Mechanically evoked pain, and in particular the

Size: px
Start display at page:

Download "Mechanically evoked pain, and in particular the"

Transcription

1 ORIGINAL ARTICLE Interrater Reliability of Algometry in Measuring Pressure Pain Thresholds in Healthy Humans, Using Multiple Raters Linda S. Chesterton, PhD,* Julius Sim, PhD,* Christine C. Wright, BSc,w and Nadine E. Foster, DPhil* Objective: To evaluate interrater reliability using 5 newly trained observers in the assessment of pressure pain threshold (PPT) using a fixed-angle algometer. Methods: The study design comprised 2 phases. Phase 1: 5 undergraduate physical therapists were trained in algometry at a predefined angle, at a rate of 5 Newtons (N)/s, to the first dorsal interosseous muscle. Each observer then underwent a competency test of the application speed. The aim was to achieve repeated applications at 5 N/s without visual feedback from the algometer. Phase 2: the 5 observers measured PPT of 13 healthy volunteers, at the first dorsal interosseous muscle. The sequence of observer measurements for each participant was randomized. Mean PPT values for each observer were analyzed using repeated measures analysis of variance, intraclass correlation coefficient (ICC 2,1 ), and standard error of measurement, with 95% confidence intervals (CIs). Results: No significant differences between observers mean values were found (P = 0.094), suggesting no bias. The ICC was 0.91 (95% CI 0.82, 0.97). The standard error of measurement value was 6.27 N/cm 2 (95% CI 5.35, 7.59). Differences in PPT measurements of more than N/cm 2 (1.77 kg/cm 2 ) are likely to exceed the magnitude of measurement error, and could be used to indicate true change. This margin of error is, however, somewhat larger than a previously proposed minimum clinically important difference in PPT of N/cm 2 (1.5 kg/cm 2 ). Discussion: This study provides new evidence that trained observers can apply an algometer at a consistent rate and provide highly reliable measures of PPT in healthy humans, when PPT is calculated as the mean of 3 trials. Key Words: reliability, multiple raters, algometry, pressure pain threshold, quantitative sensory testing (Clin J Pain 2007;23: ) Received for publication June 19, 2006; revised July 11, 2007; accepted July 15, From the *Primary Care Musculoskeletal Research Centre, Keele University, Staffordshire; and wschool of Health Sciences, University of Birmingham, Edgbaston, UK. Reprints: Julius Sim, PhD, Primary Care Musculoskeletal Research Centre, Keele University, Keele, Staffordshire ST5 5BG, UK ( j.sim@keele.ac.uk). Copyright r 2007 by Lippincott Williams & Wilkins Mechanically evoked pain, and in particular the pressure pain threshold (PPT), is a popular model for inducing acute, experimental pain. Algometry is a useful technique in determining PPT measures, and has been used widely in both clinical and laboratory settings. 1 8 Pressure algometry is predominantly a manual procedure that requires a perceptual response from the participant or patient. The reliability of PPT data is, therefore, dependent upon not only the application technique of the observer, but also the ability of the patient or participant to provide a consistent verbal indication of the PPT level. Variability seen in PPT measurements could, therefore, be due to the observer (observer error), the participant (participant error), or the measurement error (the difference between each individual rating and its true value), and these are, therefore, always confounded. 9 An important source of measurement error has been identified as the rate of pressure exerted through the algometer. Indeed, maintaining a constant rate has been cited as one of the most difficult aspects of algometry. 10 Another source of error is the angle of application of the algometer to the body part, which, if varied, will also affect the reliability of the verbal report. 11 Several studies using both clinical and healthy populations have used fixed-angle algometry techniques to overcome this source of variation Results from previous PPT reliability studies suggest, in the main, high levels of interrater reliability. 10,17 26 Where poor interrater agreement has been found, 27,28 this has largely been attributed to systematic error from unskilled observers. Despite this seemingly strong body of evidence regarding interrater reliability, there are several issues of concern in relation to these findings. First, only one study has evaluated PPT measurements using more than 2 observers. 26 Yet, on occasions it may be necessary to use a larger number of observers, for example, in a study in which there are a large number of participants, 29 or in a multicenter study. In such cases, it is important to know the likely reliability of measurements taken by multiple observers. Second, many previous studies of the reliability of PPT measurement have adopted Pearson correlation coefficient for statistical analysis. 18,19,21,24,25,27 This statistical approach has been criticized as a basis for establishing interrater reliability because it represents an index of the degree of association, not of the similarity of measures or true agreement. 30 It has been argued that a 760 Clin J Pain Volume 23, Number 9, November/December 2007

2 Clin J Pain Volume 23, Number 9, November/December 2007 Interrater Reliability of Algometry more appropriate way to analyze interrater reliability of interval or ratio data is to use intraclass correlation coefficients (ICCs) The calculation of 95% confidence intervals (CIs), to identify the precision of the estimate, has also been recommended, 34,35 accompanied by the standard error of measurement (SEM), which is a measure of the (im)precision of the measurements themselves. 35 To the best of our knowledge, only 2 previous studies have reported both ICCs and estimates of error within the statistical analysis, 10,17 and from these it is not clear that observers were trained to control the application speed at a fixed level. Finally, fixed algometers, increasingly used in laboratory and some clinical experimental studies to standardize application angles, have not been investigated. The purpose of this study was, therefore, to determine the interrater reliability of PPT measurements in healthy volunteers using 5 inexperienced but specifically trained observers, at a single anatomic site. MATERIALS AND METHODS Design This study was executed in 2 phases. The first phase comprised basic training and assessment of 5 observers in algometry and PPT measurement techniques. The primary purpose of this was to ensure a controlled rate of algometry application. The second phase was an interrater reliability study of the algometry technique using the same 5 trained observers. Before the study, University Ethics Committee approval was obtained. Instrumentation PPT was induced using a pressure algometer (Salter Abbey Weighing Machines Ltd, England) with a flat circular metal probe dressed in several layers of lint and measuring 1 cm in diameter. Force was displayed digitally in increments of 0.1 N/cm 2. The algometer was mounted vertically on a purpose-built stand to enable force to be applied perpendicular to the measurement site (Fig. 1); there was no automated control or restriction of application rate, which was controlled entirely by the observer. Phase 1-training and Testing of Observers Five female volunteer observers were recruited from a group of final year undergraduate physical therapy students (mean age 23, range: 20 to 34). We selected this population to enable the results to be generalized to observers in subsequent experimental studies investigating physical therapy interventions and to reflect the basic level of skill of a population of clinical physical therapists. Before the study, we gave each observer a full written explanation of the purpose of the experiment and the training protocol, as recommended by Goulet et al. 28 After a further verbal briefing, all observers provided written consent to take part. Observers were trained by the same trainer during a single day, in the pressure algometer technique described below, so that repeated applications at the standardized rate of 5 N/s could be FIGURE 1. Fixed-angle pressure algometer. achieved. 16,36 Practice continued until observers were able to repeat applications at the required rate without reference to the digital display of the algometer. The training and protocol incorporated suggestions made by Goulet et al 28 to increase interrater reliability. These included applying the algometer at 90 degrees to a flat skin surface, standardizing and marking the area where pressure was to be applied, covering the application tip of the algometer to avoid sharp pain from the edge of the plunger, 37 and applying pressure evenly and at a constant rate. When training was complete, we tested the observers competence in using the algometer. This test required each observer to make 5 consecutive algometry applications at the prescribed rate of 5 Newtons (N)/s, 15 seconds apart. Each application lasted 10 seconds and was made at the first dorsal interosseous muscle of a volunteer participant, who played no other part in the study than to act as a model for this procedure. Observers were masked to the application rate by covering the digital display of the algometer, such that only the study coordinator could view and record the results. The test was considered successful if all 5 applications were carried out at a rate of 5 N/s (± 1.0 N/s) over a 10-second period, r 2007 Lippincott Williams & Wilkins 761

