1 Original Research SLEEP MEDICINE Equivalence of Autoadjusted and Constant Continuous Positive Airway Pressure in Home Treatment of Sleep Apnea* Yvonne Nussbaumer, MD; Konrad E. Bloch, MD, FCCP; Therese Genser; and Robert Thurnheer, MD, FCCP Whether computerized autoadjusted continuous positive airway pressure (acpap) is effective or even superior to constant continuous positive airway pressure (ccpap) in the treatment of obstructive sleep apnea syndrome (OSAS) is still controversial. We performed a randomized, double-blind, controlled, cross-over trial comparing efficacy of sleep apnea home therapy by a novel acpap machine (REMStarAuto; Respironics; Murrysville, PA) operated in autoadjusted or constant mode. Thirty sleep apnea patients were recruited consecutively. Mean baseline Epworth sleepiness scale (ESS) score was ( SD), mean sleep resistance time was 26 2 min (Osler test; Stowood Scientific Systems; Oxford, UK), and mean apnea-hypopnea index (AHI) was h. Patients were randomly assigned to 1 month of home therapy with acpap followed by 1 month with ccpap, or vice versa. After 1 month with treatment, the mean ESS score, sleep resistance time, and AHI were significantly improved ( , 37 1 min, and h, respectively; all p < 0.05 vs baseline). Similar effects were achieved with ccpap (p not significant vs acpap). Twenty-six patients preferred acpap, and 4 patients preferred ccpap (p < 0.001). We conclude that patients with OSAS preferred acpap over ccpap in the initial phase of therapy. The effectiveness acpap in improving major outcomes was equivalent to ccpap. Since acpap does not require initial titration, it is a simple and promising modality for sleep apnea home therapy. (CHEST 2006; 129: ) Key words: auto-continuous positive airway pressure; continuous positive airway pressure; randomized trial; sleep apnea; treatment; vigilance Abbreviations: acpap autoadjusted continuous positive airway pressure; AHI apnea-hypopnea index; ccpap constant continuous positive airway pressure; CI confidence interval; ESS Epworth sleepiness scale; NS not significant; ODI oxygen desaturation index; OSAS obstructive sleep apnea syndrome; SF-36 Short Form-36 The obstructive sleep apnea syndrome (OSAS) is an important disorder affecting between 2% and 4% of adults. 1 OSAS is associated with excessive daytime sleepiness, impaired quality of life, and risk of traffic accidents. 2,3 Continuous positive airway pressure (CPAP) applied to the nose, the standard treatment for OSAS, improves vigilance, quality of *From the Pulmonary Division (Drs. Nussbaumer and Thurnheer and Ms. Genser), Department of Internal Medicine, Kantonsspital Münsterlingen, Münsterlingen; and Pulmonary Division (Dr. Bloch), Department of Internal Medicine, University Hospital of Zurich, Zurich, Switzerland. The study was supported by MEDELA AG, Switzerland, distributor of Respironics products in Switzerland. Manuscript received May 31, 2005; revision accepted September 15, life, and traffic safety. 4 6 The therapeutic mask pressure that abolishes apnea, hypopnea, and snoring in all sleep stages and body positions has conventionally been determined by manual in-laboratory titration. 7 This is expensive and responsible for long waiting lists. In addition, the therapeutic CPAP determined in a single titration night is a tradeoff between pressure-related side effects and efficacy in preventing upper-airway obstruction. This is because Reproduction of this article is prohibited without written permission from the American College of Chest Physicians ( org/misc/reprints.shtml). Correspondence to: Robert Thurnheer, MD, FCCP, Pulmonary Division, Department of Internal Medicine, Kantonsspital Münsterlingen, 8596 Münsterlingen, Switzerland; robert.thurnheer@stgag.ch 638 Original Research

2 the pressure required to maintain upper-airway patency changes during a night depending on body position, sleep stage, nasal obstruction, ingestion of alcohol or hypnotic agents, 8 and over time based on changes in body weight and upper-airway properties. This variability of CPAP requirement has generated considerable interest in devices adjusting mask pressure by feedback control according to patterns of pressure, flow, or other signals during treatment (autoadjusted CPAP [acpap] devices). Autoadjustment of treatment pressure allows for a decrease in CPAP when apneas and hypopneas disappear, and an increase when respiratory disorders reappear. This has led to the assumption that continuous adaptation of mask pressure to the actual needs of the patient might improve effectiveness of therapy by preventing sleep disturbances from unnecessarily high mask pressures, by enhancing comfort and treatment adherence. 9 Mean mask pressure applied during acpap home therapy is generally lower than the manually titrated pressure. 10,11 Compliance with acpap was higher than with constant CPAP (ccpap) in some studies 9,12 but not all studies. 10,11 Whether acpap home therapy results in similar or even greater improvement of subjective and objective sleepiness and respiratory disturbances as ccpap is still a matter of debate. 10,13,14 Since acpap has the potential to enhance comfort and simplify sleep apnea therapy, we tested the hypothesis that treatment effects are equivalent when the same machine (REM-Star Auto; Respironics; Murrysville, PA) is operated in the acpap vs ccpap mode during the initial phase of home therapy. A prospective, double-blind randomized, cross-over protocol was employed. Main outcomes were the differences between the treatment modes in subjective and objective sleepiness, quality of life, nocturnal breathing disturbances, and compliance, as well as preference for one or the other mode. Some of these results have been reported in the form of an abstract. 15 Patients Materials and Methods Consecutive patients with excessive sleepiness and apneahypopnea index (AHI) 10 obstructive events per hour were included. Patients with congestive heart failure, chronic rhinitis, and other sleep disorders were excluded. The protocol was approved by the local ethics committee. For the diagnosis of sleep apnea, unattended home sleep studies were performed with a 7-channel respiratory polygraph (PolyMESAM; MAP; Martinsried, Germany), 16 recording nasal airflow, 17 thoracoabdominal movement, ECG, oxygen saturation, body position, and leg movement. The polygraph was mounted by the study nurse in the outpatient clinic. Time periods from 11 pm to 5:30 am were taken for automated analysis. All sleep studies were manually controlled scoring 1-min epochs. Apnea was defined as a cessation of airflow at the nose for at least 10s, and hypopneas were scored when a decrease in the amplitude of airflow signal to 50% of the level prevailing before the event accompanied by a fall in oxyhemoglobin saturation of at least 3% was identified. Home monitoring has been shown to be as reliable as laboratory polysomnography in the prediction of success of nasal CPAP treatment, at least in patients without significant comorbidity. 18 Blower Machine The REMStar Auto device, an acpap machine programmable by an insertable computer chip card, was employed. In six patients, the 90th pressure percentile applied by the REM-Star Auto device was not statistically different from the ninetieth percentile applied by a previously validated device (Res Med AutoSet; Respironics) 11 during 2 titration nights, one with each device. Mean 90th pressures percentiles were 7.0 (SE, 0.37) and 7.9 (SE, 0.54) cm H 2 O, respectively (p not significant [NS]). Protocol Patients were instructed that they would sleep with a blower device working in different modes without giving further information about the modes. The patients were treated over two periods of 1 month with acpap and ccpap modes, respectively, applied by the same device in random order. At the time of randomization, a pair of two identically looking chip cards was assigned to each patient. One card contained the code to operate the CPAP device in constant mode at the 90th pressure percentile applied during 1-night acpap titration. The other card contained the code to operate the CPAP device in acpap mode with a pressure range from 5 to 15 cm H 2 O. Patients were instructed to use the CPAP device for 1 month with one of the two cards, and for another month with the other card. Patients and attending physicians were blinded to treatment modes and order of application. Measurement and Outcomes Subjective sleepiness was assessed by the Epworth sleepiness scale (ESS). 19 Health-related quality of life was assessed by the Short Form-36 (SF-36) questionnaire. 20 Symptoms, treatment benefits, and side effects were rated by patients on visual analog scales ranging from 0 to 100 mm. Daytime vigilance was evaluated by the Osler test 21,22 (Stowood Scientific Systems; Oxford, UK). The mean time until seven successive responses to a light signal were missed, and the mean number of missed stimuli per minute 23 were recorded during two test sessions on the same day at 9 am and 11 am. Statistical Analysis Data are summarized as mean SE. Repeated evaluations were performed with paired t tests or analysis of variance. Preference of CPAP mode was calculated with the Fisher exact p test. The agreement between applied mask pressure during ccpap and acpap treatment was assessed by the method of Bland and Altman. 24 The effect size was calculated as ratio of the mean difference to baseline to the SD of the mean at baseline. 25 An effect size of 0.2 is considered small, 0.5 is considered intermediate, and 0.8 is considered high. According to Jones and coworkers, 26 equivalence between treatment modes was assumed if the 95% confidence interval (CI) of an observed difference fell entirely within a specified CHEST / 129 / 3/ MARCH,

