF THE HIP 129 Fig. 1 Arbeitsgemeinschaft für Osteosynthesefragen/Orthopaedic Trauma Association classification of pertrochanteric fractures. Copyright the AO Foundation, Switzerland. (Source: AO Surgery Reference). The quality of reduction, regardless of the pattern of the fracture, is one of the most important modifiable factors in the management of intertrochanteric fractures. 10 The ability of the cortex to resist collapse in stable fractures depends on the restoration of cortical continuity. In unstable fractures, where the ability to restore this is limited by comminution or the orientation of the fracture, the boneimplant interface will be subject to increased stress. Implants used to fix unstable fractures must, therefore, be capable of bearing more load, in order to avoid loss of reduction through collapse. A valgus reduction is one way of increasing interfragmentary compression and of reducing bone-implant stresses, as well as minimising collapse and leg length discrepancy. Studies of implant loading confirm that the load borne by an implant is increased with varus mal-reduction and with decreasing stability of the fracture, and that in both cases an intramedullary device can bear greater load than an extramedullary device. 11,12 The Arbeitsgemeinschaft für Osteosynthesefragen/Orthopaedic Trauma Association classification system 13 (Fig. 1) offers both a universal descriptive language and reliable evidence on prognosis, 7 and is also useful for research and has been shown to have greater inter-observer reliability than other classification systems. 14 Simple, two-part pertrochanteric fractures with a single line of extension into the medial cortex are classified as 31-A1. These fractures are stable after reduction because bony continuity is restored and a significant proportion of the load transmitted through the proximal femur can be safely borne. Multi-fragmentary pertrochanteric fractures beginning anywhere in the greater trochanter with medial extension in two or more places creating a third fracture fragment, including the lesser trochanter, are classified as 31-A2. Many of these are unstable after reduction because the medial buttress is compromised. Fracture patterns 31-A1.1 to 31-A2.1 are typically stable and those from 31-A2.2 to 31-A3.3, which have more than one intermediate fragment, are usually unstable. 15 True intertrochanteric fractures with a fracture line that passes between the trochanters, above the lesser trochanter medially and below the vastus ridge laterally are classified as 31-A3. These fractures are unstable after reduction because both posteromedial and lateral buttresses are lost. This classification is particularly valuable in its identification of reverse oblique and intertrochanteric subtrochanteric fractures as challenging to reduce and fix with stability. Involvement of the lateral cortex is only a specific criterion of fractures in the 31-A3 category and is an important factor in loss of reduction when using an extramedullary SHS system. 16 Intra-operative fracture of the lateral cortical buttress is significantly more common in A2.2 and A2.3 fractures than in A2.1 fractures, 8 emphasising the importance of recognising the A2 sub-type when choosing the implant by which it will be managed. A surgeon s familiarity with a specific implant may have a measurable effect on the operating time, the use of fluoroscopy, intra-operative blood loss and the need for transfusion, but a systematic review of the evidence reported significant heterogeneity in published data and an implantassociated effect has not therefore been shown. 17 There is an increasing cost associated with SHSs, short nails and long nails, respectively. 18,19 Intramedullary implants are between 20% and 40% more expensive than SHSs and so, if each is appropriate for fracture fixation, the less expensive device should be used. A recent American College of Surgeons National Surgical Quality Improvement Program database study reported a shorter mean post-operative length of stay following treatment with an intramedullary implant. 20 This study did not stratify the patients by the type of fracture and hence the effect was seen across all groups, implying that the greater cost of the implant may be offset against the ongoing cost of care by using the intramedullary device in all patients. Cost-effectiveness analysis that models length of stay, rehabilitation or re-operation is dependent on the data used for the model, in particular those used to inform the rate of failure. One such study suggested that extramedullary VOL. 99-B, No. 1, JANUARY 2017

3 130 A. R. SOCCI, N. E. CASEMYR, M. P. LESLIE, M. R. BAUMGAERTNER implants are more cost-effective for A2 fractures, but assumed an equivalent rate of failure between SHSs and nails. 18 Such an assumption ignores differences in outcomes, especially when considering the A2 group as a homogenous one. Late costs are more difficult to measure and tend to relate to failure, requiring prolonged rehabilitation or reoperation. 21 First-generation intramedullary implants had a significantly higher incidence of periprosthetic fracture, thereby compounding their already increased cost. A 4.5- fold increased risk of periprosthetic fracture in 1997 decreased to 1.87 in 2005 and is no longer evident in the most recent data. 22,23 Despite a large number of randomised controlled trials (RCTs), meta-analyses have not shown differences in functional outcomes between intra- and extra-medullary devices. These studies have often used inappropriate outcome measures and reported the outcome of diverse cohorts of patients, pooled without reference to pre-operative functional status. 17,24 Some authors have, however, shown differences in the rates of recovery of function. In a RCT of intramedullary versus extramedullary fixation in 210 patients with A1 or A2 fractures, Utrilla et al 25 reported no difference in outcome in stable fractures, but better mobility at one year following intramedullary fixation of unstable fractures. Improved early functional recovery following intramedullary fixation has also been shown in several studies which also recorded sliding of SHSs, indicating shortening of the femoral neck, leg length discrepancy, and/or medialisation of the shaft Pajarinen et al 26 reported improved mobility and less shortening of the femoral neck after intramedullary fixation in agreement with the findings of Hardy et al 27 of improved early mobility, less sliding and less leg length discrepancy. The early functional recovery is better in those in whom the SHS slides less, suggesting that excessive alteration of the biomechanics of the hip can impair the recovery of mobility. A recent RCT by Reindl et al 29 found less shortening of the femoral neck in the intramedullary group, but this did not significantly alter the timed up and go scores. It is likely that the ability of a patient to tolerate malunion is related to pre-injury function, but there is currently no evidence to support this as studies tend to pool all patients. Information about functional outcomes such as return to social integration, driving and participation in recreational activities is urgently needed. Implant details The length of the nail. Early intramedullary nails used for fractures of the hip were short and associated with a risk of fracture at the tip of the nail, with an incidence of 8% to 11% historically. 30 Long nails can prevent this complication but risk anterior cortical impingement or intraoperative iatrogenic fracture with incidences of 1.5% 30 and < 1% recorded. 31,32 For this reason, the surgeon must be conscious of the design of the nail including its radius of curvature and the anterior bow of the femur. A stress riser can, in this scenario, lead to a fracture which is even more challenging to manage, especially with a nail in situ. No significant difference in the rates of periprosthetic fracture between short and long nails have been recorded at one, two or five years post-operatively. 21,30 Longer nails can cost up to 45% more than short nails. 30 Lindvall et al 21 reported that, based on a rate of periprosthetic fracture of 12% and an increased cost of $439 of long nails over short, long nails were cheaper than short ones when the cost of managing complications was taken into account. A more recent study reported equivalence between short and long nails, except for hospital costs (which can include facility fees, cost of medications given, administration of the medications, lab costs, cost of surgery and anesthesia). 19 Leaving aside the biomechanically dissimilar subtrochanteric fractures, there is little evidence relating to short versus long nails in terms of functional outcome and until these data are known, meaningful assessment of cost benefit is difficult. Distal locking. The effect of distal locking on the incidence of periprosthetic fracture is not fully understood. Lindvall et al 21 reported that distal locking may be protective against periprosthetic fracture in both short and long nails. In their cohort of 609 patients, 47% of nails were distally locked but 15 of 16 periprosthetic fractures occurred in the 53% of patients with unlocked nails at 21.5 months (standard deviation (SD) 18.6) follow-up in the short nail group and 14.8 months (SD 13.8) follow-up for long nails. They also noted that patients with unlocked long nails experienced periprosthetic fracture around the shaft, while those with locked long nails had fractures at the tip of the nail. In short unlocked nails, fractures occurred at the tip of the nail, in keeping with early findings in studies of intramedullary fixation This is important as, in relative terms, it is easier to treat a diaphyseal fracture by closed reduction with the introduction of a distal locking screw than it is to perform open reduction and internal fixation of a peri-articular fracture around a locked nail. Current evidence supports distal locking in axially or rotationally unstable fractures, including those with subtrochanteric extension or comminution, or in osteoporotic femora in which it is hard to achieve an isthmic fit. 32 Femoral head fixation: blade vs screw, one vs two Some intramedullary implants use a helical blade for fixation of the femoral head instead of a cancellous screw. Biomechanical studies in artificial, cadaveric and living bone have shown that a blade resists rotational and translational displacement better 36,37 but this has not translated into evidence of a difference in rates of cut-out, complications, or post-operative function. 38,39 Consistently, loss of fixation is more related to the quality of reduction of the fracture and a high tip-apex distance (TAD) rather than the choice of implant. 37,39 THE BONE & JOINT JOURNAL

