The accuracy of non-invasive carbon dioxide monitoring: A clinical evaluation of two transcutaneous systems

Size: px
Start display at page:

Download "The accuracy of non-invasive carbon dioxide monitoring: A clinical evaluation of two transcutaneous systems"

Transcription

1 doi: /j x APPARATUS The accuracy of non-invasive carbon dioxide monitoring: A clinical evaluation of two transcutaneous systems D. Bolliger, 1 L. A. Steiner, 2 J. Kasper, 1 O. A. Aziz, 3 M. Filipovic 2 and M. D. Seeberger 2 1 Research fellow, 2 Consultant Anaesthetist, 3 Medical Student, Department of Anaesthesia and Intensive Care Unit, University of Basel Hospital, CH-4031 Basel, Switzerland Summary We determined the accuracy of two transcutaneous carbon dioxide monitoring systems (SenTec Digital Monitor with V-Sign Sensor and TOSCA 500 with TOSCA Sensor 92) for the measurement of single values and trends in the arterial partial pressure of carbon dioxide in 122 adult patients during major surgery and in 50 adult patients in the intensive care unit. One or several paired measurements were performed in each patient. The first measurement was used to determine the accuracy of a single value of transcutaneous carbon dioxide; the difference between the first and the last measurements was used to analyse the accuracy and to track trends. We defined a 95% limit of agreement of 1 kpa as being clinically useful. There was insufficient agreement between transcutaneous carbon dioxide partial pressure values derived from the two systems and arterial carbon dioxide values for both single values and trends as defined by our suggested limit of agreement. We conclude that these systems cannot replace conventional blood gas analysis in the clinical setting studied.... Correspondence to: Dr D. Bolliger dabolliger@uhbs.ch Accepted: 2 September 2006 The arterial partial pressure of carbon dioxide (P a CO 2 ) is an important indicator of respiratory function and ventilation that is monitored in patients in the operating theatre, in the intensive care unit (ICU) and during sedation [1]. Arterial blood gas analysis is the gold standard for P a CO 2 measurement but this method has major disadvantages because each determination of P a CO 2 requires withdrawal of arterial blood and laboratory analysis. Other disadvantages are that assessment of P a CO 2 is intermittent, relatively expensive and there is a time delay in obtaining results. In addition, the need for invasive monitoring is not without complications such as infection, haemorrhage and vessel occlusion. There are alternative methods for P a CO 2 measurement but they are currently too inaccurate to replace blood gas analysis. One such method is the monitoring of end-tidal carbon dioxide values (F É CO 2 ). This can be performed non-invasively and continuously in intubated patients but typically underestimates P a CO 2 ; its dependency on ventilation-perfusion ratio and cardiac output further decreases its value as an indicator of P a CO 2 [2]. Recently, new miniaturised systems for continuous and non-invasive carbon dioxide monitoring have been introduced into clinical practice. Based on Stow- Severinghaus electrodes, these systems measure the transcutaneous partial pressure of carbon dioxide (P tc CO 2 ) after arterialisation of the skin by local warming. The sensors are attached to the ear lobe and provide continuous, non-invasive on-line information about P tc CO 2 in addition to transcutaneous oxygen saturation. Using this method, good agreement has been reported between P tc CO 2 and P a CO 2 [3 5]; however, one study only evaluated volunteers [4], and two studies were not independent of the manufacturers [3, 4]. We therefore conducted a prospective study to test the accuracy and clinical usefulness of two of these transcutaneous systems in patients undergoing major surgery and in intubated patients in the ICU. We defined the required precision of the transcutaneous 394 Journal compilation Ó 2007 The Association of Anaesthetists of Great Britain and Ireland

2 D. Bolliger et al. Æ Transcutaneous carbon dioxide monitoring systems for clinical usefulness as being 95% limit of agreement 1 kpa [6]. Methods The study protocol was approved by the Local Research Ethics Committee, and all subjects gave written, informed consent. The study consisted of two parts, one performed in the operating theatre and the other in the ICU. Patients were eligible for the operating theatre part of the study if they were scheduled for major surgery that required placement of an indwelling arterial catheter and repeated arterial blood gas analysis. Patients were eligible for the ICU part of the study if their lungs were mechanically ventilated because of a planned, delayed tracheal extubation after major surgery or for any nonsurgical reason. Four of these latter patients were included in the study after written, informed consent had been obtained from their relatives, they also gave written, informed consent themselves after tracheal extubation. The ear lobes of each patient were inspected before and after the study. Pre-existing lesions on the ear lobe were an exclusion criterion. All measurements in the operating theatre and the ICU were performed during stable cardiovascular conditions and during nasopharyngeal or tympanic normothermia, defined as a body temperature of C. In all patients in the operating theatre and in 32 (64%) patients in the ICU, measurements were performed during stable volume-controlled ventilation. In 18 (36%) patients in the ICU, measurements were performed during weaning from volume-controlled ventilation to assisted spontaneous breathing. Paired measurements could be obtained in all of these latter 18 patients with the SenTec monitor (SenTec Digital Monitor System with V-Sign TM Sensor, Sentec AG, Therwil, Switzerland) and in 15 patients with the TOSCA monitor (TOSCA 500 with TOSCA Sensor 92, Linde Medical Sensors, Basel, Switzerland). Before starting the measurements, the patients haemodynamic values, body mass index, laboratory values, medical history and medications were noted. The SenTec and TOSCA measurement systems have been described previously [3, 7, 8]. The sensors are automatically calibrated using one-point dry gas calibration with 8% carbon dioxide (SenTec) and 7% carbon dioxide (TOSCA). To induce local vasodilation and to enhance skin permeability to carbon dioxide to improve gas diffusion at the measurement sites, the sensor in the SenTec system is heated to 42 C. The TOSCA sensor is heated to 44 C for the first 20 min to prevent initial P tc CO 2 overshoot and heating then continued at 42 C [9]. The sensors were attached to the inner aspect of the ear lobe with the monitor-specific clip. Before sensor placement, the skin was cleaned with alcohol and then dried, and one drop of contact gel was applied to the centre of the sensor. If a monitor indicated bad signal quality, the sensor was repositioned and the system was recalibrated. End-tidal carbon dioxide and P tc CO 2 were measured at the same time as arterial blood gas sampling in both the operating theatre (ADU, Datex-Ohmeda, Bromma, Sweden) and the ICU (Evita 4, Dräger Medical, Lübeck, Germany). End-tidal carbon dioxide values during singlelung ventilation were excluded from analysis, as were values obtained during or after cardiopulmonary bypass in cardiac surgical patients. Arterial blood gas analyses were performed either by an ABL 700 series (Radiometer, Copenhagen, Denmark) or a GEM Premier 3000 (Instrumentation Laboratory, Barcelona, Spain) system. Arterial carbon dioxide values were compared with simultaneously obtained P tc CO 2 values from the SenTec and TOSCA systems, as well as with F É CO 2 values. Haemodynamic and respiratory parameters and the use of vasopressors at the time of each simultaneous carbon dioxide analysis were noted. For the analysis of the accuracy of the measurement of single values in the operating theatre or the ICU, the first measurement with a good signal quality in each patient was included. The proportion of P tc CO 2 values that were within the required precision of ± 1 kpa was determined. The trend, defined as the difference in P a CO 2 over time, was compared with the corresponding trends in P tc CO 2 and F É CO 2 in patients having multiple measurements. To calculate the trend, the first and the last carbon dioxide measurements from the SenTec and TOSCA systems with good quality contact as indicated by the systems were used. Receiver operator characteristic (ROC) curves for P tc CO 2 were constructed to determine the accuracy of both systems to detect hypercapnia, defined as P a CO 2 > 5.5 kpa. These ROC curves were compared with the analogous ROC curve for F É CO 2. The optimal cut-off points were determined. The proportion of hypercapnic P a CO 2 values detected by the cut-off points for each of the two systems was determined. The authors performed all measurements and analyses. The bias and 95% limit of agreement (defined as mean difference: 1.96 SD of the difference) were calculated as described by Bland and Altman [10]. The two P tc CO 2 measurement systems were compared by using the Chi-squared test. Other results are expressed as mean (SD) or as median (IQR [range]). Data were analysed using SPSS 13.0 (SPSS Inc., Chicago, IL). Results The part of the study performed in the operating theatre included 122 patients. Of these patients, 93 (76%) underwent cardiac surgery, 28 (23%) lung or mediastinal Journal compilation Ó 2007 The Association of Anaesthetists of Great Britain and Ireland 395