3 Chesterton et al Clin J Pain Volume 23, Number 9, November/December 2007 because PPT at this anatomic point is usually reported well within this timescale. All observers passed the test at the first attempt and progressed to take part in the reliability study. Phase 2-reliability Study Participants For this study, the sample size was calculated so as to detect a difference in the ICC between r = 0.9 and r = 0.7 at a 2-tailed 5% significance level and with 80% power, assuming that 5 observers rated each participant. 38,39 We chose r = 0.7 for the null value on the basis of its being the midpoint of the interval proposed by Shrout 40 to constitute moderate reliability (0.61 to 0.80). From this calculation, the minimum number of participants required was 13 (Table 1). Volunteers (n = 13; mean age 22) were recruited from the student population; they were all white and all but one was female (Table 1). Although age and sex are known to influence PPT between individuals, 4,5,41 we found no literature to suggest that these population characteristics influence the reliability of an individual s repeated verbal report of PPT. Before taking part, volunteers were screened for current and previous pain conditions, medication, peripheral neuropathy, neurologic pathologies, or injury to either hand. No participants were excluded. The experimental procedure was explained and written informed consent was obtained from each participant. PPT Measurement Procedure The reliability study was conducted in one laboratory at a mean temperature of 21.51C (SD 0.73). Participants were seated in a comfortable upright position. A single study coordinator marked the location of the PPT measurement site, at the midpoint of the muscle belly of the first dorsal interosseous muscle of the dominant hand. The hand was then placed beneath the algometer such that the algometer probe would be lowered perpendicular to the marked spot. The algometer setup was the same as used previously in training the observers. Two identical calibrated algometers were available. TABLE 1. Summary of Reliability Study Protocol Components No. participants (n) 13 Male: female 1:12 Mean (range) age 22 (20 29) y No. observers 5 Male: female 0:5 Mean (range) age 23 (20 34) y No. measures taken by each observer from a single 3 participant Interval between measurements for a single observer 15 s Interval between each observer per participant 10 min Total no. measurements per participant 15 Total no. measurements per observer 39 We explained and demonstrated the method of algometer application to each participant. Each participant then underwent 2 practice PPT measures using the nondominant hand and they were coached in differentiating their report of tactile and painful stimuli. PPT was taken as the amount of pressure required to elicit a sensation of pain distinct from pressure or discomfort. 42 We asked participants to say stop as soon as a discernible sensation of pain was felt; at this point the algometer pressure was immediately released and the plunger was retracted by the observer. The algometer probe was lowered at a constant rate of approximately 5 N/s until PPT was reached, as indicated by a participant s verbal report. Each observer took 3 PPT readings, from each participant, 15 seconds apart. This replicates recommended clinical practice. 10 Participants were given 10 minutes rest between measurements taken by each observer, which has previously been shown to be sufficient. 13 Neither the participants nor the observers were able to see the algometer display and all readings were recorded by the study coordinator. Thus, observers and participants were unaware of their own PPT recordings and those of other observers or participants. To avoid order effects, each observer tested participants in a computer-generated random sequence. All readings were taken on the same day. Data Analysis The mean of each observer s 3 PPT measurements per participant was calculated, and used as the data for the main analysis. For purposes of comparison, each observer s first measurement was also analyzed. Graphical analysis of each observer s ratings and repeated measures analysis of variance (ANOVA) were used to identify systematic bias (against an assumed null hypothesis of no difference between the 5 observers). The assumptions of this ANOVA model were checked before the analysis. Although the residuals showed some positive skew, this was considered to be within acceptable limits, given that the skew was largely unidirectional. 43 The ICC (model 2,1) was computed for interrater reliability, because the analysis of interest was to determine whether the observers could be interchangeable. 44 This ICC is based on a 2-way random effects model with rater variance included in the ICC denominator, so that differences in level across raters count as error. To express the precision of the reliability coefficient, and identify the limits likely to contain the true reliability in the population, 95% CIs were computed for the ICC values. The SEM provides a quantitative indicator of the measurement error in the units of the original measurements (which in this instance are the mean of each observer s 3 consecutive measurements per participant). The SEM is the SD p of the measurement errors, and is calculated as s ffiffiffiffiffiffiffiffiffiffi 1 r, where s is the SD of the observed ratings and r is the reliability coefficient (here, the ICC). A small value for the SEM indicates low measurement error, and thus high reliability. 45 We can be 95% confident that an interval of 1.96 SEM either side 762 r 2007 Lippincott Williams & Wilkins

4 Clin J Pain Volume 23, Number 9, November/December 2007 Interrater Reliability of Algometry of an individual s observed score contains that individual s true score (assuming that measurements are not systematically biased). The SEM can also be used to identify the minimum difference in the measure that must be exceeded to demonstrate true change, as opposed to random measurement fluctuation. Thus, for example, we can be p 95% ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi confident that a difference of greater than 1:96 2SEM 2 between 2 measurements represents true change. 35 We also calculated CIs for the SEM, via the formula for the CI of a variance. 46 Spearman rho (r S ) was used to assess the correlation between observers mean ratings and the order in which they were taken. Data were analyzed using SPSS version 14. We took Pr0.05 (2-tailed) as the cut-off for statistical significance. RESULTS The data for the reliability analysis (ie, mean values of the 3 measurements taken from each participant by each of the 5 observers) are given in Table 2. Figure 2 provides a plot of these values. This plot suggests that agreement between observers is generally high, being slightly lower for high PPT values (eg, participants 1, 6, and 10). The ICC was very high (ICC = 0.91; 95% CI 0.82, 0.97); this value is significantly larger than the null value of 0.7 (P<0.0005), and the narrow CI indicates the high precision of the estimate. The SEM was calculated as 6.27 N/cm 2 (95% CI 5.35, 7.59). On the basis of a SEM of this magnitude, we can be 95% confident that (1) an interval of N/cm 2 (1.25 kg/cm 2 ) either side of the observed PPT measurement contains the true value, and (2) a difference between 2 observers PPT measurements of more than N/cm 2 (1.77 kg/cm 2 ) represents a true difference. The ICC of 0.91 for the mean of the 3 measurements for each observer was somewhat higher than that obtained when only the observers first measurement was analyzed; for the latter analysis, ICC = 0.90 (95% CI 0.79, 0.96), and P = for the statistical test against the null value of 0.7. The SEM calculated on the observers first measurement was 6.66 N/cm 2 (95% CI 5.68, 8.06), such that we can be 95% confident that a difference between 2 observers first PPT measurements of more than N/cm 2 (1.88 kg/cm 2 ) represents a true difference. The bottom row in Table 2 shows the means for each observer across all 13 participants. Repeated measures ANOVA showed no statistically significant difference across these values (F 4,48 = 1.000, P = 0.417), which suggests that there was no bias in ratings across observers. Additionally, there was a nonsignificant correlation between the means of the 15 ratings and the sequence in which they were performed (r S = 0.343, P = 0.211), which suggests that there was no systematic change in the participants PPT values over time. DISCUSSION The value of most outcome measures and assessment techniques is in part dependent upon high intrarater and interrater reliability. The purpose of this study was to establish the interrater reliability of PPT measurement made using a pressure algometer, and with multiple observers trained in applications at a consistent rate. The results showed that the training of observers in the algometry application technique was successful, with all 5 observers passing the competency test at the first attempt and within a single day. This addresses the observer error introduced by variable rates of application and is, therefore, an important feature of the study. Most previous studies have not reported whether, or how, such testing of observers performance has taken place. The interrater ICC for the mean measurements was excellent at 0.91, with a narrow CI. These results are consistent with those of Antonaci et al, 17 Delaney and McKee, 20 and Nussbaum and Downes 10 ; notwithstanding the use of different study protocols, these authors also showed high interrater reliability (with ICCs between 0.75 and 0.88), in TABLE 2. PPT Measurements in N/cm 2 for Each Participant by Observer (Data in the Cells are the Mean of 3 Measurements Per Participant by Each Observer) Observer Participant Mean (SD) (9.6) (0.7) (5.3) (2.0) (4.4) (9.3) (6.9) (6.9) (10.3) (7.0) (5.4) (3.2) (4.8) Mean (SD) 37.7 (21.8) 34.8 (22.7) 39.7 (21.5) 38.1 (17.9) 37.0 (24.6) r 2007 Lippincott Williams & Wilkins 763