4 IMPLANT OPTIONS FOR THE TREATMENT OF INTERTROCHANTERIC FRACTURES OF THE HIP 131 The use of blades may be associated with the phenomenon of cut-through with medial perforation of the articular surface of the femoral head without loss of reduction of the fracture. 40 This is in contrast to cut-out, superior and anterior migration of the screw as the fracture displaces and the femoral head moves into varus and retroversion. In order to avoid cut-through, it has been suggested that the full length of the trajectory of the blade is not pre-drilled and that the tip of the blade should be a minimum of 10 mm from the joint surface. 40,41 There are variants with both one (uniaxial) and two (biaxial) screws in the femoral head. Greater load to failure has been reported with biaxial screws in a biomechanical study, 42 but this advantage has not been reported clinically and complications of axial and reverse axial migration (the Z and reverse Z effects) have been described. 43 A prospective RCT by Kouvidis et al 43 showed no difference in outcome (activities of daily living or mobility) between an SHS group (60 patients) and a biaxial intramedullary nail group (62 patients) at 1 year follow-up. The single systematic review of such devices suggests that biaxial fixation leads to a lower risk of post-operative periprosthetic fracture than uniaxial fixation. 44 This review examined only periprosthetic fracture however, without recording any other outcome or complication, including those specific to biaxial fixation. Proximal femoral locking plates offer a lower load to failure than cephalomedullary devices and have higher rates of mechanical failure and re-operation. 45 Much of the evidence supporting the use of newer designs of implant is biomechanical and must be interpreted carefully to understand the clinical relevance. The native hip may be loaded in different directions and different ways and the evidence should be evaluated with caution. Tip-apex distance and other targets Both cut-out and cut-through emphasise the importance of the correct placement of the screw in the femoral head, namely central and deep. 46 The TAD is an accepted and reproducible means of evaluating this. 46 A recent variation, the calcar-tad, aims to target a slightly inferior placement of the screw when using cephalomedullary devices. This reflects existing practice in both SHSs and intramedullary devices, based on there being stronger bone in the calcar and inferior femoral head. 47,48 It has not, however, been confirmed that this offers better fixation than when placing the screw in the centre of the femoral head. 49,50 In the multivariate analysis of Kashigar et al, 49 only the calcar-tad was found to be a statistically significant predictor of failure and although the authors concluded that inferior placement conferred benefit, the reduction of the fracture was classified as poor in almost half of the 77 patients at a mean follow-up of 408 days (81 days to 4.9 years). This, therefore, confounded the suggestion that the position of the screw was predictive of the outcome, as varus mal-reduction was also significantly associated with failure. Discussion Stable intertrochanteric fractures (A1 to A2.1). There is currently little evidence of the superiority of one device over another in the management of these fractures. 5,25,28,51 The quality of reduction remains paramount, with stable fractures having direct cortical contact following accurate reduction. At our institution there is a preference for SHS fixation after careful reduction. Subtrochanteric and reverse oblique fractures (A3). There is strong evidence to support the use of intramedullary fixation in subtrochanteric and reverse oblique fractures. The biomechanics of these fractures are such that fixation with a SHS is inappropriate, as the line of collapse is not perpendicular to the fracture line and the lateral cortical buttress cannot resist collapse. The use of a trochanteric sliding plate (TSP) offers increased resistance to collapse, but the fixation is less reliable than when using an intramedullary device, which offers better functional outcomes and rates of union. 27,52,53-57 Worse outcomes are also associated with the use of proximal femoral locking plates. 58 Intramedullary nails resist collapse both by their intramedullary position and by an enlarged proximal end, offering an internal buttress. The minimally invasive nature of the surgery also preserves the vastus soft tissue envelope, maintaining stability and vascularity. Unstable intertrochanteric fractures (A2.2 to A2.3). Current guidelines recommend the use of an intramedullary device for the treatment of these fractures. 5 As in subtrochanteric and reverse oblique fractures, the use of a TSP may prevent excessive collapse in unstable fractures lacking a posteromedial or lateral cortical buttress in which an SHS is used. A randomised prospective study comparing fixation using a SHS (343 patients) or intramedullary nail (341 patients) for all intertrochanteric and subtrochanteric fractures showed no long-term differences in outcome at three and 12 month follow-up, with improved early pain on mobilisation scores in the intramedullary group. 59 Importantly, a TSP was used in A3 fractures and optional in osteoporotic A1 and A2 fractures when the lateral cortical buttress was questionable. The equivalence of outcome between these constructs emphasises that maintenance of a good reduction and prevention of collapse offers optimum function. Medialisation of the femoral shaft was seen more frequently in the SHS/TSP group and was associated with increased post-operative pain. Overall, an intramedullary device appears from the evidence to offer more benefit than an SHS with TSP. Basicervical fracture. True basicervical fractures are uncommon and, although extracapsular, have similar rotational instability to those occurring more medially in the femoral neck. A simple two-part basicervical fracture requires reduction with direct cortical contact, and stable, compressed fixation. A de-rotation screw may offer increased protection against rotational loss of reduction. A comminuted fracture with basicervical extension should be treated VOL. 99-B, No. 1, JANUARY 2017