3 equivalence range. Differences in the ESS score of 2 points, 5 in the SF-36 vitality score of 10 points, 4 in AHI of 5/h, and in sleep resistance time 4 min were considered to reflect equivalence. 5 With a sample size of 30 patients, the study was powered with 80% to detect the following significant differences in major outcomes ( 0.05): ESS score, 1; SF-36 vitality score, 11; AHI, 5/min; sleep resistance time, 3 min. Patients Results Thirty-four patients were included; 4 of these patients did not complete the protocol. Two patients were given humidifiers, one stopped treatment because of claustrophobia, and another subject was unavailable for follow-up. Results from 27 men and 3 women were available for analysis. The mean age was 49 years (SE, 2 years; range, 29 to 65 years), mean body mass index was 31.1 kg/m 2 (SE, 0.6 kg/m 2 ), and mean neck circumference was 42.7 cm (SE, 2.2 cm). Outcomes The effect of treatment with acpap was compared to that with ccpap and to baseline. Main results are given in Table 1. Individual results for subjective vigilance as well as for AHI are shown in Figures 1, 2. Figure 1. Individual results of ESS scores at baseline and with the two different treatment modes. Subjective Assessment There was a significant and clinically relevant improvement in symptoms by both treatment modalities to a similar degree. Mean decrease in the ESS score was 6 points for ccpap and acpap. The effect size was 1.99 for ccpap and 2.07 for acpap. The 95% CI of the mean difference between the treatment modalities was 1.6 to 1.1, suggesting equivalence (Fig 3). The SF-36 vitality score was significantly improved with either treatment, whereas in the other domains of the SF-36 no significant differences to baseline were found (Table 1). Overall acceptance was not different between CPAP modes (Table 2). At the end of the study, all patients wished to continue CPAP therapy. When asked which treatment period they had preferred (before unblinding) 26 patients opted for the period Table 1 Effects of acpap and ccpap Therapy* Variables Baseline ccpap acpap Subjective assessment ESS SF-36 physical functioning score SF-36 role physical score SF-36 bodily pain score SF-36 general health score SF-36 vitality score SF-36 social functioning score SF-36 role emotional score SF-36 mental health score Objective vigilance Osler test sleep resistance, min Nocturnal respiration AHI, /h ODI, /h Treatment pressure Mean applied mask pressure, cm H 2 O Compliance Percentage of days with 4 h of use, % Mean use per night, h *Data are presented as mean SE. p 0.05 vs baseline. p 0.05 vs ccpap. 640 Original Research

4 Figure 2. Individual results of AHI at baseline and with the two different treatment modes. with acpap, whereas only four preferred ccpap (Fisher exact test, p 0.001). The sleep questionnaire indicated significant differences in noise perception and discomfort from high pressure in favor of acpap (Table 1). No differences in sleep quality, side effects from mask and nasal discomfort or symptoms from mouth leak were noted. More relevant side effects requiring interruption of therapy or consultations were absent. Objective Daytime Vigilance Sleep resistance improved significantly, independent of CPAP mode. With ccpap, mean sleep resistance time improved by 8.1 min (SE, 1.7 min), and with acpap by 10.9 min (SE, 2.0 min) [p vs baseline; p NS between groups]. The effect sizes were 0.69 for ccpap and 0.93 for acpap. Although the difference in sleep resistance between the two CPAP modes were not statistically significant (37.2 min with acpap vs 34.4 min with ccpap), equivalence could not be confirmed, since the 95% CI of the difference ( 0.2 to 5.8 min) extended beyond the predefined range of 4 min in favor of acpap (Fig 3). Missed stimuli per minute decreased significantly from baseline by 0.36 (SE, 0.09) with acpap and by 0.67 (SE, 0.14) with ccpap (p NS). Nocturnal Respiration, Treatment Pressure, and Compliance At baseline, mean AHI was 41.1/h (SE, 3.6/h) and mean oxygen desaturation index (ODI) was 29.0 (SE, 4.0). Mean AHI decreased significantly by 35.7/h (SE, 3.2/h) with ccpap and by 36.5/h (SE, 3.7/h) with acpap (p 0.05 vs baseline; p NS between CPAP modes). Equivalence of both treatment modes was confirmed with a 95% confidence interval of the difference from 1.7 to 3.3, which falls entirely within the range of equivalence of AHI of 5/h (Fig 3). The mean ODI was reduced by 25.0/h (SE, 3.5/h) and 24.8/h (SE, 3.8/h), respectively (p 0.05 vs baseline; p NS between CPAP modes). Treatment effect size for AHI was 1.82 for ccpap and 1.86 for acpap; corresponding values for ODI were 1.13 and 1.12, respectively. Mean mask pressure during 1 month of therapy was lower with acpap. Mean difference was 1.3 cm H 2 O (SE, 0.4 cm H 2 O; 95% CI, 2.1 to 0.5 cm H 2 O; p 0.05). Mean nightly CPAP use of 4.8 h (SE, 0.3 h) with ccpap was similar to that with acpap of 5.1 h (SE, 0.3 h) [p NS; Table 1]. In each group, there was one patient with nonadherence (use 2 h), and there were eight patients in the ccpap group and eight patients in the acpap group with mean nightly use 6h. Discussion Figure 3. To evaluate whether the effects of acpap and ccpap were equivalent, the 95% CIs of the differences in major outcomes are displayed as solid horizontal lines. Differences are computed as values during autoadjusted minus corresponding values during ccpap therapy. Vertical dashed lines delineate equivalence ranges assumed to represent no clinically relevant difference. While none of the differences were statistically significant, the 95% CIs of differences in SF-36 vitality scores, and sleep resistance time extended beyond the equivalence ranges in favor of acpap. The meaning of a positive difference is indicated at the right side of each panel. In this double-blind, randomized, controlled trial, 1 month of acpap treatment was equivalent to ccpap in improving major outcomes of sleep apnea therapy, including daytime sleepiness, impaired quality of life, and nocturnal respiratory disturbances. With both treatment modes, a significant and clinically relevant relief of the symptoms of OSAS was achieved. Our results suggest that the algorithm implemented in the acpap device per- CHEST / 129 / 3/ MARCH,