5 Chesterton et al Clin J Pain Volume 23, Number 9, November/December 2007 FIGURE 2. Mean score of 3 readings from n = 13 participants for each of 5 observers (PPT in N/cm 2 ). healthy humans. The current study, however, used a larger number of observers and reports the statistical conclusions with a stated power. Not unexpectedly, reliability was somewhat greater for the observers mean values than for their first ratings, as has been reported previously. 10 It is advisable, therefore, that mean rather than single measurements should be used in multiple-observer studies of PPT. From the SEM of 6.27 N/cm 2 (0.64 kg/cm 2 ) obtained in this study, it is highly likely that measurement error will be no greater than N/cm 2 (1.25 kg/cm 2 ), and that a difference between 2 observers measurements of PPT of more than N/cm 2 (1.77 kg/cm 2 ) will indicate a true difference, when using multiple trained raters. This value for the interobserver SEM is higher than the figure of 5.74 N/cm 2 (0.59 kg/cm 2 ) calculable for the data reported by Antonaci et al, 17 and that of 5.06 N/cm 2 (0.52 kg/cm 2 ) calculable for the data reported by Nussbaum and Downes 10 despite their reporting lower ICCs. The SDs of the PPT measurements were smaller in these studies than in our study, which is likely to have constrained the values they obtained for the ICC, but would not have affected the SEM. Fischer 47 suggested meaningful clinical differences in PPT measures would range from 1.5 to 2 kg/cm 2 (14.71 to N/cm 2 ). This suggestion was based on the comparison of PPTs in healthy tissue with that of pathologic (tender) tissue on the corresponding and contralateral anatomic site in 50 volunteers (tender spots correlated as hot spots when viewed thermographically). The magnitude of measurement error of N/cm 2 (1.77 kg/cm 2 ) in our study, using mean measurements, is therefore larger than the lower boundary of Fischer s estimate of a minimum clinically important difference. Therefore, despite the high ICC, we cannot be highly confident that a clinically important change will not be masked by measurement error, when using multiple observers and mean measurements. Further, findings from the current study show that participants are reliable in making a consistent verbal report of their own PPT with 15 repeated measures over a 1-hour period, which is consistent with findings of Chesterton et al 13 and Jensen et al, 48 but represents new information because ratings are from multiple observers. There was also evidence of heteroscedasticity (where larger differences between raters were recorded for participants showing higher PPT thresholds). This finding is of interest, and possible explanations are that some examiners were unable to maintain pressure of application at higher levels of PPT, or that participants were unable to discern reliably the threshold point at high levels of pressure and tissue distention. However, the change in angle (or possible movement/wobble) of the probe that may be produced with high pressure manual algometry clearly plays no part. The role of structured training and competency testing may also have played a part in the observed differences with previous studies, but this is difficult to quantify because Nussbaum and Downes 10 provided training but did not undertake a competency test and Antonaci et al 17 did not document any training. Hence, the consistency of application rate cannot be commented upon. The comparability in terms of developed skill in the technique is, therefore, difficult to determine. Within the data from the present study, there is a small trend toward lower PPT ratings on second assessments compared with the first and third by each observer. Variations in individual readings have been noted by other authors, 10,22,41 and mean data are, therefore, likely to give better estimates. Order effects for observers were counteracted by the random order in which observers assessed participants. The findings of this study relate to PPT measurements taken at the first dorsal interosseus muscle of the hand, using a fixed angle (via a stand). They may not necessarily be generalizable to other measurement sites, or to situations in which a purely manual method of measurement is used. With reference to the statistical analysis used in this study, some authors maintain that the ICC statistic may produce misleading results through underestimating or overestimating reliability. 30,49 Heterogeneity in individual levels of PPT was observed between participants, and this is a well documented phenomenon in PPT measurement. 4,5,48 In common with the Pearson correlation coefficient, a wide versus a narrow range of PPT measures (ie, heterogeneous population) will lead to relatively higher ICCs (though researchers should not be tempted to select a sample that is more heterogeneous than the population of interest for the sake of larger ICC values). The large interindividual differences demonstrated in the present study may, therefore, have influenced the magnitude of the ICCs obtained. For this reason, the 764 r 2007 Lippincott Williams & Wilkins

6 Clin J Pain Volume 23, Number 9, November/December 2007 Interrater Reliability of Algometry ICCs should be interpreted alongside the SEMs, 30 as the latter are not affected by the range of the measurements, and thus generalizable across samples with differing degrees of variability. An approach to analyzing agreement through bias and limits of agreement has been proposed as the method of choice, 50 and was used by Nussbaum and Downes. 10 This method, which relies on graphic display and analysis of paired differences, is impractical in the case of multiple observers; there are, for example, 10 pairwise analyses possible between the observers in this study. We feel the approach to analysis in the current study is, therefore, a useful combination of methods. The familiar skewness of PPT measures can, however, present difficulties for reliability studies. Our study provides new evidence that multiple, trained observers using a fixed-angle algometer, demonstrate high interrater reliability in the measurement of PPT in healthy humans. The goal of the training was to ensure a consistent and constant speed of application, which can be achieved through basic instruction and practice. This level of competency is easily achieved with this algometer and does not require specialist training; in this respect our findings are generalizable. Also, as our observers were undergraduate students, they can be assumed to represent a basic level of skill in the measurement of PPT; therefore, our findings may well underestimate the reliability of measurements taken by more experienced clinicians. The way in which this algometer is applied should not differ between patients and healthy humans, and it is reasonable to assume that our findings would apply to both healthy and clinical populations in the research situation. It must be acknowledged, however, that the algometer stand used in this study is limited in its application to areas that can be accommodated under the algometer probe, such as the hand or forearm. This apparatus could not be used to measure PPT in regions such as the neck or back, and our findings should not, therefore, be extrapolated to these measurement situations. We conclude that multiple trained examiners and this algometry technique have the potential to achieve highly reliable measurements of PPT in multiobserver trials. In this study, changes of greater than N/cm 2 (1.77 kg/cm 2 ) can be confidently regarded as representing true change. Notwithstanding the high value of the ICC, the measurement error estimated in this study could mask a clinically important difference. This highlights the importance of calculating measures of both reliability and measurement error. REFERENCES 1. Fischer AA. Pressure threshold measurement for diagnosis of myofascial pain and evaluation of treatment results. Clin J Pain. 1987;2: Isselee H, De Laat A, Bogaerts K, et al. Short-term reproducibility of pressure pain thresholds in masticatory muscles measured with a new algometer. J Orofac Pain. 1998;12: Isselee H, De Laat A, Bogaerts K, et al. Long-term fluctuations of pressure pain thresholds in healthy men, normally menstruating women and oral contraceptive users. Eur J Pain. 2001;5: Jensen R, Rasmussen BK, Pedersen B, et al. Cephalic muscle tenderness and pressure pain threshold in a general population. Pain. 1992;48: Jensen R, Rasmussen BK, Pedersen B, et al. Muscle tenderness and pressure pain thresholds in headache. A population study. Pain. 1993;52: Kosek E, Ekholm J, Hansson P. Pressure pain thresholds in different tissues in one body region. The influence of skin sensitivity in pressure algometry. Scand J Rehabil Med. 1999;31: Kosek E, Ordeberg G. Abnormalities of somatosensory perception in patients with painful osteoarthritis normalize following successful treatment. Eur J Pain. 2000;4: Walsh DM, Lowe AS, McCormack K, et al. Transcutaneous electrical nerve stimulation: effect on peripheral nerve conduction, mechanical pain threshold, and tactile threshold in humans. Arch Phys Med Rehabil. 1998;79: Streiner DL, Norman GR. Health Measurement Scales: A Practical Guide to Their Development and Use. 3rd ed. Oxford: Oxford University Press; Nussbaum EL, Downes L. Reliability of clinical pressure-pain algometric measurements obtained on consecutive days. Phys Ther. 1998;78: Greenspan JD, McGillis SL. Thresholds for the perception of pressure, sharpness, and mechanically evoked cutaneous pain: effects of laterality and repeated testing. Somatosens Mot Res. 1994; 11: Chesterton L, Foster N, Wright C, et al. Effects of TENS frequency, intensity and stimulation site parameter manipulation on pressure pain thresholds in healthy human subjects. Pain. 2003;106: Chesterton LS, Barlas P, Foster NE, et al. Gender differences in pressure pain threshold in healthy humans. Pain. 2003;101: MacDonald W, Atkins RM. An algometer for the automated measurement of pain threshold. Br J Rheumatol. 1990;29: Vatine JJ, Shapira SC, Magora F, et al. Electronic pressure algometry of deep pain in healthy volunteers. Arch Phys Med Rehabil. 1993;74: Walsh DM, Liggett C, Baxter D, et al. A double-blind investigation of the hypoalgesic effects of transcutaneous electrical nerve stimulation upon experimentally induced ischaemic pain. Pain. 1995;61: Antonaci F, Sand T, Lucas GA. Pressure algometry in healthy subjects: inter-examiner variability. Scand J Rehabil Med. 1998; 30: Buchanan HM, Midgley JA. Evaluation of pain threshold using a simple pressure algometer. Clin Rheumatol. 1987;6: Chung SC, Kim JH, Kim HS. Reliability and validity of the pressure pain thresholds (PPT) in the TMJ capsules by electronic algometer. Cranio. 1993;11: Delaney GA, McKee AC. Inter- and intra-rater reliability of the pressure threshold meter in measurement of myofascial trigger point sensitivity. Am J Phys Med Rehabil. 1993;72: Hogeweg JA, Langereis MJ, Bernards AT, et al. Algometry. Measuring pain threshold, method and characteristics in healthy subjects. Scand J Rehabil Med. 1992;24: Ohrbach R, Gale EN. Pressure pain thresholds in normal muscles: reliability, measurement effects and topographic differences. Pain. 1989;37: Persson AL, Brogårdh C, Sjölund B. Tender or not tender: testretest repeatability of pressure pain thresholds in the trapezius and deltoid muscles of healthy women. J Rehabil Med. 2004;36: Reeves JL, Jaeger B, Graff-Radford SB. Reliability of the pressure algometer as a measure of myofascial trigger point sensitivity. Pain. 1986;24: Takala EP. Pressure pain threshold on upper trapezius and levator scapulae muscles. Repeatability and relation to subjective symptoms in a working population. Scand J Rehabil Med. 1990;22: r 2007 Lippincott Williams & Wilkins 765