5 132 A. R. SOCCI, N. E. CASEMYR, M. P. LESLIE, M. R. BAUMGAERTNER as a comminuted, unstable intertrochanteric fracture with an intramedullary device. The role of arthroplasty. Certain unstable intertrochanteric fractures of the hip can be treated by arthroplasty. Indications include existing arthritis, osteonecrosis of the femoral head, poor quality of bone or complications following internal fixation. 60 There is limited evidence for the role of arthroplasty in treating unstable intertrochanteric fractures, with benefits including early mobilisation. 60 The surgery is technically challenging and should be undertaken by an experienced arthroplasty surgeon. In conclusion, decision-making in the management of intertrochanteric fractures requires a thorough understanding of the stability of the fracture and the implants which are available. More evidence on functional outcomes is needed, especially in higher-functioning patients. It remains of critical importance to pay careful attention to the quality of the reduction and central and deep positioning of the screw or blade, regardless of implant. Take home message: There is conflicting evidence about the choice of implant in the treatment of intertrochanteric fractures of the hip, but assessing the stability of the fracture allows one to choose between fixation using a SHS or an intramedullary device. Author contributions: A. R. Socci: Primary researcher and writer. N. E. Casemyr: Researched and wrote significant portion of introduction, Editing and revising. M. P. Leslie: Wrote a section on technical aspects of reduction which was not included, Participated in revision process. M. R. Baumgaertner: Provided necessary oversight and guidance, Focused the work on the most relevant references, concepts and controversies, Heavily involved in editing and revision. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. This article was primary edited by P. Page and first proof edited by J. Scott. References 1. No authors listed. Centers for Disease Control and Prevention: Hip Fractures Among Older Adults. (date last accessed 16 November 2016). 2. No authors listed. International Osteoporosis Foundation: Hip Fracture. (date last accessed 16 November 2016). 3. Chen F, Wang Z, Bhattacharyya T. Convergence of outcomes for hip fracture fixation by nails and plates. Clin Orthop Relat Res 2012;471: No authors listed. American Academy of Orthopaedic Surgeons: Management of Hip Fractures in the Elderly. (date last accessed 16 November 2016). 5. Roberts KC, Brox WT, Jevsevar DS, Sevarino K. Management of hip fractures in the elderly. J Am Acad Orthop Surg 2015;23: Griffen JB. The calcar redefined. Clin Orthop Relat Res 1982;164: Browner BD, Jupiter JB, Krettek C, Anderson PA. Intertrochanteric Hip Fractures. In: Skeletal Trauma: Basic Science, Management, and Reconstruction. Amsterdam: Elsevier, 2015: Palm H, Jacobsen S, Sonne-Holm S, Gebuhr P, Hip Fracture Study Group. Integrity of the lateral femoral wall in intertrochanteric hip fractures: an important predictor of a reoperation. J Bone Joint Surg [Am] 2007;89-A: Im GI, Shin YW, Song YJ. Potentially unstable intertrochanteric fractures. J Orthop Trauma 2005;19: Kaufer H. Mechanics of the treatment of hip injuries. Clin Orthop Relat Res 1980;146: Marmor M, Liddle K, Buckley J, Matityahu A. Effect of varus and valgus alignment on implant loading after proximal femur fracture fixation. Eur J Orthop Surg Traumatol 2016;26: Marmor M, Liddle K, Pekmezci M, Buckley J, Matityahu A. The effect of fracture pattern stability on implant loading in OTA type 31-A2 proximal femur fractures. J Orthop Trauma 2013;27: No authors listed. AO/OTA Fracture and Dislocation Classification. aotrauma.aofoundation.org/structure/education/self-directed-learning/referencematerials/classifications/pages/ao-ota-classification.aspx (date last accessed 17 November 2016). 14. Jin WJ, Dai LY, Cui YM, et al. Reliability of classification systems for intertrochanteric fractures of the proximal femur in experienced orthopaedic surgeons. Injury 2005;36: Lorich DG, Geller DS, Nielson JH. Osteoporotic pertrochanteric hip fractures: management and current controversies. Instr Course Lect 2004;53: Lindskog DM, Baumgaertner MR. Unstable intertrochanteric hip fractures in the elderly. J Am Acad Orthop Surg 2004;12: Parker MJ, Handoll HH. Gamma and other cephalocondylic intramedullary nails versus extramedullary implants for extracapsular hip fractures in adults. Cochrane Database of Syst Rev 2010;CD Swart E, Makhni EC, Macaulay W, Rosenwasser MP, Bozic KJ. Cost-effectiveness analysis of fixation options for intertrochanteric hip fractures. J Bone Joint Surg [Am] 2014;96: Krigbaum H, Takemoto S, Kim HT, Kuo AC. Costs and Complications of Short Versus Long Cephalomedullary Nailing of OTA 31-A2 Proximal Femur Fractures in U.S. Veterans. J Orthop Trauma 2016;30: Bohl DD, Basques BA, Golinvaux NS, et al. Extramedullary compared with intramedullary implants for intertrochanteric hip fractures: thirty-day outcomes of 4432 procedures from the ACS NSQIP database. J Bone Joint Surg [Am] 2014;96: Lindvall E, Ghaffar S, Martirosian A, Husak L. Short vs Long Intramedullary Nails in the Treatment of Pertrochanteric Hip Fractures. J Orthop Trauma 2016;30: Bhandari M, Schemitsch E, Jönsson A, Zlowodzki M, Haidukewych GJ. Gamma nails revisited: gamma nails versus compression hip screws in the management of intertrochanteric fractures of the hip: a meta-analysis. J Orthop Trauma 2009;23: Basques BA, Bohl DD, Golinvaux NS, et al. Postoperative length of stay and 30- day readmission after geriatric hip fracture: an analysis of 8434 patients. J Orthop Trauma 2015;29: Liu P, Wu X, Shi H, et al. Intramedullary versus extramedullary fixation in the management of subtrochanteric femur fractures: a meta-analysis. Clin Interv Aging 2015;10: Utrilla AL, Reig JS, Muñoz FM, Tufanisco CB. Trochanteric gamma nail and compression hip screw for trochanteric fractures: a randomized, prospective, comparative study in 210 elderly patients with a new design of the gamma nail. J Orthop Trauma 2005;19: Pajarinen J, Lindahl J, Michelsson O, Savolainen V, Hirvensalo E. Pertrochanteric femoral fractures treated with a dynamic hip screw or a proximal femoral nail. A randomised study comparing post-operative rehabilitation. J Bone Joint Surg [Br] 2005;87-B: Hardy DC, Descamps PY, Krallis P, et al. Use of an intramedullary hip-screw compared with a compression hip-screw with a plate for intertrochanteric femoral fractures. A prospective, randomized study of one hundred patients. J Bone Joint Surg [Am] 1998;80-A: Ahrengart L, Törnkvist H, Fornander P, et al. A randomized study of the compression hip screw and Gamma nail in 426 fractures. Clin Orthop Relat Res 2002;401: Reindl R, Harvey EJ, Berry GK, Rahme E; Canadian Orthopaedic Trauma Society (COTS). Intramedullary Versus Extramedullary Fixation for Unstable Intertrochanteric Fractures: A Prospective Randomized Controlled Trial. J Bone Joint Surg [Am] 2015;97: Kleweno C, Morgan J, Redshaw J, et al. Short versus long cephalomedullary nails for the treatment of intertrochanteric hip fractures in patients older than 65 years. J Orthop Trauma 2014;28: Boone C, Carlberg KN, Koueiter DM, et al. Short versus long intramedullary nails for treatment of intertrochanteric femur fractures (OTA 31-A1 and A2). J Orthop Trauma 2014;28: Kanakaris NK, Tosounidis TH, Giannoudis PV. Nailing intertrochanteric hip fractures: short versus long; locked versus nonlocked. J Orthop Trauma 2015;29:S10 S Bridle SH, Patel AD, Bircher M, et al. Fixation of intertrochanteric fractures of the femur. A randomised prospective comparison of the gamma nail and the dynamic hip screw. J Bone Joint Surg [Br] 1991;73-B: Radford PJ, Needoff M, Webb JK. A prospective randomised comparison of the dynamic hip screw and the gamma locking nail. J Bone Joint Surg [Br] 1993;75-B: THE BONE & JOINT JOURNAL

6 IMPLANT OPTIONS FOR THE TREATMENT OF INTERTROCHANTERIC FRACTURES OF THE HIP Robinson CM, Adams CI, Craig M, et al. Implant-related fractures of the femur following hip fracture surgery. J Bone Joint Surg [Am] 2002;84-A: O Neill F, Condon F, McGloughlin T, et al. Dynamic hip screw versus DHS blade: a biomechanical comparison of the fixation achieved by each implant in bone. J Bone Joint Surg [Br] 2011;93-B: Fang C, Lau TW, Wong TM, Lee HL, Leung F. Sliding hip screw versus sliding helical blade for intertrochanteric fractures: a propensity score-matched case control study. Bone Joint J 2015;97-B: Huang X, Leung F, Liu M, et al. Is helical blade superior to screw design in terms of cut-out rate for elderly trochanteric fractures? A meta-analysis of randomized controlled trials. Eur J Orthop Surg Traumatol 2014;24: Stern R, Lübbeke A, Suva D, Miozzari H, Hoffmeyer P. Prospective randomised study comparing screw versus helical blade in the treatment of low-energy trochanteric fractures. Int Orthop 2011;35: Frei HC, Hotz T, Cadosch D, Rudin M, Käch K. Central head perforation, or cut through, caused by the helical blade of the proximal femoral nail antirotation. J Orthop Trauma 2012;26: Brunner A, Jöckel JA, Babst R. The PFNA proximal femur nail in treatment of unstable proximal femur fractures--3 cases of postoperative perforation of the helical blade into the hip joint. J Orthop Trauma 2008;22: Kubiak EN, Bong M, Park SS, et al. Intramedullary fixation of unstable intertrochanteric hip fractures: one or two lag screws. J Orthop Trauma 2004;18: Kouvidis G, Sakellariou VI, Mavrogenis AF, et al. Dual lag screw cephalomedullary nail versus the classic sliding hip screw for the stabilization of intertrochanteric fractures. A prospective randomized study. Strategies Trauma Limb Reconstr 2012;7: Norris R, Bhattacharjee D, Parker MJ. Occurrence of secondary fracture around intramedullary nails used for trochanteric hip fractures: a systematic review of 13,568 patients. Injury 2012;43: Streubel PN, Moustoukas M, Obremskey WT. Locked plating versus cephalomedullary nailing of unstable intertrochanteric femur fractures. Eur J Orthop Surg Traumatol 2016;26: Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM. The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg [Am] 1995;77-A: Parker MJ. Cutting-out of the dynamic hip screw related to its position. J Bone Joint Surg [Br] 1992;74-B: Kuzyk PRT, Zdero R, Shah S, et al. Femoral head lag screw position for cephalomedullary nails: a biomechanical analysis. J Orthop Trauma 2012;26: Kashigar A, Vincent A, Gunton MJ, et al. Predictors of failure for cephalomedullary nailing of proximal femoral fractures. Bone Joint J 2014;96-B: De Bruijn K, den Hartog D, Tuinebreijer W, Roukema G. Reliability of predictors for screw cutout in intertrochanteric hip fractures. J Bone Joint Surg [Am] 2012;94- A: Varela-Egocheaga JR, Iglesias-Colao R, Suárez-Suárez MA, et al. Minimally invasive osteosynthesis in stable trochanteric fractures: a comparative study between Gotfried percutaneous compression plate and Gamma 3 intramedullary nail. Arch Orthop Trauma Surg 2009;129: Babst R, Renner N, Biedermann M, et al. Clinical results using the trochanter stabilizing plate (TSP): the modular extension of the dynamic hip screw (DHS) for internal fixation of selected unstable intertrochanteric fractures. J Orthop Trauma 1998;12: Madsen JE, Naess L, Aune AK, et al. Dynamic hip screw with trochanteric stabilizing plate in the treatment of unstable proximal femoral fractures: a comparative study with the Gamma nail and compression hip screw. J Orthop Trauma 1998;12: Miedel R, Ponzer S, Törnkvist H, Söderqvist A, Tidermark J. The standard Gamma nail or the Medoff sliding plate for unstable trochanteric and subtrochanteric fractures. A randomised, controlled trial. J Bone Joint Surg [Br] 2005;87-B: Sadowski C, Lübbeke A, Saudan M, et al. Treatment of reverse oblique and transverse intertrochanteric fractures with use of an intramedullary nail or a 95 degrees screw-plate: a prospective, randomized study. J Bone Joint Surg [Am] 2002;84- A: Schipper IB, Steyerberg EW, Castelein RM, et al. Treatment of unstable trochanteric fractures randomised comparison of the gamma nail and the proximal femoral nail. J Bone Joint Surg [Br]2004;86-B: Matre K, Havelin LI, Gjertsen JE, et al. Sliding hip screw versus IM nail in reverse oblique trochanteric and subtrochanteric fractures. A study of 2716 patients in the Norwegian Hip Fracture Register. Injury 2013;44: Kuzyk PRT, Lobo J, Whelan D, et al. Biomechanical evaluation of extramedullary versus intramedullary fixation for reverse obliquity intertrochanteric fractures. J Orthop Trauma 2009;23: Matre K, Vinje T, Havelin LI, et al. TRIGEN INTERTAN intramedullary nail versus sliding hip screw: a prospective, randomized multicenter study on pain, function, and complications in 684 patients with an intertrochanteric or subtrochanteric fracture and one year of follow-up. J Bone Joint Surg [Am] 2013;95: Mäkinen TJ, Gunton M, Fichman SG, et al. Arthroplasty for Pertrochanteric Hip Fractures. Orthop Clin North Am 2015;46: VOL. 99-B, No. 1, JANUARY 2017