3 D. Bolliger et al. Æ Transcutaneous carbon dioxide monitoring Anaesthesia, 2007, 62, pages Table 1 Patients characteristics. Values are mean (SD) or number (%). Operating theatre patients, n ¼ 122 Intensive care patients, n ¼ 50 Age; years 63.8 (11.8) 66.7 (12.5) Weight; kg 76.1 (16.3) 78.2 (15.8) Height; cm (9.3) (8.6) Body mass index; kg.m ) (4.3) 26.8 (3.9) Sex; M : F 89 : : 10 ASA physical status (8%) 0 ASA physical status (89%) 38 (76%) ASA physical status 4 4 (3%) 12 (24%) (kpa) SenTec surgery and one (1%) neurosurgery. The ICU part of the study included 50 patients, of which 46 (92%) were admitted for lung ventilation after cardiac surgery, two (4%) for severe sepsis and two (4%) for neurological reasons (one ischaemic stroke and one subarachnoid haemorrhage). A total of 42 cardiac surgical patients were studied both in the operating theatre and in the ICU. Table 1 shows the patients characteristics. To maintain haemodynamic stability in the operating theatre, 60 (49%) patients required pharmacological support within half an hour of arterial blood gas sampling: ephedrine was given to 15 patients at a median (IQR [range]) dose of 5 (2.5 5 [2.5 20]) mg and phenylephrine to 52 patients at a median (IQR [range]) dose of 0.2 ( [5 0.75]) mg; 12 of these patients were given both ephedrine and phenylephrine. Five of these patients were also given noradrenaline in a dose of 1 lg.kg )1.min )1, and one patient was given both noradrenaline and adrenaline in doses of 4 lg.kg )1.min )1. To maintain cardiovascular stability in the ICU, 26 (52%) patients were given continuous pharmacological support during the study period: noradrenaline was given to 23 (46%) patients at continuously adjusted, varying doses of up to lg.kg )1.min )1, adrenaline to 17 (34%) patients at doses of up to 0.15 lg.kg )1.min )1 ; 14 of these patients received both adrenaline and noradrenaline. The monitoring equipment used in the investigation was well tolerated by all patients, and no adverse events such as skin lesions or thermal injury to the ear lobes were observed. Some loss of data resulted from inadvertent detachment of sensors from the ear lobes, insufficient signal quality of one or both systems at the time of arterial blood sampling or failure and unavailability of a monitoring system. Figures 1 and 2 show the agreement between the P tc CO 2 values indicated by the SenTec and the TOSCA systems and P a CO 2 values measured in the operating theatre. A total of 101 and 112 paired measurements with arterial blood gas analysis were obtained for the SenTec and TOSCA systems, respectively. Bias ± 95% limit of agreement (kpa) SenTec)/2 1 Figure 1 Bland-Altman plot for single values (first measurement) from the SenTec Digital Monitor with V-Sign Sensor in the operating theatre. (kpa) TOSCA (kpa) TOSCA)/2 1 Figure 2 Bland-Altman plot for single values (first measurement) from the TOSCA Sensor 92 in the operating theatre. between P a CO 2 and P tc CO 2 was ) 0.57 ± 2 kpa for the SenTec system and ) 0.29 ± 1.43 kpa for the TOSCA system. These results were similar in the equally large groups of patients with and without administration of vasopressors (data not given in detail). In total, 76% of the measurements from the SenTec system and 85% of the measurements from the TOSCA system were within the required precision of ± 1 kpa of the corresponding P a CO 2 value (v 2 ¼ 0.10). Figures 3 and 4 show the agreement between the SenTec and the TOSCA systems and P a CO 2 values in the ICU. A total of 49 paired measurements were obtained for each system. bias ± 95% limit of agreement between P a CO 2 and P tc CO 2 was ) 0.35 ± 1.53 kpa for the 396 Journal compilation Ó 2007 The Association of Anaesthetists of Great Britain and Ireland

4 D. Bolliger et al. Æ Transcutaneous carbon dioxide monitoring (kpa) SenTec trend (kpa) trend PtcCO 2 (kpa) SenTec Operating theatre ICU (kpa) SenTec)/2 Figure 3 Bland-Altman plot for single values (first measurement) from the SenTec Digital Monitor with V-Sign Sensor in the intensive care unit. (trend (kpa) + trend (kpa) SenTec)/2 Figure 5 Bland-Altman plot for trend (difference between the last minus the first measurement) from the SenTec Digital Monitor with V-Sign Sensor. Data from the operating theatre and the intensive care unit were pooled. (kpa) TOSCA trend (kpa) trend PtcCO 2 (kpa) TOSCA Operating theatre ICU (kpa) TOSCA)/2 Figure 4 Bland-Altman plot for single values (first measurement) from the TOSCA 500 with TOSCA Sensor 92 in the intensive care unit. SenTec system and )0.29 ± 1.25 kpa for the TOSCA system. The results in the ICU were also similar in the equally large groups of patients with and without administration of vasopressors (data not given in detail). In total, 82% of the measurements from the SenTec system and 86% of the measurements from the TOSCA system were within the required precision of ± 1 kpa of the corresponding P a CO 2 value (v 2 ¼ 0.79). Figures 5 and 6 show the agreement in the trend between P a CO 2 and the two transcutaneous systems. The data from the operating theatre and the ICU parts of the study were pooled for this analysis. For the SenTec system, 24 paired measurements were performed in the (trend (kpa) + trend PtcCO 2 (kpa) TOSCA)/2 Figure 6 Bland-Altman plot for trend (difference between the last minus the first measurement) from the TOSCA 500 with TOSCA Sensor 92. Data from the operating theatre and the intensive care unit were pooled. operating theatre and 47 in the ICU. Bias ± 95% limit of agreement for the trends of P a CO 2 and the SenTec P tc CO 2 was 0.44 ± 1.82 kpa. For the TOSCA system, 26 paired measurements were performed in the operating theatre and 46 in the ICU. Bias ± 95% limit of agreement for the trends of P a CO 2 and the TOSCA P tc CO 2 was ) 0.41 ± 1.39 kpa. Bias ± 95% limit of agreement was similar when patients whose lungs were ventilated in a volumecontrolled mode at both measurement points and patients who had been switched to assisted spontaneous breathing before the second measurement point were analysed separately. In the latter patient group, bias ± 95% limit Journal compilation Ó 2007 The Association of Anaesthetists of Great Britain and Ireland 397