5 Table 2 CPAP Tolerance and Preference* Variables Constant CPAP Autoadjusted CPAP CPAP tolerance and sleep questionnaire Problem with mask leaks? Problem with mouth leaks? General fit of mask? How much refreshed do you feel this morning? Do you suffer from nasal stuffiness? Do you suffer from sore or dry mouth/throat? How uncomfortable was the air pressure? How difficult was it falling asleep? Noise of CPAP-device? Did you have difficulties in maintaining sleep? How restoring was your sleep? How well did you sleep? Overall benefit during treatment Overall, how easily did you tolerate this treatment? Preference Patients preferring CPAP mode, No *Data are presented as mean SE unless otherwise indicated. Questionnaire data are on a visual analogue scale (range, 0 to 100 mm); lower values better; questions relate to the entire 4-week periods with the two treatment modes. p 0.05 vs ccpap. p vs ccpap. forms well in respiratory event detection and automatic mask pressure adjustment, and that this is appreciated by the patients who indicated a clear preference for the acpap compared to the ccpap mode. The effect sizes achieved in major outcomes with acpap and ccpap therapy of 1.82 for AHI, 1.99 for ESS score, and 0.69 for sleep resistance time compare favorably with previous evaluations of ccpap (effect sizes 1.61, 1.40, and 0.38, respectively). 27 Effect sizes 0.8 are considered as clinically important. 25 Moreover, acpap was at least equivalent to ccpap in improvement of these outcomes (Fig 3). Mean CPAP use of 5.1 h for acpap and 4.8 h for ccpap is comparable to previous studies 5,10 and clearly exceeds a mean use of 3.4 h considered as the lower end of treatment time that provides a clinically significant treatment benefit. 28 The 90th pressure percentile titrated during a night of acpap adaptation has been taken for ccpap treatment in accordance to a previous study 29 with the ResMed AutoSet device, in which the 90th percentile best approximated manually titrated CPAP pressure. In the comparison we performed in six patients, the 90th pressure percentiles titrated with the ResMed AutoSet and the REMstar Auto showed no significant difference, and the treatment effects on respiratory disturbance and vigilance were similar to those in previous studies. 27,28 In one study, 30 autoadjusted titration has been shown to represent a valid alternative to standard in-laboratory CPAP titration. Previous studies have addressed the efficacy of acpap devices in home therapy of OSAS over several weeks. A randomized study 14 including eight patients treated for 3 weeks with the Morphée autocpap device (Pierre Medical; Verrieres le Buisson, France) in automatic mode and eight patients in constant mode revealed a better compliance in the acpap group of 6.5 h per night vs 5.1 h per night in the ccpap group. In 1997, Sériès and Mark 13 demonstrated in a small (n 10) cross-over study a similar AHI when patients were treated with the Morphée acpap device compared to manually titrated ccpap. In a single-blind parallel study, Konermann et al 9 found better compliance, sleep quality, and lower treatment pressures with acpap with the Horizon machine (De Vilbiss; Somerset, PA) compared to ccpap. However, no data on daytime vigilance were available for this cohort. 9 A similar parallel study 31 using the Somnosmart device (Weinmann; Hamburg, Germany) also reported lower treatment pressures and higher preference for acpap, but no objective data about daytime vigilance were available either. In a randomized crossover study, patients selected according to their requirements for a high therapeutic pressure 10 cm of H 2 O were treated with AutoSet T or with manually titrated fixed pressure over 6 weeks each. The AutoSet T (ResMed; North Ryde, Australia) mode was superior in sleep symptom scores, SF-36 vitality, and mental health scores. Moreover, median treatment pressures were lower and adherence per night was significantly better with the autoadjusting device, but objective vigilance was not assessed. In a randomized trial 10 of two acpap devices vs ccpap 642 Original Research

6 addressing objective and subjective sleepiness, nocturnal breathing disturbances, treatment adherence, and quality of life, no differences between ccpap and acpap by the two tested devices (AutoAdjust LT; DeVilbiss; AutoSet T; ResMed) were found. Conclusions This double-blind, randomized, controlled trial compares ccpap to acpap in the home treatment of obstructive sleep apnea. Our investigation adds to the evidence demonstrating equivalent improvements of symptoms and measured sleep resistance by acpap using a novel device. We found no advantage of acpap regarding major outcome parameters such as objective sleepiness, symptoms, or quality of life, but the majority of patients preferred treatment in the autoadjusting mode. Therefore, acpap is a promising alternative to conventional CPAP treatment with fixed mask pressure. References 1 Young T, Palta M, Dempsey J, et al. The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med 1993; 328: Baldwin CM, Griffith KA, Nieto FJ, et al. The association of sleep-disordered breathing and sleep symptoms with quality of life in the Sleep Heart Health Study. Sleep 2001; 24: Findley LJ, Weiss JW, Jabour ER. Drivers with untreated sleep apnea: a cause of death and serious injury. Arch Intern Med 1991; 151: Montserrat JM, Ferrer M, Hernandez L, et al. Effectiveness of CPAP treatment in daytime function in sleep apnea syndrome: a randomized controlled study with an optimized placebo. Am J Respir Crit Care Med 2001; 164: Jenkinson C, Davies RJ, Mullins R, et al. Comparison of therapeutic and subtherapeutic nasal continuous positive airway pressure for obstructive sleep apnoea: a randomised prospective parallel trial. Lancet 1999; 353: George CF. Reduction in motor vehicle collisions following treatment of sleep apnoea with nasal CPAP. Thorax 2001; 56: Montserrat JM, Ballester E, Olivi H, et al. Time-course of stepwise CPAP titration: behavior of respiratory and neurological variables. Am J Respir Crit Care Med 1995; 152: Teschler H, Berthon-Jones M, Wessendorf T, et al. Influence of moderate alcohol consumption on obstructive sleep apnoea with and without AutoSet nasal CPAP therapy. Eur Respir J 1996; 9: Konermann M, Sanner BM, Vyleta M, et al. Use of conventional and self-adjusting nasal continuous positive airway pressure for treatment of severe obstructive sleep apnea syndrome: a comparative study. Chest 1998; 113: Senn O, Brack T, Matthews F, et al. Randomized short-term trial of two autocpap devices versus fixed continuous positive airway pressure for the treatment of sleep apnea. Am J Respir Crit Care Med 2003; 168: Teschler H, Wessendorf TE, Farhat AA, et al. Two months auto-adjusting versus conventional ncpap for obstructive sleep apnoea syndrome. Eur Respir J 2000; 15: Massie CA, McArdle N, Hart RW, et al. Comparison between automatic and fixed positive airway pressure therapy in the home. Am J Respir Crit Care Med 2003; 167: Sériès F, Marc I. Efficacy of automatic continuous positive airway pressure therapy that uses an estimated required pressure in the treatment of the obstructive sleep apnea syndrome. Ann Intern Med 1997; 127: Meurice JC, Marc I, Sériès F. Efficacy of auto-cpap in the treatment of obstructive sleep apnea/hypopnea syndrome. Am J Respir Crit Care Med 1996; 153: Thurnheer R, Genser Th, Ochsner Y. Constant vs. automated CPAP: a prospective double-blind randomized controlled crossover trial [abstract]. Am J Respir Crit Care Med 2004; 15:A Marrone O, Salvaggio A, Insalaco G, et al. Evaluation of the POLYMESAM system in the diagnosis of obstructive sleep apnea syndrome. Monaldi Arch Chest Dis 2001; 56: Thurnheer R, Xie X, Bloch KE. Accuracy of nasal cannula pressure recordings for assessment of ventilation during sleep. Am J Respir Crit Care Med 2001; 164: Whitelaw WA, Brant RF, Flemons WW. Clinical usefulness of home oximetry compared with polysomnography for assessment of sleep apnea. Am J Respir Crit Care Med 2005; 171: Bloch KE, Schoch OD, Zhang JN et al. German version of the Epworth Sleepiness Scale. Respiration 1999; 66: Ware JE, Snow KK, Kosinski M, et al. SF-36 health survey: manual and interpretation guide. Boston, MA. The Health Institute, New England Medical Center Bennett LS, Stradling JR, Davies RJ. A behavioural test to assess daytime sleepiness in obstructive sleep apnoea. J Sleep Res 1997; 6: Mazza S, Pepin JL, Deschaux C, et al. Analysis of error profiles occurring during the OSLER test: a sensitive mean of detecting fluctuations in vigilance in patients with obstructive sleep apnea syndrome. Am J Respir Crit Care Med 2002; 166: Priest B, Brichard C, Aubert G, et al. Microsleep during a simplified maintenance of wakefulness test: a validation study of the Osler test. Am J Respir Crit Care Med 2001; 163: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 11: Kazis LE, Andersen JJ, Meenan RF. Effect sizes for interpreting changes in health status. Med Care 1989; 27:S178 S Jones B, Jarvis B, Ebutt AF. Trials to assess equivalence: the importance of rigorous methods. BMJ 1996; 313: Kingshott RN, Vennelle M, Hoy CJ, et al. Predictors of improvements in daytime function outcomes with CPAP therapy. Am J Respir Crit Care Med 2000; 161: Engleman HM, Martin SE, Deary IJ, et al. Effect of CPAP therapy on daytime function in patients with mild sleep apnoea/ hypopnoea syndrome. Thorax 1997; 52: Lloberes P, Ballester E, Montserrat JM, et al. Comparison of manual and automatic CPAP titration in patients with sleep apnea/hypopnea syndrome. Am J Respir Crit Care Med 1996; 154: Masa JF, Jimenez A, Duran J, et al. Alternative methods of titrating continuous positive airway pressure: a large multicenter study. Am J Respir Crit Care Med 2004; 170: Randerath WJ, Schraeder O, Galetke W, et al. Autoadjusting CPAP therapy based on impedance efficacy, compliance and acceptance. Am J Respir Crit Care Med 2001; 163: CHEST / 129 / 3/ MARCH,