9/24/2015. When Can I Use a SHS? When CAN T I Use a SHS? Sliding Hip Screw. Time proven. Technically simple. Cheap. Quick

9/24/2015. When Can I Use a SHS? When CAN T I Use a SHS? Sliding Hip Screw. Time proven. Technically simple. Cheap. Quick When Can I Use a SHS? Frank A. Liporace, MD Associate Professor Director of Orthopaedic Trauma Research Director of Orthopaedic Trauma Jersey City Medical Center New York University / Hospital for Joint

More information

A Comparison of the Proximal Femoral Nail Antirotation Device and Dynamic Hip Screw in the Treatment of Unstable Pertrochanteric Fracture

A Comparison of the Proximal Femoral Nail Antirotation Device and Dynamic Hip Screw in the Treatment of Unstable Pertrochanteric Fracture The Journal of International Medical Research 2010; 38: 1266 1275 A Comparison of the Proximal Femoral Nail Antirotation Device and Dynamic Hip Screw in the Treatment of Unstable Pertrochanteric Fracture

More information

The Lateral Trochanteric Wall A Key Element in the Reconstruction of Unstable Pertrochanteric Hip Fractures

The Lateral Trochanteric Wall A Key Element in the Reconstruction of Unstable Pertrochanteric Hip Fractures CLINICAL ORTHOPAEDICS AND RELATED RESEARCH Number 425, pp. 82 86 2004 Lippincott Williams & Wilkins The Lateral Trochanteric Wall A Key Element in the Reconstruction of Unstable Pertrochanteric Hip Fractures

More information

Peritroch Hip Fractures. Robert M Harris MD. Hip Fractures. Factors Influencing Construct Strength: Uncontrolled factors 4/28/2016

Peritroch Hip Fractures. Robert M Harris MD. Hip Fractures. Factors Influencing Construct Strength: Uncontrolled factors 4/28/2016 Peritroch Hip Fractures Should be treated with an IMHS Robert M Harris MD Hip Fractures General principles Approximately 250,000 hip fractures/ year Cost approximately $8.7 billion annually The number

More information

ICUC Paper. The treatment of trochanteric fractures revisited: Pietro Regazzoni, Alberto Fernandez, Dominik Heim, Stephan M. Perren.

ICUC Paper. The treatment of trochanteric fractures revisited: Pietro Regazzoni, Alberto Fernandez, Dominik Heim, Stephan M. Perren. The treatment of trochanteric fractures revisited: Pietro Regazzoni, Alberto Fernandez, Dominik Heim, Stephan M. Perren September 2016 An optimal treatment of hip fractures is crucial because of the great

More information

Technique Guide. DHS Blade. For osteoporotic bone.

Technique Guide. DHS Blade. For osteoporotic bone. Technique Guide DHS Blade. For osteoporotic bone. Table of Contents Introduction Features and Benefits 2 Indications and Contraindications 4 Clinical Cases 5 Surgical Technique Implantation 6 Implant

More information

Randomized comparative study to evaluate the role of proximal femoral nail and dynamic hip screw in unstable trochanteric fractures

Randomized comparative study to evaluate the role of proximal femoral nail and dynamic hip screw in unstable trochanteric fractures International Journal of Research in Orthopaedics Mayi SC et al. Int J Res Orthop. 2016 Sep;2(3):75-79 http://www.ijoro.org Research Article DOI: http://dx.doi.org/10.18203/issn.2455-4510.intjresorthop20162618

More information

ABSTRACT INTRODUCTION. Hui Xie 1,*, Zhan Wang 2,*, Junji Zhang 3, Langhai Xu 2 and Bao Chen 1. Research Paper

ABSTRACT INTRODUCTION. Hui Xie 1,*, Zhan Wang 2,*, Junji Zhang 3, Langhai Xu 2 and Bao Chen 1. Research Paper /, 2017, Vol. 8, (No. 47), pp: 82700-82704 Clinical outcome of dynamic hip locking plates and proximal femoral nails anti-rotation-asia for treating intertrochanteric femur fracture with lateral wall fractures

More information

HOW TO CITE THIS ARTICLE:

HOW TO CITE THIS ARTICLE: A COMPARATIVE STUDY OF FUNCTIONAL OUTCOME BETWEEN DYNAMIC HIP SCREW AND PROXIMAL FEMORAL NAIL IN SURGICAL MANAGEMENT OF PER-TROCHANTERIC FRACTURES Umesh M. Shivanna 1, Girish H. Rudrappa 2 HOW TO CITE

More information

Comparitive Study between Proximal Femoral Nailing and Dynamic Hip Screw in Intertrochanteric Fracture of Femur *

Comparitive Study between Proximal Femoral Nailing and Dynamic Hip Screw in Intertrochanteric Fracture of Femur * Open Journal of Orthopedics, 2013, 3, 291-295 Published Online November 2013 (http://www.scirp.org/journal/ojo) http://dx.doi.org/10.4236/ojo.2013.37053 291 Comparitive Study between Proximal Femoral Nailing

More information

Functional evaluation of proximal femoral fractures managed with cephalomedullary nailing by oxford hip score - A prospective study

Functional evaluation of proximal femoral fractures managed with cephalomedullary nailing by oxford hip score - A prospective study 2017; 3(3): 980-985 ISSN: 2395-1958 IJOS 2017; 3(3): 980-985 2017 IJOS www.orthopaper.com Received: 13-05-2017 Accepted: 14-06-2017 Pranav Kothiyal Senior Resident, Department of Orthopaedics, SGRRIMHS,

More information

More than 1.5 million

More than 1.5 million Clin Orthop Relat Res (2015) 473:767 774 / DOI 10.1007/s11999-014-4-7 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons Published online: 6 January

More information

Assessment of Prognosis of Patients with Intertrochanteric Fractures Undergoing Treatment with PFN: An Observational Study

Assessment of Prognosis of Patients with Intertrochanteric Fractures Undergoing Treatment with PFN: An Observational Study Original article: Assessment of Prognosis of Patients with Intertrochanteric Fractures Undergoing Treatment with PFN: An Observational Study Gajraj Singh 1, Sandhya Gautam 2 1Assistant Professor, Department

More information

Failed Subtrochanteric Fracture How I Decide What to Do?

Failed Subtrochanteric Fracture How I Decide What to Do? Failed Subtrochanteric Fracture How I Decide What to Do? Gerald E. Wozasek Thomas M. Tiefenboeck 5 October 2016, Washington Medical University of Vienna, Department of Trauma Surgery ordination @wozasek.at

More information

ORIGINAL ARTICLE. INTER TROCHANTERIC # NECK FEMUR FIXATION WITH TFN 250 CASES. Prasad Vijaykumar Joshi, Chandrashekar Yadav.