7 Chesterton et al Clin J Pain Volume 23, Number 9, November/December Tunks E, McCain GA, Hart LE, et al. The reliability of examination for tenderness in patients with myofascial pain, chronic fibromyalgia and controls. J Rheumatol. 1995;22: Chung SC, Um BY, Kim HS. Evaluation of pressure pain threshold in head and neck muscles by electronic algometer: intrarater and interrater reliability. Cranio. 1992;10: Goulet JP, Clark GT, Flack VF. Reproducibility of examiner performance for muscle and joint palpation in the temporomandibular system following training and calibration. Commun Dent Oral Epidemiol. 1993;21: Lacey RJ, Lewis M, Jordan K, et al Interrater reliability of scoring of pain drawings in a self-report health survey. Spine. 2005;30:E455 E Haas M. The reliability of reliability. J Manipulative Physiol Ther. 1991;14: Fleiss JL. The Design and Analysis of Clinical Experiments. New York: Wiley; Haas M. Statistical methodology for reliability studies. J Manipulative Physiol Ther. 1991;14: Rankin G, Stokes M. Reliability of assessment tools in rehabilitation: an illustration of appropriate statistical analyses. Clin Rehabil. 1998;12: Batterham AM, George KP. Reliability in evidence based clinical practice: a primer for allied health professionals. Phys Ther Sport. 2000;1: Eliasziw M, Young SL, Woodbury MG, et al. Statistical methodology for the concurrent assessment of interrater and intrarater reliability: using goniometric measurements as an example. Phys Ther. 1994;74: McDowell BC, McCormack K, Walsh DM, et al. Comparative analgesic effects of H-wave therapy and transcutaneous electrical nerve stimulation on pain threshold in humans. Arch Phys Med Rehabil. 1999;80: Greenspan JD, McGillis SL. Stimulus features relevant to the perception of sharpness and mechanically evoked cutaneous pain. Somatosens Mot Res. 1991;8: Donner A, Eliasziw M. Sample size requirements for reliability studies. Stat Med. 1987;6: Walter SD, Eliasziw M, Donner A. Sample size and optimal designs for reliability studies. Stat Med. 1998;17: Shrout PE. Measurement reliability and agreement in psychiatry. Stat Methods Med Res. 1998;7: Brennum J, Kjeldsen M, Jensen K, et al. Measurements of human pressure-pain thresholds on fingers and toes. Pain. 1989;38: Fischer AA. Pressure algometry over normal muscles. Standard values, validity and reproducibility of pressure threshold. Pain. 1987;30: Kirk RE. Experimental Design: Procedures for the Behavioral Sciences. 3rd ed. Pacific Grove: Brooks/Cole; Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86: Denegar CR, Ball DW. Assessing reliability and precision of measurement: an introduction to intraclass correlation and standard error of measurement. J Sports Rehabil. 1993;2: Miller I, Miller M. John E Freund s Mathematical Statistics. 6th ed. Upper Saddle River: Prentice Hall; Fischer AA. Application of pressure algometry in manual medicine. J Man Med. 1990;5: Jensen K, Andersen HO, Olesen J, et al. Pressure-pain threshold in human temporal region. Evaluation of a new pressure algometer. Pain. 1986;25: Bland JM, Altman DG. A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. Comput Biol Med. 1990;20: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1: r 2007 Lippincott Williams & Wilkins

PRESSURE ALGOMETRY IN HEALTHY SUBJECTS: INTER-EXAMINER VARIABILITY

PRESSURE ALGOMETRY IN HEALTHY SUBJECTS: INTER-EXAMINER VARIABILITY : 3 8, 1998 PRESSURE ALGOMETRY IN HEALTHY SUBJECTS: INTER-EXAMINER VARIABILITY Fabio Antonaci MD, 1;2 Trond Sand PhD 2 and Guilherme A. Lucas MD 2 From the 1 University Centre for Adaptive Disorders and

More information

TENDER OR NOT TENDER: TEST-RETEST REPEATABILITY OF PRESSURE PAIN THRESHOLDS IN THE TRAPEZIUS AND DELTOID MUSCLES OF HEALTHY WOMEN

TENDER OR NOT TENDER: TEST-RETEST REPEATABILITY OF PRESSURE PAIN THRESHOLDS IN THE TRAPEZIUS AND DELTOID MUSCLES OF HEALTHY WOMEN J Rehabil Med 2004; 36: 17 27 TENDER OR NOT TENDER: TEST-RETEST REPEATABILITY OF PRESSURE PAIN THRESHOLDS IN THE TRAPEZIUS AND DELTOID MUSCLES OF HEALTHY WOMEN Ann L. Persson, 1 Christina Brogårdh 2 and

More information

THE RELIABILITY OF PRESSURE ALGOMETRY IN ASYMPTOMATIC INDIVIDUALS OVER CONSECUTIVE DAYS. Laura Frank* Patrick McLaughlin *,^ Brett Vaughan *,^

THE RELIABILITY OF PRESSURE ALGOMETRY IN ASYMPTOMATIC INDIVIDUALS OVER CONSECUTIVE DAYS. Laura Frank* Patrick McLaughlin *,^ Brett Vaughan *,^ THE RELIABILITY OF PRESSURE ALGOMETRY IN ASYMPTOMATIC INDIVIDUALS OVER CONSECUTIVE DAYS Laura Frank* Patrick McLaughlin *,^ Brett Vaughan *,^ *Osteopathy Unit School of Biomedical & Health Sciences Victoria

More information

JCPR Journal of Clinical Physiotherapy Research

JCPR Journal of Clinical Physiotherapy Research JCPR Journal of Clinical Physiotherapy Research Original Article Intra-Examiner and Between-Day Reliability of Algometer for Pressure Pain Threshold and Pain Sensitivity in Upper Trapezius Muscle in Asymptomatic

More information

Victoria University. (c) 2004 THE EFFECT OF MANUAL PRESSURE RELEASE ON MYOFASCIAL TRIGGER POINTS IN THE UPPER TRAPEZIUS MUSCLE.

Victoria University. (c) 2004 THE EFFECT OF MANUAL PRESSURE RELEASE ON MYOFASCIAL TRIGGER POINTS IN THE UPPER TRAPEZIUS MUSCLE. THE EFFECT OF MANUAL PRESSURE RELEASE ON MYOFASCIAL TRIGGER POINTS IN THE UPPER TRAPEZIUS MUSCLE Laura Hodgson Bachelor of Science (Clinical Science) Master of Health Science (Osteopathy) Student Number:

More information

lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System

lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System Journal of Sport Rehabilitation, 1998, 7, 95-101 0 1998 Human Kinetics Publishers, Inc. lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System Randy Schmitz

More information

The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device

The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device Original Article The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device J. Phys. Ther. Sci. 15: 7 12, 2003 ASGHAR REZASOLTANI, Ph D, PT 1, 2), AMIR AHMADI, B Sc,

More information

A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total Daily Stretch Duration

A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total Daily Stretch Duration Journal of Strength and Conditioning Research, 2003, 17(2), 274 278 2003 National Strength & Conditioning Association A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total

More information

The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75

The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75 The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75 By: Brent L. Arnold and David H. Perrin * Arnold, B.A., & Perrin, D.H. (1993).