O bstructive sleep apnoea (OSA) is a common condition

O bstructive sleep apnoea (OSA) is a common condition 226 SLEEP DISORDERED BREATHING Comparison of three ways to determine and deliver pressure during nasal CPAP therapy for obstructive sleep apnoea S D West, D R Jones, J R Stradling... See end of article

More information

Comparison of three ways to determine and deliver pressure during nasal CPAP therapy for Obstructive Sleep Apnoea

Comparison of three ways to determine and deliver pressure during nasal CPAP therapy for Obstructive Sleep Apnoea Thorax Online First, published on October 27, 2005 as 10.1136/thx.2005.046300 Comparison of three ways to determine and deliver pressure during nasal CPAP therapy for Obstructive Sleep Apnoea Sophie D

More information

Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea

Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea Eur Respir J 2006; 28: 159 164 DOI: 10.1183/09031936.06.00127205 CopyrightßERS Journals Ltd 2006 Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea G.M. Nolan, S.

More information

Pressure-Relief Continuous Positive Airway Pressure vs Constant Continuous Positive Airway Pressure* A Comparison of Efficacy and Compliance

Pressure-Relief Continuous Positive Airway Pressure vs Constant Continuous Positive Airway Pressure* A Comparison of Efficacy and Compliance Original Research SLEEP MEDICINE Pressure-Relief Continuous Positive Airway Pressure vs Constant Continuous Positive Airway Pressure* A Comparison of Efficacy and Compliance Georg Nilius, MD; Andreas Happel;

More information

VARIABLE-PRESSURE VS. FIXED-PRESSURE CPAP. The obstructive sleep apnea/hypopnea syndrome

VARIABLE-PRESSURE VS. FIXED-PRESSURE CPAP. The obstructive sleep apnea/hypopnea syndrome VARIABLE-PRESSURE VS. FIXED-PRESSURE CPAP Randomized Controlled Trial of Variable-Pressure Versus Fixed-Pressure Continuous Positive Airway Pressure (CPAP) Treatment for Patients with Obstructive Sleep

More information

Key words: AutoSet; continuous positive airway pressure; obstructive sleep apnea; snoring; SOMNOsmart

Key words: AutoSet; continuous positive airway pressure; obstructive sleep apnea; snoring; SOMNOsmart Efficacy of Flow- vs Impedance-Guided Autoadjustable Continuous Positive Airway Pressure* A Randomized Cross-over Trial Dirk A. Pevernagie, MD, PhD; Pascal M. Proot, MD; Katrien B. Hertegonne, MD; Marleen

More information

Comparative Efficacy of Two Expiratory Pressure Reduction Systems in the Treatment of Obstructive Sleep Apnea

Comparative Efficacy of Two Expiratory Pressure Reduction Systems in the Treatment of Obstructive Sleep Apnea Sleep Diagnosis and Therapy 2009 V4 N1 Wolfe et al Comparative Efficacy of Two Expiratory Pressure Reduction Systems in the Treatment of Obstructive Sleep Apnea Lisa Wolfe, MD, 1 Clifford A. Massie, PhD,

More information

DECLARATION OF CONFLICT OF INTEREST

DECLARATION OF CONFLICT OF INTEREST DECLARATION OF CONFLICT OF INTEREST Obstructive sleep apnoea How to identify? Walter McNicholas MD Newman Professor in Medicine, St. Vincent s University Hospital, University College Dublin, Ireland. Potential

More information

Obstructive sleep apnoea How to identify?

Obstructive sleep apnoea How to identify? Obstructive sleep apnoea How to identify? Walter McNicholas MD Newman Professor in Medicine, St. Vincent s University Hospital, University College Dublin, Ireland. Potential conflict of interest None Obstructive

More information

Comfort and pressure profiles of two autoadjustable positive airway pressure devices: a technical report

Comfort and pressure profiles of two autoadjustable positive airway pressure devices: a technical report Respiratory Medicine (23) 97, 93 98 Comfort and pressure profiles of two autoadjustable positive airway pressure devices: a technical report Katrien B. Hertegonne, Pascal M. Proot, Romain A. Pauwels, Dirk

More information

Goals Session Putting it all Together

Goals Session Putting it all Together Goals Session Putting it all Together Adherence to therapy Case presentations Treatment issues Charles Atwood, MD, FCCP, FAASM University of Pittsburgh Disclosures Advisory - Carecore National; Philips;

More information

RESEARCH PACKET DENTAL SLEEP MEDICINE

RESEARCH PACKET DENTAL SLEEP MEDICINE RESEARCH PACKET DENTAL SLEEP MEDICINE American Academy of Dental Sleep Medicine Dental Sleep Medicine Research Packet Page 1 Table of Contents Research: Oral Appliance Therapy vs. Continuous Positive Airway

More information

(To be filled by the treating physician)

(To be filled by the treating physician) CERTIFICATE OF MEDICAL NECESSITY TO BE ISSUED TO CGHS BENEFICIAREIS BEING PRESCRIBED BILEVEL CONTINUOUS POSITIVE AIRWAY PRESSURE (BI-LEVEL CPAP) / BI-LEVEL VENTILATORY SUPPORT SYSTEM Certification Type

More information

The Epworth Sleepiness Scale (ESS) was developed by Johns

The Epworth Sleepiness Scale (ESS) was developed by Johns Clinical Reproducibility of the Epworth Sleepiness Scale Anh Tu Duy Nguyen, M.D. 1 ; Marc A. Baltzan, M.D., M.Sc. 1,2 ; David Small, M.D. 1 ; Norman Wolkove, M.D. 1 ; Simone Guillon, M.D. 3 ; Mark Palayew,

More information

Christopher D. Turnbull 1,2, Daniel J. Bratton 3, Sonya E. Craig 1, Malcolm Kohler 3, John R. Stradling 1,2. Original Article

Christopher D. Turnbull 1,2, Daniel J. Bratton 3, Sonya E. Craig 1, Malcolm Kohler 3, John R. Stradling 1,2. Original Article Original Article In patients with minimally symptomatic OSA can baseline characteristics and early patterns of CPAP usage predict those who are likely to be longer-term users of CPAP Christopher D. Turnbull

More information

Comparison of two in-laboratory titration methods to determine evective pressure levels in patients with obstructive sleep apnoea

Comparison of two in-laboratory titration methods to determine evective pressure levels in patients with obstructive sleep apnoea Thorax 2000;55:741 745 741 Centre de Recherche, Hôpital Laval, Institut Universitaire de Cardiologie et de Pneumologie de l Université Laval, Sainte-Foy, Québec G1V 4G5, Canada M P Bureau F Sériès Correspondence

More information

Patients commonly arouse from sleep and experience awake states.