ORIGINAL ARTICLE. INTER TROCHANTERIC # NECK FEMUR FIXATION WITH TFN 250 CASES. Prasad Vijaykumar Joshi, Chandrashekar Yadav. INTER TROCHANTERIC # NECK FEMUR FIXATION WITH TFN 250 CASES. Prasad Vijaykumar Joshi, Chandrashekar Yadav. 1. Assistant Professor. Department of Orthopaedics, Joshi Hospital Pvt. Ltd. Phaltan, Maharashtra.

More information

Page Proof 1 of 5. Fig. E1-A The INTERTAN nail was short or long.

Page Proof 1 of 5. Fig. E1-A The INTERTAN nail was short or long. Page 1 of 5 Fig. E1-A The INTERTAN nail was short or long. Fig. E1-B The sliding hip screw comes in different lengths, and is used with or without a trochanteric stabilizing plate. Page Proof 1 of 5 Page

More information

PFN in trochanteric fractures: Clinico radiological outcome study

PFN in trochanteric fractures: Clinico radiological outcome study 2018; 4(2): 943-947 ISSN: 2395-1958 IJOS 2018; 4(2): 943-947 2018 IJOS www.orthopaper.com Received: 14-02-2018 Accepted: 15-03-2018 R Amarnath Senior Asst. Professor, Department of Orthopaedic Surgery,

More information

Intertrochanteric hip fractures are

Intertrochanteric hip fractures are Surgical Management of Hip Fractures: An Evidence-based Review of the Literature. II: Intertrochanteric Fractures Kevin Kaplan, MD Ryan Miyamoto, MD Brett R. Levine, MD Kenneth A. Egol, MD Joseph D. Zuckerman,

More information

Salvage of failed dynamic hip screw fixation of intertrochanteric fractures

Salvage of failed dynamic hip screw fixation of intertrochanteric fractures Injury, Int. J. Care Injured (2005) xxx, xxx xxx 1 www.elsevier.com/locate/injury 2 3 4 5 6 7 8 8 9 10 11 12 13 14 15 Salvage of failed dynamic hip screw fixation of intertrochanteric s G.Z. Said, O. Farouk

More information

Journal of Orthopaedic Surgery and Research 2010, 5:62

Journal of Orthopaedic Surgery and Research 2010, 5:62 Journal of Orthopaedic Surgery and Research This Provisional PDF corresponds to the article as it appeared upon acceptance. Fully formatted PDF and full text (HTML) versions will be made available soon.

More information

JMSCR Vol. 03 Issue 08 Page August 2015

JMSCR Vol. 03 Issue 08 Page August 2015 www.jmscr.igmpublication.org Impact Factor 3.79 Index Copernicus Value: 5.88 ISSN (e)-2347-176x ISSN (p) 2455-0450 DOI: http://dx.doi.org/10.18535/jmscr/v3i8.08 Study of Functional and Radiological Outcome

More information

Intramedullary Nails Result in More Reoperations Than Sliding Hip Screws in Two-part Intertrochanteric Fractures

Intramedullary Nails Result in More Reoperations Than Sliding Hip Screws in Two-part Intertrochanteric Fractures Clin Orthop Relat Res (2013) 471:1379 1386 DOI 10.1007/s11999-012-2728-2 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons CLINICAL RESEARCH Intramedullary

More information

Clinical Results of Complex Subtrochanteric Femoral Fractures with Long Cephalomedullary Hip Nail

Clinical Results of Complex Subtrochanteric Femoral Fractures with Long Cephalomedullary Hip Nail ORIGINAL ARTICLE http://dx.doi.org/10.5371/hp.2017.29.2.113 Print ISSN 2287-3260 Online ISSN 2287-3279 Clinical Results of Complex Subtrochanteric Femoral Fractures with Long Cephalomedullary Hip Nail

More information

The Journal of the Korean Society of Fractures Vol.16, No.1, January, 2003

The Journal of the Korean Society of Fractures Vol.16, No.1, January, 2003 The Journal of the Korean Society of Fractures Vol16, No1, January, 2003 : 351 ( )463-712 TEL: (031) 780-5270/5271 FAX : (031) 708-3578 E-mail: bskima@netsgocom 16,, ( > 20mm ) 5, ) 20 % 1 ), 6,, 3 8 8

More information

Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study

Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study Nikoloski et al. Journal of Orthopaedic Surgery and Research 2013, 8:35 RESEARCH ARTICLE Open Access Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)?

More information

INTERTAN Nails Geared for Stability

INTERTAN Nails Geared for Stability Geared for stability The TRIGEN INTERTAN nail brings advanced TRIGEN nail technology to hip fractures. With a unique integrated, interlocking screw construct, TRIGEN INTERTAN nail provides all the benefits

More information

Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw

Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw Orthopaedics & Traumatology: Surgery & Research (2010) 96, 21 27 ORIGINAL ARTICLE Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw M. Güven a,, U. Yavuz

More information

A comparative study of 30 cases of trochanteric fracture femur treated with dynamic hip screw and proximal femoral nailing

A comparative study of 30 cases of trochanteric fracture femur treated with dynamic hip screw and proximal femoral nailing Original Article A comparative study of 30 cases of trochanteric fracture femur treated with dynamic hip screw and proximal femoral nailing Jaswinder Pal Singh Walia *, Himanshu Tailor**, H S Mann ***,

More information

Proximal Femoral Nail Antirotation in Treatment of Intertrochanteric Hip Fractures: a Retrospective Study in 113 Patients

Proximal Femoral Nail Antirotation in Treatment of Intertrochanteric Hip Fractures: a Retrospective Study in 113 Patients ORIGINAL PAPER doi: 10.5455/medarh.2015.69.352-356 Med Arh. 2015 Dec; 69(6): 353-356 Received: August 25th 2015 Accepted: October 05th 2015 2015 Sahmir Sadic, Svemir Custovic, Mahir Jasarevuc, Mirsad Fazlic,

More information

Comparative Study of Fixation Devices for Intertrochanteric Fractures

Comparative Study of Fixation Devices for Intertrochanteric Fractures Comparative Study of Fixation Devices for Intertrochanteric Fractures C. Sticlaru * A. Davidescu Politehnica University of Timişoara Politehnica University of Timişoara Timişoara, România Timişoara, România

More information

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia Types of Plates 1. New Dynamic Compression Plate: DCP Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia 1. Undercut adjacent to the holes low contact: less stress shield 2. Undercut at the undersurface

More information

Comparative study between proximal femoral nail and proximal femoral nail antirotation in management of unstable trochanteric fractures

Comparative study between proximal femoral nail and proximal femoral nail antirotation in management of unstable trochanteric fractures International Journal of Research in Orthopaedics Kashid MR et al. Int J Res Orthop. 2016 Dec;2(4):354-358 http://www.ijoro.org Original Research Article DOI: http://dx.doi.org/10.18203/issn.2455-4510.intjresorthop20164168

More information

International Journal of Research and Review E-ISSN: ; P-ISSN:

International Journal of Research and Review   E-ISSN: ; P-ISSN: International Journal of Research and Review www.ijrrjournal.com E-ISSN: 2349-9788; P-ISSN: 2454-2237 Original Research Article A Prospective Observational Study of Clinical, Functional and Radiological

More information

Lower risk of implant failure and reoperation

Lower risk of implant failure and reoperation The evidence is in... risk of implant failure and reoperation Faster time to fracture union High return to pre-fracture status How satisfied are you with current hip fracture outcomes? Around 1 in 4 hip

More information

Angular Malalignment in Subtrochanteric and Proximal Shaft Femur Fractures after Intramedullary Nailing using SIGN Nails

Angular Malalignment in Subtrochanteric and Proximal Shaft Femur Fractures after Intramedullary Nailing using SIGN Nails Angular Malalignment in Subtrochanteric and Proximal Shaft Femur Fractures after Intramedullary Nailing using SIGN Nails Rolando Junior L. Torres, MD Jeremiah R. Morales, MD, FPOA Subtrochanteric Femur

More information

Provision of Rotational Stability: Prevention of Collapse: Closed Fracture Reduction: Minimally Invasive Surgery with no Exposure of the Fracture:

Provision of Rotational Stability: Prevention of Collapse: Closed Fracture Reduction: Minimally Invasive Surgery with no Exposure of the Fracture: INTRODUCTION Percutaneous Compression Plating was developed by considering each of the stages in the surgical procedure for pertrochanteric fractures and the ways in which these might be improved. Primary