More information

Interrater and Intrarater Reliability of the Assisting Hand Assessment

Interrater and Intrarater Reliability of the Assisting Hand Assessment Interrater and Intrarater Reliability of the Assisting Hand Assessment Marie Holmefur, Lena Krumlinde-Sundholm, Ann-Christin Eliasson KEY WORDS hand pediatric reliability OBJECTIVE. The aim of this study

More information

Point-of-service questionnaires can reliably assess patients experiences

Point-of-service questionnaires can reliably assess patients experiences European Journal for Person Centered Healthcare Vol 2 Issue pp 85-91 ARTICLE Point-of-service questionnaires can reliably assess patients experiences Stephen D. Gill PhD Head of Safety and Quality Improvement,

More information

The intra and inter-rater reliability of a modified weightbearing lunge measure of ankle dorsiflexion

The intra and inter-rater reliability of a modified weightbearing lunge measure of ankle dorsiflexion The intra and inter-rater reliability of a modified weightbearing lunge measure of ankle dorsiflexion GRAFTON, Kate and O'SHEA, Simon Available from Sheffield Hallam University Research Archive (SHURA)

More information

The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation

The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation Lieneke J. Draaijers, M.D., Fenike R. H. Tempelman, M.D., Yvonne A. M. Botman, M.D., Wim E. Tuinebreijer,

More information

An update on the analysis of agreement for orthodontic indices

An update on the analysis of agreement for orthodontic indices European Journal of Orthodontics 27 (2005) 286 291 doi:10.1093/ejo/cjh078 The Author 2005. Published by Oxford University Press on behalf of the European Orthodontics Society. All rights reserved. For

More information

Visual analog scale and pressure pain threshold for delayed onset muscle soreness assessment

Visual analog scale and pressure pain threshold for delayed onset muscle soreness assessment Edith Cowan University Research Online ECU Publications 2013 2013 Visual analog scale and pressure pain threshold for delayed onset muscle soreness assessment Wing Y. Lau Edith Cowan University, wlau0@our.ecu.edu.au

More information

Validation of a Pressure Pain Threshold Scale in Patients Diagnosed with Myofascial Pain Syndrome and Fibromyalgia

Validation of a Pressure Pain Threshold Scale in Patients Diagnosed with Myofascial Pain Syndrome and Fibromyalgia Nova Southeastern University NSUWorks Department of Physical Therapy Student Theses, Dissertations and Capstones Department of Physical Therapy 4-1-2016 Validation of a Pressure Pain Threshold Scale in

More information

Hand grip strength is a measure of. Reliability and validity of an electronic dynamometer for measuring grip strength

Hand grip strength is a measure of. Reliability and validity of an electronic dynamometer for measuring grip strength Research Reliability and validity of an electronic dynamometer for measuring grip strength Deborah Allen, Fiona Barnett Aims: This article describes a study aiming to establish the concurrent validity

More information

Myofascial Pain Syndrome Trigger Point Detection based on Ultrasound Image

Myofascial Pain Syndrome Trigger Point Detection based on Ultrasound Image Myofascial Pain Syndrome Trigger Point Detection based on Ultrasound Image EKO SUPRIYANTO, JOANNE SOH ZI EN, SYED MOHD NOOH OMAR Advanced Diagnostics and Progressive Human Care Research Group Research

More information

Alternative scoring of the cutaneous assessment tool in juvenile dermatomyositis: Results using abbreviated formats

Alternative scoring of the cutaneous assessment tool in juvenile dermatomyositis: Results using abbreviated formats Himmelfarb Health Sciences Library, The George Washington University Health Sciences Research Commons Rheumatology Faculty Publications Medicine 3-15-2008 Alternative scoring of the cutaneous assessment

More information

Intertester Reliability of Assessing Postural Sway Using the Chattecx Balance System

Intertester Reliability of Assessing Postural Sway Using the Chattecx Balance System Intertester Reliability of Assessing Postural Sway Using the Chattecx Balance System By: Carl G. Mattacola, MEd, ATC; Denise A. Lebsack, PhD, ATC; David H. Perrin, PhD, ATC * Mattacola, C.G., Lebsack,

More information

Vibrotactile masking:,a comparison of psychophysical procedures

Vibrotactile masking:,a comparison of psychophysical procedures Perception & Psychophysics 1977, Vol. 22 (5), 471-475 Vibrotactile masking:,a comparison of psychophysical procedures RITA E. SNYDER Indiana University, Bloomington, Indiana 47401 The amounts of ipsilateral

More information

Pain Assessment in Elderly Patients with Severe Dementia

Pain Assessment in Elderly Patients with Severe Dementia 48 Journal of Pain and Symptom Management Vol. 25 No. 1 January 2003 Original Article Pain Assessment in Elderly Patients with Severe Dementia Paolo L. Manfredi, MD, Brenda Breuer, MPH, PhD, Diane E. Meier,

More information

4 Smallest detectable difference of maximal mouth opening in patients with painfully restricted temporomandibular joint function.

4 Smallest detectable difference of maximal mouth opening in patients with painfully restricted temporomandibular joint function. 4 Smallest detectable difference of maximal mouth opening in patients with painfully restricted temporomandibular joint function. Kropmans ThJB, Dijkstra PU, Stegenga B, Stewart R, de Bont LGM European

More information

Online Journal Club-Article Review

Online Journal Club-Article Review Online Journal Club-Article Review Article Citation Study Objective/Purpose (hypothesis) Brief Background (why issue is important; summary of previous literature) Study Design (type of trial, randomization,

More information

Statistical technique. Systematic error or bias (validity)

Statistical technique. Systematic error or bias (validity) Reliability of the Vertimetric vertical jump measurement device Prepared by Dr. Matt Beekley (University of Indianapolis) and Dr. Jim Brown (Safe Responder, LLC) INTRODUCTION Vertical jump is a measurement

More information

Supporting Information

Supporting Information 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 Supporting Information Variances and biases of absolute distributions were larger in the 2-line

More information

Downloaded from umj.umsu.ac.ir at 22: on Friday March 22nd 2019

Downloaded from umj.umsu.ac.ir at 22: on Friday March 22nd 2019 * 91/06/07 : 91/04/02 : :.. -. :. ICC. ICC.. :. // // :. : - - : :.(-) - (SLR) Email: eterafoskouei@tbzmed.ac.ir.(). (ULTT).().. ( ) Straight Leg Raise Upper Limb Tension Tests .().(-) Hough.().() ULTT1.()

More information

Introduction On Assessing Agreement With Continuous Measurement

Introduction On Assessing Agreement With Continuous Measurement Introduction On Assessing Agreement With Continuous Measurement Huiman X. Barnhart, Michael Haber, Lawrence I. Lin 1 Introduction In social, behavioral, physical, biological and medical sciences, reliable

More information

Journal of Sport Rehabilitation. The reliability of strength tests performed in elevated shoulder positions using a hand-held dynamometer

Journal of Sport Rehabilitation. The reliability of strength tests performed in elevated shoulder positions using a hand-held dynamometer The reliability of strength tests performed in elevated shoulder positions using a hand-held dynamometer Journal: Manuscript ID: JSR.2015-0034.R2 Manuscript Type: Technical Report Keywords: dynamometry,

More information

Assessing Agreement Between Methods Of Clinical Measurement

Assessing Agreement Between Methods Of Clinical Measurement University of York Department of Health Sciences Measuring Health and Disease Assessing Agreement Between Methods Of Clinical Measurement Based on Bland JM, Altman DG. (1986). Statistical methods for assessing

More information

The Short NART: Cross-validation, relationship to IQ and some practical considerations

The Short NART: Cross-validation, relationship to IQ and some practical considerations British journal of Clinical Psychology (1991), 30, 223-229 Printed in Great Britain 2 2 3 1991 The British Psychological Society The Short NART: Cross-validation, relationship to IQ and some practical