Patients commonly arouse from sleep and experience awake states. Sensitive to Sleep Patients commonly arouse from sleep and experience awake states. PATIENTS are intolerant of the delivered pressure and can have difficulty returning to sleep. To aid the transition back

More information

Clinical Trials in OSA

Clinical Trials in OSA 24th ANNUAL ADVANCES IN SLEEP APNEA AND SNORING February 16-17, 2018 Grand Hyatt on Union Square San Francisco, California Clinical Trials in OSA Samuel T. Kuna, MD Department of Medicine Center for Sleep

More information

Prediction of sleep-disordered breathing by unattended overnight oximetry

Prediction of sleep-disordered breathing by unattended overnight oximetry J. Sleep Res. (1999) 8, 51 55 Prediction of sleep-disordered breathing by unattended overnight oximetry L. G. OLSON, A. AMBROGETTI ands. G. GYULAY Discipline of Medicine, University of Newcastle and Sleep

More information

Critical Review Form Diagnostic Test

Critical Review Form Diagnostic Test Critical Review Form Diagnostic Test Diagnosis and Initial Management of Obstructive Sleep Apnea without Polysomnography A Randomized Validation Study Annals of Internal Medicine 2007; 146: 157-166 Objectives:

More information

Management of OSA in the Acute Care Environment. Robert S. Campbell, RRT FAARC HRC, Philips Healthcare May, 2018

Management of OSA in the Acute Care Environment. Robert S. Campbell, RRT FAARC HRC, Philips Healthcare May, 2018 Management of OSA in the Acute Care Environment Robert S. Campbell, RRT FAARC HRC, Philips Healthcare May, 2018 1 Learning Objectives Upon completion, the participant should be able to: Understand pathology

More information

Obstructive sleep apnea (OSA) is characterized by. Quality of Life in Patients with Obstructive Sleep Apnea*

Obstructive sleep apnea (OSA) is characterized by. Quality of Life in Patients with Obstructive Sleep Apnea* Quality of Life in Patients with Obstructive Sleep Apnea* Effect of Nasal Continuous Positive Airway Pressure A Prospective Study Carolyn D Ambrosio, MD; Teri Bowman, MD; and Vahid Mohsenin, MD Background:

More information

A new beginning in therapy for women

A new beginning in therapy for women A new beginning in therapy for women OSA in women Tailored solutions for Her AutoSet for Her algorithm ResMed.com Women and OSA OSA has traditionally been considered to be a male disease. However, recent

More information

Nasal pressure recording in the diagnosis of sleep apnoea hypopnoea syndrome

Nasal pressure recording in the diagnosis of sleep apnoea hypopnoea syndrome 56 Unité de Recherche, Centre de Pneumologie de l Hôpital Laval, Université Laval, Québec, Canada F Sériès I Marc Correspondence to: Dr F Sériès, Centre de Pneumologie, 2725 Chemin Sainte Foy, Sainte Foy

More information

Diabetes & Obstructive Sleep Apnoea risk. Jaynie Pateraki MSc RGN

Diabetes & Obstructive Sleep Apnoea risk. Jaynie Pateraki MSc RGN Diabetes & Obstructive Sleep Apnoea risk Jaynie Pateraki MSc RGN Non-REM - REM - Both - Unrelated - Common disorders of Sleep Sleep Walking Night terrors Periodic leg movements Sleep automatism Nightmares

More information

Role of different nocturnal monitorings in the evaluation of CPAP titration by autocpap devices

Role of different nocturnal monitorings in the evaluation of CPAP titration by autocpap devices Respiratory Medicine (2005) 99, 313 320 Role of different nocturnal monitorings in the evaluation of CPAP titration by autocpap devices Oreste Marrone, Giuseppe Insalaco, Adriana Salvaggio, Giovanni Bonsignore

More information

Comparative assessment of several automatic CPAP devices responses: a bench test study

Comparative assessment of several automatic CPAP devices responses: a bench test study ORIGINAL ARTICLE SLEEP MEDICINE Comparative assessment of several automatic CPAP devices responses: a bench test study Valentina Isetta 1,2, Daniel Navajas 1,2,3, Josep M. Montserrat 2,4,5 and Ramon Farré

More information

Effect of two types of mandibular advancement splints on snoring and obstructive sleep apnoea

Effect of two types of mandibular advancement splints on snoring and obstructive sleep apnoea European Journal of Orthodontics 20 (1998) 293 297 1998 European Orthodontic Society Effect of two types of mandibular advancement splints on snoring and obstructive sleep apnoea J. Lamont*, D. R. Baldwin**,

More information

CPAP titration by an auto-cpap device based on snoring detection: a clinical trial and economic considerations

CPAP titration by an auto-cpap device based on snoring detection: a clinical trial and economic considerations Eur Respir J 199; : 759 7 DOI:.113/09031936.9.0759 Printed in UK - all rights reserved Copyright ERS Journals Ltd 199 European Respiratory Journal ISSN 0903-1936 CPAP titration by an auto-cpap device based

More information

In-Patient Sleep Testing/Management Boaz Markewitz, MD

In-Patient Sleep Testing/Management Boaz Markewitz, MD In-Patient Sleep Testing/Management Boaz Markewitz, MD Objectives: Discuss inpatient sleep programs and if they provide a benefit to patients and sleep centers Identify things needed to be considered when

More information

Continuous positive airway pressure titration for obstructive sleep apnoea: automatic versus manual titration

Continuous positive airway pressure titration for obstructive sleep apnoea: automatic versus manual titration < Supplementary tables are published online only. To view these files please visit the journal online (http://throax.bmj. com). 1 University of Western Australia, School of Medicine and Pharmacology, Respiratory

More information

Chronic NIV in heart failure patients: ASV, NIV and CPAP

Chronic NIV in heart failure patients: ASV, NIV and CPAP Chronic NIV in heart failure patients: ASV, NIV and CPAP João C. Winck, Marta Drummond, Miguel Gonçalves and Tiago Pinto Sleep disordered breathing (SDB), including OSA and central sleep apnoea (CSA),

More information

Morbidity and mortality of sleep-disordered breathing: obstructive sleep apnoea and car crash

Morbidity and mortality of sleep-disordered breathing: obstructive sleep apnoea and car crash All course materials, including the original lecture, are available as webcasts/podcasts at www.ers-education. org/sdb2009.htm Morbidity and mortality of sleep-disordered breathing: obstructive sleep apnoea

More information

The International Palestinian Congress in Sleep Medicine

The International Palestinian Congress in Sleep Medicine The International Palestinian Congress in Sleep Medicine Temporomandibular Disorders and Sleep Apnea 26 and 27 October, 2017 Notre Dame Hotel, Jerusalem Using PAP Downloads to Manage Sleep Apnea Patients

More information

Stephanie Mazza, Jean-Louis Pepin, Chrystele Deschaux, Bernadette Naegele, and Patrick Levy

Stephanie Mazza, Jean-Louis Pepin, Chrystele Deschaux, Bernadette Naegele, and Patrick Levy Analysis of Error Profiles Occurring during the OSLER Test A Sensitive Mean of Detecting Fluctuations in Vigilance in Patients with Obstructive Sleep Apnea Syndrome Stephanie Mazza, Jean-Louis Pepin, Chrystele

More information

José Haba-Rubio, MD; Jean-Paul Janssens, MD; Thierry Rochat, MD, PhD; and Emilia Sforza, MD, PhD

José Haba-Rubio, MD; Jean-Paul Janssens, MD; Thierry Rochat, MD, PhD; and Emilia Sforza, MD, PhD Rapid Eye Movement-Related Disordered Breathing* Clinical and Polysomnographic Features José Haba-Rubio, MD; Jean-Paul Janssens, MD; Thierry Rochat, MD, PhD; and Emilia Sforza, MD, PhD Objective: The existence

More information

Upper Airway Stimulation for Obstructive Sleep Apnea

Upper Airway Stimulation for Obstructive Sleep Apnea Upper Airway Stimulation for Obstructive Sleep Apnea Background, Mechanism and Clinical Data Overview Seth Hollen RPSGT 21 May 2016 1 Conflicts of Interest Therapy Support Specialist, Inspire Medical Systems