5 D. Bolliger et al. Æ Transcutaneous carbon dioxide monitoring Anaesthesia, 2007, 62, pages of agreement for the trends of P a CO 2 and P tc CO 2 was 0.58 ± 1.23 kpa for the SenTec system (n ¼ 18) and ) 0.26 ± 1.49 for the TOSCA system (n ¼ 15). Bias ± 95% limit of agreement for single values of F É CO 2 was 0.78 ± 0.85 kpa (n ¼ 92) and 0.64 ± 1.25 kpa (n ¼ 50) in the operating theatre and in the ICU, respectively. Based on 11 paired values from the operating theatre and 49 values from the ICU, bias ± 95% limit of agreement for the trends of P a CO 2 and F É CO 2 was )0.26 ± 1.17 kpa. Receiver operator characteristic curves were constructed to study the ability of the transcutaneous systems to detect hypercapnia (P a CO 2 > 5.5 kpa). The data from the operating theatre and the ICU parts of the study were pooled for this analysis. Hypercapnia was present in 46 (27%) patients at the time of the first P a CO 2 determination. The area under the curve was 0.84 and 0.83 for the SenTec and TOSCA systems, respectively. The optimal cut-off point for the SenTec system was 5.58 kpa with a positive predictive value of 0.53 and a negative predictive value of The optimal cut-off point for the TOSCA system was 5.55 kpa with a positive predictive value of 0.50 and a negative predictive value of With these cut-off points, 93% and 86% of hypercapnic P a CO 2 levels were detected by the SenTec and TOSCA systems, respectively (v 2 ¼ 0.30). The same analysis was performed to analyse the value of F É CO 2 for detecting hypercapnia. The area under the curve was 0.75, the optimal cut-off point was 4.35 kpa, and the positive and negative predictive values were 0.40 and 0.93, respectively. This cut-off point detected 94% of hypercapnic P a CO 2 values. There was no significant difference in detection of hypercapnia compared with the transcutaneous systems. Discussion Our study found that the accuracy of the two new transcutaneous systems for non-invasive monitoring of P a CO 2 was below the level that we had defined as being clinically useful, i.e. 95% limit of agreement of ± 1 kpa [6]. The transcutaneous systems did not provide any additional information in ventilated patients with intubated tracheas because the P tc CO 2 values were no more accurate than the routinely available F É CO 2 values. We conclude that the two transcutaneous systems lack sufficient accuracy for monitoring P a CO 2 in ventilated surgical patients with intubated tracheas or critically ill patients. The gold standard for P a CO 2 monitoring, arterial blood sampling and laboratory analysis, has itself a degree of imprecision that is dependent on the P a CO 2 level. For example, the 95% limit of agreement of the ABL 700 series machine used in this study is ± 0.24 kpa at a physiological P a CO 2 of 5.5 kpa. Further imprecision can result from incorrect handling such as delayed transportation, high temperature or air in the syringe. We tried to eliminate these additional sources of imprecision by using only a small number of people to draw and immediately process the arterial blood samples according to a standardised procedure. Nevertheless, we postulated a 95% limit of agreement of ± 0.5 kpa for the arterial blood gas analysis and defined twice this range as the maximally acceptable 95% limit of agreement between P a CO 2 and P tc CO 2, i.e. ± 1.0 kpa. The findings of our study are in conflict with those of several published studies [3, 4, 11]. However, our results are in agreement with those of two recent studies performed in critically ill patients that reported 95% limits of agreement between P a CO 2 and P tc CO 2 of ± 1.6 kpa with the TOSCA system [6] and ± 1.2 kpa with the SenTec system [12]. However, we do not share the same conclusion as these investigators that these 95% limits of agreement indicate sufficient accuracy of the systems for reliable clinical use [6, 12]. The difference between our study and previous studies that found better agreement between P a CO 2 and P tc CO 2 may have several causes. One study was conducted in healthy volunteers [4], whereas all our patients underwent major surgery or were severely ill. Two previous studies were not conducted independently of the manufacturers [3, 4], whereas in our study design, data collection and manuscript preparation were. One might suspect that the use of vasopressors in our patients was another reason for the poor accuracy of the transcutaneous systems and the conflicting results, as the difference between P tc CO 2 and P a CO 2 is influenced by skin perfusion and skin temperature [13, 14]. However, we found that the use of vasopressors did not significantly affect P tc CO 2 measurements, which is in agreement with several previous studies [6, 11, 12, 15]. Moreover, fewer than half of the study patients in the ICU were given vasopressors, and higher doses of noradrenaline (> 0.15 lg.kg )1.min )1 ) were only necessary in 11% of these patients. Thus, the use of vasopressor agents did not seem to influence our P tc CO 2 measurements significantly. Hypothermia can also be excluded as a reason for differing results in our study because body temperatures were continuously monitored and all patients were normothermic. Finally, a low cardiac output also can be excluded because all patients were haemodynamically stable. The differences in P tc CO 2 measurements between the two transcutaneous systems were small. Overall, TOSCA showed slightly better, although still insufficient, accuracy in the operating theatre and in the ICU. End-tidal carbon dioxide monitoring is routinely performed to estimate P a CO 2 in intubated patients and was thus also analysed in our study. Although F É CO 2 has the disadvantages 398 Journal compilation Ó 2007 The Association of Anaesthetists of Great Britain and Ireland

6 D. Bolliger et al. Æ Transcutaneous carbon dioxide monitoring of usually underestimating P a CO 2 and being strongly dependent on ventilation-perfusion ratio and cardiac output [2], the F É CO 2 trend can be useful during stable conditions [16, 17]. Neither of the two transcutaneous systems was better than F É CO 2 in assessing single values or trends, or for detecting hypercapnia. In conclusion, the P tc CO 2 monitor systems we tested failed to indicate P a CO 2 reliably in ventilated surgical patients with intubated tracheas or in critically ill patients, and the systems were not beneficial in providing additional information when compared with routinely available F É CO 2 values. Therefore, arterial blood gas analysis still needs to be performed whenever precise assessment of P a CO 2 is needed. Acknowledgements SenTec AG, Therwil, Switzerland, provided a SenTec Digital Monitor System with V-Sign TM Sensor as well as all disposable parts for conduct of the study. Linde Medical Sensors AG, Basel, Switzerland, provided a TOSCA 500 system with a TOSCA Sensor 92 as well as all disposable parts. There are no conflicts of interest to declare. The authors thank Kai Monte RNA, for technical assistance and help with data collection in the operating theatre and Joan Etlinger BA, for editorial assistance. References 1 American Society of the Anesthesiologist Task Force. Practice guidelines for sedation and analgesia by nonanesthesiologists. Anesthesiology 2002; 96: Wahba RW, Tessler MJ. Misleading end-tidal CO 2 tensions. Canadian Journal of Anaesthesia 1996; 43: Eberhard P, Gisiger PA, Gardaz JP, Spahn DR. Combining transcutaneous blood gas measurement and pulse oximetry. Anesthesia and Analgesia 2002; 94: S Hayoz J, Rohling R, Tschupp A. World s first combined digital pulse oximetry and carbon dioxide tension ear sensor. Anesthesia and Analgesia 2002; 94 1S: Senn O, Clarenbach CF, Kaplan V, Maggiorini M, Bloch KE. Monitoring carbon dioxide tension and arterial oxygen saturation by a single earlobe sensor in patients with critical illness or sleep apnea. Chest 2005; 128: Bendjelid K, Schutz N, Stotz M, Gerard I, Suter PM, Romand JA. Transcutaneous PCO 2 monitoring in critically ill adults: clinical evaluation of a new sensor. Critical Care Medicine 2005; 33: Heuss LT, Chhajed PN, Schnieper P, Hirt T, Beglinger C. Combined pulse oximetry cutaneous carbon dioxide tension monitoring during colonoscopies: pilot study with a smart ear clip. Digestion 2004; 70: Kocher S, Rohling R, Tschupp A. Performance of a digital PCO 2 SPO 2 ear sensor. Journal of Clinical Monitoring and Computing 2004; 18: Kagawa S, Otani N, Kamide M, Gisiger PA, Eberhard P, Severinghaus JW. Initial transcutaneous PCO 2 overshoot with ear probe at 42 degrees C. Journal of Clinical Monitoring and Computing 2004; 18: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: Janssens JP, Howarth-Frey C, Chevrolet JC, Abajo B, Rochat T. Transcutaneous PCO 2 to monitor noninvasive mechanical ventilation in adults: assessment of a new transcutaneous PCO 2 device. Chest 1998; 113: Rodriguez P, Lellouche F, Aboab J, Brun Buisson C, Brochard L. Transcutaneous arterial carbon dioxide pressure monitoring in critically ill adult patients. Intensive Care Medicine 2006; 32: Burki NK, Albert RK. Noninvasive monitoring of arterial blood gases. A report of the ACCP section on respiratory pathophysiology. Chest 1983; 83: Hazinski TA, Severinghaus JW. Transcutaneous analysis of arterial PCO 2. Medical Instrumentation 1982; 16: Palmisano BW, Severinghaus JW. Transcutaneous PCO 2 and PO 2 : a multicenter study of accuracy. Journal of Clinical Monitoring 1990; 6: Hillen T, Sumpelmann R, Strauss JM. [Intraoperative changes in arterial end-tidal CO 2 partial pressure difference in interventions with constant ventilation-perfusion ratio]. Anästhesiologie und Reanimation 1999; 24: Sharma SK, McGuire GP, Cruise CJ. Stability of the arterial to end-tidal carbon dioxide difference during anaesthesia for prolonged neurosurgical procedures. Canadian Journal of Anaesthesia 1995; 42: Journal compilation Ó 2007 The Association of Anaesthetists of Great Britain and Ireland 399