More information

Statistical Methods For Assessing Measurement Error (Reliability) in Variables Relevant to Sports Medicine

Statistical Methods For Assessing Measurement Error (Reliability) in Variables Relevant to Sports Medicine REVIEW ARTICLE Sports Med 1998 Oct; 26 (4): 217-238 0112-1642/98/0010-0217/$11.00/0 Adis International Limited. All rights reserved. Statistical Methods For Assessing Measurement Error (Reliability) in

More information

COMPUTING READER AGREEMENT FOR THE GRE

COMPUTING READER AGREEMENT FOR THE GRE RM-00-8 R E S E A R C H M E M O R A N D U M COMPUTING READER AGREEMENT FOR THE GRE WRITING ASSESSMENT Donald E. Powers Princeton, New Jersey 08541 October 2000 Computing Reader Agreement for the GRE Writing

More information

A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research

A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research Journal of Chiropractic Medicine (2016) 15, 155 163 www.journalchiromed.com Cracking the Code: Providing Insight Into the Fundamentals of Research and Evidence-Based Practice A Guideline of Selecting and

More information

Test-Retest Reliability of the StepWatch Activity Monitor Outputs in Healthy Adults

Test-Retest Reliability of the StepWatch Activity Monitor Outputs in Healthy Adults Journal of Physical Activity and Health, 2010, 7, 671-676 2010 Human Kinetics, Inc. Test-Retest Reliability of the StepWatch Activity Monitor Outputs in Healthy Adults Suzie Mudge, Denise Taylor, Oliver

More information

Intramachine and intermachine reproducibility of concentric performance: A study of the Con-Trex MJ and the Cybex Norm dynamometers

Intramachine and intermachine reproducibility of concentric performance: A study of the Con-Trex MJ and the Cybex Norm dynamometers Isokinetics and Exercise Science 12 (4) 91 97 91 IOS Press Intramachine and intermachine reproducibility of concentric performance: A study of the Con-Trex MJ and the Cybex Norm dynamometers C. Bardis

More information

Original Article. Annals of Rehabilitation Medicine INTRODUCTION

Original Article. Annals of Rehabilitation Medicine INTRODUCTION Original Article Ann Rehabil Med 2013;37(4):541-546 pissn: 2234-0645 eissn: 2234-0653 http://dx.doi.org/10.5535/arm.2013.37.4.541 Annals of Rehabilitation Medicine Ischemic Compression After Trigger Point

More information

What is indirect comparison?

What is indirect comparison? ...? series New title Statistics Supported by sanofi-aventis What is indirect comparison? Fujian Song BMed MMed PhD Reader in Research Synthesis, Faculty of Health, University of East Anglia Indirect comparison

More information

2 Philomeen Weijenborg, Moniek ter Kuile and Frank Willem Jansen.

2 Philomeen Weijenborg, Moniek ter Kuile and Frank Willem Jansen. Adapted from Fertil Steril 2007;87:373-80 Intraobserver and interobserver reliability of videotaped laparoscopy evaluations for endometriosis and adhesions 2 Philomeen Weijenborg, Moniek ter Kuile and

More information

10 Intraclass Correlations under the Mixed Factorial Design

10 Intraclass Correlations under the Mixed Factorial Design CHAPTER 1 Intraclass Correlations under the Mixed Factorial Design OBJECTIVE This chapter aims at presenting methods for analyzing intraclass correlation coefficients for reliability studies based on a

More information

Convergence Principles: Information in the Answer

Convergence Principles: Information in the Answer Convergence Principles: Information in the Answer Sets of Some Multiple-Choice Intelligence Tests A. P. White and J. E. Zammarelli University of Durham It is hypothesized that some common multiplechoice

More information

Prospective Validation of Clinically Important Changes in Pain Severity Measured on a Visual Analog Scale

Prospective Validation of Clinically Important Changes in Pain Severity Measured on a Visual Analog Scale ORIGINAL CONTRIBUTION Prospective Validation of Clinically Important Changes in Pain Severity Measured on a Visual Analog Scale From the Department of Emergency Medicine, Albert Einstein College of Medicine,

More information

Goniometric Reliability in a Clinical Setting

Goniometric Reliability in a Clinical Setting Goniometric Reliability in a Clinical Setting Elbow and Knee Measurements JULES M. ROTHSTEIN, PETER J. MILLER, and RICHARD F. ROETTGER Reliability of goniometric measurements has been examined only under

More information

Reliability of Lower Extremity Functional Performance Tests

Reliability of Lower Extremity Functional Performance Tests Reliability of Lower Extremity Functional Performance Tests Lori A. Bolgla, PT' Douglas R. Keskula, Ph D, PT, ATC2 ports medicine clinicians commonly see athletes who have sustained a lower extremity injury

More information

COMMITMENT &SOLUTIONS UNPARALLELED. Assessing Human Visual Inspection for Acceptance Testing: An Attribute Agreement Analysis Case Study

COMMITMENT &SOLUTIONS UNPARALLELED. Assessing Human Visual Inspection for Acceptance Testing: An Attribute Agreement Analysis Case Study DATAWorks 2018 - March 21, 2018 Assessing Human Visual Inspection for Acceptance Testing: An Attribute Agreement Analysis Case Study Christopher Drake Lead Statistician, Small Caliber Munitions QE&SA Statistical

More information

Empirical Formula for Creating Error Bars for the Method of Paired Comparison

Empirical Formula for Creating Error Bars for the Method of Paired Comparison Empirical Formula for Creating Error Bars for the Method of Paired Comparison Ethan D. Montag Rochester Institute of Technology Munsell Color Science Laboratory Chester F. Carlson Center for Imaging Science

More information

INTRODUCTION. Ayse Ozcan, Zeliha Tulum, Lamia PInar, Ferdi Baskurt. Address for correspondence. J Korean Med Sci 2004; 19: ISSN

INTRODUCTION. Ayse Ozcan, Zeliha Tulum, Lamia PInar, Ferdi Baskurt. Address for correspondence. J Korean Med Sci 2004; 19: ISSN J Korean Med Sci 24; 19: 874-8 ISSN 111-8934 Copyright The Korean Academy of Medical Sciences Comparison of Pressure Pain Threshold, Grip Strength, Dexterity and Touch Pressure of Dominant and Non-Dominant

More information

TrichoScan-Validation with Canon Powershot A95

TrichoScan-Validation with Canon Powershot A95 TrichoScan-Validation with Canon Powershot A95 1. Introduction Hair loss or thinning hair is a common complaint in clinical dermatology, but patients seeking advice for hair loss are not necessarily completely

More information

Fatigue is widely recognized as the most common symptom for individuals with

Fatigue is widely recognized as the most common symptom for individuals with Test Retest Reliability and Convergent Validity of the Fatigue Impact Scale for Persons With Multiple Sclerosis Virgil Mathiowetz KEY WORDS energy conservation fatigue assessment rehabilitation OBJECTIVE.

More information

OBSERVATIONS ON THE REACTION TIME TO CUTANEOUS

OBSERVATIONS ON THE REACTION TIME TO CUTANEOUS J. Neurol. Neurosurg. Psychiat., 1955, 18, 120. OBSERVATIONS ON THE REACTION TIME TO CUTANEOUS THERMAL STIMULI BY P. P. LELE* and D. C. SINCLAIR From the Department of Anatomy, University of Oxford The

More information

Method Comparison for Interrater Reliability of an Image Processing Technique in Epilepsy Subjects

Method Comparison for Interrater Reliability of an Image Processing Technique in Epilepsy Subjects 22nd International Congress on Modelling and Simulation, Hobart, Tasmania, Australia, 3 to 8 December 2017 mssanz.org.au/modsim2017 Method Comparison for Interrater Reliability of an Image Processing Technique

More information

BUFFALO CONCUSSION BIKE TEST (BCBT) INSTRUCTION MANUAL

BUFFALO CONCUSSION BIKE TEST (BCBT) INSTRUCTION MANUAL Purpose BUFFALO CONCUSSION BIKE TEST (BCBT) INSTRUCTION MANUAL To investigate exercise tolerance in patients with persistent post-concussive symptoms (PCS) lasting more than 2-4 weeks. The definition of

More information

A Spreadsheet for Deriving a Confidence Interval, Mechanistic Inference and Clinical Inference from a P Value

A Spreadsheet for Deriving a Confidence Interval, Mechanistic Inference and Clinical Inference from a P Value SPORTSCIENCE Perspectives / Research Resources A Spreadsheet for Deriving a Confidence Interval, Mechanistic Inference and Clinical Inference from a P Value Will G Hopkins sportsci.org Sportscience 11,

More information

Relationship Between Intraclass Correlation and Percent Rater Agreement

Relationship Between Intraclass Correlation and Percent Rater Agreement Relationship Between Intraclass Correlation and Percent Rater Agreement When raters are involved in scoring procedures, inter-rater reliability (IRR) measures are used to establish the reliability of measures.