More information

Review of Proximal Nail Antirotation (PFNA) and PFNA-2 Our Local Experience

Review of Proximal Nail Antirotation (PFNA) and PFNA-2 Our Local Experience doi: 10.5704/MOJ.1107.001 Review of Proximal Nail Antirotation (PFNA) and PFNA-2 Our Local Experience WL Loo, M Med Orth, SYJ Loh, FRCS (Edin), HC Lee, FRCS (Edin) Department of Orthopaedic Surgery, Changi

More information

designed to advance the treatment of hip fractures.

designed to advance the treatment of hip fractures. designed to advance the treatment of hip fractures. introducing the tfn-advanced proximal femoral nailing system (tfna). the tfna system is a new system designed to solve a wide range of unmet needs for

More information

Intertrochanteric Femur Fractures. Alan Afsari, MD March 2014

Intertrochanteric Femur Fractures. Alan Afsari, MD March 2014 Intertrochanteric Femur Fractures Alan Afsari, MD March 2014 Topics epidemiology anatomy classification mechanism of injury patient assessment treatment rehabilitation complications Epidemiology 341k people

More information

Changes in Tip-Apex Distance by Position and Film Distance Measured by Picture Archiving and Communication System (PACS)

Changes in Tip-Apex Distance by Position and Film Distance Measured by Picture Archiving and Communication System (PACS) ORIGINAL ARTICLE Hip Pelvis 27(1): 36-42, 2015 http://dx.doi.org/10.5371/hp.2015.27.1.36 Print ISSN 2287-3260 Online ISSN 2287-3279 Changes in Tip-Apex Distance by Position and Film Distance Measured by

More information

PFNA-II. Proximal Femoral Nail Antirotation.

PFNA-II. Proximal Femoral Nail Antirotation. PFNA-II. Proximal Femoral Nail Antirotation. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. PFNA-II. Proximal Femoral

More information

The Mechanical Properties Of Fixing Greater Trochanter Or Lesser Trochanter In Complex Four Part Intertrochanteric Fractures

The Mechanical Properties Of Fixing Greater Trochanter Or Lesser Trochanter In Complex Four Part Intertrochanteric Fractures The Mechanical Properties Of Fixing Greater Trochanter Or Lesser Trochanter In Complex Four Part Intertrochanteric Fractures Chih-Hui Chen 1, Kui-Chou Huang 2, Cheng-Kung Cheng, PhD 3, Hung-Chan Kao 4.

More information

Original Article Comparative study of InterTAN and Dynamic Hip Screw in treatment of femoral intertrochanteric injury and wound

Original Article Comparative study of InterTAN and Dynamic Hip Screw in treatment of femoral intertrochanteric injury and wound Int J Clin Exp Med 2014;7(12):5578-5582 www.ijcem.com /ISSN:1940-5901/IJCEM0002927 Original Article Comparative study of InterTAN and Dynamic Hip Screw in treatment of femoral intertrochanteric injury

More information

Value Analysis Brief. Featuring TRAUMACEM V+ Augmentation System

Value Analysis Brief. Featuring TRAUMACEM V+ Augmentation System Value Analysis Brief Featuring TRAUMACEM V+ Augmentation System Executive Summary CLINICAL VALUE Unmet Need Cut-out is a major post-operative complication of intramedullary hip nailing. 1 It may also cause

More information

DISLOCATION AND FRACTURES OF THE HIP. Dr Károly Fekete

DISLOCATION AND FRACTURES OF THE HIP. Dr Károly Fekete DISLOCATION AND FRACTURES OF THE HIP Dr Károly Fekete 1 OUTLINE Epidemiology Incidence Anatomy Patient s examination, clinical symptons Diagnosis Classification Management Special complications 2 EPIDEMIOLOGY,

More information

VALUE ANALYSIS BRIEF

VALUE ANALYSIS BRIEF VALUE ANALYSIS BRIEF Executive Summary CLINICAL VALUE Unmet clinical need: Distal cortical impingement may occur in up to 25% of cases and is often the result of the curve of the natural femoral anatomy

More information

STUDY OF RESULTS OF ENDER NAILING AND CANNULATED CANCELLOUS SCREW IN THE TREATMENT OF INTERTROCHANTERIC FRACTURE FEMUR

STUDY OF RESULTS OF ENDER NAILING AND CANNULATED CANCELLOUS SCREW IN THE TREATMENT OF INTERTROCHANTERIC FRACTURE FEMUR RESEARCH ARTICLE STUDY OF RESULTS OF ENDER NAILING AND CANNULATED CANCELLOUS SCREW IN THE TREATMENT OF INTERTROCHANTERIC FRACTURE FEMUR Bhavik Dalal, Tarkik Amin, Archit Gandhi, Rohit Shah Smt NHL Municipal

More information

EVOS MINI with IM Nailing

EVOS MINI with IM Nailing Case Series Dr. John A. Scolaro EVOS MINI with IM Nailing A series of studies Introduction Intramedullary nailing has become the standard for many long bone fractures. Fracture reduction prior to nail

More information

Long-stem revision prosthesis for salvage of failed fixation of extracapsular proximal femoral fractures

Long-stem revision prosthesis for salvage of failed fixation of extracapsular proximal femoral fractures Acta Orthop. Belg., 2009, 75, 340-345 ORIGINAL STUDY Long-stem revision prosthesis for salvage of failed fixation of extracapsular proximal femoral fractures Rory J. SHARVILL, Nicholas A. FERRAN, Huw G.

More information

Distal femoral fracture with subsequent ipsilateral proximal femoral fracture

Distal femoral fracture with subsequent ipsilateral proximal femoral fracture Distal femoral fracture with subsequent ipsilateral proximal femoral fracture by M Agarwal, MS FRCS, AA Syed, FRCSI, PV Giannoudis (!), BSc,MB,MD,EEC(Orth) Dept. of Orthopaedics and Trauma, St.James University

More information

The standard Gamma nail or the Medoff sliding plate for unstable trochanteric and subtrochanteric fractures

The standard Gamma nail or the Medoff sliding plate for unstable trochanteric and subtrochanteric fractures Trauma The standard Gamma nail or the Medoff sliding plate for unstable trochanteric and subtrochanteric fractures A RANDOMISED, CONTROLLED TRIAL R. Miedel, S. Ponzer, H. Törnkvist, A. Söderqvist, J. Tidermark

More information

MRIMS Journal of Health Sciences 2016;4(1) pissn: , eissn:

MRIMS Journal of Health Sciences 2016;4(1) pissn: , eissn: MRIMS Journal of Health Sciences 216;4(1) pissn: 2321-76, eissn: 2321-7294 http://www.mrimsjournal.com/ Original Article A comparative study of proximal femoral nailing and dynamic hip screw in the management

More information

INTERTROCHANTERIC FEMORAL FRACTURES TREATED BY DYNAMIC HIP SCREW

INTERTROCHANTERIC FEMORAL FRACTURES TREATED BY DYNAMIC HIP SCREW 29 INTERTROCHANTERIC FEMORAL FRACTURES TREATED BY DYNAMIC HIP SCREW Muhammad Ayoub Laghari, Asadullah Makhdoom, Pir Abdul latif Qureshi, Abbass Memon, Faheem Ahmed Memon, Professor Khaleeque Ahmed Siddiqui

More information

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia. LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia. LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

More information

Pelvis injuries Fractures of the femur (proximal,shaft) Dr Tamás Bodzay

Pelvis injuries Fractures of the femur (proximal,shaft) Dr Tamás Bodzay Pelvis injuries Fractures of the femur (proximal,shaft) Dr Tamás Bodzay Pelvis anatomy Pelvis function - axial load bearing - protection: abdominal, pelvic structures Pelvic injury mechanism Falling from

More information

Journal of Orthopaedics, Trauma and Rehabilitation

Journal of Orthopaedics, Trauma and Rehabilitation Journal of Orthopaedics, Trauma and Rehabilitation 14 (2010) 3e7 Contents lists available at ScienceDirect Journal of Orthopaedics, Trauma and Rehabilitation Journal homepage: www.e-jotr.com Review Article

More information

Improving Outcomes. Intertrochanteric Hip Fractures. R. Trigg McClellan, MD

Improving Outcomes. Intertrochanteric Hip Fractures. R. Trigg McClellan, MD Intertrochanteric Hip Fractures Improving Outcomes R. Trigg McClellan, MD Associate Clinical Professor University of California San Francisco San Francisco General Hospital COA Annual Meeting Monterey,