Evaluation of a transcutaneous carbon dioxide monitor in severe obesity

Evaluation of a transcutaneous carbon dioxide monitor in severe obesity Intensive Care Med DOI 10.1007/s00134-008-1078-8 PHYSIOLOGICAL AND TECHNICAL NOTES Mauro Maniscalco Anna Zedda Stanislao Faraone Pierluigi Carratù Matteo Sofia Evaluation of a transcutaneous carbon dioxide

More information

Transcutaneous Carbon Dioxide Monitoring in Subjects With Acute Respiratory Failure and Severe Hypercapnia

Transcutaneous Carbon Dioxide Monitoring in Subjects With Acute Respiratory Failure and Severe Hypercapnia Transcutaneous Carbon Dioxide Monitoring in Subjects With Acute Respiratory Failure and Severe Hypercapnia Yolanda Ruiz MD, Eva Farrero MD PhD, Ana Córdoba MD PhD, Nuria González MD, Jordi Dorca MD PhD,

More information

Transcutaneous Monitoring and Case Studies

Transcutaneous Monitoring and Case Studies Transcutaneous Monitoring and Case Studies Objectives General concept, applications and principles of operation Role of TCM in clinical settings Role of TCM in home care settings Need for continuous TCM

More information

TITLE: Carbon Dioxide Monitoring Devices: Reliability, Clinical and Cost Effectiveness

TITLE: Carbon Dioxide Monitoring Devices: Reliability, Clinical and Cost Effectiveness TITLE: Carbon Dioxide Monitoring Devices: Reliability, Clinical and Cost Effectiveness DATE: 10 December 2008 RESEARCH QUESTIONS: 1. What is the comparative clinical effectiveness of different carbon dioxide

More information

Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis

Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis Gal S, Riskin A, Chistyakov I, Shifman N, Srugo I, and Kugelman A Pediatric Department and Pediatric Pulmonary Unit Bnai Zion

More information

British Journal of Anaesthesia 104 (6): (2010) doi: /bja/aeq092 Advance Access publication April 23, 2010

British Journal of Anaesthesia 104 (6): (2010) doi: /bja/aeq092 Advance Access publication April 23, 2010 RESPIRATION AND THE AIRWAY Detection of hypoventilation during deep sedation in patients undergoing ambulatory gynaecological hysteroscopy: a comparison between transcutaneous and nasal end-tidal carbon

More information

Is continuous transcutaneous monitoring of PCO 2 (TcPCO 2 ) over 8 h reliable in adults?

Is continuous transcutaneous monitoring of PCO 2 (TcPCO 2 ) over 8 h reliable in adults? RESPIRATORY MEDICINE (2001) 95, 331 335 doi:10.1053/rmed.2001.1045, available online at http://www.idealibrary.com on Is continuous transcutaneous monitoring of PCO 2 (TcPCO 2 ) over 8 h reliable in adults?

More information

SenTec Digital Monitoring System Transcutaneous PCO2 monitoring of patients with chronic respiratory failure PCO2 SpO2 PR

SenTec Digital Monitoring System Transcutaneous PCO2 monitoring of patients with chronic respiratory failure PCO2 SpO2 PR Digital Transcutaneous Blood Gas Monitoring SenTec Digital Monitoring System Transcutaneous PCO2 monitoring of patients with chronic respiratory failure PCO2 SpO2 PR Continuous Noninvasive Accurate Safe

More information

Effective peri-operative noninvasive PCO2 monitoring

Effective peri-operative noninvasive PCO2 monitoring Digital Transcutaneous Blood Gas Monitoring MONTHS COMPLETE WARRANTY SenTec Digital Monitoring System Effective peri-operative noninvasive PCO2 monitoring PCO2 SpO2 PR Continuous Noninvasive Accurate Safe

More information

Effective peri-operative noninvasive PCO2 monitoring

Effective peri-operative noninvasive PCO2 monitoring Digital Transcutaneous Blood Gas Monitoring MONTHS COMPLETE WARRANTY SenTec Digital Monitoring System Effective peri-operative noninvasive PCO2 monitoring PCO2 SpO2 PR Continuous Noninvasive Accurate Safe

More information

Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration

Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration The Journal of ExtraCorporeal Technology Original Articles Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration Anne Louise Bellaiche,

More information

SenTec V-Sign Illuminate Ventilation and Oxygenation. PCO2 SpO2 PR. Digital Transcutaneous Blood Gas Monitoring

SenTec V-Sign Illuminate Ventilation and Oxygenation. PCO2 SpO2 PR. Digital Transcutaneous Blood Gas Monitoring Digital Transcutaneous Blood Gas Monitoring SenTec V-Sign Illuminate Ventilation and Oxygenation MONTHS COMPLETE WARRANTY PCO2 SpO2 PR Continuous Noninvasive Easy to use Reliable Accurate Overcoming limitations

More information

Effects of the transcutaneous electrode temperature on the accuracy of transcutaneous carbon dioxide tension

Effects of the transcutaneous electrode temperature on the accuracy of transcutaneous carbon dioxide tension Scandinavian Journal of Clinical & Laboratory Investigation, 2011; 71: 548 552 ORIGINAL ARTICLE Effects of the transcutaneous electrode temperature on the accuracy of transcutaneous carbon dioxide tension

More information

Comparison of blood pressure measured at the arm, ankle and calf

Comparison of blood pressure measured at the arm, ankle and calf doi:10.1111/j.1365-2044.2008.05633.x Comparison of blood pressure measured at the arm, ankle and calf C. Moore, 1 A. Dobson, 2 M. Kinagi 2 and B. Dillon 3 1 SpR Anaesthetics, 2 Consultant Anaesthetist,

More information

European Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery

European Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery European Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery INTRODUCTION The European Board of Anaesthesiology regards it as essential that certain core

More information

http://dx.doi.org/10.1016/j.jemermed.2012.11.019 The Journal of Emergency Medicine, Vol. 45, No. 1, pp. 130 135, 2013 Copyright Ó 2013 Elsevier Inc. Printed in the USA. All rights reserved 0736-4679/$

More information

The measurement of blood gas oxygen and carbon

The measurement of blood gas oxygen and carbon Review Article The Design, Use, and Results of Transcutaneous Carbon Dioxide Analysis: Current and Future Directions Patrick Eberhard, PhD Transcutaneous carbon dioxide (CO 2 ) analysis was introduced

More information

Capnography: The Most Vital of Vital Signs. Tom Ahrens, PhD, RN, FAAN Research Scientist, Barnes-Jewish Hospital, St. Louis, MO May, 2017