More information

Electric Pulse Frequency and Magnitude of Perceived Sensation During Electrocutaneous Forearm Stimulation

Electric Pulse Frequency and Magnitude of Perceived Sensation During Electrocutaneous Forearm Stimulation 1378 ORIGINAL ARTICLE Electric Pulse Frequency and Magnitude of Perceived Sensation During Electrocutaneous Forearm Stimulation Herbert F. Jelinek, PhD, Rowan McIntyre, BPhysiotherapy ABSTRACT. Jelinek

More information

Controlled Test of the Analgesic and Relaxant Properties of Nitrous Oxide

Controlled Test of the Analgesic and Relaxant Properties of Nitrous Oxide Reprinted from the JOURNAL OF DENTAL RESEARCH, Vol. 53, No. 2, Pages: 486 490 g American Dental Association, 1974. All Rights Reserved. Controlled Test of the Analgesic and Relaxant Properties of Nitrous

More information

02a: Test-Retest and Parallel Forms Reliability

02a: Test-Retest and Parallel Forms Reliability 1 02a: Test-Retest and Parallel Forms Reliability Quantitative Variables 1. Classic Test Theory (CTT) 2. Correlation for Test-retest (or Parallel Forms): Stability and Equivalence for Quantitative Measures

More information

Development of a self-reported Chronic Respiratory Questionnaire (CRQ-SR)

Development of a self-reported Chronic Respiratory Questionnaire (CRQ-SR) 954 Department of Respiratory Medicine, University Hospitals of Leicester, Glenfield Hospital, Leicester LE3 9QP, UK J E A Williams S J Singh L Sewell M D L Morgan Department of Clinical Epidemiology and

More information

Verbal Reasoning: Technical information

Verbal Reasoning: Technical information Verbal Reasoning: Technical information Issues in test construction In constructing the Verbal Reasoning tests, a number of important technical features were carefully considered by the test constructors.

More information

Psychology, 2010, 1: doi: /psych Published Online August 2010 (

Psychology, 2010, 1: doi: /psych Published Online August 2010 ( Psychology, 2010, 1: 194-198 doi:10.4236/psych.2010.13026 Published Online August 2010 (http://www.scirp.org/journal/psych) Using Generalizability Theory to Evaluate the Applicability of a Serial Bayes

More information

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

PDF hosted at the Radboud Repository of the Radboud University Nijmegen PDF hosted at the Radboud Repository of the Radboud University Nijmegen The following full text is a publisher's version. For additional information about this publication click this link. http://hdl.handle.net/2066/23532

More information

The Immediate Effects of Lidocaine Iontophoresis on Trigger-Point Pain

The Immediate Effects of Lidocaine Iontophoresis on Trigger-Point Pain The Immediate Effects of Lidocaine Iontophoresis on Trigger-Point Pain Todd A. Evans, Jennifer R. Kunkle, Krista M. Zinz, Jessica L. Walter, and Craig R. Denegar Objective: To assess the efficacy of lidocaine

More information

Framework for Comparative Research on Relational Information Displays

Framework for Comparative Research on Relational Information Displays Framework for Comparative Research on Relational Information Displays Sung Park and Richard Catrambone 2 School of Psychology & Graphics, Visualization, and Usability Center (GVU) Georgia Institute of

More information

Intra-rater reliability of TMJ joint vibration a pilot study

Intra-rater reliability of TMJ joint vibration a pilot study Intra-rater reliability of TMJ joint vibration a pilot study Magdalena Bakalczuk 1, Marcin Berger1, Michał Ginszt 2, Jacek Szkutnik 1, Sorochynska Solomiia 3, Piotr Majcher 2 1 Department of Functional

More information

Influence of Inward Pressure Applied by the Transducer on Trunk Muscle Thickness during Ultrasound Imaging

Influence of Inward Pressure Applied by the Transducer on Trunk Muscle Thickness during Ultrasound Imaging Kawasaki Journal of Medical Welfare Vol. 19, No. 2, 2014 32-37 Original Paper Influence of Inward Pressure Applied by the Transducer on Trunk Muscle Thickness during Ultrasound Imaging Hiroshi ISHIDA *

More information

Reliability and validity of the International Spinal Cord Injury Basic Pain Data Set items as self-report measures

Reliability and validity of the International Spinal Cord Injury Basic Pain Data Set items as self-report measures (2010) 48, 230 238 & 2010 International Society All rights reserved 1362-4393/10 $32.00 www.nature.com/sc ORIGINAL ARTICLE Reliability and validity of the International Injury Basic Pain Data Set items

More information

The Effect of Inter-Trial Rest Interval on the Assessment of lsokinetic Thigh Muscle Torque

The Effect of Inter-Trial Rest Interval on the Assessment of lsokinetic Thigh Muscle Torque The Effect of Inter-Trial Rest Interval on the Assessment of lsokinetic Thigh Muscle Torque PAUL W. STRATFORD, MSc,' ANNETTE BRUULSEMA,2 BILL MAXWELL,= TIM BLACK, BSc,' BEVERLY HARDING, BSc4 Journal of

More information

G roin pain is associated with many sports and

G roin pain is associated with many sports and 446 ORIGINAL ARTICLE Clinical examination of athletes with groin pain: an intraobserver and interobserver reliability study PHölmich, L R Hölmich, A M Bjerg... Br J Sports Med 2004;38:446 451. doi: 10.1136/bjsm.2003.004754

More information

OBJECTIVE AND ACCURATE measurements of lumbar

OBJECTIVE AND ACCURATE measurements of lumbar 99 Strapped Versus Unstrapped Technique of the Prone Press-Up for Measurement of Lumbar Extension Using a Tape Measure: Differences in Magnitude and Reliability of Measurements William D. Bandy, PhD, PT,

More information

CRITICALLY APPRAISED PAPER (CAP) Evidence / Title of article

CRITICALLY APPRAISED PAPER (CAP) Evidence / Title of article CRITICALLY APPRAISED PAPER (CAP) Evidence / Title of article Sensory findings after stimulation of the thoracolumbar fascia with hypertonic saline suggest its contribution to low back pain Schilder A et

More information

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in 1 2 In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in adults: the influence of body mass index 3 4 Running title: Precision of the idxa for the measurement of visceral

More information

Chapter 7. Inter-rater reliability of a video-analysis method measuring low-back load in a field situation.

Chapter 7. Inter-rater reliability of a video-analysis method measuring low-back load in a field situation. Inter-rater reliability of a video-analysis method measuring low-back load in a field situation. P. Coenen I. Kingma C.R. Boot P.M. Bongers J.H. van Dieën Applied Ergonomics, 213, 44(5): 828-834 ABSTRACT

More information

Acupuncture and electro-acupuncture for people diagnosed with subacromial pain syndrome: a multicentre randomized trial

Acupuncture and electro-acupuncture for people diagnosed with subacromial pain syndrome: a multicentre randomized trial Acupuncture and electro-acupuncture for people diagnosed with subacromial pain syndrome: a multicentre randomized trial Jeremy Lewis 1,2,3, Julius Sim 4, Panos Barlas 5 1 Department of Clinical Therapies,

More information

The significance of sensory motor functions as indicators of brain dysfunction in children

The significance of sensory motor functions as indicators of brain dysfunction in children Archives of Clinical Neuropsychology 18 (2003) 11 18 The significance of sensory motor functions as indicators of brain dysfunction in children Abstract Ralph M. Reitan, Deborah Wolfson Reitan Neuropsychology

More information

Previously, we1 made the case that clinicians need to

Previously, we1 made the case that clinicians need to Journal of Athletic Training 2018;53(1):98 103 doi: 10.4085/1062-6050-503-16 Ó by the National Athletic Trainers Association, Inc www.natajournals.org Applied Statistics Statistical Primer for Athletic

More information

G roin pain is associated with many sports and

G roin pain is associated with many sports and 446 ORIGINAL ARTICLE Clinical examination of athletes with groin pain: an intraobserver and interobserver reliability study PHölmich, L R Hölmich, A M Bjerg... See end of article for authors affiliations...