More information

Nasal Continuous Positive Airway Pressure Treatment: Current Realities and Future

Nasal Continuous Positive Airway Pressure Treatment: Current Realities and Future Sleep, 19(9):S13I-S135 1996 American Sleep Disorders Association and Sleep Research Society Nasal Continuous Positive Airway Pressure Treatment: Current Realities and Future *Michael Berthon-Jones, tsue

More information

Methods of Diagnosing Sleep Apnea. The Diagnosis of Sleep Apnea: Questionnaires and Home Studies

Methods of Diagnosing Sleep Apnea. The Diagnosis of Sleep Apnea: Questionnaires and Home Studies Sleep, 19(10):S243-S247 1996 American Sleep Disorders Association and Sleep Research Society Methods of Diagnosing Sleep Apnea J The Diagnosis of Sleep Apnea: Questionnaires and Home Studies W. Ward Flemons

More information

Polysomnography (PSG) (Sleep Studies), Sleep Center

Polysomnography (PSG) (Sleep Studies), Sleep Center Policy Number: 1036 Policy History Approve Date: 07/09/2015 Effective Date: 07/09/2015 Preauthorization All Plans Benefit plans vary in coverage and some plans may not provide coverage for certain service(s)

More information

Removal of the CPAP Therapy Device During Sleep and Its Association With Body Position Changes and Oxygen Desaturations

Removal of the CPAP Therapy Device During Sleep and Its Association With Body Position Changes and Oxygen Desaturations Removal of the CPAP Therapy Device During Sleep and Its Association With Body Position Changes and Oxygen Desaturations Yasuhiro Yamaguchi MD PhD, Shinichiro Hibi MD PhD, Masaki Ishii MD PhD, Yoko Hanaoka

More information

OSA and COPD: What happens when the two OVERLAP?

OSA and COPD: What happens when the two OVERLAP? 2011 ISRC Seminar 1 COPD OSA OSA and COPD: What happens when the two OVERLAP? Overlap Syndrome 1 OSA and COPD: What happens when the two OVERLAP? ResMed 10 JAN Global leaders in sleep and respiratory medicine

More information

Residual subjective daytime sleepiness under CPAP treatment in initially somnolent apnea patients: A pilot study using data mining methods

Residual subjective daytime sleepiness under CPAP treatment in initially somnolent apnea patients: A pilot study using data mining methods Sleep Medicine 9 (2008) 511 516 Original Article Residual subjective daytime sleepiness under CPAP treatment in initially somnolent apnea patients: A pilot study using data mining methods Xuân-Lan Nguyên

More information

In 1994, the American Sleep Disorders Association

In 1994, the American Sleep Disorders Association Unreliability of Automatic Scoring of MESAM 4 in Assessing Patients With Complicated Obstructive Sleep Apnea Syndrome* Fabio Cirignotta, MD; Susanna Mondini, MD; Roberto Gerardi, MD Barbara Mostacci, MD;

More information

Effective Treatment for Obstructive Sleep Apnoea

Effective Treatment for Obstructive Sleep Apnoea Effective Treatment for Obstructive Sleep Apnoea The Series of Positive Airway Pressure devices from DeVilbiss Healthcare is designed to meet the varied needs of people suffering from Obstructive Sleep

More information

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea Policy Number: Original Effective Date: MM.01.009 11/01/2009 Line(s) of Business: Current Effective Date: HMO; PPO;

More information

Joel Reiter*, Bashar Zleik, Mihaela Bazalakova, Pankaj Mehta, Robert Joseph Thomas

Joel Reiter*, Bashar Zleik, Mihaela Bazalakova, Pankaj Mehta, Robert Joseph Thomas Joel Reiter*, Bashar Zleik, Mihaela Bazalakova, Pankaj Mehta, Robert Joseph Thomas Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA *ומעבדת השינה הדסה עין כרם, ירושלים חיפפ ירושלים,

More information

PEDIATRIC PAP TITRATION PROTOCOL

PEDIATRIC PAP TITRATION PROTOCOL PURPOSE In order to provide the highest quality care for our patients, our sleep disorders facility adheres to the AASM Standards of Accreditation. The accompanying policy and procedure on pediatric titrations

More information

Philip L. Smith, MD; Christopher P. O Donnell, PhD; Lawrence Allan, BS; and Alan R. Schwartz, MD

Philip L. Smith, MD; Christopher P. O Donnell, PhD; Lawrence Allan, BS; and Alan R. Schwartz, MD A Physiologic Comparison of Nasal and Oral Positive Airway Pressure* Philip L. Smith, MD; Christopher P. O Donnell, PhD; Lawrence Allan, BS; and Alan R. Schwartz, MD Study objectives: The effectiveness

More information

OBSTRUCTIVE SLEEP APNEA and WORK Treatment Update

OBSTRUCTIVE SLEEP APNEA and WORK Treatment Update OBSTRUCTIVE SLEEP APNEA and WORK Treatment Update David Claman, MD Professor of Medicine Director, UCSF Sleep Disorders Center 415-885-7886 Disclosures: None Chronic Sleep Deprivation (0 v 4 v 6 v 8 hrs)

More information

Technology Report. Auto-titrating Nasal Continuous Positive Airway Pressure Systems in the Management of Obstructive Sleep Apnea

Technology Report. Auto-titrating Nasal Continuous Positive Airway Pressure Systems in the Management of Obstructive Sleep Apnea Technology Report Issue 39 September 2003 Auto-titrating Nasal Continuous Positive Airway Pressure Systems in the Management of Obstructive Sleep Apnea Publications can be requested from: CCOHTA 600-865

More information

Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea

Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea Original Article Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea Ahmad Khajeh-Mehrizi 1,2 and Omid Aminian 1 1. Occupational Sleep

More information

O bstructive sleep apnoea (OSA) is common in the

O bstructive sleep apnoea (OSA) is common in the 945 SLEEP DISORDERED BREATHING Prevalence of obstructive sleep apnoea in men with type 2 diabetes S D West, D J Nicoll, J R Stradling... See end of article for authors affiliations... Correspondence to:

More information

Monitoring: gas exchange, poly(somno)graphy or device in-built software?

Monitoring: gas exchange, poly(somno)graphy or device in-built software? Monitoring: gas exchange, poly(somno)graphy or device in-built software? Alessandro Amaddeo Noninvasive ventilation and Sleep Unit & Inserm U 955 Necker Hospital, Paris, France Inserm Institut national

More information

Effective Treatment for Obstructive Sleep Apnoea

Effective Treatment for Obstructive Sleep Apnoea Effective Treatment for Obstructive Sleep Apnoea The Series of Positive Airway Pressure devices from DeVilbiss Healthcare is designed to meet the varied needs of people suffering from Obstructive Sleep

More information

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea Policy Number: Original Effective Date: MM.01.009 11/01/2009 Line(s) of Business: Current Effective Date: HMO; PPO

More information

Persistence of Upper-Airway Symptoms During CPAP Compromises Adherence at 1 Year

Persistence of Upper-Airway Symptoms During CPAP Compromises Adherence at 1 Year Persistence of Upper-Airway Symptoms During CPAP Compromises Adherence at 1 Year Hanna-Riikka Kreivi MD PhD, Paula Maasilta MD PhD, and Adel Bachour MD PhD BACKGROUND: The most common adverse effects of

More information

AutoAdjust CPAP Technology

AutoAdjust CPAP Technology CLINICAL OVERVIEW: AutoAdjust CPAP Technology Autotitrating positive airway pressure (PAP) devices have been on the market for 15 years. DeVilbiss Healthcare, in an effort to illuminate and clarify primary

More information

Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease

Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease Cardiovascular diseases remain the number one cause of death worldwide Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease Shaoguang Huang MD Department