Capnography: The Most Vital of Vital Signs. Tom Ahrens, PhD, RN, FAAN Research Scientist, Barnes-Jewish Hospital, St. Louis, MO May, 2017 Capnography: The Most Vital of Vital Signs Tom Ahrens, PhD, RN, FAAN Research Scientist, Barnes-Jewish Hospital, St. Louis, MO May, 2017 Assessing Ventilation and Blood Flow with Capnography Capnography

More information

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for 1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo

More information

Anesthesia Monitoring. D. J. McMahon rev cewood

Anesthesia Monitoring. D. J. McMahon rev cewood Anesthesia Monitoring D. J. McMahon 150114 rev cewood 2018-01-19 Key Points Anesthesia Monitoring: - Understand the difference between guidelines & standards - ASA monitoring Standard I states that an

More information

Non-Invasive Monitoring

Non-Invasive Monitoring Grey Nuns and Misericordia Community Hospital Approved by: Non-Invasive Monitoring Neonatal Policy & Procedures Manual : Assessment : Oct 2015 Date Effective Oct 2015 Gail Cameron Senior Director Operations,

More information

Edwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor

Edwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor Edwards FloTrac Sensor & Edwards Vigileo Monitor Performance Assessments of the FloTrac Sensor and Vigileo Monitor 1 Topics System Configuration Performance and Validation Dr. William T. McGee, Validation

More information

CLINICAL CONSIDERATIONS FOR THE BUNNELL LIFE PULSE HIGH-FREQUENCY JET VENTILATOR

CLINICAL CONSIDERATIONS FOR THE BUNNELL LIFE PULSE HIGH-FREQUENCY JET VENTILATOR CLINICAL CONSIDERATIONS FOR THE BUNNELL LIFE PULSE HIGH-FREQUENCY JET VENTILATOR 801-467-0800 Phone 800-800-HFJV (4358) Hotline TABLE OF CONTENTS Respiratory Care Considerations..3 Physician Considerations

More information

Sepsis Wave II Webinar Series. Sepsis Reassessment

Sepsis Wave II Webinar Series. Sepsis Reassessment Sepsis Wave II Webinar Series Sepsis Reassessment Presenters Nova Panebianco, MD Todd Slesinger, MD Fluid Reassessment in Sepsis Todd L. Slesinger, MD, FACEP, FCCM, FCCP, FAAEM Residency Program Director

More information

Pulse Oximeter Accuracy Study

Pulse Oximeter Accuracy Study Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709A Printed : 2014/5/26 This document is owned or controlled by Nihon Kohden and is protected by copyright law.

More information

British Journal of Anaesthesia 99 (4): (2007) doi: /bja/aem206 Advance Access publication July 25, 2007 Observational study of perioperat

British Journal of Anaesthesia 99 (4): (2007) doi: /bja/aem206 Advance Access publication July 25, 2007 Observational study of perioperat British Journal of Anaesthesia 99 (4): 567 71 (2007) doi:10.1093/bja/aem206 Advance Access publication July 25, 2007 Observational study of perioperative Ptc CO2 and Sp O2 in nonventilated patients receiving

More information

Permanent City Research Online URL:

Permanent City Research Online URL: Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement

More information

Transcutaneous CO2 Monitoring: Alerting the Anesthesia Provider to Impending Respiratory Depression

Transcutaneous CO2 Monitoring: Alerting the Anesthesia Provider to Impending Respiratory Depression Transcutaneous CO2 Monitoring: Alerting the Anesthesia Provider to Impending Respiratory Depression JEANETTE R BAUCHAT, MD, MS ASSOCIATE PROFESSOR OF ANESTHESIOLOGY DIVISION CHIEF, OBSTETRIC ANESTHESIOLOGY

More information

Intraoperative use of trancutaneous CO 2 measuring in paediatric and neonatal anaesthesia

Intraoperative use of trancutaneous CO 2 measuring in paediatric and neonatal anaesthesia Strategies and modern technologies of continous blood gas monitoring SenTec Symposium - Venice, November 1st 2017 Intraoperative use of trancutaneous CO 2 measuring in paediatric and neonatal anaesthesia

More information

A Comparison Of Transcutaneous And End-Tidal Carbon Dioxide Monitoring Among Three Devices Providing Supplemental Oxygen To Volunteers

A Comparison Of Transcutaneous And End-Tidal Carbon Dioxide Monitoring Among Three Devices Providing Supplemental Oxygen To Volunteers ISPUB.COM The Internet Journal of Anesthesiology Volume 34 Number 1 A Comparison Of Transcutaneous And End-Tidal Carbon Dioxide Monitoring Among Three Devices Providing Supplemental Oxygen To Volunteers

More information

Non-Invasive Assessment of Respiratory Function. Chapter 11

Non-Invasive Assessment of Respiratory Function. Chapter 11 Non-Invasive Assessment of Respiratory Function Chapter 11 Pulse Oximetry Laboratory measurements of ABG s are the gold standard for measuring levels of hypoxemia, however since these are performed intermittently

More information

Transcutaneous carbon dioxide in severe COPD patients during bronchoscopic lung volume reduction

Transcutaneous carbon dioxide in severe COPD patients during bronchoscopic lung volume reduction Respiratory Medicine (2011) 105, 602e607 available at www.sciencedirect.com journal homepage: www.elsevier.com/locate/rmed Transcutaneous carbon dioxide in severe COPD patients during bronchoscopic lung

More information

Comparison of central venous oxygen saturation and mixed venous oxygen saturation during liver transplantation

Comparison of central venous oxygen saturation and mixed venous oxygen saturation during liver transplantation Anaesthesia, 9,, pages 37 3 doi:1.1111/j.135-..5793.x Comparison of central venous oxygen saturation and mixed venous oxygen saturation during liver transplantation A. El Masry, 1 A. M. Mukhtar, 1 A. M.

More information

Capnography 101. James A Temple BA, NRP, CCP

Capnography 101. James A Temple BA, NRP, CCP Capnography 101 James A Temple BA, NRP, CCP Expected Outcomes 1. Gain a working knowledge of the physiology and science behind End-Tidal CO2. 2.Relate End-Tidal CO2 to ventilation, perfusion, and metabolism.

More information

FAILURE OF NONINVASIVE VENTILATION FOR DE NOVO ACUTE HYPOXEMIC RESPIRATORY FAILURE: ROLE OF TIDAL VOLUME

FAILURE OF NONINVASIVE VENTILATION FOR DE NOVO ACUTE HYPOXEMIC RESPIRATORY FAILURE: ROLE OF TIDAL VOLUME FAILURE OF NONINVASIVE VENTILATION FOR DE NOVO ACUTE HYPOXEMIC RESPIRATORY FAILURE: ROLE OF TIDAL VOLUME Guillaume CARTEAUX, Teresa MILLÁN-GUILARTE, Nicolas DE PROST, Keyvan RAZAZI, Shariq ABID, Arnaud

More information

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013

More information

Bobbia et al. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine (2015) 23:40 DOI /s

Bobbia et al. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine (2015) 23:40 DOI /s Bobbia et al. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine (2015) 23:40 DOI 10.1186/s13049-015-0120-4 ORIGINAL RESEARCH Open Access Concordance and limits between transcutaneous

More information

3/30/12. Luke J. Gasowski BS, BSRT, NREMT-P, FP-C, CCP-C, RRT-NPS

3/30/12. Luke J. Gasowski BS, BSRT, NREMT-P, FP-C, CCP-C, RRT-NPS Luke J. Gasowski BS, BSRT, NREMT-P, FP-C, CCP-C, RRT-NPS 1) Define and describe ETCO 2 2) Explain methods of measuring ETCO 2 3) Describe various clinical applications of ETCO 2 4) Describe the relationship

More information

Comparison of automated and static pulse respiratory mechanics during supported ventilation

Comparison of automated and static pulse respiratory mechanics during supported ventilation Comparison of automated and static pulse respiratory mechanics during supported ventilation Alpesh R Patel, Susan Taylor and Andrew D Bersten Respiratory system compliance ( ) and inspiratory resistance

More information

Venous ph can safely replace arterial ph in the initial evaluation of patients in the emergency department

Venous ph can safely replace arterial ph in the initial evaluation of patients in the emergency department Downloaded from emj.bmj.com on 23 November 2007 Venous ph can safely replace arterial ph in the initial evaluation of patients in the emergency department A-M Kelly, R McAlpine and E Kyle Emerg. Med. J.