More information

Dr. Balaji Douraiswami, Dr. Vijay Anand Balasubramanian, Dr. Suresh Subramani and Dr. Ramakanth Rajagopalakrishnan

Dr. Balaji Douraiswami, Dr. Vijay Anand Balasubramanian, Dr. Suresh Subramani and Dr. Ramakanth Rajagopalakrishnan 2017; 3(3): 429-435 ISSN: 2395-1958 IJOS 2017; 3(3): 429-435 2017 IJOS www.orthopaper.com Received: 19-05-2017 Accepted: 20-06-2017 Dr. Balaji Douraiswami Associate Professor, Dr. Vijay Anand Balasubramanian

More information

Treatment of extracapsular hip fractures with the proximal femoral nail (PFN) : Long term results in 45 patients

Treatment of extracapsular hip fractures with the proximal femoral nail (PFN) : Long term results in 45 patients Acta Orthop. Belg., 2004, 70, 444-454 ORIGINAL STUDY Treatment of extracapsular hip fractures with the proximal femoral nail (PFN) : Long term results in 45 patients Minos TYLLIANAKIS, Andreas PANAGOPOULOS,

More information

Dynamic Hip Screw Compared to Condylar Blade Plate in the Treatment of Unstable Fragility Intertrochanteric Fractures

Dynamic Hip Screw Compared to Condylar Blade Plate in the Treatment of Unstable Fragility Intertrochanteric Fractures Dynamic Hip Screw Compared to Condylar Blade Plate in the Treatment of Unstable Fragility Intertrochanteric Fractures CK Yong, MS Ortho, CN Tan*, MS Ortho, R Penafort**, MS Ortho, DA Singh, MS Ortho, MV

More information

Case Report Late Occurring Medial Migration of a Lag Screw in Gamma Nailing

Case Report Late Occurring Medial Migration of a Lag Screw in Gamma Nailing Case Reports in Orthopedics Volume 2016, Article ID 5201674, 4 pages http://dx.doi.org/10.1155/2016/5201674 Case Report Late Occurring Medial Migration of a Lag Screw in Gamma Nailing S. van Hoef, M. C.

More information

Geriatric Orthopaedic Trauma

Geriatric Orthopaedic Trauma Geriatric Orthopaedic Trauma EMS & Trauma Conference Seattle, WA September 2018 Lisa A. Taitsman, MD, MPH Disclosures Teach at AO courses AOA Committee OTA Committees Editorial board Journal of Orthopaedic

More information

Dynamic hip screw fixation of intertrochanteric fractures without using traction table

Dynamic hip screw fixation of intertrochanteric fractures without using traction table Acta Orthop. Belg., 2016, 82, 346-350 ORIGINAL STUDY Dynamic hip screw fixation of intertrochanteric fractures without using traction table Satish Kumar, Gaggan Naresh Chadha From the DNB orthopaedics,

More information

The Orthopaedic Enigma: A Simplified Classification

The Orthopaedic Enigma: A Simplified Classification ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 3 Number 2 The Orthopaedic Enigma: A Simplified Classification D Iyer Citation D Iyer. The Orthopaedic Enigma: A Simplified Classification. The

More information

Femoral Neck Fractures- Gold standards and breaking news. Anna Ekman, MD, PhD Stockholm South hospital Sweden

Femoral Neck Fractures- Gold standards and breaking news. Anna Ekman, MD, PhD Stockholm South hospital Sweden Femoral Neck Fractures- Gold standards and breaking news Anna Ekman, MD, PhD Stockholm South hospital Sweden Learning outcomes Learning Outcomes Epidemiology Preoperative planning Reduction techniques

More information

Standard intramedullary (IM) nails are usually used

Standard intramedullary (IM) nails are usually used 67 Effect of Piriformis Versus Trochanteric Starting Point on Fixation Stability of Short Intramedullary Reconstruction Nails Edward T. Su MD Hargovind DeWal MD Roy Sanders MD Frederick J. Kummer PhD Mohammed

More information

JOURNALOF ORTHOPAEDIC TRAUMA

JOURNALOF ORTHOPAEDIC TRAUMA JOT Special Case Report Series CASE REPORTS www.jorthotrauma.com JOURNALOF ORTHOPAEDIC TRAUMA OFFICIAL JOURNAL OF Orthopaedic Trauma Association Belgian Orthopaedic Trauma Association Canadian Orthopaedic

More information

Type I : At the level of lesser trochanter Type II : Less than 2.5 cm below lesser trochanter. Type III : cm below lesser trochanter

Type I : At the level of lesser trochanter Type II : Less than 2.5 cm below lesser trochanter. Type III : cm below lesser trochanter Type II : Major fracture line along the intertrochanteric line with communition in coronal plain. Type III : Fracture at the level of lesser trochanter with variable communition and extension in subtrochanteric

More information

A biomechanical study comparing helical blade with screw design for sliding hip fixations of unstable intertrochanteric fractures

A biomechanical study comparing helical blade with screw design for sliding hip fixations of unstable intertrochanteric fractures Title A biomechanical study comparing helical blade with screw design for sliding hip fixations of unstable intertrochanteric fractures Author(s) Luo, Q; Yuen, GWY; Lau, TW; Yeung, KWK; Leung, FKL Citation

More information

Valgus subtrochanteric osteotomy for malunited intertrochanteric fractures : Our experience in 5 cases

Valgus subtrochanteric osteotomy for malunited intertrochanteric fractures : Our experience in 5 cases Original article : Valgus subtrochanteric osteotomy for malunited intertrochanteric fractures : Our experience in 5 cases Rajendraprasad Butala *, Sunirmal Mukherjee, Prakash Samant, Ravindra Khedekar

More information

Technique Guide. DHS/DCS System. Including LCP DHS and DHS Blade.

Technique Guide. DHS/DCS System. Including LCP DHS and DHS Blade. Technique Guide DHS/DCS System. Including LCP DHS and DHS Blade. Table of Contents Introduction System Overview 2 Features and Benefits 4 Indications and Contraindications 6 Clinical Cases 8 Surgical

More information

PFNA. With Augmentation Option.

PFNA. With Augmentation Option. PFNA. With Augmentation Option. Superior anchor in osteoporotic bone Simple and reproducible procedure through standardized technique Intraoperative decision for augmentation This publication is not intended

More information

Bone&JointOutcome. The TRIGEN INTERTAN Intertrochanteric Antegrade Nail. Our Products in Clinical Studies

Bone&JointOutcome. The TRIGEN INTERTAN Intertrochanteric Antegrade Nail. Our Products in Clinical Studies Our Products in Clinical Studies Vol 04, No 01 July 2017 The TRIGEN INTERTAN Intertrochanteric Antegrade Nail A Systematic Literature Review and Meta-analysis of Clinical Outcomes Compared to Standard

More information

Plate Fixation Options

Plate Fixation Options Distal tibia extra-articular fractures can be difficult to treat Tenuous soft tissue Complex fracture patterns Plate Fixation Options Medial plating: Minimally invasive approach Technical ease Anterolateral

More information

PFNA. With Augmentation Option.

PFNA. With Augmentation Option. PFNA. With Augmentation Option. Superior anchor in osteoporotic bone Simple and reproducible procedure through standardized technique Intraoperative decision for augmentation PFNA. Proximal Femoral Nail

More information

Intramedullary fibular fixation in the operative management of fractures of the distal tibia and fibula

Intramedullary fibular fixation in the operative management of fractures of the distal tibia and fibula Royal Liverpool & Broadgreen University Hospitals NHS Foundation Trust Intramedullary fibular fixation in the operative management of fractures of the distal tibia and fibula Michael Smith MBChB, Zuned

More information

Locked Plating: Clinical Indications

Locked Plating: Clinical Indications Techniques in Orthopaedics 22(3):181 185 2007 Lippincott Williams & Wilkins, Inc. Locked Plating: Clinical Indications Kyle F. Dickson, M.D., M.B.A., John Munz, M.D. Summary: As shown in the previous article,

More information

Treatment of Femoral Intertrochanteric Fracture with Proximal Femoral Nail

Treatment of Femoral Intertrochanteric Fracture with Proximal Femoral Nail 17, 1, 20041 Journal of the Korean Fracture Society Vol. 17, No. 1, January, 2004 Treatment of Femoral Intertrochanteric Fracture with Proximal Femoral Nail Bum-Soo Kim, M.D., Sogu Lew, M.D., Sang-Hun

More information

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Technique Guide LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Table of Contents Introduction Features and Benefits 2 AO ASIF Principles 4 Indications 5 Surgical