More information

NON-INVASIVE VENTILATION MADE RIDICULOUSLY SIMPLE

NON-INVASIVE VENTILATION MADE RIDICULOUSLY SIMPLE NON-INVASIVE VENTILATION MADE RIDICULOUSLY SIMPLE Jennifer Newitt, MD 3 rd year Pulmonary/Critical Care Fellow Mentor: Patrick Strollo Jr, MD Myth or Fact?!? Myth or Fact?!? Treatment for Obstructive

More information

Sedative Use During Continuous Positive Airway Pressure Titration Improves Subsequent Compliance. A Randomized, Double-Blind, Placebo-Controlled Trial

Sedative Use During Continuous Positive Airway Pressure Titration Improves Subsequent Compliance. A Randomized, Double-Blind, Placebo-Controlled Trial CHEST Sedative Use During Continuous Positive Airway Pressure Titration Improves Subsequent Compliance A Randomized, Double-Blind, Placebo-Controlled Trial Christopher J. Lettieri, MD, FCCP; Jacob F. Collen,

More information

Controlled, prospective trial of psychosocial function before and after continuous positive airway pressure therapy

Controlled, prospective trial of psychosocial function before and after continuous positive airway pressure therapy Eur Respir J 2001; 18: 996 1002 Copyright #ERS Journals Ltd 2001 DOI: 10.1183/09031936.01.00209301 European Respiratory Journal Printed in UK all rights reserved ISSN 0903-1936 Controlled, prospective

More information

Positive Airway Pressure Therapy of Obstructive Sleep Apnea

Positive Airway Pressure Therapy of Obstructive Sleep Apnea 종 설 J Kor Sleep Soc / Volume 1 / June, 2006 Positive Airway Pressure Therapy of Obstructive Sleep Apnea Ho-Won Lee, M.D., Sung-Pa Park, M.D. Department of Neurology, Kyungpook National School of Medicine

More information

QUESTIONS FOR DELIBERATION

QUESTIONS FOR DELIBERATION New England Comparative Effectiveness Public Advisory Council Public Meeting Hartford, Connecticut Diagnosis and Treatment of Obstructive Sleep Apnea in Adults December 6, 2012 UPDATED: November 28, 2012

More information

Auto Servo Ventilation Indications, Basics of Algorithm, and Titration

Auto Servo Ventilation Indications, Basics of Algorithm, and Titration Auto Servo Ventilation Indications, Basics of Algorithm, and Titration 1 ASV Learning Objectives Understand the indications for Auto Servo Ventilation Differentiate obstructive versus central hypopneas

More information

POLICY All patients will be assessed for risk factors associated with OSA prior to any surgical procedures.

POLICY All patients will be assessed for risk factors associated with OSA prior to any surgical procedures. Revised Date: Page: 1 of 7 SCOPE All Pre-Admission Testing (PAT) and Same Day Surgery (SDS) nurses at HRMC. PURPOSE The purpose of this policy is to provide guidelines for identifying surgical patients

More information

Commissioning Policy Individual Funding Request

Commissioning Policy Individual Funding Request Commissioning Policy Individual Funding Request Continuous Positive Airway Pressure (CPAP) Treatment of Obstructive Sleep Apnoea/Hypopnoea Syndrome (OSAHS) Criteria Based Access Policy Date Adopted: 13

More information

Policy Specific Section: October 1, 2010 January 21, 2013

Policy Specific Section: October 1, 2010 January 21, 2013 Medical Policy Bi-level Positive Airway Pressure (BPAP/NPPV) Type: Medical Necessity/Not Medical Necessity Policy Specific Section: Durable Medical Equipment Original Policy Date: Effective Date: October

More information

Research Article Accuracy of Positive Airway Pressure Device Measured Apneas and Hypopneas: Role in Treatment Followup

Research Article Accuracy of Positive Airway Pressure Device Measured Apneas and Hypopneas: Role in Treatment Followup Sleep Disorders Volume 213, Article ID 314589, 6 pages http://dx.doi.org/1.1155/213/314589 Research Article Accuracy of Positive Airway Pressure Device Measured Apneas and Hypopneas: Role in Treatment

More information

Prevalence of and treatment outcomes for patients with obstructive sleep apnoea identified by preoperative screening compared with clinician referrals

Prevalence of and treatment outcomes for patients with obstructive sleep apnoea identified by preoperative screening compared with clinician referrals ERJ Express. Published on March 30, 2016 as doi: 10.1183/13993003.01503-2015 ORIGINAL ARTICLE IN PRESS CORRECTED PROOF Prevalence of and treatment outcomes for patients with obstructive sleep apnoea identified

More information

G Popescu, M Latham, V Allgar, M W Elliott

G Popescu, M Latham, V Allgar, M W Elliott Thorax 2001;56:727 733 727 Department of Respiratory Medicine, St James s University Hospital, Leeds LS9 7TF, UK G Popescu M Latham M W Elliott Department of Medical Statistics V Allgar Correspondence

More information

The most accurate predictors of arterial hypertension in patients with Obstructive Sleep Apnea Syndrome

The most accurate predictors of arterial hypertension in patients with Obstructive Sleep Apnea Syndrome The most accurate predictors of arterial hypertension in patients with Obstructive Sleep Apnea Syndrome Natsios Georgios University Hospital of Larissa, Greece Definitions Obstructive Sleep Apnea (OSA)

More information

Inspire Therapy for Obstructive Sleep Apnea. Clinical Data Update

Inspire Therapy for Obstructive Sleep Apnea. Clinical Data Update Inspire Therapy for Obstructive Sleep Apnea For OSA patients unable to tolerate or get consistent benefit from CPAP Clinical Data Update Summary of published long-term 3-year outcomes data Treatment Goals

More information

Provent Therapy for Obstructive Sleep Apnea: Impact of Nasal Obstruction

Provent Therapy for Obstructive Sleep Apnea: Impact of Nasal Obstruction The Laryngoscope VC 2015 The American Laryngological, Rhinological and Otological Society, Inc. Provent Therapy for Obstructive Sleep Apnea: Impact of Nasal Obstruction Michael Friedman, MD; Michelle S

More information

International Journal of Scientific & Engineering Research Volume 9, Issue 1, January ISSN

International Journal of Scientific & Engineering Research Volume 9, Issue 1, January ISSN International Journal of Scientific & Engineering Research Volume 9, Issue 1, January-2018 342 The difference of sleep quality between 2-channel ambulatory monitor and diagnostic polysomnography Tengchin

More information

Sleep and the Heart. Physiologic Changes in Cardiovascular Parameters during Sleep

Sleep and the Heart. Physiologic Changes in Cardiovascular Parameters during Sleep Sleep and the Heart Rami N. Khayat, MD Professor of Internal Medicine Medical Director, Department of Respiratory Therapy Division of Pulmonary, Critical Care and Sleep Medicine The Ohio State University

More information

Sleep and the Heart. Rami N. Khayat, MD

Sleep and the Heart. Rami N. Khayat, MD Sleep and the Heart Rami N. Khayat, MD Professor of Internal Medicine Medical Director, Department of Respiratory Therapy Division of Pulmonary, Critical Care and Sleep Medicine The Ohio State University

More information

T he daytime consequences of the obstructive

T he daytime consequences of the obstructive 68 REVIEW SERIES Sleep? 4: Sleepiness, cognitive function, and quality of life in obstructive sleep apnoea/hypopnoea syndrome H M Engleman, N J Douglas... Sleepiness, cognitive performance, and quality

More information

Sleep Apnea: Vascular and Metabolic Complications

Sleep Apnea: Vascular and Metabolic Complications Sleep Apnea: Vascular and Metabolic Complications Vahid Mohsenin, M.D. Professor of Medicine Yale University School of Medicine Director, Yale Center for Sleep Medicine Definitions Apnea: Cessation of

More information

What is SDB? Obstructive sleep apnea-hypopnea syndrome (OSAHS)