More information

Monitoring: gas exchange, poly(somno)graphy or device in-built software?

Monitoring: gas exchange, poly(somno)graphy or device in-built software? Monitoring: gas exchange, poly(somno)graphy or device in-built software? Alessandro Amaddeo Noninvasive ventilation and Sleep Unit & Inserm U 955 Necker Hospital, Paris, France Inserm Institut national

More information

In 1994, the American Sleep Disorders Association

In 1994, the American Sleep Disorders Association Unreliability of Automatic Scoring of MESAM 4 in Assessing Patients With Complicated Obstructive Sleep Apnea Syndrome* Fabio Cirignotta, MD; Susanna Mondini, MD; Roberto Gerardi, MD Barbara Mostacci, MD;

More information

June 2011 Bill Streett-Training Section Chief

June 2011 Bill Streett-Training Section Chief Capnography 102 June 2011 Bill Streett-Training Section Chief Terminology Capnography: the measurement and numerical display of end-tidal CO2 concentration, at the patient s airway, during a respiratory

More information

NON INVASIVE LIFE SAVERS. Non Invasive Ventilation (NIV)

NON INVASIVE LIFE SAVERS. Non Invasive Ventilation (NIV) Table 1. NIV: Mechanisms Of Action Decreases work of breathing Increases functional residual capacity Recruits collapsed alveoli Improves respiratory gas exchange Reverses hypoventilation Maintains upper

More information

NITROUS OXIDE ELIMINATION AND DIFFUSION HYPOXIA DURING NORMO- AND HYPOVENTILATION

NITROUS OXIDE ELIMINATION AND DIFFUSION HYPOXIA DURING NORMO- AND HYPOVENTILATION British Journal of Anaesthesia 1993; 71: 189-193 NITROUS OXIDE ELIMINATION AND DIFFUSION HYPOXIA DURING NORMO- AND HYPOVENTILATION S. EINARSSON, O. STENQVIST, A. BENGTSSON, E. HOULTZ AND J. P. BENGTSON

More information

Pulse oximetry in the accident and emergency department

Pulse oximetry in the accident and emergency department Archives of Emergency Medicine, 1989, 6, 137-142 Pulse oximetry in the accident and emergency department C. J. HOLBURN & M. J. ALLEN Accident and Emergency Department, Leicester Royal Infirmary, Leicester,

More information

CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE

CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE 16 CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE OMAR GALAL, MO, PhO; NEIL WILSON, MO Pulse oximetry is a noninvasive

More information

Continuous monitoring of haemoglobin concentration after in-vivo adjustment in patients undergoing surgery with blood loss*

Continuous monitoring of haemoglobin concentration after in-vivo adjustment in patients undergoing surgery with blood loss* Anaesthesia 15, 7, 83 89 doi:1.1111/anae.138 Original Article Continuous monitoring of haemoglobin concentration after in-vivo adjustment in patients undergoing surgery with blood loss* D. Frasca, 1 H.

More information

Test Bank Pilbeam's Mechanical Ventilation Physiological and Clinical Applications 6th Edition Cairo

Test Bank Pilbeam's Mechanical Ventilation Physiological and Clinical Applications 6th Edition Cairo Instant dowload and all chapters Test Bank Pilbeam's Mechanical Ventilation Physiological and Clinical Applications 6th Edition Cairo https://testbanklab.com/download/test-bank-pilbeams-mechanical-ventilation-physiologicalclinical-applications-6th-edition-cairo/

More information

Pulse Oximeter Accuracy Study

Pulse Oximeter Accuracy Study Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709 Printed : 2008/12/24 This document is owned or controlled by Nihon Kohden and is protected by copyright law.

More information

What is. InSpectra StO 2?

What is. InSpectra StO 2? What is InSpectra StO 2? www.htibiomeasurement.com What is InSpectra StO 2? Hemoglobin O 2 saturation is measured in three areas: 1) Arterial (SaO 2, SpO 2 ) Assesses how well oxygen is loading onto hemoglobin

More information

Oxygen therapy increases the arterial partial pressure of

Oxygen therapy increases the arterial partial pressure of High flow or titrated oxygen for obese medical inpatients: a randomised crossover trial Janine Pilcher 1, Michael Richards 1, Leonie Eastlake 1, Steven J McKinstry 1, George Bardsley 1, Sarah Jefferies

More information

Capnography Connections Guide

Capnography Connections Guide Capnography Connections Guide Patient Monitoring Contents I Section 1: Capnography Introduction...1 I Section 2: Capnography & PCA...3 I Section 3: Capnography & Critical Care...7 I Section 4: Capnography

More information

Bobbia et al. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine (2015) 23:77 DOI /s

Bobbia et al. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine (2015) 23:77 DOI /s Bobbia et al. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine (2015) 23:77 DOI 10.1186/s13049-015-0154-7 ERRATUM Open Access Erratum: Concordance and limits between transcutaneous

More information

Neonatal Life Support Provider (NLSP) Certification Preparatory Materials

Neonatal Life Support Provider (NLSP) Certification Preparatory Materials Neonatal Life Support Provider (NLSP) Certification Preparatory Materials NEONATAL LIFE SUPPORT PROVIDER (NRP) CERTIFICATION TABLE OF CONTENTS NEONATAL FLOW ALGORITHM.2 INTRODUCTION 3 ANTICIPATION OF RESUSCITATION

More information

NBRC Exam RPFT Registry Examination for Advanced Pulmonary Function Technologists Version: 6.0 [ Total Questions: 111 ]

NBRC Exam RPFT Registry Examination for Advanced Pulmonary Function Technologists Version: 6.0 [ Total Questions: 111 ] s@lm@n NBRC Exam RPFT Registry Examination for Advanced Pulmonary Function Technologists Version: 6.0 [ Total Questions: 111 ] https://certkill.com NBRC RPFT : Practice Test Question No : 1 Using a peak

More information

Capnography for Pediatric Procedural Sedation Learning Module Last revised: February 18, 2014

Capnography for Pediatric Procedural Sedation Learning Module Last revised: February 18, 2014 Capnography for Pediatric Procedural Sedation Learning Module Last revised: February 18, 2014 Capnography 40 Non-invasive device that continually monitors EtCO 2 While pulse oximetry measures oxygen saturation,

More information

ASPIRUS WAUSAU HOSPITAL, INC. Passion for excellence. Compassion for people. SUBJECT: END TIDAL CARBON DIOXIDE MONITORING (CAPNOGRAPHY)

ASPIRUS WAUSAU HOSPITAL, INC. Passion for excellence. Compassion for people. SUBJECT: END TIDAL CARBON DIOXIDE MONITORING (CAPNOGRAPHY) Passion for excellence. Compassion for people. P&P REF : NEW 7-2011 ONBASE POLICY ID: 13363 REPLACES: POLICY STATUS : FINAL DOCUMENT TYPE: Policy EFFECTIVE DATE: 4/15/2014 PROPOSED BY: Respiratory Therapy

More information

OBJECTIVES OF TRAINING FOR THE ANAESTHESIA TERM

OBJECTIVES OF TRAINING FOR THE ANAESTHESIA TERM College of Intensive Care Medicine of Australia and New Zealand ABN: 16 134 292 103 Document type: Training Date established: 2007 Date last reviewed: 2014 OBJECTIVES OF TRAINING FOR THE ANAESTHESIA TERM