More information

Diaphyseal Humerus Fractures. OTA Course Dallas, TX 1/20/17 Ellen Fitzpatrick MD

Diaphyseal Humerus Fractures. OTA Course Dallas, TX 1/20/17 Ellen Fitzpatrick MD Diaphyseal Humerus Fractures OTA Course Dallas, TX 1/20/17 Ellen Fitzpatrick MD OBJECTIVES TREATMENT OPTIONS SURGICAL INDICATIONS CONTROVERSIES IN MANAGEMENT Humerus Fractures Treatment Goals: Functional

More information

Surgical Technique International Version

Surgical Technique International Version Surgical Technique International Version PERI-LOC PFP 4.5mm Proximal Femur Locking Plate Surgical Technique Table of contents Product overview...2 Introduction...2 Indications...3 Case examples...4 Design

More information

Intramedullary Nailing: History & Rationale

Intramedullary Nailing: History & Rationale Intramedullary Nailing: History & Rationale Overview 1. What is IM Nailing? 2. History 3. Design Rationale & Evolution 4. Modern IM Nails 5. The Future What is IM Nailing? Method of internal fixation in

More information

Comparison of two modality of fixation in unstable trochantric fractures in elderly patients

Comparison of two modality of fixation in unstable trochantric fractures in elderly patients Original article Comparison of two modality of fixation in unstable trochantric fractures in elderly patients 1Dr. Vipin Garg, 2 Dr. Anjul Agarwal 1MS Orhtopaedics, Assistant professor, Department of orthopaedics,

More information

Predicting the Position of the Femoral Head Center

Predicting the Position of the Femoral Head Center The Journal of Arthroplasty Vol. 14 No. 1 1999 Predicting the Position of the Femoral Head Center Nobuhiko Sugano, MD, Philip C. Noble, PhD, and Emir Kamaric, MS Abstract: To find an accurate method to

More information

Orthopedics in Motion Tristan Hartzell, MD January 27, 2016

Orthopedics in Motion Tristan Hartzell, MD January 27, 2016 Orthopedics in Motion 2016 Tristan Hartzell, MD January 27, 2016 Humerus fractures Proximal Shaft Distal Objectives 1) Understand the anatomy 2) Epidemiology and mechanisms of injury 3) Types of fractures

More information

RECENT ADVANCES IN THE TREATMENT OF HIP FRACTURES IN THE ELDERLY: A REVIEW

RECENT ADVANCES IN THE TREATMENT OF HIP FRACTURES IN THE ELDERLY: A REVIEW RECENT ADVANCES IN THE TREATMENT OF HIP FRACTURES IN THE ELDERLY: A REVIEW Dr. Roshan Sah 1 * and Prof. Lui Ke bin 2 1,2 Department of Orthopaedics. The First Affiliated Peoples Hospital of Yangtze University.

More information

Locked plating constructs are creating a challenge for surgeons.

Locked plating constructs are creating a challenge for surgeons. Locked plating constructs are creating a challenge for surgeons. Three recent studies examining supracondylar femur fractures show concern for the high degree of stiffness of locked plating constructs

More information

Outcomes of Cephalomedullary Nailing in Basicervical Fracture

Outcomes of Cephalomedullary Nailing in Basicervical Fracture ORIGINAL ARTICLE Hip Pelvis 29(4): 270-276, 2017 http://dx.doi.org/10.5371/hp.2017.29.4.270 Print ISSN 2287-3260 Online ISSN 2287-3279 Outcomes of Cephalomedullary Nailing in Basicervical Fracture Seok

More information

Surgical Outcomes of Biologic Fixation for Subtrochanteric Fracture Using Locking Compression Plates

Surgical Outcomes of Biologic Fixation for Subtrochanteric Fracture Using Locking Compression Plates ORIGINAL ARTICLE Hip Pelvis 29(1): 68-76, 2017 http://dx.doi.org/10.5371/hp.2017.29.1.68 Print ISSN 2287-3260 Online ISSN 2287-3279 Surgical Outcomes of Biologic Fixation for Subtrochanteric Fracture Using

More information

Reverse LISS plate in treating femoral intertrochanteric fractures: A review of 22 cases.

Reverse LISS plate in treating femoral intertrochanteric fractures: A review of 22 cases. Biomedical Research 2017; 28 (7): 2941-2945 ISSN 0970-938X www.biomedres.info Reverse LISS plate in treating femoral intertrochanteric fractures: A review of 22 cases. Hong-Wei Chen 1*, Jun Pan 2 1 Department

More information

Aesculap Targon FN. Head Preserving Solution for Medial Femoral Neck Fractures. Aesculap Orthopaedics

Aesculap Targon FN. Head Preserving Solution for Medial Femoral Neck Fractures. Aesculap Orthopaedics Aesculap Targon FN Head Preserving Solution for Medial Femoral Neck Fractures Aesculap Orthopaedics Targon FN Operating Technique Indications for Targon FN AO 3 B. AO 3 B.2 AO 3 B.3 Undisplaced intracapsular

More information

INTERNAL FIXATION OF THE METACARPALS AND PHALANGES P. BURGE

INTERNAL FIXATION OF THE METACARPALS AND PHALANGES P. BURGE Riv Chir Mano - Vol. 43 (3) 2006 INTERNL FIXTION OF THE METCRPLS ND PHLNGES P. URGE Nuffield Orthopaedic Centre, Oxford, UK SUMMRY Techniques and instrumentation for open reduction and internal fixation

More information

Research Gaps in Care of Hip Fracture (Organized by The Hip Society and ORS) Organizers: Harry Rubash, MD Kevin Bozic, MD, MBA

Research Gaps in Care of Hip Fracture (Organized by The Hip Society and ORS) Organizers: Harry Rubash, MD Kevin Bozic, MD, MBA Research Gaps in Care of Hip Fracture (Organized by The Hip Society and ORS) Organizers: Harry Rubash, MD Kevin Bozic, MD, MBA Speakers: William Macaulay, MD Stephen Kates, MD Hip Fracture Research Gaps:

More information

We compared 54 patients treated by a Medoff

We compared 54 patients treated by a Medoff Femoral shortening in intertrochanteric fractures A COMPARISON BETWEEN THE MEDOFF SLIDING PLATE AND THE COMPRESSION HIP SCREW O. Olsson, L. Ceder, A. Hauggaard From Helsingborg Hospital, Helsingborg, Sweden

More information

Accurate and Easy Measurement of Sliding Distance of Intramedullary Nail in Trochanteric Fracture

Accurate and Easy Measurement of Sliding Distance of Intramedullary Nail in Trochanteric Fracture Original Article Clinics in Orthopedic Surgery 2015;7:152-157 http://dx.doi.org/10.4055/cios.2015.7.2.152 Accurate and Easy Measurement of Sliding Distance of Intramedullary Nail in Trochanteric Fracture

More information

Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay

Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay Reversed distal femoral locking plate for failed proximal femoral nail with nonunion of proximal femoral fractures Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay International Orthopaedics

More information

Zimmer Small Fragment Universal Locking System. Surgical Technique

Zimmer Small Fragment Universal Locking System. Surgical Technique Zimmer Small Fragment Universal Locking System Surgical Technique Zimmer Small Fragment Universal Locking System 1 Zimmer Small Fragment Universal Locking System Surgical Technique Table of Contents Introduction

More information

Cemented Hemiarthroplasty in Elderly Osteoporotic Unstable Trochanteric Fractures using Fracture Window

Cemented Hemiarthroplasty in Elderly Osteoporotic Unstable Trochanteric Fractures using Fracture Window http://dx.doi.org/10.5704/moj.1603.002 Cemented Hemiarthroplasty in Elderly Osteoporotic Unstable Trochanteric Fractures using Fracture Window Thakur A, MS Orth, Lal M, MS Orth Department of Orthopaedics,

More information

Case Report. Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology

Case Report. Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology Case Report Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology S. Robert Rozbruch, MD Hospital for Special Surgery New York, NY, USA ABSTRACT This is a case illustrating a 4.5 cm

More information

Research Article Proximal Femoral Nail versus Dynamic Hip Screw Fixation for Trochanteric Fractures: A Meta-Analysis of Randomized Controlled Trials

Research Article Proximal Femoral Nail versus Dynamic Hip Screw Fixation for Trochanteric Fractures: A Meta-Analysis of Randomized Controlled Trials Hindawi Publishing Corporation The Scientific World Journal Volume 2013, Article ID 805805, 8 pages http://dx.doi.org/10.1155/2013/805805 Research Article Proximal Femoral Nail versus Dynamic Hip Screw

More information