More information

William C Miller, PhD, FCAOT Professor Occupational Science & Occupational Therapy University of British Columbia Vancouver, BC, Canada

William C Miller, PhD, FCAOT Professor Occupational Science & Occupational Therapy University of British Columbia Vancouver, BC, Canada William C Miller, PhD, FCAOT Professor Occupational Science & Occupational Therapy University of British Columbia Vancouver, BC, Canada THE L TEST MANUAL Version: November 2014 Table of Contents Introduction...

More information

Co-localized sensations resulting from simultaneous cold and warm stimulation. By: Torsten Thunberg

Co-localized sensations resulting from simultaneous cold and warm stimulation. By: Torsten Thunberg Co-localized sensations resulting from simultaneous cold and warm stimulation By: Torsten Thunberg If one dips one hand into cold

More information

Scapulothoracic muscle strength in individuals with neck pain

Scapulothoracic muscle strength in individuals with neck pain Journal of Back and Musculoskeletal Rehabilitation 29 (2016) 549 555 549 DOI 10.3233/BMR-160656 IOS Press Scapulothoracic muscle strength in individuals with neck pain Shannon M. Petersen a,,nathana.domino

More information

La Touche et al. The Journal of Headache and Pain (2015) 16:20 DOI /s

La Touche et al. The Journal of Headache and Pain (2015) 16:20 DOI /s La Touche et al. The Journal of Headache and Pain (2015) 16:20 DOI 10.1186/s10194-015-0500-1 RESEARCH ARTICLE Open Access Masticatory sensory-motor changes after an experimental chewing test influenced

More information

Chapter 1 Chapter 1. Chapter 1 Chapter 1. Chapter 1 Chapter 1. Chapter 1 Chapter 1. Chapter 1 Chapter 1

Chapter 1 Chapter 1. Chapter 1 Chapter 1. Chapter 1 Chapter 1. Chapter 1 Chapter 1. Chapter 1 Chapter 1 psychology theory pure research applied research introspection structuralism functionalism behaviorism reinforcement Gestalt psychology a formulation of relationships underlying observed events the science

More information

The Efficacy of the Back School

The Efficacy of the Back School The Efficacy of the Back School A Randomized Trial Jolanda F.E.M. Keijsers, Mieke W.H.L. Steenbakkers, Ree M. Meertens, Lex M. Bouter, and Gerjo Kok Although the back school is a popular treatment for

More information

Department of Rehabilitation Medicine, Gyeongsang National University Hospital, 1

Department of Rehabilitation Medicine, Gyeongsang National University Hospital, 1 Original Article Ann Rehabil Med 2012; 36: 394-399 pissn: 2234-0645 eissn: 2234-0653 http://dx.doi.org/10.5535/arm.2012.36.3.394 Annals of Rehabilitation Medicine Reliability of the Pinch Strength with

More information

Model 1: Subject As Single Factor

Model 1: Subject As Single Factor 3 Model 1: Subject As Single Factor 3.1 The Model In this section, I am considering reliability experiments where each subject is scored by a different group of r raters 1. The experimenter controls which

More information

Dry Needling to a Key Myofascial Trigger Point May Reduce the Irritability of Satellite MTrPs

Dry Needling to a Key Myofascial Trigger Point May Reduce the Irritability of Satellite MTrPs Authors: Yueh-Ling Hsieh, PhD, PT Mu-Jung Kao, MD Ta-Shen Kuan, MD, MS Shu-Min Chen, MD Jo-Tong Chen, MD, PhD Chang-Zern Hong, MD Affiliations: From the Department of Physical Therapy, Hungkuang University,

More information

Pain pressure threshold of a muscle tender spot increases following local and non-local rolling massage

Pain pressure threshold of a muscle tender spot increases following local and non-local rolling massage Aboodarda et al. BMC Musculoskeletal Disorders (2015) 16:265 DOI 10.1186/s12891-015-0729-5 RESEARCH ARTICLE Pain pressure threshold of a muscle tender spot increases following local and non-local rolling

More information

RELIABILITY OF REAL TIME JUDGING SYSTEM

RELIABILITY OF REAL TIME JUDGING SYSTEM RELIABILITY OF REAL TIME JUDGING SYSTEM Maja Bučar Pajek 1, Warwick Forbes 2, Jernej Pajek 3, Bojan Leskošek 1, and Ivan Čuk 1 1 Faculty of Sport, University of Ljubljana, Slovenia 2 Australian Institute

More information

A Comparison of the Immediate Effects of Eccentric Training vs Static Stretch on Hamstring Flexibility in Basketball Players Dr.

A Comparison of the Immediate Effects of Eccentric Training vs Static Stretch on Hamstring Flexibility in Basketball Players Dr. A Comparison of the Immediate Effects of Eccentric Training vs Static Stretch on Hamstring Flexibility in Basketball Players Dr. Mandeep Thour* *Assistant Professor, Department of Physical Education SGGS

More information

PEOPLE WITH TRAUMATIC spinal cord injury (SCI) can

PEOPLE WITH TRAUMATIC spinal cord injury (SCI) can 632 ORIGINAL ARTICLE Reliability of the Clinical Outcome Variables Scale When Administered Via Telephone to Assess Mobility in People With Spinal Cord Injury Ruth N. Barker, PhD, Delena I. Amsters, MPhyt,

More information

Issues in assessing the validity of nutrient data obtained from a food-frequency questionnaire: folate and vitamin B 12 examples

Issues in assessing the validity of nutrient data obtained from a food-frequency questionnaire: folate and vitamin B 12 examples Public Health Nutrition: 7(6), 751 756 DOI: 10.1079/PHN2004604 Issues in assessing the validity of nutrient data obtained from a food-frequency questionnaire: folate and vitamin B 12 examples Victoria

More information

Evaluation of a clinical test. I: Assessment of reliability

Evaluation of a clinical test. I: Assessment of reliability British Journal of Obstetrics and Gynaecology June 2001, Vol. 108, pp. 562±567 COMMENTARY Evaluation of a clinical test. I: Assessment of reliability Introduction Testing and screening are critical parts

More information

THE WORLD HEALTH ORGANIZATION defines mobility

THE WORLD HEALTH ORGANIZATION defines mobility 9 Validity and Reliability Comparison of 4 Mobility Measures in Patients Presenting With Neurologic Impairment Philippe Rossier, MD, Derick T. Wade, MA, MD, FRCP ABSTRACT. Rossier P, Wade DT. Validity

More information

Reliability of Measuring Trunk Motions in Centimeters

Reliability of Measuring Trunk Motions in Centimeters Reliability of Measuring Trunk Motions in Centimeters MARGARET ROST, SANDRA STUCKEY, LEE ANNE SMALLEY, and GLENDA DORMAN A method of measuring trunk motion and two related motions using a tape measure

More information

Configural information is processed differently in perception and recognition of faces

Configural information is processed differently in perception and recognition of faces Vision Research 43 (2003) 1501 1505 Rapid communication Configural information is processed differently in perception and recognition of faces Adrian Schwaninger a,b, *, Stefan Ryf b, Franziska Hofer b

More information

alternate-form reliability The degree to which two or more versions of the same test correlate with one another. In clinical studies in which a given function is going to be tested more than once over

More information

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including Week 17 and 21 Comparing two assays and Measurement of Uncertainty 2.4.1.4. Explain tools used to compare the performance of two assays, including 2.4.1.4.1. Linear regression 2.4.1.4.2. Bland-Altman plots

More information

Three Different Methods of Calculating Vertical Jump Height from Force Platform Data in Men and Women

Three Different Methods of Calculating Vertical Jump Height from Force Platform Data in Men and Women Measurement in Physical Education and Exercise Science ISSN: 1091-367X (Print) 1532-7841 (Online) Journal homepage: http://www.tandfonline.com/loi/hmpe20 Three Different Methods of Calculating Vertical

More information

Assessment of Somatosensory

Assessment of Somatosensory Rivermead Assessment of Somatosensory Performance Charlotte E Winward 1, Peter W Halligan 2, Derick T Wade 3 2 nd EDITION (2012) OPERATOR MANUAL 1 Movement Science Group, Oxford Brookes University, Oxford,

More information