More information

W. J. RUSSELL*, M. F. JAMES

W. J. RUSSELL*, M. F. JAMES Anaesth Intensive Care 2004; 32: 644-648 The Effects on Arterial Haemoglobin Oxygen Saturation and on Shunt of Increasing Cardiac Output with Dopamine or Dobutamine During One-lung Ventilation W. J. RUSSELL*,

More information

IFT1 Interfacility Transfer of STEMI Patients. IFT2 Interfacility Transfer of Intubated Patients. IFT3 Interfacility Transfer of Stroke Patients

IFT1 Interfacility Transfer of STEMI Patients. IFT2 Interfacility Transfer of Intubated Patients. IFT3 Interfacility Transfer of Stroke Patients IFT1 Interfacility Transfer of STEMI Patients IFT2 Interfacility Transfer of Intubated Patients IFT3 Interfacility Transfer of Stroke Patients Interfacility Transfer Guidelines IFT 1 TRANSFER INTERFACILITY

More information

Comparison of patient spirometry and ventilator spirometry

Comparison of patient spirometry and ventilator spirometry GE Healthcare Comparison of patient spirometry and ventilator spirometry Test results are based on the Master s thesis, Comparison between patient spirometry and ventilator spirometry by Saana Jenu, 2011

More information

Frederic J., Gerges MD. Ghassan E. Kanazi MD., Sama, I. Jabbour-Khoury MD. Review article from Journal of clinical anesthesia 2006.

Frederic J., Gerges MD. Ghassan E. Kanazi MD., Sama, I. Jabbour-Khoury MD. Review article from Journal of clinical anesthesia 2006. Frederic J., Gerges MD. Ghassan E. Kanazi MD., Sama, I. Jabbour-Khoury MD. Review article from Journal of clinical anesthesia 2006 Introduction Laparoscopic surgery started in the mid 1950s. In recent

More information

How it Works. CO 2 is the smoke from the flames of metabolism 10/21/18. -Ray Fowler, MD. Metabolism creates ETC0 2 for excretion

How it Works. CO 2 is the smoke from the flames of metabolism 10/21/18. -Ray Fowler, MD. Metabolism creates ETC0 2 for excretion CO 2 is the smoke from the flames of metabolism -Ray Fowler, MD How it Works Metabolism creates ETC0 2 for excretion ETC02 and Oxygen are exchanged at the alveolar level in the lungs with each breath.

More information

POLICY. Number: Title: APPLICATION OF NON INVASIVE VENTILATION FOR ACUTE RESPIRATORY FAILURE. Authorization

POLICY. Number: Title: APPLICATION OF NON INVASIVE VENTILATION FOR ACUTE RESPIRATORY FAILURE. Authorization POLICY Number: 7311-60-024 Title: APPLICATION OF NON INVASIVE VENTILATION FOR ACUTE RESPIRATORY FAILURE Authorization [ ] President and CEO [ x ] Vice President, Finance and Corporate Services Source:

More information

Proposed presentation of data for ICU-ROX.

Proposed presentation of data for ICU-ROX. Proposed presentation of data for ICU-ROX. Version 1 was posted online on 21 November 2017 (prior to the interim analysis which occurred when the 500 th participant reached day 28). This version (version

More information

03RC1- Greif. Temperature Monitoring. Robert Greif - 1 -

03RC1- Greif. Temperature Monitoring. Robert Greif - 1 - 03RC1- Greif Temperature Monitoring Robert Greif Department of Anaesthesiology and Pain Therapy, University Hospital Bern, Inselspital Bern, Switzerland Small decreases of core body temperature during

More information

Guidelines Administrative Practice X Clinical Practice Professional Practice

Guidelines Administrative Practice X Clinical Practice Professional Practice Guidelines Administrative Practice X Clinical Practice Professional Practice End Tidal CO 2 (EtCO2) Monitoring Using the Zoll Monitor/Defibrillator Page 1 of 4 Scope: Outcome: Respiratory Therapists (RT),

More information

Materials and Methods

Materials and Methods Anesthesiology 2006; 104:701 7 2006 American Society of Anesthesiologists, Inc. Lippincott Williams & Wilkins, Inc. An Evaluation of Transcutaneous Carbon Dioxide Partial Pressure Monitoring during Apnea

More information

Assessing Agreement Between Methods Of Clinical Measurement

Assessing Agreement Between Methods Of Clinical Measurement University of York Department of Health Sciences Measuring Health and Disease Assessing Agreement Between Methods Of Clinical Measurement Based on Bland JM, Altman DG. (1986). Statistical methods for assessing

More information

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MD (ANAESTHESIOLOGY) FINAL EXAMINATION AUGUST 2013 Date : 2 nd August 2013 Time : 1.00 p.m. 4.00 p.m. Answer any three questions. Answer each question

More information

ETCO2 MONITORING NON-INTUBATED PATIENTS

ETCO2 MONITORING NON-INTUBATED PATIENTS Although the standard of care in ETC02 is well established for intubated patients, there has been little emphasis on the use of capnography in nonintubated patients till now. In addition to confirming

More information

Original Article. Postoperative hypothermia and patient outcomes after major elective non-cardiac surgery. Abstract

Original Article. Postoperative hypothermia and patient outcomes after major elective non-cardiac surgery. Abstract Original Article doi:10.1111/anae.12129 Postoperative and patient outcomes after major elective non-cardiac surgery D. Karalapillai, 1 D. Story, 2 G. K. Hart, 3 M. Bailey, 4 D. Pilcher, 5 A. Schneider,

More information

BTS Guideline for Home Oxygen use in adults Appendix 9 (online only) Key Questions - PICO 10 December 2012

BTS Guideline for Home Oxygen use in adults Appendix 9 (online only) Key Questions - PICO 10 December 2012 BTS Guideline for Home Oxygen use in adults Appendix 9 (online only) Key Questions - PICO 10 December 2012 Evidence base for Home Oxygen therapy in COPD, non-copd respiratory disease and nonrespiratory

More information

Postoperative hypothermia and patient outcomes after elective cardiac surgery

Postoperative hypothermia and patient outcomes after elective cardiac surgery doi:10.1111/j.1365-2044.2011.06784.x ORIGINAL ARTICLE Postoperative hypothermia and patient outcomes after elective cardiac surgery D. Karalapillai, 1 D. Story, 2 G. K. Hart, 3,4 M. Bailey, 5 D. Pilcher,

More information

Performance of capnometry in non-intubated infants in the pediatric intensive care unit

Performance of capnometry in non-intubated infants in the pediatric intensive care unit Coates et al. BMC Pediatrics 2014, 14:163 RESEARCH ARTICLE Performance of capnometry in non-intubated infants in the pediatric intensive care unit Bria M Coates 1*, Robin Chaize 2, Denise M Goodman 1 and

More information

NONINVASIVE CARBON DIOXIDE MONITORING IN A PORCINE MODEL OF ACUTE LUNG INJURY DUE TO SMOKE INHALATION AND BURNS

NONINVASIVE CARBON DIOXIDE MONITORING IN A PORCINE MODEL OF ACUTE LUNG INJURY DUE TO SMOKE INHALATION AND BURNS SHOCK, Vol. 39, No. 6, pp. 495 500, 2013 NONINVASIVE CARBON DIOXIDE MONITORING IN A PORCINE MODEL OF ACUTE LUNG INJURY DUE TO SMOKE INHALATION AND BURNS Slava Belenkiy,* Katherine M. Ivey,* Andriy I. Batchinsky,*

More information

Don t let your patients turn blue! Isn t it about time you used etco 2?