What is SDB? Obstructive sleep apnea-hypopnea syndrome (OSAHS) Have a Good Sleep? Estimated 70 million Americans have clinically significant sleep problems Chronic insomnias report decreased quality of life, memory and attention problems, decreased physical health

More information

Equivalence of Nasal and Oronasal Masks during Initial CPAP Titration for Obstructive Sleep Apnea Syndrome

Equivalence of Nasal and Oronasal Masks during Initial CPAP Titration for Obstructive Sleep Apnea Syndrome NASAL VERSUS ORO-NASAL MASKS DURING CPAP TITRATION FOR OSA DOI: 1.5665/SLEEP.1134 Equivalence of Nasal and Oronasal s during Initial CPAP Titration for Obstructive Apnea Syndrome Ming Teo, MBBS 2 ; Terence

More information

CPAP in the treatment of sleep apnea. Titration, fixed and intelligent systems. Marc Josep M Montserrat. Barcelona. Spain

CPAP in the treatment of sleep apnea. Titration, fixed and intelligent systems. Marc Josep M Montserrat. Barcelona. Spain CPAP in the treatment of sleep apnea. Titration, fixed and intelligent systems Marc 2015 Josep M Montserrat. Barcelona. Spain Faculty disclosure No 1. INTRODUCTION 2. TITRATION 3. AUTOMATIC CPAP 4. FOLLOW

More information

Precision Sleep Medicine

Precision Sleep Medicine Precision Sleep Medicine Picking Winners Improves Outcomes and Avoids Side-Effects North American Dental Sleep Medicine Conference February 17-18, 2017 Clearwater Beach, FL John E. Remmers, MD Conflict

More information

Snoring detection during auto-nasal continuous positive airway pressure

Snoring detection during auto-nasal continuous positive airway pressure Eur Respir J 2002; 19: 108 112 DOI: 10.1183/09031936.00213302 Printed in UK all rights reserved Copyright #ERS Journals Ltd 2002 European Respiratory Journal ISSN 0903-1936 Snoring detection during auto-nasal

More information

A Deadly Combination: Central Sleep Apnea & Heart Failure

A Deadly Combination: Central Sleep Apnea & Heart Failure A Deadly Combination: Central Sleep Apnea & Heart Failure Sanjaya Gupta, MD FACC FHRS Ohio State University Symposium May 10 th, 2018 Disclosures Boston Scientific: fellowship support, speaking honoraria

More information

OSA/OSAS Who is Fit to Drive? Stradling JR. Oxford Centre for Respiratory Medicine Churchill Hospital, Oxford

OSA/OSAS Who is Fit to Drive? Stradling JR. Oxford Centre for Respiratory Medicine Churchill Hospital, Oxford OSA/OSAS Who is Fit to Drive? Stradling JR. Oxford Centre for Respiratory Medicine Churchill Hospital, Oxford What we hear in the clinic Cost of road accidents (UK department of transport official figures)

More information

The intelligent solution. System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy

The intelligent solution. System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy The intelligent solution System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy Proven efficacy The System One REMstar Auto algorithm

More information

O bstructive sleep apnoea-hypopnoea (OSAH) is a highly

O bstructive sleep apnoea-hypopnoea (OSAH) is a highly 422 SLEEP-DISORDERED BREATHING Predictive value of automated oxygen saturation analysis for the diagnosis and treatment of obstructive sleep apnoea in a home-based setting V Jobin, P Mayer, F Bellemare...

More information

Appendix 1. Practice Guidelines for Standards of Adult Sleep Medicine Services

Appendix 1. Practice Guidelines for Standards of Adult Sleep Medicine Services Appendix 1 Practice Guidelines for Standards of Adult Sleep Medicine Services 1 Premises and Procedures Out-patient/Clinic Rooms Sleep bedroom for PSG/PG Monitoring/Analysis/ Scoring room PSG equipment

More information

Emerging Nursing Roles in Collaborative Management of Sleep Disordered Breathing and Obstructive Sleep Apnoea

Emerging Nursing Roles in Collaborative Management of Sleep Disordered Breathing and Obstructive Sleep Apnoea Emerging Nursing Roles in Collaborative Management of Sleep Disordered Breathing and Obstructive Sleep Apnoea Sigma Theta Tau International 28th International Nursing Research Congress 27-31 July 2017

More information

Management of OSA. saurabh maji

Management of OSA. saurabh maji Management of OSA saurabh maji INTRODUCTION Obstructive sleep apnea is a major public health problem Prevalence of OSAS in INDIA is 2.4% to 4.96% in men and 1% to 2 % in women In the rest of the world

More information

ResScan Report Interpretation Guide. FOR CLINICAL USE ONLY V1.0 June, 2011

ResScan Report Interpretation Guide. FOR CLINICAL USE ONLY V1.0 June, 2011 ResScan Report Interpretation Guide FOR CLINICAL USE ONLY V1.0 June, 2011 ResMed Software Support Hours 5:30 AM to 5:30 PM (Pacific Time) Monday Friday Phone +1 (800) 424-0737, Option 6 Email TechSupportUSA@resmed.com

More information

Evaluation of the Brussells Questionnaire as a screening tool

Evaluation of the Brussells Questionnaire as a screening tool ORIGINAL PAPERs Borgis New Med 2017; 21(1): 3-7 DOI: 10.5604/01.3001.0009.7834 Evaluation of the Brussells Questionnaire as a screening tool for obstructive sleep apnea syndrome Nóra Pető 1, *Terézia Seres

More information

Index. sleep.theclinics.com. Note: Page numbers of article titles are in boldface type.

Index. sleep.theclinics.com. Note: Page numbers of article titles are in boldface type. Index Note: Page numbers of article titles are in boldface type. A Acquired central hypoventilation syndrome, NPPV in children with, 475 Acute cardiogenic pulmonary edema, PAP therapy in, 394 395 Adaptive

More information

Coding for Sleep Disorders Jennifer Rose V. Molano, MD

Coding for Sleep Disorders Jennifer Rose V. Molano, MD Practice Coding for Sleep Disorders Jennifer Rose V. Molano, MD Accurate coding is an important function of neurologic practice. This section of is part of an ongoing series that presents helpful coding

More information

Intelligent CPAP systems: clinical experience

Intelligent CPAP systems: clinical experience Thorax 1998;53(Suppl 3):S49 54 S49 Ruhrlandklinik, Abteilung Pneumologie- Universitätsklinik, Tüschener Weg 40, D-45239 Essen, Germany H Teschler ResMed, 82 Waterloo Road, North Ryde, NSW 2113, Australia

More information

Sleep and the Heart Reversing the Effects of Sleep Apnea to Better Manage Heart Disease

Sleep and the Heart Reversing the Effects of Sleep Apnea to Better Manage Heart Disease 1 Sleep and the Heart Reversing the Effects of Sleep Apnea to Better Manage Heart Disease Rami Khayat, MD Professor of Internal Medicine Director, OSU Sleep Heart Program Medical Director, Department of

More information

Inspire Therapy for Sleep Apnea

Inspire Therapy for Sleep Apnea Inspire Therapy for Sleep Apnea Patient Guide Giving You the FREEDOM TO SLEEP Like Everyone Else Take Comfort. Inspire therapy can help. Inspire therapy is a breakthrough implantable treatment option for

More information

Therapy with ncpap: incomplete elimination of Sleep Related Breathing Disorder

Therapy with ncpap: incomplete elimination of Sleep Related Breathing Disorder Eur Respir J 2000; 16: 921±927 Printed in UK ± all rights reserved Copyright #ERS Journals Ltd 2000 European Respiratory Journal ISSN 0903-1936 Therapy with ncpap: incomplete elimination of Sleep Related

More information

Sleep Apnea: Diagnosis & Treatment

Sleep Apnea: Diagnosis & Treatment Disclosure Sleep Apnea: Diagnosis & Treatment Lawrence J. Epstein, MD Sleep HealthCenters Harvard Medical School Chief Medical Officer for Sleep HealthCenters Sleep medicine specialty practice group Consultant

More information