Don t let your patients turn blue! Isn t it about time you used etco 2? Don t let your patients turn blue! Isn t it about time you used etco 2? American Association of Critical Care Nurses National Teaching Institute Expo Ed 2013 Susan Thibeault MS, CRNA, APRN, CCRN, EMT-P

More information

Evaluation of oesophageal reflectance pulse oximetry in major burns patients

Evaluation of oesophageal reflectance pulse oximetry in major burns patients Burns 31 (2005) 337 341 www.elsevier.com/locate/burns Evaluation of oesophageal reflectance pulse oximetry in major burns patients S.K. Pal a, *, P.A. Kyriacou b, S. Kumaran c, S. Fadheel c, R. Emamdee

More information

Nihon Kohden America. Capnography Monitoring with the CapONE CO2 Sensor For use with the BSM 2300/4100/5100 and the BSM 6000 Series Bedside Monitors

Nihon Kohden America. Capnography Monitoring with the CapONE CO2 Sensor For use with the BSM 2300/4100/5100 and the BSM 6000 Series Bedside Monitors Nihon Kohden America Capnography Monitoring with the CapONE CO2 Sensor For use with the BSM 2300/4100/5100 and the BSM 6000 Series Bedside Monitors Self Study Training Packet August 2011 Table of Contents

More information

Sub-Study. PRotective Ventilation with Higher versus Lower PEEP during General Anesthesia for Surgery in OBESE Patients

Sub-Study. PRotective Ventilation with Higher versus Lower PEEP during General Anesthesia for Surgery in OBESE Patients PRotective Ventilation with Higher versus Lower PEEP during General Anesthesia for Surgery in OBESE Patients The PROBESE Randomized Controlled Trial Preliminary evaluation of postural reduction of peripheral

More information

Prepared by : Bayan Kaddourah RN,MHM. GICU Clinical Instructor

Prepared by : Bayan Kaddourah RN,MHM. GICU Clinical Instructor Mechanical Ventilation Prepared by : Bayan Kaddourah RN,MHM. GICU Clinical Instructor 1 Definition Is a supportive therapy to facilitate gas exchange. Most ventilatory support requires an artificial airway.

More information

Hypothermia Presentation

Hypothermia Presentation Hypothermia Presentation Thermoregulation Thermal regulation is a balance between heat production and heat loss. Despite marked changes in skin temperature, the body s homeostatic mechanisms are able to

More information

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2):209-217. 2011 Oxford University Press Effect of Phenylephrine and Ephedrine Bolus Treatment on Cerebral Oxygenation in Anaesthetized

More information

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter 1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3

More information

POSITION DESCRIPTION / PERFORMANCE EVALUATION

POSITION DESCRIPTION / PERFORMANCE EVALUATION POSITION DESCRIPTION / PERFORMANCE EVALUATION Job Title: Cardiopulmonary Services Manager Prepared by: Date: Supervised by: Cardiopulmonary Services Medical Director, CEO Approved by: Date: Job Summary:

More information

Bergen Community College Division of Health Professions Department of Respiratory Care Fundamentals of Respiratory Critical Care

Bergen Community College Division of Health Professions Department of Respiratory Care Fundamentals of Respiratory Critical Care Bergen Community College Division of Health Professions Department of Respiratory Care Fundamentals of Respiratory Critical Care Date Revised: January 2015 Course Description Student Learning Objectives:

More information

PHYSICIAN COMPETENCY FOR ADULT DEEP SEDATION (Ages 14 and older)

PHYSICIAN COMPETENCY FOR ADULT DEEP SEDATION (Ages 14 and older) Name Score PHYSICIAN COMPETENCY FOR ADULT DEEP SEDATION (Ages 14 and older) 1. Pre-procedure evaluation for moderate sedation should involve all of the following EXCEPT: a) Airway Exam b) Anesthetic history

More information

Correlation Between Partial Pressure of Arterial Carbon Dioxide and End Tidal Carbon Dioxide in Patients with Severe Alcohol Withdrawal

Correlation Between Partial Pressure of Arterial Carbon Dioxide and End Tidal Carbon Dioxide in Patients with Severe Alcohol Withdrawal ORIGINAL RESEARCH The Ochsner Journal 15:418 422, 2015 Ó Academic Division of Ochsner Clinic Foundation Correlation Between Partial Pressure of Arterial Carbon Dioxide and End Tidal Carbon Dioxide in Patients

More information

PulsioFlex Patient focused flexibility

PulsioFlex Patient focused flexibility PulsioFlex Patient focused flexibility Modular platform with intelligent visualisation for advanced patient Minimally invasive perioperative cardiac output trend with ProAQT Enables calibrated cardiac

More information

CARDIOPULMONARY EXERCISE TESTING has considerable

CARDIOPULMONARY EXERCISE TESTING has considerable 1686 Arterial Blood Gases During Exercise: Validity of Transcutaneous Measurements Carole Planès, MD, PhD, Michel Leroy, MD, Evelyne Foray, IDE, Bernadette Raffestin, MD, PhD ABSTRACT. Planès C, Leroy

More information

Extracorporeal Life Support Organization (ELSO) Guidelines for Pediatric Respiratory Failure

Extracorporeal Life Support Organization (ELSO) Guidelines for Pediatric Respiratory Failure Extracorporeal Life Support Organization (ELSO) Guidelines for Pediatric Respiratory Failure Introduction This pediatric respiratory failure guideline is a supplement to ELSO s General Guidelines for all

More information

Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context

Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context A subcommittee of the Canadian Neonatal Resuscitation Program (NRP) Steering Committee

More information

Capnography: The Most Vital Sign

Capnography: The Most Vital Sign Capnography: The Most Vital Sign Mike McEvoy, PhD, NRP, RN, CCRN Cardiac Surgical ICU RN & Chair Resuscitation Committee Albany Medical Center EMS Coordinator Saratoga County, NY www.mikemcevoy.com CO

More information

End Tidal CO2 Not All Its Cracked Up To Be The Limitations of PETCO2 In Sedation Analgesia

End Tidal CO2 Not All Its Cracked Up To Be The Limitations of PETCO2 In Sedation Analgesia End Tidal CO2 Not All Its Cracked Up To Be The Limitations of PETCO2 In Sedation Analgesia Tidal Volume Noninvasive monitoring of ventilation and exhaled carbon dioxide of a patient End Tidal CO2 Produces

More information

Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients

Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients Objectives Describe nocturnal ventilation characteristics that may indicate underlying conditions and benefits of bilevel therapy for specific

More information

Critical Care of the Post-Surgical Patient

Critical Care of the Post-Surgical Patient Critical Care of the Post-Surgical Patient, Dr med vet, DEA, DECVIM-CA Many critically ill patients require surgical treatments. These patients often have multisystem abnormalities during the immediate

More information

Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery

Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery Oliver Gödje, MD, Christian Thiel, MS, Peter Lamm, MD, Hermann Reichenspurner, MD, PhD, Christof Schmitz, MD,

More information

Cuffed Tracheal Tubes in Children - Myths and Facts. PD Dr. Markus Weiss Department of Anaesthesia University Children s Hospital Zurich Switzerland

Cuffed Tracheal Tubes in Children - Myths and Facts. PD Dr. Markus Weiss Department of Anaesthesia University Children s Hospital Zurich Switzerland Cuffed Tracheal Tubes in Children - Myths and Department of Anaesthesia University Children s Hospital Zurich Switzerland PRO Reduced gas leak, low fresh gas flow Decreased atmospheric pollution Constant

More information

7 Initial Ventilator Settings, ~05

7 Initial Ventilator Settings, ~05 Abbreviations (inside front cover and back cover) PART 1 Basic Concepts and Core Knowledge in Mechanical -- -- -- -- 1 Oxygenation and Acid-Base Evaluation, 1 Review 01Arterial Blood Gases, 2 Evaluating

More information